NIBP-1020 Series User Manual
NIBP-1020 Series User Manual
NIBP-1020 Series User Manual
BLOOD PRESSURE
SIMULATORS
NIBP-1030 NIBP-1030-BE
NIBP-1040-BE
NIBP-1020 NIBP-1010
USER MANUAL
BC BIOMEDICAL
NIBP-1000 SERIES
TABLE OF CONTENTS
DESCRIPTION ............................................................................................................. 1
OVERVIEW .................................................................................................................. 4
KEYS............................................................................................................................ 8
SCREENS .................................................................................................................... 9
SETUP ......................................................................................................................... 16
CONNECTIONS ........................................................................................................... 19
SPECIFICATIONS ....................................................................................................... 41
NOTES ......................................................................................................................... 45
i
WARNING - USERS
The NIBP-1000 Series is for use by skilled
technical personnel only.
WARNING - USE
The NIBP-1000 Series is intended for testing only
and should never be used in diagnostics,
treatment or any other capacity where it would
come in contact with a patient.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the NIBP-1000
Series. A serious hazard may occur if the patient
is connected when testing with the NIBP-1000
Series. Do not connect any leads from the
patient directly to the NIBP-1000 Series or DUT.
CAUTION - MODIFICATIONS
The NIBP-1000 Series is intended for use within
the published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.
CAUTION - ADAPTER
Turn Power Off and unplug any Battery
Eliminator before cleaning the surface
of the NIBP-1000 Series.
CAUTION - CLEANING
Do not immerse. The NIBP-1000 Series should
be cleaned by wiping gently with a damp, lint-free
cloth. A mild detergent can be used if desired.
ii
CAUTION - LIQUIDS
Do not submerge or spill liquids on the NIBP-
1000 Series. Do not operate the NIBP-1000
Series if internal components may have been
exposed to fluid.
CAUTION - SERVICE
The NIBP-1000 Series is intended to be serviced
only by authorized service personnel.
Troubleshooting and service procedures should
only be performed by qualified technical
personnel.
CAUTION - INSPECTION
The NIBP-1000 Series should be inspected before
each use for wear and should be serviced if any
parts are in question.
CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the NIBP-1000 Series. Allow the
NIBP-1000 Series to acclimate to specified
conditions for at least 30 minutes before
attempting to operate it.
iii
NOTICE – CE
This equipment has been type tested by an independent, accredited testing laboratory
and compliance was demonstrated to the above standard to the extent applicable.
EMISSIONS
Radiated and Line Conducted Emissions
IMMUNITY– CLASS C
EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and
laboratory use – General requirements”
This equipment has been type tested and compliance was demonstrated
to the above standard to the extent applicable.
iv
NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
Center Negative
NOTICE – ABBREVIATIONS
AC Alternating Current
BP Blood Pressure
BPM Beats per minute
brpm breaths per minute
C Celsius
° degree(s)
DUT Device Under Test
Dias Diastolic
ECG Electrocardiogram
Euro European
F Fahrenheit
HR Heart Rate
Hz hertz
kg kilogram
µV/V/mmHg microvolts(s) per Volt per mmHg
mA milliampere(s)
mm millimeter(s)
mmHg millimeter(s) of mercury
mV millivolt(s)
min minute(s)
Neo Neonatal
NSR Normal Sinus Rhythm
Ω ohm(s)
Lbs pounds
sec second(s)
Syst Systolic
USA United States of America
VDC Volts Direct Current
VRMS Volts Root Mean Square
v
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE
FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS.
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR
INACCURACIES.
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
[email protected]
vi
BC BIOMEDICAL
NIBP-1000 SERIES
NON-INVASIVE BLOOD PRESSURE SIMULATORS
The graphical display provides multiple screens containing pressure in mmHg, a plot of the
overall pressure, or a close-up of the BP waveform.
NIBP-1020 ADDS:
1
NIBP-1030 ADDS:
NIBP-1040 ADDS:
CARDIAC OUTPUT, FETAL/MATERNAL MODE
AVAILABLE MODELS:
NIBP-1030 BASIC UNIT WITH ECG, IBP, TEMP, PEAK & ALARM + MANY MORE
NIBP-1030-B + BATTERY
NIBP-1030-BE + UNIVERSAL PATIENT LEAD POST ECG CONNECTORS
NIBP-1030KIT NIBP-1030-B, 5 ADAPTERS & SOFT CARRYING CASE
NIBP-1030KIT-E NIBP-1030-BE, 5 ADAPTERS & SOFT CARRYING CASE
NIBP-1040KIT-E NIBP-1040-BE WITH ECG, IBP, TEMP, PEAK & ALARM, CARDIAC
OUPUT, FETAL/MATERNAL, 5 ADAPTERS & SOFT CARRYING
CASE
STANDARD ACCESSORIES:
OPTIONAL ACCESSORIES:
2
BC20-40602 NIBP FITTING KIT (HP CUFF ADAPTER,
INFLATION BULB, CPC CONNECTOR, QUICK
COUPLER WITH 1/8” MNPT AND SILICONE
TUBING AND TEE)
BC20-40605 NIBP ADAPTER KIT (11 CONNECTORS)
BC20-41333 UT-1 YSI-400 TEMPERATURE SIMULATION CABLE
BC20-41334 UT-2 YSI-700 TEMPERATURE SIMULATION CABLE
BC20-41534 FETAL ECG SCALP ELECTRODE CABLE
MSP-2100 PULSE OXIMETRY MODULE (ALLOWS NIBP TO
SIMULATE SpO2 OUTPUT WHEN USED WITH
FingerSims™)
FingerSim™ KIT COMPLETE STARTER KIT (INCLUDES 80%, 90% AND
97% FingerSims™, FingerSim™ HOLDER, AND
CARRYING CASE)
FingerSim™ SET REPLACEMENT KIT (80%, 90% AND 97%
FingerSims™ ONLY)
BC20-41337 RS-232 COMMUNICATIONS CABLE (MINI-DIN TO
DB-9F)
BC20-41339 USB COMMUNICATIONS ADAPTER (DB-9M to USB-A)
FOR USE WITH BC20-41337
BC20-41361 USB RS-232 COMMUNICATION CABLE (USB-A TO
MINI-DIN)
BC20-30111 SOFT CARRYING CASE
3
OVERVIEW
This section looks at the layout of the NIBP-1040-BE and gives descriptions of the
elements that are present.
7 pin Mini-Din
connector for
RS-232 or SpO2
Simulation
Power
2.1 mm Jack,
4
This section looks at the layout of the NIBP-1030 and gives descriptions of the elements
that are present.
NOTE: The NIBP-1030-BE model is different from the unit shown below in that it uses
universal patient lead banana posts that allow for the use of snaps and/or banana-type
patient leads.
LCD Graphical Display: Male Quick Patient Lead Connectors:
Shows Test Data and Connect for
Configuration Data Pressure Input Left Side Right Side
RA/R V2/C2
LA/L V3/C3
6 Light Touch Keys for RL/N V4/C4
Selecting Parameters LL/F V5/C5
and Settings: V1/C1 V6/C6
POWER for Turning Unit
On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options
(When No Options are
Selected, for Changing
Display Screen)
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu
RESET for resetting
peak pressure reading.
6 pin Mini-
Din
connector for
IBP
Simulation
8-Pin Mini-DIN
Plug Connector
for Temperature
Simulation
7 pin Mini-Din
connector for
RS-232 or SpO2
Simulation
Power
2.1 mm Jack,
5
This section looks at the layout of the NIBP-1020 and gives descriptions of the elements
that are present.
LCD Graphical
Display:
Shows Test Data and
Configuration Data Male Quick Patient Lead Connectors:
Connect for
Pressure Left Side Right Side
Input RA/R V2/C2
6 Light Touch Keys for LA/L V3/C3
Selecting Parameters RL/N V4/C4
and Settings: LL/F V5/C5
POWER for Turning V1/C1 V6/C6
Unit On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options
(When No Options are
Selected, for Changing
Display Screen)
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu
RESET for resetting
peak pressure reading
7 pin Mini-Din
connector for
RS-232 or SpO2
Simulation
Power
2.1 mm Jack
6
This section looks at the layout of the NIBP-1010 and gives descriptions of the elements
that are present.
LCD Graphical
Display:
Shows Test Data and
Configuration Data Male Quick
Connect for
Pressure
Input
6 Light Touch Keys for
Selecting Parameters
and Settings:
POWER for Turning Unit
On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options
(When No Options are
Selected, for Changing
Display Screen)
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu
RESET for resetting
peak pressure reading
7 pin Mini-Din
connector for
RS-232 and
SpO2 Simulation
Power
2.1 mm Jack
7
KEYS
– This key turns the unit off and on. The unit will return to the screen that was
– In the DISPLAY MODE, these keys toggle the display through the available
main screens.
In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through
– On the Main screen, this key sequences through the available NIBP or ECG
simulations. On the Setup screen, there are a number of parameters that may be selected
and changed. This key sequences the cursor (Highlight) through those parameters.
– This key is used to RESET the peak pressure reading or to begin a Leak Test.
– This key toggles the unit into and out of the Setup Mode. Depressing this key
will enter the Setup screen where the configuration can be viewed and adjusted.
Depressing the key again will exit the Setup Mode and return to the previously viewed
main screen.
8
SCREENS
MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output
Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).
PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below.
Also displayed on this screen is the peak pressure and selected output waveform.
PRESSURE
PEAK PRESSURE
(Optional)
PRESSURE
ENVELOPE
OFFSET
9
OUTPUT WAVEFORM SCREEN – This screen shows the pressure, peak pressure, output
NOTE
PRESSURE
PEAK PRESSURE
(Optional)
OUTPUT
WAVEFORM
(as generated by the
NIBP-1000 Series)
Dependant on
pressure
OUTPUT SELECTION
Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 80 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 & NIBP-1040)
Leak Test (NIBP-1030 & - - -
NIBP-1040)
Manometer - - -
10
PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as,
PRESSURE
PEAK PRESSURE
(Optional)
PRESSURE
GRAPH
2mmHg per pixel
OUTPUT SELECTION
Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 80 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 & NIBP-1040)
Leak Test (NIBP-1030 & - - -
NIBP-1040)
Manometer - - -
11
ECG OUTPUT SCREEN – This screen shows the selected ECG output mode.
NOTE: While in this mode, the NIBP simulation does not run.
NOTE: NSR ECG output is active during NIBP testing at the rate stipulated for the
selected test.
LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 & NIBP-1040
12
FETAL/MATERNAL SCREEN – This screen shows the select Fetal/Maternal output mode.
(NIBP-1040 ONLY)
NOTE: While in this mode, the NIBP simulation does not run.
IUP SIMULATION
`
MATERNAL
HEART
RATE FETAL
HEART
30 BPM RATE
40 BPM
45 BPM 60 BPM
60 BPM 90 BPM
80 BPM* 120 BPM*
90 BPM 140 BPM
100 BPM 150 BPM
120 BPM 210 BPM
140 BPM 240 BPM
160 BPM Next Contraction shows
180 BPM time remaining in NOTE: The FHR is only
200 BPM mm:ss until the next selectable when not using
220 BPM contraction begins. the IUP Wave Simulation.
240 BPM NOTE: The time is only * Indicates Default Setting When IUP Simulation is
displayed while running (See Power Up Settings) active, FHR is fixed at a
260 BPM
an IUP simulation with base rate of 140 BPM and
280 BPM an automatic trigger. then varies based on the
300 BPM IUP wave.
correlate with the monitoring equipment before simulation can begin. (See SETUP for
selection information).
13
CARDIAC OUTPUT SCREEN – This screen shows the select options and status of the
TEST TYPE
2.5 l/min*
5.0 l/min
10.0 l/min
Faulty Injection
L to R Shunt
Temp Cal Pulse
INJECTATE
TEMPERATURE
0 °C*
24 °C * Indicates Default Setting
TEST STATUS (See Power Up Settings)
Running
Stopped
14
BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage
When the level reaches 10%, the BP simulation mode will be disabled; however, the
Manometer and ECG (optional) will continue to function. Once the battery level reaches
Batteries charge from provided AC power adapter. While the batteries are charging, the
15
SETUP
The Setup Mode allows the user to adjust the configuration of the simulator. The Setup
screen can be entered using the key. The parameters can be changed by using
16
The following is a breakdown of the parameters available in the configuration of the unit
17
SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option
to shift test results to compensate for different methods of measuring Oscillometric NIBP
CAUTION
There are no absolute standards for Oscillometric NIBP readings; therefore, for a number
different method for evaluating the oscillometric pulses. Due to these varying methods,
precisely the same waveforms will give different results on different manufacturer’s units.
The normal technique used is to run the monitor against a fixed source like the NIBP-1000,
with the understanding that each manufacturer has a predictable error from this norm.
While this is generally the most direct method, users have asked for a method to correct for
this difference, making the monitors read the same as the test unit. The Systolic and
These adjustments are indicated in a line added to the main display to inform the user of
any shift that has been programmed into the system. This is done so there is no
18
CONNECTIONS
The NIBP Blood Pressure Simulators are connected between the blood pressure cuff and
the DUT (Patient Monitor). The blood pressure cuff should be disconnected from the DUT
and a ‘T’ adapter inserted between the cuff and the DUT. The NIBP simulator is then
Blood
Pressure DUT
Cuff
NIBP-
10xx
For models with ECG output, snaps or patient lead test posts (NIBP-1030-BE and NIBP-
1040-BE ONLY) are provided along the sides and are identified by the markings on the
overlay.
19
Connecting IBP Cables (NIBP-1030 & NIBP-1040 ONLY)
For models with IBP output, connect the DUT-appropriate IBP test cable between the NIBP
IBP Connector (6-pin Mini-DIN as pictured below) and the DUT. The pin-out shown below
for the Mini-DIN connector is the industry standard configuration used on most modern
patient simulators. As such, there are many existing cables that may be used or you may
20
Auxiliary Connector
The Auxiliary connector can be used for either standard RS-232 Serial communications
using an optional accessory cable or for SpO2 simulation using the optional accessory
21
THEORY OF OPERATIONS
A stepping motor and piston pressure engine is used to generate the output waveforms. A
precision pressure sensor is used to measure the pressure of the cuff. The pressure
The cuff pressure is monitored and the motor/piston assembly reacts to provide the correct
A Digital to Analog converter is used to generate the IBP waveforms. The output circuit is
fully isolated and capable of switching between the two standard sensitivities (5
Temperature
There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of
a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe
22
ECG Function
A Digital to Analog converter is used to generate the ECG outputs based on stored
Respiration Function
A Digital to Analog converter is used to generate the respiration waveform. The respiration
output is present on the LA ECG lead only. Refer to Monitor manual to determine which
lead the Monitor uses for respiration. It may be necessary to reverse the LA and LL leads
23
RUNNING A TEST
Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This
section will walk through each of the tests and their basic operation.
The Main Tests are accessible with a single key. The key will scroll through the
Adult 120/80
Adult 120/80 w/Pace
Adult High 190/120
Adult Low 80/40
Neonatal 70/40
Alarm Test
Arrhythmia Seq
Leak Test
Manometer
NIBP:
The first 5 test modes deal with various NIBP setups. To run an NIBP simulation, the cuff
and monitor are connected to the pressure input. Then the measurement is initiated by the
monitor and the NIBP Simulator will output the proper waveform based on the cuff
The NIBP output mode can be changed by pressing the key. Once the desired
operating mode is selected, the output will automatically begin when the correct pressure is
detected.
24
There are 5 selectable NIBP Basic Test Modes; Adult, Adult w/pace, Adult High, Adult Low
NOTE: The screens for the first test will have the individual components labeled. The
Adult 120/80
(Typical Screens)
PEAK PRESSURE
PRESSURE
(Optional)
PRESSURE ENVELOPE
NIBP TEST OFFSET
Type Adult 120/80 This line is only displayed
Heart Rate 80 if a non-zero value is
Respiration Rate 35 programmed into the
(Optional) Systolic and/or Diastolic
Shift. (See Setup for
(Pressure Only Screen) more information).
25
Adult 120/80 with Pace
(Typical Screens)
26
Adult Low 80/40
(Typical Screens)
Neonatal 70/40
(Typical Screens)
27
ALARM TEST:
Use the key to cycle through the Test Modes until Alarm Test appears. The
purpose of this function is to help test the alarms of the monitor under test. The ECG
output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR
with 80 brpm respiration. The time interval between alternations is 30 seconds. A count
down timer is displayed to indicate the time still remaining for the current output.
NOTE: While in this mode, the NIBP simulation output is disabled. Pressure readings in
the screens below are only manometer-type readings of static pressure applied.
TIMER
PRESSURE
ENVELOPE OFFSET
28
ARRHYTHMIA SEQUENCE: (NIBP-1030 & NIBP-1040 ONLY)
Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.
The purpose of this function is to provide a timed sequence of some of the more common
Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed
below. A count down timer is displayed to indicate the time still remaining in each step.
ARRHYTHMIA
29
LEAK TEST: (NIBP-1030 & NIBP-1040 ONLY)
Use the key to cycle through the Test Modes until Leak Test appears. The
purpose of this function is to provide a standard pressure leak test. The pressure input is
connected to the system that is to be monitored. Pressure is then applied to the system.
The Leak Test is initiated by pressing the key. The unit traces and displays the
pressure drop and the time since the test was initiated. The unit also calculates and
PRESSURE
PEAK PRESSURE
LEAK TEST
Drop in mmHg
Time in min:sec
Leak Rate in
mmHg/minute
30
MANOMETER:
Use the key to cycle through the Test Modes until Manometer appears. The
purpose of this function is to allow the unit to be used strictly as a Manometer with a full
NOTE: The Graphic Screens are also available in this mode for maximum flexibility.
PRESSURE
PEAK PRESSURE
(Optional)
PRESSURE ENVELOPE
OFFSET
This line is only displayed
if a non-zero value is
programmed into the
(Pressure Only Screen) Systolic and/or Diastolic
Shift. (See Setup for
more information).
31
SUB TEST MODES
PEAK: (Optional)
The optional Peak test feature is available whenever the display is in a pressure mode. It
will continuously monitor the pressure input and display and hold the maximum value. This
PEAK PRESSURE
(Optional)
PEAK PRESSURE
(Optional)
PEAK PRESSURE
(Optional)
32
ECG PERFORMANCE TEST: (NIBP-1020, NIBP-1030, & NIBP-1040 ONLY)
The ECG output is selected by pressing until the ECG Waveforms menu is
displayed. The key is used to choose the desired ECG output. There are 16
NOTE: While in this mode the NIBP simulation is disabled, but the IBP simulation is active
LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 & NIBP-1040
33
DYNAMIC IBP: (NIBP-1030 & NIBP-1040 ONLY)
The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG
Waveform mode.
the SETUP Mode before using IBP. (See Setup section for more information.)
A static Blood Pressure output may be selected in the SETUP mode. This fixed value will
the SETUP Mode before using IBP. (See Setup section for more information.)
34
TEMPERATURE SIMULATION: (NIBP-1030 & NIBP-1040 ONLY)
The temperature setting can be selected in the SETUP mode. This fixed value will remain
on the Temperature output until changed. The output will simulate both YSI 400 and YSI
700 Temperature probes.
NOTE: On some Patient Monitors, if ECG simulation is active on the NIBP-1030 and you
enable Temperature Simulation, the ECG trace may disappear from the Patient Monitor
screen. This is a flaw of the Patient Monitor, not the NIBP-1030 Simulator.
35
IUP SIMULATION: (NIBP-1040 ONLY)
The NIBP-1040 has the ability to simulate a combined Fetal/Maternal ECG. It assumes a
Fetal Scalp Electrode and a Maternal Thigh Electrode. Additionally, it produces a pressure
set to occur periodically. The Intrauterine-Pressure (IUP) curve causes a change in the
Fetal Heart Rate (FHR) based on the type of reaction selected. The IUP waveform is on
pressure transducer. The base 140 BPM FHR will respond to the contractions based on
36
The IUP Wave is a bell-shaped curve with a peak of 90 mmHg and duration of 90 seconds.
When active, the FHR and IUP displays continuously update to show the current output
If a Time is selected, the key is used to start a continuous cycle with a contraction
The key may be used at any point to cancel the simulation run.
37
CARDIAC OUTPUT TEST: (NIBP-1040 ONLY)
The NIBP-1040 has the ability to simulate Thermodilution Cardiac Output measurements.
Thermodilution allows the calculation of heart volume output by measuring the temperature
change of the blood after a specific volume of solution, which is room temperature or
cooler (typically 0 °C or 24 °C), is injected into the heart. The blood temperature is
curve that the Device Under Test (DUT, Cardiac Output Monitor) takes as a blood
temperature input.
The NIBP-1040 drives an external Cardiac Output module. This module (MCO-2100)
provides both a common connection point for the cabling and a manual simulation setting
for the injectate temperature. Abnormal waveforms are provided to simulate Injectate
Failure and Left-to-Right Shunt conditions. A special Calibration Pulse that puts out a
Setting Up a Test
1) The DUT (Cardiac Output Monitor) needs to be setup to match the NIBP-1040. The
3) Connect the Blood Temperature Sensor Line (BT Thermistor Cable) from the DUT to
38
4) Connect the Injectate Temperature Sensor Line for the DUT to the larger 4-pin
5) Turn on the DUT and the NIBP-1040. The DUT should show a blood temperature
of about 37 °C.
6) Adjust the trim pot on the MCO-2100 Module until the DUT shows the desired injectate
temperature (0 °C or 24 °C)
39
MANUAL REVISIONS
LIMITED WARRANTY
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
P:\Manuals\BCGroup\…\NIBP1000_UM_Rev16.doc
40
SPECIFICATIONS
41
PACEMAKER SIMULATION WAVEFORMS (NIBP-1020,
NIBP-1030 & NIBP-1040)
AMPLITUDE 3 mV ± 10%
WIDTH 6 ms ± 5%
42
CARDIAC OUTPUT (NIBP-1040)
INJECTATE VOLUME 10 cc
INJECTATE 0 or 24 °C
TEMPERATURE ACCURACY ± 2%
43
ELECTRICAL
12 VDC, 500 mA
AC POWER ADAPTER
BC20-21112 (Interchangeable)
Rechargeable NiMH pack
BATTERY (OPTIONAL)
(Not user serviceable)
BATTERY RUNTIME 500 Test Cycles between charges
44
NOTES
45
NOTES
46
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
[email protected]
Copyright © 2014
Made in the USA