Operating Manual

General Monitors S5000


Gas Monitor

Order No.: MANS5000/06

Print Spec: 10000005389 (EO)

CR:800000042524

MSAsafety .com
WARNING!

Read this manual carefully before using or maintaining the device. The device will perform as designed only if it is used and
maintained in accordance with the manufacturer's instructions. Otherwise, it could fail to perform as designed, and persons
who rely on this device could sustain serious injury or death.

The warranties made by MSA with respect to the product are voided if the product is not installed and used in accordance
with the instructions in this manual. Please protect yourself and your employees by following the instructions.

Please read and observe the WARNINGS and CAUTIONS inside. For additional information relative to use or repair, call 1-
800-MSA-2222 during regular working hours.

For countries of Russian Federation, Republic of Kazakhstan and Republic of Belarus, the gas detector will be delivered
with a passport document that includes valid approval information. On the CD with manual instruction attached to the gas
detector the user will find the documents "Type Description" and "Test Method" - appendixes to Pattern Approval
Certificate of Measuring instrument, valid in the countries of use.

MSA is a registered trademark of MSA Technology, LLC in the US, Europe and other Countries. For all other trademarks
visit https://us.msasafety.com/Trademarks.

This product incorporates Bluetooth® wireless technology. The Bluetooth word mark and logos are registered trademarks
owned by Bluetooth SIG, Inc., and any use of such marks by MSA is under license. Other trademarks and trade names are
those of their respective owners.

This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device
may not cause harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.

You are cautioned that changes or modifications not expressly approved by the party responsible for compliance could void
the user's authority to operate the equipment.

English:

This device complies with RSS-210 of the Industry Canada Rules. Operation is subject to the following two conditions: (1)
This device may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.

French:
Ce dispositif est conforme à la norme CNR-210 d'Industrie Canada applicable aux appareils radio exempts de licence. Son
fonctionnement est sujet aux deux conditions suivantes: (1) le dispositif ne doit pas produire de brouillage préjudiciable, et
(2) ce dispositif doit accepter tout brouillage reçu, y compris un brouillage susceptible de provoquer un fonctionnement
indésirable.

© MSA 2020. All rights reserved


US

26776 Simpatica Circle


Lake Forest, CA 92630
USA

For your local MSA contacts please go to our website www.MSAsafety.com

General Monitors S5000 3


US

Contents

1 Safety Regulations 7
1.1 Correct Use 7
1.2 Product Warranty 9

2 Description 10
2.1 Display 10
2.2 No Tool Interface 10
2.3 Bluetooth® Wireless Technology 11
2.4 Dual Sensing 11
2.5 Retrofit Installations 12
2.6 XCell Sensors Optimized for Fixed Gas Applications 12
2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors 12
2.8 SafeSwap 13
2.9 Housing 13
2.10 Component Overview 14
2.11 Label Overview 15

3 Installation 17
3.1 Installation Warnings - Read Before Installation 17
3.2 Reviewing Shipment and Identifying Product Model 17
3.3 Product Installation Check List 18
3.4 Mounting 18
3.4.1 Sensor Mounting Location 19
3.4.2 Transmitter Mounting Location 19
3.4.3 Sensor Orientation 19
3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box 20
3.4.5 Integrated Mounting Points 22
3.4.6 Adjustable Pipe Mount 23
3.4.7 Duct Mount 24
3.4.8 Mounting with a Sunshield 24
3.4.9 Mounting with an SM5000 Sampling Module 25
3.5 Installing a Remote Sensor Junction Box 26
3.6 Electrical Power Connections 27
3.6.1 Electrical Warnings - Read before Connecting Power 27
3.6.2 Retrofit Applications with S4000CH, S4000TH, or TS4000H 28
3.6.3 Electrical Hardware Requirements 28
3.6.4 Power Load Requirements and Maximum Mounting Distances 29
3.6.5 Instructions for Power and Analog Output 35
3.6.6 Relay Electrical and Power Connections 37

4 Operation 39
4.1 Startup 39
4.1.1 Initial Startup 39
4.1.2 Sensor Warm Up Times 39
4.1.3 Startup after Power Failure 40
4.2 Settings 40
4.2.1 Instrument Settings 41
4.2.2 Sensor Setup 53

General Monitors S5000 4


US

4.2.3 Relay Settings 56


4.2.4 HART Settings 58
4.2.5 HART AO 58
4.2.6 Calibration Alert 59
4.2.7 Bluetooth 59
4.2.8 Modbus Settings 60
4.2.9 EZ Touch Button 61
4.2.10 Swap Delay 61
4.2.11 UI Password 62
4.2.12 Transmitter Setting Reset 63
4.3 Info Menu - Viewing Device Status 63
4.3.1 AO Type 63
4.3.2 View or Change System Time 64
4.3.3 Last Calibration 64
4.3.4 Non-Critical Fault Log 64
4.3.5 Device Tag 64
4.3.6 Bluetooth FCC/IC ID 65
4.3.7 Sensor Life and Health Status (only displayed if XCell Sensor is connected) 65
4.4 Setting only configurable via Bluetooth®, Modbus, or HART 65
4.4.1 AO Custom Levels 66
4.4.2 Relay Zone - Horn Mode 66
4.4.3 Unit Device ID/Tag 66
4.4.4 Alarm Direction 67
4.4.5 Alarm Enable/Disable 67

5 Calibration 68
5.1 Calibration Equipment 68
5.2 Calibration Frequency 70
5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only) 71
5.3.1 XCell Sensor with TruCal and Diffusion Supervision (CO&H2S Only) 71
5.3.2 XCell Sensor with TruCal without Diffusion Supervision (CO&H2S Only) 71
5.4 Calibration Types: Zero vs. Span 71
5.5 How to Zero Calibrate XCell and IR Sensors 72
5.6 How to Calibrate XCell Sensors 73
5.7 How to Calibrate an Oxygen XCell Sensor 75
5.8 How to Calibrate an IR Sensor 75
5.9 How to Calibrate a Passive Sensor (Catalytic Bead or MOS) 75
5.10 XCell Catalytic Bead Failsafe 76
5.11 Calibration Confirmation 76

6 Maintenance 77
6.1 IR Sensor Cleaning Procedure 77
6.2 Replacing an XCell Sensor 78
6.3 Clearing a Blockage 80
6.4 Replacing a Passive Sensor (Cat Bead or MOS) 81
6.5 Troubleshooting 81

7 Ordering Information 87
7.1 Replacement Parts 87

General Monitors S5000 5


US

8 Appendix: Specifications 90

9 Appendix: General Certification Information 94

10 Appendix: HART Specific Information 101

General Monitors S5000 6


US 1 Safety Regulations

1 Safety Regulations
1.1 Correct Use

The S5000 Gas Monitor, hereafter also called device, is a gas monitor for measuring toxic and combustible gases as well
as oxygen. Using sensors, the device tests the ambient air and triggers the alarm as soon as the gas exceeds a specific
concentration level.

WARNING!

Do not use silicone-type lubricants in assembling the device and do not allow silicone vapors to be drawn into the flow
system while in operation. Silicone can desensitize the combustible gas sensor, thereby giving erroneously low readings.

Use only genuine MSA replacement parts when performing any maintenance procedures provided in this manual. Failure to
do so may seriously impair sensor and gas monitoring performance, alter flameproof/explosionproof characteristics or void
agency approvals.

Failure to follow these warnings can result in serious personal injury or death.

WARNING!

As with all gas monitors of this type, high levels of, or long exposure to, certain compounds in the tested atmosphere could
contaminate the sensors. In atmospheres where an S5000 Gas Monitor may be exposed to such materials, calibration
must be performed frequently for dependable operation and to confirm that display indications are accurate.

These effects include, but are not limited to:


• Passive MOS sensors may be adversely affected by prolonged exposure to certain substances. Loss of sensitivity or
corrosion may be gradual if such agents are present in low concentrations or it may be rapid at high concentrations.
Examples of these substances are as follows:
– Halides: compounds containing fluorine, chlorine, bromine and iodine
– Heavy metals, e.g. tetraethyl lead
– Caustic and Acidic liquids and vapors
– Glycol
• The H2S Digital Sensor may be adversely affected by the following substances:
– Alcohols (methanol, ethanol, isopropanol)
– Nitrogen dioxide (NO2)
– Chlorine (Cl2)
– Paint solvents (acetone, turpentine, toluene, mineral spirits, etc.)
• The CO Digital Sensor may be adversely affected by the following substances:
– Alcohols (methanol, ethanol, isopropanol)
– Paint solvents (acetone, turpentine, toluene, mineral spirits, etc.)
• The O2 Digital Sensor may be adversely affected by the following substances:
– Long term exposure to low levels of Acetylene
– Paint solvents (acetone, turpentine, toluene, mineral spirits, etc.) in high concentrations larger than 1000 ppm or
prolonged exposure to lower concentrations

General Monitors S5000 7


US 1 Safety Regulations

• Prolonged exposure of the H2S Digital Sensor to humidity levels of 5% RH or lower will result in gas measurement
readings of H2S that are greater than the actual gas concentration present.
• It is not recommended to expose the O2, H2S and CO Digital Sensors to environments containing oxygen levels above
30% (v/v) or below 5% (v/v). The O2 sensor will operate at concentrations below 5% (v/v) but not for prolonged periods
of time.

Failure to follow these warnings can result in serious personal injury or death.

This device complies with Part 15 of the FCC Rules. Operation is subject to the -following two conditions:
• this device may not cause harmful interference, and
• this device must accept any interference received, including interference that may cause undesired
operation.

This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to
Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference when the equipment is operated in a commercial environment. This equipment generates, uses,
and can radiate radio frequency energy and, if not installed and used in accordance with the instruction
manual, may cause harmful interference to radio communications. Operation of this equipment in a residential
area is likely to cause harmful interference in which case the user will be required to correct the interference at
his own expense.

WARNING!

The Digital Sensor Xcell sensor module utilizes thread locker suitable to an ambient temperature of -55°C to +74°C. If the
Digital Sensor is exposed to temperatures outside of its listed ratings, reapplication of thread locker when changing out the
Xcell sensor module may be required.

Failure to follow this warning can result in serious personal injury or death.

NOTICE

This is a Class A product in accordance with CISPR 22. In a domestic environment, this product may cause radio
interference, in which case the user may be required to take adequate -measures.

NOTICE

The XCell sensor refers to the sensor portion of the Digital Sensor throughout this manual.

FCC Warning Statements

Changes or modifications not expressly approved by the manufacturer could void the user's -authority to operate the
equipment.

Industry Canada (IC) Warning Statements

The installer of this radio equipment must ensure that the antenna is located or pointed such that it does not emit RF field in
excess of Health Canada limits for the general population; consult -Safety Code 6, obtainable from Health Canada's
website www.hc-sc.gc.ca.

General Monitors S5000 8


US 1 Safety Regulations

1.2 Product Warranty


The warranties made by GM with respect to the product are voided if the product is not installed, used and serviced in
accordance with the instructions in this manual. Please protect yourself and your employees by following the instructions.

ITEM WARRANTY PERIOD


S5000 Gas Monitor MSA warrants that this product will be free from mechanical defects and faulty workmanship for
the period specified in this table for each component, provided it is maintained and used in
accordance with MSA’s instructions and/or recommendations. Warranty shall not exceed.
Main Transmitter 2 years from date of shipment.
Housing and PCBA
Shall not exceed 2 years and 6 months from date of manufacture.
XCell Sensors 3 years from date of shipment.

Shall not exceed 3 years and 6 months from date of manufacture.


IR Sensor 2 years from date of shipment.

Shall not exceed 2 years and 6 months from date of manufacture.


Passive Catalytic 2 years from date of shipment.
Bead
Shall not exceed 2 years and 6 months from date of manufacture.
Passive MOS 2 years from date of shipment.

Shall not exceed 2 years and 6 months from date of manufacture.

This warranty does not cover filters, fuses, etc. Certain other accessories not specifically listed here may have different
warranty periods. This warranty is valid only if the product is maintained and used in accordance with Seller's instructions
and/or recommendations. The Seller shall be released from all obligations under this warranty in the event repairs or
modifications are made by persons other than its own or authorized service personnel or if the warranty claim results from
physical abuse or misuse of the product. No agent, employee or representative of the Seller has any authority to bind the
Seller to any affirmation, representation or warranty concerning this -product. Seller makes no warranty concerning
components or accessories not manufactured by the Seller, but will pass on to the Purchaser all warranties of
manufacturers of such components.

THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED, IMPLIED OR STATUTORY, AND
IS STRICTLY LIMITED TO THE TERMS HEREOF. SELLER SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE.

Exclusive Remedy
It is expressly agreed that Purchaser's sole and exclusive remedy for breach of the above warranty, for any tortious
conduct of Seller, or for any other cause of action, shall be the replacement at Seller's option, of any equipment or parts
thereof, which after examination by Seller is proven to be defective. Replacement equipment and/or parts will be provided
at no cost to Purchaser, F.O.B. Seller's Plant. Failure of Seller to successfully replace any nonconforming equipment or
parts shall not cause the remedy established hereby to fail of its essential purpose.

Exclusion of Consequential Damage


Purchaser specifically understands and agrees that under no circumstances will seller be liable to purchaser for economic,
special, incidental or consequential damages or losses of any kind whatsoever, including but not limited to, loss of
anticipated profits and any other loss caused by reason of non-operation of the goods. This exclusion is applicable to
claims for breach of warranty, tortious conduct or any other cause of action against seller.

General Monitors S5000 9


US 2 Description

2 Description
2.1 Display

The S5000 utilizes a dot matrix LED display, capable of displaying four alphanumeric characters at a time. The display will
scroll words that exceed four letters. Most of the messages scroll twice across the screen before moving onto next
selection.

Figure 1 S5000 Main Display

In addition to the red LED display, the S5000 uses six icons to indicate status. Green LED indicates power supply status.
A yellow triangle and red bell indicate fault and warning or alarm conditions respectively. The Bluetooth® icon indicates that
the Bluetooth® wireless technology is enabled on the device. Yellow "1" and "2" icon indicate which sensor gas reading is
being displayed, or during configuration which sensor's options are being accessed.

2.2 No Tool Interface

The S5000 does not require any tools or third party devices to change settings, reset alarms or perform any maintenance
operation. The EZ touch button works through the glass and does not require opening the explosion proof enclosure. The
EZ touch button works with bare fingers or with gloved hands, so long as the gloves are not black. See 4 Operation for more
information on navigating the menu with the EZ touch button.

General Monitors S5000 10


US 2 Description

Figure 2 Interface

The user menu can also be accessed using the round GM magnet on the General Monitors logo.

2.3 Bluetooth® Wireless Technology

The S5000 comes by default with Bluetooth communication. Using the X/S Connect App on an appropriate smart phone or
tablet, users are able to interface with the S5000 menu options in a larger and more user friendly setting. The Bluetooth
communication can interact with the device within a maximum transmission distance of 70 feet (21 m).

WARNING!

Bluetooth operation is dependent upon signal availability of the wireless service(s) necessary to maintain the
communication link. Loss of wireless signal will prevent communication of alarms and other information to linked devices.
Take appropriate precautions in the event a loss of wireless signal occurs.

Failure to follow this warning can result in serious personal injury or death.

If ordered with Bluetooth, the device will be shipped with Bluetooth enabled. See 4 Operation for instructions on
disabling Bluetooth.

If the device was not ordered with Bluetooth it cannot be upgraded.

The S5000 and user provided communication device will need to be paired. This requires both devices to be in range and for
a pairing sequence inputting a 6 digit pairing code. The instructions will be displayed on both the S5000 and communication
device.

There are communication devices capable of being used in classified areas. Please contact your MSA
representative for additional information.

2.4 Dual Sensing

The S5000 supports two Digital Sensors, or one IR Sensor point IR detector and one Digital Sensor simultaneously with
four wire connections. However, the device will only support one passive sensor, either combustible catalytic bead or
metal oxide semiconductor (MOS) sensors, based on the ATO configuration. Passive catalytic bead uses three wires,
passive MOS uses four wires.

The S5000 Gas Monitor generates two independent analog outputs; one for each sensor connected to the transmitter. The
analog output associated with Sensor 1 also has the digital HART (Highway Addressable Remote Transducer)

General Monitors S5000 11


US 2 Description

communication superimposed on the analog signal. If two sensors are connected, the digital HART communication carries
information for both sensors.

2.5 Retrofit Installations

The S5000 has the conduit entries located in the exact same orientation and distance from the wall and the mounting holes
for attaching to a wall are identical to the S4000 Series detectors. Users will be able to re-use the existing wiring for the
remotely mounted sensors.

2.6 XCell Sensors Optimized for Fixed Gas Applications

XCell toxic and combustible catalytic bead sensors are developed and manufactured by MSA. Now optimized for fixed gas
applications, the XCell sensor platform is available in the S5000 and provides multiple benefits, including a standard 3-year
warranty on all XCell sensors.

One important optimization for fixed gas was incorporating the GM catalytic bead into the XCell sensor.

The XCell Oxygen sensor does not use lead, but rather a non-consuming reaction chemistry. The XCell Oxygen sensor is
expected to last well over 3 years and can be safely stored on the shelf for at least 1 year without sensor performance
degradation. Changes in barometric pressure across the range of 86 kPA - 108 kPA have a negligible effect on the
operation of the sensor.

2.7 TruCal Sensing Technology for CO and H2S Electrochemical Sensors

Using patented pulse check technology and proprietary Adaptive Environmental Compensation (AEC) algorithms, all XCell
sensors with TruCal verify operation by actively adjusting the sensor output for changes in sensitivity. Some XCell sensors
with TruCal also include Diffusion Supervision, which monitors the sensor inlet for obstructions that could prevent gas from
reaching the sensor.

Every six hours, an electrical pulse stimulates the XCell sensor similar to having actual calibration gas applied, providing a
snapshot of the sensor's sensitivity at the time of the pulse. Using this sensitivity snapshot, the sensor can diagnose
sensor failures like electrode poisoning, electrolyte leaking, or electrical connectivity issues.

AEC uses the sensitivity snapshots provided by the pulse check to adjust sensor output, compensating for environmental
impacts on sensor accuracy. If the AEC adjustment is greater than expected based on typical environmental impact
variations, the transmitter LED’s will slowly flash GREEN, alerting users that the sensor should be calibrated to reset the
AEC cycle. Users can also enable a Calibration Alert function that will send a milli-amp signal on the analog output to the
control room. The result is a sensor that actively self-monitors for operation and accuracy, with far fewer manual
calibrations.

Diffusion supervision actively monitors the sensor inlet for obstructions. If an obstruction is detected, the sensor will go
into a fault mode to alert users and the control room that it is not seeing gas due to an obstruction. Objects residing directly
on or in the sensor inlet that result in a significant impact to the gas path are very likely to be detected by Diffusion
Supervision. Examples include paint, tape, water, and dirt. Small amounts of these materials can be visible on the inlet
while not impacting the gas path enough to trigger a Diffusion Supervision Fault. A fault signal will only be sent out when
the system determines that the amount of material that has accumulated on or inside the sensor inlet is negatively affecting
the gas path.

Actual TruCal sensor performance will depend on the application, background gas exposure, and environment. To validate
XCell sensors with TruCal, it is recommended that users follow their regular calibration cycle and record the "as found" and
"as left" values. This data can be used to extend the time between calibrations depending on the required specification of
the application.

General Monitors S5000 12


US 2 Description

2.8 SafeSwap

The S5000 comes with patented SafeSwap technology, which allows users to change or replace XCell sensors without
needing to power down the instrument. Swap delay is enabled on the S5000 by default; a feature that gives users a
2 minute window to change sensors without triggering a fault condition. SafeSwap and Swap Delay are only applicable for
XCell sensors. For more information on SafeSwap and Swap Delay, see 6 Maintenance.

WARNING!
• As part of the product certification, it was verified that optional communication functions of this gas detection
instrument while operating at the maximum transaction rate do not adversely affect the gas detection operation and
functions of the instrument. The product certification, however, does not include or imply approval of the SafeSwap
feature, communications protocol or functions provided by the software of this instrument or of the communications
apparatus and software connected to this instrument.
• Follow the warnings below when removing or replacing sensors. Reference Figure 3 for component overview.
– Never remove or replace a Sensor Body Assembly or an IR Sensor while under power or when explosive hazards
are present.
– Confirm that the area is free of explosive hazards before removing or replacing an XCell Sensor under power.
– To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then remove the XCell
Sensor completely.

Failure to follow these warnings can result in serious personal injury or death.

2.9 Housing

The S5000 comes in 316 Stainless Steel for the highest corrosion resistance. All housings have ¾” NPT conduit entries.
Custom tags are available and easily attach to an integral ring. The JB5000 junction box comes in 316 Stainless Steel for
the highest corrosion resistance. The housing is offered in ¾" NPT of M25 conduit entries.

General Monitors S5000 13


US 2 Description

2.10 Component Overview


Figure 3 Exploded View

General Monitors S5000 14


US 2 Description

2.11 Label Overview


Figure 4 Transmitter - Position of Labels

Figure 5 Board stack - Position of Labels

Figure 6 Digital Sensor - Position of Labels

General Monitors S5000 15


US 2 Description

Figure 7 IR Sensor - Position of Labels

Figure 8 Passive Sensor - Position of Labels

General Monitors S5000 16


US 3 Installation

3 Installation
3.1 Installation Warnings - Read Before Installation

WARNING!
• Refer to 9 Appendix: General Certification Information before installation and operation.
• Some Digital Sensors are provided in a fritless sensor housing. The fritless sensor housing is labeled as Div 2 or Zone 2
and is approved for Div 2 or Zone 2 installations only. The protection method is Nonincendive or Type n respectively.
Ensure all components are approved for the wiring method being used and in accordance with the National Electric
Code of the country of use, any applicable local regulations, and this manual.
• As part of the product certification, it was verified that optional communication functions of this gas detection
instrument while operating at the maximum transaction rate do not adversely affect the gas detection operation and
functions of the instrument. The product certification, however, does not include or imply approval of the SafeSwap
feature, communications protocol or functions provided by the software of this instrument or of the communications
apparatus and software connected to this instrument.
• Follow the warnings below when removing or replacing sensors. Reference Figure 3 for component overview.
– Never remove or replace a sensor body assembly or an IR Sensor while under power or when explosive hazards
are present.
– Confirm that the area is free of explosive hazards before removing or replacing an XCell Sensor under power.
– To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then remove the XCell
Sensor completely.
• Plug all unused conduit entries with a suitably certified blanking/stopping plug.
• Do not paint the device. Avoid painting in areas where the S5000 and remote sensor junction box are located. If painting
is required in an area where an S5000 or remote sensor has been installed, exercise caution to ensure paint is not
deposited on the sensor inlet fitting. Paint solvents can also cause an alarm condition to occur or potentially poison
electrochemical sensors.
• Protect the device from extreme vibration.
• Do not mount the sensing head in direct sunlight without a sunshield (P/N 10180254).
• IR Sensors contain no user- or field-serviceable parts and must be returned to the factory for repair. Any attempt to open
the monitor will damage the unit and void the warranty.

Failure to follow these warnings can result in serious personal injury or death.

NOTICE

When installing the IR Sensor, under no circumstances should a pry-bar be applied to the two legs that support the unit's
reflectors during installation or removal of the sensor. Applying force to the legs can permanently damage the IR Sensor.

3.2 Reviewing Shipment and Identifying Product Model

To determine the sensor type and options, check the shipping carton.

General Monitors S5000 17


US 3 Installation

Figure 9 Shipping Label

Passive, Digital, and IR sensors are shipped attached to the S5000. Passive and IR Sensors are one-piece sensors.
Digital sensors comprise two parts: the Sensor Body Assembly and the XCell Sensor. Sensor Body Assemblies must be
installed and tightened using a strap wrench.

Check the sensor details before attaching to the S5000 housing. The sensor details are listed on the inside of the XCell
Sensor. Unscrew the XCell Sensor from the sensor body assembly and check the label on the inside for gas type, range,
configuration ordered, serial number, and firmware revision number.

3.3 Product Installation Check List

Before Installation
• Review national electrical codes
• Review local procedural and building codes
• Determine optimum transmitter placement
• Determine wire requirements
• Determine mounting hardware requirements
• Review approvals and ensure suitability for installation

Mounting
• Attach appropriate sensor to housing or junction box (see 3.4.3 Sensor Orientation for proper sensor orientation)
• Mount transmitter or junction box using appropriate mounting hardware
• Confirm free air flow around the sensor

3.4 Mounting

WARNING!

Refer to 9 Appendix: General Certification Information before installation.

Some toxic gases are provided in a fritless sensor housing. The fritless sensor housing is labeled as Div 2 or Zone 2 and is
approved for Div 2 or Zone 2 installations only. The protection method is Nonincendive or Type n respectively. Ensure all

General Monitors S5000 18


US 3 Installation

components are approved for the wiring method being used and in accordance with the National Electrical Code of the
country of use, any applicable local regulations, and this manual.

Failure to follow this warning can result in serious personal injury or death.

3.4.1 Sensor Mounting Location

The best location for the transmitter and the sensor may not be the same location. Sensors should be placed in a location
where a gas leak is most likely to be detected. When the best sensor placement would not allow the transmitter display to
be easily viewed or accessed, a remote junction box can be used to mount the sensor remotely from the transmitter,
allowing both to be installed in the optimum location.

Two main factors should be considered when choosing a sensor location. The first is the density of the target gas relative
to the air. Gases, such as propane, that are heavier than air should be placed near ground level while gases that are lighter
than air should be placed above potential leak sources.

Optimum sensor placement will depend on the surrounding processing equipment, such as pipes, valves, or turbines. MSA
offers a gas and flame mapping service that systematically evaluates potential sources of leaks and recommends detector
quantity and placement to create the most effective detection system.

3.4.2 Transmitter Mounting Location

The transmitter display should be mounted so that the screen is visible and easily accessed after installation. The
electronics assembly inside the enclosure have one orientation inside the cylindrical housing. Take care to position the
conduit entries and display so that the display is read in the correct orientation.

3.4.3 Sensor Orientation

WARNING!

Mount the IR Sensor with the sensor inlet fitting extended horizontally from the main enclosure (see Figure 10 ) to prevent
the build-up of particulate or liquid matter on the monitor's optical surfaces.

Mount the digital sensor with the sensor inlet fitting (see Figure 11 ) pointed downward; otherwise, the inlet may become
clogged with particulate matter or liquids.

Failure to follow this warning can result in serious personal injury or death.

Sensor orientation will depend on the sensor type. If mounting an IR Sensor, whether locally on the transmitter or via
remote junction box, the sensor should be mounted horizontally. If the IR Sensor is not mounted horizontally, the sensor
will be prone to more frequent beam blocking issues due to accumulated dust and condensation on the surface of the
IR Sensor. Figure 10 shows the correct and incorrect mounting orientations for the IR Sensor.

General Monitors S5000 19


US 3 Installation

Figure 10 Correct and Incorrect Mounting Orientations for IR Sensor

All other sensors, including electrochemical, catalytic bead, oxygen, passive catalytic bead, and passive MOS should be
mounted vertically with the gas inlet pointed downward. If the sensor is not mounted with the gas inlet facing down, it is
more likely to become clogged with particulate matter or liquids. Figure 11 shows the correct and incorrect mounting
orientation for digital sensors. Passive catalytic bead and MOS sensors come already installed on the transmitter housing.

Figure 11 Correct and Incorrect Mounting Orientation for Digital Sensors

3.4.4 Connecting Sensor to Transmitter Housing or Remote Junction Box

Digital and IR Sensors are shipped attached to the main enclosure. The main sensor input is provided via a four-terminal
connection that provides a digital interface for all sensor modules. Up to two sensors (excluding passive sensors) can be
connected to a single transmitter with two analog outputs capable of representing the readings of the individual sensors.

Passive sensors are shipped already attached and electrically wired to the device. Only one passive sensor can be used
on a single S5000, and they are not interchangeable with other passive sensors or digital sensors.

Consider the sensor dimensions when choosing a mounting location for the transmitter or junction box.

To connect the sensor:

(1) Loosen the set screw located on the lid using a 1.5 mm Allen wrench.

(2) Turn the lid counter-clockwise to remove.

(3) Pull out display module to expose terminal connections.

General Monitors S5000 20


US 3 Installation

Figure 12 Terminal Connections

(4) Route the cable from the sensor through a conduit entry hole in the enclosure so that the sensor is oriented in the
correct position (see 3.6 Electrical Power Connections for details).

(Repeat to attach a second sensor to the S5000 transmitter).

(5) Connect the sensor to the "Sensor 1" position on the electronics assembly.

a) If using a second sensor, connect it to the second sensor position.

NOTICE

If only using one sensor and it is connected to “Sensor 2” position, the S5000 will enter Sensor Missing fault. See Disable
Sensor in 4.2.2 Sensor Setup for details on how to clear this fault.

General Monitors S5000 21


US 3 Installation

Figure 13 Connecting Sensor to the Figure 14 SOURCE/SINK Switch


Stack and Grounding Terminal Location

(6) Verify the sensor connector is firmly seated on the terminal board.

(7) Attach the sensor's ground to either of the grounding screws inside the S5000 housing.

(8) Set the analog output to SOURCE or SINK using tweezers, flat head screwdriver or similar tool.

a) With the display board removed, locate “S1” switch on the main board (see Figure 14 ).

b) Set “S1” switch in the required position:

For SINK, set switch to the right side.

For SOURCE, set switch to the left side.

(9) Replace the display module. Push firmly on the board stack.

NOTICE

Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated, the touch interface
performance can be negatively affected.

(10) Replace the cover by turning clockwise.

(11) Tighten the set screw located on the lid using a 1.5 mm Allen wrench.

NOTE: The JB5000 junction box is not compatible with IR400 and passive sensors.

3.4.5 Integrated Mounting Points

The S5000 transmitter can be surface mounted without any additional brackets using the integrated mounting tabs.

General Monitors S5000 22


US 3 Installation

Figure 15 Integral Mounting Tabs

Mounting Points JB5000 Junction Box

The JB5000 junction box can be mounted directly using the 4 integrated 10-32 threaded holes on the back of the enclosure
or with the use of a mounting bracket (P/N 10206570).

3.4.6 Adjustable Pipe Mount

A Universal Pipe Mount Kit (P/N 10176946) can be used to mount the S5000 on pipes ranging from 20-150 mm in diameter.
Two brackets are mounted over the top of the integrated mounting tabs and fitted with an adjustable pipe band (not
included).

Figure 16 Adjustable Pipe Mount

General Monitors S5000 23


US 3 Installation

3.4.7 Duct Mount

Duct mount kits are available for monitoring atmosphere inside flat or round ducts. Round duct mount kits are available for
small ducts 12-20" in diameter (P/N 10179323) and large ducts 20-40" in diameter (P/N 10179324). The flat duct mount
(P/N 10179322) is universal for flat ducts.

NOTICE

Consider the sensor type before choosing a duct mount location. IR Sensors should be mounted horizontally and all other
sensors should be mounted vertically.

NOTICE

Air flow in the duct must be zero to ensure proper calibration.

Figure 17 Flat Duct Mount

Figure 18 Round Duct Mount

3.4.8 Mounting with a Sunshield

A sunshield is required to protect the S5000 from direct sun light (P/N 10180254). The sunshield can be used in any of the
mounting configurations.

General Monitors S5000 24


US 3 Installation

Figure 19 Sunshield with Surface Mount

Figure 20 Sunshield with Universal Pipe Mount

3.4.9 Mounting with an SM5000 Sampling Module

An aspirated (P/N 10158101) and a DC pump (P/N 10043264) model are available for use with the S5000 with the Digital
Sensor. For more information on mounting requirements and use with the SM5000 sampling modules, see the SM5000
operating manual(s).

General Monitors S5000 25


US 3 Installation

Figure 21 Mounting the SM5000 with a Digital Sensor

SM5000 is not for sale in EU.

Diffusion Supervision must be disabled when using the SM5000.

3.5 Installing a Remote Sensor Junction Box

Sensors mounted remotely must use the S5000 junction box. The junction box housing is the same construction as the
transmitter. The mounting options and instructions for connecting the sensor are the same for sensors connected directly
to the S5000 transmitter housing. The junction box is available in 316 Stainless Steel.

Sensors can be remote mounted up to 100 m from the transmitter housing, as long as the S5000 transmitter is mounted
within maximum distance from the power supply, as indicated in Table 1 .

The junction box does not have an illuminated display and has two connectors for attaching a single sensor input and an
output connects to the transmitter. A 16 AWG 4 element cable with a braided shield should be used for the electrical
connection between the junction box and the S5000 transmitter. Specific cable recommendations are available upon
request.

General Monitors S5000 26


US 3 Installation

Figure 22 Junction Box

Figure 23 Junction Box Electrical Connections

3.6 Electrical Power Connections

3.6.1 Electrical Warnings - Read before Connecting Power

WARNING!
• Before wiring the S5000 transmitter, disconnect the power source supplying the transmitter and ensure no hazardous
atmosphere present; otherwise, electrical shock or ignition of hazardous gases could occur.
• Install wiring in accordance with the electrical code of the country in use, the local authority having jurisdiction and
these installation instructions, as applicable.

General Monitors S5000 27


US 3 Installation

• Do not make any connections to the S5000 main board or junction box input, output, and relay connections while under
power. Making connections under power could lead to electrical shock or ignition of a hazardous atmosphere.
• Ensure that water and dirt are not able to enter the unit via the wire or conduit. If the unit is installed in a location known
to be wet or damp, it is good practice to loop or bend the entry into the unit that prevents water incursion.
• The internal grounding terminal (located on the interior metal board stack plate) must be used for equipment grounding.
The external grounding terminal is only to be used as a supplemental bonding connection where local authorities permit
or require such a connection.
• As part of the product certification, it was verified that optional communication functions of this gas detection
instrument while operating at the maximum transaction rate do not adversely affect the gas detection operation and
functions of the instrument. The product certification, however, does not include or imply approval of the SafeSwap
feature, communications protocol or functions provided by the software of this instrument or of the communications
apparatus and software connected to this instrument.
• Follow the warnings below when removing or replacing sensors. Reference Figure 3 for component overview.
– Never remove or replace a sensor body assembly or an IR Sensor while under power or when explosive hazards
are present.
– Confirm that the area is free of explosive hazards before removing or replacing an XCell Sensor under power.
– To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then remove the XCell
Sensor completely.

Failure to follow these warnings can result in serious personal injury or death.

3.6.2 Retrofit Applications with S4000CH, S4000TH, or TS4000H

The S5000 was designed to be easily retrofitted with existing S4000CH, S4000TH, and TS4000H wiring. When replacing
an existing S4000CH, S4000TH, or TS4000H with the equivalent S5000 sensor technology, the following items need to be
checked in order for the S5000 to operate:

(1) Wire gauge needs to be 18-14 AWG

(2) Sufficient power must be supplied to the S5000 in accordance with the maximum wire lengths. (See the tables in
3.6.4 Power Load Requirements and Maximum Mounting Distances)

If these requirements are met, performance of the S5000 should meet the noise immunity standard equivalent of the
S4000CH, S4000TH, and TS4000H using the existing wiring; However, the installation may not meet the latest EMC
EN50270 noise immunity standard that the S5000 meets with the grounding and wiring scheme as indicated in this manual
and corresponding I/O drawing.

3.6.3 Electrical Hardware Requirements

Braided shielded, twisted pair, instrument quality wire or cable should be used to minimize the possibility of noise
interference and contact with other voltages. Selection of shielded cable must comply with local requirements.

Conduit, in addition to braided shielded wire, may also be needed in areas where a large amount of electrical noise is
expected. All cable shields should be terminated to earth ground at one end only.

The S5000 has a four-wire power terminal, one four-wire communication terminal, and three four-wire sensor terminals.
Relays can be added as an option. Terminals for power and relays can take wires up to 12 AWG while all other terminals
take wires up to 14 AWG. Four conductors are also required for the S5000 remote junction boxes.

General Monitors S5000 28


US 3 Installation

Incoming power and signal cables should be a braided shield cable such as Alpha Wire 3248 or equivalent. The braided
shield must be terminated to the board stack as shown in Figure 27 , or alternatively, the earth ground at the user’s power
source location.

An external Class 2 power supply is required to supply 12-30 VDC to the S5000. Incoming power and signal cables should
be a braided shield cable such as Alpha Wire 3248 or equivalent.

3.6.4 Power Load Requirements and Maximum Mounting Distances

Consider future needs when selecting cable size and power supply. The maximum distance between the S5000 transmitter
and the power supply depends on the sensor configuration (sensing technology and one or two sensors), wire gauge, and
the power supply voltage. The table below outlines the maximum transmitter mounting distances. First determine if the
sensor(s) will be locally or remotely mounted. Then choose sensor type(s). The corresponding maximum power and
mounting distances by wire gauge are shown.

Figure 24 Local Sensors

General Monitors S5000 29


US 3 Installation

Table 1 Maximum Mounting Distance for Local Sensors, Imperial Units

Max. Max. Distance (ft)


Local Sensor 1 Local Sensor 2 Power1
(W) 18 AWG 16 AWG 14 AWG 12 AWG
Passive CB None 6.0 1280 2030 3230 5130
Passive MOS None 10.8 710 1130 1790 2850
Digital CB None 6.0 1280 2030 3220 5130
Digital CB 8.4 910 1450 2300 3660
Digital Toxic 6.7 1140 1810 2880 4580
Digital Toxic None 3.6 2130 3380 5370 8550
Digital Toxic 4.3 1770 2820 4480 7120
Digital CB 6.7 1140 1810 2880 4580
IR Sensor None 8.9 860 1370 2180 3470
Digital CB 11.8 650 1040 1650 2620
Digital Toxic 9.6 800 1270 2020 3210
1- When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply

Assumes transmitter was ordered with relays

Table 2 Maximum Mounting Distance for Local Sensors, Metric Units

Max. Max. Distance (m)


Local Sensor 1 Local Sensor 2 Power1
(W) 1 mm2 1.5 mm2 2.5 mm2 4 mm2

Passive CB None 6.0 470 710 1180 1890


Passive MOS None 10.8 260 390 660 1050
Digital CB None 6.0 470 710 1180 1890
Digital CB 8.4 340 510 840 1350
Digital Toxic 6.7 420 630 1050 1690
Digital Toxic None 3.6 790 1180 1970 3150
Digital Toxic 4.3 650 980 1640 2620
Digital CB 6.7 420 630 1050 1690
IR Sensor None 8.9 320 480 800 1280
Digital CB 11.8 240 360 600 960
Digital Toxic 9.6 290 440 740 1180
1- When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply

Assumes transmitter was ordered with relays

General Monitors S5000 30


US 3 Installation

Figure 25 Local and Remote Sensors

General Monitors S5000 31


US 3 Installation

Table 3 Maximum Mounting Distance for Local and Remote Sensors, Imperial Units

Max. Max. Distance (ft)


Remote Sensor Local Sensor Power1
(W) 18 AWG 16 AWG 14 AWG 12 AWG
Passive CB None 6.2 1240 1970 3140 4990
Passive MOS None 11.2 690 1090 1730 2760
Digital CB None 6.2 1240 1970 3130 4980
Digital CB 8.6 890 1420 2260 3590
Digital Toxic 6.9 1110 1770 2810 4460
IR Sensor 11.9 640 1020 1620 2580
Digital Toxic None 3.6 2130 3380 5370 8550
Digital Toxic 4.3 1770 2820 4480 7120
Digital CB 6.7 1140 1810 2880 4580
IR Sensor 9.6 800 1270 2020 3210
IR Sensor None 9.0 850 1350 2140 3410
Digital CB 11.9 640 1020 1620 2580
Digital Toxic 9.7 790 1250 1980 3160
1- When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply

Assumes transmitter was ordered with relays

General Monitors S5000 32


US 3 Installation

Table 4 Maximum Mounting Distance for Local and Remote Sensors, Metric Units

Max. Max. Distance (m)


Remote Sensor Local Sensor Power1
(W) 1 mm2 1.5 mm2 2.5 mm2 4 mm2

Passive CB None 6.2 460 690 1150 1840


Passive MOS None 11.2 250 380 630 1010
Digital CB None 6.2 460 690 1150 1840
Digital CB 8.6 330 490 830 1320
Digital Toxic 6.9 410 620 1030 1640
IR Sensor 11.9 240 360 590 950
Digital Toxic None 3.6 790 1180 1970 3150
Digital Toxic 4.3 650 980 1640 2620
Digital CB 6.7 420 630 1050 1690
IR Sensor 9.6 290 440 740 1180
IR Sensor None 9.0 310 470 780 1260
Digital CB 11.9 240 360 590 950
Digital Toxic 9.7 290 440 730 1160
1- When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply

Assumes transmitter was ordered with relay

General Monitors S5000 33


US 3 Installation

Figure 26 Remote Sensors

Table 5 Maximum Mounting Distance for Remote Sensors, Imperial Units

Max. Max. Distance (ft)


Remote Sensor Remote Sensor
Power1
1 2 18 AWG 16 AWG 14 AWG 12 AWG
(W)
IR Sensor Digital CB 12.4 620 980 1560 2480
IR Sensor Digital Toxic 9.8 780 1240 1980 3150
Digital CB Digital CB 9.1 840 1340 2130 3380
Digital CB Digital Toxic 6.9 1100 1760 2790 4440
Digital Toxic Digital Toxic 4.3 1770 2810 4470 7110
1- When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply

Assumes transmitter was ordered with relays

General Monitors S5000 34


US 3 Installation

Table 6 Maximum Mounting Distance for Remote Sensors, Metric Units

Max. Max. Distance (m)


Remote Sensor Remote Sensor
Power1
1 2 1 mm2 1.5 mm2 2.5 mm2 4 mm2
(W)
IR Sensor Digital CB 12.4 230 340 570 910
IR Sensor Digital Toxic 9.8 290 430 720 1160
Digital CB Digital CB 9.1 310 470 780 1250
Digital CB Digital Toxic 6.9 410 610 1020 1640
Digital Toxic Digital Toxic 4.3 650 980 1640 2620
1- When sizing a system's 24 V supply, a 1 A inrush current with a 1 ms duration should be
considered for each device on the power supply

Assumes transmitter was ordered with relays

3.6.5 Instructions for Power and Analog Output

WARNING!

Read all electrical warnings and wiring requirements before connecting power to the S5000.

Failure to follow this warning can result in serious personal injury or death.

Connector for HART analog output and power are provided to increase ease of connecting power.

Connect power and remote sensor cable shields to shield terminals on the main PC board. Provide shield terminations
inside he sensor housing as indicated on the installation outline drawing.

(1) Remove the cover by turning counter-clockwise.

(2) Remove the display module to expose the wiring terminations and sensor connections.

(3) Remove the 5.08 mm pitch connector for power supply. The power connector is larger than other 3.81 mm pitch
connectors.

(4) Use a small flat head screw driver to open wire entries on connector.

(5) Insert wires to connector so that when installed each wire is in the correct location.

1. +DC

2. -DC

3. mA1 - analog output of sensor 1

4. mA2 - analog output of sensor 2

(6) Tighten screws on connector and tug gently on wires to ensure they are secure.

(7) Attach the connector to the board stack.

General Monitors S5000 35


US 3 Installation

(8) Make sure the appropriate wires are in the correct terminals.

(9) Remove enough of the wire housing to expose the 3-4 inches of the cable shielding, but do not expose so much that it
goes beyond the cable entry.

(10) Attach the cable, shielding exposed, to the grounding point.

(11) Replace the display module. Push firmly on the board stack where indicated.

(12) Replace the S5000 cover by turning clockwise. Be sure to align threads to avoid cross-threading.

Figure 27 Connecting Power and Grounding Cable

NOTICE

Ensure that the electronics assembly is fully engaged in the mounting holes. If not fully seated, the touch interface
performance can be negatively affected.

Care must be taken to ensure the S5000 inside glass surface glass is free of smudges/dirt and grease. Dirt and grease can
interfere with the touch interface of the display.

S5000 Installation Outline Drawings

Table 7 Installation Outline Drawings

Model Document No.


S5000 324102

General Monitors S5000 36


US 3 Installation

3.6.6 Relay Electrical and Power Connections

Relay Board Stack Overview

The S5000 can be purchased with three relays. Two of the relays can be configured for either de-energized (default) or
energized and latching or non-latching. The third relay is a dedicated fault relay.

All electrical connections to internal relays can be made directly on the PC board. The board is labeled for Normally Open 
(NO) and Normally Closed (NC) de-energized state.

Figure 28 PC Board with Relays

Relay Specifications

Table 8 Relay Specifications

Temperature Range -40 to +85 °C (-40 to 185 °F)


Relays SPDT (Single Pole Double
Throw)
Fault Normally Energized
Warning Configurable
Alarm Configurable
Relay Rating
125 or 250 VAC (Resistive) 5A, 100K Cycles

1.6 HP @ 250VAC
30 VDC (Resistive) 5A, 100K cycles

General Monitors S5000 37


US 3 Installation

If using AC power, the relay wires should not be run within the same conduit or cable tray as the DC power supplied to the
S5000 or the S5000 junction box. A separate wire entry on the device should be used for AC power connected to the relays.
The S5000 is built with an additional wire entry to allow this.

Exceeding the volt-amp rating of the relay can cause damage to the switching contacts.

Relay Connections to Inductive Loads

If connecting the relays to motors, fluorescent lighting, or other inductive loads, it is necessary to suppress any sparks or
inductive feedback that may occur at the relay contact. These effects may render the unit inoperative.

One way to reduce these effects is to install a Quencharc® (P/N 630413) across the load being switched.

Fault Relay Wiring and Configurations

The Fault relay state in non-fault operating condition is Energized and terminal connections are supplied for Normally
Closed and Normally Open.

The energized fault relay setting provides an electrical path for fail-safe relay operation. In the event of any failure, including
loss of power, the relay will change to the de-energized state to indicate a fault condition.

The Fault relay state cannot be reconfigured.

Relay Energy State and Terminal Connections

The S5000 relay states are labeled for the default de-energized state. The alarm/warning relay energy state can be changed
on the device, which will exchange the normally open and normally closed terminals. The preferred relay energy state
should be determined before making connections. Table 9 shows the terminal connections by energy state and is
applicable to both relay 1 and relay 2.

Table 9 Relay Terminal Connections by Energy State

Energy State NC (Normally Closed) NO (Normally Open)


De-Energized (default) Closed Open
Energized Open Closed

General Monitors S5000 38


US 4 Operation

4 Operation

WARNING!

Refer to 9 Appendix: General Certification Information before installation and operation.

Failure to follow this warning can result in serious personal injury or death.

The S5000 Series gas monitor is factory calibrated and shipped with the most common default options to minimize set up
effort. Using sensors, the device tests the ambient air and triggers an alarm as soon as the gas exceeds a specific
concentration level.

4.1 Startup

4.1.1 Initial Startup

The first time the S5000 is powered on, the following will appear on the display:
• S5000
• Software Version No.
• Sensor Warm-up
• Wait

The S5000 will remain in Start-up mode in which the fault relay is de-energized and the analog output is 3.5 mA by default.
The time the S5000 stays in Start-up mode depends on the sensor.

Oxygen and Carbon Monoxide sensors require a 30 minute warm-up time before being fully functional. The device will be in
fault during the 30 minute start-up time. All other sensor start-up times vary, the unit will be in fault for the first 2 minutes of
the start-up time and the analog output signal will be at the maintenance level (3.5 mA default).

A full calibration is recommended after one hour of a sensor being installed and acclimated to the environmental conditions.
See 5 Calibration for calibration details.

4.1.2 Sensor Warm Up Times

WARNING!

For optimal sensor performance, allow sensors 24 hours to acclimate to application conditions before performing an initial
calibration

Failure to follow this warning can result in serious personal injury or death.

IR Sensor: ≤ 10 min

H2S: ≤ 5 min

SO2: ≤ 5 min

Cl2: ≤ 10 min

NH3: ≤ 5 min

Cat Bead: ≤ 5 min

General Monitors S5000 39


US 4 Operation

Passive MOS: ≤ 5 min

Passive Cat Bead: ≤ 5 min

CO: 30 min (see note)

O2: 30 min

Carbon Monoxide sensor gas codes D10, D11, D12, and D14 may require initial warm-up periods longer than 30 minutes. If
the 30-minute warm up is over, the sensor may show a positive reading that exceeds alarm levels. In the event of a short-
term power outage, the warm-up time for carbon monoxide gas codes D10, D11, D12, and D14 will be significantly less.
See the table below to estimate required warm-up time.

Table 10 Warm Up Times CO

Reading Level < 10ppm1 < 1ppm


Time left unpowered Time to warm up
1 min ≤ 5 min ≤ 5 min
8 hours ≤ 30 min ≤ 30 min
5 days ≤ 30 min 2-4 hrs
1 month ≤ 2 hrs 6-10 hrs
1- The minimum alarm level of all three CO sensors is 10ppm. Below
this level would not cause an alarm on the device.

4.1.3 Startup after Power Failure

If the S5000 loses power, all of the settings are saved to the internal memory. When power is restored to the device, it will
go back to the same settings as before the loss of power. To check the settings, go through the menu or view on the X/S
Connect App.

4.2 Settings

The S5000 is a tool free transmitter. The infrared EZ touch button on the face of the display can be used to navigate through
the menu structure. The button is designed for use with fingers with a “press” and “release” action, and works best without
gloves.

The button works the same as the magnet does with the S4000 menu.

Changing a value

(1) Press and hold EZ Touch button.

(2) Wait for relevant menu to scroll (each menu scrolls twice).

(3) Release to enter menu while it is scrolling.

(4) Use Press and Release function to change values.

The EZ touch interface can be disabled, but will require a password. Menu settings can also be activated using a magnet
on the General Monitors logo.

Values that are changed in the menu are saved after main setup loop “Finished”, except for sensor range selection under
sensor setup. Each menu ends with "Finished?" scrolling twice on the display. If the button is not touched during

General Monitors S5000 40


US 4 Operation

"Finished?", the menu will begin at the start again and scroll through the options and values again. The new values that
were entered will be displayed on the first pass.

When user exits a menu by touching "Finished?" and there isn't a second sub menu, the previous menu will be displayed
starting at the menu that had just been used.

Some instrument settings are only configurable through Bluetooth® X/S Connect App, Modbus, or HART. See
4.4 Setting only configurable via Bluetooth®, Modbus, or HART.

4.2.1 Instrument Settings

The following settings are saved to the device memory and will not change if the sensor type is changed.

(1) Scroll to Settings.

(2) Select Instrument.

(3) Select to enter the menu.

General Monitors S5000 41


US 4 Operation

Table 11 Default Device Settings

Setting Menu 1 Default Range of options


Sensor Setup Sensor 1/Sensor 2
Disable Enabled Enable/Disable
Gas Label and Unit see Table 12 see Table 12
Range see Table 12 see Table 12
Cal Level see Table 12 see Table 12
Warning Level see Table 12 see Table 12
Alarm Level see Table 12 see Table 12
Alarm Action see Table 12 see Table 12
Sensor Reset No Yes/No
Relay Settings Relay Zone/Mapping Common Common/Discrete
Relay 1/Relay 2 De-energized Relay 1/Relay 2

Energized/De-energized Energized/De-energized
HART Enabled/Disabled Enabled Enable/Disable
HART AO Analog Output Enabled 1.25 or 3.5
Cal Alert Enable/Disable Enabled Enable/Disable
Bluetooth® Enable/Disable Enabled Enable/Disable
Modbus Baud Rate 19200 2400-115200 BPS
Format 8-n-1
Address 1 1-247
EZ Touch Enable/Disable Enabled
Swap Delay Enabled Enabled Enable/Disable
UI Password Enabled/Disabled Disabled Enable/Disable

New Password No Yes/No


Transmitter Reset No No Yes/No

General Monitors S5000 42


US 4 Operation

Table 12 Digital Sensor Default Settings

Warn Alarm
Range Display Alarm Cal Cal Cal
Thread Unit Warn Default Alarm Default Alarm Alarm Range Range
Gas (Code) TruCal Default Resolution Action Level Level Level
Type3 2 5 Default Default AO4 Default AO4 Min Max Min max
Default Default Min Max
(mA) (mA)
Ammonia Coarse 0-100 1 PPM 10 5.6 20 7.2 Increasing 25 2 100 0-10 0-100 1 FS1
(D40)
Non-Latch
Ammonia Coarse 0-1000 10 PPM 100 5.6 200 7.2 Increasing 300 100 1000 190 1000 290 FS1
(D41)
Non-Latch
Carbon Fine 0-100 1 PPM 10 5.6 30 8.8 Increasing 60 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
(D10)
Carbon Fine 0-500 1 PPM 50 5.6 150 8.8 Increasing 300 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
(D11)
Carbon Fine 0-1000 1 PPM 100 5.6 300 8.8 Increasing 400 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
(D12)
Carbon Fine 0-100 1 PPM 10 5.6 30 8.8 Increasing 60 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
H2 Resistant
(D14)
Carbon Fine 0-100 1 PPM 10 5.6 30 8.8 Increasing 60 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
(D36)
Carbon Fine 0-500 1 PPM 50 5.6 150 8.8 Increasing 300 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
(D37)

General Monitors S5000 43


US 4 Operation

Warn Alarm
Range Display Alarm Cal Cal Cal
Thread Unit Warn Default Alarm Default Alarm Alarm Range Range
Gas (Code) TruCal Default Resolution Action Level Level Level
Type3 2 5 Default Default AO4 Default AO4 Min Max Min max
Default Default Min Max
(mA) (mA)
Carbon Fine 0-1000 1 PPM 100 5.6 300 8.8 Increasing 400 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
(D38)
Carbon Fine 0-100 1 PPM 10 5.6 30 8.8 Increasing 60 10 1000 0-10 0- 5 FS1
Monoxide 1000
Non-Latch
(D39)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 50 5 60 0-202 0-100 10 100
Bead
Non-Latch
5% Methane
(D60)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 57 5 60 0-202 0-100 10 100
Bead
Non-Latch
4.4%
Methane
(D65)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 29 5 60 0-202 0-100 10 100
Bead
Non-Latch
2.1%
Propane
(D61)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 36 5 60 0-202 0-100 10 100
Bead
Non-Latch
1.7%
Propane
(D66)

General Monitors S5000 44


US 4 Operation

Warn Alarm
Range Display Alarm Cal Cal Cal
Thread Unit Warn Default Alarm Default Alarm Alarm Range Range
Gas (Code) TruCal Default Resolution Action Level Level Level
Type3 2 5 Default Default AO4 Default AO4 Min Max Min max
Default Default Min Max
(mA) (mA)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 45 5 60 0-202 0-100 10 100
Bead 1.05%
Non-Latch
Heptane
(D62)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 56 5 60 0-202 0-100 10 100
Bead 0.85%
Non-Latch
Heptane
(D67)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 61 5 60 0-202 0-100 10 100
Bead 0.8%
Non-Latch
Nonane
(D63)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 70 5 60 0-202 0-100 10 100
Bead 0.7%
Non-Latch
Nonane
(D68)
Catalytic Fine 0-100 1 % LEL 10 5.6 30 8.8 Increasing 20 5 60 0-202 0-100 10 100
Bead 4%
Non-Latch
Hydrogen
(D64)
Chlorine Coarse 0-5 0.1 PPM 0.5 5.6 1.0 7.2 Increasing 2.0 0.3 5.0 0-1.0 0-5.0 0.1 FS
(D30)
Non-Latch
Chlorine Coarse 0-10 0.1 PPM 0.5 4.8 1.0 5.6 Increasing 2.0 0.3 10.0 0-1.0 0-10.0 0.1 FS1
(D31)
Non-Latch
Chlorine Coarse 0-20 0.1 PPM 2.0 5.6 4.0 7.2 Increasing 10.0 0.6 20.0 0-1.0 0-20.0 0.1 FS1
(D32)

General Monitors S5000 45


US 4 Operation

Warn Alarm
Range Display Alarm Cal Cal Cal
Thread Unit Warn Default Alarm Default Alarm Alarm Range Range
Gas (Code) TruCal Default Resolution Action Level Level Level
Type3 2 5 Default Default AO4 Default AO4 Min Max Min max
Default Default Min Max
(mA) (mA)
Non-Latch
Hydrogen Fine 0-10.0 0.1 PPM 1.0 5.6 3.0 8.8 Increasing 5.0 1.0 100 0-10 0-100 5 FS1
Sulfide (D20,
Non-Latch
D83)
Hydrogen Fine 0-50.0 0.1 PPM 5.0 5.6 15.0 8.8 Increasing 40.0 1.0 100 0-10 0-100 5 FS1
Sulfide (D21,
Non-Latch
D84)
Hydrogen Fine 0- 0.1 PPM 10.0 5.6 30.0 8.8 Increasing 40.0 1.0 100 0-10 0-100 5 FS1
Sulfide (D22, 100.0
Non-Latch
D85)
Hydrogen Fine 0-20.0 0.1 PPM 6.0 8.8 12.0 13.6 Increasing 10.0 1.0 19 0-10 0-20 5 FS1
Sulfide (D24,
Non-Latch
D77)
Hydrogen Fine 0-50.0 0.1 PPM 15.0 8.8 30.0 13.6 Increasing 25.0 3.0 48 0-10 0-50 5 FS1
Sulfide (D25,
Non-Latch
D78)
Hydrogen Fine 0- 0.1 PPM 30.0 8.8 60.0 13.6 Increasing 50.0 5.0 95 0-10 0-100 5 FS1
Sulfide (D26, 100.0
Non-Latch
D79)
Hydrogen Fine 0-10.0 0.1 PPM 1.0 5.6 3.0 8.8 Increasing 5.0 1.0 100 0-10 0-100 5 FS1
Sulfide (D27,
Non-Latch
D86)
Hydrogen Fine 0-50.0 0.1 PPM 5.0 5.6 15.0 8.8 Increasing 40.0 1.0 100 0-10 0-100 5 FS1
Sulfide (D28,
Non-Latch
D87)
Hydrogen Fine 0- 0.1 PPM 10.0 5.6 30.0 8.8 Increasing 40.0 1.0 100 0-10 0-100 5 FS1

General Monitors S5000 46


US 4 Operation

Warn Alarm
Range Display Alarm Cal Cal Cal
Thread Unit Warn Default Alarm Default Alarm Alarm Range Range
Gas (Code) TruCal Default Resolution Action Level Level Level
Type3 2 5 Default Default AO4 Default AO4 Min Max Min max
Default Default Min Max
(mA) (mA)
Sulfide (D29, 100.0 Non-Latch
D88)
Hydrogen Fine 0-20.0 0.1 PPM 6.0 8.8 12.0 13.6 Increasing 10.0 1.0 19 0-10 0-20 5 FS1
Sulfide (D42,
Non-Latch
D80)
Hydrogen Fine 0-50.0 0.1 PPM 15.0 8.8 30.0 13.6 Increasing 25.0 3.0 48 0-10 0-50 5 FS1
Sulfide (D43,
Non-Latch
D81)
Hydrogen Fine 0- 0.1 PPM 30.0 8.8 60.0 13.6 Increasing 50.0 5.0 95 0-10 0-100 5 FS1
Sulfide (D44, 100.0
Non-Latch
D82)
Oxygen FM Fine 0-25.0 0.1 % 19.5 16.48 18.0 5.15 Increasing 20.8 17.0 24 5-25 5-25 15 25
(D15)
Non-Latch
Oxygen Fine 0-25.0 0.1 %VOL 19.5 5.248 18.0 5.152 Decreasing 20.8 15.0 25 15-25 15-25 15 25
(D16)
Non-Latch
Low Oxygen Fine 0-25.0 0.1 % 1 4.64 5.28 Increasing 20.8 0.2 25 0-2 0-25 20.8 FS1
(D17) Non-Latch (Room
Air)
Sulfur Coarse 0-25.0 0.1 PPM 2.0 5.28 5.0 7.2 Increasing 10.0 0.4 25 0-5 0-25 2.5 25
Dioxide Non-Latch
(D50)
Sulfur Coarse No 0-100 1 PPM 10 5.6 20 7.2 Increasing 10 1 100 0-25 0-100 7 FS1
Dioxide
(D51) Non-Latch
1- FS = Full Scale range

General Monitors S5000 47


US 4 Operation

Warn Alarm
Range Display Alarm Cal Cal Cal
Thread Unit Warn Default Alarm Default Alarm Alarm Range Range
Gas (Code) TruCal Default Resolution Action Level Level Level
Type3 2 5 Default Default AO4 Default AO4 Min Max Min max
Default Default Min Max
(mA) (mA)
2- The maximum range value on catalytic bead cannot be set below 20%.
3- Class 1 Division 2 only sensors do not have a flame arrestor (aka Frit). Course threads on the sensor assembly and sensor body are used to prevent a customer from
installing into a Class 1 Division 1 sensor body.
4- Analog Output for default range and alarms
5- Display resolution is not a configurable option
- TruCal not available

- Adaptive Environmental Compensation (AEC) Equipped

- AEC and Diffusion Supervision Equipped

General Monitors S5000 48


US 4 Operation

Table 13 IR Sensor Default Settings

Alarm Alarm Cal Cal Display


Range Unit Warn Warn Default Alarm Cal Level Alarm Alarm
Gas (Code) Range Default AO1 Action Level Level Resolution
Default Default Default AO1 (mA) Default Default Min Max 2
(mA) Default Min Max
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% 1

5% Methane Non-Latch
(R31)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% 1

4.4% Non-Latch
Methane
(R43)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

2.1% Non-Latch
Propane
(R32)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

1.7% Non-Latch
Propane
(R44)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Hexane Non-Latch
(R34)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Pentane Non-Latch
(R35)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Ethylene Non-Latch

General Monitors S5000 49


US 4 Operation

Alarm Alarm Cal Cal Display


Range Unit Warn Warn Default Alarm Cal Level Alarm Alarm
Gas (Code) Range Default AO1 Action Level Level Resolution
Default Default Default AO1 (mA) Default Default Min Max 2
(mA) Default Min Max
(R36)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Butane Non-Latch
(R37)
IR400 0-100 0-10.0 PPM 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Ethane Non-Latch
(R38)
IR400 0-100 0-50.0 PPM 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Hexane IEC Non-Latch


(R45)
IR400 0-100 0-100.0 PPM 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Pentane Non-Latch
IEC (R46)
IR400 0-100 0-25.0 %VOL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Butane IEC Non-Latch


(R47)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Ethane IEC Non-Latch


(R48)
IR400 0-100 0-100 % LEL 30 8.8 60 13.6 Increasing 50 5 60 Fixed at 50% LEL 1

Ethylene Non-Latch
IEC (R50)
IR700 0-2000 0-2000 PPM 600 8.8 1200 13.6 Increasing 1000 100 1900 1000 20

General Monitors S5000 50


US 4 Operation

Alarm Alarm Cal Cal Display


Range Unit Warn Warn Default Alarm Cal Level Alarm Alarm
Gas (Code) Range Default AO1 Action Level Level Resolution
Default Default Default AO1 (mA) Default Default Min Max 2
(mA) Default Min Max
Carbon Non-Latch
Dioxide
(R51)
IR700 0-5000 0-5000 PPM 1500 8.8 3000 13.6 Increasing 2500 250 4750 2500 50

Carbon Non-Latch
Dioxide
(R52)
IR700 0- 0-10000 PPM 3000 8.8 6000 13.6 Increasing 5000 500 9500 5000 100
10000
Carbon Non-Latch
Dioxide
(R53)
IR700 0- 0-30000 PPM 9000 8.8 18000 13.6 Increasing 15000 1500 28500 15000 300
30000
Carbon Non-Latch
Dioxide
(R54)
IR700 0- 0-50000 PPM 15000 8.8 30000 13.6 Increasing 25000 2500 47500 25000 500
50000
Carbon Non-Latch
Dioxide
(R55)
1- Analog Output for default range and alarms
2- Display resolution is not a configurable option

General Monitors S5000 51


US 4 Operation

Table 14 Passive Sensor Default Settings

Warn Alarm Alarm Cal Cal Cal


Range Unit Range Warn Alarm Alarm Alarm Display
Gas (Code) Range Default Default AO Action Level Level Level
Default Default Options Default Default Min Max Resolution
AO (mA) (mA) Default Default Min Max
Passive CB 0- 0-100 % LEL Fixed at 30 8.8 60 13.6 Increasing 50 5 60 20 90 1%
(C04, C07, C11) 100% 0-100
Non-Latch
Passive CB Low 0-20% 0-20 % LEL Fixed at 6 8.8 12 13.6 Increasing 10 1 19.8 4 18 0.10%
Range (C09, 0-10
Non-Latch
C10) Fixed at 30 60 5 0.5 9.9 2 9
0-20
Passive MOS 0-100 0-100 PPM Fixed at 30 8.8 60 13.6 Increasing 50 5 95 Fixed at 1 ppm
(M04) 0-100 50 ppm
Non-Latch
ppm
Passive MOS 0-50 0-50 PPM Fixed at 15 8.8 30 13.6 Increasing 25 2 47 Fixed at 1 ppm
(M05) 0-50 25 ppm
Non-Latch
ppm
Passive MOS 0-20 0-20 PPM Fixed at 6 8.8 12 13.6 Increasing 10 1 19 Fixed at 1 ppm
(M06) 0-20 10 ppm
Non-Latch
ppm

General Monitors S5000 52


US 4 Operation

4.2.2 Sensor Setup

Configure gas unit, range, calibration level (i.e. span value), warning and alarm levels and whether they are latching or non-
latching.

The Sensor Setup menu will go through each sub menu before exiting Sensor Setup.

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when Sensor Setup appears (first option).

(4) The following menus are available under Sensor Setup.

When "Finished?" is selected under each of the following menus, the next menu will start.

a) Sensor 1/Sensor 2

b) Disable

c) Gas Label & Unit

d) Range

e) Cal Level (Span Value)

f) Warning Level

g) Warning Action

h) Alarm Level

i) Alarm Action

j) Diffusion Supervision

k) Reset

(5) Select "Finished?" after the Reset menu to go back to the main Setup menu.

Disable Sensor (for Digital Sensors Only)

When removing a sensor from the transmitter while under power, the S5000 will enter a Sensor Missing fault condition after
the two minute Swap Delay period has expired (if enabled). If Swap Delay is disabled, the transmitter will go into Sensor
Missing fault immediately after removing the sensor from the transmitter. If the system is off at the time a sensor is
removed, the transmitter will go into fault after its startup sequence. This fault condition can be removed by disabling the
affected sensor position.

Disabling a sensor removes the fault and stops communications with the sensor, the sensor’s reading on the display is
removed, and the mA channel for that sensor position is set to 0 mA. By default, the S5000 has the Sensor 2 position
disabled. If at any time a sensor is connected to a position that is disabled, the device will automatically enable that sensor
position.

To disable the sensor after removal:

(1) Go to Sensor Setup menu.

(2) Touch button when Disable appears on screen.

General Monitors S5000 53


US 4 Operation

(3) Select Sensor 1 or Sensor 2.

Current status (enabled/disabled) displays.

(4) Touch button to toggle to desired status.

(5) Select “Finished?”.

To remove a sensor that is already installed, the sensor must first be removed. The sensor cannot be disabled
while it is installed on the transmitter.

The device only allows a sensor to be disabled after the transmitter has gone into Sensor Missing fault.

Only one sensor can be disabled at a time. The transmitter will not allow both sensors to be disabled at the same
time. This only applies to digital sensors (gas codes starting with D).

Gas Label and Unit

Change the gas unit displayed. For toxic sensors, select from ppm, mg/m3, and µm/mol.

(1) Go to Sensor Setup menu.

(2) Touch button when Gas Label & Unit appears on screen.

The Gas type and unit will appear on the screen.

(3) Touch button while the units scroll to change to another unit.

(4) Touch button when “Finished?” is displayed to exit.

Sensor Range

Set the range of the sensor. To see what ranges are possible for the sensor type, see Table 12 , Table 13 , and Table 14 .

Ensure that the IR700 range is configured to match that of the sensor.

(1) Go to Sensor Setup menu and select “Finished?” until Range displays.

Range will scroll across the screen and the current Range will be displayed.

(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move through values more slowly.

(3) When the desired value is displayed, wait until “Finished?” scrolls across screen.

(4) Touch button when “Finished?” is displayed to exit.

Cal Level (i.e. Span Value)

Set gas concentration used during calibration.

(1) Go to Sensor Setup menu and select “Finished?” until Cal Level displays.

Cal Level will scroll across the screen and the current Cal Level will be displayed.

(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move through values more slowly.

(3) When the desired value is displayed, wait until "Finsihed?" scrolls across screen.

General Monitors S5000 54


US 4 Operation

(4) Touch button when “Finished?” is displayed to exit to save the sensor parameter.

Warning Settings

Set Warning level and whether warning is latching or non-latching.

(1) Go to Sensor Setup menu and select “Finished?” until Warning displays.

Warning will scroll across the screen and the current Warning level will be displayed.

(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move through values more slowly.

(3) When the desired value is displayed, wait until Latching option appears.

(4) To change warning to latching or non-latching, touch button when the option is displayed.

(5) Touch button when “Finished?” is displayed to exit.

Alarm Settings

Set Alarm level and whether warning is latching or non-latching.

(1) Go to Sensor Setup menu and select “Finished?” until Alarm displays.

Alarm will scroll across the screen and the current Alarm level will be displayed.

(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move through values more slowly.

(3) When the desired value is displayed, wait until Latching option appears.

(4) To change alarm to latching or non-latching, touch button when the option is displayed.

(5) Touch button when “Finished?” is displayed to exit.

Alarm Relay Settings is the last Sensor Setup menu. “Finished?” will save changes and take the user back to
Setup menus.

If the Alarm Relay is configured to be non-latching, the Alarm Relay output must be connected to an auxiliary
device that performs the latching function.

Diffusion Supervision

Diffusion supervision actively monitors the sensor inlet for obstructions. If an obstruction is detected, the sensor will go
into a fault mode to alert users and the control room that it is not seeing gas due to an obstruction. Objects residing directly
on or in the sensor inlet that result in a significant impact to the gas path are very likely to be detected by Diffusion
Supervision. Examples include paint, tape, water, and dirt. Small amounts of these materials can be visible on the inlet
while not impacting the gas path enough to trigger a Diffusion Supervision Fault. A fault signal will only be sent out when
the system determines that the amount of material that has accumulated on or inside the sensor inlet is negatively affecting
the gas path.

Even if a Diffusion Supervision Fault has not been triggered, it is good practice to clear any foreign material from
the sensor inlet if any is observed while inspecting the sensor.

To enable or disable Diffusion Supervision:

(1) Go to Sensor Setup menu.

General Monitors S5000 55


US 4 Operation

(2) Touch button when Diffusion Supervision appears.

(3) Select Sensor 1 or Sensor 2.

Current status (enabled/disabled) displays.

(4) Touch button to toggle to desired status.

(5) Select "Finished?".

Sensor Reset - Last option in Sensor Setup

NOTICE

The sensor goes into a Sensor Configuration Reset fault (F007) and must be calibrated after a sensor reset.

 All settings, including Alarm Set Point and Calibration Values, will be returned to factory defaults.

The sensor default values can be restored by resetting the sensor. During sensor reset, the analog output displays the
sensors current gas value (i.e. 0 % LEL = 4 mA) during the 120 s count down.

To reset sensor to factory default:

(1) Go to Sensor Setup menu.

(2) Go through menus selecting “Finished?” at the end of each option.

(3) Touch button when Reset Setting scrolls across screen.

(4) Touch button to change to yes.

(5) Touch button when “Finished?” is displayed to exit.

This is the last Sensor Setup menu. “Finished?” will take the user back to Setup Menus and will save any
changes made.

4.2.3 Relay Settings

Relay Setup is used to change the relay zone mapping and energized/de-energized settings.

Relay Zone/Mapping

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when Relay Setup is displayed.

Current setting of Zone/Mapping will be displayed (Common/Discrete).

(4) Touch button again will toggle the setting.

(5) If setting is changed, touch button when “Finished?” is displayed to save the change and move on to next setup.
If no change is made, the menu will move on to next setup after two screen scroll rotation.

Relay 1 and Relay 2 can be configured for common and discrete modes via the device display menu or X/S Connect app.

General Monitors S5000 56


US 4 Operation

Common mode is the default relay mapping setting. In Common mode, Relay 1 is actuated by Alarm 1 on either sensor,
and Relay 2 is actuated by Alarm 2 on either sensor.

Figure 29 Common Mode Relay Map and Alarm Actuation

Discrete mode allows a separate action for each sensor. Relay 1 is actuated by Sensor 1 alarms and Relay 2 is actuated
by Sensor 2 alarms

Figure 30 Discrete Mode Relay Map and Alarm Actuation

Horn mode is designed to allow local acknowledgment of a relay-triggered horn, while the alarm state is still present. All
alarms on both sensors trigger both relays, however the second relay can be acknowledged through the "Reset" entry in UI
menu by touching the EZ touch button or applying magnet to the GMI logo.

General Monitors S5000 57


US 4 Operation

Figure 31 Horn Mode Relay Map and Alarm Actuation

Energized/De-energized

(1) Relay 1 and Relay 2 will scroll across screen alternatively.

(2) Touch button when Relay 1 or Relay 2 is scrolling.

(3) Touch button to show the current setting (energized/de-energized).

(4) Touch button again will toggle the setting.

(5) Touch button when “Finished?” is displayed to exit from Relay setting.

4.2.4 HART Settings

There are two HART menus under Setup. HART Setup is used to Enable/Disable HART. HART AO Setup is used to
configure the analog output signal for HART.

Enable/Disable HART

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when HART Setup is displayed.

The current HART setting is displayed (Enabled or Disabled).

(4) To change the availability of HART, touch button to change to the opposite state.

(5) Touch button when "Finished?" is displayed to exit.

4.2.5 HART AO
(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when HART AO Setup is displayed.

The current HART analog output signal is displayed.

(4) To change the HART output signal, touch button to change to either 1.25 or 3.5.

General Monitors S5000 58


US 4 Operation

(5) Touch button when "Finished?" is displayed to exit.

4.2.6 Calibration Alert

Sensors with TruCal technology actively monitor sensor and adjust sensitivity without any manual intervention. When
equipped with Diffusion Supervision, TruCal will also monitor the sensor inlet for obstructions while Diffusion Supervision
is enabled. These sensors do not need to be calibrated on a static maintenance cycle. When a manual calibration is
recommended, the sensor will detect this and slow flash either the left green LED or right green LED indicating that
calibration is recommended for sensor 1 or sensor 2 respectively. Users can also enable Calibration Alert so that an analog
output signal is sent to the control room when a calibration is recommended. Whether or not the calibration alert is enabled,
if gas is detected by the sensor, the S5000's analog output and display will follow the gas reading.

To Enable or Disable Calibration Alert:

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup appears.

(3) Touch button when Cal Alert is displayed.

The current Cal Alert setting is displayed (Enabled/Disabled).

(4) Touch button to toggle desired status.

(5) Touch button when "Finished?" is displayed to exit.

4.2.7 Bluetooth

Enable/Disable Bluetooth

If the device is ordered with Bluetooth, it will come enabled by default. If unit is not ordered with Bluetooth, it cannot be
retroactively added. This menu will not be displayed if the device was ordered without Bluetooth.

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when Bluetooth Setup is displayed.

The current Bluetooth setting is displayed (Enabled or Disabled).

(4) To change the Bluetooth availability, touch button to change to the opposite state.

(5) Touch button when "Finished?" is displayed to exit.

Bluetooth Pairing

The device memory has the ability to store up to 25 mobile devices in its memory. If a 26th device is paired, the memory
will erase the first device stored in its memory. Each time a device is paired, it is logged as an event that is viewable in the
X/S Connect App.

The S5000 blue LED in the middle left of the display will flash when a device is paired with it as a visual indication of the
pairing. Once paired with an S5000, the user will be able to connect to the same S5000 remotely and without a password,
unless over 25 other devices are paired with the same S5000 afterwards.

To pair with the S5000:

(1) Download the X/S Connect App from the Google Play Store or the iOS App Store.

General Monitors S5000 59


US 4 Operation

(2) Open the X/S Connect App.

(3) Select “Pair” for the S5000 that you would like to connect with.

(4) (First Time Only) Enter Pairing Code on S5000 display.

(5) (First Time Only) Accept pairing by touching button on S5000 display.

Bluetooth Security

The Bluetooth connection is encrypted and secured with a unique six digit pin that must be confirmed on the mobile device
and acknowledged on the detector display. All of the previously paired devices can be erased from the S5000 to provide
additional security and control.

To reset all device pairings:

(1) Scroll to Settings.

(2) Scroll and select Bluetooth

(3) Scroll and select “Reset All”.

(4) Select “Continue”.

NOTICE

Reset All will delete all paired device memory. All devices will have to re-initiate pairing at the device.

Bluetooth Tag ID

See 4.3.6 Bluetooth FCC/IC ID to view Bluetooth Tag ID.

4.2.8 Modbus Settings

The Modbus Setup menu consists of three configurable sub menus: Baud Rate, Format and Address.

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when Modbus Setup is displayed.

Modbus Baudrate

(1) Go to Modbus Setup menu. Baudrate option is the first submenu.

Baudrate will scroll across display followed by the current Baudrate setting.

(2) Touch the button to change to one of the following baudrate options:

a) 2400

b) 4800

c) 9600

d) 19K2

e) 38K4

General Monitors S5000 60


US 4 Operation

f) 115K

(3) Touch button when “Finished?” is displayed to exit.

Modbus Data Format

(1) Go to Modbus Setup menu and select “Finished?” until Format is displayed.

Format will scroll across display followed by the current Format setting.

(2) Touch the button to change to one of the following format options:

a) 8N1

b) 8N2

c) 8E1

d) 8O1

(3) Touch button when “Finished?” is displayed to exit.

Modbus Address

(1) Go to Modbus Setup menu and select “Finished?” until Address is displayed.

Address will scroll across display followed by the current Address setting.

(2) Hold finger on button to scroll quickly through values or touch button repeatedly to move through values more slowly.
Address options range from 1-247.

(3) Touch button when “Finished?” is displayed to exit.

4.2.9 EZ Touch Button

The device has a single button that allows the user to change configurations without the use of a magnet or any other tool.
Disabling the EZ Touch button requires the user to access menus with a magnet or via HART, Modbus, or Bluetooth.

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when EZ Touch is displayed.

The current EZ Touch setting is displayed (Enabled or Disabled).

(4) To change the EZ Touch availability, touch button to change to the opposite state.

After disabling and selecting “Finished?”, the user will need a magnet to enable the setting. Otherwise it will not be
possible to navigate the S5000 menus.

(5) Touch button when "Finished?" is displayed to exit.

4.2.10 Swap Delay

Swap Delay allows the user a brief window to change an XCell sensor without the device going into a fault condition. Once
a sensor is disconnected from the transmitter, the user will have 2 minutes to reconnect a sensor. During this time, the
device display and analog output will go to its Maintenance level. If a sensor is reconnected or replaced during the 2 minute

General Monitors S5000 61


US 4 Operation

window, the new sensor’s countdown sequence will begin and the analog output will remain at the Maintenance level. After
the sensor countdown is complete, the analog output will return to reporting a live gas reading. If a sensor is not
reconnected after the 2 minute window, the S5000 will enter a fault condition. All XCell Sensors have Swap Delay and do
not need to be disconnected from power while changing sensors. For more details on how to change sensors, see 6.2
Replacing an XCell Sensor. Swap Delay is enabled on all S5000 transmitters by default.

NOTICE

The transition to maintenance mode during the 2 minute Swap Delay window and sensor countdown will not trigger the
Fault Relay. The Fault Relay will only be triggered when the device enters a fault condition.

To change Swap Delay status:

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when Swap Delay is displayed.

The Swap Delay status is displayed (enabled or disabled).

(4) To change the Swap Delay status, touch button to change to either enabled or disabled.

(5) Touch button when "Finished?" is displayed to exit.

4.2.11 UI Password

Enable/Change Password

Enabling and changing the password is within the same menu. Enabling password will require the user to enter the
password before entering any of the settings menus. The password entry screen defaults to 0000 and is disabled by
default.

If the password is lost, call GM Customer Service at +1-949-581-4464.

(1) Hold finger on button while menu selections scroll across screen.

(2) Remove finger when Setup menu appears.

(3) Touch button when UI Password Setup is displayed.

The current UI Password setting is displayed (Enabled or Disabled).

(4) Touch button to change to the opposite state.

(5) Touch button when “Finished?” is displayed to save the enable/disable setting.

Next menu "New Password?" is displayed is for setting password. Default password is 0000 if enabled.

(6) To change password, touch button when “No” appears on screen to change option to “Yes”.

(7) Touch button while “Finished?” is displayed to save change.

(8) The password entry screen will display a current password and each digit will blink for 4 seconds starting from left.

(9) While blinking, touch button to change desired numbers. The range is from 0 to 9 for each digit and 0000 to 9999 for all
4 digits.

General Monitors S5000 62


US 4 Operation

(10) After changing the password, select “Finished?” to store the new password in the system.

Enter Password (if password is enabled)

If a password is enabled, the user must enter the password before entering any of the settings menus. The password entry
screen defaults to 000 and is disabled by default

(1) Holding finger on button will display rotating left arrow then “password”.

(2) Remove finger when UI Password appears.

(3) The password entry screen will display with 0 and each digit will blink for 4 seconds starting from left.

(4) While blinking, touch button to change the desired numbers. The range is from 0 to 9 for one digit and 0000 to 9999 for
all 4 digits.

(5) If the password is correct, holding finger on the button again will display rotating left arrow then go into the menu
items.

As long as the setup menu is active, no password is required to enter again. If menu was not active for
30 seconds, the password will be required again.

4.2.12 Transmitter Setting Reset

The transmitter default values can be restored by resetting the main board. The unit will restart after the reset and the unit
may go into sensor fault. During transmitter reset, the analog output displays 1.25 mA.

NOTICE

Verify all sensor settings (calibration level and alarm values) after resetting transmitter. Sensors may require recalibration
to clear a fault after resetting transmitter.

4.3 Info Menu - Viewing Device Status

This section describes how to view the following options:


• Display Current System Time
• Change System Time
• Sensor Life and Health Status
• Last Calibration Dates
• Non-critical Faults
• Device Tag
• Bluetooth ID

4.3.1 AO Type
(1) Go to Info menu and select AO Type.

The AO type selected for this unit, Source or Sink, will scroll across the screen twice.

(2) Touch button when "Finished?" is displayed to exit.

General Monitors S5000 63


US 4 Operation

4.3.2 View or Change System Time


(1) Go to Info menu and select System Time.

(2) Hold finger on button while menu selections scroll across screen.

(3) Remove finger when Info is displayed.

(4) Touch button when System Time is displayed.

The current date followed by the current time will scroll across the screen twice.

(5) Touch button when Change Time displays if the current date and time are not correct.

(6) If selected, change the following values using the button and selecting "Finished?" after each entry:

a) Month

b) Day

c) Year

d) Hour

e) Minutes

(7) Touch button when "Finished?" is displayed to exit.

You can also use the X/S connect app to sync time and date with a mobile device.

4.3.3 Last Calibration


(1) Go to Info menu and select Last Calibration.

The last Calibration and Zero Calibration dates will scroll across the screen twice.

(2) Touch button when "Finished?" is displayed to exit.

4.3.4 Non-Critical Fault Log


(1) Go to Info menu and select Non-Critical Fault.

Each of the last 10 faults will scroll across the screen twice.

(2) At the end of the fault log, menu goes back to Info menu options.

The user will not be asked to exit with “Finished?” at the end of this menu.

4.3.5 Device Tag


(1) Go to Info menu and select Device Tag.

The current Device Tag name will scroll across screen.

• DEFAULT - GM_S5K_0
• The device tag name can be changed via Modbus or HART and is limited to 8 characters.

(2) After the Device Tag name scrolls across screen twice, menu goes back to Info menu options.

General Monitors S5000 64


US 4 Operation

The user will not be asked to exit with “Finished?” at the end of this menu.

4.3.6 Bluetooth FCC/IC ID


(1) Go to Info menu and select Bluetooth FCC/IC ID.

The current Bluetooth FCC/IC ID will scroll across screen.

The Bluetooth FCC/IC ID cannot be changed and is unique to the device.

(2) Touch button when "Finished?" is displayed to exit.

4.3.7 Sensor Life and Health Status (only displayed if XCell Sensor is connected)
(1) Select Info.

(2) Select Sensor Status.

If only one sensor is installed, the device displays the Life & Health status for Sensor 1.

If two sensors are installed, "Sensor Selection" will be displayed.

(3) Select the desired sensor (Sensor 1 or Sensor 2) and touch button when "Finished?" is displayed.

The Life and Health information reports the general health of the sensor.

XCell sensors with TruCal (Hydrogen Sulfide and Carbon Monoxide) calculate current sensor sensitivity using automated
pulse checks. The pulse stimulates the sensor with a response similar to having actual calibration gas applied. The
stimulated response is compared to the last calibration and will make adjustments to sensitivity to match the last
calibration. When the required adjustment is greater than the accuracy of the algorithm's adjustment, the sensor will call for
a calibration.

All Digital Sensors, including catalytic bead and oxygen, will show life health status as 'Good' or 'Fair' with the
following calculation:

Good condition occurs when the current calibrated span sensitivity is greater than 50% of the way between the
initial calibration sensitivity and the end-of-life span sensitivity. Fair condition occurs when the current calibrated
span sensitivity is less than 50% of the way from the initial calibration sensitivity and the end of life span
sensitivity.

NOTICE

Using expired calibration gas or the incorrect calibration gas can result in a premature "Fair" status.

4.4 Setting only configurable via Bluetooth®, Modbus, or HART

The following settings are only configurable through Bluetooth®, Modbus, or HART.
• AO Custom Levels
• Relay Zone – Horn Mode
• Unit Device ID/Tag
• Alarm Direction

General Monitors S5000 65


US 4 Operation

• Alarm Enable/Disable

4.4.1 AO Custom Levels

S5000 offers three factory defined analog outputs (AO) level sets, i.e., 0 mA with HART disabled, or 3.5 mA and 1.25 mA
with HART enabled. Users can also specify their own custom AO levels for each transmitter mode. The
Maintenance AO level is used during start up, Reset Main Unit, and Transmitter Settings Reset. Output setting for oxygen
sensor calibration is not configurable. All changeable settings can be modified through Modbus/Bluetooth/HART.

Table 15 Default AO Values

HART HART Default Enabled: Custom 1 Custom 2


Mode
Disabled (3.5 mA) HART (1.25 mA) Default Default

Fault 0 mA 3.5 mA 1.25 mA 1.25 mA*


Calibration & 1.5 mA 3.5 mA 1.5 mA 1.5 mA* 1.5 mA*
Gas Check (All Other Sensors)

21.0 mA
(Oxygen Sensor)
Cal Alert 3.0 3.5 3.0 3.0 3.0
IR Cleaning 2.0 mA 2.0 mA*
Needed
Maintenance 3.5 mA 3.5 mA*
Zero Reading 4.04 mA
Over Range 21.7 mA

The range for the changeable AO settings is 0.000 mA to 3.750 mA. The change step is 0.025 mA. Please note the HART
communication may not work reliably if the current level is set to below 1.25 mA.

4.4.2 Relay Zone - Horn Mode

Horn mode is designed to allow local acknowledgment of a relay-triggered horn, while the alarm state is still present. All
alarms on both sensors trigger both relays, however the second relay can be acknowledged over
Modbus/Bluetooth/HART.

4.4.3 Unit Device ID/Tag

This is device tag/name used to identify the unit during HART and Bluetooth communication. Default is “GM_S5K_0 “.
Users can change this string to any characters from the following character set. The string length is limited to eight
characters.

General Monitors S5000 66


US 4 Operation

4.4.4 Alarm Direction

Relays can be triggered by an alarm threshold as gas readings increase or decrease. Most applications require increasing
alarm thresholds, except for Oxygen monitoring, which is most often a decreasing alarm.

The default alarm direction for Oxygen sensor is decreasing, or downward.

The default alarm direction for all other sensors is increasing, or upward.

The alarm direction can be set to increasing/upward or decreasing/downward.

Alarm direction for Oxygen sensor can be changed through UI directly, or through Modbus/Bluetooth/HART.

The alarm direction setting for all other sensors can only be changed through Modbus/Bluetooth/HART.

4.4.5 Alarm Enable/Disable

Users can disable alarms by changing this setting.

Default value is Enabled for this setting.

Users can change this setting to Enabled or Disabled.

General Monitors S5000 67


US 5 Calibration

5 Calibration
Calibration is the process of applying a known quantity of gas to the transmitter so that the transmitter can adjust the
precision and accuracy of the measurements made in normal operating mode. This process ensures that gas
measurements are as accurate as possible.

Calibration Warnings - Read before Calibrating

Although S5000 sensors are factory calibrated, another calibration is recommended once the unit is installed in its final
environmental destination.

WARNING!

Use zero gas when zeroing the S5000 transmitter if there is any possibility of background gas. Otherwise, improper
calibration could occur.

Use zero gas when zeroing IR700 sensors. Air contains CO2 and cannot be used. Nitrogen is recommended as a zero gas.
Otherwise, improper calibration could occur.

For optimal sensor performance, allow sensor to acclimate to application conditions for 24 hours before performing an initial
calibration.

Perform calibrations 24 hours after initial start up and at a frequency identified in 5.2 Calibration Frequency and 5.3
Calibration Frequency for XCell Sensors with TruCal (H2S & CO only) .

XCell Sensors with TruCal (CO & H2S) with Calibration Alert enabled provide an indication for “Calibration Recommended”
and “Calibration Required’. When either indication is provided by the device, calibrate the sensor.

Failure to follow this warning can result in serious personal injury or death.

5.1 Calibration Equipment

A gas cylinder with a known concentration of gas appropriate for the range of measurement is needed. Sensors come with
preset span gas values appropriate for the measurement range.

All of the calibration hardware for XCell sensors is included in a convenient calibration kit. It includes a carrying case that
can fit two cylinders, a regulator, tubing, and any other accessories required for each sensor type. The calibration kit can be
ordered separately or as a single line item with the gas cylinder. See Table 16 for a complete list of calibration cylinders
and kits.

The Digital Toxic, Oxygen, and Catalytic Bead sensors share the same sensor body design and use the same calibration
hardware. The calibration kit for IR400 sensors is ordered separately (P/N 1400270-2) and contains a regulator, tubing, and
other calibration accessories specific to the IR400. The calibration kit for IR700 sensors is P/N 31478-1.

Passive sensors use different calibration hardware to accommodate the different sized sensor housing. Passive MOS
sensors can be calibrated with either calibration gas cylinders or ampules. See Table 16 for P/N details.

The sensor guard and rubber calibration cap should be removed after calibration.

General Monitors S5000 68


US 5 Calibration

Table 16 Calibration Kits

Calibration Kit
P/N with
Gas Type Range Concentration Cylinder P/N Only
Calibration Kit
(No Cylinder)
Carbon 0-100 PPM 60 PPM 710882 710882-KIT1 CALKIT1
Monoxide 0-500 PPM 300 PPM 10027938 10027938-KIT1
0-1000 PPM 400 PPM 10028048 10028048-KIT1
Hydrogen 0-10 PPM 5 PPM 710414 710414-KIT1 CALKIT1
Sulfide 0-50 PPM 40 PPM 10028062 10028062-KIT1
0-100 PPM
0-20 PPM 10 PPM 1400255-1 1400250-1 N/A
0-50 PPM 25 PPM 1400255-3 1400250-3
0-100 PPM 50 PPM 1400255-5 1400250-5
0-500 PPM 250 PPM 10089547 10089547-KIT1 CALKIT1
Oxygen 0-25% 20.8% 10028028 10028028-KIT1 CALKIT1
Sulfur 0-25 PPM 10 PPM1 10028070 10028070-KIT2 CALKIT2
Dioxide
Chlorine 0-5 PPM 2 PPM2 710331 710331-KIT1 CALKIT1
0-10 PPM
0-20 PPM 10 PPM1 10028066 10028066-KIT1 CALKIT1
Ammonia 0-100 PPM 25 PPM2 10028076 10028076-KIT1 CALKIT1
XCell 0-100% LEL 2.5% Methane 10028032 10028032-KIT1 CALKIT1
Combustible (50 % LEL)
5% Methane
0-100% LEL 2.5% Methane
(57% LEL)
4% Methane
0-100% LEL 0.6% Propane 10028034 10028034-KIT1
(29% LEL)
2.1% Propane
0-100% LEL 0.6% Propane
(35% LEL)
1.7% Propane
Passive Cat 0-20% LEL 0.5% Methane 1400274-25 1400276-25 1400152-1 or
Bead 0-100% LEL 2.5% Methane 1400155-M 1400150-M 1400154-1

Passive Ampules Gas Ampules Gas N/A


MOS 0-100 PPM 50 PPM 50004-13 1400255-5 50009-9 1400250-5
0-50 PPM 25 PPM 5004-21 1400255-3 5009-16 1400250-3
0-20 PPM 10 PPM 50004-3 1400255-1 50009-10 1400250-1

General Monitors S5000 69


US 5 Calibration

Calibration Kit
P/N with
Gas Type Range Concentration Cylinder P/N Only
Calibration Kit
(No Cylinder)
Methane 0-100% 50% 1400275-1A 32575-1 1400270-2
(IR400)
Propane 0-100% 50% 1400275-6A 32575-6 1400270-2
(IR400)
Hexane 0-100% 50% 31486-1A 31468-7 1400270-2
(IR400)
Pentane 0-100% 50% 14002754-10A 32575-10 1400270-2
(IR400)
Ethylene 0-100% 50% 1400275-9A 32575-9 1400270-2
(IR400)
Butane 0-100% 50% 1400275-4A 32575-4 1400270-2
(IR400)
Ethane 0-100% 50% 1400275-8A 32575-8 1400270-2
(IR400)
0-2000 1000 PPM 1400262-41A 31478-5 1400270-2
Carbon 0-5000 2500 PPM 1400262-16A 31478-1 1400270-2
Dioxide 0-10000 5000 PPM 1400262-37A 31478-2 1400270-2
(IR700)
0-30000 15000 PPM 1400262-27A 31478-3 1400270-2
0-50000 25000 PPM 1400262-40A 31478-4 1400270-2
Note that for zeroing of the IR700 sensor, zero gas is required.
Air contains CO2 and cannot be used. Nitrogen is recommended
as the zero gas. The nitrogen cylinder P/N is 1400262-17.

Flow Rates:
CALKIT1 and 1400270-2 = 1 liter/min
1400152-1 and 1400154-1 = 0.5 liter/min
CALKIT2 = 0.25 liter/min

5.2 Calibration Frequency

The frequency of calibration gas testing depends on the operating time, chemical exposure, and type of sensor. Especially
in new installations or applications, it is recommended that the first sensors be calibrated more often to establish the
sensor performance in this particular environment.

For this, record the "as found" and "as left" values and track the percent adjustment over time. Then, gradually extend
calibration intervals until the percent adjustment is greater than the expected accuracy of the sensor.

General Monitors S5000 70


US 5 Calibration

5.3 Calibration Frequency for XCell Sensors with TruCal (H2S & CO only)

WARNING!

Certain conditions may prevent the performance of Diffusion Supervision and/or TruCal testing. If such a condition persists
for 90 days, the XCell sensor with TruCal will recommend a calibration which is communicated through the transmitter
display and LEDs. When Calibration Alert is enabled, the milli-amp signal output will also be driven to calibration alert
output. Perform a calibration when the device recommends it.

Failure to follow this warning can result in serious personal injury or death.

5.3.1 XCell Sensor with TruCal and Diffusion Supervision (CO&H2S Only)

Sensors with TruCal technology will adjust sensitivity without any manual intervention or calibration, unless called for by
the sensor. If the adjusted sensitivity from TruCal deviates too far from the last gas calibration sensitivity, the sensor will
recommend or, in extreme cases, require a calibration. When a calibration is recommended, the transmitter LED status
indicators slowly pulse green. Users can also enable a Calibration Alert function that will send a milli-amp signal on the
analog output to the control room when a calibration is recommended. When a calibration is required, the status LEDs will
flash yellow and the transmitter will go into a Calibration Required Fault.

With Diffusion Supervision and Calibration Alert enabled, the time between scheduled calibrations can be extended to 24
Months.

If calibration alert and/or diffusion supervision are disabled follow the calibration frequency recommendation set forth in
5.3.2 XCell Sensor with TruCal without Diffusion Supervision (CO&H2S Only).

5.3.2 XCell Sensor with TruCal without Diffusion Supervision (CO&H2S Only)

Sensors with TruCal technology will adjust sensitivity without any manual intervention or calibration, unless called for by
the sensor. If the adjusted sensitivity from TruCal deviates too far from the last gas calibration sensitivity, the sensor will
recommend or, in extreme cases, require a calibration. When a calibration is recommended, the transmitter LED status
indicators slowly pulse green. Users can also enable a Calibration Alert function that will send a milli-amp signal on the
analog output to the control room when a calibration is recommended. When a calibration is required, the status LEDs will
flash yellow and the transmitter will go into a Calibration Required Fault. Without Diffusion Supervision, regular calibration
of the sensor must still be conducted to confirm the sensor inlet is not obstructed.

Actual TruCal sensor performance will depend on the application, background gas exposure, and environment. To validate
XCell sensors with TruCal, it is recommended that users follow their regular calibration cycle and record the “as found” and
“as left” values, tracking the percent adjustment over time. Once a baseline is established, the calibration intervals can be
extended until the percent adjustment is greater than the expected accuracy of the sensor.

5.4 Calibration Types: Zero vs. Span

There are two types of calibration; zero and span calibration.

The zero calibration resets the baseline level reading to zero. If the target gas is suspected to be occasionally present, it is
best to also use a zero gas cylinder during the zero calibration. If the target gas is not present in the atmosphere, an
additional calibration cylinder is not required.

The span calibration is a process that involves first zeroing the baseline reading, then applying a known amount of target
gas so that the accuracy and precision of the transmitter can be adjusted to the known value. The known concentration of
gas is called a "Span Value" because it represents the span or reach of the calibration curve from zero to that value.

General Monitors S5000 71


US 5 Calibration

Figure 32 Calibration Curve

The Sensor Span Value (i.e. Cal Level) in the device menu should be the same as the concentration listed on the calibration
gas cylinder.

5.5 How to Zero Calibrate XCell and IR Sensors

NOTICE

If a password is enabled, you will not be able to proceed with the calibration without the password.

To abort, press the EZ touch button or apply magnet over GM logo before applying target gas.

In the event that a calibration cannot be completed, the user can acknowledge the FAIL by pressing the EZ
touch button or applying magnet to the GMI logo area until "Reset" is displayed and then release. The unit will
revert to the settings of the last successful calibration.

WARNING!

The regulator used with the zero gas cylinders should not be the same as the regulator used for the target gas. The zero gas
regulator can be contaminated with the target gas over time, and thus raise the detection baseline and make the sensor
less sensitive to the target gas.

Failure to follow this warning can result in serious personal injury or death.

If there is no target gas in the atmosphere around the sensor, using a zero gas cylinder is optional. Note that,
since air contains CO2, a zero gas cylinder is required for zeroing an IR 700 sensor. For IR700 sensors, nitrogen
is recommended as the zero gas.

To zero calibrate the sensor:

General Monitors S5000 72


US 5 Calibration

(1) Screw the sensor guard to the bottom of the sensor housing.

(2) Place the green calibration cap over the sensor guard inlet so that it is flush with the bottom of the sensor guard and
completely covers the sensor guard inlet.

(3) Attach the tubing to the plastic stem protruding through the green calibration cap.

(4) Screw the regulator onto the top of the zero gas cylinder.

If a password is enabled, you will need to enter it to access the menu system.

NOTICE

If the calibration cap cannot be used (such as for a remote sensor application), ambient wind conditions must be minimized
during calibration to avoid a calibration with increased sensitivity.

(5) Press EZ touch until Zero is displayed, and release.

(6) Once the screen displays Zero in Progress, turn on zero gas flow by turning knob on the regulator.

(7) Wait while the device displays Zero in Progress.

(8) Once the zero calibration is complete, a Zero PASS or FAIL is displayed.

If PASS displays, the procedure is complete.

If FAIL displays, the procedure was unsuccessful (see 6.5 Troubleshooting).

(9) Remove the zero calibration equipment from the device including the sensor guard and calibration cap.

Sensors with TruCal and have Diffusion Supervision enabled can go into Diffusion Supervision Fault if the green
Calibration Cap is left on after completing a zero calibration.

WARNING!

The green Calibration Cap must be removed from the sensor after a zero calibration. Failure to do so could restrict gas flow
to the sensor and result in erroneously low readings.

Failure to follow this warning can result in serious personal injury or death.

5.6 How to Calibrate XCell Sensors

See 5.7 How to Calibrate an Oxygen XCell Sensor for calibrating oxygen sensors.

NOTICE

If a password is enabled, you will not be able to proceed with the calibration without the password.

To abort, press the EZ touch button or apply magnet over GM logo before applying target gas.

In the event that a calibration cannot be completed, the user can acknowledge the FAIL by pressing the EZ
touch button or applying magnet to the GMI logo area until "Reset" is displayed and then release. The device will
revert to the settings of the last successful calibration.

General Monitors S5000 73


US 5 Calibration

If it is suspected that gases are present, it will be necessary to purge the sensor environment with zero air.

To calibrate the sensor:

(1) Attach a regulator to the zero gas cylinder (if using) and the calibration cylinder.

(2) Attach the plastic sensor guard to the bottom of the sensor housing if not already installed.

(3) Place the green calibration cap over the sensor guard inlet so that it is flush with the bottom of the sensor guard and
completely covers the sensor guard inlet.

(4) Attach the tubing to the plastic stem protruding through the green calibration cap.

(5) Push the other end of the tubing over the zero cylinder regulator. Ensure the tubing completely covers the gas outlet.

(6) Press EZ touch until Calibration is displayed, and release.

If a password is enabled, you will need to enter it to access the menu system.

NOTICE

If the calibration cap cannot be used (such as for a remote sensor application), ambient wind conditions must be minimized
during calibration to avoid a calibration with increased sensitivity.

(7) Select Sensor #1 or Sensor #2, and then select “Finished?” to start the calibration process.

(8) Once the screen displays Zero in Progress, turn on zero gas flow by turning knob on the regulator (if using).

(9) Wait while the device displays Zeroing.

(10) Once the Zero Calibration is complete, remove the tubing from the sensor guard inlet.

(11) The display will show Apply Gas. Attach tubing for calibration gas and turn on the regulator

Display will show “Calibration in Progress”.

“Calibration Complete Remove Gas” will show when span is complete.

If “PASS” displays, the procedure is complete.

If “FAIL” displays, the procedure was unsuccessful. See 6.5 Troubleshooting.

(12) Remove the calibration equipment from the device, including the sensor guard and calibration cap.

Sensors with TruCal and have Diffusion Supervision enabled can go into Diffusion Supervision Fault if the green
Calibration Cap is left on after completing a calibration.

WARNING!

The green Calibration Cap must be removed from the sensor after calibration. Failure to do so could restrict gas flow to the
sensor and result in erroneously low readings.

Failure to follow this warning can result in serious personal injury or death.

General Monitors S5000 74


US 5 Calibration

5.7 How to Calibrate an Oxygen XCell Sensor

An oxygen span gas cylinder is not needed if the sensor is in an area that maintains ambient air conditions. Follow the
same process for XCell sensors as described in 5.6 How to Calibrate XCell Sensors. When the display prompts "Apply
Span Gas", simply allow the countdown to occur without applying gas.

If the sensor is located in an area of normally low or enriched oxygen, then a 20.8% oxygen sample must be applied.

5.8 How to Calibrate an IR Sensor

A full span calibration is not required for the IR Sensors. Any degradation of the sensor's performance is associated with
slight drifts in its zero response. Restoring the sensor's zero is typically sufficient. Note that for zeroing of the IR700
sensor, zero gas is required. Air contains CO2 and cannot be used. Nitrogen is recommended as the zero gas. The nitrogen
cylinder P/N is 1400262-17.

See 5.5 How to Zero Calibrate XCell and IR Sensors for zero calibration instructions.

If a full calibration of the IR Sensor is required, a larger calibration cap is used and covers the entire sensor guard
during calibration. The calibration cap must be removed after completing the zeroing and/or spanning procedure.

NOTICE

If the calibration cap cannot be used (such as for a remote sensor application), ambient wind conditions must be minimized
during calibration to avoid a calibration with increased sensitivity.

5.9 How to Calibrate a Passive Sensor (Catalytic Bead or MOS)

WARNING!

The sensor calibration cup must be removed from the sensor after calibration. Failure to do so could restrict gas flow to the
sensor and result in erroneously low readings.

Failure to follow this warning can result in serious personal injury or death.

If a password is enabled, you will need to enter it to access the menu system.

If the sensor is fitted with a sensor guard, calibration must be performed with the sensor guard in place.

If it is suspected that gases are present, it will be necessary to purge the sensor environment with zero air.

To calibrate sensor:

(1) Attach a regulator to the zero air cylinder (if using) and the calibration cylinder.

(2) Attach the plastic calibration cup (P/N 1400152-1 or 1400154) to the bottom of the sensor if not already installed.

(3) Attach the tubing to the plastic stem protruding through the calibration cup.

(4) Push the other end of the tubing over the zero cylinder regulator. Ensure the tubing completely covers the gas outlet.

(5) Press EZ touch until Calibration is displayed, and release the touch.

General Monitors S5000 75


US 5 Calibration

(6) Turn on zero gas flow by turning knob on the regulator (if using).

(7) Unit will start flashing the sensor life. Tap the EZ Touch if you want to reset the sensor life.

(8) Wait while the device displays “Zero in Progress”.

(9) Once the Zero Calibration is complete, remove the tubing from the calibration cup inlet.

(10) The display will show “Apply Gas”. Attach tubing for calibration gas and turn on the regulator.

(11) Display will show “Calibration in Progress”.

(12) “Calibration Complete. Remove Gas” will show when span is complete.

(13) Turn off the calibration gas regulator. Remove the calibration equipment from the device, including the calibration cup.
Unit will return to normal operation.

If “Cal Fail” and fault code displayed, the procedure was unsuccessful. See 6.5 Troubleshooting.

5.10 XCell Catalytic Bead Failsafe

Catalytic bead sensors require the presence of oxygen in order to sense combustible gas. In the event of very large
combustible gas leaks that exceed 100 % LEL, enough oxygen can be displaced so that the sensor's response to gas is no
longer proportional to the calibration profile. The XCell catalytic bead has a locking fail safe mechanism that prevents the
false reporting of a safe condition while the % LEL concentration is still above 100 % LEL. When the gas concentration
exceeds 100 % LEL, the sensor will go into LOC over range.

To clear the LOC over rage, the user needs to acknowledge and calibrate the sensor. To acknowledge LOC over range,
press the EZ touch button or apply magnet to the GMI logo area until "Reset" is displayed and then release. This will allow
a recalibration of the sensor to clear the LOC condition.

WARNING!

Ensure that the area has been cleared of gas before acknowledging the LOC Over Range and recalibrating the sensor.

Failure to follow this warning can result in serious personal injury or death.

5.11 Calibration Confirmation

The S5000 Series Monitor records the date of the last successful calibration. This date can then be displayed via the
Status Menu.

General Monitors S5000 76


US 6 Maintenance

6 Maintenance

WARNING!

Use only genuine MSA replacement parts when performing any maintenance procedures provided in this manual. Failure to
do so may seriously impair sensor and gas monitoring performance, alter flameproof/explosionproof characteristics or void
agency approvals. Failure to follow this warning could cause the product to fail to perform as designed and persons who
rely on this product for their safety could sustain serious personal injury or loss of life.

Repair or alteration of the S5000 Gas Monitor, beyond the scope of the maintenance procedures provided in this manual or
by anyone other than authorized MSA service personnel, could cause the product to fail to perform as designed and
persons who rely on this product for their safety could sustain serious personal injury or loss of life.

The S5000 Gas Monitor is constantly performing a self-check. When a problem is found, it displays the appropriate error
message (see 6.5 Troubleshooting). When a critical error is detected within the device, the 4-20 mA output signal goes to a
fault condition of 0 mA, 1.25 mA, 3.5 mA or user defined level.

6.1 IR Sensor Cleaning Procedure

The presence of particulate matter, oil films, liquid water, or the residue from water drops on the two monitor windows can
adversely affect its performance. The sensor guard is designed to prevent foreign solids or liquids from reaching the
monitor's optical system. Additionally, heating elements are incorporated into the device to prevent water condensation.
Under severe conditions, however, some material may collect on these surfaces and it may be necessary to occasionally
check and clean the windows.

While both windows are made of a highly durable material that is not easily scratched, avoid excessive pressure when
cleaning them. Clean, cotton-tipped applicators are the most convenient tool to remove material collected on the windows.

(1) Remove the sensor guard.

(2) Use a dry applicator or one moistened with distilled water to wipe the window and remove dust.

(3) Use an additional clean, dry applicator to remove any residual water.

(4) Use an applicator moistened with isopropyl alcohol to remove heavy deposits of solids, liquids or oil films. Clean the
window again with a second applicator moistened with distilled water; then, dry the window with a final applicator.

Avoid using excessive amounts of water or alcohol in the cleaning procedure, and inspect the window to ensure that
the entire surface is clean.

(5) The device will go into a “Cleaning required” fault with an analog output of 2.0 mA when cleaning is needed.

The S5000 analog output may go down to 1.25 mA while the optical path is blocked, and display will indicate “Beam
Block”.

While in "Beam Block" fault, the sensor will not respond to the presence of gas.

(6) When cleaning is done and the objects are removed from the sensor window, the device returns to normal operation. If
water or isopropyl alcohol was used, allow the device to operate for 15 minutes to completely dry before replacing the
sensor guard continuing to monitor for combustible gas.

(7) Replace the sensor guard.

General Monitors S5000 77


US 6 Maintenance

When the cleaning process is complete, be sure to remove all objects from the light path. It is recommended to
check the sensor’s response to zero and calibration gas after cleaning.

WARNING!

Do not place foreign objects in the sensor's analytical region (except per the “S5000 IR Sensor Cleaning Procedure” as
described above); otherwise, the infrared beam can be partially blocked, causing the sensor to generate false readings. All
objects must be removed from the sensor's analytical region for it to function properly.

Failure to follow this warning can result in serious personal injury or death.

6.2 Replacing an XCell Sensor

The only routine maintenance item is the sensor, which has a limited lifetime. The S5000 sensors with TruCal technology
will indicate when the sensor is near end of life through the Status Menu. When the Sensor Life & Health status is "Fair",
there is approximately 2 month time frame to replace the sensor before it will no longer function. Actual time between “Fair”
status and sensor end of life depends on exact environmental conditions. When a TruCal sensor is no longer capable of
sensing it will go into fault and LEDs will flash yellow. It is good practice to obtain a replacement sensor before the sensor
within the unit becomes inoperative.

There is no need to open the main enclosure. Simply unscrew the sensor assembly located on the bottom of the main
assembly.

WARNING!
• Do not disassemble the sensor assembly. The sensing element inside may contain hazardous material.
• Handle the sensor carefully; the electrochemical version is a sealed unit which contains a corrosive electrolyte.
• Any leaked electrolyte that comes in contact with skin, eyes or clothes can cause burns.
• If any contact with the electrolyte does occur, immediately rinse with a large quantity of water. On contact with the
eyes, rinse thoroughly with water for 15 minutes and consult a doctor.
• Do not install a leaking sensor in the sensing head assembly.
• A leaking sensor must be disposed of in accordance with local, state and federal laws.
• As part of the product certification, it was verified that optional communication functions of this gas detection
instrument while operating at the maximum transaction rate do not adversely affect the gas detection operation and
functions of the instrument. The product certification, however, does not include or imply approval of the SafeSwap
feature, communications protocol or functions provided by the software of this instrument or of the communications
apparatus and software connected to this instrument.
• Follow the warnings below when removing or replacing sensors. Reference Figure 3 for component overview.
– Never remove or replace a Sensor Body Assembly or an IR Sensor while under power or when explosive hazards
are present.
– Confirm that the area is free of explosive hazards before removing or replacing an XCell Sensor under power.
– To remove an XCell Sensor, unscrew XCell Sensor three full turns, wait 10 seconds, and then remove the XCell
Sensor completely.

Failure to follow this warning can result in serious personal injury or death.

General Monitors S5000 78


US 6 Maintenance

Identify the sensor assembly needed via the A-5K-SENS code on the interior sensor label and obtain the appropriate
sensor assembly; replace sensor assembly.

Figure 33 Removing the XCell Sensor

Figure 34 Digital Sensor - Position of Labels

Alarm set points, span value, full scale limit and alarm direction will not change when replacing a sensor with the
same gas type.

Alarm set points, span value, full scale limit and alarm direction will change to the new sensor’ settings when
replacing a sensor with a different gas type.

The S5000 Gas Monitor is shipped with the Sensor Swap Delay enabled. IF enabled, the 4-20 mA output signal goes to
3.5 mA and the FAULT relay will hold off a fault indication for 2 minutes. This setting allows the operator to exchange
sensor modules without a FAULT indication.

It is recommended that all other maintenance be performed at an MSA factory-authorized service center.

General Monitors S5000 79


US 6 Maintenance

6.3 Clearing a Blockage

If using a Digital Sensor with Diffusion Supervision, and if the Diffusion Supervision feature has been enabled in the Sensor
Settings, the sensor will notify the user if it detects foreign material obstructing the sintered metal frit where gas would
normally enter the sensor (see Frit Location). When this condition is detected, the transmitter will go into fault and show
“Diffusion Supervision Fault” on the display. Follow these recommended steps to attempt to clear this fault:

(1) Inspect and clear the sensor inlet and frit.

(2) Inspect and clear any accessories attached to the sensor.

(3) Ensure Sensor is fully mated to the Sensor Body Assembly.

(4) Attempt to recalibrate the Sensor.

(5) Disable Diffusion Supervision or replace the Sensor.

First, with the sensor still installed, visually inspect the sensor to see if any objects are covering the gas inlet and remove
them. Example objects may include tape or the green Calibration Cap. Next, check to make sure that there is no foreign
material stuck to the frit. If any material is found, gently remove it while considering any hazards that may arise from such
an action, like the accumulation of a static charge, and taking the necessary steps to avoid them. Take care to ensure that
the frit is not damaged during the cleaning process. DO NOT use water or liquids to try to clean the frit. The frit acts like a
sponge and will retain liquids, which will keep the gas path blocked.

Figure 35 Frit Location

WARNING!

The sensor relies on the frit as a part of its Flameproof/Explosionproof approval. If the frit is damaged while attempting to
clean it, immediately remove power from the device. Ensure the area is free of a hazardous atmosphere, then follow all
steps as outlined in this manual to properly replace the sensor.

Failure to follow this warning can result in serious personal injury or death.

If no material is visible, the frit may still be blocked by a liquid which has absorbed into the frit. Once absorbed, liquids are
very difficult to remove. For these situations, relying on natural evaporation is recommended. Due to the time it can take for
liquids to evaporate sufficiently from the frit, users may want to consider swapping out the affected sensor with one that is
known to be clear. Placing absorbent material against the outer surface of the frit can help accelerate this process.

General Monitors S5000 80


US 6 Maintenance

If a sensor is in Diffusion Supervision Fault, the device will still allow the user to attempt to calibrate the sensor. In the rare
event that a sensor calibration is successful, but the Diffusion Supervision system continues to detect a blockage in the
gas path, the Diffusion Supervision Fault will return in approximately 24 hours. In this scenario, there is likely an
environmental factor affecting the Diffusion Supervision detection. Provided the sensor can successfully calibrate, the
sensor is still capable of detecting gas. To prevent the Diffusion Supervision Fault from reoccurring, a user can either
disable Diffusion Supervision in the Sensor Settings (see 4.2.2 Sensor Setup) or replace the sensor.

6.4 Replacing a Passive Sensor (Cat Bead or MOS)

When the passive Catalytic Bead or MOS sensor is damaged or gradually losing sensitivity and eventually fails to respond
to gas adequately or fails to pass calibration, the sensor needs to be replaced.

The mounting threads of the sensor head may require lubrication. Grease must not be used. As an alternate, PTFE (Teflon)
tape may be used on sensor threads.

To replace the passive sensors:

(1) Declassify the area and make sure there is no hazardous gas/vapor present.

(2) Power down the S5000 transmitter.

(3) Unscrew the lid.

(4) Remove the User Interface Module.

(5) Refer to the sensor wiring diagram label inside the S5000 transmitter to locate the sensor connector and unplug it from
the transmitter.

(6) Unscrew the sensor head until it is totally off. Guide the sensor wires so they will not get caught by the internal parts
of the S5000 while sensor head is turning.

(7) Guide the connector and wires of the new sensor head through the conduit entry for sensor. Do not plug in the sensor
connector at this time.

(8) Screw in the new sensor head. Guide the sensor wires so they will not get caught by the internal parts of the S5000
while sensor head is turning.

(9) Verify the sensor connector wiring against the wiring diagram.

(10) Plug in the connector to the indicated header and make sure they are mated correctly.

6.5 Troubleshooting
Table 17 Troubleshooting

Fault Code Latch Description Action


ER01 Main Board Non-Latch This fault indicates the Ensure the display module is properly
Communication Error communication between the seated in the board stack fully. Return
main board and the display the unit to the factory or authorized
module is abnormal. service center if the problem persists.
F001 Supply Voltage Non-Latch This fault occurs if the supply Ensure that the supply voltage is
Out-of-Range Fault voltage at the S5000 is below within the range of 12-30 VDC at the
+10.5 VDC or higher than S5000.
31.5 VDC.

General Monitors S5000 81


US 6 Maintenance

Fault Code Latch Description Action


F002 RAM Checksum Latch This fault indicates that the Cycle the power to the unit. If the fault
Error contents of the S5000's persists, the unit must be returned to
internal RAM memory have the factory or authorized service
corrupted. center for repair.
F003 Flash Checksum Latch This fault indicates that the The unit must be returned to the
Error contents of the S5000's factory or authorized service center for
program memory have repair.
changed. This usually occurs
when powering the unit up
after a lightning strike or large
voltage transient on the power
or signal lines to the unit.
F004 EEPROM Error Latch In the event of an EEPROM (1) Unit returns to normal.
error, the user must recycle
a) This indicates the EEPROM
the power to potentially clear
writing did not keep up to
the error. After power reset,
changing events or the write
the following may occur:
cycle is too fast.

(2) Unit goes to sensor faults. The


user must recalibrate after 1-2
minutes sensor warm up.

a) This means the non-critical


part of the EEPROM was
corrupted.

b) This is probably caused by an


event-logging problem.

c) There is a possibility any of


the following may be
corrupted: Event logging data,
Modbus settings, HART
settings, Calibration
information.

(3) Unit returns to F004.

a) This is a critical error. This


fault occurs when an attempt
to verify the setup/calibration
parameters just written to the
EEPROM memory fails.

b) Reset transmitter settings to


reload the factory default to
the EEPROM. Sensor

General Monitors S5000 82


US 6 Maintenance

Fault Code Latch Description Action


recalibration is required after
this operation. All settings
including, but limited to,
alarm/warning, setpoint, relay,
Modbus, AO output level
should be verified and
reprogrammed.
F005 Internal Circuit Non-Latch An internal error has occurred. The
Error unit must be returned to the factory for
service. The possible errors are
internal voltages are not at their proper
values.
F006 Relay Fault Latch This fault indicates that either Cycle the power to the unit. If the fault
the relays or the relay driver is persists, the unit must be returned to
not working properly. the factory or authorized service
center for repair.
F007 Invalid Sensor Latch This fault occurs when the Verify at least one sensor channel is
Configuration Fault sensor channel configuration enabled. Cycle the power to the unit. If
is invalid. the fault persists, the unit must be
returned to the factory or authorized
service center for repair.
F008 Other System Non-Latch This fault occurs when other The unit must be returned to the
level Fault fault conditions not covered factory or authorized service center for
above happen. repair.
F009 EZ Touch User Non-Latch This fault indicates that the Verify the EZ Touch button area is free
Button Fault EZ Touch user button is of debris such as metal pieces or ice.
constant on or faulty. Return the unit to the factory or
authorized service center if the
problem persists.
F010 RTC Battery Non-Latch This fault indicates that the Return the unit to the factory or
Voltage Low Fault battery for Real Time Clock is authorized service center for RTC
running low. battery replacement.
F011 Display Board Non-Latch This fault indicates that the Verify the User Interface Module is
Communication Fault communication between the properly placed into the socket and the
Main Control Board and connectors on Main Control Board and
Display Board in User Display Board are intact. Return the
Interface Module is faulty, or unit to the factory or authorized
the Display Board is faulty. service center if the problem persists.
F012 Menu Setup Non-Latch This fault indicates that the N/A.
Timeout Fault user menu was timed out and
not properly exited when it
was accessed last time.
F013 Bluetooth Fault Non-Latch This fault indicates that the Return the unit to the factory or

General Monitors S5000 83


US 6 Maintenance

Fault Code Latch Description Action


communication between the authorized service center if the
Main Control Board and the problem persists.
Bluetooth module is faulty.
F014 BCM Fault (Not N/A N/A N/A
used, reserved for
future)
F015 HART Reference Non-Latch This fault indicates that the Return the unit to the factory or
Power Supply Fault reference power supply to authorized service center if the
HART modem circuity is problem persists.
faulty.
F016 (Not used, N/A N/A N/A
reserved for future)
F017 Magnet User Non-Latch This fault indicates that the Verify there is no magnet left on the
Button Fault EZ Touch user button is unit front window nor magnetic field
constant on or faulty. present right next to the unit window.
Return the User Interface Module to
the factory or authorized service
center if the problem persists.
F101/F201 Sensor Non-Latch This fault occurs when the Check the wiring and sensor
Missing Fault sensor is disconnected or connection.
wiring is faulty or damaged.
F102/F202 Sensor Non-Latch This fault occurs when the Check the wiring and sensor
Supply Voltage Fault power supply to the sensor is connection.
faulty.
F103/F203 Invalid Latch This fault occurs when the Check the sensor parameters,
Sensor Parameters in sensor parameters stored in including Calibration Level, Full Scale,
Main EEPROM Fault the main EEPROM is Warning and Alarm set points. Correct
incorrect. any mis-configured parameters.
F104/F204 Sensor Non-Latch This fault occurs when the Check the wiring and sensor
Element Fault sensor sensing element is connection. Replace the sensor if
faulty or damaged. needed.
F105/F205 Sensor Non-Latch This fault occurs when the Check the wiring and sensor
Heater Fault sensor heater is faulty or connection. Replace the sensor if
damaged. needed.
F106/F206 Other Non-Latch This fault occurs when the Check the wiring and sensor
Sensor Internal Fault sensor has other internal connection. Replace the sensor if
faults. needed.
F107/F207 Sensor Non-Latch This fault occurs when the Replace the sensor.
Internal Data CRC Error internal data of the sensor has
Fault CRC error.
F108/F208 Sensor EOL Non-Latch This fault occurs when the Replace the sensor.
Fault sensing element of the sensor

General Monitors S5000 84


US 6 Maintenance

Fault Code Latch Description Action


is at the end of life state.
F109/F209 Sensor Non-Latch This fault occurs when the Clean out any debris or foreign objects
Blockage Fault sensor gas sensing path blocking the sensing path/opening.
and/or opening is blocked. For more details, see 6.1 IR Sensor
Cleaning Procedure or 6.3 Clearing a
Blockage, depending on sensor type.
F110/F210 Negative Latch This fault occurs when the Re-calibrate the sensor. Replace the
Drift Fault sensor has severe negative sensor if this fault persists.
drift.
F111/F211 Cal Line Non-Latch This fault occurs when the Check the wiring and sensor
Shortage Fault IR400/IR700 sensor Cal Line connection.
(brown wire) is shorted to
ground.
F112/F212 Zero Latch This fault occurs when last Use the “Reset Sensor” entry in main
Calibration Failed Fault zero calibration has failed. menu to reset the fault. Re-calibrating
the sensor will also clear the fault.
Replace the sensor if this fault
persists.
113/F213 Full Span Latch This fault occurs when last full Use the "Reset" entry in UI menu to
Calibration Failed Fault span calibration has failed reset the fault. Re-calibrate the sensor
will also clear the fault. Replace the
sensor if this fault persists.
F114/214 Cal Check Non-Latch This fault occurs when the Check whether test gas is left on for
Timeout Fault last Cal Check has failed due Calibration Check mode. Remove the
to timeout. test gas if it is left on. Once it is
cleared, the unit will return to normal
operation.
F115/F215 Sensor Latch This fault occurs when the Check the sensor parameters,
Configuration Reset configuration parameters for including calibration level, Range,
Fault sensor have been reset. Warning and Alarm set points. Verify
all these parameters are configured as
intended. Recalibrate the sensor to
clear the fault if needed. Running
through the Calibration Check is
recommended.
F116/F216 Calibration Latch This fault occurs when the Recalibrate the sensor.
Required Fault sensor is in need of
calibration.
F117/F217 Bead Off Latch Indicates that the combustible Make sure the environment is free of
Fault sensors beads are OFF. dangerous combustible gas and then
use the “Reset” entry in UI menu to
acknowledge the fault or cycle power
to the sensor. Let the sensor warmup,

General Monitors S5000 85


US 6 Maintenance

Fault Code Latch Description Action


then recalibrate the sensor. If this
does not resolve the issue, replace the
sensor.
F118/F218 Analog Non-Latch This fault occurs when the 4- Check the 4-20 mA analog output loop
Output Mismatch Fault 20 mA analog output loop is is closed properly with nominal
malfunctioning. 250 ohm load resistor installed. The
fault will be cleared automatically
once the loop is closed and wires are
properly connected.

General Monitors S5000 86


US 7 Ordering Information

7 Ordering Information
7.1 Replacement Parts

See Table 18 for replacement sensor kits. To obtain a replacement sensor, address the order or inquiry to:

General Monitors
26776 Simpatica Circle
Lake Forest, CA 92630

or call, toll-free, 1-800-446-4872.

Inquiries can also be e-mailed to [email protected].

WARNING!

Use only genuine MSA replacement parts when performing any maintenance procedures provided in this manual. Failure to
do so may seriously impair sensor and gas monitoring performance, alter flameproof/explosionproof characteristics or void
agency approvals. Failure to follow this warning could cause the product to fail to perform as designed and persons who
rely on this product for their safety could sustain serious personal injury or loss of life.

Repair or alteration of the S5000 Gas Monitor, beyond the scope of the maintenance procedures provided in this manual or
by anyone other than authorized MSA service personnel, could cause the product to fail to perform as designed and
persons who rely on this product for their safety could sustain serious personal injury or loss of life.

Failure to follow these warnings can result in serious personal injury or death.

Table 18 Replacement Parts

Type Description P/N


PCBA Assemblies S5000-PCB
Sensor Body, XCell A-5K-SENS
Sensors, XCell A-5K-SENS
Sensor Guard, XCell Sensor Guard Assembly, S5000 10177624
Sensor Guard, IR400 Sensor Guard, IR400, with Screen 32545-1
CB 0-100% LEL (C04 – Screened, Div Approvals) 10058-1

CB 0-20% LEL (C10 – Sintered, Zone Approvals) 11159-8

CB 0-100% LEL (C11 – Sintered, Zone Approvals) 11159-1


MOS 0-100ppm (M04 – Screened, Div Approvals) 50448-1
Sensors, Passive MOS 0-100ppm (M14 – Sintered, Zone Approvals) 51457-1
MOS 0-50ppm (M05 – Screened, Div Approvals) 50448-5

MOS 0-50ppm (M15 – Sintered, Zone Approvals) 51457-5


MOS 0-20ppm (M06 – Screened, Div Approvals) 50448-9

MOS 0-20ppm (M16 – Sintered, Zone Approvals) 51457-9

General Monitors S5000 87


US 7 Ordering Information

Type Description P/N


Methane, NFPA IR400-4-01-2-2-0-1-0
Propane, NFPA IR400-4-02-2-2-0-1-0
Hexane, NFPA IR400-4-04-2-2-0-1-0
Pentane, NFPA IR400-4-05-2-2-0-1-0
Sensor, IR4001 Ethylene, NFPA IR400-4-06-2-2-0-1-0
Butane, NFPA IR400-4-07-2-2-0-1-0
Ethane, NFPA IR400-4-08-2-2-0-1-0
Methane, IECEx IR400-4-13-2-2-0-1-0
Propane, IECEx IR400-4-14-2-2-0-1-0
Hexane, IECEx IR400-4-15-2-2-0-1-0
Pentane, IECEx IR400-4-16-2-2-0-1-0
Butane, IECEx IR400-4-17-2-2-0-1-0
Ethane, IECEx IR400-4-18-2-2-0-1-0
Ethylene, IECEx IR400-4-22-2-2-0-1-0
0-2000ppm IR700-4-1-2-2-1
0-5000ppm IR700-4-2-2-2-1
Sensor, IR7001 0-10000ppm IR700-4-3-2-2-1
0-30000ppm IR700-4-4-2-2-1
0-50000ppm IR700-4-5-2-2-1
Duct mount kit, rectangular 10179322
Duct mount kit, round, small, 10179323

Duct Mount Kit (12"-20" / 305-508 mm diameter)


Duct mount kit, round, large, 10179324

(20"-40" / 508-1016 mm diameter)


Calibration Kit (XCell) CALKIT1
Pipe Mount Kit, Pipe Mount, X5000/S5000 10176946
Universal
Junction Box, SST, IIB + H2 324240-1
Junction Box
Junction Box, SST, IIC 324240-3
3/4" NPT 10213879
JB5000 Junction Box
M25 10213893
Calibration Cap, XCell Calibration Cap, X5000/S5000, packaged 10181450
Calibration Cap, IR400 Calibration Cap, IR400 packaged 1400270-2
Sun Shield Sunshield, X5000/S5000 10180254
1- Only IR Sensor P/Ns shown can be used with the S5000.

General Monitors S5000 88


US 7 Ordering Information

For position of labels, see 2.11 Label Overview

The following accessories have been validated for use when installed in accordance with the instructions in this manual:

Accessory Validation Method


Calibration Kit CSA/Sira, FM Approvals
Junction Box CSA/Sira, FM Approvals
JB5000 Junction Box CSA/Sira
Calibration Cap (XCell, IR400) CSA/Sira, FM Approvals

General Monitors S5000 89


US 8 Appendix: Specifications

8 Appendix: Specifications
Table 19 Digital Sensor Specifications

Digital XCell IR
Sensor Options Toxics
Oxygen Combustible IR400 IR700
CO H 2S SO2 Cl2 NH3
Operating Range2 -40 °C to -40 °C to -40 °C to -40 °C to -40 °C to -40 °C to -55 °C to 60 °C -40 °C to -40 °C to
+60 °C +60 °C +60 °C +60 °C +60 °C +60 °C +75° C +50° C
Storage -40 °C to -40 °C to -40 °C to -40 °C to -40 °C to -40 °C to -40 °C to 60 °C -50 °C to -60 °C to
+60 °C +60 °C +60 °C +60 °C +60 °C +60 °C +85 °C +85 °C
Zero Drift1 ≤ 1% FS per ≤ 1% FS per ≤ 1% FS per ≤ 1% FS per ≤ 1% FS per ≤ 0.2% Vol per ≤ 5% LEL per Not Applicable Not Applicable
year year month month month year year
Span Drift1 ≤ 2% per year ≤ 2% per year ≤ 2% FS per ≤ 2% FS per ≤ 2% FS per ≤ 0.2% Vol per ≤ 5% LEL per Not Applicable Not Applicable
month month month year year
Noise1 < 2 display < 2 display < 2 display < 2 display < 1 display unit < 2 display 1% LEL 1% LEL 1% LEL
units units units units units
Repeatability1 < ±1% < ±1% < ±1% < ±1% < ±1% < ±1% < ±3% LEL ±3% LEL at ±5% FS at
≤ 50% LEL ≤ 50% FS
±5% LEL at ±10% FS at
> 50% LEL > 50% FS
Resolution 1 ppm 0.1 ppm 0.1 ppm 0.1 ppm 0.1 ppm 0.1 % Vol 1 % LEL 1 % LEL 1 % LEL
T90 < 9 s < 23 s < 6 s < 12 s < 280 s < 11 s < 22 s ≤ 3 s ≤ 9 s
Humidity 10-95% RH 10-95% RH 10-95% RH 10-95% RH 10-95% RH 10-95% RH 10-95% RH 10-95% RH 10-95% RH
Sensor Life 5 years 5 years 5 years 5 years 5 years 5 years 5 years > 5 years > 5 years
Warranty 3 years 3 years 3 years 3 years 3 years 3 years 3 years 2 years 2 years
Enabling - XCell - non- XCell - non- XCell - non- XCell - non- XCell - non- XCell - non- XCell - IR IR
Technology consuming consuming consuming consuming consuming consuming GM cat bead

General Monitors S5000 90


US 8 Appendix: Specifications

Digital XCell IR
Sensor Options Toxics
Oxygen Combustible IR400 IR700
CO H 2S SO2 Cl2 NH3
TruCal3 Yes Yes No No No No No No No
SafeSwap Yes Yes Yes Yes Yes Yes Yes No No
Remote Mount 100 m 100 m 100 m 100 m 100 m 100 m 100 m 1100 m
Distance
Power: 3.6 W 3.6 W 3.6 W 3.6 W 3.6 W 3.6 W 6.0 W 4.8 W
Single Sensor
Power: 4.3 W 4.3 W 4.3 W 4.3 W 4.3 W 4.3 W 8.4 W NA
Dual Sensing
1- Typical response at room temperature
2- S5000 transmitter has an operating temperature range of -55°C to +75°C.
3- Adaptive Environmental Compensation (AEC) and Diffusion Supervision (DS) will remain active between -20°C and +60°C.
4- These specifications are internally generated data. See the FM Approval section for data that conforms to the FM certification in accordance to FM 6340.

General Monitors S5000 91


US 8 Appendix: Specifications

Table 20 Passive Sensor Specifications

Passive Catalytic Bead Passive MOS


Sensor Options Sintered Screened Sintered Screened
(Zone Approvals) (Division Approvals) (Zone Approvals) (Division Approvals)
Operating Range -40 °C to +70 °C -40 °C to +75 °C -40 °C to +70 °C -40 °C to +75 °C
Storage -50 °C to +85 °C -50 °C to +85 °C -50 °C to +85 °C -50 °C to +85 °C
Zero Drift1 < 5% of full scale < 5% of full scale per None None
per year year
Noise 1% LEL 1% LEL 1 ppm 1 ppm
Repeatability1 ±3% LEL at ≤ 50% ±3% LEL at ≤ 50% ±2 ppm or 10% of ±2 ppm or 10% of
LEL LEL applied gas applied gas

±5% LEL at > 50% ±5% LEL at > 50%


LEL LEL
Resolution 1% LEL 1% LEL 1 ppm 1 ppm
T50 < 10 s < 10 sec < 30 s < 14 s
T90 < 30 s < 30 s NA NA
Humidity 5-90% RH 5-90% RH 0-95% RH 0-95% RH
Sensor Life 3-5 years 3-5 years 3-5 years 3-5 years
Warranty 2 years 2 years 2 years 2 years
Enabling - Catalytic Bead Catalytic Bead MOS MOS
Technology
TruCal No No No No
SafeSwap No No No No
Remote Mount 307 m (12 AWG wiring) 1200 m (14 AWG wiring)
Distance
Power: 6.0 W 10.8 W
Single Sensor
Power: Transmitters purchased with passive sensors are not capable of dual sensing and
Dual Sensing cannot be changed to any other sensor type than the one with which it was originally
purchase.
1- Typical response at room temperature.
2- S5000 transmitter has an operating temperature range of -55°C to +75°C.
3- Adaptive Environmental Compensation (AEC) and Diffusion Supervision (DS) will remain active

between -20°C and +60°C.


4- These specifications are internally generated data. See the FM Approval section for data that conforms
to the FM certification in accordance to FM 6340.

General Monitors S5000 92


US 8 Appendix: Specifications

Figure 36 Housing Dimensions

Table 21 Housing Material Specifications

Material Weight
316 Stainless Steel 7.30 lbs

7.92 lbs w/ relays

General Monitors S5000 93


US 9 Appendix: General Certification Information

9 Appendix: General Certification Information


The S5000 Transmitter, S5000 Junction Box, and Digital Sensor have the following certifications/approvals: ATEX,
IECEx, CSA, FM,INMETRO, DNV·GL Type Approval. They also comply to the EMC, Radio Equipment and Marine
Equipment Directives. The S5000 Transmitter and Junction Box have a Type 4X and IP66 environment rating.

S5000 Transmitter, Junction Box, and JB5000 Junction Box

U.S. Division Hazardous Locations as defined by the NEC


Class I, Division 1, Groups A, B, C, D; T5/T6*

Class I, Division 2, Groups A, B, C, D; T4/T6*

Class II, Division 1,2 Groups E, F, G; Class III, Division 1,2; T6 (T5 under FM Approval)

NOTE: See CSA Certificate for detailed certification information including possible
limitations to the above information. FM Approval does not include Zone classified locations.

*Applicable to S5000 Junction Box only


U.S. Zone Hazardous Locations as defined by the NEC
Class I, Zone 1, AEx db IIB+H2 T5/T6* Gb

Class I, Zone 2, AEx nA nC** IIC T4/T6* Gc

Zone 21, AEx tb IIIC T85°C Db

NOTE: See CSA Certificate for detailed certification information including possible
limitations to the above information. FM Approval does not include Zone classified locations.

*Applicable to S5000 Junction Box only

**nC not applicable to S5000 Junction Box


Canadian Division Hazardous Locations as defined by the CEC
Class I, Division 1, Groups A**, B, C, D; T5/T6*

Class I, Division 2, Groups A, B, C, D; T4/T6*

Class II, Division 1,2 Groups E, F, G; Class III, Division 1,2; T6

NOTE: See CSA Certificate for detailed certification information including possible
limitations to the above information.

*Applicable to S5000 Junction Box only

**Applicable for cemented window covers only


Canadian Zone Hazardous Locations as defined by the CEC
Ex db IIB+H2 T5/T6* Gb

Ex nA nC** IIC T4/T6* Gc

Ex tb IIIC T85°C Db

NOTE: See CSA Certificate for detailed certification information including possible
limitations to the above information.

*Applicable to S5000 Junction Box only

General Monitors S5000 94


US 9 Appendix: General Certification Information

**nC not applicable to S5000 Junction Box


ATEX/IECEx/INMETRO Hazardous Locations
(Sira 17ATEX1048X, Sira 17ATEX4052X, IECEx SIR 17.0016X, NCC 18.0097X)
Ex db IIB+H2 T5/T6* Gb

Ex nA nC** IIC T4/T6* Gc

Ex tb IIIC T85°C Db

NOTE: See ATEX/IECEx Certificates for detailed certification information including


possible limitations to the above information.

*Applicable to S5000 Junction Box only

**nC not applicable to S5000 Junction Box


NOTE: For configurations incorporating cemented covers, IIB+H2 is replaced by IIC.

Digital Sensor with Frit (Fine Threads) Digital Sensor without Frit (Coarse Threads)
U.S. Division Hazardous Locations as defined U.S. Division Hazardous Locations as defined
by the NEC: by the NEC:

Class I, Division 1 & 2, Groups A, B, C, D; T5 Class I, Division 2, Groups A, B, C, D; T5

Class II, Division 1, Groups E, F, G; Class III;


T5
U.S. Zone Hazardous Locations as defined by U.S. Zone Hazardous Locations as defined by
the NEC: the NEC:

Class I, Zone 1, AEx db IIC T5 Gb Class I, Zone 2, AEx nA IIC T5 Gc

Class I, Zone 2, AEx db nA IIC T5 Gc

Zone 21, AEx tb IIIC T85°C Db


Canadian Division Hazardous Locations as Canadian Division Hazardous Locations as
defined by the CEC: defined by the CEC:

Class I, Division 1 & 2, Groups A, B, C, D; T5 Class I, Division 2, Groups A, B, C, D; T5

Class II, Division 1, Groups E, F, G; Class III,


T5
Canadian Zone Hazardous Locations as defined Canadian Zone Hazardous Locations as defined
by the CEC: by the CEC:

Ex db IIC T5 Gb Ex nA IIC T5 Gc

Ex db nA IIC T5 Gc

Ex tb IIIC T85°C Db
ATEX/IECEx/INMETRO Hazardous Locations ATEX/IECEx/INMETRO Hazardous Locations
(Sira 17ATEX1048X, Sira 17ATEX4052X, (Sira 17ATEX4052X, IECEx SIR 17.0016X,
IECEx SIR 17.0016X, NCC 18.0097 X) NCC 18.0097 X)

General Monitors S5000 95


US 9 Appendix: General Certification Information

Digital Sensor with Frit (Fine Threads) Digital Sensor without Frit (Coarse Threads)
Ex db IIC T5 Gb Ex nA IIC T5 Gc

Ex db nA IIC T5 Gc IP55

Ex tb IIIC T85°C Db

IP65

WARNING!

Some toxic gases are provided in a fritless sensor housing. The fritless sensor housing is labeled as Div 2 or Zone 2 and is
approved for Div 2 or Zone 2 installations only. The protection method is Nonincendive or Type n respectively. Ensure all
components are approved for the wiring method being used and in accordance with the National Electric Code of the
country of use, any applicable local regulations, and this manual.

Failure to follow this warning can result in serious personal injury or death.

Figure 37 Digital Sensor for Div 1 & 2, Zone 1 & 2 (Left) vs Digital Sensor for Div 2, Zone 2 only (right)

IR Sensor

See the applicable IR Sensor Instruction Manual (IR400 Sensor: P/N MANIR400; IR700 Sensor: P/N MANIR700), or
certificates Sira 07ATEX1253 and IECEx SIR 07.0080.

Passive Sensors

See certificates Sira 00ATEX1039U and IECEx SIR 07.0007U.

FM Approval

This approval does not include or imply approval of apparatus to which the subject instrument may be connected. In order
to maintain an FM Approved system, the apparatus to which this instrument is connected, must also be approved by FM
Approvals. As part of this approval, it was verified that optional communication functions of this gas detection instrument
while operating at the maximum transaction rate do not adversely affect the gas detection operation and functions of the
instrument. This approval, however, does not include or imply approval of the communications protocol or functions
provided by the software of this instrument or of the communications apparatus or software connected to this instrument.

The S5000 Transmitter, S5000 Junction Box, Digital Sensors (H2S, 500 ppm H2S, CO, O2), and Passive MOS H2S
sensors also have FM Performance Approval for use in ordinary (nonhazardous) or hazardous locations.

General Monitors S5000 96


US 9 Appendix: General Certification Information

The following information shall be noted for FM Approved configurations:


• The FM Approved Digital Sensors have an IP65 and TYPE 3X environmental rating, however if water is splashed onto
the frit (also referred to as a flame arrestor), the sensor may not be able to detect gas for a period of up to 12 minutes. If
water is splashed onto this element, use a cloth or paper towel to dab the frit in order to absorb any excess water.
• The FM Approval requires that password protection for Bluetooth be enabled.
• The Digital Sensor (O2) is not to be used in environments containing 5% (v/v) or more of carbon dioxide.
• The Digital Sensor (O2) is unaffected by air velocities of up to 5 m/s.
• The response and recovery times for the Digital Sensor (H2S, 500 ppm H2S, CO, O2) and Passive MOS sensor are as
follows:

Digital Sensor Passive MOS Sensor


Specification
H2S H2S 500 ppm CO H2S
T20 4 sec 4 sec 6 sec 1 sec
T50 8 sec 7 sec 8 sec 1 sec
T90 16 sec 28 sec 15 sec 2 sec
T10 (recovery) 23 sec 30 sec 28 sec 5 sec

Specification Digital Sensor (O2)


Tlowest alarm setpoint 5 sec
Thighest alarm setpoint 5 sec

• The operating and storage temperature ranges are as follows:

Storage Temperature Operating Temperature


Range Range Pressure
Limits
min. °C max. °C min. °C max. °C
S5000 without relays -50 +85 -55* +75 86 kPa – 108 kPa
S5000 with relays -50 +85 -40 +75 86 kPa – 108 kPa
Junction Box -50 +85 -55* +75 86 kPa – 108 kPa
Digital Sensor (H2S, CO) -40 +60 -40 +60 86 kPa – 108 kPa
Digital Sensor (H2S 500 -40 +60 -40 +50 86 kPa – 108 kPa
ppm)
Digital Sensor (O2) -40 +60 -40 +60 86 kPa – 108 kPa
Passive MOS Sensor -50 +85 -40 +60 86 kPa – 108 kPa
(H2S)
*Limited to -50°C if installed in Hazardous Locations

General Monitors S5000 97


US 9 Appendix: General Certification Information

• The Accuracy spec. is as follows:

Accuracy
Digital Sensor (H2S) ±3 ppm or 10% of reading
Digital Sensor (H2S 500 ppm) ±3 ppm or 10% of reading (-20°C to +40°C)

-20% of reading (-20°C to -40°C)

+25% of reading (+40°C to +50°C)


Digital Sensor (CO) ±6 ppm or ±10% of reading (-10°C to +60°C)

+15% of reading (-10°C to -20°C)

-30% of reading (-20°C to -40°C)


Digital Sensor (O2) ±0.5% O2 (FM 6340)

±0.2% (v/v) or ±2.5% of the measuring range (EN 50104)


Passive MOS Sensor (H2S) ±3 ppm or 10% of reading

Accuracy
Digital Sensor (H2S) ±3 ppm or 10% of reading
Digital Sensor (H2S 500 ppm) ±3 ppm or 10% of reading (-20°C to +40°C)

-20% of reading (-20°C to -40°C)

+25% of reading (+40°C to +50°C)


Digital Sensor (CO) ±6 ppm or ±10% of reading (-10°C to +60°C)

+15% of reading (-10°C to -20°C)

-30% of reading (-20°C to -40°C)


Digital Sensor (O2) ±0.5% O2 (FM 6340)

±0.2% (v/v) or ±2.5% of the measuring range (EN 50104)


Passive MOS Sensor (H2S) ±3 ppm or 10% of reading

WARNING!
• Read and understand all instructions, warnings and cautions prior to installation of any components of this system.
• Open circuit before removing cover.
• Sealing requirements:
– Aluminum S5000 transmitter and Junction Box: A seal shall be installed within 18 in (450 mm) of the enclosure.
– Stainless Steel S5000 transmitter, Junction Box, and JB5000: Seal is not required.
– Stainless Steel S5000 transmitter and Junction Box with FM Approval: A seal shall be installed within 18 inches
(450mm) of the enclosure.
• Transmitter and Junction Box (cemented or non-cemented covers) - do not open when energized or when an explosive
atmosphere is present.

General Monitors S5000 98


US 9 Appendix: General Certification Information

• Transmitter and Junction Box (cemented or non-cemented covers) - live parts behind cover, open circuit before
removing cover - do not contact.
• Transmitter and Junction Box (cemented or non-cemented covers) - potential electrostatic charging hazard. Clean
equipment only with damp cloth. Failure to follow this warning can result in severe personal injury or loss of life.
• Digital Sensor - do not separate when an explosive atmosphere is present.
• No ingress protections claims are made for combustible gas detection performance.
• For safety reasons this equipment must be operated and serviced by qualified personnel only. Do not operate this
equipment until after the instruction manual is read and understood for proper installation and operation.
• High off-scale reading may indicate explosive concentration.
• Wiring to or from this device, which enters or leaves the system enclosure, must utilize wiring methods suitable for
Class I, Division 2/Zone 2 hazardous locations, as appropriate for the installation.
• Explosion hazard: Do not connect or disconnect when energized.
• Use cable with insulation rated at least 24 °C above the surrounding ambient.
• As part of this Approval, it was verified that optional communication functions of this gas detection instrument while
operating at the maximum transaction rate do not adversely affect the gas detection operation and functions of the
instrument. This Approval, however, does not include or imply Approval of the "SafeSwap" feature, communications
protocol or functions provided by the software of this instrument or of the communications apparatus or software
connected to this instrument.
• Use of some accessories are not covered under the performance approval. Contact General Monitors for details.

Specific Conditions of Use S5000 Transmitter


• Under certain extreme circumstances, the non-metallic parts incorporated in the enclosure of this equipment may
generate an ignition-capable level of electrostatic charge. Therefore, the equipment shall only be cleaned with a damp
cloth.
• This fixed equipment apparatus is exclusively designed for field mounting in the vertical orientation with restrictions
placed around the conduit entry locations permitted for connection of both the Digital Sensor and IR Sensor. The
equipment is subject to the installation and orientation requirements defined in the product manual.
• The flameproof joints shall not be repaired.
• The S5000 Gas Monitor fixed gas detection system complies with EN 50271 (clause 4.8, safety integrity assessment
excluded from the assessment).

S5000 Junction Box


• Under certain extreme circumstances, the non-metallic parts incorporated in the enclosure of this equipment may
generate an ignition-capable level of electrostatic charge. Therefore, the equipment shall only be cleaned with a damp
cloth.
• The flameproof joints shall not be repaired.
• The S5000 Gas Monitor fixed gas detection system complies with EN 50271 (clause 4.8, safety integrity assessment
excluded from the assessment).

Digital Sensor
• Under certain extreme circumstances, the non-metallic parts incorporated in the enclosure of this equipment may
generate an ignition-capable level of electrostatic charge. Therefore, the equipment shall only be cleaned with a damp
cloth.
• The flameproof joints shall not be repaired.
• If the sensor is uninstalled, the equipment instruction manual shall be referenced prior to reinstalling.

General Monitors S5000 99


US 9 Appendix: General Certification Information

• The Digital Sensor is provided with a ¾“ NPT thread and shall only be connected to a suitably certified enclosure. The
installation to the certified enclosure shall be with five fully engaged threads, tightened wrench-tight.
• The Digital Sensor shall be connected directly to a suitably certified junction box or instrument for the hazardous area of
installation and thereby provide Ex protection for the flying lead connections.
• For combustible gas detection performance applications, the appropriate Digital Sensor model number shall only be
used to construct the S5000 Gas Monitor fixed gas detection system; mounted onto either the S5000 transmitter or
S5000 Junction Box enclosures and receive power and control from the transmitter.
• The Ingress Protection rating is exclusively based upon the installation instruction for orientation specified in the
operating manual.
• The Digital Sensor shall only be installed for external connection to suitably certified equipment (transmitters) providing
transient protection set at a maximum transient overvoltage of 119 V (140% of 85 Vpeak).
• Not performance approved for Class II, Class III, Zone 21. The Digital Sensor may become clogged and not detect gas
or warn user of inability to detect gas. Regularly visually inspect the sensor and apply gas to ensure a clear gas path to
the sensor if there is potential that blockage has occurred. Conditions that may lead to blockage, including but not
limited to snow, ice, water, dirt, dust, or insects necessitate more frequent inspections.

Failure to follow these warnings can result in serious personal injury or death.

General Monitors S5000 100


US 10 Appendix: HART Specific Information

10 Appendix: HART Specific Information


The S5000 Gas Monitor is available with an optional HART (Highway Addressable Remote Transducer) output
communications protocol. With this option, the S5000 complies with HART Protocol Revision 7 and uses the 16-bit
manufacturer and device codes. This document specifies all the device specific features and documents HART Protocol
implementation details (e.g., the Engineering Unit Codes supported). These specifications assume the reader is somewhat
familiar with HART Protocol requirements and terminology.

This specification is a technical reference for HART-capable HOST Application Developers, System Integrators and
knowledgeable End Users. It also provides functional specifications (e.g., commands, enumerations and performance
requirements) used during Field Device deployment, maintenance, testing, and operations. It is recommended that the 4-
20 mA output be the primary gas monitoring signal. The HART signal can be the secondary method.

Table 22 Device Identification

Manufacturer Name General Monitors Model Name S5000 Transmitter


HART ID Code 227 (0xE3) Device Type Code 115 (0x73)
HART Protocol -Revision 7 Device Revision 1
Number of Device Variables 1/2
Physical Layers Supported FSK, Wireless
Physical Device Category Transmitter

The 3.5 mA version of the IR400 is not compatible with the S5000.