Aisys User's Reference Manual
Aisys User's Reference Manual
Aisys User's Reference Manual
1 Introduction
2078169-001 i
Aisys
3 Operation
ii 2078169-001
Table of Contents
2078169-001 iii
Aisys
4 Preoperative checkout
5 Preoperative tests
iv 2078169-001
Table of Contents
6 Airway modules
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Alarm priorities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
Pausing alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Cancelling audio pause . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Display changes during alarms . . . . . . . . . . . . . . . . . . . . 7-3
De-escalating alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Battery indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
Internal failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
Informational tones . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
List of alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
Sustained pressure threshold . . . . . . . . . . . . . . . . . . . . 7-12
Alarm ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-15
Breathing system problems . . . . . . . . . . . . . . . . . . . . . . . . .7-17
Electrical problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
Pneumatic problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-19
2078169-001 v
Aisys
9 User maintenance
vi 2078169-001
Table of Contents
10 Parts
2078169-001 vii
Aisys
viii 2078169-001
Table of Contents
2078169-001 ix
Aisys
13 Vaporizer cassettes
Vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-2
Aladin2 cassette variants . . . . . . . . . . . . . . . . . . . . . . . 13-2
Aladin cassette variants . . . . . . . . . . . . . . . . . . . . . . . . 13-4
Changing a cassette during a case . . . . . . . . . . . . . . . . . . .13-7
Removing a cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-8
Installing a cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-9
Cassette maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-10
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13-10
Draining cassettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-10
Draining halothane cassettes . . . . . . . . . . . . . . . . . . . 13-11
Filling Aladin2 cassettes . . . . . . . . . . . . . . . . . . . . . . . . . . 13-12
Filling with Easy-Fil system . . . . . . . . . . . . . . . . . . . . . 13-12
Filling with Quik-Fil system . . . . . . . . . . . . . . . . . . . . . 13-14
Filling with Saf-T-Fill bottle . . . . . . . . . . . . . . . . . . . . . 13-15
Filling Aladin cassettes . . . . . . . . . . . . . . . . . . . . . . . . . . . 13-17
Filling with keyed filler system . . . . . . . . . . . . . . . . . . .13-17
Filling with Quik-Fil system . . . . . . . . . . . . . . . . . . . . . 13-18
Filling with Saf-T-Fil bottle . . . . . . . . . . . . . . . . . . . . . . 13-19
Index
x 2078169-001
1 Introduction
Introduction
2078169-001 1-1
Aisys
Intended use
The Aisys anesthesia system is scalable, flexible, and functionally
integrated, featuring the most advanced design, ventilation,
respiratory monitoring, and breathing system. Module bays allow for
the physical integration of legacy Datex-Ohmeda patient monitors
and supports mounting of other GE Healthcare monitors. Optionally,
the open architecture design supports mounting of non-Datex-
Ohmeda patient monitors, record keeping, and connections to the
hospital information system. The INview movable display arm helps
keep the anesthetist’s focus on the patient by offering control of all
hemodynamic, gas delivery, anesthetic agent, and ventilation
parameters.
This anesthesia system is designed for mixing and delivering
inhalation anesthetics, Air, O2, and N2O.
The anesthesia system’s small breathing system volume allows an
anesthetic agent delivery response time of less than 7 seconds.
(Agent response time defined as achieving 90% of the setting
change measured while in non-circle circuit with a 2 l/min fresh gas
flow.)
This anesthesia system uses SmartVent ventilation technology
offering Volume Control Ventilation with tidal volume compensation
and electronic PEEP. The proven SmartVent technology also
features optional Pressure Control Ventilation, Pressure Support
Ventilation with an Apnea Backup (PSVPro) that is used for
spontaneously breathing patients, Synchronized Intermittent
Mandatory Ventilation (SIMV) modes, Pressure Control Ventilation-
Volume Guarantee (PCV-VG), Continuous Positive Airway Pressure
+ Pressure Support Ventilation (CPAP + PSV), and VCV Cardiac
Bypass. In Volume Control Ventilation, a patient can be ventilated
using a minimal tidal volume of 20 ml. In Pressure Control
Ventilation, volumes as low as 5 ml can be measured. These
advanced features allow for the ventilation of a broad patient range.
General information
1-2 2078169-001
1 Introduction
2078169-001 1-3
Aisys
AB.75.277
Serial numbers
1-4 2078169-001
1 Introduction
AAAX11111 AAAXX111111AA
The X represents an alpha character The XX represents a number
indicating the year the product was indicating the year the product was
manufactured; H = 2004, J = 2005, manufactured; 04 = 2004, 05 = 2005,
etc. I and O are not used. etc.
Trademarks
2078169-001 1-5
Aisys
Air Air
Maximum Vacuum
1-6 2078169-001
1 Introduction
Lock Unlock
Systems with this mark agree with the Authorized representative in the
European Council Directive (93/42/ European Community
EEC) for Medical Devices when they
are used as specified in their User’s
Reference manuals. The xxxx is the
certification number of the Notified
Body used by Datex-Ohmeda’s Quality
Systems.
Date of manufacture Manufacturer
2078169-001 1-7
Aisys
1-8 2078169-001
1 Introduction
Pediatric Adult
Pipeline Cylinder
Test indicator: red for failure, yellow for Alarm low and alarm high limit indicator
conditional outcome, and green for
pass.
Manual ventilation Enhanced temperature sensing
2078169-001 1-9
Aisys
1-10 2078169-001
1 Introduction
Abbreviations
Abbreviation Definition
A
AA Anesthetic agent
ABS Advanced breathing system
ACGO Auxiliary common gas outlet
AGSS Anesthesia gas scavenging system
Alt O2 Alternate O2
APL Adjustable pressure-limiting
APN Apnea
ATPD Ambient temperature and pressure, dry humidity
conditions
B
BTPS Body temperature, ambient pressure, saturated humidity
conditions
C
CGO Common gas outlet
CO Carbon monoxide
CO2 Carbon dioxide
Compl Compliance
CPAP + PSV Continuous positive airway pressure + pressure support
ventilation
E
EMC Electromagnetic Compatibility
ET End-tidal concentration
EtCO2 End-tidal carbon dioxide
EtO2 End-tidal oxygen
Exp Expiratory
F
FI Fraction of inspired gas
FiCO2 Fraction of inspired carbon dioxide
FiO2 Fraction of inspired oxygen
Flow-Vol Flow-volume loop
I
I:E Inspiratory-expiratory ratio
Insp Inspiratory
Insp Pause Inspiratory pause time
K
kg Kilogram
M
2078169-001 1-11
Aisys
Abbreviation Definition
MAC Minimum alveolar concentration
MV Minute volume
MVexp Expired minute volume
MVinsp Inspired minute volume
N
N2 Nitrogen
N2O Nitrous oxide
O
O2 Oxygen
P
Paw Airway pressure
PCV Pressure control ventilation
PCV-VG Pressure control ventilation - volume guaranteed
PEEP Positive end expiratory pressure
Paw-Flow Pressure-flow loop
Pinsp Inspiratory pressure
Pmax Maximum pressure
Pmean Mean pressure
Ppeak Peak pressure
Pplat Plateau pressure
Psupport Support pressure
PSV Pressure support ventilation
PSVPro Pressure support ventilation with apnea backup
Paw-Vol Pressure-volume loop
R
Raw Airway resistance
RF Radio frequency
RR Respiratory rate
S
SIMV PCV Synchronized intermittent mandatory ventilation -
pressure control ventilation
SIMV PCV-VG Synchronized intermittent mandatory ventilation -
pressure control ventilation - volume guaranteed
SIMV VCV Synchronized intermittent mandatory ventilation - volume
control ventilation
T
Texp Expiratory time
Tinsp Inspiratory time
Tpause Time where breath is paused with no flow
TV Tidal volume
TVexp Expired tidal volume
TVinsp Inspired tidal volume
1-12 2078169-001
1 Introduction
Abbreviation Definition
V
VCV Volume control ventilation
Vol Volume
2078169-001 1-13
Aisys
System information
System classification
This system is classified as follows:
• Class I Equipment.
• Type B Equipment.
• Type BF Equipment (airway modules).
• Ordinary Equipment.
• Not for use with flammable anesthetics.
• Continuous operation.
Devices used with this anesthesia system shall comply with the
following standards where applicable:
Devices used with this anesthesia system shall comply with the
following standards where applicable:
1-14 2078169-001
1 Introduction
• CO2 monitor.
• Anesthetic agent monitor.
• O2 monitor (when O2 cell is not installed).
• Suction regulator.
• EZchange canister system.
• Condenser.
System accessories
These devices can be used as accessories on this anesthesia
system:
• Suction regulator.
• EZchange canister system.
• Condenser.
• Aladin and Aladin2 vaporizer cassettes.
2078169-001 1-15
Aisys
System safety
1-16 2078169-001
1 Introduction
2078169-001 1-17
Aisys
Electrical safety
1-18 2078169-001
1 Introduction
2078169-001 1-19
1-20 2078169-001
2 System controls and menus
2078169-001 2-1
Aisys
System overview
14
13
12
11
1
2
10 3
4
5
6
7
15
8
AB.75.278
2-2 2078169-001
2 System controls and menus
AB.75.279
11 7
9 8
10
1. Serial port 7. Mains inlet
2. Collection bottle connection 8. System circuit breaker
3. Cylinder wrench (key) storage 9. Equipotential stud
4. Cylinder yoke 10. Outlet circuit breaker
5. Anesthesia Gas Scavenging System 11. Isolated electrical outlet
6. Pipeline connections
2078169-001 2-3
Aisys
AB.72.014
1. Push down on the brake pedal to lock the casters.
2. Lift up on the brake pedal to release the brake. The casters are
unlocked.
2-4 2078169-001
2 System controls and menus
13
12
11
10
1
2
9
8
7
6 3
4
AB.75.280
2078169-001 2-5
Aisys
2
3
AB.75.281
7
Use the optional bag support arm to hold the breathing circuit bag.
AB.82.023
1. To raise the bag support arm, squeeze the button and rotate the
arm up the top position.
2. To lower the bag support arm, squeeze the button and rotate the
arm down to the lower position.
2-6 2078169-001
2 System controls and menus
AB.60.038
4 3
2078169-001 2-7
Aisys
Display controls
The system uses touchscreen technology, hard keys, and a
ComWheel to access system functions, menus, and settings.
The touchscreen has numerous touch point areas that make
accessing menus and settings quick and easy. The buttons on the
right side of the screen provide direct access to commonly used
functions. The ventilation quick keys enable setup of ventilation
modes. The gas control quick keys provide a method to set up the
gas used for a case.
Touch only one touch point at a time to ensure the correct selection
is made.
AC.22.001
6 5 4 3 2 1
1. ComWheel Selects a menu item or confirms a setting. Turn clockwise or counterclockwise to scroll
through menu items or change settings.
2. Home key Removes all menus from the screen.
3. Screen Lock/Unlock Locks the touchscreen. Toggles between lock and unlock functions.
key
4. Start/End Case key Initiates Start or End Case function.
5. Touchscreen Activates functions when touch areas on the screen are selected.
2-8 2078169-001
2 System controls and menus
6. Audio Pause key Stops audio for 120 seconds for any active, eligible high and medium priority alarms.
Prevents audio (audio off) for 90 seconds when no medium or high priority alarms are
active. Allows the operator to acknowledge any non-active medium or high priority latched
alarms.
Touch points
1 2
3
7
4
AB.75.256
6 5
1. Wave fields 5. Ventilator quick keys
2. Measured values 6. Gas quick keys
3. Function keys 7. Split screen values
4. Digit fields
Figure 2-7 • Normal/Full screen view with shaded touch point areas
2078169-001 2-9
Aisys
2-10 2078169-001
2 System controls and menus
1 2 3 5 6
12 7
8
9
AB.75.258
11 10
1. Audio pause symbol and Indicates when alarm audio is paused and the countdown clock until audio is
countdown clock on.
2. Alarm message fields Displays the active alarms.
3. Waveform fields Displays the waveforms of measured values. For example: Paw, Flow, and
CO2.
4. General message fields or lock Displays general messages and the touchscreen lock indicator.
touchscreen indicator
5. Measured values fields Displays the measured values. For example: Paw, Flow, and CO2.
6. Clock Displays the current time.
7. Function keys Functions available are: Audio Pause, Alarm Setup, Alarms On/Off, Auto
Limits, System Setup, Next Page, Trends, Spirometry, Procedures, Timer,
Start, and End Case.
8. Digit fields Contains information for Spirometry, Resp, Agent, and Gases.
9. Ventilation mode Displays the selected ventilation mode. For example: Ventilator On, and
Volume Control.
10. Ventilator quick keys Displays Mode, associated ventilation parameters, and More Settings. For
example: Mode, TV, RR, I:E, PEEP, and More Settings.
2078169-001 2-11
Aisys
11. Gas quick keys Displays O2, Total Flow, and Gas Setup.
12. Split screen Contains airway pressure, gas flow values, compliance, trends, and optional
ecoFLOW information.
Digit fields
The digit field can be set to show specific information such as gas
types, gas supply, flow, agent, respiration, and spirometry loops. If
the digit field is set to show agent and no airway module is inserted,
the area is blank.
Paw, O2, and either TVexp or CO2 must show on the display during
a case. If any of these parameters are not selected to show on the
display, the right most digit field information is replaced with the
missing parameter.
See "Screen setup menu" in the "Operation" section for more
information.
Waveform fields
The split screen field can be set to show gas metabolics, trends,
spirometry loops, Paw gauge, airway compliance, and optional
2-12 2078169-001
2 System controls and menus
2078169-001 2-13
Aisys
Display navigation
Use the touchscreen and ComWheel to navigate the display.
AB.75.257
1. Menu Displays the title of the open menu. For example: Start Case.
2. Instructions or help information This shows any additional instructions or help messages.
3. Menu items Shows Case Defaults, Volume Apnea Alarm, CO2 Alarms, Age, Ideal
Weight, and Start Case Now.
Using menus
2-14 2078169-001
2 System controls and menus
Use the ComWheel to scroll through the quick key settings and
function keys, make selections, change settings, and confirm
settings.
The gas settings and the main ventilator settings for each ventilation
mode can be changed using the quick keys.
2078169-001 2-15
Aisys
2-16 2078169-001
3 Operation
Operation
2078169-001 3-1
Aisys
3-2 2078169-001
3 Operation
1. Plug the power cord into an electrical outlet. Make sure the
system circuit breaker is on.
AB.91.044
1. System switch
2. Mains indicator
Note The system must perform a power-up self test after 12 hours of
remaining on. If the system has been on longer than 12 hours
without a power-up self test, the ‘Turn power Off and On for self
tests’ alarm occurs. Turn the power off and then back on between
cases to resolve the alarm.
2078169-001 3-3
Aisys
Start a case
Use the Start Case menu to set the case data and to start the gas
flow.
A case can be started using default settings or using custom
settings. The default settings are configured by the Super User. See
the "Super user mode" section for information on the Start Case
menu defaults.
Default Settings selection shows the first of five default case types
when the Start Case menu is accessed. Four of the default case
types are configured by the Super User. The fifth default case is Last
Case.
The Ideal Weight, Age, and Volume Apnea Alarm values are set to
the pre-selected settings defined by the Super User corresponding to
the case type.
Note The TV for Ideal Body Weight menu item from the Patient
Demographics menu can only be accessed when the ventilation
mode is set to VCV, PCV-VG, SIMV VCV, and SIMV PCV-VG. Use
this setting for breath rate and tidal volume calculations based on the
set patient weight.
3-4 2078169-001
3 Operation
Start a case using the default settings by case type defined by the
Super User.
Case Defaults contain five case type selections. Each case type has
preset values for Ideal Weight, Age, and Volume Apnea Alarm.
The first four default case types are configured and named by the
Super User. The fifth default case is Last Case.
Ideal Weight, Age, CO2 Alarms, and Volume Apnea Alarm can be
custom set on the Start Case menu before starting a case.
Additional ventilator settings, ventilation mode, alarm settings, and
gas settings can be custom set through the Vent Mode menu and
other ventilation quick keys, Alarm Setup menu, and Gas Setup
menu.
2078169-001 3-5
Aisys
See the "Ventilator setup" section for information on the Vent Mode
menu.
See the "Gas setup" section for information on the Gas Setup menu.
See the "Alarm setup" section for information on the Alarm Setup
menu.
End a case
Use the End Case menu to stop gas flow and end the patient
alarms.
3-6 2078169-001
3 Operation
2078169-001 3-7
Aisys
Ventilator setup
Use the Vent Mode menu to set the ventilation mode. Use ventilator
quick keys and More Settings to change ventilator settings.
1. Select the Mode quick key. The Vent Mode menu shows.
2. Select the desired ventilation mode.
3. Set and confirm the primary ventilation setting to activate the
ventilation mode.
Controls that are frequently used in the ventilation mode can be
adjusted with the ventilator quick keys and the More Settings
quick key.
Change the ventilator settings for the ventilation mode when a case
is running.
3-8 2078169-001
3 Operation
Auto limits
Use the Auto Limits menu to quickly set alarm ranges for ‘MV’, ‘TV’,
and ‘EtCO2’ during mechanical ventilation.
Note The proposed low and high alarm limits are shown in highlighted
text. The alarm limits that are not highlighted do not change.
2078169-001 3-9
Aisys
Gas setup
Use the Gas Setup menu to adjust the Agent, O2% and total flow, to
change the balance gas, and to change the circuit type.
3-10 2078169-001
3 Operation
2078169-001 3-11
Aisys
3-12 2078169-001
3 Operation
System setup
Use System Setup to access menus and settings for Patient
Demographics, Screen Setup, Fresh Gas Usage, System Status,
Calibration, and Checkout.
Note System Status shows the status of gas supplies, electrical supplies,
and the software version.
Patient demographics
Use the Screen Setup menus to customize the screen view. Areas
of the screen can be customized to show specific information.
Screen Setup contains the Layout, Scales, Time and Date, and
More Settings submenus.
2078169-001 3-13
Aisys
Note The Time and Date menu cannot be changed when a case is
running.
4. Select Close.
3-14 2078169-001
3 Operation
2078169-001 3-15
Aisys
Note Both O2% and Flow are available if User is selected by the Super
User. If User is not selected by the Super User, either O2% or Flow
will be unavailable.
Use Fresh Gas Usage to view the volume of O2, Air, N2O, and
agents used for the three most recent cases.
• Data only shows for gases available on the system.
• Agent data shows the three most recently used agents.
3-16 2078169-001
3 Operation
Alarm setup
Use the Alarm Setup menu to set and adjust alarm limits, alarm
volume, and other alarm settings and to view alarm history. The
Alarm Setup menu contains the Primary Limits, More Limits,
Alarm History, and Configure submenus.
Setting Leak Audio to Off silences audio alarms for small leaks.
Leak Audio is automatically set to On and cannot be changed when
either the Low MV alarm limits are off or the MV/TV Alarms is set to
Off.
Note Selecting Change to Default Limits loads the default settings as set
by the Super User or the factory defaults if no Super User settings
have been entered.
Use the CO2 Alarms setting to turn off the CO2 ‘Apnea’ alarm,
‘EtCO2 low’, ‘EtCO2 high’, and ‘FiCO2 high. Absorbent OK?’ alarms
during manual ventilation.
CO2 information is obtained from the airway module installed in the
anesthesia system module bay. The CO2 alarms setting has no
effect if there is no airway module in the anesthesia system.
2078169-001 3-17
Aisys
2. To turn the volume apnea alarms off, select Vol Apnea Alarm
to Off.
‘Volume Apnea Off’ shows in the general message field.
3. To turn the volume apnea alarms on, set Vol Apnea Alarm to
On.
4. Select Close.
Use the MV/TV Alarms setting to turn off the MV and TV alarms.
Settings made during manual ventilation are not retained when
mechanical ventilation starts. Settings made during mechanical
ventilation are retained when manual ventilation starts.
For example, if MV/TV Alarms is set to Off during manual
ventilation, the alarms remain off until the Bag/Vent switch is set to
Vent or the MV/TV Alarms is set to On.
For example, if the MV/TV Alarms is set to Off during mechanical
ventilation, the alarms remain off when manual ventilation starts.
3-18 2078169-001
3 Operation
Use the Alarm History tab to view the list of the 12 most recent high
and medium priority alarms that occurred since the start of the case.
The alarm history clears at the start of a new case.
Use the Apnea Delay setting to set the desired apnea time delay.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
4. Select Close.
2078169-001 3-19
Aisys
Use the Leak Audio setting to silence audio alarms of small leaks.
3-20 2078169-001
3 Operation
4. Select Close.
2078169-001 3-21
Aisys
Alarms On Off
Use the Alarms On/Off menu to turn On or turn Off the CO2 Alarms
Limits and MV/TV Alarm Limits during manual ventilation. Alarm
limits are enabled at the start of mechanical ventilation.
Note The CO2 ‘Apnea’ alarm cannot be turned off through the Alarms
On/Off menu. Only the ‘EtCO2’ and ‘FiCO2’ alarms are turned off
using this menu.
3-22 2078169-001
3 Operation
Next page
Select Next Page to change the screen view. A default view and four
configurable screen views are available. A general message displays
identifying the page number of the screen view.
See the "Super user mode" section for information on setting the
preset screen views.
2078169-001 3-23
Aisys
Trends
Use the Trends menu to view patient trends and set the time scale.
There are three views for patient trends: measured (numerical),
settings, and graphical. Trend information is saved every 15 seconds
for the most recent 24 hours.
Setting trends
1. Select Trends.
2. Select the desired view.
3. Select Scroll to move through the current trend view.
4. Select Time Scale to select the desired scale from the drop-
down menu.
5. Select Next Page to view additional parameters.
6. Select Close.
3-24 2078169-001
3 Operation
Spirometry
Use the Spirometry menu to:
AB.98.039
1. Volume axis
2. Pressure axis
3. Real-time loop
4. Reference loop (appears on display in gray)
1. Select Spirometry.
2. Select Loop Type and select the loop from the drop-down list.
3. Select Close.
2078169-001 3-25
Aisys
WARNING Make sure that the set sensor type corresponds to the
type of airway adapter in use. If the sensor type is not set
correctly, the information displayed may not be accurate.
3-26 2078169-001
3 Operation
1. Select Spirometry.
2. To store a loop to memory, select Save Loop. Up to six loops
can be saved.
3. To view a saved loop, set Show Ref. Loop to the time at which
it was saved.
4. To delete a saved loop, set Delete Ref. Loop to the time at
which it was saved.
5. Select Close.
2078169-001 3-27
Aisys
Procedures
Use the Procedures menu to pause the gas flow, start cardiac
bypass, perform or change settings for a vital capacity procedure, or
perform or change the settings for a cycling procedure.
Note Vital Capacity shows in the menu if it is set to Yes by the Super
User. Vital Capacity is only selectable during mechanical ventilation.
Note Cycling shows in the menu if it is set to Yes by the Super User.
Cycling is only selectable during mechanical ventilation.
Use Pause Gas Flow to temporarily suspend the flow of gas during
a case. Using Pause Gas Flow while the breathing circuit is
disconnected prevents the flow of gas into the room. Pause Gas
Flow is available during both mechanical ventilation and manual
ventilation.
1. Select Procedures.
2. Select Pause Gas Flow.
The amount of time remaining in the gas flow pause shows in
the window.
Gas flow stops for 1 minute and automatically resumes after 1
minute.
If mechanical ventilation is on, mechanical ventilation stops for 1
minute and then automatically resumes after 1 minute.
3. Resume the flow of gas at any time during the pause by
selecting Restart Gas Flow.
Cardiac bypass
3-28 2078169-001
3 Operation
alarms are suspended. The alarms are enabled when VCV cardiac
bypass is turned off or mechanical ventilation is stopped.
Note PEEP and TV settings can be changed after entering cardiac bypass
mode.
4. Select Close.
Vital capacity
2078169-001 3-29
Aisys
The Pressure Hold, Hold Time, and PEEP on Exit settings can be
preset by the Super User. These settings can be changed by the
user before starting the procedure.
Cycling
Using cycling
1. Select Procedures.
3-30 2078169-001
3 Operation
2. Select Cycling.
3. Select a Procedure to perform.
4. Select Start Cycling.
The procedure begins.
Procedure progress shows in the procedure window.
5. Stop the procedure anytime by selecting Stop Cycling.
6. Select Close.
2078169-001 3-31
Aisys
Timer function
Use Start as a timer function. When selected a clock will display
counting up from zero.
3-32 2078169-001
3 Operation
ecoFLOW
This feature provides a split screen view that shows the approximate
minimum O2 flow to maintain a preset inspired O2 concentration.
Also shown is the approximate agent used per hour and the cost.
40
20 60
0 80
-20 100
15 21.91 2
9
3
100
9 2
8
10
1
7
O2 Total Iso
6 3
1.30 4 AC.22.041
Fi25 O2 flow
5 0.60
4
1. Agent Shows Agent cost and flow information.
2. Agent cost The cost of the current agent flow. This value is determined by the agent flow multiplied
by the agent cost set in Super user mode.
3. Agent flow The measured value of the liquid agent flow from the vaporizer. The agent flow may have
a delayed response. For example: Iso.
4. FiO2 flow marker The graphical representation on the flow tube of the FiO2 flow value. This marker can be
removed by disabling it in Super user mode.
5. FiO2 flow The minimum O2 flow needed to maintain the set inspired O2 flow. This item can be
disabled in Super user mode. For example: Fi25 O2 flow.
6. O2 total (numeric) The numeric representation of the total O2 flow. If N2O is the balance gas, this equals the
set O2 flow. If Air is the balance gas, this is the set O2 flow plus 21% of the Air flow. For
example: O2 Total.
7. O2 total (graphical) The graphical representation of the total O2 flow. If N2O is the balance gas, this equals
the set O2 flow. If Air is the balance gas, this is the set O2 flow plus 21% of the Air flow.
8. Remaining gas flow If N2O is the balance gas, this equals the set N2O flow. If Air is the balance gas, this N2O
is 79% of the Air flow.
9. Flow bobbin The height of this represents the total fresh gas flow delivered to the patient.
10. Total flow Shows Total Flow information.
2078169-001 3-33
Aisys
Using ecoFLOW
3-34 2078169-001
3 Operation
Alternate O2 control
2
AB.91.077
1. ON/OFF button
2. Flow control
3. Flow tube
2078169-001 3-35
Aisys
3-36 2078169-001
3 Operation
AB.82.042
Figure 3-5 • EZchange canister holder label
1
AC.22p.010
1. Side rails
2078169-001 3-37
Aisys
Note Make sure that the absorber canister has side rails. If the canister
does not have side rails, it will not work in the EZchange canister
holder.
3-38 2078169-001
3 Operation
Condenser
Use the optional condenser to remove water in the system that is
produced from the reaction of CO2 gas with the absorbent. The
condenser is connected between the outlet of the absorber canister
and the inlet of the circuit module. Moisture in the gas is condensed
into water droplets, which run into the condenser’s reservoir.
AB.82.043
2 1
1. Drain button
2. Reservoir
3. Condenser
2078169-001 3-39
Aisys
3-40 2078169-001
3 Operation
AC.20.027
Figure 3-8 • ACGO switch position
WARNING The Bag/Vent switch and the APL valve do not control
the ACGO or any breathing circuit connected to the
AGCO. Do not use these controls when using a breathing
circuit with fresh gas from the ACGO. Patient injury may
occur.
• Volume and pressure monitoring are not available
through the system when using the ACGO port. Monitor
the patient using other methods.
• Do not use an external ventilator on the ACGO. Do not
use the ACGO to drive external ventilators or for jet
ventilation.
• The maximum pressure at the ACGO can be up to 55
kPa (8 psi). Use a breathing circuit with a pressure
limiting device to limit the pressure at the patient
connection port, during normal and single-fault
2078169-001 3-41
Aisys
Note A sample of the fresh gas is diverted to the O2 cell in the breathing
system. The sample flow to the O2 cell is dependent on the pressure
in the external circuit. The sample flow reduces the fresh gas flow
rate to the auxiliary breathing circuit equal to the amount diverted to
the O2 cell.
3-42 2078169-001
4 Preoperative checkout
Preoperative checkout
2078169-001 4-1
Aisys
□ Turn the system circuit breaker off and check that mechanical
ventilation continues while the system is running on battery
power. After completing the check, turn the system circuit
breaker on. The mains indicator is lit when AC power is
connected.
□ Set the appropriate controls and alarm limits for the case.
4-2 2078169-001
4 Preoperative checkout
Note This check does not need to be done before the first case of the day
if the "Every day before your first patient" checklist was done.
□ Set the appropriate controls and alarm limits for the case.
2078169-001 4-3
4-4 2078169-001
5 Preoperative tests
Preoperative tests
2078169-001 5-1
Aisys
1. Using the liquid level indicator, check that the cassette is filled to
the appropriate level.
2. If using Aladin2, unlock the cassette handle before installing it
into the active bay.
3. Insert the cassette into the active bay until a click is heard. An
audible click indicates that the cassette is in position.
4. Turn the lock on the handle to the horizontal position (Aladin2
cassettes only).
5. The cassette is properly inserted when the agent is identified on
the display. Make sure that the displayed agent matches the
cassette.
Note Store the cassettes in the cassette storage bay when they are not in
use. See the "Vaporizer cassettes" section for additional cassette
information.
5-2 2078169-001
5 Preoperative tests
Important Room temperature fluctuations of more than 5°C may affect sensor
measurements. Recalibrate the flow sensors if the room temperature
changes by more than 5°C.
Calibrate the flow sensors by removing the flow sensor module from
the system. On-screen instructions are available through System
Setup - Calibration - Flow and Pressure.
2078169-001 5-3
Aisys
WARNING Perform a Vent and Gas test after changing the patient
tube type. Changing the patient breathing circuit after
completing a Vent and Gas test affects the volume
measurements in all modes.
5-4 2078169-001
5 Preoperative tests
Checkout menu
The Checkout menu shows on the display after turning on the
system. To access the Checkout menu between cases, select
Checkout. Step-by-step instructions show in the Checkout menu
during the tests. Use the Checkout menu to:
• Perform a Full Test.
• Perform any of the individual tests.
• View the Test Log.
• Start a case.
2078169-001 5-5
Aisys
Full test
The Full Test or the individual tests must be performed at least once
within every 24-hour period.
Perform the Full Test at the start of each day. The full test runs
automatically and beeps to indicate when it is finished or if
interaction is required.
The Full Test does the following tests: Vent and Gas, Circuit Leak,
and Circuit O2 Cell (if circuit O2 cell is present). When one of the
tests is completed, the next test begins.
1. From the Checkout menu, select Full Test and follow the
instructions.
2. If a test fails, follow the instructions to perform a retest or accept
the results.
3. When the Full Test is completed, start a case.
Note Any cassette may be used during the test. Use a non-desflurane
cassette during the test to check the full functionality of the internal
electronic control unit.
Circuit leak
The Circuit Leak test checks the Bag/Vent switch, proper gas
supply pressures, airway pressure measurement transducer, APL
valve, and manual circuit leak.
To run this test, follow the on-screen instructions. When the test
passes, the next test starts.
5-6 2078169-001
5 Preoperative tests
Circuit O2 cell
When External Gas Monitor is set to Yes by the Super User, the
External Gas Monitor reminder occurs. This is not a test. This is a
reminder to connect a respiratory gas monitor.
2078169-001 5-7
Aisys
Individual tests
The Full Test or all of the individual tests must be completed with
passing results at least once within every 24-hour period.
Individual tests allow the user to perform any combination of single
tests. These tests are helpful if there is a specific problem/alarm and
the user wishes to test only that portion of the system.
The tests do not automatically move on to the next test. After
completing a test, do another or start a case. If a test fails, follow the
instructions to perform a recheck or accept the results.
Note Any cassette may be used during the test. Use a non-desflurane
cassette during the test to check the full functionality of the internal
electronic control unit.
Circuit leak
The Circuit Leak test checks the Bag/Vent switch, proper gas
supply pressures, airway pressure measurement transducer, APL
valve, and manual circuit leak.
To run this test, follow the on-screen instructions.
Circuit O2 cell
Low P leak
5-8 2078169-001
5 Preoperative tests
Note Any cassette may be used during the check. Using a desflurane
cassette only checks for leaks between the vaporizer and common
gas outlet.
Agent delivery
The agent delivery test checks the agent delivery systems and the
cassette. This check releases agent to the circuit.
Note Any cassette may be used during the test. Use a non-desflurane
cassette during the test to check the full functionality of the internal
electronic control unit.
2078169-001 5-9
Aisys
Note For ACGO machines, perform either a negative Low P Leak test or
a positive low-pressure leak test depending on local requirements.
1. Connect the leak test device to the ACGO port with the positive-
pressure leak test adapter. Push the adapter into the ACGO port
throughout the test to get a good seal.
2. Fully open the needle valve on the test device. Keep the test
device flow tube vertical for accurate results.
CAUTION If the needle valve is not fully open, this test can damage
the pressure gauge on the test device.
3. Turn the ACGO switch to the ACGO position.
4. Set the O2 flow to 500 ml.
5. Make sure that the total flow through the flowmeter on the test
device is 0.5 l/min.
6. Make sure that the pressure gauge on the test device reads
zero.
7. Close the needle valve on the test device until the test gauge
reads 20 kPa (3 psi) (BSI) or 3 kPa (0.4 psi) (ISO).
8. If the flow through the test device is less than 0.45 l/min (ISO) or
0.4 l/min (BSI), there is a low pressure leak in the anesthesia
machine. See the "Alarms and troubleshooting" section for more
information.
9. Remove the adapter and leak test device.
10. Set the O2 flow to 1 l/min and continue flow for one minute to
flush out any residual agent.
WARNING Agent mixtures from the low-pressure leak test stay in the
system. Always flush the system with O2 after the low-
pressure leak test (1 l/min for at least one minute).
5-10 2078169-001
6 Airway modules
Airway modules
2078169-001 6-1
Aisys
Airway modules
The optional compact airway modules measure and monitor gases
delivered to the patient and exhaled through the breathing circuit.
The modules consist of:
Systems with both an airway module and an O2 cell will display the
patient inspired O2 value obtained from the airway module.
Respiratory rate is the frequency of peak (end tidal) CO2
measurements per minute. A breath is defined as a change in the
CO2 signal that exceeds 1% (8 mmHg). All concentrations are
measured and displayed breath by breath.
WARNING Remove the airway sampling line for the patient’s airway
and seal the sample port while nebulized medications are
being delivered. Nebulized medications interfere with
accurate gas reading.
• In the Spirometry menu, if the Data Source is set to Patient,
no Paw, Agent, or CO2 waveforms or numeric
information is displayed during the airway gas module
warm-up period (approximately 2 minutes). Set the Data
Source to Vent to display the waveforms and numeric
information from the ventilator during the airway gas
module warm-up period.
• Leaks in the gas sampling circuit (water trap and
sampling line) may cause inaccurate readings.
CAUTION Use only cables and accessories approved by GE
Healthcare Finland Oy. Other cables and accessories
may damage the system or interfere with measurement.
Single-use accessories are not designed to be reused.
• Strong scavenging suction on the monitor exhaust port
may change the operation pressure of the monitor and
cause inaccurate readings or internal damage.
Use only airway modules that have anesthetic agent monitoring and
O2 monitoring on this system. Refer to the manual supplied with the
airway modules for detailed information on the modules. The
following modules can be used on this system:
6-2 2078169-001
6 Airway modules
6
2
3
4
5
MD.60.008
2078169-001 6-3
Aisys
1
2
3
MD.60.013
5
6-4 2078169-001
6 Airway modules
3
AC.22.010
1. Airway module
2. Gas sampling line
3. Airway adapter with sampling line connector
2078169-001 6-5
Aisys
Parameters setup
Use the Spirometry menu to change the monitoring settings of the
data source, CO2, O2, agent, and spirometry. To access the
Spirometry menu, push the Spirometry button and then select
Setup Loops.
Data source
Set the Data Source to Patient or Vent to select the primary source
for information. If Patient is selected, the airway module will be the
first source for information. If Vent is selected, the internal sensors of
the ventilator will be the first source for information.
When setting the airway module as the data source, make sure that
a D-lite or Pedi-lite sensor is properly connected on the airway
module. If the sensor is not properly connected, but the airway
module is installed, the waveform shows no flow. The gas monitor
samples and displays room air.
If information is not available through the airway module, information
comes from the internal ventilator sensors.
6-6 2078169-001
6 Airway modules
2078169-001 6-7
Aisys
6-8 2078169-001
7 Alarms and troubleshooting
2078169-001 7-1
Aisys
Alarms
Alarms may be high priority, medium priority, or low priority. When an
alarm occurs during a case, an alarm tone sounds and the alarm
message is displayed in the alarm message field. The system
checks for alarm conditions at 1 second intervals. The alarm tone is
from 47 to 78 db(A) depending on the alarm volume setting.
Alarm priorities
Alarm priority is indicated by the color of the alarm message and the
audio sequence.
Pausing alarms
Selecting Audio Pause for an active alarm stops the audible tone for
120 seconds. The alarm message shows in the alarm message field.
Selecting Audio Pause when no medium or high priority alarms are
active prevents the audible alarm tones (audio off) for 90 seconds.
Alarms in the apnea alarm family have special silence behavior to
reduce apnea nuisance alarms. Apnea family alarms include
‘Apnea’, ‘EtCO2 low’, ‘MVexp low’, ‘RR low’, and ‘TVexp low’.
When pausing an apnea family alarm, the audio tone for the active
alarm is paused for 120 seconds. The audible tone for any additional
apnea family alarm that occurs during the audio paused period is
silenced for the remaining time shown on the audio pause
countdown. Only the audible alarm tone is silenced. The alarm
messages still show in the alarm message fields. ‘APN’ shows above
the audio pause countdown when the audible tone silence is in effect
for the apnea family alarms.
7-2 2078169-001
7 Alarms and troubleshooting
Selecting and holding Audio Pause for 2 seconds will cancel the
audio pause function.
De-escalating alarms
Some device related alarms, such as ‘No insp flow sensor’ will de-
escalate priority when the alarm is acknowledged by selecting Audio
Pause. The alarm message shows at the low-priority alarm level until
the alarm condition is resolved and the alarm is cleared. If that alarm
reoccurs after it has been resolved, the alarm occurs at its standard
priority level.
Battery indicator
The color and fill amount of the battery in use symbol indicates the
amount of battery power remaining.
2078169-001 7-3
Aisys
Internal failure
Informational tones
7-4 2078169-001
7 Alarms and troubleshooting
List of alarms
If corrective action does not resolve the alarm message, contact an
authorized service representative.
Circuit pressures and volumetric flows are measured by the
ventilator and airway gas module. If the Data Source is set to
Patient the displayed waveforms and numeric information are
measured by the airway gas module. Although not displayed, the
ventilator measurements continue and if a measured value violates
an alarm setting, the appropriate alarm occurs. The value highlighted
in the parameter numeric box may not appear to have violated the
alarm setting. Changing the Data Source to Vent will display the
circuit pressures and volumetric flows measured by the ventilator.
.
2078169-001 7-5
Aisys
7-6 2078169-001
7 Alarms and troubleshooting
2078169-001 7-7
Aisys
7-8 2078169-001
7 Alarms and troubleshooting
2078169-001 7-9
Aisys
7-10 2078169-001
7 Alarms and troubleshooting
2078169-001 7-11
Aisys
Mechanical Ventilation with For Pmax less than 30 cmH2O, the sustained pressure limit is 6 cmH2O.
PEEP Off: For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of
Pmax.
For Pmax greater than 60 cmH2O, the sustained pressure limit is 12 cmH2O.
Mechanical Ventilation with For Pmax less than 30 cmH2O, the sustained pressure limit is 6 cmH2O plus “set
PEEP On: PEEP” minus 2 cmH2O.
For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of
Pmax plus “set PEEP” minus 2 cmH2O.
For Pmax greater than 60 cmH2O, the sustained pressure limit is 12 cmH2O plus
“set PEEP” minus 2 cmH2O.
Mechanical Ventilation Off: For Pmax between 12 and 60 cmH2O, the sustained pressure limit is 50% of
Pmax.
For Pmax greater than 60 cmH2O, the sustained pressure limit is 30 cmH2O.
7-12 2078169-001
7 Alarms and troubleshooting
Alarm ranges
The alarm names are listed in the Primary Limits and More Limits
tabs on the Alarm Setup menu. See the "Super user mode" section
for more information on the alarm default settings.
2078169-001 7-13
Aisys
7-14 2078169-001
7 Alarms and troubleshooting
Alarm tests
Test the system to verify that alarms are functioning.
Note If an airway module is installed, the FiO2 readings are taken from the
module instead of from the O2 cell. A sample line must be connected
from the airway module to the breathing circuit to test the O2 alarms.
• Set the FiO2 Low alarm limit to 50%. Make sure an FiO2
low alarm occurs.
• Set the FiO2 Low alarm limit back to 21% and make sure
that the FiO2 low alarm cancels.
• Set the FiO2 High alarm limit to 50%.
• Push the O2 flush.
• Make sure the FiO2 high alarm occurs.
• Set the FiO2 High alarm limit back to 100%. Make sure that
the FiO2 high alarm cancels.
2078169-001 7-15
Aisys
7-16 2078169-001
7 Alarms and troubleshooting
2078169-001 7-17
Aisys
Electrical problems
7-18 2078169-001
7 Alarms and troubleshooting
Pneumatic problems
.
2078169-001 7-19
7-20 2078169-001
8 Setup and connections
2078169-001 8-1
Aisys
Setup safety
8-2 2078169-001
8 Setup and connections
2078169-001 8-3
Aisys
8-4 2078169-001
8 Setup and connections
6 1
5
2
AB.74p042
4
3
7
AB.74p.043
2078169-001 8-5
Aisys
8-6 2078169-001
8 Setup and connections
Removing a canister
1. Hold the canister by the handle and push on the release latch to
unlock the canister.
AB.74p058
2. Remove the canister by tilting it downward and off the two
support pins.
1. Hold the canister by the handle and push the canister cradle
release latch to unlock the canister cradle.
AB.75p088
2078169-001 8-7
Aisys
AB.75p089
Filling the Reusable Multi Absorber canister
1. Turn the canister upside down and, using your thumbs, turn the
cover locking ring counterclockwise to unlock it.
AB.74p044
8-8 2078169-001
8 Setup and connections
AB.74p047
5. To clean and disinfect the canister, refer to “Absorbent canister
cleaning” in the “Cleaning and Sterilization” manual.
6. Place a new filter in the bottom of the canister, pour absorbent
into the canister and place a new filter over the absorbent before
closing and locking the cover. Wipe off any absorbent dust.
7. Align the cover slots with the canister locking tabs and press the
cover down into place. Turn the cover locking ring clockwise to
lock the cover in place. Ensure the cover is properly sealed to
prevent leaks and spillage. Alignment of the arrows helps to
indicate correct assembly.
AB.82p001
2078169-001 8-9
Aisys
Electrical connections
Mains inlet
The arrow shows the mains power inlet and cord.
AB.75p008
Equipotential stud
Outlets
Labels show outlet voltage ratings and circuit breaker amp ratings.
These are isolated outlets. Regularly test the leakage current.
8-10 2078169-001
8 Setup and connections
AB.75p134
Serial port
The system has a Universal Serial Bus (USB) port on the back of the
display unit. This port is for use by authorized service personnel to
load software and access log files. It should not be used for other
purposes.
AB.75p133
2078169-001 8-11
Aisys
Pneumatic connections
CAUTION Use only medical grade gas supplies. Other types of gas
supplies may contain water, oil, or other contaminants
which could affect the operation of the pneumatic system.
The gas supplies provide gas to these devices through internal
connections:
Pipeline inlets
AB.75p010
Scavenging
1
2
3
AC.22p011
8-12 2078169-001
8 Setup and connections
1. Auxiliary inlet
2. Sample gas scavenging port
3. Outlet connection
2078169-001 8-13
Aisys
AB.75p127
Connect the sample gas exhaust tube to the gas return port. Exhaust
gas will be directed to the scavenging system.
AB.75p109
8-14 2078169-001
8 Setup and connections
AB.75p007
Vacuum suction regulator (optional)
2078169-001 8-15
Aisys
1. Venturi muffler
2. Overflow safety trap
3. Splash guard
4. Collection bottle connection
1. Auxiliary O2 outlet
8-16 2078169-001
8 Setup and connections
2078169-001 8-17
Aisys
8-18 2078169-001
8 Setup and connections
2078169-001 8-19
Aisys
WARNING Fully tighten the straps. If straps are not fully tightened,
equipment can fall off the top of the machine.
8-20 2078169-001
8 Setup and connections
Passive AGSS
2078169-001 8-21
Aisys
Active AGSS
• The active low flow system is for use with high vacuum disposal
systems. It requires a vacuum system capable of a continuous
nominal flow of 36 l/min and 300 mmHg (12 inHg) or greater
vacuum pressure. A flow indicator on the system indicates when
the unit is in operation.
• The active high flow system is for use with low vacuum (blower
type) disposal systems. This requires a system capable of
providing a continuous nominal flow of 50 l/min. A flow indicator
on the system indicates when the unit is in operation.
• The active adjustable flow system provides the ability to adjust
the flow with a needle valve (located in a receiver underneath
the breathing system) and a visual indicator bag which should
be properly inflated. It requires a vacuum system capable of a
continuous nominal flow of 30 l/min and 300 mmHg (12 inHg) or
greater vacuum pressure.
• The active low flow system with a 12.7 mm hose barb connector
is for use with low vacuum disposal systems. It requires an
external system with flowmeter.
• The active low flow system with a 25 mm barb connector is for
use with low vacuum disposal systems. It requires an external
venturi/ejector system with 40 to 50 l/min extract flow. A flow
indicator on the system indicates when the unit is in operation.
• The active low flow system with 30 mm ISO taper is for use with
low vacuum disposal systems. It requires an external venturi/
ejector system with 40 to 50 l/min extract flow. A flow indicator
on the system indicates when the unit is in operation.
8-22 2078169-001
8 Setup and connections
AC.22p002
Note The ball in the upper red zone indicates excessively high extraction
flow. The ball in the lower red zone indicates extraction flow rate is
too low or a blocked filter.
4. Complete the tests in the "Preoperative tests" section of this
manual.
The active adjustable AGSS flow rate is limited to 30 l/min with this
option.
To use the optional active AGSS installed on the system which uses
the three-liter bag as a visual indicator, connect it as follows:
2078169-001 8-23
Aisys
AB.75p.110
5. Complete the tests in the "Preoperative tests" section.
8-24 2078169-001
9 User maintenance
User maintenance
2078169-001 9-1
Aisys
Maintenance safety
9-2 2078169-001
9 User maintenance
Repair policy
Do not use malfunctioning equipment. Make all necessary repairs or
have the equipment serviced by an authorized service
representative. After repair, test the equipment to ensure that it is
functioning properly, in accordance with the manufacturer’s
published specifications.
To ensure full reliability, have all repairs and service done by an
authorized service representative. If this cannot be done,
replacement and maintenance of those parts listed in this manual
may be undertaken by a competent, trained individual having
experience in the repair of devices of this nature.
2078169-001 9-3
Aisys
Monthly
• Perform 21% and 100% O2 cell calibrations.
As necessary
• Install new cylinder gaskets on cylinder yokes.
• Empty the water reservoir and replace the absorbent in the canister.
• Empty and clean the overflow trap on the optional suction regulator.
• Replace the circuit O2 cell. (Under typical use the cell meets
specifications for 1 year.)
• Replace the disposable flow sensor (plastic). (Under typical use the
sensor meets specifications for a minimum of 3 months.)
• Replace the autoclavable flow sensors (metal). (Under typical use the
sensor meets specifications for a minimum of 1 year.)
• Replace the receiver filter (active gas scavenging only).
• Calibrate the airway modules every 6 months or when there are
indications of errors in the gas readings. Calibrate airway modules that
get extensive usage every 2 months.
• Inspect and clean the fan filters (display, lower electrical enclosure,
and airway module).
9-4 2078169-001
9 User maintenance
Comply with all local regulations which meet or exceed this minimum
level of maintenance.
.
2078169-001 9-5
Aisys
1. Pull the latch to unlock the flow sensor module from the
breathing system.
AB.74p052
2. Pull the flow sensor module from the breathing system.
3. Remove the O2 cell cable connector from the O2 cell and
unscrew the cell counterclockwise.
AB.74p065
5. Replace the flow sensor module on the system and push the
latch closed to secure the module.
6. Calibrate the new O2 cell according to the "Circuit O2 cell"
procedure.
9-6 2078169-001
9 User maintenance
Calibration menu
Access the Calibration menu by selecting the System Setup
button. Follow the instructions on the screen.
Important Room temperature fluctuations of more than 5°C may affect sensor
measurements. Recalibrate the flow sensors if the room temperature
changes by more than 5°C.
Calibrate the flow sensors by removing the flow sensor module from
the system. On-screen instructions are available through System
Setup - Calibration - Flow and Pressure.
Circuit O2 cell
21% O2 calibration
1. Select System Setup.
2. Select Calibration on the System Setup menu.
3. Select Circuit O2 Cell 21%.
4. Follow the instructions on the screen.
100% O2 calibration
1. Ensure the wye-piece, or patient connection, is not plugged or
there is no patient tubing connected to the system.
2. (ACGO option only.) Set the ACGO switch to Circle.
3. Set the Bag/Vent switch to Vent.
4. Select Circuit O2 Cell 100%.
2078169-001 9-7
Aisys
Water buildup
9-8 2078169-001
10 Parts
Parts
2078169-001 10-1
Aisys
AB.82.019
2
10-2 2078169-001
10 Parts
AB.82.021
5
2078169-001 10-3
Aisys
Bellows
3
12
4
11 5
6
10
9
8
AB.82.018
7
10-4 2078169-001
10 Parts
AB.82.001
2078169-001 10-5
Aisys
Absorber canister
3
1
AB.82.017
Item Description Stock number
1 Multi absorber, reusable (includes 40 pack of 1407-7004-000
foam; does not include absorbent)
2 Cover assembly, CO2 canister 1009-8240-000
3 Foam, CO2 canister (pack of 40) 1407-3201-000
4 O-ring 1407-3204-000
5 Canister, CO2 with handle 1407-3200-000
- Multi absorber, disposable, white to violet (pack 8003138
of six)
- Multi absorber, disposable, pink to white (pack 8003963
of six)
10-6 2078169-001
10 Parts
AB.82.035
Description Stock number
Exhalation valve assembly 1407-7005-000
2078169-001 10-7
Aisys
AGSS
10-8 2078169-001
10 Parts
2078169-001 10-9
Aisys
1
3
2 4
AB.82.131
Item Description Stock number
1 EZchange canister module, includes valve and 1407-7021-000
cap
2 Valve 1407-7023-000
3 Cap 1407-3130-000
4 Condenser 1407-7024-000
- EZchange canister module with condenser 1407-7027-000
10-10 2078169-001
10 Parts
Condenser
AB.82.045
Item Description Stock number
- Condenser assembly (includes module and 1407-7026-000
condenser)
1 Condenser module 1407-7025-000
2 Condenser 1407-7024-000
2078169-001 10-11
Aisys
10-12 2078169-001
11 Specifications and theory of
operation
2078169-001 11-1
Aisys
AB.75.282
11-2 2078169-001
11 Specifications and theory of operation
1. Auxiliary O2, 0-10 l/min (optional) 46. Patient wye (patient connection)
2. 241 kPa (35 psi) secondary O2 regulator 47. Airway transducer
3. Pipeline: O2, Air, N2O 48. Inspiratory flow transducer
4. Cylinder: O2, Air, N2O 49. Expiratory flow transducer
5. 758 kPa (110 psi) relief 50. Manifold pressure transducer
6. Venturi drive gas connection 51. Passive gas scavenging interface
7. Selector valve, O2, Air, N2O 52. 30 mm male to disposal system
8. O2 flush 53. 0.3 cmH2O entrainment
9. 0-120 l/min flow 54. Reservoir
10. Alternate O2 55. Adjustable gas scavenging interface
11. System switch 56. DISS EVAC connector
12. Alternate O2 disable valve 57. Active gas scavenging interface
13. ACGO/non-circle port 58. Room air
14. Vent drive gas select 59. Filter
15. O2 flow controller 60. High or low flow restrictor
16. Balance flow controller 61. Flow indicator
17. Gas inlet valve 62. To disposal system
18. Vent drive gas regulator 172 kPa (25 psi) at 15 63. ACGO variant
l/min
19. Inspiratory flow control valve 64. Circle or ACGO mode switch
20. Drive gas check valve (3.5 cmH2O bias) 65. 22 mm ACGO port
21. Mechanical over pressure valve (110 cmH2O) 66. O2 cell
22. Free breathing check valve 67. Non-circle variant
23. Atmosphere 68. Circle or non-circle mode switch
24. Pressure relief valve 69. Non-circle circuit pressure limiting valve
25. Exhalation valve (2 cmH2O bias) 70. Bag/Vent switch
26. Relief valve 10 cmH2O 71. Bag
27. 0-10 l/min drive gas, 0-10 l/min patient and fresh 72. APL valve 0-70 cmH2O
gas, 0-10 l/min total typical flow
28. 200 ml reservoir 73. Default factory connection
29. Control bleed to ambient approximately 1 l/min at 3 74. Gas monitor
cmH2O if continuous (rate dependent)
30. Vent to ambient 75. Negative pressure relief valve
31. Flush switch 37.2 kPa (5.4 psi) 76. Absorber
32. Electronic vaporizer 77. Drain
33. ACGO port relief valve 78. EZchange canister
34. Cassette inflow measure 79. Condenser
35. Cassette inflow valve 80. Bypass back-pressure valve
36. Cassette identification 81. Inflow check valve
37. Cassette outflow measure 82. Cassette pressure sensor
38. Cassette flow control valve 83. Cassette connection valve
39. Liquid flow prevention valve 84. Aladin cassette
40. Cassette outflow valve 85. Pneumatic power outlet
41. Scavenging valve 86. Embedded temperature sensor (not available on all
cassettes)
42. Cassette temperature sensor 87. Fresh gas sample relief valve
43. Liquid level indicator (not available on all cassettes) 88. E or M-series airway module
44. Expiratory flow sensor 89. CARESCAPE airway module
45. Inspiratory flow sensor
2078169-001 11-3
Aisys
Gas supplies
O2 flow
11-4 2078169-001
11 Specifications and theory of operation
Mixed gas
Either Air or N2O can be selected as the balance gas. The balance
gas flow is controlled by the mixer. When Alternate O2 control is in
use, only O2 is flowing and balance gas is disabled.
The mixed gas goes from the mixer outlet through the vaporizer to
the fresh gas outlet and into the breathing system. A pressure-relief
valve sets the maximum outlet pressure.
Note When the total is adjusted toward the lower flow limit and the
calculated O2 flow is required to go below 100 ml, a message shows
in the gas settings area of the screen. The anesthesia system
calculates the required O2% to maintain the user-set total flow.
EZchange canister
Condenser
2078169-001 11-5
Aisys
Pneumatic specifications
Gas supplies
A relief valve on the vaporizer manifold limits the fresh gas pressure
at the ACGO port to 55 kPa (8 psi) at 45 l/min.
A relief valve limits the fresh gas pressure at the inspiratory port to
27 kPa (4 psi) at 55 l/min when non-circle circuit is selected.
The pneumatic outlet is supplied with the same gas supply pressure
as the attached oxygen medical gas pipeline supply. The anesthesia
11-6 2078169-001
11 Specifications and theory of operation
system is rated for supply pressures between 280 kPa - 600 kPa.
The anesthesia system is designed to function normally with a
maximum flow of 8 l/min from the pneumatic outlet. Higher flows
from this connection are possible and may result in decreased
anesthesia system performance.
2078169-001 11-7
Aisys
5
9 10
1 2 3 4 6 7 8
20 22
25
19 21 23 24
26
18 27
17 14 28
16
15 29
56
11 12
13
55 57
58
40
41 42 45 49
50 51
52
30
53
31 54
46 47 48
32
33
34 37
35 38 43 44
36 39 63
59 60
69
70
71
62 72
61
73
74 AB.75.265
64 65 66 67 68
11-8 2078169-001
11 Specifications and theory of operation
2078169-001 11-9
Aisys
6
9 10
1 2 3 4 5 7 8
25 27 12
30 11
24 26 28 29 13
31
23 32
22 33
21 19 34
20
15
35 18 14
37 38 16
36 17
39
40 49 50
46
41
47
42
48
43 58
54
44 51 59 60
45 61
52 53
55 56 57 62
63
64
65 66
69
70
71
67 72
68
73
74 AB.75.283
79 78 77 76 75
11-10 2078169-001
11 Specifications and theory of operation
13. Battery backup for monitor 53. Gas inlet valve with cable
14. Display connector board 54. Expiratory and inspiratory flow sensors
15. I/O port 55. O2 cell
16. Airway module power supply board 56. Bag/Vent switch
17. Airway module 57. ABS connected switch
18. Anesthesia control board 58. Light strip boards
19. Agent delivery board 59. Light strip switch
20. Cassette temperature sensor 60. CGO valve
21. Inflow zero valve 61. ACGO/non-circle circuit switch
22. Outflow zero valve 62. CGO switch
23. Inflow pressure sensor board 63. O2 flush switch
24. AC mains LED 64. Serial I/O port
25. Outflow pressure sensor board 65. Communication to system
26. Inflow shutoff valve 66. DC power in
27. Outflow shutoff valve 67. LCD with touchscreen
28. Cassette interface board 68. Display carrier board and CPU
29. Cassette 69. USB I/O port (software upgrade only)
30. Liquid level sensor (not available on all cassettes) 70. USB I/O port
31. Embedded temperature sensor (not available on all 71. Network I/O port - 1 (Ethernet)
cassettes)
32. Scavenging valve 72. Network I/O port - 2 (Ethernet)
33. Cassette pressure sensor 73. VGA output
34. Proportional valve 74. Patient monitoring on/standby
35. Pan connector board 75. ComWheel encoder
36. Ventilator interface board 76. Keypad
37. Electronic gas mixer board 77. Display unit fan with cable
38. Pan fan with cable 78. Speaker 2
39. Air cylinder pressure transducer 79. Speaker 1
40. N2O cylinder pressure transducer
2078169-001 11-11
Aisys
Electrical power
Power cord
Length 5 meters
Voltage rating 100 to 240 Vac
Current capacity 10 A for 220-240 Vac
15 A for 100-120 Vac
10 A for 120/220-240 Vac
Type Three conductor power supply cord (medical
grade where required).
Battery information
The system is not a portable unit. A sealed lead acid battery supplies
backup power in the event of a power failure.
11-12 2078169-001
11 Specifications and theory of operation
2078169-001 11-13
Aisys
Flow specifications
Alternate O2
Minimum flow 500 ml/min to 700 ml/min
Maximum flow 10 l/min to 13 l/min
Indicator Flow tube
Indicator accuracy +/- 5% full scale
Fresh gas
Flow range 0 and 200 ml/min to 15 l/min
Minimum total flow O2 and balance gas is
200 ml/min
Total flow accuracy +/- 10% or +/- 40 ml/min of setting
(whichever is larger)
O2 flow accuracy +/- 5% or +/- 20 ml/min of setting (whichever
is larger)
Balance gas flow accuracy +/- 5% or +/- 20 ml/min of setting (whichever
is larger) Air/N2O
O2 concentration range 21%, 25% to 100%
O2 concentration accuracy +/- 2.5% of setting for total flow of greater
than or equal to 1 l/min and less than or
equal to 15 l/min
+/- 5% of setting for total flow of greater than
or equal to 0.4 l/min and less than 1 l/min
+/- 6.5% of setting for total flow of greater
than or equal to 0.2 l/min and less than 0.4
l/min
Electronic mixer response 500 ms (10% to 90% flow step)
time
Compensation Temperature and atmospheric pressure
compensated to standard conditions of 20°C
and 101.3 kPa (14.7 psi)
Hypoxic guard Electronic
Note: Leakage from one gas inlet to another gas inlet is less than 10 ml per
hour.
Note: The stated flow accuracy may not be met for total flows between 200
and 400 ml/min. The total flow will maintain a minimum of 21% O2.
11-14 2078169-001
11 Specifications and theory of operation
2078169-001 11-15
Aisys
Gas scavenging
All scavenging
Positive pressure relief 10 cmH2O
Passive scavenging
Negative pressure relief 0.3 cmH2O
Outlet connector 30 mm male taper ISO
11-16 2078169-001
11 Specifications and theory of operation
Active scavenging
Disposal system Outlet Flow range Pressure
type connector*
Adjustable DISS EVAC Up to 30 l/min not applicable
High flow, low BSI 30 mm 50 to 80 l/min not applicable
vacuum threaded
Low flow, high DISS EVAC 36 +/- 4 l/min 305 mmHg (12
vacuum inHg) minimum
vacuum
Low flow, low 12.7 mm hose 36 +/- 4 l/min not applicable
vacuum barb
Low flow, low 25 mm hose barb 40 to 50 l/min not applicable
vacuum
Low flow, low 30 mm ISO taper 40 to 50 l/min not applicable
vacuum
*Other market-specific connectors may be available.
Particle filter at the outlet has a pore size of 225 microns. All flow data uses
a new filter.
2078169-001 11-17
Aisys
Physical specifications
All specifications are approximate values and can change without
notice.
11-18 2078169-001
11 Specifications and theory of operation
25
20
15
10
AC.22.039
0
0 10 20 30 40 50 60 70 80
12
11
10
6 AC.22.020
5
40 45 50 55 60 65 70 75
2078169-001 11-19
Aisys
Environmental requirements
.
11-20 2078169-001
11 Specifications and theory of operation
E- and M-series
Airway humidity 0 to 100% condensing
Sampling delay 2.5 seconds typical with a 3 m sampling line
Total system response time 2.9 seconds typical with a 3 m sampling line, including sampling delay and rise time
Warm-up time 2 minutes for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
30 minutes for full specifications
Respiration rate 4 to 60 breaths/min
Diversion flow 200 ml/min +/- 20 ml/min
Airway pressure -20 cmH2O to 100 cmH2O
2078169-001 11-21
Aisys
CARESCAPE
Airway humidity 0 to 100% condensing
Sampling delay 3.0 seconds typical with a 3 m sampling line
Total system response time Less than 3.8 seconds with a 3 m sampling line
Warm-up time 1 minute for operation with CO2, O2, and N2O
5 minutes for operation of anesthetic agents
20 minutes for full specifications
Respiration rate 4 to 100 breaths/min
Diversion flow 120 +/- 20 ml/min
Airway pressure -20 cmH2O to 100 cmH2O
11-22 2078169-001
11 Specifications and theory of operation
Typical performance
.
2078169-001 11-23
Aisys
11-24 2078169-001
11 Specifications and theory of operation
Ventilator theory
The ventilator pneumatics are at the rear of the breathing system. A
precision valve controls gas flow to the patient. During inspiration,
this gas flow closes the exhalation valve and pushes the bellows
down. During expiration, a small flow pressurizes the exhalation
diaphragm to supply PEEP pressure. If the maximum pressure
(pmax) is reached during inspiration, the ventilator will cycle to
expiration.
Volume and flow measurements come from flow sensors in the flow
sensor module. Two tubes from each sensor connect to a transducer
that measures the pressure change across the sensor, which
changes with the flow. A third transducer measures airway pressures
at the inspiratory flow sensor.
The ventilator uses the data from the flow sensors for volume-related
numerics and alarms. The ventilator also uses the flow sensors to
adjust its output for changes in fresh gas flow, small leaks, and gas
compression upstream of the breathing circuit. There is adjustment
for compression in the patient circuit.
In volume ventilation modes, certain alarm conditions prevent the
automatic adjustment of ventilator delivery based on measured flow
values. In these cases, ‘TV accuracy decreased. Adjust manually.’
displays above the ventilator setting area of the screen. When this
message shows, the ventilator may not be able to deliver within the
accuracy range specified. When this occurs, the tidal volume must
be manually adjusted until the volume delivered reaches the desired
level. If compensation stops for a number of breaths, the condition
causing the hold shows as an alarm. Automatic volume
compensation resumes when the alarm conditions are resolved.
For better precision a small quantity of gas bleeds through a resistor
to keep pressure on the exhalation valve constant. At high airway
pressures, this can cause a slight hiss during inspiration.
2078169-001 11-25
Aisys
The ecoFLOW option provides a split screen view that shows the
approximate minimum O2 flow to maintain a preset FiO2 value. It
also shows the approximate agent used per hour and the cost. Some
information associated with the ecoFLOW option is not provided
when the airway module is warming up, not present, or non-
functional.
The split screen shows the Paw gauge in the upper area and the
ecoFLOW gauge in the lower portion of the screen. The ecoFLOW
gauge consists of a fresh gas flow tube, an agent flow indicator, and
related parameters.
The fresh gas flow tube is a stacked flow tube showing the total O2
flow on the bottom and the remaining gas (N2 or N2O) on top. These
gases determine the total fresh gas flow setting. Below the fresh gas
flow tube is the measured total O2 flow to the patient and the
calculated FiO2 flow value. The FiO2 flow value is based on the
FiO2 Setting specified in Super User mode. This is the minimum O2
flow needed to deliver a preset inspired O2 concentration. The FiO2
flow value is specific to each patient and case. It is calculated using
the fresh gas settings, the patient O2 uptake, the dilution effect of
agent being delivered, and the effects of the circle breathing system.
The agent flow indicator shows the amount of liquid agent flow as
related to the fresh gas setting. The calculated cost of the agent
shows above this indicator. This cost is based on agent flow and the
values entered in the Agent Costs menu set in Super User mode.
Ventilation modes
11-26 2078169-001
11 Specifications and theory of operation
• TV
• RR
• I:E
• Tpause
• PEEP
• Pmax
2
3 4
1
5
7
6
AB.98.036
1. Paw waveform
2. Tinsp
3. Insp Pause
2078169-001 11-27
Aisys
4. Texp
5. PEEP
6. Flow waveform
7. TV
• Pinsp
• RR
• I:E
• PEEP
• Pmax
• Rise Rate
2 3
4
1
5
6
AB.98.037
1. Paw waveform
2. Tinsp
3. Texp
4. Pinsp
5. PEEP
11-28 2078169-001
11 Specifications and theory of operation
6. Flow waveform
PCV-VG mode
In PCV-VG, a tidal volume is set and the ventilator delivers that
volume using a decelerating flow and a constant pressure. The
ventilator will adjust the inspiratory pressure needed to deliver the
set tidal volume breath-by-breath so that the lowest pressure is used.
The pressure range that the ventilator will use is between the PEEP
+ 2 cmH2O level on the low end and 5 cmH2O below Pmax on the
high end. The inspiratory pressure change between breaths is a
maximum of +/- 3 cmH2O. If a high airway pressure alarm is active
due to the current breath, the next breath’s target will be 0.5 cmH2O
less than the current breath’s pressure target.
This mode will deliver breaths with the efficiency of pressure
controlled ventilation, yet still compensate for changes in the
patient’s lung characteristics. PCV-VG begins by first delivering a
volume breath at the set tidal volume. The patient’s compliance is
determined from this volume breath and the inspiratory pressure
level is then established for the next PCV-VG breath.
PCV-VG mode setting:
• TV
• RR
• I:E
• PEEP
• Pmax
• Rise Rate
2 3
1
5
7
6
AB.98.034
1. Paw waveform
2. Tinsp
2078169-001 11-29
Aisys
3. Texp
4. Variable pressure to deliver desired TV
5. PEEP
6. Flow waveform
7. TV
• TV
• RR
• Tinsp
• Tpause
• Psupport
• PEEP
• Pmax
• Trig Window
• Flow Trigger
• End of Breath
• Rise Rate
11-30 2078169-001
11 Specifications and theory of operation
1
4
5 2
AB.91.040
TI
6
• Pinsp
• RR
• Tinsp
• Psupport
• PEEP
• Pmax
• Trig Window
• Flow Trigger
• End of Breath
• Rise Rate
• Exit Backup
2078169-001 11-31
Aisys
2 3 4
6
1 5
AB.98.038
1. Paw waveform
2. Tinsp
3. Spontaneous breathing period
4. Trigger window
5. Pressure supported breath
6. Pinsp
7. Flow waveform
11-32 2078169-001
11 Specifications and theory of operation
• TV
• RR
• Flow Trigger
• Psupport
• PEEP
• Tinsp
• Pmax
• Trig Window
• End of Breath
• Rise Rate
• PSV Rise Rate
2 3 4
1 6
8
7
AB.98.153
1. Paw waveform
2. Tinsp
3. Spontaneous breathing period
4. Trig Window
5. Variable pressure to deliver desired TV
2078169-001 11-33
Aisys
6. PEEP
7. Flow waveform
8. TV
PSVPro mode
PSVPro is pressure supported ventilation with apnea backup.
PSVPro is a spontaneous mode of ventilation that provides a
constant support pressure once the ventilator senses that the patient
has made an inspiratory effort. In this mode, the clinician sets the
Pressure Support (Psupport) and PEEP levels. The patient
establishes the rate, inspiratory flow, and inspiratory time. The tidal
volume is determined by the pressure, lung characteristics, and
patient effort.
PSVPro uses an inspiration termination level that establishes when
the ventilator will stop the pressure-supported breath and cycle to the
expiratory phase. The inspiration termination level is user adjustable
from 5 to 75%. This parameter sets the percent of peak inspiratory
flow that the ventilator uses to end the inspiratory phase of the
breath and to cycle into the expiratory phase. If the inspiration
termination is set to 30% then the ventilator will stop inspiration when
the flow decelerates to a level equal to 30% of the measured peak
inspiratory flow. The lower the setting, the shorter the inspiratory
phase.
An apnea backup mode is provided if the patient stops breathing.
When setting this mode the clinician adjusts the inspiratory pressure
(Pinsp), respiratory rate (RR), and the inspiratory time (Tinsp). As
long as the patient is triggering the ventilator and the apnea alarm
does not activate, the patient will get pressure-supported breaths
and the ventilator will not deliver machine breaths.
If the patient stops triggering the ventilator for the set apnea delay
time, the apnea alarm will activate and the ventilator will
automatically switch to the backup mode that is SIMV PCV. Once in
this mode the ventilator will begin delivering machine pressure
controlled breaths at the inspiratory pressure level, inspiratory time,
and rate that the user has set. If, during this time, the patient takes
spontaneous breaths in between the machine breaths, the patient
will receive pressure supported breaths.
PSVPro mode settings:
• Psupport
• PEEP
• Trig Window
• Flow Trigger
• End of Breath
• Pmax
• Backup Time
11-34 2078169-001
11 Specifications and theory of operation
• Pinsp
• RR
• Tinsp
• Rise Rate
• Exit Backup
AB.91.041
TI
3
1. Paw
2. PEEP
3. Time
When the ventilator switches to the backup mode, the ‘Backup Mode
active’ alarm shows until PSVPro is reinstated or until another
ventilation mode is selected. PSVPro mode automatically resumes
when the ventilator registers the number of consecutive patient-
triggered breaths set for the Exit Backup setting. The factory default
setting for Exit Backup is 2. When Exit Backup is set to Off, the user
must reselect the PSVPro mode to reactivate PSVPro. Upon
returning to PSVPro the ventilator immediately begins providing
pressure supported breaths to the patient using the established
settings.
Spontaneous breaths that occur during this mode are indicated by a
color change in the waveform.
2078169-001 11-35
Aisys
• Psupport
• PEEP
• Flow Trigger
• End of Breath
• Pmax
• Pinsp
• Minimum RR
• Tinsp
• Rise Rate
2
AB.91.041
TI
3
1. Paw
2. PEEP
3. Time
11-36 2078169-001
11 Specifications and theory of operation
The ventilation mode factory default settings for the ADULT case
default show in the following table. An * indicates the setting is not
used for the ventilation mode. See the "Super user mode" section for
all other case default values.
.
2078169-001 11-37
Aisys
11-38 2078169-001
11 Specifications and theory of operation
Pneumatics
Pressure
Volume
2078169-001 11-39
Aisys
Oxygen
11-40 2078169-001
11 Specifications and theory of operation
Delivery accuracy
Volume delivery accuracy Greater than 210 ml tidal volume - accuracy better than 7%.
Less than or equal to 210 ml but greater than or equal to 60 ml tidal
volume - accuracy better than 15 ml.
Less than 60 ml tidal volume - accuracy better than 10 ml.
Volume monitoring accuracy Greater than 210 ml tidal volume - accuracy better than 9%.
Less than or equal to 210 ml but greater than or equal to 60 ml tidal
volume - accuracy better than 18 ml.
Less than 60 ml tidal volume - accuracy better than 10 ml.
Inspiratory pressure delivery accuracy Greater of +/- 10% or +/- 3 cmH2O
PEEP delivery accuracy +/- 1.5 cmH2O
Pressure monitoring accuracy Greater of +/- 5% or +/- 2 cmH2O
Note: Gas composition errors may be in additions to the above normalized accuracy. When adding errors, positive
errors can have the effect of nulling out negative errors.
Note: Use of anesthetic agent could affect the errors by approximately -0.95%/% volume agent in normal mode.
Note: Temperatures greater than or less than 24°C may decrease BTPS accuracy.
AC.22.029
2078169-001 11-41
Aisys
Response time
To 90% of step; measured at fresh gas outlet Less than 7 seconds at fresh gas flow of 2 l/min
AB.60.055
+20%, +1.0% v/v
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
12000
13000
14000
15000
11000
11-42 2078169-001
11 Specifications and theory of operation
3 4 5 6
AB.60.056
+20%, +1.0% v/v
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
12000
13000
14000
15000
11000
2
AB.60.057
18-25C
10-35C
-20%, -0.4% v/v
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
12000
13000
14000
15000
11000
2078169-001 11-43
Aisys
AB.60.059
+20%, +0.4% v/v
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
12000
13000
14000
15000
11000
2
11-44 2078169-001
11 Specifications and theory of operation
3 4 5
AB.60.060
+20%, +0.4% v/v
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
12000
13000
14000
15000
11000
2
Note The effect of the fresh gas composition, back pressure (including
ventilation), and ambient pressure on the agent concentration are
included in the accuracy specifications.
Note The system delivers the same output in volumetric percent (% v/v)
within the specified altitude range.
2078169-001 11-45
Aisys
Aladin2 cassettes
.
Filling
Filling system Easy-Fil: Adapter filler system for isoflurane, enflurane, and sevoflurane.
Quick-Fil: Filler system compatible to Abbot’s system for sevoflurane.
Saf-T-Fill: Filler system compatible to the Datex-Ohmeda Saf-T-Fil bottle for desflurane.
Filling speed Greater than 2 ml/s
Overfilling protection Overfilling prevention systems built into the cassettes.
Cassette
Empty weight 2.8 kg
Height 7 cm
Depth 24 cm
Width 14 cm
Aladin cassettes
.
Filling
Filling system Adapter filling: Rectangular keyed adapter filler system for halothane, isoflurane,
enflurane, and sevoflurane.
Quik-Fil: filler system compatible to Abbot’s system for sevoflurane.
Cylindrical keyed adapter: Filler system compatible to the Datex-Ohmeda Saf-T-Fil bottle
for desflurane.
Filling speed Greater than 2 ml/s
Overfilling protection Overfilling prevention systems built into the cassettes.
Liquid capacity
Maximum 250 ml
11-46 2078169-001
11 Specifications and theory of operation
Liquid capacity
Normal fill when indicator shows empty 150 ml (residual volume 100 ml)
Cassette
Empty weight 2 kg: enflurane, isoflurane, sevoflurane with keyed filler
2.5 kg: halothane with keyed filler, sevoflurane with Quik-Fil
3 kg: desflurane
Height 7 cm
Depth 23 cm
Width 14 cm
16 cm with keyed filler
2078169-001 11-47
Aisys
Essential performance
Cables
11-48 2078169-001
11 Specifications and theory of operation
2078169-001 11-49
Aisys
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment guidance
Electrostatic +/- 6 kV contact +/- 6 kV contact Floors should be wood, concrete, or ceramic tile.
discharge (ESD) +/- 8 kV air +/- 8 kV air If floors are covered with synthetic material, the
IEC 61000-4-2 relative humidity should be at least 30%.
Electrical fast +/- 2 kV for power supply +/- 2 kV for power Mains power quality should be that of a typical
transient/burst lines supply lines commercial and/or hospital environment.
IEC 61000-4-4 +/- 1 kV for input/output +/- 1 kV for input/
lines output lines
Surge IEC +/- 1 kV line(s) to line(s) +/- 1 kV differential Mains power quality should be that of a typical
61000-4-5 +/- 2 kV line(s) to earth mode commercial and/or hospital environment.
+/- 2 kV common
mode
Voltage dips, <5% UT (>95% dip in UT) <5% UT (>95% dip Mains power quality should be that of a typical
short for 0.5 cycle in UT) for 0.5 cycle commercial and/or hospital environment. If the
interruptions, and 40% UT (60% dip in UT) 40% UT (60% dip user of the system requires continued operation
voltage variations for 5 cycles in UT) for 5 cycles during power mains interruptions, it is
on power supply 70% UT (30% dip in UT) 70% UT (30% dip recommended that the system be powered from
input lines IEC for 25 cycles in UT) for 25 an uninterruptible power supply or battery.
61000-4-11 <5% UT (>95% dip in UT) cycles
for 5 seconds <5% UT (>95% dip
in UT) for 5
seconds
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at
(50/60 Hz levels characteristic of a typical location in a
magnetic field typical commercial or hospital environment.
IEC 61000-4-8
Note: UT is the AC mains voltage prior to application of the test level.
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment guidance
recommended separation distance
Portable and mobile RF communications
equipment should be used no closer to any
part of the system, including cables, than
the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
11-50 2078169-001
11 Specifications and theory of operation
Immunity test IEC 60601-1-2 test level Compliance level Electromagnetic environment guidance
recommended separation distance
Conducted RF IEC 3 Vrms 10 Vrms (V1) D=0.35√P
61000-4-6 150 kHz to 80 MHz outside
ISM bands
10 Vrms 10 Vrms (V2) D=1.2√P
150 kHz to 80 MHz in ISM
bands
Radiated RF IEC 10 V/m 10 V/m (E1) D=1.2√P 80 MHz to 800 MHz
61000-4-3 80 MHz to 2.5 GHz D=2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and D is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey, should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
• The ISM (industrial, scientific, and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz,
13.553 MHz to 13.567 MHz, 26.957 MHz to 27.283 MHz, and 40.66 MHz to 40.70 MHz.
• The compliance levels in the ISM frequency bands between 150 kHz and 80 MHz and in the frequency range
80 MHz to 2.5 GHz are intended to decrease the likelihood that a mobile/portable communications equipment
could cause interference if it is inadvertently brought into patient areas. For this reason, an additional factor of
10/3 has been incorporated into calculating the recommended separation distance for transmitters in these
frequency ranges.
• Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the system is used exceeds
the applicable RF compliance level above, the system should be observed to verify normal operation. If
abnormal performance is observed, additional measures may not be necessary, such as re-orienting or
relocating the system.
• Over the frequency range of 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
2078169-001 11-51
Aisys
11-52 2078169-001
12 Super user mode
2078169-001 12-1
Aisys
Note The menu selections shown in the examples are the factory default
values. The available settings are shown to the right of the menu.
12-2 2078169-001
12 Super user mode
Gas usage
Use the Gas Usage menu to view the total fresh gas usage since the
last reset, to reset the gas usage to zero, and view optional
ecoFLOW and agent cost information.
Setting ecoFLOW
Use the ecoFLOW menu to set the split screen defaults for FiO2
Setting, FiO2 Line, and Agent Cost. This menu item is only
available on systems with the ecoFLOW option installed.
1. From Super User mode, select Gas Usage and then ecoFLOW.
2. Select the item to change. Make the change.
For FiO2 Setting, set the numeric value for FiO2 on the split
screen.
For FiO2 Line, select whether to show or hide the FiO2 line on
the split screen.
For Agent Cost, select whether to show or hide the agent
delivered costs on the split screen.
3. Select Back to return to the Gas Usage menu.
Use the Agent Costs menu to modify the currency and agent cost
defaults used to show the cost of agent delivered on the split screen.
This menu item is only available on systems with the ecoFLOW
option installed.
1. From Super User mode, select Gas Usage and then Agent
Costs.
2. Select the Currency from the drop-down list.
2078169-001 12-3
Aisys
12-4 2078169-001
12 Super user mode
System configuration
Use the System Config. menu to set default information for the
system. Use the System Config. menu to access menus and
settings for Colors, Units, Ventilator Settings, Alarm Settings,
Parameter Settings, Trends Setup, and Pages Setup.
Display settings
Use the Colors menu to set the default colors for waveforms, digit
fields, and trends.
Use the Units menu to set the default units for parameters and
currency.
Setting colors
Set Paw, Flow, Respiratory, and CO2 parameters.
Setting units
Set Patient Weight, CO2, Gas Supply Pressure, Paw, and
Currency to the desired units. Currency is only available for
systems with the ecoFLOW option installed.
1. From Super User mode, select System Config. and then Units.
2. For each parameter, select the desired unit from the drop-down
list.
3. Select Back to return to the System Config. menu.
Ventilator settings
Use the Ventilator Settings menu to set the default user settings for
VCV Cardiac Bypass, Circuit Compliance, Sample
Compensation, and TV for Ideal Body Weight.
2078169-001 12-5
Aisys
Alarm settings
Use the Alarm Settings menu to set External Gas Monitor, Apnea
Alarm Filter, and Show Alarm Limits default settings.
Parameter settings
12-6 2078169-001
12 Super user mode
Note When set to ATPD, the volume calculation conditions are based on
ambient temperature and pressure, dry humidity conditions. When
set to BTPS, the volume calculation conditions are based on body
temperature, ambient pressure, saturated humidity conditions.
2078169-001 12-7
Aisys
Trends setup
Use the Trends Setup menu to set the default trend graphical page
layout for the top, middle, and bottom trends for each page shown in
the user’s Trends menu.
1. From the Super User mode, select System Config. and then
Trends Setup.
2. Use the drop-down lists to set Top Trend, Middle Trend, and
Bottom Trend to show the desired parameters.
3. Select Next Page to set up graphical trend defaults for another
page or Back to return to the System Config. menu.
Page setup
Use the Pages Setup menu to set the default screen views for
waveform fields, digit fields, and split screen areas. Four default
screen views can be configured.
12-8 2078169-001
12 Super user mode
1. From Super User mode, select System Config. and then Pages
Setup.
2. Select the page to set up.
3. Select an item to change. Make the change.
Waveform fields can be set to AA, CO2, Flow, Paw, or Off.
Digit fields can be set to Gases, Respiration, Gas supplies,
Flow, Agent, Loops, or Off.
Split screen can be set to None, Trends, Spirometry, Paw,
Compliance, or optional ecoFLOW.
4. Select another item. Make the change.
5. When done, select Back to select another page to set up.
2078169-001 12-9
Aisys
Case defaults
Use the Case Defaults menu to set the default case types that show
on the Start Case menu and access the Volume Apnea Setup
menu.
Use the Volume Apnea Setup menu to enable or disable the user’s
ability to turn the volume apnea alarms off during manual ventilation.
12-10 2078169-001
12 Super user mode
Each case has multiple settings. The default settings for the default
case types are shown in the following table. An * indicates that the
setting is not used for the default ventilation mode in the case type.
Use the empty columns to write in facility changes.
Note VCV ADULT settings are used as the default if the system does not
have an optional ventilation mode shown.
.
2078169-001 12-11
Aisys
12-12 2078169-001
12 Super user mode
2078169-001 12-13
Aisys
12-14 2078169-001
12 Super user mode
Gas controls
Use the Gas Controls menu to set the user’s Agents, Total Flow,
and O2% quick select preset values and the Fresh Gas Controls
setting.
2078169-001 12-15
Aisys
Procedures setup
Use the Procedures menu to set the procedures available to the
user. From this menu select the default settings for Vital Capacity
and Cycling procedures.
12-16 2078169-001
13 Vaporizer cassettes
Vaporizer cassettes
2078169-001 13-1
Aisys
Vaporizer
The electronically controlled vaporizer consists of the internal
electronic control unit and the Aladin agent cassette. The agent
cassettes are color coded, have indexed filling ports, and are
magnetically coded for each agent. The electronic control unit
governs the flow through the agent cassette and the agent
concentration in the fresh gas flow.
Both the Aladin2 and the Aladin cassettes can be used on this
system.
Remove the cassette from the active bay when not administering
anesthetic agent. Store cassettes in the cassette storage bay when
they are not in use. Store the cassettes and agent at the same
temperature as the system.
13-2 2078169-001
13 Vaporizer cassettes
AB.60.050
4 3
2
AB.60.038
4 3
2078169-001 13-3
Aisys
13-4 2078169-001
13 Vaporizer cassettes
AB.60.004
4 3 2
1
AB.60.005
3 2
2078169-001 13-5
Aisys
AB.60.003
3 2
13-6 2078169-001
13 Vaporizer cassettes
2078169-001 13-7
Aisys
Removing a cassette
Use the Changing a cassette during a case procedure when
changing a cassette while a case is in progress.
13-8 2078169-001
13 Vaporizer cassettes
Installing a cassette
1. Using the liquid level indicator, check that the cassette is filled to
the appropriate level.
2. If using Aladin2, unlock the cassette handle before installing it in
the active bay.
3. Insert the cassette into the active bay until a click is heard,
ensuring the cassette is in the correct position.
4. Turn the lock on the handle to the horizontal position (Aladin2
cassettes only).
5. The cassette is properly inserted when the agent is identified on
the display. Make sure that the displayed agent matches the
cassette.
Note Store the cassettes in the cassette storage bay when they are not in
use.
2078169-001 13-9
Aisys
Cassette maintenance
Clean and drain the cassettes as recommended.
Cleaning
Draining cassettes
• Make sure that the valve pins do not come in contact with
the system or any other object.
13-10 2078169-001
13 Vaporizer cassettes
2078169-001 13-11
Aisys
AB.60.054
Filling with Easy-Fil system
13-12 2078169-001
13 Vaporizer cassettes
AB.60.040
Figure 13-6 • Aladin2 cassette with Easy-Fil system
WARNING Do not open or press the cassette filling port or the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
• When the liquid level indicator reaches the full mark, slowly
remove the bottle from the filling port.
2078169-001 13-13
Aisys
Note Removing the bottle quickly may cause the agent to splash or
squirt out of the filling port.
AB.60.041
Figure 13-7 • Aladin2 cassette with Quik-Fil system
WARNING Do not open or press the cassette filling port or the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
• When the liquid level indicator reaches the full mark, slowly
remove the bottle from the filling port.
13-14 2078169-001
13 Vaporizer cassettes
Note Removing the bottle quickly may cause the agent to splash or
squirt out of the filling port.
AB.60.050
AB.60.049
2078169-001 13-15
Aisys
AA.43.129
1. Bottle cap
2. O-ring
3. Insert the nozzle into the filling port, and push the bottle firmly
against the spring pressure until it stops.
4. Keep the bottle firmly inserted, and lift the bottle upwards.
5. Watch the liquid level indicator. Do not leave the bottle
unattended while it is attached to the cassette.
6. When the liquid level reaches the full mark, lower the bottle to
the lower stop position.
7. Slowly remove the bottle from the filling port. Removing the
bottle quickly may cause the agent to splash or squirt out of the
filling port.
8. Put the cap back on the agent bottle.
13-16 2078169-001
13 Vaporizer cassettes
AB.60.019
WARNING Do not open or press the cassette filling port of the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
2078169-001 13-17
Aisys
6. Watch the liquid level indicator. When the cassette is full, lower
the bottle to stop the flow of liquid.
7. Turn the lock and fill wheel a half turn counterclockwise to close
the filling port. Wait for the residual anesthetic agent to trickle
back into the bottle.
8. Turn the lock and fill wheel fully counterclockwise, and remove
the keyed filler from the filling port.
9. Remove the bottle adapter from the anesthetic agent bottle. Put
the cap back on the agent bottle.
AB.60.018
WARNING Do not open or press the cassette filling port or the gas
connection valves with fingers or any kind of instrument.
Anesthetic agent liquid or gas may squirt into the air.
13-18 2078169-001
13 Vaporizer cassettes
• When the liquid level indicator reaches the full mark, slowly
remove the bottle from the filling port.
Note Removing the bottle quickly may cause the agent to splash or
squirt out of the filling port.
AB.60.017
2078169-001 13-19
Aisys
2
AA.43.129
3. Insert the nozzle into the filling port, and push the bottle firmly
against the spring pressure until it stops.
4. Keep the bottle firmly inserted, and lift the bottle upwards.
5. Watch the liquid level indicator. Do not leave the bottle
unattended while it is attached to the cassette.
6. When the liquid level reaches the full mark, lower the bottle to
the lower stop position.
7. Slowly remove the bottle from the filling port. Removing the
bottle quickly may cause the agent to splash or squirt out of the
filling port.
8. Put the cap back on the agent bottle.
13-20 2078169-001
Index
A tests 7-15
volume 3-19
Abbreviations 1-11 Alarm history 3-19
Absorbent canister Alarm setup
canister removal 8-7 Auto MV limit 3-20
EZchange canister removal 8-7 CO2 alarms 3-17
filling the canister 8-8 default limits 3-20
Absorbent color change 8-6 leak audio alarms 3-20
Absorber canister MV/TV alarms 3-18
parts 10-6 setting alarm limits 3-18
setup 8-5 setting apnea delay 3-19
when to change the absorbent 8-6 volume apnea 3-17
ACGO Alarms
port relief 11-6 viewing alarm history 3-19
Active AGSS 8-22 Alternate O2 control 3-35
Advanced breathing system Anesthesia system display 2-11
parts 10-5 Apnea delay 3-19
Advanced Breathing System Auto MV Limit 3-20
components 2-5 Automatic agent identification 6-7
AdvancedBreathingSystem Auxiliary Common Gas Outlet 3-41
definition 1-2 Auxiliary O2 flowmeter 8-16
AGSS
active 8-22
connecting active adjustable 8-23
B
connecting active with a flow indicator 8-22 Battery indicator 7-3
parts 10-8 Battery information 11-12
passive 8-21 Bellows assembly parts 10-4
Airway module Breathing circuit module
specifications 11-21 parts 10-3
Airway modules Breathing system
agent identification 6-7 gas scavenging 11-16
calibration 6-8 problems 7-17
connection to a patient 6-5 specifications 11-15
parameter setup 6-6
setting the data source 6-6 C
Alarm
battery indicator 7-3 Calibration
de-escalating 7-3 100% O2 9-7
information 7-2 21% O2 9-7
internal failure 7-4 airway module 6-8
list of 7-5 circuit O2 cell 9-7
pausing audio 7-2 flow and pressure 5-3, 9-7
priorities 7-2 Cardiac bypass
ranges 7-13 manual ventilation 3-29
sustained pressure threshold 7-12 VCV 3-29
2078169-001 I-1
Aisys
I-2 2078169-001
Index
2078169-001 I-3
Aisys
specifications 11-24 U
Super User menus
alarm settings 12-6 USB port 8-11
case defaults 12-10 Using menus 2-14
display settings 12-5 Using quick keys 2-15
gas controls 12-15 Using the ComWheel 2-15
gas usage 12-3
parameter settings 12-6 V
procedures setup 12-16 Vacuum suction regulator 8-15
system configuration 12-5 Vent and gas test 5-8
ventilator settings 12-5 Ventilation
volume apnea 12-10 modes 11-26
Super user mode Ventilation modes
access to 12-2 CPAP+ PSV 11-35
exit 12-2 factory defaults 11-37
Sweep speed 3-15 PCV 11-28
Symbols 1-6, 1-9 PCV-VG 11-29
System PSVPro 11-34
classification 1-14 SIMV PCV 11-31
device standards 1-14, 1-14 SIMV PCV-VG 11-32
information 1-2, 1-14 SIMV VCV 11-30
integral system components 1-15 transition 11-38
not integral system components 1-15 VCV 11-27
operation safety 3-2 Ventilator
overview 2-2 accuracy data 11-41
preparing for use 1-16 changing ventilator mode 3-8
safety 1-16 changing ventilator settings 3-8
symbols 1-6 operating specifications 11-39
turning on 3-3 setup 3-8
System controls and menus 2-1 Ventilator specifications
System pneumatic circuits 11-2 oxygen 11-40
pneumatics 11-39
T pressure 11-39
Test tools volume 11-39
parts 10-12 Ventilator theory
Time and date O2 monitoring 11-25
setup 3-14 Venturi suction regulator 8-16
Timer 3-32 Vital capacity
Touchscreen changing settings 3-30
alarms 2-10 Volume apnea 3-17
digit fields 2-12
measured values 2-9 W
split screen field 2-12 Water buildup 9-8
waveform fields 2-12 Waveform
Trends setup 3-13
setup 3-24 Waveform fields 2-12
Turning system on 3-3
Typeface conventions 1-10
I-4 2078169-001
Warranty
This Product is sold by Datex-Ohmeda under the warranties set forth in the following paragraphs. Such
warranties are extended only with respect to the purchase of this Product directly from Datex-Ohmeda or
Datex-Ohmeda’s Authorized Dealers as new merchandise and are extended to the Buyer thereof, other
than for the purpose of resale.
For a period of twelve (12) months from the date of the original delivery to the Buyer or Buyer’s order, but
in no event for a period of more than two years from the date of original delivery by Datex-Ohmeda to a
Datex-Ohmeda Authorized Dealer, this Product, other than its expendable parts, is warranted against
functional defects in materials and workmanship and to conform to the description of the Product
contained in this User’s Reference manual and accompanying labels and/or inserts, provided that the
same is properly operated under the conditions of normal use, that regular periodic maintenance and
service is performed and that replacements and repairs are made in accordance with the instructions
provided. This same warranty is made for a period of thirty (30) days with respect to expendable parts.
The foregoing warranties shall not apply if the Product has been repaired other than by Datex-Ohmeda,
or altered by anyone other than Datex-Ohmeda, or if the Product has been subject to abuse, misuse,
negligence, or accident.
Datex-Ohmeda’s sole and exclusive obligation and Buyer’s sole and exclusive remedy under the above
warranties is limited to repairing or replacing, free of charge, at Datex-Ohmeda’s option, a Product, which
is telephonically reported to the nearest Datex-Ohmeda Customer Service Center and which, if so
advised by Datex-Ohmeda, is thereafter returned with a statement of the observed deficiency, not later
than seven (7) days after the expiration date of the applicable warranty, to Datex-Ohmeda Customer
Service and Distribution Center during normal business hours, transportation charges prepaid, and which,
upon Datex-Ohmeda’s examination, is found not to conform with above warranties. Datex-Ohmeda shall
not be otherwise liable for any damages including but not limited to incidental damages, consequential
damages, or special damages.
There are no express or implied warranties which extend beyond the warranties hereinabove set forth.
Datex-Ohmeda makes no warranty of merchantability or fitness for a particular purpose with respect to
the product or parts thereof.
Aisys
User's Reference Manual
English
2078169-001
10 14 D