Introduction
In today’s process plants, most new field instruments
are smart digital instruments. Smart implies a How are HART
microprocessor-based instrument with extra functionality instruments properly calibrated?
and digital compensation. These instruments generally To calibrate a HART instrument consistent with its
offer better accuracy, long-term stability, and reliability application, it is very helpful to understand the functional
than conventional analog instruments. structure of a typical HART transmitter. HART instruments
consist of three distinct sections (see Figure 1).
1. Shifts in performance of electronic instruments over Leave the Communicator in the Shop!
time, due to exposure of the electronics and the The Fluke-744 HART Calibrator requires no external box
primary sensing element to temperature, humidity, or communicator for everyday HART field calibration and
vibration, and other field environmental factors. maintenance. It supports many popular models of HART
2. Regulations governing occupational safety, consumer transmitters, with more device-specific command support
safety, and environmental protection. than any other HART field calibrator. With the 744,
3. Quality programs such as ISO 9000 standards for all you can:
instruments that impact product quality. • Interrogate HART devices to determine type,
4. Commercial requirements such as weights, measures manufacturer, model, tag-ID, PV, and PVAO
and custody transfer. • Perform automated HART sensor trim and output trim
for selected devices
• Adjust ranging, damping, and other basic
A calibration procedure consists of a verification (As process–configuration settings
Found) test, adjustment to within acceptable tolerance if • Read and write HART tag and message fields to
necessary, and a final verification (As Left) test if an re-label smart transmitters
adjustment has been made. Data from the calibration is • Change sensor configuration on most temperature
collected and used to complete a report of calibration, transmitters
documenting instrument performance over time.
• Clone additional transmitters with basic HART
configuration data