ESU-2000A UM Rev08

Download as pdf or txt
Download as pdf or txt
You are on page 1of 20

ELECTROSURGICAL UNIT

ANALYZER

ESU-2000A

USER MANUAL
BC BIOMEDICAL
ESU-2000A
TABLE OF CONTENTS

WARNINGS, CAUTIONS, NOTICES ............................................................................ ii

DESCRIPTION ............................................................................................................. 1

OVERVIEW .................................................................................................................. 3

OPERATING INSTRUCTIONS .................................................................................... 4

MANUAL REVISIONS .................................................................................................... 6

LIMITED WARRANTY ................................................................................................... 6

SPECIFICATIONS ......................................................................................................... 7

NOTES ........................................................................................................................... 9

i
WARNING - USERS
The ESU-2000A is for use by skilled technical
personnel only.

WARNING - USE
The ESU-2000A is intended for testing only and
should never be used in diagnostics, treatment
or any other capacity where it could come in
contact with a patient.

WARNING - USE
Never touch exposed metal surfaces on test
leads or other current-carrying parts while the
DUT is activated.

WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the ESU-2000A. A
serious hazard may occur if the patient is
connected when testing with the ESU-2000A. Do
not connect any leads from the patient directly to
the ESU-2000A or DUT.

WARNING - MODIFICATIONS
The ESU-2000A is intended for use within the
published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.

ii
WARNING - LIQUIDS
Do not submerge or spill liquids on the ESU-
2000A. Do not operate the ESU-2000A if internal
components not intended for use with fluids may
have been exposed to fluid, as the internal
leakage may have caused corrosion and be a
potential hazard.

CAUTION - INSPECTION
The ESU-2000A should be inspected before each
use for wear and should be serviced if any parts
are in question.

CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the ESU-2000A. Allow ESU-2000A
to acclimate to specified conditions for at least
30 minutes before attempting to operate it.

CAUTION - SERVICE
The ESU-2000A is intended to be serviced only
by authorized service personnel.
Troubleshooting and service procedures should
only be performed by qualified technical
personnel. Never open the ESU-2000A while
connected to the DUT, as a hazardous condition
may exist.

CAUTION - FUSE
Only replace the ESU-2000A fuse with the
specified type and rating.

CAUTION - VENTILATION
The ESU-2000A includes ventilation slots to help
prevent overheating during operation and should
not be blocked.
iii
CAUTION - CLEANING
Do not immerse. The ESU-2000A should be
cleaned by wiping gently with a damp, lint-free
cloth. A mild detergent can be used if desired.

NOTICE SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)

NOTICE ABBREVIATIONS
A Ampere(s)
C Celsius
Degree(s)
DUT Device Under Test
FS Full Scale
Hz Hertz
kg kilogram(s)
lbs pounds
M Mega- (106)
MHz Megahertz
m milli- (10-3)
mA milliampere(s)
mm millimeter(s)
ohm(s)
RF Radio Frequency
RMS Root Mean Square
s second(s)
USA United States of America
W Watt(s)

NOTICE PERFORMING TESTS


REFER TO DUT MANUFACTURERS SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS .

iv
NOTICE DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE RESPONSIBLE
FOR ANY INJURIES SUSTAINED DUE TO UNAUTHORIZED
EQUIPMENT MODIFICATIONS OR APPLICATION OF EQUIPMENT
OUTSIDE OF THE PUBLISHED INTENDED USE AND
SPECIFICATIONS.

NOTICE DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL HAS
BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED FOR
INACCURACIES.

NOTICE CONTACT INFORMATION


BC BIOMEDICAL
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DR.
SAINT CHARLES, MO 63301
USA

1-800-242-8428
1-314-638-3800

www.bcgroupintl.com
[email protected]

ESU-2000A User Manual Copyright 2016


www.bcgroupintl.com Made in the USA
04/16 Rev 08

v
This Page Intentionally Left Blank

vi
BC BIOMEDICAL
ESU-2000A
ELECTROSURGICAL UNIT ANALYZER

The Model ESU-2000A is a precision Electro-Surgery Unit Analyzer (ESU). It tests the RF
power output of Electro-Surgery Units. The power reading is taken from a large 4- inch
analog display. It offers six resistive loads for RF power testing from 50 to 500 ohms,
along with a 200 ohm load for RF leakage current testing. The patient return electrode
resistance simulation is adjustable in one ohm increments from 0 to 999 ohms.

The following are highlights of some of the main features:

LARGE 4 INCH ANALOG DISPLAY


2% OF FULL SCALE CURRENT ACCURACY
0-10 MHz BANDWIDTH
RETURN CONTACT MONITOR TEST
RF LEAKAGE TEST
ISOLATED OSCILLOSCOPE OUTPUT
PORTABLE
EASY OPERATION
NO BATTERIES OR POWER SUPPLY REQUIRED
STANDARD ACCESSORY KIT INCLUDED
HIGH IMPACT PLASTIC ENCLOSURE

1
STANDARD ACCESSORIES:

BC20 00121 ACCESSORY KIT (INCLUDES TEST LEADS,


ALLIGATOR CLIPS, AND REPLACEMENT FUSE)

OPTIONAL ACCESSORIES:

BC20 30108 SOFT CARRYING CASE

2
OVERVIEW

This section looks at the layout of the ESU-2000A and gives descriptions of the elements
that are present.

4- INCH Patient Plate Return Electrode


ANALOG Monitor Test Terminals
DISPLAY

Fast Blow Potentiometer


Fuse Ohms

Oscilloscope
Output

Carrying
Handle

RF Leakage
Input Terminal

ESU Ground RF
Leakage Terminal

ESM-1 Switching Module


RF Leakage Testing
or
ESU Power Testing

3
OPERATING INSTRUCTIONS

NOTICE PERFORMING TESTS


REFER TO DUT MANUFACTURERS SERVICE MANUAL FOR
TEST PROCEDURES AND MEASUREMENT LIMITS .

DUT RF Power Measurements

1. Consult the DUT manufacturers manual for load specifications.

2. Remove all connections between the DUT and the ESU-2000A Analyzer.

3. Install the Switching Module on the ESU-2000A front panel so that the words ESU
Power Testing are right-side-up. When installed correctly, the Active Electrode
connectors should be exposed, and the RF Leakage connectors should be
inaccessible.

4. Connect a single electrode REM cable from the ESU-2000A terminal labeled
Patient Plate to the DUTs Patient Plate or Return Electrode.

5. Connect a yellow safety lead from the DUTs RF Active Output to the desired load
terminal of the ESU-2000A. The Active Electrode input connectors are color coded,
and each corresponds to a color-coded range on the analog meter.

6. Set the DUT for proper output mode and power level, being careful that the power
setting does not exceed the full-scale Power range selected on the ESU-2000A.

7. Activate the DUT output. The Analyzer will indicate both mA RMS and Power (W)
on the appropriate color-coded wattage meter range.

8. Deactivate the DUT output before disconnecting from ESU-2000A, or before moving
the Active Electrode test lead to a different input range connector.

DUT RF Leakage Current Measurements

1. Remove all connections between the DUT and the ESU-2000A Analyzer.

2. Install the Switching Module on the ESU-2000A front panel so that the words RF
Leakage Testing are right-side-up. When installed correctly, the Active Electrode
connectors should be inaccessible, and the RF Leakage connectors should be
exposed.

3. Connect a yellow safety lead from the ESU-2000A terminal labeled RF Leakage to
the DUTs RF Active Output. DO NOT MEASURE ACTIVE ELECTRODE LEAKAGE
ON NON-ISOLATED ESUs with the ESU-2000A. Damage may occur to the ESU-
2000A.

4
4. Connect a green safety test lead from the ESU-2000A terminal labeled EARTH to
the DUTs Earth Ground; an alligator clip with safety jack may be necessary. If no
accessible Earth Ground point exists on the DUT, it is acceptable in most cases to
use Receptacle Earth.

5. Set the DUT for proper output mode and power level.

6. Activate the DUT output. The Analyzer will indicate the RF Leakage current from
the DUT Active Electrode on the color-colored RF Leakage scale.

7. Deactivate the DUT output.

8. Repeat Active Output RF Leakage Testing for all available Output Modes and
Output Electrodes as called for in the DUT Leakage Test Procedures.

9. Remove the yellow safety lead from the DUTs RF Active Output and the ESU-
2000A.

10. Connect the single electrode REM cable to the DUTs Patient Plate or Return
Electrode and the Leakage Input on the ESU-2000A.

11. Set the DUT for proper output mode and power level.

12. Activate the DUT output. The Analyzer will indicate the RF Leakage current from
the DUT Active Electrode on the color-colored RF Leakage scale.

13. Deactivate the DUT output.

14. Repeat Return Output RF Leakage Testing for all available Output Modes and
Output Electrodes as called for in the DUT Leakage Test Procedures.

DUT Patient Plate or Return Electrode Monitor Test

1. Remove all connections between the DUT and the ESU-2000A Analyzer.

2. Connect the dual electrode REM cable from the ESU-2000A terminals labeled
PATIENT PLATE to the DUT Patient Plate or Return Electrode.

3. Adjust ESU-2000A Patient Plate Resistance as necessary to verify DUT Patient


Plate or Return Electrode Monitor mode as specified by the DUT manufacturer.

5
MANUAL REVISIONS

Revision # Revisions Made

Rev 01 New Manual Format


Rev 02 Accessory Fuse Updated
Rev 03 Notices Updated
Rev 04 Specifications Updated
Rev 05 Format Updated, Misc. Edits
Rev 06 Format Updated
Rev 07 Specifications Updated
Rev 08 Accessory Kit Updated

LIMITED WARRANTY

WARRANTY: BC GROUP INTERNATIONAL, INC. WARRANTS ITS NEW PRODUCTS TO BE FREE


FROM DEFECTS IN MATERIALS AND WORKMANSHIP UNDER THE SERVICE FOR WHICH THEY
ARE INTENDED. THIS WARRANTY IS EFFECTIVE FOR TWELVE MONTHS FROM THE DATE OF
SHIPMENT.

EXCLUSIONS: THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTY EXPRESSED OR


IMPLIED, INCLUDING, BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY
OR FITNESS FOR A PARTICULAR PURPOSE.

BC GROUP INTERNATIONAL, INC. IS NOT LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL


DAMAGES.

NO PERSON OTHER THAN AN OFFICER IS AUTHORIZED TO GIVE ANY OTHER WARRANTY OR


ASSUME ANY LIABILITY.

REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR OR
REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE OPTION
OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.

6
SPECIFICATIONS

RF MEASUREMENT
CURRENT 0 to 1000 mA RMS
RANGE
POWER 0 to 500 W
CURRENT 2% FS
ACCURACY
POWER 5% FS
FREQUENCY 0 to 10 MHz
FUSE 500 mA Fast-Blow FWH-1/2
Isolated
OUTPUT
OSCILLOSCOPE OUTPUT Uncalibrated
CONNECTION BNC (50 )
50
100
200
VALUES
300
400
500
TYPE Non-Inductive
ACCURACY 2% 1
VARIABLE LOAD 50 50 W
100 100 W

POWER RATING 200 200 W


(MAX) 300 300 W
400 400 W
500 500 W
DUTY CYCLE 10 seconds on, 20 seconds off
VALUE 200
TYPE Non-Inductive
LEAKAGE LOAD
ACCURACY 2% 1
POWER RATING 50 W
CONNECTIONS 4 mm Banana

7
RETURN ELECTRODE MONITOR TEST LOAD
RANGE 0 to 999
RESOLUTION 1
ACCURACY 3% + 3
CONNECTIONS 4 mm Banana

PHYSICAL & ENVIRONMENTAL


4-1/2 Analog Meter
DISPLAY
Color-coded display (W and mA)
ENCLOSURE ABS Plastic
CONSTRUCTION
FACE PLATE Back-printed Lexan
5.36 x 8.68 x 9.10 Inches
SIZE
(136.1 x 220.5 x 231.1 mm)
WEIGHT 4 Lbs (1.82 kg)
15 to 30 C (59 to 86 F)
OPERATING RANGE
20 to 80% RH, Non-Condensing
STORAGE RANGE -40 to 60 C (-40 to 140 F)

8
NOTES

9
NOTES

10
BC GROUP INTERNATIONAL, INC.
3081 ELM POINT INDUSTRIAL DRIVE
ST. CHARLES, MO 63301
USA

1-800-242-8428
1-314-638-3800

www.bcgroupintl.com
[email protected]

ESU-2000A User Manual


04/16 Rev 08

Copyright 2016
Made in the USA

You might also like