Philips LR - v60 - Pocketguide PDF
Philips LR - v60 - Pocketguide PDF
Philips LR - v60 - Pocketguide PDF
respiratory care
Pocket guide
Philips Respironics V60 ventilator
General operation
Mask and port settings 6
Setting changes 6
Setting rate 6
Rise time 6
I-Time 7
Mode changes 7
Batch changes 8
Alarm message navigation 8
Alarm settings 9
Low rate 9
Menu tab 10
Screen Lock 10
Patient data and waveforms 11
Standby 12
Help button 12
Mask leak symbols 13
Exhalation port settings 14
Exhalation port test 14
Navigation ring
and Accept
Back
Option labels
Power cord
Cooling and retainer
fan filter
Remote
alarm and
nurse call
RS-232 serial and analog I/O connector connector
Side
Bracket
holding
filter
Air inlet
filter Air inlet
Patient circuits
Assemble the patient circuit, including main flow
bacterial filter, proximal line, and humidifier (if desired).
Setting changes
1. In the Settings window, touch the setting to be changed.
2. Adjust the setting by using either the arrow keys or the
navigation ring.
3. Once the correct value has been chosen, press Accept.
Setting rate
Because the V60 is intended to augment ventilation in patients
who are spontaneously breathing, the rate should be set
as a back-up rate in the case of apnea. If the patient fails
to trigger a breath through Auto-Trak within the interval
determined by the rate setting, the ventilator triggers a
mandatory breath.
I-Time
Setting I-Time adjusts the inspiratory time for a machine-
triggered breath, therefore influencing the I:E ratio in V60
machine-triggered breaths. Inspiratory time is controlled
by the patient in a patient-triggered breath.
Mode changes
The active ventilation mode is displayed in the upper
left corner of the screen. To set or change a mode,
do the following:
Low rate
On the V60 ventilator, the low rate alarm may also serve as
an apnea alarm. It is recommended to set the low rate alarm
higher than the backup rate but lower than the patient’s
spontaneous rate. If the low rate alarm value is set at or
below the set rate, the low rate alarm cannot be triggered,
and the alarm is essentially disabled. A text warning will
appear on the left side of the Settings screen if the user
sets the low rate at or below the set (backup) rate.
Screen Lock
Screen Lock deactivates all buttons and tabs on the
touchscreen except the 100% O2 key (available in 2.30
software*), Alarm Silence, Alarm Reset, the Alarm Message
button, and Help icon. The tabs will be grayed out.
To activate Standby
1. Press the Standby tab. The Entering Standby
window appears.
2. Disconnect the patient from the ventilator.
Help button
Press the Help icon (?) to display additional information.
Touch the screen anywhere to return to normal operation.
Leak symbol
(printed on mask) Patient interface
Respironics
Disposable
No
Exhalation
Port (DEP)
Respironics
Whisper No
Swivel
Respironics
Plateau
Yes
Exhalation
Valve (PEV)
Other
exhalation Yes
port
100% O2 key*
Upon pressing the 100% O2 key the V60 delivers 100% O2
for 2 minutes. The clinician is also given the option to cancel
the action or add an additional 2 minutes of 100% O2.
Ramp
The ramp time allows the patient to adapt to ventilation
gradually by increasing inspiratory and expiratory pressures
(IPAP and EPAP/CPAP) from sub-therapeutic to user-set
pressures over a user-set interval (5-45 minutes).
The
illustration
correlates
with the
C-Flex
setting
Pressure-controlled ventilation
In pressure-controlled ventilation (PCV), breaths with a
user-set IPAP and I-Time are delivered to the patient. The
patient can trigger an inspiration and, therefore, control the
rate. However, the patient does not control the inspiratory
time. Also be aware that any changes in EPAP without an
equal change in IPAP will change the pressure support.
Target VT
Starting AVAPS
1. When switching from S/T mode to AVAPS, set the Min P
at the current IPAP pressure.
2. During AVAPS startup, there may be a period of time
before the target tidal volume is achieved. If the target
VT is not achieved at the current Min P setting, increase
the Min P until the target VT is reached. Remember,
the VT displayed on the V60 is a six-breath average,
so the effect of a settings change may not be fully
reflected for several breaths.
3. Once the target VT is reached, reduce Min P slightly
to allow the AVAPS algorithm to adjust.
4. If target VT is not achieved due to a low Min P setting,
an informational message will appear. Adjust Min P
accordingly unless the maximum pressure for the patient
has been reached.
5. If target VT is exceeded because Min P is set too high,
an informational message will appear. Adjust Min P
accordingly unless the minimum pressure for the patient
has been reached.
6. Set the high and low VT alarms appropriately.
7. In AVAPS, the EPAP setting must be at least 1 cmH2O below
the Min P setting. In some cases, an increase in the Min P
setting is required before increasing the EPAP setting.
90
80
70
60
50
40
30
20
10
0
30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
FIO2 (%)
140
120
100
80
60
40
20
0 30 35 40 45 50 55 60 65 70 75 80 85 90 95 100
FIO2 (%)