User’s manual
Release 4 January 2006 (Rev. 0)
USER’S MANUAL
Contents
Contents
1 INTRODUCTION 1
1.1 Icon appearing in the manual ................................................................. 1
2 SAFETY ASPECT 2
2.1 Warning.................................................................................................. 3
2.2 Environmental risks and displacement ................................................... 5
2.3 Symbols used ......................................................................................... 6
3 DESCRIPTION 7
3.1 Identification labels ................................................................................ 7
3.2 Description ............................................................................................. 8
3.2.1 DIVA-D composition ................................................................... 9
3.2.2 Monitors..................................................................................... 9
3.2.3 Accessories................................................................................. 9
4 TECHNICAL DATA 10
4.1 DIVA-D AS 25....................................................................................... 10
4.2 CRT Monitor ......................................................................................... 15
4.3 LCD Monitor ......................................................................................... 17
4.4 Environmental conditions..................................................................... 18
4.5 Standards............................................................................................. 18
6 UNIT SWITCHING ON 22
6.1 CCD camera not connected .................................................................. 24
6.2 System fault ......................................................................................... 25
7 PATIENT DATA 26
8 FLUOROSCOPY 28
8.1 Fluoro special function ......................................................................... 30
9 MODALITY SELECTION 31
10 GRAPHY 33
(Rev. 0) i DIVA-D
USER’S MANUAL
Contents
11 POST PROCESSING 35
11.1 Special function .................................................................................... 39
11.2 Multi image........................................................................................... 41
11.3 Image print ........................................................................................... 42
12 DICOM FUNCTIONS 43
12.1 Waiting list ........................................................................................... 44
12.2 Queue view and flush ........................................................................... 45
12.3 Print ..................................................................................................... 47
12.4 Store ..................................................................................................... 49
12.5 CD - ROM ............................................................................................. 51
1 INTRODUCTION
G
NOTE:
The present manual is updated for the product it is sold with in order to
grant an adequate reference to use properly and safely the product.
The manual may not reflect changes to the product not impacting
operating modes or safety.
The purpose of this manual is to give the User instructions for a safe and
efficient use of the DIVA-D.
G
This icon indicates a NOTE: please read thoroughly the items marked
by this picture.
This icon indicates a WARNING message; the items marked by this icon
refer to the safety aspects of the patient and/or of the operator.
(Rev. 0) 1 DIVA-D
USER’S MANUAL
Safety aspect
2 SAFETY ASPECT
Only the authorized personnel can remove the covers and/or have
access to the components under voltage.
DIVA-D 2 (Rev. 0)
USER’S MANUAL
Safety aspect
2.1 Warning
WARNING:
If during the self-test phase the equipment shows a failure message
caused by the electronic circuits or the Ram disks, the images of a given
patient retrieved for post processing may not belong to the requested
patient.
In this case, immediately contact an authorized Villa Sistemi Medicali
engineer and do not make any diagnosis or evaluation on the stored
images.
In addition, as the failure may have occurred during the last working
session, diagnosis carried out after the last exam cannot be
considered valid.
(Rev. 0) 3 DIVA-D
USER’S MANUAL
Safety aspect
During the use of the equipment, the operator must comply with the
local work safety regulation, in particular with the ones referring to the
use of video terminals.
The cathode X-ray tube (CRT) of the monitors, when present, is
manufactured in order to minimize the risk of implosion. However, if
heavily hit it may break thus projecting out pieces of glass, powder and
toxic gas. Avoid hitting the CRT with any object and make sure that it
does not hit any object.
These risks are not present when LCD monitors are used.
DIVA-D 4 (Rev. 0)
USER’S MANUAL
Safety aspect
The DIVA-D contains in some of its parts, materials that at the end of the
units life, must be recycled at an approved location.
Particularly the device contains the following materials and/or
components:
WARNING
Glass vacuum tube may implode with projection of glass pieces, powder
and toxic gas.
(Rev. 0) 5 DIVA-D
USER’S MANUAL
Safety aspect
Symbol Description
∼ A.C.
Dangerous voltage
DIVA-D 6 (Rev. 0)
USER’S MANUAL
Description
3 DESCRIPTION
1
4
2
3
1
Main label (rack)
2
Control keyboard label
3
CRT Monitor label
4
LCD Monitor label
(Rev. 0) 7 DIVA-D
USER’S MANUAL
Description
3.2 Description
The images coming from the image intensifier are acquired through a
CCD camera, then fed to the digitalization stage (A to D Converter) after
that they can be stored in the permanent memory (Ram disk).
From this moment on the images can be processed thanks to the
dedicated software programs.
The system has a large accessible database, which allows for the entry
and retrieval of patient information along with exam description, remark
and notes.
DIVA-D RF 12
DIVA-D AS 25
DIVA-D 8 (Rev. 0)
USER’S MANUAL
Description
3.2.2 Monitors
All DIVA-D models are available with CRT or LCD monitors for operators
and "in room".
The size of these monitors depends on customer request.
3.2.3 Accessories
(Rev. 0) 9 DIVA-D
USER’S MANUAL
Technical data
4 TECHNICAL DATA
G
NOTE:
Technical data, performance and instructions provided in this manual
are referred to the DIVA-D AS 25 model. Therefore, some of these
characteristics are not applicable for DIVA-D RF 12 systems.
4.1 DIVA-D AS 25
Identification
Type DIVA-D AS 25
Model NDR 632 E00 + NDR transf 02
Components
Central unit Rack composed by:
Power supply; Computer; Pcb's; Ram
disk
Control system Flat control keyboard
Acquisition by CCD Camera 1024 x 1024
Interface system Generator and Remote Control Table
interface PCB
Connection cables Prepared interconnection cables
Operator: 1 or 2 on request
Monitor
"In room": 1 or 2 on request
Dimension and weight Central Unit (Rack)
Weight 15 Kg
Height 254 mm
Depth 345 mm
Width 440 mm
Dimension and weight Control keyboard
Weight 6 Kg
Height 50 mm
Depth 235 mm
Width 460 mm
DIVA-D 10 (Rev. 0)
USER’S MANUAL
Technical data
(Rev. 0) 11 DIVA-D
USER’S MANUAL
Technical data
DIVA-D 12 (Rev. 0)
USER’S MANUAL
Technical data
Working configuration
Remote R\ F control table - Apollo
- Mercury 332
R\ F tilting table - Vision
Integration with generator Genius 65/80 HF-A with serial mode
(RS 232) interface
Clinical application
Fluoroscopy Continuous and pulsed HCF
Standard exams and with contrast agent Gastroenterological, bone exams, ERCP,
Urology, Mielography, etc.
Tomography In low dose modality
Vascular exams Thoracic angiography, carotid,
peripheral, phlebography.
Exams with on-line subtraction and
post processing.
Peripheral steps Angiography Optional. Possible application with
Apollo remote control table only.
Working characteristics
User interface Windows representation
Fluoroscopy mode
Storage Manual start on Ram disk
Regulation of the parameter Auto video level regulation.
Auto radiological parameter regulation.
Manual selection of six field video levels
Edge enhancement Two selectable levels
Positive/Negative function (grey scale In line/off line
inversion)
Noise reducer Two selectable level
Road mapping Direct masking
Imaging comparison Automatic or manual transfer of the
image
Last image hold (LIH) With integrated image
Manual regulation of the video level Three selectable levels
Motion detection filter Automatic exclusion of the recursive
filter during the movement of the image
(no blurring effect)
Dynamic recursive filter DRF Automatic noise/blurring optimization
effect offers zero persistence
(Rev. 0) 13 DIVA-D
USER’S MANUAL
Technical data
Radiography mode
Storage On Ram disk
Five standard acquisition modality with Spot – Bone – Angio - Peripheral angio –
different functions. Tomo.
One optional acquisition modality Stepping angio (optional)
Anatomic program for each acquisition 20
modality
Low dose acquisition For pediatric exams
Maximum opacification function Track due to the storing of the contrast
(maxop) liquid flow with subtraction acquisition
mode
Acquisition time selection Auto-self in according with the selected
program.
Manual in real time.
Positive/Negative (grey scale inversion) In line/off line
Last image hold (LIH) Each exposure
Post processing mode
Electronic collimator With scrolling
Edge enhancement Nine regulation levels
Windowing Auto/manual
Road map contrast Nine regulation levels
Positive/negative (grey scale inversion) Default or manual
Image visualization Full screen or multi format size
Angle measurement Displayed data
Length measurement Displayed data and ratio
Editing on the image Available in full screen
Zoom Zoom x 4 with scrolling
Exams review With acquisition speed; image stand;
single reply
Visualization with subtraction Possible with standard acquisition
exams
Exams with subtraction review With Remasking and Maxop functions
Optimization of the subtraction Pixel shift function
Image printing Full screen
DIVA-D 14 (Rev. 0)
USER’S MANUAL
Technical data
G
NOTE:
The characteristics of the monitors are the same for both DIVA-D's
model.
Technical data
Type Monitor CRT B&W
Manufacturer Nical
Model Refer to optional
Dimensions and weight Monitor 17” Monitor 21”
Weight 17,5 Kg 21,8 Kg
Height 335 (359) mm 394 (418) mm
Depth 350 mm 380 mm
Width 405 mm 489 mm
Screen dimensions 17" (414 mm) 21" (511 mm)
Electrical characteristics
Standard power supply 230 V ac ±15%
Frequency 50-60 Hz
Current absorbed 350 mA
Power absorbed 80 W
Scanning
Type Interlaced
Line number 1049
"Refresh" frequency 60 Hz
Standards EIA
Video input signal
Type Video input connected ac
Connection type BNC
Impedance 75 Ω
Input video signal 0,5 to 1,5 V pp (negative
synchronism)
(Rev. 0) 15 DIVA-D
USER’S MANUAL
Technical data
Display characteristics
Deflection angle 90°
Aspect ratio 4:3
Band width 40 MHz ±3 dB
Raster 0.25 mm
Phosphor White P45
Geometric error <1.5 %
Vertical/horizontal linear error <1.5 %
Maximum brightness 850 cd/m2
Light transmission 46%
Dynamic focus Yes
Stabilized EHT Yes
Optional
Contrast and brightness control with digital Available on request on all models
push-button on front panel and rear remote
control connection
Self luminosity sensor on front panel
Contrast and brightness control only with Available on request on all models
rear remote control connection
DIVA-D 16 (Rev. 0)
USER’S MANUAL
Technical data
G
NOTE:
The characteristics of the monitors are the same for both DIVA-D's
model.
Technical data
Type Monitor LCD B&W
Manufacturer Nical
Model Refer to optionals
Dimensions and weight Monitor 18” Monitor 20”
Weight 5,5 Kg 6,0 Kg
Height 340 mm 355 mm
Depth 60 mm 60 mm
Width 430 mm 457 mm
Screen dimensions 18" 20"
Electrical characteristics
Standard power supply 230 V ac ±15%
Frequency 50-60 Hz
Current absorbed 250 mA
Power absorbed 50 W
Display characteristic
Visibility angle 180°
Contrast 600:1
Resolution 1280x1024
Gray scale 768 levels
Max luminance 700 Cd/m2
Calibration automatic
Optional
Contrast and brightness control with digital Available on request on all models
push-button on front panel and rear remote
control connection
Contrast and brightness control only with Available on request on all models
rear remote control connection
(Rev. 0) 17 DIVA-D
USER’S MANUAL
Technical data
4.5 Standards
DIVA-D system with accessories and optional complying with the relevant
standards:
NICAL S.p.A.
Via Soffredini, 43
20126 - Milano - Italia
Tel.: +39-02-25.711.10
Fax: +39-02-25.722.07
DIVA-D 18 (Rev. 0)
USER’S MANUAL
Keyboard and functions
(Rev. 0) 19 DIVA-D
USER’S MANUAL
Keyboard and functions
DIVA-D 20 (Rev. 0)
USER’S MANUAL
Keyboard and functions
(Rev. 0) 21 DIVA-D
USER’S MANUAL
Unit switching ON
6 UNIT SWITCHING ON
Any time that DIVA-D is switched on, it performs an auto test to verify if
the CCD camera is connected and correctly working and if the electronics
of the main rack, including RAM disks are normally operating.
SYSTEM READY
DIVA-D 22 (Rev. 0)
USER’S MANUAL
Unit switching ON
SYSTEM READY
DICOM READY
(Rev. 0) 23 DIVA-D
USER’S MANUAL
Unit switching ON
During auto test, the unit could recognize a fault in the CCD camera
connection.
In this condition the window shows:
DIVA-D 24 (Rev. 0)
USER’S MANUAL
Unit switching ON
(Rev. 0) 25 DIVA-D
USER’S MANUAL
Patient data
7 PATIENT DATA
Before starting with a new exam is necessary to insert into DIVA-D the
data of the new patient.
If this is not done, as soon the exposure images acquisition is required,
the message:
is displayed.
It is necessary to press NEW PATNT key to open the patient data
entering window.
G
NOTE:
If fluoroscopy mode is activated, without storing any LIH or any fluoro
sequence, the images visualization is present also if no patient data is
inserted. This because a patient could perform a fluoroscopic
investigation only. In all other cases, because all radiographic images are
automatically stored on RAM disks, the patient data must be entered into
the data base.
When the key NEW PATNT is activated the new patient window is
displayed:
DIVA-D 26 (Rev. 0)
USER’S MANUAL
Patient data
G
NOTE:
If the birth date is not entered, the patient age in the data base will be
recorded as 0.
If sex is not inserted, O (other) will be present in the data base).
Confirming by ENTER, the data are stored into the data base and the
modality window is displayed to permit the acquisition mode selection.
If "last patient" is selected, the PATIENT DATA window appears with the
data relevant to the last patient stored.
This selection is used when during exam on a patient, the unit has been
switched off.
This feature is very useful to avoid 2 files for the same patient.
(Rev. 0) 27 DIVA-D
USER’S MANUAL
Fluoroscopy
8 FLUOROSCOPY
If DIVA-D is equipped with 2 monitors in the control desk area, one of
them is reserved to fluoroscopy images only, otherwise the fluoro window
is present any time the fluoro command is activated.
DIVA-D 28 (Rev. 0)
USER’S MANUAL
Fluoroscopy
• Store
In normal use the condition is OFF. To activate the fluoro storing
modality it is necessary to press STORE. The relevant led turns on
and the images number stored are indicated at the bottom of the
display.
When fluoro acquisition is interrupted also the storage is stopped.
At new fluoro acquisition it will be necessary to press the STORE key
again to perform a new storage.
In LIH condition, pressing the STORE key, the LIH will be stored into
the hard disk.
• Edge
By default the value is OFF. It means the edge level used is the
default one. Using E key, the visualization is ON and the edge level
selected is the one configured at the installation.
• Noise
By default the value is OFF. It means the noise suppression level
used is the default one. Using N key, the visualization is ON and the
noise level selected is the one configured at the installation.
• Road map
Road mapping function is available on DIVA-D AS25 only.
The function is activated by using the ROAD MAP key. Pressing the
key, the relevant led turns on and the fluoro images will be in
subtractions mode.
Releasing fluoro, on the display the LIH will be in sub mode and
following fluoro images will have the mask overlapped.
To reset the road mapping mode and the mask, press the key with
fluoro off.
The Led will be turned off and next fluoro will be performed in
standard mode.
• Kv – mA – Dose
In these fields the actual kV and mA are displayed.
Dose field is not yet activated.
• Run – Frame
In these fields the fluoro run number stored for the present patient
and the images contained in the run are displayed.
The dose level can be modified by using the keys kV+ and kV-.
The effect of this action is the image contrast level becoming higher or
lower depending by the key activated. 3 decreasing steps and 3
increasing steps are available starting from the standard level.
The radiological parameters are automatically adjusted by the system,
during fluoro acquisition to optimize the image video level.
This adjustment is calculated based on ROI selection.
(Rev. 0) 29 DIVA-D
USER’S MANUAL
Fluoroscopy
In fluoro off line mode, by pressing the SPEC FUNCT key, the relevant
window is displayed:
• Store XX Fr/s
The storage rate in fluoro can be selected to preserve disk space.
This selection modifies the storage speed only without modifying the
visualization rate.
• Standard / Pulsed
This selection can also be done in on line mode by using the pulsed
key as described above.
• Ref image
This selection appears when DIVA-D is equipped with 2 monitors in
control desk area only.
When this selection is active the graphy monitor is used to display
the LIH while on fluoro monitor the live images are displayed.
• Pattern 1 / 2
Allows to display 2 different patterns in order to adjust correctly the
monitor brightness and contrast.
DIVA-D 30 (Rev. 0)
USER’S MANUAL
Modality selection
9 MODALITY SELECTION
When patient data are inserted, the modality and relevant anatomic
programs must be chosen.
The access to this function is automatic when exiting from patient data
window with ENTER key or any time the MODE key is activated.
This because when a new patient is inserted it is necessary to select the
modality for the first acquisition, and to change this modality if a patient
requires different exams type.
The acquisition modality can be:
• SPOT
• BONE
• ANGIO
• PERIF
• STEP
• TOMO.
SPOT, BONE and TOMO are allowed for both DIVA-D models RF and AS,
ANGIO and PERIF on AS only, STEP could be present on a AS model only
as supplementary option.
The step modality is allowed only if the system is based on APOLLO table
only.
The modality window looks as follows:
It is divided in 3 areas:
• modality selection
• anatomic selection
• parameters modification.
By default the left area is active when modality is selected. To inform
about the area selected, the background is light gray.
(Rev. 0) 31 DIVA-D
USER’S MANUAL
Modality selection
G
NOTE:
The not available modality (depending by the DIVA-D license level) will be
displayed in gray and not reachable by selection.
DIVA-D 32 (Rev. 0)
USER’S MANUAL
Graphy
10 GRAPHY
When the modality selection is done and confirmed, DIVA-D is ready for
graphy acquisition.
The window displayed is:
The area “Run” is reserved to the number of exams stored for the present
patient.
During acquisition the areas “Run”, “Frame”, “mSec” and “kV” are
updated.
(Rev. 0) 33 DIVA-D
USER’S MANUAL
Graphy
DIVA-D 34 (Rev. 0)
USER’S MANUAL
Post processing
11 POST PROCESSING
To review the exams contained for the current “patient”, 2 options are
possible:
• by MENU key
• by STOP key.
(Rev. 0) 35 DIVA-D
USER’S MANUAL
Post processing
The window reporting the data relevant to the different runs of the
patient is displayed.
Select the desired run using UP and DOWN keys.
Press ENTER to confirm.
DIVA-D 36 (Rev. 0)
USER’S MANUAL
Post processing
All of them and other settings can be modified using the dedicated keys:
(Rev. 0) 37 DIVA-D
USER’S MANUAL
Post processing
DIVA-D 38 (Rev. 0)
USER’S MANUAL
Post processing
• Overwrite
Allows to report comments on the image. To position the cursor on
the required place use UP, DOWN, LEFT and RIGHT arrow keys.
Press ENTER key to confirm when the text insertion is completed.
Changing the image or using STOP key, the text inserted is cancelled.
• Angle measure
Allows to take a measure of the angle of a part of image.
Using UP and DOWN keys select the approx level from the table top
on which the part to be measured is located between the 3 levels
proposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT
and RIGHT arrow keys.
Press ENTER key to confirm the starting measuring point. Move the
cursor on the angle vertex and confirm whit ENTER. Move the cursor
(Rev. 0) 39 DIVA-D
USER’S MANUAL
Post processing
WARNING:
Because of different reasons ( II deformation, error in the level organ
position and others) the angle value obtained are only indicative. Do not
use this value for diagnostic purpose.
• Length measure
Allows to take a measure of a part of the image.
Using UP and DOWN keys select the approx level from the table top
on which the part to be measured is located between the 3 levels
proposed 0, 10 or 20 cm. Press ENTER to confirm.
To position the cursor on the required place use UP, DOWN, LEFT
and RIGHT arrow keys.
Press ENTER key to confirm the starting measuring point. Move the
cursor on the final measuring point and confirm with ENTER, the
length value is displayed.
Changing the image or using STOP key, the text inserted is cancelled.
More than one measure can be done on the same image. To do that
repeat the above described steps.
Ratio between 2 measure can be done selecting the relevant window.
WARNING:
Because of different reasons (II deformation, error in the level organ
position and others) the measures obtained are only indicative. Do not
use this value for diagnostic purpose.
• Test pattern
Allow to display a test pattern to adjust correctly the monitor
brightness and contrast.
Starting with any modality the test pattern is reset from monitor.
DIVA-D 40 (Rev. 0)
USER’S MANUAL
Post processing
When the patient menu window is displayed, multi image function can be
used.
It is a short and easy way how to select the run to be review.
Using the key, on monitor the first image of any run present into
the selected patient is displayed.
By means of the UP, DOWN, LEFT and RIGHT arrow keys, select the
desired run and press ENTER to confirm.
The run selected will be displayed in full screen mode.
(Rev. 0) 41 DIVA-D
USER’S MANUAL
Post processing
G
NOTE:
Do not change the image displayed until the laser printer acquisition is
completed.
DIVA-D 42 (Rev. 0)
USER’S MANUAL
DICOM functions
12 DICOM FUNCTIONS
If the DIVA-D is equipped with DICOM module, the DICOM key is active.
By means of this key the following window is displayed:
(Rev. 0) 43 DIVA-D
USER’S MANUAL
DICOM functions
is displayed.
On this window the patients present on the worklist are displayed.
By means of the UP and DOWN arrow key select the desired one, press
ENTER to confirm.
The patient data window will be displayed (chapter 7) reporting all data
inserted in HIS/RIS.
By this function any manual patient data entering is avoided.
DIVA-D 44 (Rev. 0)
USER’S MANUAL
DICOM functions
will be displayed.
If no queue error or messages are present, the last used function will be
proposed, anyway the queue view will be reachable by means of ENTER
key directly.
In any case UP and DOWN arrow keys can be used to reach the function
desired or queue view.
If errors or messages are present in the queue, the relevant field is with
white background; pressing ENTER key the window
will appears.
(Rev. 0) 45 DIVA-D
USER’S MANUAL
DICOM functions
DIVA-D 46 (Rev. 0)
USER’S MANUAL
DICOM functions
12.3 Print
When DICOM window is selected, the last function used in DICOM mode
is proposed.
By means of UP and DOWN arrow keys select the required function and
confirm with ENTER key.
If print function is selected the printer window selection is showed.
With UP and DOWN arrow keys, select the required printer and confirm.
With LEFT and RIGHT arrow key select the film subdivision and confirm.
The DICOM print window
is displayed.
Select the image required and modify, if necessary the characteristic as
in standard post processing mode.
In the lower section of the window, the frame number and position are
displayed.
(Rev. 0) 47 DIVA-D
USER’S MANUAL
DICOM functions
To add the image, the key must be pressed 2 times. After the first
time the window
is displayed.
Pressing once more the addition is confirmed and the image stored into
DICOM print temporary archive.
The previous window will be showed allowing a new image addition.
When the above window is present, 3 other functions are available (use
UP and DOWN arrow keys to select and press ENTER to confirm):
• Delete & exit
To reset the temporary DICOM print archive and exit from DICOM
functions.
• Study print
To print the image stored.
DIVA-D 48 (Rev. 0)
USER’S MANUAL
DICOM functions
12.4 Store
By means of UP and DOWN arrow keys select the store function and
confirm with ENTER key.
When store function is selected the window is showed:
On the lower part of the window the following information are listed:
• Total Mbytes
Mbytes used by the images stored in the present DICOM store study.
This is referred to the temporary DICOM archive.
• Total Patient
This value can be 0 or 1.
As matter of fact a DICOM study to be stored cannot contain more
than 1 patient.
• Total runs
Is the number of runs stored to be transferred in the network.
• Total frames
Is the number of single image to be transferred in the network.
• Total sub-runs
Is the number of sub-runs stored to be transferred in the network.
(Rev. 0) 49 DIVA-D
USER’S MANUAL
DICOM functions
To add the image, the key must be pressed two times. After the
first time the window
• Study send
To send the study stored.
DIVA-D 50 (Rev. 0)
USER’S MANUAL
DICOM functions
12.5 CD - ROM
By means of UP and DOWN arrow keys select the CD-ROM function and
confirm with ENTER key.
When CD-ROM function is selected the media type window is showed:
(Rev. 0) 51 DIVA-D
USER’S MANUAL
DICOM functions
On the lower part of the window the following information are listed:
• CD free space
The disk percentage space still available.
• Total Patient
This value can be 0 or 1.
As matter of fact a DICOM study to be stored cannot contain more
than 1 patient.
• Total runs
Is the number of runs stored to be transferred in the network.
• Total frames
Is the number of single image to be transferred in the network.
• Total sub-runs
Is the number of sub-runs stored to be transferred in the network.
DIVA-D 52 (Rev. 0)
USER’S MANUAL
DICOM functions
To store into temporary DICOM archive, select the patient, the run or the
image required and press ENTER. The window:
is displayed.
By means of UP and DOWN arrow keys select if the memory addition will
be of the showed frame only, of the run or of the total patient.
Press ENTER to confirm.
Before to select "burn disk", put a new CD-ROM into driver and press
key.
The dicomization procedure will start and at the end of this operation the
CD-ROM burning start.
This sequence is displayed on the window with a waiting notice.
G
NOTE:
If no CD-ROM or not valid one is present into the driver, an error will be
present in the queue.
DIVA-D will try again to access the CD after 30 sec.
Insert a valid CD-ROM into the driver before a new writing attempt.
(Rev. 0) 53 DIVA-D
USER’S MANUAL
Maintenance and cleaning
WARNING:
• Before proceeding with the cleaning of the equipment, disconnect it
from the mains.
• Make sure that water and any other liquids are not spilled inside the
equipment as this may cause corrosion and short circuits.
• Clean the painted surfaces by using a moist cloth and neutral
detergent, taking care to wipe them by using a dry cloth. Do not use
solvents (alcohol, petrol, and trichloroethilene) or corrosive or
abrasive products.
Frequency Description
DIVA-D 54 (Rev. 0)
Cod. 6995001003_Rev.0 0123