ULT-2000 Series User Manual
ULT-2000 Series User Manual
ULT-2000 Series User Manual
LEAKAGE TESTER
ULT-2000 SERIES
USER MANUAL
BC BIOMEDICAL
ULT-2000 SERIES
TABLE OF CONTENTS
DESCRIPTION ............................................................................................................... 1
BACKGROUND .............................................................................................................. 4
TEST PROCESS............................................................................................................. 8
LAYOUT ......................................................................................................................... 9
KEYS............................................................................................................................. 10
SCREENS .................................................................................................................... 11
SETUP .......................................................................................................................... 18
PC SOFTWARE ............................................................................................................ 21
LIMITED WARRANTY................................................................................................... 26
SPECIFICATIONS ........................................................................................................ 27
NOTES .......................................................................................................................... 29
i
CALIBRATION INTERVAL
To ensure the accuracy of the ULT-2000 Series,
BC Group International, Inc. recommends
calibration at least once every 12 months.
Calibration must be done by qualified personnel.
Contact BC Group International, Inc. for
calibration.
PREVENTIVE MAINTENANCE
To ensure the proper operation of the ULT-PA-XX
Series, and ULT-PC-XX Series, BC Group
International, Inc. recommends preventive
maintenance at least once every 12 months.
Maintenance must be done by qualified
personnel. Contact BC Group International, Inc.
for Maintenance.
WARNING - USERS
The ULT-2000 is for use by
skilled technical personnel only.
WARNING - USE
The ULT-2000 is intended for testing only and
should never be used in diagnostics, treatment
or any other capacity where it would come in
contact with a patient.
WARNING - MODIFICATIONS
The ULT-2000 is intended for use within the
published specifications. Any application
beyond these specifications or any unauthorized
user modifications may result in hazards or
improper operation.
ii
WARNING - CONNECTIONS
All connections to patients must be removed
before connecting the DUT to the ULT-2000. A
serious hazard may occur if the patient is
connected when testing with the ULT-2000.
Do not connect any leads from the patient
directly to the ULT-2000 or DUT.
WARNING - LIQUIDS
Do not submerge or spill liquids on the ULT-
2000.
Do not operate the ULT-2000 if internal
components may have been exposed to fluid.
WARNING - VOLTAGE
High Voltages are generated by the ULT-2000
when running tests. Do not touch any surface
that is in contact with or connected to the ULT-
2000, including the ultrasound transducer,
adapter, conductivity probe, basin or liquid
medium contained in the basin, as it might be at
an electrical potential of 90 to 275 VAC.
iii
CAUTION - SERVICE
The ULT-2000 is intended to be serviced only by
authorized service personnel. Troubleshooting
and service procedures should only be
performed by qualified technical personnel.
CAUTION - ENVIRONMENT
Exposure to environmental conditions outside
the specifications can adversely affect the
performance of the ULT-2000. Allow ULT-2000 to
acclimate to specified conditions for at least 30
minutes before attempting to operate it.
CAUTION - CLEANING
Do not immerse. The ULT-2000 should be
cleaned by wiping gently with a damp, lint-free
cloth. A mild detergent can be used if desired.
CAUTION - INSPECTION
The ULT-2000 should be inspected before each
use for obvious signs of abuse or wear. The
ULT-2000 should not be used and should be
serviced if any parts are in question.
iv
v
NOTICE – SYMBOLS
Symbol Description
Caution
(Consult Manual for Further Information)
Electrical Caution
(Consult Manual for Further Information)
Center Negative
(Refers to Battery Eliminator Connector)
NOTICE – ABBREVIATIONS
ANSI American National Standards Institute
C Celsius
° degree(s)
DC Direct Current
DUT Device Under Test
Euro European
FS Full Scale
Hz Hertz
kg kilogram(s)
µA microampere(s)
mA milliampere(s)
mm millimeter(s)
NEDA National Electronic Distributors Association
Ω ohm(s)
PC Personal Computer
Lbs pounds
sec second(s)
TEE Transesophageal Echocardiography
ULT Ultrasound (Transducer) Leakage Tester
USA United States of America
V Volt(s)
VAC Volts Alternating Current
VDC Volts Direct Current
vi
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. WILL NOT BE
RESPONSIBLE FOR ANY INJURIES SUSTAINED DUE TO
UNAUTHORIZED EQUIPMENT MODIFICATIONS OR
APPLICATION OF EQUIPMENT OUTSIDE OF THE PUBLISHED
INTENDED USE AND SPECIFICATIONS.
NOTICE – DISCLAIMER
BC GROUP INTERNATIONAL, INC. RESERVES THE RIGHT TO
MAKE CHANGES TO ITS PRODUCTS OR SPECIFICATIONS AT
ANY TIME, WITHOUT NOTICE, IN ORDER TO IMPROVE THE
DESIGN OR PERFORMANCE AND TO SUPPLY THE BEST
POSSIBLE PRODUCT. THE INFORMATION IN THIS MANUAL
HAS BEEN CAREFULLY CHECKED AND IS BELIEVED TO BE
ACCURATE. HOWEVER, NO RESPONSIBILITY IS ASSUMED
FOR INACCURACIES.
1-800-242-8428
1-314-638-3800
www.bcgroupintl.com
[email protected]
vii
BC BIOMEDICAL
ULT-2000 SERIES
ULTRASOUND TRANSDUCER LEAKAGE TESTER
1
ULT-2020 (METER, DATALOG)
STANDARD ACCESSORIES:
OPTIONAL ACCESSORIES:
2
CONDUCTIVITY PROBES: (not included, order separately)
ULT-PC-10 DUAL CONDUCTIVITY PROBE (Short)
ULT-PC-15 DUAL CONDUCTIVITY PROBE (Medium)
ULT-PC-20 DUAL CONDUCTIVITY PROBE (Short & Long)
ULT-PC-25 FOR USE WITH GUS CLEANING SYSTEM
ULT-PC-30 DUAL CONDUCTIVITY PROBE (Flexible)
GE YMS/RT ULT-PA-27
GE: 4C-D, 6VT-D, Voluson E8, Vivid E9
GE / Samsung ULT-PA-30
Samsung: RS80A, HS70A
HI VISION 900, 5500, 6500, 8500
Hitachi ULT-PA-21
EUB-2000, EUB-525, EUB-405 Plus
Philips iE33 and iU22 diagnostic TEE - S7-2
Philips / ATL / Mindray (TEE), S7-3t (TEE), S3-1, C8-4v, C9-5, et all with ULT-PA-17
bellhousing, Mindray DC8
HP/Agilent/Philips 21202A, 21364A, 21365A,
Philips / HP ULT-PA-18
21366A, 21367A
Mindray M9, M7, TE7; Philips Sparq, CX50, Affiniti
Philips / Mindray /
50, Affiniti 70, Epiq 5, Epiq 7, Epiq CVx; Samsung ULT-PA-19
Samsung / Koelis SAS
Medison HM70A; Koelis - All Transducers
ICT7-4, ITC8-5, C60, L38/10-5, TEEx For
Sonosite Sonosite Titan and Micromaxx Ultrasound ULT-PA-11
Systems
Acuson/Toshiba for use with Acuson/Siemens XP,
Aspen, Capasee, 3-needle guide C3 Transducers;
Toshiba ATL 3.5 DFT Transducers; Toshiba PSF-37HT ULT-PA-20
and F series Transducers For Toshiba SSH-140A,
SSA-270A, and 340A systems
Zonare E9-4 ULT-PA-29
The following is the minimal equipment needed to test the electrical safety of ultrasound
transducers:
transducers, independent of the ultrasound machines on which they are typically used.
Although the ULT-2000 can be used on virtually any type of ultrasound transducer, it is
The ULT-2000 tests the integrity of the outer insulation barrier of the transducer and
transducer cable as well as the existing capacitive leakage currents. Due to the proximity
of the TEE transducer to the human heart during a normal procedure, abnormally
elevated electrical leakage currents can be hazardous to the patient. Excessive electrical
routinely test TEE (and other types of) ultrasound transducers prior to their use.
All ultrasound transducers have inherent electrical leakage characteristics, and the
maximum leakage currents and associated PASS / FAIL limits as prescribed by the
manufacturer. These are different than the acceptable electrical leakage current limits
4
for the actual ultrasound machines. The ULT-2000 is the only battery-operated handheld
tester on the market today that tests according to these established protocols, which have
been adopted by diagnostic ultrasound manufacturers. It tests both the upper and lower
Typical electrical safety (leakage) testing of the diagnostic ultrasound transducer should
occur as part of the routine cleaning and disinfecting activity that is performed between
immersed in a basin (see Item 4 in Figure 1) containing conductive liquid (see Item 7 in
Figure 1) suitable for performing electrical safety tests. This liquid can be the routine
current flow. Cidex is an example of a widely used disinfecting agent used for cleaning
ultrasound transducers, and is suitably conductive for performing electrical safety tests
on transducers.
The ultrasound transducer is immersed in this conductive liquid and the electrical
connector of the transducer (see Item 6 in Figure 1) is attached to a suitable adapter (see
Item 3 in Figure 1) for that particular manufacturer and model transducer. This creates
one electrical “pole” for the test. The transducer adapter is then attached to the ULT-
2000 as shown in Figure 1. A special conductive probe (see Item 2 in Figure 1) is then
attached to the ULT-2000. This conductive probe is also immersed in the basin of
conductive liquid, creating the second electrical pole. The setup is now complete for the
5
During testing, the transducer is subjected to a user selectable source voltage. This
voltage is typically set to the normal operating voltage of the transducer’s ultrasound
machine. In North America, the source voltage is typically set to 120 VAC @ 60 Hz. For
countries where the normal operating voltage is 230 VAC, the source voltage can be set
4 – Test Basin
5 – Transducer (DUT)
6 – Transducer
Electrical Connector
1 – ULT-2000
3 – Transducer Adapter
Figure 1
Typical Test Setup for ULT-2000
6
NOTICE – TESTING
CONDUCTIVE SURFACES SUCH AS METAL CARTS OR
TABLETOPS CAN CAUSE ERRONEOUS READINGS DUE TO
ALTERNATE LEAKAGE PATHS FROM TRANSDUCER
ADAPTERS, CABLES, ETC. ENSURE TEST IS DONE ON A NON-
CONDUCTIVE SURFACE FOR BEST RESULTS.
7
TEST PROCESS
The ULT-2000 series completes four intermediate tests as part of the Full Test to fully
evaluate the integrity of an ultrasound probe. The following is the details of each test
step:
Source Voltage Test - The first step is to read the actual source (challenge) voltage that
will be applied during the testing to ensure that it is within range. If it is not, an alarm is
Self Test - The second step checks the Leakage Measuring Circuitry. A relay switches
to a dummy internal load. The source (challenge) voltage is then applied to this load. The
unit must correctly read the known leakage current. If it is not read correctly, an alarm is
Bath Conductivity Test – The third step tests the conductivity of the liquid in the test
basin. The ULT-2000 accomplishes this task quite easily and reports a simple PASS /
FAIL, or the actual numerical conductivity of the liquid (depending on the system
leakage current test, the ULT-2000 will report this and will not allow the probe test to be
performed.
Probe Leakage Test – The fourth step tests the electrical leakage of the ultrasound
transducer. The measured leakage current is compared to the selected upper and lower
limits. Again, the results of the test will be reported as a simple PASS / FAIL, or the actual
leakage current values. This step is only included when the user performs a Full Test.
8
LAYOUT
This section looks at the physical layout of the ULT-2000 Series and gives descriptions
of the elements.
9V Battery
Compartment
(Back)
9
KEYS
Six tactile-touch keys with audio feedback are provided for system operation:
– This key turns the unit on and off. The unit will initiate with the Main Screen.
– At the Main Screen, this key initiates the Full Test, which includes a Source
Voltage Test, a Self Test, a Bath Test, and a transducer Probe Test.
– All other screens, this key scrolls up through the selected
parameter options.
– At the Datalog Screen, this key prints the latest test results to the serial port.
– All other screens, this key selects the next available parameter.
– When not in a SETUP menu, this key resets the system to the main
screen.
– When in a SETUP menu, this key has no function.
– This key toggles the unit through operating modes. Pressing this key toggles
from the Main Screen, to the Meter Screen (ULT-2020 Only), to the Datalog
Screen (ULT-2020 only), and then to the Device Configuration Screen.
– Pressing and holding this key allows entry to the SETUP menus where system
configurations can be viewed and adjusted. When in a SETUP menu, this key
exits the SETUP menu and returns to the previously viewed screen. This also
saves the system settings to the internal EEPROM memory so they are
retained with the power turned off or battery removed.
10
SCREENS
MAIN SCREEN – The main screen indicates that the ULT-2000 Series unit is initialized
and ready for testing. This screen displays after power-up initialization, and can be
accessed by pressing the key at any time other than during setup mode.
FULL TEST SCREEN – This screen is accessed from the Main Screen by pressing the
11
BATH TEST SCREEN – This screen is accessed from the Main Screen by pressing the
Test Measurements
Numerical values will not
Progress Indicator
appear if Test Mode is set
to “Pass/Fail”. See
System Configuration in
SETUP.
TEST MESSAGES – Messages appear after each FULL or BATH test is completed.
TEST PASSED – This message indicates a successful test within the selected limits.
TEST FAILED – This message indicates a test failure, and provides failure details. The
example shown below indicates that the probe leakage current is above the selected
upper limit.
12
METER MODE SCREEN (ULT-2020 ONLY) – The meter mode screen allows extended
transducer probes and cables. This screen displays the current leakage current reading
and user selectable Source Voltage, Source Frequency, and Output Control. This screen
is accessed from the Main Screen by pressing the key. The Source Voltage,
Frequency, and Output Control are selected by pressing the key until the desired
parameter is highlighted. Scroll through the highlighted parameter options using the
keys.
13
DATALOG SCREEN (ULT-2020 ONLY) – The Datalog Screen displays the test results
of up to 99 test records. The Datalog Screen includes a Record Number, Date / Time
Stamp, Probe Information, Test Results, Test Measurements, and Test Limits. This
screen is accessed from the Meter Screen by pressing the key. Scroll through
Probe Information
Manufacturer: Model. Test Results
Pass / Fail
Blank if test was configured
for Fixed or User Test
Limits. Test Limits
1 - 500 A
Test Measurements
Measurements taken
during test.
14
Use the key to print the displayed record to the serial printer. Below is a sample
print.
15
DEVICE CONFIGURATION SCREEN – The Device Configuration displays the current
test configuration. Device Configuration parameters include Test Limit Mode, Probe
Manufacturer, and Probe Model. This screen is accessed from the Datalog Screen by
pressing the key. The parameters are selected by pressing the key until
the desired parameter is highlighted. Scroll through the highlighted parameter options
NOTE: If the “Allow User Config” parameter is set to NO, the user will not be able to
modify the test parameters. See SETUP for more details.
16
LOW BATTERY – When the battery life is 10% or less, the LOW BATTERY message
Battery Eliminator (BC20-21111) that may be used for continuous run applications. It
NOTE: The unit is shipped with a Red Battery Lock-Out plug installed into the line power
connector as shown below. Its purpose is to prevent the unit from accidentally being
turned on during handling and transport, subsequently depleting the battery. This plug
17
SETUP
Two user selectable setup menus are provided, SYSTEM CONFIGURATION and USER
TEST CONFIGURATION. Enter a setup menu by pressing and holding the key
until the Access Code Screen appears (5 sec). The arrows are then used
NOTE: The access codes may be changed once you have entered the selected configuration.
See System Configuration and User Configuration for access code details.
When the desired access code is displayed, press the key again to access the
selected setup menu. The configuration parameters are selected by pressing the
key until the desired parameter is highlighted. Scroll through the highlighted
SYSTEM CONFIGURATION
Below is the typical System Configuration Screen, followed by a table of the available
18
System Configuration
19
USER TEST CONFIGURATION
Below is the typical User Test Configuration Screen, followed by a table of the available
20
PC SOFTWARE
The included PC Software allows the user to create custom limit configurations including
Manufacturer, Model, Leakage Limits, Test Voltage, and Frequency. Install and launch
the PC Software, then connect the PC and ULT-2000 via cable BC20-41361 (See
software allows the user to download/read custom configurations to/from the ULT-2000.
NOTE: The PC Software only manipulates the “Custom Test Limits” portion of the
21
COMMUNICATION PROTOCOL
The communication protocol provides a means to completely configure and use the ULT-
Communication Port
The Serial port is configured as 115,200 Baud Rate, 8 Data Bits, 1 Stop Bit, and No Parity.
Command Syntax
The command description is broken into columns; the KEYWORD, the NODE, and the
VALUE.
The KEYWORD provides the name of the command. The actual name of the command
consists of one or more keywords since SCPI commands are based on a hierarchical
structure, also known as a tree system.
In such a system, associated commands are grouped together under a common node in
the hierarchy, analogous to the way leaves at a same level are connected at a common
branch. This, and similar branches are connected to fewer and thicker branches, until
they meet at the root of the tree. The closer to the root, the higher a node is considered
in the hierarchy. To activate a particular command, the full path to it must be specified.
This path is represented in the following table by placing the highest node in the left-most
position. Further nodes are indented one position to the right of the parent node.
The highest level node of a command is called the Keyword, followed by the Node(s),
and then the value. Keywords and Nodes are separated by a colon to create the parent
structure. The parent structure is separated from the value by one space, and the
command is executed with a carriage-return (<cr>). For example, to change the system
display contrast to a value of ten, use “SYSTem:CONtrast 10<cr>”.
Some commands allow for reading and writing data and some commands are Read Only.
To indicate a read function, a question mark (?) is placed at the end of the command path.
For example, to read the system version, use “SYSTem:VERsion?<cr>”. This command
would return the current system version.
NOTE: Commands can be entered in either upper or lowercase or a mixture of the two,
uppercase and lowercase. Commands sent to the device are not case sensitive. Upper
and lower cases are only used when documenting the commands.
22
ULT-2000 Communication Command Summary
Keywords Nodes Subnodes Values
OUTput VOLTage 90-275
FREQuency 50 hz, 60 hz
TEST MODE NUMerical, PASSfail
LIMits FIXed, DEFaults, CUSTom
DMANufacturer 1-12 (Selects Manufacturer 1-12)
DMODel 1-10 (Selects Model 1-10)
CONFigure
LIMits CONDuctivity 0 - 500 (uA)
ULEAkage 0 - 500 (uA)
LLEAkage 0 - 500 (uA)
METer VOLTage 90-275
(ULT-2020 ONLY) FREQuency 50 hz, 60 hz
OUTput OFF, ON
INITiate FULLtest, BATHonly, METermode
ABORt
TIME hh,mm (24hr mode only)
TFORmat AMpm, 24hr
DATE yy,mm,dd
DFORmat MMddyy,DDmmyy
BATlife? [read only]
CONtrast 0-20
BTIMe OFF, 1-20, ON
AOFF 0-30
SYSTem
VERsion? [read only]
HEADer String - Up to 40 Chars
MODEL? [read only]
DMANufacturer <m>, xxx <m> = 1-12, xxx = String data up to 10 chars
DMODel <m>, <n> <m> = 1-12 (Manuf), <n> = 1-10 (model), Model String
DLIMits <m>, <n> , Limits <m> = 1-12 (Manuf), <n> = 1-10 (model), Limits
DSAVe Saves DUT tables to Eeprom
KEY 1-6
VOLTage? 90 - 275 VAC [read only]
SENSe
LEAKage? 0 - 500 uA [read only]
23
ULT-2000 Communication Command Summary (cont.)
Keywords Nodes Values
Bit Value Definition
0 1 Test Running
1 2 Bath Only Test
2 4 Full Test
3 8
4 16
5 32
6 64 Test Failed
STATus? 7 128 Test Passed
8 256
9 512
10 1024
11 2048 Program Mode
12 4096 Meter Mode
13 8192
14 16384
15 32768 Calibration Mode
VTOLerance 0-25 %
FVOLtage 90 - 275 VAC
FACTory FCONductivity 0 - 500 uA
FULeakage 0 - 500 uA
FLLeakage 0 - 500 uA
DATAlog NUMrecords 1 - 100
(ULT- ERASelog [Set to '1' to erase datalog]
2020) RECord <n>? Returns Record <n> (Omit <n> to read last record)
24
MANUAL REVISIONS
Revision # Program # Revisions Made
Rev 01 DT7331CA Origination
Rev 02 DT7331CA Misc. Edits
Rev 03 DT7331CA Adapter Information Updated
Rev 04 DT7331CB Model Information Updated
Rev 05 DT7331CC Dynamic Device Configuration Screen
Rev 06 DT7331CC Edits to Euro Transformer, Conductivity Probe and
Transducer Adapter List Updated, and Misc. Edits
Rev 07 DT7331CE Address Updated
Rev 08 DT7331CF Transducer Adapter List Update, Accessories List
Updated, Communication Protocol Added, Sample
Data Print Added, PC Software Section Added,
Misc. Edits
Rev 09 DT7331CF Euro Transformer Edited
Rev 10 DT7331CG Specifications Updated, Misc. Edits
Rev 11 DT7331CG Misc. Edits
Rev 12 DT7331CG Format Updated, Printer cable part number
updated, Misc. Edits
Rev 13 DT7331CG Specifications Updated
Rev 14 DT7331CJ Transducer Adapter List Update, Conductivity
Probe List Update, Printer Feature Update
Rev 15 DT7331CJ Battery Eliminator Updated, Misc. Edits
Rev 16 DT7331CL Remote command for model fixed, Datalog for low
leakage fixed, Default settings updated.
Rev 17 DT7331CO Calibration Interval added for PA and PC Series
Rev 18 DT7331CP Declaration of Conformity updated
Rev 19 DT7331CP Added Preventive Maintenance Section
Rev 20 DT7331CQ Changed number of Manufacturers to 12 and
Models to 10.
25
LIMITED WARRANTY
REMEDIES: THE PURCHASER'S SOLE AND EXCLUSIVE REMEDY SHALL BE: (1) THE REPAIR
OR REPLACEMENT OF DEFECTIVE PARTS OR PRODUCTS, WITHOUT CHARGE. (2) AT THE
OPTION OF BC GROUP INTERNATIONAL, INC., THE REFUND OF THE PURCHASE PRICE.
26
SPECIFICATIONS
CONNECTIONS
27
ELECTRICAL & MISC.
9V Lithium Battery
BATTERY
(ANSI/NEDA 1604LC or equivalent)
10 VDC, 300mA
BATTERY
ELIMINATOR
BC20-21111 (Universal)
> 100 Full Tests
CONTINUOUS
(Note: Backlight set to OFF)
BATTERY LIFE
OFF 1 year
BAUD 115200
DATA BITS 8
START BITS 1
STOP BITS 1
PARITY none
HANDSHAKING none
Seven (7) pin Mini-DIN
RS232 Pinout:
COMMUNICATIONS
CONNECTIONS
28
NOTES
29
NOTES
30