NIBP-1020 Series User Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 56

NON-INVASIVE

BLOOD PRESSURE
SIMULATORS

NIBP-1030 NIBP-1030-BE

NIBP-1040-BE

NIBP-1020 NIBP-1010

USER MANUAL
BC BIOMEDICAL
NIBP-1000 SERIES
TABLE OF CONTENTS

WARNINGS, CAUTIONS, NOTICES ........................................................................... ii

DESCRIPTION ............................................................................................................. 1

OVERVIEW .................................................................................................................. 4

KEYS............................................................................................................................ 8

SCREENS .................................................................................................................... 9

SETUP ......................................................................................................................... 16

CONNECTIONS ........................................................................................................... 19

THEORY OF OPERATIONS ........................................................................................ 22

RUNNING A TEST ....................................................................................................... 24

MANUAL REVISIONS .................................................................................................. 40

LIMITED WARRANTY ................................................................................................. 40

SPECIFICATIONS ....................................................................................................... 41

NOTES ......................................................................................................................... 45

i
WARNING - USERS
The NIBP-1000 Series is for use by skilled
technical personnel only.

WARNING - USE
The NIBP-1000 Series is intended for testing only
and should never be used in diagnostics,
treatment or any other capacity where it would
come in contact with a patient.

WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the NIBP-1000
Series. A serious hazard may occur if the patient
is connected when testing with the NIBP-1000
Series. Do not connect any leads from the
patient directly to the NIBP-1000 Series or DUT.

CAUTION - MODIFICATIONS
The NIBP-1000 Series is intended for use within
the published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.

CAUTION - ADAPTER
Turn Power Off and unplug any Battery
Eliminator before cleaning the surface
of the NIBP-1000 Series.

CAUTION - CLEANING
Do not immerse. The NIBP-1000 Series should
be cleaned by wiping gently with a damp, lint-free
cloth. A mild detergent can be used if desired.

ii
CAUTION - LIQUIDS
Do not submerge or spill liquids on the NIBP-
1000 Series. Do not operate the NIBP-1000
Series if internal components may have been
exposed to fluid.

CAUTION - SERVICE
The NIBP-1000 Series is intended to be serviced
only by authorized service personnel.
Troubleshooting and service procedures should
only be performed by qualified technical
personnel.

CAUTION - INSPECTION
The NIBP-1000 Series should be inspected before
each use for wear and should be serviced if any
parts are in question.

CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the NIBP-1000 Series. Allow the
NIBP-1000 Series to acclimate to specified
conditions for at least 30 minutes before
attempting to operate it.

iii
NOTICE – CE

The NIBP-1000 Series Simulators bear the mark


based on the following testing standards:

ELECTROMAGNETIC COMPATIBILITY DIRECTIVE


EMC – Directive 89/336/EEC and 2004/108/EC as amended by
92/31/EEC, 93/68/EEC and Directive 91/263/EEC [ TTE/SES ]

EN 61326-1:1997 + A1:1998 + A2:2001 + A3:2003


“Electrical equipment for measurement, control and
laboratory use – EMC requirements”

This equipment has been type tested by an independent, accredited testing laboratory
and compliance was demonstrated to the above standard to the extent applicable.

EMISSIONS
Radiated and Line Conducted Emissions

EN 61000-3-2 Harmonic Current Emissions


EN 61000-3-3 Voltage Fluctuation and Flicker

IMMUNITY– CLASS C

EN 61000-4-2 Electrostatic Discharge


EN 61000-4-3 Radiated Electric Field Immunity
EN 61000-4-4 Electrical Fast Transients / Bursts
EN 61000-4-5 Surge Voltage
EN 61000-4-6 Conducted Disturbance
EN 61000-4-11 Voltage Dips and Short Interrupts

LOW VOLTAGE DIRECTIVE


EC – Directive 73/23/EC

EN 61010-1:2001
“Safety requirements for electrical equipment for measurement, control, and
laboratory use – General requirements”

This equipment has been type tested and compliance was demonstrated
to the above standard to the extent applicable.

iv
NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)

Center Negative

Per European Council Directive 2002/95/EC, do


not dispose of this product as unsorted
municipal waste.

NOTICE – ABBREVIATIONS
AC Alternating Current
BP Blood Pressure
BPM Beats per minute
brpm breaths per minute
C Celsius
° degree(s)
DUT Device Under Test
Dias Diastolic
ECG Electrocardiogram
Euro European
F Fahrenheit
HR Heart Rate
Hz hertz
kg kilogram
µV/V/mmHg microvolts(s) per Volt per mmHg
mA milliampere(s)
mm millimeter(s)
mmHg millimeter(s) of mercury
mV millivolt(s)
min minute(s)
Neo Neonatal
NSR Normal Sinus Rhythm
Ω ohm(s)
Lbs pounds
sec second(s)
Syst Systolic
USA United States of America
VDC Volts Direct Current
VRMS Volts Root Mean Square

v
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE
FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS.

NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR
INACCURACIES.

NOTICE – CONTACT INFORMATION


BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA

1-800-242-8428
1-314-638-3800

www.bcgroupintl.com
[email protected]

NIBP-1000 Series User Manual Copyright © 2014


www.bcgroupintl.com Made in the USA
08/14 Rev 16

vi
BC BIOMEDICAL
NIBP-1000 SERIES
NON-INVASIVE BLOOD PRESSURE SIMULATORS

The NIBP-1000 Series is a family of microprocessor-based, high-precision Non-Invasive


Blood Pressure (NIBP) Simulators. The units are small, easy to use and have multiple
features to fit many different applications. The NIBP-1010 is the basic model, which
includes built-in manometer function. The NIBP-1020 adds realistic, full QRS and
respiration ECG waveforms. The NIBP-1030 additionally offers Invasive Blood Pressure,
Temperature, Arrhythmias and a Leak Rate test mode. The NIBP-1040 adds
Fetal/Maternal ECG and an IUP waveform, as well as Cardiac Output simultaions.

The graphical display provides multiple screens containing pressure in mmHg, a plot of the
overall pressure, or a close-up of the BP waveform.

The following are highlights of the main features:

NIBP-1010 (BASIC FEATURES, NIBP ONLY):

 LARGE BACKLIT GRAPHICAL LCD DISPLAY WITH CURSOR SELECTION


OF OPTIONS AND PARAMETERS
 FULL RANGE MANOMETER
 ADULT, NEONATAL, HYPERTENSIVE, AND HYPOTENSIVE MODES
 ± 500 mmHg PRESSURE RANGE
 DIGITAL PRESSURE ENVELOPE OFFSET
 OPTIONAL PEAK PRESSURE DETECT WITH SIMPLE RESET
 SpO2 READY – COMPATIBLE WITH MSP-2100 MODULE
 DIGITAL CALIBRATION – NO POTS TO TURN
 SELECTABLE DISPLAY OPTIONS AND DIGIT SIZES
 SOFTWARE ADJUSTABLE CONTRAST
 FLASH PROGRAMMABLE
 BATTERY LIFE DISPLAY (0 TO 100%)
 OPTIONAL RECHARGEABLE NiMH BATTERIES
 BATTERY ELIMINATOR
 RS-232 INTERFACE

NIBP-1020 ADDS:

 ECG OUTPUT WITH FULL NSR WAVEFORM


 SINUSOIDAL RESPIRATION SIMULATION
 ECG TEST WAVEFORMS
 PACE WAVEFORM
 OPTIONAL PEAK PRESSURE DETECT WITH ECG ALARM TEST
 ECG SYNCHRONIZED WITH BLOOD PRESSURE

1
NIBP-1030 ADDS:

 SYNCHRONIZED INVASIVE BLOOD PRESSURE OUTPUT


 SELECTABLE IBP SENSITIVITY 5 OR 40 μV/V/mmHg
 STATIC IBP SIMULATION –10mmHg TO 400 mmHg
 LEAK RATE TEST
 ECG ARRHYTHMIA WAVEFORMS
 ECG ARRHYTHMIA SEQUENCE
 YSI 400 AND 700 SERIES TEMPERATURE SIMULATION

NIBP-1040 ADDS:
 CARDIAC OUTPUT, FETAL/MATERNAL MODE

AVAILABLE MODELS:

NIBP-1010 BASIC UNIT


NIBP-1010-P + PEAK
NIBP-1010-BP + BATTERY & PEAK
NIBP-1010KIT NIBP-1010-BP, 5 ADAPTERS & SOFT CARRYING CASE

NIBP-1020 BASIC UNIT WITH ECG


NIBP-1020-PA + PEAK & ALARM
NIBP-1020-BPA + BATTERY, PEAK & ALARM
NIBP-1020KIT NIBP-1020-BPA, 5 ADAPTERS & SOFT CARRYING CASE

NIBP-1030 BASIC UNIT WITH ECG, IBP, TEMP, PEAK & ALARM + MANY MORE
NIBP-1030-B + BATTERY
NIBP-1030-BE + UNIVERSAL PATIENT LEAD POST ECG CONNECTORS
NIBP-1030KIT NIBP-1030-B, 5 ADAPTERS & SOFT CARRYING CASE
NIBP-1030KIT-E NIBP-1030-BE, 5 ADAPTERS & SOFT CARRYING CASE

NIBP-1040KIT-E NIBP-1040-BE WITH ECG, IBP, TEMP, PEAK & ALARM, CARDIAC
OUPUT, FETAL/MATERNAL, 5 ADAPTERS & SOFT CARRYING
CASE

STANDARD ACCESSORIES:

 MCO-2100 CARDIAC OUTPUT MODULE (NIBP-1040 ONLY)


 BC20-21112 AC POWER ADAPTER (INTERCHANGEABLE)

OPTIONAL ACCESSORIES:

 BC20-40714 DINAMAP/CRITIKON ADAPTER


 BC20-40715 QUICK DISCONNECT ADAPTER
 BC20-40716 LUER ADAPTER
 BC20-40717 MARQUETTE ADAPTER
 BC20-40608 BULB ADAPTER

2
 BC20-40602 NIBP FITTING KIT (HP CUFF ADAPTER,
INFLATION BULB, CPC CONNECTOR, QUICK
COUPLER WITH 1/8” MNPT AND SILICONE
TUBING AND TEE)
 BC20-40605 NIBP ADAPTER KIT (11 CONNECTORS)
 BC20-41333 UT-1 YSI-400 TEMPERATURE SIMULATION CABLE
 BC20-41334 UT-2 YSI-700 TEMPERATURE SIMULATION CABLE
 BC20-41534 FETAL ECG SCALP ELECTRODE CABLE
 MSP-2100 PULSE OXIMETRY MODULE (ALLOWS NIBP TO
SIMULATE SpO2 OUTPUT WHEN USED WITH
FingerSims™)
 FingerSim™ KIT COMPLETE STARTER KIT (INCLUDES 80%, 90% AND
97% FingerSims™, FingerSim™ HOLDER, AND
CARRYING CASE)
 FingerSim™ SET REPLACEMENT KIT (80%, 90% AND 97%
FingerSims™ ONLY)
 BC20-41337 RS-232 COMMUNICATIONS CABLE (MINI-DIN TO
DB-9F)
 BC20-41339 USB COMMUNICATIONS ADAPTER (DB-9M to USB-A)
FOR USE WITH BC20-41337
 BC20-41361 USB RS-232 COMMUNICATION CABLE (USB-A TO
MINI-DIN)
 BC20-30111 SOFT CARRYING CASE

 INVASIVE BLOOD PRESSURE (IBP) CABLES:


FOR A COMPLETE LIST CONSULT THE BC GROUP WEBSITE AT
WWW.BCGROUPINTL.COM

3
OVERVIEW

This section looks at the layout of the NIBP-1040-BE and gives descriptions of the
elements that are present.

Patient Lead Connectors:

LCD Graphical Display: Male Quick Left Side Right Side


Shows Test Data and Connect for RA/R V2/C2
Configuration Data Pressure Input LA/L V3/C3
6 Light Touch Keys for RL/N V4/C4
Selecting Parameters LL/F V5/C5
and Settings: V1/C1 V6/C6
POWER for Turning Unit
On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options
(When No Options are
Selected, for Changing
Display Screen)
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu
RESET for resetting
peak pressure reading, 6 pin Mini-
starting and stopping Din
IUP/Cardiac Output connector for
Simulations IBP
Simulation
8-Pin Mini-DIN
Plug Connector for
Temperature and
Cardiac Output
Simulation

7 pin Mini-Din
connector for
RS-232 or SpO2
Simulation

Power
2.1 mm Jack,

4
This section looks at the layout of the NIBP-1030 and gives descriptions of the elements
that are present.

NOTE: The NIBP-1030-BE model is different from the unit shown below in that it uses
universal patient lead banana posts that allow for the use of snaps and/or banana-type
patient leads.
LCD Graphical Display: Male Quick Patient Lead Connectors:
Shows Test Data and Connect for
Configuration Data Pressure Input Left Side Right Side
RA/R V2/C2
LA/L V3/C3
6 Light Touch Keys for RL/N V4/C4
Selecting Parameters LL/F V5/C5
and Settings: V1/C1 V6/C6
POWER for Turning Unit
On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options
(When No Options are
Selected, for Changing
Display Screen)
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu
RESET for resetting
peak pressure reading.
6 pin Mini-
Din
connector for
IBP
Simulation
8-Pin Mini-DIN
Plug Connector
for Temperature
Simulation

7 pin Mini-Din
connector for
RS-232 or SpO2
Simulation

Power
2.1 mm Jack,

5
This section looks at the layout of the NIBP-1020 and gives descriptions of the elements
that are present.

LCD Graphical
Display:
Shows Test Data and
Configuration Data Male Quick Patient Lead Connectors:
Connect for
Pressure Left Side Right Side
Input RA/R V2/C2
6 Light Touch Keys for LA/L V3/C3
Selecting Parameters RL/N V4/C4
and Settings: LL/F V5/C5
POWER for Turning V1/C1 V6/C6
Unit On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options
(When No Options are
Selected, for Changing
Display Screen)
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu
RESET for resetting
peak pressure reading

7 pin Mini-Din
connector for
RS-232 or SpO2
Simulation

Power
2.1 mm Jack

6
This section looks at the layout of the NIBP-1010 and gives descriptions of the elements
that are present.

LCD Graphical
Display:
Shows Test Data and
Configuration Data Male Quick
Connect for
Pressure
Input
6 Light Touch Keys for
Selecting Parameters
and Settings:
POWER for Turning Unit
On and Off
UP and DOWN Arrows
for Scrolling Through
Selected Options
(When No Options are
Selected, for Changing
Display Screen)
SELECT for Choosing
Setting or Parameter to
Change
SETUP/RETURN for
Entering and Exiting
Setup Menu
RESET for resetting
peak pressure reading

7 pin Mini-Din
connector for
RS-232 and
SpO2 Simulation

Power
2.1 mm Jack

7
KEYS

Six tactile-touch keys are provided for system operation:

– This key turns the unit off and on. The unit will return to the screen that was

active when it was turned off.

– In the DISPLAY MODE, these keys toggle the display through the available

main screens.

In the SELECT MODE, if a parameter has been highlighted, these keys will scroll through

the available settings.

– On the Main screen, this key sequences through the available NIBP or ECG

simulations. On the Setup screen, there are a number of parameters that may be selected

and changed. This key sequences the cursor (Highlight) through those parameters.

– This key is used to RESET the peak pressure reading or to begin a Leak Test.

– This key toggles the unit into and out of the Setup Mode. Depressing this key

will enter the Setup screen where the configuration can be viewed and adjusted.

Depressing the key again will exit the Setup Mode and return to the previously viewed

main screen.

8
SCREENS

MAIN SCREENS – There are five main screens: Pressure Only, Pressure with Output

Waveform, Pressure with Pressure Graph, ECG (optional) and Battery Indicator (optional).

The available screens can be toggled using .

PRESSURE ONLY SCREEN – This screen has a large pressure display, as shown below.

Also displayed on this screen is the peak pressure and selected output waveform.

The display will resemble the following:

PRESSURE

PEAK PRESSURE
(Optional)

PRESSURE
ENVELOPE
OFFSET

This line is only


displayed if a non-
zero value is
programmed into
OUTPUT SELECTION the Systolic and/or
Diastolic Shift. (See
Respiration Setup for more
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional) information).
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 80 35
Adult High 190/120 150 80 35
Adult Low 80/40 53 80 35
Neonatal 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 & NIBP-1040)
Leak Test (NIBP-1030 & - - -
NIBP-1040)
Manometer - - -

9
OUTPUT WAVEFORM SCREEN – This screen shows the pressure, peak pressure, output

selection, and output waveform.

NOTE

This waveform is not intended to be


physiologically correct.

The display will resemble the following:

PRESSURE

PEAK PRESSURE
(Optional)

OUTPUT
WAVEFORM
(as generated by the
NIBP-1000 Series)

Dependant on
pressure

OUTPUT SELECTION

Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 80 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 & NIBP-1040)
Leak Test (NIBP-1030 & - - -
NIBP-1040)
Manometer - - -

10
PRESSURE GRAPH SCREEN – This screen provides a graph of the pressure, as well as,

the pressure, peak pressure and output selection.

PRESSURE
PEAK PRESSURE
(Optional)

PRESSURE
GRAPH
2mmHg per pixel

OUTPUT SELECTION

Respiration
Systolic/Diastolic (mean) Mean Heart Rate
Rate (Optional)
Adult 120/80 93 80 35
Adult 120/80 w/Pace 93 Pacemaker 80 35
High (Adult) 190/120 150 80 35
Low (Adult) 80/40 53 80 35
Neo (Neonatal) 70/40 55 94 47
Alarm Test (Optional) 47 00
-
(Alternates every 30 sec) 160 80
Arrhythmia Seq - 80 35
(NIBP-1030 & NIBP-1040)
Leak Test (NIBP-1030 & - - -
NIBP-1040)
Manometer - - -

11
ECG OUTPUT SCREEN – This screen shows the selected ECG output mode.

(NIBP-1020, NIBP-1030, & NIBP-1040 ONLY)

NOTE: While in this mode, the NIBP simulation does not run.

NOTE: NSR ECG output is active during NIBP testing at the rate stipulated for the

selected test.

The display will resemble the following:

LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 & NIBP-1040

12
FETAL/MATERNAL SCREEN – This screen shows the select Fetal/Maternal output mode.

(NIBP-1040 ONLY)

NOTE: While in this mode, the NIBP simulation does not run.

IUP SIMULATION

Uniform Deceleration* TRIGGER


Early Deceleration
Late Deceleration Manual*
Uniform Acceleration 2 Min
3 Min
5 Min

`
MATERNAL
HEART
RATE FETAL
HEART
30 BPM RATE
40 BPM
45 BPM 60 BPM
60 BPM 90 BPM
80 BPM* 120 BPM*
90 BPM 140 BPM
100 BPM 150 BPM
120 BPM 210 BPM
140 BPM 240 BPM
160 BPM Next Contraction shows
180 BPM time remaining in NOTE: The FHR is only
200 BPM mm:ss until the next selectable when not using
220 BPM contraction begins. the IUP Wave Simulation.
240 BPM NOTE: The time is only * Indicates Default Setting When IUP Simulation is
displayed while running (See Power Up Settings) active, FHR is fixed at a
260 BPM
an IUP simulation with base rate of 140 BPM and
280 BPM an automatic trigger. then varies based on the
300 BPM IUP wave.

NOTE: The Transducer Sensitivity (5 mV/V/mmHg or 40 mV/V/mmHg) must be set to

correlate with the monitoring equipment before simulation can begin. (See SETUP for

selection information).

13
CARDIAC OUTPUT SCREEN – This screen shows the select options and status of the

Cardiac Output Test. (NIBP-1040 ONLY)

TEST TYPE

2.5 l/min*
5.0 l/min
10.0 l/min
Faulty Injection
L to R Shunt
Temp Cal Pulse

INJECTATE
TEMPERATURE

0 °C*
24 °C * Indicates Default Setting
TEST STATUS (See Power Up Settings)

Running
Stopped

14
BATTERY INDICATOR SCREEN – This screen shows the status of the battery voltage

level with Built-In Battery Option only.

NOTE: It is only an estimate of the battery life remaining.

When the level reaches 10%, the BP simulation mode will be disabled; however, the

Manometer and ECG (optional) will continue to function. Once the battery level reaches

0%, the unit will automatically turn itself off.

The display will resemble the following:

Batteries charge from provided AC power adapter. While the batteries are charging, the

display will flash “charging.”

15
SETUP

The Setup Mode allows the user to adjust the configuration of the simulator. The Setup

screen can be entered using the key. The parameters can be changed by using

key to highlight the line and to toggle the available options.

The Setup screen can be exited using the key.

NIBP-1010 and NIBP-1020

NIBP-1030 and NIBP-1040

16
The following is a breakdown of the parameters available in the configuration of the unit

and their available options:

System Setup Configuration

Parameter Description Range


Adjusts the Systolic Output of the NIBP
Systolic Shift Simulation. This is not a direct mmHg ±50.0
adjustment.
Adjusts the Diastolic Output of the NIBP
Diastolic Shift Simulation. This is not a direct mmHg ±50.0
adjustment.
Static BP
(NIBP-1030 & Adjusts the Static Blood Pressure Output. -10 to 400 mmHg
NIBP-1040)
IBP Sens
Selects the Invasive Blood Pressure Output 5 μV/V/mmHg
(NIBP-1030 & sensitivity. 40 μV/V/mmHg
NIBP-1040)
Temp
0, 24, 30, 35, 37, 40,
(NIBP-1030 & Selects the simulated Temperature output.
42 C
NIBP-1040)
This parameter determines whether the SpO2
SpO2 Output output pulse is active. The output drives an Off/On
MSP-2100 FingerSim Module.
Determines the period of inactivity before the unit is
turned OFF. A timer is started when the unit is
turned ON and is reset each time a key is pressed.
When the timer reaches the value set in this
Auto Off Timer parameter, the power is automatically turned OFF. 0-30 Minutes
(NOTE: Setting this parameter to 0 disables the
Auto Off timer. When running from line power, the
unit does not automatically shut off. Auto Off timer
is inactive during a test.)
Contrast Adjust Sets the contrast of the display screen. 0-20
Off – Always off
1-30 sec – The elapsed time after which the backlight
Backlight will automatically turn off. Off, 1-30 sec, ON
ON – Always ON.
The Default setting is 30 seconds.
Available only with Battery Option installed.
Displays current life of the battery. 0-100%
Battery Life At 10%, a warning screen will appear. (Read Only)
At 0%, the unit will power down automatically.
Software Displays current software program. (Read Only)

17
SYSTOLIC AND DIASTOLIC SHIFT – The NIBP-1000 Series is equipped with the option

to shift test results to compensate for different methods of measuring Oscillometric NIBP

by various manufacturers and models of devices under test.

CAUTION

These adjustments must be used with caution as they will


allow the user to adjust the output results to invalid values.

These adjustments should only be used to aid in the


simplification of testing and with documented controls.

There are no absolute standards for Oscillometric NIBP readings; therefore, for a number

of reasons (including patents, technology, etc.), each manufacturer has established a

different method for evaluating the oscillometric pulses. Due to these varying methods,

precisely the same waveforms will give different results on different manufacturer’s units.

The normal technique used is to run the monitor against a fixed source like the NIBP-1000,

with the understanding that each manufacturer has a predictable error from this norm.

While this is generally the most direct method, users have asked for a method to correct for

this difference, making the monitors read the same as the test unit. The Systolic and

Diastolic Shift settings allow for just such correction.

These adjustments are indicated in a line added to the main display to inform the user of

any shift that has been programmed into the system. This is done so there is no

misunderstanding of the meaning of the results.

18
CONNECTIONS

Connecting Pressure Cuffs

The NIBP Blood Pressure Simulators are connected between the blood pressure cuff and

the DUT (Patient Monitor). The blood pressure cuff should be disconnected from the DUT

and a ‘T’ adapter inserted between the cuff and the DUT. The NIBP simulator is then

connected to the open side of the ‘T’ adapter as pictured below:

Blood
Pressure DUT
Cuff
NIBP-
10xx

NOTE: Roll the cuff tightly on itself or around a mandrel.

Connecting Patient Leads

For models with ECG output, snaps or patient lead test posts (NIBP-1030-BE and NIBP-

1040-BE ONLY) are provided along the sides and are identified by the markings on the

overlay.

19
Connecting IBP Cables (NIBP-1030 & NIBP-1040 ONLY)

For models with IBP output, connect the DUT-appropriate IBP test cable between the NIBP

IBP Connector (6-pin Mini-DIN as pictured below) and the DUT. The pin-out shown below

for the Mini-DIN connector is the industry standard configuration used on most modern

patient simulators. As such, there are many existing cables that may be used or you may

obtain the proper cable from BC Group.

NOTE: Mini-DIN pinout viewed from NIBP Connector Side

Connecting Temperature Simulation Cables (NIBP-1030 & NIBP 1040 ONLY)


For the models with Temperature output, connect the DUT-appropriate Temperature
Simulation test cable between the NIBP Temperature Connector (8-pin Mini-DIN as
pictured below) and the DUT.

NOTE: Mini-DIN pinout viewed from NIBP Connector Side

20
Auxiliary Connector

The Auxiliary connector can be used for either standard RS-232 Serial communications

using an optional accessory cable or for SpO2 simulation using the optional accessory

MSP-2100 SpO2 module available from BC Biomedical.

NOTE: Mini-DIN pinout viewed from NIBP Connector Side

21
THEORY OF OPERATIONS

NIBP (Non-Invasive Blood Pressure) Function

A stepping motor and piston pressure engine is used to generate the output waveforms. A

precision pressure sensor is used to measure the pressure of the cuff. The pressure

sensor is read by a 16-bit differential Analog to Digital converter.

The cuff pressure is monitored and the motor/piston assembly reacts to provide the correct

response based on the selected mode.

IBP (Invasive Blood Pressure) Function

A Digital to Analog converter is used to generate the IBP waveforms. The output circuit is

fully isolated and capable of switching between the two standard sensitivities (5

µV/V/mmHg and 40 µV/V/mmHg).

Temperature

There is an 8-pin mini-DIN plug connector on the right side of the unit for connection of

a Temperature cable. Temperatures are simulated for both YSI 400 and YSI 700 probe

types. There are seven different temperatures selectable for each.

22
ECG Function

A Digital to Analog converter is used to generate the ECG outputs based on stored

waveform data for standard ECG and Arrhythmias.

Respiration Function

A Digital to Analog converter is used to generate the respiration waveform. The respiration

output is present on the LA ECG lead only. Refer to Monitor manual to determine which

lead the Monitor uses for respiration. It may be necessary to reverse the LA and LL leads

for respiration to be detected.

23
RUNNING A TEST

Depending on the model, there are up to 9 Basic Test Modes and 3 Sub-Test Modes. This

section will walk through each of the tests and their basic operation.

The Main Tests are accessible with a single key. The key will scroll through the

following tests in a continuous loop:

Adult 120/80
Adult 120/80 w/Pace
Adult High 190/120
Adult Low 80/40
Neonatal 70/40
Alarm Test
Arrhythmia Seq
Leak Test
Manometer

BASIC TEST MODES

NIBP:

The first 5 test modes deal with various NIBP setups. To run an NIBP simulation, the cuff

and monitor are connected to the pressure input. Then the measurement is initiated by the

monitor and the NIBP Simulator will output the proper waveform based on the cuff

pressure provided by the monitor and the selected simulation.

The NIBP output mode can be changed by pressing the key. Once the desired

operating mode is selected, the output will automatically begin when the correct pressure is

detected.

24
There are 5 selectable NIBP Basic Test Modes; Adult, Adult w/pace, Adult High, Adult Low

and Neonatal. The displays will resemble the following examples:

NOTE: The screens for the first test will have the individual components labeled. The

component labels for subsequent tests are the same.

Adult 120/80
(Typical Screens)
PEAK PRESSURE
PRESSURE
(Optional)

PRESSURE ENVELOPE
NIBP TEST OFFSET
Type Adult 120/80 This line is only displayed
Heart Rate 80 if a non-zero value is
Respiration Rate 35 programmed into the
(Optional) Systolic and/or Diastolic
Shift. (See Setup for
(Pressure Only Screen) more information).

PRESSURE PEAK PRESSURE


(Optional)

NIBP TEST OUTPUT


WAVEFORM
Type Adult 120/80 (as generated by the
Heart Rate 80 NIBP-1000 Series)
Respiration Rate 35
(Optional) Dependant on
pressure
(Output Waveform Screen)

PRESSURE PEAK PRESSURE


(Optional)

NIBP TEST PRESSURE


GRAPH
Type Adult 120/80
Heart Rate 80 2mmHg per
Respiration Rate 35 pixel
(Optional)

(Pressure Graph Screen)

25
Adult 120/80 with Pace
(Typical Screens)

Adult High 120/80


(Typical Screens)

26
Adult Low 80/40
(Typical Screens)

Neonatal 70/40
(Typical Screens)

27
ALARM TEST:

Use the key to cycle through the Test Modes until Alarm Test appears. The

purpose of this function is to help test the alarms of the monitor under test. The ECG

output will alternate from a 47 BPM NSR with apnea respiration (0 brpm) to 160 BPM NSR

with 80 brpm respiration. The time interval between alternations is 30 seconds. A count

down timer is displayed to indicate the time still remaining for the current output.

NOTE: While in this mode, the NIBP simulation output is disabled. Pressure readings in

the screens below are only manometer-type readings of static pressure applied.

The displays will resemble the following examples: (Typical Screens)


PRESSURE PEAK PRESSURE
(NIBP Simulation disabled, (NIBP Simulation disabled,
this is pressure of the static this is peak pressure of the
volume only) static volume only)

TIMER

PRESSURE
ENVELOPE OFFSET

ALARM TEST This line is only


displayed if a non-zero
Heart Rate 47/160 value is programmed
Respiration Rate 0/80 into the Systolic and/or
Diastolic Shift. (See
(Pressure Only Screen) Setup for more
information).

(Output Waveform Screen) (Pressure Graph Screen)

28
ARRHYTHMIA SEQUENCE: (NIBP-1030 & NIBP-1040 ONLY)

Use the key to cycle through the Test Modes until Arrhythmia Sequence appears.

The purpose of this function is to provide a timed sequence of some of the more common

Arrhythmias. The test will continually cycle through the 6 arrhythmias in the sequence listed

below. A count down timer is displayed to indicate the time still remaining in each step.

NOTE: While in this mode, the NIBP simulation output is disabled.

Step Display Total Time ECG Output


1 Single PVCs 18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC
2 Pair of PVCs 9 sec 10 NSR, Pair of PVCs
3 Single PVCs 18 sec 11 NSR, 1 PVC, 11 NSR, 1 PVC
4 Run of 5 PVCs 12 sec 11 NSR, Run of 5 PVCs
5 Vtach 30 sec 15 NSR, VTach
6 Vfib 30 sec 15 NSR, VFib
7 Asystole 30 sec 15 NSR, Asystole

The displays will resemble the following examples: (Typical Screens)


PRESSURE PEAK PRESSURE
(NIBP Simulation disabled, (NIBP Simulation disabled,
this is pressure of the static this is peak pressure of the
volume only) static volume only)
TIMER

ARRHYTHMIA

Single PVCs PRESSURE ENVELOPE


Pair of PVCs OFFSET
Single PVCs
Run of 5 PVCs This line is only displayed
Vtach if a non-zero value is
Vfib programmed into the
Asystole Systolic and/or Diastolic
(Pressure Only Screen) Shift. (See Setup for
more information).

(Output Waveform Screen) (Pressure Graph Screen)

29
LEAK TEST: (NIBP-1030 & NIBP-1040 ONLY)

Use the key to cycle through the Test Modes until Leak Test appears. The

purpose of this function is to provide a standard pressure leak test. The pressure input is

connected to the system that is to be monitored. Pressure is then applied to the system.

The Leak Test is initiated by pressing the key. The unit traces and displays the

pressure drop and the time since the test was initiated. The unit also calculates and

displays the leak rate in mmHg per minute.

NOTE: While in this mode, the NIBP simulation output is disabled.

The displays will resemble the following examples: (Typical Screens)

PRESSURE

PEAK PRESSURE
LEAK TEST

Drop in mmHg
Time in min:sec
Leak Rate in
mmHg/minute

(Pressure Only Screen)

(Output Waveform Screen) (Pressure Graph Screen)

30
MANOMETER:

Use the key to cycle through the Test Modes until Manometer appears. The

purpose of this function is to allow the unit to be used strictly as a Manometer with a full

scale range of ± 500 mmHg.

NOTE: The Graphic Screens are also available in this mode for maximum flexibility.

The displays will resemble the following examples: (Typical Screens)

PRESSURE
PEAK PRESSURE
(Optional)

PRESSURE ENVELOPE
OFFSET
This line is only displayed
if a non-zero value is
programmed into the
(Pressure Only Screen) Systolic and/or Diastolic
Shift. (See Setup for
more information).

(Output Waveform Screen) (Pressure Graph Screen)

31
SUB TEST MODES

PEAK: (Optional)

The optional Peak test feature is available whenever the display is in a pressure mode. It

will continuously monitor the pressure input and display and hold the maximum value. This

value may be reset at any time by pressing the key.

The displays will resemble the following examples: (Typical Screens)

PEAK PRESSURE
(Optional)

PEAK PRESSURE
(Optional)

PEAK PRESSURE
(Optional)

32
ECG PERFORMANCE TEST: (NIBP-1020, NIBP-1030, & NIBP-1040 ONLY)

The ECG output is selected by pressing until the ECG Waveforms menu is

displayed. The key is used to choose the desired ECG output. There are 16

fixed waveforms available.

NOTE: While in this mode the NIBP simulation is disabled, but the IBP simulation is active

(NIBP-1030 & NIBP-1040 ONLY).

The display will resemble the following example:

LIST OF AVAILABLE
OUTPUTS
SELECTED
OUTPUT NSR 30 Bpm
NSR 60 Bpm
NSR 120 Bpm
NSR 240 Bpm
Sine Wave 10 Hz
Sine Wave 60 Hz
Sine Wave 100 Hz
Square Wave 0.125 Hz
Square Wave 2 Hz
Triangle Wave 2 Hz
PVC Arrhythmia*
Pair of PVC’s*
Run of 5 PVC’s*
Ventricular Tachycardia*
Ventricular Fibrillation*
Asystole*
*NIBP-1030 & NIBP-1040

33
DYNAMIC IBP: (NIBP-1030 & NIBP-1040 ONLY)

The IBP output matches the NIBP setting in the NIBP mode or the ECG setting in the ECG

Waveform mode.

NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in

the SETUP Mode before using IBP. (See Setup section for more information.)

STATIC IBP: (NIBP-1030 & NIBP-1040 ONLY)

A static Blood Pressure output may be selected in the SETUP mode. This fixed value will

remain on the IBP output until changed.

NOTE: The IBP sensitivity (5 µV/V/mmHg or 40 µV/V/mmHg) must be correctly selected in

the SETUP Mode before using IBP. (See Setup section for more information.)

34
TEMPERATURE SIMULATION: (NIBP-1030 & NIBP-1040 ONLY)

The temperature setting can be selected in the SETUP mode. This fixed value will remain
on the Temperature output until changed. The output will simulate both YSI 400 and YSI
700 Temperature probes.

NOTE: Both outputs are available at the output connector simultaneously.

NOTE: On some Patient Monitors, if ECG simulation is active on the NIBP-1030 and you

enable Temperature Simulation, the ECG trace may disappear from the Patient Monitor

screen. This is a flaw of the Patient Monitor, not the NIBP-1030 Simulator.

35
IUP SIMULATION: (NIBP-1040 ONLY)

The NIBP-1040 has the ability to simulate a combined Fetal/Maternal ECG. It assumes a

Fetal Scalp Electrode and a Maternal Thigh Electrode. Additionally, it produces a pressure

waveform to simulate uterine contractions. The waveform may be triggered manually or

set to occur periodically. The Intrauterine-Pressure (IUP) curve causes a change in the

Fetal Heart Rate (FHR) based on the type of reaction selected. The IUP waveform is on

Blood Pressure Channel 1 (BP1) and simulates an Intra-Amniotic catheter connected to a

pressure transducer. The base 140 BPM FHR will respond to the contractions based on

the setting as follows:

IUP Simulation Type FHR Response


The FHR goes from 140 BPM down to 100
BPM and back again. The rate follows the
Uniform Deceleration
90 seconds bell curve of the contraction but
is delayed by 30 seconds.
The FHR goes from 140 BPM down to 100
Early Deceleration BPM and back again. The rate follows the
90 seconds bell curve of the contraction.
The FHR goes from 140 BPM down to 100
BPM and back again. The rate follows the
Late Deceleration
90 seconds bell curve of the contraction but
is delayed by 45 seconds.
The FHR goes from 140 BPM up to 175
Uniform Acceleration BPM and back again. The rate follows the
90 seconds bell curve of the contraction.

36
The IUP Wave is a bell-shaped curve with a peak of 90 mmHg and duration of 90 seconds.

When active, the FHR and IUP displays continuously update to show the current output

values. The Trigger is set by selecting Manual, 2, 3 or 5 minutes.

If Manual is selected, the key is used to start one contraction.

If a Time is selected, the key is used to start a continuous cycle with a contraction

IUP wave starting every 2, 3 or 5 minutes.

The key may be used at any point to cancel the simulation run.

37
CARDIAC OUTPUT TEST: (NIBP-1040 ONLY)

The NIBP-1040 has the ability to simulate Thermodilution Cardiac Output measurements.

Thermodilution allows the calculation of heart volume output by measuring the temperature

change of the blood after a specific volume of solution, which is room temperature or

cooler (typically 0 °C or 24 °C), is injected into the heart. The blood temperature is

measured by a thermistor on the catheter. The NIBP-1040 simulation provides an output

curve that the Device Under Test (DUT, Cardiac Output Monitor) takes as a blood

temperature input.

The NIBP-1040 drives an external Cardiac Output module. This module (MCO-2100)

provides both a common connection point for the cabling and a manual simulation setting

for the injectate temperature. Abnormal waveforms are provided to simulate Injectate

Failure and Left-to-Right Shunt conditions. A special Calibration Pulse that puts out a

1.5 °C drop in the temperature is also provided.

Setting Up a Test

1) The DUT (Cardiac Output Monitor) needs to be setup to match the NIBP-1040. The

following settings are required:

Catheter Baxter Edwards 93a – 131 – 7f


0.542 for 0 °C Injectate
Calibration Coefficient
0.595 for 24 °C Injectate
Injectate Volume 10 cc
Injectate Temperature 0 °C or 24 °C

2) Connect MCO-2100 Module to the CO/Temp port on the NIBP-1040.

3) Connect the Blood Temperature Sensor Line (BT Thermistor Cable) from the DUT to

the small 4-pin connector on the MCO-2100.

38
4) Connect the Injectate Temperature Sensor Line for the DUT to the larger 4-pin

connector on the MCO-2100.

5) Turn on the DUT and the NIBP-1040. The DUT should show a blood temperature

of about 37 °C.

6) Adjust the trim pot on the MCO-2100 Module until the DUT shows the desired injectate

temperature (0 °C or 24 °C)

39
MANUAL REVISIONS

Revision # Program # Revisions Made


Rev 01 DT7355CA Origination
Rev 02 DT7355CA Specification Information Updated
Rev 03 DT7355CA Misc. Edits
Rev 04 DT7355CF Add Systolic/Diastolic adjustments
Rev 05 DT7355CG Output Selection edited
Rev 06 DT7355CI Misc. Edits
Rev 07 DT7355CI Warnings, Cautions, Notices Updated
Rev 08 DT7359CA NIBP-1030, IBP, ECG Arrhythmia Waveforms Added
Rev 09 DT7359CA Temperature Added
Rev 10 DT7359CA Temp Connector Updated, Static 120 Added
Rev 11 DT7355CO, Address Updated, Screen Captures Updated, Output
DT7359CH Selection Charts Updated
Rev 12 DT7355CO, Misc. Edits
DT7359CH
Rev 13 DT7355CO, Added NIBP-1030-BE model, revised accessories
DT7359CH listing, and other miscellaneous edits.
Rev 14 DT7355CO, Format Updated, Misc. Edits
DT7359CH
Rev 15 DT7355CO, Misc. Edits
DT7359CH
Rev 16 DT7355CO,
DT7359CH,
DT735BCA NIBP-1040-BE, Fetal/Maternal, Cardiac Output Added

LIMITED WARRANTY

WARRANTY: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE


FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.

EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR


IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE.

BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL


DAMAGES.

NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR


ASSUME ANY LIABILITY.

REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.

P:\Manuals\BCGroup\…\NIBP1000_UM_Rev16.doc
40
SPECIFICATIONS

MANOMETER (PRESSURE MEASUREMENT)


RANGE ± 500 mmHg @ 20 °C
ACCURACY ± (1% of Reading + 0.5 mmHg @ 20 °C)

NON-INVASIVE BLOOD PRESSURE (NIBP) SIMULATION


80, 94 BPM
RATE
(NOTE: Synchronized to ECG on NIBP-1020/1030)
ACCURACY ± 1%

INVASIVE BLOOD PRESSURE (IBP) SIMULATION (NIBP-1030 & 1040)


-10, -5, 0, 20, 40, 50, 60, 80, 100, 120, 150, 160, 200,
STATIC PRESSURE
240, 250, 300, 320, 400 mmHg
± (1% Full Range + 1 mmHg) or
ACCURACY
± (2% Setting + 2 mmHg) whichever is better
IMPEDANCE 300 Ω, ± 10%
EXCITATION RANGE 2 to 16 VRMS
EXCITATION FREQUENCY DC to 5 kHz
SENSITIVITY 5 or 40 µV/V/mmHg

ECG SIMULATION (NIBP-1020, NIBP-1030 & NIBP-1040)


NSR RATE 30, 60, 120, 240 BPM
SINE RATE 10, 60, 100 Hz
PERFORMANCE
SQUARE RATE 0.125, 2 Hz
WAVEFORMS
TRIANGLE RATE 2 Hz
RATE ACCURACY ± 1%
AMPLITUDE 2.75 mV ± 2% @ Lead II

41
PACEMAKER SIMULATION WAVEFORMS (NIBP-1020,
NIBP-1030 & NIBP-1040)
AMPLITUDE 3 mV ± 10%
WIDTH 6 ms ± 5%

RESPIRATION SIMULATION (NIBP-1020, NIBP-1030 & NIBP-1040)


RATE ACCURACY ± 1%
DELTA 3.0 Ω ± 10%
IMPEDANCE
BASELINE 1000 Ω ± 5%

TEMPERATURE SIMULATION (NIBP-1030 & NIBP-1040)


SENSOR TYPE YSI Series 400 and 700
0, 24, 30, 35, 37, 40, 42 °C
POINTS
(32, 75.2, 86, 95, 98.6, 104, 107.6 °F)
ACCURACY ± 0.1 °C

FETAL / MATERNAL SIMULATION (NIBP-1040)


FETAL HEART RATE 60, 90, 120, 140, 150, 210, 240 BPM
Uniform Deceleration
INTER UTERINE
Early Deceleration
PRESSURE RESPONSE
Late Deceleration
TYPES
Uniform Acceleration
IUP WAVE Bell Curve with 90 mmHg Peak and 90 second Width
Manual
IUP TRIGGER
Auto: 2, 3, 5 minutes

42
CARDIAC OUTPUT (NIBP-1040)
INJECTATE VOLUME 10 cc

INJECTATE 0 or 24 °C
TEMPERATURE ACCURACY ± 2%

INJECTATE CALIBRATION 0 °C 0.542


COEFFICIENT 24 °C 0.595
37 °C (98.6 °F)
BLOOD TEMPERATURE
ACCURACY ± 2%
2.5, 5.0, 10.0 L/min,
CARDIAC OUTPUT
ACCURACY ± 5%
Normal Flow
Faulty Injectate
SIMULATIONS
Left-to-Right Shunt
Temperature Calibration Pulse
TEMPERATURE 1.5 °C Down for 1 sec
CALIBRATION PULSE ACCURACY ± 1%
CATHETER TYPE Baxter Edwards, 93a-131-7f

43
ELECTRICAL
12 VDC, 500 mA
AC POWER ADAPTER
BC20-21112 (Interchangeable)
Rechargeable NiMH pack
BATTERY (OPTIONAL)
(Not user serviceable)
BATTERY RUNTIME 500 Test Cycles between charges

PHYSICAL & ENVIRONMENTAL


128 X 64 Pixels Graphical LCD,
DISPLAY
White LED Backlight
NIBP-1000 through 7.70 x 5.36 x 4.07 Inches
NIBP-1030 (195.6 x 136.1 x 103.4 mm)
SIZE
NIBP-1030-BE & 7.70 x 6.15 x 4.07 Inches
NIBP-1040-BE (195.6 x 156.2 x 103.4 mm)
WEIGHT < 3.5 Lbs (1.6 kg)
ENCLOSURE ABS Plastic
CONSTRUCTION
OVERLAY Back-Printed Lexan
15 to 40 °C
OPERATING RANGE
(59 to 104 °F)
-20 to 65 °C
STORAGE RANGE
(-4 to 149 °F)

44
NOTES

45
NOTES

46
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA

1-800-242-8428
1-314-638-3800

www.bcgroupintl.com
[email protected]

NIBP-1000 Series User Manual


08/14 – Rev 16

Copyright © 2014
Made in the USA