Ventilador de Anestecia AV 800 Service Manual PDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 88

AV800

Anaesthesia Ventilator
Service Manual

Quality and Assurance in Anaesthesia


IMPORTANT

Servicing and Repairs

In order to ensure the full operational life of this


ventilator, servicing by a Penlon-trained
engineer should be undertaken periodically.

The ventilator must be serviced to the following


schedule:

(a) Six monthly service - inspection and


function testing.
(b) Annual service.
(c) Five year major service including battery
replacement.

Details of these operations are given in this


AV800 Service Manual, available only for
Penlon trained engineers.

For any enquiry regarding the servicing or


repair of this product, contact the nearest
accredited Penlon agent:

or communicate directly with:

Technical Support
Penlon Limited
Abingdon
OX14 3PH
UK

Tel: +44 (0) 1235 547076


Fax: +44 (0) 1235 547062
E-mail: [email protected]

Always give as much of the following


information as possible:

1. Type of equipment
2. Product name
3. Serial number
4. Approximate date of purchase
5. Apparent fault

(i)
FOREWORD

This manual has been produced to provide


authorised personnel with information on the
function, routine performance, servicing,
maintenance checks and repairs applicable
to the AV800 Anaesthesia Ventilator.

Information contained in this manual is


correct at the date of publication.
The policy of Penlon Limited is one of
continued improvement to its products.
Because of this policy, Penlon Limited
reserves the right to make any changes
which may affect instructions in this manual,
without giving prior notice.

Personnel must make themselves familiar


with the contents of this manual and the
machine’s function before servicing the
apparatus.

Copyright © Penlon Limited, 2003.


All rights reserved.

(ii)
CONTENTS
Page No.
USER RESPONSIBILITY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1. WARNINGS AND CAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2. PURPOSE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

3. DESCRIPTION
3.1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Ventilation Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.3 Pneumatic System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.4 Control Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

3.5 Operating Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17


3.5.1 Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.5.2 Tidal Volume Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.3 Ventilation Rate Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.4 Ventilation I:E Ratio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.5.5 Airway Pressure Limit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5.6 Tidal Volume Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5.7 Inspiratory Pause . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5.8 Print . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

3.6 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20


3.7 Message Displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.8 Back-up Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.9 Operational Capability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.10 System software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25

4. SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

5. PRE-OPERATION PROCEDURES
5.1 Set-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.1 Components Supplied with the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.2 Mounting the Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.3 Ventilator Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.4 Ventilator Gas Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.1.5 Spirometer Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
5.2 Bellows Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5.3 Pre-use Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.3.1 Daily Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.3.2 Weekly Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

6. FAULT FINDING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

7. SERVICE SCHEDULE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

8. SERVICE PROCEDURES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.1 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
8.2 Sterilisation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8.3 Tubing Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.4 Fitting the Annual Preventive Maintenance Kit . . . . . . . . . . . . . . . . . . . . . . 55

(iii)
CONTENTS

8.5 Fitting the Five Year Overhaul Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56


8.6 Manifold Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.7 Electrical Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
8.8 Front Panel and PCB Assembly, and Gauge . . . . . . . . . . . . . . . . . . . . . . . 59
8.9 Main PCB Tray Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

9. PARTS LIST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61

10. APPENDIX
Back-up Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73

(iv)
USER RESPONSIBILITY

This anaesthesia ventilator has been built to Statements in this manual preceded by the
conform with the specification and operating following words are of special significance:
procedures stated in this manual and/or
accompanying labels and notices when WARNING means there is a
checked, assembled, operated, maintained possibility of injury to
and serviced in accordance with these yourself or others.
instructions.
CAUTION means there is a possibility
To ensure the safety of this device it must be of damage to the apparatus
checked and serviced to at least the or other property.
minimum standards laid out in this manual.
A defective, or suspected defective, product NOTE indicates points of
must not under any circumstances be used. particular interest for more
efficient and convenient
The user must accept responsibility for any
operation.
malfunction which results from non-
compliance with the servicing requirements
detailed in this manual. Always take particular notice of the
warnings, cautions and notes provided
Additionally, the user must accept throughout this manual.
responsibility for any malfunction which may
result from misuse of any kind or non-
compliance with other requirements detailed
in this manual.

Worn, broken, distorted, contaminated or


missing components must be replaced
immediately. Should such a repair become
necessary it is recommended that a request
for service advice be made to the nearest
accredited Penlon agent.

This device and any of its constituent parts


must be repaired only in accordance with
written instructions issued by Penlon
Limited and must not be altered or modified
in any way without the written approval of
Penlon Limited. The user of this equipment
shall have the sole responsibility for any
malfunction which results from improper
use, maintenance, repair, damage or
alteration by anyone other than Penlon or its
appointed agents.

USA and Canadian Federal Law restricts the


sale and use of this device to, or on the order
of, a licensed practitioner.

1
1. WARNINGS AND CAUTIONS

The following WARNINGS and CAUTIONS cannot be guaranteed.


must be read and understood before See section 8 for battery
servicing or repairing this ventilator. maintenance.
See also CAUTION No. 7.
WARNINGS
6. No oil, grease or other flammable
General Information lubricant or sealant must be used
on any part of the machine in close
proximity to medical gas
1. Personnel must make themselves distribution components.
familiar with the contents of this There is a risk of fire or explosion.
manual and the machine’s function
before using the ventilator. 7. Exterior panels must not be
removed by unauthorised
Servicing and Repair personnel and the apparatus must
not be operated with such panels
2. Opening the control unit by missing.
unauthorised personnel There is a possible electric shock
automatically voids all warranties hazard.
and specifications. Always disconnect the ventilator
from the mains electrical supply
Prevention of tampering with the and drive gas supply before
control unit is exclusively the removing the cover.
user’s responsibility. If the control
unit seal is broken, the
Before Using the Ventilator
manufacturer assumes no liability
for any malfunction or failure of the
ventilator. 8. Before the ventilator is used
clinically for the first time, verify
3. Great care must be taken not to that the hospital engineering
damage the precision surface of department has carried out an earth
the valve seat on the patient gas continuity test.
exhalation diaphragm valve in the
base of the bellows assembly. 9. Excessive electronic noise caused
by other poorly regulated devices,
Never use any hard object or such as an electrocautery unit, may
abrasive detergent to clean it; use adversely interfere with the proper
only a soft cloth. functioning of the ventilator.
If the valve seat is damaged, the
valve will leak and may cause To avoid this problem, do not
serious ventilator malfunction. connect the ventilator’s power cord
into the same electrical wall outlet
4. For continued protection against or adaptor strip into which an
fire hazards, replace the two fuses electrocautery unit is connected.
only with the identical type and
rating of fuse. 10. If used with a mains extension cord,
See section 4.39 for fuse rating. the unit may be subject to electro-
magnetic interference.
5. If the internal battery is fully
discharged, the ventilator will not 11. The driving gas supply must be
function in the event of mains clean and dry to prevent ventilator
power failure. The battery must be malfunction.
recharged before the ventilator is
used clinically, otherwise backup
2
WARNINGS AND CAUTIONS

12. This ventilator is designed to be 15. The breathing system which


driven by oxygen or medical air conveys gases from the
only. It is calibrated during anaesthetic machine to the patient,
manufacture for use with either and disposes of expired gases, is a
gas. vital part of the anaesthetic delivery
Before the ventilator is used system.
clinically for the first time, the Because breathing systems require
commissioning engineer must frequent cleaning and disinfection
confirm that the internal they are not a permanent part of the
Air/Oxygen switch is set correctly anaesthetic ventilator and
for the gas that is to be used. therefore cannot be directly under
The use of any other gas will cause the control of the anaesthetic
inaccurate operation and may ventilator manufacturer.
damage the ventilator, resulting in However, we strongly recommend
potential injury to the patient. that only breathing systems which
have been approved and
authorised by Penlon for use with
13. The driving gas is discharged AV800 should be employed.
through the opening in the back of
the ventilator control unit. This Do not use conductive or anti-static
opening is labelled: EXHAUST DO breathing system hoses.
NOT BLOCK and must be
completely free of any obstruction 16. Applying negative or positive
and nothing should be connected pressure to the exhaust port may
to it. result in positive or negative
pressure in the patient breathing
The discharged gas does not system. Therefore the scavenging
contaminate the environment, but system must not generate more
on machines using oxygen as the than 0.5 cmH2O positive or
drive gas it can cause a fire hazard negative pressure when connected
if allowed to accumulate. to the ventilator.

In addition, do not block the PRV Any problem arising from an


outlet on the back of the control improperly functioning scavenging
unit, system is solely the user’s
and also the pneumatic system responsibility.
outlet on the base of the unit.
17. Do not connect a spirometer to the
14. The bellows can only support exhaust port on the bellows base.
approximately 1 kPa (10 cmH2O) The device will not measure
differential positive pressure, exhaled volumes.
above which it may be dislodged
from the mounting ring, resulting in 18. The operation of each alarm
dangerous malfunction of the function should be verified daily.
ventilator. See section 5.3.1.
Periodically check the alarms at
Do not connect a PEEP valve or suitable intervals. If the audible
other restrictive device to the alarm or the visual indicator of any
exhaust port on the bellows base. alarm function fails to activate
during any alarm condition or fails
This would increase the pressure to reset after the alarm has been
inside the bellows and the bellows cleared, refer the unit to an
could detach from the base, authorised service technician.
causing serious malfunction.
3
WARNINGS AND CAUTIONS

19. Before using the ventilator check monitored continuously, at or near


that all connections are correct, the proximal airway with an oxygen
and verify that there are no leaks. monitor that includes high/low
alarms.
Patient circuit disconnects are a
hazard to the patient. Extreme care 25. If the drive gas supply pressure
should be taken to prevent such drops below a nominal 262 kPa (38
occurrences. psig), the LOW DRIVE GAS
SUPPLY alarm will activate both
It is recommended that Penlon audibly and visually. Patient
Safelok fittings are used minute volume may be reduced due
throughout the breathing circuit. to lowered flow rates

26. An audible alarm indicates an


Using the Ventilator anomalous condition and should
never go unheeded.
20. This apparatus must not be used
with, or in close proximity to, 27. The characteristics of the breathing
flammable anaesthetic agents. circuit connected between the
ventilator and the patient can
There is a possible fire or modify or change patient
explosion hazard. ventilation.

21. Anaesthesia apparatus must be To assist the maintenance of the


connected to an anaesthetic gas delivered patient tidal volume, the
scavenging system to dispose of ventilator control system software
waste gas and prevent possible includes a compliance
health hazards to operating room compensation algorithm.
staff. This requirement must be
observed during test procedures However, patient ventilation must
as well as during use with a patient. be monitored independently from
Any problem arising from an the ventilator.
improperly functioning scavenging It is the responsibility of the user to
system is solely the user’s monitor patient ventilation.
responsibility.
28. On models with spirometry, the
22. When the ventilator is connected to spirometer flow sensor must be
a patient, it is recommended that a installed in the expiratory limb of
qualified practitioner is in the breathing circuit.
attendance at all times to react to
an alarm or other indication of a A breathing system filter, or heat
problem. and moisture exchanger (HME),
must be connected upstream of the
23. In compliance with good sensor to prevent blockage of the
anaesthesia practice, an alternative side-stream sample lines
means of ventilation must be
available whenever the ventilator is 29. The Vent Inop (ventilator
in use. inoperative) alarm indicates that
one of the following conditions has
24. The AV800 is not equipped with an occurred:
oxygen analyser. It is A) A solenoid has failed.
recommended that the patient B) The flow control valve has failed.
oxygen concentration be C) Internal electrical fault.
D) Internal communications error
4
WARNINGS AND CAUTIONS

E) Battery disconnected 4. The exhalation valve located in the


F) Software error bellows base assembly and the
paediatric bellows adaptor must be
Note that if a ventilator error is cleaned and sterilised separately. See
detected, an error code will be sections 7.2 and 7.3.
displayed in the Rate BPM display
window on the front control panel. 5. Care must be taken not to let any
liquid run into the control unit; serious
30. The High and Low Airway Pressure damage may result.
Alarms are important for patient
care. 6. Always check for correct fitment, and
The ventilator is designed to be carry out a full function test before
used with a distal sensing tee only. clinical use, if the bellows has been
(Catalogue No. 53194, Breathing removed and refitted for any reason.
System Tee Assembly - see section See sections 5.2 and 7.2.
8 in the user manual).
7. Damage may occur to the battery if it
The distal sensing tee must be is allowed to remain in a discharged
located close to the circle system, state.
in the inspiratory limb of the circuit
between the patient and the circle Check the battery frequently if the
system inspiratory one way valve. ventilator is in storage (see Appendix
See section 5.1.4. 1).

CAUTIONS

1. Do not sterilise the ventilator control


unit.
The internal components are not
compatible with sterilisation
techniques and damage may result.

2. For ventilator components which


require sterilisation, peak sterilisation
temperatures should not exceed
136oC (275oF) to prevent possible
damage. (See sections 7.2 and 7.3).

3. Those parts suitable for ethylene


oxide sterilisation should, following
sterilisation, be quarantined in a well
ventilated area to allow dissipation of
residual gas absorbed by the
components.

Follow the steriliser manufacturer’s


recommendations for any special
aeration periods required.

5
6
2. PURPOSE

The AV800 Ventilator is a software


controlled, multi-mode ventilator, designed
for mechanical ventilation of adult and
paediatric patients under general
anaesthesia.
In addition, in spontaneous mode, it can be
used to monitor spontaneously breathing
patients
It is designed for use in closed-circuit
anaesthesia and also to drive a Mapleson D
circuit.

7
3. DESCRIPTION

Bellows Housing

Bellows Base

Control Unit

Fig. 1 AV800 Ventilator

3.1 General The ventilator drive gas supply can be


oxygen or air, and the supply must be at 38
The AV800 is a multi-mode, time-cycled, to 100 psig. Note that the drive gas is
volume/pressure controlled, and pressure specified by the customer prior to delivery.
limited ventilator for closed circuit ventilation To change the drive gas, refer to a Penlon-
or for use with a Mapleson D circuit. trained service engineer.
The ventilator is compliance compensated
and has a user selectable option of an Models are available with a manometer
inspiratory pause fixed at 25% of the and/or spirometry, and the ventilator can be
inspiratory time. used with power supplies from 110 to 240
The print function provides a permanent VAC.
record of function activity for up to eight
hours during a procedure. All components used in the ventilator,
The bellows unit can be easily detached and including the bellows, are latex-free ( but
then refitted to the bellows base assembly to note that the optional paediatric bellows
facilitate cleaning. is not latex free).

8
Driving Gas
Port

Exhaust Port
(patient gas)
Do NOT connect
spirometer
Breathing
System
Port

Spirometer
connectors

DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE
USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS
! COVER. REFER TO QUALIFIED SERVICE PERSONNEL.

RS232 PORT PRINTER PORT ANALOGUE/ALARM OUTPUTS

SPIRO

EXHAUST VALVE MAINS SUPPLY


OXYGEN/AIR DO NOT BLOCK
38 - 100 psi Drive Gas
2.6 - 6.9 Bar
262 - 689 kPa

FUSE RATING/TYPE
MANOMETER/ PRV 80 cmH2O 3.15 A 240 V ANTI SURGE HRC
PRESSURE Penlon Limited
TRANSDUCER Abingdon, Oxon
-20 To 100 cmH2O OX14 3PH CLASS 1 TYPE B
Tel 01235 547000
Fax 01235 547031
VOLTAGE 90 - 264 VAC
Tested by: Serial No. FREQUENCY 47 - 63 Hz
MADE IN UK MAX CURRENT 1 A

Fig. 2 Ventilator Rear Panel and Gas Ports

9
DESCRIPTION

3.2 Ventilation Cycle

This section provides a simplified


description of the ventilation cycle.

1. Beginning of
Inspiratory Phase

Drive gas pressure builds up


above the bellows, which starts
to move down, forcing patient
PATIENT gas into the breathing system.
GAS OUT TO
BREATHING
CIRCUIT

DRIVE GAS IN

2. End of Inspiratory
Phase

The main drive gas valve


closes and the bellows stops
moving.

MAIN DRIVE
GAS VALVE
CLOSED

10
DESCRIPTION

3. Beginning of
Expiratory Phase

The discharge valve opens,


allowing the drive gas above
the bellows to escape to
atmosphere.

The bellows starts to rise and


exhaled gas enters the bellows.

DISCHARGE
VALVE

BELLOWS
EXHALATION
DIAPHRAGM 4. End of Expiratory Phase
VALVE

The bellows exhalation


diaphragm valve in the base of
BELLOWS the bellows assembly opens
EXHAUST PORT
when the bellows reaches the
top of the chamber.
Patient circuit gas exits through
the bellows assembly exhaust
port.

11
PNEUMATIC SYSTEM COMPONENTS

1. Gas supply inlet


2. Input gas filter
3. Low supply pressure detector
4. Input pressure regulator
5. Test point
17 6. Inspiratory valve
7. Drive gas flow metering valve
8. Stepper motor and feedback
potentiometer (for 7)
9. Expiratory valve
10. Exhaust outlets (outlet on rear panel
as illustrated, the outlet through the
base of the control unit is not shown)
11. Pressure relief valve
12. Exhaust valve
13. Bellows drive gas connector
14. Driving gas over-pressure switch
15. Bellows driving gas port
5 16. Pressure gauge
17. Pressure transducer
18. Inlet from breathing circuit /
manometer connection

7 6
3

15

1
14

12

18
11 10 13

12
DESCRIPTION

3
4 5 6
8 16
7
1

9 17
11
10

10 12
14
13 18

15

PNEUMATIC SYSTEM DIAGRAM


1. Gas supply 10. Exhaust outlets
2. Input gas filter 11. Pressure relief valve
3. Low supply pressure detector 12. Exhaust valve
4. Input pressure regulator 13. Bellows drive gas connector
5. Test point 14. Driving gas over-pressure switch
6. Inspiratory valve 15. Bellows driving gas port
7. Drive gas flow metering valve 16. Pressure gauge
8. Stepper motor and feedback 17. Pressure transducer
potentiometer (for 7) 18. Inlet from breathing circuit / manometer
9. Expiratory valve connection

3.3 Pneumatic System The drive gas is filtered with a 40-micron


Input Gas Filter which protects the
pneumatic components from incoming
Refer to the system diagrams above.
particulate matter.
The AV800 Ventilator is designed to operate
on a 38-100 psig drive gas supply (oxygen or The Low Supply Pressure Detector is a
air - to customer’s requirement). pressure switch set at a predetermined level
The gas source is connected to the Drive to detect a loss or reduction of the input gas
Gas Supply DISS fitting on the rear of the source pressure. When the pressure falls
ventilator control unit and should be capable below 38 psig (262 kPa), the LOW SUPPLY
of a flow rate of 75 L/min while maintaining a PRESSURE indicator will illuminate and the
minimum pressure in excess of 38 psig. high priority alarm will activate.

13
DESCRIPTION

The Input Pressure Regulator conditions the spirometry, the actual tidal volume
input drive gas to a stable 35 psig pressure exhaled will be displayed.
which will operate the internal pneumatic
system. The patient must be monitored
independently from the ventilator
The Inspiratory Valve is a large orifice, It is the responsibility of the user to
electro-pneumatically-driven valve which monitor the patient for adequate
supplies the drive gas to the Drive Gas Flow ventilation.
Metering Valve .
A compliance compensation algorithm is
The Drive Gas Flow Metering Valve is a built into the control software which
variable-orifice needle valve which monitors the volume of gas delivered and
determines the drive gas flow rate of the the rate of pressure rise.
bellows during inspiration. It calculates an additional volume to deliver
into the breathing system to compensate for
The Valve Position Feedback Potentiometer the reduced volume delivered to the
and the Flow Control Motor function patient as a result of the increased
together to set a flow rate as required by the breathing system pressure.
front panel controls. As a safety feature, this additional volume is
The flow from the Drive Gas Flow Metering restricted to a maximum of 15% of the set
Valve goes to the Bellows Assembly, via the tidal volume irrespective of the maximum
drive connector, closes the exhalation pressure rise.
diaphragm valve and pushes the bellows
downward. User Selectable Inspiratory Pause
As the bellows moves downwards, the gas A user selectable inspiratory pause is
inside the bellows is forced into the provided which, when activated, holds the
Breathing System. inspiratory phase at the end of inspiration,
for a period of 25% of the inspiratory time
At the end of inspiration the exhaust valve before reverting to the expiratory phase.
opens and allows the drive gas in the top of
the bellows housing to exhaust out through NOTE
the Exhaust Outlet. This function is not available in
As the pressure in the top of the bellows PRESSURE controlled cycle
housing is reduced to zero, the patient
exhales into the breathing system and the
bellows rises. Automatic altitude compensation
Ambient pressure is monitored and the
Compliance Compensation ventilator automatically compensates the
delivered volume according to the local
atmospheric pressure.
WARNING
The AV800 has compliance
Manometer (if fitted)
compensation but the actual tidal volume
The manometer connection is connected
delivered to the patient may be different
to the pressure gauge.
to the ventilation parameters set by the
user due to:
A) an extreme compliance condition,
Patient Gas Pressure Transducer
B) a substantial system leak, or The Patient Gas Pressure Transducer is
C) patient circuit pressure effects. connected to the patient breathing system
In addition, high fresh gas flows will lead via the manometer connection.
to an increased Vt being delivered to the
patient. In VOL CYCLE mode whenever this
Note that on models fitted with pressure exceeds the maximum working

14
DESCRIPTION

pressure as set on the front panel an alarm 3.4 Control Unit


is activated and the following occurs:

1. The HIGH AIRWAY PRESSURE Ventilator parameters:


visual indicator is illuminated (high
priority alarm). VENTILATION MODE
2. The audible alarm is activated (high TIDAL VOLUME
priority alarm). RATE,
3. The Inspiratory Valve is closed, and I:E RATIO,
the inspiration cycle is ended. INSPIRATORY PAUSE, and
Note that the rate of breaths per AIRWAY PRESSURE CONTROL
minute will be affected in this
condition as the ventilator will cease These parameters are set on the front panel
to deliver flow above the maximum by five rotary controls and one momentary
working pressure. action push button.
Based on the control settings, the system:
WARNING 1. Calculates the INSPIRATORY FLOW
If this warning is ignored, the patient may and the INSPIRATORY and
receive an insufficient minute volume. EXPIRATORY times (see section 3.7).
In PRESSURE controlled mode, when the 2. Controls the flow metering valve.
set pressure is reached the following 3. Displays the value of the tidal volume,
occurs: rate, and I:E Ratio parameters on
three digital displays.
1. The Inspiratory valve is closed. 4. Generates the appropriate messages
2. The ventilator maintains the set and alarms.
pressure until the end of the
inspiration cycle.

High Pressure Protection


High pressure in the ventilator is limited by
three independent protective systems.
The pressure transducer has already been
described.
In addition, the Driving Gas Over-pressure
switch is set at 80 cmH2O and will shut off
drive gas flow at this value.
Further, the Pressure Relief Valve is a
mechanical over-pressure relief which will
open at 80 cmH2O, diverting the driving gas
to atmosphere through the PRV exhaust
port.
The Exhaust Outlets on (a) the back of the
control unit, and (b) the base of the control
unit, accept the drive gas exhaust from all
internal pneumatic components.

15
DESCRIPTION

Front panel - Models with Manometer and Spirometry

Front panel - Models without Manometer and Spirometry

16
DESCRIPTION

3.5 Front Panel Operating (in STANDBY mode only):


1. Press the MODE SELECT switch.
Controls 2. Rotate the I:E ratio control fully
clockwise for ON, or anti-clockwise for
A description of each control is given in the OFF.
following paragraphs. 3. Press MODE SELECT to return to
STANDBY.
3.5.1 Power Switch

OFF SPONT (spontaneous)


(mains power connected) Applicable for Vt of ¦200 ml.
All power is removed from the ventilator The indicator SPONT will be illuminated
circuitry (except the battery charger circuit). with a green indicator and the ventilator
will monitor and display the functions
The text display ‘OFF’ will be illuminated with listed below.
a yellow light when the mains electrical No gas will be delivered in this mode.
power is connected and the battery is being
recharged. VENT INOP
HIGH AIRWAY PRESSURE
LOW SUPPLY PRESSURE
STANDBY
LOW AIRWAY PRESSURE
Allows for ventilator function check and set-
(Apnoea)
up.
The text display ‘STANDBY’ will be LOW BATTERY
illuminated with an amber light. MAINS FAILURE

If the ventilator is fitted with the spirometer


1. During the first two seconds in this
option the ventilator will also display the
position, all LED segments and alarm
tidal or minute volume value, when
indicators are lit and the audible
selected.
alarms will activate.
2. After the first two seconds, the
VOLUME CYCLE MODE
following control/display functions are
Ventilator cycling will commence and all
available for set-up without cycling of
alarms will be activated.
the ventilator.
In this mode the VOL CYCLE indicator will
illuminate with a green indicator.
TIDAL VOLUME Maximum breathing system pressure will
RATE be monitored and when maximum
I:E RATIO pressure is reached the HIGH AIRWAY
AIRWAY PRESSURE LIMIT PRESSURE alarm will be activated and the
ventilator will immediately revert to the
3. The following alarms are functional in expiratory phase, irrespective of the function
the STANDBY mode: of the inspiratory pause.
The inspiratory pause function operates in
LOW SUPPLY PRESSURE this mode.
MAINS FAILURE
VENT INOP PRESSURE CONTROL MODE
LOW BATTERY In this mode the PRESSURE indicator will
HIGH AIRWAY PRESSURE illuminate with a green indicator.
INCORRECT RATE/RATIO Ventilator cycling will commence and all
alarms will be activated, except high airway
Spirometry pressure.
Spirometry (if fitted) can be enabled / The ventilator will attempt to deliver the set
disabled by using the following procedure tidal volume until the pressure limit is
reached.
17
DESCRIPTION

At this point the ventilator will stop In addition,high fresh gas flows will lead
delivering gas and pause in this condition to an increased Vt being delivered to the
until the calculated end of inspiration time patient.
has been reached. Note that on models fitted with
The ventilator will then return to the spirometry, the actual tidal volume
exhalation phase. exhaled will be displayed.
The patient must be monitored
The inspiratory pause function does not independently from the ventilator.
operate in this mode. It is the responsibility of the user to
monitor patient ventilation.

3.5.2 Tidal Volume Control


3.5.3 Ventilation Rate Control
This dial controls the requested tidal volume
(LITRES). RATE BPM
Adjustment is only possible when the This dial controls the requested ventilator
ventilator is in volume set mode. rate.
The respiratory RATE is set by a control
Models without spirometry: knob and the system accepts the setting as
The digital display indicates the requested long as the set parameters are within normal
tidal volume. limits
The digital display indicates the requested
Models with spirometry: RATE.
In any of the operating modes, press the
MODE SELECT button to obtain a display of If during adjustment the required flow rate is
the tidal volume (set or measured) and less than 2 L/min or exceeds 75 L/min, the
minute volume (measured). set rate will be limited accordingly.

NOTE
When the ventilator is switched from standby 3.5.4 Ventilator I:E Ratio
to the required operating mode, the
spirometer will take 1 - 2 minutes before WARNING
stabilising The ventilator settings can allow for and
inverse I:E ratio up to 1:0.3. The clinician
The TIDAL VOLUME is set by the control must always ensure that sufficient time is
knob. If the set parameters are within normal allowed for the patient to adequately
limits, the system will deliver the set volume exhale.
indicated.
If during adjustment the required flow rate is I:E RATIO
less than 2 L/min or exceeds 75 L/min, the This dial and digital display controls and
set tidal volume will be limited accordingly. indicates the requested ventilator I:E RATIO.

If during adjustment the required flow rate is


WARNING less than 2 L/min or exceeds 75 L/min, the
The AV800 has compliance set I:E ratio will be limited accordingly.
compensation but the actual tidal volume
delivered to the patient may be different
NOTE
to the ventilation parameters set by the
Minute Volume = Tidal Volume x Rate
user due to:
(see section 3.7)
A) an extreme compliance condition,
B) a substantial system leak, or
C) patient circuit pressure effects.

18
DESCRIPTION

3.5.5 Airway Pressure Limit 3.5.8 Print


(cmH2O) A printed copy of the ventilator conditions for
This dial sets the maximum breathing up to eight hours of the procedure can be
system pressure as sensed by the pressure provided.
transducer via the patient breathing circuit
via the manometer connection. Connect a HPL 2 (or greater) format printer
to the printer output port on the rear of the
In the Volume Cycle mode, the inspiratory ventilator.
cycle is terminated whenever this set At the end of the clinical procedure switch
pressure is achieved. The ventilator then the ventilator to standby, and press the print
reverts to the exhalation phase. push button once.

In the PRESSURE CONTROLLED mode Pressing the print button a second time will
the inspiration is terminated whenever cancel the print function.
this set pressure is achieved and the Turn the ventilator to OFF, and switch the
ventilator pauses until the end of the ventilator back on again to clear the stored
calculated inspiratory time before reverting print information.
to the exhalation phase. The printer port must only be connected to
devices that comply to EN 60950.

3.5.6 Alarm Mute


When the alarm mute push button is
depressed during an alarm condition the
audible alarm can be muted for the
following periods:

Low Airway pressure 30 seconds


Low Tidal Volume 30 seconds
Incorrect rate or ratio 120 seconds
Low battery 120 seconds only

No other alarm conditions are silenceable.


Mutable alarms will be immediately
reinstated upon fresh alarm conditions.

3.5.7 Inspiratory Pause


In volume ventilation mode depressing the
Insp Pause button will allow a 25%
inspiratory pause to be included at the end
of the inspiratory phase.
The I:E ratio will be maintained.
To compensate for the reduced inspiratory
flow time period, the inspiratory flow rate is
increased (up to a maximum flow rate of 75
litres per minute).

The inspiratory pause does not function in


Pressure Ventilation mode.

19
DESCRIPTION

3.6 Rear Panel

3.6.1 Labelling Terminology


The terms Class 1, and Type B are
defined in IEC 601-1 (the standard for
electrical medical equipment).

NOTE
This symbol denotes
Type B equipment.

Type B equipment calls for a particular


degree of protection against electric shock.

Class 1 equipment has additional protection


such that metal parts of the unit that are
accessible to the user (e.g. the metal casing
of the AV800 control unit) cannot become
live in the event of failure of the basic
insulation of the electrical components within
the unit.
Spirometer
connectors

DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE
USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS
! COVER. REFER TO QUALIFIED SERVICE PERSONNEL.

RS232 PORT PRINTER PORT ANALOGUE/ALARM OUTPUTS

SPIRO

EXHAUST VALVE MAINS SUPPLY


DO NOT BLOCK
OXYGEN/AIR Drive Gas
38 - 100 psi
2.6 - 6.9 Bar
262 - 689 kPa

FUSE RATING/TYPE
MANOMETER/ PRV 80 cmH2O 3.15 A 240 V ANTI SURGE HRC
PRESSURE Penlon Limited
Abingdon, Oxon
TRANSDUCER
-20 To 100 cmH2O
OX14 3PH CLASS 1 TYPE B
Tel 01235 547000
Fax 01235 547031
VOLTAGE 90 - 264 VAC
Tested by: Serial No. FREQUENCY 47 - 63 Hz
MADE IN UK MAX CURRENT 1 A

Ventilator Rear Panel


20
DESCRIPTION

3.6.2 Mains supply 3.6.9 Printer Port


The mains supply inlet is designed for This standard Centronics 36 way parallel
connection to any mains voltage from 90 to port is configured to output to any HPL 2 or
264 VAC and a frequency of 47 to 63 Hz greater compatible parallel printer.
without any adjustment. The port must only be connected to devices
The connector is a standard IEC type that comply to EN 60950.
dismountable connector.

3.6.10 Spirometer connections


3.6.3 Oxygen/Air Male and female Luer lock connections are
Ventilator driving gas is attached to this provided for the connection of the side
connector at a pressure of 38 - 100 psi stream spirometer system. Correct polarity
(262 - 689 kPa) and capable of supplying must be observed to ensure that the correct
gas at a flow rate of not less than 80 L/min flow is measured.
at the minimum pressure. Air or Oxygen To maintain the accuracy of the flow
may be used as the drive gas but the supply measurement only the specified sample
must be clean and dry. lines may be used.
Note that the drive gas is specified by the
customer. To change the drive gas, refer to a
Penlon-trained service engineer.

3.6.4 Drive Gas


The drive gas outlet is a special 17 mm
diameter taper connector and delivers the
A
drive gas from the control unit to the
ventilator bellows assembly.

3.6.5 Exhaust Valve


Exhaust bellows drive gas, either air or
oxygen, returned from the bellows assembly
is exhausted through this outlet. The port
must not be blocked.
Latest type tubing assemblies have a
restrictor fitted to the tube fitted to the lower
3.6.6 Manometer / Pressure connector on the ventilator rear panel.
Transducer This tube is labelled ‘VENT END’ (A) to
The breathing system is connected to this assist correct fitment.
port with a 6 mm flexible sample tube.
CAUTION
A) The lines must not be trapped during use.
3.6.7 PRV 80 cmH2O B) Do not alter the length of the sample lines.
The drive gas over pressure relief valve is
set at a non-adjustable 80 cmH2O.
The PRV outlet port must not be blocked. 3.6.11 Analogue/alarm Outputs
This port may be configured to interface
with a medical information bus system by
3.6.8 RS 232 Port special order.
Configuration details for this port are The port must only be connected to devices
available by special order from Penlon that comply to EN 60950. See also section
Limited.
10.
The port must only be connected to devices
that comply to EN 60950.

21
DESCRIPTION

3.7 Message Displays HIGH CONT (Continuing) PRESSURE


(High priority)
A visual indicator and audible alarm which
3.7.1 Alarm Mute activate when the pressure sensed at the
For those alarms which can be silenced, the patient tee exceeds 30 cmH2O just prior to
mute button will provide 30 seconds of the next inspiratory cycle.
muting for high priority alarms and 120 Once activated, this alarm remains on until
seconds of muting for medium priority the pressure falls below the required
alarms. baseline at the beginning of a breath.
This alarm is not silenceable.

3.7.2 Alarm Indicators LOW AIRWAY PRESSURE


Visual indicators are backlit, no warnings are (High priority)
visible when the power is turned off. (Disconnect Alarm)
All alarms are self-cancelling (with a A visual indicator and an audible alarm
minimum activation period of 2 seconds), which activate if the pressure sensed at the
except VENT INOP (ventilator inoperative). patient tee-piece in the expiratory limb of
Alarms comply with EN475. the breathing circuit does not increase by at
NOTE least 4-14 cmH2O during a ventilator
There are no means provided to defeat any inspiratory cycle, depending on tidal volume
visual alarm. setting.
Alarms are priority configured. The highest Once this alarm is activated, it remains on
priority alarm will always take precedence. until the required pressure difference is
reached.
Normal conditions must be restored or the This alarm can be silenced for 30 seconds.
unit turned off in order to cancel the alarms.
LOW TIDAL VOLUME
LOW DRIVE GAS SUPPLY (High priority)
(High priority) The AV800 can be fitted with an external
A visual indicator and audible alarm which volume spirometer which provides signals to
activate if the input drive gas pressure has the volume display. If the apparent tidal
dropped below 38 psig (262 kPa). volume as measured by the spirometer falls
This alarm is not silenceable. below 50% of the set tidal volume, a visual
indicator and audible alarm are activated.
HIGH AIRWAY PRESSURE This alarm can be silenced for 30 seconds.
(High priority)
Standby, Spontaneous, and Volume
Spirometer Disconnect
Cycle Mode
If the spirometer tube becomes
A visual indicator and audible alarm which
disconnected, causing an erroneous
activate when the pressure sensed at the
measured volume greater than 50% of the
patient tee exceeds the setting of the
set tidal volume, the Tv display will flash.
AIRWAY PRESSURE LIMIT control.
This alarm will remain on until the pressure INCORRECT RATE OR RATIO
falls below the control setting for a minimum (Low priority)
of 1 second. A visual indicator and audible alarm which
This alarm will also activate if the drive gas indicates that the required inspiratory flow
supply overpressure switch operates (and rate is below 2 L/min, or has reached the
also operates in Pressure control mode). upper limit of 75 L/min as determined by the
This alarm is not silenceable. settings for TIDAL VOLUME, RATE , I.E.
RATIO, and INSP PAUSE.
This alarm can be silenced for 120 seconds.

22
DESCRIPTION

MAINS FAILURE 3.8 Back-up Battery


(Low priority)
A visual indicator and an audible alarm (low In the event of mains electrical failure, the
priority) that is battery driven. back-up battery cuts in automatically.
The alarm activates when electrical power is A fully charged battery will power the
disconnected when the ventilator is ventilator for 60 minutes.
operating. See Appendix 1 for battery care procedures.
Note that if fully charged, the back-up battery
will power the ventilator for 60 minutes.

LOW BATTERY
(Low priority / Medium priority)
A visual indicator and an audible alarm
which activate if the internal battery is almost
exhausted, indicating that ventilation will
soon cease.

When approximately ten minutes of battery


backup time remains, the alarm will change
from low to medium priority.
To prevent damage to the battery, the
ventilator will shut down before the battery is
fully discharged.

VENT INOP
(High priority)
The VENT INOP (ventilator inoperative)
alarm indicates that one of the following
conditions has occurred:

A) A solenoid has failed,


B) The flow control valve has failed,
C) Internal electrical fault,
D) Internal Communications error,
E) Battery disconnected,
F) Software error

In the event of a ventilator error being


detected an error code will be displayed in
the RATE / BPM window and additionally will
be stored in a non volatile memory which
can be accessed by the service engineer
to assist with fault finding subsequently.

WARNING
The only way to reset a VENT INOP alarm
is to turn the Power switch OFF for a
minimum of one second and then back
ON.
If the VENT INOP alarm occurs again,
remove the ventilator from use and refer
to an authorised service technician.

23
DESCRIPTION

3.9 Operational Capability

I:E Ratio
1:6 1:5 1:4 1:3 1:2 1:1 1:0.3

1.6
1.5
1.4
1.3
1.2
1.1
1.0
Tidal
Volume 0.9

(litres) 0.8
(Vt) 0.7 X
0.6

0.5
0.4

0.3
0.2
0.1

0
0 5 10 15 20 25 30 35 40 45 50 55 60

Rate (bpm)

The ventilator is capable of operating at the volumes and rates below each I:E ratio curve.

Note a) The minimum tidal volume setting is 50 ml.


b) The minimum rate setting is 4 bpm.

Example
1. Select required volume (Vt) (e.g. 0.7 litres)
2. Select rate (e.g. 10 bpm).
3. Select I:E ratio of 1:2.

The point X on the graph lies beneath the 1:2 ratio curve, and is therefore within the
ventilator’s capability.

24
DESCRIPTION

3.10 System Software

The software controls all ventilator functions


including performance accuracy and safety
checks.

In addition, the software controls the


ventilator/user interface, apart from the
manometer.

The software is held in 256 kbytes of Flash


EPROM, and is copied into RAM at power-up
and runs there.

Upgrades
Any upgrades will be offered to customers for
downloading via a connector on the main PCB.

Digital to Analogue Converter


Provides analogue output of the following
parameters:
Breathing circuit pressure
Average breathing circuit pressure
Airway pressure limit
Tidal volume
BPM
I:E ratio
Measured volume

These are all generated using an 8 bit, 8 channel


D/A converter (TLC5628). This is driven using
synchronous serial communications using three
I/0 bits on the processor.

15 pin D-type Description Range Analogue


connector pin scaling
layout

1 Ground
2 Breathing Circuit Pressure -20 to 80 cmH20 0.05 V/cmH20
3 Average Breathing Circuit Pressure -20 to 80 cmH20 0.05 V/cmH20
4 Airway Pressure Limit 10 to 70 cmH20 0.05 V/cmH2O
5 Tidal Volume 0-1.6 litres 0.3 V/100ml

6 BPM 0-60 bpm 0.08 V/bpm


7 I:E Ratio 1:0.3 to 1:6.0 0.08 V/O.IE
8 Measured Volume 0-1.6 litres 0.3 V/100ml

25
DESCRIPTION

Alarm Outputs

Alarm outputs are provided on the system


for various system parameters
Alarm outputs are on pin 9-15.

15 pin D-type Alarm Outputs Logic level


connector pin
layout

9 System OK High
10 Power problem - mains fail, Low
battery missing or battery low
11 Incorrect rate/ratio Low
12 Low measured tidal volume Low
13 Low supply pressure Low

14 Low airway pressure Low


15 High airway pressure / high Low
continuous pressure

Real Time Clock


Provides a time stamp for the error log and any
printouts.
The clock is set at the factory to GMT (Greenwich Mean
Time).

Engineering Mode
A facility to allow Penlon-trained service engineers
access to the following functions:

1. Software revision information - indicates the


versions currently installed .
2. Error log - displays a log of the last 40 events
which caused a system operational error.
Note that after power-up, only the most recent
operational error is displayed.
3. Set date and time on the real time clock.
4. Abort to the FORTH interpreter to allow
communication to the unit through the serial port,
for possible software debug.

To gain access to Engineering Mode, switch the power


control to STANDBY, and then press the INSP PAUSE
and MODE SELECT buttons simultaneously.

Rotate the Tidal Volume control knob clockwise / anti-


clockwise to move between the above functions.
Press the INSP PAUSE button to exit Engineering
Mode.
26
4. SPECIFICATION

4.1 Application Ventilation for use in anaesthesia.

4.2 Internal Compliance 3 ml/cmH2O (nominal)

4.3 Minute Volume Range This parameter cannot be set on the


machine, but can be calculated if required.
(see section 3.5.4).

4.4 Tidal Volume Range 50 to 1600 ml (±10%)

4.5 Frequency (Rate) Range 4 to 60 bpm, limited by tidal volume setting

4.6 Inspiratory Phase Time Range 0.14 to 11.54 seconds

4.7 Expiratory Phase Time Range 0.23 to 12.9 seconds

4.8 Inspiratory/Expiratory Phase 1: 0.3 to 1: 6.0 - limited by tidal


Time Ratio Range volume and rate settings
(I:E ratio)

4.9 Pressure Control Range 10 to 70 cmH2O (±10%)


(pressure ventilation mode)

4.10 Inspiratory Flow Range 2 to 75 L/ min

4.11 Airway Pressure Limit 10 to 70 cmH2O (±10%)


(volume ventilation mode)

4.12 Inspiratory Triggering No trigger

4.13 Inspiratory Triggering No trigger


- Response Time

4.14 Maximum Safety Pressure 80 cmH2O

4.15 Maximum Working Pressure 70 cm H2O

4.16 Minimum Safety Pressure -10 cm H2O

4.17 Minimum Working Pressure Atmospheric

4.18 Sub-atmospheric Pressure None


Range

4.19 Expiratory Resistance 5 cmH2O/Ls-1


(60 L/min)

4.20 Sigh Characteristics None

4.21 Inspiratory Mixture No mixture controls

27
SPECIFICATION

4.22 Flowmeters None

4.23 Manual Changeover None

4.24 Inspiratory Relief Valve Fixed, 80 cmH2O

4.25 Humidifiers and Monitors No humidifier.


Volume monitor optional (spirometer)
Pressure monitor built in

4.26 System Pressure Gauge None

4.27 Airway Pressure Gauge Optional

4.28 Spirometer Optional


Accuracy Varies with gas composition.
Note that the spirometer is calibrated for
typical clinical gas mixtures. Differing gas
composition will cause variances from actual
values. Maximum variation is ±20%.
Range 300 ml Vt minimum in Spontaneous mode.
50 to 1600 ml in all other operating modes.

4.29 Alarms (in priority order) See section 3 for detailed descriptions.

Vent. inop. (ventilator inoperative)


High airway pressure
High cont. (continuing) airway pressure
(PEEP)
Low supply pressure
Low airway pressure
Low tidal volume
Incorrect rate or ratio
Low battery
Mains failure

4.30 Power Source


Electrical 90 to 264 VAC
47/63 Hz, universal input.
Input Drive Gas Oxygen or air (dry, and oil free) at 38 to 100
psig (262 to 689 kPa).

4.31 Power Consumption


Electrical 37 watts maximum
Gas 75 L/min intermittent maximum inspiratory
flow.
37.5 L/min minute volume
4.32 Dimensions
Height 370 mm
Height of control unit only 135 mm
Width 235 mm
Depth 260 mm

28
SPECIFICATION

4.33 Weight
Weight with adult bellows 9.0 kg
Weight with paediatric bellows 8.7 kg
Weight of control unit only 7.6 kg

4.34 Method of Disinfection Bellows base assembly and inside of


or Sterilisation bellows require sterilisation - section 7.3

4.35 Bacterial Filter None (see section 5.1.4 for


recommendations for breathing system)

4.36 Fail Safe Mechanism Battery back-up in case of mains electricity


failure

4.37 Reliability MTBF: 5x106 to 50x106 cycles

4.38 Waveform Tests Not applicable

4.39 Volume Tests Not applicable

4.40 (A) Mobility Secure mounting on anaesthesia machine


required.
(B) Mounting Mounting bracket available as optional extra.
Bellows assembly can be separated from the
control unit for remote mounting.

4.41 Fuses 220/240 V: two fuses, 3.15 A, 20 mm, anti


surge

4.42 Environmental
Ambient Temperature
Storage -5 to 50oC (23 to 122oF).
Refer to Appendix 1 for battery care during
storage.
Operating 10 to 38oC (50 to 100oF)

Humidity 10-95% RH (relative humidity), non-


condensing

Altitude Up to 2775 m (9000 feet)

Ingress protection Conforms to EN 794-1 Clause 44 (spillage


test)

29
5. PRE-OPERATION PROCEDURES

5.1 Set-up WARNING


Excessive electronic noise caused by
other, poorly regulated devices, such as
5.1.1 Components supplied with electrocautery, may adversely interfere
the ventilator with the proper functioning of the
Check that the ventilator has been supplied ventilator.
with: To avoid this problem, do not connect the
(a) A 210 mm long, 16 mm diameter, ventilator power cord into the same
corrugated hose to connect the electrical wall outlet or strip into which an
control unit to the bellows assembly electrocautery unit is connected.
(b) Drive gas hose (to customer
specification)
(c) Electrical mains supply cable (to 5.1.4 Ventilator gas connections
customer specification). Before the ventilator can be used, the
(d) Pressure sampling line and T-piece. following preparation must be made:
(e) Spirometer sampling line and T-piece
(supplied with spirometer option 1. Verify the drive gas specified for the
machines only) ventilator (oxygen or air) and connect
the drive gas inlet port on the rear of
the control unit to a dry, oil free supply
5.1.2 Mounting the ventilator at 38 to 100 psi (2.6-6.9 bar, 262-689
The AV800 can be mounted on the kPa)
anaesthetic machine as a single, complete To change the drive gas, refer to a
unit. Penlon-trained engineer.
The drive gas inlet port is labelled:
Locate the ventilator in a safe place. OXYGEN / AIR 38 -100 PSI
Preferably, mount it permanently on the
anaesthesia shelf of the machine or on a OXYGEN SUPPLY - O2 cylinder,
strong bracket to protect the ventilator from anaesthetic machine O2 auxiliary gas
accidental fall and accidental disconnection outlet, or O2 pipeline supply from a
of the hoses. wall outlet.
To mount the ventilator permanently on a
Penlon bracket: AIR SUPPLY - Air cylinder,
1. Align the four mounting feet over the anaesthetic machine air auxiliary gas
mating holes in the bracket. outlet or air pipeline supply from a wall
2. Use four M4 screws inserted through outlet.
the bracket and rubber feet and
screwed into into the threaded inserts 2. Connect the 16 mm diameter
in the base of the ventilator. corrugated hose (provided), between
the control unit drive gas outlet
Pole-mount type mounting brackets are (labelled: BELLOWS DRIVING GAS)
available, see section 8. and the bellows base DRIVING GAS
port.

5.1.3 Ventilator electrical 3. Connect the bellows base EXHAUST


port, with a 30 mm hose, to a properly
connections
functioning scavenging system. The
Before connecting the ventilator to the mains
diaphragm valve under the bellows is
supply, check that the power supply is within
connected internally to the EXHAUST
the correct rating as stated on the label on
port to facilitate the discharge of
the rear of the control unit.
excess breathing gas at the end the
expiratory phase.

30
Driving Gas
Port

Exhaust Port
(patient gas)
Do NOT connect
spirometer
Breathing
System
Port

Spirometer
connectors

DANGER - POSSIBLE EXPLOSION HAZARD. DO NOT CAUTION - ELECTRIC SHOCK HAZARD. DO NOT REMOVE
USE IN THE PRESENCE OF FLAMMABLE ANAESTHETICS ! COVER. REFER TO QUALIFIED SERVICE PERSONNEL.

RS232 PORT PRINTER PORT ANALOGUE/ALARM OUTPUTS

SPIRO

EXHAUST VALVE MAINS SUPPLY


OXYGEN/AIR DO NOT BLOCK
38 - 100 psi Drive Gas
2.6 - 6.9 Bar
262 - 689 kPa

MANOMETER/ PRV 80 cmH2O


PRESSURE
FUSE RATING/TYPE
TRANSDUCER Penlon Limited
3.15 A 240 V ANTI SURGE HRC
-20 To 100 cmH2O Abingdon, Oxon
OX14 3PH
Tel 01235 547000 CLASS 1 TYPE B
Fax 01235 547031 VOLTAGE 90 - 264 VAC
Tested by: Serial No.
FREQUENCY 47 - 63 Hz
MADE IN UK
MAX CURRENT 1 A

Ventilator Rear Panel and Gas Ports

31
PRE-OPERATION PROCEDURES

AV800 VENTILATOR
(REAR VIEW)

EXHAUST TO
SCAVENGE AV800 ventilator connected as part of a
SYSTEM
circle breathing system

BREATHING CIRCUIT
BACTERIAL FILTER
PRESSURE MONITOR LINE

SPIROMETER

FRESH GAS SUPPLY


(FROM MACHINE CGO)

HEAT AND MOISTURE ANAESTHETIC MACHINE


EXCHANGER (HME)
CIRCLE
SYSTEM
VENTILATOR
ABSORBER
DRIVE GAS
(FROM MACHINE
AUXILIARY OUTLET)
PATIENT

4. Connect the spirometer (if fitted) - see expiratory limb to protect the
Section 5.1.5. breathing system.

WARNING WARNING
Applying negative or positive pressure to The High and Low Airway Pressure
the bellows exhaust port results in Alarms are important for patient care.
positive pressure in the patient breathing The ventilator is designed to be used with
system. a distal sensing tee only. (Catalogue No.
Therefore, the scavenging system must 53194, Breathing System Tee Assembly.)
not generate more than 0.5 cmH2O It is important that the distal sensing tee
positive or negative pressure when is properly located in the inspiratory limb
connected to the ventilator. of the breathing system.
Any problem arising from an improperly 6. Set the AIRWAY PRESSURE LIMIT
functioning scavenging system is solely control to 50 cmH2O.
the user`s responsibility.
7. Attach a printer to the printer port if a
5. Connect the MANOMETER/ printed output of the ventilator
PRESSURE TRANSDUCER port on function is required.
the rear panel of the control unit to the
breathing system. 8. Connect the ventilator bellows base
The recommended placement for the BREATHING SYSTEM port to the
distal sensing tee is in the inspiratory breathing system.
limb of the breathing system close to Use a breathing system bacterial filter
the circle system inspiratory valve. in the expiratory limb of the breathing
Use a Heat and moisture exchanger circuit or a heat and moisture
(HME) at the patient Y piece, or a exchanger (HME) at the patient Y
breathing system filter in the piece. See illustration above.
32
PRE-OPERATION PROCEDURES

9. Connect a 2-litre breathing bag to the


patient connection as a test lung.

10. Close the anaesthetic machine APL or


PRV valve in the breathing system.

11. Turn the ventilator POWER switch to


STANDBY and :
set the tidal VOLUME to 800 ml,
set RATE to 10 bpm,
set I:E RATIO to 1:2.0.

12. Use the O2 flush button on the


anaesthetic machine to fill the
bellows.

13. Turn the ventilator POWER switch to


VOL CYCLE.

14. The delivered tidal volume read on the


scale of the bellows housing should
be 800 ml.

Before using the ventilator clinically:


a) Check that all connections are
correct, and verify that there are no
leaks.
b) Carry out the function checks
listed in section 5.3.1.

33
PRE-OPERATION PROCEDURES

A1
B2
B1 D
C1

C2
Spirometer
E
Head

Pressure
A2
Monitor Line
Notes
a) Carefully twist each sample
line anti-clockwise before
connecting to the ventilator and
5.1.5 Spirometer Connections Tee-piece.
WARNING The lumen assembly must be Turn each connector fully
clockwise to ensure a leak-free
connected as illustrated. Do not connect to the
joint.
expiratory limb of the breathing circuit. b) Latest type tubing assemblies
have a restrictor fitted to tube
The triple lumen assembly incorporates the C1/C2.
ventilator pressure monitor line (shown as This tube is labelled ‘VENT END’
Tube A1/A2). (E) to assist correct fitment.
This line must be connected to the breathing
circuit at the absorber expiratory hose
connector.

1. Use a breathing system bacterial filter 4. Check for secure connection at each
in the expiratory limb, or a heat and connector.
moisture exchanger (HME) at the NOTE
patient Y piece. See section 5.1.4, A) If the connection is incorrectly made at the
operation 8. spirometer sensor, the ventilator will alarm LOW
TIDAL VOLUME. To allow the ventilator to be
used in the event of damage or or non-functioning
2. Fit the spirometer head to the Safelock of the spirometer head, turn off the spirometry
connector at the ventilator port on the function - see STANDBY mode.
back of the Circle Absorber. B) Connection of the spirometer into the exhaled
limb also provides an indication of breathing
3. Connect the tubing from the spirometer system disconnect.
head to the connectors at the rear of the
C) When testing the spirometer prior to clinical
AV800 ventilator as shown. use, it is preferable to use a 70% N2O / 30% O2
For additional identification: mixture.
i) The centre tube (C1/C2) has a RED If 100% oxygen is used, the Measured Vt value
display may flash. This is not significant, and
connector at C2 which must be fitted to should be disregarded.
the lower connector on the ventilator
(fitted with a RED washer ,D). Absorber switched to Bag mode
Later assemblies have a label (E) (VENT NOTE If the absorber bag/ventilator switch
END) on tube C1/C2 is set to ‘Bag’, the spirometer will not
ii) Connector B1 is BLUE. measure tidal volume.
34
PRE-OPERATION PROCEDURES

5.2 Bellows Assembly 3. Fit the bellows, and then fit the
bellows housing to the base by
CAUTION pushing it down, then twisting it
Always ensure correct fitment of bellows slightly clockwise until the bayonet
(see illustration above), and carry out a full tabs completely engage.
function test before clinical use, if a bellows 4. Function test the ventilator - section
is removed and refitted. 5.3.1.

1. Remove the bellows housing (1) by NOTE


twisting it slightly counterclockwise If there is any malfunction, the ventilator
until the bayonet tabs become free, must NOT be used.
then lift it up from the base (2).
2. Remove the bellows (3) from the If the problem cannot be located, refer to
base. section 6 - Fault Finding, or return the
ventilator for repair to a Penlon accredited
agent, or to the manufacturer.

35
PRE-OPERATION PROCEDURES

5.3 Pre-use Checklist 3. Connect the ventilator bellows base


BREATHING SYSTEM port to the
5.3.1 Daily Checklist breathing system.

The following tests should be done at least 4. Connect a 2-litre breathing bag to the
at the beginning of every working day: patient connection as a test lung.

WARNING 5. Close the anaesthetic machine APL or


The operation of each alarm function PRV valve in the breathing system,
should be verified daily. Periodically
check the alarms at suitable intervals. 6. Set the tidal VOLUME to 800 ml;
If the audible alarm or the visual indicator RATE to 10 bpm, and I:E RATIO to
of any alarm function fails to activate 1:2.0.
during any alarm condition or fails to
reset after the alarm has been cleared, 7. Use the O2 flush button on the
refer the unit to an authorised service anaesthetic machine to fill the
technician. bellows.

WARNING 8. Turn the ventilator MODE switch to


If the internal battery is fully discharged, VOL/CYCLE.
the ventilator will not function.
Recharge the battery before the ventilator is 9. The delivered tidal volume read on
used clinically. the scale of the bellows housing
Charging the battery for 14 hours from a should be 800 ml.
discharged state will allow a minimum of 45
minutes of continuous operation. Connect If the delivered tidal volume is less
the ventilator to a mains power supply. The than 700 or greater than 900 ml, refer
OFF indicator will turn yellow when the to a Penlon trained engineer.
battery is being charged (it is not necessary
to run the ventilator). 10. Fill the bellows using the O2 flush
again.
Ventilator internal test
Turn the POWER switch to STANDBY. 11. Occlude the patient connection port of
A two-second internal test is initiated. the breathing system, distal end.
The following must occur if everything
performs correctly: 12. The HIGH AIRWAY PRESSURE
alarm should be activated. The peak
1. All three digital displays must read 8. pressure read on the breathing
2 All ALARMS messages must be lit. system pressure gauge is the
3 The audible alarm drivers must sound maximum working pressure limit and
- one high tone, and one low tone. should agree with the setting.

Functional checkout: 13. Open the patient connection port to


1. Set the AIRWAY PRESSURE LIMIT ambient pressure. At the second
control to 50 cmH2O. cycle, the LOW AIRWAY PRESSURE
alarm should be activated.
2. Check that the MANOMETER /
PRESSURE TRANSDUCER port on
the rear of the control unit is correctly 14. Turn the ventilator MODE switch to
connected to the inspiratory limb of STANDBY.
the breathing circuit, close to the Before using the ventilator clinically,
patient tee piece (see operation 5, check that all connections are correct,
section 5.1.4). and verify that there are no leaks.

36
PRE-OPERATION PROCEDURES

5.3.2 Weekly Checklist

At least every week, in addition to the daily


functional check:

1. Turn the ventilator MODE switch to


STANDBY.

2. Unplug the mains power cable from


the AC outlet.

The MAINS FAILURE alarm should


activate.

3. Reconnect the mains power cable to


the AC outlet. The alarm should turn
off.

4. Turn the MODE switch to


VOL/CYCLE.

5. Disconnect the drive gas supply hose.


The LOW SUPPLY PRESSURE
alarm should activate.

NOTE
If there is any malfunction, the ventilator
must NOT be used.

If the problem cannot be located, refer to


section 6 - Fault Finding, or return the
ventilator for repair to a Penlon authorised
service centre or to the manufacturer.

37
6. FAULT FINDING

CONDITION POSSIBLE CAUSE / REMEDY

No power at switch on 1. Check mains supply/plug.


(no VENT INOP alarm)
2. Check yellow OFF LED (mains indicator).

3. Check fuses - mains input socket.

4. Check voltage from power supply unit - 14.3 V.

5. Replace ON/OFF switch.

LOW BATTERY message is on 1. Charge battery by plugging unit into the mains
supply.

2. Defective battery - replace.

3. Mains fuse blown.

Bellows does not fill, or, collapses. 1. No fresh gas flow.

2. Breathing system hose leak or hose disconnected.

3. Broken bellows base.

4. Defective or detatched bellows.

5. Damaged exhalation diaphragm valve.

6. Missing or defective O-ring seals.

Bellows progressively becomes less 1. Check breathing system for leaks.


full despite usual fresh gas flow setting.
2. Check and close anaesthesia system APL or PRV
valve.

3. Check exhalation diaphragm valve seat under the


bellows for damage.

4. Damaged O-ring on bellows base - replace.

LOW BREATHING SYSTEM PRESSURE 1. Check pressure pick-up tubing.


alarm is on but ventilation is normal.
2. Look for disconnect or kink.

3. Control unit internal failure

LOW DRIVE GAS SUPPLY alarm activated 1. Check drive gas supply pressure (> 38 psi)

38
FAULT FINDING

CONDITION POSSIBLE CAUSE / REMEDY

Displayed values incorrect 1. Check with stopwatch / Vt measurement.

2. Check that the inlet filter is not blocked.

3. Replace/reset pressure switch.

Continuous high airway pressure limit 1. Potentiometer set too low (or knob out of
alarm calibraion) - adjust.

2. Error code 12
Potentiometer failure - replace display assembly
PCB.

3. Error code 8
Potentiometer sensor failure - replace main PCB.

Unwanted PEEP and bulging, full bellows 1. Defective or poorly regulated scavenging system,
creating too high resistance or too much vacuum.

2. Exhaust partially obstructed.

3. Bellows diaphragm defective.

Pressure limit fails to operate 1. Error code 8


Pressure sensor failure - replace main PCB.

2. Internal tube disconnected - reconnect and secure.

Stacking breaths (inadvertent PEEP) 1. Error code 3


Expiratory valve solenoid failed to open, resulting
in high airway pressure alarm.- replace.

Continuous flow of drive gas, or, bellows 1. Error code 2


driven to bottom. Faulty inspiratory solenoid valve - replace.

Tidal volume not delivered and 1. Check/increase AIRWAY PRESSURE LIMIT


HIGH AIRWAY PRESSURE alarm is on. setting.

2. Check breathing system for obstruction.

Zero tidal volume. 1. Error code 2


Inspiratory solenoid valve faulty - replace.

39
FAULT FINDING

CONDITION POSSIBLE CAUSE / REMEDY

Zero tidal volume and drive gas leak from 1. Error code 3
exhaust port. Expiratory solenoid failed open - replace.

2. Exhaust valve failed - replace.

Low tidal volume 1. Error code 2


Inspiratory solenoid valve faulty - replace.

2. Pressure relief valve leaking - replace O ring.

3. Inlet filter blocked - replace.

4. Defective spirometry head or leak in spirometry


system.

5. Breathing system leak or hose disconnected.

6. Exhaust valve leaking - replace O ring / silicon


washer.

High tidal volume 1. Leaking spirometer connector.

2. Incorrect calibration of drive gas flow - recalibrate


needle valve / stepper motor position.

Tidal volume incorrect or inconsistent 1. Missing small O-ring under the exhalation
diaphragm valve.

2. Partially detached or defective bellows.

3. Damaged exhaust valve seal or seat.

4. Very high/low breathing system resistance or


compliance.

5. High/low fresh gas flow - check volume delivered


using standard calibration method.

6. Incorrectly regulated gas pressure - check


pressure / replace regulator.

7. Error code 1
Flow valve fails to adjust correctly due to
mechanism binding - realign and check calibration.

8. Component failure on main PCB - replace PCB.

40
FAULT FINDING

CONDITION POSSIBLE CAUSE / REMEDY

Tidal volume incorrect or inconsistent 9. Spirometer faulty - replace spirometer head and
(continued) lead.

10. Loose grub screws on motor drive / needle shaft /


potentiometer drive shaft.
- check security of grub screws.
- return ventilator to accredited agent or
manufacture for calibration.

11. Broken or damaged potentiometer flexible coupling


- return ventilator to accredited agent or
manufacture for calibration.

12. Worn needle valve - return ventilator to accredited


agent or manufacture for calibration.

41
FAULT FINDING

42
7. SERVICE SCHEDULE

The AV800 ventilator must be serviced to the following schedules.


Only Penlon-trained engineers should undertake servicing and repairs..

6 Month Service

1. Full service performance check.

12 Month Service

1. Fit Preventive Maintenance Kit components :

AV800 12 month service Kit Part No. 57633

Part No. 042819 Quad Seal ( Exhaust valve )


Part No. 0314 ‘O’ Seal Viton ( PRV )
Part No. 041204 ‘O’ Seal Viton ( Bellows base diaphragm valve )
Part No. 041226 ‘O’ Seal Viton ( Bellows Base to housing )
Part No. 300045 Silicon Sealing Washer ( Exhaust valve )
Part No. 0762 Filter - W9151 - 40 Micron.
Part No. 15446 Warranty sticker
Part No. 300329 Spirometer tubing assembly

2. Full service performance check.

5 year Overhaul
1. Fit overhaul kit components :

5 year overhaul kit Part No. 57635


Parts listed in 12 month kit, plus:
Part No. 104019 Lithium Battery ( Real time clock )
Part No. 103996 Lead Acid Battery ( Mains back up )
Part No. 045438 Internal Regulator
Part No. 300116 Internal Tubing Set
Part No. 011052 Adhesive tape (battery and speaker)

2. Full service performance check.

Software Upgrades

Software upgrades will be offered to distributors and customers by means of a software


program which will be loaded via the main PC.104 connector using a specially developed
ROM Card
For distributors and customers who do not have access to a computer, a PCB board and front
panel pic chip replacement will be available for early machines.

Special Tools
Part No. Description
------ Multimeter
300128 PRV and Exhaust - cap removal tool
------ Test Gauge

43
SERVICE SCHEDULE

AV800 SERVICE PROCEDURE (version 2.0 software)

Item Operation
Service frequency
1 Initial Checks
1.1 Check and record serial number and determine service required.
1.2 Check general condition of ventilator assembly Six months
1.3 Check configuration of attachments and note. (ALL item 1
1.4 Check and record front panel control settings. operations)

2 Engineering mode test


2.1 Switch ventilator selector switch to ‘OFF’. Six months
Press MODE SELECT and INSP PAUSE simultaneously, and (ALL item 2
hold. operations)
Switch Ventilator selector switch to ‘STANDBY’.
The ventilator is now in Engineering Mode.
NOTE To set a non spirometer version in engineering mode
press the panel where the ‘MODE SELECT’ switch is hidden from
view.

2.2 Check date and time


Turn the I:E ratio knob to move between parameters.
Turn the RATE knob to adjust, if necessary,
If any adjustments have been made, press ‘MODE SELECT’ at the
end of the procedure (i.e., in Check Time - Minute mode) to save
new settings.
Turn the I:E ratio knob to move between parameters.

2.3 Verify correct revision level.

2.4 Note all errors and investigate.


Error codes :-
0 - Watchdog reset.
1 - Stepper position error.
2 - Inspiratory solenoid fail.
3 - Expiratory solenoid fail.
4 - Front panel comms fail.
5 - Wild reset detected.
6 - Stepper feedback pot fail.
7 - System power fail.
8 - Airway pressure transducer fail.
9 - Ambient pressure transducer fail.
10 - Software Exception error - Parameter gives the exception
number.
11 - Stack underflow or overflow.
12 - Front panel pressure limit pot fail.
13 - Flash checksum changed.

2.5 Press MODE SELECT to clear errors,


Press INSP PAUSE to exit engineering mode.

44
Check Date - Check Date -
Year Month
Year shown - 1998 Month shown -
Turn RATE knob December
to adjust. Turn RATE knob
Turn I:E knob to adjust.
clockwise to get to Turn I:E knob
Check Date - clockwise to get
Month to Check Date -
Day

Check Date - Day in Check Date - Day of


Month Week

Day shown - 10th Day shown - Sunday

Turn RATE knob to adjust. Turn RATE knob to adjust.


Turn I:E knob clockwise to Turn I:E knob clockwise to
get to Check Date - Day of get to Check Time - Hour
Week

Check Time - Minute


Check Time - Hour Time shown - two
minutes past the hour
Time shown - 15.00
Turn RATE knob to
Turn RATE knob to adjust. adjust.
Turn I:E knob clockwise to Turn I:E knob clockwise
get to Check Time - Minute to get to Check Revision
Press Mode Select button
to save any new settings.

45
Check Revision - Check Revision -
Main Board Front Panel

Revision shown - 1.2.1 Revision shown - O.O.A


Turn I:E knob clockwise Turn I:E knob clockwise
to get to Check Revision to get to Error codes.
- Front Panel

Error Codes Error Codes

Error 13 shown All errors cleared

Press Mode Select to clear


error Press INSP PAUSE to
exit engineering mode.

Gas supply filter


Bellows and
diaphragm
assembly

Spirometer tubing
and housing

Exhaust valve
Pressure relief valve
46 (PRV)
SERVICE SCHEDULE
Service frequency
3 Electrical Safety Checks Six months
Carry out the electrical safety checks with the ventilator power switch (ALL item 3
in the ‘OFF’ position. operations)
3.1 Earth continuity (max 0.2 ohms )
3.2 Insulation resistance ( not less than 10 m ohms )
3.3 Earth leakage ( max 100 micro amps )

4 Bellows Assembly (refer to illustrations on previous page)

4.1 Remove and clean bellows housing. One year


4.2 Remove bellows assembly from base. Six months
4.3 Unscrew the three thumb screws and remove the diaphragm valve. Six months
Inspect the valve seat for damage, check the valve disc lies
horizontal. ( Do NOT attempt to dismantle diaphragm assembly ).
Clean the valve seat and valve disc, use an alcohol wipe.
4.4 Remove and discard the O seals from the bellows base and One year
diaphragm valve.
4.5 Reassemble using correct service kit. One year

5 Pneumatic System (refer to illustrations on previous page)


5.1 Remove covers and unlock the main PCB Tray to gain access. Six months
5.2 Using special tool, replace the Quad seal in the exhaust valve. Six months
5.3 Replace Exhaust valve silicon seal. One year
5.4 Using special tool replace ‘O’ seal in PRV One year
5.5 Check internal spirometer tubing, and replace if necessary. One year
5.6 Check external spirometer tubing, and replace if necessary One year
5.7 Replace O seal in spirometer housing Safelock connector One year
5.8 Replace gas supply filter. One year
5.9 Replace pressure reducing regulator. Five years
5.10 Replace all internal tubing. Five years
5.11 Connect a test gauge to the pneumatic assembly, ensure regulator is Six months
set to 35 psi (240 kPa) at 5 L/min.
5.12 Pressure test the gas inlet to inspiratory valve Six months
5.13 Connect the ventilator to a power supply, turn the power switch to Six months
STANDBY.
Set Litres, Rate and I:E Ratio to minimum, and check that at this
position the needle valve is almost shut.

6 Control Unit
6.1 Check all electrical connections and components for security. Six months
6.2 Replace Back up battery. Five years
6.3 Replace real time clock battery. Five years

7 Set Up Six months


7.1 Select drive hose O2 or Air and attach to the OXYGEN/AIR (ALL item 7
connection to the rear panel and ensure the drive gas switch is in operations)
the correct position. (Air down O2 up) connect mini-Schrader to
anaesthetic machine.
7.2 Connect the small corrugated hose to the Driving Gas Ports.
7.3 Connect a hose from the Breathing System Port to a CO2 absorber.
7.4 Connect the pressure tube to MANOMETER/PRESSURE
TRANSDUCER connection on the rear panel and to the To Patient
port of the CO2 absorber

47
SERVICE SCHEDULE

Item Operation Service frequency

7.5 Connect the spirometer tubing to SPIRO connections on the rear


panel and to the spirometer ‘Tee’ piece which should be connected to
the ‘From Patient’ port of the CO2 absorber.
7.6 Connect the breathing tubes to the CO2 absorber and attach the
patient ‘Y’ to a test lung.
7.7 Attach the FRESH GAS SUPPLY to the CO2 absorber and set 02
flow at 0.25 L/m or min flow.
7.8 Connect the gas scavenging system to the EXHAUST PORT.
7.9 Connect the mains cable from the main socket in the rear panel to a
240 VAC supply.

8 Standby mode test Six months


8.1 Switch the ventilator power switch to ‘ON’ verify that two bleeps (ALL item 8
sound and all LEDs light up. operations)
8.2 Check that previous settings are returned to display, (Litres, Rate
and I:E Ratio)
8.3 Set an incorrect rate, i.e. increase Litres and/or Rate controls
clockwise, check ‘INCORRECT RATE OR RATIO’ alarm
8.4 Verify setting lock, and check I:E Ratio display flashes.
8.5 Press ‘MODE SELECT’ check ‘SPIRO’ can be set ‘ON’ or ‘OFF’,
leave set to ‘ON’.

9 Spontaneous Six months


9.1 Switch ventilator switch to SPONTANEOUS and press MODE (ALL item 9
SELECT to access TIDAL VOLUME (MEASURED). operations)
9.2 Switch CO2 absorber to BAG
9.3 Operate Bag Squeezing and check ‘LITRES / MIN’ and ‘RATE B.P.M’
displays indicate readings.
9.4 Stop Bag Squeezing, ensure pressure gauge falls to zero, after 30
seconds delay.
Check ‘LOW AIRWAY PRESSURE’ alarm, and also check that
displays show = =
9.5 Press ‘MUTE’ and verify audible alarm is muted for 30 seconds.

10 Volume cycle Six months


10.1 Switch Ventilator switch to ‘ VOL CYCLED’ (ALL item 10
10.2 Set ‘TIDAL VOLUME LITRES / MIN’ to 0.8, set ‘RATE BPM’ to 10, operations)
‘I:E RATIO’ to 1:2 and ‘AIRWAY PRESSURE LIMIT’ to max.
10.3 Verify 800 ml is delivered (indicated on bellows canister).
10.4 Press ‘INSP PAUSE’ check 25% pause during the inspiratory phase.
10.5 Press ‘MODE SELECT’ check displays indicate values on ‘LITRES /
MIN’ and ‘RATE BPM. Note these will show incorrect values due to
gas and agent variations.
10.6 Remove and occlude expiratory breathing hose and check that the
following alarms operate.
‘HIGH AIRWAY PRESSURE’
‘HIGH CONTINUOUS PRESSURE’
‘LOW AIRWAY PRESSURE’
‘LOW TIDAL VOLUME’
10.7 Disconnect mains supply, verify ventilator continues to operate on
battery, check ‘MAINS FAIL’ alarm
48
SERVICE SCHEDULE
Service frequency
11 Pressure ventilation Six months
11.1 Switch Ventilator selector switch to ‘PRESSURE’ (ALL item 11
11.2 Set ‘AIRWAY PRESSURE LIMIT’ to 20 cmH2O and verify that the operations)
pressure is held at setting.
11.3 Carry out leak test. Set ‘LITRES’ to 1 L/m, ‘RATE’ to 5 bpm and
‘AIRWAY PRESSURE LIMIT’ to 20 cmH2O, ensure rotameter is set
to 200 ml or min. Fill bellows using O2 flush, allow bellows to cycle
and verify bellows remain full.
Maximum permissible leak 200 ml.
11.4 Remove mini schrader from gas supply pressure and check ‘LOW
SUPPLY PRESSURE’ alarm, verify ‘LITRES / MIN’ display flashes.

12 Completion Six months


12.1 Check in Engineering Mode for error messages (ALL item 12
12.2 Sign and date service card, indicate Commissioning, attach card and operations)
plastic wallet to equipment.
12.3 Fill out service report.
12.4 Leave equipment ready for use as requested by the customer.

49
8. SERVICE PROCEDURES

8.1 Cleaning inspected at least every six months or after


1200 hours of use, whichever comes first.
The bellows must be replaced if it shows
8.1.1 Outside surfaces and signs of aging
bellows housing The bellows can be removed by carefully
pulling it off the base.
CAUTION
Care must be taken not to allow liquids to run Exhalation Diaphragm Valve
into the control unit; serious damage may The exhalation diaphragm valve is under the
result. bellows and can be removed by loosening
Check that the unit is disconnected from the the three thumbscrews.
electrical supply before cleaning. The valve seat is now visible.
Do not use cleaning solutions containing
alcohol; the bellows housing may be WARNING
damaged. Great care must be taken not to damage
the precision surface of the diaphragm
To clean the outside surface of the ventilator, valve seat (A).
use a damp cloth. Never use any hard object or abrasive
If necessary use a warm, mild detergent agent to clean it; use only a soft cloth
solution to remove resistant grime. Make (alcohol wipe recommended).
sure that all detergent residues are fully If the valve seat is damaged, the
removed after cleaning. diaphragm valve will leak and may cause
serious malfunction.
Never use any harsh abrasive cleaning
agent. The transparent acrylic bellows Using an alcohol wipe, clean the seat (A),
housing (canister) and, in general, the and the metal disk (B) attached to the base
surfaces of the control unit are not scratch of the diaphragm valve, thoroughly and
resistant. remove all contamination from the surfaces
of both components.
The inside of the canister, under normal
conditions, is not in contact with the NOTE
If excessive contamination is discovered,
breathing gas and therefore only needs check that a bacterial filter is used in the
cleaning as described above. expiratory limb of the breathing circuit (or an
HME at the patient tee-piece).
Remove the canister by slightly twisting it See section 5.1.4.
counter-clockwise until the tabs at the
bottom clear the bayonet locks, then lift it After cleaning, check that the small O-ring
straight up from the base. (C) located in the bellows base under the
diaphragm valve is in place. The ventilator
8.1.2 Bellows and exhalation will not function if the O-ring is missing.
diaphragm valve See section 8.2 for information on
sterilisation procedures.
Refer to illustration on next page.
Each time the bellows assemblies are Refit the diaphragm valve assembly to the
opened for cleaning, all visible components bellows base and reassemble the bellows
must be carefully inspected and damaged assembly.
parts must be replaced.
CAUTION
Bellows Always check for correct fitment of the
As with all elastomers, the bellows material bellows (see illustration) and carry out a full
deteriorates with aging and should be function test of the ventilator before clinical
use.

50
SERVICE PROCEDURES

A C

Exhalation Diaphragm
Valve Assembly

Bellows Base

Refitting the Bellows

51
SERVICE PROCEDURES

8.1.3 Spirometer Sensor and


Tubing Assembly

On a regular basis (at least weekly), the sensor


assembly and sample lines must be thoroughly
cleaned, and sterilised as necessary.

a) Carefully detach the sample lines from the


Luer taper connectors on the rear panel of the A
ventilator, and remove the spirometer assembly
(A) from the absorber.

b) Wash the components, and sample lines


by rinsing with clean (distilled) water poured
into the outlet (not under pressure).
Ensure that all traces of solid contamination
are flushed out.
Note
Latest tubing assemblies have a restrictor
fitted to the tube marked ‘VENT END’ (B)
After washing, check that the restrictor is not
blocked.
B
c) Sterilisation - see section 8.2, below.

d) When the components are dry, refit the


sample lines to the spirometer sensor
assembly and connectors on the rear panel of
the ventilator.
The tube labelled ‘VENT END’ (B), must be
fitted to the ventilator rear panel, as illustrated
above.

52
SERVICE PROCEDURES

8.2 Sterilisation
Recommended guidelines for sterilisation:

CAUTION
To prevent possible damage to components, peak sterilisation temperatures must not exceed
54oC (130oF) for gas (ethylene oxide) or, 134oC (275oF) for steam autoclave.

Do not sterilise the ventilator control unit. The internal components are not compatible with
sterilisation techniques and may be damaged.
Following sterilisation with ethylene oxide, components must be quarantined in a well
ventilated area to allow dissipation of any residual gases.
Follow the recommendations given by the steriliser manufacturer for aeration periods.

ITEM METHOD
Bellows * Gas, liquid, autoclave (20 cycles max.)
Hoses Gas, liquid, autoclave
O rings Gas, liquid, autoclave.
Bellows base Gas, liquid, autoclave.
Exhalation valve assembly Gas, liquid, pasteurise, low temperature autoclave.
Control unit Do not sterilise - see section 9.1.
Bellows canister Liquid.
Spirometer sensor tee piece Gas, liquid, autoclave
Spirometer sample lines Liquid
Distal sensing Tee Liquid

*Applies to standard size bellows only, the optional paediatric bellows must not be steam
autoclaved.

NOTE
Examples of suitable liquid agents are: Nu-Cidex, Sporicidin, and Sonacide.
The exhalation diaphragm valve must be removed, cleaned and sterilised sep
arately.

53
SERVICING PROCEDURES

8.3 Tubing Connectors

To connect and disconnect the tubing from the push-in


connectors, follow the procedures detailed below.

Do not use excess force.

Disconnecting tubing from connector

Push in the tube and the connector


end piece ‘A’.
A
Hold the end piece ‘A’ in place.

Pull the tube out to disconnect.

Fitting tubing to connector

Push in the tube as far as possible


Do not use excessive force.

The connector end piece ‘A’ will also move inwards.

Pull the tube carefully outwards.

The end piece ‘A’ will be pulled outwards to the


‘locked’ position

54
SERVICING PROCEDURES

1
041204
O-ring 6
7
3
0762 Filter
2 (drive gas
inlet)

5
4
041226 042819
O-ring O-ring

8.4 Fitting the Annual Preventive 10 9 300045


Sealing washer
Maintenance Kit
1. Disconnect the mains electrical and gas supply.
8
Exhaust valve
Bellows assembly
1. Remove the bellows housing and bellows (1).
2. Loosen the three screws securing the exhalation
diaphragm valve (2) and remove.
3. Fit a new O rings (3 and 4) to the bellows base (5).
4. Check the condition of the exhalation valve disk
and valve seat (see section 9.1.2)

NOTE If a ventilator is equipped with an optional 11


paediatric bellows assembly, the O-ring on the paediatric
adaptor must also be renewed annually.
Order O-ring, Part No. 041225 0314 O-ring
(pressure relief
Drive gas inlet filter valve)
1. Remove the chassis unit cover.
2. Undo the hose connector (6).
3. Fit a new filter (7), and refit the hose connector.

Exhaust valve
1. Use service tool to unscrew the exhaust valve
cover (8).
2. Remove the valve components and fit a new
sealing washer (9) and O-ring (10). A

Pressure relief valve (PRV)


1. Use service tool to unscrew the PRV cover.
2. Fit a new O-ring (11) to the valve plunger.

Spirometer Tubing Assembly


1. Fit the tubing with label ‘VENT END’ (A) to the
lower connector at the rear of the ventilator 300329
Spirometer Tubing Assembly
55
SERVICING PROCEDURES

8.5 Fitting the Five Year


Overhaul Kit 4
1. Disconnect the ventilator from 1
electrical and gas supplies. Remove
the cover from the ventilator chassis.
2. Remove the two screws (1) to allow 5
the main PCB / tray assembly to be
pivotted upwards.
3. Disconnect the leads from the battery
(2).
2
Regulator assembly
4. Remove the securing screws (3), and
detach the regulator (4) from the 1
manifold assembly.

Mains back-up battery

5. Carefully break the adhesive seal and


detach the alarm system speaker (5)
from the top of the battery.
6. Break the adhesive seal between the
battery and the base of the ventilator 3
chassis and remove the battery.
7. Fit a new battery, using a new
adhesive pad to hold in place.
8. Refit the speaker (5), using a new
adhesive pad to hold in place. 4
9. Reconnect the battery leads.

Real time clock battery


CAUTION
To prevent possible contamination and
subsequent corrosion, do not handle the
battery.

10. Fit a new battery (6).


Check for real time errors which may
occur because of battery
replacement.

Internal Tubing 6
11. Renew all internal tubing (including
spirometer tubing if fitted).
Cut the new tubing to the exact
lengths specified below.
Spirometer accuracy is dependant
on correct tubing lengths.

56
4
64 mm 45 mm
40 mm

Plan
View

1 (160 mm)

1 (115 mm) Ref Part No.


2
1 462541 (623 mm / 4 mm o.d.)
2 054541 (Adaptor)
3 462544 (220 mm / 8 mm o.d.)
4 462545 (149 mm 10 mm o.d.)
5 011092 (430 mm / 6 mm o.d.)
6 053211 (Plug - fitted only to
models without gauge)
7 054947 (Tee piece)
8 054949 (Connector)
9 011092 (510 mm / 6 mm o.d.
(models with spirometer)
3

5 (350 mm) 1 (18 mm)


6 1 (330 mm)
Side View

5
5 (80 mm)
8

57
SERVICING PROCEDURES

8.6 Manifold Assembly


3
1. Disconnect the ventilator from 1
electrical and gas supplies. Remove
the cover.
2. Remove the two screws (1) to allow
the main PCB / tray assembly to be
pivotted upwards.
3. Disconnect the back-up battery wiring
(2).
4. Disconnect all tubing and wiring 2
connectors (3) from the manifold
assembly.
5. From underneath the ventilator, 1
remove the three screws (4) to allow
removal of the manifold assembly (5).
5

8.7 Electrical Power Supply

1. Disconnect the ventilator from


electrical and gas supplies. Remove 1
the cover.
2. Remove the two screws (1) to allow
the main PCB / tray assembly to be
pivotted upwards.
3. Disconnect the back-up battery wiring
(2).
4. Remove the two screws and detach
the mains in/out socket / filter unit (3)
from the chassis, but do not 2
disconnect the wiring.
5. From underneath the ventilator,
remove the four screws and lift the 1 4
power supply unit (4) from the
chassis.
Disconnect the wiring from the power 3
supply unit to allow removal.

58
SERVICING PROCEDURES

8.8 Front Panel and PCB


Assembly, and Gauge
1
NOTE
The display PCB and the control PCB are
not available as separate items. A front
panel and PCB assembly must be ordered
if either PCB is faulty.
2
1. Disconnect the ventilator from
electrical and gas supplies. Remove
the cover. 1
2. Remove the two screws (1) to allow
the main PCB / tray assembly to be 3
pivotted upwards.
3. Disconnect the back-up battery wiring
(2).
4. Disconnect the wiring connectors (3) 4
from the PCB. 6
5. Remove the four screws (4) and 4
detach the front panel assembly (5)
from the ventilator.
6. On models with a manometer:
Disconnect the tubing (6) from the
back of the gauge.
Slacken the screws on the retaining
ring (7) and detach the gauge (8) from 4 5
the assembly.
7. Carefully remove the caps from each
control knob (9), then use the special
tool to remove each knob. 7
8. To reassemble, reverse the removal
procedure
9. Fitting a new front panel assembly -
On models with a manometer, the
new front panel must be cut before the 8
gauge can be refitted.to allow fitment
of the gauge.

59
SERVICING PROCEDURES

8.9 Main PCB Tray Assembly

NOTE 1
The main PCB is not available as a
separate item. A tray assembly complete
with pre-programmed PCB must be ordered
if the PCB is faulty.

1. Disconnect the ventilator from


electrical and gas supplies. Remove 2
the cover.
2. Remove the two screws (1) to allow
1
the main PCB / tray assembly to be
pivotted upwards.
3. Disconnect the back-up battery wiring 6
(2), then relocate the tray assembly. 4
4. Disconnect the wiring connectors (3) 3
from the PCB .
5. Disconnect the spirometer tubing
from the connectors (4) on the rear
panel (models with spirometry). 6
6. Disconnect the pressure transducer 4
tubing from the connector (5) on the
PCB.
7. Remove the two remaining screws (6) 5
and detach the tray assembly (7) from
the ventilator chassis.
8. To reassemble, reverse the removal
7
procedure.

60
9. PARTS LIST

Annual Preventive Maintenance Kit Cat No 57633

041204 O-ring - bellows base


041226 O-ring - bellows base
042819 Quad seal - exhaust valve
300045 Sealing washer - exhaust valve
0314 O-ring - Pressure relief valve
0762 Filter - drive gas inlet
043 O-ring (Safelock connector - spirometer housing)
300329 Spirometer Tubing Assembly

042819
Quad seal

300045
Sealing
washer
0762 Filter
(drive gas inlet)
Exhaust valve

0314 O-ring
(pressure relief valve)

61
PARTS LIST

043
O-ring
(Safelock 300329 Spirometer Tubing
connector - Spirometer Assembly - 300329
spirometer Tubing
housing) Assembly The spirometer sample
line fitted to the lower
connector at the rear of
the ventilator has a
restrictor fitted.
This tube is labelled
‘VENT END’ (A), and this
end of the tube must be
fitted to the ventilator
041204 rear panel, as illustrated
O-ring above.

041226
O-ring
Note
If a paediatric bellows
assembly is used with the
AV800, the O-ring on the
bellows adaptor must be
renewed annually.
Order O-ring, Part No 41225.

62
PARTS LIST

Five Year Overhaul Kit Cat No 57635

104019 Battery - real time clock


103996 Battery - mains back-up, 12 V
045538 Regulator
Tubing - internal (see next page)

103996
Battery -
104019 mains back-up
Battery -
real time
clock

045538
Regulator

63
PARTS LIST

Internal Tubing
NOTE : Replacement tubing must be to the length as specified on illustrations

Ref Part No. Qty. Description

1 462541 623 mm Tubing 4 mm o.d.

2 054541 1 Adaptor
3 462544 220 mm Tubing 8 mm o.d.
4 462545 149 mm Tubing 10 mm o.d.
5 011092 430 mm Tubing 6 mm o.d.
6 053211 1 Plug (fitted only to models without gauge)
7 054947 1 Tee piece (3 mm / 2.25 mm)
8 054949 1 Connector

4
Plan View 64 mm 45 mm
40 mm

1 (160 mm)

1 (115 mm)
2

3
64
5 Side View

8
1 (18 mm)

5 (350 mm) 1 (330 mm)


6

3
5
5 (80 mm)
7

65
PARTS LIST
1
8

10

11
3 4

Bellows and Base Assembly

Ref Part No. Qty. Description

1 57551 1 Canister
2 57550 1 Bellows
3 406020 1 Diaphragm valve assembly
4 041204 1 O ring
5 57548 1 Bellows base assembly
6 041226 1 O ring
7 019103 4 Screw - M5
011107 4 Knurled knob

8 57553 1 Canister - paediatric (available as an option)


9 57552 1 Bellows - paediatric (available as an option)
10 57544 1 Adapter assembly - paediatric (available as an option)
11 041225 1 O ring

66
PARTS LIST

4
2
7

6 8
7 av8ser12.tif

PCBs and Electrical System Components

Ref Part No. Qty. Description

1 300xxx 1 AV80 Main PCB tray assembly


2 300xxx 1 AV800 Display PCB and front panel assembly

3 104019 1 Battery - PCB, LiMn, 3 V, 260 mAh, 24.5 mm


4 103996 1 Battery - mains back-up, 12 V
5 104007 1 Speaker - alarm systems
300006 1 Speaker and wiring loom
6 105715 1 Mains inlet / Filter assembly
7 1040008 2 Fuse, 3.15 A, 20 mm , anti-surge
8 103998 1 Electrical power supply assembly

67
PARTS LIST

Rear panel Connectors, Valve Assemblies, and Manometer

Ref Part No. Qty. Description

1 35150 1 Hose connector - manometer / pressure transducer

Pressure Relief Valve


2 300073 1 PRV body
3 300072 1 Valve plunger
4 300074 1 Cap
5 0314 1 O seal, Viton
- 031047 1 Spring
- 054540 1 Connector

6 300038 1 O2 connector - DISS


020013 1 Locking nut, M12
025205 1 Washer, 0.5 inch i.d.
7 0762 1 Filter
8 054519 1 Straight adaptor. 1/8 BSP to 8 mm tube

Exhaust valve
9 300041 1 Cap
10 300044 1 Spacer
11 300045 1 Washer
12 031045 1 Spring
13 300043 1 Piston
14 042819 1 Seal, quad ring
15 054541 1 Straight adaptor, M5 to 4 mm tube

16 300040 1 Tapered connector - bellows drive gas

17 049040 1 Gauge assembly (includes locking ring)

18 104001 1 Knob - off/standby/mode control


104010 1 Nut cover
104009 1 Cap

19 104000 1 Knob - airway pressure limit control


104010 1 Nut cover
104009 1 Cap

20 103999 1 Knob
104002 1 Nut cover
104003 1 Cap

68
PARTS LIST

2 15
1 3 4

14
12

16 13
11
8 7 10
9

17

18

19

20
69
PARTS LIST
1

2 3

4
5
8

Pneumatic System Assemblies


Ref Part No. Qty. Description

1 300027 1 Manifold assembly (includes wiring - not shown, and items 2 to 7:


2 045438 1 Regulator assembly (includes O rings)
3 102952 1 Pressure switch
4 054946 1 Coupling (flexible)
5 054948 1 Coupling (includes washer)
6 045053 1 Solenoid valve
7 045052 1 Solenoid valve
8 300011 1 Connector assembly

70
PARTS LIST

4
6
5

3 Original Specification

7
2 2
8

Revised Specification
3 (May 2000 onwards)

Spirometer System Components

Original Specification Revised Specification - May 2000 onwards


Ref Part No. Qty. Description Ref Part No. Qty. Description

1 300109 1 Sensor and PCB assembly 1 300117 1 Sensor and PCB assembly
(includes tubing and fittings) (includes tubing and fittings)

2 57631 1 Spirometer assembly,


includes: 2 300124 1 Spirometer assembly,
34331 1 Safelock nut includes:
043 1 O seal 34331 1 Safelock nut
043 1 O seal
3 57632 1 Hose assembly
(tubing and connectors) 3 300329 1 Hose assembly
Panel connectors (tubing and connectors)
4 011245 1 Luer connector Panel connectors
5 011244 1 Luer connector 4 011245 1 Luer connector
6 300059 1 Nut - tube adaptor 5 011244 1 Luer connector
7 011249 1 Luer connector 6 300059 1 Nut - tube adaptor
7 011249 1 Luer connector
8 011260 1 Identification Ring (Red)

71
PARTS LIST

72
10. APPENDIX

APPENDIX 1
CARE OF BACK-UP BATTERY
CAUTION
Damage may occur if the battery is allowed to remain in
a discharged state. Never discharge the battery to below 10.2 volts.

A. Battery installed in ventilator


The battery must be charged before the machine is released
for use with an 14 hour charge from the ventilator’s internal power
supply (ventilator connected to the mains supply, but not
running).
Note that the OFF indicator will show a yellow light
during charging.

Subsequently the recharge periods for a battery on a machine


in store are similar to those in B, below.

Batteries in machines in normal use will be kept charged by the


internal power supply and will only require special charging care
following the discharge test carried out during function testing,
or if the battery is fully discharged following use during mains
failure.

B. Battery care/storage requirements.


During storage batteries will require a periodic recharge, the
frequency of which is determined by the storage temperature,
which must not exceed 50oC (120oF).

Storage Recharge
temperature period

38 to 50oC (100 to 122oF) 1 month

21 to 38oC (70 to 100oC) 3 months

7 to 21oF (45 to 70oF) 6 months

0 to 7oC (32 to 45oF) 9 months

-5 to 0oC (23 to 32oF) 12 months

Duration - recharge until the charge current is less than 25 mA (typically overnight).
It is recommended that at each charge an updated label is affixed to each battery to indicate
date of the last charge.

C Battery disposal
Dispose of used batteries according to hospital, local, state, and federal regulations.

73
74
Cat No 52704
Doc No AV8 0103SM
March 2003

Penlon Limited
Abingdon
OX14 3PH
UK

Technical Support
Tel: +44 (0) 1235 547076
Fax: +44 (0) 1235 547062
E-mail: [email protected]

International Sales
Tel: +44 1235 547001
Fax: +44 1235 547021
E-mail: [email protected]

UK Sales
Tel: 01235 547036
Fax: 01235 547023
E-mail: [email protected]

You might also like