AS3000 Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 180

S e r v i ce M a n u a l

Datascope
AS ™

Anesthesia Delivery System

0070-01-0683-02_revB_AS3000 srv color.indd 1 3/10/10 5:13:28 PM


S e r v i ce M a n u a l

Datascope
AS ™

Anesthesia Delivery System


AS3000™ is a U.S. trademark of Mindray DS USA, Inc.

Krytox® is a U.S. registered trademark of E. I. du Pont de Nemours and Company

Selectatec® is a U.S. trademark of Datex Ohmeda

Glyptal® is a U.S. registered trademark of Glyptal, Inc.

Dräger Vapor® and plug-in system S-2000 are registered trademarks of Dräger Medical.

Copyright © Mindray DS USA, Inc., 2008. All rights reserved. Contents of this publication may not be reproduced in any
form without permission of Mindray DS USA, Inc.

AS3000™ Service Manual 0070-10-0683


Table of Contents

Foreword .......................................................................................................................................................vii
Warnings, Cautions and Notes ........................................................................................................................vii
Warnings ......................................................................................................................................................vii
Cautions .......................................................................................................................................................viii
Notes ............................................................................................................................................................ ix
Theory of Operation .....................................................................................................1 - 1
Introduction ................................................................................................................................................ 1 - 1
Microprocessor-controlled Ventilator.............................................................................................................. 1 - 2
Components............................................................................................................................................... 1 - 3
Front View .......................................................................................................................................... 1 - 3
Side View........................................................................................................................................... 1 - 5
Rear View........................................................................................................................................... 1 - 7
Other components (not identified in graphics) ......................................................................................... 1 - 8
Breathing System ................................................................................................................................. 1 - 9
The Ventilator Unit ............................................................................................................................... 1 - 9
Adjustable Alarms ............................................................................................................................. 1 - 10
Compliance Compensation................................................................................................................. 1 - 10
Electrical Supply ....................................................................................................................................... 1 - 11
Electrical components......................................................................................................................... 1 - 11
Ventilator Control and Drive ....................................................................................................................... 1 - 12
BDU (Basic Digital Unit)...................................................................................................................... 1 - 12
Amplifier Board ........................................................................................................................ 1 - 14
Drive Gas Pressure Sensor Board ................................................................................................ 1 - 19
PAW Pressure Sensor Board ....................................................................................................... 1 - 20
Breathing System Heater ............................................................................................................ 1 - 21
Sensor Board............................................................................................................................ 1 - 23
Power Management........................................................................................................................... 1 - 26
Battery ............................................................................................................................................. 1 - 28
Power Supply.................................................................................................................................... 1 - 28
Anesthesia System Components .................................................................................................................. 1 - 29
Auxiliary Outlets................................................................................................................................ 1 - 29
Absorber Heater Wire Board .............................................................................................................. 1 - 29
Work Light Board .............................................................................................................................. 1 - 29
Ventilator UI ............................................................................................................................................. 1 - 31
Keyboard Board................................................................................................................................ 1 - 31
Display............................................................................................................................................. 1 - 34
Communication Interface / RS232 Isolate Board ................................................................................... 1 - 35
Fuses ............................................................................................................................................... 1 - 35
Ventilator Pneumatic - O2 Drive Gas ........................................................................................................... 1 - 36
Ventilator pneumatic drive .................................................................................................................. 1 - 36
Drive Pressure-High pressure regulator ................................................................................................. 1 - 36
Gas Box Assembly............................................................................................................................. 1 - 36
Tube color coding.............................................................................................................................. 1 - 36
The Breathing System ................................................................................................................................ 1 - 37
CMV mode, inspiration ...................................................................................................................... 1 - 37
CMV mode, expiration....................................................................................................................... 1 - 37
Manual mode, inspiration................................................................................................................... 1 - 38
Manual mode, expiration ................................................................................................................... 1 - 38
Pneumatic PEEP ................................................................................................................................. 1 - 39
Ventilator in Standby.......................................................................................................................... 1 - 39
Breathing System Components ............................................................................................................ 1 - 39
Ventilation Bellows System.......................................................................................................... 1 - 39

AS3000™ Service Manual 0070-10-0683 i


Table of Contents

Manual Breathing Bag ............................................................................................................... 1 - 39


CO2 Absorber .......................................................................................................................... 1 - 39
Inspiratory and Expiratory Valves ................................................................................................ 1 - 39
APL (Airway Pressure Limiting) valve ............................................................................................ 1 - 39
Installation Guide .........................................................................................................2 - 1
Delivery of The New Anesthesia Machine ...................................................................................................... 2 - 1
Assembly ................................................................................................................................................... 2 - 2
Unpacking.......................................................................................................................................... 2 - 2
Breathing System and Breathing System Accessories ................................................................................ 2 - 2
Attaching the Mounting Arms........................................................................................................ 2 - 3
Breathing System Connections ...................................................................................................... 2 - 3
Anesthesia System Connections..................................................................................................... 2 - 3
Tank Wrench and Pre-operation Checklist............................................................................................... 2 - 3
Patient Suction Regulator and Arm ......................................................................................................... 2 - 4
Suction Canister Bracket ....................................................................................................................... 2 - 4
Utility Tray, Monitor Mounting Arm with Utility Hook(s)............................................................................. 2 - 4
Vaporizers.......................................................................................................................................... 2 - 4
High Pressure Hoses ............................................................................................................................ 2 - 4
Emergency Cylinder(s) ......................................................................................................................... 2 - 5
Breathing Circuit, CO2 Absorbent, and Liquid Vaporizer Agent ................................................................ 2 - 5
Monitoring Products - Mounting and Electrical Connection ........................................................................ 2 - 5
Agent Monitor Waste Gas Scavenging .................................................................................................. 2 - 6
Installation Checkout Procedure .................................................................................................................... 2 - 7
Repair Information .......................................................................................................3 - 1
Introduction ................................................................................................................................................ 3 - 1
Warnings and Cautions............................................................................................................................... 3 - 2
Warnings ........................................................................................................................................... 3 - 2
Cautions............................................................................................................................................. 3 - 2
Troubleshooting Guidelines .......................................................................................................................... 3 - 2
Special Tools Required ................................................................................................................................ 3 - 3
Troubleshooting Chart ................................................................................................................................. 3 - 4
Common Symptoms and Corrective Actions for Field Service Technicians ................................................... 3 - 4
Leak Troubleshooting................................................................................................................................. 3 - 11
Test Pneumatics ........................................................................................................................................ 3 - 12
Leak Test - Manual Ventilation Test ....................................................................................................... 3 - 12
Safety Valve Test ............................................................................................................................... 3 - 13
Leak Test - Automatic Ventilation Test.................................................................................................... 3 - 14
Compliance Test................................................................................................................................ 3 - 15
Pneumatic Hose and Wiring Diagrams ........................................................................................................ 3 - 16
Pneumatic Hose Labeling.................................................................................................................... 3 - 16
Sampling Pipeline Module Interface Labeling ........................................................................................ 3 - 17
Electrical Cable Labeling .................................................................................................................... 3 - 18
Replacement Parts and Accessories ...............................................................................4 - 1
Introduction ................................................................................................................................................ 4 - 1
Available Replacement Parts and Sub-Assemblies ........................................................................................... 4 - 1
Exchange Program .............................................................................................................................. 4 - 1
Replacement Parts Pricing Information .................................................................................................... 4 - 1
Ordering Information ........................................................................................................................... 4 - 2
Isometric Drawings...................................................................................................................................... 4 - 3
Chassis .............................................................................................................................................. 4 - 3
Chassis Parts List ......................................................................................................................... 4 - 5
Breathing System ................................................................................................................................. 4 - 6

ii 0070-10-0683 AS3000™ Service Manual


Table of Contents

Breathing System Parts List.......................................................................................................... 4 - 11


Electric Box....................................................................................................................................... 4 - 12
Electric Box Parts List.................................................................................................................. 4 - 12
Gas Circuit Box................................................................................................................................. 4 - 13
Gas Circuit Box Parts List............................................................................................................ 4 - 13
User Interface.................................................................................................................................... 4 - 14
User Interface Parts List............................................................................................................... 4 - 15
Flowmeter......................................................................................................................................... 4 - 16
Flowmeter Parts List.................................................................................................................... 4 - 16
Calibration ...................................................................................................................5 - 1
Introduction ................................................................................................................................................ 5 - 1
Calibration Warnings, Precautions, and Notes ............................................................................................... 5 - 2
Warnings ........................................................................................................................................... 5 - 2
Cautions............................................................................................................................................. 5 - 2
Notes................................................................................................................................................. 5 - 3
General Guidelines ..................................................................................................................................... 5 - 4
Test Equipment and Special Tools Required.................................................................................................... 5 - 4
Calibration Procedures ................................................................................................................................ 5 - 4
Oxygen Sensor Calibration .................................................................................................................. 5 - 4
Proportional Valve Regulator Calibration ................................................................................................ 5 - 5
Flow Sensor Calibration ....................................................................................................................... 5 - 7
Flow Valve Calibration......................................................................................................................... 5 - 9
Paw Sensor....................................................................................................................................... 5 - 10
PEEP Valve ....................................................................................................................................... 5 - 11
Flow Meter ....................................................................................................................................... 5 - 12
Leakage detection.............................................................................................................................. 5 - 13
Startup leakage detection........................................................................................................... 5 - 13
APL Valve leakage detection....................................................................................................... 5 - 13
Safety Valve leakage detection ................................................................................................... 5 - 13
Breathing System leakage detection............................................................................................. 5 - 13
Compliance detection ........................................................................................................................ 5 - 13
Periodic Maintenance....................................................................................................6 - 1
Maintenance Schedule ................................................................................................................................ 6 - 1
Periodic Maintenance Consumable Parts Kits.................................................................................................. 6 - 1
Periodic Maintenance Schedule of Service Activities........................................................................................ 6 - 2
Visual Inspection Checklist ........................................................................................................................... 6 - 2
Replacement of Consumable Parts................................................................................................................. 6 - 3
Check Valve Cleaning ................................................................................................................................. 6 - 4
Tools and Materials ............................................................................................................................. 6 - 4
Cleaning Procedure ............................................................................................................................. 6 - 5
Battery Maintenance and Replacement ........................................................................................................ 6 - 13
Battery Maintenance .......................................................................................................................... 6 - 13
Battery Replacement .......................................................................................................................... 6 - 13
Functional Tests ........................................................................................................................................ 6 - 13
Test Equipment and Special Tools Required........................................................................................... 6 - 13
Pressure Regulator Checks .................................................................................................................. 6 - 14
Proportional Valve Regulator ...................................................................................................... 6 - 14
Gas Delivery System Tests................................................................................................................... 6 - 14
O2 Flush Verification ................................................................................................................. 6 - 14
O2:N2O Ratio System ............................................................................................................... 6 - 15
Start-Up Tests .................................................................................................................................... 6 - 15
System Self Test......................................................................................................................... 6 - 15

AS3000™ Service Manual 0070-10-0683 iii


Table of Contents

Leak/Safety Valve Test............................................................................................................... 6 - 15


Leak Test .................................................................................................................................. 6 - 17
Compliance Test ....................................................................................................................... 6 - 20
Manual Leak Test ...................................................................................................................... 6 - 21
Oxygen Sensor Calibration ........................................................................................................ 6 - 22
Pneumatic Leak Tests .......................................................................................................................... 6 - 24
N2O Cylinder Leak Test ............................................................................................................. 6 - 24
O2 Cylinder Leak Test................................................................................................................ 6 - 24
AIR Cylinder Leak Test ............................................................................................................... 6 - 24
Line Supply Check Valves Test .................................................................................................... 6 - 24
N2O Line Pressure Leak Test ....................................................................................................... 6 - 25
O2 Line Pressure Leak Test.......................................................................................................... 6 - 25
AIR Line Pressure Leak Test ......................................................................................................... 6 - 25
Cylinder Supply Check Valves Test .............................................................................................. 6 - 25
Breathing System Checks.................................................................................................................... 6 - 26
Waste Gas Scavenger Test (if available) ...................................................................................... 6 - 26
Internal Gas Connections Test ..................................................................................................... 6 - 26
Drive Gas Pressure Loss Alarm, N2O Cutoff Test ........................................................................... 6 - 26
Performance Verification .................................................................................................................... 6 - 27
Standby Mode Ventilation Test.................................................................................................... 6 - 27
Manual Mode Ventilation Test .................................................................................................... 6 - 27
APNEA Alarm Test .................................................................................................................... 6 - 27
Alarm MUTE Test....................................................................................................................... 6 - 27
CMV Adult Ventilation Mode Test................................................................................................ 6 - 28
CMV Child Ventilation Mode Test................................................................................................ 6 - 29
Airway Disconnect Alarm Test..................................................................................................... 6 - 29
PCV Adult Ventilation Mode Test ................................................................................................. 6 - 30
Pressure Support (PS) Ventilation Mode Test.................................................................................. 6 - 31
Alarms and Failsafe Functions ............................................................................................................. 6 - 32
Set Up ..................................................................................................................................... 6 - 32
Low FiO2 Alarm Test.................................................................................................................. 6 - 32
High FiO2 Alarm Test ................................................................................................................ 6 - 33
Peak Pressure Alarms Test .......................................................................................................... 6 - 33
Minute Volume Alarm Test .......................................................................................................... 6 - 34
Miscellaneous Tests............................................................................................................................ 6 - 34
Test the Line Voltage Alarm ........................................................................................................ 6 - 34
Wheel Brakes Test..................................................................................................................... 6 - 34
Work Light Test ......................................................................................................................... 6 - 34
Auxiliary Flowmeter ................................................................................................................... 6 - 35
Patient Suction Regulator (if available) ......................................................................................... 6 - 35
Vaporizers........................................................................................................................................ 6 - 35
Vaporizer Interlock Test.............................................................................................................. 6 - 35
Vaporizer Accuracy Test ............................................................................................................ 6 - 36
Vaporizer Leak Test ................................................................................................................... 6 - 36
Dräger Vapor 2000 Operating Instructions ARRB-F001 ................................................................. 6 - 37
Electrical Tests ................................................................................................................................... 6 - 38
Convenience AC Outlets Test...................................................................................................... 6 - 38
Electrical Safety Inspection Test ................................................................................................... 6 - 38
AS3000 Installation Checklist ............................................................................................................. 6 - 38
Cleaning.................................................................................................................................................. 6 - 39
Cleaning and Disinfecting the AS3000 ................................................................................................ 6 - 39
Cleaning and Sterilizing the Breathing System and Components.............................................................. 6 - 39

iv 0070-10-0683 AS3000™ Service Manual


Table of Contents

Preoperative Checklist ............................................................................................................................... 6 - 39


Phone Numbers and How To Get Assistance................................................................................................ 6 - 47
Warranty .....................................................................................................................7 - 1
Warranty Statements ................................................................................................................................... 7 - 1
Disclaimers ................................................................................................................................................ 7 - 2
Product Improvements .......................................................................................................................... 7 - 2
Manufacturer’s Responsibility ....................................................................................................................... 7 - 2

AS3000™ Service Manual 0070-10-0683 v


Table of Contents

This page intentionally left blank.

vi 0070-10-0683 AS3000™ Service Manual


Foreword Introduction

Foreword
This Service Manual is intended as a guide for technically qualified personnel performing
repair and calibration procedures.

Warnings, Cautions and Notes


Please read and adhere to all warnings, cautions and notes listed here and in the
appropriate areas throughout this manual.

A WARNING is provided to alert the user to potential serious outcomes (death, injury, or
serious adverse events) to the patient or the user.

A CAUTION is provided to alert the user to use special care necessary for the safe and
effective use of the device. They may include actions to be taken to avoid effects on patients
or users that may not be potentially life threatening or result in serious injury, but about which
the user should be aware. Cautions are also provided to alert the user to adverse effects on
this device of use or misuse and the care necessary to avoid such effects.

A NOTE is provided when additional general information is applicable.

Warnings
WARNING: Whenever using anesthetic gases, nitrous oxide, oxygen, or
any hospital gas always follow the appropriate agent
evacuation/collection procedures. Use the hospital gas
evacuation system.

WARNING: For continued protection against fire hazard, replace all


fuses with the specified type and rating.

WARNING: In order to prevent an electric shock, the machine


(protection class I) may only be connected to a correctly
grounded mains connection (socket outlet with grounding
contact).

WARNING: Remove all accessory equipment from the shelf before


moving the anesthesia machine over bumps or on any
inclined surface. Heavy top loading can cause the machine
to tip over causing injury.

WARNING: Possible explosion hazard. Do not operate machine near


flammable anesthetic agents or other flammable
substances. Do not use flammable anesthetic agents (i.e.,
ether or cyclopropane.)

WARNING: The use of anti-static or electrically conductive respiration


tubes, when utilizing high frequency electric surgery
equipment, may cause burns and is therefore not
recommended in any application of this machine.

WARNING: Possible electric shock hazard. The machine may only be


opened by authorized service personnel.

AS3000™ Service Manual 0070-10-0683 vii


Introduction Cautions

WARNING: Compressed gasses are considered Dangerous Goods/


Hazardous Materials per I.A.T.A. and D.O.T. regulations. It is
a violation of federal and international law to offer any
package or over pack of dangerous goods for
transportation without the package being appropriately
identified, packed, marked, classified, labeled and
documented according to D.O.T. and I.A.T.A. regulations.
Please refer to the applicable I.A.T.A. Dangerous Goods
Regulations and /or the Code of Federal Regulations 49
(Transportation, Parts 171-180) for further information.

WARNING: Avoid exposure to respiratory gases by always directing


the fresh gas flow from the fresh gas outlet to the waste gas
scavenger.

Cautions
CAUTION: This device uses high pressure compressed gas. When
attaching or disconnecting backup gas cylinders, always
turn the cylinder valves slowly. Use the AS3000 flow meters
to bleed down the pressure, watching the cylinder gauge
indicate the depleting cylinder pressure, before
disconnecting the cylinder from the yoke. Always open and
close cylinder valves fully.

CAUTION: This device operates using compressed gas at high


pressures from the hospital central supply. When connecting
gas supply lines attach the hose connection to the machine
before connecting the quick disconnect fitting to the hospital
source. Disconnect the supply hose from the hospital source
connection prior to disconnecting it from the AS3000 gas
connection fittings.

CAUTION: Refer to the “Periodic Maintenance Schedule of Service


Activities” on page 6-2, in the Periodic Maintenance section
for assistance when performing scheduled periodic
maintenance.

CAUTION: Do not leave gas cylinder valves open if the pipeline supply
is in use and the system master switch is turned to 'ON'. If
used simultaneously, cylinder supplies could be depleted,
leaving an insufficient reserve supply in the event of
pipeline failure.

CAUTION: Use cleaning agent sparingly. Excess fluid could enter the
machine, causing damage.

CAUTION: This machine must only be operated by trained, skilled


medical staff.

CAUTION: Perform the electrical safety inspection as the last step after
completing a repair or after routine maintenance. Perform
this inspection with all covers, panels, and screws installed.

CAUTION: After changing the CO2 absorbent, carry out a system leak
test.

CAUTION: Only Selectatec™ compatible vaporizers with Interlock-


System may be used with the AS3000 unit.

CAUTION: After each exchange of a vaporizer, carry out a system Leak


test.

viii 0070-10-0683 AS3000™ Service Manual


Notes Introduction

CAUTION: The bellows dome cannot be autoclaved.

CAUTION: Do not clean the machine while it is on and/or plugged in.

CAUTION: Pressing “Quit” at any time during the procedure will cancel
the session's settings and reload the previously-stored
calibration coefficients.

CAUTION: Depleted soda lime changes color. Replace the soda lime if
approximately 2/3 of the absorber content is discolored.
CO2 absorbent can be safely changed without stopping
mechanical ventilation.

Notes
NOTE: Unauthorized servicing may void the remainder of the
warranty. Check with the factory or with a local authorized
distributor to determine the warranty status of a particular
instrument.

AS3000™ Service Manual 0070-10-0683 ix


Introduction Notes

This page intentionally left blank.

x 0070-10-0683 AS3000™ Service Manual


1.0 Theory of Operation

1.1 Introduction
The AS3000 is a continuous flow anesthesia system which offers manual or automatic
ventilation, easily adjustable fresh gas delivery, anesthetic agent delivery, ventilation
monitoring, convenient ergonomics, and state-of-the-art safety systems. The components of the
AS3000 Anesthesia System are described this chapter.

AS3000™ Service Manual 0070-10-0683 1-1


Microprocessor-controlled Ventilator Theory of Operation

1.2 Microprocessor-controlled Ventilator


The Microprocessor-controlled ventilator, with its dedicated Breathing System, allows time-
controlled, pressure limited, constant volume ventilation for all patient groups within a tidal
volume range of 40 mL (4 kg infant) to 1400 mL (large adult).

Time-controlled, pressure limited, and compliance compensated constant volume ventilation


is provided through the Controlled Mandatory Ventilation (CMV) mode. The CMV mode
delivers a viable ventilation method for complicated lung conditions. The ventilator also
provides time-controlled, volume dependant ventilation, targeting a set (adjustable) target
pressure provided through the Pressure Controlled Ventilation (PCV) mode. Automatic and
comprehensive system startup tests and alarm management systems ensure controlled
ventilation conditions in every mode of operation.

The durable and ergonomically designed user interface and Navigator™ Knob enables easy
operation. The display provides the selected ventilation modes (CMV, PCV, PS and SIMV)
and the following values: Tidal Volume, Peak Pressure, Mean Pressure, FiO2, Breath Rate, I:E
Ratio, PEEP, Plateau Pressure, Alarm limits, and real-time Airway Pressure and flow
waveform.

P
Exp. Valve

Exp. Flow
APL Valve
Sensor
Breathing Bag

Absorber Exhaust
Valve
Patient Canister
Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve
Fresh Gas

Pressure AIR•
Regulator
O2
Flow Flow
Meter Meter

O2
Outlet

Pressure
O2 Switch
Flush
P

Cylinder
O2
O

Fresh Gas Flow


Pressure
Flow Sensor

Tubes Inlet Filter


P

Vaporizer
Vaporizer
Regulator
P

V
Cylinder
Gas Inlet
AIR

Block
O

Pressure Inlet Filter


P

Regulator
P

Cylinder
N20

Ratio•
O

Controller
N20• Pressure Inlet Filter
P

Cuttoff Regulator

FIGURE 1-1 The AS3000 Pneumatic System

1-2 0070-10-0683 AS3000™ Service Manual


Theory of Operation Components

1.3 Components
1.3.1 Front View
1

11

3 10

FIGURE 1-2 AS3000 Front View

1. Monitor Arm
The Monitor Arm provides support for a bedside monitor. It can easily rotate for more
convenient viewing.

2. User Interface
The display of the user Interface provides waveforms, numeric data and menu tabs. The keys
and Navigator Knob enable the user to power up the system, silence alarms, access menu
tabs, and switch between manual and mechanic ventilation.

AS3000™ Service Manual 0070-10-0683 1-3


Components Theory of Operation

3. UI Arm
The UI Arm provides support for the user interface assembly. It can easily rotate for more
convenient viewing.

4. Oxygen Sensor
The Oxygen Sensor monitors the oxygen concentration of the inspired gas of the Breathing
System.

5. Breathing System
The Breathing System’s main function is to store anesthetic gas, oxygen, and air; vent exhaust
gas; and absorb carbon dioxide. It connects directly to the respiratory passage to help
complete the breathing process.

6. CGO (Common Gas Outlet) Subassembly


Mixed gas composed of O2, AIR, N2O, and anesthetic agent connects to the patient's
Breathing System via a flexible tube from the CGO Subassembly.

7. O2 Flush Valve
The O2 Flush Valve is located on the front of the AS3000. The supplied gas does not pass
the flowmeter and vaporizer. It is directly sent to the fresh gas outlet. Press this button to
supply gas (35 - 50 L/min). Release this switch to automatically close the gas supply.

8. AGSS (Anesthetic Gas Scavenging System)


The AGSS (Anesthetic Gas Scavenging System) reclaims exhausted gas generated during
anesthesia.

9. Drawer Subassembly
The AS3000 has three drawers for storage, which can be locked and fixed through the
uppermost drawer lock.

10. Vaporizer Mounting Manifold


The Vaporizer Mounting Manifold provides support for up to two Selectatec® compatible
anesthetic vaporizers.

11. Flowmeter
The Flowmeter displays gas flow values for N2O, O2, AIR, Auxiliary O2, and Auxiliary AIR.
It consists of coarse and fine flow tubes used to accurately measure gas flow with additional
flow tubes to measure auxiliary O2 and AIR gas flow.
Flow tube measurement values:

COARSE FLOW TUBE


GAS FINE FLOW TUBE -01 UNITS -02 UNITS AUXILIARY FLOW TUBE
N2 O 0 - 1 L/min 1 - 12 L/min 1 - 10 L/min N/A
O2 0 - 1 L/min 1 - 10 L/min 1 - 10 L/min 0 - 15 L/min
AIR 0 - 1 L/min 1 - 15 L/min 1 - 12 L/min 0 - 15 L/min

1-4 0070-10-0683 AS3000™ Service Manual


Theory of Operation Components

1.3.2 Side View

12

13

14

15

16

17

18

19

20

21

22

23

FIGURE 1-3 AS3000 Side View

12. Upper Mainframe Subassembly


The Upper Mainframe Subassembly provides support for the pressure gauges, the Flowmeter,
the gas box assembly, the mains supply assembly, and the YOKE assembly.

13. User Interface Cable


The User Interface Cable provides signal transmission between the User Interface and the
main unit.

AS3000™ Service Manual 0070-10-0683 1-5


Components Theory of Operation

14. Auxiliary Gas Outlet Assembly


The Auxiliary Gas Outlet Assembly provides the patient with auxiliary oxygen and air mixtures
(of different concentrations). It also contains an O2 connection for use with other equipment.

15. Breathing System Interface


The Breathing System Interface provides a connection interface for gas signal acquisition
between the Breathing System and gas circuit.

16. Breathing System Pneumatic Hose


The Breathing System Pneumatic Hose provides multiple connections including: PEEP control,
Auto/Manual control, and four pressure sampling connections between the Breathing System
and mainframe.

17. O2 Sensor Cable


The O2 Sensor Cable provides a connection between the oxygen sensor component in the Breathing
System and the oxygen concentration signal acquisition port on outlet module of the mainframe.

18. Breathing System Interface


The Breathing System Interface provides the main unit with an interface for connection with
drive gas, heating system and the oxygen concentration sensor of the Breathing System.

19. The Breathing System Support Arm


The Breathing System Support Arm provides connection between the Breathing System and
the mainframe, in order to support the Breathing System.

20. Heater Wire


The Heater Wire provides connection to the heating system of the Breathing System. By
heating the Breathing System during anesthesia, accumulation of water in the Breathing
System is minimized. It also provides comfortable gas to the patient.

21. AGSS Transfer Hose


The AGSS Transfer Hose provides pipeline connection between the exhaust gas outlet of the
Breathing System and the AGSS evacuation system.

22. Lower Mainframe Assembly


The Lower Mainframe Assembly provides support for the drawer subassembly, outlet module,
CGO subassembly, O2 flush valve and the electric box. Together with the Upper Mainframe
Subassembly and Base Assembly, forms the mainframe of the machine.

23. Base Assembly


The Base Assembly provides support to the whole machine. The four casters provide
movement for the machine in any direction. The front casters have a locking function.

1-6 0070-10-0683 AS3000™ Service Manual


Theory of Operation Components

1.3.3 Rear View

24

25

26

27

28

29

30

31

32

33

FIGURE 1-4 AS3000 Rear View

24. Top Shelf Assembly


The Top shelf Assembly includes the work light and its switch. It also serves as a storage
area.

25. Gas Input Assembly


The Gas Input Assembly allows gas in the pipeline and backup gas cylinder to enter the
AS3000 for regulation. The Pressure regulator reduces pressure to 36 psi (250 kPa). The
O2 operated valve can open allowing N2O to enter the flowmeter only when O2 pressure is
above 7.3 psi (50 kPa). Otherwise, N2O cannot enter the gas circuit. The pressure switch
generates a signal when the input pressure is below 29 psi (200 kPa), and the system will
provide an audible alarm.

AS3000™ Service Manual 0070-10-0683 1-7


Components Theory of Operation

26. Vaporizer Storage Mount


The Vaporizer Storage Mount provides the AS3000 with auxiliary placement for a
vaporizer.

27. Mains Supply Assembly


The Mains Supply Assembly attaches the AS3000 to an external AC wall outlet. It also
supplies power throughout the unit.

28. Electronic Flowmeter Assembly


The Electronic Flowmeter Assembly provides fresh gas flow measurement for the AS3000.

29. Gas Box Assembly


The Gas Box Assembly provides drive gas to the Breathing System and includes a safety
valve to insure that pressure stays below 85 cmH2O. It also contains the PEEP valve, and the
differential pressure sensors.

30. YOKE Assembly


The YOKE Assembly provides connection for three backup gas cylinders. The pressure
regulator for the backup gas cylinder reduces the pressure of a high pressure gas cylinder
down to 43.5 psi (300 kPa).

31. AMP Cables


The AMP Cables provide signal transmission between the electric box and gas circuit box.

32. Electric Box Assembly


The Electric Box Assembly provides, and distributes power to the AS3000.

33. Rear Panel Assembly


The Rear Panel Assembly protects, and provides heat dissipation for the electric box and
auxiliary support for the backup gas cylinder.

1.3.4 Other components (not identified in graphics)


34. Inspiratory and Expiratory Valves
The Inspiratory and Expiratory Valves are unidirectional valves that allow air to flow in only
one direction.

35. Absorber Canister


The AS3000 uses two Absorber Canisters which can contain 1500 mL of soda lime each.
They can be used for 6 - 8 hours each if full. Water generated from the reaction with CO2 is
drained from a valve on the lower side of the canister.

36. APL (Airway Pressure Limiting) Valve


The APL Valve is used for limiting maximum airway pressure during manual ventilation. The
adjustable range of the APL is 0 - 70 cmH2O.

1-8 0070-10-0683 AS3000™ Service Manual


Theory of Operation Components

37. Bellows Assembly


The AS3000 employs ascending ventilation bellows, in which mixed gas is stored. Drive
gas supplied by the ventilator forces the bellows to descend sending mixed gas into the
inspiratory passage of patient's airway. If a patient's airway suffers from gas leakage, the
bellows will collapse, informing the operator of a possible problem. A tidal volume scale is
provided on the transparent dome, through which a patient's tidal volume can be estimated.

38. Pressure-relief Valve


The Pressure-relief Valve is located at the base of the bellows. When end-expiration airway
pressure reaches 1 - 3 cmH2O, the pressure-relief valve opens and redundant gas is
expelled.

39. Exhaust Gas Outlet


The Exhaust Gas Outlet is located on the lower part of the Breathing System. It is connected
to the AGSS or via the AGSS transfer tube.

40. Absorber Heating System


The Breathing System is heated to body temperature to avoid humidified gases condensing
within the Breathing System thus improving airway climatization for the patient's re-breathing
of respiratory gases.

1.3.5 Breathing System


The Breathing System is integrated into a compact aluminum block. This block is heated to
body temperature to prevent condensing of humidified gases within the Breathing System,
thus improving airway climatization for the patient's re-breathing of respiratory gases. The
heated Breathing System contains: an inspiratory valve with O2 adapter for FiO2
measurement, expiratory valve, APL Valve, breathing bag connection, and internal
inspiratory and expiratory flow sensors.

1.3.6 The Ventilator Unit


The AS3000 ventilator offers multiple ventilation modes: Controlled Mandatory Ventilation
with volume control (CMV), Pressure Control Ventilation (PCV), Synchronized Intermittent
Mandatory Ventilation (SIMV), and Pressure Support (PS) ventilation. Electronic PEEP is
available in all ventilation modes. User control over inspiratory flow (SLOPE) is possible in
PCV, SIMV, and PS modes. Automatic fresh gas compensation limits the effect of user
changes in fresh gas flow rate on the patient. The traditional bellows system is driven by
oxygen and makes patient disconnections clearly visible.

AS3000™ Service Manual 0070-10-0683 1-9


Components Theory of Operation

1.3.7 Adjustable Alarms


Minimum and maximum alarm limits can be set for Peak Pressure, Mean Pressure and FiO2.
Minimum alarm limits can be set for Tidal Volume and Minute Volume. Exceeding the peak
pressure alarm limit automatically halts the inspiratory phase preventing airway pressure
from exceeding the high alarm setting. In the CMV mode, when reaching this pressure limit,
a “High Airway Pressure” alarm is displayed, and inspiration is discontinued. The next
inspiration occurs at the regular time interval, preventing increase of the respiratory rate. The
result is a decreased tidal volume (T Vol.) and minute volume (M Vol.). During pressure
limitation, the ventilator displays the alarm message until the condition is corrected.

1.3.8 Compliance Compensation


Compliance compensation automatically corrects for the expansion of the circuit in CMV
ventilation mode. System compliance is measured by the ventilator to maintain the set tidal
volume (±15%). The compliance test may be bypassed at machine power up. When
bypassing the compliance test, the default settings are used.

1 - 10 0070-10-0683 AS3000™ Service Manual


Theory of Operation Electrical Supply

1.4 Electrical Supply


1.4.1 Electrical components
Ventilator UI Anesthesia
10.4” TFT-LCD Backlight Inverter

FM Back Light
Amplifier for

Work Light
audio alarm Keypad
PC104
Communication Knob
and Interface

Cable SW

Ventilator Control and Drive


Switch

BDU PWR MNG

AC Heater
Pwr

Amplifier DC/DC Ctrl &


Protect
Switch
DA
Gas Source Pressure

Power Supply
Manual/Mech Val.

PS

Filter
Airway Pressure

Inlet
FM Flow Sensor

Internal Signals
Val. Exp. PEEP

BAT
FiO2 Sensor

P. Val. Insp.
Probe Flow

AC/DC
Cal. Valve

Flow Insp.

Aux Mains
Outlet

FIGURE 1-5 Electrical Components Overview

AS3000™ Service Manual 0070-10-0683 1 - 11


Ventilator Control and Drive Theory of Operation

1.5 Ventilator Control and Drive


1.5.1 BDU (Basic Digital Unit)
The BDU serves as the active ventilator control. The BDU controls the actions of the PEEP,
calibration, and proportional valves, and reads the signal flow, and pressure sensors and
valves

FIGURE 1-6 BDU Control Board, Top View

1 - 12 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

FIGURE 1-7 BDU Control Board, Bottom View

AS3000™ Service Manual 0070-10-0683 1 - 13


Ventilator Control and Drive Theory of Operation

1.5.1.1 Amplifier Board

FIGURE 1-8 Amplifier Board, Top View

FIGURE 1-9 Amplifier Board, Bottom View

1 - 14 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

O2 Sensor Input, S4
PIN NAME FUNCTION
1. O2- O2 Sensor Input -
2. O2- O2 Sensor Input -
3. O2+ O2 Sensor Input +
4. O2+ O2 Sensor Input +

DC Power Input, J7
PIN NAME FUNCTION
1. GND Power Ground
2. +5V Controller Logic Power
3. P12V Power 12V
4. PWBUS Power Bus 24V
5. GND Analog Ground
6. A12V Analog 12V
7. 5V/4.8V Power 5V/4.8V

Power Control, J20


PIN NAME FUNCTION
1. GND Ground
2. /PWON Power On Enable, Low active
3. PWEN Power On Enable, High active
4. 5VAUX 5V Auxiliary
5. DC_M AC/DC Monitor signal
6. PWBUS Power bus
7. BAT_M Battery Monitor signal
8. APE Auxiliary Power Enable
9. CHAR_SIGNAL Charge Status, No connection
10. /MUTE Mute, No connection

Signals and Power to Keyboard, J5


PIN NAME FUNCTION
1. ENPW Power Enable, High active
2. /PWON / Power on Enable, Low active
3. 5VAUX 5V Auxiliary
4. APE Auxiliary Power Enable
5. GND Ground
6. TTLRX TTL Receive
7. 232ARX RS232A Receive
8. 232BRX RS232B Receive
9. GND Ground

AS3000™ Service Manual 0070-10-0683 1 - 15


Ventilator Control and Drive Theory of Operation

Signals and Power to Keyboard, J5 (Continued)


PIN NAME FUNCTION
10. PWBUS PowerBus
11. PWBUS PowerBus
12. PWBUS PowerBus
13. GND Ground
14. GND Ground
15. DC24V DC24V monitor
16. BATV+ Battery monitor
17. /MUTE No connection
18. GND Ground
19. TTLTX TTL Transmit
20. 232ATX RS232A Transmit
21. 232BTX RS232B Transmit
22. GND Ground
23. PWBUS PowerBus
24. PWBUS PowerBus
25. GND Ground

Signals and Power to Sensor Board, J30


PIN NAME FUNCTION
1. A+12V Analogue +12V
2. FL_INS Import flow
3. FL_EXP Export flow
4. FM_FL Flowmeter flow
5. Pair Pressure of air
6. Paw Pressure inside air way
7. Pgas Pressure gas supply status
8. GND Ground
9. AGND Analog Ground
10. AGND Analog Ground
11. P24V Power 24V
12. P24V Power 24V
13. P24V Power 24V
14. P12V Power 12V
15. P12V Power 12
16. P5V Power 5V
17. P5V Power 5V
18. Inhale Inspire Valve
19. Man Manual/AUTO Valve
20. Exhale Expire Valve
21. Exhale Expire Valve
22. AGND Analog Ground

1 - 16 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

Signals and Power to Sensor Board, J30 (Continued)


PIN NAME FUNCTION
23. AGND Analog Ground
24. AGND Analog Ground
25. AGND Analog Ground

Bus to BDU, J96


PIN NAME FUNCTION
1-3 5V 5V Power
4-6 GND Ground

Test Point Definition


DESIGNATOR NAME FUNCTION RANGE
T1 FIO2 ADC input of O2 Sensor 0.25-0.51V (In AIR)
T2 Pgas Pressure switch of gas supply,
low active
T3 FL_INSP Inspire Flow signal 0.20-0.30V (0 flow)
T4 Pair Absolute Pressure of air way 1.7-2.1V (atmospheric
pressure)
T5 FL_EXP Expire flow signal 0.20-0.30V (0 flow)
T6 Fm_fl Flow of flowmeter 0.25-0.51V (flowmeter off)
T7 AD-Exhale PEEP valve, current feedback Level depends on PEEP
setting and vent mode. No
activity may indicate open
connection from driver to
valve (ie. bad cable)
T8 Volt Internal power inspect
T9 MUTE Not used
T10 M-EN Manual/Auto mode select
T11 V-EN Valves enable Input Global valve enable signal.
Must be active for valves to
operate.
T12 EX-DA Exhale DA output
T13 In-DA Inhale DA output
T14 GS_ST gas supply status, high active
T15 XVI unamplified version of T7
T16 IVI unamplified version of T28
T17 MVI Manual-Valve current feedback
T18 IVG Proportional valve drive signal, Level depends on machine
drives T16 and T28 settings.
T19 XVG PEEP valve drive signal - drives Level depends on PEEP
T15 and T7 setting and vend mode.
T20 PADJ Power 4.8V 4.8V
T21 P24V Power 24V 24V
T22 P12V Power 12V 12V

AS3000™ Service Manual 0070-10-0683 1 - 17


Ventilator Control and Drive Theory of Operation

Test Point Definition


DESIGNATOR NAME FUNCTION RANGE
T23 -10V Negative Voltage -10V -10V
T24 A12V Analogue 12V 12V
T25 VCC Power 5V 5V
T26 5VAUX 5V Auxiliary 5V
T27 10V Analogue +10V +10V
T28 AD-Inhale Inhale-STATE feedback No activity may indicate
open connection from driver
to valve (ie. bad cable)
T29 AGND Analogue Ground
T30 GND Ground
T31 PAW Pressure inside air way 0.45-0.55 (open to
atmosphere)
T32 Manual Manual/Auto Valve feedback

1 - 18 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

1.5.1.2 Drive Gas Pressure Sensor Board

FIGURE 1-10 Drive Gas Pressure Sensor Board, Top View

FIGURE 1-11 Drive Gas Pressure Sensor Board, Bottom View

Pair Board
Pair, J1
PIN NAME FUNCTION
1. AGND Analog Ground
2. Pair Pressure of air
3. NC No connection
4. AR10V 10V
5. NC No connection

AS3000™ Service Manual 0070-10-0683 1 - 19


Ventilator Control and Drive Theory of Operation

1.5.1.3 PAW Pressure Sensor Board

FIGURE 1-12 Paw Pressure Sensor Board, Top View

FIGURE 1-13 Paw Pressure Sensor Board, Bottom View

Paw Board
Paw, J2
PIN NAME FUNCTION
1. AGND Analog Ground
2. Paw Pressure inside airway
3. AR5V 5V
4. NC No connection
5. NC No connection

1 - 20 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

1.5.1.4 Breathing System Heater

FIGURE 1-14 Breathing System Heater Board, Top View

FIGURE 1-15 Breathing System Heater Board, Bottom View

J2
PIN NAME FUNCTION
2. GND Power Ground
3. 5V Controller Power, Output Enable
4. 24V Heater Power Supply
5. NC No Connection

AS3000™ Service Manual 0070-10-0683 1 - 21


Ventilator Control and Drive Theory of Operation

J3
PIN NAME FUNCTION
1. RET Heater Power Ground
2. DQ Sensor signal
3. 24V Heater Output

Test point definition


DESIGNATOR NAME FUNCTION
T10 GND Power Ground
T1 5V Controller Power
T7 24V Heater Power input
T9 POUT Heater Output

Indicating lamp definition


DESIGNATOR STATUS FUNCTION
LT2 RED Heater Output available
LT3 GREEN Heater Ready

1 - 22 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

1.5.1.5 Sensor Board

FIGURE 1-16 Sensor Board, Top View

FIGURE 1-17 Sensor Board, Bottom View

AS3000™ Service Manual 0070-10-0683 1 - 23


Ventilator Control and Drive Theory of Operation

Sensor Board
Pair, J1
PIN NAME FUNCTION
1. AGND Analog Ground
2. Pair Pressure of air
3. NC No connection
4. AR10V 10V
5. NC No connection

PAW, J2
PIN NAME FUNCTION
1. AGND Analog Ground
2. Paw Pressure inside airway
3. AR5V 5V
4. NC No connection
5. NC No connection

Work Light Power, J11


PIN NAME FUNCTION
1. 24V Power 24V
2. GND Power Ground

Backlight Power, J15


PIN NAME FUNCTION
1. 24V Power 24V
2. GND Power Ground

Switch of Gas Supply, J6


PIN NAME FUNCTION
1. GND Ground
2. NC No connection
3. Pgas Switch input

Expire Valve, J14


PIN NAME FUNCTION
1. 5V 5V
2. Valexp Expire Valve Drive

1 - 24 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

Inspire Valve, J12


PIN NAME FUNCTION
1-2 24V 24V
3-4 Valexp inspire Valve Drive

Man/Auto Valve, J14


PIN NAME FUNCTION
1. 24V 24V
2. Valman Manual/Auto Valve Drive

Signals and Power of Sensor Board, J13


PIN NAME FUNCTION
1. A+12V Analogue +12V
2. FL_INS Import flow
3. FL_EXP Export flow
4. FM_FL Flowmeter flow
5. Pair Pressure of air
6. Paw Pressure inside air way
7. Pgas Pressure gas supply status
8. GND Ground
9. AGND Analog Ground
10. AGND Analog Ground
11. P24V Power 24V
12. P24V Power 24V
13. P24V Power 24V
14. P12V Power 12V
15. P12V Power 12
16. P5V Power 5V
17. P5V Power 5V
18. Inhale Inspire Valve
19. Man MANUAL/AUTO Valve
20. Exhale Expire Valve
21. Exhale Expire Valve
22. AGND Analog Ground
23. AGND Analog Ground
24. AGND Analog Ground
25. AGND Analog Ground

AS3000™ Service Manual 0070-10-0683 1 - 25


Ventilator Control and Drive Theory of Operation

1.5.2 Power Management


Power management is located behind the Rear Panel Assembly. This module serves as the
voltage supply for the ventilator control and drive BDU, the flowmeter backlight, the work
light, the Breathing System Heater, and the charging/discharging control for the battery.

FIGURE 1-18 Power Board, Top View

FIGURE 1-19 Power Board, Bottom View

1 - 26 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator Control and Drive

Power Board
Battery input, J4
PIN NAME FUNCTION
1. BAT Battery Input +
2. 5V Power Ground
3. BATTEST Battery test input

DC24V Power input, J3


PIN NAME FUNCTION
4. GND Power Ground
5. GND Power Ground
6. 24V Power input
7. 24V Power input

DC Power output, J2
PIN NAME FUNCTION
1. GND Power Ground
2. +5V Controller Logic Power
3. P12V Power 12V
4. PWBUS Power Bus 24V
5. GND Analog Ground
6. A12V Analog 12V
7. 5V/4.8V Power 5V/4.8V

Heater Power, J1
PIN NAME FUNCTION
6. GND Power Ground
7. 5V Controller Power, Output Enable
8. 24V Heater Power Supply
9. PBUS Power Bus, No Connection

Power Control, J2
PIN NAME FUNCTION
1. GND Ground
2. /PWON Power On Enable, Low active
3. PWEN Power On Enable, High active
4. 5VAUX 5V Auxiliary
5. DC_M AC/DC Monitor signal
6. PWBUS Power bus
7. BAT_M Battery Monitor signal

AS3000™ Service Manual 0070-10-0683 1 - 27


Ventilator Control and Drive Theory of Operation

Power Control, J2 (Continued)


PIN NAME FUNCTION
8. APE Auxiliary Power Enable
9. CHAR_SIGNAL Charge Status, No connection
10. /MUTE Mute, No connection

Power Switch, J5
PIN NAME FUNCTION
1. GND Ground
2. NC No Connection
3. PWON Power On Enable, Low active

Test Point definitions


DESIGNATOR NAME FUNCTION
T1 24V 24V DC Input
T2 PWBUS Power Bus
T3 A12V Analogue 12V
T4 P12V Power 12V
T5 +5V Logic Power 5V
T6 5V/4.8V Power 5V/4.8V
T7 BAT+ Battery Input +
T8 30V 30V for charger
T9 PWAUX Power Auxiliary
T10 CHARGE Charge Status
T11 5VAUX 5V Auxiliary
T12 VRs Charger Feedback

Led Designation
DESIGNATOR LED COLOR STATUS FUNCTION
DS1 GREEN 24V DC available
DS2 GREEN Power Bus On
DS3 GREEN Charge

1.5.3 Battery
The sealed lead-acid battery is maintenance free, and has a maximum recharge time of 8
hours when fully discharged. The run time is at least 45 minutes with a fully charged battery.
To prevent unintended loss of battery operation, the recommended replacement period is
every 3 years.

1.5.4 Power Supply


The power supply provides power to the machine and relevant controls.

1 - 28 0070-10-0683 AS3000™ Service Manual


Theory of Operation Anesthesia System Components

1.6 Anesthesia System Components


1.6.1 Auxiliary Outlets
The AS3000 has four auxiliary outlets (120 VAC, 60 Hz, 2A maximum each). There are
two 2A fuses for each outlet.

1.6.2 Absorber Heater Wire Board

FIGURE 1-20 Absorber Heater Wire Board, Top View

FIGURE 1-21 Absorber Heater Wire Board, Bottom View

1.6.3 Work Light Board

FIGURE 1-22 Work Light Board, Top View

FIGURE 1-23 Work Light Board, Bottom View

AS3000™ Service Manual 0070-10-0683 1 - 29


Anesthesia System Components Theory of Operation

Work light Power, J11


PIN NAME FUNCTION
1. 24V Power
2. NC No Connection
3. GND Power Ground

Work light switch, J11


PIN NAME FUNCTION
1. 24V Power 24V
2. NC No Connection
3. LOW Low dim
4. HIGH High dim

1 - 30 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator UI

1.7 Ventilator UI
1.7.1 Keyboard Board

FIGURE 1-24 Keyboard Board, Top View

AS3000™ Service Manual 0070-10-0683 1 - 31


Ventilator UI Theory of Operation

FIGURE 1-25 Keyboard Board, Bottom View

1 - 32 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator UI

Keyboard
Signals and Power of Keyboard, J1
PIN NAME FUNCTION
1. ENPW Power Enable, High active
2. GND Ground
3. /PWON / Power on Enable, Low active
4. DC24V DC24V monitor
5. 5VAUX 5V Auxiliary
6. BATV+ Battery monitor
7. APE Auxiliary Power Enable
8. /MUTE No connection
9. GND Ground
10. GND Ground
11. TTLRX TTL Receive
12. TTLTX TTL Transmit
13. 232ARX RS232A Receive
14. 232ATX RS232A Transmit
15. 232BRX RS232B Receive
16. 232BTX RS232B Transmit
17. GND Ground
18. GND Ground
19. PWBUS PowerBus
20. PWBUS PowerBus
21. PWBUS PowerBus
22. PWBUS PowerBus
23. PWBUS PowerBus
24. GND Ground
25. GND Ground

PC104 Power output, J6


PIN NAME FUNCTION
1. 5V Controller Logic Power
2. GND Power Ground
3. GND Power Ground
4. 12V Power 12V, No connection

Backlight Power, J2
PIN NAME FUNCTION
1. 12V Power output
2. 12V Power output
3. GND Power Ground
4. GND Power Ground

AS3000™ Service Manual 0070-10-0683 1 - 33


Ventilator UI Theory of Operation

Backlight Power, J2 (Continued)


PIN NAME FUNCTION
5. ON/OFF Backlight ENABLE
6. DIM Contrast adjust

Power of 232 Isolator, J9


PIN NAME FUNCTION
1. 5V Power 5V
2. NC No connection
3. GND Ground

COM1/2
PIN NAME FUNCTION
2 RXD 232RXD TO PC104 INPUT
3 TXD 232TXD TO PC104 OUTPUT
5 GND Ground

Test Point Definition


DESIGNATOR NAME FUNCTION
T5 +5V Logic Power 5V
T6 P12V Power 12V
T9 PWBUS Power Bus

LED Designation
DESIGNATOR STATUS FUNCTION
LT1 GREEN/BLINK Keyboard ready

1.7.2 Display
The display is a 10.4 inch TFT LCD

1 - 34 0070-10-0683 AS3000™ Service Manual


Theory of Operation Ventilator UI

1.7.3 Communication Interface / RS232 Isolate Board

FIGURE 1-26 RS232 Isolate Board, Top View

FIGURE 1-27 RS232 Isolate Board, Bottom View

RS232 Isolator
Power, J3
Pin Name Function
4. 5V Power 5V
5. NC No connection
6. GND Ground

COM1/2
Pin Name Function
2 RXD 232RXD TO PC INPUT
3 TXD 232TXD TO PC OUTPUT
5 GND Ground

COM1A/B
Pin Name Function
2 TXD 232TXD TO PC OUTPUT
3 RXD 232RXD TO PC INPUT
5 GND Ground

1.7.4 Fuses
The AS3000 has 10 fuses on the back panel of the unit. There are two 2A fuses for each of
the four outlets and two 10A fuses for the line cord.

AS3000™ Service Manual 0070-10-0683 1 - 35


Ventilator Pneumatic - O2 Drive Gas Theory of Operation

1.8 Ventilator Pneumatic - O2 Drive Gas


1.8.1 Ventilator pneumatic drive
Oxygen is the driving gas for the ventilator. In addition to the flowmeter block, a high
pressure regulator reduces the supply pressure to 25.4 psi (175 kPa). This pressure
represents the drive gas for the ventilator.

The drive pressure regulator is placed ahead of the proportional valve that generates the
driving gas flow during the inspiratory phase. This flow fills the bellows dome that surrounds
the bellows.

1.8.2 Drive Pressure-High pressure regulator


The drive pressure regulator stabilizes the supply pressure provided to the proportional valve.
The flow generated by the proportional valve is therefore independent of pressure variations
at the supply.

Setting the drive pressure regulator at 25.4 psi (175 kPa) allows for a maximum inspiratory
flow of 70 L/min at the ventilator.

1.8.3 Gas Box Assembly


The driving module consists of the proportional valve and a solenoid valve for the vent mode
switch. The proportional valve pressure regulator, and generates a driving gas flow of 0 - 70
L/min in relation to the control voltage of the proportional valve of 0 - 5VDC.

The control voltage of the proportional valve, required for the pre-selected parameter
settings, is generated by the BDU board. The driving gas flow Qdrive gas is in the following
relationship with the tidal volume:

Qdrive gas = VT/T

with: Qdrive gas = driving gas flow

VT = generated tidal volume

T = time

1.8.4 Tube color coding


All the pneumatic tubes used in the AS3000 are color coded for use in the United States
only.

GAS US STANDARD
O2 Green
N2 O Blue
AIR Yellow

1 - 36 0070-10-0683 AS3000™ Service Manual


Theory of Operation The Breathing System

1.9 The Breathing System


1.9.1 CMV mode, inspiration
Tidal volume (T Vol.) compensates for variations in gas flow. This is to ensure that the set tidal
volume is delivered to the patient.

P
Exp. Valve

Exp. Flow
APL Valve
Sensor

Breathing Bag
Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve
Fresh Gas

Pressure
Regulator

FIGURE 1-28 Breathing System Pneumatics, CMV Mode, Inspiration

1.9.2 CMV mode, expiration


As the patient exhales tidal volume into the expiratory limb, fresh gas enters the bellows.
Fresh gas mixes with exhaled gas after the Absorber removes CO2. Excess fresh gas passes
through the exhaust valve to the AGSS

P
Exp. Valve

Exp. Flow
APL Valve
Sensor
Breathing Bag

Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve
Fresh Gas

Pressure
Regulator

FIGURE 1-29 Breathing System Pneumatics, CMV Mode, Expiration

AS3000™ Service Manual 0070-10-0683 1 - 37


The Breathing System Theory of Operation

1.9.3 Manual mode, inspiration


As the breathing bag is compressed, the gas is directed to the patient. Pressures exceeding
the set value of the APL Valve will pass through the APL Valve the AGSS.

P
Exp. Valve

Exp. Flow
APL Valve
Sensor

Breathing Bag
Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow 2 cmH2O
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve
Fresh Gas

Pressure
Regulator

FIGURE 1-30 Breathing System Pneumatics, Manual Mode, Inspiration

1.9.4 Manual mode, expiration


As the patient exhales tidal volume into the expiratory limb, fresh gas enters the Breathing
System. Fresh gas mixes with exhaled gas after the Absorber removes CO2. Excess fresh gas
passes through the exhaust valve to the AGSS

P
Exp. Valve

Exp. Flow
APL Valve
Sensor
Breathing Bag

Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve
2 cm H2O
System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve
Fresh Gas

Pressure
Regulator

FIGURE 1-31 Breathing System Pneumatics, Manual Mode, Expiration

1 - 38 0070-10-0683 AS3000™ Service Manual


Theory of Operation The Breathing System

1.9.5 Pneumatic PEEP


The PEEP valve regulates the pressure at which the exhaust valve opens, therefore the exhaust
valve opens only when the pressure exceeds the set PEEP pressure.

1.9.6 Ventilator in Standby


When the AS3000 is in the standby mode, monitoring will be inactive, The patient should
not be ventilated when the system is in standby mode.

1.9.7 Breathing System Components


1.9.7.1 Ventilation Bellows System
The ventilator's driving system is a flow generator. Driving gas fills the bellows dome to
compress the bellows. The breathing gas is pressed out of the bellows into the patient
breathing circuit. The bellows is refilled with fresh gas and the expired gas from the patient.

1.9.7.2 Manual Breathing Bag


In manual mode, this device acts as a normal breathing bag, enabling the user to ventilate
the patient manually. In mechanical ventilation mode, this bag is cut off from the breathing
circuit by the reversing valve.

1.9.7.3 CO2 Absorber


The soda lime inside the absorber retains the carbon dioxide from the exhaled gas. The
AS3000 accommodates standard sized Pre-Paks or loose-fill CO2 absorbent.

1.9.7.4 Inspiratory and Expiratory Valves


To ensure correct gas flow direction to and from the patient, one-way-valves are integrated in
the inspiratory and expiratory limb of the Breathing System.

1.9.7.5 APL (Airway Pressure Limiting) valve


In manual mode, the APL Valve acts as a normal spring loaded pressure relief valve, limiting
the maximum pressure in the Breathing System.

AS3000™ Service Manual 0070-10-0683 1 - 39


The Breathing System Theory of Operation

This page intentionally left blank.

1 - 40 0070-10-0683 AS3000™ Service Manual


2.0 Installation Guide

2.1 Delivery of The New Anesthesia Machine


The following customer supplied material must be present prior to installation. Missing
equipment can result in delays, incomplete installations and/or extra visits.

• Compatible emergency O2, N2O, and AIR cylinders


• Agent vaporizers and key fillers, if not purchased with the AS3000
• Liquid agent medication
• CO2 absorbent Pre-Paks or loose fill
• Active O2, N2O, and AIR, lines at 50 psi
• Drop down hoses for ceiling mounted medical gas utilities, compatible with quick-
disconnect hoses if not purchased with the AS3000
• Activated medical gases (O2, N2O, AIR, VAC, and EVAC)

AS3000™ Service Manual 0070-10-0683 2-1


Assembly Installation Guide

2.2 Assembly
NOTE: The AS3000 Breathing System Block is matched to the
AS3000 it is attached to by calibration and installation.
After removing the Breathing System Block from the
Mounting Arm of the AS3000, assure that the Breathing
System Block is returned to the same AS3000 that it came
from originally.

2.2.1 Unpacking
1. Inspect the two boxes for any damage, check the tip watches, and report any damage
or sign of tipping to the dispatcher.
2. Ensure the serial numbers on both carton labels match prior to assembling the system.
3. Remove the shipping straps and caps from the 2 boxes.
4. Remove the foam from inside the top of the box.
5. Remove the Corrugated box from around AS3000.
6. Remove the bag and shrink wrap from around AS3000.
7. Remove the foam from AS3000 casters.

2.2.2 Breathing System and Breathing System Accessories


5
4 6 7

8 9 10 11 12

FIGURE 2-1 The AS3000 and accessories

1. AGSS 2. Breathing System


3. User Interface Mounting Arm 4. User Interface
5. Patient Monitor Mounting Arm 6. O2 Cable
7. Pneumatic Hose Assembly 8. Heater Cable
9. Drive Gas 10. Breathing System Mounting Arm
11. AGSS Transfer Hose 12. CGO Port

2-2 0070-10-0683 AS3000™ Service Manual


Installation Guide Assembly

2.2.2.1 Attaching the Mounting Arms


(see FIGURE 2-1)

1. Remove the mounting arms from small box and install them onto the AS3000 using a 3
mm allen wrench.
2. Remove the keys from the parts box, and unlock the AS3000 drawers and put parts
box into top drawer.
3. Remove the AGSS and Breathing System mount from the foam packaging, and install
them onto the AS3000.
4. Remove the User Interface and slide it onto the User Interface mounting arm.

2.2.2.2 Breathing System Connections


(see FIGURE 2-1)

1. Remove the foam and bag from around the Breathing System.
2. Ensure that the serial number tag attached to the bag arm matches the serial number
found on the package carton.
3. Install the Breathing System onto the arm on the left side of the AS3000 by attaching it
to the installed bracket.
4. Attach the Drive Gas Hose to the bottom port of the Breathing System
5. Attach the Heater cable into the port located on the bottom of the Breathing System
6. Install the O2 Cell to the O2 Interface
7. Attach the O2 cell to the O2 cable and insert the cell into its designated port on the
Breathing System.
8. Attach the Pneumatic Hose Assembly to the designated port on the rear of the Breathing
System.
9. Attach the AGSS Transfer Hose to its designated port on the bottom of the Breathing
System.

2.2.2.3 Anesthesia System Connections


(see FIGURE 2-1)

1. Attach the Fresh Gas Hose of the Breathing System to the CGO port on the system.
2. Attach the Drive Gas Hose to the designated port on the Breathing System Interface.
3. Attach the Heater Cable to the designated port on the Breathing System Interface.
4. Attach the O2 Cable to the designated port on the Breathing System Interface.
5. Attach the Pneumatic Hose Assembly to the designated connector labeled Breathing
System Pneumatics.
6. Attach the AGSS Transfer Hose.

2.2.3 Tank Wrench and Pre-operation Checklist


1. Mount the tank wrench on the rear of the AS3000 so that it can be used to open or
close each cylinder, without disconnecting it from the machine.
2. Attach the pre-operation checklist to a location on the AS3000 where the operator can
access it.

AS3000™ Service Manual 0070-10-0683 2-3


Assembly Installation Guide

2.2.4 Patient Suction Regulator and Arm


1. Remove the Patient Suction Regulator and mounting arm from its packaging.
2. Attach the mounting arm to the customers desired channel. It mounts to either side
channel.
3. Tighten the locking screw.
4. Thread the patient suction regulator into its mating port on the mounting arm.
5. Attach a barbed fitting adapter to the suction regulator's VAC fitting as necessary to
attach a hospital grade suction hose from the regulator output to the suction canister.

2.2.5 Suction Canister Bracket


1. Remove the suction canister bracket from its packaging.
2. Attach the bracket to the right lower rail and secure.

2.2.6 Utility Tray, Monitor Mounting Arm with Utility Hook(s)


1. Remove the items from their packing material.
2. Slide the utility tray into the desired channel.
3. Tighten the locking screw.
4. Remove the plastic cap from under the mounting arm's swivel head.
5. Mount the utility hook(s) under the swivel head of the arm using the screws and tool
provided with the hook.
6. Slide the mounting arm into the desired channel. Allow enough room for mounting
monitors at the desired height.
7. Tighten the locking screw.

2.2.7 Vaporizers
1. Prior to mounting, set the vaporizer to the T position. (Does not apply for all vaporizers.)
2. Discard the washers that came with the vaporizer. Use only approved O-rings that come
mounted on the vaporizer mounting ports of the AS3000.
3. Mount a mechanical vaporizer to either side of the AS3000 vaporizer mount by
placing the vaporizer on the ports and locking down the lever.
4. Mount an electronic vaporizer in the same manner. Follow any installation instructions
that come with the specific vaporizer.

2.2.8 High Pressure Hoses


1. Remove the hoses from their packing material.
2. Attach each hose to its mating connector by hand. Do not use any tools.
3. Use extreme care while attaching the EVAC hose to the waste gas scavenger (WGS); as
it is extremely delicate.
4. Attach an EVAC hose to a VAC source connection only if an EVAC source connection is
not available.

2-4 0070-10-0683 AS3000™ Service Manual


Installation Guide Assembly

2.2.9 Emergency Cylinder(s)


1. Remove the cover from a new O2, N2O, and AIR cylinder.
2. Mount one at a time onto the rear of the anesthesia machine.
3. Discard the cylinder's tank washer. Always use the approved tank washer provided with
the AS3000.
4. Open the bail of each yoke and mount the cylinder over the tank washer.
5. Ensure the O2 cylinder mates to the O2 Pin Index Safety System (PISS) connection on the
O2 yoke. Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke
port.
6. Open and close the cylinder valve and observe that the cylinder gauge on the
anesthesia machine rises to the colored range.
7. Ensure that the N2O cylinder mates to the N2O PISS connection on the N2O yoke.
Close the yoke bail and use the hand-screw to tighten the cylinder to the yoke port.
8. Open and close the cylinder valve and observe that the cylinder gauge on the
anesthesia machine rises to the colored range.
9. Ensure that the AIR cylinder mates to the AIR PISS connection on the AIR yoke. Close the
yoke bail and use the hand-screw to tighten the cylinder to the yoke port.
10. Open and close the cylinder valve and observe the cylinder gauge on the anesthesia
machine rise to the colored range.

2.2.10 Breathing Circuit, CO2 Absorbent, and Liquid Vaporizer Agent


1. Attach a breathing circuit to the inspiratory and expiratory ports as detailed in the
directions for use. Attach the breathing bag and any other respiratory accessories as
described.
2. Insert the Pre-Paks one on top of the other with the wider side facing up on both Pre-
Paks. (alternatively the absorber canisters may be filled with loose fill absorbent.)
3. Install the absorber canister with a quarter turn of the lever at the bottom the absorber
assembly, this ensures a tight seal.

2.2.11 Monitoring Products - Mounting and Electrical Connection


1. Any monitoring system compatible with the GCX mounting system's swivel head may be
mounted to the AS3000’s arm.

NOTE: Use of other monitors and mounting hardware is the


responsibility of the installer.

2. Always make full use of all mounting fasteners and strap capturing devices when
mounting monitors to the AS3000.
3. After mounting a monitor to the AS3000, connect it to one of the AC outlets located on
the rear of the AS3000.
a. Turn on each monitor one at a time and ensure that the circuit breaker holds without
tripping.
b. Dress each line cord neatly along the side of the anesthesia machine so that it can
not be easily pulled or extend far from the main chassis.

AS3000™ Service Manual 0070-10-0683 2-5


Assembly Installation Guide

2.2.12 Agent Monitor Waste Gas Scavenging


1. Respiratory gas monitoring products have an exhaust port from which waste gas expels.
The exhaust port on the gas monitor must be connected to the open barbed fitting on the
waste gas scavenger.
2. Ensure that a tight connecting fitting is attached to the rear of the gas monitor. Ensure
that the other end of the same tube has a tight fitting connection attached to the waste
gas scavenger’s (WGS) barbed fitting.
3. Dress the exhaust tubing neatly along the side of the anesthesia machine so that it can
not be easily pulled or extend far from the main chassis.

2-6 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

2.3 Installation Checkout Procedure


Complete each step to verify the functionality of the AS3000 prior to clinical use.

Also, perform this checkout after installation, reinstallation, servicing or after any periodic
maintenance activity. This checklist does not replace periodic maintenance actions that must
be performed to maintain peak performance.

1. Verify that all components are present, and inspect for physical damage.
a. Ensure that the Operating Instructions is present.
b. Verify unit is free from cosmetic defects. Plastics, labeling and display window are
free from inclusions, pitting, bulges, sink marks, nicks, scratches, gouges dents,
discolorations, etc. Verify all major hardware is fastened properly.
c. Verify that there are Tank Washers for each of the external cylinder yokes.
d. Verify that each vaporizer's locking spring is intact on the manifold by looking
through the opening and checking for the visible wire spring.
e. Verify that each connector of the vaporizer mount has an O-ring.
f. Verify that breathing circuit consumables and CO2 absorbent are present.
g. Inspect the O2, N2O, AIR, VAC, and EVAC supply hoses for damages.
h. Turn the flow control knobs for O2, AIR and N2O and verify that the floats rise and
spin as the flow is increased and fall freely as the flow is decreased.
i. inspect the AC line cord for fraying and damages.

2. Verify that the laminated Preoperative Checkout card is attached.

3. Verify that the tank wrench is attached.

4. Verify that the waste scavenger flow rate is set and that the hose is
intact, connected and dry.
a. Attach the waste gas hose from the waste gas port on the Breathing System to one of
the waste gas ports on the AGSS.
b. Attach the EVAC hose from the waste gas assembly DISS (Diameter Index Safety
System) fitting to a source of vacuum.
c. Adjust the position of the float to be between the Min and Max lines by turning its
flow adjustment knob (counterclockwise increases flow, clockwise decreases flow).

AS3000™ Service Manual 0070-10-0683 2-7


Installation Checkout Procedure Installation Guide

5. Check N2O, O2 and AIR Lines for leaks.


a. Remove the cylinders from the AS3000.
b. Connect the O2, AIR and N2O hoses to the gas line inlets.

NOTE: If a specific gas source is not available, skip the


corresponding leak test.

c. Verify O2, AIR and N2O Line Pressure gauges read the corresponding line pressure.
d. Set the AIR flow controller to minimum.
e. Pinch to occlude the O2, AIR and N2O hoses.
f. Disconnect each hose from its line pressure source.
g. Verify that the pressures do not drop more than 2 psi over a time period of 20
seconds.

6. Verify that all available line and cylinder gauges are operational.
a. Connect the O2 supply hose and remove the AIR and N2O hoses.
b. Verify the O2 Line Pressure gauge shows the reading of the line pressure.
c. Connect the AIR supply hose and remove the O2 and N2O hoses.
d. Verify that the AIR Line Pressure gauges show the reading of the line pressure.
e. Connect the O2 and N2O hoses and remove the AIR hose.
f. Verify that the N2O and O2 Line Pressure gauge shows the reading of the line
pressure.
g. Attach the cylinders to the AS3000's yokes (O2 and N2O or AIR). In the case
where multiple cylinders of the same gas are required, attach one at a time.
h. Open all attached cylinders including the O2 cylinder. If a cylinders pressure is less
than 500 psi, replace the respective cylinder.
i. Verify that the cylinder pressure gauges register a pressure for each type of gas
attached.

7. Test O2 Flush operation with only O2 connected.


a. Connect the O2 supply hose and remove the AIR and N2O hoses.
b. Attach a calibrated flow meter to the Common Gas Outlet.
c. Push the O2 Flush button and verify a reading between 35 L/min and 50 L/min at
the flow meter.

NOTE: O2 flow may be greater than 50 L/min at higher altitudes.

d. Reconnect the O2, AIR and N2O hoses.

8. Test the O2:N2O ratio system (applicable to -01 units only).


a. Set all flow control knobs to minimum.
b. Increase the flow of N2O until the O2 reaches 1 L/min
c. Verify that the N2O flow is not higher than 3.7 L/min.
d. Return flow control knobs to minimum.

2-8 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

9. Test the O2:N2O ratio system (applicable to -02 units only).


a. Set all flow control knobs to minimum.
b. Set the flow of O2 to 1 L/min.
c. Open the N2O flow knob and verify that the N2O flow will not increase higher than
3.7 L/min.
d. Return flow control knobs to minimum.

10. Perform System Test.


a. Power ON the AS3000.
b. Wait until the System Self Test is complete

FIGURE 2-2 Startup Self Test

c. Select Continue to prompt to the Safety Valve test

AS3000™ Service Manual 0070-10-0683 2-9


Installation Checkout Procedure Installation Guide

11. Perform Leak/Safety Valve Test


a. Follow the on-screen instructions.

FIGURE 2-3 Leak Test Setup

b. Select Continue to start the test.

FIGURE 2-4 Safety Valve Test Setup

2 - 10 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

FIGURE 2-5 Safety Valve Test in Progress

FIGURE 2-6 Safety Valve Test Passed Message

c. After 3 seconds, the screen will prompt to the Leak Test.

12. Perform Leak Test

AS3000™ Service Manual 0070-10-0683 2 - 11


Installation Checkout Procedure Installation Guide

a. Follow the on-screen instructions.


b. Select Continue to start the test.

FIGURE 2-7 Leak Test Setup

2 - 12 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

FIGURE 2-8 Leak Test in Progress

c. Select Continue to prompt to the Compliance Test.

FIGURE 2-9 Leak Test Results

AS3000™ Service Manual 0070-10-0683 2 - 13


Installation Checkout Procedure Installation Guide

13. Perform Compliance Valve Test


a. Follow the on-screen instructions.
b. Select Continue to start the test.

FIGURE 2-10 Compliance Test Setup

FIGURE 2-11 Compliance Test in Progress

2 - 14 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

c. Select Continue to prompt to the Normal Screen.

FIGURE 2-12 Compliance Test Results

14. Manual Leak Test

NOTE: The Manual Leak Test detects smaller leaks than can be
detected in the Automatic Leak Test.

a. Ensure that the gas pressure for O2, N2O, and AIR are at 50 ±10 psi.
b. Power ON the AS3000.
c. Attach a reusable-silicone-rubber breathing circuit to the Breathing System.

NOTE: For testing purposes always use a reusable breathing


circuit.

d. Tightly connect the Y-fitting on the breathing circuit to the test port.
e. Attach a breathing bag to the bag arm.
f. Set the APL Valve to the fully closed position (70 cmH2O).
g. Rotate the O2 Flow Control Valve until 50 cmH2O pressure is observed on the
Airway Pressure Gauge.
h. Verify that the flow required to stabilize the pressure is less than 300 mL/min.

AS3000™ Service Manual 0070-10-0683 2 - 15


Installation Checkout Procedure Installation Guide

15. Oxygen Sensor Calibration

NOTE: Oxygen Sensor Calibration can be performed in all


ventilation modes.

a. Allow the breathing system to warm up and reach thermal equilibrium


(approximately 30-60 minutes).
b. Press the MENU key and then use the Navigator™ Knob to scroll to the
Calibrate menu tab (see FIGURE 2-13). Select the Start Calibration button.

FIGURE 2-13 Calibrate Menu Tab

c. After the Start Calibration button has been selected, the screen shown in
FIGURE 2-14 or FIGURE 2-16will be displayed, instructing the user to remove the
oxygen sensor from the Breathing System and expose it to room air for at least 3
minutes before proceeding. The O2 sensor voltage is displayed during the
calibration. This is the amplified O2 cell voltage at the A/D converter for the oxygen
sensor.

NOTE: The O2 sensor voltage is not displayed for UI versions 2.24


and lower.

d. Flush the O2 sensor with air from the auxiliary output for 5-10 seconds to ensure that
no O2 bubbles are trapped in the sensor.

NOTE: Do not shake the O2 sensor during calibration.

NOTE: Keep the O2 sensor in a vertical position, connector side up,


during calibration.

NOTE: Place the O2 sensor on top of the heated block during


calibration to minimize the temperature difference from
within the heated block.

NOTE: If the system is going to be used during the calibration,


insert the O2 cell plug into the port from which the oxygen
sensor was removed using a push and turn motion.

2 - 16 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

e. After at least 3 minutes have passed, select the Next button to initiate the
calibration process. The progress bar shown in FIGURE 2-15 or FIGURE 2-17will be
displayed.

FIGURE 2-14 Oxygen Sensor FIGURE 2-15 Oxygen Sensor Calibration


Calibration Instructions Progress Bar

FIGURE 2-16 Oxygen Sensor FIGURE 2-17 Oxygen Sensor Calibration


Calibration Instructions (UI Progress Bar (UI versions
versions 2.24 and lower) 2.24 and lower)

16. Proceed based on one of the following two conditions:


• If the calibration is successful, the screen shown in FIGURE 2-18 will be
displayed, instructing the user to reinstall the oxygen sensor into the Breathing System.
Select the Done button to complete the process.

AS3000™ Service Manual 0070-10-0683 2 - 17


Installation Checkout Procedure Installation Guide

• If the calibration fails, the screen shown in FIGURE 2-19 will be displayed,
instructing the user to either repeat the calibration (by selecting the Repeat Cal
button) or to replace the oxygen sensor. If the oxygen sensor must be replaced, select
the Exit button, replace the oxygen sensor using a push and turn motion, and then
repeat the calibration.

FIGURE 2-18 Oxygen Sensor FIGURE 2-19 Oxygen Sensor


Calibration Successful Calibration Failed

17. Verify that the APNEA alarm activates in Manual mode


a. Press the MANUAL/AUTO key.
b. After 60 seconds, verify that:
• an alarm tone sounds
• the alarm message APNEA is displayed in red text

18. Verify that CMV ADULT ventilator mode operates and that the tidal
volume display reads within 15% of the CMV set value.
a. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

b. Attach an adult test lung to the Y-fitting of the breathing circuit.


c. Attach a Vent Tester between the EXP port and the expiratory hose.
d. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
e. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Patient Type Adult
Ventilation Mode CMV
Tidal Volume - VT 600
Breath Rate - freq 8

2 - 18 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

VENTILATOR CONTROLS VENTILATOR SETTINGS


I:E Ratio - I:E 1:2
Plateau - TP 10
PEEP - PEEP Off

f. Select CMV again to begin ventilation.


g. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp.
rate and minute volume values appear on the screen.
h. Verify the Tidal Volume display is within 15% of the delivered volume measured with
the Vent Tester within approximately 1 minute from the start of ventilation.
i. Verify the delivered volume at the Vent Tester is within 15% of the Tidal Volume set
value within approximately 1 minute from the start of ventilation.
j. Verify the measured O2 concentration is at least 97% after 5 minutes.
k. Set the AIR flow to 3 L/min and set the N2O and O2 flow rates to minimum flow.
l. Verify the measured O2 concentration is 21% ±3% vol. % after 5 minutes.

19. Verify that CMV CHILD ventilator mode operates and that the tidal
volume display reads within 10% of the CMV set value.
a. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

b. Attach an adult test lung to the Y-fitting of the breathing circuit.

NOTE: Limit the volume in the test lung to provide sufficient airway
pressure to satisfy the Low Peak Pressure alarm. Or reduce
the Peak Pressure alarm limit to a lower valve to prevent
the alarm when using an adult test lung.

AS3000™ Service Manual 0070-10-0683 2 - 19


Installation Checkout Procedure Installation Guide

c. Attach a Vent Tester between the EXP port and the expiratory hose.
d. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
e. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Patient Type Child
Ventilation Mode CMV
Tidal Volume - VT 120
Breath Rate - freq 20
I:E Ratio - I:E 1:2
Plateau - TP 10
PEEP - PEEP Off

f. Select CMV to begin ventilation.


g. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp.
rate and minute volume values appear on the screen.
h. Verify the Tidal Volume display is within 25 ml or ±15% of the delivered volume
(whichever is greater) measured within approximately 1 minute from the start of
ventilation.
i. Verify the delivered volume as measured by a Vent Tester at the expiratory port, is
within 17% of the Tidal Volume set value within approximately 1 minute from the
start of ventilation.
j. Verify that the PEEP reading is 4 ±2 within approximately 1 minute from the start of
ventilation.

20. Verify that the PCV ADULT ventilator mode operates.


a. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

b. Attach an adult test lung to the Y-fitting of the breathing circuit.


c. Attach a Vent Tester between the EXP port and the expiratory hose.
d. Set the O2 flow to 3 L/min and set the N2O and AIR flow rates to minimum flow.

2 - 20 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

e. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Patient Type Adult
Ventilation Mode PCV
Target Pressure - PTARGET 20
Breath Rate - freq 8
I:E Ratio - I:E 1:2
PEEP - PEEP Off
Inspiratory Slope - Tslope 0.5

f. Select PCV to begin ventilation.


g. Verify the Peak Pressure reading of the display is ±20%, at least 2 cmH2O of the
Peak Pressure measured with the Vent Tester.
h. Verify that the pressure waveform, Tidal Volume, Resp. Rate and minute volume
values appear on the screen.
i. Verify that the PEAK Value reaches 20 ±4 cmH2O within five breaths from the start
of ventilation.

21. Verify that the Low Airway Pressure alarm activates.


a. While the ventilator is running disconnect the Inspiratory hose.
b. Verify that the “Low Airway Pressure” and the “APNEA” message appear on the
screen.
c. Reconnect the hose to the Inspiratory port and verify that the alarm is cleared.

22. Verify that the O2 concentration alarm signals activate.


a. While the ventilator is running, press the ALARM LIMITS key.
b. Set the O2 Minimum Alarm Limit to a value higher than the current O2 reading.
c. Verify that the FiO2 alarm activates indicated by an audio tone is present and that
the alarm message “Low FiO2” appears on the screen within three ventilation cycles.
d. Set the O2 Minimum Alarm Limit back to a value lower than the current O2 reading.
e. Verify that the FiO2 alarm resets.

AS3000™ Service Manual 0070-10-0683 2 - 21


Installation Checkout Procedure Installation Guide

23. Verify that the drive gas pressure loss alarm signals activate and N2O
stops and AIR continues to flow.
a. Set the O2 flow for 1 L/min.
b. Set the N2O flow to 1 L/min.
c. Set the AIR flow to 1 L/min.
d. Disconnect the O2 line supply to the AS3000, and Close the O2 cylinder.
e. Verify that the flow of N2O stops when O2 is lost.
f. Verify that AIR continues to flow.
g. Verify the “O2 Supply Failure” message appears on screen, a steady audio tone
should sound.
h. Reconnect O2 line pressure source to the AS3000.
i. Verify that the alarms are reset.
j. Verify N2O flow has resumed once O2 pressure returns.

24. Verify that the battery operation and charging icons appears.
a. Set the vent mode to Standby.
b. Disconnect AC line cord.
c. Verify that the battery icon appears on the screen and indicates the charge level of
the battery.
d. Verify that the line cord icon is flashing.
e. Verify that the flow meter backlight is illuminated during battery operation.
f. When the battery icon is in a not full state, reconnect the AC line cord.
g. Verify that the battery charge icon appears on the screen.
h. Verify that the flow meter backlight is illuminated during AC voltage operation.

25. Verify operation of the work light.


a. Turn on the work light located on the bottom side of the top panel.
b. Verify that it lights in both on positions.

26. Verify that the Auxiliary O2 and AIR Flowmeters operates.


a. Verify an AIR flow of 15 L/min can be obtained by connecting the auxiliary AIR hose
to the pressure source and opening the flow meter.
b. Verify an O2 flow of 15 L/min can be obtained by connecting the auxiliary O2 hose
to the pressure source and opening the flow meter.

27. Verify that the Suction Regulator operates in LINE and REGULATE
positions.
a. Set the suction regulator's selection dial to LINE.
b. Verify maximum suction vacuum on the regulator's gauge.
c. Set the suction regulator's selection dial to REGULATE.
d. Verify the suction vacuum is adjustable on the regulator's gauge.
e. Set the suction regulator's selection dial to OFF.

2 - 22 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

28. Verify that the Breathing System heats to body temperature.


a. Operate the AS3000 on AC operation for approximately 40 minutes.
b. Verify the Breathing System has heated up to body temperature

29. Complete electrical safety inspection.

NOTE: Perform the electrical safety inspection as the last step after
completing a repair or after routine maintenance. Perform
this inspection with all covers, panels, and screws installed.

a. Withdraw the Power cable(s) from the convenience receptacles at the rear of the
AS3000.
b. Plug the AS3000 into a Safety Analyzer.
c. Connect the case ground lead of the analyzer to the U-blade ground of one of the
convenience receptacles. Perform the following tests with the case grounded:
• Normal polarity
• Normal polarity with open neutral
d. Perform the following tests with the case ungrounded:
• Normal polarity
• Normal polarity with open neutral
• Reverse polarity
e. Verify that the maximum leakage current does not exceed 300 µA (0.3 mA).

NOTE: Ground Resistance (between the U-blade ground on any


convenience outlet to the U-blade ground on the AC line
cord).

f. Plug the AS3000 into the safety analyzer.


g. Attach the resistance-measuring probe on the analyzer to the AS3000 U-blade
ground on any convenience outlet.
h. Invoke the resistance function on the safety analyzer, following the instructions for the
analyzer.
i. Verify that resistance to ground is less than 0.2 Ohms (200 mOhms).

30. Check the vaporizer interlock.


a. Attach two vaporizers to the Vaporizer Mounting Manifold and lock them in place.
b. Rotate either of the vaporizer’s dial to 3% agent.
c. Verify that the other vaporizer dial cannot be rotated to a setting.
d. Set both vaporizer’s dials to 0.
e. Rotate the other vaporizer dial to 3%.
f. Verify that the first vaporizer dial cannot be rotated.
g. Rotate both vaporizer dials to T and remove both vaporizers.
h. Verify that the locking spring is intact.
i. Reconnect both vaporizers to the Vaporizer Mounting Manifold

AS3000™ Service Manual 0070-10-0683 2 - 23


Installation Checkout Procedure Installation Guide

31. Check each vaporizer's agent concentration output and accuracy.


a. Insert the agent measuring device sampling tube inside the common gas outlet port.
b. Fill the Vaporizer with anesthetic agent.

NOTE: Do not overfill by filling past the indicator line on the


vaporizer.

c. Test the vaporizer accuracy per the Vapor 2000 instructions or see the appropriate
vaporizer manual for testing details.
d. Test each vaporizer in turn.
e. Test any vaporizer on the Vaporizer Storage Mount.
f. Remove the measuring device from the common gas outlet port.
g. Connect the fresh gas tube to the common gas outlet port.

32. Dräger Vapor 2000 Operating Instructions ARRB-F001.


a. Fill Vaporizer - at least half full between minimum and maximum mark.
b. Allow the filled Vapor to warm up to room temperature of 20-24C. Wait long
enough for the temperature to equalize - the time will vary depending on the
temperature differential 'T'.
• 1 hour = up top 2C
• 3 hours = ±6C
• 4 hours = ±10C
• 5 hours = ±20C
c. Check anesthetic agent monitor. Perform zero calibration of monitor with the desired
gas (AIR or O2)
d. Connect monitor to fresh gas outlet or Y-fitting. Make sure that all connections are
leak-tight.
e. Connect and start scavenging system.
f. Switch OFF ventilator or set vent pressure to less than 5 cmH2O.
g. Set monitor to anesthetic agent being used and to continuous measurement.
h. Set flow between 2.5 and 4 L/min AIR. Use O2 if AIR is not available.
i. Check 0 and T marks, 1 vol.% 4 vol % and at least three other concentrations.
j. Adjust control dial on the vaporizer.
k. Read concentration after it has reached steady state.
l. Correct measured values for the effect of carrier gas used. If AIR no correction
required. If O2 use following correction factor:
• Measured value vol.% = <1.0, correction = -0.05 vol.%
• Measured value vol.% = 1.0 - 2.0, correction = -0.1 vol.%
• Measured value vol.% = 2.5 - 4.0, correction = -0.2 vol.%
• Measured value vol.% = 5.0 - 8.0, correction = -0.3 vol.%
m. If the value displayed on the monitor is in % partial pressure, no correction is
required. If in vol.% convert to partial pressure:
• Concentration [% partial pressure] = measured value [vol.%] x atmospheric
pressure [cmH2O] / 1013 cmH2O

2 - 24 0070-10-0683 AS3000™ Service Manual


Installation Guide Installation Checkout Procedure

n. For setting 0 and T there should be no output of anesthetic agent.


• At 1 vol.%; 0.8 - 1.2 vol.% *
• At 2 vol.%; 1.8 - 2.2 vol.% *
• At 3 vol.%; 2.8 - 3.2 vol.% *
• At 4 vol.%; 3.8 - 4.2 vol.% *
• At 5 vol.%; 4.8 - 5.2 vol.% *
• At 6 vol.%; 5.7 - 6.3 vol.% *
• At 7 vol.%; 6.7 - 7.3 vol.% *
• At 8 vol.%; 7.7 - 8.3 vol.% *

* = Correct for temperature and carrier gas if necessary.


o. Switch off the vaporizer until 0 engages.
p. Switch off the AIR or O2 flow.

AS3000™ Service Manual 0070-10-0683 2 - 25


Installation Checkout Procedure Installation Guide

This page intentionally left blank.

2 - 26 0070-10-0683 AS3000™ Service Manual


3.0 Repair Information

3.1 Introduction
This chapter of the Service Manual provides the necessary technical information to perform
repairs to the system. The most important prerequisites for effective troubleshooting are a
thorough understanding of the system’s functions, as well as understanding its principles of
operation.

AS3000™ Service Manual 0070-10-0683 3-1


Warnings and Cautions Repair Information

3.2 Warnings and Cautions


In the event the instrument covers are removed, observe the following warnings and cautions:

3.2.1 Warnings
WARNING: Whenever using anesthetic gases, nitrous oxide, oxygen, or
any hospital gas always follow the appropriate agent
evacuation/collection procedures. Use the hospital gas
evacuation system.

3.2.2 Cautions
CAUTION: This device uses high pressure compressed gas. When
attaching or disconnecting backup gas cylinders, always
turn the cylinder valves slowly. Use the AS3000 flow meters
to bleed down the pressure, watching the cylinder gauge
indicate the depleting cylinder pressure, before
disconnecting the cylinder from the yoke. Always open and
close cylinder valves fully.

CAUTION: This device operates using compressed gas at high


pressures from the hospital central supply. When connecting
gas supply lines attach the hose connection to the machine
before connecting the quick disconnect fitting to the hospital
source. Disconnect the supply hose from the hospital source
connection prior to disconnecting it from the AS3000 gas
connection fittings.

3.3 Troubleshooting Guidelines


1. Identify the problem — Due to the wide variety of potential symptoms, certain
problems may be more subtle than others. Following the guidelines of the tests will help
determine the problem, if one exists.
2. Avoid shorting component leads together — During repair procedures, it can
be tempting to make a quick series of measurements. Always turn the power off before
connecting and disconnecting the test leads and probes. The accidental shorting of
leads can easily stress the components and cause a second failure (aside from the safety
risk).
3. Use the proper equipment — The equipment listed in “Special Tools Required” on
page 3-3, is suggested to fulfill a wide range of troubleshooting requirements. It is
imperative to use the designated equipment in order to ensure proper results of any and
all test procedures.
4. Clean up the repair area — After any repair, clean off the repair area.

3-2 0070-10-0683 AS3000™ Service Manual


Repair Information Special Tools Required

3.4 Special Tools Required


PART NUMBER DESCRIPTION SPECIFICATION
1 0070-00-0683 AS3000 Service Manual NA
2 not applicable Vaporizer Instruction Manual NA
3 not applicable Safety Analyzer Dempsey 430 or equivalent
4 not applicable Digital Volt Meter 3 1/2 digit
5 not applicable Agent (and N2O) Analyzer ±0.3 V/V% + 5% of reading
6 0138-00-0012 Test Lung, Adult NA
7 not applicable Digital Pressure Meter BC Biomedical DPM-2301751
NMC Digital Pressure Meter
8 not applicable Central supplied O2, N2O, AIR Minimum of 35 psi, DISS
connections.
9 not applicable Cylinder gases O2, N2O, AIR Full PISS yoke connections.
10 not applicable Hand tools, Allen wrench set Metric
11 0367-00-0080 Tank Wrench NA
12 0103-00-0508 Y-Fitting 15 mm connection
13 0004-00-0076 Respiration Tube (2 required) 0.6 meter silicone, 15 mm
14 0992-00-0139 Breathing Bag 2.3 L silicone
15 0510-00-0020 Krytox Lubricant NA
16 not applicable Vent Tester Certifier-FA PLUS Respiratory
Vent.Tester 4080
17 0997-00-0641 AS3000 Calibration hose NA
18 0103-00-0691 AS3000 test plug NA
19 0040-00-0448 Plug Kit NA
20 0453-00-1216 Regulator Calibration Hose NA

AS3000™ Service Manual 0070-10-0683 3-3


3.5 Troubleshooting Chart
3-4

Troubleshooting Chart
The following table shows common symptoms and corrective actions for problems when troubleshooting the AS3000. The information given
indicates failures during startup and runtime.

3.5.1 Common Symptoms and Corrective Actions for Field Service Technicians

MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION


“BDU Communication . . . Startup test 1 The power supply switch is turned to ON, 1 Turn the power supply switch to OFF and
FAIL” immediately after turned to OFF, which power on again after waiting for a while (1 - 2
makes the time too short for the BDU control seconds).
board power to reboot.
2 The data wire is disconnected or damaged. 2 Check whether the data wire (signal wire at
backside of display) is disconnected or is
damaged.
3 The BDU control board is defective. 3 Replace the BDU control board and perform
each item of calibration and startup test again.
“Software Version . . . FAIL” Startup test 1 Software versions of GUI, BDU and 1 Update software version.
0070-10-0683

Keyboard are incompatible.


2 The power supply switch was turned ON, 2 Turn the power supply switch to OFF, wait 1 -2
and OFF quickly. seconds, then power on again.
3 The data wire is disconnected or damaged. 3 Check whether the data wire (signal wire at
backside of display) is disconnected or is
damaged.
4 The BDU control board is defective. 4 Replace the BDU control board and perform
each item of calibration and startup test again.
5 The keyboard board is defective. 5 Replace the user interface assembly.
“BDU EEPROM Data . . . Startup test 1 Data stored in EEPROM on BDU control 1 Startup. Then perform each item of the
FAIL” board is lost (checksum error) calibration and startup test again.
2 The BDU is defective 2 Replace the BDU board
“EEPROM IC . . . FAIL” Startup test 1 The BDU control board is defective or its 1 Check that the EEPROM (BDU U12) is
EEPROM is damaged. connected correctly, or Replace the EEPROM
and BDU control board, and perform each
item of the calibration and startup test again.
AS3000™ Service Manual

“BDU WDT . . . FAIL” Startup test 1 The Watch Dog Timer in the digital signal 1 Replace the BDU control board, and perform
processor on the BDU control board is each item of calibration and startup test again.

Repair Information
damaged.
“AD/DA . . . FAIL” Startup test 1 The AD or 4052 in the digital signal 1 Replace the BDU control board, and perform
processor on BDU control board is each item of the calibration and startup test
damaged. again.
AS3000™ Service Manual

Repair Information
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION
“Vent/Manual Valve Failure” Startup test 1 The wire that connects to the vent/manual 1 Check whether the connection wire of the
valve is disconnected or is damaged. vent/manual valve is disconnected or is
damaged.
2 The vent/manual solenoid valve is 2 Replace the vent/manual solenoid valve.
damaged.
3 The amplifier board or connection cable 3 Check the connection cable and replace the
defective. amplifier board. Then perform each item of the
calibration and startup tests again.
“Expiration Sensor Failure” Startup test 1 Fresh gas flow is present at startup. 1 Turn off all fresh gas flow and reboot the unit.

2 The sensor board is defective. 2 Replace the sensor board, and then perform
each item of the calibration and startup test
again.
“Keyboard Communication . Startup test 1 The data wire is disconnected or is 1 Replace the user interface assembly.
. . FAIL” damaged.
2 The keyboard board is defective. 2 Replace the user interface assembly.
“Pressure Sensor . . . FAIL” Startup test 1 The connection wire of the PAW board is 1 Check whether the connection wire of the PAW
disconnected or is damaged. board is disconnected or damaged.
0070-10-0683

2 The connection of the PAW board or sensor 2 Check the power supply and connection of the
board is defective. sensor board. Replace the Paw board and
sensor board.
3 The tubing kinked or occluded. 3 Check the tubing for kinks and remove any
occlusions.
“O2 Sensor . . . FAIL” Startup test 1 The O2 concentration is below 16%. 1 Increase O2 concentration.
2 The oxygen sensor is out of range. 2 Replace the oxygen sensor.
3 The main unit-oxygen sensor cable is 3 Check whether the main unit-oxygen sensor
disconnected or damaged. cable is disconnected or damaged.
4 The amplifier board is defective. 4 Replace the amplifier board, and then perform
each item of the calibration and startup test
again.
“PEEP Valve . . . FAIL” Startup test 1 The connection wire of the PEEP valve is 1 Check whether the PEEP valve is disconnected
disconnected or damaged. or damaged.
2 The PEEP valve is defective. 2 Replace the PEEP valve.
3 The Amplifier board is defective. 3 Replace the amplifier board and then perform
each item of the calibration and startup test

Troubleshooting Chart
again.
“Inspiration Valve . . . FAIL” Startup test 1 The connection wire of the inspiration valve 1 Check whether the inspiration valve is
is disconnected or damaged. disconnected or damaged.
2 The inspiration valve is defective. 2 Replace the inspiration valve.
3 The amplifier board is defective. 3 Replace the amplifier board and then perform
each item of the calibration and startup test
3-5

again.
3-6

Troubleshooting Chart
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION
“Inspiration Sensor . . . FAIL” Startup test 1 Fresh gas is flowing during the startup 1 Verify that no fresh gas is flowing during the
or procedure. startup procedure.
“Inspiration Sensor . . . FAIL 2 The tubing kinked or occluded. 2 Check the tubing for kinks and remove any
Try re-starting the system with occlusions.
all gas flows OFF” (UI 3 The sensor board is defective. 3 Replace the sensor board and then perform
version 2.25 and higher) each item of the calibration and startup test
again.
“O2 Supply . . . FAIL” Startup test 1 The gas source is not connected. 1 Check gas source.
Runtime 2 The gas supply pressure switch or its cable 2 Check and replace the switch or cable of the
is damaged. gas source.
3 The amplifier board is defective. 3 Replace the amplifier board, and then perform
each item of the calibration and startup test
again.
“AC Power Failure” Startup test 1 AC power is not connected. 1 Check the connection of AC power.
Runtime 2 A fuse is tripped 2 Replace the fuse.
“Low Battery” Runtime 1 Battery capacity is low. 1 Charge the battery by connecting AC power.
2 The battery is defective. 2 Replace the battery.
3 The connection wire or power board is 3 Check the connection wire or replace the
0070-10-0683

defective. power board.


“Alarm Speaker . . . FAIL” Startup test 1 The connection wire to the speaker is 1 Check whether the connection wire of the
disconnected or damaged. speaker is disconnected or damaged.
2 The speaker is damaged. 2 Replace the user interface assembly.
3 The keyboard board is defective. 3 Replace the user interface assembly.
“Software Mismatch” Startup test 1 Power is recycled to the unit too quickly, not 1 Shut down the unit and wait until the backlight
or allowing the memory to fully clear. is fully extinguished before restoring power.
“Software Mismatch or
Failure to Shut Down
Completely” (UI version 2.25
and higher)
Charging state is always Runtime 1 Battery capacity is low. 1 Charge the battery by connecting AC power.
displayed. 2 The connection wire to the battery is 2 Check whether the connection wire to the
disconnected or damaged. battery is disconnected or damaged.
3 The battery is defective. 3 Replace the battery.
AS3000™ Service Manual

During startup leak test, the Startup test 1 A solenoid failure or wiring error. 1 Correct the wiring or replace the solenoid
bellows moves when it valve.
shouldn’t

Repair Information
In leakage detection, The APL Startup test 1 The Breathing System pneumatic hose leaks 1 Replace the Breathing System pneumatic hose
Valve detection pressure gas when in manual mode.
displayed on the pressure 2 Pressure gauge failure. 2 Replace the pressure gauge
gauge does not stay within 3 The APL Valve is out of spec 3 Replace the APL Valve
28 - 32 cmH2O.
AS3000™ Service Manual

Repair Information
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION
In leakage detection, Safety Startup test 1 The proportional valve on the airway 1 Calibrate the inspiration valve again.
valve detection failed. module failed.
2 The safety valve in the airway module 2 Replace the airway module.
failed.
In leakage detection, the Startup Test 1 The tubing connection is not tight. 1 Check the connection on the tubing.
leakage volume is 501 - 2 Incorrect installation of the absorber 2 Check the installation of the absorber canister.
999mL. canister.
3 The pressure gauge or oxygen sensor on the 3 Check the installation of the pressure gauge or
circuit is not installed tightly. oxygen sensor on circuit.
4 There is a leakage in the fresh gas supply. 4 Check the fresh gas supply for leakage
In leakage detection, the Startup Test 1 The tubing connection is not tight. 1 Check the connection on the tubing.
leakage volume is larger than 2 Incorrect installation of the absorber 2 Check the installation of the absorber canister.
1L. canister.
3 The pressure gauge or oxygen sensor on the 3 Check the installation of the pressure gauge or
circuit is not installed tightly. oxygen sensor on circuit.
4 There is a leakage in the fresh gas supply. 4 Check the fresh gas supply for leakage
In compliance test, Startup Test 1 The Y-fitting is not sealed or the Breathing 1 Check the connection and leakage of
compliance test failed. System has leakage. respiration tubes.
0070-10-0683

2 The respiration tubes are too long. 2 The respiration tubes cannot be too long.
3 Calibration of the pressure sensor is 3 Calibrate the pressure oxygen again.
inaccurate.
4 The flow meter is set incorrectly 4 Set the flow meter to the correct value
5 The calibration of the flow meter is 5 Recalibrate the flow meter
inaccurate
The power supply switch on All 1 The power supply switch on the main unit or 1 Check the connection cable of the power
the main unit is defective. the cable connection failed. supply switch of the main unit, or replace the
power supply switch of main unit.
2 Communication board is defective. 2 Replace the communication board.
All or some keys are non All 1 The keyboard is defective. 1 Replace the user interface assembly.
responsive. 2 The communication board is defective. 2 Replace the user interface assembly.
The encoder is non All 1 The encoder failed. 1 Replace the user interface assembly.
responsive. 2 The communication board is defective. 2 Check and replace the user interface assembly.
All or some of the work lights All 1 The work light circuit board is defective. 1 Replace the work light circuit board.
are defective. 2 The cable is defective 2 Check the connection to the power supply.

Troubleshooting Chart
The switch on the work light All 1 The switch on the work light is defective. 1 Replace the work light switch.
is defective.
The flow meter light is non All 1 The power supply to the light is defective 1 Replace the power supply to the light.
operational. 2 The foil of light is defective. 2 Replace the foil of light.
3 The inverter is defective. 3 Replace the inverter
3-7
3-8

Troubleshooting Chart
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION
Heating of circuit is non Runtime 1 Line power is not connected. 1 Check the line power connection.
functional. 2 The circuit cable is defective. 2 Replace the cable connection to the circuit.
3 Heating control board is defective. 3 Replace the heating control board.
AC Outlets are non All 1 A fuse is tripped 1 Replace the fuse
functional
The O2 reading is out of Runtime 1 The calibration of the oxygen sensor is 1 Calibrate the oxygen sensor again, or replace
tolerance inaccurate, or the oxygen sensor is the oxygen sensor.
defective.
The display screen has a All 1 The display screen is defective. 1 Replace the user interface assembly.
hotspot or uneven brightness 2 PC104 is defective. 2 Replace the user interface assembly.
The display screen is a solid Startup Test 1 The GUI program has crashed. 1 Reboot the system
color Runtime 2 The signal cable is disconnected. 2 Check the signal cable.
3 PC104 is defective. 3 Replace the user interface assembly.
4 The communication board or power supply 4 Replace the user interface assembly.
for the light is defective.
5 The display screen is damaged. 5 Replace the user interface assembly.
The pressure reading is out of Runtime 1 Calibration of the pressure sensor is out of 1 Calibrate the pressure sensor again.
0070-10-0683

tolerance spec.
The PEEP reading is out of Runtime 1 Calibration of expiration valve (PEEP) is 1 Calibrate the expiration valve (PEEP).
tolerance inaccurate.
2 Calibration of pressure sensor is inaccurate. 2 Calibrate the pressure sensor.
the tidal volume reading is Runtime 1 The compliance test after startup is 1 Perform the compliance test after startup again.
out of tolerance inaccurate.
2 Calibration of the flow sensor is inaccurate. 2 Calibrate the flow sensor again.
3 Calibration of the inspiration valve is 3 Calibrate the inspiration valve again.
inaccurate.
4 Calibration of the pressure sensor is 4 Calibrate the pressure sensor again.
inaccurate.
“System Self Test” Startup Test 1 Self-test during each startup. 1 No action needed.
“NON-FUNCTIONAL Call Startup Test 1 A failure has been found (Displayed on the 1 Handle it according to specific information on
Service Representative” screen), which leads to the abnormal the failure.
operation of anesthesia machine.
AS3000™ Service Manual

After the startup test, only the Startup Test 1 A failure has been found (Displayed on the 1 Handle it according to specific information on
MANUAL/AUTO key is screen), which leads to the abnormal the failure.
displayed, with the alarm: operation of anesthesia machine.

Repair Information
(except O2 supply fail)
After the startup test, there Startup Test 1 A failure has been found (Displayed on the 1 Handle it according to specific information on
are Bypass and Continue screen), but it does not affect the normal the failure.
keys, but with an alarm operation of each ventilation mode.
displayed
AS3000™ Service Manual

Repair Information
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION
“Ventilator setting is not Runtime 1 A parameter has been set to a limit that can 1 The parameter set cannot exceed the
possible!” be set. adjustable range. If the limit is not reached,
another parameter may be limiting the range
and needs to be adjusted accordingly
“Displayed parameter values Runtime 1 The AS3000 is in the pre-selection mode, 1 Set parameters 15 seconds in advance; or
are for the pending mode” but it is still working according to the directly confirm and enter ventilation mode; or
original operating mode. press the NORMAL SCREEN button to return
to the original ventilation mode interface; or do
not perform any operation and it will
automatically return to the original ventilation
mode interface after 15 seconds.
“Pressure, Volume and Runtime 1 The functional alarm in Manual mode is 1 Enable the functional alarm in Manual mode.
Apnea Alarms are OFF!” disabled.
“Automatic Ventilation not Startup Test 1 A failure has been found (Displayed on the 1 Handle it according to specific information on
available!” Runtime screen); the AS3000 can work only in the failure.
Standby/Manual mode. This information is
displayed after entering the Main Screen.
0070-10-0683

“BDU Communication Startup Test 1 The communication cable is disconnected. 1 Check the communication cable.
Failure!” Runtime 2 The BDU software has crashed or reset. 2 Power on again.
“O2 Sensor failure” Runtime 1 The main unit-oxygen sensor cable is 1 Check the unit-oxygen sensor cable.
disconnected.
“O2 Supply failure” Runtime 1 The gas source is closed or disconnected. 1 Check the gas source.
Continuous Pressure Runtime 1 Airway pressure is too high. 1 Relieve the airway pressure
2 The expiration valve (PEEP) is defective. 2 Repair/replace the expiration valve
3 The airway is occluded. 3 Clear the occlusion
High Airway Pressure Runtime 1 Airway pressure is high. 1 Adjust the upper limit set for the airway
pressure alarm.
Low Airway Pressure Runtime 1 Airway pressure is low. 1 Adjust the lower limit set for the airway
pressure alarm.
Negative Pressure Runtime 1 Airway pressure is continuously negative. 1 Check the airway and fresh gas setting
Low FiO2 Runtime 1 Oxygen concentration is lower than the 1 Adjust the lower limit set for the Oxygen
lower limit set for Oxygen Concentration Concentration alarm or the actual oxygen
alarm. concentration.

Troubleshooting Chart
2 Oxygen calibration is inaccurate. 2 Calibrate the oxygen concentration sensor
again.
3-9
3 - 10

Troubleshooting Chart
MESSAGE/ISSUE OCCURRENCE CAUSE SOLUTION
High FiO2 Runtime 1 Oxygen concentration is higher than the 1 Adjust the upper limit set for the Oxygen
upper limit set for the Oxygen Concentration Concentration alarm or the actual oxygen
alarm. concentration.
2 Oxygen calibration is inaccurate. 2 Calibrate the oxygen concentration sensor
again.
APNEA Runtime 1 Breathing hoses are disconnected. 1 Check the breathing hoses.
Low Battery Runtime 1 Battery capacity is low. 1 Connect the unit to AC power.
High Minute Volume Runtime 1 Minute ventilation is higher than the upper 1 Adjust the upper limit set for the Minute
limit set for the Oxygen Concentration Ventilation alarm.
alarm.
Low Minute Volume Runtime 1 Minute ventilation is lower than the lower 1 Adjust the lower limit set for the Minute Volume
limit set for the Minute Volume alarm. alarm.
APNEA Backup Runtime 1 PS mode is not triggered and APNEA 1 No action needed.
backup ventilation is given according to the
minimum breathing frequency that has been
set.
High PEEP Runtime 1 The expiration time is too short or expiration 1 Check the time setting and expiration pipeline.
0070-10-0683

limb is occluded 2 Check the expiration valve (PEEP).


2 The expiration valve (PEEP) is defective.
Keyboard communication Startup Test 1 The signal wire is disconnected or 1 Check the signal wire.
Failure Runtime damaged. 2 Power on again.
2 The keyboard board program crashed.
AC Power Failure All 1 AC power is not connected or a fuse is 1 Connect AC and check the fuses.
tripped.
O2 Cal Due Startup Test 1 The oxygen sensor has not been calibrated 1 Calibrate the oxygen sensor.
Runtime for more than 3 days.
Tidal volume displayed by Runtime 1 This caused by the compliance and 1 No action needed.
the AS3000 is smaller than compensation of resisting force between the
the value displayed on airway and patient.
bellow.
Bellows is deflated and at the Runtime 1 This is caused by a disconnected breathing 1 Install a breathing circuit and connect the Y-
AS3000™ Service Manual

bottom of its travel during circuit, a removed absorber canister, a piece to the Test port. Select CMV mode and
runtime, after initial power-up removed O2 sensor or sensor adapter, confirm the selection. Immediately push the O2
and Leak Test. check valve dome rings, or no fresh gas Flush button until the bellows is completely

Repair Information
flow while ventilator is running. filled. Return to STANDBY mode and remove Y-
piece from Test port.
Repair Information Leak Troubleshooting

3.6 Leak Troubleshooting


Manual No Leaks
Leak Test PASS Leak Test PASS Detected

FAIL FAIL

Seal the
Manual scavenger port Check APL
Leak Test PASS PASS Valve
and repeat the
test

Check the
FAIL FAIL Manual bag
and bag arm

Check the
Check the Check the Check the PEEP
Bellows Dome Exhaust valve valve Proportional
Valve

Pinch the CGO Remove the Check the


Hose and Vaporizers vaporizers and
PASS PASS
repeat the Leak and repeat the the vaporizer
test Leak test mount

Check the
Flowmeter
FAIL FAIL
module /
CGO port

Check and if Check Check Verify the I/E Rings Check the
Remove
sampling necessary the O2 the are tight and not PAW
exchange the cell Absorber cracked (do not gauge and
line
breathing hoses o-ring assembly over tighten!) its o-rings

FIGURE 3-1 Leak Troubleshooting Flow Chart

AS3000™ Service Manual 0070-10-0683 3 - 11


3.7 Test Pneumatics
3 - 12

Test Pneumatics
The following section shows the pneumatic components involved while performing the AS3000 startup tests.

3.7.1 Leak Test - Manual Ventilation Test


P
Exp. Valve

Exp. Flow
APL Valve
Sensor

Breathing Bag
Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
0070-10-0683

Proportional
Valve
Fresh Gas

Pressure AIR• O2
Regulator
Flow Flow
Meter Meter

O2
Outlet

Pressure
O2 Switch
Flush

P
Cylinder

O2
O
Fresh Gas Flow
Pressure
Flow Sensor

Tubes Inlet Filter

P
Vaporizer
Vaporizer
Regulator

P
V
Cylinder
Gas Inlet

AIR
Block

O
AS3000™ Service Manual

Pressure Inlet Filter

P
Regulator

P
Cylinder

N20

Repair Information
Ratio•

O
Controller
N20• Pressure Inlet Filter

P
Cuttoff Regulator

FIGURE 3-2 Manual Ventilation Test Pneumatics


3.7.2 Safety Valve Test
AS3000™ Service Manual

Repair Information
P
Exp. Valve

Exp. Flow
APL Valve
Sensor

Breathing Bag
Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve

Fresh Gas
Pressure AIR•
Regulator
O2
Flow Flow
Meter Meter
0070-10-0683

O2
Outlet

Pressure
O2 Switch
Flush

P
Cylinder

O2
O
Fresh Gas Flow
Pressure
Flow Sensor

Tubes Inlet Filter

P
Vaporizer
Vaporizer
Regulator

P
V
Cylinder
Gas Inlet

AIR
Block

O
Pressure Inlet Filter

P
Regulator

P
Cylinder

N20
Ratio•

O
Controller
N20• Pressure Inlet Filter

P
Cuttoff Regulator

FIGURE 3-3 Safety Valve Test Pneumatics

Test Pneumatics
3 - 13
3.7.3 Leak Test - Automatic Ventilation Test
3 - 14

Test Pneumatics
P
Exp. Valve

Exp. Flow
APL Valve
Sensor

Breathing Bag
Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve

Fresh Gas
Pressure AIR•
Regulator
O2
Flow Flow
Meter Meter
0070-10-0683

O2
Outlet

Pressure
O2 Switch
Flush

P
Cylinder

O2
O
Fresh Gas Flow
Pressure
Flow Sensor

Tubes Inlet Filter

P
Vaporizer
Vaporizer
Regulator

P
V
Cylinder
Gas Inlet

AIR
Block

O
Pressure Inlet Filter

P
Regulator

P
Cylinder

N20
Ratio•

O
AS3000™ Service Manual

Controller
N20• Pressure Inlet Filter

P
Cuttoff Regulator

Repair Information
FIGURE 3-4 Automatic Ventilation Test Pneumatics
3.7.4 Compliance Test
AS3000™ Service Manual

Repair Information
P
Exp. Valve

Exp. Flow
APL Valve
Sensor

Breathing Bag
Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

Electronic PEEP
Control Valve
Solenoid
Valve

Ventilator
Proportional
Valve

Fresh Gas
Pressure AIR• O2
Regulator
Flow Flow
Meter Meter
0070-10-0683

O2
Outlet

Pressure
O2 Switch
Flush

P
Cylinder

O2
O
Fresh Gas Flow
Pressure
Flow Sensor

Tubes Inlet Filter

P
Vaporizer
Vaporizer
Regulator

P
V
Cylinder
Gas Inlet

AIR
Block

O
Pressure Inlet Filter

P
Regulator

P
Cylinder

N20
Ratio•

O
Controller
N20• Pressure Inlet Filter

P
Cuttoff Regulator

FIGURE 3-5 Compliance Test Pneumatics

Test Pneumatics
3 - 15
3.8 Pneumatic Hose and Wiring Diagrams
3 - 16

Pneumatic Hose and Wiring Diagrams


3.8.1 Pneumatic Hose Labeling
P
Exp. Valve

Exp. Flow
APL Valve
Sensor

Breathing Bag
Absorber Exhaust
Valve
Patient Canister

Bellows

Reversing
Valve AGSS
Breathing
O2 Sensor
Insp. Valve System
Insp. Flow
Sensor

8•
9
10
11
29 12 1 7
28
Electronic PEEP
Control
37 33 Valve
Solenoid
Valve

Ventilator
Proportional
Valve 3
0070-10-0683

34 31
30 32
Fresh Gas

36
38 Pressure AIR• O2
Regulator 20
Flow Flow
Meter Meter
17
O2
Outlet
5 6
15 Pressure
O2 21 16 Switch
Flush

P
14 41
Cylinder
4

O2
18
22

O
Fresh Gas Flow
Pressure 13
Flow Sensor

42 Tubes Inlet Filter

P
Vaporizer
Vaporizer
Regulator

P
24 40
V
Cylinder
Gas Inlet

AIR
19
Block

O
Pressure Inlet Filter 25

P
Regulator
AS3000™ Service Manual

P
39
Cylinder

N20
Ratio•23

Repair Information
O
Controller
N20• Pressure Inlet Filter 27

P
Cuttoff Regulator 26 2

FIGURE 3-6 Pneumatic Hose Labeling Diagram


3.8.2 Sampling Pipeline Module Interface Labeling
AS3000™ Service Manual

Repair Information
11 Insp. High 12 Manual/Auto
Switch
9 Exp. Low 10 Insp. Low

7 PEEP Control 8 Exp. High

12 Manual/Auto
Switch
0070-10-0683

10 Insp. Low
11 Insp. High

8 Exp. High

Pneumatic Hose and Wiring Diagrams


7 PEEP Control 9 Exp. Low

FIGURE 3-7 Sampling Pipeline Module Interface Diagram


3 - 17
3.8.3 Electrical Cable Labeling
3 - 18

Pneumatic Hose and Wiring Diagrams


GAS UINT UI

INVERTER LCD
SW2 20
TopLight-Switch TopLight Sensor
5 J1 J2
4 J11
PC104
J15 LCD
30
BackLight J13

J10

J14

J12
15 KeyBoard

J6

J1

J2

J4
J7
22 Speaker J5 SPK
23 PWR

J2
J6 J7 PWR
24 LPT 232 ISOLATOR
FSU3 LJ3 J5 PRN
FSfm DB25 25 COM1 27 COMA

12
14
43
42
16
17
18
COM1 COM1 COM4 COM1
32 J1 26 COM2 COM3 COM2
COM2 COM2
31 COMB
Mains input LJ3S 44
Mains Auxiliary SW-GP IV1 Vinsp J9 J3
Speaker

J4

J8
S1 RL1
21

S2
Paw IV2 Vexp
S3 J1
Encoder
28
S4
Pair LV1 man/cmv
J1
KeyPad
29

35
0070-10-0683

Power
AC/DC
LJ4
36 2 J3 10 19
LJ4S Amplifier

J2
8 J7
BDU
BDU

J30

J5
J6
9 J20
BAT LJ1
SLA-24V-3

S4
40 3 J4 J10

SW1
1 J5

J8 J1

LJ6 LJ2 HeaterCtrl


Heater
Breathing System

38 7 J3 J2
6

LJ6S LJ2S SCI


11

O2-Sensor LJ5 CONTROL UINT


39 13

LJ5S

FIGURE 3-8 Electrical Cable Labeling Diagram


AS3000™ Service Manual

Repair Information
4.0 Replacement Parts and Accessories

4.1 Introduction
This section of the manual provides information that is necessary to identify replacement
parts and accessories.

4.2 Available Replacement Parts and Sub-Assemblies


The following part lists are divided into two sections. The Isometric Drawings and the
accompanying part lists identify the available chassis mounted components.

4.2.1 Exchange Program


Mindray DS USA, Inc. offers an exchange policy for many of the printed circuit board
assemblies. This program may provide the most expedient method of servicing the
equipment. A standard charge is associated with this service. Contact the Service
Department for details concerning this exchange program.

Many circuit boards make extensive use of multi-layer and surface mount technology.
Individual component replacement is not recommended on these boards. Board exchange or
replacement is the most efficient method of repair for these types of assemblies. Component
level repair is not recommended.

Circuit boards, returned as parts of the exchange program, that show evidence of improper
repair techniques are not considered for exchange. Damaged boards will be invoiced at full
value and no exchange credit will be applied.

4.2.2 Replacement Parts Pricing Information


Current parts prices and exchange charges can be determined by contacting the Customer
Support Department.

AS3000™ Service Manual 0070-10-0683 4-1


Available Replacement Parts and Sub-Assemblies Replacement Parts and Accessories

4.2.3 Ordering Information


Replacement parts and assemblies are available. Please follow these guidelines when
ordering replacement items for the product:

1. Include the Model and Serial Number of the product.


2. Include the Part Number exactly as it appears in the Parts List under the column, “Part
Number.”
3. Include a description of the item.
EXAMPLE ORDERS: (1) ea. P/N 0334-00-2611-01
Label, Fuse Replacement, Serial No. XXXX
(2) ea. P/N 0213-07-0404
Screw, Self Tap, #4 x 0.25", Serial No. XXXX

NOTE: Mindray DS USA, Inc. maintains a policy of continuous


development for product improvement and reserves the
right to change materials, specifications, and prices without
notice.

4-2 0070-10-0683 AS3000™ Service Manual


Replacement Parts and Accessories Isometric Drawings

4.3 Isometric Drawings


4.3.1 Chassis

17

16

15

14

13
5
12

11

10

FIGURE 4-1 AS3000 Frontview

AS3000™ Service Manual 0070-10-0683 4-3


Isometric Drawings Replacement Parts and Accessories

18

19
31

20 30

29

21

28

27
22
26

25

23

24

FIGURE 4-2 AS3000 Rearview

4-4 0070-10-0683 AS3000™ Service Manual


Replacement Parts and Accessories Isometric Drawings

4.3.1.1 Chassis Parts List

FIG. NO. DESCRIPTION PART NUMBER


1 Arm, Monitor 0436-00-0248
2 User Interface Assembly 0997-00-0620 (M31)
021-000026-00 (M32)
115-005982-00 (M33)
3 Aux Gas Assembly 0997-00-0634
4 Channel, Left Side 0436-00-0250
5 Breathing System 115-005967-00
6 Absorber Canister Assembly 0997-00-0629 (original)
115-005724-00 (revised)
7 O2 Flush Valve Assembly 0104-00-0064
8 Scavenger System, AS3000 0992-00-0279
9 Drawer Assembly 0441-00-0206
10 Gauge, N2O, Line Pressure 0118-00-0039
11 Gauge, AIR, Line Pressure 0118-00-0040
12 Gauge, O2, Line Pressure 0118-00-0041
13 Gauge, N2O High Pressure 0118-00-0042
14 Gauge, AIR High Pressure 0118-00-0036
15 Gauge, O2 High Pressure 0118-00-0037
16 Switch, Mains (external) 0261-00-0220
Switch, Mains (internal) 0261-00-0218
17 Vaporizer Mount 0997-00-0616
18 Rail, Right Side 0436-00-0251
19 Fuse, 2A, 250V 0159-00-0057
20 Vaporizer Storage Mount 0436-00-0221
21 Fuse, 10A, 250V 0159-00-0058
22 Lock, Key 0366-60-0137
23 Caster, without Brake 0401-00-0054
24 Caster, with Brake 0401-00-0053
25 Tank Wrench Assembly 0367-00-0080
26 Yoke, O2 0103-00-0676
27 Yoke, AIR 0103-00-0677
28 Yoke, N2O 0103-00-0678
29 User Interface Cable 0012-00-1778
30 User Interface Mounting Arm 0436-00-0249
31 Gas Input Assembly 115-006314-00
NS Kit Check Valve 0040-00-0458
NS Not Shown

AS3000™ Service Manual 0070-10-0683 4-5


Isometric Drawings Replacement Parts and Accessories

4.3.2 Breathing System

32

33

34

35

36

37

FIGURE 4-3 Breathing System, Top View

4-6 0070-10-0683 AS3000™ Service Manual


Replacement Parts and Accessories Isometric Drawings

38

39

40

41

42

43

FIGURE 4-4 Breathing System Bottom View

AS3000™ Service Manual 0070-10-0683 4-7


Isometric Drawings Replacement Parts and Accessories

44

53

54

45

46

47
55
48

49

50

51

52

FIGURE 4-5 Breathing System Exploded Top View

4-8 0070-10-0683 AS3000™ Service Manual


Replacement Parts and Accessories Isometric Drawings

56

57

FIGURE 4-6 Breathing System Exploded Bottom View

AS3000™ Service Manual 0070-10-0683 4-9


Isometric Drawings Replacement Parts and Accessories

58

59

60

61

FIGURE 4-7 Absorber Canister Exploded View

4 - 10 0070-10-0683 AS3000™ Service Manual


Replacement Parts and Accessories Isometric Drawings

4.3.2.1 Breathing System Parts List

FIG. NO. DESCRIPTION PART NUMBER


32 Common Gas Outlet Assembly 0997-00-0633
33 O2 Sensor 0600-00-0149
34 Airway Pressure Gauge 0118-00-0038
35 Bag Arm Tall 0009-00-0013
36 APL Valve 0104-00-0065
37 Absorber Canister Assembly 0997-00-0629 (original)
115-005724-00 (revised)
38 Breathing System Pneumatics Hose 082-000387-00
39 Breathing System Support Arm 0436-00-0253
40 O2 Sensor cable 0012-00-1775
41 Cable, Heater 0012-00-1780
42 Drive Gas Hose 0004-00-0098
43 AGSS Transfer Hose 0004-00-0095
44 O2 Sensor Interface 0198-00-0080
45 O Ring, 6 x 1.8 0354-00-0188
46 O Ring, 4 x 1.8 0354-00-0189
47 Ring, I/E Valve 0219-00-0018 (plastic)*
0219-00-0020 (metal)*
48 Cover, I/E Valves 0352-00-0068
49 Gasket, I/E Valves 0354-00-0192
50 Frame, I/E Valve 0352-00-0069
51 Disc, Ceramic I/E Valve 0354-00-0182
52 O2 Port Plug 0198-00-0083
53 Cover, Bellows 0198-00-0079
54 Bellows Assembly 115-005968-00
55 Gasket, Bellows Canister base 0354-00-0191
56 Absorber Hose 0004-00-0097
57 Absorber Mount Assembly 0436-00-0256
58 Absorber Gasket 0354-00-0199 (original)
047-002849-00 (revised)
59 Canister, Soda Lime 0352-00-0070 (original)
045-000254-00 (revised)
60 Absorber Screen, Metal 0378-00-0073
61 Absorber Canister Gasket 0354-00-0183 (original)
045-000252-00 (revised)
047-002850-00 (revised)
NS Bag Arm Short 0009-00-0012
NS Bellows Support Cage 082-000389-00
NS Alignment Bracket Kit 115-007641-00
NS Not Shown
* Parts are not one-to-one interchangeable.

AS3000™ Service Manual 0070-10-0683 4 - 11


Isometric Drawings Replacement Parts and Accessories

4.3.3 Electric Box


68
62

67

63
66

64
65

FIGURE 4-8 Electric Box Assembly

4.3.3.1 Electric Box Parts List

FIG. NO. DESCRIPTION PART NUMBER


62 PC Board Assembly, Power 0671-00-0264 (M31)
051-000484-00 (M32)
051-000484-00 (M33)
63 Power Supply, SNP-B209, Switching 0014-00-0091
64 PC Board Assembly, Breathing System Heater 0671-00-0107
65 Cable, Amp PC Board to Sensor PC Board 0012-00-1777
66 PC Board Assembly, Amplifier 0671-00-0263
67 PC Board Assembly, BDU 0671-00-0262 (M31)
051-000517-00 (M32)
051-000493--00 (M33)
68 Battery Assembly SLA-5AH24V-3 0997-00-0619

4 - 12 0070-10-0683 AS3000™ Service Manual


Replacement Parts and Accessories Isometric Drawings

4.3.4 Gas Circuit Box


73

69
72

71

70

FIGURE 4-9 Gas Circuit Box Assembly

4.3.4.1 Gas Circuit Box Parts List

FIG. NO. DESCRIPTION PART NUMBER


69 Electronic Flow Meter Assembly 0154-00-0010
70 PC Board Assembly, Sensor 0671-00-0265
71 PEEP Valve 0104-00-0062
72 Gas Drive Module Assembly 0997-00-0632
73 Valve, Pressure Reducer 0103-00-0681
NS Exhaust Gas Outlet Port 082-000681-00
NS Tube, Green, 6 mm OD, 4 mm ID 0008-00-0338
NS Tube, Green, 8 mm OD, 6 mm ID 0008-00-0337
NS Tube, Yellow, 6 mm OD, 4 mm ID 0008-00-0356
NS Tube, Blue, 6 mm OD, 4 mm ID 0008-00-0340
NS Tube, Black, 8 mm OD, 6 mm ID 0008-00-0358-15
NS Tube, Silicone, 20 mm OD, 15 mm ID 0008-00-0371
NS Tube, Silicone, 5 mm OD, 2 mm ID 0008-00-0370
NS Tube, Silicone, 8 mm OD, 4 mm ID 0008-00-0369
NS Not Shown

AS3000™ Service Manual 0070-10-0683 4 - 13


Isometric Drawings Replacement Parts and Accessories

4.3.5 User Interface


74

79

75 78

76

77

FIGURE 4-10 User Interface Assembly

80

82

83

81

FIGURE 4-11 User Interface Rearview

4 - 14 0070-10-0683 AS3000™ Service Manual


Replacement Parts and Accessories Isometric Drawings

4.3.5.1 User Interface Parts List

FIG. NO. DESCRIPTION PART NUMBER


74 LCD Display 0160-00-0119
75 Keypad Overlay 0331-00-0139
76 Knob, Navigator™ 0366-00-0136
77 PC Board RS232 0671-00-0111
78 Cable, Comm PC Board to Display Panel 0012-00-1774
79 Speaker Assembly 0119-00-0230
80 Inverter, LCD 0671-00-0112
81 Cable, PC104 PC Board to Display 0012-00-1776
82 Communications Board 0671-00-0109 (M31)
051-000483-00 (M32, M33)
83 PC104 Board 0671-00-0276 (M31)
051-000518-00 (M32)
050-000359-00 (M33)
PC104 Board (old) 0671-00-0110 (M31)
N/A (M32)
051-000564-00 (M33)
— Display Assembly 0997-00-0620 (M31)
021-000026-00 (M32)
115-005982-00 (M33)

AS3000™ Service Manual 0070-10-0683 4 - 15


Isometric Drawings Replacement Parts and Accessories

4.3.6 Flowmeter

84

85

86

87

88

89

90

FIGURE 4-12 Flowmeter Module

4.3.6.1 Flowmeter Parts List

FIG. NO. DESCRIPTION PART NUMBER


84 Flow Meter FM400 Assembly 0997-00-0617 (for -01 units)
115-005969-00 (for -02 units)
85 Label, Flowmeter N2O 0334-00-1780
86 N2O Knob 0366-00-0134
87 Label, Flowmeter AIR 0334-00-1781
88 AIR Knob 0366-00-0133
89 Label, Flowmeter O2 0334-00-1779
90 O2 Knob 0366-00-0135

4 - 16 0070-10-0683 AS3000™ Service Manual


5.0 Calibration

5.1 Introduction
This section provides detailed information required to properly test and calibrate the
AS3000. Calibration consists of making mechanical and electrical adjustments with the
proper test equipment. The instrument should be tested and calibrated after repairs have
been completed or at regular intervals as part of a periodic maintenance procedure.

NOTE: Both calibration and a functional test must be performed to


verify complete and proper operation.

NOTE: Calibration is conducted in normal operating mode and in


Service Diagnostic mode.

Ensure that all testing materials, including drive gas, breathing circuits, test fixtures, tools and
documents are available and current, calibrated and in good working order prior to
beginning.

Testing and programming requires utilizing internal service software. A password is required
to enter these screens. This password is provided during biomedical technician training.

AS3000™ Service Manual 0070-10-0683 5-1


Calibration Warnings, Precautions, and Notes Calibration

5.2 Calibration Warnings, Precautions, and Notes


5.2.1 Warnings
WARNING: For continued protection against fire hazard, replace all
fuses with the specified type and rating.

WARNING: In order to prevent an electric shock, the machine


(protection class I) may only be connected to a correctly
grounded mains connection (socket outlet with grounding
contact).

WARNING: Remove all accessory equipment from the shelf before


moving the anesthesia machine over bumps or on any
inclined surface. Heavy top loading can cause the machine
to tip over causing injury.

WARNING: Possible explosion hazard. Do not operate machine near


flammable anesthetic agents or other flammable
substances. Do not use flammable anesthetic agents (i.e.,
ether or cyclopropane.)

WARNING: The use of anti-static or electrically conductive respiration


tubes, when utilizing high frequency electric surgery
equipment, may cause burns and is therefore not
recommended in any application of this machine.

WARNING: Possible electric shock hazard. The machine may only be


opened by authorized service personnel.

WARNING: Compressed gasses are considered Dangerous Goods/


Hazardous Materials per I.A.T.A. and D.O.T. regulations. It is
a violation of federal and international law to offer any
package or over pack of dangerous goods for
transportation without the package being appropriately
identified, packed, marked, classified, labeled and
documented according to D.O.T. and I.A.T.A. regulations.
Please refer to the applicable I.A.T.A. Dangerous Goods
Regulations and /or the Code of Federal Regulations 49
(Transportation, Parts 171-180) for further information.

5.2.2 Cautions
CAUTION: Refer to the “Periodic Maintenance Schedule of Service
Activities” on page 6-2, in the Periodic Maintenance section
for assistance when performing scheduled periodic
maintenance.

CAUTION: Do not leave gas cylinder valves open if the pipeline supply
is in use and the system master switch is turned to 'ON'. If
used simultaneously, cylinder supplies could be depleted,
leaving an insufficient reserve supply in the event of
pipeline failure.

CAUTION: Use cleaning agent sparingly. Excess fluid could enter the
machine, causing damage.

CAUTION: This machine must only be operated by trained, skilled


medical staff.

5-2 0070-10-0683 AS3000™ Service Manual


Calibration Calibration Warnings, Precautions, and Notes

5.2.3 Notes
NOTE: Only bacterial filters with a low flow resistance must be
connected to the patient module and/or the patient
connection.

NOTE: Use surgical gloves whenever touching or disassembling


valves or other internal components of the Breathing
System.

NOTE: Ensure that the gas supply of the machine always complies
with the technical specification.

NOTE: The APL Valve and PAW gauge numerics are for reference
only. Calibrated patient airway pressure is available on the
ventilator screen.

NOTE: If the machine should show faults during the initial


calibration or testing, the machine should not be operated
until the fault has been repaired by a qualified service
technician.

NOTE: After servicing, functional, sensor and system tests must be


carried out before clinical use.

NOTE: To accommodate additional monitors and other equipment


the AS3000 offers up to two vertical mounting tracks. Use
of unauthorized mounting accessories is not recommended.

NOTE: Always secure any equipment placed on the AS3000’s top


shelf.

AS3000™ Service Manual 0070-10-0683 5-3


General Guidelines Calibration

5.3 General Guidelines


1. Before disconnecting any pneumatic hoses, the hoses and mating fittings should be
tagged to show the proper connections. When reconnecting, all hoses must be checked
for proper connection. To further assure proper connection, all pneumatic calibrations
and tests defined in this manual should be accomplished before use on a patient.
2. Once the instrument covers have been removed, an electric shock hazard may exist.
Therefore, calibration should only be performed by qualified service personnel who
proceed with care and follow proper servicing techniques.
3. Do not attempt to calibrate the instrument without the test equipment and tools listed in
this manual.
4. Exercise care when reaching into the opened instrument which contains line (mains)
voltage.
5. When making adjustments and measurements, avoid accidental shorting of component
leads that can result in component failure.
6. Perform all steps in the order given. Do not skip any steps unless noted.
7. Understand each step of the procedure thoroughly before performing the procedure.
8. Before removing or replacing any circuit boards, disconnect the AS3000 from line
power, switch the Mains Switch to the OFF position, and disconnect the negative
terminal from the battery.

5.4 Test Equipment and Special Tools Required


See “Special Tools Required” on page 3-3, for a list of tools required for calibration.

5.5 Calibration Procedures


NOTE: The following calibration procedures must be performed in
sequence.

5.5.1 Oxygen Sensor Calibration


1. Preparing the unit
a. Allow the breathing system to warm up and reach thermal equilibrium
(approximately 30-60 minutes).
b. In Standby mode, press the MENU button. The menu screen will appear.
c. Select Service, then input the password 2010 to enter the Service screen.
d. Select Calibration to enter the Calibration screen.
e. Select Oxygen Sensor to enter the Oxygen Sensor Calibration screen.
2. Calibration
a. Remove the oxygen sensor from the Breathing System and expose it to room air for
at least 3 minutes.

NOTE: The O2 sensor voltage is displayed during the calibration.


This is the amplified O2 cell voltage at the A/D converter for
the oxygen sensor. The O2 sensor voltage is not displayed
for UI versions 2.24 and lower.

b. Flush the O2 sensor with air from the auxiliary output for 5-10 seconds to ensure that
no O2 bubbles are trapped in the sensor.

5-4 0070-10-0683 AS3000™ Service Manual


Calibration Calibration Procedures

NOTE: Do not shake the O2 sensor during calibration.

NOTE: Keep the O2 sensor in a vertical position, connector side up,


during calibration.

NOTE: Place the O2 sensor on top of the heated block during


calibration to minimize the temperature difference from
within the heated block.

NOTE: If the system is going to be used during the calibration,


insert the O2 cell plug into the port from which the oxygen
sensor was removed using a push and turn motion.

c. Select 21% to start 21% oxygen concentration calibration.


d. Install the oxygen sensor into the Breathing System.

NOTE: The oxygen sensor must be installed in the Breathing


System for 10 minutes prior to 100% oxygen concentration
calibration.

e. Press the O2 Flush button for 5-10 seconds to clear out any non-O2 gases from the
system.
f. Expose the oxygen sensor to 100% pure oxygen for at least 3 minutes.
g. Select 100% to start 100% oxygen concentration calibration.

5.5.2 Proportional Valve Regulator Calibration


1. Open the gas cover panel.

AS3000™ Service Manual 0070-10-0683 5-5


Calibration Procedures Calibration

2. Remove the drive pressure source from the AS3000.


3. Disconnect the input hose from the Proportional Valve.
4. Connect a Digital Pressure Meter to the proportional valve input hose.
5. Reconnect the drive pressure source to the AS3000.
6. Set the pressure on the pressure regulator to between 25.1 and 25.7 psi by pulling
down on the pressure regulator knob and rotating it until the desired pressure is
reached. (see FIGURE 5-1)
7. Push the regulator knob back up once the desired pressure is reached to lock it in place.
8. Remove the drive pressure source from the AS3000.
9. Remove the Pressure Meter and reconnect the input hose to the Proportional Valve.
10. Reconnect the drive pressure source to the AS3000.

Proportional valve input hose

FIGURE 5-1 The Proportional Valve Regulator

5-6 0070-10-0683 AS3000™ Service Manual


Calibration Calibration Procedures

5.5.3 Flow Sensor Calibration


1. Preparing the unit

FIGURE 5-2 Connecting the FIGURE 5-3 Blocking the Bag Port
Respiration Tube

a. Remove the Inspiratory and Expiratory valve disks, and reinstall the valve covers.
b. Connect a Respiration Tube between the Inspiratory and Expiratory Ports. (see
FIGURE 5-2)
c. Use the AS3000 Test Plug to block the Bag Port. (see FIGURE 5-3)
d. Set the APL Valve to 70 cmH2O.
e. With the Fresh Gas flow set to 100 mL/min, press the O2 Flush button. Verify that the
Paw gauge reads a minimum of 40 cmH2O after releasing the button.
f. If 40 cmH2O cannot be attained, open and close the APL Valve and repeat step e
prior to leakage troubleshooting.
g. Set the Fresh Gas flow to 0 mL/min.
h. Select Continue.

AS3000™ Service Manual 0070-10-0683 5-7


Calibration Procedures Calibration

2. Calibration

FIGURE 5-4 Connecting the Vent Tester to the AS3000

a. Connect one end of the Calibration Hose to the High Pressure port on the AS3000
and the other to the inlet on the Vent Tester.
b. Connect one end of the Respiration Tube to the expiratory port on the AS3000 and
the other to the outlet of the Vent Tester.
c. Connect an unterminated tube (at least 1 meter long) to the inspiratory port on the
AS3000.
d. Select Continue to start calibration.
e. Ensure that there is 0 flow on the Flow Analyzer.
f. Set the on screen calibration to 0, and select Done to begin calibration. Successful
calibration will be confirmed by illuminating the corresponding calibration value
yellow. If the calibration value illuminates red, recalibrate the value until it
successfully calibrated.
g. Set the onscreen calibration value to 1.
h. Raise the flow until it reads 1.00 on the Flow Analyzer using the calibration hose
valve.
i. Select Done to begin calibration.
j. Repeat steps g. through i. for all calibration values listed on the screen. Ensure all
successfully calibrate and illuminate yellow.
k. After all values are successfully calibrated select Finish Cal.
l. The AS3000 will confirm that calibration was successfully completed.
m. Select Done to return to the calibration screen.

NOTE: The flow should not be fluctuating in during the calibration.

5-8 0070-10-0683 AS3000™ Service Manual


Calibration Calibration Procedures

5.5.4 Flow Valve Calibration


1. Preparing the unit

FIGURE 5-5 Connecting the breathing circuit

a. Reinstall the Inspiratory and Expiratory valve disks.


b. Connect a breathing circuit to the Breathing System (see FIGURE 5-5).
c. Connect a 3L bag to the Y-fitting.

NOTE: A 2.3L bag can be used in place of a 3L bag.

d. Select Continue.
2. Calibration
a. Press the O2 Flush Valve to fully fill bellows.
b. Select Continue to start calibration.
c. Follow the on-screen instructions.

NOTE: The flow Valve calibration duration is approximately 5


minutes.

AS3000™ Service Manual 0070-10-0683 5-9


Calibration Procedures Calibration

5.5.5 Paw Sensor


1. 0 cmH2O Calibration

FIGURE 5-6 Removing the 3L bag

a. Remove the 3L bag from the breathing circuit.


b. Set the APL Valve to 0 cmH2O.
c. Select Continue to start calibration.
d. Follow the on-screen instructions.

5 - 10 0070-10-0683 AS3000™ Service Manual


Calibration Calibration Procedures

2. 30 cmH2O Calibration

FIGURE 5-7 Connecting the Y-fitting to the Test Port

a. Set the APL Valve to 30 cmH2O.


b. Connect the Y-fitting to the Test Port (see FIGURE 5-7)
c. Connect a Pressure Meter between the Y-fitting and the AS3000 (use an adapter if
necessary).
d. Use the O2 Flush Valve and O2 Flow Meter to maintain pressure at 30.0 ±0.2
cmH2O on the Pressure Meter.
e. Select Continue to start calibration.
f. Follow the on-screen instructions.

5.5.6 PEEP Valve

FIGURE 5-8 Disconnect the plastic flaps from the Adult Test Lung

1. Calibration
a. Disconnect the plastic flaps from the Adult Test Lung (P/N 0138-00-0012) and
connect it to the Y-fitting.
b. Set the Fresh Gas flow to 0 mL/min.
c. Press the O2 Flush Valve to fully fill bellows.
d. Select Continue to start calibration.
e. Follow the on-screen instructions.

AS3000™ Service Manual 0070-10-0683 5 - 11


Calibration Procedures Calibration

5.5.7 Flow Meter


1. Calibration

FIGURE 5-9 Connecting the Flow Analyzer

a. Connect one end of the Respiration Tube to the Common Gas Outlet on the
AS3000, and the other to the inlet on the Flow Analyzer (Certifier-FA PLUS).
b. Set the Fresh Gas flow to 0 mL/min.
c. Select Continue to start calibration.
d. Ensure that there is 0 flow on the Flow Analyzer.
e. Set the on screen calibration to 0, and select Done to begin calibration. Successful
calibration will be confirmed by illuminating the corresponding calibration value
yellow. If the calibration value illuminates red, recalibrate the value until it
successfully calibrated.
f. Set the onscreen calibration value to 1.
g. Raise the flow until it reads 1.00 on the Flow Analyzer using the Flow Meter.
h. Select Done to begin calibration.
i. Repeat steps f. through h. for all calibration values listed on the screen. Ensure all
successfully calibrate and illuminate yellow.
j. After all values are successfully calibrated select Finish Cal.
k. The AS3000 will confirm that calibration was successfully completed.
l. Select Done to return to the calibration screen.

NOTE: When calibrating the Flow Meter, use AIR if available,


otherwise use a combination of O2 and N2O to attain a flow
of 15 L/min.

5 - 12 0070-10-0683 AS3000™ Service Manual


Calibration Calibration Procedures

5.5.8 Leakage detection


5.5.8.1 Startup leakage detection
After restarting, select Continue to enter the Leakage Detection screen. Connect the Y-fitting
to the AS3000’s Test Port, and use the AS3000 Test Plug to block the Bag Port. Adjust the
APL Valve to 30 cmH2O and close the Flowmeter. Then, select Continue to prompt to the
next screen.

5.5.8.2 APL Valve leakage detection


Follow the on-screen instructions and use the O2 flush valve to pressurize the system. Ensure
that the bellows does not move while pressurizing the system. Meanwhile, observe the
pressure gauge to ensure that the pressure exceeds 30 cmH2O. Release the O2 Flush Valve
and observe whether the pressure displayed on pressure gauge stays within 25 - 35 cmH2O.
After passing the test, select Continue to prompt to the next screen.

5.5.8.3 Safety Valve leakage detection


The Safety Valve leak detection test will start automatically. After finishing, the result of the
test will be displayed on the screen. After passing the test, select Continue to prompt to the
next screen.

5.5.8.4 Breathing System leakage detection


Use the O2 Flush Valve to fill the bellow full with oxygen according to the prompt on the
screen. Select Continue to start the test. After finishing, the results of the test will be
displayed on screen. Leakage detection must be less than 400 mL/min. After passing the test,
select Continue to prompt to the next screen.

5.5.9 Compliance detection


Adjust the O2 Flowmeter to make the flow equal to 5 ±0.1 L/min according to the prompt on
the screen. Select Continue to start the Compliance Detection test. After finishing, the results
of the test will be displayed on screen. Compliance detection must be within 1 - 11mL/
cmH2O. After finishing compliance detection successfully, select Continue to enter Standby
Screen.

AS3000™ Service Manual 0070-10-0683 5 - 13


Calibration Procedures Calibration

This page intentionally left blank.

5 - 14 0070-10-0683 AS3000™ Service Manual


6.0 Periodic Maintenance

6.1 Maintenance Schedule


The following is a list of activities required for periodic maintenance of the AS3000
Anesthesia System. Physical inspection, replacement of consumables and performance
checks should be periodically performed per the schedule listed below. Certain calibration
adjustments are required only after replacing one or both of the active devices. The
manufacturer is not responsible for component failure or loss resulting from the use of stated
consumables beyond their recommended replacement interval. These are noted in the
Periodic Maintenance Schedule (See section 6.3).

6.2 Periodic Maintenance Consumable Parts Kits


Consumable parts are available in the PM kits listed below.

• AS3000, 12-Month PM Kit: 0040-00-0446


• AS3000, 36-Month PM Kit: 0040-00-0445
Parts must be replaced at periodic intervals according to the schedule.

AS3000™ Service Manual 0070-10-0683 6-1


Periodic Maintenance Schedule of Service Activities Periodic Maintenance

6.3 Periodic Maintenance Schedule of Service Activities


AFTER EACH EVERY 12 EVERY 36
REQUIRED ACTION SERVICE MONTHS MONTHS
Check list before surgery X X
Visual Inspection Checklist X X
Replace of Consumable Parts X X
Battery Maintenance and Replacement X
Functional Tests X X
Preoperative Checklist X X X

6.4 Visual Inspection Checklist


1. Verify that the AS3000 has no physical damage that would prevent operation.
2. Verify that the breathing circuit and CO2 absorbent are present.
3. Verify that the vaporizers are filled but not overfilled.
4. Verify that the Preoperative Checkout List is attached.
5. Verify that the tank wrench is attached. (0367-00-0080)
6. Verify that the waste gas scavenger hose to the APL Valve (0004-00-0095) is not
damaged. Drain any moisture.
7. Verify that the O2, N2O, AIR, VAC and EVAC Supply Hoses (as applicable) are not
damaged.
8. Verify that the AC line cord is not frayed or damaged.

6-2 0070-10-0683 AS3000™ Service Manual


6.5 Replacement of Consumable Parts
AS3000™ Service Manual

Periodic Maintenance
Parts are replaced at multiple intervals from the date of installation.

CONSUMABLE PARTS AMOUNT 12 MONTH PM 36 MONTH PM PART NUMBER


1 O2 Sensor 1 check check 0600-00-0149
2 Battery Assembly SLA-5AH24V 1 check replace 0997-00-0619
3 Drive Gas Tube 1 check replace 0004-00-0098
4 Bellows Assembly 1 replace replace 115-005968-00
5 Scavenger Hose 1 check replace 0004-00-0095
6 I/E Valve gasket 2 replace replace 0354-00-0192
7 O-ring, 6 x 1.8 (for airway pressure gauge) 1 replace replace 0354-00-0188
8 O-ring, 4 x 1.8 (for airway pressure gauge) 1 replace replace 0354-00-0189
9 Bag Arm retainer ring 1 check check 0354-00-0039
10 Disc, Ceramic I/E Valve 2 check replace 0354-00-0182
0070-10-0683

11 Gasket, Bellows canister base 1 replace replace 0354-00-0191


3 check replace 0354-00-0183 (original)
12 Absorber canister Seal 1 check replace 047-002850-00 (revised)
2 check replace 047-000252-00 (revised)
13 O-ring O2 Interface / O2 Plug 2 replace replace 0354-00-0187
14 O-ring Vaporizer mount 14x2.65 4 replace replace 0354-00-0193
15 Cylinder Washer, Black 3 replace replace 0348-00-0185
16 Absorber assembly 1 clean clean 0997-00-0629 (original)
115-005724-00 (revised)
17 AGSS filter 1 clean replace 0378-00-0074
18 Absorber Gasket 2 clean replace 0354-00-0199 (original)

Replacement of Consumable Parts


047-002849-00 (revised)
19 Check Valve 1 clean clean Non-replaceable
(Applicable to original model P/N 0997-00-0618
only. See “Check Valve Cleaning” on page 6-4.)
6-3
Check Valve Cleaning Periodic Maintenance

6.6 Check Valve Cleaning


Check valve cleaning is applicable only the original beathing system model (P/N 0997-00-
0618) only. It does not need to be performed on the revised beathing system model (P/N
115-005967-00) (see FIGURE 6-1).

Constant contact with moist patient gas can result in the breathing system check valve
becoming sticky. Cleaning the check valve will resolve this issue. Perform this cleaning
procedure at least every 12 months.

NOTE: Plastic gloves should be worn when handling breathing


systems

New Style Breathing Systems Old Style Breathing Systems


(115-005967-00) (0997-00-0618)

FIGURE 6-1 Breathing System Styles

6.6.1 Tools and Materials


• Philips screw driver
• 2.5 mm hex wrench
• 3 mm hex wrench
• 4 mm hex wrench
• 12 mm standard socket
• 13 mm open end wrench
• Silicone Lubricant (PN 0510-00-0021)
• Tape measure (metric)
• Loctite 243
• Isopropyl alcohol

6-4 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Check Valve Cleaning

6.6.2 Cleaning Procedure


1. Remove Absorber Locking Mechanism and Absorber Mount Assembly:
1. Rotate the cam handle to the unlocked position.
2. Remove (2) Absorber Canister Assemblies.
3. Remove Absorber Hose from port on Absorber Mount Assembly.
4. Remove (2) 4mm hex screws securing the Absorber Locking Mechanism to the rods.
5. Remove the Absorber Locking Mechanism and Absorber Mount Assembly.

Absorber
Locking
Mechanism
Absorber Mount
Assembly

Absorber Canister Absorber


Assembly Hose

FIGURE 6-2 Absorber Assembly

AS3000™ Service Manual 0070-10-0683 6-5


Check Valve Cleaning Periodic Maintenance

2. Remove Absorber Guide Rods and Mounting Plate:


1. Loosen the locking nuts on (2) Absorber Guide Rods.
2. Remove the (2) Absorber Guide Rods by unscrewing them from the Breathing System
Block.
3. Remove Gasket and Screen from Absorber Mounting Plate.
4. Remove (3) philips head screws with (3) associated washers securing the Absorber
Mounting Plate.
5. Remove the Absorber Mounting Plate

Absorber Guide
Rod

Gasket and
Screen

Locking Nuts

FIGURE 6-3 Absorber guide rods and mounting plate

6-6 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Check Valve Cleaning

3. Remove Breathing System parts:


1. Remove the following parts from the Breathing System:
• APL Valve
• Breathing Circuit
• Manual Bag
• Bag Arm
• Bellows Cover
• Airway Pressure (PAW) Gauge
• Bellows
• Bellows Cover Base Gasket
• O2 Sensor Assembly
• Inspiratory Check Valve Assy
• Drive Gas Tube
• CGO Hose
• Heater Cable
• Pneumatc Hose
• Scavenger Hose
• Bellows Support Cage (if available)

Bellows Cover
O2 Sensor
Bellows
Assembly

PAW Gauge
Bag Arm

Inspiratory
Check Valve,
Expiratory
Valve Holder,
Check Valve,
Valve Holder, Cover, and Nut
Cover, and Nut

APL Valve

FIGURE 6-4 Breathing System

AS3000™ Service Manual 0070-10-0683 6-7


Check Valve Cleaning Periodic Maintenance

4. Remove Breathing System bottom block:


1. Remove (15) 3mm hex screws securing the bottom block of the Breathing System.
Save the screws.
2. Remove the bottom block of the Breathing System by inserting a standard 12mm
socket into the Absorber Port to hold down the block. Pull the installed Absorber Hose
Port to separate bottom block.
3. Remove the 12mm socket from the Breathing System.

Absorber Port

FIGURE 6-5 Absorber port

6-8 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Check Valve Cleaning

5. Remove and clean check valve:


1. Using needle nose pliers, remove the bellows check valve by gently inserting into
round holes, then prying loose.
2. Separate valve holder, check valve and valve seat.
3. Clean check valve, valve seat, and valve holder using a cloth or equivalent moistened
with isopropyl alcohol .
4. Re-assemble check valve, valve seat, and valve holder.
5. Apply silicone lubricant PN 0510-00-0021 to valve seat o-ring.
6. Gently re-install bellows check valve assembly into port, assuring that there is no
damage to the o-ring. Push down until it bottoms out.

Bellows Check Valve

FIGURE 6-6 Bellows check valve

AS3000™ Service Manual 0070-10-0683 6-9


Check Valve Cleaning Periodic Maintenance

6. Inspect gaskets and secure bottom block:


1. Inspect all internal gaskets and make sure they are seated properly.
2. Secure the bottom block with the original (15) 3mm hex screws:
• Loosely install all (15) screws
• Tighten the screws using the numbered sequence shown below.

11

FIGURE 6-7 Bottom block screws

6 - 10 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Check Valve Cleaning

7. Install absorber canister assembly:


1. Apply Loctite 243 to (3) screws of the Absorber Mounting Plate.
2. Secure mounting plate to Breathing System Block using (3) screws and (3) associated
washers.
3. Apply Loctite 243 to the threads of (2) Absorber Guide Rods.
4. Install rods into block until rod height is 314 +/- 0.5 mm (old style breathing systems)
or 303+/-0.25 mm (new style breathing systems) (see FIGURE 6-9).

Absorber
Mounting Plate Holder,
Absorber
Canister

Gasket,
Absorber Gasket,
Canister Absorber

Rod, Support
Absorber Ring Seal
Canister (2 pcs) (inside
Absorber
Canister)

FIGURE 6-8 Absorber Canister Assembly

New Style Breathing Systems Old Style Breathing Systems


(115-005967-00) (0997-00-0618)

FIGURE 6-9 Breathing System Styles

AS3000™ Service Manual 0070-10-0683 6 - 11


Check Valve Cleaning Periodic Maintenance

8. Secure rods and install final assembly:


1. Both rods must be equal in height.
Secure rods by tightening the locking nuts.
2. Install an Absorber Gasket and a Screen onto the Absorber Mounting Plate.
3. Insert Absorber Mount Assembly onto guide rods, noting the position of the Absorber
Hose Port and Absorber Hose.
4. Install Absorber Hose onto Absorber Hose Port.
5. Install Absorber Locking Mechanism onto guide rods, noting the silk screening must
face the front of the block. When assembling, put the locking mechanism lever in the
"locked" position.
6. Secure the locking mechanism using the (2) original 4mm hex screws. Install the two
Absorber Canister Assemblies.
7. Measure the distance between the absorber mount assembly plate at four different
points. The difference in length should not be more than 1 mm.
9. Perform tests on the check valve:
1. Perform "Start-Up Tests" on page 6-15 and "Performance Verification" on page 6-27.

6 - 12 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Battery Maintenance and Replacement

6.7 Battery Maintenance and Replacement


6.7.1 Battery Maintenance
1. The AS3000 uses a sealed lead-acid battery. Due to the self-discharge characteristics
of this type of battery, it is imperative that it is charged after 3 months of storage (or after
extended periods of non-use). If not charged, a permanent loss of capacity may occur as
a result of sulfation.
2. Check the battery run time every 12 months.
3. Replace batteries when operating time is less than 30 minutes or when the battery is 3
years old.

NOTE: The sealed lead-acid batteries used in the system are


maintenance free. They will perform reliably provided that
they are kept in the charged state.

4. Disposal of batteries should be conducted in accordance with local recycling statues


and labeling shown on the battery pack.

6.7.2 Battery Replacement


1. Open the access door located at the rear of the AS3000.
2. Disconnect the battery connections.
3. Remove the old battery.
4. Ensure the new battery is electrically connected and secured in the same manner as the
original one.
5. Close the access door.
6. Use only Approved Batteries (P/N 0997-00-0619).

6.8 Functional Tests


Refer to “Calibration” on page 5-1 if any values are out of specification.

6.8.1 Test Equipment and Special Tools Required


Refer to “Special Tools Required” on page 3-3 for a list of tools required for Running the
functional tests.

AS3000™ Service Manual 0070-10-0683 6 - 13


Functional Tests Periodic Maintenance

6.8.2 Pressure Regulator Checks


6.8.2.1 Proportional Valve Regulator
1. Open the gas cover panel.
2. Remove the drive pressure source from the AS3000.
3. Disconnect the input hose from the Proportional Valve.
4. Connect a Digital Pressure Meter to the proportional valve input hose.
5. Reconnect the drive pressure source to the AS3000.
6. Verify the pressure displayed on the Pressure Meter is between 23.9 and 26.8 psi.
7. Remove the drive pressure source from the AS3000.
8. Remove the Pressure Meter and reconnect the input hose to the Proportional Valve.
9. Reconnect the drive pressure source to the AS3000.

NOTE: If the proportional Valve Regulator is out of specification,


calibration is required.

Proportional valve input hose

FIGURE 6-10 The Proportional Valve Regulator

6.8.3 Gas Delivery System Tests


6.8.3.1 O2 Flush Verification
1. Attach the Digital Flow Meter to the Common Fresh Gas Outlet.
2. Verify that the O2 flush flow is between 35 to 50 L/min when pressing the O2 flush valve.

6 - 14 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.3.2 O2:N2O Ratio System


For -01 units:
1. Set the O2 and N2O Flow Control Valves to minimum.
2. Rotate the N2O Flow Control Valve until the top of the O2 flow meter's float rises to 1L/min.
3. Verify that the N2O flow is not higher than 3.7 L/min.
4. Lower the N2O flow to minimum.

For -02 units:


1. Set all flow control knobs to minimum.
2. Set the flow of O2 to 1 L/min.
3. Open the N2O flow knob and verify that the N2O flow will not increase higher than
3.7 L/min.
4. Return flow control knobs to minimum.

6.8.4 Start-Up Tests


6.8.4.1 System Self Test.
1. Power ON the AS3000.
2. Wait until the System Self Test is complete

FIGURE 6-11 Startup Self Test

3. Press Continue to prompt to the Safety Valve Test

6.8.4.2 Leak/Safety Valve Test


1. Follow the on-screen instructions.

AS3000™ Service Manual 0070-10-0683 6 - 15


Functional Tests Periodic Maintenance

FIGURE 6-12 Leak Test Setup

2. Select Continue to start the test.

FIGURE 6-13 Safety Valve Test Setup

6 - 16 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

FIGURE 6-14 Safety Valve Test in Progress

FIGURE 6-15 Safety Valve Test Passed Message

3. After 3 seconds, the screen will prompt to the Leak Test.

6.8.4.3 Leak Test


1. Follow the on-screen instructions.

AS3000™ Service Manual 0070-10-0683 6 - 17


Functional Tests Periodic Maintenance

2. Select Continue to start the test.

FIGURE 6-16 Leak Test Setup

FIGURE 6-17 Leak Test in Progress

6 - 18 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

3. Select Continue to prompt to the Compliance Test.

FIGURE 6-18 Leak Test Results

AS3000™ Service Manual 0070-10-0683 6 - 19


Functional Tests Periodic Maintenance

6.8.4.4 Compliance Test


1. Follow the on-screen instructions.
2. Select Continue to start the test.

FIGURE 6-19 Compliance Test Setup

FIGURE 6-20 Compliance Test in Progress

6 - 20 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

3. Select Continue to prompt to the Normal Screen.

FIGURE 6-21 Compliance Test Results

6.8.4.5 Manual Leak Test


NOTE: The Manual Leak Test detects smaller leaks than can be
detected in the Automatic Leak Test.

1. Ensure that the gas pressure for O2, N2O, and AIR are at 50 ±10 psi.
2. Power ON the AS3000.
3. Attach a reusable-silicone-rubber breathing circuit to the Breathing System.

NOTE: For testing purposes always use a reusable breathing


circuit.

4. Tightly connect the Y-fitting on the breathing circuit to the test port.
5. Attach a breathing bag to the bag arm.
6. Set the APL Valve to the fully closed position (70 cmH2O).
7. Rotate the O2 Flow Control Valve until 50 cmH2O pressure is observed on the Airway
Pressure Gauge.
8. Verify that the flow required to stabilize the pressure is less than 300 mL/min.

AS3000™ Service Manual 0070-10-0683 6 - 21


Functional Tests Periodic Maintenance

6.8.4.6 Oxygen Sensor Calibration


NOTE: Oxygen Sensor Calibration can be performed in all
ventilation modes.

1. Press the MENU key and then use the Navigator™ Knob to scroll to the Calibrate
menu tab (see FIGURE 6-22). Select the Start Calibration button.

FIGURE 6-22 Calibrate Menu Tab

2. After the Start Calibration button has been selected, the screen shown in
FIGURE 6-23 will be displayed, instructing the user to remove the oxygen sensor from
the Breathing System and expose it to room air for at least 3 minutes before proceeding.
After at least 3 minutes have passed, select the Next button to initiate the calibration
process. The progress bar shown in FIGURE 6-24 will be displayed.

NOTE: If the system is going to be used during the calibration,


insert the O2 cell plug into the port from which the oxygen
sensor was removed using a push and turn motion.

6 - 22 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

FIGURE 6-23 Oxygen Sensor FIGURE 6-24 Oxygen Sensor Calibration


Calibration Instructions Progress Bar

3. Proceed based on one of the following two conditions:


• If the calibration is successful, the screen shown in FIGURE 6-25 will be displayed,
instructing the user to reinstall the oxygen sensor into the Breathing System. Select the
Done button to complete the process.
• If the calibration fails, the screen shown in FIGURE 6-26 will be displayed,
instructing the user to either repeat the calibration (by selecting the Repeat Cal button)
or to replace the oxygen sensor. If the oxygen sensor must be replaced, select the Exit
button, replace the oxygen sensor using a push and turn motion, and then repeat the
calibration.

FIGURE 6-25 Oxygen Sensor FIGURE 6-26 Oxygen Sensor


Calibration Successful Calibration Failed

AS3000™ Service Manual 0070-10-0683 6 - 23


Functional Tests Periodic Maintenance

6.8.5 Pneumatic Leak Tests


6.8.5.1 N2O Cylinder Leak Test
1. Remove the N2O line pressure hose from the line pressure inlet on the AS3000.
2. Mount a full N2O cylinder to the rear panel yoke. If necessary, place a new clean tank
washer between the cylinder and the yoke to minimize any leaks at the yoke connection.
3. Open the N2O cylinder until its pressure gauge indicates cylinder pressure.
4. Close the N2O cylinder.
5. The N2O cylinder pressure gauge should not drop more than 10% of its initial pressure
over 1 minute.

6.8.5.2 O2 Cylinder Leak Test


1. Remove the O2 line pressure hose from the line pressure inlet on the AS3000.
2. Mount a full O2 cylinder to the rear panel yoke. If necessary, place a new clean tank
washer between the cylinder and the yoke to minimize any leaks at the yoke connection.
3. Open the O2 cylinder until its pressure gauge indicates cylinder pressure.
4. Close the O2 cylinder.
5. The O2 cylinder pressure gauge should not drop more than 10% of its initial pressure
over 1 minute.

6.8.5.3 AIR Cylinder Leak Test


1. Remove the AIR line pressure hose from the line pressure inlet on the AS3000.
2. Mount a full AIR cylinder to the rear panel yoke. If necessary, place a new clean tank
washer between the cylinder and the yoke to minimize any leaks at the yoke connection.
3. Open the AIR cylinder until its pressure gauge indicates cylinder pressure.
4. Close the AIR cylinder.
5. The AIR cylinder pressure gauge should not drop more than 10% of its initial pressure
over 1 minute.

6.8.5.4 Line Supply Check Valves Test


1. Connect and open a full gas cylinder on each yoke.
2. Verify the cylinder gauges operate.
3. Remove each line pressure hose one at a time from the wall supply inlet.
4. Verify no gas is escaping from any of the three central supply high pressure hoses
connected to the AS3000 using a Digital Flow Meter.
5. Verify there is no external damage to the gas supply inlet coupling on the AS3000.
6. Reattach each central supply gas hose to the wall supply inlets.

6 - 24 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.5.5 N2O Line Pressure Leak Test


1. Remove the N2O cylinder from the AS3000.
2. Connect the N2O line pressure hose to the line pressure inlet on the AS3000.
3. Pinch the N2O line pressure hose to stop N2O line flow.
4. Remove the N2O line pressure hose from the N2O line source.
5. The N2O line pressure gauge should not fall more than 2 psi in 20 seconds.
6. Reconnect the N2O line pressure and remove the pinch in the hose.

6.8.5.6 O2 Line Pressure Leak Test


1. Remove the O2 cylinder from the AS3000.
2. Connect the O2 line pressure hose to the line pressure inlet on the AS3000.
3. Pinch the O2 line pressure hose to stop O2 line flow.
4. Remove the O2 line pressure hose from the O2 line source.
5. The O2 line pressure gauge should not fall more than 2 psi in 20 seconds.
6. Reconnect the O2 line pressure and remove the pinch in the hose.

6.8.5.7 AIR Line Pressure Leak Test


1. Remove the AIR cylinder from the AS3000.
2. Connect the AIR line pressure hose to the line pressure inlet on the AS3000.
3. Pinch the AIR line pressure hose to stop AIR line flow.
4. Remove the AIR line pressure hose from the AIR line source.
5. The AIR line pressure gauge should not fall more than 2 psi in 20 seconds.
6. Reconnect the AIR line pressure and remove the pinch in the hose.

6.8.5.8 Cylinder Supply Check Valves Test


1. Close and remove all gas cylinders from the AS3000.
2. Verify the cylinder gauges return to zero.
3. Using the Digital Flow Meter, verify that no gas is escaping from any of the three yoke
connections on the AS3000.
4. Verify there is no external damage to the gas cylinder or yoke pins.
5. Reconnect the cylinder(s) as necessary.

AS3000™ Service Manual 0070-10-0683 6 - 25


Functional Tests Periodic Maintenance

6.8.6 Breathing System Checks


6.8.6.1 Waste Gas Scavenger Test (if available)
1. Connect one end of the low pressure waste gas hose to the port on the Waste Gas
Scavenger Assembly. Connect the other end of the hose to the EVAC port.

NOTE: If operating the AS3000 with other types of waste gas


scavenging, ensure that waste gases are directed from the
EVAC port to that scavenging system.

2. Connect the respiratory gas monitor exhaust output to the barbed fitting port on the
Waste Gas Scavenger Assembly.
3. Cap any open ports on the waste gas scavenger assembly.
4. Ensure that the waste gas scavenger flow adjustment is able to be set throughout its full
range.

6.8.6.2 Internal Gas Connections Test


1. Close and remove all gas cylinders from the AS3000
2. Connect only the AIR line pressure hose to the AS3000 from the wall supply. Leave all
other line pressure hoses disconnected.
3. With the AS3000 powered OFF, rotate the AIR flow control knob to ensure a
continuous flow increase throughout its full range
4. Disconnect the AIR line pressure hose from the AS3000, and connect the O2 line
pressure hose from the wall supply.
5. Fully rotate the N2O flow control knob and verify that there is no flow.
6. Fully rotate the AIR flow control knob and verify that there is no flow.

6.8.6.3 Drive Gas Pressure Loss Alarm, N2O Cutoff Test


1. Set the O2 flow to 2 L/min using the flow control valve.
2. Set the N2O flow to 2 L/min using the flow control valve.
3. Set the AIR flow to 2 L/min using the flow control valve.
4. Interrupt the O2 supply to the AS3000.
5. Verify that the flow of N2O and O2 stops within 2 minutes and that the flow of AIR (if
available) continues to flow at 2 L/min.
6. Verify the following alarms are activated:
• O2 Supply Failure appears on the screen
• An alarm tone sounds.

6 - 26 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.7 Performance Verification


6.8.7.1 Standby Mode Ventilation Test
1. Ensure that the gas pressure for O2, N2O, and AIR are at 50 ±10 psi.
2. Power ON the AS3000.
3. Perform the start up tests per the on-screen instructions. Ensure successful completion.
4. Attach a breathing circuit and test lung to the Y-fitting of the breathing circuit.

NOTE: For testing purposes always use a reusable breathing


circuit.

5. Set the APL Valve to approximately 15 cmH2O.


6. Set the AIR flow to 5 L/min using the flow control valve.
7. Squeeze the breathing bag once every 10 seconds to inflate and deflate the test lung to
approximately 20 cmH2O of pressure.
8. Verify the inflation and deflation of the test lung.

6.8.7.2 Manual Mode Ventilation Test


1. Set the Ventilation Mode to MANUAL.
2. Set the APL Valve to approximately 25 cmH2O. Push the O2 Flush button to fill the
breathing bag.
3. Set the AIR flow to 1 L/min using the flow control valve.
4. Squeeze the breathing bag once every 3 seconds.
5. Verify the inflation and deflation of the test lung.
6. Verify that an airway pressure waveform and all numeric values appear on screen
during bag compressions.
7. Stop squeezing the breathing bag and set the APL Valve to the open position (SP).

6.8.7.3 APNEA Alarm Test


1. While in the Manual Ventilation Mode, stop ventilating the test lung.
2. Verify that the following APNEA alarm signals activate at approximately 60 seconds
from the last bag compression.
• APNEA appears on the screen.
• An alarm tone sounds.

6.8.7.4 Alarm MUTE Test


1. While the APNEA alarm is sounding, press the MUTE key.
2. Verify the audio portion of the alarm stops and resumes after 2 minutes.

AS3000™ Service Manual 0070-10-0683 6 - 27


Functional Tests Periodic Maintenance

6.8.7.5 CMV Adult Ventilation Mode Test


1. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

2. Attach an adult test lung to the Y-fitting of the breathing circuit.


3. Attach a Vent Tester between the EXP port and the expiratory hose.
4. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Patient Type Adult
Ventilation Mode CMV
Tidal Volume - VT 600
Breath Rate - freq 8
I:E Ratio - I:E 1:2
Plateau - TP 10
PEEP - PEEP Off

6. Select CMV again to begin ventilation.


7. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp. rate
and minute volume values appear on the screen.
8. Verify the Tidal Volume display is within 15% of the set value within approximately 1
minute from the start of ventilation.
9. Verify the Tidal Volume display is within 15% of the delivered volume measured with the
Vent Tester within approximately 1 minute from the start of ventilation.
10. Verify the measured O2 concentration is at least 97% after 5 minutes.
11. Set the AIR flow to 3 L/min and set the N2O and O2 flow rates to minimum flow.
12. Verify the measured O2 concentration is 21% ±3% vol. % after 5 minutes.

6 - 28 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.7.6 CMV Child Ventilation Mode Test


1. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

2. Attach an adult test lung to the Y-fitting of the breathing circuit.

NOTE: Limit the volume in the test lung to provide sufficient airway
pressure to satisfy the Low Peak Pressure alarm. Or reduce
the Peak Pressure alarm limit to a lower valve to prevent
the alarm when using an adult test lung.

3. Attach a Vent Tester between the EXP port and the expiratory hose.
4. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Patient Type Child
Ventilation Mode CMV
Tidal Volume - VT 120
Breath Rate - freq 20
I:E Ratio - I:E 1:2
Plateau - TP 10
PEEP - PEEP Off

6. Select CMV to begin ventilation.


7. Verify that the pressure waveform, Tidal Volume, Mean or Plateau Pressure, Resp. rate
and minute volume values appear on the screen.
8. Verify the Tidal Volume display is within 25ml or ±15% of the delivered volume
(whichever is greater) measured with the Vent Tester within approximately 1 minute from
the start of ventilation.
9. Verify the delivered volume as measured by a Vent Tester at the expiratory port, is within
17% of the Tidal Volume set value within approximately 1 minute from the start of
ventilation.

6.8.7.7 Airway Disconnect Alarm Test


1. While the ventilator is running, disconnect the expiratory limb from the Expiratory Port
on the Breathing System.
2. Verify the following airway pressure disconnect alarm signals activate:
• APNEA message appears on the screen.
• Low Airway Pressure message appears on the screen.
• An alarm tone sounds.

AS3000™ Service Manual 0070-10-0683 6 - 29


Functional Tests Periodic Maintenance

6.8.7.8 PCV Adult Ventilation Mode Test


1. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

2. Attach an adult test lung to the Y-fitting of the breathing circuit.


3. Attach a Vent Tester between the EXP port and the expiratory hose.
4. Set the O2 flow to 3 L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Patient Type Adult
Ventilation Mode PCV
Target Pressure - PTARGET 20
Breath Rate - freq 8
I:E Ratio - I:E 1:2
PEEP - PEEP Off
Inspiratory Slope - Tslope 0.5

6. Select PCV to begin ventilation.


7. Verify the Peak Pressure reading of the display is within ±4 cmH2O of the Peak Pressure
measured with the Vent Tester.
8. Verify that the pressure waveform, Tidal Volume, Resp. Rate and minute volume values
appear on the screen.
9. Verify that the PEAK Value reaches 20 ±4 cmH2O within five breaths from the start of
ventilation.

6 - 30 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.7.9 Pressure Support (PS) Ventilation Mode Test


1. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

2. Attach an adult test lung to the Y-fitting of the breathing circuit.


3. Attach a Vent Tester between the EXP port and the expiratory hose.
4. Set the O2 flow to 1 L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Vent Dial Patient Type Adult
Vent Mode Ventilation Mode PS
Differential Pressure - P 20
PEEP - PEEP Off
Flow Trigger - Trigger 3
Inspiratory Slope - Tslope 0.5
Minimum frequency - freqMIN 2

6. Select PS to begin ventilation.


7. Begin triggering breaths by slightly squeezing the test lung and releasing. Maintain a
continuous breath rate.
8. Verify that a pressure waveform and all ventilation parameters appear on the screen.
9. Verify that the Peak Pressure reading on the display is ±4 the value of P + PEEP.
10. Stop triggering breaths.
11. Verify that after 30 seconds the ventilator delivers a breath and displays the message
APNEA BACKUP.
12. Verify the system ventilates with a frequency of 2 bpm

AS3000™ Service Manual 0070-10-0683 6 - 31


Functional Tests Periodic Maintenance

6.8.8 Alarms and Failsafe Functions


6.8.8.1 Set Up
1. Ensure that the gas pressure for O2, N2O, and AIR are at 50 ±10 psi.
2. Power ON the AS3000.
3. Perform the Startup Tests per the on-screen instructions. Ensure successful completion.
4. Attach a breathing circuit and breathing bag.

NOTE: For testing purposes always use a reusable breathing


circuit.

5. Attach an adult test lung to the Y-fitting of the breathing circuit.


6. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
7. Set the ventilator controls to:

VENTILATOR CONTROLS VENTILATOR SETTINGS


Patient Type Adult
Ventilation Mode CMV
Tidal Volume - VT 600
Breath Rate - freq 8
I:E Ratio - I:E 1:2
Plateau - TP 10
PEEP - PEEP Off

8. Select CMV to begin ventilation.

6.8.8.2 Low FiO2 Alarm Test


1. Set the low FiO2 Alarm limit to 50% O2.
2. Set the AIR flow control valve to 5 L/min.
3. Set the FiO2 flow controller to minimum flow.
4. Verify the following Low FiO2 alarm signals activate, within three ventilation cycles:
• Low FiO2 message appears on the screen.
• An alarm tone sounds.
5. Set the Low FiO2 alarm limit to 18%.
6. Verify the alarm signals cease.

6 - 32 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.8.3 High FiO2 Alarm Test


1. Set the high FiO2 Alarm limit to 49% O2.
2. Set the FiO2 flow control valve to 5 L/min.
3. Set the AIR flow controller to minimum.
4. Verify the following High FiO2 alarm signals activate:
• High FiO2 message appears on the screen.
• An alarm tone sounds.
5. Set the high FiO2 alarm limit to the max setting.
6. Verify the alarm signals cease.

6.8.8.4 Peak Pressure Alarms Test


1. Set the PAW low alarm to the lowest setting.
2. Set the PAW high alarm limit set point about 5 to 8 digits below the Peak Pressure
displayed on the upper left of the screen.
3. Verify the following (high) peak pressure alarms activate:
a. High Airway Pressure message appears on the screen.
b. An alarm tone sounds.
c. Inspiration ends and expiration begins as the pressure meets the high alarm limit.
4. Set the PAW high alarm limit set point to 65 (cmH2O).
5. Verify the alarms signals cease.
6. Set the PAW low alarm limit set point to 50 (cmH2O).
7. Verify the following (low) peak pressure alarms activate:
a. Low Airway Pressure message appears on the screen.
b. An alarm tone sounds.
8. Set the PAW low alarm limit to 12 (cmH2O).
9. Verify the alarm signals cease.

AS3000™ Service Manual 0070-10-0683 6 - 33


Functional Tests Periodic Maintenance

6.8.8.5 Minute Volume Alarm Test


1. Set the MV Low alarm limit set point to the highest value.
2. Verify the following alarms activate:
• Low MV message appears on the screen.
• An alarm tone sounds.
3. Set the MV Low alarm limit to minimum setting.
4. Verify the the alarm signals cease.
5. Set the MV High alarm limit set point to the lowest value.
6. Verify the following alarms activate:
• High MV message appears on the screen.
• An alarm tone sounds.
7. Set the MV High alarm limit set point to the highest value
8. Verify that the alarm signals cease.

6.8.9 Miscellaneous Tests


6.8.9.1 Test the Line Voltage Alarm
1. Before starting this test, verify that the battery is fully charged. (The battery icon in the
upper right corner of the screen is solid when the battery is fully charged.)
2. Interrupt AC line voltage.
3. Verify that the following alarms activate:
• An alarm tone sounds.
• AC Power Failure message appears on the screen.
4. Verify that a fully charged battery operates the ventilator for a minimum of 45 minutes.
5. Plug the AS3000 into AC line voltage.
6. Verify that the alarm signals cease.
7. Verify the presence of the battery charging icon in the upper right corner of the screen
(status bar moving from right to left).

6.8.9.2 Wheel Brakes Test


1. Verify that each front wheel brake operates.

6.8.9.3 Work Light Test


1. Turn on the work light located on the bottom side of the top panel.
2. Verify that it lights in both on positions.

6 - 34 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.9.4 Auxiliary Flowmeter


1. Verify an AIR flow of 15 L/min can be obtained by connecting the auxiliary AIR hose to
the pressure source and opening the flow meter.
2. Verify an O2 flow of 15 L/min can be obtained by connecting the auxiliary O2 hose to
the pressure source and opening the flow meter.

6.8.9.5 Patient Suction Regulator (if available)


1. Set the suction regulator’s selection dial to LINE.
2. Verify maximum suction vacuum on the regulator's gauge.
3. Set the suction regulator's selection dial to REGULATE.
4. Verify the suction vacuum is adjustable on the regulator's gauge.
5. Set the suction regulator's selection dial to OFF.

6.8.10 Vaporizers
6.8.10.1 Vaporizer Interlock Test
1. Attach two vaporizers to the Vaporizer Mounting Manifold and lock them in place.
2. Rotate either of the vaporizer dial to 3% agent.
3. Verify that the other vaporizer dial cannot be rotated to a setting.
4. Set both vaporizer dials to 0.
5. Rotate the other vaporizer dial to 3%.
6. Verify that the first vaporizer dial cannot be rotated.
7. Rotate both vaporizer dials to T and remove both vaporizers.
8. Verify that the locking spring is intact.
9. Reconnect both vaporizers to the Vaporizer Mounting Manifold.

AS3000™ Service Manual 0070-10-0683 6 - 35


Functional Tests Periodic Maintenance

6.8.10.2 Vaporizer Accuracy Test


WARNING: Avoid exposure to respiratory gases by always directing
the fresh gas flow from the fresh gas outlet to the waste gas
scavenger.

1. Remove the Fresh Gas Hose from the Common Gas Outlet.
2. Insert an endo-tracheal tube mask elbow adapter with monitoring port, into the Common
Gas Outlet.
3. Set the APL Valve to 70 cmH2O.
4. Remove the Waste Gas Scavenger Hose from the bottom of the Breathing System EVAC
port. Leave the connection to the waste gas scavenger attached.
5. Set the waste gas scavenger flow above the minimum setting.
6. Adapt the Waste Gas Scavenger Hose to connect to the endo-tracheal tube mask elbow,
now connected to the Common Gas Outlet port.
7. Attach the Agent Analyzer meter sampling tube to the endo-tracheal tube mask elbow's
monitoring port.
8. Fill the vaporizer with anesthetic agent (if necessary).

NOTE: Do not overfill by filling past the indicator line on the


vaporizer.

9. Test the vaporizer accuracy per the Dräger Vapor 2000 instructions (See
section 6.8.10.4), or see the appropriate vaporizer manual for testing details.
10. Test each vaporizer in turn.
11. Test any vaporizer on the Vaporizer Storage Mount.
12. Remove the measuring equipment from the Common Gas Outlet.
13. Reconnect the Fresh Gas, and Waste Gas Scavenger Hoses.

6.8.10.3 Vaporizer Leak Test


NOTE: Verify the system has passed the leak test without the
vaporizers prior to this test.

1. Verify that vaporizers are mounted correctly.


2. Ensure that the gas pressure for O2, N2O, and AIR are at 50 ±10 psi.
3. Power ON the AS3000.
4. Attach a reusable-silicone-rubber breathing circuit to the Breathing System.

NOTE: For testing purposes always use a reusable breathing


circuit.

5. Tightly connect the Y-fitting on the breathing circuit to the test port.
6. Attach a breathing bag to the bag arm.
7. Set the APL Valve to the fully closed position (70 cmH2O).
8. Rotate the O2 Flow Control Valve until 50 cmH2O pressure is observed on the Airway
Pressure Gauge.
9. Verify that the necessary flow to stabilize the pressure is less than 300 mL/min.

6 - 36 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Functional Tests

6.8.10.4 Dräger Vapor 2000 Operating Instructions ARRB-F001


1. Fill Vaporizer - at least half full between minimum and maximum mark.
2. Allow the filled Vapor to warm up to room temperature of 20-24C. Wait long enough
for the temperature to equalize - the time will vary depending on the temperature
differential 'T'.
• 1 hour = up top 2C
• 3 hours = ±6C
• 4 hours = ±10C
• 5 hours = ±20C
3. Check anesthetic agent monitor. Perform zero calibration of monitor with the desired
gas (AIR or O2)
4. Connect monitor to fresh gas outlet or Y-fitting. Make sure that all connections are leak-tight.
5. Connect and start scavenging system.
6. Switch OFF ventilator or set vent pressure to less than 5 cmH2O.
7. Set monitor to anesthetic agent being used and to continuous measurement.
8. Set flow between 2.5 and 4 L/min AIR. Use O2 if AIR is not available.
9. Check 0 and T marks, 1 vol.% 4 vol % and at least three other concentrations.
10. Adjust control dial on the vaporizer.
11. Read concentration after it has reached steady state.
12. Correct measured values for the effect of carrier gas used. If AIR no correction required.
If O2 use following correction factor:
• Measured value vol.% = <1.0, correction = -0.05 vol.%
• Measured value vol.% = 1.0 - 2.0, correction = -0.1 vol.%
• Measured value vol.% = 2.5 - 4.0, correction = -0.2 vol.%
• Measured value vol% = 5.0 - 8.0, correction = -0.3 vol.%
13. If the value displayed on the monitor is in % partial pressure, no correction is required. If
in vol.% convert to partial pressure:
• Concentration [% partial pressure] = measured value [vol.%] x atmospheric
pressure [cmH2O] / 1013 cmH2O
14. For setting 0 and T there should be no output of anesthetic agent.
• At 1 vol.%; 0.8 - 1.2 vol.% *
• At 2 vol.%; 1.8 - 2.2 vol.% *
• At 3 vol.%; 2.8 - 3.2 vol.% *
• At 4 vol.%; 3.8 - 4.2 vol.% *
• At 5 vol.%; 4.8 - 5.2 vol.% *
• At 6 vol.%; 5.7 - 6.3 vol.% *
• At 7 vol.%; 6.7 - 7.3 vol.% *
• At 8 vol.%; 7.7 - 8.3 vol.% *
* = Correct for temperature and carrier gas if necessary.
15. Switch off the vaporizer until 0 engages.
16. Switch off the AIR or O2 flow.

AS3000™ Service Manual 0070-10-0683 6 - 37


Functional Tests Periodic Maintenance

6.8.11 Electrical Tests


CAUTION: Perform the electrical safety inspection as the last step after
completing a repair or after routine maintenance. Perform
this inspection with all covers, panels, and screws installed.

6.8.11.1 Convenience AC Outlets Test


1. Verify AC voltage is present at each AC outlet with the AS3000 Mains switch in both
the ON and OFF positions.

6.8.11.2 Electrical Safety Inspection Test


NOTE: Perform the electrical safety inspection as the last step after
completing a repair or after routine maintenance. Perform
this inspection with all covers, panels, and screws installed.

1. Unplug the Power cable(s) from the convenience receptacles at the rear of the AS3000.
2. Plug the AS3000 into a Safety Analyzer.
3. Connect the case ground lead of the analyzer to the U-blade ground of one of the
convenience receptacles. Perform the following tests with the case grounded:
• Normal polarity
• Normal polarity with open neutral
4. Perform the following tests with the case ungrounded:
• Normal polarity
• Normal polarity with open neutral
• Reverse polarity
5. Verify that the maximum leakage current does not exceed 300 µA (0.3 mA).

6.8.12 AS3000 Installation Checklist


Refer to section 2.0 (pg. 2-1) ‘‘Installation Guide’’ for the installation checkout procedure.
Complete each step to check the functionality of the anesthesia machine prior to clinical use.
Also, perform the installation checkout procedure after installation, reinstallation, servicing or
after any periodic maintenance activity. This checklist does not replace periodic maintenance
actions that must be performed to maintain peak performance.

6 - 38 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Cleaning

6.9 Cleaning
6.9.1 Cleaning and Disinfecting the AS3000
Before cleaning, switch off the AS3000 and disconnect it from the mains.

NOTE: See the “AS3000 Operating Instructions”


(P/N: 0070-00-0684-XX) for more in-depth cleaning
instructions.

6.9.2 Cleaning and Sterilizing the Breathing System and Components


Cleaning method and chemical
STERILIZATION MAXIMUM
SOAP CHEMICALS STEAM TEMPERATURE
COMPONENT WATER CIDEX AUTOCLAVE (°F/°C)
Absorber canister yes yes no
Airway pressure gauge no no no
APL Valve yes yes no
Valve cover and valve nut yes yes no
Breathing System module yes yes yes 273°F/134°C
(without the above
mentioned)
AGSS yes yes no
The components in the above table should be rinsed in warm water thoroughly and dried in air after cleaning
or sterilizing.

6.10 Preoperative Checklist


This checklist should be conducted before administering anesthesia.

NOTE: This is a guideline which can be modified to accommodate


variations in local clinical practice. Such local modifications
should have appropriate peer review.

If an anesthetist uses the same machine in successive cases, this checkout need not be
repeated or may be abbreviated after the initial checkout.

1. Inspect the system for:


a. Identification number
b. Valid inspection sticker
c. Damage to flowmeters, vaporizers, gauges, and supply hoses
d. Complete Breathing System with adequate CO2 absorbent Pre-Paks or loose fill
e. Correct mounting of gas cylinders in the yokes
f. Presence of the Tank Wrench

AS3000™ Service Manual 0070-10-0683 6 - 39


Preoperative Checklist Periodic Maintenance

2. Per manufacturers’ specifications, turn ON the patient monitors to allow


time for their warm-up (ECG, Blood Pressure, SpO2, Gas Monitoring,
etc.).

3. Prepare the Anesthetic Gas Scavenging System (AGSS).


a. Remove the AGSS from the AS3000. While viewing the float, turn the AGSS upside
down to verify whether the float moves freely along its shaft. Replace the float as
necessary. Reconnect the AGSS to the AS3000.
b. Connect the vacuum hose to the vacuum port on the AGSS. Adjust the position of the
float to be between the Min and Max lines by turning its flow adjustment knob
(counterclockwise increases flow, clockwise decreases flow).
c. Drain any moisture from the waste gas hose. Connect the waste gas hose to the
AGSS waste gas port.

4. Verify that:
a. Flow-control valves are off
b. Vaporizers are off
c. Vaporizers are filled.
d. Vaporizer Filler caps are sealed tightly.

5. Check oxygen (O2) cylinder supply:


a. Disconnect line pressure hoses (if connected) and return the cylinder and line
pressure gauges to zero using O2 flush valve.
b. Open the O2 cylinder, and check for pressure.
c. Close the O2 cylinder and observe the pressure gauge for evidence of high-pressure
leaks.
d. Press the O2 flush valve to empty the piping.
e. A typical full O2 cylinder’s pressure is 1900 psi. Replace the cylinder if its pressure
is less than 1000 psi.
f. Reopen the cylinder.

6. Check nitrous oxide (N2O) cylinder supply:


a. Disconnect line pressure hoses (if connected) and return the cylinder and line
pressure gauges to zero using the flow control knobs.
b. Open the N2O cylinder, and check for pressure.
c. Close the N2O cylinder and observe the pressure gauge for evidence of high-
pressure leaks.
d. A typical full N2O cylinder’s pressure is 745 psi. Replace the cylinder if its pressure
is less than 600 psi.
e. Reopen the cylinder.

6 - 40 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Preoperative Checklist

7. Check AIR cylinder supply:


a. Disconnect line pressure hoses (if connected) and return the cylinder and line
pressure gauges to zero using the flow control knobs.
b. Open the AIR cylinder, and check for pressure.
c. Close the AIR cylinder and observe the pressure gauge for evidence of high-pressure
leaks.
d. A typical full AIR cylinder’s pressure is 1900 psi. Replace the cylinder if its pressure
is less than 1000 psi.
e. Reopen the cylinder.
8. Power up the system and follow the on-screen prompts to perform the
Leak and Compliance Tests.

9. Test Flowmeters:
a. Check that all floats are at the bottom of the flow tubes with the flow control valves
closed.
b. Adjust the flow of all gases through their full range and check the floats for erratic
movements.

10. Test Hypoxic-Guard System:

For -01 units:


a. Attempt to create hypoxic O2/ N2O mixture by slowly opening the N2O Flow
Control Valve.
b. Continue to increase the N2O flow and observe O2 and N2O rise in proportion to
maintain a minimum concentration of 21% O2 in fresh gas.

For -02 units:


a. Attempt to create hypoxic O2/ N2O mixture by opening the N2O flow control valve
completely.
b. Increase the O2 flow and observe O2 and N2O rise in proportion to maintain a
minimum concentration of 21% O2 in fresh gas.

11. Test Line Pressure Gas Supplies:


a. Inspect the supply hoses (should not be cracked or worn).
b. Connect the supply hoses, verifying correct color coding.
c. Adjust both the O2 and N2O flows to at least mid-range.
d. Verify that the O2 and N2O supply pressures hold (45–55 psi).
e. Shut off the O2 and N2O flow control valves.
f. Adjust the AIR flow to at least mid-range.
g. Verify that the AIR supply pressure hold (45–55 psi).
h. Shut off the AIR flow control valve.

12. Accessories Connection:


a. Connect a breathing circuit to the Breathing System.
b. Connect a breathing bag to the bag arm.

AS3000™ Service Manual 0070-10-0683 6 - 41


Preoperative Checklist Periodic Maintenance

13. Check Unidirectional Valves:


a. Set the Ventilation mode to STANDBY.
b. Set the APL Valve to 20 cmH2O
c. Attach a test lung to the Y-fitting of the breathing circuit.
d. Set the AIR flow to 5 L/min.
e. Press the O2 Flush Valve to fill the breathing bag with volume.
f. Slowly squeeze the breathing bag once every 10 seconds.
g. Verify that the test lung inflates and deflates.
h. Verify that the inspiratory unidirectional valve opens when the test lung inflates.
i. Verify that the expiratory unidirectional valve opens when the test lung deflates.

14. Test drive gas pressure failure system:


a. Set the O2, N2O, and AIR gas flows to mid-range using the flow control knobs.
b. Close the O2 gas cylinder using the Tank Wrench.
c. Verify that the O2 and N2O flows fall to zero.
d. Verify that:
• An alarm tone sounds.
• The alarm message O2 Supply Failure is displayed on the screen.
e. Open the O2 gas cylinder. Verify that the alarm tone stops and the alarm message is
removed from the screen.
f. Close all gas cylinders.
g. Remove the O2 cylinder from the AS3000.
h. Verify that:
• An alarm tone sounds.
• the alarm message O2 Supply Failure is displayed on the screen.
i. Shut off all flow control valves.

15. High Pressure Leak Test:


a. Connect a breathing circuit and breathing bag to the Breathing System.
b. Set APL Valve to 70 cmH2O.
c. Connect the breathing circuit Y-fitting to the test port on the AS3000.
d. Inflate breathing bag by pressing the O2 Flush Valve until the pressure gauge
reaches 40 cmH2O.
e. Verify that breathing circuit holds pressure for a minimum of 10 seconds.

6 - 42 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Preoperative Checklist

16. Test the Ventilator in STANDBY mode:


a. Set the APL Valve to 20 cmH2O.
b. Attach a test lung to the breathing circuit Y-fitting.
c. Set the AIR flow to 5 L/min.
d. Press the O2 Flush Valve to fill the breathing bag with volume.
e. Squeeze the breathing bag once every 10 seconds.
f. Verify that the test lung inflates and deflates.
g. Set the APL Valve to the open position (SP) and stop squeezing the breathing bag.

17. Test the Ventilator in MANUAL mode:


a. Set the ventilation mode to MANUAL.
b. Set the APL Valve to 20 cmH2O.
c. Set the AIR flow to 5 L/min.
d. Squeeze the breathing bag once every 10 seconds.
e. Verify that the test lung inflates and deflates to approximately 20 cmH2O pressure.
f. Verify that tidal volume, resp. rate, minute volume, and PEEP values appear on the
screen.
g. Verify that the FiO2 reading is approximately 21.
h. Verify a pressure waveform appears on the screen along with the bag compressions.
i. Set the APL Valve to the open position (SP) and stop squeezing the breathing bag.

18. Test the Tidal Volume in CMV mode:


a. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
b. Set the ventilator controls to:

ATTRIBUTE SETTING
Patient Type Adult
Ventilation Mode CMV
Tidal Volume - VT 600
Breathing Rate - freq 8
I:E Ratio - I:E 1:2
Plateau - TP 10
PEEP - PEEP OFF

c. Select CMV to begin ventilation.


d. Verify that the Tidal Volume display is within 15% of the set value within 5 breaths.
e. Verify that the O2 display reads greater than 95% within 5 minutes.

AS3000™ Service Manual 0070-10-0683 6 - 43


Preoperative Checklist Periodic Maintenance

19. Test the Tidal Volume in the PCV ventilation mode:


a. Set the O2 flow to 2 L/min and set the N2O and AIR flow rates to minimum flow.
b. Set the ventilator controls to:

ATTRIBUTE SETTING
Patient Type Adult
Ventilation Mode PCV
Target Pressure - PTARGET 20
Breathng Rate - freq 8
I:E Ratio - I:E 1:2
PEEP - PEEP Off
Inspiratory Slope - Tslope 0.5

c. Select PCV to begin ventilation.


d. Verify that the PEAK Pressure settles within ±4 of the set value within 3 breaths.
e. Re-activate CMV ventilation mode.

20. Low FiO2 Alarm Test


a. Set the low FiO2 Alarm limit to 50% O2.
b. Set the AIR flow control valve to 5 L/min.
c. Set the FiO2 flow controller to minimum flow.
d. Verify the following Low FiO2 alarm signals activate, within three ventilation cycles:
• Low FiO2 message appears on the screen.
• An alarm tone sounds.
e. Set the Low FiO2 alarm limit to 18%.
• Verify the alarm signals stop activating.

21. High FiO2 Alarm Test


a. Set the high FiO2 Alarm limit to 49% O2.
b. Set the FiO2 flow control valve to 5 L/min.
c. Set the AIR flow controller to minimum.
d. Verify the following High FiO2 alarm signals activate:
• High FiO2 message appears on the screen.
• An alarm tone sounds.
e. Set the high FiO2 alarm limit to the max setting.
• Verify the alarm signals stop activating.

6 - 44 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Preoperative Checklist

22. High and Low PAW Alarm Test


a. Set the PAW low alarm to the lowest setting.
b. Set the PAW high alarm limit set point about 5 to 8 digits below the Peak Pressure
displayed on the upper left of the screen.
c. Verify the following (high) peak pressure alarms activate:
• High Airway Pressure message appears on the screen.
• An alarm tone sounds.
• Inspiration ends and expiration begins as the pressure meets the high alarm limit.
a. Set the PAW high alarm limit set point to 65 (cmH2O).
b. Verify the alarms signals cease.
c. Set the PAW low alarm limit set point to 50 (cmH2O).
d. Verify the following (low) peak pressure alarms activate:
• Low Airway Pressure message appears on the screen.
• An alarm tone sounds.
a. Set the PAW low alarm limit to 12 (cmH2O).
b. Verify the alarm signals cease.

23. Low MV and APNEA Alarm Test


a. Set the PAW low alarm to the lowest setting.
b. Set the PAW high alarm limit set point about 5 to 8 digits below the Peak Pressure
displayed on the upper left of the screen.
c. Verify the following (high) peak pressure alarms activate:
• High Airway Pressure message appears on the screen.
• An alarm tone sounds.
• Inspiration ends and expiration begins as the pressure meets the high alarm limit.
d. Set the PAW high alarm limit set point to 65 (cmH2O).
e. Verify the alarms signals cease.
f. Set the PAW low alarm limit set point to 50 (cmH2O).
g. Verify the following (low) peak pressure alarms activate:
• Low Airway Pressure message appears on the screen.
• An alarm tone sounds.
h. Set the PAW low alarm limit to 12 (cmH2O).
i. Verify the alarm signals cease.
j. Press the MANUAL/AUTO key.
k. After 60 seconds, verify that:
• An alarm tone sounds.
• the alarm message “APNEA” is displayed on screen.

AS3000™ Service Manual 0070-10-0683 6 - 45


Preoperative Checklist Periodic Maintenance

24. Place the system in STANDBY mode.

25. Check for appropriate level of patient suction.

26. Check, connect, and calibrate other electronic monitors.

27. Turn on and set other appropriate alarms for equipment to be used.

NOTE: The following step should be performed every 3 days or


when prompted by the machine.

28. Perform the Oxygen Sensor Calibration (Refer to section 6.8.4.6 (pg. 6-
22) ‘‘Oxygen Sensor Calibration’’).

NOTE: The following step should be performed weekly or


whenever a new vaporizer is installed or when CO2
absorbent is replaced.

29. Test for leaks in the machine and vaporizers by performing the High
Pressure Leak Test as described in step 15 of this section.

6 - 46 0070-10-0683 AS3000™ Service Manual


Periodic Maintenance Phone Numbers and How To Get Assistance

6.11 Phone Numbers and How To Get Assistance


A network of service representatives and factory-trained distributors is available. Prior to
requesting service, perform a complete operational check of the instrument to verify proper
control settings. If operational problems continue to exist, contact the Service Department at
(800) 288-2121, ext: 8116 for Technical Support or (201) 995-7875 for assistance in
determining the nearest field service location.

Please include the instrument model number, the serial number, and a description of the
problem with all requests for service.

Warranty questions should be directed to a local representative. A list of offices, along with
their phone numbers, is provided at the end of this manual.

Upon request, calibration instructions or other information will be provided to assist


the user’s appropriately qualified technical personnel in repairing those parts of the
AS3000 which are designated as repairable.

AS3000™ Service Manual 0070-10-0683 6 - 47


Phone Numbers and How To Get Assistance Periodic Maintenance

This page intentionally left blank.

6 - 48 0070-10-0683 AS3000™ Service Manual


7.0 Warranty

7.1 Warranty Statements


Mindray DS USA, Inc. warrants that components within the anesthesia system will be free
from defects in workmanship and materials for the number of years shown on the invoice.
Under this extended warranty, Mindray DS USA, Inc. will repair or replace any defective
component at no charge for labor and/or materials. This extended warranty does not cover
consumable items such as (but not limited to) batteries and external cables.

Recommended preventative maintenance, as prescribed in the Service Manual, is the


responsibility of the user, and is not covered by this warranty.

Except as otherwise provided herein, the terms, conditions, and limitations of Mindray DS
USA, Inc.’s standard warranty will remain in effect.

Mindray DS USA, Inc. warrants that its products will be free from defects in workmanship
and materials for a period of one (1) year from the date of purchase except that disposable
or one-time use products are warranted to be free from defects in workmanship and materials
up to a date one year from the date of purchase or the date of first use, whichever is sooner.
This warranty does not cover consumable items such as, but not limited to, batteries, external
cables, sensors, cuffs, hoses, or mounts.

Mindray DS USA, Inc. will not be liable for any incidental, special, or consequential loss,
damage, or expense directly or indirectly arising from the use of its products, liability under
this warranty and the buyer’s exclusive remedy under this warranty is limited to servicing or
replacing at Mindray DS USA, Inc.’s option at the factory or at an authorized distributor, any
product which shall under normal use and service appear to the Company to have been
defective in material or workmanship.

AS3000™ Service Manual 0070-10-0683 7-1


Disclaimers Warranty

No agent, employee, or representative of Mindray DS USA, Inc. has any authority to bind
Mindray DS USA, Inc. to any affirmation, representation, or warranty concerning its
products, and any affirmation, representation or warranty made by any agent, employee, or
representative shall not be enforceable by buyer.

This warranty is expressly in lieu of any other express or implied warranties, including any
implied warranty or merchantability or fitness, and of any other obligation on the part of the
seller.

Damage to any product or parts through misuse, neglect, accident, or by affixing any non-
standard accessory attachments or by any customer modification voids this warranty.
Mindray DS USA, Inc. makes no warranty whatever in regard to trade accessories, such
being subject to the warranty of their respective manufacturers.

A condition of this warranty is that this equipment or any accessories which are claimed to
be defective be returned when authorized, freight prepaid to Mindray DS USA, Inc.,
Mahwah, New Jersey 07430. Mindray DS USA, Inc. shall not have any responsibility in the
event of loss or damage in transit.

Calibration may be performed without the need to disassemble the instrument. It is the
responsibility of the purchaser to perform calibration as necessary, in accordance with the
instructions provided in this manual.

7.2 Disclaimers
7.2.1 Product Improvements
Mindray DS USA, Inc. retains the right to modify the machine and/or operating instructions
without prior notification. These operating instructions explain all features of the AS3000
system and are correct at time of manufacture. Instructions and models produced at a later
stage, may contain improvements or modifications that were not included in previous
models.

7.3 Manufacturer’s Responsibility


The effects on safety, reliability, and performance of the equipment are the manufacturer’s
responsibility only if:

a. assembly operations, extensions, readjustments, modifications or repairs are carried


out by authorized personnel; and
b. the electrical installation of the relevant room complies with the appropriate
requirements; and
c. the equipment is used in accordance with the instructions for use

7-2 0070-10-0683 AS3000™ Service Manual


Warranty Manufacturer’s Responsibility

This page intentionally left blank.

AS3000™ Service Manual 0070-10-0683 7-3


0070-10-0683 Rev H November 24, 2010
Mindray DS USA, Inc. • 800 MacArthur Boulevard • Mahwah, NJ 07430 • USA •
Dom. Customer Service: 1.800.288.2121 • Intl. Customer Service: +1.201.995.8000 •
Dom. Fax: 1.800.926.4275 • Intl. Fax: +1.201.995.8680 • www.mindray.com

Mindray Medical Netherlands B.V.• P.O. Box 26 • 3870 CA Hoevelaken • The Netherlands •
Tel: +31 33 25 44 911 • Fax: +31 33 25 37 621

Mindray (UK) Limited • 3 Percy Road • St. John’s Park • Huntingdon • Cambridgeshire PE29 6SZ •
United Kingdom • Tel: 01480 416840 • Fax: 01480 436588

Mindray Medical France SARL • Europarc Créteil •123, Chemin des Bassins •
94035 Créteil Cedex • France • Tel: (0)1.45.13.91.50 • Fax: (0)1.45.13.91.51

Mindray Medical German GmbH • Zwischen den Bächen 4 • 64625 Bensheim • Germany •
Tel: +49.6251.17524-0 • Fax: +49.6251.17524-20

Mindray Medical International Ltd. • 2813 Office Tower, Convention Plaza • No 1 Harbour Road •
Wanchai • Hong Kong • Tel: +852 2793 5596 • Fax: +852 2344 8824

0002-08-8918

You might also like