Breas - PV401 - Service - Manual (Ventilador)
Breas - PV401 - Service - Manual (Ventilador)
Breas - PV401 - Service - Manual (Ventilador)
TABLE OF CONTENTS
1. INTRODUCTION.....................................................................................................1
1.1 Intended use of the ventilator................................................................................ 1
1.2 Design and function of the ventilator..................................................................... 1
1.3 The scope of this manual...................................................................................... 1
1.4 Intended audience ................................................................................................ 1
1.5 Service personnel's training requirements............................................................. 1
4. REPLACEMENT PARTS.........................................................................................1
4.1 General view ........................................................................................................ 1
4.2 Casings and decals............................................................................................... 2
4.3 Drive unit.............................................................................................................. 3
4.4 Exploded view A, Motor and Transformer ............................................................. 4
4.5 Exploded view B, Casing and Base Plate ............................................................. 5
4.6 Exploded view C, Circuit Boards and Connectors ................................................. 6
4.7 Exploded view D,.Bellows and Drive Screw .......................................................... 7
4.8 Part No. list........................................................................................................... 8
5. FUNCTIONAL DIAGRAMS
5.1 Pneumatic Diagram .............................................................................................. 1
5.2 Functional Block Diagram..................................................................................... 2
7. MOTOR UNIT..........................................................................................................1
7.1 Construction ......................................................................................................... 1
7.2 Inspecting/Replacing the drive belts...................................................................... 2
7.3 Lubricating the drive screw ................................................................................... 2
7.4 Replacing the membranes in the check valves ..................................................... 3
7.5 Tubes and bellows, leakage check........................................................................ 6
7.6 Replacing the drive screw assembly ..................................................................... 7
8. ELECTRONICS ...................................................................................................... 1
8.1 Function and construction......................................................................................1
8.2 Wiring diagram, test points....................................................................................6
8.3 Main cabling diagram ............................................................................................7
8.4 Test points.............................................................................................................8
8.5 Calibration of pressure sensors..............................................................................9
8.6 Check of battery operation...................................................................................11
8.7 Replacing the alarm batteries ..............................................................................12
8.8 Electrical safety precautions................................................................................14
8.9 Upgrading the software........................................................................................15
8.10 Upgrading the software to revision MXG..............................................................16
8.10 Circuit diagrams ..................................................................................................18
8.11 Component positions...........................................................................................28
8.12 Component list ....................................................................................................33
10. APPENDICES
10.1 Engineering Change History PV 401......................................................................1
10.2 Service Record For PV 401...................................................................................1
10.3 Returning Products To Breas.................................................................................1
10.4 Breas Addresses ...................................................................................................1
1 INTRODUCTION
PV 401
Operating Manual PV 401
CONTENTS
1 INTRODUCTION.......................................................................................................1
3 INSTALLATION ........................................................................................................3
3.1 Unpacking..................................................................................................................... 3
3.2 Placement ..................................................................................................................... 3
3.3 Connecting the tubes................................................................................................... 3
3.4 Mains power supply ..................................................................................................... 3
3.5 Replacing the fuse........................................................................................................ 4
3.6 Power cord.................................................................................................................... 4
3.7 External battery power ................................................................................................. 4
4 PANEL DESCRIPTIONS...........................................................................................5
4.1 Front panel.................................................................................................................... 5
4.2 Setting panel description............................................................................................. 6
4.3 Rear panel description ................................................................................................. 7
4.4 Underside description.................................................................................................. 8
7 USAGE ................................................................................................................... 28
7.1 Adjustment of the mask and patient circuit...............................................................28
7.2 Settings in PSV mode .................................................................................................28
7.3 Settings in PCV mode .................................................................................................28
8 BATTERY OPERATION......................................................................................... 29
8.1 External Battery...........................................................................................................29
8.2 Low battery voltage alarm ..........................................................................................29
9 OXYGEN CONNECTOR......................................................................................... 30
12 ACCESSORIES ...................................................................................................... 33
1 INTRODUCTION
The PV 401 is a pressure controlled and pressure support ventilator.
It has two modes of operation; PSV and PCV. Both modes have an adjustable trigger
sensitivity setting which allows the patient to initiate ventilator-assisted breaths.
In PSV mode (Pressure Support Ventilation), the ventilator’s expiratory sense can also be
adjusted allowing the ventilator to more easily match each patient's needs.
Body floating.
• Clinic personnel and patients must read this operating manual and understand how the
PV 401 works before setting up and using the ventilator.
• To ensure maximum operational reliability, the PV 401 must be serviced at the specified
intervals by authorised service personnel.
• Oxygen may only be supplied through the patient hose, use BREAS oxygen adapter.
• Do not use the PV 401 in environments where there are explosive gases or other
inflammable anaesthetic agents present.
• Do not use patient hoses or tubes made of static or electrically conducting material.
• Electromagnetic fields greater than 10 V/m can affect operation. Examples of equipment
that produce electromagnetic fields are defibrillators, diathermia equipment, cellular
telephones, etc.
• The ventilator’s performance may deteriorate at temperatures below 5°C and above
50°C.
• Do not place the PV 401 on soft surfaces so that the air inlet on the underside can be
blocked.
• Always clean all parts that come in contact with the exhalation air before use on a new
patient.
• If an air humidifier is used, always place it lower than the PV401 to prevent water from
running into the ventilator if the humidifier tips over.
3 INSTALLATION
3.1 Unpacking
• Check the ventilator for any damage and that all ordered accessories are included.
3.2 Placement
• Place the PV 401 on its feet, on a hard, flat surface. Make sure that nothing can block
the patient air inlet under the ventilator.
• Connect the patient circuit (A) to the “PATIENT AIR” outlet on the front panel.
• Connect the thin tube (B) for the exhalation valve to the “EXH. VALVE” nipple.
• Connect the pressure measuring tube (green) (C) to the “PRESSURE” connector.
• If a bacteria filter/oxygen adapter (D) is used, connect it between hose (A) and the
patient air outlet.
PAT IEN T A IR
Exh. Valve B
Pressure
D A
• IMPORTANT! Make sure that the power cord is undamaged before it is connected to the
ventilator.
• Check that the voltage selector is set for the voltage the
P O W E R S U P P LY ventilator is to be connected to. Different countries have
different power supplies, the USA has, for example 115 V.
POWER All European countries have 230 V.
Voltage range 115 V covers 110 – 120 V
Voltage range 230 V covers 220 – 240 V
• Check that the power supply fuse has the correct rating for
the set voltage.
2
For 230 V, use a glass fuse F 315 mAL
For 115 V, use a glass fuse F 630 mAL
1 • Change the fuse by pulling the fuse holder straight out.
4 PANEL DESCRIPTIONS
4.1 Front panel
PATIE N T A IR
EX H.
VA LVE
PR ES SUR E
1 2 3
1 2 3 4
P O W E R S U P P LY O P E R ATIO N H O U R S
PO W ER
A N A L O G O U TP U T D IG ITA L O U T P U T
FUS E E X T. B AT TE R Y FUS E
5 6 7 8 9 10
(1) Air inlet and filter for patient air. See Chapter 10 for replacing/cleaning the filter.
(2) Model and serial number label.
3 4 5 6 7
The green LED shows a steady light when the ventilator is operating from
the mains supply. The LED flashes when operating from an external battery.
O N /O F F
Press and hold the button for 2 seconds to switch on or off.
The ventilator should be kept connected to the mains supply, even when not in use, to
maintain charging of the alarm batteries.
%
E S T. V O L. Liter
E S T. V O L. Liter
Pressing the function button will light the background illumination of the display.
The buttons are also used to set the date and time and as
Yes/No selection buttons when setting the alarm levels for
pressure and tidal volume, where
+ = Yes and - = No.
As part of the data storage function, the date and time is recorded to allow analysis of the
patient data in the different printouts available.
PR ESS URE m b ar
• Switch off the ventilator.
R AT E B PM The new date and/or time will be saved.
INS P. T IM E se c
T R IG G E R m b ar
TU R N
P LAT E AU :
O FF
POWER %
E S T. V O L. Liter
5.7 Alarms
Power (red LED)
Should the power supply fail during operation, the ventilator will
A LA R M alarm by the red LED for “POWER” being lit and an audible alarm.
POW ER Attempts to start the ventilator without any power source
PAT IE N T connected will also cause this alarm.
The alarm is reset once the pressure and/or the tidal volume reaches
the set value.
Alarm mute
The alarm can be muted for 2 minutes by pressing the Function
(alarm mute) button.
The alarm for low tidal volume can also be adjusted, see 5.7.7.
This section describes the PCV and PSV modes. The ventilator always starts in the mode
which was active when it was switched off.
PSV • To change between PSV and PCV mode, press the MODE button for one
second.
MODE
PCV
In PCV mode, the ventilation is controlled by the ventilator. The RATE setting corresponds to
the patient’s breathing rate expressed as the number of breaths per minute (BPM). The length
of each inspiration is controlled by the INSPIRATION TIME setting. The PLATEAU parameter
sets the speed at which the pressure increases to the set value. The lower the setting, the
slower the pressure increase will be, a higher setting will give a faster pressure increase and
therefore a longer plateau.The setting has 9 steps where 1 is the lowest and 9 is the highest.
The figure below shows how the Pressure and Flow settings control the ventilator's
function in PCV mode, where the Pressure setting is 20 cm H2O, the Inspiration Time is
1.8 seconds and the Plateau setting is 5.
A breath is started when either the set rate starts a breath or the patient triggers a breath
(if trigger is activated). The ventilator tries to reach and maintain the set pressure until
either the inspiration time setting expires or the bellows reaches its end position.
Pressure
cm H O 2
20 P atient ex hales
P re ss ure s etting
P lateau s etting
S tart of b re ath
seconds
In sp.tim e P C V
Flow
%
P atient ex hales
seconds
In PSV mode, the patient normally controls both inspiration through the trigger function,
and exhalation by the expiratory sense. If the patient’s own “triggering” ceases, the
ventilator’s rate setting will take over where the RATE setting corresponds to the patient’s
breathing rate expressed as the number of breaths per minute (BPM).
The PLATEAU parameter sets the speed at which the pressure increases to the set
value. The lower the setting, the slower the pressure increase will be, a higher setting will
give a faster pressure increase. The setting has 9 steps where 1 is the lowest and 9 is
the highest.
The figure below shows how the Pressure and Plateau (Flow) settings control the
ventilator's function in PSV mode, where the Pressure setting is 20 cm H2O, the
Expiration Sense is 25% and the Plateau setting is 5.
A breath is started when either the patient triggers a breath (if trigger is activated) or the set
back-up Rate takes over. The ventilator tries to reach and maintain the set pressure until
either the flow drops to the “EXP.SENSE” setting (percentage of maximum insp. flow), the
bellows reaches its end position or Inspiration is longer than 3 seconds.
P re s s u re
cm H O 2
P res su re setting
20
M ax . de livered press ure
S ta rt o f breath
s ec o n d s
In sp .tim e P S V
F lo w
%
10 0% M ax . flow
The figure below shows the previous two examples superimposed to illustrate how the PCV
and PSV modes differ.
P ressure
cm H O PSV
2
PCV
20
seconds
Insp.tim e P S V
Insp.tim e P C V
F lo w
%
10 0%
2 5%
seconds
5.9 Settings
This section describes the different parameters and how they are set, both in PSV and PCV
modes. Note! Certain parameters are not accessible in the PSV or PCV modes repectively.
10
-- 80
% 2. Lit LED shows which parameter is active.
EST. VOL. 0
X.X 1.8
Liter 3. Adjust to the required setting with the
– or + button.
5.9.2 Pressure
PLATEA U
Pressure meter
P R ES S U R E
Displays the pressure measured at the exhalation valve.
Range 0 to 40 mbar (cm H2O).
0 10 20 30 40
m bar / cm H 2 O
The ventilator has a Pressure alarm function which is activated if the set percentage of the
pressure setting is not reached within 15 seconds, e.g. due to leakage from the mask.
TRIGGER mbar
O F F,-2 -0.5 8
The alarm is reset once the pressure reaches its set value again.
A L A R M m b ar
• Make sure the setting panel is not locked.
PR ESS URE
R AT E SET B PM
• Press the Function button until Est. Vol. is selected.
INS P. T IM E se c The display will change as shown in the figure.
T R IG G E R m b ar
P LAT E AU : Note! The alarm setting display will only be active for 5 seconds. If
NO YES % no other button is pressed during this time the display will return
E S T. V O L. - + Liter to standard display. Press the Function button once more to
return to the Alarm Setting function.
%
(Quit). (Note! If required, the Tidal Volume alarm can also be
E S T. V O L. 1.40 Liter
set before quitting).
The value is automatically saved when Quit is selected or the
display switches back to the standard setting display.
Note! If no other button is pressed within 5 seconds the display will return to standard display.
In this case, the entire alarm setting procedure must be repeated.
5.9.4 Rate
RATE 6 10 40
BPM Setting range: 6 to 40 breaths per minute (BPM).
INSP.TIME sec
0.5 1.8 5
Press – to decrease, or + to increase the value of the
TRIGGER -0.5 mbar
OFF,-2 8
selected setting. The ventilator always gives the set number of
PLATEAU
breaths, plus any triggered breaths.
RATE 6 10 40
BPM Lit LED shows parameter selected.
IN SP.TIM E sec
0 .5 1.8 5 Setting range: 0.5 – 5 seconds
TRIGG E R OF F,-2
-0.5 8
mbar
P LATE AU Press – to decrease or + to increase the value of the selected
setting.
I/E Table
I/E Table
Table for calculating the I/E ratio based on inspirations time and rate
6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 22 24 26 28 30 32 34 36 38
0.5 1:19 1:16 1:14 1:12 1:11 1:10 1:9 1:8 1:8 1:7 1:7 1:6 1:6 1:5 1:5 1:4 1:4 1:4 1:3 1:3 1:3 1:3 1:2 1:2
0.6 1:16 1:13 1:12 1:10 1:9 1:8 1:7 1:7 1:6 1:6 1:5 1:5 1:5 1:4 1:4 1:4 1:3 1:3 1:3 1:2 1:2 1:1.9 1:1.8 1:1.6
0.7 1:13 1:11 1:10 1:9 1:8 1:7 1:6 1:6 1:5 1:5 1:4 1:4 1:4 1:4 1:3 1:3 1:3 1:2 1:2 1:1.9 1:1.7 1:1.5 1:1.4 1:1.3
0.8 1:12 1:10 1:8 1:7 1:7 1:6 1:5 1:5 1:4 1:4 1:4 1:3 1:3 1:3 1:3 1:2 1:2 1:1.9 1:1.7 1:1.5 1:1.3 1:1.2 1:1.1
0.9 1:10 1:9 1:7 1:6 1:6 1:5 1:5 1:4 1:4 1:3 1:3 1:3 1:3 1:3 1:2 1:2 1:1.8 1:1.6 1:1.4 1:1.2 1:1.1
1.0 1:9 1:8 1:7 1:6 1:5 1:4 1:4 1:4 1:3 1:3 1:3 1:3 1:2 1:2.2 1:2 1:1.7 1:1.5 1:1.3 1:1.1 1:1.0
1.2 1:7 1:6 1:5 1:5 1:4 1:4 1:3 1:3 1:3 1:2 1:2 1:1.9 1:1.8 1:1.6 1:1.5 1:1.3 1:1.1
1.5 1:6 1:5 1:4 1:3 1:3 1:3 1:2 1:2 1:1.9 1:1.7 1:1.5 1:1.4 1:1.2 1:1.1 1:1.0
2.0 1:4 1:3 1:3 1:2 1:2 1:1.7 1:1.5 1:1.3 1:1.1 1:1.0
5.0 1:1.0
5.9.6 Trigger
RATE 6
10 40
BPM Lit LED shows parameter selected.
INSP.TIME 0 .5
-- 5
sec Setting range: Off / -2 to +8 mbar.
TRIGGER OFF,-2
-0.5 8
mbar
PLATEAU
Press – to decrease or + to increase the value of the selected
-- %
setting.
When the patient starts a breath, a negative pressure is
created in the patient circuit which the ventilator registers
and immediately starts an inspiration.
• Choose a setting comfortable for the patient; normally between -0.2 to -0.5 mbar.
If the patient cannot trigger a breath, the ventilator automatically takes over and delivers
breaths according to the set rate. For each breath the patient triggers, the green LED
“TRIGGER” in the setting panel lights.
When the trigger is set to OFF, the patient cannot trigger any extra breaths.
The trigger can be set to positive values if an external PEEP valve is used.
Note! In the range 0 to +8 mbar there is a risk of the ventilator “self triggering”
5.9.7 Plateau
Press – to decrease or + to increase the value of the setting. The setting has 9 steps
between the end positions where 1 is the lowest and 9 is the highest.
As the pressure increase is experienced differently by each patient, it should be adjusted to
give each patient the best possible breathing comfort.
P LAT E A U
Can only be selected in PSV mode.
10
30 80
% Lit LED shows parameter selected.
EST. VOL. 0 1.8
Liter Setting range: 10 to 80 % of max. flow.
10 % gives the largest tidal volume.
80 % gives the lowest tidal volume.
Normal setting is 25 to 30%
Press – to decrease or + to increase the value of the setting.
The ventilator measures the internal maximum flow during inspiration. When the flow drops
below the set percentage of the maximum flow, the inspiration phase stops and an exhalation
phase starts.
This setting allows the tidal volume to be matched to different types of diagnosis/compliance.
The ventilator has a low tidal volume alarm function which is activated if the delivered tidal
volume does not reach the set value.
The alarm is reset once the tidal volume reaches its set value.
R AT E SET B PM
• Make sure the setting panel is not locked.
INS P. T IM E se c
T R IG G E R m b ar • Press the Function button until Est. Vol. is selected.
P LAT E AU : The display will change as shown in the figure.
NO YES %
E S T. V O L. - + Liter Note! The alarm setting display will only be active for 5
seconds. If no other button is pressed during this time the
display will return to standard display. Press the Function
button once more to return to the Alarm Setting function.
Note! If no other button is pressed within 5 seconds the display will return to standard display.
In this case, the entire alarm setting procedure must be repeated.
flash. R ATE 4 20 40
BPM
In PSV mode, only the Rate LED IN SP.TIM E se c
flashes. 0.5 3.3 5
Pin 2 = RX
Pin 3 = TX
Pin 5 = Ground
6 FUNCTION CHECKS
PS V In the PSV mode, the ventilator is normally controlled by the patient; inspiration
by the trigger function and exhalation by the expiratory sense.
MODE If the patient’s own “triggering” ceases, the set rate takes over the ventilation.
PCV
supply. 0 10 20 30 40
m bar / cm H 2 O
RATE 10 BPM
• Start the ventilator by pressing the
4 40
IN SP.TIM E 0.5
-- 5
sec
On/Off button for 2 seconds. The green TRIGGER OFF,-2 -0.5 8
m bar
LED comes on and the ventilator starts. PLATEAU
10 80
%
• If the Lock LED is lit, the setting panel is EST. VOL. 0
30 1.6
Liter
15
• Make sure that the red LED “PATIENT” starts
ALAR M PR ESS UR E 10 40
m bar
POWER RATE 10 BPM
to flash after 15 seconds, the flashing arrow
6 40
PAT IE N T INSP.TIME 1.8 sec
alarm indicator is shown against the Pressure
0. 5 5
TRIGGER mbar
field in the display and an audible alarm is
O F F,-2 -0.5 8
PLATEAU heard.
The alarm should stop when the pressure exceeds the set percentage of the pressure
setting.
• The alarm can be muted for 2 minutes by pressing the alarm mute
button.
6.1.2 Inspiration
• Connect a 0.5 litre test lung. Check that the test lung inflates. The pressure gauge
should show a pressure of approx. 15 mbar before inspiration stops. Depending on the
compliance of the test lung, breaths can be stopped more quickly for different test lungs.
Check that the membrane in the exhalation valve seals properly.
6.1.3 Exhalation
• The air in the test lung should easily flow out through the exhalation valve.
6.1.4 Trigger
INSP.TIME 18 sec make sure the green LED “TRIGGER” comes on and a
0 .5 5
PLATEAU
-- %
There are three parameters that can stop an inspiration and start an exhalation.
P L ATE A U (1) The inspiration flow has dropped to the value set for
10
30 80
% EXP.SENSE .
EST. VOL. Liter
0 1 .8
(2) The inspiration is longer than 3 seconds.
(3) The bellows in the ventilator reaches its end position.
P LAT EA U • If the tidal volume now drops below the set alarm
A LA R M 30 % limit for more than 15 seconds, an audible alarm is
10 80
• Switch off the ventilator by pressing the On/Off button for 2 seconds.
O N /O F F
Note! Adjust the ventilator settings to the values prescribed by the doctor.
If anything is unclear or there are doubts, perform a new function check, reading the detailed
description for each function.
In the PCV mode, ventilation is controlled by the ventilator. The rate setting
PS V controls the patient’s breathing rate. The length of each breath is controlled by
MO DE the inspiration time setting. If the trigger function is used, the patient can start
PC V an “extra” breath.
• Make sure the patient circuit is correctly connected (see section 3.3).
• Start the ventilator by pressing the On/Off button for 2 seconds. The green LED lights
and the ventilator starts.
• If the Lock LED is lit, the setting panel is locked. Press both the – and + buttons for
5 seconds to unlock.
• If necessary, change to PCV mode by pressing the MODE button for 1 second.
P R ES S U R E
PRESSURE: 15 mbar
0 10 20 30 40
RATE: 10 breaths/min m b ar / c m H 2 O
ON / OFF
A LA R M
P O W ER
PATIE N T
PSV
M O DE
PCV PCV
The alarm should stop once the pressure exceeds the set percentage of the pressure
setting.
• The alarm can be muted for 2 minutes by pressing the alarm mute button.
• Adjust to a suitable sound level ( in figure) by using a small
screwdriver on the ventilator’s rear panel.
6.2.2 Inspiration
• Connect a 0.5 litre test lung. Check that the test lung inflates. The pressure gauge
should show a pressure of 15 mbar. Make sure that the membrane in the exhalation
valve seals properly.
6.2.3 Exhalation
• Check that the air in the test lung easily flows out through the exhalation valve.
6.2.4 Trigger
RATE 6
10 40
BPM • Create a negative pressure in the patient circuit and make
INSP.TIME 0.5
18 5
sec sure the green LED “TRIGGER” comes on and a breath is
TRIGGER OFF,-2
-0.5 8
mbar started.
PLATEAU
-- %
A LA R M
PLATE AU
30 %
• If the tidal volume now drops below the set alarm limit
10 80
• Switch off the ventilator by pressing the On/Off button for 2 seconds.
O N /O FF
Note! Adjust the ventilator settings to the values prescribed by the doctor.
If anything is unclear or there are doubts, perform a new function check, reading the
detailed description for each function.
7 USAGE
7.1 Adjustment of the mask and patient circuit
• Adjust the mask straps so that it feels comfortable for the patient.
IMPORTANT! Don′t strap the mask too tight. Start with loose straps and tighten as
necessary to prevent leakage from the mask.
• Route the patient circuit in a way convenient for the patient. Try different ways to find
the best one.
• Place the PV 401 on a hard flat surface. Make sure that the air inlet on the underside
cannot be blocked.
The setting panel should normally be “locked” when the ventilator is used away
from the clinic.
In PCV mode, a more accurate setting of the rate and inspiration time is required for
maximum patient comfort and the most effective result.
Always document the set values before the patient returns home.
The setting panel should normally be “locked” when the ventilator is used away
from the clinic.
8 BATTERY OPERATION
The mains supply is always selected first if both mains power and battery power are
available. Should the mains power supply fail during operation, the ventilator will
automatically switch over to the external battery supply, if connected, and give a short
audible alarm. The ventilator automatically returns to mains power once the supply is
restored.
When battery power is being used, the green LED in the ON/OFF switch
will flash to indicate that battery power is being used.
If, during operation, the battery voltage drops below 22.5 V, an audible alarm
(short signals) is given.
If the voltage drops below 22 V, the ventilator will switch off.
9 OXYGEN CONNECTOR
Oxygen can be added via the patient circuit using oxygen adapter SA 2222F-6.
Always use an oxygen analyser to make sure that the correct oxygen level is obtained.
Measure it as close as possible to the exhalation valve (see illustration).
PAT IEN
PAT IENTT LAIR
UFT
E XH .
EXP.
VEN
VA TLVILE
PTRRYC
E SKS U R E
O xygen
analyser
SA 222 2F-6
The patient/staff should be observant if the ventilator shows signs of abnormal operation.
If in the least doubt, contact service personnel.
The patient air filter consists of three filters; a grey foam rubber inner and outer filter which
can be washed and re-used, and the middle, white filter that must not be washed and re-
used. The white filter must be replaced. The filters are located underneath the ventilator.
Wash the two grey foam rubber filters in warm water and, if necessary, use a little washing-
up liquid. Let them dry before refitting them.
The yearly service should be carried out by authorised personnel according to the separate
maintenance scheme.
11 FAULT TRACING
11.1 Fault tracing chart
The ventilator does Leakage from patient Check the tubes, mask
not give adequate circuit/mask. and exhalation valve for
pressure/volume. leakage.
Fault code “FAIL 39” is Green pressure tube Connect the patient circuit 3.3
shown in the display. in the patient circuit is correctly.
correctly connected.
A fault code “FAIL XX” The ventilator does Send the ventilator for
where XX can be 00 to not function. service.
45, except 39, is
shown in the display.
12 ACCESSORIES
Part Number Description
204 030 Patient circuit PV 401, with exhalation valve and flextube
200 040 Arm for patient circuit, incl. mount (table mounting)
Display/Outputs
Patient pressure................................ 0 to 40 mbar
Delivered tidal volume....................... 0.1 to 1.6 litres
Analog output.................................... 0 to 6V
Digital output..................................... PC connector
Indications/Alarms
Mains ................................................ Green LED
Ext. battery........................................ Green LED, flashing
Inspiration ......................................... Green LED
Trigger .............................................. Green LED
Low Pressure/Leakage alarm ........... Red LED, flashing arrow indicator, audible alarm
50 - 80 % of set pressure
Power failure alarm ........................... Red LED/audible alarm
Low Tidal volume .............................. Red LED, flashing arrow indicator, audible alarm
Power supplies
Mains supply ..................................... 115–120/230–240 VAC, 50–60 Hz, 50VA
External battery operation................. 24 VDC, max. 30W
Dimensions/Weight
Dimensions (W x H x D).................... 350 x 175 x 260 mm
Weight .............................................. 5.5 kg
14 PNEUMATIC DIAGRAM
6
2
9
1 10
3
4 13 12
5
11
7 8
Note! These instructions must be thoroughly explained to the patient together with the doctor before
the patient/care giver takes over the operation at home.
PAT IEN T A IR
Exh. Valve
Leakage test
• Check that there is no leakage from the patient circuit. This is easiest done by blocking the tube
when the ventilator delivers an inspiration. Listen for any leakage. Make sure that all hoses are
connected properly.
If a mask is used, perform a leakage test after the mask has been fitted. Instructions on how this
is done should be given by the doctor depending on what type of mask is used.
Cleaning
• Wipe the outside of the ventilator using a damp cloth with a little mild washing-up liquid or
window cleaning fluid.
• The fluid must not be allowed to enter into the ventilator.
• The responsible clinic must always instruct on the usage and how the patient circuit and mask
are best cleaned.
• Patient hose, exhalation valve, flex tube and mask are cleaned, when required, in warm water
with a mild fragrance-free soap. The thin tube to the exhalation valve does not need to be
washed. Remove these tubes before cleaning – it is important that no water enters into the
tubes. Always be careful when cleaning so as not to damage anything.
• Let all parts dry before reassembling them. (See illustration above).
Always perform a leakage test after cleaning.
SETTINGS
Patient:
Date:
Clinic:
Set by:
PSV
MODE: PSV MODE
PCV
PCV
PRESSURE: _____________mbar
RATE: _____________BPM
PRESSURE 6 40
m ba r
INSPIRATION TIME: _____________seconds RATE BPM
6 40
TRIGGER OF F,-2 8
m ba r
PLATEAU: _____________ PL ATE AU :
IMPORTANT!
A complete maintenance service (as described in this chapter) must be done every
12th month. If the ventilator is used for continuous operation (24 hours per day)
a complete maintenance service must be done every 6th month.
Every 12th month or every 6th month if the ventilator is used for See Ch./
continuous operation (24 hours per day) Section No.
Motor Unit
Replace drive belts 7.2
Lubricate felt sleeve for drive screw 7.3
Replace check valve membrane in patient air outlet 7.4.1
Test of motor unit and tube for leakage 7.5
Electronics
Calibrate the pressure sensor 8.6
Check operation using external battery 8.7
Check electrical safety levels 8.9
Accessories (where applicable)
Inspect patient circuit 3.9.9
Replace membrane in exhalation valve 3.9.9
Clean PEEP adapter, change O-ring 3.9.9
Every 5th year
Replace the alarm batteries 8.8
Every 10,000 operating hours
Replace the complete motor unit 6.5
3.6.1 Open a new Service Record, identify and register the ventilator
• Make a copy of the Service Record in the Appendices chapter.
• Fill out the model, serial number and any inventory number on the Service Record.
• Check the service schedule to see whether the alarm batteries or the complete motor
unit are due to be replaced.
3.6.3 Markings
Make sure that all markings on labels can be read:
• Make, model description, serial number.
• Warning texts.
• How does the patient check the function of the ventilator. How often?
• How many filters will the patient need of until the next service?
• Other observations?
• Check if any modification or upgrading of the ventilator needs to be done at the same
time as the service.
• Check for any visible damage to the casings and other components.
• Make sure that the strain relief for the power cord is not damaged.
3.8.2 Cabling
• Inspect all cables and their connectors. Check the front and rear panels to make
sure that cables and wires are not pinched.
• Make sure that all components, such as the motor, printed circuit boards,
connectors, etc are fastened securely.
• Make sure that the power socket is undamaged and is securely in place.
• Make sure that the touch-protection is undamaged and properly tensioned over the socket.
• Check the accuracy (±10%). Check that the LED for PATIENT lights and an audible
alarm is given.
• Check the accuracy (±10%). Check that the LED for PATIENT lights, that the text
LOW is shown in the display and that an audible alarm is given.
3.9.5 Trigger
• Set the trigger to -0.5 mbar.
• Create a negative pressure and make sure a triggered breath is given. The green LED
should light.
P V 401 m bar
• Measure the pressure, rate and insp. time and check that they are correct,
(accuracy ± 10%). The measuring should be done with a test lung or a test bladder
connected. (If these are not available, block the exhalation valve in the patient circuit).
When using a volume monitor, an exhalation valve with a PEEP valve connector is required.
The volume monitor is then connected to the PEEP valve outlet of the exhalation valve.
Volu m e m on itor
• Check that the ventilator automatically switches over to external battery operation,
that an audible alarm is given and the LED in the ON/OFF button starts to flash.
• Reconnect the power cord and make sure that the LED in the ON/OFF button shows
a steady light and the audible alarm stops.
• Clean the inside of the exhalation valve using a moist rag or wash in hot water
using a washing-up liquid. Let it air dry.
• Connect the exhalation valve to a test lung. Check that no leakage occurs during
the exhalation phase.
Note! Do not fit the O-ring to the exhalation valve cover before screwing it on.
Fit the PEEP adapter to the exhalation valve and screw on the plastic nut
4 REPLACEMENT PARTS
1 Upper casing
2 Fastening plate, rear panel
3 Printed circuit board - Hour meter
4 LCD-board
5 CPU-MDA printed circuit board package
6 Motor unit (also available as exchange unit, part No. 204740
7 Printed circuit board - Filter
8 Lower casing
10
11
12
13
1
14
15
1 Bellows
2 Clamp
3 Base plate
4 Air pressure nipple for regulator sensor G2
5 Air pressure nipple for exhalation valve, internal
6 Cover plate
7 Air pressure nipple for exhalation valve, patient circuit
8 Nipple for measuring pressure in patient circuit
9 Patient air outlet
10 Bellows top cover
11 Breaker for light beam in home position
12 Opto-switch
13 Slotted disc
14 Encoder
15 Motor
C
(2 x)
114 5
11 46
437
C
11 3 7
11 3 8
11 3 9
11 4 0
11 2 9
11 3 0
11 3 1
11 3 2
11 2 8
11 3 4
11 3 5
11 3 6
117 0
Issue: K-2
1564
1 5 6 3 (4x )
11 7 4
11 68 1 5 62 (2 x)
11 72 11 6 9
11 73
6 7 8(4 x )
11 7 4 678
117 3
4 71
4 48
4.7 Exploded view, Bellows and Drive Screw
BREAS MEDICAL
11 4 7
7 13 (4 x)
472
7 68 (3 x)
Page 7
REPLACEMENT PARTS
REPLACEMENT PARTS Service Manual PV 401
668 M4 x 30
669 M4 x 20
670 M4 x 16 F
671 M4 x 16
672 M4 x 12
673 M4 x 10 F
674 M4 - wing nut
675 M4 - serrated washer
676 M4 - flat washer
677 M4 - nut
678 M4 - lock nut
679 M3 x 8
680 M3 x 6 F
681 M3 x 6
682 M3 x 16 F
683 M3 x 12 black cap socket
684 M3 x 12
685 M3 x 10 F
686 M3 x 10
687 M3 - serrated washer
688 M3 - flat washer
689 M3 - nut
690 M3 - threaded stud
691 M2 x 10
692 M10 - serrated washer
693 KFXS 2,9 x 9,5 F
694 Fibre washer
695 WAGO - marking 301
696 WAGO - marking 1-8
697 WAGO - marking 1-6
698 WAGO - marking 1-4
699 Wire - 0,22mm blue
701 Plastic spacer- 4 x 15
702 Plastic spacer- 4 x 10
703 Plastic spacer- 3 x 10
704 Blind plug - securing bracket hole
705 Double-sided adhesive tape
706 Power cord strain relief
707 Nuts - 201
709 Fuse - T 1,25 A
711 Cable tie - mini
713 M4 x 6 cap socket
714 IC - MDA
715 IC - 33035
716 M8 x 45 stop screw
717 Cone - counterpressure when test running
718 M5 x 12 F
719 Fuse - T 6,3A
720 Shrink tube 4,8mm
721 Velcro fastening
722 BREAS - label (small transparent)
723 Magnetic strip
724 Cable tie - alarm
725 Shrink tube - 19,1mm (transp.)
5 FUNCTIONAL DIAGRAMS
5.1 Pneumatic Diagram
P V 40 1
6
2 G rey/Transparen t 9
R ed
3 12
13 10
1 4
5
G ree n G ree n 11
G ree n B la ck
B lu e B lu e
7 8
B lu e
The following tools are required to dismantle and assemble the different units in the PV 401.
• Remove the four cap head socket screws (2 mm Allen key) from the front and rear
panels (2 pcs in each panel) and the wing bolt for the power cord strain relief.
• Carefully lift the upper casing upwards and disconnect the ribbon cable and the black
ground wire from the CPU board. Put the casing and panel plates to one side.
• Disconnect the ribbon cable and the black ground wire from the LCD board.
• Reassemble with the new board in the reverse order.
• Carefully bend the rear edge of the lower casing in the direction of the arrow (see figure
above) so that the motor unit can be lifted out. Holding the motor, lift the complete unit
upwards and backwards from the lower casing.
The complete motor unit must be replaced after 10,000 operating hours. To replace the
motor unit, the motor unit must first be removed from the casing.
6.5.1 Removing the Circuit Board Package from the Motor Unit
Important! When performing this operation be careful to note how the different
coloured tubes are fitted to ensure their correct location when reassembling.
• Remove the motor unit, see operation Nos. 6.2 and 6.4.
• Without disconnecting any of the air tubes, proceed to loosen the CPU board by removing
the three screws (1 in figure). Take care of the fiber washer (2) under the centre screw.
2
C PU board
Connector CN 4
Connector CN 5
M DA boa rd
3
4
14
R ed tube c onnec tor
5
12
G ray tube con nec tor
11
G ree n tu be c onn ec to r
Filter board
10
8
9
• Lift the CPU board at the rear edge enough to disconnect the two connectors CN4 and
CN 5, see figure.
• Remount the CPU board using only the two outer screws but with the connectors
disconnected.
• Remove the screw (3) and washer (4) for the rear panel.
• Turn the motor unit over taking care not to damage the boards, etc.
• Remove the screw (5) and washer (6) that hold the rear panel.
• Cut and remove the plastic cable tie for the external battery input cable.
• Disconnect the green and gray tubes from their connectors at the rear of the patient
air outlet.
• Proceed to loosen the Filter board. Start by removing the screw and washer (8 and 9).
Loosen nut (10).
• Using a 5.5 mm socket remove the three nuts and washers holding the circuit board
package to the base plate (11, 12 and 13). Swing the Filter board out a little to access
the nut behind the patient air outlet.
• Turn the Motor Unit over taking care not to damage any of the components.
• Remove the two screws (14) that hold the cooling bracket for the MDA board.
The bracket can remain attached to the board.
• Finally, disconnect the red and black tubes from their connectors above the patient air
outlet, see figure.
• The circuit board package can now be separated from the Motor Unit.
• Fasten the bracket for the MDA board in place with the two screws (14).
• Connect the red and black tubes to their respective connectors above the patient air outlet.
• Take a good firm grip around the Motor Unit and Circuit Board Package and turn it upside
down.
• Mount the Filter board and screw on the three nuts and washers (11, 12 and 13) that hold
the Circuit Board Package. Tighten the two outer nuts using appropriate hand pressure.
Do not tighten the nut holding the Filter board at this stage.
• Swing the Filter board in place. Fasten it with screw (9) and washer (8) and tighten the
third nut.
• Mount the transformer and fasten it with the centre bolt (7) and washer. Tighten using
appropriate hand pressure.
• Mount the rear panel. Start by only fitting the screw (5). The Motor Unit must be turned over
before screw (3) can be fitted.
• Turn the motor unit and circuit board package the right way up.
• Lift the board a little so that the two connectors, CN4 and CN5 can be connected. CN4 is
connected to the Sensor connector and CN5 to the Motor connector. Make sure that all the
pins locate in their respective holes.
• Mount the CPU board and fasten it with the three screws (1). Make sure that the tubes are
NOT kinked or pinched. Do not forget the fibre washer (2) under the centre screw.
• Check, by looking from the side between the CPU and MDA boards, that the black and red
tubes run at a 45° angle down behind the large capacitor.
• Fit a new plastic cable tie to the external battery input cable.
• Carefully bend out the rear edge of the lower casing in the direction of the arrow
(see figure) fit the motor unit into the lower casing. Secure in place with the three
screws inserted from underneath the lower casing.
• Connect the ribbon cable and the black ground wire to the CPU board. Fit the upper
casing. Fit the front and rear panels and secure with the four cap head socket screws
(2 mm key) (2 pcs in each panel) and the wing bolt for the power cord strain relief.
7 MOTOR UNIT
7.1 Construction
Pos. Description
1 Bellows
2 Bellows clamps
3 Air channel block assembly
4 Air pressure nipple for regulator sensor G2
5 Air pressure nipple for exhalation valve, internal
6 Base plate
7 Air pressure nipple for exhalation valve, patient circuit
8 Nipple for measuring pressure in patient circuit
9 Patient air outlet
10 Bellows end cover
13 Slotted disc
14 Opto-switch
15 Motor
This operation is carried out with the main unit removed from the casing.
See Chapter 6 for instructions on removing the main unit.
• Check that the drive belts are undamaged and have not become slack.
Replace the belts at each service.
• Reassemble the main unit and casing. See Chapter 6.
• Remove the protection plug (3) from underneath the lower casing.
• Remove the aluminium plug (2) using two screwdrivers.
• Apply grease of type BREAS 240 AZ to the inside of the felt sleeve (1).
• Fit the felt sleeve and press the protection plug back into place.
This operation is carried out with the motor unit removed from the casing. See Chapter 6
for instructions.
Before starting work make sure you have at hand a Service kit for check valves which
contains, 1 pce membrane and 1 pce O-ring for 22 mm hose connector and 2 drive belts.
2 (4x) 3
• Remove the patient air outlet (1) by removing the four screws (2).
• Remove the membrane (3) by carefully pulling it over the screw head.
• Make sure that the new membrane lies flat against its surface.
3
5 (4x)
• Remove the circuit board package comprising the filter, MDA and CPU boards. See
Chapter 6 for instructions.
• Leave all tubes and cables in place.
• Remove the hose clamp (1) from the top cover.
• Loosen the screw (2) for the lower clamp that holds the rubber damper band (3). Unhook
the damper band and let it hang from the top clamp.
• Unscrew and remove the top cover (4) complete with the drive screw.
• Remove the four Allen bolts (5) and bolt (6).
• The air channel block assembly (7) can now be removed.
• Remove the membrane (1) by carefully pulling it over the screw head.
• Make sure that the new membrane lies flat against its surface.
• Reassemble in reverse order. Do not forget to fit the O-ring (3) and rubber seal (4).
This operation is carried out with the main unit removed from the casing. See Chapter 6
for instructions.
• Connect the main unit with the patient circuit to a test lung/reservoir bag.
PV 401
• Check that the bellows has not worked loose from its end covers.
• Connect the mains power supply or an external 24 V battery and switch on.
Pressure 40 mbar
Rate 6 BPM
Insp.time 5.0 seconds
Mode PCV
• Check that the motor stops when the pressure has reached 40 mbar. If there is any
leakage, the motor will continue to work to compensate for the air leakage. The tidal
volume should be less than 0.05 litre.
This operation should only be performed by personnel who have attended an appropriate
Breas training course and have access to the special tools required. BREAS MEDICAL
recommends the use of exchange units rather than attempting to replace the drive screw
assembly.
The circuit board package should be removed from the Motor Unit before starting this
operation, see Chapter 6 for instructions.
• Separate the Motor unit from the circuit board package, see Chapter 6.
3
5 (4 x)
• Remove the circuit board package comprising the filter, MDA and CPU boards.
See Chapter 6 for instructions.
• Loosen the screw (2) for the lower clamp that holds the rubber damper band (3).
Unhook the damper band and let it hang from the top clamp.
• Unscrew and remove the top cover (4) complete with the drive screw.
• Turn the unit over and cut the plastic cable tie for the external battery input cable.
3 16
5 (4x)
15
2
14
13 (4 x ) 12
6
11 (3 x )
10
8 (2 x )
7
• Unscrew the three stop screws (11) (Allen key 2 mm) so that approx. 2 mm of thread
is showing.
• Using a suitable puller and a counterhold for the drive sleeve, approx. 18 mm diameter,
pull off the pulley wheel (12) taking care that none of the components are damaged.
• The drive sleeve assembly can now be knocked out from underneath using a plastic
hammer.
• To remove the lower bearing, unscrew the four screws (13) holding the black plastic
ring (14). Work the bearing (15) up with a flat screwdriver.
• To remove the drive screw from the top cover, grip the screw firmly in a vice (with
protective jaw covers) and unscrew the centre screw (16). Note! Lock fluid is used at
assembly which can make loosening difficult.
• Fit the lower bearing (15) in the recess under the bottom cover. Fit the black plastic
ring (14) and fasten with the four screws (13).
• Arrange good support for the base plate. Apply a little grease to the sleeve assembly
and from above, press it down through the lower bearing making sure the upper
bearing seats properly in the recess provided. Note! If the sleeve is being reused,
remove any rough edges caused by the locking screws for the pulley wheel.
• To fit the pulley wheel (12), turn the Motor unit upside down. Note! Inspect the pulley
wheel grooves for damage before fitting. Remove one of the stop screws. Arrange for
a suitable counterhold for the drive sleeve so that the red plastic insert does not get
damaged. Press on the pulley wheel. Align the screw hole with one of the recesses in
the drive sleeve. With the hole aligned, tighten the other two stop screws. Screw in
the third screw and tighten.
• Before installing the air channel block assembly, replace the membranes for the check
valve and O-rings. See Section 7.4.
• Install the air channel block assembly and fasten with the four Allen bolts (5) from
above the bottom cover and bolt (6).
• Grip the new screw firmly in a vice (with protective jaw covers). Apply a little lock fluid
to the centre screw threads. Align the top cover centre screw hole and fit the centre
screw. Tighten using normal hand pressure.
• Apply a little grease to the drive screw. Screw in the top cover assembly all the way
down. This is to ensure that grease gets applied to the threads in the drive sleeve.
• Unscrew the top cover assembly and remove any excess grease from the top of the
drive sleeve. Make sure the drive sleeve threads are greased.
• Hook the rubber damper band under the lower clamp and tighten the clamp screw.
• Pull the bellows up around the top cover and fit the hose clamp.
• Before fitting, apply grease type BREAS 240 AZ to the inside of the felt sleeve (10).
• The Motor unit should be tested for 24 hours before it is installed in a machine.
8 ELECTRONICS
8.1 Function, construction
The electronics, optics, mechanics and pneumatics in the PV 401 are integrated. To fully
understand the electronics used in the Breas PV 401, you must know how to use the
ventilator, have studied the air flow diagram and acquainted yourself with the mechanical
construction.
The block diagram (PV401T1) below shows how the electronics are arranged and how they
are connected to other components.
The electronics of the PV 401 comprises 5 printed circuit boards, the motor with sensor, rear
panel with connectors, and the transformer.
R74
All pushbuttons, the LCD display, LEDs and bar indicator for the pressure gauge are mounted
on this board.
The LCD display contrast can be adjusted by R74 (underneath the board) without removing
the board.
The master processor U2 has its program stored in EPROM U6. U2 reads all button presses,
stores all settings in its integrated eeprom memory, presents the settings on the display and
controls the LEDs.
The values are saved just prior to the next breath and the display flashes once. U2 manages
all alarm functions, such as low and high pressure, low volume, low power, etc.
The pressure sensors, G1, which monitors and presents the patient air pressure, and G2,
which is used for regulating the pressure, are located here.
The magnetic valves MV1 and MV2 are located underneath the board.
MV1 is used for calibration of the zero pressure level for the pressure sensor G1.
When the trigger function is activated, G1 is auto-calibrated between each breath.
MV2 is used as a safety valve at high pressure alarm.
Motor control, charging of alarm batteries and part of the power supply is done from the MDA
board.
The transformer supplies CN6 on the MDA board, via the filter board, with 18 VAC rectified. If
an external 24 VDC supply is connected, the supply is fed via the filter board to CN6.
At switch-on, relay K1 supplies the rest of the MDA board and other boards with power.
Motor control is managed by the slave processor U28, control circuit U30 and the transistors
MF1-MF6.
U 28 calculates the moment the motor started based on the settings and regulates towards
the set patient pressure with the help of G2.
The reflection sensor for the motor registers the home position. The opto switch counts the
number of pulses given through the rotating slotted disc on the shaft of the motor to register
many revolutions the motor has made and from this, calculate how far the bellows has
travelled. The estimated tidal volume, which is shown in the display, is calculated using the
number of pulses counted after each breath
This board comprises the hour meter LCD display (1), the alarm buzzer (2), the trim
potentiometer (3) to adjust the sound level with, and the analog (4) and digital (5) connectors.
1 2 3
4 5
This board comprises the filter used to prevent electro-magnetic interference both to and
from the ventilator. Also located here is the relay which disconnects the external battery
when mains power is available.
9 P O L E P a n d u it C E 10 0 F 2 4 -9
1 A node RE D
1
Hom e 2 C athode
4 E mitter
BLA CK
2
p o sitio n 3 C ollector W H IT E
3
G P 1L03
4
CN 5
BLU E 5
W H IT E 6 CN 2 CN 7
E nc o d er RE D 7
BLA CK 8
O PB930W 51 G REE N 9
5 P O L E P a n d u it C E 10 0 F 2 4 -5
MDA CPU LC D
RE D 1
CM 2 CN 1
BLA CK 2
O RAN G E 3
Y E L LO W 4
R E L IA N C E
M O TO R CN 4
W H IT E 5
3 P O L E P a n d u it C E 10 0 F 2 4 -3
PU RP LE 6
BLU E 7
M a in s CN 6
CN 3
BR OW N 8
inp u t
FUS E
F 315 m AL TR ANS FOR M ER
F ilte r b oa r d 4 P O L E P a n d u it
C E 1 0 0 F 24 - 4 CN 1
Y E L LO W RE D RE D
1 3
11 5 /2 3 0 V
B atte ry G REE N CN 8 TM
RE D
inp u t BLU E
BLA CK
2 4 RE D
1 3 RE D DB 1
F U S E F 3 .1 5 m A L
RE D CN 9
BLA CK 2 4 BLA CK DB 2
R74
MDA board:
TP 1 (TC2, 9&10) GND
TP 2 (TC2, 2&6) Unregulated on 20-32 VDC -
TP 3 (TC2, 3) 24 V on 24 VDC -
TP 4 (TC2, 4) 12 V on 12 VDC -
TP 5 (TC2, 5) 5V on 5 VDC -
(TC2, 7) +NET mains 20 - 32VDC -
BT1 (TC2, 1) Alarm battery off, mains 7.8 ± 0.5 VDC -
(TC2, 8) +EXT ext.batt connected 24VDC
CPU board:
TP 6 (TC1, 2) 5V on 5 VDC -
TP 7 (TC1, 1) 9V on 9 VDC -
TP 8 (TC1, 5-10) GND
TP 9 (TC1, 3) Pressure amplifier G1 on P=0 mbar 1.0 VDC R31,(R45)
P=30 mbar 4.0 VDC
TP 10 (TC1, 4) Pressure amplifier G2 on P=10 mbar 1.0 VDC R32,(R46)
P=30 mbar 3.0 VDC
PV 401 has two pressure sensors, G1 and G2. G1 monitors the pressure at the exhalation
valve and displays the value on the pressure gauge. G2 is the pressure regulator sensor.
Let the ventilator run with a test lung connected (with the compliance high enough so
that at a pressure of 40 mbar, the volume is no larger than 1.2 litre) for 30 minutes so
that it is well warmed-up.
PV 401 m bar
T P8 : G n d
T P9 : G 1
R 31 : G 1 O ffs et
R 45 : G 1 G a in
R 32 : G 2 O ffs et
R 46 : G 2 G a in
• Repeat steps 1-4 until both pressures are maintained without any adjustment.
• Make sure that the pressure, at each breath, stabilises at approx. 30 mbar.
If not, adjust R45.
PRESSURE
0 10 20 30 40
mbar cm H 2O
• Repeat steps 1-4 until both pressures are maintained without any further adjustment.
This procedure checks battery operation and low external battery voltage alarm.
Pressure: 6 mbar
Rate: 8
Insp.time: 3 seconds
Mode PCV
• Connect an adjustable DC power supply unit to the external 24V socket of the ventilator.
• Switch on the ventilator, and make sure the LED in the ON/OFF button flashes.
• When the external battery voltage drops to 22.5V (± 0.2V) the low battery voltage alarm will
start.
• Check that a “power alarm” is given as an audible signal every 5th second at 22.5 V (± 0.2 V).
• Again, slowly reduce the voltage down to 22.0 V. Check that the ventilator switches off at
22.0 V(± 0.2 V) and that the red LED for “POWER” is lit.
The alarm battery should be replaced after 5 years from the delivery date and then every
5th year.
The alarm batteries are located on the MDA board and provide power for the Mains power
failure and Low voltage alarms.
A larm b attery
Jum per J1
• Disconnect the mains power supply and any external battery. Remove jumper J1 from
the MDA board.
• Leave all the wires in place. This makes it easier to get access with a soldering iron.
• Clip the two plus pins of the outer battery (Y). Unsolder the battery from the minus pole.
C le an so ld er fro m holes
• Unsolder the clipped-off pins and suck away any remaining solder from the holes.
• The PV 401 checks, at each start cycle, the condition of the alarm batteries.
• If a fault occurs with the alarm batteries, the ventilator will not start and the LED for
“POWER” will light and error message FAIL 08 is displayed. If this should happen after
replacing the batteries, let the ventilator remain switched on for a few hours, running
from the mains supply, so that the batteries are charged.
• With the ventilator running from the mains supply, pull out the power cord. The red LED
“POWER” should light and an audible alarm be given. When the power supply is
reconnected, the ventilator should re-start normally.
Electrical safety measuring must be taken according to IEC 601. However, the insulation
resistance can be measured instead of the voltage test specified by the standard.
The measurements can be done using an automatic electrical safety tester. All tests must be
done according to class II type BF.
Supply voltage
Note the power voltage reading. As the currents measured are directly in relation to the supply
voltage, the voltage must be noted for each service check. This allows all measurements made
on the same ventilator to be compared with measurements made on different occasions.
Insulation
The insulation resistance is measured using a 500 VDC power supply. The most suitable
method is to connect the plus lead to both the ventilator power socket pins and the minus lead
to the casing or patient air connector. The measurements taken during the delivery inspection
provide the reference values with which measurements taker during later services are
compared. If no reference values are available, then the value for the insulation resistance
should be > 20MΩ.
Leakage currents
Leakage currents are measured at different parts of the ventilator using an RC circuit to
ground.
The measurements are taken partly at normal case (NC) and at the single fault condition
(SFC). Reverse the polarity of the power supply and note the lowest value.
Leakage currents to ground must not exceed the stated limit values.
Leakage currents with mains power supply at the patient connected part
This test must be done using an automatic electrical safety tester that has this function.
See the safety instruction for the tester.
• Make sure you have the correct tool to pull the PLCC capsule from its socket.
• Disconnect the mains supply, and if connected, the external battery supply.
• Pull out U28 using a PLCC puller and replace with the new version software.
• Check that the magnetic valves underneath the CPU board are not pinched.
The software revision MXG contains an internal memory and an adjustable low pressure
alarm level.
CPU board
Besides replacing memory chip U6, memory chip U41 must also be installed. Make sure
the chips are installed in the right directions, see figure.
U 6 M XG
U 41 D allas m em ory ch ip
MDA board
Check that the slave processor has software revision SGP. This can be checked on the
display at start up. If not, replace the chip with version SGP.
Sh ould be revisio n SG P
After installation:
• Start the ventilator.
• Press and hold the Function button and start the ventilator in order to access
"Set date and time" mode.
• Set the date and time using the Function key and the - and + buttons.
• Switch off the ventilator when the display shows "TURN OFF PWR"
• To erase the memory, press and hold the + button and start the ventilator. The display
shows the text "DUMP MEM PUSH FUNC TO DUMP". Press the Function button.
The display will now show the text "WAIT DUMP IS ON". When the display shows
"DUMP RDY TURN OFF PWR", press the PSV/PCV Mode button for 10 seconds until
the display shows "MEM CLR RDY TURN OFF PWR". Switch off the ventilator.
• Check that the correct version of the Operating Manual accompanies the ventilator.
This covers the additional functions available with the MXG software.
• For full information about the internal memory, please refer to the PV 401 Calendar
Data Analysis Program manual (Doc. No. 1399).
9 FAULT TRACING
9.1 Fault tracing
Pressure indicator shows Internal supply tube Check the tubes and connectors. 5
no pressure reading. blocked.
If the PV 401 has the MXG software installed (see section 8.10) an Error Code memory
is provided which can store the last 20 error codes generated. (Earlier software will only
show the current error code and this information is lost once the machine is switched off.)
The error codes provide useful information when fault-finding or carrying out service
work. To read the error codes, proceed as follows:
• Press and hold the minus button.
• Start the ventilator, keeping the minus button pressed.
• The display will now show the message FAIL MEM. PUSH FUNC TO SHOW
• Press the Function button.
• The latest error code stored is shown first. The information given is YEAR, MONTH,
DAY, HOUR, MINUTE, the text FAIL and the error code itself, see below for
explanations.
• Press the function button to page back through the error codes to the earliest code
stored.
• After the last code has been displayed, the message; SHOW FAIL RDY, is shown.
Switch the ventilator off.
The table below shows each error code, its meaning and the action required to remedy
the situation. Where the Action Required column contains more than one action they are
listed in the order they should be carried out e.g. if recommended action No. 1 does not
solve the problem proceed with action No. 2 and so on.
Refer to Chapter 6 for information about removing the different boards etc.
Error
code Explanation Action required
00 Slave processor does not answer. 1 Replace ribbon cable between
CPU - MDA board.
2 Replace MDA board.
3 Replace CPU board.
01 Movement of bellows failed to start. 1 If ventilator is cold, let it warm up.
2 Replace motor.
3 Replace MDA board.
02 Slave processor answer fail. 1 Replace ribbon cable between
CPU - MDA board.
2 Replace MDA board.
3 Replace CPU board.
03 Movement of bellows failed. 1 If ventilator is cold, let it warm up.
2 Replace motor.
3 Replace MDA board.
Issue: K-2
Comments and notes:
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
Issue: K-2
Service Manual PV 401 RETURNING PRODUCTS TO BREAS
If a product needs to be returned to Breas for any reason, e.g. for service, warranty,
repair or upgrade, the following routine must be followed to ensure that the correct action
is taken and to avoid unnecessary delays.
Pack the product in its original packaging. If this is not available, pack the product in
packaging suitable for the transportation to Breas.
Fill out the Customer part of a Service Report and pack it together with the product. The
Service Report, completed by Breas, will be returned with the product.
A copy of the Service Report is provided on the next page. This can be photocopied and
used as required.
..........................................................................................................................................
Our Ref. No:
Reference: ............................................. Ref. No. ....................................................
..........................................................................................................................................
..........................................................................................................................................
..........................................................................................................................................
Action taken:
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
______________________________________________________________________
Issue: K-2
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
__________________________________________________________________________
Issue: K-2