Ventilador de Anestecia AV 800 Service Manual PDF
Ventilador de Anestecia AV 800 Service Manual PDF
Ventilador de Anestecia AV 800 Service Manual PDF
Anaesthesia Ventilator
Service Manual
Technical Support
Penlon Limited
Abingdon
OX14 3PH
UK
1. Type of equipment
2. Product name
3. Serial number
4. Approximate date of purchase
5. Apparent fault
(i)
FOREWORD
(ii)
CONTENTS
Page No.
USER RESPONSIBILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1. WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3. DESCRIPTION
3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Ventilation Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4 Control Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4. SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
5. PRE-OPERATION PROCEDURES
5.1 Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.1 Components Supplied with the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.2 Mounting the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.3 Ventilator Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.4 Ventilator Gas Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.5 Spirometer Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2 Bellows Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.3 Pre-use Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.3.1 Daily Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.3.2 Weekly Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
6. FAULT FINDING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
7. SERVICE SCHEDULE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
8. SERVICE PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.2 Sterilisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.3 Tubing Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.4 Fitting the Annual Preventive Maintenance Kit . . . . . . . . . . . . . . . . . . . . . . 55
(iii)
CONTENTS
9. PARTS LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
10. APPENDIX
Back-up Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
(iv)
USER RESPONSIBILITY
This anaesthesia ventilator has been built to Statements in this manual preceded by the
conform with the specification and operating following words are of special significance:
procedures stated in this manual and/or
accompanying labels and notices when WARNING means there is a
checked, assembled, operated, maintained possibility of injury to
and serviced in accordance with these yourself or others.
instructions.
CAUTION means there is a possibility
To ensure the safety of this device it must be of damage to the apparatus
checked and serviced to at least the or other property.
minimum standards laid out in this manual.
A defective, or suspected defective, product NOTE indicates points of
must not under any circumstances be used. particular interest for more
efficient and convenient
The user must accept responsibility for any
operation.
malfunction which results from non-
compliance with the servicing requirements
detailed in this manual. Always take particular notice of the
warnings, cautions and notes provided
Additionally, the user must accept throughout this manual.
responsibility for any malfunction which may
result from misuse of any kind or non-
compliance with other requirements detailed
in this manual.
1
1. WARNINGS AND CAUTIONS
CAUTIONS
5
6
2. PURPOSE
7
3. DESCRIPTION
Bellows Housing
Bellows Base
Control Unit
8
Driving Gas
Port
Exhaust Port
(patient gas)
Do NOT connect
spirometer
Breathing
System
Port
Spirometer
connectors
DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE
USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS
! COVER. REFER TO QUALIFIED SERVICE PERSONNEL.
SPIRO
FUSE RATING/TYPE
MANOMETER/ PRV 80 cmH2O 3.15 A 240 V ANTI SURGE HRC
PRESSURE Penlon Limited
TRANSDUCER Abingdon, Oxon
-20 To 100 cmH2O OX14 3PH CLASS 1 TYPE B
Tel 01235 547000
Fax 01235 547031
VOLTAGE 90 - 264 VAC
Tested by: Serial No. FREQUENCY 47 - 63 Hz
MADE IN UK MAX CURRENT 1 A
9
DESCRIPTION
1. Beginning of
Inspiratory Phase
DRIVE GAS IN
2. End of Inspiratory
Phase
MAIN DRIVE
GAS VALVE
CLOSED
10
DESCRIPTION
3. Beginning of
Expiratory Phase
DISCHARGE
VALVE
BELLOWS
EXHALATION
DIAPHRAGM 4. End of Expiratory Phase
VALVE
11
PNEUMATIC SYSTEM COMPONENTS
7 6
3
15
1
14
12
18
11 10 13
12
DESCRIPTION
3
4 5 6
8 16
7
1
9 17
11
10
10 12
14
13 18
15
13
DESCRIPTION
The Input Pressure Regulator conditions the spirometry, the actual tidal volume
input drive gas to a stable 35 psig pressure exhaled will be displayed.
which will operate the internal pneumatic
system. The patient must be monitored
independently from the ventilator
The Inspiratory Valve is a large orifice, It is the responsibility of the user to
electro-pneumatically-driven valve which monitor the patient for adequate
supplies the drive gas to the Drive Gas Flow ventilation.
Metering Valve .
A compliance compensation algorithm is
The Drive Gas Flow Metering Valve is a built into the control software which
variable-orifice needle valve which monitors the volume of gas delivered and
determines the drive gas flow rate of the the rate of pressure rise.
bellows during inspiration. It calculates an additional volume to deliver
into the breathing system to compensate for
The Valve Position Feedback Potentiometer the reduced volume delivered to the
and the Flow Control Motor function patient as a result of the increased
together to set a flow rate as required by the breathing system pressure.
front panel controls. As a safety feature, this additional volume is
The flow from the Drive Gas Flow Metering restricted to a maximum of 15% of the set
Valve goes to the Bellows Assembly, via the tidal volume irrespective of the maximum
drive connector, closes the exhalation pressure rise.
diaphragm valve and pushes the bellows
downward. User Selectable Inspiratory Pause
As the bellows moves downwards, the gas A user selectable inspiratory pause is
inside the bellows is forced into the provided which, when activated, holds the
Breathing System. inspiratory phase at the end of inspiration,
for a period of 25% of the inspiratory time
At the end of inspiration the exhaust valve before reverting to the expiratory phase.
opens and allows the drive gas in the top of
the bellows housing to exhaust out through NOTE
the Exhaust Outlet. This function is not available in
As the pressure in the top of the bellows PRESSURE controlled cycle
housing is reduced to zero, the patient
exhales into the breathing system and the
bellows rises. Automatic altitude compensation
Ambient pressure is monitored and the
Compliance Compensation ventilator automatically compensates the
delivered volume according to the local
atmospheric pressure.
WARNING
The AV800 has compliance
Manometer (if fitted)
compensation but the actual tidal volume
The manometer connection is connected
delivered to the patient may be different
to the pressure gauge.
to the ventilation parameters set by the
user due to:
A) an extreme compliance condition,
Patient Gas Pressure Transducer
B) a substantial system leak, or The Patient Gas Pressure Transducer is
C) patient circuit pressure effects. connected to the patient breathing system
In addition, high fresh gas flows will lead via the manometer connection.
to an increased Vt being delivered to the
patient. In VOL CYCLE mode whenever this
Note that on models fitted with pressure exceeds the maximum working
14
DESCRIPTION
15
DESCRIPTION
16
DESCRIPTION
At this point the ventilator will stop In addition,high fresh gas flows will lead
delivering gas and pause in this condition to an increased Vt being delivered to the
until the calculated end of inspiration time patient.
has been reached. Note that on models fitted with
The ventilator will then return to the spirometry, the actual tidal volume
exhalation phase. exhaled will be displayed.
The patient must be monitored
The inspiratory pause function does not independently from the ventilator.
operate in this mode. It is the responsibility of the user to
monitor patient ventilation.
NOTE
When the ventilator is switched from standby 3.5.4 Ventilator I:E Ratio
to the required operating mode, the
spirometer will take 1 - 2 minutes before WARNING
stabilising The ventilator settings can allow for and
inverse I:E ratio up to 1:0.3. The clinician
The TIDAL VOLUME is set by the control must always ensure that sufficient time is
knob. If the set parameters are within normal allowed for the patient to adequately
limits, the system will deliver the set volume exhale.
indicated.
If during adjustment the required flow rate is I:E RATIO
less than 2 L/min or exceeds 75 L/min, the This dial and digital display controls and
set tidal volume will be limited accordingly. indicates the requested ventilator I:E RATIO.
18
DESCRIPTION
In the PRESSURE CONTROLLED mode Pressing the print button a second time will
the inspiration is terminated whenever cancel the print function.
this set pressure is achieved and the Turn the ventilator to OFF, and switch the
ventilator pauses until the end of the ventilator back on again to clear the stored
calculated inspiratory time before reverting print information.
to the exhalation phase. The printer port must only be connected to
devices that comply to EN 60950.
19
DESCRIPTION
NOTE
This symbol denotes
Type B equipment.
DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE
USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS
! COVER. REFER TO QUALIFIED SERVICE PERSONNEL.
SPIRO
FUSE RATING/TYPE
MANOMETER/ PRV 80 cmH2O 3.15 A 240 V ANTI SURGE HRC
PRESSURE Penlon Limited
Abingdon, Oxon
TRANSDUCER
-20 To 100 cmH2O
OX14 3PH CLASS 1 TYPE B
Tel 01235 547000
Fax 01235 547031
VOLTAGE 90 - 264 VAC
Tested by: Serial No. FREQUENCY 47 - 63 Hz
MADE IN UK MAX CURRENT 1 A
21
DESCRIPTION
22
DESCRIPTION
LOW BATTERY
(Low priority / Medium priority)
A visual indicator and an audible alarm
which activate if the internal battery is almost
exhausted, indicating that ventilation will
soon cease.
VENT INOP
(High priority)
The VENT INOP (ventilator inoperative)
alarm indicates that one of the following
conditions has occurred:
WARNING
The only way to reset a VENT INOP alarm
is to turn the Power switch OFF for a
minimum of one second and then back
ON.
If the VENT INOP alarm occurs again,
remove the ventilator from use and refer
to an authorised service technician.
23
DESCRIPTION
I:E Ratio
1:6 1:5 1:4 1:3 1:2 1:1 1:0.3
1.6
1.5
1.4
1.3
1.2
1.1
1.0
Tidal
Volume 0.9
(litres) 0.8
(Vt) 0.7 X
0.6
0.5
0.4
0.3
0.2
0.1
0
0 5 10 15 20 25 30 35 40 45 50 55 60
Rate (bpm)
The ventilator is capable of operating at the volumes and rates below each I:E ratio curve.
Example
1. Select required volume (Vt) (e.g. 0.7 litres)
2. Select rate (e.g. 10 bpm).
3. Select I:E ratio of 1:2.
The point X on the graph lies beneath the 1:2 ratio curve, and is therefore within the
ventilator’s capability.
24
DESCRIPTION
Upgrades
Any upgrades will be offered to customers for
downloading via a connector on the main PCB.
1 Ground
2 Breathing Circuit Pressure -20 to 80 cmH20 0.05 V/cmH20
3 Average Breathing Circuit Pressure -20 to 80 cmH20 0.05 V/cmH20
4 Airway Pressure Limit 10 to 70 cmH20 0.05 V/cmH2O
5 Tidal Volume 0-1.6 litres 0.3 V/100ml
25
DESCRIPTION
Alarm Outputs
9 System OK High
10 Power problem - mains fail, Low
battery missing or battery low
11 Incorrect rate/ratio Low
12 Low measured tidal volume Low
13 Low supply pressure Low
Engineering Mode
A facility to allow Penlon-trained service engineers
access to the following functions:
27
SPECIFICATION
4.29 Alarms (in priority order) See section 3 for detailed descriptions.
28
SPECIFICATION
4.33 Weight
Weight with adult bellows 9.0 kg
Weight with paediatric bellows 8.7 kg
Weight of control unit only 7.6 kg
4.42 Environmental
Ambient Temperature
Storage -5 to 50oC (23 to 122oF).
Refer to Appendix 1 for battery care during
storage.
Operating 10 to 38oC (50 to 100oF)
29
5. PRE-OPERATION PROCEDURES
30
Driving Gas
Port
Exhaust Port
(patient gas)
Do NOT connect
spirometer
Breathing
System
Port
Spirometer
connectors
DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE
USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS ! COVER. REFER TO QUALIFIED SERVICE PERSONNEL.
SPIRO
31
PRE-OPERATION PROCEDURES
AV800 VENTILATOR
(REAR VIEW)
EXHAUST TO
SCAVENGE AV800 ventilator connected as part of a
SYSTEM
circle breathing system
BREATHING CIRCUIT
BACTERIAL FILTER
PRESSURE MONITOR LINE
SPIROMETER
4. Connect the spirometer (if fitted) - see expiratory limb to protect the
Section 5.1.5. breathing system.
WARNING WARNING
Applying negative or positive pressure to The High and Low Airway Pressure
the bellows exhaust port results in Alarms are important for patient care.
positive pressure in the patient breathing The ventilator is designed to be used with
system. a distal sensing tee only. (Catalogue No.
Therefore, the scavenging system must 53194, Breathing System Tee Assembly.)
not generate more than 0.5 cmH2O It is important that the distal sensing tee
positive or negative pressure when is properly located in the inspiratory limb
connected to the ventilator. of the breathing system.
Any problem arising from an improperly 6. Set the AIRWAY PRESSURE LIMIT
functioning scavenging system is solely control to 50 cmH2O.
the user`s responsibility.
7. Attach a printer to the printer port if a
5. Connect the MANOMETER/ printed output of the ventilator
PRESSURE TRANSDUCER port on function is required.
the rear panel of the control unit to the
breathing system. 8. Connect the ventilator bellows base
The recommended placement for the BREATHING SYSTEM port to the
distal sensing tee is in the inspiratory breathing system.
limb of the breathing system close to Use a breathing system bacterial filter
the circle system inspiratory valve. in the expiratory limb of the breathing
Use a Heat and moisture exchanger circuit or a heat and moisture
(HME) at the patient Y piece, or a exchanger (HME) at the patient Y
breathing system filter in the piece. See illustration above.
32
PRE-OPERATION PROCEDURES
33
PRE-OPERATION PROCEDURES
A1
B2
B1 D
C1
C2
Spirometer
E
Head
Pressure
A2
Monitor Line
Notes
a) Carefully twist each sample
line anti-clockwise before
connecting to the ventilator and
5.1.5 Spirometer Connections Tee-piece.
WARNING The lumen assembly must be Turn each connector fully
clockwise to ensure a leak-free
connected as illustrated. Do not connect to the
joint.
expiratory limb of the breathing circuit. b) Latest type tubing assemblies
have a restrictor fitted to tube
The triple lumen assembly incorporates the C1/C2.
ventilator pressure monitor line (shown as This tube is labelled ‘VENT END’
Tube A1/A2). (E) to assist correct fitment.
This line must be connected to the breathing
circuit at the absorber expiratory hose
connector.
1. Use a breathing system bacterial filter 4. Check for secure connection at each
in the expiratory limb, or a heat and connector.
moisture exchanger (HME) at the NOTE
patient Y piece. See section 5.1.4, A) If the connection is incorrectly made at the
operation 8. spirometer sensor, the ventilator will alarm LOW
TIDAL VOLUME. To allow the ventilator to be
used in the event of damage or or non-functioning
2. Fit the spirometer head to the Safelock of the spirometer head, turn off the spirometry
connector at the ventilator port on the function - see STANDBY mode.
back of the Circle Absorber. B) Connection of the spirometer into the exhaled
limb also provides an indication of breathing
3. Connect the tubing from the spirometer system disconnect.
head to the connectors at the rear of the
C) When testing the spirometer prior to clinical
AV800 ventilator as shown. use, it is preferable to use a 70% N2O / 30% O2
For additional identification: mixture.
i) The centre tube (C1/C2) has a RED If 100% oxygen is used, the Measured Vt value
display may flash. This is not significant, and
connector at C2 which must be fitted to should be disregarded.
the lower connector on the ventilator
(fitted with a RED washer ,D). Absorber switched to Bag mode
Later assemblies have a label (E) (VENT NOTE If the absorber bag/ventilator switch
END) on tube C1/C2 is set to ‘Bag’, the spirometer will not
ii) Connector B1 is BLUE. measure tidal volume.
34
PRE-OPERATION PROCEDURES
5.2 Bellows Assembly 3. Fit the bellows, and then fit the
bellows housing to the base by
CAUTION pushing it down, then twisting it
Always ensure correct fitment of bellows slightly clockwise until the bayonet
(see illustration above), and carry out a full tabs completely engage.
function test before clinical use, if a bellows 4. Function test the ventilator - section
is removed and refitted. 5.3.1.
35
PRE-OPERATION PROCEDURES
The following tests should be done at least 4. Connect a 2-litre breathing bag to the
at the beginning of every working day: patient connection as a test lung.
36
PRE-OPERATION PROCEDURES
NOTE
If there is any malfunction, the ventilator
must NOT be used.
37
6. FAULT FINDING
LOW BATTERY message is on 1. Charge battery by plugging unit into the mains
supply.
LOW DRIVE GAS SUPPLY alarm activated 1. Check drive gas supply pressure (> 38 psi)
38
FAULT FINDING
Continuous high airway pressure limit 1. Potentiometer set too low (or knob out of
alarm calibraion) - adjust.
2. Error code 12
Potentiometer failure - replace display assembly
PCB.
3. Error code 8
Potentiometer sensor failure - replace main PCB.
Unwanted PEEP and bulging, full bellows 1. Defective or poorly regulated scavenging system,
creating too high resistance or too much vacuum.
39
FAULT FINDING
Zero tidal volume and drive gas leak from 1. Error code 3
exhaust port. Expiratory solenoid failed open - replace.
Tidal volume incorrect or inconsistent 1. Missing small O-ring under the exhalation
diaphragm valve.
7. Error code 1
Flow valve fails to adjust correctly due to
mechanism binding - realign and check calibration.
40
FAULT FINDING
Tidal volume incorrect or inconsistent 9. Spirometer faulty - replace spirometer head and
(continued) lead.
41
FAULT FINDING
42
7. SERVICE SCHEDULE
6 Month Service
12 Month Service
5 year Overhaul
1. Fit overhaul kit components :
Software Upgrades
Special Tools
Part No. Description
------ Multimeter
300128 PRV and Exhaust - cap removal tool
------ Test Gauge
43
SERVICE SCHEDULE
Item Operation
Service frequency
1 Initial Checks
1.1 Check and record serial number and determine service required.
1.2 Check general condition of ventilator assembly Six months
1.3 Check configuration of attachments and note. (ALL item 1
1.4 Check and record front panel control settings. operations)
44
Check Date - Check Date -
Year Month
Year shown - 1998 Month shown -
Turn RATE knob December
to adjust. Turn RATE knob
Turn I:E knob to adjust.
clockwise to get to Turn I:E knob
Check Date - clockwise to get
Month to Check Date -
Day
45
Check Revision - Check Revision -
Main Board Front Panel
Spirometer tubing
and housing
Exhaust valve
Pressure relief valve
46 (PRV)
SERVICE SCHEDULE
Service frequency
3 Electrical Safety Checks Six months
Carry out the electrical safety checks with the ventilator power switch (ALL item 3
in the ‘OFF’ position. operations)
3.1 Earth continuity (max 0.2 ohms )
3.2 Insulation resistance ( not less than 10 m ohms )
3.3 Earth leakage ( max 100 micro amps )
6 Control Unit
6.1 Check all electrical connections and components for security. Six months
6.2 Replace Back up battery. Five years
6.3 Replace real time clock battery. Five years
47
SERVICE SCHEDULE
49
8. SERVICE PROCEDURES
50
SERVICE PROCEDURES
A C
Exhalation Diaphragm
Valve Assembly
Bellows Base
51
SERVICE PROCEDURES
52
SERVICE PROCEDURES
8.2 Sterilisation
Recommended guidelines for sterilisation:
CAUTION
To prevent possible damage to components, peak sterilisation temperatures must not exceed
54oC (130oF) for gas (ethylene oxide) or, 134oC (275oF) for steam autoclave.
Do not sterilise the ventilator control unit. The internal components are not compatible with
sterilisation techniques and may be damaged.
Following sterilisation with ethylene oxide, components must be quarantined in a well
ventilated area to allow dissipation of any residual gases.
Follow the recommendations given by the steriliser manufacturer for aeration periods.
ITEM METHOD
Bellows * Gas, liquid, autoclave (20 cycles max.)
Hoses Gas, liquid, autoclave
O rings Gas, liquid, autoclave.
Bellows base Gas, liquid, autoclave.
Exhalation valve assembly Gas, liquid, pasteurise, low temperature autoclave.
Control unit Do not sterilise - see section 9.1.
Bellows canister Liquid.
Spirometer sensor tee piece Gas, liquid, autoclave
Spirometer sample lines Liquid
Distal sensing Tee Liquid
*Applies to standard size bellows only, the optional paediatric bellows must not be steam
autoclaved.
NOTE
Examples of suitable liquid agents are: Nu-Cidex, Sporicidin, and Sonacide.
The exhalation diaphragm valve must be removed, cleaned and sterilised sep
arately.
53
SERVICING PROCEDURES
54
SERVICING PROCEDURES
1
041204
O-ring 6
7
3
0762 Filter
2 (drive gas
inlet)
5
4
041226 042819
O-ring O-ring
Exhaust valve
1. Use service tool to unscrew the exhaust valve
cover (8).
2. Remove the valve components and fit a new
sealing washer (9) and O-ring (10). A
Internal Tubing 6
11. Renew all internal tubing (including
spirometer tubing if fitted).
Cut the new tubing to the exact
lengths specified below.
Spirometer accuracy is dependant
on correct tubing lengths.
56
4
64 mm 45 mm
40 mm
Plan
View
1 (160 mm)
5
5 (80 mm)
8
57
SERVICING PROCEDURES
58
SERVICING PROCEDURES
59
SERVICING PROCEDURES
NOTE 1
The main PCB is not available as a
separate item. A tray assembly complete
with pre-programmed PCB must be ordered
if the PCB is faulty.
60
9. PARTS LIST
042819
Quad seal
300045
Sealing
washer
0762 Filter
(drive gas inlet)
Exhaust valve
0314 O-ring
(pressure relief valve)
61
PARTS LIST
043
O-ring
(Safelock 300329 Spirometer Tubing
connector - Spirometer Assembly - 300329
spirometer Tubing
housing) Assembly The spirometer sample
line fitted to the lower
connector at the rear of
the ventilator has a
restrictor fitted.
This tube is labelled
‘VENT END’ (A), and this
end of the tube must be
fitted to the ventilator
041204 rear panel, as illustrated
O-ring above.
041226
O-ring
Note
If a paediatric bellows
assembly is used with the
AV800, the O-ring on the
bellows adaptor must be
renewed annually.
Order O-ring, Part No 41225.
62
PARTS LIST
103996
Battery -
104019 mains back-up
Battery -
real time
clock
045538
Regulator
63
PARTS LIST
Internal Tubing
NOTE : Replacement tubing must be to the length as specified on illustrations
2 054541 1 Adaptor
3 462544 220 mm Tubing 8 mm o.d.
4 462545 149 mm Tubing 10 mm o.d.
5 011092 430 mm Tubing 6 mm o.d.
6 053211 1 Plug (fitted only to models without gauge)
7 054947 1 Tee piece (3 mm / 2.25 mm)
8 054949 1 Connector
4
Plan View 64 mm 45 mm
40 mm
1 (160 mm)
1 (115 mm)
2
3
64
5 Side View
8
1 (18 mm)
3
5
5 (80 mm)
7
65
PARTS LIST
1
8
10
11
3 4
1 57551 1 Canister
2 57550 1 Bellows
3 406020 1 Diaphragm valve assembly
4 041204 1 O ring
5 57548 1 Bellows base assembly
6 041226 1 O ring
7 019103 4 Screw - M5
011107 4 Knurled knob
66
PARTS LIST
4
2
7
6 8
7 av8ser12.tif
67
PARTS LIST
Exhaust valve
9 300041 1 Cap
10 300044 1 Spacer
11 300045 1 Washer
12 031045 1 Spring
13 300043 1 Piston
14 042819 1 Seal, quad ring
15 054541 1 Straight adaptor, M5 to 4 mm tube
20 103999 1 Knob
104002 1 Nut cover
104003 1 Cap
68
PARTS LIST
2 15
1 3 4
14
12
16 13
11
8 7 10
9
17
18
19
20
69
PARTS LIST
1
2 3
4
5
8
70
PARTS LIST
4
6
5
3 Original Specification
7
2 2
8
Revised Specification
3 (May 2000 onwards)
1 300109 1 Sensor and PCB assembly 1 300117 1 Sensor and PCB assembly
(includes tubing and fittings) (includes tubing and fittings)
71
PARTS LIST
72
10. APPENDIX
APPENDIX 1
CARE OF BACK-UP BATTERY
CAUTION
Damage may occur if the battery is allowed to remain in
a discharged state. Never discharge the battery to below 10.2 volts.
Storage Recharge
temperature period
Duration - recharge until the charge current is less than 25 mA (typically overnight).
It is recommended that at each charge an updated label is affixed to each battery to indicate
date of the last charge.
C Battery disposal
Dispose of used batteries according to hospital, local, state, and federal regulations.
73
74
Cat No 52704
Doc No AV8 0103SM
March 2003
Penlon Limited
Abingdon
OX14 3PH
UK
Technical Support
Tel: +44 (0) 1235 547076
Fax: +44 (0) 1235 547062
E-mail: [email protected]
International Sales
Tel: +44 1235 547001
Fax: +44 1235 547021
E-mail: [email protected]
UK Sales
Tel: 01235 547036
Fax: 01235 547023
E-mail: [email protected]