ESU-2000A UM Rev08
ESU-2000A UM Rev08
ESU-2000A UM Rev08
ANALYZER
ESU-2000A
USER MANUAL
BC BIOMEDICAL
ESU-2000A
TABLE OF CONTENTS
DESCRIPTION ............................................................................................................. 1
OVERVIEW .................................................................................................................. 3
SPECIFICATIONS ......................................................................................................... 7
NOTES ........................................................................................................................... 9
i
WARNING - USERS
The ESU-2000A is for use by skilled technical
personnel only.
WARNING - USE
The ESU-2000A is intended for testing only and
should never be used in diagnostics, treatment
or any other capacity where it could come in
contact with a patient.
WARNING - USE
Never touch exposed metal surfaces on test
leads or other current-carrying parts while the
DUT is activated.
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the ESU-2000A. A
serious hazard may occur if the patient is
connected when testing with the ESU-2000A. Do
not connect any leads from the patient directly to
the ESU-2000A or DUT.
WARNING - MODIFICATIONS
The ESU-2000A is intended for use within the
published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.
ii
WARNING - LIQUIDS
Do not submerge or spill liquids on the ESU-
2000A. Do not operate the ESU-2000A if internal
components not intended for use with fluids may
have been exposed to fluid, as the internal
leakage may have caused corrosion and be a
potential hazard.
CAUTION - INSPECTION
The ESU-2000A should be inspected before each
use for wear and should be serviced if any parts
are in question.
CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the ESU-2000A. Allow ESU-2000A
to acclimate to specified conditions for at least
30 minutes before attempting to operate it.
CAUTION - SERVICE
The ESU-2000A is intended to be serviced only
by authorized service personnel.
Troubleshooting and service procedures should
only be performed by qualified technical
personnel. Never open the ESU-2000A while
connected to the DUT, as a hazardous condition
may exist.
CAUTION - FUSE
Only replace the ESU-2000A fuse with the
specified type and rating.
CAUTION - VENTILATION
The ESU-2000A includes ventilation slots to help
prevent overheating during operation and should
not be blocked.
iii
CAUTION - CLEANING
Do not immerse. The ESU-2000A should be
cleaned by wiping gently with a damp, lint-free
cloth. A mild detergent can be used if desired.
NOTICE SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
NOTICE ABBREVIATIONS
A Ampere(s)
C Celsius
Degree(s)
DUT Device Under Test
FS Full Scale
Hz Hertz
kg kilogram(s)
lbs pounds
M Mega- (106)
MHz Megahertz
m milli- (10-3)
mA milliampere(s)
mm millimeter(s)
ohm(s)
RF Radio Frequency
RMS Root Mean Square
s second(s)
USA United States of America
W Watt(s)
iv
NOTICE DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE
FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS.
NOTICE DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR
INACCURACIES.
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
[email protected]
v
This Page Intentionally Left Blank
vi
BC BIOMEDICAL
ESU-2000A
ELECTROSURGICAL UNIT ANALYZER
The Model ESU-2000A is a precision Electro-Surgery Unit Analyzer (ESU). It tests the RF
power output of Electro-Surgery Units. The power reading is taken from a large 4- inch
analog display. It offers six resistive loads for RF power testing from 50 to 500 ohms,
along with a 200 ohm load for RF leakage current testing. The patient return electrode
resistance simulation is adjustable in one ohm increments from 0 to 999 ohms.
1
STANDARD ACCESSORIES:
OPTIONAL ACCESSORIES:
2
OVERVIEW
This section looks at the layout of the ESU-2000A and gives descriptions of the elements
that are present.
Oscilloscope
Output
Carrying
Handle
RF Leakage
Input Terminal
ESU Ground RF
Leakage Terminal
3
OPERATING INSTRUCTIONS
2. Remove all connections between the DUT and the ESU-2000A Analyzer.
3. Install the Switching Module on the ESU-2000A front panel so that the words ESU
Power Testing are right-side-up. When installed correctly, the Active Electrode
connectors should be exposed, and the RF Leakage connectors should be
inaccessible.
4. Connect a single electrode REM cable from the ESU-2000A terminal labeled
Patient Plate to the DUTs Patient Plate or Return Electrode.
5. Connect a yellow safety lead from the DUTs RF Active Output to the desired load
terminal of the ESU-2000A. The Active Electrode input connectors are color coded,
and each corresponds to a color-coded range on the analog meter.
6. Set the DUT for proper output mode and power level, being careful that the power
setting does not exceed the full-scale Power range selected on the ESU-2000A.
7. Activate the DUT output. The Analyzer will indicate both mA RMS and Power (W)
on the appropriate color-coded wattage meter range.
8. Deactivate the DUT output before disconnecting from ESU-2000A, or before moving
the Active Electrode test lead to a different input range connector.
1. Remove all connections between the DUT and the ESU-2000A Analyzer.
2. Install the Switching Module on the ESU-2000A front panel so that the words RF
Leakage Testing are right-side-up. When installed correctly, the Active Electrode
connectors should be inaccessible, and the RF Leakage connectors should be
exposed.
3. Connect a yellow safety lead from the ESU-2000A terminal labeled RF Leakage to
the DUTs RF Active Output. DO NOT MEASURE ACTIVE ELECTRODE LEAKAGE
ON NON-ISOLATED ESUs with the ESU-2000A. Damage may occur to the ESU-
2000A.
4
4. Connect a green safety test lead from the ESU-2000A terminal labeled EARTH to
the DUTs Earth Ground; an alligator clip with safety jack may be necessary. If no
accessible Earth Ground point exists on the DUT, it is acceptable in most cases to
use Receptacle Earth.
5. Set the DUT for proper output mode and power level.
6. Activate the DUT output. The Analyzer will indicate the RF Leakage current from
the DUT Active Electrode on the color-colored RF Leakage scale.
8. Repeat Active Output RF Leakage Testing for all available Output Modes and
Output Electrodes as called for in the DUT Leakage Test Procedures.
9. Remove the yellow safety lead from the DUTs RF Active Output and the ESU-
2000A.
10. Connect the single electrode REM cable to the DUTs Patient Plate or Return
Electrode and the Leakage Input on the ESU-2000A.
11. Set the DUT for proper output mode and power level.
12. Activate the DUT output. The Analyzer will indicate the RF Leakage current from
the DUT Active Electrode on the color-colored RF Leakage scale.
14. Repeat Return Output RF Leakage Testing for all available Output Modes and
Output Electrodes as called for in the DUT Leakage Test Procedures.
1. Remove all connections between the DUT and the ESU-2000A Analyzer.
2. Connect the dual electrode REM cable from the ESU-2000A terminals labeled
PATIENT PLATE to the DUT Patient Plate or Return Electrode.
5
MANUAL REVISIONS
LIMITED WARRANTY
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
6
SPECIFICATIONS
RF MEASUREMENT
CURRENT 0 to 1000 mA RMS
RANGE
POWER 0 to 500 W
CURRENT 2% FS
ACCURACY
POWER 5% FS
FREQUENCY 0 to 10 MHz
FUSE 500 mA Fast-Blow FWH-1/2
Isolated
OUTPUT
OSCILLOSCOPE OUTPUT Uncalibrated
CONNECTION BNC (50 )
50
100
200
VALUES
300
400
500
TYPE Non-Inductive
ACCURACY 2% 1
VARIABLE LOAD 50 50 W
100 100 W
7
RETURN ELECTRODE MONITOR TEST LOAD
RANGE 0 to 999
RESOLUTION 1
ACCURACY 3% + 3
CONNECTIONS 4 mm Banana
8
NOTES
9
NOTES
10
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
[email protected]
Copyright 2016
Made in the USA