Ventilador de Anestecia AV 800 Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 88
At a glance
Powered by AI
The document discusses servicing requirements for an anaesthesia ventilator and provides information in the service manual.

The ventilator must be serviced every 6 months for inspection and function testing, annually, and undergo a major service every 5 years including battery replacement.

The service manual provides information on the function, routine performance, servicing, maintenance checks and repairs applicable to the AV800 Anaesthesia Ventilator.

AV800

Anaesthesia Ventilator
Service Manual

Quality and Assurance in Anaesthesia

IMPORTANT
Servicing and Repairs
In order to ensure the full operational life of this
ventilator, servicing by a Penlon-trained
engineer should be undertaken periodically.
The ventilator must be serviced to the following
schedule:
(a)
(b)
(c)

Six monthly service - inspection and


function testing.
Annual service.
Five year major service including battery
replacement.

Details of these operations are given in this


AV800 Service Manual, available only for
Penlon trained engineers.
For any enquiry regarding the servicing or
repair of this product, contact the nearest
accredited Penlon agent:

or communicate directly with:


Technical Support
Penlon Limited
Abingdon
OX14 3PH
UK
Tel:
+44 (0) 1235 547076
Fax:
+44 (0) 1235 547062
E-mail: [email protected]
Always give as much of the following
information as possible:
1.
2.
3.
4.
5.

Type of equipment
Product name
Serial number
Approximate date of purchase
Apparent fault

(i)

FOREWORD
This manual has been produced to provide
authorised personnel with information on the
function, routine performance, servicing,
maintenance checks and repairs applicable
to the AV800 Anaesthesia Ventilator.
Information contained in this manual is
correct at the date of publication.
The policy of Penlon Limited is one of
continued improvement to its products.
Because of this policy, Penlon Limited
reserves the right to make any changes
which may affect instructions in this manual,
without giving prior notice.

Personnel must make themselves familiar


with the contents of this manual and the
machines function before servicing the
apparatus.

Copyright Penlon Limited, 2003.


All rights reserved.

(ii)

CONTENTS
Page No.

1.
2.

USER RESPONSIBILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1
2
7

3.
3.1
3.2
3.3
3.4

DESCRIPTION
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ventilation Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Control Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

8
10
13
15

3.5
3.5.1
3.5.2
3.5.3
3.5.4
3.5.5
3.5.6
3.5.7
3.5.8

Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tidal Volume Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ventilation Rate Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ventilation I:E Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Airway Pressure Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tidal Volume Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Inspiratory Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

17
17
18
18
18
19
19
19
19

3.6
3.7
3.8
3.9
3.10

Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Message Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Back-up Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Operational Capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
System software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

20
22
23
24
25

4.

SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

27

5.
5.1
5.1.1
5.1.2
5.1.3
5.1.4
5.1.5
5.2
5.3
5.3.1
5.3.2

PRE-OPERATION PROCEDURES
Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Components Supplied with the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . .
Mounting the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ventilator Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ventilator Gas Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spirometer Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Bellows Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Pre-use Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Daily Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Weekly Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

30
30
30
30
30
34
35
36
36
37

6.

FAULT FINDING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

38

7.

SERVICE SCHEDULE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

43

8.
8.1
8.2
8.3
8.4

SERVICE PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sterilisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tubing Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Fitting the Annual Preventive Maintenance Kit . . . . . . . . . . . . . . . . . . . . . .

50
50
53
54
55

(iii)

CONTENTS
8.5
8.6
8.7
8.8
8.9

Fitting the Five Year Overhaul Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .


Manifold Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Electrical Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Front Panel and PCB Assembly, and Gauge . . . . . . . . . . . . . . . . . . . . . . .
Main PCB Tray Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

56
58
58
59
60

9.

PARTS LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

61

10.

APPENDIX
Back-up Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

73

(iv)

USER RESPONSIBILITY
This anaesthesia ventilator has been built to
conform with the specification and operating
procedures stated in this manual and/or
accompanying labels and notices when
checked, assembled, operated, maintained
and serviced in accordance with these
instructions.

Statements in this manual preceded by the


following words are of special significance:

To ensure the safety of this device it must be


checked and serviced to at least the
minimum standards laid out in this manual.
A defective, or suspected defective, product
must not under any circumstances be used.
The user must accept responsibility for any
malfunction which results from noncompliance with the servicing requirements
detailed in this manual.

WARNING

means there is a
possibility of injury to
yourself or others.

CAUTION

means there is a possibility


of damage to the apparatus
or other property.

NOTE

indicates points of
particular interest for more
efficient and convenient
operation.

Always take particular notice of the


warnings, cautions and notes provided
throughout this manual.

Additionally, the user must accept


responsibility for any malfunction which may
result from misuse of any kind or noncompliance with other requirements detailed
in this manual.
Worn, broken, distorted, contaminated or
missing components must be replaced
immediately. Should such a repair become
necessary it is recommended that a request
for service advice be made to the nearest
accredited Penlon agent.
This device and any of its constituent parts
must be repaired only in accordance with
written instructions
issued by Penlon
Limited and must not be altered or modified
in any way without the written approval of
Penlon Limited. The user of this equipment
shall have the sole responsibility for any
malfunction which results from improper
use, maintenance, repair, damage or
alteration by anyone other than Penlon or its
appointed agents.
USA and Canadian Federal Law restricts the
sale and use of this device to, or on the order
of, a licensed practitioner.

1. WARNINGS AND CAUTIONS


cannot be guaranteed.
See
section
8
for
maintenance.
See also CAUTION No. 7.

The following WARNINGS and CAUTIONS


must be read and understood before
servicing or repairing this ventilator.

WARNINGS

6.

No oil, grease or other flammable


lubricant or sealant must be used
on any part of the machine in close
proximity
to
medical
gas
distribution components.
There is a risk of fire or explosion.

7.

Exterior panels must not be


removed
by
unauthorised
personnel and the apparatus must
not be operated with such panels
missing.
There is a possible electric shock
hazard.
Always disconnect the ventilator
from the mains electrical supply
and drive gas supply before
removing the cover.

General Information
1.

Personnel must make themselves


familiar with the contents of this
manual and the machines function
before using the ventilator.

Servicing and Repair


2.

Opening the control unit by


unauthorised
personnel
automatically voids all warranties
and specifications.
Prevention of tampering with the
control unit is exclusively the
users responsibility. If the control
unit seal is broken,
the
manufacturer assumes no liability
for any malfunction or failure of the
ventilator.

3.

Before Using the Ventilator

Great care must be taken not to


damage the precision surface of
the valve seat on the patient gas
exhalation diaphragm valve in the
base of the bellows assembly.

8.

Before the ventilator is used


clinically for the first time, verify
that the hospital engineering
department has carried out an earth
continuity test.

9.

Excessive electronic noise caused


by other poorly regulated devices,
such as an electrocautery unit, may
adversely interfere with the proper
functioning of the ventilator.

Never use any hard object or


abrasive detergent to clean it; use
only a soft cloth.
If the valve seat is damaged, the
valve will leak and may cause
serious ventilator malfunction.
4.

5.

To avoid this problem, do not


connect the ventilators power cord
into the same electrical wall outlet
or adaptor strip into which an
electrocautery unit is connected.

For continued protection against


fire hazards, replace the two fuses
only with the identical type and
rating of fuse.
See section 4.39 for fuse rating.
If the internal battery is fully
discharged, the ventilator will not
function in the event of mains
power failure. The battery must be
recharged before the ventilator is
used clinically, otherwise backup

battery

10.

If used with a mains extension cord,


the unit may be subject to electromagnetic interference.

11.

The driving gas supply must be


clean and dry to prevent ventilator
malfunction.

WARNINGS AND CAUTIONS


12.

This ventilator is designed to be


driven by oxygen or medical air
only. It is calibrated during
manufacture for use with either
gas.
Before the ventilator is used
clinically for the first time, the
commissioning engineer must
confirm
that
the
internal
Air/Oxygen switch is set correctly
for the gas that is to be used.
The use of any other gas will cause
inaccurate operation and may
damage the ventilator, resulting in
potential injury to the patient.

13.

The driving gas is discharged


through the opening in the back of
the ventilator control unit. This
opening is labelled: EXHAUST DO
NOT BLOCK and must be
completely free of any obstruction
and nothing should be connected
to it.

15.

Do not use conductive or anti-static


breathing system hoses.
16.

The discharged gas does not


contaminate the environment, but
on machines using oxygen as the
drive gas it can cause a fire hazard
if allowed to accumulate.

Applying negative or positive


pressure to the exhaust port may
result in positive or negative
pressure in the patient breathing
system. Therefore the scavenging
system must not generate more
positive or
than 0.5 cmH2O
negative pressure when connected
to the ventilator.
Any problem arising from an
improperly functioning scavenging
system is solely the users
responsibility.

In addition, do not block the PRV


outlet on the back of the control
unit,
and also the pneumatic system
outlet on the base of the unit.
14.

The breathing system which


conveys
gases
from
the
anaesthetic machine to the patient,
and disposes of expired gases, is a
vital part of the anaesthetic delivery
system.
Because breathing systems require
frequent cleaning and disinfection
they are not a permanent part of the
anaesthetic
ventilator
and
therefore cannot be directly under
the control of the anaesthetic
ventilator manufacturer.
However, we strongly recommend
that only breathing systems which
have
been
approved
and
authorised by Penlon for use with
AV800 should be employed.

The bellows can only support


approximately 1 kPa (10 cmH2O)
differential positive pressure,
above which it may be dislodged
from the mounting ring, resulting in
dangerous malfunction of the
ventilator.
Do not connect a PEEP valve or
other restrictive device to the
exhaust port on the bellows base.
This would increase the pressure
inside the bellows and the bellows
could detach from the base,
causing serious malfunction.
3

17.

Do not connect a spirometer to the


exhaust port on the bellows base.
The device will not measure
exhaled volumes.

18.

The operation of each alarm


function should be verified daily.
See section 5.3.1.
Periodically check the alarms at
suitable intervals. If the audible
alarm or the visual indicator of any
alarm function fails to activate
during any alarm condition or fails
to reset after the alarm has been
cleared, refer the unit to an
authorised service technician.

WARNINGS AND CAUTIONS


19.

Before using the ventilator check


that all connections are correct,
and verify that there are no leaks.

monitored continuously, at or near


the proximal airway with an oxygen
monitor that includes high/low
alarms.

Patient circuit disconnects are a


hazard to the patient. Extreme care
should be taken to prevent such
occurrences.

25.

If the drive gas supply pressure


drops below a nominal 262 kPa (38
psig), the LOW DRIVE GAS
SUPPLY alarm will activate both
audibly and visually.
Patient
minute volume may be reduced due
to lowered flow rates

26.

An audible alarm indicates an


anomalous condition and should
never go unheeded.

27.

The characteristics of the breathing


circuit connected between the
ventilator and the patient can
modify
or
change
patient
ventilation.

It is recommended that Penlon


Safelok
fittings
are
used
throughout the breathing circuit.

Using the Ventilator


20.

This apparatus must not be used


with, or in close proximity to,
flammable anaesthetic agents.
There is a possible
explosion hazard.

21.

22.

23.

24.

fire

or

To assist the maintenance of the


delivered patient tidal volume, the
ventilator control system software
includes
a
compliance
compensation algorithm.

Anaesthesia apparatus must be


connected to an anaesthetic gas
scavenging system to dispose of
waste gas and prevent possible
health hazards to operating room
staff. This requirement must be
observed during test procedures
as well as during use with a patient.
Any problem arising from an
improperly functioning scavenging
system is solely the users
responsibility.

However, patient ventilation must


be monitored independently from
the ventilator.
It is the responsibility of the user to
monitor patient ventilation.
28.

When the ventilator is connected to


a patient, it is recommended that a
qualified
practitioner
is
in
attendance at all times to react to
an alarm or other indication of a
problem.

On models with spirometry, the


spirometer flow sensor must be
installed in the expiratory limb of
the breathing circuit.
A breathing system filter, or heat
and moisture exchanger (HME),
must be connected upstream of the
sensor to prevent blockage of the
side-stream sample lines

In
compliance
with
good
anaesthesia practice, an alternative
means of ventilation must be
available whenever the ventilator is
in use.

29.

The AV800 is not equipped with an


oxygen
analyser.
It
is
recommended that the patient
oxygen
concentration
be
4

The
Vent
Inop
(ventilator
inoperative) alarm indicates that
one of the following conditions has
occurred:
A) A solenoid has failed.
B) The flow control valve has failed.
C) Internal electrical fault.
D) Internal communications error

WARNINGS AND CAUTIONS


E) Battery disconnected
F) Software error
Note that if a ventilator error is
detected, an error code will be
displayed in the Rate BPM display
window on the front control panel.
30.

The High and Low Airway Pressure


Alarms are important for patient
care.
The ventilator is designed to be
used with a distal sensing tee only.
(Catalogue No. 53194, Breathing
System Tee Assembly - see section
8 in the user manual).
The distal sensing tee must be
located close to the circle system,
in the inspiratory limb of the circuit
between the patient and the circle
system inspiratory one way valve.
See section 5.1.4.

Do not sterilise the ventilator control


unit.
The internal components are not
compatible
with
sterilisation
techniques and damage may result.

2.

For ventilator components which


require sterilisation, peak sterilisation
temperatures should not exceed
136oC (275oF) to prevent possible
damage. (See sections 7.2 and 7.3).

3.

Those parts suitable for ethylene


oxide sterilisation should, following
sterilisation, be quarantined in a well
ventilated area to allow dissipation of
residual gas absorbed by the
components.

The exhalation valve located in the


bellows base assembly and the
paediatric bellows adaptor must be
cleaned and sterilised separately. See
sections 7.2 and 7.3.

5.

Care must be taken not to let any


liquid run into the control unit; serious
damage may result.

6.

Always check for correct fitment, and


carry out a full function test before
clinical use, if the bellows has been
removed and refitted for any reason.
See sections 5.2 and 7.2.

7.

Damage may occur to the battery if it


is allowed to remain in a discharged
state.
Check the battery frequently if the
ventilator is in storage (see Appendix
1).

CAUTIONS
1.

4.

Follow the steriliser manufacturers


recommendations for any special
aeration periods required.

2. PURPOSE

The AV800 Ventilator is a software


controlled, multi-mode ventilator, designed
for mechanical ventilation of adult and
paediatric patients
under general
anaesthesia.
In addition, in spontaneous mode, it can be
used to monitor spontaneously breathing
patients
It is designed for use in closed-circuit
anaesthesia and also to drive a Mapleson D
circuit.

3. DESCRIPTION

Bellows Housing

Bellows Base

Control Unit

Fig. 1 AV800 Ventilator

3.1

General

The ventilator drive gas supply can be


oxygen or air, and the supply must be at 38
to 100 psig. Note that the drive gas is
specified by the customer prior to delivery.
To change the drive gas, refer to a Penlontrained service engineer.

The AV800 is a multi-mode, time-cycled,


volume/pressure controlled, and pressure
limited ventilator for closed circuit ventilation
or for use with a Mapleson D circuit.
The ventilator is compliance compensated
and has a user selectable option of an
inspiratory pause fixed at 25% of the
inspiratory time.
The print function provides a permanent
record of function activity for up to eight
hours during a procedure.
The bellows unit can be easily detached and
then refitted to the bellows base assembly to
facilitate cleaning.

Models are available with a manometer


and/or spirometry, and the ventilator can be
used with power supplies from 110 to 240
VAC.
All components used in the ventilator,
including the bellows, are latex-free ( but
note that the optional paediatric bellows
is not latex free).
8

Driving Gas
Port
Exhaust Port
(patient gas)
Do NOT connect
spirometer

Breathing
System
Port

Spirometer
connectors
DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT

USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS

RS232 PORT

CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE


COVER.

REFER TO QUALIFIED SERVICE PERSONNEL.

PRINTER PORT

ANALOGUE/ALARM OUTPUTS

SPIRO

OXYGEN/AIR
38 - 100 psi
2.6 - 6.9 Bar
262 - 689 kPa

MANOMETER/
PRESSURE
TRANSDUCER
-20 To 100 cmH2O

EXHAUST VALVE
DO NOT BLOCK

Drive Gas

MAINS SUPPLY

FUSE RATING/TYPE
3.15 A 240 V ANTI SURGE HRC

PRV 80 cmH2O
Penlon Limited
Abingdon, Oxon
OX14 3PH
Tel 01235 547000
Fax 01235 547031
Tested by:
Serial No.
MADE IN UK

CLASS 1 TYPE B
VOLTAGE 90 - 264 VAC
FREQUENCY 47 - 63 Hz
MAX CURRENT 1 A

Fig. 2 Ventilator Rear Panel and Gas Ports

DESCRIPTION
3.2

Ventilation Cycle

This section provides a simplified


description of the ventilation cycle.

1.

Beginning of
Inspiratory Phase
Drive gas pressure builds up
above the bellows, which starts
to move down, forcing patient
gas into the breathing system.

PATIENT
GAS OUT TO
BREATHING
CIRCUIT
DRIVE GAS IN

2.

End of Inspiratory
Phase
The main drive gas valve
closes and the bellows stops
moving.

MAIN DRIVE
GAS VALVE
CLOSED

10

DESCRIPTION

3.

Beginning of
Expiratory Phase
The discharge valve opens,
allowing the drive gas above
the bellows to escape to
atmosphere.
The bellows starts to rise and
exhaled gas enters the bellows.

DISCHARGE
VALVE

BELLOWS
EXHALATION
DIAPHRAGM
VALVE

4.

End of Expiratory Phase


The
bellows
exhalation
diaphragm valve in the base of
the bellows assembly opens
when the bellows reaches the
top of the chamber.
Patient circuit gas exits through
the bellows assembly exhaust
port.

BELLOWS
EXHAUST PORT

11

PNEUMATIC SYSTEM COMPONENTS


1.
2.
3.
4.
5.
6.
7.
8.

17

9.
10.

11.
12.
13.
14.
15.
16.
17.
18.

Gas supply inlet


Input gas filter
Low supply pressure detector
Input pressure regulator
Test point
Inspiratory valve
Drive gas flow metering valve
Stepper motor and feedback
potentiometer (for 7)
Expiratory valve
Exhaust outlets (outlet on rear panel
as illustrated, the outlet through the
base of the control unit is not shown)
Pressure relief valve
Exhaust valve
Bellows drive gas connector
Driving gas over-pressure switch
Bellows driving gas port
Pressure gauge
Pressure transducer
Inlet from breathing circuit /
manometer connection

4
8

9
7

15
2

1
14
12
18

11

10

13

12

DESCRIPTION
3
4

16

1
2
9

11

17

10

10

12
14
13

18
15

PNEUMATIC SYSTEM DIAGRAM


1.
2.
3.
4.
5.
6.
7.
8.
9.

Gas supply
Input gas filter
Low supply pressure detector
Input pressure regulator
Test point
Inspiratory valve
Drive gas flow metering valve
Stepper motor and feedback
potentiometer (for 7)
Expiratory valve

10.
11.
12.
13.
14.
15.
16.
17.
18.

3.3 Pneumatic System

Exhaust outlets
Pressure relief valve
Exhaust valve
Bellows drive gas connector
Driving gas over-pressure switch
Bellows driving gas port
Pressure gauge
Pressure transducer
Inlet from breathing circuit / manometer
connection

The drive gas is filtered with a 40-micron


Input Gas Filter
which protects the
pneumatic components from incoming
particulate matter.

Refer to the system diagrams above.


The AV800 Ventilator is designed to operate
on a 38-100 psig drive gas supply (oxygen or
air - to customers requirement).
The gas source is connected to the Drive
Gas Supply DISS fitting on the rear of the
ventilator control unit and should be capable
of a flow rate of 75 L/min while maintaining a
minimum pressure in excess of 38 psig.

The Low Supply Pressure Detector is a


pressure switch set at a predetermined level
to detect a loss or reduction of the input gas
source pressure. When the pressure falls
below 38 psig (262 kPa), the LOW SUPPLY
PRESSURE indicator will illuminate and the
high priority alarm will activate.

13

DESCRIPTION
The Input Pressure Regulator conditions the
input drive gas to a stable 35 psig pressure
which will operate the internal pneumatic
system.

spirometry, the actual tidal volume


exhaled will be displayed.
The patient must be monitored
independently from the ventilator
It is the responsibility of the user to
monitor the patient for adequate
ventilation.

The Inspiratory Valve is a large orifice,


electro-pneumatically-driven valve which
supplies the drive gas to the Drive Gas Flow
Metering Valve .

A compliance compensation algorithm is


built into the control software which
monitors the volume of gas delivered and
the rate of pressure rise.
It calculates an additional volume to deliver
into the breathing system to compensate for
the reduced volume delivered to the
patient as a result of the increased
breathing system pressure.
As a safety feature, this additional volume is
restricted to a maximum of 15% of the set
tidal volume irrespective of the maximum
pressure rise.

The Drive Gas Flow Metering Valve is a


variable-orifice
needle
valve
which
determines the drive gas flow rate of the
bellows during inspiration.
The Valve Position Feedback Potentiometer
and the Flow Control Motor
function
together to set a flow rate as required by the
front panel controls.
The flow from the Drive Gas Flow Metering
Valve goes to the Bellows Assembly, via the
drive connector, closes the exhalation
diaphragm valve and pushes the bellows
downward.
As the bellows moves downwards, the gas
inside the bellows is forced into the
Breathing System.

User Selectable Inspiratory Pause


A user selectable inspiratory pause is
provided which, when activated, holds the
inspiratory phase at the end of inspiration,
for a period of 25% of the inspiratory time
before reverting to the expiratory phase.

At the end of inspiration the exhaust valve


opens and allows the drive gas in the top of
the bellows housing to exhaust out through
the Exhaust Outlet.
As the pressure in the top of the bellows
housing is reduced to zero, the patient
exhales into the breathing system and the
bellows rises.

NOTE
This function is not available in
PRESSURE controlled cycle

Automatic altitude compensation


Ambient pressure is monitored and the
ventilator automatically compensates the
delivered volume according to the local
atmospheric pressure.

Compliance Compensation
WARNING
The
AV800
has
compliance
compensation but the actual tidal volume
delivered to the patient may be different
to the ventilation parameters set by the
user due to:
A) an extreme compliance condition,
B) a substantial system leak, or
C) patient circuit pressure effects.
In addition, high fresh gas flows will lead
to an increased Vt being delivered to the
patient.
Note that on models fitted with

Manometer (if fitted)


The manometer connection is connected
to the pressure gauge.

Patient Gas Pressure Transducer


The Patient Gas Pressure Transducer
is
connected to the patient breathing system
via the manometer connection.
In VOL CYCLE mode whenever this
pressure exceeds the maximum working
14

DESCRIPTION
pressure as set on the front panel an alarm
is activated and the following occurs:

3.4

1.

Ventilator parameters:

2.
3.

The HIGH AIRWAY PRESSURE


visual indicator is illuminated (high
priority alarm).
The audible alarm is activated (high
priority alarm).
The Inspiratory Valve is closed, and
the inspiration cycle is ended.
Note that the rate of breaths per
minute will be affected in this
condition as the ventilator will cease
to deliver flow above the maximum
working pressure.

VENTILATION MODE
TIDAL VOLUME
RATE,
I:E RATIO,
INSPIRATORY PAUSE, and
AIRWAY PRESSURE CONTROL
These parameters are set on the front panel
by five rotary controls and one momentary
action push button.
Based on the control settings, the system:
1.
Calculates the INSPIRATORY FLOW
and the INSPIRATORY and
EXPIRATORY times (see section 3.7).
2.
Controls the flow metering valve.
3.
Displays the value of the tidal volume,
rate, and I:E Ratio parameters on
three digital displays.
4.
Generates the appropriate messages
and alarms.

WARNING
If this warning is ignored, the patient may
receive an insufficient minute volume.
In PRESSURE controlled mode, when the
set pressure is reached the following
occurs:
1.
2.

Control Unit

The Inspiratory valve is closed.


The
ventilator maintains the set
pressure until the end of the
inspiration cycle.

High Pressure Protection


High pressure in the ventilator is limited by
three independent protective systems.
The pressure transducer has already been
described.
In addition, the Driving Gas Over-pressure
switch is set at 80 cmH2O and will shut off
drive gas flow at this value.
Further, the Pressure Relief Valve is a
mechanical over-pressure relief which will
open at 80 cmH2O, diverting the driving gas
to atmosphere through the PRV exhaust
port.
The Exhaust Outlets on (a) the back of the
control unit, and (b) the base of the control
unit, accept the drive gas exhaust from all
internal pneumatic components.

15

DESCRIPTION

Front panel - Models with Manometer and Spirometry

Front panel - Models without Manometer and Spirometry

16

DESCRIPTION
3.5

Front Panel Operating


Controls

(in STANDBY mode only):


1.
Press the MODE SELECT switch.
2.
Rotate the I:E ratio control fully
clockwise for ON, or anti-clockwise for
OFF.
3.
Press MODE SELECT to return to
STANDBY.

A description of each control is given in the


following paragraphs.

3.5.1 Power Switch

SPONT (spontaneous)

OFF

Applicable for Vt of 200 ml.


The indicator SPONT will be illuminated
with a green indicator and the ventilator
will monitor and display the functions
listed below.
No gas will be delivered in this mode.

(mains power connected)


All power is removed from the ventilator
circuitry (except the battery charger circuit).
The text display OFF will be illuminated with
a yellow light when the mains electrical
power is connected and the battery is being
recharged.

VENT INOP
HIGH AIRWAY PRESSURE
LOW SUPPLY PRESSURE
LOW AIRWAY PRESSURE
(Apnoea)
LOW BATTERY
MAINS FAILURE

STANDBY
Allows for ventilator function check and setup.
The text display STANDBY will be
illuminated with an amber light.
1.

2.

If the ventilator is fitted with the spirometer


option the ventilator will also display the
tidal or minute volume value, when
selected.

During the first two seconds in this


position, all LED segments and alarm
indicators are lit and the audible
alarms will activate.
After the first two seconds, the
following control/display functions are
available for set-up without cycling of
the ventilator.

VOLUME CYCLE MODE


Ventilator cycling will commence and all
alarms will be activated.
In this mode the VOL CYCLE indicator will
illuminate with a green indicator.
Maximum breathing system pressure will
be monitored and when maximum
pressure is reached the HIGH
AIRWAY
PRESSURE alarm will be activated and the
ventilator will immediately revert to the
expiratory phase, irrespective of the function
of the inspiratory pause.
The inspiratory pause function operates in
this mode.

TIDAL VOLUME
RATE
I:E RATIO
AIRWAY PRESSURE LIMIT
3.

The following alarms are functional in


the STANDBY mode:
LOW SUPPLY PRESSURE
MAINS FAILURE
VENT INOP
LOW BATTERY
HIGH AIRWAY PRESSURE
INCORRECT RATE/RATIO

PRESSURE CONTROL MODE

In this mode the PRESSURE indicator will


illuminate with a green indicator.
Ventilator cycling will commence and all
alarms will be activated, except high airway
pressure.
The ventilator will attempt to deliver the set
tidal volume until the pressure limit is
reached.

Spirometry
Spirometry (if fitted) can be enabled /
disabled by using the following procedure
17

DESCRIPTION
At this point the ventilator will stop
delivering gas and pause in this condition
until the calculated end of inspiration time
has been reached.
The ventilator will then return to the
exhalation phase.

In addition,high fresh gas flows will lead


to an increased Vt being delivered to the
patient.
Note that on models fitted with
spirometry, the actual tidal volume
exhaled will be displayed.
The patient must be monitored
independently from the ventilator.
It is the responsibility of the user to
monitor patient ventilation.

The inspiratory pause function does not


operate in this mode.

3.5.2 Tidal Volume Control

3.5.3 Ventilation Rate Control

This dial controls the requested tidal volume


(LITRES).
Adjustment is only possible when the
ventilator is in volume set mode.

RATE BPM

Models with spirometry:

This dial controls the requested ventilator


rate.
The respiratory RATE is set by a control
knob and the system accepts the setting as
long as the set parameters are within normal
limits
The digital display indicates the requested
RATE.

In any of the operating modes, press the


MODE SELECT button to obtain a display of
the tidal volume (set or measured) and
minute volume (measured).

If during adjustment the required flow rate is


less than 2 L/min or exceeds 75 L/min, the
set rate will be limited accordingly.

Models without spirometry:


The digital display indicates the requested
tidal volume.

NOTE

3.5.4 Ventilator I:E Ratio

When the ventilator is switched from standby


to the required operating mode, the
spirometer will take 1 - 2 minutes before
stabilising

WARNING
The ventilator settings can allow for and
inverse I:E ratio up to 1:0.3. The clinician
must always ensure that sufficient time is
allowed for the patient to adequately
exhale.

The TIDAL VOLUME is set by the control


knob. If the set parameters are within normal
limits, the system will deliver the set volume
indicated.
If during adjustment the required flow rate is
less than 2 L/min or exceeds 75 L/min, the
set tidal volume will be limited accordingly.

I:E RATIO
This dial and digital display controls and
indicates the requested ventilator I:E RATIO.
If during adjustment the required flow rate is
less than 2 L/min or exceeds 75 L/min, the
set I:E ratio will be limited accordingly.

WARNING
The
AV800
has
compliance
compensation but the actual tidal volume
delivered to the patient may be different
to the ventilation parameters set by the
user due to:
A) an extreme compliance condition,
B) a substantial system leak, or
C) patient circuit pressure effects.

NOTE
Minute Volume = Tidal Volume x Rate
(see section 3.7)

18

DESCRIPTION
3.5.5 Airway Pressure Limit
(cmH2O)

3.5.8

This dial sets the maximum breathing


system pressure as sensed by the pressure
transducer via the patient breathing circuit
via the manometer connection.

Connect a HPL 2 (or greater) format printer


to the printer output port on the rear of the
ventilator.
At the end of the clinical procedure switch
the ventilator to standby, and press the print
push button once.

In the Volume Cycle mode, the inspiratory


cycle is terminated whenever this set
pressure is achieved. The ventilator then
reverts to the exhalation phase.

Pressing the print button a second time will


cancel the print function.
Turn the ventilator to OFF, and switch the
ventilator back on again to clear the stored
print information.
The printer port must only be connected to
devices that comply to EN 60950.

In the PRESSURE CONTROLLED mode


the inspiration is terminated whenever
this set pressure is achieved and the
ventilator pauses until the end of the
calculated inspiratory time before reverting
to the exhalation phase.

3.5.6

Alarm Mute

When the alarm mute push button is


depressed during an alarm condition the
audible alarm can be muted for the
following periods:
Low Airway pressure
Low Tidal Volume
Incorrect rate or ratio
Low battery

30 seconds
30 seconds
120 seconds
120 seconds only

No other alarm conditions are silenceable.


Mutable alarms will be immediately
reinstated upon fresh alarm conditions.

3.5.7

Print

A printed copy of the ventilator conditions for


up to eight hours of the procedure can be
provided.

Inspiratory Pause

In volume ventilation mode depressing the


Insp Pause
button will allow a 25%
inspiratory pause to be included at the end
of the inspiratory phase.
The I:E ratio will be maintained.
To compensate for the reduced inspiratory
flow time period, the inspiratory flow rate is
increased (up to a maximum flow rate of 75
litres per minute).
The inspiratory pause does not function in
Pressure Ventilation mode.

19

DESCRIPTION
3.6

Rear Panel

3.6.1 Labelling Terminology


The terms
Class 1, and Type B are
defined in IEC 601-1 (the standard for
electrical medical equipment).

NOTE
This symbol denotes
Type B equipment.

Type B equipment calls for a particular


degree of protection against electric shock.
Class 1 equipment has additional protection
such that metal parts of the unit that are
accessible to the user (e.g. the metal casing
of the AV800 control unit) cannot become
live in the event of failure of the basic
insulation of the electrical components within
the unit.

DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT


USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS

RS232 PORT

Spirometer
connectors

CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE


COVER.

REFER TO QUALIFIED SERVICE PERSONNEL.

PRINTER PORT

ANALOGUE/ALARM OUTPUTS

SPIRO

OXYGEN/AIR
38 - 100 psi
2.6 - 6.9 Bar
262 - 689 kPa

MANOMETER/
PRESSURE
TRANSDUCER
-20 To 100 cmH2O

EXHAUST VALVE
DO NOT BLOCK

Drive Gas

PRV 80 cmH2O

FUSE RATING/TYPE
3.15 A 240 V ANTI SURGE HRC
Penlon Limited
Abingdon, Oxon
OX14 3PH
Tel 01235 547000
Fax 01235 547031
Tested by:
Serial No.
MADE IN UK

Ventilator Rear Panel


20

MAINS SUPPLY

CLASS 1 TYPE B
VOLTAGE 90 - 264 VAC
FREQUENCY 47 - 63 Hz
MAX CURRENT 1 A

DESCRIPTION
3.6.2

3.6.9

Mains supply

This standard Centronics 36 way parallel


port is configured to output to any HPL 2 or
greater compatible parallel printer.
The port must only be connected to devices
that comply to EN 60950.

The mains supply inlet is designed for


connection to any mains voltage from 90 to
264 VAC and a frequency of 47 to 63 Hz
without any adjustment.
The connector is a standard IEC type
dismountable connector.

3.6.3

Printer Port

3.6.10

Oxygen/Air

Spirometer connections

Male and female Luer lock connections are


provided for the connection of the side
stream spirometer system. Correct polarity
must be observed to ensure that the correct
flow is measured.
To maintain the accuracy of the flow
measurement only the specified sample
lines may be used.

Ventilator driving gas is attached to this


connector at a pressure of 38 - 100 psi
(262 - 689 kPa) and capable of supplying
gas at a flow rate of not less than 80 L/min
at the minimum pressure. Air or Oxygen
may be used as the drive gas but the supply
must be clean and dry.
Note that the drive gas is specified by the
customer. To change the drive gas, refer to a
Penlon-trained service engineer.

3.6.4 Drive Gas


The drive gas outlet is a special 17 mm
diameter taper connector and delivers the
drive gas from the control unit to the
ventilator bellows assembly.

3.6.5 Exhaust Valve


Exhaust bellows drive gas,
either air or
oxygen, returned from the bellows assembly
is exhausted through this outlet. The port
must not be blocked.

Latest type tubing assemblies have a


restrictor fitted to the tube fitted to the lower
connector on the ventilator rear panel.
This tube is labelled VENT END (A) to
assist correct fitment.

3.6.6 Manometer / Pressure


Transducer
The breathing system is connected to this
port with a 6 mm flexible sample tube.

3.6.7

CAUTION
A) The lines must not be trapped during use.
B) Do not alter the length of the sample lines.

PRV 80 cmH2O

The drive gas over pressure relief valve is


set at a non-adjustable 80 cmH2O.
The PRV outlet port must not be blocked.

3.6.8

3.6.11

Analogue/alarm Outputs

This port may be configured to interface


with a medical information bus system by
special order.
The port must only be connected to devices
that comply to EN 60950. See also section
10.

RS 232 Port

Configuration details for this port are


available by special order from Penlon
Limited.
The port must only be connected to devices
that comply to EN 60950.

21

DESCRIPTION
3.7

Message Displays

HIGH CONT (Continuing) PRESSURE


(High priority)
A visual indicator and audible alarm which
activate when the pressure sensed at the
patient tee exceeds 30 cmH2O just prior to
the next inspiratory cycle.
Once activated, this alarm remains on until
the pressure falls below the required
baseline at the beginning of a breath.
This alarm is not silenceable.

3.7.1 Alarm Mute


For those alarms which can be silenced, the
mute button will provide 30 seconds of
muting for high priority alarms and 120
seconds of muting for medium priority
alarms.

3.7.2 Alarm Indicators

LOW AIRWAY PRESSURE


(High priority)
(Disconnect Alarm)
A visual indicator and an audible alarm
which activate if the pressure sensed at the
patient tee-piece in the expiratory limb of
the breathing circuit does not increase by at
least 4-14 cmH2O during a ventilator
inspiratory cycle, depending on tidal volume
setting.
Once this alarm is activated, it remains on
until the required pressure difference is
reached.
This alarm can be silenced for 30 seconds.

Visual indicators are backlit, no warnings are


visible when the power is turned off.
All alarms are self-cancelling (with a
minimum activation period of 2 seconds),
except VENT INOP (ventilator inoperative).
Alarms comply with EN475.

NOTE
There are no means provided to defeat any
visual alarm.
Alarms are priority configured. The highest
priority alarm will always take precedence.
Normal conditions must be restored or the
unit turned off in order to cancel the alarms.

LOW TIDAL VOLUME


(High priority)
The AV800 can be fitted with an external
volume spirometer which provides signals to
the volume display. If the apparent tidal
volume as measured by the spirometer falls
below 50% of the set tidal volume, a visual
indicator and audible alarm are activated.
This alarm can be silenced for 30 seconds.

LOW DRIVE GAS SUPPLY


(High priority)
A visual indicator and audible alarm which
activate if the input drive gas pressure has
dropped below 38 psig (262 kPa).
This alarm is not silenceable.
HIGH AIRWAY PRESSURE
(High priority)
Standby, Spontaneous, and Volume
Cycle Mode
A visual indicator and audible alarm which
activate when the pressure sensed at the
patient tee exceeds the setting of the
AIRWAY PRESSURE LIMIT control.
This alarm will remain on until the pressure
falls below the control setting for a minimum
of 1 second.
This alarm will also activate if the drive gas
supply overpressure switch operates (and
also operates in Pressure control mode).
This alarm is not silenceable.

Spirometer Disconnect
If
the
spirometer
tube
becomes
disconnected, causing an erroneous
measured volume greater than 50% of the
set tidal volume, the Tv display will flash.
INCORRECT RATE OR RATIO
(Low priority)
A visual indicator and audible alarm which
indicates that the required inspiratory flow
rate is below 2 L/min, or has reached the
upper limit of 75 L/min as determined by the
settings for TIDAL VOLUME, RATE , I.E.
RATIO, and INSP PAUSE.
This alarm can be silenced for 120 seconds.

22

DESCRIPTION
3.8

MAINS FAILURE
(Low priority)
A visual indicator and an audible alarm (low
priority) that is battery driven.
The alarm activates when electrical power is
disconnected when the ventilator is
operating.
Note that if fully charged, the back-up battery
will power the ventilator for 60 minutes.

In the event of mains electrical failure, the


back-up battery cuts in automatically.
A fully charged battery will power the
ventilator for 60 minutes.
See Appendix 1 for battery care procedures.

LOW BATTERY
(Low priority / Medium priority)
A visual indicator and an audible alarm
which activate if the internal battery is almost
exhausted, indicating that ventilation will
soon cease.
When approximately ten minutes of battery
backup time remains, the alarm will change
from low to medium priority.
To prevent damage to the battery, the
ventilator will shut down before the battery is
fully discharged.
VENT INOP
(High priority)
The VENT INOP (ventilator inoperative)
alarm indicates that one of the following
conditions has occurred:
A)
B)
C)
D)
E)
F)

Back-up Battery

A solenoid has failed,


The flow control valve has failed,
Internal electrical fault,
Internal Communications error,
Battery disconnected,
Software error

In the event of a ventilator error being


detected an error code will be displayed in
the RATE / BPM window and additionally will
be stored in a non volatile memory which
can be accessed by the service engineer
to assist with fault finding subsequently.

WARNING
The only way to reset a VENT INOP alarm
is to turn the Power switch OFF for a
minimum of one second and then back
ON.
If the VENT INOP alarm occurs again,
remove the ventilator from use and refer
to an authorised service technician.
23

DESCRIPTION
3.9

Operational Capability
I:E Ratio
1:6

1:5 1:4 1:3

1:2

1:1

1:0.3

1.6
1.5
1.4
1.3
1.2
1.1

Tidal
Volume
(litres)
(Vt)

1.0
0.9
0.8

0.7
0.6
0.5
0.4
0.3
0.2
0.1
0
0

10

15

20

25

30

35

40

45

50

55

Rate (bpm)

The ventilator is capable of operating at the volumes and rates below each I:E ratio curve.
Note

a) The minimum tidal volume setting is 50 ml.


b) The minimum rate setting is 4 bpm.

Example
1.
Select required volume (Vt) (e.g. 0.7 litres)
2.
Select rate (e.g. 10 bpm).
3.
Select I:E ratio of 1:2.
The point X on the graph lies beneath the 1:2 ratio curve, and is therefore within the
ventilators capability.

24

60

DESCRIPTION
3.10 System Software
The software controls all ventilator functions
including performance accuracy and safety
checks.
In addition, the software controls
ventilator/user interface, apart from
manometer.

the
the

The software is held in 256 kbytes of Flash


EPROM, and is copied into RAM at power-up
and runs there.

Upgrades
Any upgrades will be offered to customers for
downloading via a connector on the main PCB.

Digital to Analogue Converter


Provides analogue output of the following
parameters:
Breathing circuit pressure
Average breathing circuit pressure
Airway pressure limit
Tidal volume
BPM
I:E ratio
Measured volume
These are all generated using an 8 bit, 8 channel
D/A converter (TLC5628). This is driven using
synchronous serial communications using three
I/0 bits on the processor.

15 pin D-type
connector pin
layout

Description

Range

Analogue
scaling

Ground

Breathing Circuit Pressure

-20 to 80 cmH20

0.05 V/cmH20

Average Breathing Circuit Pressure

-20 to 80 cmH20

0.05 V/cmH20

Airway Pressure Limit

10 to 70 cmH20

0.05 V/cmH2O

Tidal Volume

0-1.6 litres

0.3 V/100ml

BPM

0-60 bpm

0.08 V/bpm

I:E Ratio

1:0.3 to 1:6.0

0.08 V/O.IE

Measured Volume

0-1.6 litres

0.3 V/100ml

25

DESCRIPTION
Alarm Outputs
Alarm outputs are provided on the system
for various system parameters
Alarm outputs are on pin 9-15.
15 pin D-type
connector pin
layout

Alarm Outputs

Logic level

System OK

High

10

Power problem - mains fail,


battery missing or battery low

Low

11

Incorrect rate/ratio

Low

12

Low measured tidal volume

Low

13

Low supply pressure

Low

14
15

Low airway pressure


High airway pressure / high
continuous pressure

Low
Low

Real Time Clock


Provides a time stamp for the error log and any
printouts.
The clock is set at the factory to GMT (Greenwich Mean
Time).

Engineering Mode
A facility to allow Penlon-trained service engineers
access to the following functions:
1.
2.

3.
4.

Software revision information - indicates the


versions currently installed .
Error log - displays a log of the last 40 events
which caused a system operational error.
Note that after power-up, only the most recent
operational error is displayed.
Set date and time on the real time clock.
Abort to the FORTH interpreter to allow
communication to the unit through the serial port,
for possible software debug.

To gain access to Engineering Mode, switch the power


control to STANDBY, and then press the INSP PAUSE
and MODE SELECT buttons simultaneously.
Rotate the Tidal Volume control knob clockwise / anticlockwise to move between the above functions.
Press the INSP PAUSE button to exit Engineering
Mode.
26

4. SPECIFICATION
4.1

Application

Ventilation for use in anaesthesia.

4.2

Internal Compliance

3 ml/cmH2O (nominal)

4.3

Minute Volume Range

This parameter cannot be set on the


machine, but can be calculated if required.
(see section 3.5.4).

4.4

Tidal Volume Range

50 to 1600 ml (10%)

4.5

Frequency (Rate) Range

4 to 60 bpm, limited by tidal volume setting

4.6

Inspiratory Phase Time Range

0.14 to 11.54 seconds

4.7

Expiratory Phase Time Range

0.23 to 12.9 seconds

4.8

Inspiratory/Expiratory Phase
Time Ratio Range
(I:E ratio)

1: 0.3 to 1: 6.0 - limited by tidal


volume and rate settings

4.9

Pressure Control Range


(pressure ventilation mode)

10 to 70 cmH2O (10%)

4.10

Inspiratory Flow Range

2 to 75 L/ min

4.11

Airway Pressure Limit


(volume ventilation mode)

10 to 70 cmH2O (10%)

4.12

Inspiratory Triggering

No trigger

4.13

Inspiratory Triggering
- Response Time

No trigger

4.14

Maximum Safety Pressure

80 cmH2O

4.15

Maximum Working Pressure

70 cm H2O

4.16

Minimum Safety Pressure

-10 cm H2O

4.17

Minimum Working Pressure

Atmospheric

4.18

Sub-atmospheric Pressure
Range

None

4.19

Expiratory Resistance
(60 L/min)

5 cmH2O/Ls-1

4.20

Sigh Characteristics

None

4.21

Inspiratory Mixture

No mixture controls
27

SPECIFICATION
4.22

Flowmeters

None

4.23

Manual Changeover

None

4.24

Inspiratory Relief Valve

Fixed, 80 cmH2O

4.25

Humidifiers and Monitors

No humidifier.
Volume monitor optional (spirometer)
Pressure monitor built in

4.26

System Pressure Gauge

None

4.27

Airway Pressure Gauge

Optional

4.28

Spirometer
Accuracy

Optional
Varies with gas composition.
Note that the spirometer is calibrated for
typical clinical gas mixtures. Differing gas
composition will cause variances from actual
values. Maximum variation is 20%.
300 ml Vt minimum in Spontaneous mode.
50 to 1600 ml in all other operating modes.

Range
4.29

Alarms (in priority order)

See section 3 for detailed descriptions.


Vent. inop. (ventilator inoperative)
High airway pressure
High cont. (continuing) airway pressure
(PEEP)
Low supply pressure
Low airway pressure
Low tidal volume
Incorrect rate or ratio
Low battery
Mains failure

4.30

Power Source
Electrical

90 to 264 VAC
47/63 Hz, universal input.
Oxygen or air (dry, and oil free) at 38 to 100
psig (262 to 689 kPa).

Input Drive Gas


4.31

4.32

Power Consumption
Electrical
Gas

37 watts maximum
75 L/min intermittent maximum inspiratory
flow.
37.5 L/min minute volume

Dimensions
Height
Height of control unit only
Width
Depth

370
135
235
260
28

mm
mm
mm
mm

SPECIFICATION
4.33

Weight
Weight with adult bellows
Weight with paediatric bellows
Weight of control unit only

9.0 kg
8.7 kg
7.6 kg

4.34

Method of Disinfection
or Sterilisation

Bellows base assembly and inside of


bellows require sterilisation - section 7.3

4.35

Bacterial Filter

None (see section 5.1.4 for


recommendations for breathing system)

4.36

Fail Safe Mechanism

Battery back-up in case of mains electricity


failure

4.37

Reliability

MTBF: 5x106 to 50x106 cycles

4.38

Waveform Tests

Not applicable

4.39

Volume Tests

Not applicable

4.40

(A) Mobility

Secure mounting on anaesthesia machine


required.
Mounting bracket available as optional extra.
Bellows assembly can be separated from the
control unit for remote mounting.

(B) Mounting

4.41

Fuses

4.42

Environmental
Ambient Temperature

220/240 V: two fuses, 3.15 A, 20 mm, anti


surge

-5 to 50oC (23 to 122oF).


Refer to Appendix 1 for battery care during
storage.
10 to 38oC (50 to 100oF)

Storage

Operating
Humidity

10-95% RH (relative humidity), noncondensing

Altitude

Up to 2775 m (9000 feet)

Ingress protection

Conforms to EN 794-1 Clause 44 (spillage


test)

29

5. PRE-OPERATION PROCEDURES
5.1

Set-up

WARNING
Excessive electronic noise caused by
other, poorly regulated devices, such as
electrocautery, may adversely interfere
with the proper functioning of the
ventilator.
To avoid this problem, do not connect the
ventilator power cord into the same
electrical wall outlet or strip into which an
electrocautery unit is connected.

5.1.1 Components supplied with


the ventilator
Check that the ventilator has been supplied
with:
(a)
A 210 mm long, 16 mm diameter,
corrugated hose to connect the
control unit to the bellows assembly
(b)
Drive gas hose (to customer
specification)
(c)
Electrical mains supply cable (to
customer specification).
(d)
Pressure sampling line and T-piece.
(e)
Spirometer sampling line and T-piece
(supplied with spirometer option
machines only)

5.1.4 Ventilator gas connections


Before the ventilator can be used, the
following preparation must be made:
1.

5.1.2 Mounting the ventilator


The AV800 can be mounted on the
anaesthetic machine as a single, complete
unit.
Locate the ventilator in a safe place.
Preferably, mount it permanently on the
anaesthesia shelf of the machine or on a
strong bracket to protect the ventilator from
accidental fall and accidental disconnection
of the hoses.
To mount the ventilator permanently on a
Penlon bracket:
1.
Align the four mounting feet over the
mating holes in the bracket.
2.
Use four M4 screws inserted through
the bracket and rubber feet and
screwed into into the threaded inserts
in the base of the ventilator.

Verify the drive gas specified for the


ventilator (oxygen or air) and connect
the drive gas inlet port on the rear of
the control unit to a dry, oil free supply
at 38 to 100 psi (2.6-6.9 bar, 262-689
kPa)
To change the drive gas, refer to a
Penlon-trained engineer.
The drive gas inlet port is labelled:
OXYGEN / AIR 38 -100 PSI
OXYGEN SUPPLY - O2 cylinder,
anaesthetic machine O2 auxiliary gas
outlet, or O2 pipeline supply from a
wall outlet.
AIR SUPPLY - Air cylinder,
anaesthetic machine air auxiliary gas
outlet or air pipeline supply from a wall
outlet.

2.

Connect the 16 mm diameter


corrugated hose (provided), between
the control unit drive gas outlet
(labelled: BELLOWS DRIVING GAS)
and the bellows base DRIVING GAS
port.

3.

Connect the bellows base EXHAUST


port, with a 30 mm hose, to a properly
functioning scavenging system. The
diaphragm valve under the bellows is
connected internally to the EXHAUST
port to facilitate the discharge of
excess breathing gas at the end the
expiratory phase.

Pole-mount type mounting brackets are


available, see section 8.

5.1.3 Ventilator electrical


connections
Before connecting the ventilator to the mains
supply, check that the power supply is within
the correct rating as stated on the label on
the rear of the control unit.

30

Driving Gas
Port
Exhaust Port
(patient gas)
Do NOT connect
spirometer

Breathing
System
Port

Spirometer
connectors

DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT

USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS

RS232 PORT

CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE


COVER.

REFER TO QUALIFIED SERVICE PERSONNEL.

ANALOGUE/ALARM OUTPUTS

PRINTER PORT

SPIRO

OXYGEN/AIR
38 - 100 psi
2.6 - 6.9 Bar
262 - 689 kPa

MANOMETER/
PRESSURE
TRANSDUCER
-20 To 100 cmH2O

EXHAUST VALVE
DO NOT BLOCK

Drive Gas

MAINS SUPPLY

PRV 80 cmH2O
Penlon Limited
Abingdon, Oxon
OX14 3PH
Tel 01235 547000
Fax 01235 547031
Tested by:
Serial No.
MADE IN UK

FUSE RATING/TYPE
3.15 A 240 V ANTI SURGE HRC

CLASS 1 TYPE B
VOLTAGE 90 - 264 VAC
FREQUENCY 47 - 63 Hz
MAX CURRENT 1 A

Ventilator Rear Panel and Gas Ports

31

PRE-OPERATION PROCEDURES

AV800 VENTILATOR
(REAR VIEW)
EXHAUST TO
SCAVENGE
SYSTEM

AV800 ventilator connected as part of a


circle breathing system

BREATHING CIRCUIT
BACTERIAL FILTER

PRESSURE MONITOR LINE


SPIROMETER

FRESH GAS SUPPLY


(FROM MACHINE CGO)

HEAT AND MOISTURE


EXCHANGER (HME)

ANAESTHETIC MACHINE
CIRCLE
SYSTEM
ABSORBER

VENTILATOR
DRIVE GAS
(FROM MACHINE
AUXILIARY OUTLET)

PATIENT

4.

Connect the spirometer (if fitted) - see


Section 5.1.5.

expiratory limb to
breathing system.

WARNING

the

WARNING

Applying negative or positive pressure to


the bellows exhaust port results in
positive pressure in the patient breathing
system.
Therefore, the scavenging system must
not generate more than 0.5 cmH2O
positive or negative pressure when
connected to the ventilator.
Any problem arising from an improperly
functioning scavenging system is solely
the user`s responsibility.
5.

protect

The High and Low Airway Pressure


Alarms are important for patient care.
The ventilator is designed to be used with
a distal sensing tee only. (Catalogue No.
53194, Breathing System Tee Assembly.)
It is important that the distal sensing tee
is properly located in the inspiratory limb
of the breathing system.
6.
Set the AIRWAY PRESSURE LIMIT
control to 50 cmH2O.

Connect
the
MANOMETER/
PRESSURE TRANSDUCER port on
the rear panel of the control unit to the
breathing system.
The recommended placement for the
distal sensing tee is in the inspiratory
limb of the breathing system close to
the circle system inspiratory valve.
Use a Heat and moisture exchanger
(HME) at the patient Y piece, or a
breathing
system filter
in the
32

7.

Attach a printer to the printer port if a


printed
output of the ventilator
function is required.

8.

Connect the ventilator bellows base


BREATHING SYSTEM port to the
breathing system.
Use a breathing system bacterial filter
in the expiratory limb of the breathing
circuit or a heat and moisture
exchanger (HME) at the patient Y
piece. See illustration above.

PRE-OPERATION PROCEDURES
9.

Connect a 2-litre breathing bag to the


patient connection as a test lung.

10.

Close the anaesthetic machine APL or


PRV valve in the breathing system.

11.

Turn the ventilator POWER switch to


STANDBY and :
set the tidal VOLUME to 800 ml,
set RATE to 10 bpm,
set I:E RATIO to 1:2.0.

12.

Use the O2 flush button on the


anaesthetic machine to fill the
bellows.

13.

Turn the ventilator POWER switch to


VOL CYCLE.

14.

The delivered tidal volume read on the


scale of the bellows housing should
be 800 ml.
Before using the ventilator clinically:
a)
Check that all connections are
correct, and verify that there are no
leaks.
b)
Carry out the function checks
listed in section 5.3.1.

33

PRE-OPERATION PROCEDURES

A1

B2

B1

C1

C2
Spirometer
Head

E
Pressure
Monitor Line

A2
Notes
a) Carefully twist each sample
line anti-clockwise before
connecting to the ventilator and
Tee-piece.
Turn each connector fully
clockwise to ensure a leak-free
joint.
b) Latest type tubing assemblies
have a restrictor fitted to tube
C1/C2.
This tube is labelled VENT END
(E) to assist correct fitment.

5.1.5 Spirometer Connections


WARNING
The lumen assembly must be
connected as illustrated. Do not connect to the
expiratory limb of the breathing circuit.

The triple lumen assembly incorporates the


ventilator pressure monitor line (shown as
Tube A1/A2).
This line must be connected to the breathing
circuit at the absorber expiratory hose
connector.
1.

2.

3.

4.

Use a breathing system bacterial filter


in the expiratory limb, or a heat and
moisture exchanger (HME) at the
patient Y piece. See section 5.1.4,
operation 8.

Check for secure connection at each


connector.

NOTE
A) If the connection is incorrectly made at the
spirometer sensor, the ventilator will alarm LOW
TIDAL VOLUME. To allow the ventilator to be
used in the event of damage or or non-functioning
of the spirometer head, turn off the spirometry
function - see STANDBY mode.

Fit the spirometer head to the Safelock


connector at the ventilator port on the
back of the Circle Absorber.

B) Connection of the spirometer into the exhaled


limb also provides an indication of breathing
system disconnect.

Connect the tubing from the spirometer


head to the connectors at the rear of the
AV800 ventilator as shown.
For additional identification:
i) The centre tube (C1/C2) has a RED
connector at C2 which must be fitted to
the lower connector on the ventilator
(fitted with a RED washer ,D).
Later assemblies have a label (E) (VENT
END) on tube C1/C2
ii) Connector B1 is BLUE.

C) When testing the spirometer prior to clinical


use, it is preferable to use a 70% N2O / 30% O2
mixture.
If 100% oxygen is used, the Measured Vt value
display may flash. This is not significant, and
should be disregarded.

Absorber switched to Bag mode


NOTE If the absorber bag/ventilator switch
is set to Bag, the spirometer will not
measure tidal volume.
34

PRE-OPERATION PROCEDURES
1

5.2

Bellows Assembly

3.

CAUTION
Always ensure correct fitment of bellows
(see illustration above), and carry out a full
function test before clinical use, if a bellows
is removed and refitted.
1.

2.

4.

Fit the bellows, and then fit the


bellows housing to the base by
pushing it down, then twisting it
slightly clockwise until the bayonet
tabs completely engage.
Function test the ventilator - section
5.3.1.

NOTE

Remove the bellows housing (1) by


twisting it slightly counterclockwise
until the bayonet tabs become free,
then lift it up from the base (2).
Remove the bellows (3) from the
base.

If there is any malfunction, the ventilator


must NOT be used.
If the problem cannot be located, refer to
section 6 - Fault Finding, or return the
ventilator for repair to a Penlon accredited
agent, or to the manufacturer.

35

PRE-OPERATION PROCEDURES
5.3

Pre-use Checklist

3.

Connect the ventilator bellows base


BREATHING SYSTEM port to the
breathing system.

The following tests should be done at least


at the beginning of every working day:

4.

Connect a 2-litre breathing bag to the


patient connection as a test lung.

WARNING

5.

Close the anaesthetic machine APL or


PRV valve in the breathing system,

6.

Set the tidal VOLUME to 800 ml;


RATE to 10 bpm, and I:E RATIO to
1:2.0.

7.

Use the O2 flush button on the


anaesthetic machine to fill the
bellows.

8.

Turn the ventilator MODE switch to


VOL/CYCLE.

9.

The delivered tidal volume read on


the scale of the bellows housing
should be 800 ml.

5.3.1 Daily Checklist

The operation of each alarm function


should be verified daily. Periodically
check the alarms at suitable intervals.
If the audible alarm or the visual indicator
of any alarm function fails to activate
during any alarm condition or fails to
reset after the alarm has been cleared,
refer the unit to an authorised service
technician.

WARNING

If the internal battery is fully discharged,


the ventilator will not function.
Recharge the battery before the ventilator is
used clinically.
Charging the battery for 14 hours from a
discharged state will allow a minimum of 45
minutes of continuous operation. Connect
the ventilator to a mains power supply. The
OFF indicator will turn yellow when the
battery is being charged (it is not necessary
to run the ventilator).

If the delivered tidal volume is less


than 700 or greater than 900 ml, refer
to a Penlon trained engineer.
10.

Fill the bellows using the


again.

11.

Occlude the patient connection port of


the breathing system, distal end.

12.

The HIGH AIRWAY PRESSURE


alarm should be activated. The peak
pressure read on the breathing
system pressure gauge is the
maximum working pressure limit and
should agree with the setting.

13.

Open the patient connection port to


ambient pressure. At the second
cycle, the LOW AIRWAY PRESSURE
alarm should be activated.

14.

Turn the ventilator MODE switch to


STANDBY.
Before using the ventilator clinically,
check that all connections are correct,
and verify that there are no leaks.

Ventilator internal test

Turn the POWER switch to STANDBY.


A two-second internal test is initiated.
The following must occur if everything
performs correctly:
1.
2
3

All three digital displays must read 8.


All ALARMS messages must be lit.
The audible alarm drivers must sound
- one high tone, and one low tone.

Functional checkout:
1.

Set the AIRWAY PRESSURE LIMIT


control to 50 cmH2O.

2.

Check that the


MANOMETER /
PRESSURE TRANSDUCER port on
the rear of the control unit is correctly
connected to the inspiratory limb of
the breathing circuit, close to the
patient tee piece (see operation 5,
section 5.1.4).
36

O2 flush

PRE-OPERATION PROCEDURES
5.3.2 Weekly Checklist
At least every week, in addition to the daily
functional check:
1.

Turn the ventilator MODE switch to


STANDBY.

2.

Unplug the mains power cable from


the AC outlet.
The MAINS FAILURE alarm should
activate.

3.

Reconnect the mains power cable to


the AC outlet. The alarm should turn
off.

4.

Turn
the
MODE
VOL/CYCLE.

5.

Disconnect the drive gas supply hose.


The LOW SUPPLY PRESSURE
alarm should activate.

switch

to

NOTE
If there is any malfunction, the ventilator
must NOT be used.
If the problem cannot be located, refer to
section 6 - Fault Finding, or return the
ventilator for repair to a Penlon authorised
service centre or to the manufacturer.

37

6. FAULT FINDING
CONDITION
No power at switch on
(no VENT INOP alarm)

LOW BATTERY message is on

Bellows does not fill, or, collapses.

Bellows progressively becomes less


full despite usual fresh gas flow setting.

LOW BREATHING SYSTEM PRESSURE


alarm is on but ventilation is normal.

LOW DRIVE GAS SUPPLY alarm activated

POSSIBLE CAUSE / REMEDY


1.

Check mains supply/plug.

2.

Check yellow OFF LED (mains indicator).

3.

Check fuses - mains input socket.

4.

Check voltage from power supply unit - 14.3 V.

5.

Replace ON/OFF switch.

1.

Charge battery by plugging unit into the mains


supply.

2.

Defective battery - replace.

3.

Mains fuse blown.

1.

No fresh gas flow.

2.

Breathing system hose leak or hose disconnected.

3.

Broken bellows base.

4.

Defective or detatched bellows.

5.

Damaged exhalation diaphragm valve.

6.

Missing or defective O-ring seals.

1.

Check breathing system for leaks.

2.

Check and close anaesthesia system APL or PRV


valve.

3.

Check exhalation diaphragm valve seat under the


bellows for damage.

4.

Damaged O-ring on bellows base - replace.

1.

Check pressure pick-up tubing.

2.

Look for disconnect or kink.

3.

Control unit internal failure

1.

Check drive gas supply pressure (> 38 psi)

38

FAULT FINDING
CONDITION
Displayed values incorrect

POSSIBLE CAUSE / REMEDY


1.

Check with stopwatch / Vt measurement.

2.

Check that the inlet filter is not blocked.

3.

Replace/reset pressure switch.

1.

Potentiometer set too low (or knob out of


calibraion) - adjust.

2.

Error code 12
Potentiometer failure - replace display assembly
PCB.

3.

Error code 8
Potentiometer sensor failure - replace main PCB.

1.

Defective or poorly regulated scavenging system,


creating too high resistance or too much vacuum.

2.

Exhaust partially obstructed.

3.

Bellows diaphragm defective.

1.

Error code 8
Pressure sensor failure - replace main PCB.

2.

Internal tube disconnected - reconnect and secure.

Stacking breaths (inadvertent PEEP)

1.

Error code 3
Expiratory valve solenoid failed to open, resulting
in high airway pressure alarm.- replace.

Continuous flow of drive gas, or, bellows


driven to bottom.

1.

Error code 2
Faulty inspiratory solenoid valve - replace.

Tidal volume not delivered and


HIGH AIRWAY PRESSURE alarm is on.

1.

Check/increase AIRWAY PRESSURE LIMIT


setting.

2.

Check breathing system for obstruction.

1.

Error code 2
Inspiratory solenoid valve faulty - replace.

Continuous high airway pressure limit


alarm

Unwanted PEEP and bulging, full bellows

Pressure limit fails to operate

Zero tidal volume.

39

FAULT FINDING
CONDITION
Zero tidal volume and drive gas leak from
exhaust port.

Low tidal volume

High tidal volume

Tidal volume incorrect or inconsistent

POSSIBLE CAUSE / REMEDY


1.

Error code 3
Expiratory solenoid failed open - replace.

2.

Exhaust valve failed - replace.

1.

Error code 2
Inspiratory solenoid valve faulty - replace.

2.

Pressure relief valve leaking - replace O ring.

3.

Inlet filter blocked - replace.

4.

Defective spirometry head or leak in spirometry


system.

5.

Breathing system leak or hose disconnected.

6.

Exhaust valve leaking - replace O ring / silicon


washer.

1.

Leaking spirometer connector.

2.

Incorrect calibration of drive gas flow - recalibrate


needle valve / stepper motor position.

1.

Missing small O-ring under the exhalation


diaphragm valve.

2.

Partially detached or defective bellows.

3.

Damaged exhaust valve seal or seat.

4.

Very high/low breathing system resistance or


compliance.

5.

High/low fresh gas flow - check volume delivered


using standard calibration method.

6.

Incorrectly regulated gas pressure - check


pressure / replace regulator.

7.

Error code 1
Flow valve fails to adjust correctly due to
mechanism binding - realign and check calibration.

8.

Component failure on main PCB - replace PCB.

40

FAULT FINDING
CONDITION
Tidal volume incorrect or inconsistent
(continued)

POSSIBLE CAUSE / REMEDY


9.

Spirometer faulty - replace spirometer head and


lead.

10.

Loose grub screws on motor drive / needle shaft /


potentiometer drive shaft.
- check security of grub screws.
- return ventilator to accredited agent or
manufacture for calibration.

11.

Broken or damaged potentiometer flexible coupling


- return ventilator to accredited agent or
manufacture for calibration.

12.

Worn needle valve - return ventilator to accredited


agent or manufacture for calibration.

41

FAULT FINDING

42

7. SERVICE SCHEDULE
The AV800 ventilator must be serviced to the following schedules.
Only Penlon-trained engineers should undertake servicing and repairs..
6 Month Service
1.

Full service performance check.

12 Month Service
1.

Fit Preventive Maintenance Kit components :

AV800 12 month service Kit


Part
Part
Part
Part
Part
Part
Part
Part
2.

No.
No.
No.
No.
No.
No.
No.
No.

042819
0314
041204
041226
300045
0762
15446
300329

Part No. 57633


Quad Seal ( Exhaust valve )
O Seal Viton ( PRV )
O Seal Viton ( Bellows base diaphragm valve )
O Seal Viton ( Bellows Base to housing )
Silicon Sealing Washer ( Exhaust valve )
Filter - W9151 - 40 Micron.
Warranty sticker
Spirometer tubing assembly

Full service performance check.

5 year Overhaul
1.
Fit overhaul kit components :
5 year overhaul kit
Part No. 57635
Parts listed in 12 month kit, plus:
Part No. 104019
Lithium Battery ( Real time clock )
Part No. 103996
Lead Acid Battery ( Mains back up )
Part No. 045438
Internal Regulator
Part No. 300116
Internal Tubing Set
Part No. 011052
Adhesive tape (battery and speaker)
2.

Full service performance check.

Software Upgrades
Software upgrades will be offered to distributors and customers by means of a software
program which will be loaded via the main PC.104 connector using a specially developed
ROM Card
For distributors and customers who do not have access to a computer, a PCB board and front
panel pic chip replacement will be available for early machines.
Special Tools
Part No.

-----300128
------

Description

Multimeter
PRV and Exhaust - cap removal tool
Test Gauge
43

SERVICE SCHEDULE
AV800 SERVICE PROCEDURE (version 2.0 software)

Item

Operation

Service frequency

1
1.1
1.2
1.3
1.4

Initial Checks
Check and record serial number and determine service required.
Check general condition of ventilator assembly
Check configuration of attachments and note.
Check and record front panel control settings.

2
2.1

Engineering mode test


Switch ventilator selector switch to OFF.
Press MODE SELECT and INSP PAUSE simultaneously, and
hold.
Switch Ventilator selector switch to STANDBY.
The ventilator is now in Engineering Mode.
NOTE To set a non spirometer version in engineering mode
press the panel where the MODE SELECT switch is hidden from
view.

2.2

Check date and time


Turn the I:E ratio knob to move between parameters.
Turn the RATE knob to adjust, if necessary,
If any adjustments have been made, press MODE SELECT at the
end of the procedure (i.e., in Check Time - Minute mode) to save
new settings.
Turn the I:E ratio knob to move between parameters.

2.3

Verify correct revision level.

2.4

Note all errors and investigate.


Error codes :0 - Watchdog reset.
1 - Stepper position error.
2 - Inspiratory solenoid fail.
3 - Expiratory solenoid fail.
4 - Front panel comms fail.
5 - Wild reset detected.
6 - Stepper feedback pot fail.
7 - System power fail.
8 - Airway pressure transducer fail.
9 - Ambient pressure transducer fail.
10 - Software Exception error - Parameter gives the exception
number.
11 - Stack underflow or overflow.
12 - Front panel pressure limit pot fail.
13 - Flash checksum changed.

2.5

Press MODE SELECT to clear errors,


Press INSP PAUSE to exit engineering mode.

44

Six months
(ALL item 1
operations)
Six months
(ALL item 2
operations)

Check Date Month

Check Date Year

Month shown December


Turn RATE knob
to adjust.
Turn I:E knob
clockwise to get
to Check Date Day

Year shown - 1998


Turn RATE knob
to adjust.
Turn I:E knob
clockwise to get to
Check Date Month

Check Date - Day in


Month

Check Date - Day of


Week

Day shown - 10th

Day shown - Sunday

Turn RATE knob to adjust.


Turn I:E knob clockwise to
get to Check Date - Day of
Week

Turn RATE knob to adjust.


Turn I:E knob clockwise to
get to Check Time - Hour

Check Time - Minute

Check Time - Hour

Time shown - two


minutes past the hour

Time shown - 15.00


Turn RATE knob to adjust.
Turn I:E knob clockwise to
get to Check Time - Minute

45

Turn RATE knob to


adjust.
Turn I:E knob clockwise
to get to Check Revision
Press Mode Select button
to save any new settings.

Check Revision Main Board

Check Revision Front Panel

Revision shown - 1.2.1

Revision shown - O.O.A

Turn I:E knob clockwise


to get to Check Revision
- Front Panel

Turn I:E knob clockwise


to get to Error codes.

Error Codes

Error Codes

Error 13 shown

All errors cleared

Press Mode Select to clear


error

Press INSP PAUSE to


exit engineering mode.

Gas supply filter


Bellows and
diaphragm
assembly

Spirometer tubing
and housing

Exhaust valve

46

Pressure relief valve


(PRV)

SERVICE SCHEDULE
3
3.1
3.2
3.3

Electrical Safety Checks


Carry out the electrical safety checks with the ventilator power switch
in the OFF position.
Earth continuity (max 0.2 ohms )
Insulation resistance ( not less than 10 m ohms )
Earth leakage ( max 100 micro amps )

Bellows Assembly (refer to illustrations on previous page)

4.1
4.2
4.3

Remove and clean bellows housing.


Remove bellows assembly from base.
Unscrew the three thumb screws and remove the diaphragm valve.
Inspect the valve seat for damage, check the valve disc lies
horizontal. ( Do NOT attempt to dismantle diaphragm assembly ).
Clean the valve seat and valve disc, use an alcohol wipe.
Remove and discard the O seals from the bellows base and
diaphragm valve.
Reassemble using correct service kit.

4.4
4.5
5
5.1
5.2
5.3
5.4
5.5
5.6
5.7
5.8
5.9
5.10
5.11

Pneumatic System (refer to illustrations on previous page)


Remove covers and unlock the main PCB Tray to gain access.
Using special tool, replace the Quad seal in the exhaust valve.
Replace Exhaust valve silicon seal.
Using special tool replace O seal in PRV
Check internal spirometer tubing, and replace if necessary.
Check external spirometer tubing, and replace if necessary
Replace O seal in spirometer housing Safelock connector
Replace gas supply filter.
Replace pressure reducing regulator.
Replace all internal tubing.
Connect a test gauge to the pneumatic assembly, ensure regulator is
set to 35 psi (240 kPa) at 5 L/min.
5.12 Pressure test the gas inlet to inspiratory valve
5.13 Connect the ventilator to a power supply, turn the power switch to
STANDBY.
Set Litres, Rate and I:E Ratio to minimum, and check that at this
position the needle valve is almost shut.
6
6.1
6.2
6.3

Control Unit
Check all electrical connections and components for security.
Replace Back up battery.
Replace real time clock battery.

7
7.1

Set Up
Select drive hose O2 or Air and attach to the OXYGEN/AIR
connection to the rear panel and ensure the drive gas switch is in
the correct position. (Air down O2 up) connect mini-Schrader to
anaesthetic machine.
Connect the small corrugated hose to the Driving Gas Ports.
Connect a hose from the Breathing System Port to a CO2 absorber.
Connect the pressure tube to MANOMETER/PRESSURE
TRANSDUCER connection on the rear panel and to the To Patient
port of the CO2 absorber

7.2
7.3
7.4

47

Service frequency
Six months
(ALL item 3
operations)

One year
Six months
Six months

One year
One year
Six months
Six months
One year
One year
One year
One year
One year
One year
Five years
Five years
Six months
Six months
Six months

Six months
Five years
Five years
Six months
(ALL item 7
operations)

SERVICE SCHEDULE
Item
7.5
7.6
7.7
7.8
7.9
8
8.1
8.2
8.3
8.4
8.5
9
9.1
9.2
9.3
9.4

9.5

Operation

Service frequency

Connect the spirometer tubing to SPIRO connections on the rear


panel and to the spirometer Tee piece which should be connected to
the From Patient port of the CO2 absorber.
Connect the breathing tubes to the CO2 absorber and attach the
patient Y to a test lung.
Attach the FRESH GAS SUPPLY to the CO2 absorber and set 02
flow at 0.25 L/m or min flow.
Connect the gas scavenging system to the EXHAUST PORT.
Connect the mains cable from the main socket in the rear panel to a
240 VAC supply.
Standby mode test
Switch the ventilator power switch to ON verify that two bleeps
sound and all LEDs light up.
Check that previous settings are returned to display, (Litres, Rate
and I:E Ratio)
Set an incorrect rate, i.e. increase Litres and/or Rate controls
clockwise, check INCORRECT RATE OR RATIO alarm
Verify setting lock, and check I:E Ratio display flashes.
Press MODE SELECT check SPIRO can be set ON or OFF,
leave set to ON.

Six months
(ALL item 8
operations)

Spontaneous
Switch ventilator switch to SPONTANEOUS and press MODE
SELECT to access TIDAL VOLUME (MEASURED).
Switch CO2 absorber to BAG
Operate Bag Squeezing and check LITRES / MIN and RATE B.P.M
displays indicate readings.
Stop Bag Squeezing, ensure pressure gauge falls to zero, after 30
seconds delay.
Check LOW AIRWAY PRESSURE alarm, and also check that
displays show = =
Press MUTE and verify audible alarm is muted for 30 seconds.

Six months
(ALL item 9
operations)

10
Volume cycle
10.1 Switch Ventilator switch to VOL CYCLED
10.2 Set TIDAL VOLUME LITRES / MIN to 0.8, set RATE BPM to 10,
I:E RATIO to 1:2 and AIRWAY PRESSURE LIMIT to max.
10.3 Verify 800 ml is delivered (indicated on bellows canister).
10.4 Press INSP PAUSE check 25% pause during the inspiratory phase.
10.5 Press MODE SELECT check displays indicate values on LITRES /
MIN and RATE BPM. Note these will show incorrect values due to
gas and agent variations.
10.6 Remove and occlude expiratory breathing hose and check that the
following alarms operate.
HIGH AIRWAY PRESSURE
HIGH CONTINUOUS PRESSURE
LOW AIRWAY PRESSURE
LOW TIDAL VOLUME
10.7 Disconnect mains supply, verify ventilator continues to operate on
battery, check MAINS FAIL alarm
48

Six months
(ALL item 10
operations)

SERVICE SCHEDULE
11
Pressure ventilation
11.1 Switch Ventilator selector switch to PRESSURE
11.2 Set AIRWAY PRESSURE LIMIT to 20 cmH2O and verify that the
pressure is held at setting.
11.3 Carry out leak test. Set LITRES to 1 L/m, RATE to 5 bpm and
AIRWAY PRESSURE LIMIT to 20 cmH2O, ensure rotameter is set
to 200 ml or min. Fill bellows using O2 flush, allow bellows to cycle
and verify bellows remain full.
Maximum permissible leak 200 ml.
11.4 Remove mini schrader from gas supply pressure and check LOW
SUPPLY PRESSURE alarm, verify LITRES / MIN display flashes.
12
Completion
12.1 Check in Engineering Mode for error messages
12.2 Sign and date service card, indicate Commissioning, attach card and
plastic wallet to equipment.
12.3 Fill out service report.
12.4 Leave equipment ready for use as requested by the customer.

49

Service frequency
Six months
(ALL item 11
operations)

Six months
(ALL item 12
operations)

8. SERVICE PROCEDURES
8.1
8.1.1

Cleaning

inspected at least every six months or after


1200 hours of use, whichever comes first.
The bellows must be replaced if it shows
signs of aging
The bellows can be removed by carefully
pulling it off the base.

Outside surfaces and


bellows housing

CAUTION
Care must be taken not to allow liquids to run
into the control unit; serious damage may
result.
Check that the unit is disconnected from the
electrical supply before cleaning.
Do not use cleaning solutions containing
alcohol; the bellows housing may be
damaged.

Exhalation Diaphragm Valve


The exhalation diaphragm valve is under the
bellows and can be removed by loosening
the three thumbscrews.
The valve seat is now visible.

WARNING
Great care must be taken not to damage
the precision surface of the diaphragm
valve seat (A).
Never use any hard object or abrasive
agent to clean it; use only a soft cloth
(alcohol wipe recommended).
If the valve seat is damaged, the
diaphragm valve will leak and may cause
serious malfunction.

To clean the outside surface of the ventilator,


use a damp cloth.
If necessary use a warm, mild detergent
solution to remove resistant grime. Make
sure that all detergent residues are fully
removed after cleaning.
Never use any harsh abrasive cleaning
agent. The transparent acrylic bellows
housing (canister) and, in general, the
surfaces of the control unit are not scratch
resistant.

Using an alcohol wipe, clean the seat (A),


and the metal disk (B) attached to the base
of the diaphragm valve, thoroughly and
remove all contamination from the surfaces
of both components.

The inside of the canister, under normal


conditions, is not in contact with the
breathing gas and therefore only needs
cleaning as described above.

NOTE
If excessive contamination is discovered,
check that a bacterial filter is used in the
expiratory limb of the breathing circuit (or an
HME at the patient tee-piece).
See section 5.1.4.

Remove the canister by slightly twisting it


counter-clockwise until the tabs at the
bottom clear the bayonet locks, then lift it
straight up from the base.

8.1.2

After cleaning, check that the small O-ring


(C) located in the bellows base under the
diaphragm valve is in place. The ventilator
will not function if the O-ring is missing.
See section 8.2 for information on
sterilisation procedures.

Bellows and exhalation


diaphragm valve

Refer to illustration on next page.


Each time the bellows assemblies are
opened for cleaning, all visible components
must be carefully inspected and damaged
parts must be replaced.

Refit the diaphragm valve assembly to the


bellows base and reassemble the bellows
assembly.

CAUTION
Always check for correct fitment of the
bellows (see illustration) and carry out a full
function test of the ventilator before clinical
use.

Bellows
As with all elastomers, the bellows material
deteriorates with aging and should be
50

SERVICE PROCEDURES

B
C

A
Exhalation Diaphragm
Valve Assembly

Bellows Base

Refitting the Bellows

51

SERVICE PROCEDURES
8.1.3

Spirometer Sensor and


Tubing Assembly

On a regular basis (at least weekly), the sensor


assembly and sample lines must be thoroughly
cleaned, and sterilised as necessary.
a) Carefully detach the sample lines from the
Luer taper connectors on the rear panel of the
ventilator, and remove the spirometer assembly
(A) from the absorber.
b) Wash the components, and sample lines
by rinsing with clean (distilled) water poured
into the outlet (not under pressure).
Ensure that all traces of solid contamination
are flushed out.
Note
Latest tubing assemblies have a restrictor
fitted to the tube marked VENT END (B)
After washing, check that the restrictor is not
blocked.
c)

Sterilisation - see section 8.2, below.

d) When the components are dry, refit the


sample lines to the spirometer sensor
assembly and connectors on the rear panel of
the ventilator.
The tube labelled VENT END (B), must be
fitted to the ventilator rear panel, as illustrated
above.

52

SERVICE PROCEDURES
8.2

Sterilisation

Recommended guidelines for sterilisation:

CAUTION
To prevent possible damage to components, peak sterilisation temperatures must not exceed
54oC (130oF) for gas (ethylene oxide) or, 134oC (275oF) for steam autoclave.
Do not sterilise the ventilator control unit. The internal components are not compatible with
sterilisation techniques and may be damaged.
Following sterilisation with ethylene oxide, components must be quarantined in a well
ventilated area to allow dissipation of any residual gases.
Follow the recommendations given by the steriliser manufacturer for aeration periods.
ITEM

METHOD

Bellows *
Hoses
O rings
Bellows base

Gas,
Gas,
Gas,
Gas,

Exhalation valve assembly


Control unit
Bellows canister
Spirometer sensor tee piece
Spirometer sample lines

Gas, liquid, pasteurise, low temperature autoclave.


Do not sterilise - see section 9.1.
Liquid.
Gas, liquid, autoclave
Liquid

Distal sensing Tee

Liquid

liquid,
liquid,
liquid,
liquid,

autoclave (20 cycles max.)


autoclave
autoclave.
autoclave.

*Applies to standard size bellows only, the optional paediatric bellows must not be steam
autoclaved.

NOTE
Examples of suitable liquid agents are: Nu-Cidex, Sporicidin, and Sonacide.
The exhalation diaphragm valve must be removed, cleaned and sterilised sep
arately.

53

SERVICING PROCEDURES
8.3

Tubing Connectors

To connect and disconnect the tubing from the push-in


connectors, follow the procedures detailed below.
Do not use excess force.
Disconnecting tubing from connector
Push in the tube and the connector
end piece A.

A
Hold the end piece A in place.

Pull the tube out to disconnect.

Fitting tubing to connector

Push in the tube as far as possible


Do not use excessive force.

The connector end piece A will also move inwards.

A
Pull the tube carefully outwards.
The end piece A will be pulled outwards to the
locked position

54

SERVICING PROCEDURES
1

041204
O-ring

0762 Filter
(drive gas
inlet)

5
4
042819
O-ring

041226
O-ring

8.4 Fitting the Annual Preventive


Maintenance Kit
1.

3.
4.

Exhaust valve

Remove the bellows housing and bellows (1).


Loosen the three screws securing the exhalation
diaphragm valve (2) and remove.
Fit a new O rings (3 and 4) to the bellows base (5).
Check the condition of the exhalation valve disk
and valve seat (see section 9.1.2)

NOTE
If a ventilator is equipped with an optional
paediatric bellows assembly, the O-ring on the paediatric
adaptor must also be renewed annually.
Order O-ring, Part No. 041225

Drive gas inlet filter


1.
2.
3.

300045
Sealing washer

Disconnect the mains electrical and gas supply.

Bellows assembly
1.
2.

10

11
0314 O-ring
(pressure relief
valve)

Remove the chassis unit cover.


Undo the hose connector (6).
Fit a new filter (7), and refit the hose connector.

Exhaust valve
1.
2.

Use service tool to unscrew the exhaust valve


cover (8).
Remove the valve components and fit a new
sealing washer (9) and O-ring (10).

Pressure relief valve (PRV)


1.
2.

Use service tool to unscrew the PRV cover.


Fit a new O-ring (11) to the valve plunger.

Spirometer Tubing Assembly


1.

Fit the tubing with label VENT END (A) to the


lower connector at the rear of the ventilator
55

300329
Spirometer Tubing Assembly

SERVICING PROCEDURES
8.5 Fitting the Five Year
Overhaul Kit
1.
2.
3.

4
1

Disconnect the ventilator from


electrical and gas supplies. Remove
the cover from the ventilator chassis.
Remove the two screws (1) to allow
the main PCB / tray assembly to be
pivotted upwards.
Disconnect the leads from the battery
(2).

Regulator assembly
4.

Remove the securing screws (3), and


detach the regulator (4) from the
manifold assembly.

Mains back-up battery


5.
6.
7.
8.
9.

Carefully break the adhesive seal and


detach the alarm system speaker (5)
from the top of the battery.
Break the adhesive seal between the
battery and the base of the ventilator
chassis and remove the battery.
Fit a new battery, using a new
adhesive pad to hold in place.
Refit the speaker (5), using a new
adhesive pad to hold in place.
Reconnect the battery leads.

Real time clock battery


CAUTION
To prevent possible contamination and
subsequent corrosion, do not handle the
battery.
10.

Fit a new battery (6).


Check for real time errors which may
occur
because
of
battery
replacement.

Internal Tubing
11.

Renew all internal tubing (including


spirometer tubing if fitted).
Cut the new tubing to the exact
lengths
specified
below.
Spirometer accuracy is dependant
on correct tubing lengths.
56

64 mm
40 mm

45 mm

Plan
View

1 (160 mm)

1 (115 mm)

Ref

Part No.

462541 (623 mm / 4 mm o.d.)

2
3
4
5
6

054541 (Adaptor)
462544 (220 mm / 8 mm o.d.)
462545 (149 mm 10 mm o.d.)
011092 (430 mm / 6 mm o.d.)
053211 (Plug - fitted only to
models without gauge)
054947 (Tee piece)
054949 (Connector)
011092 (510 mm / 6 mm o.d.
(models with spirometer)

7
8
9

3
9
5

1 (18 mm)

5 (350 mm)
6

1 (330 mm)

Side View

3
5 (80 mm)
8
57

SERVICING PROCEDURES
8.6 Manifold Assembly
1.
2.
3.
4.
5.

3
1

Disconnect the ventilator from


electrical and gas supplies. Remove
the cover.
Remove the two screws (1) to allow
the main PCB / tray assembly to be
pivotted upwards.
Disconnect the back-up battery wiring
(2).
Disconnect all tubing and wiring
connectors (3) from the manifold
assembly.
From underneath the ventilator,
remove the three screws (4) to allow
removal of the manifold assembly (5).

2
1
5

8.7 Electrical Power Supply


1.
2.
3.
4.

5.

Disconnect the ventilator from


electrical and gas supplies. Remove
the cover.
Remove the two screws (1) to allow
the main PCB / tray assembly to be
pivotted upwards.
Disconnect the back-up battery wiring
(2).
Remove the two screws and detach
the mains in/out socket / filter unit (3)
from the chassis, but do not
disconnect the wiring.
From underneath the ventilator,
remove the four screws and lift the
power supply unit (4) from the
chassis.
Disconnect the wiring from the power
supply unit to allow removal.

2
1

4
3

58

SERVICING PROCEDURES
8.8 Front Panel and PCB
Assembly, and Gauge

NOTE
The display PCB and the control PCB are
not available as separate items. A front
panel and PCB assembly must be ordered
if either PCB is faulty.
1.
2.
3.
4.
5.
6.

7.
8.
9.

Disconnect the ventilator from


electrical and gas supplies. Remove
the cover.
Remove the two screws (1) to allow
the main PCB / tray assembly to be
pivotted upwards.
Disconnect the back-up battery wiring
(2).
Disconnect the wiring connectors (3)
from the PCB.
Remove the four screws (4) and
detach the front panel assembly (5)
from the ventilator.
On models with a manometer:
Disconnect the tubing (6) from the
back of the gauge.
Slacken the screws on the retaining
ring (7) and detach the gauge (8) from
the assembly.
Carefully remove the caps from each
control knob (9), then use the special
tool to remove each knob.
To reassemble, reverse the removal
procedure
Fitting a new front panel assembly On models with a manometer, the
new front panel must be cut before the
gauge can be refitted.to allow fitment
of the gauge.

1
3

4
6

4
7

59

SERVICING PROCEDURES
8.9

Main PCB Tray Assembly

NOTE

The main PCB is not available as a


separate item. A tray assembly complete
with pre-programmed PCB must be ordered
if the PCB is faulty.
1.
2.
3.
4.
5.
6.
7.
8.

Disconnect the ventilator from


electrical and gas supplies. Remove
the cover.
Remove the two screws (1) to allow
the main PCB / tray assembly to be
pivotted upwards.
Disconnect the back-up battery wiring
(2), then relocate the tray assembly.
Disconnect the wiring connectors (3)
from the PCB .
Disconnect the spirometer tubing
from the connectors (4) on the rear
panel (models with spirometry).
Disconnect the pressure transducer
tubing from the connector (5) on the
PCB.
Remove the two remaining screws (6)
and detach the tray assembly (7) from
the ventilator chassis.
To reassemble, reverse the removal
procedure.

2
1
4

6
3

6
4
5
7

60

9. PARTS LIST
Annual Preventive Maintenance Kit
041204
041226
042819
300045
0314
0762
043
300329

Cat No 57633

O-ring - bellows base


O-ring - bellows base
Quad seal - exhaust valve
Sealing washer - exhaust valve
O-ring - Pressure relief valve
Filter - drive gas inlet
O-ring (Safelock connector - spirometer housing)
Spirometer Tubing Assembly

042819
Quad seal
300045
Sealing
washer
0762 Filter
(drive gas inlet)
Exhaust valve

0314 O-ring
(pressure relief valve)

61

PARTS LIST

A
043
O-ring

300329
Spirometer
Tubing
Assembly

(Safelock
connector spirometer
housing)

041204
O-ring

041226
O-ring
Note
If a paediatric bellows
assembly is used with the
AV800, the O-ring on the
bellows adaptor must be
renewed annually.
Order O-ring, Part No 41225.

62

Spirometer Tubing
Assembly - 300329
The spirometer sample
line fitted to the lower
connector at the rear of
the ventilator has a
restrictor fitted.
This tube is labelled
VENT END (A), and this
end of the tube must be
fitted to the ventilator
rear panel, as illustrated
above.

PARTS LIST
Five Year Overhaul Kit
104019
103996
045538

Cat No 57635

Battery - real time clock


Battery - mains back-up, 12 V
Regulator
Tubing - internal (see next page)

103996
Battery mains back-up

104019
Battery real time
clock

045538
Regulator

63

PARTS LIST
Internal Tubing
NOTE : Replacement tubing must be to the length as specified on illustrations
Ref

Part No.

Qty.

Description

462541

623 mm

Tubing 4 mm o.d.

054541

Adaptor

462544

220 mm

Tubing 8 mm o.d.

462545

149 mm

Tubing 10 mm o.d.

011092

430 mm

Tubing 6 mm o.d.

053211

Plug (fitted only to models without gauge)

054947

Tee piece (3 mm / 2.25 mm)

054949

Connector

Plan View

40 mm

1 (160 mm)

1 (115 mm)

64 mm

3
64

45 mm

Side View
8

1 (18 mm)

1 (330 mm)

5 (350 mm)
6

3
5

5 (80 mm)
7

65

PARTS LIST
1

9
2

10

11
4

5
6
7

Bellows and Base Assembly


Ref

Part No.

Qty.

Description

1
2
3
4
5
6
7

57551
57550
406020
041204
57548
041226
019103
011107

1
1
1
1
1
1
4
4

Canister
Bellows
Diaphragm valve assembly
O ring
Bellows base assembly
O ring
Screw - M5
Knurled knob

8
9
10
11

57553
57552
57544
041225

1
1
1
1

Canister - paediatric (available as an option)


Bellows - paediatric (available as an option)
Adapter assembly - paediatric (available as an option)
O ring
66

PARTS LIST

5
4
2
7

PCBs and Electrical System Components


Ref

Part No.

Qty.

Description

1
2

300xxx
300xxx

1
1

AV80 Main PCB tray assembly


AV800 Display PCB and front panel assembly

3
4
5

104019
103996
104007
300006
105715
1040008
103998

1
1
1
1
1
2
1

Battery - PCB, LiMn, 3 V, 260 mAh, 24.5 mm


Battery - mains back-up, 12 V
Speaker - alarm systems
Speaker and wiring loom
Mains inlet / Filter assembly
Fuse, 3.15 A, 20 mm , anti-surge
Electrical power supply assembly

6
7
8

67

8
av8ser12.tif

PARTS LIST
Rear panel Connectors, Valve Assemblies, and Manometer
Ref

Part No.

Qty.

Description

35150

Hose connector - manometer / pressure transducer

Pressure Relief Valve


2
300073
1
3
300072
1
4
300074
1
5
0314
1
031047
1
054540
1

PRV body
Valve plunger
Cap
O seal, Viton
Spring
Connector

300038
020013
025205
0762
054519

1
1
1
1
1

O2 connector - DISS
Locking nut, M12
Washer, 0.5 inch i.d.
Filter
Straight adaptor. 1/8 BSP to 8 mm tube

Exhaust valve
9
300041
10
300044
11
300045
12
031045
13
300043
14
042819
15
054541

1
1
1
1
1
1
1

Cap
Spacer
Washer
Spring
Piston
Seal, quad ring
Straight adaptor, M5 to 4 mm tube

16

300040

Tapered connector - bellows drive gas

17

049040

Gauge assembly (includes locking ring)

18

104001
104010
104009

1
1
1

Knob - off/standby/mode control


Nut cover
Cap

19

104000
104010
104009

1
1
1

Knob - airway pressure limit control


Nut cover
Cap

20

103999
104002
104003

1
1
1

Knob
Nut cover
Cap

6
7
8

68

PARTS LIST

15

14

16
8

12

13
11
10
9

17

18
19
20

69

PARTS LIST
1

4
5

Pneumatic System Assemblies


Ref

Part No.

Qty.

Description

1
2
3
4
5
6
7
8

300027
045438
102952
054946
054948
045053
045052
300011

1
1
1
1
1
1
1
1

Manifold assembly (includes wiring - not shown, and items 2 to 7:


Regulator assembly (includes O rings)
Pressure switch
Coupling (flexible)
Coupling (includes washer)
Solenoid valve
Solenoid valve
Connector assembly
70

PARTS LIST
1

4
4

Original Specification

7
2

Revised Specification
(May 2000 onwards)

Spirometer System Components


Original Specification

Revised Specification - May 2000 onwards

Ref

Part No.

Qty.

Description

Ref

Part No.

Qty.

Description

300109

Sensor and PCB assembly


(includes tubing and fittings)

300117

Sensor and PCB assembly


(includes tubing and fittings)

57631

300124

34331
043

1
1

Spirometer assembly,
includes:
Safelock nut
O seal

34331
043

1
1

Spirometer assembly,
includes:
Safelock nut
O seal

57632

300329

Hose assembly
(tubing and connectors)

1
1
1
1
1

Luer connector
Luer connector
Nut - tube adaptor
Luer connector
Identification Ring (Red)

Panel connectors
4
011245
5
011244
6
300059
7
011249

1
1
1
1

Hose assembly
(tubing and connectors)

Luer connector
Luer connector
Nut - tube adaptor
Luer connector

Panel connectors
4
011245
5
011244
6
300059
7
011249
8
011260

71

PARTS LIST

72

10. APPENDIX
APPENDIX 1
CARE OF BACK-UP BATTERY
CAUTION
Damage may occur if the battery is allowed to remain in
a discharged state. Never discharge the battery to below 10.2 volts.

A. Battery installed in ventilator


The battery must be charged before the machine is released
for use with an 14 hour charge from the ventilators internal power
supply (ventilator connected to the mains supply, but not
running).
Note that the OFF indicator will show a yellow light
during charging.
Subsequently the recharge periods for a battery on a machine
in store are similar to those in B, below.
Batteries in machines in normal use will be kept charged by the
internal power supply and will only require special charging care
following the discharge test carried out during function testing,
or if the battery is fully discharged following use during mains
failure.

B. Battery care/storage requirements.


During storage batteries will require a periodic recharge, the
frequency of which is determined by the storage temperature,
which must not exceed 50oC (120oF).
Storage
temperature

Recharge
period

38 to 50oC (100 to 122oF)

1 month

21 to 38oC (70 to 100oC)

3 months

7 to 21oF (45 to 70oF)

6 months

0 to 7oC (32 to 45oF)

9 months

-5 to 0oC (23 to 32oF)

12 months

Duration - recharge until the charge current is less than 25 mA (typically overnight).
It is recommended that at each charge an updated label is affixed to each battery to indicate
date of the last charge.

C Battery disposal
Dispose of used batteries according to hospital, local, state, and federal regulations.

73

74

Cat No 52704
Doc No AV8 0103SM
March 2003

Penlon Limited
Abingdon
OX14 3PH
UK

Technical Support
Tel:
+44 (0) 1235 547076
Fax:
+44 (0) 1235 547062
E-mail: [email protected]

International Sales
Tel:
+44 1235 547001
Fax:
+44 1235 547021
E-mail: [email protected]

UK Sales
Tel:
01235 547036
Fax:
01235 547023
E-mail: [email protected]

You might also like