2024 - Product Catalog Ultracentrifuge

Download as pdf or txt
Download as pdf or txt
You are on page 1of 15

PREPARATIVE & ANALYTICAL

ULTRACENTRIFUGES
PRECISION-MADE PERFORMANCE

P R O D U C T C ATA L O G
ULTRA POWERFUL SO QUIET
SOPHISTICATED MULTILINGUAL
INTELLIGENT ALWAYS SECURE
SYSTEMATIC EXELLENCE
INTRODUCTION

Since the introduction of the first commercial ultracentrifuge in 1947 Rotors with the BioCertified symbol have been tested to demonstrate
— the classic Beckman Model L — Beckman Coulter Life Sciences has containment of aerosols under normal operating conditions of the
been at the forefront of centrifuge innovation. Although the physics associated Beckman Coulter centrifuge when used and maintained
of this basic separation technique never change, we continually as instructed.
design new and innovative rotors and accessories, and develop
advanced methods that allow the forces of centrifugation to be
applied in new ways.

This centrifuge product selection guide is designed to


help you determine the most efficient centrifuge tools for
your laboratory.

Each section begins with a brief description of instruments that


Beckman Coulter Life Sciences offers within that centrifuge
category. Because biocontainment and sterility are major concerns
in today’s laboratories, we provide a number of options that
address this issue. Special BioSafe instruments and accessories
are available across our centrifuge product line and are identified
with the following icons:
Following the centrifuge descriptions, listings of their rotors are
BioSafe and BioSafety are terms intended to describe
included with information on speed and g-force capability. Also
the enhanced biocontainment features of our products.
included is information on tubes and bottles that can be used and
BioCertified is a term used to describe our products the adapters they require.
which have been tested and validated to demonstrate
Tubes and bottles are cross-referenced in a separate section that
containment of microbiological aerosols by an
provides details on tube materials, chemical compatibility, tube
independent, third-party facility (Health Protection
designs, and tube closure options.
Agency, Porton Down, UK). Improper use or maintenance
may affect seal integrity and, thus, containment. A reference section at the back of the guide includes
quick-reference charts on instrument and tube selection, as well
BioEnhanced is a term intended to describe
as frequently used formulas.
our products’ enhanced level of biocontainment
by design.

Contact your local Beckman Coulter office


Please visit us at beckman.com/contact-us

Additional centrifugation resources:


beckman.com/centrifugation

Disclaimer: The content included in this catalog is for reference purposes only. The most accurate and up-to-date information
is located in Instrument and Rotor Instructions For Use which are available at beckman.com/techdocs

Beckman Coulter Ultracentrifuges | beckman.com/centrifuges i


1. Analytical UC

FLOOR PREP ULTRACENTRIFUGATION

Optima X Series Floor Model Preparative Ultracentrifuge


2. Floor Prep UC

Experience the visibility and control of our intelligent user interface that makes the Optima X Series
a breeze to operate. Our entry-level model, the XE delivers all of the features required to quickly set
up and complete a run. Our premium model, the XPN, boasts networking capability and a variety of
customizable security and tracking features, ensuring chain of custody and compliance. Both models
feature multi-layered BioSafety features in order to maintain a safe working environment. And, as
you have come to expect, all instruments are designed as a total system with our innovative rotors
and labware.
3. Tabletop UC

Large touchscreen display Energy efficient


Easy to read and navigate. Viewable from across the lab. Intuitive Maintains runs during power fluctuations. Regenerative braking
graphical interface with on-screen help. Multilingual software returns energy to local circuit. Highly efficient thermoelectric
supporting 9 major languages. Touchscreen size approximately cooling results in lower power consumption. Idle state energy usage
15in (38cm). equivalent to a 60 Watt light bulb.
4. Micro-UC

Remote monitoring and control (XPN only) Powerful on-board simulation and calculation
Saves time. Start, stop, and monitor from computer or smart phone.
tools (XPN only)
Receive diagnostic alerts via e-mail. iPhone or Android compatible. Saves time and samples. Proprietary software optimizes protocols
before running to ensure maximum efficiency. Quickly perform
common calculations and conversions.

GMP supporting and security features (XPN only)


Password protection and electronic signatures help you maintain
5. Tubes

chain of custody and compliance. Export run and system logs for
record keeping. Manage rotor life by serial number.

Quiet-drive technology
Designed for low drive noise. Operates at less than 51 dBA.

Part Numbers*
6. Accessories

Instrument BioSafe Package**


Model
IVD1 Non-IVD2 IVD1 Non-IVD2

Optima XPN-100 A99846 A94469 B10053 B10048

Optima XPN-90 A99842 A94468 B10052 B10047

Optima XPN-80 A99839 A95765 B10051 B10046


7. Reference

Optima XE-100 A99836 A94516 B10050 B10045

Optima XE-90 A99833 A94471 B10049 B10044

* Availability varies by geography. Contact your local sales representative to


determine the correct part number for ordering.
** BioSafe Package includes centrifuge and HEPA/Pharmaceutical-Grade
Sterilizing filter kit PN 354457.
1 For In Vitro Diagnostic Use.
2 For Research Use Only, not for use in diagnostic procedures.
8. Support

2-1 Beckman Coulter Ultracentrifuges | beckman.com/centrifuges


1. Analytical UC
FLOOR PREP ULTRACENTRIFUGATION

Features Optima XPN Optima XE

2. Floor Prep UC
Giant LCD touchscreen display. Approximately 15 inch (38 cm) • •
Intuitive, multilingual interface • •
Improved power tolerance • •
Compatible with existing Beckman Coulter rotors • •
Step-by-step zonal/CF operation screens • •
Rotor tracking by serial number •
User-defined programs •
Remote monitoring and control •

3. Tabletop UC
Password protection and three user access levels •
Exportable run logs and electronic signatures •
eXPert simulation software •
Speed/temperature vs. time plot •

Specifications

Optima XPN Optima XE

4. Micro-UC
Maximum speed (rpm) 100,000 100,000
90,000 90,000
80,000

Maximum RCF (x g) 802,400 802,400


694,000 694,000
548,300

Speed control ± 2 rpm of set speed (above 1,000 rpm)

Set temperature 0 to 40°C in 1°C increments

Temperature control ± 0.5°C of set temperature∞

Ambient operating range 10 to 35°C

5. Tubes
Accel/decel profiles 10 accel/11 decel

Braking Regenerative, power reclamation

Operating languages English, French, Italian, German, Spanish, Japanese, Russian, Korean, and Chinese (Simplified)

Advanced software Optima eXPert calculations, simulations, references No

User-defined programs 1,000 with 30 steps each No

User profiles 50 unique users and passwords No

6. Accessories
Electronic signature Yes No

Rotor tracking By serial number No

Data transfer (3) USB 2.0 type A connectors, (1) RJ-45 connector No

Heat output 3400 Btu/hr (1.0 kW)

Drive cooling Air-cooled

Refrigeration system Thermoelectric - no CFCs, ODCs

Sound level <51 dBA

Remote control Available No


7. Reference

HEPA filter Available (Pharmaceutical-Grade Sterilizing Filter)

Power requirement 200-240 VAC, 30 A, 50/60 Hz

125.7 x 94 x 68.1 cm
Dimensions (H x W x D)
49.5 x 37 x26.8 in

Weight 485 kg (1068 lb)

∞ ± 0.8ºC when using SW 32 swinging-bucket rotor.


8. Support

Beckman Coulter Ultracentrifuges | beckman.com/centrifuges 2-2


1. Analytical UC

FLOOR PREP ULTRACENTRIFUGATION

Quick-Reference Guide for Rotor Specifications for Floor Model Ultracentrifuges


2. Floor Prep UC

No of Tubes x Nominal Nominal Rotor


Rotor P/N Max. RPM Max.g k Factor
Tube volume (mL) Capacity (mL)

Fixed-Angle Rotors

Type 100 Ti 363013 100 000 802 000 15 8 x 6.8 54

Type 90 Ti 355530 90 000 694 000 25 8 x 13.5 108

Type 70.1 Ti 342184 70 000 450 000 36 12 x 13.5 162


3. Tabletop UC

Type 70 Ti 337922 70 000 504 000 44 8 x 39 312

Type 50.4 Ti 347299 50 000 270 000 39 44 x 6.5 286


312 000 33

Type 50.2 Ti 337901 50 000 302 000 69 12 x 39 468

Type 45 Ti 339160 45 000 235 000 133 6 x 94 564

Type 42.2 Ti 343007 42 000 223 000 12 72 x 230 μL 16.5

Type 25 Ti 347261 25 000 70 300 84 100 x 1 100


81 400 71
4. Micro-UC

92 500 62

Type 19 Ti 325620 19 000 53 900 951 6 x 250 1500

Near-Vertical Rotors

NVT 100 365898 100 000 750 000 8 8 x 5.1 40.8

NVT 90 362752 90 000 645 000 10 8 x 5.1 40.8

NVT 65.2 361073 65 000 416 000 15 16 x 5.1 81.6

NVT 65 362755 65 000 402 000 21 8 x 13.5 108


5. Tubes

Vertical Rotors

VTi 90 362751 90 000 645 000 6 8 x 5.1 40.8

VTi 65.2 362754 65 000 416 000 10 16 x 5.1 81.6

VTi 65.1 362759 65 000 402 000 13 8 x 13.5 108

VTi 50.1 C78759 50 000 251 000 34 12 x 39 468


6. Accessories

Swinging-Bucket Rotors

SW 60 Ti 335649 60 000 485 000 45 6x4 24

SW 55 Ti 342194 55 000 368 000 48 6x5 30

SW 41 Ti 331362 41 000 288 000 124 6 x 13.2 79.2

SW 40 Ti 331302 40 000 285 000 137 6 x 14 84

SW 32.1 Ti 369651 32 000 187 000 229 6 x 17 102


7. Reference

SW 32 Ti 369650 32 000 175 000 204 6 x 38.5 231

SW 28.1 342216 28 000 150 000 275 6 x 17 102

SW 28 342207 28 000 141 000 246 6 x 38.5 231

Continuous Flow & Zonal Rotors

CF-32 Ti 350700 32 000 102 000 42 N/A 430

Ti-15 969312 32 000 102 000 481 N/A 1675


8. Support

2-3 Beckman Coulter Ultracentrifuges | beckman.com/centrifuges


1. Analytical UC

SW 32 Ti 6 x 38.5 mL

Swinging-Bucket Rotor and Buckets, Titanium


For use in instruments classified: H R S
2. Floor Prep UC

Major applications: Separation of subcellular particles and


viruses in density gradients.
rmin (66.8 mm)

(109.7 mm) Number of Tubes Rotor


rav Max. RPM Max. g k Factor
Volume/Size Capacity
rmax (152.5 mm)
32 000 175 000 204 6 x 38.5 mL 231 mL
Axis of Rotation 1 x 3.5 in
25 x 89 mm

No. 369694. SW 32 Ti Rotor Package. Includes Rotor, 1 box 326823 Polypropylene Tubes, 2 boxes 344058 Ultra-Clear Tubes, 1 box 355631 Polycarbonate Tubes, 1 each
3. Tabletop UC

331186 Bucket Holder Rack, 306812 Spinkote Lubricant, 335148 Vacuum Grease, 335456 Overspeed Disk, 978354 hooked O-ring Removal Tool, and 24 each
812715 O-rings.
No. 369650. SW 32 Ti Rotor Assembly.

Tubes and Bottles

Tube Style/ Vol. (mL) Size Required Caps /


Part No. Quantity Adapters (qty. 6) g-Force k Factor Max. Speed
Material g-Max (mm) Spacers (qty. 1)

OptiSeal 32.4 361625 50 26 x 77 392833§§ - 175 000 155 32 000


Polypropylene

Quick-Seal 15.0 343664 50 25 x 38 355536 - 175 000 81 32 000


4. Micro-UC

Polypropylene 27.0 343665 50 25 x 64 355536 - 175 000 146 32 000


Bell-Top 33.5 344623 50 25 x 83 355536 - 175 000 192 32 000

Quick-Seal 8.4 358652 50 25 x 38 355536 358156 175 000 74 32 000


konical 28.0 358651 50 25 x 83 355536 358156 175 000 198 32 000
Polypropylene
Bell-Top

Thinwall konical 25.5 358125 50 25 x 76 - 358156 175 000 156 32 000


Polypropylene 31.5 358126* 50 25 x 89 - 358156 175 000 199 32 000

Thickwall 31.0 355642 25 25 x 89 - - 175 000 198 32 000


Polypropylene

Thinwall 38.5 326823* 50 25 x 89 - - 175 000 204 32 000


5. Tubes

Polypropylene

Thickwall 31.0 355631 25 25 x 89 - - 175 000 202 32 000


Polycarbonate

Ultra-Clear 38.5 344058* 50 25 x 89 - - 175 000 204 32 000

Sterile and Certified-Free Tubes

Vol. Beckman Certified Sterile Required Required


Tube Style/ Size
(mL) Standard Free Part Qty Part Qty Caps / Spacers Adapters g-Force k Factor Max. Speed
Material (mm)
6. Accessories

g-Max Part Number Number Number (qty. 1) (qty. 1)

Ultra-Clear 38.5 25 x 89 344058 C13926 50 C14292 48 - - 175 000 204 32 000

Polypropylene 38.5 25 x 89 326823 C14285 50 C14301 48 - - 175 000 204 32 000

Polypropylene 30 25 x 89 358126 C14291 50 C14307 48 - 358156 175 000 199 32 000


konical

Note: To help you locate g-Max tubes in the rotor listings, these tubes are
Rotor Replacement Parts
highlighted with bold, italic type.
§§ Set of 2. 331186 Bucket Holder Rack
C46871 Rotor Stand
* Sterile and C-Free options available. 335456 Overspeed Disk, 32 000 rpm
7. Reference

369643 Bucket Cap (each)


369692 Bucket Assembly, with Caps and O-rings, set of 6
812715 O-ring for Bucket

Extra Buckets for the SW 32 Ti

The buckets of the SW 32.1 can be used with the SW 32 rotor head.
369693 Bucket Assembly, with Caps and O-rings, set of 6

Adapters/Spacers

355536 358156 392833


8. Support

2-27 Beckman Coulter Ultracentrifuges | beckman.com/centrifuges


1. Analytical UC
TUBES AND BOTTLES

Tubes and Bottles For Every Application


No single tube design or material will meet all application requirements. A number of factors should be considered at the time a supply

2. Floor Prep UC
of tubes is ordered: the particular technique to be used, the nature of the sample and any solvent or gradient media, the desirability
of reusing the tubes, and certain convenience factors. The properties listed below provide a guide for anyone involved in the tube
selection process.

• Strength and Flexibility, to resist permanent deformation even when run in fixed-angle rotors without tube caps.

• Chemical Resistance to a wide range of bases, acids, and solvents.

• Transparency, to permit a clear view of fractions and bands after centrifugation.

3. Benchtop UC
• Thin enough to be sliced or punctured after centrifugation for fraction collection.

• Impermeable to Water, to prevent aqueous solutions from permeating the tube wall and reaching the rotor cavity.

• Surface Properties that prevent the adherence of nucleic acids and proteins.

• Temperature Tolerance throughout a wide range of operating temperatures, without deforming at high temperatures or cracking
when used close to 0°C.

• Autoclavable, for convenient sterilization and reuse.

The full line of Beckman Coulter tubes includes a number of tube materials, each with its own distinct combination of properties, to meet a

4. Micro-UC
variety of application requirements. Available are transparent, translucent, and opaque tubes, tubes that can be sliced or punctured, tubes
that can be sterilized and reused, and tubes that are resistant to a variety of chemical compounds.

5. Tubes
6. Accessories
7. Reference
8. Support

Beckman Coulter Ultracentrifuges | beckman.com/centrifuges 5-1


1. Analytical UC
TUBES AND BOTTLES

Sterile and Certified Free Tubes


Whether you are helping to solve complex biological problems, investigating the causes of diseases or

2. Floor Prep UC
finding potential new cures, protecting your sample from degradation and contamination during your
study is paramount.

Certified Free Certified Free + Sterile


Certified Free tubes exhibit no detectable DNA, DNase, RNase, Beckman sterile ultracentrifuge tubes are sterilized via
PCR inhibitors, or endotoxins.* ethylene oxide.

Ideal for Genomics, Exosomes, and Proteomics workflows Ideal for Protein Prep and Virus Purification workflows

3. Benchtop UC
• Prevents sample degradation and contamination • Includes all advantages of being certified free

• Double-bagged to prevent contamination during • Pre-sterilized ready-to-use for improved


cleanroom transfer workflow efficiency

• Improve QC data quality • Eliminate need for developing sterilization protocol

• Certificates available on beckman.com • Packaged in workflow-based quantities

*Based on sample results below detectable limits. • Sterilized and packaged compliant with
ISO 11135 and 11607

4. Micro-UC
• Certificates available on beckman.com
Selection Guide
Nominal
Nominal size Beckman Certified Certified
Tube type Capacity Compatible rotors
(mm) Standard Free Free + Sterile
(mL)

Type 70 Ti, Type 50.2 Ti,


38.5 25x89 344058 C13926 C14292
SW 32 Ti, SW 28, JS-24.38

17 16x102 344061 C14281 C14297 SW 32.1 Ti, SW 28.1

14 14x95 344060 C14278 C14294 SW 40 Ti

5. Tubes
Ultra-clear Open-top
13.2 14x89 344059 C14277 C14293 SW 41 Ti
Round Bottom
Type 90 Ti, Type 70.1 Ti, Type 70 Ti, Type 50.4 Ti,
6.5 13x64 344088 C14280 C14296
Type 50.2 Ti, Type 45 Ti, MLA-55

5 13x51 344057 C14279 C14295 SW 55 Ti

4 11x60 344062 C14282 C14298 SW 60 Ti

Ultra-clear Quick Seal


39 25x89 344326 C14283 C14299 Type 70 Ti, Type 50.2 Ti, VTi 50, VTi 50.1
Round Bottom

6. Accessories
38.5 25x89 326823 C14285 C14301 Type 70 Ti, 50.2 Ti, SW 32 Ti, SW 28, JS-24.38

Polypropylene 14 14x95 331374 C14287 C14303 SW 40 Ti


Open-top Round Bottom 13.2 14x89 331372 C14286 C14302 SW 41 Ti

4 11x60 328874 C18669 C18674 SW 60 Ti

100 38x102 345776 C14289 C14305 Type 45 Ti, JA-18


Polypropylene
39 25x89 342414 C14288 C14304 Type 70 Ti, Type 50.2 Ti, VTi 50, VTi 50.1
Quick Seal Round Bottom
13.5 16x76 342413 C14290 C14306 Type 90 Ti, Type 70.1 Ti, NVT 65, VTi 65.1, MLA-55

Polypropylene
30 25x89 358126 C14291 C14307 SW 32 Ti, SW 28, JS-24.38
Open-top konical
7. Reference
8. Support

Beckman Coulter Ultracentrifuges | beckman.com/centrifuges 5-5


1. Analytical UC

TUBES AND BOTTLES

Nominal Filling Rotors


Nominal Size (mm) Part No.
Capacity (mL) See specific rotor pages for required spacers/adapters/accessories
2. Floor Prep UC

OptiSeal Polypropylene Tubes

3.3 13 x 33 361627 SW-55 Ti, SW-50.1, TLN-100, MLS-50

4.7 13 x 48 361621 Types 50.4 Ti, 50.3 Ti, TLA-110, TLA-100.4

4.9 13 x 51 362185 VTi 90, VTi 65.2, NVT 90, NVT 65.2

8.9 16 x 60 361623 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti, 50, MLA-55

11.2 16 x 70 362181 VTi 65.1, NVT 65

32.4 25 x 77 361625 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, 42.1, 30, SW 32 Ti, SW 28, MLA-50

36.2 25 x 89 362183 VTi 50.1


3. Benchtop UC

Quick-Seal Polypropylene Tubes

1.0 8 x 51 345831 Type 25

1.2 8 x 35 361082 TLN-120

1.5 11 x 25 344624 SW 60 Ti, TLA-120.2, TLA-100.2, MLA-130, MLA-150, TLS-55

2.0 11 x 32 344625 SW 60 Ti, TLA-120.2, TLA-100.2, MLA-130, MLA-150, TLS-55

2.0 13 x 25 345829 Types 100 Ti, 50.4 Ti, 50.3, NVT 100, NVT 90, NVT 65.2, VTi 90, VTi 65.2, SW 55 Ti,
SW 50, TLA-110, TLA-100.4, TLA-100.3, MLS-50

3.5 13 x 32 349621 Types 100 Ti, TLA-110, TLA-100.3


4. Micro-UC

3.5 14 x 25 355870 SW 41 Ti, SW 40 Ti

3.9 13 x 38 358980 TLN-100

4.2 16 x 32 356562 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti, 40, SW 30.1, SW 32.1 Ti, SW 28.1, MLA-80, MLN-80, MLA-55

5.1 13 x 51 342412 VTi 90, VTi 65.2, NVT 100, NVT 90, NVT 65.2

5.1 13 x 51 362248 Type 100 Ti, TLA-110

5.9 14 x 47 355537 SW 41 Ti, SW 40 Ti

6.0 13 x 64 344619 Types 100 Ti, 50.4 Ti, 50.3 Ti

6.3 16 x 45 345830 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti, 50, 40, NVT 65, VTi 65.1, SW 30.1, SW 32.1 Ti,
SW 28.1, MLA-80, MLN-80, MLA-55
5. Tubes

8.0 16 x 58 344621 Type 50, NVT 65, VTi 65.1, SW 32.1 Ti, SW-28.1, MLA-80, MLN-80

10.0 16 x 67 344622 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti ,40, NVT 65, VTi 65.1, SW 32.1 Ti, SW 28.1, MLA-55

13.5 16 x 76 342413* Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti ,40, NVT 65, VTi 65.1, MLA-55

15.0 25 x 38 343664 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, VTi 50, VTi 50.1, SW 30, SW 32 Ti, SW 28, MLA-50

17.0 16 x 102 356291 SW 32.1 Ti, SW 28.1

27.0 25 x 64 343665 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, 42.1, VTi 50, VTi 50.1, SW 32 Ti, SW 28, MLA-50

33.0 25 x 83 344623 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, SW 32 Ti, SW 28
6. Accessories

39.0 25 x 89 342414* Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, VTi 50, VTi 50.1

100.0 38 x 102 345776* Type 45 Ti

Quick-Seal Polypropylene konical Tubes

1.3 11 x 35 358655 SW 60 Ti

3.0 11 x 60 358648 SW 60 Ti

4.0 14 x 48 358650 SW 41 Ti, SW 40 Ti

8.0 14 x 89 358649 SW 41 Ti, SW 40 Ti

8.5 25 x 38 358652 SW 30, SW 32 Ti, SW 28


7. Reference

28.0 25 x 83 358651 SW 32 Ti, SW 32 Ti, SW 28

Quick-Seal Ultra-Clear Tubes

5.1 13 x 51 344075 NVT 90, NVT 65.2, VTi 90, VTi 65.2

6.0 13 x 64 344320 Types 50.4 Ti, 50.3 Ti

13.5 16 x 76 344322 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti, 40, NVT 65, VTi 65.1, MLA-55
8. Support

* Sterile and Certified Free Options Available

5-6 Beckman Coulter Ultracentrifuges | beckman.com/centrifuges


1. Analytical UC
TUBES AND BOTTLES

Nominal Filling Nominal Size Rotors


Part No.
Capacity (mL) (mm) See specific rotor pages for required spacers/adapters/accessories

2. Floor Prep UC
Quick-Seal Ultra-Clear Tubes (cont’d)

15.0 25 x 38 344324 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, VTi 50, VTi 50.1, MLA-50

27.0 25 x 64 344323 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, VTi 50, VTi 50.1, MLA-50

39.0 25 x 89 344326* Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, VTi 50, VTi 50.1

100.0 38 x 102 345778 Type 45 Ti

Ultra-Clear Tubes

175 μL 5 x 20 344718 A-110, A-100/30, A-100/18

3. Benchtop UC
450 μL 8 x 20 345843 A-95

0.8 5 x 41 344090 SW 55 Ti, SW 50.1, MLS-50

2.0 8 x 49 344091 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti, 50.4 Ti, 50.3 Ti, 50, 40, MLA-55

2.2 11 x 34 347356 TLS-55

3.0 13 x 32 344092 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti, 50, 40, MLA-55

4.0 11 x 60 344062* SW 60 Ti

4.0 13 x 41 344093 Types 90 Ti, 80 Ti, 70.1 Ti, 65, 50 Ti, 50, 40, MLA-55

5.0 13 x 51 344057* SW 55 Ti, SW 50.1, MLS-50

6.5 13 x 64 344088* Types 90 Ti, 80 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.4 Ti, 50.3 Ti, 50.2 Ti, 50 Ti,

4. Micro-UC
45 Ti, 40.3, 40.2, 40, 21, MLA-55

10.0 16 x 64 344089 Type 50

10.5 13 x 89 344087 Types 45 Ti, 21

13.2 14 x 89 344059* SW 41 Ti

13.5 16 x 76 344085 Types 90 Ti, 80 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.2 Ti, 50 Ti, 45 Ti, 40, 21, MLA-55

14.0 14 x 95 344060* SW 40 Ti

17.0 16 x 102 344061* SW 32.1 Ti, SW 28.1

38.5 25 x 89 344058* Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, SW 32 Ti, SW 28

94.0 38 x 102 345777 Types 45 Ti, 21

5. Tubes
Thinwall Polypropylene Tubes

175 μL 5 x 20 342630 TLS-55, A-110, A-100/30, A-100/18

2.2 11 x 34 347357 TLS-55

4.0 11 x 60 328874* SW 60 Ti

5.0 13 x 51 326819 SW 55 Ti, SW 50.1, MLS-50

6.5 13 x 64 326820 Types 90 Ti, 80 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.4 Ti, 50.3 Ti, 50.2 Ti, 50 Ti,
45 Ti, 40, 40.3, 40.2, 21, MLA-55, MLA-50

10.5 13 x 89 326822 Types 45 Ti, 21

6. Accessories
13.2 14 x 89 331372* SW 41 Ti

13.5 16 x 76 326814 Types 90 Ti, 80 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.2 Ti, 50 Ti, 45 Ti, 40, 30, 21, MLA-55, MLA-50

14.0 14 x 95 331374* SW 40 Ti

17.0 16 x 102 337986 SW 32.1 Ti, SW 28.1

35.5 25 x 83 344367 Type 70 Ti

38.5 25 x 89 326823* Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, SW 32 Ti, SW 28

94.0 38 x 102 345775 Types 45 Ti, 21

Thinwall Polypropylene, konical Tubes


7. Reference

1.5 11 x 35 358117 SW 60 Ti

3.0 13 x 51 358119 SW 55 Ti, SW 50.1 Ti, MLS-50

10.0 14 x 89 358120 SW 41 Ti, SW 40 Ti

14.5 16 x 102 358123 SW 32.1 Ti, SW 28.1

25.0 25 x 76 358125 SW 32 Ti, SW 28

30.0 25 x 89 358126* SW 32 Ti, SW 28


8. Support

* Sterile and Certified Free Options Available

Beckman Coulter Ultracentrifuges | beckman.com/centrifuges 5-7


1. Analytical UC

TUBES AND BOTTLES

Nominal Filling Nominal Size Rotors


Part No.
Capacity (mL) (mm) See specific rotor pages for required spacers/adapters/accessories
2. Floor Prep UC

Thickwall Polypropylene Tubes

230 μL 7 x 20 343621 Type 42.2 Ti, TLA-100

0.5 8 x 34 343777 TLA-120.1, TLA-100.1

1.0 11 x 34 347287 TLA-120.2, TLA-100.2, MLA-130, MLA-150, TLS-55

3.0 11 x 60 355636 SW 60 Ti

3.2 13 x 56 362333 TLA-110, TLA-100.4


3. Benchtop UC

3.5 13 x 51 349623 SW 55 Ti, SW 50.1, TLA-100.3, MLS-50

4.0 13 x 64 355644 "Types 90 Ti, 80 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.4 Ti, 50.3
Ti, 50.2 Ti, 50 Ti, 45 Ti, 40, 21, MLA-55, MLA-50

6.5 16 x 64 355646 Type 50, MLA-80

10.0 16 x 76 355640 "Types 90 Ti, 80 Ti, 75 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.2
Ti, 50 Ti,45 Ti, 42.1, 40, 35, 30, 21, SW 25.1, MLA-55, MLA-50

10.5 13 x 89 355639 Types 45 Ti, 21

32.0 25 x 89 355642 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, SW 32 Ti, SW 28
4. Micro-UC

94.0 38 x 102 355643 Types 45 Ti, 21

Thickwall Polycarbonate Tubes

230 μL 7 x 20 343775 Type 42.2 Ti, TLA-100, TLS-55

0.5 8 x 34 343776 TLA-120.1, TLA-100.1

1.0 8 x 51 355657 Types 50.4 Ti, 25

1.0 11 x 34 343778 TLA-120.2, TLA-100.2, MLA-130, MLA-150, TLS-55

3.0 11 x 60 355635 SW 60 Ti
5. Tubes

3.2 13 x 56 362305 TLA-110, TLA-100.4

3.5 13 x 51 349622 SW 55 Ti, SW 50.1, TLA-100.3, MLS-50

4.0 13 x 64 355645 "Types 90 Ti, 80 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.4 Ti,
50.3 Ti, 50.2 Ti, 50 Ti, 45 Ti, 40, 21, MLA-55, MLA-50

6.5 16 x 64 355647 Type 50, MLA-80

10.0 16 x 76 355630 "Types 90 Ti, 80 Ti, 75 Ti, 70.1 Ti, 70 Ti, 65, 60 Ti, 55.2 Ti, 50.2
6. Accessories

Ti, 50 Ti, 45 Ti, 40, 21, MLA-55, MLA-50

10.5 13 x 89 355629 Types 45 Ti, 21

32.0 25 x 89 355631 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, SW 32 Ti, SW 28

94.0 38 x 102 355628 Types 45 Ti, 21

Cellulose Propionate Tubes

230 μL 7 x 20 342303 Type 42.2 Ti

Stainless Steel Tubes


7. Reference

38.5 25 x 89 301112 Types 70 Ti, 60 Ti, 55.2 Ti, 50.2 Ti, 42.1,

94.0 38 x 102 303380 Types 45 Ti, 21

Polyethylene Tubes

175 μL 5 x 20 343622 TLS-55, A-110, A-100/30, A-100/18

Polypropylene Tubes with Snap-on Caps

1.5 11 x 40 357448 TLA-110, TLA-100.4, TLA-100.3, TLA-55, TLA-45


8. Support

(Natural)

5-8 Beckman Coulter Ultracentrifuges | beckman.com/centrifuges


1. Analytical UC

TUBES AND BOTTLES

Tube Cap Assemblies and Parts


2. Floor Prep UC

O-ring or
Cap Material Tube Cap Hex Nut Crown Set Screw Insert Stem
Gasket

Set Screw

Insert

5⁄16 in. (8mm) Hex Nut


Al 303624 303379 303809 - - 303730 303377
Washer
3. Benchtop UC

Al 303658 303379 303810 - - 303730 303377

1⁄2 in. (13 mm) Crown


Al 303113 301870 307004 - - 344672 307005
O-ring
Al C20178 301870 307004 803543 302312 344672 346246 or Gasket
Al/SS 305022 301870 307004 344672 302331

5⁄8 in. (16 mm) Stem

Al 303319 301870 307006 338864 302312 301869 302266

Al 330860 301870 330774 803543 302312 858046 330788 Tube


4. Micro-UC

Al 338907 301870 338911 338864 302312 878572 338910

Ti 341968 335320 335319 338864 302312 858046 341969

1 in. (25 mm)

Al 302359 301870 302169 338864 302312 301473 302168

Al 302133 301870 302169 338864 302312 301473 302138

Al C20182 330791 331153* 338864 302312 334280 331152

Al/Ti 337927 330791 338863* 338864 302312 - 338865

Al 338901 330791 338912* 338864 302312 878188 338908


5. Tubes

Al 338906 330791 338915* 338864 302312 878188 338908

11⁄2 in. (38 mm)

Al 326891 301870 326890 808482 302312 346242 326889

Al 326905 301870 326890 338864 302312 801761 326899

Al 330901 330791 330793* 338864 302312 346242 330900

Al 338903 330791 338914* 338864 302312 341767 338909


6. Accessories

Al 338905 330791 338913* 338864 302312 341767 338909

* Also order 330899 Washer.


7. Reference
8. Support

5-10 Beckman Coulter Ultracentrifuges | beckman.com/centrifuges


1. Analytical UC

SUPPORT SERVICES

Field Service Support Sometimes rotor corrosion is so bad it can be seen with the
naked eye, while at other times it is not so apparent. If you have
Most of us appreciate the need for careful design, handling, and
doubt about the condition of a single Beckman Coulter rotor, you
2. Floor Prep UC

maintenance of certain kinds of laboratory equipment—the optics of


may return it to the factory where it will be inspected free of charge
a spectrophotometer, for example. It may not be obvious, however,
by our metallurgists. The rotor will be examined by a variety of
that anything as substantial looking as a centrifuge rotor must
nondestructive methods, including ultrasound which is capable of
be properly maintained during use in order for it to fulfill its life
detecting internal flaws. Before shipping the rotor, you must contact
expectancy.
the nearest Beckman Coulter Sales and Service office for specific
If you consider that, during rotation, an Ultracentrifuge Rotor instructions. A written statement must accompany each rotor,
may experience more than 1,000,000 times the force of gravity, indicating that it is safe to handle, (i.e. free of any pathogenic or
it becomes apparent that seemingly minor flaws will assume radioactive contamination).
3. Benchtop UC

much greater significance at these g-forces. In effect, one gram


Rotor Safety Seminars
will “weigh” 1,000 kilograms, and a tiny flaw in a critical part
may generate stresses greater than the rotor was designed To help you get the most from your Beckman Coulter rotors, Rotor
to withstand. Safety Seminars are frequently presented by our Field Service
engineers. Seminars are designed to remind centrifuge users about
We at Beckman Coulter Life Sciences maintain stringent quality the importance of proper rotor care and maintenance.
standards and rigorously test every rotor design. Specific
instructions for care and maintenance are included in the rotor When you attend a Beckman Coulter Rotor Safety Seminar, you
manual that accompanies each rotor. And our Field Rotor Inspection will have an opportunity to ask questions that are specific to your
Program is available at no charge to all users of Beckman Coulter particular situation, and you will learn:
4. Micro-UC

Ultracentrifuges and High-Speed Centrifuges.


• What causes corrosion in a rotor, where it starts, what it does to
Field Rotor Inspection Program. What Is It? the rotor, and how to avoid it.

Our Field Rotor Inspection Program (FRIP) has two purposes: to • The proper way to clean and store rotors.
prevent premature rotor failures by detection of stress corrosion, • How to get longer life from your rotor.
metal fatigue, wear or damage to anodized coatings; and to instruct
laboratory personnel in the proper care of rotors. On-Site Rotor Inspection and Maintenance
Specially trained Beckman Coulter Field Service engineers are
Contact your local Beckman Coulter Service Office for details on
5. Tubes

available at your request for on-site inspection and minor rotor


the Field Rotor Inspection Program. Using nondestructive methods
maintenance. Rotors that are free of biological and radiological
(fiber optics borescopy and dyepenetrant analysis), they may find
contamination can be brought to these sessions for immediate
signs of corrosion or other damage. If so, they will recommend
inspection so you can continue operating your centrifuge with
repair or replacement, and a potentially costly failure may
confidence in the condition of your rotors.
be prevented.

To give a fuller understanding of rotors, a comprehensive slide


A Comprehensive Approach to
presentation is also offered. Centrifuge users are informed about
Centrifuge System Care
6. Accessories

maintenance procedures, rotor damage, and its significance. Throughout the world, Beckman Coulter Field Service engineers are
The stress corrosion of aluminum rotors, in particular, can be greatly on hand to provide the support you need. So you can operate your
reduced by good laboratory practice. centrifuges worry-free, a number of programs are in place that cover
preventive maintenance of your system, including parts, labor, drive
For more information about the Field Rotor Inspection Program systems, and rotors. It all adds up to a comprehensive approach to
(FRIP), please contact your Beckman Coulter Life Sciences maximizing your system investment.
representative.
Contact your local Beckman Coulter Sales and Service office to find
out the specifics about the program available in your area.
7. Reference
8. Support

8-1 Beckman Coulter Ultracentrifuges | beckman.com/centrifuges


ULTRACENTRIFUGATION
WITHOUT LIMITATION.
Starting with the innovations created by our
founder, Arnold O. Beckman Ph.D., Beckman
Coulter Life Sciences has lived at the forefront of
the technology that continues to help improve life
for all of humankind.

Our foray into the world of ultracentrifugation


is no different. In fact, we introduced the first
commercial ultracentrifuge in 1947. Since then,
our elite line of ultracentrifuges has attained
incredible levels of speed and sophistication,
without sacrificing usability.

Not unlike the rest of our portfolio of products,


our ultracentrifugation solutions certainly live up
to the Beckman Coulter name—a name in which
laboratories around the world continue to have
the utmost confidence.

You might also like