WG-602 Free Chlorine User Manual V5.15-2US Jan 2018
WG-602 Free Chlorine User Manual V5.15-2US Jan 2018
WG-602 Free Chlorine User Manual V5.15-2US Jan 2018
User Manual
1
Table of Contents
1 General Safety Precautions ............................................................................... 4
2 Preface .............................................................................................................. 5
2.1 Intended Use............................................................................................... 5
2.2 Safety Precautions ...................................................................................... 5
3 Overview ........................................................................................................... 6
3.1 Features...................................................................................................... 6
3.2 System Components ................................................................................... 6
4 Installation ....................................................................................................... 13
4.1 Location Considerations ............................................................................ 13
4.2 Site Requirements and Installation ............................................................ 13
4.2.1 Mechanical Installation ....................................................................... 13
4.3 Plumbing Requirements and Installation ................................................... 14
4.3.1 Water supply and drain ...................................................................... 14
4.4 Electrical Requirements and Installation .................................................... 15
4.4.1 Connecting the Main Electrical Power ................................................ 16
4.4.2 Input Switches.................................................................................... 16
4.5 Installing electrodes .................................................................................. 17
4.5.1 Free Chlorine electrode OCS140 ....................................................... 17
4.5.2 pH electrode....................................................................................... 19
4.5.3 Temperature sensor ........................................................................... 19
5 First Time Operation and Calibration ............................................................... 20
5.1 Menu Setup............................................................................................... 20
5.2 Operator Menu .......................................................................................... 21
5.3 Technician Menu ....................................................................................... 24
5.4 Calibration .................................................... Error! Bookmark not defined.
5.4.1 Temperature Calibration..................................................................... 27
5.4.2 pH Calibration .................................................................................... 28
5.4.3 Chlorine Electrode Calibration ............................................................ 29
5.5 4 to 20 mA Output configuration ................................................................ 30
6 Routine Operation and Maintenance ............................................................... 31
6.1 Filter Cleaning ........................................................................................... 31
6.2 Shut-down and Winterizing ....................................................................... 31
6.3 Start-up and Preventive Maintenance ....................................................... 31
6.4 Chlorine Electrode Maintenance ............................................................... 31
6.4.1 Cleaning the electrode ....................................................................... 31
6.4.2 Electrolyte Change ............................................................................. 32
6.4.3 Membrane Cap Change ..................................................................... 32
6.4.4 Storage .............................................................................................. 33
7 Technical Specification .................................................................................... 33
2
Table of figures
3
1 General Safety Precautions
This section presents important information intended to ensure safe
and effective use of this product.
Warning: Only properly trained and licensed electricians should attempt to wire or service
the electronic components of the analyzer/controller.
Always verify that all electrical power source(s) are off before opening the
analyzer/controller unit or attempting to service electronic components or wiring.
Caution: Extreme caution should be used when installing, operating, and maintaining the
WATERGUARD® Analyzer. Only properly trained technicians are authorized to install and
maintain the analyzer/controller.
Always follow local health and safety regulations when performing any service on the
analyzer/controller unit or when changing chemical dosing settings.
Each relay connection is limited to 8 amps, to prevent overheating. The relays may show a
higher rating but do not connect equipment exceeding 8 amps.
All electrical connections should comply with National Electrical Code (NEC) and all local
regulations.
Caution: Do not use chemicals that reduce the surface tension. When using hydrochloric
acid, observe all safety regulations.
Electrodes:
Warning: Do not swallow the electrolyte. Avoid electrolyte contact with skin or eyes. In
case of accidental contact, wash with cold water! In case of eye inflammation, contact a
doctor immediately. Wear safety glasses and gloves when working with the electrolyte
solution.
Caution: Do not touch or damage the electrodes. The electrolyte is sensitive to oxidation:
Always keep the electrolyte bottle closed after use. Do not transfer the electrolyte to other
containers. The electrolyte should not be stored for more than one year and should be
clear (not yellow) in appearance (for use by date, see label). Avoid forming air bubbles
when pouring the electrolyte into the measuring chamber.
Caution: WATERGUARD's controller unit should not be opened except for initial
installation and troubleshooting, and should only be opened by a trained and approved
technician.
4
2 Preface
The WaterGuard WG-602 Free chlorine (FCL) is an amperometric chlorine analyzer
that measures free chlorine in water applications.
WARNING
Only properly trained and licensed electricians should attempt to wire or service
the electronic components of the analyzer. There is an Electrical Shock Hazard
when servicing this system. Always verify that all electrical power source(s) are
off before opening the analyzer unit or attempting to service electronic
components or wiring.
CAUTION
Extreme caution should be used when installing, operating, and maintaining the
WaterGuard WG-602 Water Quality Analyzer and Controller. Only properly
trained technicians are authorized to install and maintain the analyzer. Only
properly trained and licensed electricians should attempt any change to the
system’s electrical components. Only properly trained and licensed operators
should attempt to make any changes to chemical dosing levels.
NOTE
Always follow local health and safety regulations when performing any service
on the WaterGuard unit or changing chemical dosing settings.
5
3 Overview
3.1 Features
The WaterGuard WG602 Free chlorine (FCL) is an amperometric chlorine analyzer
that measures free chlorine in water applications. The analyzer includes a free
chlorine electrode, pH and temperature sensors.
The WG602 FCL analyzer includes 2 built-in 4/20 mA analog outputs, and can
support additional 4 analog outputs (total of 6 analog outputs) when required.
The WG602 FCL analyzer includes RS485 digital output that support MODBUS
communication protocol OR BLUE I communication protocol for remote
communication options. For more information about remote communication and
WATERGUARD ON-LINE remote management tools please contact your local
distributor or representative.
The WG602 FCL analyzer supports connection of external conductivity and turbidity
sensors, fully displayed. For more information about conductivity and turbidity
measurements options please contact your local distributor or representative.
6
Figure 1: WG602 FCL Analyzer - General overview
1. Controller
2. Free chlorine electrode
3. pH electrode
4. Acrylic flow cell
5. Water inlet (on/off ball valve)
6. Pre-filter (washable disc filter)
7. Pressure regulator
8. Water drain
9. Power connection
7
Figure 2: Acrylic flow cell overview
2. pH electrode
5. Flow regulator
6. Temperature sensor
8. Sampling valve
9. Sampling connection
8
Figure 3: Controller overview
1. Electrodes card
2. I/O card
3. Control panel card
9
Figure 4: Electrodes card
1. pH electrode [+]
2. pH electrode [-] sheath
3. ORP jumper
4. ORP jumper
5. FCL electrode wire A
6. FCL electrode wire K
7. FCL electrode wire GROUND
8. Temp sensor (color coded)
9. Temp sensor (color coded)
10. Temp sensor (color coded)
11. Temp sensor (color coded)
12. Flat cable from I/O card
10
Figure 5: I/O card
4. Control output pH
11
Figure 6: Control panel card
4. 12 VDC
5. 24 VDC
12
4 Installation
4.1 Location Considerations
Take extra time in selecting a location since the installation location will determine
the ease of the installation and future operation and maintenance. The location
where the WG-602 FCL is installed is dependent on various considerations:
Convenient access – The WG-602 FCL should be installed where it can easily be
viewed and operated.
Dry area – The WG-602 FCL requires a clean, dry area for safe operation and
electrical connections.
Minimum distance from supply pipe - The water sampling line that is connected to
the main pipe should be as short as possible. A long sample line creates as
unnecessary delay between supply and measurement.
Drainage – The installation location requires the sample water outlet of the acrylic
flow cell to be disposed at gravity drain.
The WG-602 FCL unit and its mounting panel are not shipped with mounting screws
or anchors. The installer must provide screws and anchors that can hold the weight
of the complete unit. The screws and anchors must be compatible with the wall
where it will be installed.
13
Figure 7: Mounting panel holes and mechanical installation
The WG-602 FCL requires a pressurized water supply to the flow cell. An isolating
valve must be installed between the main line and the pipe (tube) to the analyzer.
The water supply need to be connected to the pre-filter with a 6 mm (1/4'') feed tube.
The minimum pressure to the analyzer should be 0.5 bar (7 psi) and should not
exceed 1 bar (14.5 psi). Set the right pressure using the pressure regulator that
attached to the outlet of the pre-filter. Pull up the pressure regulating adjusting valve
and set the right pressure by turning the valve. Push down and lock it when finished.
Adjust the flow rate through the acrylic flow cell so the metal float will be in front of
the proximity flow switch. Use the flow regulating valve (on the bottom left of the
acrylic flow cell) to regulate the flow properly. A gravity drain is recommended for the
water outlet from the acrylic flow cell.
NOTE:
The distance from the main pipe to the analyzer should be as short as possible, in
order to minimize the delay time between the water being sampled and the analyzer
measurement.
14
Figure 8: water connections, pressure regulating and flow regulating
3. Pressure regulator
6. Metal float
15
4.4.1 Connecting the Main Electrical Power
CAUTION:
Verify that the power switch or circuit breaker to the non-dependent power source is
OFF.
NOTE:
The main power supply may be connected to either 100-115 or 200-230 VAC 50/60
Hz.
1. Connect the line (live) wire to the I/O board connector marked line.
2. Connect the neutral wire to the I/O board connector marked Neutral.
3. Connect the earth wire to the I/O board connector marked Ground.
4. Continue with the other electrical connections.
5. Turn on electrical power only after all electrical connections have been
completed.
1. Proximity flow switch (internal): Flow switch connected to the flow cell of the
analyzer. Connect the Black wire to "IN", Blue wire to "GND", and Brown wire
to "VCC".
2. External flow switch: Connection for an external 2-wire flow switch. If an
external switch is not connected, a jumper must be installed for the analyzer
to operate properly.
3. Flow meter: Connection of 2 or 3 wire flow meter. The analyzer will not look
for the flow meter connection unless the option is turned ON in the technician
menu; therefore, no jumper is required if a meter in not installed.
CAUTION:
Electrical connections depicted in this section are ONLY recommendations. All
electrical connections should comply with National Electrical Code (NEC) and all
local regulations.
16
4.5 Installing electrodes
General Information
Free chlorine is defined as the sum of Hypochlorous acid (HOCL), hypochlorite ions
(OCl ̄ ) and Cl2, and the proportion of the different free chlorine species is pH and
temperature dependent.
The hypochlorous acid (HOCl) contained in the medium diffuses through the sensor
membrane and is reduced to chloride ions (Cl-) on the gold cathode. On the silver
anode, silver is oxidised to silver chloride. The electron release of the gold cathode
and electron acceptance on the silver anode result in a current flow which is
proportional to the HOCL concentration in the medium under constant conditions.
Since the concentration of hypochlorous acid in the medium depends on the pH
value, pH has to be measured in order to perform an accurate compensation and
present a free chlorine value.
NOTE:
The free chlorine sensor is shipped wet, filled with electrolyte from the factory and
does not requires electrolyte refilling before using.
NOTE:
The free chlorine cable will arrive connected to the electrodes card. For more
information about wiring the free chlorine cable look at figure 4.
1. Verify that the chlorine sensor is full with electrolyte and the membrane is not
damaged. Additional information about membrane replacement or electrolyte
refilling exist in chapter 6.4 (Chlorine electrode maintenance).
2. Install the sensor into it's position in the acrylic flow cell and HAND-tighten.
3. Route the wires (if not wired) through an open hole in the gasket and connect
to the electrodes card (figure 3). Connect wire K to CL-K terminal, wire A to
CL-A terminal and the unmarked wire (ground) to the ground terminal.
Safety hint:
Some electrolyte contains diluted acids. Please read the warnings on the electrolyte
bottle.
Safety hint:
Do not Swallow the electrolyte. Avoid contact of the electrolyte with skin and eyes.
Otherwise, wash with a lot of water. in case of eye inflammation, contact a doctor.
NOTE:
It is recommended to read the complete instruction manual provided with the sensor.
17
Figure 9: Free chlorine Electrode
1. Electrode cable
3. Gold cathode
4. Measuring chamber
5. Membrane cap
18
4.5.2 pH electrode
pH electrode is shipped from the factory in the original package. Open the package
and install the electrode into its position in the acrylic flow cell and HAND-tighten.
Connect the pH cable on the top of the electrode and HAND-tighten.
NOTE:
The pH electrode cable will arrive connected to the electrodes card. For more
information about wiring the pH electrode cable look at figure 4.
Temperature sensor is shipped from the factory installed within the flow cell, and
wired to the electrodes card.
NOTE:
The Temperature sensor will arrive connected to the electrodes card. For more
information about wiring the temperature cable look at figure 4.
19
5 First Time Operation and Calibration
5.1 Menu Setup
This section describes how to configure the various settings (set points, alarms, and
calibrations) using the WG-602 control panel.
CAUTION:
The WG-602 FCL controller should not be opened except for initial installation and
troubleshooting and should only be opened by a trained and approved technician.
The WG-602 FCL menu is a simple and intuitive interface with a large display. The
control panel includes the following components:
WG-602 FCL includes operator menu and technician menu. Each menu has a
separate password.
20
5.2 Operator Menu
Each of the parameters in the operator Menu is configured in the same way. The
following procedure describes how to configure a typical setting:
1. Locate the desired parameter in the menu:
a) Press Menu until the desired parameter name appears in the LCD display.
2. Press OK. “Enter Password 100” appears in the LCD display.
3. Enter the operator password 123 (or technician password 456; both are
accepted).
4. Press the up arrow or down arrow until the password number is reached.
5. Press OK to accept the password. The parameter’s name and current setting
appear in the LCD display.
6. Press OK, again. The LCD display shows the parameter and the current setting.
7. Enter the new parameter setting:
a) Press the up arrow or down arrow until the desired value is reached.
b) The second row of the menu display, below the value that is being changed,
shows the current value.
8. Press Enter to save the new setting or Esc to abort without saving the new setting.
To change the settings of additional parameters, press menu until the desired
parameter appears in the LCD display. Repeat steps 6-8 above to set the new
parameter.
NOTE:
The factory-set operator password is 123. The operator password can be changed
in the technician menu (see technician menu setup).
NOTE:
Holding Menu while pressing up or down will advance the first digit. Holding up or
down for an extended period of time will proceed through the numbers more quickly.
NOTE:
The Menu button displays the next parameter in the list, so that the operator can
check every parameter in the menu. There is no scroll-back option. To view or
change a previous parameter in the menu, you must exit the menu by pressing
Esc, and start the above procedure from the beginning.
21
OPERATOR MENU FUNCTIONS AND DESCRIPTIONS
# Function name Description
#1 Shock chlorination Additional Cl dosing mode (On/OFF)
#2 Menu relays Manual activation of relays
#3 Cl set point 1 Control Cl relay 1 – On/OFF or proportional
#4 Cl shock set point Control Cl relay shock mode
#5 Cl set point 2 Control Cl relay 2 – On/OFF only
#6 FCl calibrated to FCl calibration
#7 Cl low alarm Cl low limit / alarm
#8 Cl high alarm Cl High limit / alarm
#9 pH set point Control pH relay 3 – On/Off or proportional
# 10 pH 7 calibrated to pH calibration
# 11 pH 4, 10 calibrated to pH calibration
# 12 pH low alarm pH low limit / alarm
# 13 pH high alarm pH high limit / alarm
# 14 ORP calibrated to ORP calibration
# 15 ORP low alarm ORP low limit / alarm
# 16 ORP high alarm ORP high limit / alarm
# 17 Temp. calibrat. to Temp calibration
# 18 Temp. low alarm Temp low limit / alarm
# 19 Temp. high alarm Temp high limit / alarm
# 20 Turbidity set point Control turbidity relay 4
# 21 NTU1 calibrated to Low Turbidity calibration
# 22 NTUh calibrated to High Turbidity calibration
# 23 Turb. high alarm Turbidity high limit / alarm
# 24 Conduc. set point Control conductivity relay 6
# 25 Conduc. calibrated to Conductivity calibration
# 26 Conduc. low alarm Conductivity Low limit/ alarm
# 27 Conduc. high alarm Conductivity High limit/ alarm
# 28 Conduc. active time Activation time of conductivity relay
# 29 Alarm delay Time delay for relay 5 (alarm)
# 30 Flow low limit Low limit for external flow meter
# 31 Flow K-factor K-factor for external flow meter
# 32 Total alkalinity Manually entered for Langelier Index
# 33 Total hardness Manually entered for Langelier Index
# 34 TDS Manually entered for Langelier Index
# 35 Language Choice of language
# 36 SYSTEM RESET On/Off for controller power
22
OPERATOR MENU AND VARIABLES LIMITS
# Function name Min value Max value Default Units
#1 Shock chlorination Off On Off ---
#2 Menu relays OFF ON OFF ---
#3 Cl set point 1 0 9.99 1.50 ppm
#4 Cl shock set point 0 9.99 3.00 ppm
#5 Cl set point 2 0.1 9.99 0.50 ppm
#6 FCl calibrated 0.1 9.99 1.50 ppm
#7 Cl low alarm 0 9.99 0.50 ppm
#8 Cl high alarm 0 10 2.00 ppm
#9 pH set point 4 14 7.40 ---
# 10 pH 7 calibrated to --- --- 7.40 ---
# 11 pH 4, 10 calibrated to --- --- 7.40 ---
# 12 pH low alarm 4 9.99 7.00 ---
# 13 pH high alarm 7 9.99 7.8 ---
# 14 ORP calibrated to 0 1999 740 mV
# 15 ORP low alarm 0 1999 600 mV
# 16 ORP high alarm 0 1999 850 mV
# 17 Temp. calibrat. to --- --- 22.0/ 71.6 C/F
# 18 Temp. low alarm 1/34 49.9/121.9 0.0/ 32.0 C/F
# 19 Temp. high alarm 1/34 49.9/121.9 100/ 212 C/F
# 20 Turbidity set point 0 200 0.60 NTU
# 21 NTU1 calibrated to --- --- 0.50 NTU
# 22 NTUh calibrated to --- --- 1.01 NTU
# 23 Turb. high alarm 0 99.9 99.99 NTU
# 24 Conduc. set point 1 10000 1000 µS/cm
# 25 Conduc. calibrated to 1 5000 740 µS/cm
# 26 Conduc low alarm 1 5000 0 µS/cm
# 27 Conduc. high alarm 1 5000 4900 µS/cm
# 28 Conduc active time --- --- 0:01 Min
# 29 Alarm delay 0 10 0:30 Min
# 30 Flow low limit 0 200 or 999 0=OFF m³/h / GPM
# 31 Flow K-factor 0.001 655 0.01 ---
# 32 Total alkalinity 0 600 0 ppm
# 33 Total Hardness 0 600 0 ppm
# 34 TDS 0 5000 0 ppm
# 35 Language --- --- English ---
# 36 SYSTEM RESET --- --- --- ---
23
5.3 Technician Menu
NOTE:
The technician menu includes advanced parameter settings and accessible
separately from the operational menu. Anyone can view the settings, however,
only someone with the technician password can make changes in the technician
menu.
NOTE:
Entering into the technician menu is simple. Press menu to enter into the operator
menu. When in the operator menu, press the up and down arrows simultaneously.
The first function in the technician menu will be #51.
NOTE:
Changing settings in the technician menu is similar to changing settings in the
operator menu. Only technician password in allowed in order to make changes.
Each of the parameters in the technician Menu is configured in the same way. The
following procedure describes how to configure a typical setting:
1. Press menu to enter into the operator menu. When in the operator menu, press
the up and down arrows simultaneously.
2. Locate the desired parameter in the menu:
a) Press Menu until the desired parameter name appears in the LCD display.
3. Press OK. “Enter Password 100” appears in the LCD display.
4. Enter the technician password 456 (only technician password is accepted).
5. Press the up arrow or down arrow until the password number is reached.
6. Press OK to accept the password. The parameter’s name and current setting
appear in the LCD display.
7. Press OK, again. The LCD display shows the parameter and the current setting.
8. Enter the new parameter setting:
a) Press the up arrow or down arrow until the desired value is reached.
b) The second row of the menu display, below the value that is being changed,
shows the current value.
9. Press Enter to save the new setting or Esc to abort without saving the new setting.
To change the settings of additional parameters, press menu until the desired
parameter appears in the LCD display. Repeat steps 7-9 above to set the new
parameter.
NOTE:
The factory-set technician password is 456. Restoration of the technician password
is not possible. If lost, replacing the control panel card is required.
NOTE:
Holding Menu while pressing up or down will advance the first digit. Holding up or
down for an extended period of time will proceed through the numbers more quickly.
NOTE:
The Menu button displays the next parameter in the list, so that the operator can
check every parameter in the menu. There is no scroll-back option. To view or
change a previous parameter in the menu, you must exit the menu by pressing
Esc, and start the above procedure from the beginning.
24
TECHNICIAN MENU FUNCTIONS AND DESCRIPTIONS
# Function name Description
# 51 Cl P factor Proportional Factor for Cl Relay 1
# 52 Cl pump period 1 Pump cycle (on+off time) for Cl relay
# 53 Cl pump freq. Cl pump max pulses/min, 0 for on/off pump
# 54 Cl averaging Displays an average of the last 4 Cl readings
# 55 Cl sensor type Not relevant for Free Cl.
# 56 pH P factor Proportional Factor for pH Relay 3
# 57 pH pump period 1 pump cycle (on + off time) for pH relay 3
# 58 pH pump freq. pH pump max pulses/min, 0 for on/off pump
# 59 pH balance type Selection of Acid or Base to adjust pH
# 60 Flow sensor Turns Flow sensor on/off (optional module)
# 61 Flow rate Choose between metric and US units
# 62 Celsius / Fahrenheit Choose between metric and US units
# 63 pH P factor Proportional Factor for pH Relay 3
# 64 Turbidity Turbidity module on/off (optional module)
# 65 NTU wiper interval Cleaning interval for turbidity sensor
# 66 Turbidity range Selection of turbidity measurement range
# 67 Conductivity range Selection of conductivity measurement range
# 68 Minutes Current time for Data Logger
# 69 Hour Current time for Data Logger
# 70 Day Current time for Data Logger
# 71 Month Current time for Data Logger
# 72 Year Current time for Data Logger
# 73 Recording interval Time interval for Data Logger
# 74 View Free Chlor Displays measurement value on LCD
# 75 View pH Displays measurement value on LCD
# 76 View Conductivity Displays measurement value on LCD
# 77 View ORP Displays measurement value on LCD
# 78 Address Controller ID – used with remote comm.
# 79 Software ver. Current SW version
# 80 Max Flow Range Maximum Flow Rate of 4-20mA module
# 81 4-20 mA Output Settings Configures 4/20 mA outputs
# 82 On Alarm go to Configures 4/20 mA outputs
# 83 Blue I /Modbus. Format Communication protocol
# 84 Operator password Change operator’s password
25
# 85 Technical password Change technician’s password
26
# 85 Technical password 1 999 456 ---
5.4 Calibration
NOTE:
Always take water for calibration from the sampling valve, NOT from the process
line directly. The analyzer should always be calibrated with water from the same
source.
1. Place the external thermometer in the flow cell or place both the PT-100 and
thermometer in the same sample.
2. Wait for temperature readings to stabilize.
3. Press menu (Operator menu) until "Temp Calibrated to" (#17) appears in the
LCD display.
4. Press OK.
5. Enter the password (operator or technician password). Press the up arrow or
down arrow until the password is reached.
6. Press OK.
7. The value that appears is the last calibrated value.
8. Press OK again.
9. Press the up arrow or down arrow until the value is the same as the
independent digital thermometer.
10. Press OK to save the new calibration or Esc to abort without saving.
11. Press Esc to return to the main display.
27
5.4.2 pH Calibration
1. Shut off the water inlet and outlet from the flow cell.
2. Remove the pH sensor and temperature probe from the flow cell.
3. Wipe the sensor probe with a dry cloth and submerge it and the (PT-100)
temperature sensor into a cup with a buffer 7 solution. Wait for the reading to
stabilize.
4. Press menu until “pH 7 Calibrated to” (#10) appears in the LCD display.
5. Press OK.
6. Enter the password. Press the up arrow or down arrow until the password is
reached.
7. Press OK.
8. Press OK again.
9. Press the up arrow or down arrow until the value is the same as the value
printed on the buffer 7 solution.
10. Press OK to save the new calibration or Esc to abort without saving.
11. Press menu until “pH 4,10 Calibrated to” (#11) appears in the LCD display.
12. Press OK.
13. Enter the password. Press the up arrow or down arrow until the password is
reached.
14. Press OK.
15. Press OK again.
16. Press the up arrow or down arrow until the value is the same as the value
printed on the buffer 4 or 10 solution.
17. Press OK to save the new calibration or Esc to abort without saving.
18. Press Esc to return to the main display.
NOTE:
When calibrating pH using buffer solutions, the reading of the pH electrode will not
stabilize if the temperature probe is not also in the buffer solution.
28
5.4.3 Chlorine Electrode Calibration
NOTE:
Calibrate the free chlorine sensor after operating at least 1 hour. Check for
calibration after 24 hours of operation. If required, perform another calibration.
NOTE:
Reliable chlorine testing device is required for proper calibration of chlorine
electrode.
The top line will display “Cl Calibrated to” and a number. The number displayed
is the last value entered for the calibration.
The bottom line will display “Cl Sensor was” and a number. This number is the
sensor reading without any calibration at the time of the last calibration.
If there is a large discrepancy between the above two numbers - the sensor
was calibrated improperly or there is a problem with the analyzer.
The analyzer uses the calibrated value, as shown on the main screen, to
determine dosing rates.
4. Press OK.
5. Enter the password. Press the up arrow or down arrow until the password is
reached.
6. Press OK.
7. Press OK again.
The display will now show “Calibrate Cl to” on the top line and “Sensor
Reading” on the bottom line.
The “Sensor Reading” is the current reading of the sensor with no calibration.
The “Calibrate Cl to” value is the new value which you want to set.
8. Press the up arrow or down arrow until the value is the same as the value given
by the digital photometer.
9. Press OK to save the new calibration or Esc to abort without saving.
10. Press Esc to return to the main display.
NOTE:
Chlorine calibration should always be performed within 25% of the set point. If
current chlorine level is 25% above or below the set point, do not perform
calibration until the chlorine level is closer to the set point.
29
5.5 4 to 20 mA Output configuration
The WG-602 TCL analyzer includes 2 built-in 4/20 mA analog outputs that can be
used for any of the measured parameters. If required, there is an option to add
additional 4 outputs on the "Conductivity / NTU card". For additional information
about the "Conductivity / NTU card" contact your local representative.
1. Enter the technician menu and scroll until “4-20mA output settings” appears in
the LCD display (#81).
2. Press OK.
3. Enter the technician password and press OK.
4. Select "Built-in ch. # (1 or 2)" and press OK.
5. Select parameter (Cl for free chlorine reading, pH, etc.) and press OK.
6. Select "Set value for 4mA" and press OK.
7. Select "Set value for 20mA" and press OK.
8. Select "set test channel" (to be used for troubleshooting purposes or initial
tests): 2mA, 4mA, 12mA, 20mA, and press OK.
9. Press Esc to return to the set-up menu and repeat the above steps for all active
4/20 mA outputs.
1. Enter the technician menu and scroll until “On alarm go to” appears in the LCD
display (#82).
2. Press OK.
3. Enter the technician password and press OK.
4. Press OK again.
5. Select the desired output 2mA, 4mA, 20mA, or hold and press OK.
6. Press Esc to return to the main display.
NOTE:
The alarm will not activate at low or high chlorine levels. It will only go on in case of
low or no flow and /or when there is a communication failure between the analyzer
and the mA output i.e. power failure.
30
6 Routine Operation and Maintenance
6.1 Filter Cleaning
The WG-602 FCL analyzer has a pre-filter installed in its water supply pipe, which
collects particles and impurities. It should be cleaned or replaced periodically,
depending on water quality. The filter is a washable disc filter. Open the filter cap and
remove the discs. Separate the discs and wash them under flowing water. Place the
disc back in the filter and close the filter cap.
31
with a gentle water jet or swirl solution of 2% hydrochloric acid (no other additives). If
the membrane cannot be cleaned properly, replace the membrane.
NOTE:
Refill the electrode with every 6-12 months, or when unstable values displayed.
Depending on the chlorine content on site, this period can be reduced or
expended.
NOTE:
If the chlorine electrode continues to display unstable values, a new membrane
cap must be used.
WARNING:
Do not swallow the electrolyte. Avoid contact of the electrolyte with skin or eyes. In
case of accidental contact, wash with a lot of cold water!
In case of eye inflammation contact a doctor immediately. Wear safety glasses and
gloves when working with the electrolyte solution.
CAUTION:
Do not touch or damage the electrolyte. The electrolyte is sensitive to oxidation:
Always keep the electrolyte bottle closed after use. Do not transfer the electrolyte
into other containers. The electrolyte should not be stored for more than one year
and should be clear (not yellow) in appearance (use by date, see label). Avoid
forming air bubbles when pouring the electrolyte into the measuring chamber.
NOTE:
Change the membrane cap once a year, or when unstable values displayed.
NOTE:
If the chlorine electrode continues to display unstable values, a check /
reconditioning by the manufacturer has to be done.
32
6.4.4 Storage
The chlorine electrode can remain in the flow cell for short term, as long as the flow
cell is not drained. If this is not ensured, remove the electrode from the flow cell and
place the yellow protection cap on top of the electrode to keep it wet. Make sure that
the inner sponge inside the yellow protection cap is wet.
NOTE:
When maintaining chlorine electrodes, follow guidelines and information in user
manuals and operating instructions that supplied together with the total chlorine
electrode.
7 Technical Specification
33
MEASUREDPARAMETERS
Chlorine electrodes Free Chlorine standard range 0.05-10 ppm or
Free Chlorine low range 0.03-5* ppm
Flow Cell Chlorine, pH, Temp, ORP
External connection Turbidity or Conductivity
FREECHLORINE(FC) MEASUREMENT
Passive-operated Chlorine sensor with gold
Free Chlorine Electrode
cathode & silver/silver chloride anode
Measuring Range 0.05-10 ppm (standard)
Accuracy ± 2 %or ± 0.01 ppm whichever is greater
Minimum Detection Limit 0.05 ppm
Resolution 0.01 ppm
Repeatability 1%span
Response time Approx. 2 min
pHoperation range 4 to 8
Body material PVC
Membrane material PTFE
pHMEASUREMENT
Electrode Ceramic diaphragm and gel filling
Measurement Range 0 to 12
Input impedance 0.5 x 1.12k Ω
ORP (REDOX) MEASUREMENT
Sensor Ceramic diaphragm and gel filling
Measurement range 0 to 2000 mV
TEMPERATUREMEASUREMENT
Sensor PT-100
Measurement range 0°Cto 100°C(32°F to 212°F)
ANLYZERFLOWMONITORING
Flow sensor Inductive proximity switch
FLOWMEASUREMENT (Main line)
Measurement range 0-1000 Cu.m/h (0-11 Mgpd)
Frequency input Via I/Ocard
Or 4-20 mA input Via NTUcard
34
ELECTRICAL CONNECTION
Power supply 100-115 VAC, 50/60 Hz, 1.0 Amp
200-230 VAC, 50/60 Hz, 0.5 Amp
Power consumption Approx. 60 VA
Power supply for RTC 3.6V Lithium Battery memory (CR2032)
MECHANICAL DATA & DIMENSIONS
Dimensions (controller) 340 x 220 x 120 mm
(L x W x D) (14.0" x 7.0" x 5.0")
Dimensions (Mounting board) 800 x 550 x 5 mm
(L x W x D) (31.5" x 21.7" x 0.2")
Weight (approx.) 9 kg (22 lbs.)
Display 5.5'' graphic monochromatic display
Cable entries PG9 cable Glands
Enclosure rating IP 65 (NEMA 4 equivalent)
OPERATIONAL REQUIREMENTS
Sample and drain connection Pressurized sample inlet and gravity drain
Inlet Pressure 0.35-1 bar (5-14.5 psi)
Outlet pressure (Closed cell) up to 0.9 bar (13 psi)
Measuring cell flow rate 35-60 l/h (9-16 gph)
Ambient temperature 2°Cto 50°C(35.6°F to 122°F)
Sample temperature 1°Cto 45°C(33.8°F to 113°F)
DATAOUTPUT
Digital communication RS485 Modbus ORBLUEI protocol
Local I/O 2 Standard 4-20 mA outputs
4 or 6 Optional 4-20 mA outputs
SECURITY
Operation password Yes
Technician password Yes
35
RELAYS
Cl (Chlorine) set point 1 250 VAC/DC8 Amp max**
Cl (Chlorine) set point 2 250 VAC/DC8 Amp max**
pH1 250 VAC/DC8 Amp max**
Turbidity control*1 250 VAC/DC8 Amp max**
Temperature control 250 VAC/DC8 Amp max**
General Alarm 250 VAC/DC8 Amp max**
CHLORINECONTROL #1
Control function PI or On/Off or frequency
Proportional band Yes
Relay function Pulse length proportional controller
Pulse frequency proportional controller
CHLORINECONTROL #2
Control function On/Off
Proportional band No
Relay function Pulse length proportional controller
Pulse frequency proportional controller
pHVALUECONTROL
Control function P or PI or On/Off or frequency
Characteristics Normal / Inverted
Relay function Pulse length proportional controller
Pulse frequency proportional controller
ORP CONTROL
Control function High alarm as chlorine override
36
No part of this publication may be reproduced, transmitted, transcribed,
stored in a retrieval system, or translated into any language or any
computer language, in any form or by any third party, without the prior
written permission of Chemical Injection Technologies Inc. Disclaimer
Chemical Injection Technologies, Inc. does not accept any responsibility
for any damage caused to its products by unauthorized personnel. Use of
non-SUPERIOR reagents and/or replacement parts will void all
warranties.
37