Aim: Apparatus:: To Perform Thermo-Gravimetric Analysis of Calcium Oxalate
Aim: Apparatus:: To Perform Thermo-Gravimetric Analysis of Calcium Oxalate
Aim: Apparatus:: To Perform Thermo-Gravimetric Analysis of Calcium Oxalate
Apparatus: thermal analysis set up, Al2O3 crucibles, spatula, tweezers, gloves, weighing
machine.
Principle/ Theory:
Thermal Analysis
In destructive technique the whole or part of the sample under testing is destroyed.
Examples: Chemical analysis; thermal analysis; mechanical testing like ball milling, tensile and
compressive strength measurement, scratch testing, frictional coefficient measurement, etc.
In non-destructive technique the sample is preserved and is not destroyed during the
characterization process. Example: Infra red spectroscopy, Uv-visible spectroscopy, X-ray
photoelectron spectroscopy, Raman spectroscopy, etc.
Thermal analysis helps in gaining information regarding the sample. The formal
definition of Thermal analysis as given by the International Confederation for Thermal Analysis
and Calorimetry) and Japanese Industrial Standards (JIS) is
For the analysis to be carried out in effective way it is necessary to have a high degree of
precision in the measurement of weight, temperature and temperature change. As many weight
loss curves look similar, the weight loss curve may require transformation before results may be
interpreted. A derivative weight loss curve can be used to tell the point at which weight loss is
most apparent.
Measuring Signal:
Temperature dependence
M = m(T) or ΔM = m(T) - mo
Time dependence
Where
m: sample mass
T: temperature
t: time
Procedure:
A) Zero baseline curve:
• Clean the crucible or replace it if the old sample can not be cleaned from the crucible.
• Inspect the Al2O3 reference material. If it is contaminated replace with new Al 2O3.
• The normal amount is 1/3 full (max. 150 mg for sample and reference).
• Close the “plug-in” on the furnace top.
4. Start the software STA data collection software using the desktop icon.
c) Enter the data collection information . The data collection termination is control by
the Duration and End Temp. Value. Once either value is reached the data collection
will be terminated. Always select a time or temperature higher than the final data
point needed. Click “OK” to complete settings
g) Check the current furnace and sample temperature. The temperature must be a
room temperature before starting the system.
h) select “OPTIONS”, “SETUP SCALES” and Click to TARA ADJUSTMENT to calibrate the
Tara
i) Start the measurement by selecting “Start”, Check data collection and controller.
Click “OK” to start the measurement
j) Monitor the run using the real time graphic and current values windows.
k) Once the run is completed monitor the current values and do not lift the furnace and
o
protection tube until the temperature is less than 100 C.
B) Sample curve
• Clean the crucible or replace it if the old sample can not be cleaned from the
crucible.
3) Start the software STA data collection software using the desktop icon
• Operator, Atmosphere,
• Laboratory,
• Flow Rate,
• Comments
The data acquisition termination is determined by the Duration and End Temp. value. Once
either value is reached the data collection will be terminated. Always select a time or
temperature higher than the final data point needed. Click “OK” to complete settings
9) Check the current furnace and sample temperature. The temperature must be a
room temperature before starting the system.
10) select “OPTIONS”, “SETUP SCALES” and Click to TARA ADJUSTMENT to calibrate
the Tara
11) Start the measurement by selecting “Start”, Check data collection and controller.
Click “OK” to start the measurement
12) Monitor the run using the real time graphic and current values windows.
13) Once the run is completed monitor the current values and do not lift the furnace
o
and protection tube until the temperature is less than 100 C.
Example: TGA is a useful technique for assessing the effectiveness of oven drying solution
polymerized polymer samples. The volatilization of residual solvent is typically associated with
the initial weight loss process in a TGA heating run. In some cases, absorbed moisture may be
liberated over this same temperature range, though. After the initial solvent (or moisture) weight
loss process, TGA profiles will typically plateau to some constant weight level until the polymer
degradation temperature range is reached. The weight fraction of residual solvent (or moisture)
and the onset and maximum rate weight loss degradation temperatures are readily determined by
TGA.
Applications: TGA measures the weight change in materials as a function of time and
temperature. The measurements provide basic information about the thermal stability of a
chemical and its composition. TGA is one of the most commonly requested thermal techniques
and it is used to characterize both inorganic and organic materials, including polymers. The
measurement is typically carried out on milligram size samples with a heat rate of 10°C/min
from 20° to 1000°C in a nitrogen atmosphere.
Calcium-oxalate: CaC2O4+1H2O
(it is useful to start the measuring without the controller to see if the TG and the DTA
signal has no drift anymore. if it is ok stop measurement and start the whole experiment)
- after run is finished switch SLIDER in the menu OPTIONS to ON √ for faster cooling
Measurement zero file (1000 °C)
- One heating rate to the maximum, same gas conditions, and heating rate like in the
sample run
- zero file name (should follow DOS conventions max. 8 characters or numbers only),
duration, end temperature, and ranges for TG 25mg and DTA 250uV, switch filter to √
(it is useful to start the measuring without the controller to see if the TG and the DTA
signal has no drift anymore. if it is ok stop measurement and start the whole experiment)
- monitor measurement with CONTROLLER CURRENT VALUES or
- after run is finished switch SLIDER in the menu OPTIONS to ON √ for faster cooling