Good One - TEM - MSC2100Manual1 PDF
Good One - TEM - MSC2100Manual1 PDF
Good One - TEM - MSC2100Manual1 PDF
1
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
If the program is not open click double-click on the TEM Controller icon in
the PC monitor screen.
Check the points bellow:
1
4
1)
1) High Voltage (HT) : Acc. = 160kV (Start) or 200kV (Middle of the Day);
2) Spot Size = 1, Alpha = 3;
3) X = Y = 0, Z = 0, TX = 0 (TY doesnt matter);
4) Select the sample-holder (SH) type that you use:
Single Tilt: EM-21010 = Single Tilt Holder;
Double Tilt: EM-31630 = Specimen Tilt Holder (DTC);
Single Tilt Rotation: EM-31650 = Rotation Holder (STR);
Double Tilt Beryllium: EM-31640 = Specimen Tilt Holder Be (DTB);
2
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
Make sure that the HT Status = Ready (HT allowed) on the right upper
corner of the window High Voltage
160.00
5
4
6.
For single tilt holder, loosen two screws TWO turns, rotate the cover
plate away and mount specimen with film side UP. For double tilt holder,
loosen two screws TWO turns, rotate the two plate-clamping fingers and
mount sample facing DOWN. NEVER OVERTIGHTEN THE SCREWS!!!
After to put firmly your sample, always check that the SHs o-rings are
completely free of dust or lint, under the optical microscope. Use tweezers
to remove the contamination.
Be Careful!!! Dont introduce the SH on the microscope until
be sure that it is completely free of dust or lint. One can
induce vacuum leak!!!
4
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
Normal Setup:
Degas Setup:
Most of the commands on the control panels (right and left) related to
Standard Operation, Apertures and Movements of the Stages can be accessed
in Operation menu.
b) Focus the beam using the BRIGHTNESS knob to obtain the crossover
and bring it to the center of the phosphor screen using the BRIGHTNESS
SHIFT knobs X and Y.
c) Defocus (clockwise) the beam till approximately the same screen
diameter. If the beam is not centered use the aperture position adjust to
center it (left panel).
Important: Remember to check the condenser aperture
alignment from time to time. It is mechanical and it will
misalign more often than you imagine.
a) Repeat 9a to 9d.
b) Set SPOT SIZE to 5. Focus the beam with BRIGHTNESS knob, and
center it by using the SHIFT knobs.
c) Set SPOT SIZE to 1 and refocus the beam with BRIGHTNESS knob.
d) Press F4 button (Gun command) and center the beam with the SHIFT
knobs. Disable F4 button.
e) Repeat the procedures a to c several times till the focused beam dont
move from the center when the spot size is modified from 5 to 1. Disable
F4.
e) Press COND STIG and improve the image quality with DEF X and Y
knobs to obtain the smallest and brightest spot.
f) With BRIGHTNESS knob check if there is astigmatism (i.e. the image is
distorted in perpendicular directions if the BRIGHTNESS knob is turned
from under-focus to over-focus).
g) Repeat e till a good correction. Disable COND STIG.
8
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
Objective Lens
If you prefer, you can use the ESW 500 camera to adjustment the
voltage center. To use the camera, adjust the BRIGHTNESS knob until
the illumination larger than the phosphor screen. Also lower the screen
before you change the TEM magnification or its mode. In the
9
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
During the training do not save acquired images to HDD: the acquired
images are stored as 16 bit TIFF files in the specified folder. Unselect the button
with the ellipsis (...) to select the folder. It is also possible to create a new folder
by clicking on the "Create folder ..." button
Be sure that the images are not being saved on the hard drive (HD)
images;
11
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
Figure 1: Image with strong astigmatism: (a) under-focus, (b) focus, and (c) over-focus. After
almost fully correcting the astigmatism: (a) under-focus, (b) focus, and (c) over-focus
Press
12
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
4. Acquiring Images
Click on camera1
Press the "Viewport options" button and it will open a popup menu.
- Checked Auto increment: each click on the exposure button will store
the image in the next available image buffer in the image manager. If it is not
checked, the image in the current image buffer will be overwritten without a
warning.
- Checked "Scale bar", a scale bar is shown above the image data.
The position and appearance of the scale bar can be modified in EMMENU 4's option
dialog. There are 4 different predefined positions to display the scale bar: upper left,
upper right, lower left and lower right corner.
13
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
Just turn off the filament. You can leave the TEM HT at 200kV.
Click on FILAMENT OFF on HIGH VOLTAGE CONTROL window.
BEAM CURRENT must decrease gradually from 112 A down to 102 A
(means that just HT is ON = 200 kV).Wait!!!
TEM phosphor screen must be down (no beam going to the cameras).
Remove the objective and the SAD apertures if you did use them.
CCD Camera (Gatan MSC794): Leave it on, but the CAMERA key must
be down (position OUT).
To remove the SH follow the instructions on the page at the end of this manual. A
copy of it is also on the TEM. Read it before removing the SH.
REMEMBER: The vacuum will pull the SH into the TEM. You
should never rotate and pull the SH simultaneously. Do only one
movement at time. If not, you may leak air and crash the TEM.
14
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
15
Elaborado: Marina Magnani
Reviso: Jefferson Bettini
Julho/2011
To remove the SH follow the instructions on the page at the end of this manual. A
copy of it is also on the TEM. Read it before removing the SH.
REMEMBER: The vacuum will pull the SH into the TEM. You
should never rotate and pull the SH simultaneously. Do only one
movement at time. If not, you may leak air and crash the TEM.
PUMP/AIR
~70o
STOP
~20o
Black Pin
Guide Pin
17
STOP
~20o
PUMP/AIR
4
6. Set the switch to AIR.
7. Wait for ~30 seconds.
8. Remove the SH (step 5).
18