Instruction Manual: pH/mV/°C/°F Bench Meter
Instruction Manual: pH/mV/°C/°F Bench Meter
Instruction Manual: pH/mV/°C/°F Bench Meter
The 700 series benchtop meters replace our popular 510 series
meters that were introduced in 2000.
1
2. GETTING STARTED
2.1 Keypad Functions
Powers the meter on and off. Upon power on, the meter
automatically begins in the mode that was last used.
Calibration and memory values are retained even if meter
is unplugged.
2
2.2 LCD Annunciators
DC Power supply.
3
3. CALIBRATION
3.1 pH Calibration
For best results, periodic calibration with known accurate standards is
recommended. Calibrate with standards that bracket your intended
measuring range while including a neutral standard (pH 7.00 or 6.86).
For example, if you expect to measure samples from pH 6.2 to pH 9.5,
calibration with 4.01, 7.00, and 10.01 will work well.
3) Rinse the pH and ATC electrodes then dip into the next pH buffer. The
secondary display will lock on the appropriate buffer value. When the
READY indicator appears, press to accept. The primary reading
will flash briefly then display the percent efficiency (slope) before the
secondary display begins scrolling the remaining available buffers.
4
4) To calibrate another buffer repeat step 3) or press to return to the
measurement mode. Note: The meter will automatically return to
measurement mode upon successful completion of the number of
specified calibration points. To specify a different number of pH
calibration points see Section 5.6.
Additional notes:
A single point (offset) calibration is only allowed with pH 7.00 or
pH 6.86 buffers.
1) Connect the temperature probe to the meter and place into a solution
with a known accurate temperature such as a constant temperature
bath. Note: To adjust the manual temperature compensation (MTC)
value, do not connect the temperature probe.
2) Press as needed to select pH or mV/R.mV.
3) Press followed by . The primary display shows the
measured temperature while the secondary display shows the factory
default temperature.
4) Adjust the temperature using or . Press to accept or
to cancel. The meter allows an adjustable maximum value of ± 5ºC
(or ± 0.9ºF) from the factory default temperature.
5
3.3 Millivolt (mV) Offset Adjustment
Oxidization Reduction Potential (ORP or Redox) is not a precise
measurement, but is useful as a relative indicator. As such, mV offset
adjustment is not meant to enhance accuracy, but rather to make readings
comparable to a reference. Commercial ORP solutions are often used as
a check standard in which a meter/electrode system are verified to be
close to a given value, instead of being used as a calibration standard in
which adjustments are made in an attempt to match the ORP value.
Note: When an offset has been stored successfully, R.mV replaces mV.
6
1) Connect the ISE and press as needed to select ppm. Note: The
primary reading will show “- - - -” without a stored ion calibration. The
secondary value is the corresponding mV reading of the ISE.
2) Dip the ISE into your lowest standard solution value and stir. Press
when the secondary reading is stable.
3) Press or to match the primary display to your corresponding
ion calibration value (0.10, 1.0, 10, etc).
4) Press to accept or to abort. The primary display will show the
next highest calibration standard value.
5) Rinse the ISE then dip it into the corresponding calibration standard
and stir.
6) Press to accept or to cancel. The mV/decade slope value will
be displayed briefly if the calibration is successful. “SLP Err” indicates
that the calibration for the current point was not successful. This occurs
when the slope (mV difference between two consecutive points) is
lower than 15mV/decade or higher than 90mV/decade.
7) Revert to step 3) to calibrate additional points or press as needed
to return to measurement mode at any time. Note: The meter will
automatically return to measurement mode upon successful completion
of the number of specified calibration points. To change the number of
ion calibration points see Section 5.6.
7
2) To recall data from memory, press . The location of the most recent
stored data is displayed first. Press or to select the location of
the desired data, then press to accept.
5. SETUP FUNCTIONS
Use the set up mode to customize your instrument operation. During
measurement, press and hold for 5 seconds to enter SETUP mode.
Press or to change programs or change options.
Press to select the program or confirm selection.
Press to go back a level or to return to measurement mode.
8
Note: When multiple slopes exist, the slope that is available to view will
depend on the measured value before entering SETUP mode.
For example, if pH calibration values pH 4.01, 7.00 and 10.01 were
completed, there will be two slope values; one that corresponds to pH
values below 7.00 and another that corresponds to pH values above 7.00.
If the measured pH value was 5.23 when setup mode was entered, the pH
slope that corresponds to values lower than 7.00 will be available to view.
If no calibration is stored, the offset will be 0.0 mV and the slope is 100%.
Press .
Press .
Press or to choose READY “On”, READY “OFF”, or
Auto HOLd.
Press to confirm.
Press .
Press or to select ºC or ºF.
Press to confirm.
9
5.6 P3.3 buFF (pH Buffers & Calibration Points)
Press .
Press or to select “USA” or “NIST” buffer group
(pH mode only).
Press or to select number of Calibration Points
(pH and Ion only).
Press to confirm.
Press .
Press or to select “Yes” or “No”.
If “Yes”, press or to select “Cal” (calibration reset only) or
“FCt” (complete reset to factory default settings).
Press to confirm.
Press .
Press or to select “Yes” (erase memory) or “No”.
Press to confirm.
10
6. TROUBLESHOOTING GUIDE
11
CAUSE SOLUTION
12
7. ION SELECTIVE ELECTRODES
Contact our partners for more details regarding ISE and related solutions.
The table below lists the most common electrodes available.
13
8. SPECIFICATIONS
14
9. WARRANTY
This meter is supplied with a warranty against significant deviations in
material and workmanship for a period of THREE years from date of
purchase whereas probe with a SIX month warranty.
If repair or adjustment is necessary and has not been the result of abuse
or misuse within the designated period, please return – freight prepaid –
and correction will be made without charge. Eutech Instruments/Oakton
Instruments will determine if the product problem is due to deviations or
customer misuse.
Out of warranty products will be repaired on a charged basis.
The warranty on your instrument shall not apply to defects resulting from:
15
For more information on our products, please contact our channel
partner or visit our websites listed below:
Distributed by: