Datex-Ohmeda Avance - User Reference Manual
Datex-Ohmeda Avance - User Reference Manual
Datex-Ohmeda Avance - User Reference Manual
User Responsibility
This Product will perform in conformity with the description thereof
contained in this User’s Reference manual and accompanying labels
and/or inserts, when assembled, operated, maintained, and repaired
in accordance with the instructions provided. This Product must be
checked periodically. A defective Product should not be used. Parts
that are broken, missing, plainly worn, distorted, or contaminated
should be replaced immediately. Should repair or replacement
become necessary, Datex-Ohmeda recommends that a telephonic or
written request for service advice be made to the nearest
Datex-Ohmeda Customer Service Center. This Product or any of its
parts should not be repaired other than in accordance with written
instructions provided by Datex-Ohmeda and by Datex-Ohmeda
trained personnel. The Product must not be altered without the prior
written approval of Datex-Ohmeda. The user of this Product shall
have the sole responsibility for any malfunction which results from
improper use, faulty maintenance, improper repair, damage, or
alteration by anyone other than Datex-Ohmeda.
AAAX11111 AAAXX111111AA
The X represents an alpha The XX represents a number
character indicating the year indicating the year the product
the product was manufactured; was manufactured; 04 = 2004,
H = 2004, J = 2005, etc. I and 05 = 2005, etc.
O are not used.
1 Introduction
M1145956 i
Avance
3 Operation
ii M1145956
Table of Contents
M1145956 iii
Avance
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Alarm priorities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Silencing alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Display changes during alarms . . . . . . . . . . . . . . . . . . . 7-2
De-escalating alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Battery indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Internal failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
List of alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Alarm ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Breathing system problems . . . . . . . . . . . . . . . . . . . . . . . . 7-16
Electrical problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17
Pneumatic problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
iv M1145956
Table of Contents
M1145956 v
Avance
vi M1145956
Table of Contents
M1145956 vii
Avance
Warranty
viii M1145956
1 Introduction
M1145956 1-1
Avance
Intended use
The Avance anesthesia system combines our highly advanced
anesthesia delivery, the very best in patient monitoring, and care
information management. The contemporary, compact design
addresses many ergonomic considerations including an efficient
cable management solution and storage system and an extra large
work surface area. Optional, integrated features include auxiliary O2
and suction control. The Avance system offers improved usability
through integration of ventilation, gas delivery and gas monitoring (E
and M-Series respiratory gas module capable) on a full-color 12-inch
display. This system also features electronic gas mixing of up to three
gases.
The Avance system uses SmartVent ventilation technology offering
Volume Control Ventilation with tidal volume compensation and
electronic PEEP. The proven SmartVent technology also features
optional Pressure Control Ventilation, Pressure Support Ventilation
with an Apnea Backup (PSVPro) that is used for spontaneously
breathing patients, Synchronized Intermittent Mandatory Ventilation
(SIMV) modes, Pressure Control Ventilation-Volume Guarantee
(PCV-VG), and VCV cardiac bypass. These advanced features allow
for the ventilation of a broad patient range: from neonates and
compromised patients to routine cases.
The Avance system uses the Advanced Breathing System (ABS).
This integrated breathing system is easy to remove and disassemble,
is made of highly durable polymers, and is fully autoclavable. Its fully
integrated design enhances the system’s elegance while minimizing
tube connections, minimizing circuit volume, and increasing the work
surface area.
The anesthesia system is designed for expansion and upgrades, so it
is easy to add new technologies and ventilation capabilities without
investing in a new system.
This anesthesia system is not suitable for use in an MRI environment.
This system must only be operated by authorized medical personnel
well trained in the use of this product. It must be operated according
to the instructions in this User’s Reference manual.
1-2 M1145956
1 Introduction
AB.91.108
M1145956 1-3
Avance
Standby
O2 + O2 Flush button
1-4 M1145956
1 Introduction
SN Serial number
REF Stock number
Maximum Vacuum
Lock Unlock
This way up
O2% O2 cell connection
M1145956 1-5
Avance
Caution: federal law prohibits Alarm silence touch key (Tec 6 Plus)
dispensing without prescription.
O2% indicator on left and balance gas Gas indicator. Color associated with gas
indictor on right. Colors associated settings.
with gas settings.
Alarm silence Submenu
Pipeline Cylinder
Systems with this mark agree with the Authorized representative in the
European Council Directive (93/42/ European Community
EEC) for Medical Devices when they
are used as specified in their User’s
Reference manuals. The xxxx is the
certification number of the Notified
Body used by Datex-Ohmeda’s
Quality Systems.
Date of manufacture Manufacturer
1-6 M1145956
1 Introduction
Abbreviations
Abbreviation Definition
A
AA Anesthetic agent
ABS Advanced breathing system
ACGO Auxiliary Common Gas Outlet
AGSS Anesthesia Gas Scavenging System
Alt O2 Alternate O2
APL Adjustable pressure-limiting
APN Apnea
C
CGO Common Gas Outlet
CO2 Carbon dioxide
Compl Compliance
E
ET End-tidal concentration
EtCO2 End-tidal carbon dioxide
EtO2 End-tidal oxygen
Exp Expiratory
F
FI Fraction of inspired gas
FiCO2 Fraction of inspired carbon dioxide
FI-ET Difference between inspiratory and expiratory
concentrations
FiO2 Fraction of inspired oxygen
F-V Flow-volume loop
M1145956 1-7
Avance
Abbreviation Definition
I
I:E Inspiratory-expiratory ratio
Insp Inspiratory
Insp Pause Inspriatory pause time
M
MAC Minimum Alveolar Concentration
MV Minute volume
MVexp Expired minute volume
MVinsp Inspired minute volume
N
N2O Nitrous oxide
O
O2 Oxygen
P
Pair Air supply pressure
Paux Auxiliary pressure
Paw Airway pressure
PCV Pressure controlled ventilation
PCV-VG Pressure controlled ventilation - volume guaranteed
PEEP Positive end expiratory pressure
PEEPe Extrinsic positive end expiratory pressure
Pexp Expiratory pressure
P-F Pressure-flow loop
Pinsp Inspiratory pressure
Plimit High pressure limit
Pmax Maximum pressure
Pmean Mean pressure
PO2 Oxygen supply pressure
Ppeak Peak pressure
Pplat Plateau pressure
Psupp Support pressure
PSV Pressure supported ventilation
PSVPro Pressure supported ventilation with apnea backup
P-V Pressure-volume loop
R
Rate Respiratory rate
Raw Airway resistance
RR Respiratory rate
1-8 M1145956
1 Introduction
Abbreviation Definition
S
SIMV/PSV Synchronized intermittent mandatory ventilation with
pressure supported ventilation
SIMV-PC Synchronized intermittent mandatory ventilation -
pressure controlled
T
TV Tidal volume
TVexp Expired tidal volume
TVinsp Inspired tidal volume
V
VCO2 Carbon dioxide production
VCV Volume controlled ventilation
Vol Volume
M1145956 1-9
Avance
1-10 M1145956
2 System Controls and Menus
M1145956 2-1
Avance
System overview
12
11
2
3
10
4
5
6
7
8
AB.91.111
2-2 M1145956
2 System Controls and Menus
5 System switch Set the switch to the On (|) position to permit gas flow and to turn on the system.
6 Mains indicator The mains indicator will come on when AC power is connected.
7 Integrated suction Turn the switch to MAX for full vacuum. Turn the switch to Off (0) for no vacuum. Turn
(optional) the switch to On (|) for adjustable vacuum. When in adjustable vacuum, turn the knob
clockwise to increase the vacuum and counterclockwise to decrease the vacuum.
9 O2 flush button Push the O2 flush button to supply high flows of O2 to the breathing system.
11 Auxiliary O2 flow Turn the knob counterclockwise to increase the flow. Turn the knob clockwise to
control (optional) decrease the flow.
M1145956 2-3
Avance
1
2
3
4
5
6
11
10
AB.91.029
9
8
2-4 M1145956
2 System Controls and Menus
15
14
13
12
11
10
1
2
9
3
4
AB.91.110
6 5
M1145956 2-5
Avance
10 Adjustable pressure-limiting Adjusts breathing system pressure limit during manual ventilation. The
(APL) valve scale shows approximate pressures. Above 30 cmH2O, the knob will click
as it turns.
70
MI
N
30
20
2-6 M1145956
2 System Controls and Menus
Optional ABS
components
1 4
2
3
AB.82.043
AB.91.109
6
2 Auxiliary Common Gas Set switch to ACGO position for fresh gas to flow through the ACGO port. The
Outlet (ACGO) switch ACGO may be used to provide fresh gas to an auxiliary manual breathing circuit.
5 EZchange canister Push to drop the canister to EZchange position. This seals the breathing circuit,
release permitting continued ventilation and rebreathing of exhaled gases. Be sure to
hold the canister by the handle before releasing the canister.
6 Condenser drain button Push to drain water out of the condenser.
M1145956 2-7
Avance
Non-circle circuit Fresh gas flow is diverted around the inspiratory check valve and out
through the inspiratory port when non-circle ventilation is selected.
This fresh gas source may be used with circuits without CO2
absorbent capability (for example, Mapleson variants). Mechanical
ventilation is not available when using the non-circle circuit. Tidal
volume monitoring is not available.
O2 monitoring of fresh gas is available automatically when using the
non-circle circuit if the system has the airway module option or the O2
cell monitoring option.
Fresh gas oxygen concentration is displayed on the screen. Set the
alarm limits appropriately. Note that fresh gas oxygen concentration
may not reflect FiO2 when using these types of circuits. Use an
external O2 monitor when using a rebreathing circuit with the non-
circle circuit.
Systems with both an airway module and an O2 cell will display the
circuit O2 value obtained from the airway module.
Do not use an external ventilator when using the non-circle circuit. Do
not use the non-circle circuit to drive external ventilators or for jet
ventilation.
2-8 M1145956
2 System Controls and Menus
ACGO (optional) Fresh gas flow is directed through the Auxiliary Common Gas Outlet
(ACGO) on the front of the machine when the ACGO switch is in the
ACGO position. Mechanical ventilation is not available when
operating an auxiliary manual breathing circuit with fresh gas from the
ACGO. The Bag/Vent switch, APL valve, and bag arm are not part of
the external circuit. Volume and pressure monitoring are not
available.
O2 monitoring of fresh gas is available automatically when the ACGO
is selected if the system has the airway module option or the O2 cell
monitoring option. A sample of the fresh gas is diverted to the O2 cell
in the breathing system. The sample flow to the O2 cell is dependent
on the pressure in the external circuit. The sample flow reduces the
fresh gas flow rate to the auxiliary breathing circuit equal to the
amount diverted to the O2 cell.
Fresh gas oxygen concentration is displayed on the screen. Set the
alarm limits appropriately. Note that fresh gas oxygen concentration
may not reflect FiO2 during spontaneous breathing or in rebreathing
circuits. Use an external O2 monitor if using a rebreathing circuit on
ACGO.
Systems with both an airway module and an O2 cell will display the
circuit O2 value obtained from the airway module.
Do not use an external ventilator on the ACGO. Do not use the ACGO
to drive external ventilators or for jet ventilation.
Scavenging the A sample of the fresh gas is diverted to the airway module or the O2
ACGO sample flow cell in the breathing system to show the O2 numerics on the screen.
This sample flow should be scavenged when an auxiliary manual
breathing circuit is used with N2O or volatile anesthetics. If
scavenging is not connected, the sample flow is emptied into the
room. To connect the scavenging:
1. Attach a circle breathing circuit to the inspiratory and expiratory
ports.
2. Occlude the breathing circuit by connecting the Y-piece to the
leak test plug located behind the expiratory port.
3. Check for clinically correct settings.
4. Check the position of the Bag/Vent switch.
• If the Bag/Vent switch is set to mechanical ventilation mode,
the bellows fills slowly with the sample flow. When the
bellows is full, the sample flow goes to the AGSS.
(Mechanical ventilation does not start when the ACGO switch
is set to ACGO.)
• If the Bag/Vent switch is set to the bag mode, set the APL
valve to MIN, and attach a bag. The bag fills slowly with the
sample flow. When the bag is full, the sample flow goes to the
AGSS.
M1145956 2-9
Avance
Scavenging from an Scavenge the exhaust if an auxiliary manual breathing circuit is used
auxiliary manual with N2O or volatile anesthetics.
breathing circuit An auxiliary inlet is available for active and passive AGSS units. It
provides a female connection with 30 mm - 30 mm male connector
(or a 30 mm -19 mm male connector) into the auxiliary port under the
breathing system. Do not use these connectors as an outlet for
exhaust flow.
The auxiliary inlet is a convenience inlet to the air brake of active
AGSS units. There is a reservoir to capture exhaust flows higher than
the extract flow.
A separate exhaust hose is needed from the auxiliary manual
breathing circuit to the disposal point for all AGSS units.
Scavenging a gas Sample gas from a gas monitor can be scavenged using the sample
monitor sample flow gas return port or the AGSS.
• To scavenge from a gas monitor using the sample gas return
port, connect the tubing from the monitor to the sample gas return
port.
• To scavenge from a gas monitor using the AGSS, connect tubing
from the monitor to the male luer inlet on the bottom of the AGSS
underneath the breathing system.
2-10 M1145956
2 System Controls and Menus
Vaporizer controls
Refer to the vaporizer User’s Reference manual for more detailed
information on the vaporizer.
3 3
4 4
AA.43.051
AB.80.009
5
1 2
1. Tec 6 Plus
2. Tec 7
3. Lock lever
4. Concentration control and release
5. Indicators (Tec 6 Plus)
6. Silence alarm touch key (Tec 6 Plus)
Figure 2-5 • Vaporizer controls
M1145956 2-11
Avance
4 Concentration control and Push the release and turn the concentration control to set the agent
release concentration. The Tec 6 Plus concentration control does not turn as long
as the warm-up indicator is on.
n
Åbe
5 Indicators (Tec 6 Plus) All indictors come on briefly at the start. The warm-up indicator goes off
after approximately 10 minutes and the operational indicator comes on.
Other indicators come on to advise the user of required action.
6 Silence alarm touch key Push to silence alarms. Hold for 4 seconds to sound the speaker and light
(Tec 6 Plus) all indicators (alarm test).
2-12 M1145956
2 System Controls and Menus
Display controls
7 1
6 2
5
3
AB.91.107
2 4
1. Silence Alarms key Push to silence any active, silenceable high and medium priority alarms or to
suspend/acknowledge any non-active medium or high priority alarms. Alarm is
silenced for 120 seconds or alarm is suspended for 90 seconds.
2. Menu keys Push to show corresponding menu.
3. ComWheel Push to select a menu item or confirm a setting. Turn clockwise or counterclockwise
to scroll menu items or change settings.
4. Normal Screen key Push to remove all menus from the screen.
5. Quick keys Push to change corresponding gas setting or ventilator setting. Turn the ComWheel
to make a change. Push the ComWheel to activate the change.
6. Timer keys Push to start or stop the timer. Push to reset the timer back to zero.
7. MV/TV Alarms key Push to turn off the MV and TV alarms. Push again to turn the MV and TV alarms on.
8. Alarm LEDs Turn on solid or flash to indicate alarm priority.
M1145956 2-13
Avance
2 3 4 5 6
PEEP RR /min
Et % Fi %
Agent
8
MAC 40
CO2
EtCO2 %s FiCO2 %s
Air O2
EtO2 % FiO2 %
AB.91.125
12 11 10 9
1. Split screen area showing electronic gas flow indicators
2. Alarm silence symbol and countdown
3. Alarm message fields
4. Waveform fields
5. General message field or timer field
6. Clock
7. Battery indicator field
8. Measured values field
9. Pipeline and cylinder supply or respiratory data or digit field
10. Ventilator settings
11. Ventilation mode
12. Gas settings
Figure 2-7 • Normal view
2-14 M1145956
2 System Controls and Menus
When a menu key is selected, the menu field overlays the gas flow
indicators and the waveform fields start at the right edge of the menu.
cmH2O
1 Main Menu Ppeak Pmean
Trends
System Status
Cardiac Bypass Off PEEP RR /min
Fresh Gas Usage
Screen Setup
Parameters Setup Et % Fi %
Calibration
Normal Screen
MAC 40
EtO2 % FiO2 %
AB.91.124
1. Menu
2. Waveform fields
Figure 2-8 • Menu view
M1145956 2-15
Avance
Waveform fields Up to three waveforms can be shown on the normal screen view.
Each waveform can be set to show specific information such as Paw,
agent, flow, or CO2 data. The corresponding numeric information
shows in the measured values field to the right of the waveform. If the
waveform is set to show the agent and no airway module is inserted,
that waveform and numeric areas are blank.
When one waveform is turned off, that waveform and the
corresponding numerics information are removed from the normal
screen view. The remaining waveforms and numerics increase in size
to fill the waveform area. When two waveforms are turned off, those
waveforms and the corresponding numerics information are removed
from the normal screen view. The remaining waveform increases in
size and is centered in the waveform area.
See “Screen configuration” in the “Operation” section for more
information.
Digit field The digit field can be set to show specific information such as gas
supply, flow, or agent. If the digit field is set to show agent and no
airway module is inserted, the area is blank.
Paw, O2, and either TVexp or CO2 must show on the display during a
case. If any of these parameters is not selected to show on the
display, the digit field information is replaced with the missing
parameter.
See “Screen configuration” in the “Operation” section for more
information.
2-16 M1145956
2 System Controls and Menus
Using menus
Push a menu key to display the corresponding menu. Use the
ComWheel to navigate the menu.
1 2
Main Menu
Trends 3
System Status On
Cardiac Bypass On Off
6 Fresh Gas Usage 4
Screen Setup
Parameters Setup
Calibration
Normal Screen
AB.91.126
1. Menu title
2. Current selection
3. Adjustment window
4. Submenu
5. Instructions or help information
6. Menu items
Figure 2-9 • Menu example
M1145956 2-17
Avance
2-18 M1145956
3 Operation
M1145956 3-1
Avance
AB.91.044
1. System switch
2. Mains indicator
Figure 3-1 • Mains indicator and system switch
3-2 M1145956
3 Operation
Start Case
Default Settings
Patient Weight
Set Vent By Weight
Patient Age
Other Gas
CO2 Alarms
Volume Apnea
The Default Settings has five selections. The first four selections are
case types. The fifth selection is Last Case. The case type in the first
slot shows as the default when the Start Case menu is opened.
Case types are configured in the Install/Service - Case Defaults
menu. The Install/Service menu is only accessible by a Super User
or a service person. See the “Super User Mode” section for more
information on the Start Case menu defaults.
The values of Patient Weight, Patient Age, Other Gas, and Volume
Apnea are configured to the default case type by the Super User.
CO2 Alarms setting is set to ‘On’ when the Start Case menu is
opened.
Note Volume Apnea is not shown on the Start Case menu when the
Volume Apnea Selection is set to Disable in the Install/Service
menu.
The Set Vent By Weight menu item is accessible when the
ventilation mode is set to VCV or PCV-VG. Use this setting for breath
rate and tidal volume calculations based on the set patient weight.
The Default Settings selection shows ‘Weight’ when Set Vent By
Weight is selected.
M1145956 3-3
Avance
Using customized The case type, ventilator settings, ventilation mode, alarm settings,
settings and gas settings can be changed before starting a case. Use the
Vent Setup menu, Alarm Setup menu, or Gas Setup menu to
change settings not shown on the Start Case menu before starting a
case. Return to the Start Case menu from these menus to make
additional adjustments or to start a case.
1. Set the Bag/Vent switch to Bag.
2. Push the Start/End Case key.
3. Verify or change the settings on the Start Case menu.
• Select a case type or Last Case from the Default Settings.
• Make adjustments to the other settings as appropriate for the
case.
• The Default Settings selection changes to show ‘Preset’ if
any changes to the settings are made.
4. To change the alarm settings:
• Push the Alarm Setup key.
• Select Adjust Settings.
• Change the settings.
• Select Back.
• Make any other changes to the alarm setup.
• Select Previous Menu to return to the Start Case menu to
make additional adjustments.
5. To change ventilator settings:
• Push the Vent Setup key.
• Select Adjust Settings.
• Change the settings.
• Select Exit to return to the Start Case menu.
6. To change the ventilation mode:
• Push the Vent Setup key.
• Select the desired mode.
• Select Confirm or adjust the settings and select Confirm to
return to the Start Case menu.
3-4 M1145956
3 Operation
M1145956 3-5
Avance
End case
Use the End Case menu to stop gas flow and end the patient alarms.
End Case
End Case Now
Normal Screen
3-6 M1145956
3 Operation
Ventilator setup
Access the Vent Setup menu by pushing the Vent Setup key. The
system has up to six modes of mechanical ventilation:
• Volume Control Ventilation (VCV).
• Pressure Control Ventilation (PCV) (optional).
• Synchronized Intermittent Mandatory Ventilation/Pressure
Support (SIMV/PSV) (optional).
• Pressure Support Ventilation (PSVPro) (optional).
• Synchronized Intermittent Mandatory Ventilation-Pressure
Control (SIMV-PC) (optional).
• Pressure Control Ventilation-Volume Guarantee (PCV-VG)
(optional).
Vent Setup
Adjust Settings
Mode:
VCV
PCV
Spirometry
Normal Screen
Using quick keys The four main ventilator settings for each mode can easily be
changed using the ventilator quick keys.
1. Push a ventilator quick key to select the corresponding ventilator
setting.
2. Turn the ComWheel to make a change.
3. Push the ComWheel to activate (confirm) the change.
M1145956 3-7
Avance
3-8 M1145956
3 Operation
Gas setup
Access the Gas Setup menu by pushing the Gas Setup key. Use
the Gas Setup menu to adjust the O2% and total flow, to change the
balance gas, and to change the circuit type.
Gas Setup
Adjust Settings
Gas:
N2O
Air
Circuit:
Circle
Non-Circle
Normal Screen
Using quick keys The O2% and total flow can easily be changed using the gas quick
keys.
1. Push a gas quick key to select the corresponding gas setting.
2. Turn the ComWheel to make a change.
3. Push the ComWheel to activate (confirm) the change.
M1145956 3-9
Avance
3-10 M1145956
3 Operation
Spirometry
There are three types of spirometry loops: Pressure-Volume (P-V),
Flow-Volume (F-V) and Pressure-Flow (P-F). View, save, and erase
spirometry loops and set the loop scaling in the Spirometry menu.
Access the Spirometry menu by pushing the Vent Setup key and
selecting Spirometry.
Use the Spirometry menu to:
• View a specific loop type; select Loop Type and set the loop
type.
• Adjust the loop scaling; select Scaling and set the scale type.
• Access the Spirometry Setup menu; select Spiro Setup.
• Store a loop to memory; select Save Loop. Up to six loops can
be saved.
• View a saved loop; select Reference Loop and the time at which
the loop was saved.
• Erase a saved loop; select Erase Loop and the time at which the
loop was saved.
Spirometry
Loop Type P-V
Scaling
Spiro Setup
Save Loop
Reference Loop None
Erase Loop None
Previous Menu
M1145956 3-11
Avance
Setting loop type The spirometry loops show in the spirometry window and can be set
to show alongside the waveforms as the split screen.
AB.98.039
1. Volume axis
2. Pressure axis
3. Real-time loop
4. Reference loop (appears on display in white)
Figure 3-2 • Example of a P-V loop
3-12 M1145956
3 Operation
Scaling The scales shown in the spirometry loop graph are set in the Scaling
menu. The available settings for the volume, Paw, and flow graph
axes are dependent on the set patient type of adult or pediatric.
The AUTO selection automatically adjusts the volume, Paw, and flow
axes of the loop graph based on the minimum and maximum breath
reading shown in the waveform.
The Linked selection links the adjustment of the volume, Paw, and
flow axes of the loop graph together. Change one of the scales and
the remaining two scales automatically change based on the one set
scale.
The Indep selection allows the axes of the loop graph to be changed
separately for the volume, Paw, and flow axes.
Scaling
Scaling AUTO
Vol Scale
Paw Scale
Flow Scale
Previous Menu
M1145956 3-13
Avance
Spirometry setup
Push the Vent Setup key and select Spirometry - Spiro Setup to
access the Spirometry Setup menu. Use the Spirometry Setup
menu to:
• Set the patient and sensor type.
• Select a data source.
• Set the loop type.
• Change the volume shown on the spirometry split screen to
MVexp or TVexp.
• Set split-screen view.
• Access the Paw and MVexp alarm limits.
Note The Spirometry Setup menu is also accessible through Main Menu
- Parameters Setup - Spirometry Setup.
Spirometry Setup
Patient and Sensor Type Adult
Data Source Vent
Loop Type P-V
TV or MV TV
Split Screen None
Paw Alarm
MVexp Alarm
Previous Menu
Setting patient and Patient and sensor type refer to the style of airway adapter used with
sensor type the airway module. If spirometry data is obtained from the airway
module, ensure that the sensor type matches the airway adapter
used.
If the sensor type is not set correctly, the information displayed may
not be accurate.
1. Push the Vent Setup key.
2. Select Spirometry.
3. Select Spiro Setup - Patient and Sensor Type.
4. Select Adult or Pedi depending on the sensor used.
• Adult refers to the D-lite sensor.
• Pedi refers to the Pedi-lite sensor.
3-14 M1145956
3 Operation
Selecting a data Several monitoring parameters can be obtained from the ventilator or
source the airway module.
When setting the airway module as the data source, make sure that a
D-lite or Pedi-lite sensor is properly connected on the airway module.
If the sensor is not properly connected, but the airway module is
installed, the waveform shows no flow. The gas monitor samples and
displays room air.
If information is not available through the airway module, information
comes from the internal ventilator sensors. If the internal ventilator
sensor readings disagree with the airway module sensor readings,
the associated alarm occurs.
Information that is retrieved from the airway module is identified with
the module data indicator. See the “Airway Modules” section for more
information.
Setting spirometry Spirometry loops can be viewed alongside the waveforms on the
split screen normal screen. To set up the spirometry split screen:
1. Push the Vent Setup key.
2. Select Spirometry.
3. Select Spiro Setup.
4. Select Split Screen - Spiro.
5. Push the Normal Screen key.
M1145956 3-15
Avance
Main Menu
Menus and settings available through the Main Menu include
Trends, System Status, Cardiac Bypass, Fresh Gas Usage,
Screen Setup, Parameters Setup, and Calibration.
System Status shows the status of gas supplies, electrical supplies,
and software settings.
See “Screen configuration” for information on the Screen Setup
menu item.
See “Parameters setup” in the “Airway Modules” section for more
information on the Parameters Setup menu item.
See the “User Maintenance” section for more information on
calibrations.
Main Menu
Trends
System Status
Cardiac Bypass Off
Fresh Gas Usage
Screen Setup
Parameters Setup
Calibration
Normal Screen
3-16 M1145956
3 Operation
Trends There are three views for patient trends: measured (numerical),
settings, and graphical. Trend information is saved every 1 minute for
the most recent 24 hours.
Trends
Cursor
Next Page
View
Measured
Settings
Graphical
Time Scale
Previous Menu
M1145956 3-17
Avance
Cardiac bypass There are two types of cardiac bypass. Manual ventilation cardiac
bypass is standard. VCV cardiac bypass is optional.
Manual ventilation Manual ventilation cardiac bypass suspends alarms for patients on
cardiac bypass cardiac bypass when the ventilator is not mechanically ventilating.
The volume, apnea, low agent, CO2, and respiratory rate alarms are
suspended. The alarms are enabled when cardiac bypass is turned
off or mechanical ventilation is started.
VCV cardiac bypass Systems with VCV cardiac bypass enabled can mechanically
(optional) ventilate while in VCV mode. The VCV mode is the only ventilation
mode available while using VCV cardiac bypass. The volume, apnea,
low agent, CO2, low Paw, and respiratory rate alarms are suspended.
The alarms are enabled when VCV cardiac bypass is turned off or
mechanical ventilation is stopped.
3-18 M1145956
3 Operation
Fresh gas usage Use the Fresh Gas Usage selection to view the volume of O2, Air,
N2O, and agents used for the three most recent cases. Data only
shows for gases available on the system. Agent data shows the three
most recently used agents.
1. Push the Main Menu key.
2. Select Fresh Gas Usage.
3. Select the patient case to view.
4. Push the Normal Screen key or select Previous Menu.
M1145956 3-19
Avance
Screen configuration
Each case type selected on the Start Case menu has a
corresponding screen configuration for the normal screen. Four
additional normal screen views are available through the Select Page
menu. The case type screen configurations and the normal screen
views are set by the Super User in the Install/Service menu.
Areas of the normal screen can be customized to show specific
information. Use the Screen Setup menu to customize the normal
screen view.
Select page Use the Select Page menu to quickly change or return the normal
screen view to a case type view.
Select Page
Page:
Default Screen
PAW TUBES
LOOPS
BIG WAVE
LOCAL
Normal Screen
3-20 M1145956
3 Operation
Screen setup Waveform, digit field, fresh gas control style, split screen, sweep
speed, and brightness are adjusted in the Screen Setup menu.
Changes made to all settings (except Brightness) during a case are
not saved when the case is ended. The set Brightness remains until
it is changed again or until the system power is turned off.
Screen Setup
Waveform Field 1 Paw
Waveform Field 2 Flow
Waveform Field 3 CO2
Digit Field AA
Fresh Gas Controls O2%
Split Screen None
Sweep Speed Fast
Brightness 4
Install/Service
Previous Menu
Waveform fields The waveforms can be set to show agent, CO2, flow, Paw, or can be
set to off. Waveforms cannot be set to the same value (except for off).
If waveform is set to the same value as another waveform, the
previously set waveform changes to off and is removed from the
normal screen
1. Push the Main Menu key.
2. Select Screen Setup.
3. Select the waveform field and make the change.
4. Push the Normal Screen key or select Previous Menu.
Digit field The digit field can be set to show gas supply, flow, or agent.
1. Push the Main Menu key.
2. Select Screen Setup - Digit Field.
3. Select the Supply, Flow, or AA.
• If AA is selected, an airway module must be inserted or the
digit field will be blank.
4. Push the Normal Screen key or select Previous Menu.
M1145956 3-21
Avance
Setting gas controls and There are two ways to configure gas controls: O2% with total flow or
screen configuration individual gas flow. The Super User can set the gas control for the
facility. Or the Super User can set the gas control to allow the user to
select the setting.
Selecting O2% shows O2% as the first quick key and total flow l/min
as the second quick key. This gas control configuration adjusts the
balance automatically when either the O2% or the total flow is
changed.
Selecting Flow shows balance gas l/min as the one quick key and O2
l/min as the other quick key. This gas control/screen configuration
allows individual control of the gasses.
1. Push the Main Menu key.
2. Select Screen Setup - Fresh Gas Controls.
3. Select the desired view.
4. Push the Normal Screen key or select Previous Menu.
Split screen Gas and agent delivery, trends, spirometry loops, or a Paw gauge
can be shown as a split screen alongside the waveforms on the
normal screen. To change the normal screen to show a split screen:
1. Push the Main Menu key.
2. Select Screen Setup - Split Screen.
3. Select the desired view.
4. Push the Normal Screen key or select Previous Menu.
Sweep speed The waveform draw rate can be set to fast (6.25 mm/s) or slow
(0.625 mm/s). When the sweep speed is changed, the waveforms are
redrawn at the new rate.
1. Push the Main Menu key.
2. Select Screen Setup - Sweep Speed.
3. Set the speed.
4. Push the Normal Screen key or select Previous Menu.
3-22 M1145956
3 Operation
Alarm setup
Alarm limits, alarm volume, and other alarm settings are adjusted in
the Alarm Setup menu. Alarm history is also accessed through this
menu. Selecting Default Limits loads the default settings as set by
the Super User or the factory defaults if no Super User settings have
been entered.
Setting Leak Audio to Off silences audio alarms for small leaks. Leak
Audio is automatically set to On and cannot be changed when either
the Low MV alarm limits are off or the MV/TV Alarms is set to Off.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
The apnea time delay range is 10 to 30 seconds. Set the Apnea
Time in 1 second increments.
Alarm Setup
Adjust Settings
Volume Apnea Off
MV/TV Alarms On
Leak Audio On
CO2 Alarms On
Auto MV Limit Off
Default Limits
Alarm Volume 3
Apnea Time 30
Alarm History
Previous Menu
M1145956 3-23
Avance
Volume apnea Volume Apnea is not shown on the Alarm Setup menu when the
Volume Apnea Selection is set to Disable in the Install/Service
menu.
Set Volume Apnea to Off to disable the volume apnea alarm during
manual ventilation. ‘Volume Apnea Off’ shows in the general
message field. The volume apnea alarm remains disabled until the
Bag/Vent switch is set to ventilator or Volume Apnea is set to On.
1. Push the Alarm Setup key.
2. Select Volume Apnea and set to On or Off.
3. Push the Normal Screen key or select Previous Menu.
Note When Volume Apnea is set to Off during manual ventilation and then
mechanical ventilation is started, the volume apnea alarms are active.
Volume Apnea - Off requires confirmation through a pop-up window
when manual ventilation is resumed.
MV/TV alarms The MV and TV alarms can be turned off. When the volume alarms
are disabled, ‘MV/TV Alarms Off’ appears in the general message
field. The volume alarm limits waveform numerics show as dashes
during a case.
If the MV and TV alarms are set to Off during manual ventilation, the
alarms remain off until the Bag/Vent switch is set to ventilator or the
MV/TV Alarms is set to On.
The MV and TV alarms retain the last value set during mechanical
ventilation. If MV/TV Alarms is set to On during mechanical
ventilation, it remains on when manual ventilation starts. If MV/TV
Alarms is set to Off during mechanical ventilation it remains off when
manual ventilation starts.
1. Push the Alarm Setup key.
2. Select MV/TV Alarms and set to On or Off.
3. Push the Normal Screen key or select Previous Menu.
CO2 alarms CO2 information is obtained from the airway module installed in the
anesthesia system module bay. The CO2 alarms setting has no affect
if there is no airway module in the anesthesia system.
Set CO2 Alarms to Off during manual ventilation to disable the CO2
‘Apnea’ alarm, ‘EtCO2 low,’ ‘EtCO2 high,’ and ‘FiCO2 high.
Absorbent OK?’ alarms. The ‘CO2 Alarms Off’ message shows in the
general message field. The CO2 and O2 alarm limits waveform
numerics show as dashes during a case.
If CO2 Alarms is set to Off, the alarms remain disabled until the Bag/
Vent switch is set to ventilator, the case is ended, or the CO2 Alarms
is set to On.
1. Push the Alarm Setup key.
2. Select CO2 Alarms and set to On or Off.
3. Push the Normal Screen key or select Previous Menu.
3-24 M1145956
3 Operation
M1145956 3-25
Avance
Alternate O2 control
3
AB.91.077
1. ON/OFF button
2. Flow tube
3. Flow control
Figure 3-4 • Alternate O2 control
3-26 M1145956
3 Operation
AB.82.042
To return to absorber mode, reinsert the canister into the holder and
push the canister back up and snap it into absorber position. When
the canister is in the absorber position, the exhaled gas flows through
the absorber, removing CO2.
Note Check the absorber canister to ensure it has side rails. If the canister
does not have side rails, it will not work in the EZchange canister
holder.
1
AB.74p.043
1. Side rails
M1145956 3-27
Avance
Condenser (optional)
The condenser removes water in the system that is produced from
the reaction of CO2 gas with the absorbent. When fresh gas flow
settings of less than the patient minute volume are used during
ventilation, the amount of re-breathed gas increases. More CO2 flows
through the breathing system increasing the absorbent consumption.
The moisture buildup in the breathing system is also increased.
The condenser is connected between the outlet of the absorber
canister and the inlet of the circuit module. Moisture in the gas is
condensed into water droplets, which run into the condenser’s
reservoir.
Visually check the condenser reservoir daily. Drain the reservoir daily.
1. Place a container under the reservoir.
2. Push the drain button to empty any water in the condenser.
AB.75p080
2 1
1. Drain button
2. Reservoir
3. Condenser
Figure 3-5 • Condenser
3-28 M1145956
3 Operation
M1145956 3-29
Avance
3-30 M1145956
3 Operation
Connecting active To use the optional active AGSS on a system that has a flow
AGSS with a flow indicator, connect it as follows.
indicator 1. Connect the proper hose to the AGSS outlet connector on the
bottom of the AGSS underneath the breathing system. Attach the
other end to the hospital disposal system.
2. With the AGSS operating, verify that the flow indicator ball on the
flow indicator rises to the green zone, indicating adequate flow.
AB.91p093
Note The ball in the upper red zone indicates excessively high extraction
flow. The ball in the lower red zone indicates extraction flow rate is
too low or a blocked filter.
3. Complete the tests in the “Preoperative Tests” section of this
manual.
Connecting active The active adjustable AGSS option flow rate is limited to 30 l/min with
adjustable AGSS this option.
To use the optional active AGSS installed on the system which uses
the three-liter bag as a visual indicator, connect it as follows:
1. Connect a disposal hose to the DISS connector on the needle
valve on the bottom of the AGSS (underneath). The hose should
be flexible and reinforced to help prevent kinking and crushing.
2. Attach the other end of the hose to the hospital disposal system.
3. Attach the three-liter bag to the 30 mm auxiliary 1 port on the
bottom of the AGSS.
M1145956 3-31
Avance
4. Use the needle valve to adjust the flow rate to match the amount
of gas being scavenged. Use the visual indicator bag when
adjusting the flow rate. The bag should remain partially inflated
when the flow rate is adequate.
AB.75p.110
5. Complete the tests in the “Preoperative Tests” section of this
manual.
3-32 M1145956
4 Preoperative Checkout
M1145956 4-1
Avance
Check that the equipment is not damaged and that components are
correctly attached.
Check that pipeline gas supplies are connected and cylinders are
installed.
Set the appropriate controls and alarm limits for the case.
4-2 M1145956
4 Preoperative Checkout
Set the appropriate controls and alarm limits for the case.
M1145956 4-3
Avance
4-4 M1145956
5 Preoperative Tests
M1145956 5-1
Avance
5-2 M1145956
5 Preoperative Tests
Vaporizer installation
M1145956 5-3
Avance
5-4 M1145956
5 Preoperative Tests
Checkout menu
The Checkout menu shows on the display after turning on the
system. To access the Checkout menu between cases, push the
Checkout key. Step-by-step instructions show in the right window
next to the Checkout menu during the checks. Use the Checkout
menu to:
• Perform a Machine Check.
• Perform any of the individual checks.
• Set the Leak < 250 ml setting.
• View the Check Log.
• Start a case.
Checkout
Machine Check
Leak < 250 ml No
Individual Checks
System
Circuit
Circuit O2 Cell
Low P Leak
Check Log
Start Case
No When No is selected, the leak test will pass for leaks below 250 ml at
3 kPa (30 cmH2O) pressure with no user interaction required. For
leaks between 250 ml and 750 ml, the user can fix the leak and rerun
the test or accept the leak and continue. For leaks above 750 ml, the
test will fail and the user must fix the leak and rerun the test.
Yes Set to Yes to measure small leaks above 100 ml during the checkout
procedures. Selecting Yes will display the measured leak at 3 kPa
(30 cmH2O) pressure and result in the test taking somewhat longer.
M1145956 5-5
Avance
Machine check
The Machine Check or the individual checks must be performed at
least once within every 24-hour period.
Perform the Machine Check at the start of each day. The machine
check runs automatically and beeps to indicate when it is finished or if
interaction is required.
The Machine Check does a Machine Check-System check,
Machine Check-Circuit check, and a Machine Check-Circuit O2
cell check (if circuit O2 cell is present). When one of the checks is
completed, the next check begins.
1. Turn the System switch to On.
2. Select Machine Check and follow the instructions.
3. If a check fails, follow the instructions to perform a recheck or
accept the results.
4. When the Machine Check is completed, start a case.
Note In case of a patient emergency, the Machine Check may be
bypassed by selecting Start Case from the Checkout menu. The
general message ‘Please Do Checkout’ is displayed if a Machine
Check is not completed with passing results within 24 hours.
Machine check - The Machine Check-System checks the Bag/Vent switch, proper
system gas supply pressures, ventilator operation and leak, battery and
electrical power, circuit compliance, and flow control operation. This
is a two-step check.
1. Set the Bag/Vent switch to Vent.
2. Open the patient Y.
3. (ACGO option only.) Set the ACGO switch to Circle.
4. Select Start. The display shows the checks being run.
• The system beeps when this portion of the check is done.
• The results are shown on the display.
5. Make sure the bellows is fully collapsed.
6. Occlude the patient Y.
7. Select Continue. The display shows the checks being run.
8. When the check passes, the next check starts.
5-6 M1145956
5 Preoperative Tests
Machine check - The Machine Check-Circuit checks the Bag/Vent switch, proper gas
circuit supply pressures, airway pressure measurement transducer, APL
valve, and manual circuit leak.
1. Occlude the patient Y.
2. Set Bag/Vent switch to Bag.
3. Set the APL valve halfway between 30 and 70.
4. (ACGO option only.) Set the ACGO switch to Circle.
5. Select Start. The display shows the checks being run.
• The system beeps when the check is done.
• The results are shown on the display.
6. When the check passes, the next check starts.
Machine check - When External Gas Monitor is set to Yes by the Super User, the
monitor Machine Check-Monitor check occurs. This check is not a test. This
check is a reminder to connect a respiratory gas monitor.
M1145956 5-7
Avance
Individual checks
The Machine Check or the individual checks must be performed at
least once within every 24-hour period.
Individual checks allow the user to perform any combination of single
checks. These checks are helpful if there is a specific problem/alarm
and the user wishes to test only that portion of the system.
The checks do not automatically move on to the next check. After
completing a check, do another check or start a case. If a check fails,
follow the instructions to perform a recheck or accept the results.
System The System check checks the Bag/Vent switch, proper gas supply
pressures, ventilator operation and leak, battery and electrical power,
circuit compliance, and flow control operation. This is a two-step
check.
1. Set the Bag/Vent switch to Vent.
2. Open the patient Y.
3. (ACGO option only.) Set the ACGO switch to Circle.
4. Select Start. The display shows the checks being run.
• The system beeps when this portion of the check is done.
• The results are shown on the display.
5. Make sure the bellows is fully collapsed.
6. Occlude the patient Y.
7. Select Continue. The display shows the checks being run.
8. When the check passes, select Back.
9. Select another check or select Start Case to go to the Start Case
menu.
Circuit The Circuit check checks the Bag/Vent switch, proper gas supply
pressures, airway pressure measurement transducer, APL valve, and
manual circuit leak.
1. Occlude the patient Y.
2. Set Bag/Vent switch to Bag.
3. Set the APL valve halfway between 30 and 70.
4. (ACGO option only.) Set the ACGO switch to Circle.
5. Select Start. The display shows the checks being run.
6. When the check passes, select Back.
7. Select another check or select Start Case to go to the Start Case
menu.
5-8 M1145956
5 Preoperative Tests
Low P leak The positive pressure Low P Leak check measures machine leaks
before the breathing system, between the common gas outlet and the
high pressure pneumatics and includes the gas mixer and vaporizer.
It measures low pressure pneumatic leaks with a pass or fail limit of
50 ml.
1. Make sure that the vaporizers to be used during the case are
mounted on the machine and turned off.
2. Occlude the inspiratory (right-hand) port.
3. Select Start.
4. The display shows the checks being run. The system beeps
when the check is done.
5. Repeat the check once for each vaporizer with one vaporizer
turned on at a time.
6. When the checks pass, turn the vaporizer off.
7. Open the inspiratory port and reconnect the breathing circuit.
Select another check or select Start Case to go to the Start Case
menu.
Low P leak The negative Low P Leak check measures machine leaks before the
(machines with breathing system, between the common gas outlet and the high
pressure pneumatics and includes the gas mixer and vaporizer. It
ACGO) measures low pressure pneumatic leaks with a pass/fail limit of 50 ml.
1. Make sure that the vaporizers to be used during the case are
mounted on the machine and turned off.
2. Make sure the ACGO switch is set to ACGO.
3. Attach the squeeze bulb to the ACGO outlet.
4. Squeeze (collapse) the bulb.
5. If the bulb inflates in less than 30 seconds, select Fail.
6. If the bulb remains collapsed, repeat the check once for each
vaporizer with one vaporizer turned on at a time.
7. When the checks pass, turn the vaporizers off and remove the
squeeze bulb from the ACGO outlet.
M1145956 5-9
Avance
CAUTION If the needle valve is not fully open, this test can damage
the pressure gauge on the test device.
3. Turn the ACGO switch to the ACGO position.
4. Set the O2 flow to 500 ml.
5. Make sure that the total flow through the flowmeter on the test
device is 0.5 l/min.
6. Make sure that the pressure gauge on the test device reads zero.
7. Close the needle valve on the test device until the test gauge
reads 20 kPa (3 psi) (BSI) or 3 kPa (0.4 psi) (ISO).
8. If the flow through the test device is less than 0.45 l/min (ISO)
or 0.4 l/min (BSI), there is a low pressure leak in the anesthesia
machine. See the “Alarms and Troubleshooting” section for more
information.
9. Repeat this low-pressure leak test for each vaporizer.
• Set the applicable vaporizer to 1%.
• Turn the vaporizer off after the test.
WARNING Agent mixtures from the low-pressure leak test stay in the
system. Always flush the system with O2 after the low-
pressure leak test (1 l/min for at least one minute).
5-10 M1145956
5 Preoperative Tests
WARNING Anesthetic agent comes out of the circuit during this test.
Use a safe, approved procedure to collect and remove the
agent.
1. Set the System switch to On.
2. Start a case.
3. Set the O2 flow to 6 l/min.
4. Slowly adjust the vaporizer concentration from 0 to 1%.
• Make sure that the O2 flow stays constant.
• Verify that the system continues to operate without issuing
any related alarms.
5. Repeat this test for both vaporizer positions.
M1145956 5-11
Avance
5-12 M1145956
6 Airway Modules
M1145956 6-1
Avance
Airway modules
The optional compact airway modules measure and monitor gases
delivered to the patient and exhaled through the breathing circuit. The
modules consist of an infrared sensor for measuring CO2, N2O, and
anesthetic agents; a paramagnetic O2 sensor; and a gas sampling
system with the D-fend water separation system.
Systems with both an airway module and an O2 cell will display the
patient inspired O2 value obtained from the airway module.
Respiratory rate is the frequency of peak (end tidal) CO2
measurements per minute. A breath is defined as a change in the
CO2 signal that exceeds 1% (8 mmHg). All concentrations are
measured and displayed breath by breath.
6-2 M1145956
6 Airway Modules
Use only airway modules that have anesthetic agent monitoring and
O2 monitoring on this system. The following modules can be used on
this system: E-CAiO, E-CAiOV, E-CAiOVX, M-CAiO, M-CAiOV, and
M-CAiOVX. (E series modules must be software version 4.5 and
above. M series modules must be software version 3.2 and above.)
Letters in the name of the airway modules stand for:
• E = plug-in gas module
• M = plug-in gas module
• C = CO2 and N2O
• A = anesthetic agents
• i = agent identification
• O = patient O2
• V = patient spirometry
• X = Gas exchange
2
3
4
6
5 MD.60.008
M1145956 6-3
Avance
3
AB.91.010
1. Airway module
2. Gas sampling line
3. Airway adapter with sampling line connector
Figure 6-2 • Airway gases setup with compact airway module
6-4 M1145956
6 Airway Modules
Parameters setup
Use the Parameters Setup menu to change the monitoring settings
of the data source, CO2, O2, agent, and spirometry. Push the Main
Menu key. Select Parameters Setup.
Data source Several monitoring parameters can be obtained from the ventilator or
the airway module. Information that is retrieved from the airway
module is identified with the module data indicator.
Set the Data Source to Patient or Vent to select the primary source
for information. If Patient is selected, the airway module will be the
first source for information. If Vent is selected, the internal sensors of
the ventilator will be the first source for information.
When setting the airway module as the data source, make sure that a
D-lite or Pedi-lite sensor is properly connected on the airway module.
If the sensor is not properly connected, but the airway module is
installed, the waveform shows no flow. The gas monitor samples and
displays room air.
If information is not available through the airway module, information
comes from the internal ventilator sensors. If the internal ventilator
sensor readings disagree with the airway module sensor readings,
the associated alarm occurs.
CO2 setup Change the size of the CO2 waveform by changing the scale height.
The scale units (%, kPa, and mmHg) are set by the Super User.
Select CO2 Alarm or Resp Rate Alarm to access and change the
alarm limits for the corresponding alarm.
Agent setup Change the size of the agent waveform by changing the scale height.
Select Agent Alarm to access and change the agent alarm limits.
Spirometry setup Change the size of the Paw and Flow waveforms by changing the
corresponding scale heights. Select Paw Alarm or MVexp Alarm to
access and change the alarm limits for the corresponding alarm.
M1145956 6-5
Avance
Calibration
Calibrate airway modules once every six months or whenever there
are indications of errors in the gas readings. Use a Datex-Ohmeda
calibration gas and regulator to calibrate the modules. See the
“Parts” section for the stock numbers of the calibration gas and
regulator.
6-6 M1145956
6 Airway Modules
M1145956 6-7
Avance
6-8 M1145956
7 Alarms and Troubleshooting
M1145956 7-1
Avance
Alarms
Alarms are divided into technical alarms and parameter alarms.
These alarms may be high priority, medium priority, or informational.
When an alarm occurs during a case, an alarm tone sounds and the
alarm message is displayed in the alarm message field.
Technical alarms result from a technical problem and occur whether
or not a patient is connected to the system. Parameter alarms are
calculated limits and limits that are set by the user on the Alarm
Setup menu. Parameter alarms occur only during a case.
Alarm priorities Alarm priority is indicated by the color of the alarm message and the
alarm LED located next to the alarm silence button.
High-priority alarm messages appear in white text on a red
background. During a high-priority alarm, the red LED flashes.
Medium-priority alarm messages appear in yellow text on a gray
background. During a medium-priority alarm, the yellow LED flashes.
Informational alarms appear in white text on a gray background.
During an informational alarm, the yellow LED is on solid.
When a high-priority alarm is active at the same time as a medium-
priority alarm or an informational alarm, the red and yellow LEDs
flash. When a high-priority alarm is active at the same time as an
informational alarm, the red LED flashes and the yellow LED is on
solid. When a medium-priority alarm is active at the same time as an
informational alarm, the yellow LED flashes. Pushing the Silence
Alarms key changes the LED from flashing to on solid until the end
of the alarm silence.
Silencing alarms Silencing an alarm stops the audible tone for 120 seconds. The alarm
message shows in the alarm message field. Pushing the Silence
Alarms key when no medium or high priority alarms are active
suspends audible alarm tones for 90 seconds.
Alarms in the apnea alarm family have special silence behavior to
reduce apnea nuisance alarms. Apnea family alarms include ‘Apnea,’
‘TVexp low,’ ‘EtCO2 low,’ ‘MVexp low,’ and ‘RR low.’
When silencing an apnea family alarm, the audio tone for the active
alarm is silenced for 120 seconds. The audible tone for any additional
apnea family alarm that occurs during the silence period is silenced
for the remain time shown on the alarm silence countdown. Only the
audible alarm tone is silenced. The alarm messages still show in the
alarm message fields. ‘APN’ shows above the alarm silence
countdown when the audible tone silence is in affect for the apnea
family alarms.
Display changes Messages may appear in the waveform field during some alarms. If
during alarms more than one alarm has a message, the message for the highest
priority alarm is displayed. The message is removed when the alarm
is resolved.
7-2 M1145956
7 Alarms and Troubleshooting
The color of the alarm text shown in the alarm message fields is
dependent on the alarm priority. Messages for high-priority alarms
use red text. Messages for medium-priority alarms use yellow text.
Informational messages use white text.
When the O2 pipeline supply pressure drops to less than 252 kPa (36
psi), the lower-right corner of the display toggles between the O2
supply information and the set digit field.
Some patient parameter alarms, such as ‘Ppeak high’ and ‘FiO2 low,’
will latch when the alarm condition is corrected. When an alarm is
latched, it is displayed in white text on a black background. The
parameter box will stop flashing. All the associated waveform,
numeric, and digit field messages are removed from the display. The
flashing LED associated with that alarm changes from flashing to on
solid. The alarm will remain in this condition until it is acknowledged
by pushing the Silence Alarms key or until the alarm re-occurs.
When the alarm is acknowledged, it is removed from the screen. If an
alarm has latched and the alarm re-occurs before it is acknowledged,
the alarm will revert to an active state.
De-escalating alarms Some device related alarms, such as ‘Reverse Flow’ and ‘No insp
flow sensor,’ will de-escalate priority when the alarm is acknowledged
by pushing the Silence Alarms key. The audible alarm tone is
stopped for that active alarm until the alarm condition is resolved. The
alarm message shows at the informational alarm level until the alarm
condition is resolved and the alarm is cleared. If that alarm reoccurs
after it has been resolved, the alarm occurs at its standard priority
level.
Battery indicator The color and fill amount of the battery in use symbol indicates the
amount of battery power remaining. Green indicates greater than 10
minutes of battery power remaining. Yellow indicates less than 5
minutes battery power remaining.
Internal failure ‘Internal problem prevents normal operation.’ shows on the display
during a software or hardware failure that requires service. If this
message occurs, contact a Datex-Ohmeda trained service
representative.
M1145956 7-3
Avance
List of alarms
If the corrective action does not resolve the alarm message, contact a
Datex-Ohmeda trained service representative
Circuit pressures and volumetric flows are measured by the ventilator
and airway gas module. If the Data Source is set to Patient, the
displayed waveforms and numeric information are measured by the
airway gas module. Although not displayed, the ventilator
measurements continue and if a measured value violates an alarm
setting, the appropriate alarm occurs. The value highlighted in the
parameter numeric box may not appear to have violated the alarm
setting. Changing the Data Source to Vent will display the circuit
pressures and volumetric flows measured by the ventilator.
AA, CO2 Medium External Gas Monitor setting is Set External gas monitor setting to Yes if the
monitoring not set to No on the Install/Service system uses a stand-alone monitor for O2,
connected menu. External Gas Monitor AA, and CO2.
setting is No on the Install/
Service menu and the monitor
condition changes to
disconnected.
Air pressure low. High O2% is set to Air at 21%. Air Ensure the air pipeline and cylinder are
Increase O2 %. pipeline pressure is less than properly connected. Increase the O2% flow.
252 kPa (36 psi) and the air
cylinder pressure dropped below
2633 kPa (381 psi) for one
second.
Air supply Medium Air pipeline pressure is less than Ensure the air pipeline and cylinder are
pressure low 252 kPa (36 psi) and the air properly connected.
cylinder pressure dropped below Contact a Datex-Ohmeda trained service
2633 kPa (381 psi) for one representative.
second.
Apnea Medium Apnea time delay (10-30 Check for leaks in the patient circuit. Check
seconds) has passed without a for patient disconnection.
measured breath.
Apnea time delay (20 - 30
seconds) has passed without a
changed in the measured CO2
by at least 1%.
Apnea >120 s High Apnea time exceeds 120 Check for leaks or blockages in the breathing
seconds. circuit. Ensure the Bag/Vent switch is in the
Vent position. Check the patient.
Backup Mode Informational No spontaneous breaths in set Select a new ventilation mode.
active period of time (Backup Time) The number of consecutive patient triggered
and 30 seconds have passed breaths reaches the Exit Backup setting.
since starting PSVPro mode.
Breathing Informational The breathing system is not Push the breathing system onto the frame
system loose latched. and ensure it latches.
7-4 M1145956
7 Alarms and Troubleshooting
Calibrate Informational Flow calibration failure or more Calibrate the flow sensors.
(remove) flow than 24 hours have passed since
sensors the last flow sensor calibration.
Calibrate O2 Informational Calibration failure or measured Calibrate the O2 cell. Replace the O2 cell if
sensor O2 is greater than 110%. necessary.
Calibrate, dry, or Informational Patient volume mismatch Calibrate, dry or replace the flow sensors.
replace flow occurred during the last case. Start a new case.
sensors
Cannot monitor Medium Air pipeline pressure is invalid. Check pipeline supply pressure.
Air pipeline Contact a Datex-Ohmeda trained service
representative.
Cannot monitor Medium O2 pipeline pressure is invalid. Check pipeline supply pressure.
O2 pipeline
Cannot read gas Medium Transducer failure. Contact a Datex-Ohmeda trained service
supply representative.
pressures
Check circuit Medium Breaths detected in circle circuit Check circle circuit connections and settings.
connections while non-circle circuit is
selected.
Check D-Fend Medium Water trap not attached. Check that the water trap is properly attached
to the airway module.
Check flow Medium System has detected an Ensure the internal flow sensors are
sensors improper flow pattern in the connected correctly.
breathing circuit.
Check sample Medium Possible blockage in airway Check for blockage in the airway module
gas out module sample gas outlet. sample gas outlet. Remove blockage.
Circuit leak Medium Vent TVexp is less than half of Check for leaks in the patient circuit.
vent TVinsp for at least 30 Calibrate flow sensors. If problem persists,
seconds. replace flow sensors.
Circuit leak Informational Setting on Alarm Setup menu. Vent TVexp is less than 50% of vent TVinsp
silenced for at least 30 seconds. Message indicates
that the Leak Audio alarm is turned off.
Circuitry >75C Medium Power supply temperature Shut down system as soon as possible.
shutdown exceeds 75°C. Then, check cooling fans and filters.
possible
Cooling fan Medium Fan reporting error. Shut down system as soon as possible.
needs service. Then, check cooling fans and filters.
System OK.
Cooling fans Medium Fan reporting error. Shut down system as soon as possible.
failed. May Contact a Datex-Ohmeda trained service
overheat. representative.
M1145956 7-5
Avance
Display panel Medium Communication lost between Turn the system off and back on.
controls failure panel and key pad.
EtCO2 high High EtCO2 is greater than high alarm Check the patient and EtCO2 settings. Check
limit. if absorbent needs to be changed.
EtCO2 low Informational EtCO2 is less than alarm limit. Ensure the patient is properly intubated.
Medium after 1 Priority escalates to Medium if Check for leaks or blockages in the patient
minute alarm is not resolved after 1 circuit.
minute.
EtDES high Medium1 EtDES is greater than alarm Set the alarm limits appropriately. Decrease
limit. the agent concentration.
EtDES low Informational EtDES is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
EtENF high Medium1 EtENF is greater than alarm Set the alarm limits appropriately. Decrease
limit. the agent concentration.
EtENF low Informational EtENF is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
EtHAL high Medium1 EtHAL is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
EtHAL low Informational EtHAL is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
EtISO high Medium1 EtISO is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
EtISO low Informational EtISO is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
EtO2 high Medium EtO2 is greater than high alarm Set the alarm limits appropriately. Decrease
limit. the O2 concentration.
EtO2 low Medium EtO2 is less than low alarm limit. Set the alarm limits appropriately. Increase
the O2 concentration.
EtSEV high Medium EtSEV is greater than alarm Set the alarm limits appropriately. Decrease
limit. the agent concentration.
EtSEV low Informational EtSEV is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
FiCO2 high. High FiCO2 is greater than alarm limit. Check if absorbent needs to be changed.
Absorbent OK? Check the patient.
FiDES high Medium1 FiDES is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
7-6 M1145956
7 Alarms and Troubleshooting
FiDES low Informational FiDES is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
FiENF high Medium1 FiENF is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiENF low Informational FiENF is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
FiHAL high Medium1 FiHAL is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiHAL low Informational FiHAL is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
FiISO high Medium1 FiISO is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiISO low Informational FiISO is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
FiO2 high Medium FiO2 is greater than high alarm Check the O2 setting. Recalibrate the O2 cell
limit. and the airway module.
FiO2 low High FiO2 is less than low alarm limit. Check O2 setting. Check for leaks or
blockages in the patient circuit.
FiSEV high Medium FiSEV is greater than alarm limit. Set the alarm limits appropriately. Decrease
the agent concentration.
FiSEV low Informational FiSEV is less than alarm limit. Check the fill level on the vaporizers. Set
alarm limit appropriately. Increase the agent
concentration.
For mech vent, Informational Bag/Vent switch is set to Bag Move switch to the Vent position to start
set Bag/Vent and the ACGO switch is set to mechanical ventilation.
switch Circle.
Gas monitoring Medium Airway module hardware failure. Replace airway module. Then, turn power off
not available and back on between cases to clear the
alarm and receive module data.
Increase low MV Medium Low MV limit is off in SIMV/PSV, Increase Low MV alarm limit to improve
limit SIMV-PC, or PSVPro modes. patient disconnection detection.
Internal failure. High Power controller software failure. Contact a Datex-Ohmeda trained service
System may representative.
shut down.
Internal failure. Medium Power controller software failure. Contact a Datex-Ohmeda trained service
System may representative.
shut down.
M1145956 7-7
Avance
Module not Informational The monitoring module detected Remove the incompatible module. Use a
compatible is not compatible with system compatible module.
software.
Move Bag/Vent Medium Bag/Vent switch is in the wrong Move switch to the Bag position.
Switch to Bag position.
MVexp high Medium MVexp is greater than MVexp Change TV, RR, I:E, or PEEP to reset the
high alarm limit (for nine breaths minute volume to below the MVexp high
or one minute). alarm limit.
MVexp low Medium MVexp is less than MVexp low Change TV, RR, I:E, or PEEP to reset the
alarm limit (for nine breaths or minute volume to above the MVexp low alarm
one minute). limit.
N2O supply Medium N2O pipeline pressure is less Ensure the N2O pipeline and cylinder are
pressure low than 252 kPa (36 psi) and the properly connected.
N2O cylinder pressure is less
than 2633 kPa (381 psi).
Negative airway High Paw is less than -10 cm H2O. Check for blockages in the patient circuit.
pressure
No battery Medium Battery or charging failure. Between cases turn the system circuit
backup breaker off, then back on after 15 seconds to
reset the system.
No exp flow Medium Electrical signals show the flow Connect the flow sensor. Replace the flow
sensor sensor is not connected. sensor if necessary.
No fresh gas High Possible patient detected while Disconnect the patient or start a case.
flow! system is in checkout state.
No fresh gas High Possible fresh gas flow occlusion Switch to circle circuit or bag the patient.
flow? or loss of gas pressure. Check pipeline supply connection.
No insp flow Medium Electrical signals show the flow Connect the flow sensor. Replace the flow
sensor sensor is not connected. sensor if necessary.
O2 flush stuck Informational Switch is detected “on” Check flush valve. Ensure flush valve is not
on? continuously for more than 30 sticking.
seconds.
O2 monitoring Medium O2 cell not connected. Install airway gas module or connect the O2
not connected cell.
O2 supply High O2 pipeline pressure is less than Ensure the O2 pipeline and cylinder are
pressure low 252 kPa (36 psi) and the O2 properly connected.
cylinder pressure dropped below
2633 kPa (381 psi) for one
second.
7-8 M1145956
7 Alarms and Troubleshooting
PEEP high. High Paw greater than or equal to Check for blockages in the patient circuit.
Blockage? sustained limit for 15 seconds.2
Plug in power Medium The mains supply is not Ventilate manually to save power. Make sure
cable. On connected or has failed and the the power cable is plugged in and system
battery. system is using battery power. circuit breaker is on.
Ppeak high High Paw is greater than Pmax alarm Check for blockages in the patient circuit.
limit.
Ppeak low. Medium Peak airway pressure is less Check for leaks in the patient circuit.
Leak? than low Pmin + 4 cmH2O for 20
consecutive seconds if the set
respiratory rate is four or higher
and 35 seconds if the set
respiratory rate is less than four
breaths/min.
Replace exp Informational EEPROM calibration data read Replace the expiratory flow sensor.
flow sensor failure.
Replace insp Informational EEPROM calibration data read Replace the inspiratory flow sensor.
flow sensor failure.
Replace O2 Informational Measured O2 is less than 5%. Calibrate the O2 cell. Replace the O2 cell if
sensor necessary.
Reverse exp Medium Flow toward the patient seen in Check the flow sensor condition. Replace the
flow. Check the expiratory flow sensor during expiratory check valve between cases. Zero
valves OK? inspiration for six breaths in a flow transducers between cases.
row.
Reverse insp Medium Flow away from the patient seen Check the flow sensor condition. Replace the
flow. Check in the inspiratory sensor during inspiratory check valve between cases. Zero
valves OK? expiration for six breaths in a flow transducers between cases.
row.
RR high Medium RR is greater than high alarm Set the alarm limits appropriately or adjust
limit. the RR setting.
RR low Medium RR is less than low alarm limit. Set the alarm limits appropriately or adjust
the RR setting.
Sample line Medium Airway module sample line is Replace airway module sample line.
blocked blocked.
Set Alt O2 flow! Medium Software or hardware failure Contact a Datex-Ohmeda trained service
Check agent prevents mixed gas delivery. representative.
setting!
System leak? Informational Leak detected between Check for leaks in the breathing system.
ventilator and patient circuit.
M1145956 7-9
Avance
System High Remaining battery power is Plug in the power cable. Check that the
shutdown in <5 between zero and five minutes. system circuit breaker is on.
min
Turn power Off Informational System has been operating for Turn power off and back on between cases to
and On for self longer than 12 hours without a perform a self test.
tests power-up self test.
Turn switch on High System is in therapy when Turn the System switch to On to continue
to continue use System switch is turned to therapy. The system will return to normal
Standby. function. If System switch is not turned to On
within 8 seconds, the system will shut down.
TV not achieved Informational Measured tidal volume is less Check for leaks in the patient circuit. Check
than set tidal volume. for leaks in the breathing system.
TVexp high Medium TVexp is greater than TVexp Change TV, RR, I:E, or PEEP or change the
high alarm limit (for nine tidal volume to reset TVexp high alarm limit.
breaths).
TVexp low Medium TVexp is less than TVexp low Change TV, RR, I:E, or PEEP or change the
alarm limit (for nine breaths). tidal volume to reset TVexp low alarm limit.
Unable to drive Informational Bellows is collapsed. Check the drive gas. Increase fresh gas flow
bellows (or push the O2 flush button) to fill the
bellows.
Using battery. Medium Mains supply is OK, but the Shut down the system as soon as possible.
Power system is running on the battery. Contact a Datex-Ohmeda trained service
Controller fail. representative.
Ventilate High Software or hardware failure Use a manual bag to ventilate the patient or
manually! prevents mechanical ventilation. use an alternate machine. Shutdown the
No pressure, no flow, and no system as soon as possible and contact a
volume monitoring from Datex-Ohmeda trained service
ventilator. representative.
Ventilate Medium Software or hardware failure Use a manual bag to ventilate the patient or
manually! prevents mechanical ventilation. use an alternate machine. Shutdown the
Pressure, flow, and volume system as soon as possible and contact a
monitoring from ventilator still Datex-Ohmeda trained service
available. representative.
Ventilator has High Drive gas supply is not sufficient Check drive gas supply. Use a manual bag to
no drive gas to mechanically ventilate. ventilate the patient until the drive gas supply
is restored.
Vol and Apnea Informational Non circle circuit is selected. Message will clear when circle circuit is
monitoring off selected.
Vol vent only. Medium Manifold pressure error. Use volume control ventilation mode. Shut
No PEEP or Pressure control unavailable. down system as soon as possible. Contact a
PSV. Medium priority alarm when Bag/ Datex-Ohmeda trained service
Vent switch is in Vent and representative.
running PCV, PSVPro, or SIMV-
PC mode.
7-10 M1145956
7 Alarms and Troubleshooting
Vol vent only. Informational Manifold pressure error. Continue to use volume control ventilation
No PEEP or Pressure control unavailable. mode or ventilate manually. Shut down
PSV. Informational priority alarm when system as soon as possible. Contact a
Bag/Vent switch is in Vent and Datex-Ohmeda trained service
not running PCV, PSVPro, or representative.
SIMV-PC mode; Bag/Vent switch
is in Bag; or Non-circle circuit or
ACGO is selected.
Volume sensors Informational TVexp is greater than TVinsp for Calibrate the flow sensors. Replace the flow
disagree six breaths. sensors if the message does not clear.
1 The alarm priority escalates to High if the alarm is not resolved after 2 minutes.
2The sustained pressure threshold is calculated from the pressure limit setting. The sustained limit is calculated as
follows:
Mechanical For Pmax less than 30 cmH2O, the sustained pressure limit is 6 cmH2O.
Ventilation with For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of Pmax.
PEEP Off: For Pmax greater than to 60 cmH2O, the sustained pressure limit is 12 cmH2O.
Mechanical For Pmax less than 30 cmH2O, the sustained pressure limit is 6 cmH2O plus “set PEEP”
Ventilation with minus 2 cmH2O.
PEEP On: For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of Pmax plus “set
PEEP” minus 2 cmH2O.
For Pmax greater than 60 cmH2O, the sustained pressure limit is 12 cmH2O plus “set PEEP”
minus 2 cmH2O.
Mechanical For Pmax between 12 and 60 cmH2O, the sustained pressure limit is 50% of Pmax.
Ventilation Off: For Pmax greater than 60 cmH2O, the sustained pressure limit is 30 cmH2O.
M1145956 7-11
Avance
Alarm ranges
The alarms names are listed in order they appear on the Alarm
Limits page accessible from the Alarm Setup menu. The alarms
setting ranges show in the order they appear in the setting selections
in the Alarm Settings menu. See the “Super User Mode” section for
more information on the alarm default settings.
7-12 M1145956
7 Alarms and Troubleshooting
M1145956 7-13
Avance
Alarm tests
Test the system to verify that alarms are functioning.
Note If an airway module is installed, the FiO2 readings are taken from the
module instead of from the O2 cell. A sample line must be connected
from the airway module to the breathing circuit in order to test the O2
alarms.
1. Connect a test lung to the patient connection.
2. Start a case.
3. Set the Bag/Vent switch to Vent.
4. Set the O2 concentration to 30%, and allow the O2 reading to
stabilize.
• For machines configured to individual gas control, set the O2
flow to approximately 500 ml/min and Air flow to
approximately 5 l/min.
5. Test the O2 alarms:
• Set the FiO2 Low alarm limit to 50%. Make sure an FiO2 low
alarm occurs.
• Set the FiO2 Low alarm limit back to 21% and make sure
that the FiO2 low alarm cancels.
• Set the FiO2 High alarm limit to 50%.
• Push the O2 flush button.
• Make sure the FiO2 high alarm occurs.
• Set the FiO2 High alarm limit back to 100%. Make sure that
the FiO2 high alarm cancels.
6. Test the MVexp low alarm:
• Go to the Alarm Setup menu.
• Set the MV Low alarm limit to greater than the measured
minute volume.
• Make sure that a MVexp low alarm occurs.
• Set the MV Low alarm limit to off.
7. Test the Ppeak high alarm:
• Set the Pmax to less than the peak airway pressure.
• Make sure that the Ppeak high alarm occurs.
• Set the Pmax to the desired level.
8. Test the PEEP high. Blockage? alarm:
• Close the APL valve.
• Set the Bag/Vent switch to Bag. Mechanical ventilation stops.
• Block the patient connection and push the O2 flush button.
• Make sure that the PEEP high. Blockage? alarm occurs
after approximately 15 seconds.
7-14 M1145956
7 Alarms and Troubleshooting
M1145956 7-15
Avance
Filter blockage. Active systems Replace the filter. Refer to “Remove the
have a flow indicator. AGSS receiver filter” in the “Advanced
Breathing System Cleaning and Sterilization”
User’s Reference manual.
The bellows fills when the Bag/ Leak through Bag/Vent switch. Contact a Datex-Ohmeda trained service
Vent switch is set to Bag or the representative to repair the system.
bag fills when the switch is set to
Vent.
The ventilator does not read the Ventilator or absorber malfunction. Ventilate manually.
position of the Bag/Vent switch. Contact a Datex-Ohmeda trained service
representative to repair the system.
APL valve does not operate APL valve problem. Replace APL valve seal and diaphragm.
correctly.
Large breathing system leak not Bag hose not connected properly. Ensure that the bag hose is connected to the
quickly located (in bag mode). bag port (below the APL valve).
Absorber canister not installed Reinstall the absorber canister, ensure both
correctly. pins are engaged.
Bellows falls below top of Leak in the breathing system. Check, clean, or reposition the pressure
indicator during “Bellows relief valve. If the problem persists, replace
assembly test.” the pressure relief valve, bellows base, or
bellows assembly.
7-16 M1145956
7 Alarms and Troubleshooting
Electrical problems
M1145956 7-17
Avance
Pneumatic problems
7-18 M1145956
8 Setup and Connections
M1145956 8-1
Avance
Setup warnings
See “Standards” and “System components” in the
“Specifications and Theory of Operation” section for
information on specific monitoring requirements.
w Always make sure that the pipeline supply hoses and the
breathing circuit components are not toxic and will not:
• Cause an allergic reaction in the patient.
• React with the anesthetic gases or agent to produce
dangerous by-products.
8-2 M1145956
8 Setup and Connections
M1145956 8-3
Avance
Canister setup
The absorber canister is available in two versions: Disposable Multi
Absorber and Reusable Multi Absorber. Both are removed and
installed on the breathing system in the same way.
Each canister holds 800 grams of loose absorbent. The manufacturer
recommends MedisorbTM absorbent.
Both absorber versions should only be used with mixtures of air,
oxygen, nitrous oxide, halothane, enflurane, isoflurane, desflurane
and sevoflurane.
6 1
5 2
4
3
7
AB.74p043
AB.74p042
8-4 M1145956
8 Setup and Connections
When to change the A gradual color change of the absorbent in the canister indicates
absorbent absorption of carbon dioxide. The color change of the absorbent is
only a rough indicator. Use carbon dioxide monitoring to determine
when to change the canister.
Discard the absorbent when it has changed color. If left standing for
several hours, absorbent may regain its original color giving a
misleading indication of activity.
Read the absorbent manufacturer’s instructions completely before
using the product.
M1145956 8-5
Avance
Removing a canister 1. Hold the canister by the handle and push on the release latch to
unlock the canister.
AB.74p058
2. Remove the canister by tilting it downward and off the two
support pins.
Removing an 1. Hold the canister by the handle and push the canister cradle
EZchange canister release latch to unlock the canister cradle.
AB.75p088
8-6 M1145956
8 Setup and Connections
Reusable Multi 1. Turn the canister upside down and, using your thumbs, turn the
Absorber canister cover locking ring counterclockwise to unlock it.
filling
AB.74p044
2. Push up to release the seal.
3. Lift off the cover to remove it.
AB.74p046
4. Remove and properly discard the foam filters, the absorbent, and
any water in the reservoir.
AB.74p047
M1145956 8-7
Avance
8-8 M1145956
8 Setup and Connections
Electrical connections
Mains inlet Arrow shows the mains power inlet and cord.
AB.74p048
Outlets Labels show outlet voltage ratings and circuit breaker amp ratings.
These are isolated outlets. Regularly test the leakage current.
AB.91p048
M1145956 8-9
Avance
Serial port The system has an RS-232C electrical interface. The RS-232C
connector allows serial input/output of commands and data. The 15-
pin connector is located on the back of the display unit.
The 15-pin female D connector - Data Communications Equipment
configuration (DCE):
• Pin 1 - Monitor On/Standby
• Pin 5 - Signal ground
• Pin 6 - Receive data
• Pin 9 - Monitor On/Standby Return
• Pin 13 - Transmit data
AB.91p050
AB.91p085
8-10 M1145956
8 Setup and Connections
Pneumatic connections
CAUTION Use only medical grade gas supplies. Other types of gas
supplies may contain water, oil, or other contaminants
which could affect the operation of the pneumatic system.
The gas supplies provide gas to these devices through internal
connections:
• venturi suction regulator (optional).
• auxiliary O2 flowmeter (optional).
Pipeline inlets
Scavenging AB.91p040
The scavenging assembly is located below the bellows on the
breathing system. Adapters may be necessary to interface to the
scavenging connector.
See “Passive AGSS (optional)” and “Active AGSS (optional)” in the
“Operation” section for more scavenging information.
AB.91p045
M1145956 8-11
Avance
Sample gas return Connect the Datex-Ohmeda sample gas exhaust tube to the gas
port return port. Exhaust gas will be directed to the scavenging system.
AB.91p093
Vacuum suction The vacuum suction regulator uses an external vacuum supply.
regulator (optional) Connect the vacuum connection to the source vacuum supply.
Connect the collection bottle connection to the collection bottle.
3
2
AB.74p032
8-12 M1145956
8 Setup and Connections
Venturi suction The venturi suction regulator uses the system air or O2 supply
regulator (optional) source. Connect the collection bottle connection to the collection
bottle.
3
2
AB.74p189
1. Venturi muffler
2. Overflow safety trap
3. Splash guard
4. Collection bottle connection
Figure 8-3 • Venturi suction
Auxiliary O2
flowmeter (optional)
2
AB.91p039
1. Auxiliary O2 outlet
2. Auxiliary O2 flow control
Figure 8-4 • Auxiliary O2 flowmeter
M1145956 8-13
Avance
8-14 M1145956
8 Setup and Connections
M1145956 8-15
Avance
WARNING The top of the machine has a weight limit of 34 kg (75 lb).
WARNING Fully tighten the straps. If straps are not fully tightened,
equipment can fall off the top of the machine.
8-16 M1145956
9 User Maintenance
M1145956 9-1
Avance
Repair policy
Do not use malfunctioning equipment. Make all necessary repairs or
have the equipment serviced by a Datex-Ohmeda trained service
representative. After repair, test the equipment to ensure that it is
functioning properly, in accordance with the manufacturer’s published
specifications.
To ensure full reliability, have all repairs and service done by a
Datex-Ohmeda trained service representative. If this cannot be done,
replacement and maintenance of those parts listed in this manual
may be undertaken by a competent, trained individual having
experience in the repair of devices of this nature.
9-2 M1145956
9 User Maintenance
M1145956 9-3
Avance
AB.74p052
2. Pull the flow sensor module from the breathing system.
3. Remove the O2 cell cable connector from the O2 cell and
unscrew the cell counterclockwise.
AB.74p065
5. Replace the flow sensor module on the system and push the
latch closed to secure the module.
6. Calibrate the new O2 cell according to the “Circuit O2 cell
calibration” procedure.
9-4 M1145956
9 User Maintenance
Calibration menu
Access the Calibration menu by pushing the Main Menu key and
selection Calibration from the Main Menu. Select the calibration
procedure and follow the instructions shown.
See “Calibration” in the “Airway Modules” section for information on
calibrating the airway modules.
See “Backlight test” for information on testing the backlights.
Calibration
Flow and Pressure
Circuit O2 Cell
Airway Gas
Backlight Test
Previous Menu
M1145956 9-5
Avance
100% O2 calibration 1. Ensure patient Y-piece is not plugged or there is no patient tubing
connected to the system.
2. (ACGO option only.) Set the ACGO switch to Circle.
3. Set the Bag/Vent switch to Vent.
4. Select 100% O2 and push the ComWheel to start the calibration.
5. System will flow O2 to calibrate.
9-6 M1145956
9 User Maintenance
Backlight test
1. Push the Main Menu key.
2. Select Calibration.
3. Select Backlight Test.
4. Select Start Test.
5. The display will show the test running on light 1 and then on light
2. If the display goes completely blank or flickers during the test,
one of the lights has failed. Contact a Datex-Ohmeda trained
service representative to replace the backlights.
M1145956 9-7
Avance
9-8 M1145956
10 Parts
Note This section lists user-replaceable parts only. For other components,
refer to the Technical Reference manual.
M1145956 10-1
Avance
AB.82.019
2
10-2 M1145956
10 Parts
AB.82.021
5
M1145956 10-3
Avance
Bellows
3
7
4
5
AB.82.018
10-4 M1145956
10 Parts
AB.82.001
Item Description Stock number
Complete breathing system assembly includes:
• Flow sensor module (does not include flow sensors)
• Breathing circuit module (does not include APL valve,O2 cell, o-ring,
or cable)
• Complete bellows base
• Canister not included
- Autoclavable breathing assembly 1407-7008-000
- Autoclavable breathing assembly, Australia 1407-7016-000
M1145956 10-5
Avance
Absorber canister
3
1
AB.82.017
Item Description Stock number
1 Multi absorber, reusable (includes 40 pack of 1407-7004-000
foam) (does not include absorbent)
2 Cover assembly, CO2 canister 1009-8240-000
3 Foam, CO2 canister (pack of 40) 1407-3201-000
4 O-ring 1407-3204-000
5 Canister, CO2 with handle 1407-3200-000
- Multi absorber, disposable, white to violet, (pack 8003138
of 6)
- Multi absorber, disposable, pink to white (pack 8003963
of 6)
10-6 M1145956
10 Parts
M1145956 10-7
Avance
AGSS
10-8 M1145956
10 Parts
1
3
2 4
AB.82.044
AB.82.057
Item Description Stock number
1 EZchange canister module, includes valve and 1407-7021-000
cap
2 Valve 1407-3126-000
3 Cap 1407-3130-000
4 Condenser 1407-7024-000
- EZchange canister module with condenser 1407-7027-000
M1145956 10-9
Avance
Condenser
AB.82.045
Item Description Stock number
Condenser assembly (includes module and condenser) 1407-7026-000
1 Condenser module 1407-7025-000
2 Condenser 1407-7024-000
10-10 M1145956
10 Parts
M1145956 10-11
Avance
10-12 M1145956
11 Specifications and Theory of
Operation
Note All specifications are nominal and subject to change without notice.
Note All displayed values are shown at ambient temperature and pressure
dry.
M1145956 11-1
Avance
AB.91.103
11-2 M1145956
11 Specifications and Theory of Operation
M1145956 11-3
Avance
Gas supplies Pressurized gas supplies enter the system through a pipeline or
cylinder connection. All connections have indexed fittings, filters, and
check valves.
A regulator decreases the cylinder pressures to the appropriate
system pressure. A pressure relief valve helps protect the system
from high pressures.
To help prevent problems with the gas supplies:
• Install yoke plugs on all empty cylinder connections.
• When a pipeline supply is connected, keep the cylinder valve
closed.
• Disconnect the pipeline supplies when the system is not in use.
Air and N2O Pipeline or regulated cylinder pressure supplies Air directly to the air
channel of the gas mixer and to the ventilator if Air is configured as
drive gas. If the pressure is too low, an alarm appears on the display.
Pipeline or regulated cylinder pressure supplies N2O directly to the
N2O channel of the gas mixer. If the pressure is too low, an alarm
appears on the display.
Mixed gas Either Air or N2O can be selected as the balance gas. The balance
gas flow is controlled by the mixer. When Alternate O2 control is in
use, only O2 is flowing and balance gas is disabled.
The mixed gas goes from the mixer outlet through the vaporizer to the
fresh gas outlet and into the breathing system. A pressure-relief valve
sets the maximum outlet pressure.
EZchange canister When activated, this mode permits continued ventilation and
rebreathing of exhaled gases without any gas passing through the
absorbent.
11-4 M1145956
11 Specifications and Theory of Operation
Condenser The condenser removes water in the system that is produced from
the reaction of CO2 gas with the absorbent. When fresh gas flow
settings of less than the patient minute volume are used during
ventilation, the amount of re-breathed gas increases. More CO2 flows
through the breathing system increasing the absorbent consumption.
The moisture buildup in the breathing system is also increased.
The condenser is connected between the outlet of the absorber
canister and the inlet of the circuit module. Moisture in the gas is
condensed into water droplets, which run into the condenser’s
reservoir.
Pneumatic specifications
Gas supplies
Pipeline gases O2, Air, N2O
Cylinder gases O2, Air, N2O (2 cylinder maximum)
Cylinder connections Pin indexed
Nut and gland DIN-477
Large cylinder kit available for O2 and N2O
Primary regulator output Pin indexed: The primary regulator is set to
pressure pressure less than 345 kPa (50 psi).
DIN-477: The primary regulator is set to
pressure less than 414 kPa (60 psi).
Pressure-relief valve Approximately 758 kPa (110 psi)
Pipeline connections DISS - Male; DISS-Female; DIN 13252; AS
(filtered) 4059 (Australian); S90-116 (French Air
Liquide); BSPP 3/8 (Scandinavian) or NIST
(ISO 5359). All fittings available for O2, Air,
and N2O.
Pressure displays On system display.
Pipeline inlet pressure 280-600 kPa (41-87 psi)
Flush flow 35 l/min to 50 l/min
O2 supply pressure at which O2 pipeline less than 252 kPa (36 psi).
N2O shutoff O2 cylinder less than 2633 kPa (381 psi).
ACGO Port relief A relief valve on the vaporizer manifold limits the fresh gas pressure
at the ACGO port to 55 kPa (8 psi) at 45 l/min.
Non-circle circuit A relief valve limits the fresh gas pressure at the inspiratory port to
relief 27 kPa (4 psi) at 55 l/min when non-circle circuit is selected.
M1145956 11-5
Avance
AB.91.114
11-6 M1145956
11 Specifications and Theory of Operation
M1145956 11-7
Avance
Electrical power
Power cord
Length 5 meters
Voltage rating 100 to 240 Vac
Current capacity 10 A for 220-240 Vac
15 A for 100-120 Vac
Type Three conductor power supply cord (medical grade
where required).
WARNING Unplug the system power cord to run the system on the
battery power if the integrity of the protective earth
conductor is in doubt.
11-8 M1145956
11 Specifications and Theory of Operation
Battery information The system is not a portable unit; a sealed lead acid battery supplies
backup power in the event of a power failure.
• Capacity to operate for 90 minutes under typical operating
conditions; 30 minutes under extreme conditions.
• The system functions to specifications through the transition to
battery power.
Only trained service representatives are to replace the battery.
Batteries must be disposed of in accordance with applicable
regulatory requirements in effect at the time and place of disposal.
Contact a trained service representative to disconnect the battery if
the equipment is not likely to be used for an extended time.
Flow specifications
Alternate O2
Flow range 500 ml/min to 10 l/min
Indicator Flow tube
Indicator accuracy +/- 5% full scale
Fresh gas
Flow range 0 and 150 ml/min to 15 l/min
Minimum total flow O2 and balance gas
is 150 ml/min
Minimum single gas flow 100 ml/min
(maintaining the total flow of 150 ml/
min)
Total flow accuracy +/- 10% or +/- 40 ml/min of setting
(larger of)
O2 flow accuracy +/- 5% or +/- 20 ml/min of setting (larger
of)
Balance gas flow accuracy +/- 5% or +/- 20 ml/min of setting (larger
of) Air/N2O
O2 concentration range 21%, 25% to 100%
O2 concentration accuracy +/- 5% of setting
Electronic mixer response time 500 ms (10% to 90% flow step)
Compensation Temperature and atmospheric pressure
compensated to standard conditions of
20°C and 101.3 kPa (14.7 psi)
Hypoxic guard Electronic
Note: Leakage from one gas inlet to another gas inlet is less than 10 ml per
hour.
M1145956 11-9
Avance
11-10 M1145956
11 Specifications and Theory of Operation
Gas scavenging
All scavenging
Positive pressure relief 10 cmH2O
Passive scavenging
Negative pressure relief 0.3 cmH2O
Outlet connector 30 mm male taper ISO
Active scavenging
Disposal system Outlet
Flow range Pressure
type connector*
Adjustable DISS EVAC Up to 30 l/min not applicable
High flow, low BS6834 50 to 80 l/min 1.6 kPa
vacuum (12 mmHg)
Low flow, high DISS EVAC 36 +/- 4 l/min 305 mmHg
vacuum (12 inHg)
minimum
vacuum
Low flow, low 12.7 mm hose 36 +/- 4 l/min not applicable
vacuum barb
Low flow, low 25 mm barb 36 +/- 4 l/min not applicable
vacuum
Low flow, low 30 mm ISO 36 +/- 4 l/min not applicable
vacuum taper
*Other market-specific connectors may be available.
Particle filter at the outlet has a pore size of 225 microns. All flow data uses
a new filter.
M1145956 11-11
Avance
Physical specifications
All specifications are approximate values and can change without
notice.
Environmental requirements
11-12 M1145956
11 Specifications and Theory of Operation
Gas specifications
Airway humidity 0 to 100% condensing
Sampling delay 2.5 seconds typical with a 3 m sampling line
Total system response time 2.5 seconds typical with a 3 m sampling line,
including a sampling delay and rise time
Warm-up time 2 minutes for operation with CO2, O2, and
N2O
5 minutes for operation of anesthetic agents
30 minutes for full specifications
Diversion flow 200 ml/min
M1145956 11-13
Avance
Typical performance
CO2 Measurement range 0 to 15 vol% (0 to 15 kPa, 0 to 113
mmHg).
Measurement rise time less than 400 ms typical.
Accuracy +/- (0.2 vol% + 2% of reading).
Gas cross effects less than 2 vol% (O2, N2O,
anesthetic agents).
O2 Measurement range 0 to 100 vol%.
Measurement rise time less than 400 ms typical.
Accuracy +/- (1 vol% + 2% of reading).
Gas cross effects less than 1 vol% anesthetic agents,
less than 2 vol% N2O.
N2O Measurement range 0 to 100 vol%.
Measurement rise time less than 450 ms typical.
Accuracy +/- (2 vol% + 2% of reading).
Gas cross effects less than 2 vol% anesthetic agents.
Anesthetic agents Measurement range Hal, Enf, Iso 0 to 6 vol%.
Measurement range Sev 0 to 8 vol%.
Measurement range Des 0 to 20 vol%.
Measurement rise time less than 400 ms typical.
Accuracy +/- (0.15 vol% + 5% of reading).
Gas cross effects less than 0.15 vol% N2O.
11-14 M1145956
11 Specifications and Theory of Operation
M1145956 11-15
Avance
Ventilator theory
The ventilator pneumatics are at the rear of the breathing system. A
precision valve controls gas flow to the patient. During inspiration, this
gas flow closes the exhalation valve and pushes the bellows down.
During expiration, a small flow pressurizes the exhalation diaphragm
to supply PEEP pressure.
Volume and flow measurements come from flow sensors in the flow
sensor module. Two tubes from each sensor connect to a transducer
that measures the pressure change across the sensor, which
changes with the flow. A third transducer measures airway pressures
at the inspiratory flow sensor.
The ventilator uses the data from the flow sensors for volume-related
numerics and alarms. The ventilator also uses the flow sensors to
adjust its output for changes in fresh gas flow, small leaks, and gas
compression upstream of the breathing circuit. There is adjustment
for compression in the patient circuit.
In volume ventilation modes, certain alarm conditions prevent the
automatic adjustment of ventilator delivery based on measured flow
values. In these cases, ‘Vent mode: TV accuracy decreased. Adj.
manually.’ shows above the ventilator setting area on the screen.
When this message shows, the ventilator may not be able to deliver
within the accuracy range specified. When this occurs, manually
adjust the tidal volume until the volume delivered reaches the desired
level. If compensation stops for a number of breaths, the condition
causing the hold shows as an alarm. Automatic volume
compensation resumes when alarm conditions are resolved.
For better precision a small quantity of gas bleeds through a resistor
to help keep the pressure on the exhalation valve constant. At high
airway pressures, this can cause a slight hiss during inspiration.
11-16 M1145956
11 Specifications and Theory of Operation
M1145956 11-17
Avance
Volume control mode Volume control supplies a set tidal volume. The ventilator calculates a
flow based on the set tidal volume and the length of the inspiratory
time (TI) to deliver that tidal volume. It then adjusts that output by
measuring delivered volumes at the flow sensors. Since the ventilator
adjusts output, it can compensate for breathing system compliance,
fresh gas flow, and moderate breathing system leaks.
A typical volume-controlled pressure waveform increases throughout
the entire inspiratory period, and rapidly decreases at the start of
expiration. An optional inspiratory pause is available to improve gas
distribution.
Volume control mode settings:
• TV
• RR
• I:E
• Tpause
• PEEP
• Pmax
AB.98.036
1. Paw waveform
2. Tinsp
3. Insp Pause
4. Texp
5. PEEP
6. Flow waveform
7. TV
Figure 11-1 • Volume control diagram
11-18 M1145956
11 Specifications and Theory of Operation
Pressure control mode Pressure control supplies a constant set pressure during inspiration.
The ventilator calculates the inspiratory time from the frequency and
I:E ratio settings. A high initial flow pressurizes the circuit to the set
inspiratory pressure. The flow then decreases to maintain the set
pressure (Pinspired).
Pressure sensors in the ventilator measure patient airway pressure.
The ventilator automatically adjusts the flow to maintain the set
inspiratory pressure.
Pressure control mode settings:
• Pinsp
• RR
• I:E
• PEEP
• Pmax
• Rise Rate
AB.98.037
1. Paw waveform
2. Tinsp
3. Texp
4. Pinsp
5. PEEP
6. Flow waveform
Figure 11-2 • Pressure control diagram
M1145956 11-19
Avance
4
5
AB.91.040
6
1. Mandatory SIMV breath
2. Spontaneous pressure supported breath
3. Paw
4. Psupport
5. PEEP
6. Time
Figure 11-3 • SIMV/PSV diagram
11-20 M1145956
11 Specifications and Theory of Operation
M1145956 11-21
Avance
AB.91.041
3
1. Paw
2. PEEP
3. Time
Figure 11-4 • PSVPro diagram
When the ventilator switches to the backup mode, the alarm text
‘Backup Mode active’ shows as an information message until PSVPro
is reinstated or until another ventilation mode is selected. PSVPro
mode automatically resumes when the ventilator registers the
number of consecutive patient-triggered breaths set for the Exit
Backup setting. When Exit Backup is set to off, the user must reselect
the PSVPro mode to reactivate PSVPro. Upon returning to PSVPro
the ventilator immediately begins providing pressure supported
breaths to the patient using the established settings.
Spontaneous breaths that occur during this mode are indicated by a
color change in the waveform.
11-22 M1145956
11 Specifications and Theory of Operation
AB.91.038
1. Paw waveform
2. Tinsp
3. Spontaneous breathing period
4. Trigger window
5. Pressure supported breath
6. Pinsp
7. Flow waveform
Figure 11-5 • SIMV-PC waveforms
M1145956 11-23
Avance
PCV-VG mode In PCV-VG, a tidal volume is set and the ventilator delivers that
volume using a decelerating flow and a constant pressure. The
ventilator will adjust the inspiratory pressure needed to deliver the set
tidal volume breath-by-breath so that the lowest pressure is used.
The pressure range that the ventilator will use is between the PEEP +
2 cmH2O level on the low end and 5 cmH2O below Pmax on the high
end. The inspiratory pressure change between breaths is a maximum
of ± 3 cmH2O.
This mode will deliver breaths with the efficiency of pressure
controlled ventilation, yet still compensate for changes in the patient’s
lung characteristics. PCV-VG begins by first delivering a volume
breath at the set tidal volume. The patient’s compliance is determined
from this volume breath and the inspiratory pressure level is then
established for the next PCV-VG breath.
PCV-VG mode settings:
• TV
• RR
• I:E
• PEEP
• Pmax
• Rise Rate
AB.98.034
1. Paw waveform
2. Tinsp
3. Texp
4. Variable pressure to deliver desired TV
5. PEEP
6. Flow waveform
7. TV
Figure 11-6 • PCV-VG waveforms
11-24 M1145956
11 Specifications and Theory of Operation
Pneumatics
Gas source Anesthesia system
Gas composition Medical Air or O2
Nominal supply pressure 350 kPa (50 psi)
Pressure range at inlet 240 to 700 kPa (35 to 102 psi)
Peak gas flow 120 l/min at 240 kPa (35 psi), 0.75
seconds
Continuous gas flow 80 l/min at 240 kPa (35 psi)
Flow valve range 1 to 120 l/min at 240 kPa (35 psi)
Fresh gas
compensation Flow compensation range 150 ml/min to 15 l/min
Gas composition O2, N2O, Air, anesthetic agents
Pressure
Patient airway pressure range -20 to +120 cmH2O, +/- 1 cmH2O
resolution
High pressure alarm set range 12 to 100 cmH2O, 1 cm increment
Sustained pressure alarm range 6 to 30 cmH2O, 1 cm increment
Display range -20 to 120 cmH2O
Volume
Tidal volume display range 0 to 9999 ml, 1 ml resolution
Setting range 20 to 1500 ml
Minute volume 0.0 to 99.9 liters
Breath rate 4 to 100 bpm (non-spontaneous)
2 to 60 bpm (spontaneous)
1 bpm resolution
Volume sensor type Variable flow orifice
Oxygen
Display range 0 to 110% O2
Display resolution 1% increments
Sensor type Galvanic fuel cell
Measurement range 0 to 100% O2
Measurement accuracy Better than ± 3% of full scale
Cell response time 35 seconds
Note: Response time of cell and
adapters is measured using the test
method described in ISO 7767 (1997).
Low O2 alarm range 18% to 99%
High O2 alarm setting 19% to 100% or Off
Note: Low O2 limit may not be set
above high O2 limit. High O2 limit may
not be set below the low O2 limit.
Expected cell life Four months of shelf life (23°C room
air) and one year of normal operation.
M1145956 11-25
Avance
Delivery accuracy
Volume delivery accuracy greater than 210 ml tidal volume -
accuracy better than 7%
less than 210 ml but greater than 60 ml
tidal volume - accuracy better than 15 ml
less than 60 ml tidal volume - accuracy
better than 10 ml
Volume monitoring accuracy greater than 210 ml tidal volume -
accuracy better than 9%
less than 210 ml but greater than 60 ml
tidal volume - accuracy better than 18 ml
less than 60 ml tidal volume - accuracy
better than 10 ml
Inspiratory pressure delivery accuracy greater of +/- 10% or +/- 3 cmH2O
PEEP delivery accuracy +/- 1.5 cmH2O
Pressure monitoring accuracy greater of +/- 5% or +/- 2 cmH2O
Note: Gas composition errors may be in addition to the above normalized
accuracy. When adding errors, positive errors can have the effect of nulling out
negative errors.
Note: Use of anesthetic agent could affect the errors by approximately -0.95%/%
volume agent in normal mode.
AB.74.027
11-26 M1145956
11 Specifications and Theory of Operation
Guidance and The system is suitable for use in the specified electromagnetic
manufacturer’s environment. The customer and/or the user of the system should
assure that is used in an electromagnetic environment as described
declaration - below.
electromagnetic
emissions
Emissions test Compliance Electromagnetic environment guidance
RF emissions Group 1 The system uses RF energy only for its internal function. Therefore, its RF
CISPR 11 emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions Class B The system is suitable for use in all establishments, including domestic
CISPR 11 establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
Harmonic Class A
emissions
IEC 61000-3-2
Voltage Complies
fluctuations/
flicker emissions
IEC 61000-3-3
M1145956 11-27
Avance
Guidance and The system is suitable for use in the specified electromagnetic
manufacturer’s environment. The customer and/or the user of the system should
assure that it is used in an electromagnetic environment as described
declaration - below.
electromagnetic
immunity
Power immunity
Immunity test IEC 60601-1-2 test Level Compliance level Electromagnetic environment guidance
Electrostatic discharge ± 6 kV contact ± 6 kV contact Floors should be wood, concrete, or
(ESD) IEC 61000-4-2 ± 8 kV air ± 8 kV air ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast ± 2 kV for power supply ± 2 kV for power supply Mains power quality should be that of a
transient/burst IEC lines lines typical commercial and/or hospital
61000-4-4 ± 1 kV for input/output ± 1 kV for input/output environment.
lines lines
Surge IEC 61000-4-5 ± 1 kV differential mode ± 1 kV differential modeMains power quality should be that of a
± 2 kV common mode ± 2 kV common mode typical commercial and/or hospital
environment.
Voltage dips, short <5% UT (>95% dip in <5% UT (>95% dip in UT) Mains power quality should be that of a
interruptions and UT) for 0.5 cycle for 0.5 cycle) typical commercial and/or hospital
voltage variations on 40% UT (60% dip in UT) 40% UT (60% dip in UT) environment. If the user of the system
power supply input for 5 cycles for 5 cycles requires continued operation during
lines IEC 61000-4-11 70% UT (30% dip in UT) 70% UT (30% dip in UT) power mains interruptions, it is
for 25 cycles for 25 cycles recommended that the system be
<5% UT (>95% dip in <5% UT (>95% dip in UT) powered from an uninterruptible power
UT) for 5 sec. for 5 sec. supply or a battery.
Power frequency (50/ 3 A/m 3 A/m If display distortion or other abnormalities
60 Hz) magnetic field occur, it may be necessary to position the
IEC 61000-4-8 3 Anesthetic System further from sources
of power frequency magnetic fields or to
install magnetic shielding. The power
frequency magnetic field should be
immersed in the intended installation
location to assure that it is sufficiently low.
Note: UT is the AC mains voltage before application of the test level.
11-28 M1145956
11 Specifications and Theory of Operation
Radiated immunity
Electromagnetic environment guidance
Immunity test IEC 60601-1-2 test Level Compliance level
Recommended separation distance
Portable and mobile RF communications
equipment should be used no closer to
any part of the system, including cables,
than the recommended separation
distance calculated from the equation
appropriate for the frequency of the
transmitter.
Conducted RF 3 Vrms 3 Vrms (V1) D=3.5√P
IEC 61000-4-6 150 kHz to 80 MHz
outside ISM bands
10 Vrms 10 Vrms (V2) D=12√P
150 kHz to 80 MHz in
ISM bands
Radiated RF 10 V/m 10 V/m (E1) D=1.2√P 80 MHz to 800 MHz
IEC 61000-4-6 80 MHz to 2.5 GHz D=3.5√P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and D is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
frequency range.
The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553
MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range 80 MHz
to 2.5 GHz are intended to decrease the likelihood that a portable communications device could cause interference if
it is inadvertently brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating the
recommended separation distance for transmitters in these frequency ranges.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the system is used exceeds the applicable RF
compliance level above, the system should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the system.
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 1 V/m.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
M1145956 11-29
Avance
11-30 M1145956
11 Specifications and Theory of Operation
Electrical safety
M1145956 11-31
Avance
Standards
Devices used with this anesthesia system shall comply with the
following standards where applicable:
• Breathing system and breathing system components ISO 8835-2.
• Anesthetic vapor delivery devices ISO 8835-4.
• Anesthetic agent monitors ISO 11196.
• Oxygen monitors ISO 7767.
• Carbon dioxide monitors ISO 9918.
• Exhaled volume monitors IEC 60601-2-13.
11-32 M1145956
11 Specifications and Theory of Operation
System components
Not integral These devices are not integral to this anesthesia system:
• CO2 monitor.
• Anesthetic agent monitor.
• O2 monitor.
M1145956 11-33
Avance
11-34 M1145956
12 Super User Mode
M1145956 12-1
Avance
Install/Service menu
Install/Service - Page 1
Cumulative Gas Usage
Show Alarm Limits Yes Yes, No
Fresh Gas Controls O2% O2%, Flow, User
VCV Cardiac Bypass No Yes, No
External Gas Monitor No Yes, No
Volume Apnea Setup
Time and Date
Factory Defaults
Install/Service - Page 2
Exit
Install/Service - Page 2
Trends Setup
Colors and Units
Parameter Settings
Page Setup
Case Defaults
Installation
Service
Install/Service - Page 1
Exit
12-2 M1145956
12 Super User Mode
Using super user Select a menu item showing the submenu symbol to go to the
mode corresponding submenu. Access the Cumulative Gas Usage,
Volume Apnea Setup, Time and Date, Trends Setup, Colors and
Units, Parameter Settings, Page Setup, and Case Defaults menus
through the Install/Service menu.
Set Show Alarm Limits to Yes to display the alarm limits next to the
measured values on the normal screen.
Set the default style for the fresh gas controls. Set to O2% for O2%
and total flow controls. Set to Flow for individual gas flow controls.
Set to User to allow the user to select the flow control style through
the Screen Setup menu.
Set the VCV Cardiac Bypass to Yes to enable alveolar support
during cardiac bypass. This menu item is only available on systems
with the VCV Cardiac Bypass option installed.
Set the External Gas Monitor to Yes only if the system uses a stand-
alone monitor for O2, agent, and CO2. Selecting Yes disables the O2
limit alarms and the ‘O2 monitoring not connected’ alarm when there
is not O2 cell or airway module detected in the system. The ‘AA, CO2
monitoring not connected’ alarm is also disabled.
Factory defaults are the settings installed by the manufacturer. These
cannot be changed. Select Factory Defaults to return the system to
the factory defaults.
Installation is a password-protected menu selection.
Service is a password-protected menu selection.
M1145956 12-3
Avance
Menus
Not all menu items are available depending on the system
configuration. Inactive menu items are grayed out and are not
selectable.
Menu selections shown are the factory default values. The options
are listed to the right of the menu shown.
Cumulative gas View the total system fresh gas usage since the last reset or reset the
usage gas usage to 0 through the Cumulative Usage menu.
Cumulative Usage
Reset Usage
Desflurane
Enflurane
Halothane
Isoflurane
Sevoflurane
O2 (*1000 l)
Air (*1000 l)
N2O (*1000 l)
Previous Menu
12-4 M1145956
12 Super User Mode
Volume apnea setup Use the Volume Apnea Setup menu to enable or disable the user’s
ability to turn the volume apnea alarms off during manual ventilation.
Set the Volume Apnea Selection to Enable to allow the user to turn
the volume apnea alarm on or off from the Start Case menu and the
Alarm Setup menu. Set the Volume Apnea Selection for each case
type default through this menu.
Set the Volume Apnea Selection to Disable to set the volume
apnea alarms to be always on during manual ventilation. When set to
Disable, the Volume Apnea menu selection is not available from the
Start Case menu and the Alarm Setup menu.
Menu selection names shown in all capital letters are configured from
the Case Defaults menu
M1145956 12-5
Avance
Setting time and date Use the Time and Date menu to set the time and date. The Hour
field is dependent on the Clock Format setting. The Clock Format
can be set to a 12 or 24 hour format.
12-6 M1145956
12 Super User Mode
Trends setup Use the Trends Setup menu to set the default trend that shows in the
right window on the user Trends menu. Access the Graphical
Trends menu from the Trends Setup menu.
Trends Setup
Default Trend Num Num, Graph, Set
Graphical Trends
Previous Menu
Setting graphical trends Use the Graphical Trends menu to configure the graphical trends
pages pages that show in the right window on the user Trends menu.
Graphical Trends
Page 1
Page 2
Page 3
Page 4
Page 5
Previous Menu
M1145956 12-7
Avance
Page 1
Field 1 Pres
Field 2 TVexp
Field 3 CO2
Previous Menu
The available selections for Field 1, Field 2, and Field 3 are the same
for each field on each page. Selections are Off, rr+co2, Pres, MVexp,
CO2, O2, Bal, AA1, AA2, N2O, MAC, TVexp, Pmean, Spont, Compl,
VO2. The default settings for Field 1, Field 2, and Field 3 of each
page show in the table.
12-8 M1145956
12 Super User Mode
Setting colors and Set the units of the parameters through the Colors and Units menu.
units Access the Colors menu through Colors and Units menu.
Setting colors Set the colors of the parameter waveform information, digit fields, and
trends through the Colors menu.
Colors
Paw Yellow Yellow, White, Green, Red, Blue
Flow Green Yellow, White, Green, Red, Blue
Resp White Yellow, White, Green, Red, Blue
CO2 White Yellow, White, Green, Red, Blue
Previous Menu
M1145956 12-9
Avance
Parameter settings Set the volume conditions and the CO2 humidity compensation
through the Parameter Settings menu. Set the volume calculation
conditions to be based on ATPD (ambient temperature and pressure,
dry humidity conditions) or BTPS (body temperature, ambient
pressure, saturated humidity condition). Set the humidity
compensation type in CO2 partial pressure values to wet or dry.
Parameter Settings
TV Based on ATPD ATPD, BTPS
CO2 Numbers Dry Dry, Wet
Previous Menu
Page Setup Set the preset normal screen layouts through the Page Setup menu.
Menu selection names shown in all capital letters are configured from
the corresponding submenu.
12-10 M1145956
12 Super User Mode
The default settings for each page type show in the following table.
The fields for each page type have the same selections available:
• Waveform Fields are AA, CO2, Flow, Paw, Off.
• Digit Field is Supply, Flow, AA.
• Split Screen is None, Gas, Trend, Spiro, Metab, Paw.
M1145956 12-11
Avance
Configuring case Set the user selectable default case types that show in the user’s
defaults Start Case menu through the Case Defaults menu. The Set to Last
Case selection is only available on submenus if a case has occurred.
12-12 M1145956
12 Super User Mode
Each case type has multiple settings. The default settings for the
default case types show in the following table. Values in bold are
different from the default case type ADULT. A * indicates that the
setting is not available for the default ventilation mode.
M1145956 12-13
Avance
12-14 M1145956
Index
M1145956 I-1
Avance
Circuit F
circle 3-10
non-circle 2-8, 3-10 Flow and pressure calibration 9-5
non-circle relief 11-5 Flow sensors
Circuit compliance compensation 5-4 parts 10-2
CO2 alarms 3-24 prevent water buildup 9-7
Condenser Flow specifications 11-9
operation 3-28 Fresh gas usage 3-19
parts 10-10
theory 11-5 G
Configuring case defaults 12-12 Gas scavenging
Connections specifications 11-11
electrical 8-9 Gas setup 3-9
pneumatic 8-11 Gas supplies 11-4
Controls specifications 11-5
alternate O2 3-26
on the display 2-13 H
vaporizer 2-11
Cylinder High-pressure leak test 8-15
installation 8-14 How to attach equipment to the top of the
machine 8-16
D How to install gas cylinders 8-14
I-2 M1145956
M Preoperative Tests 5-1
Problems
MAC 3-5 breathing system 7-16
Machine check 5-6 electrical 7-17
circuit 5-7 pneumatic 7-18
circuit O2 5-7
monitor 5-7
Q
system 5-6
Main menu 3-16 Quick keys
Mains inlet 8-9 changing gas settings using 3-9
Maintenance schedule 9-2 changing ventilator settings using 3-7
Minimum Alveolar Concentration 3-5
MV/TV alarms 3-24 R
Repair policy 9-2
N
Negative low pressure leak test 5-9 S
Non-circle circuit 2-8
Sample gas return port 8-12
Scavenging 8-11
O
a gas monitor sample flow 2-10
O2 cell ACGO sample flow 2-9
calibration 9-6 from an auxiliary manual breathing circuit
replacement 9-4 2-10
theory 11-17 specifications 11-11
O2 flow Screen configuration 3-20
adjust 3-9 Screen setup 3-21
start 3-3 Select page 3-20
Outlets 8-9 Serial port 8-10
Overview Service schedule 9-3
system 2-2 Setting alarm limits 3-23
Setting gas controls 3-22
P Setup
changing circuit type 3-10
Parameters setup 6-5 changing gas and settings 3-9
agent setup 6-5 gas 3-9
CO2 setup 6-5 vent 3-7
data source 6-5 Setup warnings 8-2
O2 setup 6-5 Silencing alarms 7-2
spirometry setup 6-5 Specifications
Passive AGSS 3-29 breathing system 11-10
Pin indexed cylinder yokes 8-14 flow 11-9
Pipeline inlets 8-11 gas scavenging 11-11
Pneumatic physical 11-12
connections 8-11 pneumatic 11-5
problems 7-18 ventilator operating 11-25
specifications 11-5 Spirometry
system pneumatic circuits 11-2 menu functions 3-11
Positive low pressure leak test (ACGO scaling 3-13
systems only) 5-10 setting loop type 3-12
Preoperative Checkout 4-1
M1145956 I-3
Avance
T
Test tools and system parts 10-11
Trends 3-17
Troubleshooting
breathing system 7-16
electrical 7-17
pneumatic 7-18
Turning system on 3-2
U
Using menus 2-16
I-4 M1145956
Warranty
This Product is sold by Datex-Ohmeda under the warranties set forth
in the following paragraphs. Such warranties are extended only with
respect to the purchase of this Product directly from Datex-Ohmeda
or Datex-Ohmeda’s Authorized Dealers as new merchandise and are
extended to the Buyer thereof, other than for the purpose of resale.
For a period of twelve (12) months from the date of original delivery to
Buyer or to Buyer’s order, but in no event for a period of more than
two years from the date of original delivery by Datex-Ohmeda to a
Datex-Ohmeda Authorized Dealer, this Product, other than its
expendable parts, is warranted against functional defects in materials
and workmanship and to conform to the description of the Product
contained in this User’s Reference manual and accompanying labels
and/or inserts, provided that the same is properly operated under the
conditions of normal use, that regular periodic maintenance and
service is performed and that replacements and repairs are made in
accordance with the instructions provided. This same warranty is
made for a period of thirty (30) days with respect to expendable parts.
The foregoing warranties shall not apply if the Product has been
repaired other than by Datex-Ohmeda or in accordance with written
instructions provided by Datex-Ohmeda, or altered by anyone other
than Datex-Ohmeda, or if the Product has been subject to abuse,
misuse, negligence, or accident.
Datex-Ohmeda’s sole and exclusive obligation and Buyer’s sole and
exclusive remedy under the above warranties is limited to repairing or
replacing, free of charge, at Datex-Ohmeda’s option, a Product,
which is telephonically reported to the nearest Datex-Ohmeda
Customer Service Center and which, if so advised by Datex-Ohmeda,
is thereafter returned with a statement of the observed deficiency, not
later than seven (7) days after the expiration date of the applicable
warranty, to the Datex-Ohmeda Customer Service and Distribution
Center during normal business hours, transportation charges prepaid,
and which, upon Datex-Ohmeda’s examination, is found not to
conform with above warranties. Datex-Ohmeda shall not be
otherwise liable for any damages including but not limited to
incidental damages, consequential damages, or special damages.
There are no express or implied warranties which extend beyond the
warranties hereinabove set forth. Datex-Ohmeda makes no warranty
of merchantability or fitness for a particular purpose with respect to
the product or parts thereof.
Avance
User’s Reference Manual
English
M1145956
05 08 001 18 05 02
Printed in USA