Oscilloscope PDF
Oscilloscope PDF
Oscilloscope PDF
2000 X-Series
Oscilloscopes
User's Guide
Notices
© Agilent Technologies, Inc. 2005-2011 Warranty defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
No part of this manual may be reproduced in The material contained in this docu-
any form or by any means (including elec- no greater than Limited Rights as defined in
ment is provided “as is,” and is sub- FAR 52.227-14 (June 1987) or DFAR
tronic storage and retrieval or translation ject to being changed, without notice, 252.227-7015 (b)(2) (November 1995), as
into a foreign language) without prior agree- in future editions. Further, to the max-
ment and written consent from Agilent applicable in any technical data.
imum extent permitted by applicable
Technologies, Inc. as governed by United law, Agilent disclaims all warranties,
States and international copyright laws. Safety Notices
either express or implied, with regard
to this manual and any information
Manual Part Number contained herein, including but not
limited to the implied warranties of CAUTION
75015-97012
merchantability and fitness for a par-
ticular purpose. Agilent shall not be A CAUTION notice denotes a haz-
Edition
liable for errors or for incidental or ard. It calls attention to an operat-
Third edition, June 2011 consequential damages in connec- ing procedure, practice, or the like
Printed in Malaysia tion with the furnishing, use, or per-
formance of this document or of any that, if not correctly performed or
Agilent Technologies, Inc. information contained herein. Should adhered to, could result in damage
1900 Garden of the Gods Road Agilent and the user have a separate to the product or loss of important
Colorado Springs, CO 80907 USA written agreement with warranty data. Do not proceed beyond a
terms covering the material in this
document that conflict with these CAUTION notice until the indicated
Print History terms, the warranty terms in the sep- conditions are fully understood and
arate agreement shall control. met.
75015-97000, January 2011
75015-97011, February 2011 Technology Licenses
75015-97012, June 2011 The hardware and/or software described in
this document are furnished under a license WA R N IN G
Trademarks and may be used or copied only in accor-
Java is a U.S. trademark of Sun Microsys- dance with the terms of such license. A WARNING notice denotes a
tems, Inc. hazard. It calls attention to an
Restricted Rights Legend operating procedure, practice, or
Sun, Sun Microsystems, and the Sun Logo
are trademarks or registered trademarks of If software is for use in the performance of a the like that, if not correctly per-
Sun Microsystems, Inc. in the U.S. and other U.S. Government prime contract or subcon- formed or adhered to, could result
countries. tract, Software is delivered and licensed as
“Commercial computer software” as
in personal injury or death. Do not
defined in DFAR 252.227-7014 (June 1995), proceed beyond a WARNING
or as a “commercial item” as defined in FAR notice until the indicated condi-
2.101(a) or as “Restricted computer soft- tions are fully understood and
ware” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or met.
contract clause. Use, duplication or disclo-
sure of Software is subject to Agilent Tech-
nologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
When unpacking and using the • Chapter 1, “Getting Started,” starting on page 19
oscilloscope for the first time, see:
When displaying waveforms and • Chapter 2, “Horizontal Controls,” starting on page 43
acquired data, see: • Chapter 3, “Vertical Controls,” starting on page 55
• Chapter 4, “Math Waveforms,” starting on page 63
• Chapter 5, “Reference Waveforms,” starting on page
75
• Chapter 6, “Digital Channels,” starting on page 79
• Chapter 7, “Display Settings,” starting on page 99
• Chapter 8, “Labels,” starting on page 105
When setting up triggers or changing • Chapter 9, “Triggers,” starting on page 111
how data is acquired, see: • Chapter 10, “Trigger Mode/Coupling,” starting on
page 133
• Chapter 11, “Acquisition Control,” starting on page 141
Making measurements and analyzing • Chapter 12, “Cursors,” starting on page 157
data: • Chapter 13, “Measurements,” starting on page 167
• Chapter 14, “Mask Testing,” starting on page 187
When using the built-in waveform • Chapter 15, “Waveform Generator,” starting on page
generator, see: 199
When saving, recalling, or printing, • Chapter 16, “Save/Recall (Setups, Screens, Data),”
see: starting on page 205
• Chapter 17, “Print (Screens),” starting on page 219
When using the oscilloscope's utility • Chapter 18, “Utility Settings,” starting on page 223
functions or web interface, see: • Chapter 19, “Web Interface,” starting on page 241
For reference information, see: • Chapter 20, “Reference,” starting on page 255
1 Getting Started
Inspect the Package Contents 19
Install the Optional LAN/VGA or GPIB Module 22
Tilt the Oscilloscope for Easy Viewing 22
Power-On the Oscilloscope 23
Connect Probes to the Oscilloscope 24
3 Vertical Controls
To turn waveforms on or off (channel or math) 56
To adjust the vertical scale 57
To adjust the vertical position 57
To specify channel coupling 57
To specify bandwidth limiting 58
To change the vertical scale knob's coarse/fine adjustment
setting 58
To invert a waveform 59
Setting Analog Channel Probe Options 59
To specify the channel units 60
To specify the probe attenuation 60
To specify the probe skew 61
5 Reference Waveforms
To save a waveform to a reference waveform location 75
To display a reference waveform 76
To scale and position reference waveforms 77
To adjust reference waveform skew 77
To display reference waveform information 78
To save/recall reference waveform files to/from a USB storage
device 78
6 Digital Channels
To connect the digital probes to the device under test 79
7 Display Settings
To adjust waveform intensity 99
To set or clear persistence 101
To clear the display 102
To select the grid type 102
To adjust the grid intensity 103
To freeze the display 103
8 Labels
To turn the label display on or off 105
To assign a predefined label to a channel 106
9 Triggers
Adjusting the Trigger Level 112
Forcing a Trigger 113
Edge Trigger 113
Pattern Trigger 116
Hex Bus Pattern Trigger 118
Pulse Width Trigger 119
Video Trigger 121
To trigger on a specific line of video 125
To trigger on all sync pulses 126
To trigger on a specific field of the video signal 127
To trigger on all fields of the video signal 128
To trigger on odd or even fields 129
10 Trigger Mode/Coupling
To select the Auto or Normal trigger mode 134
To select the trigger coupling 136
To enable or disable trigger noise rejection 137
To enable or disable trigger HF Reject 137
To set the trigger holdoff 138
External Trigger Input 138
12 Cursors
To make cursor measurements 158
Cursor Examples 161
13 Measurements
To make automatic measurements 168
Measurements Summary 169
Snapshot All 171
Voltage Measurements 171
14 Mask Testing
To create a mask from a "golden" waveform (Automask) 187
Mask Test Setup Options 189
Mask Statistics 192
To manually modify a mask file 193
15 Waveform Generator
To select generated waveform types and settings 199
To output the waveform generator sync pulse 202
To specify the waveform generator output load 203
To use waveform generator logic presets 203
To restore waveform generator defaults 204
18 Utility Settings
I/O Interface Settings 223
Setting up the Oscilloscope's LAN Connection 224
To establish a LAN connection 225
Stand-alone (Point-to-Point) Connection to a PC 226
File Explorer 227
Setting Oscilloscope Preferences 229
To choose "expand about" center or ground 229
To disable/enable transparent backgrounds 230
To load the default label library 230
To set up the screen saver 230
To set AutoScale preferences 231
Setting the Oscilloscope's Clock 232
Setting the Rear Panel TRIG OUT Source 232
Performing Service Tasks 233
To perform user calibration 234
To perform hardware self test 236
To perform front panel self test 236
To display oscilloscope information 236
To display the user calibration status 237
To clean the oscilloscope 237
To check warranty and extended services status 237
To contact Agilent 237
19 Web Interface
Accessing the Web Interface 242
Browser Web Control 243
Real Scope Remote Front Panel 244
Simple Remote Front Panel 245
Remote Programming via the Web Interface 246
Remote Programming with Agilent IO Libraries 247
Save/Recall 247
Saving Files via the Web Interface 247
Recalling Files via the Web Interface 249
Get Image 249
Identification Function 250
Instrument Utilities 251
Setting a Password 252
20 Reference
Specifications and Characteristics 255
Measurement Category 255
Oscilloscope Measurement Category 256
Measurement Category Definitions 256
Transient Withstand Capability 257
Index
1
Getting Started
Inspect the Package Contents 19
Tilt the Oscilloscope for Easy Viewing 22
Power-On the Oscilloscope 23
Connect Probes to the Oscilloscope 24
Input a Waveform 25
Recall the Default Oscilloscope Setup 25
Use Auto Scale 26
Compensate Passive Probes 27
Learn the Front Panel Controls and Connectors 28
Learn the Rear Panel Connectors 37
Learn the Oscilloscope Display 38
Access the Built-In Quick Help 40
This chapter describes the steps you take when using the oscilloscope for
the first time.
19
1 Getting Started
N2862B probes
(Qty 2 or 4)
Documentation CD
LAN/VGA Module
Module Slot
GPIB Module
L6GC>C</B6>CI6>C
<GDJC9ID 6KD>9
:A:8IG>8H=D8@
s &%%"&'%K!*%$+%$)%%=o
s &%%"')%K!*%$+%=o
&%% LViihB6M
The LAN/VGA or GPIB module must be installed before powering on the oscilloscope.
NOTE
Flip-Out Tabs
Always use a grounded power cord. Do not defeat the power cord ground.
WARNIN G
Do not negate the protective action of the ground connection to the oscilloscope. The
WARNIN G
oscilloscope must remain grounded through its power cord. Defeating the ground
creates an electric shock hazard.
Input a Waveform
The first signal to input to the oscilloscope is the Demo 2, Probe Comp
signal. This signal is used for compensating probes.
1 Connect an oscilloscope probe from channel 1 to the Demo 2 (Probe
Comp) terminal on the front panel.
2 Connect the probe's ground lead to the ground terminal (next to the
Demo 2 terminal).
Horizontal Normal mode, 100 µs/div scale, 0 s delay, center time reference.
Other Acquire mode normal, [Run/Stop] to Run, cursors and measurements off.
Labels All custom labels that you have created in the Label Library are preserved (not
erased), but all channel labels will be set to their original names.
In the Save/Recall Menu, there are also options for restoring the complete
factory settings (see “Recalling Default Setups" on page 216) or performing
a secure erase (see “Performing a Secure Erase" on page 217).
If you see the waveform, but the square wave is not shaped correctly as
shown above, perform the procedure “Compensate Passive Probes" on
page 27.
If you do not see the waveform, make sure the probe is connected
securely to the front panel channel input BNC and to the left side,
Demo 2, Probe Comp terminal.
How AutoScale Auto Scale analyzes any waveforms present at each channel and at the
Works external trigger input. This includes the digital channels, if connected.
Auto Scale finds, turns on, and scales any channel with a repetitive
waveform that has a frequency of at least 25 Hz, a duty cycle greater than
0.5%, and an amplitude of at least 10 mV peak- to- peak. Any channels that
do not meet these requirements are turned off.
The trigger source is selected by looking for the first valid waveform
starting with external trigger, then continuing with the lowest number
analog channel up to the highest number analog channel, and finally (if
digital probes are connected) the highest number digital channel.
During Auto Scale, the delay is set to 0.0 seconds, the horizontal time/div
(sweep speed) setting is a function of the input signal (about 2 periods of
the triggered signal on the screen), and the triggering mode is set to Edge.
Perfectly compensated
Over compensated
Under compensated
1. Power switch Press once to switch power on; press again to switch power off. See
“Power-On the Oscilloscope" on page 23.
2. Softkeys The functions of these keys change based upon the menus shown on the
display directly above the keys.
The Back
Back/Up key moves up in the softkey menu hierarchy. At the
top of the hierarchy, the Back Back/Up key turns the menus off, and
oscilloscope information is shown instead.
3. [Intensity] key Press the key to illuminate it. When illuminated, turn the Entry knob to
adjust waveform intensity.
You can vary the intensity control to bring out signal detail, much like an
analog oscilloscope.
Digital channel waveform intensity is not adjustable.
More details about using the Intensity control to view signal detail are on
“To adjust waveform intensity" on page 99.
4. Entry knob The Entry knob is used to select items from menus and to change values.
The function of the Entry knob changes based upon the current menu
and softkey selections.
Note that the curved arrow symbol above the entry knob
illuminates whenever the entry knob can be used to select a value. Also,
note that when the Entry knob symbol appears on a softkey, you
can use the Entry knob, to select values.
Often, rotating the Entry knob is enough to make a selection. Sometimes,
you can push the Entry knob to enable or disable a selection. Pushing the
Entry knob also makes popup menus disappear.
9. [Default Setup] Press this key to restore the oscilloscope's default settings (details on
key “Recall the Default Oscilloscope Setup" on page 25).
10. [Auto Scale] When you press the [AutoScale] key, the oscilloscope will quickly
key determine which channels have activity, and it will turn these channels
on and scale them to display the input signals. See “Use Auto Scale" on
page 26.
11. Additional The additional waveform controls consist of:
waveform • [Math] key — provides access to math (add, subtract, etc.) waveform
controls functions. See Chapter 4, “Math Waveforms,” starting on page 63.
• [Ref] key — provides access to reference waveform functions.
Reference waveforms are saved waveforms that can be displayed and
compared against other analog channel or math waveforms. See
Chapter 5, “Reference Waveforms,” starting on page 75.
• [Digital] key — Press this key to turn the digital channels on or off
(the arrow to the left will illuminate).
When the arrow to the left of the [Digital] key is illuminated, the
upper multiplexed knob selects (and highlights in red) individual
digital channels, and the lower multiplexed knob positions the
selected digital channel.
If a trace is repositioned over an existing trace the indicator at the left
edge of the trace will change from Dn designation (where n is a one
digit channel number from 0 to 7) to D*. The "*" indicates that two
channels are overlaid.
You can rotate the upper knob to select an overlaid channel, then
rotate the lower knob to position it just as you would any other
channel.
For more information on digital channels see Chapter 6, “Digital
Channels,” starting on page 79.
• [Serial] key — This key is not currently used with the 2000 X-Series
oscilloscopes.
• Multiplexed scale knob — This scale knob is used with Math, Ref, or
Digital waveforms, whichever has the illuminated arrow to the left.
For math and reference waveforms, the scale knob acts like an analog
channel vertical scale knob.
• Multiplexed position knob — This position knob is used with Math,
Ref, or Digital waveforms, whichever has the illuminated arrow to the
left. For math and reference waveforms, the position knob acts like an
analog channel vertical position knob.
13. Waveform keys The [Acquire] key lets you select Normal, Peak Detect, Averaging, or
High Resolution acquisition modes (see “Selecting the Acquisition
Mode" on page 147) and use segmented memory (see “Acquiring to
Segmented Memory" on page 153).
The [Display] key lets you access the menu where you can enable
persistence (see “To set or clear persistence" on page 101), clear the
display, and adjust the display grid (graticule) intensity (see “To adjust
the grid intensity" on page 103).
14. File keys Press the [Save/Recall] key to save or recall a waveform or setup. See
Chapter 16, “Save/Recall (Setups, Screens, Data),” starting on page 205.
The [Print] key opens the Print Configuration Menu so you can print the
displayed waveforms. See Chapter 17, “Print (Screens),” starting on page
219.
15. [Help] key Opens the Help Menu where you can display overview help topics and
select the Language. See also “Access the Built-In Quick Help" on
page 40.
18. Demo 2, • Demo 2 terminal — This terminal outputs the Probe Comp signal
Ground, and which helps you match a probe's input capacitance to the
Demo 1 oscilloscope channel to which it is connected. See “Compensate
terminals Passive Probes" on page 27. With certain licensed features, the
oscilloscope can also output demo or training signals on this terminal.
• Ground terminal — Use the ground terminal for oscilloscope probes
connected to the Demo 1 or Demo 2 terminals.
• Demo 1 terminal — With certain licensed features, the oscilloscope
can output demo or training signals on this terminal.
19. USB Host port This port is for connecting USB mass storage devices or printers to the
oscilloscope.
Connect a USB compliant mass storage device (flash drive, disk drive,
etc.) to save or recall oscilloscope setup files and reference waveforms
or to save data and screen images. See Chapter 16, “Save/Recall
(Setups, Screens, Data),” starting on page 205.
To print, connect a USB compliant printer. For more information about
printing see Chapter 17, “Print (Screens),” starting on page 219.
You can also use the USB port to update the oscilloscope's system
software when updates are available.
You do not need to take special precautions before removing the USB
mass storage device from the oscilloscope (you do not need to "eject"
it). Simply unplug the USB mass storage device from the oscilloscope
when the file operation is complete.
6. EXT TRIG IN
connector
5. Calibration
protect
button 3. GPIB
option module
4. TRIG OUT
connector
L6GC>C</B6>CI6>C
<GDJC9ID 6KD>9
:A:8IG>8H=D8@
3. Module slot
s &%%"&'%K!*%$+%$)%%=o
s &%%"')%K!*%$+%=o
&%% LViihB6M
2. Kensington lock hole
1. Power cord connector
4. TRIG OUT Trigger output BNC connector. See “Setting the Rear Panel TRIG OUT
connector Source" on page 232.
5. Calibration See “To perform user calibration" on page 234.
protect button
6. EXT TRIG IN External trigger input BNC connector. See “External Trigger Input" on
connector page 138 for an explanation of this feature.
8. USB Device This port is for connecting the oscilloscope to a host PC. You can issue
port remote commands from a host PC to the oscilloscope via the USB device
port. See “Remote Programming with Agilent IO Libraries" on page 247.
7. USB Host port This port functions identically to the USB host port on the front panel.
USB Host Port is used for saving data from the oscilloscope and loading
software updates. See also USB Host port (see page 35).
Measurements
Menu line
Softkeys
Status line The top line of the display contains vertical, horizontal, and trigger setup
information.
Display area The display area contains the waveform acquisitions, channel identifiers, and
analog trigger, and ground level indicators. Each analog channel's information
appears in a different color.
Signal detail is displayed using 256 levels of intensity. For more information
about viewing signal detail see “To adjust waveform intensity" on page 99.
For more information about display modes see Chapter 7, “Display Settings,”
starting on page 99.
Information area The information area normally contains acquisition, analog channel,
automatic measurement, and cursor results.
Menu line This line normally contains menu name or other information associated with
the selected menu.
Softkey labels These labels describe softkey functions. Typically, softkeys let you set up
additional parameters for the selected mode or menu.
Pressing the Back Back/Up key at the top of the menu hierarchy turns off
softkey labels and displays additional status information describing channel
offset and other configuration parameters.
Quick Help
message
Quick Help remains on the screen until another key is pressed or a knob
is turned.
To select the user To select the user interface and Quick Help language:
interface and 1 Press [Help], then press the Language softkey.
Quick Help
language 2 Repeatedly press and release the Language softkey or rotate the Entry
knob until the desired language is selected.
The following languages are available: English, French, German, Italian,
Japanese, Korean, Portuguese, Russian, Simplified Chinese, Spanish, and
Traditional Chinese.
2
Horizontal Controls
To adjust the horizontal (time/div) scale 44
To adjust the horizontal delay (position) 45
Panning and Zooming Single or Stopped Acquisitions 46
To change the horizontal time mode (Normal, XY, or Roll) 47
To display the zoomed time base 50
To change the horizontal scale knob's coarse/fine adjustment setting 52
To position the time reference (left, center, right) 52
Navigating the Time Base 53
• The zoom key for quickly enabling/disabling the split- screen zoom
display.
• The [Search] key for finding events on analog channels.
• The [Navigate] keys for navigating time, search events, or segmented
memory acquisitions.
The following figure shows the Horizontal Menu which appears after
pressing the [Horiz] key.
43
2 Horizontal Controls
Sample rate
XY or Roll
mode
The Horizontal Menu lets you select the time mode (Normal, XY, or Roll),
enable Zoom, set the time base fine control (vernier), and specify the time
reference.
The current sample rate is displayed above the Fine and Time Ref softkeys.
stopped, adjusting the horizontal scale knob lets you zoom into acquired
data. See "Panning and Zooming Single or Stopped Acquisitions" on
page 46.
Note that the horizontal position knob has a different purpose in the
Zoom display. See "To display the zoomed time base" on page 50.
XY Time Mode
The XY time mode converts the oscilloscope from a volts- versus- time
display to a volts- versus- volts display using two input channels. Channel 1
is the X- axis input, channel 2 is the Y- axis input. You can use various
transducers so the display could show strain versus displacement, flow
versus pressure, volts versus current, or voltage versus frequency.
Example This exercise shows a common use of the XY display mode by measuring
the phase difference between two signals of the same frequency with the
Lissajous method.
1 Connect a sine wave signal to channel 1, and a sine wave signal of the
same frequency but out of phase to channel 2.
2 Press the [AutoScale] key, press the [Horiz] key; then, press Time Mode
and select "XY".
3 Center the signal on the display with the channel 1 and 2 position ( )
knobs. Use the channel 1 and 2 volts/div knobs and the channel 1 and
2 Fine softkeys to expand the signal for convenient viewing.
The phase difference angle (θ) can be calculated using the following
formula (assuming the amplitude is the same on both channels):
A or C
sinθ = B D
Signal must
be centered in
“ X”
D A B
1 Press the zoom key (or press the [Horiz] key and then the Zoom
softkey).
Normal
window
Signal
anomaly
expanded
in zoom
window
Zoom
window
Select
Zoom
The area of the normal display that is expanded is outlined with a box
and the rest of the normal display is ghosted. The box shows the portion
of the normal sweep that is expanded in the lower half.
To change the time/div for the Zoom window, turn the horizontal scale
(sweep speed) knob. As you turn the knob, the zoomed window time/div
is highlighted in the status line above the waveform display area. The
Horizontal scale (sweep speed) knob controls the size of the box.
The Horizontal position (delay time) knob sets the left- to- right position of
the zoom window. The delay value, which is the time displayed relative to
the trigger point) is momentarily displayed in the upper- right portion of
the display when the delay time ( ) knob is turned.
The time reference position sets the initial position of the trigger event
within acquisition memory and on the display, with delay set to 0.
Turning the Horizontal scale (sweep speed) knob expands or contracts the
waveform about the time reference point (∇). See "To adjust the horizontal
(time/div) scale" on page 44.
Turning the Horizontal position ( ) knob in Normal mode (not Zoom)
moves the trigger point indicator (▼) to the left or right of the time
reference point (∇). See "To adjust the horizontal delay (position)" on
page 45.
To navigate time
When acquisitions are stopped, you can use the navigation controls to play
through the captured data.
1 Press [Navigate].
2 In the Navigate Menu, press Navigate; then, select Time.
To navigate segments
When the segmented memory acquisition is enabled and acquisitions are
stopped, you can use the navigation controls to play through the acquired
segments.
1 Press [Navigate].
2 In the Navigate Menu, press Navigate; then, select Segments.
3
Vertical Controls
To turn waveforms on or off (channel or math) 56
To adjust the vertical scale 57
To adjust the vertical position 57
To specify channel coupling 57
To specify bandwidth limiting 58
To change the vertical scale knob's coarse/fine adjustment setting 58
To invert a waveform 59
Setting Analog Channel Probe Options 59
55
3 Vertical Controls
Channel 1
ground
level
Channel 2
ground
level
The ground level of the signal for each displayed analog channel is
identified by the position of the icon at the far- left side of the display.
1 Turn the large knob above the channel key marked to set the
vertical scale (volts/division) for the channel.
The vertical scale knob changes the analog channel scale in a 1- 2- 5 step
sequence (with a 1:1 probe attached) unless fine adjustment is enabled
(see "To change the vertical scale knob's coarse/fine adjustment setting" on
page 58).
The analog channel Volts/Div value is displayed in the status line.
The default mode for expanding the signal when you turn the
volts/division knob is vertical expansion about the ground level of the
channel; however, you can change this to expand about the center of the
display. See "To choose "expand about" center or ground" on page 229.
If the channel is DC coupled, you can quickly measure the DC component of the signal by
TIP simply noting its distance from the ground symbol.
If the channel is AC coupled, the DC component of the signal is removed, allowing you to
use greater sensitivity to display the AC component of the signal.
When Fine adjustment is selected, you can change the channel's vertical
sensitivity in smaller increments. The channel sensitivity remains fully
calibrated when Fine is on.
The vertical scale value is displayed in the status line at the top of the
display.
When Fine is turned off, turning the volts/division knob changes the
channel sensitivity in a 1- 2- 5 step sequence.
To invert a waveform
1 Press the desired channel key.
2 In the Channel Menu, press the Invert softkey to invert the selected
channel.
When Invert is selected, the voltage values of the displayed waveform are
inverted.
Invert affects how a channel is displayed. However, when using basic
triggers, the oscilloscope attempts to maintain the same trigger point by
changing trigger settings.
Inverting a channel also changes the result of any math function selected
in the Waveform Math Menu or any measurement.
The Probe Check softkey guides you through the process of compensating
passive probes (such as the N2862A/B, N2863A/B, N2889A, N2890A,
10073C, 10074C, or 1165A probes).
See Also • "To specify the channel units" on page 60
• "To specify the probe attenuation" on page 60
• "To specify the probe skew" on page 61
3 Turn the Entry knob to set the attenuation factor for the connected
probe.
When measuring voltage values, the attenuation factor can be set from
0.1:1 to 1000:1 in a 1- 2- 5 sequence.
When measuring current values with a current probe, the attenuation
factor can be set from 10 V/A to 0.001 V/A.
When specifying the attenuation factor in decibels, you can select values
from - 20 dB to 60 dB.
4
Math Waveforms
To display math waveforms 63
To perform a transform function on an arithmetic operation 64
To adjust the math waveform scale and offset 65
Add or Subtract 66
Multiply 65
FFT Measurement 67
Units for Math Waveforms 74
63
4 Math Waveforms
2 If f(t) is not already shown on the Function softkey, press the Function
sofkey and select f(t): Displayed.
3 Use the Operator softkey to select an operator.
For more information on the operators, see:
• "Add or Subtract" on page 66
• "Multiply" on page 65
• "FFT Measurement" on page 67
4 Use the Source 1 softkey to select the analog channel on which to
perform math. You can rotate the Entry knob or repetitively press the
Source 1 softkey to make your selection. If you choose a transform
function (FFT) the result is displayed.
5 If you select an arithmetic operator, use the Source 2 softkey to select
the second source for the arithmetic operation. The result is displayed.
6 To re- size and re- position the math waveform, see "To adjust the math
waveform scale and offset" on page 65.
Multiply
When you select the multiply math function, the Source 1 and Source 2
values are multiplied point by point, and the result is displayed.
Multiply is useful for seeing power relationships when one of the channels
is proportional to the current.
Add or Subtract
When you select add or subtract, the Source 1 and Source 2 values are
added or subtracted point by point, and the result is displayed.
You can use subtract to make a differential measurement or to compare
two waveforms.
If your waveforms' DC offsets are larger than the dynamic range of the
oscilloscope's input channels you will need to use a differential probe
instead.
FFT Measurement
FFT is used to compute the fast Fourier transform using analog input
channels or an arithmetic operation g(t). FFT takes the digitized time
record of the specified source and transforms it to the frequency domain.
When the FFT function is selected, the FFT spectrum is plotted on the
oscilloscope display as magnitude in dBV versus frequency. The readout
for the horizontal axis changes from time to frequency (Hertz) and the
vertical readout changes from volts to dB.
Use the FFT function to find crosstalk problems, to find distortion
problems in analog waveforms caused by amplifier non- linearity, or for
adjusting analog filters.
To display a FFT waveform:
1 Press the [Math] key, press the Function softkey and select f(t), press the
Operator softkey and select FFT.
• Source 1 — selects the source for the FFT. (See "To perform a
transform function on an arithmetic operation" on page 64 for
information about using g(t) as the source.)
• Span — sets the overall width of the FFT spectrum that you see on
the display (left to right). Divide span by 10 to calculate the number
of Hertz per division. It is possible to set Span above the maximum
available frequency, in which case the displayed spectrum will not
take up the whole screen. Press the Span softkey, then turn the Entry
knob to set the desired frequency span of the display.
• Center — sets the FFT spectrum frequency represented at the center
vertical grid line of the display. It is possible to set the Center to
values below half the span or above the maximum available
frequency, in which case the displayed spectrum will not take up the
whole screen. Press the Center softkey, then turn the Entry knob to
set the desired center frequency of the display.
• Scale — lets you set your own vertical scale factors for FFT expressed
in dB/div (decibels/division). See "To adjust the math waveform scale
and offset" on page 65.
• Offset — lets you set your own offset for the FFT. The offset value is
in dB and is represented by the center horizontal grid line of the
display. See "To adjust the math waveform scale and offset" on
page 65.
• More FFT — displays the More FFT Settings Menu.
2 Press the More FFT softkey to display additional FFT settings.
3 To make cursor measurements, press the [Cursors] key and set the
Source softkey to Math: f(t).
Use the X1 and X2 cursors to measure frequency values and difference
between two frequency values (∆X). Use the Y1 and Y2 cursors to
measure amplitude in dB and difference in amplitude (∆Y).
4 To make other measurements, press the [Meas] key and set the Source
softkey to Math: f(t).
You can make peak- to- peak, maximum, minimum, and average dB
measurements on the FFT waveform. You can also find the frequency
value at the first occurrence of the waveform maximum by using the X
at Max Y measurement.
The following FFT spectrum was obtained by connecting a 4 V, 75 kHz
square wave to channel 1. Set the horizontal scale to 50 µs/div, vertical
sensitivity to 1 V/div, Units/div to 20 dBV, Offset to - 60.0 dBV, Center
frequency to 250 kHz, frequency Span to 500 kHz, and window to
Hanning.
FFT Resolution
NOTE
The FFT resolution is the quotient of the sampling rate and the number of FFT points (fS/N).
With a fixed number of FFT points (up to 65,536), the lower the sampling rate, the better the
resolution.
FFT Units
0 dBV is the amplitude of a 1 Vrms sinusoid. When the FFT source is
channel 1 or channel 2 (or channel 3 or 4 on 4- channel models), FFT
units will be displayed in dBV when channel units is set to Volts and
channel impedance is set to 1 MΩ.
FFT units will be displayed in dBm when channel units is set to Volts and
channel impedance is set to 50Ω.
FFT units will be displayed as dB for all other FFT sources or when a
source channel's units has been set to Amps.
FFT DC Value
The FFT computation produces a DC value that is incorrect. It does not
take the offset at center screen into account. The DC value is not
corrected in order to accurately represent frequency components near DC.
FFT Aliasing
When using FFTs, it is important to be aware of frequency aliasing. This
requires that the operator have some knowledge as to what the frequency
domain should contain, and also consider the sampling rate, frequency
span, and oscilloscope vertical bandwidth when making FFT
measurements. The FFT resolution (the quotient of the sampling rate and
the number of FFT points) is displayed directly above the softkeys when
the FFT Menu is displayed.
Figure 7 Aliasing
Because the frequency span goes from ≈ 0 to the Nyquist frequency, the
best way to prevent aliasing is to make sure that the frequency span is
greater than the frequencies of significant energy present in the input
signal.
add or subtract V or A
multiply V2, A2, or W (Volt-Amp)
FFT dB* (decibels). See also "FFT Units" on
page 72.
* When the FFT source is channel 1, 2, 3 or 4, FFT units will be displayed in dBV when channel
units is set to Volts and channel impedance is set to 1 MΩ. FFT units will be displayed in dBm
when channel units is set to Volts and channel impedance is set to 50Ω. FFT units will be
displayed as dB for all other FFT sources or when a source channel's units has been set to Amps.
5
Reference Waveforms
To save a waveform to a reference waveform location 75
To display a reference waveform 76
To scale and position reference waveforms 77
To adjust reference waveform skew 77
To display reference waveform information 78
To save/recall reference waveform files to/from a USB storage
device 78
75
5 Reference Waveforms
2 In the Reference Waveform Menu, press the Ref softkey and turn the
Entry knob to select the desired reference waveform location.
3 Press the Source softkey and turn the Entry knob to select the source
waveform.
4 Press the Save to R1/R2 softkey to save the waveform to the reference
waveform location.
Reference waveforms are non-volatile — they remain after power cycling or performing a
NOTE default setup.
6
Digital Channels
To connect the digital probes to the device under test 79
Acquiring waveforms using the digital channels 83
To display digital channels using AutoScale 83
Interpreting the digital waveform display 84
To switch all digital channels on or off 86
To switch groups of channels on or off 86
To switch a single channel on or off 86
To change the displayed size of the digital channels 85
To reposition a digital channel 87
To change the logic threshold for digital channels 86
To display digital channels as a bus 88
Digital channel signal fidelity: Probe impedance and grounding 91
To replace digital probe leads 96
This chapter describes how to use the digital channels of a Mixed- Signal
Oscilloscope (MSO).
The digital channels are enabled on MSOX2000 X- Series models and
DSOX2000 X- Series models that have the DSOX2MSO upgrade license
installed.
79
6 Digital Channels
Turning off power to the device under test would only prevent damage
that might occur if you accidentally short two lines together while
connecting probes. You can leave the oscilloscope powered on because
no voltage appears at the probes.
Off
3 Connect the ground lead on each set of channels (each pod), using a
probe grabber. The ground lead improves signal fidelity to the
oscilloscope, ensuring accurate measurements.
Channel
Pod Ground
Circuit
Ground
4 Connect a grabber to one of the probe leads. (Other probe leads are
omitted from the figure for clarity.)
Grabber
6 For high- speed signals, connect a ground lead to the probe lead,
connect a grabber to the ground lead, and attach the grabber to ground
in the device under test.
Signal
Ground
Grabber
7 Repeat these steps until you have connected all points of interest.
Signals
Ground
Any digital channel with an active signal will be displayed. Any digital
channels without active signals will be turned off.
• To undo the effects of AutoScale, press the Undo AutoScale softkey before
pressing any other key.
This is useful if you have unintentionally pressed the [AutoScale] key or do
not like the settings AutoScale has selected. This will return the
oscilloscope to its previous settings. See also: "How AutoScale Works" on
page 27.
To set the instrument to the factory- default configuration, press the
[Default Setup] key.
Digital Activity
channel indicators
identifiers
Activity indicator When any digital channels are turned on, an activity indicator is displayed
in the status line at the bottom of the display. A digital channel can be
always high ( ), always low ( ), or actively toggling logic states ( ). Any
channel that is turned off will be grayed out in the activity indicator.
2 Press the size ( ) softkey to select how the digital channels are
displayed.
The sizing control lets you spread out or compress the digital traces
vertically on the display for more convenient viewing.
3 Press the D7 - D0 softkey, then select a logic family preset or select User
to define your own threshold.
CMOS +2.5 V
ECL –1.3 V
User Variable from –8 V to +8 V
The threshold you set applies to all channels within the selected D7 - D0
group. Each of the two channel groups can be set to a different threshold
if desired.
Values greater than the set threshold are high (1) and values less than the
set threshold are low (0).
If the Thresholds softkey is set to User, press the User softkey for the
channel group, then turn the Entry knob to set the logic threshold. There
is one User softkey for each group of channels.
Next, select a bus. Rotate the Entry knob, then press the Entry knob or
the Bus1/Bus2 softkey to switch it on.
Use the Channel softkey and the Entry knob to select individual channels
to be included in the bus. You can rotate the Entry knob and push it or
push the softkey to select channels. You can also press the Select/Deselect
D7-D0 softkey to include or exclude groups of eight channels in each bus.
If the bus display is blank, completely white, or if the display includes "...",
you need to expand the horizontal scale to allow space for the data to be
shown, or use the cursors to display the values (see "Using cursors to read
bus values" on page 89).
The Base softkey lets you choose to display the bus values in hex or
binary.
The buses are shown at the bottom of the display.
X1 cursor
X2 cursor
Bus values
When you press the [Digital] key to display the Digital Channel Menu, the
digital activity indicator is shown where the cursor values were and the
bus values at the cursors are displayed in the graticule.
Bus values are The bus values are also displayed when using the Pattern trigger function.
displayed when Press the [Pattern] key on the front panel to display the Pattern Trigger
using Pattern Menu and the bus values will be displayed on the right, above the
trigger softkeys.
The dollar sign ($) will be displayed in the bus value when the bus value
cannot be displayed as a hex value. This occurs when one or more "don't
cares" (X) are combined with low (0) and high (1) logic levels in the
pattern specification, or when a transition indicator — rising edge ( ) or
falling edge ( ) — are included in the pattern specification. A byte that
consists of all don't cares (X) will be displayed in the bus as a don't care
(X).
Input Impedance
The logic probes are passive probes, which offer high input impedance and
high bandwidths. They usually provide some attenuation of the signal to
the oscilloscope, typically 20 dB.
Passive probe input impedance is generally specified in terms of a parallel
capacitance and resistance. The resistance is the sum of the tip resistor
value and the input resistance of the test instrument (see the following
figure). The capacitance is the series combination of the tip compensating
capacitor and the cable, plus instrument capacitance in parallel with the
stray tip capacitance to ground. While this results in an input impedance
specification that is an accurate model for DC and low frequencies, the
high- frequency model of the probe input is more useful (see the following
figure). This high- frequency model takes into account pure tip capacitance
to ground as well as series tip resistance, and the cable's characteristic
impedance (Zo).
The impedance plots for the two models are shown in these figures. By
comparing the two plots, you can see that both the series tip resistor and
the cable's characteristic impedance extend the input impedance
significantly. The stray tip capacitance, which is generally small (1 pF),
sets the final break point on the impedance chart.
100 k
10 k High
Frequency
Model
Impedance
1k
Typical
100 Model
10
1
10 kHz 100 kHz 1 MHz 10 MHz 100 MHz 1 GHz
Frequency
The logic probes are represented by the high- frequency circuit model
shown above. They are designed to provide as much series tip resistance
as possible. Stray tip capacitance to ground is minimized by the proper
mechanical design of the probe tip assembly. This provides the maximum
input impedance at high frequencies.
Probe Grounding
A probe ground is the low- impedance path for current to return to the
source from the probe. Increased length in this path will, at high
frequencies, create large common mode voltages at the probe input. The
voltage generated behaves as if this path were an inductor according to
the equation:
di
V = L dt
Probe 1 Z in
i1
L (GND)
i2 +iÊ n
Probe 2 Z in
in
Probe N Z in
7
Display Settings
To adjust waveform intensity 99
To set or clear persistence 101
To clear the display 102
To select the grid type 102
To adjust the grid intensity 103
To freeze the display 103
99
7 Display Settings
3 To erase the results of previous acquisitions from the display, press the
Clear Persistence softkey.
The oscilloscope will start to accumulate acquisitions again.
4 To return the oscilloscope to the normal display mode, turn off
persistence; then, press the Clear Persistence softkey.
Turning off persistence does not clear the display. The display is
cleared if you press the Clear Display softkey or if you press the
[AutoScale] key (which also turns off persistence).
For another method of seeing worst- case extremes of varying waveforms,
see "Glitch or Narrow Pulse Capture" on page 149.
1 Press [Display].
2 Press the Grid softkey; then, turn the Entry knob to select the grid
type.
2 Press the Intensity softkey; then, turn the Entry knob to change the
intensity of the displayed grid.
The intensity level is shown in the Intensity softkey and is adjustable from
0 to 100%.
Each major vertical division in the grid corresponds to the vertical
sensitivity shown in the status line at the top of the display.
Each major horizontal division in the grid corresponds to the time/div
shown in the status line at the top of the display.
8
Labels
To turn the label display on or off 105
To assign a predefined label to a channel 106
To define a new label 107
To load a list of labels from a text file you create 108
To reset the label library to the factory default 109
You can define labels and assign them to each analog input channel, or
you can turn labels off to increase the waveform display area. Labels can
also be applied to digital channels on MSO models.
105
8 Labels
The figure above shows the list of channels and their default labels. The
channel does not have to be turned on to have a label assigned to it.
3 Press the Library softkey, then turn the Entry knob or successively press
the Library softkey to select a predefined label from the library.
4 Press the Apply New Label softkey to assign the label to your selected
channel.
5 Repeat the above procedure for each predefined label you want to
assign to a channel.
Turning the Entry knob selects a character to enter into the highlighted
position shown in the "New label =" line above the softkeys and in the
Spell softkey. Labels can be up to ten characters in length.
4 Press the Enter softkey to enter the selected character and to go to the
next character position.
5 You may position the highlight on any character in the label name by
successively pressing the Enter softkey.
6 To delete a character from the label, press the Enter softkey until the
letter you want to delete is highlighted, then press the Delete Character
softkey.
7 When you are done entering characters for the label, press the
Apply New Label softkey to assign the label to the selected channel.
When you define a new label, it is added to the nonvolatile label list.
Label Assignment When you assign a label ending in a digit, such as ADDR0 or DATA0, the
Auto-Increment oscilloscope automatically increments the digit and displays the modified
label in the "New label" field after you press the Apply New Label softkey.
Therefore, you only need to select a new channel and press the Apply New
Label softkey again to assign the label to the channel. Only the original
label is saved in the label list. This feature makes it easier to assign
successive labels to numbered control lines and data bus lines.
3 Load the list into the oscilloscope using the File Explorer (press [Utility]
> File Explorer).
Pressing the Default Library softkey will remove all user-defined labels from the library and
NOTE set the labels back to the factory default. Once deleted, these user-defined labels cannot be
recovered.
9
Triggers
Adjusting the Trigger Level 112
Forcing a Trigger 113
Edge Trigger 113
Pattern Trigger 116
Pulse Width Trigger 119
Video Trigger 121
A trigger setup tells the oscilloscope when to acquire and display data.
For example, you can set up to trigger on the rising edge of the analog
channel 1 input signal.
You can adjust the vertical level used for analog channel edge detection by
turning the Trigger Level knob.
In addition to the edge trigger type, you can also set up triggers on pulse
widths, patterns, and video signals.
You can use any input channel or the "External Trigger Input" on
page 138 BNC as the source for most trigger types.
Changes to the trigger setup are applied immediately. If the oscilloscope is
stopped when you change a trigger setup, the oscilloscope uses the new
specification when you press [Run/Stop] or [Single]. If the oscilloscope is
running when you change a trigger setup, it uses the new trigger
definition when it starts the next acquisition.
You can use the [Force Trigger] key to acquire and display data when
triggers are not occurring.
You can use the [Mode/Coupling] key to set options that affect all trigger
types (see Chapter 10, “Trigger Mode/Coupling,” starting on page 133).
111
9 Triggers
You can save trigger setups along with the oscilloscope setup (see
Chapter 16, “Save/Recall (Setups, Screens, Data),” starting on page 205).
Triggers - General A triggered waveform is one in which the oscilloscope begins tracing
Information (displaying) the waveform, from the left side of the display to the right,
each time a particular trigger condition is met. This provides stable
display of periodic signals such as sine waves and square waves, as well
as nonperiodic signals such as serial data streams.
The figure below shows the conceptual representation of acquisition
memory. You can think of the trigger event as dividing acquisition memory
into a pre- trigger and post- trigger buffer. The position of the trigger event
in acquisition memory is defined by the time reference point and the
delay (horizontal position) setting (see "To adjust the horizontal delay
(position)" on page 45).
Trigger Event
Acquisition Memory
The trigger level for a selected digital channel is set using the threshold
menu in the Digital Channel Menu. Press the [Digital] key on the front
panel, then press the Thresholds softkey to set the threshold level (TTL,
CMOS, ECL, or user defined) for the selected digital channel group. The
threshold value is displayed in the upper- right corner of the display.
The line trigger level is not adjustable. This trigger is synchronized with
the power line supplied to the oscilloscope.
You can also change the trigger level of all channels by pressing [Analyze] > Features and
NOTE then selecting Trigger Levels.
Forcing a Trigger
The [Force Trigger] key causes a trigger (on anything) and displays the
acquisition.
This key is useful in the Normal trigger mode where acquisitions are made
only when the trigger condition is met. In this mode, if no triggers are
occurring (that is, the "Trig'd?" indicator is displayed), you can press [Force
Trigger] to force a trigger and see what the input signals look like.
In the Auto trigger mode, when the trigger condition is not met, triggers
are forced and the "Auto?" indicator is displayed.
Edge Trigger
The Edge trigger type identifies a trigger by looking for a specified edge
(slope) and voltage level on a waveform. You can define the trigger source
and slope in this menu. The slope can be set to rising edge or falling edge,
and can be set to alternating edges or either edge on all sources except
Line. The trigger type, source, and level are displayed in the upper- right
corner of the display.
1 On the front panel, in the Trigger section, press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey, and use the Entry knob
to select Edge.
3 Select the trigger source:
Alternating edge mode is useful when you want to trigger on both edges of a clock (for
NOTE example, DDR signals).
Either edge mode is useful when you want to trigger on any activity of a selected source.
All modes operate up to the bandwidth of the oscilloscope except Either edge mode, which
has a limitation. Either edge mode will trigger on Constant Wave signals up to 100 MHz, but
can trigger on isolated pulses down to 1/(2*oscilloscope's bandwidth).
Using AutoScale The easiest way to set up an Edge trigger on a waveform is to use
to Set Up Edge AutoScale. Simply press the [AutoScale] key and the oscilloscope will
Triggers attempt to trigger on the waveform using a simple Edge trigger type. See
"Use Auto Scale" on page 26.
Pattern Trigger
The Pattern trigger identifies a trigger condition by looking for a specified
pattern. This pattern is a logical AND combination of the channels. Each
channel can have a value of 0 (low), 1 (high), or don't care (X). A rising
or falling edge can be specified for one channel included in the pattern.
You can also trigger on a hex bus value as described on "Hex Bus Pattern
Trigger" on page 118.
1 Press the [Trigger] key.
2 In the Trigger Menu, press the Trigger softkey; then, turn the Entry knob
to select Pattern.
3 For each analog or digital channel you want to include in the desired
pattern, press the Channel softkey to select the channel.
This is the channel source for the 0, 1, X, or edge condition. As you
press the Channel softkey (or rotate the Entry knob), the channel you
select is highlighted in the Pattern = line directly above the softkeys
and in the upper- right corner of the display next to "Pat".
Adjust the trigger level for the selected analog channel by turning the
Trigger Level knob. Press the [Digital] key and select Thresholds to set
the threshold level for digital channels. The value of the trigger level or
digital threshold is displayed in the upper- right corner of the display.
4 For each channel you select, press the Pattern softkey; then, turn the
Entry knob to set the condition for that channel in the pattern.
If a digit is made up of less than four bits, then the value of the digit will be limited to the
NOTE value that can be created by the selected bits.
5 You can use the Set all Digits softkey to set all digits to a particular
value.
When a hex bus digit contains one or more don't care (X) bits and one or
more bit with a value or 0 or 1, the "$" sign will be displayed for the digit.
For information regarding digital bus display when Pattern triggering see
"Bus values are displayed when using Pattern trigger" on page 90.
3 Press the Source softkey; then, rotate the Entry knob to select a channel
source for the trigger.
The channel you select is shown in the upper- right corner of the
display next to the polarity symbol.
The source can be any analog or digital channel available on your
oscilloscope.
4 Adjust the trigger level:
• For analog channels, turn the Trigger Level knob.
• For digital channels, press the [Digital] key and select Thresholds to set
the threshold level.
The value of the trigger level or digital threshold is displayed in the
upper- right corner of the display.
10 ns 10 ns Trigger
10 ns 10 ns Trigger
10 ns 15 ns 12 ns Trigger
7 Select the qualifier time set softkey (< or >), then rotate the Entry
knob to set the pulse width qualifier time.
The qualifiers can be set as follows:
• 2 ns to 10 s for > or < qualifier (5 ns to 10 s for 350 MHz
bandwidth models).
• 10 ns to 10 s for >< qualifier, with minimum difference of 5 ns
between upper and lower settings.
Pulse width • When the less than (<) qualifier is selected, the Entry knob sets the
trigger < qualifier oscilloscope to trigger on a pulse width less than the time value
time set softkey displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry knob sets the
upper time range value.
Pulse width • When the greater than (>) qualifier is selected, the Entry knob sets the
trigger > qualifier oscilloscope to trigger on a pulse width greater than the time value
time set softkey displayed on the softkey.
• When the time range (><) qualifier is selected, the Entry knob sets the
lower time range value.
Video Trigger
Video triggering can be used to capture the complicated waveforms of
most standard analog video signals. The trigger circuitry detects the
vertical and horizontal interval of the waveform and produces triggers
based on the video trigger settings you have selected.
The oscilloscope's MegaZoom IV technology gives you bright, easily viewed
displays of any part of the video waveform. Analysis of video waveforms is
simplified by the oscilloscope's ability to trigger on any selected line of the
video signal.
It is important, when using a 10:1 passive probe, that the probe is correctly compensated.
NOTE The oscilloscope is sensitive to this and will not trigger if the probe is not properly
compensated, especially for progressive formats.
3 Press the Source softkey and select any analog channel as the video
trigger source.
The selected trigger source is displayed in the upper- right corner of the
display. Turning the Trigger Level knob does not change the trigger level
because the trigger level is automatically set to the sync pulse. Trigger
coupling is automatically set to TV in the Trigger Mode and Coupling
Menu.
4 Press the sync polarity softkey to set the Video trigger to either positive
( ) or negative ( ) sync polarity.
5 Press the Settings softkey.
6 In the Video Trigger Menu, press the Standard softkey to set the video
standard.
The oscilloscope supports triggering on the following television (TV) and
video standards.
7 Press the Auto Setup softkey to automatically set up the oscilloscope for
the selected Source and Standard:
• Source channel vertical scaling is set to 140 mV/div.
• Source channel offset is set to 245 mV.
• Source channel is turned on.
• Trigger type is set to Video.
• Video trigger mode is set to All Lines.
• Display Grid type is set to IRE (when Standard is NTSC) or mV (see "To
select the grid type" on page 102).
Video Triggering The following are exercises to familiarize you with video triggering. These
Examples exercises use the NTSC video standard.
• "To trigger on a specific line of video" on page 125
Alternate Triggering
NOTE
If Line:Alternate is selected, the oscilloscope will alternately trigger on the selected line
number in Field 1 and Field 2. This is a quick way to compare the Field 1 VITS and Field 2
VITS or to check for the correct insertion of the half line at the end of Field 1.
starts on Line 4. In the case of NTSC video, the oscilloscope will trigger
on color field 1 alternating with color field 3 (see the following figure).
This setup can be used to measure the envelope of the reference burst.
If a more detailed analysis is required, then only one color field should be
selected to be the trigger. You can do this by using the Field Holdoff softkey
in the Video Trigger Menu. Press the Field Holdoff softkey and use the Entry
knob to adjust the holdoff in half- field increments until the oscilloscope
triggers on only one phase of the color burst.
A quick way to synchronize to the other phase is to briefly disconnect the
signal and then reconnect it. Repeat until the correct phase is displayed.
When holdoff is adjusted using the Field Holdoff softkey and the Entry knob,
the corresponding holdoff time will be displayed in the Trigger Mode and
Coupling Menu.
Standard Time
NTSC 8.35 ms
PAL 10 ms
PAL-M 10 ms
SECAM 10 ms
10
Trigger Mode/Coupling
To select the Auto or Normal trigger mode 134
To select the trigger coupling 136
To enable or disable trigger noise rejection 137
To enable or disable trigger HF Reject 137
To set the trigger holdoff 138
External Trigger Input 138
Noisy Signals If the signal you are probing is noisy, you can set up the oscilloscope to
reduce the noise in the trigger path and on the displayed waveform. First,
stabilize the displayed waveform by removing the noise from the trigger
path. Second, reduce the noise on the displayed waveform.
1 Connect a signal to the oscilloscope and obtain a stable display.
2 Remove the noise from the trigger path by turning on high- frequency
rejection ("To enable or disable trigger HF Reject" on page 137),
low- frequency rejection ("To select the trigger coupling" on page 136),
or "To enable or disable trigger noise rejection" on page 137.
3 Use "Averaging Acquisition Mode" on page 151 to reduce noise on the
displayed waveform.
133
10 Trigger Mode/Coupling
Trigger Indicator The trigger indicator at the top right of the display shows whether triggers
are occurring.
In the Auto trigger mode, the trigger indicator can show:
• Auto? (flashing) — the trigger condition is not found (after the
pre- trigger buffer has filled), and forced triggers and acquisitions are
occurring.
• Auto (not flashing) — the trigger condition is found (or the pre- trigger
buffer is being filled).
In the Normal trigger mode, the trigger indicator can show:
• Trig'd? (flashing) — the trigger condition is not found (after the
pre- trigger buffer has filled), and no acquisitions are occurring.
• Trig'd (not flashing) — trigger condition is found (or pre- trigger buffer is
being filled).
When the oscilloscope is not running, the trigger indicator area shows
Stop.
When to Use Auto The Auto trigger mode is appropriate when:
Trigger Mode • Checking DC signals or signals with unknown levels or activity.
• When trigger conditions occur often enough that forced triggers are
unnecessary.
When to Use The Normal trigger mode is appropriate when:
Normal Trigger • You only want to acquire specific events specified by the trigger
Mode settings.
• Triggering on an infrequent signal from a serial bus (for example, I2C,
SPI, CAN, LIN, etc.) or another signal that arrives in bursts. The Normal
trigger mode lets you stabilize the display by preventing the oscilloscope
from auto- triggering.
• Making single- shot acquisitions with the [Single] key.
Often with single- shot acquisitions, you must initiate some action in the
device under test, and you don't want the oscilloscope to auto- trigger
before that happens. Before initiating the action in the circuit, wait for
the trigger condition indicator Trig'd? to flash (this tells you the
pre- trigger buffer is filled).
See Also • "Forcing a Trigger" on page 113
• "To set the trigger holdoff" on page 138
0 dB
3 dB down point
Pass
Band
DC 50 kHz
0 dB
-3 dB Pass
Band
DC 50 kHz
200 ns 600 ns
The external trigger input impedance is 1M Ohm. This lets you use passive
probes for general- purpose measurements. The higher impedance
minimizes the loading effect of the oscilloscope on the device under test.
To set the EXT TRIG IN units and probe attenuation:
1 Press the [Mode/Coupling] key in the Trigger section of the front panel.
2 In the Trigger Mode and Coupling Menu, press the External softkey.
3 In the External Trigger Menu, press the Units softkey to select between:
• Volts — for a voltage probe.
• Amps — for a current probe.
Measurement results, channel sensitivity, and the trigger level will
reflect the measurement units you have selected.
4 Press the Probe softkey; then, turn the entry knob to specify the probe
attenuation.
The attenuation factor can be set from 0.1:1 to 1000:1 in a 1- 2- 5
sequence.
The probe attenuation factor must be set properly for measurements to
be made correctly.
11
Acquisition Control
Running, Stopping, and Making Single Acquisitions (Run Control) 141
Overview of Sampling 143
Selecting the Acquisition Mode 147
Acquiring to Segmented Memory 153
This chapter shows how to use the oscilloscope's acquisition and run
controls.
141
11 Acquisition Control
When you press [Single], the display is cleared, the trigger mode is
temporarily set to Normal (to keep the oscilloscope from auto- triggering
immediately), the trigger circuitry is armed, the [Single] key is
illuminated, and the oscilloscope waits until a trigger condition occurs
before it displays a waveform.
When the oscilloscope triggers, the single acquisition is displayed and
the oscilloscope is stopped (the [Run/Stop] key is illuminated in red).
Press [Single] again to acquire another waveform.
If the oscilloscope doesn't trigger, you can press the [Force Trigger] key to
trigger on anything and make a single acquisition.
To display the results of multiple acquisitions, use persistence. See "To set
or clear persistence" on page 101.
Single vs. The maximum data record length is greater for a single acquisition than
Running and when the oscilloscope is running (or when the oscilloscope is stopped
Record Length after running):
• Single — Single acquisitions always use the maximum memory available
— at least twice as much memory as acquisitions captured when
running — and the oscilloscope stores at least twice as many samples.
At slower time/div settings, because there is more memory available for
a single acquisition, the acquisition has a higher effective sample rate.
• Running — When running (versus taking a single acquisition), the
memory is divided in half. This lets the acquisition system acquire one
record while processing the previous acquisition, dramatically improving
the number of waveforms per second processed by the oscilloscope.
When running, a high waveform update rate provides the best
representation of your input signal.
To acquire data with the longest possible record length, press the [Single]
key.
For more information on settings that affect record length, see "Length
Control" on page 211.
Overview of Sampling
To understand the oscilloscope's sampling and acquisition modes, it is
helpful to understand sampling theory, aliasing, oscilloscope bandwidth
and sample rate, oscilloscope rise time, oscilloscope bandwidth required,
and how memory depth affects sample rate.
Sampling Theory
The Nyquist sampling theorem states that for a limited bandwidth
(band- limited) signal with maximum frequency fMAX, the equally spaced
sampling frequency fS must be greater than twice the maximum frequency
fMAX, in order to have the signal be uniquely reconstructed without
aliasing.
fMAX = fS/2 = Nyquist frequency (fN) = folding frequency
Aliasing
Aliasing occurs when signals are under- sampled (fS < 2fMAX). Aliasing is
the signal distortion caused by low frequencies falsely reconstructed from
an insufficient number of sample points.
Figure 21 Aliasing
%Y7
"(Y7
6iiZcjVi^dc
[C [H
;gZfjZcXn
%Y7
"(Y7
6iiZcjVi^dc
6a^VhZY[gZfjZcXn
XdbedcZcih
[H$) [C [H
;gZfjZcXn
An oscilloscope's rise time is not the fastest edge speed that the
oscilloscope can accurately measure. It is the fastest edge speed the
oscilloscope can possibly produce.
See Also Choosing an Oscilloscope with the Right Bandwidth for your
Application, Agilent Application Note 1588
("http://cp.literature.agilent.com/litweb/pdf/5989- 5733EN.pdf")
a Press the zoom key (or press the [Horiz] key and then the Zoom
softkey).
b To obtain a better resolution of the glitch, expand the time base.
Use the horizontal position knob ( ) to pan through the waveform to set
the expanded portion of the normal window around the glitch.
4 9
16 10
64 11
≥ 256 12
The higher the number of averages, the slower the displayed waveform
responds to waveform changes. You must compromise between how quickly
the waveform responds to changes and how much you want to reduce the
displayed noise on the signal.
To use the Averaging mode:
1 Press the [Acquire] key, then press the Acq Mode softkey until the
Averaging mode is selected.
2 Press the #Avgs softkey and turn the Entry knob to set the number of
averages that best eliminates the noise from the displayed waveform.
The number of acquisitions being averaged is displayed in the # Avgs
softkey.
Displayed Sample Rate (sr, per Displayed Sample Rate (sr, Bits of
channel, 1 Gsa/s max) interleaved, 2 Gsa/s max) resolution
62.5 MSa/s < sr ≤ 250 MSa/s 125 MSa/s < sr ≤ 500 MSa/s 9
Progress indicator
Sample rate
Navigating Segments
1 Press the Current Seg softkey and turn the Entry knob to display the
desired segment along with a time tag indicating the time from the first
trigger event.
You can also navigate segments using the [Navigate] key and controls. See
"To navigate segments" on page 53.
12
Cursors
To make cursor measurements 158
Cursor Examples 161
Cursors are horizontal and vertical markers that indicate X- axis values
and Y- axis values on a selected waveform source. You can use cursors to
make custom voltage, time, phase, or ratio measurements on oscilloscope
signals.
Cursor information is displayed in the right- side information area.
Cursors are not always limited to the visible display. If you set a cursor,
then pan and zoom the waveform until the cursor is off screen, its value
will not be changed. It will still be there when you return to its original
location.
X Cursors X cursors are vertical dashed lines that adjust horizontally and can be
used to measure time (s), frequency (1/s), phase (°), and ratio (%).
The X1 cursor is the short- dashed vertical line, and the X2 cursor is the
long- dashed vertical line.
When used with the FFT math function as a source, the X cursors
indicate frequency.
In XY horizontal mode, the X cursors display channel 1 values (Volts or
Amps).
The X1 and X2 cursor values for the selected waveform source are
displayed in the softkey menu area.
The difference between X1 and X2 (∆X) and 1/∆X are displayed in the
Cursors box in the right- side information area.
157
12 Cursors
Y Cursors Y cursors are horizontal dashed lines that adjust vertically and can be
used to measure Volts or Amps, dependent on the channel Probe Units
setting, or they can measure ratios (%). When math functions are used as
a source, the measurement units correspond to that math function.
The Y1 cursor is the short- dashed horizontal line and the Y2 cursor is the
long- dashed horizontal line.
The Y cursors adjust vertically and typically indicate values relative to the
waveform's ground point, except math FFT where the values are relative
to 0 dB.
In XY horizontal mode, the Y cursors display channel 2 values (Volts or
Amps).
When active, the Y1 and Y2 cursor values for the selected waveform
source are displayed in the softkey menu area.
The difference between Y1 and Y2 (∆Y) is displayed in the Cursors box in
the right- side information area.
Manual and Track Waveform modes can be used on waveforms that are
displayed on the analog input channels (including math functions).
Binary and Hex modes apply to digital signals (of MSO oscilloscope
models).
In Hex and Binary modes, a level can be displayed as 1 (higher than
trigger level), 0 (lower than trigger level), indeterminate state ( ), or X
(don't care).
In Binary mode, X is displayed if the channel is turned off.
In Hex mode, the channel is interpreted as 0 if turned off.
4 Press Source (or X1 Source, X2 Source in the Track Waveform mode); then,
select the input source for cursor values.
5 Select the cursor(s) to be adjusted:
• Push the Cursors knob; then, turn the Cursors knob. To finalize your
selection, either push the Cursors knob again or wait about five
seconds for the popup menu to disappear.
Or:
Cursor Examples
Figure 28 Cursors used to measure pulse widths other than middle threshold points
Expand the display with Zoom mode, then characterize the event of
interest with the cursors.
Put the X1 cursor on one side of a pulse and the X2 cursor on the other
side of the pulse.
Press the X1 X2 linked softkey and move the cursors together to check for
pulse width variations in a pulse train.
13
Measurements
To make automatic measurements 168
Measurements Summary 169
Voltage Measurements 171
Time Measurements 179
Measurement Thresholds 184
Measurement Window with Zoom Display 186
167
13 Measurements
2 Press the Source softkey to select the channel, running math function,
or reference waveform to be measured.
Only channels, math functions, or reference waveforms that are
displayed are available for measurements.
If a portion of the waveform required for a measurement is not
displayed or does not display enough resolution to make the
measurement, the result will display "No Edges", "Clipped", "Low Signal",
"< value", or "> value", or a similar message to indicate that the
measurement may not be reliable.
3 Press the Type softkey then rotate the Entry knob to select a
measurement to be made.
Measurements Summary
The automatic measurements provided by the oscilloscope are listed in the
following table. All measurements are available for analog channel
waveforms. All measurements except Counter are available for math
waveforms other than FFT. A limited set of measurements is available for
math FFT waveforms and for digital channel waveforms (as described in
the following table).
Snapshot All
The Snapshot All measurement type displays a popup containing a
snapshot of all the single waveform measurements.
You can also configure the [Quick Action] key to display the Snapshot All
popup. See "Configuring the [Quick Action] Key" on page 238.
Voltage Measurements
The following figure shows the voltage measurement points.
Maximum Top
Amplitude Peak-Peak
Base
Minimum
Measurement units for each input channel can be set to Volts or Amps
using the channel Probe Units softkey. See "To specify the channel units" on
page 60.
Peak-Peak
The peak- to- peak value is the difference between Maximum and Minimum
values. The Y cursors show the values being measured.
Maximum
Maximum is the highest value in the waveform display. The Y cursor
shows the value being measured.
Minimum
Minimum is the lowest value in the waveform display. The Y cursor shows
the value being measured.
Amplitude
The Amplitude of a waveform is the difference between its Top and Base
values. The Y cursors show the values being measured.
Top
The Top of a waveform is the mode (most common value) of the upper
part of the waveform, or if the mode is not well defined, the top is the
same as Maximum. The Y cursor shows the value being measured.
See Also • "To isolate a pulse for Top measurement" on page 173
Base
The Base of a waveform is the mode (most common value) of the lower
part of the waveform, or if the mode is not well defined, the base is the
same as Minimum. The Y cursor shows the value being measured.
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being
measured (edge closest to the trigger reference point).
Overshoot
local Maximum
Top
Base
local Minimum Overshoot
Preshoot
Preshoot is distortion that precedes a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being
measured (edge closest to the trigger reference point).
Preshoot
local Maximum
Top
Base
Preshoot local Minimum
Average
Average is the sum of the levels of the waveform samples divided by the
number of samples.
∑ xi
Average = n
DC RMS
DC RMS is the root- mean- square value of the waveform over one or more
full periods.
n
∑i=1xi2
RMS (dc) = n
AC RMS
AC RMS is the root- mean- square value of the waveform, with the DC
component removed. It is useful, for example, for measuring power supply
noise.
The N Cycles measurement interval measures the value on an integral
number of periods of the displayed signal. If less than three edges are
present, the measurement shows "No edges".
The X cursors show the interval of the waveform being measured.
The Full Screen (Std Deviation) measurement interval variation is an RMS
measurement across the full screen with the DC component removed. It
shows the standard deviation of the displayed voltage values.
The standard deviation of a measurement is the amount that a
measurement varies from the mean value. The Mean value of a
measurement is the statistical average of the measurement.
The following figure graphically shows the mean and standard deviation.
Standard deviation is represented by the Greek letter sigma: σ. For a
Gaussian distribution, two sigma (± 1σ) from the mean, is where 68.3
percent of the measurement results reside. Six sigma (± 3σ) from is where
99.7 percent of the measurement results reside.
bZVc
N
∑i=1 xi
x̄ = N
where:
• x = the mean.
• N = the number of measurements taken.
• xi = the ith measurement result.
∑i=1
N
(xi − x̄)2
σ= N
where:
• σ = the standard deviation.
• N = the number of measurements taken.
• xi = the ith measurement result.
• x = the mean.
Time Measurements
The following figure shows time measurement points.
The default lower, middle, and upper measurement thresholds are 10%,
50%, and 90% between Top and Base values. See "Measurement
Thresholds" on page 184 for other percentage threshold and absolute value
threshold settings.
• "Period" on page 179
• "Frequency" on page 180
• "+ Width" on page 181
• "– Width" on page 181
• "Burst Width" on page 181
• "Duty Cycle" on page 181
• "Rise Time" on page 182
• "Fall Time" on page 182
• "Delay" on page 182
• "Phase" on page 183
Period
Period is the time period of the complete waveform cycle. The time is
measured between the middle threshold points of two consecutive,
like- polarity edges. A middle threshold crossing must also travel through
the lower and upper threshold levels which eliminates runt pulses. The X
cursors show what portion of the waveform is being measured. The Y
cursor shows the middle threshold point.
Frequency
Frequency is defined as 1/Period. Period is defined as the time between
the middle threshold crossings of two consecutive, like- polarity edges. A
middle threshold crossing must also travel through the lower and upper
threshold levels which eliminates runt pulses. The X cursors show what
portion of the waveform is being measured. The Y cursor shows the
middle threshold point.
See Also • "To isolate an event for frequency measurement" on page 180
+ Width
+ Width is the time from the middle threshold of the rising edge to the
middle threshold of the next falling edge. The X cursors show the pulse
being measured. The Y cursor shows the middle threshold point.
– Width
– Width is the time from the middle threshold of the falling edge to the
middle threshold of the next rising edge. The X cursors show the pulse
being measured. The Y cursor shows the middle threshold point.
Burst Width
The Burst Width measurement is the time from the first edge to the last
edge on screen.
Burst width
Duty Cycle
The duty cycle of a repetitive pulse train is the ratio of the positive pulse
width to the period, expressed as a percentage. The X cursors show the
time period being measured. The Y cursor shows the middle threshold
point.
Rise Time
The rise time of a signal is the time difference between the crossing of the
lower threshold and the crossing of the upper threshold for a
positive- going edge. The X cursor shows the edge being measured. For
maximum measurement accuracy, set the horizontal time/div as fast as
possible while leaving the complete rising edge of the waveform on the
display. The Y cursors show the lower and upper threshold points.
Fall Time
The fall time of a signal is the time difference between the crossing of the
upper threshold and the crossing of the lower threshold for a
negative- going edge. The X cursor shows the edge being measured. For
maximum measurement accuracy, set the horizontal time/div as fast as
possible while leaving the complete falling edge of the waveform on the
display. The Y cursors show the lower and upper threshold points.
Delay
Delay measures the time difference from the selected edge on source 1
and the selected edge on source 2 closest to the trigger reference point at
the middle threshold points on the waveforms. Negative delay values
indicate that the selected edge of source 1 occurred after the selected
edge of source 2.
Source 1
Delay
Source 2
Phase
Phase is the calculated phase shift from source 1 to source 2, expressed in
degrees. Negative phase shift values indicate that the rising edge of source
1 occurred after the rising edge of source 2.
Delay
Phase = Source 1 Period × 360
Period
Source 1
Delay
Source 2
Measurement Thresholds
Setting measurement thresholds defines the vertical levels where
measurements will be taken on an analog channel or math waveform.
1 From the Measurement Menu, press the Settings softkey; then, press the
Thresholds softkey to set analog channel measurement thresholds.
You can also open the Measurement Threshold Menu by pressing
[Analyze] > Features and then selecting Measurement Thresholds.
2 Press the Source softkey to select the analog channel or math waveform
source for which you want to change measurement thresholds.
Each analog channel and the math waveform can be assigned unique
threshold values.
4 Press the Lower softkey; then, turn the Entry knob to set the lower
measurement threshold value.
Increasing the lower value beyond the set middle value will
automatically increase the middle value to be more than the lower
value. The default lower threshold is 10% or 800 mV.
If threshold Type is set to %, the lower threshold value can be set from
5% to 93%.
5 Press the Middle softkey; then, turn the Entry knob to set the middle
measurement threshold value.
The middle value is bounded by the values set for lower and upper
thresholds. The default middle threshold is 50% or 1.20 V.
• If threshold Type is set to %, the middle threshold value can be set
from 6% to 94%.
6 Press the Upper softkey; then, turn the Entry knob to set the upper
measurement threshold value.
Decreasing the upper value below the set middle value will
automatically decrease the middle value to be less than the upper value.
The default upper threshold is 90% or 1.50 V.
• If threshold Type is set to %, the upper threshold value can be set
from 7% to 95%.
14
Mask Testing
To create a mask from a "golden" waveform (Automask) 187
Mask Test Setup Options 189
Mask Statistics 192
To manually modify a mask file 193
Building a Mask File 196
187
14 Mask Testing
5 Press Automask.
6 In the Automask Menu, press the Source softkey and ensure the desired
analog channel is selected.
9 To clear the mask and switch off mask testing, press the Back Back/Up
key to return to the Mask Test Menu, then press the Clear Mask softkey.
If infinite persistence display mode (see "To set or clear persistence" on
page 101) is "on" when mask test is enabled, it stays on. If infinite
persistence is "off" when mask test is enabled, it is switched on when
mask test is switched on, then infinite persistence is switched off when
mask test is switched off.
Troubleshooting If you press Create Mask and the mask appears to cover the entire screen,
Mask Setup check the ± Y and ± X settings in the Automask Menu. If these are set to
zero the resulting mask will be extremely tight around the waveform.
If you press Create Mask and it appears that no mask was created, check
the ± Y and ± X settings. They may be set so large that the mask is not
visible.
Run Until The Run Until softkey lets you specify a condition on which to terminate
testing.
• Forever — The oscilloscope runs continuously. However, if an error occurs
the action specified using the On Error softkey will occur.
• Minimum # of Tests — Choose this option and then use the # of Tests
softkey to select the number of times the oscilloscope will trigger, display
the waveform(s), and compare them to the mask. The oscilloscope will
stop after the specified number of tests have been completed. The
specified minimum number of tests may be exceeded. If an error occurs the
action specified using the On Error softkey will occur. The actual number of
tests completed is displayed above the softkeys.
• Minimum Time — Choose this option and then use the Test Time softkey
to select how long the oscilloscope will run. When the selected time has
passed the oscilloscope will stop. The specified time may be exceeded. If
an error occurs the action specified using the On Error softkey will occur.
The actual test time is displayed above the softkeys.
• Minimum Sigma — Choose this option and then use the Sigma softkey to
select a minimum sigma. The mask test runs until enough waveforms are
tested to achieve a minimum test sigma. (If an error occurs the
oscilloscope will perform the action specified by the On Error softkey.)
Note that this is a test sigma (the max achievable process sigma,
assuming no defects, for a certain number of tested waveforms) as
opposed to a process sigma (which is tied to the amount of failures per
test). The sigma value may exceed the selected value when a small sigma
value is chosen. The actual sigma is displayed.
On Error The On Error setting specifies the action(s) to take when the input waveform
does not conform to the mask. This setting supersedes the Run Until setting.
• Stop — The oscilloscope will stop when the first error is detected (on the
first waveform that does not conform to the mask). This setting
supersedes the Minimum # of Tests and Minimum Time settings.
• Save — The oscilloscope saves the screen image when an error is
detected. In the Save Menu (press [Save/Recall] > Save), select an
image format (*.bmp or *.png), destination (on a USB storage device), and
file name (which can be auto-incrementing). If errors occur too frequently
and the oscilloscope spends all its time saving images, press the [Stop]
key to stop acquisitions.
• Print — The oscilloscope prints the screen image when an error is
detected. This option is only available when a printer is connected as
described in "To print the oscilloscope's display" on page 219.
• Measure — Measurements (and measurement statistics if your
oscilloscope supports them) run only on waveforms that contain a mask
violation. Measurements are not affected by passing waveforms. This
mode is not available when the acquisition mode is set to Averaging.
Note that you can choose to Print or Save, but you cannot select both at the
same time. All other actions may be selected at the same time. For example,
you can select both Stop and Measure to cause the oscilloscope to measure
and stop on the first error..
You can also output a signal on the rear panel TRIG OUT BNC connector when
there is a mask test failure. See "Setting the Rear Panel TRIG OUT Source" on
page 232.
Source Lock When you turn on Source Lock using the Source Lock softkey, the mask is
redrawn to match the source whenever you move the waveform. For example,
if you change the horizontal timebase or the vertical gain the mask is redrawn
with the new settings.
When you turn off Source Lock, the mask is not redrawn when horizontal or
vertical settings are changed.
Source If you change the Source channel, the mask is not erased. It is re-scaled to the
vertical gain and offset settings of the channel to which it is assigned. To
create a new mask for the selected source channel, go back up in the menu
hierarchy; then, press Automask, and press Create Mask.
The Source softkey in the Mask Setup Menu is the same as the Source
softkey in the Automask Menu.
Test All When enabled, all displayed analog channels are included in the mask test.
When disabled, just the selected source channel is included in the test.
Mask Statistics
From the Mask Test Menu, press the Statistics softkey to enter the Mask
Statistics Menu.
Show Stats When you enable Show Statistics the following information is displayed:
• Current mask, name of mask, Channel number, date and time.
• # of Tests (total number of mask tests executed).
• Status (Passing, Failing, or Untested).
• Accumulated test time (in hours, minutes, seconds, and tenths of
seconds).
And for each analog channel:
• Number of failures (acquisitions in which the signal excursion went
beyond the mask).
• Failure rate (percentage of failures).
• Sigma (the ratio of process sigma to maximum achievable sigma, based on
number of waveforms tested).
Transparent Enable Transparent mode to write measurement values and statistics on the
screen without a background. Disable the Transparent mode to show them
with a gray background. The Transparent setting affects mask test statistics,
measurement statistics, and reference waveform information display.
Region 1
Region 2
Region number 1 is the top mask region. The vertices in Region 1 describe
points along a line; that line is the bottom edge of the top portion of the
mask.
Similarly, the vertices in Region 2 describe the line that forms the top of
the bottom part of the mask.
The vertices in a mask file are normalized. There are four parameters that
define how values are normalized:
• X1
• ∆X
• Y1
• Y2
These four parameters are defined in the Oscilloscope Setup portion of the
mask file.
The Y- values (normally voltage) are normalized in the file using the
following equation:
Ynorm = (Y - Y1)/∆Y
where ∆Y = Y2 - Y1
To convert the normalized Y- values in the mask file to voltage:
Y = (Ynorm * ∆Y) + Y1
where ∆Y = Y2 - Y1
The X- values (normally time) are normalized in the file using the
following equation:
Xnorm = (X - X1)/∆X
Oscilloscope The keywords "setup" and "end_setup" (appearing alone on a line) define
Setup Information the beginning and end of the oscilloscope setup region of the mask file.
The oscilloscope setup information contains remote programming language
commands that the oscilloscope executes when the mask file is loaded.
Any legal remote programming command can be entered in this section.
The mask scaling controls how the normalized vectors are interpreted.
This in turn controls how the mask is drawn on the display. The remote
programming commands that control mask scaling are:
:MTES:SCAL:BIND 0
:MTES:SCAL:X1 -400.000E-06
:MTES:SCAL:XDEL +800.000E-06
:MTES:SCAL:Y1 +359.000E-03
:MTES:SCAL:Y2 +2.35900E+00
"All Regions"
/* Region Number */ 1
/* Number of vertices */ 4
-12.50, MAX
-10.00, 1.750
10.00, 1.750
12.50, MAX
/* Region Number */ 2
/* Number of vertices */ 5
-10.00, 1.000
-12.50, 0.500
-15.00, 0.500
-15.00, 1.500
-12.50, 1.500
/* Region Number */ 3
/* Number of vertices */ 6
-05.00, 1.000
-02.50, 0.500
02.50, 0.500
05.00, 1.000
02.50, 1.500
-02.50, 1.500
/* Region Number */ 4
/* Number of vertices */ 5
10.00, 1.000
12.50, 0.500
15.00, 0.500
15.00, 1.500
12.50, 1.500
/* Region Number */ 5
/* Number of vertices */ 5
-10.00, -1.000
-12.50, -0.500
-15.00, -0.500
-15.00, -1.500
-12.50, -1.500
/* Region Number */ 6
/* Number of vertices */ 6
-05.00, -1.000
-02.50, -0.500
02.50, -0.500
05.00, -1.000
02.50, -1.500
-02.50, -1.500
/* Region Number */ 7
/* Number of vertices */ 5
10.00, -1.000
12.50, -0.500
15.00, -0.500
15.00, -1.500
12.50, -1.500
/* Region Number */ 8
/* Number of vertices */ 4
-12.50, MIN
-10.00, -1.750
10.00, -1.750
12.50, MIN
setup
:MTES:ENAB 1
15
Waveform Generator
To select generated waveform types and settings 199
To output the waveform generator sync pulse 202
To specify the waveform generator output load 203
To use waveform generator logic presets 203
To restore waveform generator defaults 204
199
15 Waveform Generator
Waveform Characteristics
Type
Sine Use the Frequency/Frequency Fine/Period/Period Fine,
Amplitude/High-Level, and Offset/Low-Level softkeys to set the sine signal
parameters.
The frequency can be adjusted from 100 mHz to 20 MHz.
Square Use the Frequency/Frequency Fine/Period/Period Fine,
Amplitude/High-Level, Offset/Low-Level, and Duty Cycle softkeys to set the
square wave signal parameters.
The frequency can be adjusted from 100 mHz to 10 MHz.
The duty cycle can be adjusted from 20% to 80%.
Ramp Use the Frequency/Frequency Fine/Period/Period Fine,
Amplitude/High-Level, Offset/Low-Level, and Symmetry softkeys to set the
ramp signal parameters.
The frequency can be adjusted from 100 mHz to 100 kHz.
Symmetry represents the amount of time per cycle that the ramp waveform is
rising and can be adjusted from 0% to 100%.
Notice that you can select between coarse and fine adjustments for
frequency, period, and width. Also, pushing the Entry knob is a quick
way to toggle between coarse and fine adjustments.
The Settings softkey opens the Waveform Generator Settings Menu which
lets you make other settings related to the waveform generator.
See:
• "To output the waveform generator sync pulse" on page 202
• "To specify the waveform generator output load" on page 203
• "To use waveform generator logic presets" on page 203
• "To restore waveform generator defaults" on page 204
Noise N/A
Softkey (logic levels) Low level High level, 50 ohm High level, high-Z
expected output load expected output load
TTL 0V +2.5 V (TTL compatible) +5 V
CMOS (5.0V) 0V Not Available +5 V
CMOS (3.3V) 0V +2.5 V (CMOS +3.3 V
compatible)
16
Save/Recall (Setups, Screens, Data)
Saving Setups, Screen Images, or Data 205
Recalling Setups, Masks, or Reference Waveforms 215
Recalling Default Setups 216
Performing a Secure Erase 217
205
16 Save/Recall (Setups, Screens, Data)
When saving screen images, the oscilloscope uses the last menu visited before pressing
NOTE the [Save/Recall] key. This lets you save any relevant information within the softkey menu
area.
To save a screen image showing the Save/Recall Menu at the bottom, press the
[Save/Recall] key twice before saving the image.
You can also save the oscilloscope's display image using a web browser. See "Get
NOTE Image" on page 249 for details.
• Length — sets the number of data points that will be output to the
file. For more information, see "Length Control" on page 211.
• Alb Format — Available ALB serial formats include:
• Default
• CAN
• I2C
• LIN
• UART/RS232
• SPI (2 Wire Serial)
• SPI (3 Wire Serial)
• SPI (4 Wire Serial)
When you choose any ALB format other than Default, waveforms
from the analog channels (if displayed) are also digitized and
displayed as digital waveforms. The analog channel's trigger level
determines the point at which the voltage is considered a logic 1 or
0.
If you connect your serial bus signals to the oscilloscope as shown in
the following table, the bus/signal names in the Agilent Logic
Analyzer application will be correct. Otherwise, you must re- map the
signals in the Agilent Logic Analyzer application.
Clk Ch2 D1
Data Ch3 D2
DataIn Ch3 D2
DataOut Ch4 D3
Length Control
The Length control is available when saving data to CSV, ASCII XY, BIN, or
ALB format files. It sets the number of data points that will be output to
the file. Only displayed data points are saved.
The maximum number of data points depends on these things:
• Whether acquisitions are running. When stopped, data comes from the
raw acquisition record. When running, data comes from the smaller
measurement record.
• Whether the oscilloscope was stopped using [Stop] or [Single]. Running
acquisitions split memory to provide fast waveform update rates. Single
acquisitions use full memory.
• Whether only one channel of a pair is turned on. (Channels 1 and 2 are
one pair, channels 3 and 4 are the other.) Acquisition memory is
divided among the channels in a pair.
• Whether reference waveforms are on. Displayed reference waveforms
consume acquisition memory.
• Whether digital channels are on. Displayed digital channels consume
acquisition memory.
• Whether segmented memory is on. Acquisition memory is divided by
the number of segments.
• The horizontal time/div (sweep speed) setting. At faster settings, fewer
data points appear on the display.
• When saving to a CSV format file, the maximum number of data points
is 50,000.
When necessary, the Length control performs a "1 of n" decimation of the
data . For example: if the Length is set to 1000, and you are displaying a
record that is 5000 data points in length, four of each five data points will
be decimated, creating an output file 1000 data points in length.
When saving waveform data, the save times depend on the chosen format:
ASCII XY medium
CSV slowest
To save masks
Mask files can be saved to one of four internal (\Agilent Flash) locations
or to an external USB storage device.
1 Press [Save/Recall] > Save > Format; then, turn the Entry knob to select
Mask (*.msk).
2 Press the softkey in the second position and use the Entry knob to
navigate to the save location. See "To navigate storage locations" on
page 214.
3 Finally, press the Press to Save softkey.
A message indicating whether the save was successful is displayed.
Mask files have the extension MSK.
Masks are also saved as part of setup files. See "To save setup files" on page 207.
NOTE
17
Print (Screens)
To print the oscilloscope's display 219
To set up network printer connections 220
To specify the print options 221
To specify the palette option 222
You can print the complete display, including the status line and softkeys,
to a USB printer or a network printer when the DSOXLAN LAN/VGA
module is installed.
The Print Configuration Menu is displayed when you press the [Print] key.
The print option softkeys and the Press to Print softkey are ghosted (not
available) until a printer is connected.
219
17 Print (Screens)
The oscilloscope will print the last menu visited before you pressed the [Print] key.
NOTE Therefore, if you have measurements (Amplitude, Frequency, etc.) showing on the display
before you press [Print], the measurements will be shown on the printout.
To print the display showing the Print Configuration Menu at the bottom, press the [Print]
key twice; then, press the Press to Print softkey.
You can also configure the [Quick Action] key to print the display. See
"Configuring the [Quick Action] Key" on page 238.
• Invert Graticule Colors — Select this to reduce the amount of black ink it
takes to print oscilloscope images by changing the black background to
white. Invert Graticule Colors is the default mode.
• Form Feed — Select this to send a form feed command to the printer
after the waveform is printed and before the setup information is
printed. Switch Form Feed off if you want setup information printed on
the same sheet of paper with the waveform. This option only has an
effect when the Setup Information option is selected. Also, if the amount
of setup information will not fit on the same page with the waveform, it
will be printed on a new page regardless of the Form Feed setting.
• Landscape — Select this to print horizontally on the page instead of
vertically (portrait mode).
18
Utility Settings
I/O Interface Settings 223
Setting up the Oscilloscope's LAN Connection 224
File Explorer 227
Setting Oscilloscope Preferences 229
Setting the Oscilloscope's Clock 232
Setting the Rear Panel TRIG OUT Source 232
Performing Service Tasks 233
Configuring the [Quick Action] Key 238
223
18 Utility Settings
When you connect the oscilloscope to a LAN it is a good practice to limit access to the
NOTE oscilloscope by setting a password. By default, the oscilloscope is not password protected.
See "Setting a Password" on page 252 to set a password.
Any time you modify the oscilloscope's hostname, it breaks the connection between the
NOTE oscilloscope and the LAN. You need to re-establish communication to the oscilloscope
using the new hostname.
File Explorer
The File Explorer lets you navigate the oscilloscope's internal file system
and the file systems of connected USB storage devices.
From the internal file system, you can load oscilloscope setup files or
mask files.
From a connected USB storage device, you can load setup files, mask files,
license files, firmware update (*.cab) files, label files, etc. Also, you can
delete files on a connected USB storage device.
The USB port on the front panel, and the USB port on the rear panel labeled "HOST" are
NOTE USB Series A receptacles. These are the receptacles to which you can connect USB mass
storage devices and printers.
The square receptacle on the rear panel labeled "DEVICE" is provided for controlling the
oscilloscope over USB. See the Programmer's Guide for more information.
You do not need to "eject" the USB mass storage device before removing it.
Simply ensure that any file operation you've initiated is done, and remove
the USB drive from the oscilloscope's host port.
Do not connect USB devices that identify themselves as hardware type
"CD" because these devices are not compatible with the InfiniiVision
X- Series oscilloscopes.
If two USB mass storage devices are connected to the oscilloscope, the
first one is designated "\usb" and the second one is designated "\usb2".
See Also • Chapter 16, “Save/Recall (Setups, Screens, Data),” starting on page 205
If the ground level is off screen, the waveform will expand about the
top or bottom edge of the screen based on where the ground is off
screen.
• Center— The displayed waveform will expand about the center of the
display.
3 Press the Wait softkey; then, turn the Entry knob to select the number
of minutes to wait before the selected screen saver activates.
When you turn the Entry knob, the number of minutes is displayed on
the Wait softkey. The default time is 180 minutes (3 hours).
4 Press the Preview softkey to preview the screen saver you have selected
with the Saver softkey.
5 To view the normal display after the screen saver has started, press any
key or turn any knob.
2 Press the Year, Month, Day, Hour or Minute softkey; then, rotate the Entry
knob to set to the desired number.
The hours are shown in the 24- hour format. So 1:00 PM is hour 13.
The real- time clock only allows selection of valid dates. If a day is
selected and the month or year is changed so the day is invalid, the day
is automatically adjusted.
Long cable
to TRIG OUT
To Channel 1
To Channel 2
To Channel 1
To Channel 2
To Channel 3
Longer cable
to TRIG OUT
To Channel 4
4 Connect a BNC cable (40 inches maximum) from the TRIG OUT
connector on the rear panel to the BNC barrel connector.
5 Press the [Utility] key; then, press the Service softkey.
6 Begin the Self Cal by pressing the Start User Cal softkey.
To contact Agilent
Information on contacting Agilent Technologies can be found at:
"www.agilent.com/find/contactus"
19
Web Interface
Accessing the Web Interface 242
Browser Web Control 243
Save/Recall 247
Get Image 249
Identification Function 250
Instrument Utilities 251
Setting a Password 252
241
19 Web Interface
If Java is not installed on your PC, you will be prompted to install the Sun Microsystems
NOTE Java Plug-in. This plug-in must be installed on the controlling PC for the web interface's
Remote Front Panel or Remote Programming operations.
Main Menu
Function Keys
Hint appears
when you roll
over with
mouse
Softkeys
(left-click to
select, right-click
for Quick Help)
Scrolling and When using a monitor resolution of 800 x 600 or less on the remote
Monitor computer, you need to scroll to access the full remote front panel. To
Resolution display the remote front panel without scroll bars, use a monitor
resolution greater than 800 x 600 on your computer's display.
Save/Recall
You can save setup files, screen images, waveform data files, or mask files
to your PC via the oscilloscope's web interface (see "Saving Files via the
Web Interface" on page 247).
You can recall setup files, reference waveform data files, or mask files
from your PC via the oscilloscope's web interface (see "Recalling Files via
the Web Interface" on page 249).
You can click Preview to view the oscilloscope's current screen image.
When previewing, you can use the New Acquisition check box to force
a new acquisition before the preview.
With some formats, you can click Save Setup Info to save setup
information to an ASCII .txt format file.
c Click Save.
The current acquisition is saved.
d In the File Download dialog, click Save.
e In the Save As dialog, navigate to the folder where you want to save
the file; then, click Save.
d Click Recall.
Get Image
To save (or print) the oscilloscope's display from the web interface:
1 Access the oscilloscope's web interface (see "Accessing the Web
Interface" on page 242).
2 When the oscilloscope's web interface is displayed, select the Get Image
tab from the left side of the Welcome screen. After a delay of several
seconds, the oscilloscope's screen image will be displayed.
3 Right- click on the image and select Save Picture As... (or Print Picture...).
4 Select a storage location for the image file and click Save.
Identification Function
The Identification web interface feature is useful when trying to locate a
specific instrument in a rack of equipment.
1 Access the oscilloscope's web interface (see "Accessing the Web
Interface" on page 242).
2 When the oscilloscope's web interface Welcome Page is displayed, select
the Identification on radio button.
An "Identify" message is displayed on the oscilloscope; you can either
select Identification off or press the OK softkey on the oscilloscope to
continue.
Identification option
Instrument Utilities
The Instrument Utilities page of the web interface lets you:
• View installed options.
• View firmware versions.
• Install firmware upgrade files.
• View calibration status.
You can select these cababilities via a drop- down menu.
Setting a Password
Whenever you connect the oscilloscope to a LAN, it is good practice to set
a password. The password prevents someone from remotely accessing the
oscilloscope via a Web browser and changing parameters. Remote users
can still view the Welcome screen, view network status, etc., but they can't
operate the instrument or change its setup without the password.
To set a password:
1 Access the oscilloscope's web interface (see "Accessing the Web
Interface" on page 242).
2 When the oscilloscope's web interface is displayed, select the Configure
Network tab from the instrument's Welcome page.
Modify Configuration
Configure
Network
tab
Enter
password
When accessing the password protected oscilloscope, the user name is the
IP address of the oscilloscope.
To reset the Do one of these things to reset the password:
password • Using the keys on the front panel of the oscilloscope, press [Utility] > I/O
> LAN Reset.
• Using the web browser select the Configure Network tab, select Modify
Configuration, erase the Password, and select Apply Changes.
20
Reference
Specifications and Characteristics 255
Measurement Category 255
Environmental Conditions 257
Probes and Accessories 258
Loading Licenses and Displaying License Information 261
Software and Firmware Updates 262
Binary Data (.bin) Format 262
CSV and ASCII XY files 269
Acknowledgements 271
Measurement Category
• "Oscilloscope Measurement Category" on page 256
• "Measurement Category Definitions" on page 256
255
20 Reference
Use this instrument only for measurements within its specified measurement
WARNIN G
category.
Environmental Conditions
Pollution Degree Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
Definitions The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.
Passive Probes
Passive probes are included with the 2000 X- Series oscilloscopes for each
analog channel.
The following passive probes can be used with the InfiniiVision
2000 X- Series oscilloscopes. Any combination of passive probes can be
used.
Model Description
1165A Passive probe, 10:1, 600 MHz, 1.5 m
Model Description
N2771A/B Passive probe, 1000:1, 30 kV, 50 MHz
N2862A/B Passive probe, 10:1, 150 MHz, 1.2 m
Differential Probes
The following differential probes can be used with the InfiniiVision
2000 X- Series oscilloscopes.
Model Description
1141A Active differential probe, 200 MHz, 200 VDC + peak AC max (requires
1142A power supply)
1144A Active probe, 800 MHz (requires 1142A power supply)
1145A Active probe, 750 MHz 2-ch (requires 1142A power supply)
N2772A Active differential probe, 20 MHz, 1.2 kVDC + peak AC max (requires
N2773A power supply)
Current Probes
The following current probes can be used with the InfiniiVision
2000 X- Series oscilloscopes.
Model Description
N2781A Current probe, 10 MHz, 150 A, AC/DC (use with N2779A power supply)
N2782A Current probe, 50 MHz, 30 A, AC/DC (use with N2779A power supply)
N2783A Current probe, 100 MHz, 30 A, AC/DC (use with N2779A power supply)
Accessories Available
In addition to passive proves ("Passive Probes" on page 258), differential
probes ("Differential Probes" on page 259), and current probes ("Current
Probes" on page 259), the following accessories are available for the
InfiniiVision 2000 X- Series oscilloscopes.
Model/Part # Description
Option Order
A6J ANSI Z540 Compliant Calibration
Upgrading to an MSO
A license can be installed to activate the digital channels of an
oscilloscope that was not originally ordered as a mixed- signal oscilloscope
(MSO). A mixed signal oscilloscope has analog channels plus 8
time- correlated digital timing channels.
For information about upgrading your oscilloscope through licensing,
contact your local Agilent Technologies representative or see
"www.agilent.com/find/2000X- Series".
Because the data is in binary format, the size of the file is approximately
5 times smaller than the ASCII XY format.
If more than one source is on, all displayed sources will be saved, except
math functions.
When using segmented memory, each segment is treated as a separate
waveform. All segments for a channel are saved, then all segments of the
next (higher numbered) channel are saved. This continues until all
displayed channels are saved.
When the oscilloscope is in the Peak Detect acquisition mode, the
minimum and maximum value waveform data points are saved to the file
in separate waveform buffers. The minimum value data points are saved
first; then, the maximum value data points are saved.
BIN data - using When saving all segments, each segment has its own waveform header (see
segmented "Binary Header Format" on page 264).
memory
In BIN file format, data are presented as follows:
• Channel 1 data (all segments)
• Channel 2 data (all segments)
• Channel 3 data (all segments)
• Channel 4 data (all segments)
• Digital channel data (all segments)
• Math waveform data (all segments)
When not saving all segments, the number of waveforms is equivalent to
the number of active channels (including math and digital channels, with
up to seven waveforms for each digital pod). When saving all segments,
the number of waveforms is equal to the number of active channels
multiplied by the number of segments acquired.
Cookie Two byte characters, AG, that indicate the file is in the Agilent Binary Data file
format.
Version Two bytes that represent the file version.
File Size A 32-bit integer that is the number of bytes that are in the file.
Number of A 32-bit integer that is the number of waveforms that are stored in the file.
Waveforms
Waveform Header It is possible to store more than one waveform in the file, and each
waveform stored will have a waveform header. When using segmented
memory, each segment is treated as a separate waveform. The waveform
header contains information about the type of waveform data that is
stored following the waveform data header.
Header Size A 32-bit integer that is the number of bytes in the header.
Waveform Type A 32-bit integer that is the type of waveform stored in the file:
• 0 = Unknown.
• 1 = Normal.
• 2 = Peak Detect.
• 3 = Average.
• 4 = Not used in InfiniiVision oscilloscopes.
• 5 = Not used in InfiniiVision oscilloscopes.
• 6 = Logic.
Number of A 32-bit integer that is the number of waveform buffers required to read the
Waveform data.
Buffers
Points A 32-bit integer that is the number of waveform points in the data.
Count A 32-bit integer that is the number of hits at each time bucket in the waveform
record when the waveform was created using an acquisition mode like
averaging. For example, when averaging, a count of four would mean every
waveform data point in the waveform record has been averaged at least four
times. The default value is 0.
X Display Range A 32-bit float that is the X-axis duration of the waveform that is displayed. For
time domain waveforms, it is the duration of time across the display. If the
value is zero then no data has been acquired.
X Display Origin A 64-bit double that is the X-axis value at the left edge of the display. For time
domain waveforms, it is the time at the start of the display. This value is
treated as a double precision 64-bit floating point number. If the value is zero
then no data has been acquired.
X Increment A 64-bit double that is the duration between data points on the X axis. For
time domain waveforms, this is the time between points. If the value is zero
then no data has been acquired.
X Origin A 64-bit double that is the X-axis value of the first data point in the data
record. For time domain waveforms, it is the time of the first point. This value
is treated as a double precision 64-bit floating point number. If the value is
zero then no data has been acquired.
X Units A 32-bit integer that identifies the unit of measure for X values in the acquired
data:
• 0 = Unknown.
• 1 = Volts.
• 2 = Seconds.
• 3 = Constant.
• 4 = Amps.
• 5 = dB.
• 6 = Hz.
Y Units A 32-bit integer that identifies the unit of measure for Y values in the acquired
data. The possible values are listed above under X Units.
Date A 16-byte character array, left blank in InfiniiVision oscilloscopes.
Time Tags A 64-bit double, only used when saving multiple segments (requires
segmented memory option). This is the time (in seconds) since the first
trigger.
Segment Index A 32-bit unsigned integer. This is the segment number. Only used when saving
multiple segments.
Waveform Data A waveform may have more than one data set. Each waveform data set
Header will have a waveform data header. The waveform data header consists of
information about the waveform data set. This header is stored
immediately before the data set.
Waveform Data A 32-bit integer that is the size of the waveform data header.
Header Size
Buffer Type A 16-bit short that is the type of waveform data stored in the file:
• 0 = Unknown data.
• 1 = Normal 32-bit float data.
• 2 = Maximum float data.
• 3 = Minimum float data.
• 4 = Not used in InfiniiVision oscilloscopes.
• 5 = Not used in InfiniiVision oscilloscopes.
• 6 = Digital unsigned 8-bit char data (for digital channels).
Bytes Per Point A 16-bit short that is the number of bytes per data point.
Buffer Size A 32-bit integer that is the size of the buffer required to hold the data points.
File Header
Number of Waveforms = N
12 bytes
Waveform Data
Buffer Type = 1 (floating point)
Header 1
Bytes per Point = 4
12 bytes
Voltage Data 1
buffer size
Waveform Data
Buffer Type = 1 (floating point)
Header 2
Bytes per Point = 4
12 bytes
Voltage Data 2
buffer size
Waveform Data
Buffer Type = 1 (floating point)
Header N
Bytes per Point = 4
12 bytes
Voltage Data N
buffer size
Single The following picture shows a binary file of a single acquisition with all
Acquisition All pods for the logic channels saved.
Pods Logic
Channels
File Header
Number of Waveforms = 2
12 bytes
Waveform Data
Buffer Type = 6 (unsigned char)
Header 1
Bytes per Point = 1
12 bytes
Waveform Data
Buffer Type = 6 (unsigned char)
Header 2
Bytes per Point = 1
12 bytes
File Header
Number of Waveforms = N = Number of Segments
12 bytes
Waveform Data
Buffer Type = 1 (floating point)
Header 1
Bytes per Point = 4
12 bytes
Voltage Data 1
buffer size
Waveform Data
Buffer Type = 1 (floating point)
Header 2
Bytes per Point = 4
12 bytes
Voltage Data 2
buffer size
Waveform Data
Buffer Type = 1 (floating point)
Header N
Bytes per Point = 4
12 bytes
Voltage Data N
buffer size
Acknowledgements
RealVNC RealVNC is licensed under the GNU General Public License. Copyright (C)
2002- 2005 RealVNC Ltd. All Rights Reserved.
This is free software; you can redistribute it and/or modify it under the
terms of the GNU General Public License as published by the Free
Software Foundation; either version 2 of the License, or (at your option)
any later version.
This software is distributed in the hope that it will be useful, but
WITHOUT ANY WARRANTY; without even the implied warranty of
MERCHANTABILITY or FITNESS FOR A PARTICULAR PURPOSE. See the
GNU General Public License for more details. The license is located on the
Agilent InfiniiVision Oscilloscopes Documentation CD- ROM.
RealVNC source code can be obtained from RealVNC or by contacting
Agilent. Agilent will charge for the cost of physically performing the
source distribution.
HDF5 Reference Waveform files use HDF5.
HDF5 was developed by "The HDF Group" and by the National Center for
Supercomputing Applications at the University of Illinois at
Urbana- Champaign.
GPIB module, 22, 37 invert graticule colors, 208 low-frequency noise rejection, 136
GPIB module installation, 22 invert waveform, 59
grabber, 81, 82 IP address, 225, 241 M
graphical user interface language, 41 Italian front panel overlays, 36
graticule intensity, 103 Italian user interface and Quick Help, 41 mask files, recall, 216
graticule type, 102 MASK license, 261
grid intensity, 103 J mask test, trigger output, 191, 233
grid type, 102 mask testing, 187
ground level, 56 Japanese front panel overlays, 36 mask, TRIG OUT signal, 233
Ground terminal, 34 Japanese user interface and Quick Math key, 32
Help, 41 math, 1*2, 65
H math, addition, 66
K math, FFT, 67
Hanning FFT window, 68 math, functions, 63
hardware self test, 236 keys, front panel, 28 math, multiply, 65
Help key, 33 knobs, front panel, 28 math, offset, 65
help, built-in, 40 Korean front panel overlays, 36 math, scale, 65
hex bus trigger, 118 Korean user interface and Quick Help, 41 math, subtract, 66
Hex softkey, 118 math, transform functions on arithmetic
HF Reject, 137 L operations, 64
high-frequency noise rejection, 137 math, units, 65, 74
high-resolution mode, 147, 153 Label key, 34 math, using waveform math, 63
holdoff, 138 label list, 109 MATLAB binary data, 263
Horiz key, 31, 43, 48, 50, 150 label list, loading from text file, 108 Maximum measurement, 172
Horizontal controls, 31, 47 labels, 105 maximum sample rate, 147
horizontal Navigate key, 31 labels, auto-increment, 108 Meas key, 33, 167
horizontal position control, 31 labels, default library, 109 Measure controls, 33
horizontal position knob, 45 LAN connection, 225 measure, Quick Measure All, 238
horizontal scale fine adjustment, 52 LAN interface, remote control, 223 measurement category, definitions, 256
horizontal Search key, 31 LAN port, 37 measurement definitions, 169
horizontal sweep speed control, 31 LAN Settings softkey, 225, 226 measurement record, 212
horizontal time/div control, 31 LAN/VGA module, 22, 37 measurement thresholds, 184
horizontal Zoom key, 31 LAN/VGA module installation, 22 measurement window with zoom
host name, 225, 241 landscape mode, 222 display, 186
Host name softkey, 226 language, user interface and Quick measurements, 169
Help, 41 measurements, automatic, 167
I length control, 211 measurements, delay, 170
length softkey, 209, 210 measurements, overshoot, 170
I/O interface settings, 223 level, trigger, 112 measurements, phase, 170
identification function, web interface, 250 LF Reject, 136 measurements, preshoot, 170
impedance, digital probes, 91 library, labels, 107 measurements, time, 179
indeterminate state, 159 licenses, 261, 262 measurements, voltage, 171
infinite persistence, 101, 142, 149 line voltage, 23 MegaZoom IV, 4
information area, 39 load file, 227 memory depth and sample rate, 147
installed licenses, 236 Load from, 214 memory, segmented, 153
installed options, 251 localized front panel overlay, 35 menu line, 39
Instrument Utilities web page, 251 Location, 214, 228 Minimum measurement, 172
intensity control, 99 logic presets, waveform generator, 203 Mode/Coupling key, trigger, 133
Intensity key, 29 logic threshold, 86 model number, 236, 241