Hill Rom Care Assist Service Manual
Hill Rom Care Assist Service Manual
Hill Rom Care Assist Service Manual
CareAssist™
Bed Errata
138755
From Hill-Rom
CareAssist™ Bed
Service Manual
Revisions
MAN330 REV 2
Second Edition
To order additional copies of this manual, call (800) 445-3720, and place a
parts order for part number MAN330.
NOTES:
Chapter 1: Introduction
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Chapter 1: Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Chapter 2: Troubleshooting Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Chapter 3: Theory of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Chapter 4: Removal, Replacement, and Adjustment Procedures . . . . . . . . . . . 1 - 2
Chapter 5: Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Chapter 6: General Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Chapter 7: Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Document Symbol Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Physical Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 5
Model Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 10
Safety Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 11
Warning and Caution Labels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 13
Chapter 2: Troubleshooting Procedures
Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Initial Actions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Problem/Solution Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Function Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Scale and Bed Exit Function Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 9
Final Actions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 11
No Functions Work. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 12
Lockout Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 15
Hilow Malfunction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 16
NOTES:
Purpose
This manual provides requirements for the CareAssist™ Bed normal operation
and maintenance. It also includes parts lists (in chapter 5) for ordering
replacement components.
Audience
This manual is intended for use by only facility-authorized personnel. Failure
to observe this restriction can result in severe injury to people and serious
damage to equipment.
Organization
This manual contains seven chapters.
Chapter 1: Introduction
In addition to a brief description of this service manual, chapter 1 also provides
a product overview.
Chapter 7: Accessories
A list of additional products, that can be used in conjunction with the
CareAssist™ Bed, is available in chapter 7. Installation procedures for these
accessories are also included.
Overview
The CareAssist ™ Bed is intended for low to moderate acuity patients in the
medical/surgical area of the hospital. The CareAssist™ Bed can also be used
as a general-purpose variable height hospital bed for general care, post-
operative care, and general medicine wards.
For more information refer to the CareAssist™ Bed User Manual (usr116).
Before you service the CareAssist™ Bed, make sure you have read and
understood in detail the contents of this manual. It is important that you read
and strictly obey the aspects of safety contained in the this manual.
Feature Dimension
Total length 100" (254 cm)
Maximum width (siderails stored) 40" (102 cm)
Maximum width (siderails up) 40" (102 cm)
Mattress dimensions:
Mattress width 36.0" (91.4 cm)
Mattress length 80" (203.2 cm)
Caster size 6" (15 cm)
Total weight 420 lb (191 kg) without surface, options, or
accessories
Feature Dimension
Head section inclination (maximum) 65°
Knee section inclination (maximum) 20°
Foot section inclination (maximum) -23°
Maximum height (to top of sleep deck) 32.5" (82.6 cm)
Minimum height (to top of sleep deck) 16.75" (42.55 cm)
Trendelenburg position (maximum) 16°
Safe working load 400 lb (181 kg)
(safe working load includes: patient, acces
sories, mattress, and etc.)
Siderail opening size 4.34" (11.02 cm)
Distance between siderails < 2.3" (58.4 mm)
Condition Range
Temperature -40°F to 122°F (-40°C to 50°C)
Relative humidity 95% non-condensing
Pressure 50 kPa to 106 kPa
Mattress Flammability
The sleep surface mattress meets the following specifications:
• CAL TB-117, Requirements, Test Procedure and Apparatus for
Testing the Flame Retardance of Resilient Filling Materials Used in
Upholstered Furniture (foam)
• 16 CFR 1632, Standard for the Flammability of Mattresses and
Mattress Pads
• CAL TB-129, Flammability Test Procedures for Mattresses for Use in
Public Buildings (models with fire barrier only)
• BFD IX-II, Boston Fire Department Mattress Fire Test (models with
fire barrier only)
Condition Range
Temperature 32°F to 104°F (0°C to 40°C) ambient tem
perature
Relative humidity range 30% to 85% non-condensing
Atmospheric Pressure 70 kPa to 106 kPa
120 60 6.0
a
Condition Range
Battery fuse 10 A, 32 V~, ATO
Mains fuse 2 each 2.5 A, 250 V~, 5 x 20 mm, UL
a
198G Slo-Blo® or equivalent (220, 230,
and 240 V beds)
2 each 5 A, 250 V~, .25" x 1.25", UL 198G
Slo-Blo® or equivalent (100 V through
127 V beds)
a. Slo-Blo® is a registered trademark of Littlefuse, Inc.
IEC 60601-2-38
EN 60601-1
IEC 60601-1-2
EN ISO 9001 and EN 13485
Equipment Classification per IEC 60601-1 Class I equipment, internally powered
equipment
Degree of Protection Against Electric Type B
Shock
Degree of Protection Against Ingress of Protection against spraying or splashing
Water water- IPX4
Degree of Protection Against the Presence Not for use with flammable anaesthetics.
of Flammable Anaesthetic Mixtures
Mode of Operation Continuous operation with intermittent
loading, 3 minutes ON/27 minutes OFF
Sound level (measured 1 meter from < 60 dBA
patient’s ear)
a. CSA® is a registered trademark of Canadian Standards Association, Inc.
RF emissions Group 1 The model P1170 uses RF energy only for its internal
CISPR 11 functions. Therefore, its RF emissions are low and are
not likely to cause any interference in nearby electronic
equipment.
RF Emissions Class A The model P1170 is suitable for use in all establishments
CISPR 11 other than domestic and those directly connected to the
public low-voltage power supply network that supplies
Harmonic Emissions Not Applicable buildings used for domestic purposes.
IEC 61000-3-2
Electrical Fast ± 2 kV on Power Supply ± 2 kV on Power Supply Mains power quality should be that
Transient/Burst Lines Lines of a typical commercial or hospital
IEC 61000-4-4 ± 1 kV on Input/ Output ± 1 kV on Input/ Output environment.
Lines Lines
Surge ± 1 kV Differential Mode ± 1 kV Differential Mode Mains power quality should be that
IEC 61000-4-5 (Line-Line) (Line-Line) of a typical commercial or hospital
± 2 kV Common Mode ± 2 kV Common Mode environment.
(Line-Ground) (Line-Ground)
Voltage Dips, Short < 5% UT < 5% UT Mains power quality should be that
Interrupts, & Varia (95% dip in UT for 0.5 (95% dip in UT for 0.5 of a typical commercial or hospital
tions On Power Sup- cycles ) cycles ) environment. If operation is required
ply Lines < 40% U T < 40% U T during an extended power outage or
IEC 61000-4-11 interruption, the model P1170
(60% dip in UT for 5 (60% dip in UT for 5
should be switched to operate from
cycles ) cycles ) the backup battery.
< 70% U T < 70% U T
(30% dip in UT for 25 (30% dip in UT for 25
cycles ) cycles )
< 5% UT < 5% UT
(95% dip in UT for 5 sec (95% dip in UT for 5 sec
onds ) onds )
(See Note 1)
Note 2: The compliance levels in the ISM frequency range 150 kHz to 2.5 GHz are intended to decrease the likelihood
that mobile/portable communications equipment could cause interference if it is inadvertently brought into the patient
area. However, emission limits, IEC 60601 test levels, and tests specified in IEC 60601-1-2:2001 do not address electro
magnetic compatibility of electrical equipment at very close distances. Care should always be exercised when using any
electrical or RF equipment in the immediate patient area.
Model Identification
Chapter 1: Introduction
Model Identification
For CareAssist™ Bed model identification, see table 1-10 on page 1-10.
WARNING:
Only facility-authorized personnel should service the CareAssist™ Bed.
Servicing performed by unauthorized personnel could result in personal
injury or equipment damage.
WARNING:
Adhere to appropriate infection control policies and procedures. Failure
to do so could result in the spread of infection.
WARNING:
To prevent injury from sharp edges and rough surfaces, wear protective
gloves.
WARNING:
To reduce the risk of eye injury, wear eye protection.
WARNING:
Do not work under an unsupported load. Install applicable temporary
supports. Failure to do so could cause personal injury or equipment
damage.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture. Personal injury or
equipment damage could occur.
CAUTION:
Do not use harsh cleaners, solvents, or detergents. Equipment damage
could occur.
CAUTION:
Do not use silicone-based lubricants. Equipment damage could occur.
CAUTION:
To prevent component damage, make sure that your hands are clean,
and only handle the P.C. board by its edges.
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
CAUTION:
For shipping and storage, place the removed P.C. board in an antistatic
protective bag. Equipment damage can occur.
m330_2_170
IV Pole Warning
NOTES:
WARNING:
Only facility-authorized personnel should service the CareAssist™ Bed.
Servicing by unauthorized personnel could result in personal injury or
equipment damage.
Begin each procedure in this chapter with step 1. Follow the sequence outlined
(each step assumes the previous step has been completed). In each step, the
normal operation of the product can be confirmed by answering Yes or No to
the statement. Your response will lead to another step in the procedure, a repair
analysis procedure (RAP), or a component replacement. If more than one
component is listed, replace them in the given order.
To begin gathering information about the problem, start with Initial Actions.
To isolate or identify a problem and to verify the repair after completing each
corrective action (replacing or adjusting a part, seating a connector, etc.),
perform the Function Checks.
To verify the repair, perform the Final Actions after the Function Checks.
Initial Actions
Use Initial Actions to gather information from operators concerning problems
with the CareAssist™ Bed. Note the symptoms or other information
concerning the problem that the operator describes. This information helps
identify the probable cause.
2. Ask this person to demonstrate or explain the problem. The problem can be
duplicated.
Yes No
↓ → Go to “Function Checks” on page 2-4.
Problem/Solution Table
If the problem can be easily identified, use the following tables to determine
the appropriate troubleshooting procedure.
Problem Solution
No Functions Work Go to RAP 2.1 on page 2-12.
Lockout Malfunction Go to RAP 2.2 on page 2-15.
Hilow Malfunction Go to RAP 2.3 on page 2-16.
Knee Section Malfunction Go to RAP 2.4 on page 2-17.
Head Section Malfunction (excluding Automatic Go to RAP 2.5 on page 2-18.
Contour)
Battery Backup Malfunction Go to RAP 2.6 on page 2-19.
Foot Section Malfunction Go to RAP 2.7 on page 2-20.
Problem Solution
Trendelenburg/Reverse Trendelenburg Malfunction Go to RAP 2.8 on page 2-22.
Bed Connected to AC Power, Brakes Not Applied Go to RAP 2.9 on page 2-23.
Detection Malfunction
CPR Malfunction
Braking Malfunction
Go to RAP 2.10 on page 2-26.
Go to RAP 2.11 on page 2-27. 2
Steering Malfunction Go to RAP 2.12 on page 2-28.
Problem Solution
Scale Error 0 Go to RAP 2.13 on page 2-29.
Scale Error 1 Go to RAP 2.14 on page 2-30.
Scale Error 2 Go to RAP 2.15 on page 2-31.
Scale Error 3 Go to RAP 2.16 on page 2-32.
Scale Error 5 Go to RAP 2.17 on page 2-33.
Function Checks
Chapter 2: Troubleshooting Procedures
Function Checks
1. The "Initial Actions" have been performed.
Yes No
↓ → Go to “Initial Actions” on page 2-2.
2. Set the brakes, raise and lock the siderails in the high position. Plug in the
power cord into an appropriate power source.
Yes No
↓ → Go to RAP 2.3.
15. Unplug the power from the power source, and test the functions checked in
step 12 to step 14.
The bed functions operate properly.
Yes No
↓ → Go to RAP 2.6.
16. Press the Knee Up control until it reaches the high position and then press
the Foot Up control.
The foot section rises to the high position without stopping or the audible
alarm sounding.
Yes No
↓ → Go to RAP 2.7.
18. Press the Knee Down control until knee section is in the horizontal
position. Press the Dining Chair™ Position control.
The head section and knee section rise until the required position is
reached.
Yes No
↓ → Go to RAP 2.5.
20. Press the Bed Up control until the intermediate position is reached, and
then operate the Trendelenburg control.
The sleep surface gradually tilts to maximum Trendelenburg without any
problem or abnormal noise.
Yes No
↓ → Go to RAP 2.8.
22. Plug the bed into an appropriate power source. Set the brakes to the neutral
position.
An audible alarm can be heard.
2
Yes No
↓ → Go to RAP 2.9.
24. Press the Head Up control until the section reaches the high position.
NOTE:
Someone should be lying on the bed so that this test can be carried out
conclusively.
Pull one of the CPR control handles.
The head section descends quickly to the mid position, then slows until the
horizontal position is reached. The movement occurs without any problem
or abnormal noise. Repeat for the other CPR control handle.
Yes No
↓ → Go to RAP 2.10.
25. Release the CPR control handle, and press the Head Up control to check
that the head section motor drive mechanism is working correctly.
The head section rises without any problem or abnormal noise.
Yes No
↓ → Go to RAP 2.10.
26. Try to move the bed with the brake still applied.
The four wheels are locked and prevent any movement.
Yes No
↓ → Go to RAP 2.11.
27. Set the brakes to Steer, and move the bed far enough to lock the steer caster
into position.
NOTE:
The Brake Not Set alarm will sound.
The steer caster locks into position.
Yes No
↓ → Go to RAP 2.12.
4. Put 40 lb (18 kg) on the center of the bed. Press the Zero control.
Weight display shows 0.0.
Yes No
↓ → Replace scale P.C. board.
Final Actions
1. Perform the required preventive maintenance procedures. See “Preventive
Maintenance” on page 6-4.
NOTE:
This procedure assumes that no functions work from the pendant or caregiver
controls on the siderails.
SHOCK HAZARD:
The potential for electrical shock exists with electrical equipment.
Failure to follow facility protocols may cause death or serious personal
injury.
1. Verify the two AC power fuses and one interface P.C. board in-line cable
fuse are serviceable (A model beds only).
The fuses are serviceable.
Yes No
↓ → Replace the fuses (refer to procedure 4.16) (A model beds) or
(refer to procedure 4.34) (B model beds), and then go to
“Function Checks” on page 2-4.
2. Verify all the cables on the AC power P.C. board and motor control P.C.
board (A model beds) or siderail interface P.C. board (B model beds) are
connected correctly.
All cables are connected properly.
Yes No
↓ → Connect the cables as required, and then go to “Function
Checks” on page 2-4.
4. Unplug the bed from its power source and use an appropriate test device to
check for the presence of the voltage at the power outlet.
The voltage is correct.
Yes No
↓ → Inform the customer of the problem concerning the power outlet.
Connect the bed to a working power outlet, and then go to
“Function Checks” on page 2-4.
6. Using the CPR control, adjust the head section to the high position.
2
SHOCK HAZARD:
The potential for electrical shock exists with electrical equipment.
Failure to follow facility protocols may result in death or serious
personal injury.
a. Remove the power supply cover.
b. Plug the unit into an appropriate power source.
c. Using a digital multimeter set to AC volts, on the AC power cable,
check for the presence of the AC voltage.
The voltage is correct.
Yes No
↓ → Replace the AC power cable (refer to procedure 4.2) (A model
beds) or (refer to procedure 4.33) (B model beds), and then go to
“Function Checks” on page 2-4.
7. Using a digital multimeter set to AC volts, check for the presence of the
AC voltage on the transformer connection on the motor control P.C. board.
The voltage is correct.
Yes No
↓ → Replace the power supply (A model beds) (refer to procedure
4.3), or replace the motor control P.C. board (refer to procedure
4.27) and then go to “Function Checks” on page 2-4.
Yes No
↓ → Replace the original interface P.C. board. If this solves the
problem perform the “Final Actions” on page 2-11, otherwise
contact Hill-Rom Technical Support.
10. Make sure that the patient pendant and caregiver siderail controls operate
correctly.
The functions are operational.
Yes No
↓ → Replace the patient pendant or siderail as required. If this solves
the problem perform the “Final Actions” on page 2-11,
otherwise contact Hill-Rom Technical Support.
NOTE:
The symptom of this fault is after pressing the lockout button, none of the
indicators flash.
NOTE:
The Bed Not Down indicator does not flash in the steps that include the 2
asterisk symbol (*).
3. Replace the siderail interface P.C. board (refer to procedure 4.15) (A model
beds) or (refer to procedure 4.31) (B model beds) on the affected siderail.
3. Swap the hilow column connectors #6 and #4 (yellow) on the power supply
P.C. board (see table 4-1 on page 4-11).
The faulty column operates properly.
Yes No
↓ → Replace the defective column: head hilow (refer to procedure 4.9
on page 4-31) or foot hilow (refer to procedure 4.21 on page
4-63).
4. Replace the power supply (refer to procedure 4.3 on page 4-9) or motor
control P.C board (B model beds) (refer to procedure 4.27 on page 4-85).
6. Replace the power supply (A model beds) (refer to procedure 4.3 on page
4-9) or motor control P.C board (B model beds) (refer to procedure 4.27 on
page 4-85).
This solves the problem.
Yes No
↓ → Contact Hill-Rom technical support.
4. Remove the power supply cover (A model beds), or battery cover (B model
beds), and make sure the batteries are properly connected.
The battery is connected properly.
Yes No
↓ → Connect the battery properly and go to “Function Checks” on
page 2-4.
7. Make sure the battery backup is not defective by replacing with a new
battery.
The bed operates correctly with the new battery backup.
Yes No
↓ → Contact Hill-Rom Technical Support.
8. Replace the power supply (A model beds) (refer to procedure 4.3) or the
motor control P.C. board (B model beds) (refer to procedure 4.27). If this
solves the problem perform the “Final Actions” on page 2-11, otherwise
contact Hill-Rom Technical Support.
4. Connect one control at a time until the bad control is located. Replace the
affected control.
6. Go to step 11.
7. Do the following:
a. Remove the power supply cover.
b. Swap the connectors of the knee section and foot section motors (#5
brown, #2 green) on the power board.
c. Activate the CPR function.
The foot section function operates correctly, the knee section motor rod
extends to mid-travel.
Yes No
↓ → Replace the foot motor (refer to procedure 4.8 on page 4-27).
2
m330_108
10. Replace the knee motor (refer to procedure 4.7 on page 4-23), then go to
“Final Actions” on page 2-11, otherwise contact Hill-Rom Technical
Support.
11. Using an ohmmeter, check the state of the foot section motor intermediate
switch.
• When the rod is extended, contacts 3 and 4 (see Figure 2-1 on page
2-21) of the connector (socket #2, green) are short-circuited.
• When the rod is retracted, the circuit is open.
The foot section motor intermediate switch is working.
Yes No
↓ → Replace the power supply (A model beds) (refer to procedure
4.3) or the motor control P.C. board (B model beds) (refer to
procedure 4.27). If this solves the problem perform the “Final
Actions” on page 2-11, otherwise contact Hill-Rom Technical
Support.
12. Replace the foot motor (refer to procedure 4.8 on page 4-27).
This solves the problem.
Yes No
↓ → Contact Hill-Rom Technical Support.
5. Do the following to make sure the foot column cables are connected
correctly.
a. Reconnect the 2-pin connector to the bottom of the foot column.
b. Disconnect the foot column extension cable in the central rail of the
bed frame.
c. The same verification applies to the terminals of the 8-pin connector in
the central rail of the bed frame by checking the operation of the brake
detection switch from terminals 1 and 6 of the foot column (see figure
2-2 on page 2-24).
m330_107
The foot column cable connections and brake detection switch continuity
are correct.
Yes No
↓ → Replace the foot column (refer to procedure 4.10 on page 4-34)
(A model beds) or (refer to procedure 4.22 on page 4-68) (B
model beds), then go to “Final Actions” on page 2-11.
7. Replace the power supply (A model beds) (refer to procedure 4.3) or the
motor control P.C. board (B model beds) (refer to procedure 4.27).
This solves the problem.
Yes No
2
↓ → Contact Hill-Rom technical support.
3. Check that the adjustment of the release cable is correct (refer to procedure
4.6 on page 4-20).
The adjustment is correct.
Yes No
↓ → Adjust the release cable (refer to procedure 4.6 on page 4-20).
4. Replace the head section motor (refer to procedure 4.5 on page 4-15) and
release cable subassembly (refer to procedure 4.6 on page 4-20).
This solves the problem.
Yes No
↓ → Contact Hill-Rom technical support.
2. Check the steering function. Set the brake/steer bar to the steer position,
push the bed lengthways, and check that the foot left caster locks in the
lengthways direction of the bed.
The caster locks in the lengthways direction.
Yes No
↓ → Replace the caster (refer to procedure 4.13 on page 4-44). If this
solves the problem perform the “Final Actions” on page 2-11,
otherwise contact Hill-Rom technical support.
NOTES:
Introduction
The Theories of Operation for the CareAssist™ Bed are as follows:
• Mechanical System
• Electrical System (A model beds)
• Electrical System (B model beds)
3
The interaction between these subsystems lets all the operational functions be
used.
Mechanical System
The mechanical subsystem consists of two connected systems:
• Bed frame
• Sleep surface
Hilow System
WARNING:
The column must be replaced if it becomes defective. The column must
never be dismantled. Failure to observe this warning could result in
personal injury.
The hilow system features two columns each composed of three sections that
slide in relation with each other by a pantograph system. This system is driven
by a fixed motor installed at the base of the mechanism. This system lets the
unit travel up or down 15.7" (400 mm). Upper and lower end of travel (EOT)
switches that are backed up to make sure of electrical security. The system
does not require any special maintenance during its designed period of use.
However, the columns can be replaced in the event of a failure, but should
never be dismantled or repaired except by Hill-Rom personnel.
Casters
The bed has four casters: three brake and one brake/steer. Each caster has a 30°
inclination activation mechanism and three positions:
• Brake caster—Brake/Neutral/Neutral
• Steer caster—Brake/Neutral/Steer
The brake/steer caster is located on the left side, foot end. Antistatic brake
casters are standard for the remaining caster positions. They are identified by a
3
yellow dot on the tire.
Braking
The braking system comprises four brake casters that are controlled by the
brake/steer bar. The bar consists of two hexagonal bars with a connecting bar
situated in the right-hand box-section of the lower frame. The brake/steer bar is
directly mounted on the foot end hexagonal bar of the bed.
Steering
The left foot end caster is the steering caster. It is controlled by the brake/steer
bar.
Description Specification
Type CareAssist™ Bed
Type of protection against electric shock Class I
(according to IEC 60601-1)
Degree of protection against electric shock Type B
(according to IEC 60601-1)
Protection against harmful ingress of water IPX4
(according to IEC 60529)
Intermittent operation 10% 3 min/30 min
Voltage 120 V (+10%, -15%) AC
voltage and earth
Frequency 60 Hz
Maximum power load 4.2 A
Refer to the schematic diagram at the end of this section to obtain specific
information for the cabling.
The bed is operated by a power supply unit. This unit controls the electrical
functions of the bed: hilow, head section and knee section with automatic
contour, and foot section.
Power Stage
The power stage is composed of a 120 V AC input transformer, protected at the
input by 2 antisurge fuses of 2.2 A and an internal thermal cutout that cannot
be reset. There is one non-fused secondary for the main control board and 1
fused secondary for the rest of the circuitry both followed at the outputs by
bridge rectifiers and filter capacitors. The value of this antisurge fuse is 1 A.
The non-fused secondary, through the bridge rectifier and filter capacitor,
produces an output voltage VDC1 which is approximately 24 V DC and
unregulated. The fused secondary produces a soft regulated 12 V DC for use in
the SideCom® Communication System P.C. board and Bed Exit System. The
available power enables several functions to be controlled simultaneously.
Function Control
The motors are controlled by relays. In order to protect these relays from
electrical sparks during switching, the second FET transistor checks the delay
power when a function is activated, thus allowing the switches to close with no
voltage. In the same way, when a function control is released, the FET
transistor cuts the power before the contact opens.
Battery Circuit
The bed is equipped with battery backup. This battery backup is internal to the
power supply unit. The charging circuit, situated in the power supply unit, has
current limiting according to the level of the battery charge. Charging begins as
soon as the unit is switched on. At the end of charging, the load current is very
low.
When the power supply is disconnected from the AC power, the battery
backup is activated by pressing the Battery control on the caregiver controls.
The battery charge level is controlled. In this way, a flashing indicator and an
audible low battery alarm by buzzer are activated during movement if the
battery voltage drops (this is a continuous beep of several seconds that is heard
each time a control is pressed).
After connecting to the power outlet, the indicator continues to flash until the
battery has charged.
Battery operation is automatically stopped 30 seconds after the end of the last
movement.
Battery Characteristics
Table 3-3. Battery Backup Characteristics
Description Specification
Components 2 12 V batteries
Capacity of each battery 1.2 Ah
Voltage 24 V DC
Maximum charging time 12 hours
Discharging time (storage), battery Minimum 10 days
connected to bed
Discharging time (storage), battery not 3 months
connected to bed
The batteries are a sealed lead acid and are connected in series on the main
3
board and protected by a 10 Amp fuse on the connection wires.
The battery type has no memory effect and can be charged regularly without
any particular constraints.
The capacity of the battery backup after full charging is at least three raise and
lower cycles for hilow, head section, knee section and foot section with the
maximum safe working load 400 lb (182 kg) on the bed.
Lockout management
Patient pendant or caregiver siderail control unit interfacing
Indication of Bed Not in Low Position (see “Detection of Bed Not in Low
Position” on page 3-9)
Indication of Bed Connected to AC Power but Brakes not Applied (see
“Detection of Bed Connected to AC Power but Brakes not Applied” on page
3-16)
All of these functions are integrated into a complex programmable logic device
(CPLD).
Patient Pendant
The patient pendant controls operate at 3.3 V DC supplied by the main bed
control and integrate three main principles:
• Filtering and amplification of inputs for electrostatic protection and signal
formatting
• Multiplexing for the management of the controls
• Demultiplexing for the control of the indicators
Data from the controls and indicators are transmitted in series (link type - SPI)
so as to reduce the number of wires. The clock signal that starts these
exchanges is controlled by the main bed control P.C board. All the data is sent
in tens of milliseconds. When the system is in stand-by mode, the power is cut
off and the clock signal is no longer controlled. The battery control is therefore
directly wired so as to have continuous operation.
Motor Characteristics
For motor characteristics, see table 3-4 on page 3-8.
Hilow column
Description Head motor Knee motor Foot motor
motor
Power 96 W (x2) 120 W 84 W 84 W
Voltage 18 to 40 V DC 18 to 40 V DC 18 to 40 V DC 18 to 40 V DC
Intermittent 10% 10% 10% 10%
service 3 min/30 min 2 min/18 min 6 min/60 min 6 min/60 min
Actuator/ 400 mm 239 mm 60 mm 45 mm
column travel ± 5 mm (15.75") ± 2 mm (9.4") ± 2 mm (2.36") ± 2 mm (1.77")
Retracted rod/ 445 mm 310 mm 325 mm
center distance ± 2mm (17.5") ± 2mm (12.2") ± 2mm (12.8")
The two hilow columns also have upper and lower internal EOT devices. For
safety reasons, they are both backed up at each end. If one of these safety
devices is activated, the column cannot be used; however the other functions
remain operational.
Automatic Contour
3
The power supply, when it received a command from the control unit, controls
the automatic contour (head section and thigh section) operation of the bed.
This function is carried out through an intermediate switch of the thigh section
actuator. When the Head Up function is activated from the horizontal position,
the head section and knee section rise until the intermediate internal
microswitch of the knee section actuator cuts its power at approximately 20°.
The head section actuator continues to operate until its upper EOT is reached
or the control is released. When the Head Down function is activated and the
head section reaches the low position, the knee section actuator continues
operating until its lower EOT is reached or the control is released.
Trendelenburg/Reverse Trendelenburg
The Trendelenburg/Reverse Trendelenburg is a fully electric function accessed
by the caregiver. The function is enabled by the Enable control on the
caregiver siderail controls. The principle involves controlling the two hilow
columns in opposing directions according to the position required, regardless
of the initial height and position of the sleep surface. As long as one of the two
controls remains pressed, the two columns continue to move until their internal
EOT (end of travel) is reached. The maximum inclination can be reached as
long as AC power or battery backup is available.
Description Specification
Type of protection against electric shock Class I
(according to IEC 60601-1)
Degree of protection against electric shock Type B
(according to IEC 60601-1)
Protection against harmful ingress of water IPX4
(according to IEC 60529)
Intermittent operation 10% 3 min/30 min
3
Maximum power load 4.2 A
Refer to the schematic diagram at the end of this section to obtain specific
information for the cabling.
The bed is operated by a power supply unit. This unit controls the electrical
functions of the bed: hilow, head section and knee section with automatic
contour, and foot section.
The functions may be locked out using the caregiver controls. When they are
not locked out, they be used by one of the following:
• Patient pendant
• Caregiver controls
Power Stage
The power stage is composed of an AC input transformer, protected at the
input by two antisurge fuses and an internal self-resetting thermal cutout.
There are two transformer options: one having a single winding primary,
120 V, 60 Hz for use in North America, and the other having two primaries,
multi-tapped, for use in 100, 110, 115, 127, 220, 230, and 240 V, 50/60 Hz and
120 V, 50 Hz applications.
Function Control
The motors are controlled by relays and/or FET transistors. In order to protect
these relays from electrical arcing during switching, the FET transistor
operates such that when a function is activated, the relay closes with no
current. In the same way, when a function control is released, the FET cuts the
power before the relay opens.
Battery Circuit
The bed is equipped with battery backup. This battery backup is internal to the
power supply unit. The charging circuit, has current limiting according to the
level of the battery charge. Charging begins as soon as the unit is connected to
AC power. At the end of charging, the load current is very low.
When the power supply is disconnected from the AC power, the battery
backup is activated by pressing the Battery control on the caregiver controls.
The battery charge level is controlled. In this way, a flashing indicator and an
audible low battery alarm by buzzer are activated during movement if the
battery voltage drops (this is a continuous beep of several seconds that is heard
each time a control is pressed).
After connecting to the power outlet, the indicator continues to flash until the
battery has charged.
Battery operation is automatically stopped 30 seconds after the end of the last
movement.
Battery Characteristics
Table 3-9. Battery Backup Characteristics
Description Specification
Components two 12 V, 1.2 Ah batteries
Voltage 24 V DC
Maximum charging time 12 hours
Discharging time (storage), battery Minimum 10 days
connected to bed
Discharging time (storage), battery not 3 months
connected to bed
The batteries are sealed lead acid, connected in series, and protected by a 10 A
fuse on the connection wires.
The battery type has no memory effect and can be charged regularly without
any particular constraints.
The capacity of the battery backup after full charging is at least three raise and
lower cycles for hilow, head section, knee section and foot section with the
maximum safe working load 400 lb (182 kg) on the bed.
Lockout management
Patient Pendant
The patient pendant controls operate at 3.3 V DC supplied by the motor control
P.C. board and integrate three main principles:
• Filtering and amplification of inputs for electrostatic protection and signal
3
formatting
• Multiplexing for the management of the controls
• Demultiplexing for the control of the indicators
Data from the controls and indicators are transmitted in series (link type - SPI)
so as to reduce the number of wires. The clock signal that starts these
exchanges is controlled by the motor control P.C board. All the data is sent in
tens of milliseconds. When the system is in stand-by mode, the power is cut off
and the clock signal is no longer controlled. The battery control is therefore
directly wired so as to have continuous operation.
Motor Characteristics
Hilow column
Description Head motor Knee motor Foot motor
motor
Power 96 W (x2) 120 W 84 W 84 W
Voltage 18 to 40 V DC 18 to 40 V DC 18 to 40 V DC 18 to 40 V DC
Intermittent 10% 10% 10% 10%
service 3 min/30 min 2 min/18 min 6 min/60 min 6 min/60 min
Actuator/ 400 mm 239 mm 60 mm 45 mm
column travel ± 5 mm (15.75") ± 2 mm (9.4") ± 2 mm (2.36") ± 2 mm (1.77")
Retracted rod/ 445 mm 310 mm 325 mm
center distance ± 2mm (17.5") ± 2mm (12.2") ± 2mm (12.8")
The two hilow columns also have upper and lower internal EOT devices. For
safety reasons, they are both backed up at each end. If one of these safety
devices is activated, the column cannot be used; however, the other functions
remain operational.
Automatic Contour
The motor control P.C. board, upon receiving a command from the control
unit, controls the automatic contour (head section and knee section) operation
of the bed. This function is carried out through an intermediate switch of the
knee section actuator. When the Head Up function is activated while in the
horizontal position, the head section and knee section rise until the
intermediate internal microswitch of the knee section actuator cuts its power at
approximately 20°. The head section actuator continues to operate until it
reaches its upper EOT or its control is released. When the Head Down function
is activated and the head section reaches the low position, the knee section
actuator continues operating until it reaches its lower EOT or its control is
released.
Trendelenburg/Reverse Trendelenburg
The Trendelenburg/Reverse Trendelenburg is a fully electric function accessed
by the caregiver. The function is enabled by the Enable control on the
caregiver siderail controls. The principle involves controlling the two hilow
columns in opposing directions according to the position required, regardless
of the initial height and position of the sleep surface. As long as one of the two
controls remains pressed, the two columns continue to move until their internal
EOT is reached. The maximum inclination can be reached as long as AC
3
power or battery backup is available.
The scale P.C. board has a single chip analog solution specially designed for
scale transducers. The load beams are resistive bridge sensors excited by a
regulated DC voltage. This regulated voltage is supplied to the beams and to
the analog to digital converter (ADC) reference input to minimize the effects
of variations.
Each beam is connected to the differential input on the ADC. The multiplexer
and gain (up to 128) are internally configured through software. The ADC has
the ability to detect open or shorted sensors. It also has a second stage sigma-
delta converter. The serial interface is SPI compatible and transfers between
the micro controller on the scale P.C. board and the microcontroller on the
display P.C. boards.
The Scale/Bed Exit System user interface consists of 5 seven segment LED
digits, LED indicators, and switches to start scale and bed exit functions. The
scale P.C. board sends controller area network (CAN) messages to the scale
display P.C. board which, illuminates the LEDs. The scale display P.C. board
sends CAN messages to the scale P.C. board to start activation of the switches
in the user interface.
All of the scale functions are done by microcontrollers on the scale P.C. board.
These include:
• Weigh
• Zero scale
• Add/Delete items
• Select LB/KG display mode
The Bed Exit System functions are shared between the scale P.C. board and the
scale display P.C. board. These include:
• Arm/Disarm bed exit
• Alarm tone select
• Alarm volume select
• Alarm
If the Bed Exit System is armed, AC power removed, and the bed equipped
with the SideCom® Communication System, the display P.C. board will cause
the SideCom® Communication System to send a nurse call. If the Bed Exit
System is armed and the display P.C. board loses communication with the
scale P.C. board, the display P.C. board will cause the SideCom®
Communication System to send a nurse call.
When the bed is unplugged from AC power, neither scale or the Bed Exit
System will function.
Adjustment Procedures
• Tape
• (2) 4" x 4" x 2' piece of wood
• Rags
• Isopropyl alcohol
• Wire cutters
1
• ViceGrip®
• Antistatic strap
• String 10' (305 cm)
• Needle nose pliers
• Small-bladed screwdriver (2)
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
4. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
5. For the foot section, grip the section (A) on either end and remove it from
the bed (see figure 4-1 on page 4-4).
4
6. For the thigh section, grip the section (B) on either end and remove it from
the bed.
8. For beds equipped with bed exit, remove the tape switches (refer to
procedure 4.20).
d. Remove the screw (E) securing the CPR handle (F) to the CPR mount
bracket (I).
e. Remove the CPR handle (F) from the CPR cable (G).
f. Loosen, do not remove, the jam nut (H) securing the CPR cable (G) to
the CPR mount bracket (I).
g. Repeat step d through step f for the other side.
h. Remove the E-ring (L) from one side of the lower hinge point (M).
i. Remove the plastic washer (N) from the lower hinge point (M).
j. Repeat step h and step i for the other side.
k. Move the head section to one side to disengage the hinge bracket (O)
from the hinge point (M).
l. Move the head section to the other side to disengage the hinge bracket
(O) from the hinge point (M).
m. Slide the head section (D) toward the head end of the bed to disengage
it from the head section frame.
n. Remove the head section (D) from the bed.
Replacement
1. Do the removal procedure in reverse order.
4
NOTE:
The seat section can only be installed when the head section is in the full up
position.
2. Make sure the CPR cable is adjusted correctly (refer to procedure 4.6)
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds so the
battery will time out.
5. Remove the two screws (A) and washers (B) from the power supply cover
(C) (see figure 4-2 on page 4-7).
7. Disconnect the power cord connector (D) from the fuse board (E).
8. Remove the screw (F) securing the power cord ground wire (G) to the bed
frame (H).
9. Remove the two screws (I) securing the power cord strain relief (J) to the
bed frame (H).
Replacement
1. Perform the removal procedure in reverse order.
3. Raise and lower the bed fully to make sure the power cord expands and
contracts without any problems.
4. Check that the earth continuity of the accessible metal parts is correct (see
“Leakage current” on page 6-5 in the Preventive Maintenance Schedule).
SHOCK HAZARD:
An abnormally high leakage current is symptomatic of a degradation
within the AC power cable and the power supply. Personal injury could
result if the value of the leakage current is above 500 microamperes.
5. Check that the leakage current of the bed is compliant (see “Leakage
current” on page 6-5 in the Preventive Maintenance Schedule).
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) and washers (B) securing the cover (C) to the
power supply (D) (see figure 4-3 on page 4-9).
6. Note the routing and connection of the cables in the power supply.
7. Disconnect the AC power cable (E) from the fuse board (F) (see figure 4-4
on page 4-10).
8. Disconnect the cables (H) from the power supply P.C. board (I).
9. Disconnect the two ground wires (G) from the fuse board (F).
10. Disconnect the battery cable (J) from power supply P.C. board (I).
11. Disconnect the transformer power supply cable (L) from the interface P.C.
board cable.
13. Remove the power supply (D) from the bed (see figure 4-3 on page 4-9).
Replacement
1. Do the removal procedures in reverse order. For cable connections, refer to
table 4-1 on page 4-11.
Connector
Marking Color Identification
number
0 #0 Red Not used
1a #1a Gray Right hand siderail
1b #1b Gray Left hand siderail
2 #2 Green Foot section motor
3 #3 Blue Head section motor
4 #4 Yellow Foot hilow column
5 #5 Brown Knee section motor
6 #6 Yellow Head hilow column
2. Check the dates shown on the batteries and replace them if they are older
than three years.
3. Check that the earth continuity of the accessible metal parts is correct (see
4
“Leakage current” on page 6-5 in the Preventive Maintenance Schedule).
SHOCK HAZARD:
An abnormally high leakage current is symptomatic of a degradation
within the AC power cable and the power supply. Personal injury could
result if the value of the leakage current is above 500 microamperes.
4. Check that the leakage current of the bed is compliant (see “Leakage
current” on page 6-5 in the Preventive Maintenance Schedule).
NOTE:
It is required to change both batteries at the same time.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) and washers (B) from the power supply cover
(C) (see figure 4-5 on page 4-13).
6. Disconnect the battery cable (D) from the power supply P.C. board (E).
9. Disconnect the battery cables (D) and (G) from the batteries (F).
4
Replacement
CAUTION:
The polarities must be respected when installing the elements of the
battery backup. Disregarding this caution will permanently damage the
power supply P.C. board of the bed.
1. Perform the removal in reverse order, ensure the correct polarities are used
when connecting new batteries.
NOTE:
When connecting the batteries, the red wire must be connected last. A spark is
produced when it is connected, but is not dangerous for the power supply P.C.
board or the technical personnel.
2. Raise and lower the bed one time to make sure the battery is functioning
correctly.
NOTE:
A technician may lay on the bed during these tests.
NOTE:
The batteries should be fully recharged at the end of the test or if the low
battery signal is heard during the tests.
Removal
1. Set the brakes.
NOTE:
If the head section motor is defective, the CPR control can be used to raise the
head section.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
4. Unplug the bed from its power source. Let the bed sit for 60 seconds so the
battery will time out.
4
5. Remove the headboard.
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
e. Disconnect the battery cable from the power supply P.C. board.
f. Remove the power supply cover (S) from the bed.
9. On the power supply P.C. board, disconnect the head section motor
4
connector (number 3).
10. Remove the cable cover on the central rail and remove the head section
motor cable.
11. Using the CPR control, lower the head section to the flat position.
12. Disconnect the CPR cable (F) from the head motor actuator arm (G) (see
figure 4-8 on page 4-18).
13. Count the number of threads (E) on the CPR cable (F) that extend past the
cable mount (H) on the head motor (I).
14. Loosen the jam nut (J) on the CPR cable (F).
15. Remove the CPR cable (F) from the cable mount (H).
NOTE:
The CPR cable is threaded into the cable mount.
16. Remove the E-ring (K) from the pin (L) at the rod end of the motor (I).
17. Remove the E-ring (M) from the pin (N) at the body end of the motor (I).
18. While supporting the motor (I), remove the pin (L) .
19. While supporting the motor (I), remove the pin (N).
Replacement
CAUTION:
Ensure that the elastic found on the new motors is not wound around
the motor body before removing it (cut). If necessary, turn the motor
end rod in the appropriate direction to unwind this elastic. It is used to
indicate the correct position of the switches and internal end of travels
of the motor. The incorrect position of these elements may damage the
motor or the structure of the bed.
3. Activate the CPR function. The CPR function should activate when the
CPR handle is approximately half way through its travel.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
4. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
5. Remove the screw (A) securing the right CPR handle (B) to the CPR
mounting bracket (C) (see figure 4-9 on page 4-20).
6. Remove the right CPR handle (B) from the CPR cable (D).
7. Loosen the jam nut (not shown) on the CPR cable (D) at the CPR cable
mount bracket (C).
8. Remove the screw (E) securing the left CPR handle (F) to the CPR
mounting bracket (G).
9. Remove the left CPR handle (F) from the CPR cable (D).
10. Loosen the jam nut (not shown) on the CPR cable (D) at the CPR cable
mount bracket (G).
11. Loosen the two jam nuts (J) securing the head section CPR cables to the
adjuster bracket (K) (see figure 4-10 on page 4-21).
12. Loosen the jam nuts (L) securing the CPR cable (D) to the adjuster bracket
(K).
13. Disconnect the CPR cable (D) from the release mechanism (M).
14. Count the number of threads (O) on the CPR cable (D) at the CPR cable
mount (P).
15. Loosen the jam nut (Q) on the CPR cable (D) on the CPR cable mount (P).
16. Unscrew the CPR cable (D) from the CPR cable mount (P).
Replacement
WARNING:
Make sure the CPR cable is routed through the head section guide
bracket correctly: right side to right handle, left side to left handle.
Failure to do so could cause patient injury or equipment damage.
2. Make sure to route through the head section guide bracket correctly, right
side to right handle, left side to left handle.
Adjustment
1. To add tension in the cable, loosen the two jam nuts (L) and move the cable
sheath (R) toward the foot end of the bed to the point where the release
mechanism on the head motor begins to move from the neutral position
(see figure 4-10 on page 4-21).
2. To relieve tension in the cable, loosen the two jam nuts (L) and move the
cable sheath (R) toward the head end of the bed to the point where the
release mechanism on the head motor begins to move from the neutral
position.
3. Stop at the point where the release mechanism begins to move from the
neutral position.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
4. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
e. Disconnect the battery cable from the power supply P.C. board.
f. Remove the power supply cover (M) from the bed.
8. On the power supply P.C. board, disconnect the knee section motor
4
connector (number 5).
9. Remove the cable cover on the central rail and remove the knee section
motor cable.
10. Remove the knee section motor cable from the bed.
11. Remove the E-ring (E) from the rod end pin (F) (see figure 4-13 on page
4-26).
12. Remove the E-ring (G) from the motor end pin (H).
13. While supporting the motor (J), remove the rod end pin (F).
14. While supporting the motor (J), remove the motor end pin (H).
Replacement
1. Perform the removal procedure in reverse order.
CAUTION:
Ensure that the elastic found on the new motors is not wound around
the motor body before removing it (cut). If necessary, turn the motor
end rod in the appropriate direction to unwind this elastic. Indeed, it is
used to indicate the correct position of the switches and internal end of
travels of the motor. The incorrect position of these elements may
damage the motor or the structure of the bed.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
5. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out. 4
6. For A model beds, do as follows, otherwise go to step 7.
a. Remove the two screws (A) and washers (B) securing the power supply
cover (C) to the power supply (D) (see figure 4-14 on page 4-28).
b. Remove the power supply cover (C).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
e. Disconnect the battery cable from the power supply P.C. board.
f. Remove the power supply cover (N) from the bed.
8. On the power supply P.C. board, disconnect the foot section motor
4
connector (number 2).
9. Remove the cable cover on the central rail and remove the foot section
motor cable.
10. Remove the foot section motor cable from the bed.
11. Remove the E-ring (E) from the rod end pin (F) (see figure 4-16 on page
4-30).
12. Remove the E-ring (G) from the motor end pin (H).
13. Remove the pin (F) and two plastic spacers (I).
NOTE:
Note the position of the plastic spacers for installation later.
Replacement
1. Perform the removal procedure in the reverse order.
CAUTION:
Ensure that the elastic found on the new motors is not wound around
the motor body before removing it (cut). If necessary, turn the motor
end rod in the appropriate direction to unwind this elastic. Indeed, it is
used to indicate the correct position of the switches and internal end of
travels of the motor. The incorrect position of these elements may
damage the motor or the structure of the bed.
Removal
1. Set the brakes.
WARNING:
Do not work under an unsupported load. Install applicable temporary
supports. Failure to do so could cause personal injury or equipment
damage.
6. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
8. Disconnect the head hilow column cable from the power supply P.C.
board.
9. Remove the screw (A) securing the hilow column ground wire (B) to the
hilow column (L).
10. Remove the head end base frame cover (C) (see figure 4-17 on page 4-32).
11. Remove the four screws (D) and washers (E) securing the hilow column
(H) to the frame (G).
12. Slide the hilow column (H) towards the head end of the bed to remove it
from the guide rails (F).
13. Remove the hilow column (H) from the bed. frame (G).
Replacement
1. Position the new hilow column (H) on the bed frame (G) with the cable
facing the foot end of the bed.
2. Install the two plastic bearings (I) onto the new column.
3. Connect the cable to the power supply P.C. board (connector 6).
4. Disconnect the foot hilow column from the power supply P.C. board.
6. Use the Bed Hilow control, to raise the hilow column to the same height as
the bed.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
7. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
8. Connect the foot hilow column to the power supply P.C. board.
11. Operate the Trendelenburg function to make sure there is no tension on the
power supply cable of the column.
SHOCK HAZARD:
The equipotential link and the grounding of the metal parts of the bed
depend on the correct installation of the earth conductor and grounding
spring blades. Personal injury could result if the installation is not
correct and/or if the earth continuity is compromised.
4
12. Check that the earth continuity of the accessible metal parts is correct (see
“Leakage current” on page 6-5 in the Preventive Maintenance Schedule).
13. To make sure the CareAssist™ Bed operates correctly, do the “Function
Checks” on page 2-4.
Removal
1. Set the brakes.
WARNING:
Do not work under an unsupported load. Install applicable temporary
supports. Failure to do so could cause personal injury or equipment
damage.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
8. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
10. Disconnect the head hilow column cable from the power supply P.C.
board.
11. Install the Vice Grips® into the head hilow channel (see figure 4-18 on
page 4-35).
13. Remove the cable cover (B) from the central rail.
14. Disconnect the foot hilow column cable from the extension cable inside the
central rail.
15. Remove the cover (C) from the bottom of the foot hilow column (D).
17. Remove the four screws (F) and washers (G) securing the foot hilow
19. Remove the four screws (J) securing the hilow column (D) to the
21. Tilt the column (D) towards the foot end of the bed.
22. Slide the upper section to remove it from the guide rails (M).
Replacement
1. Position the new column (D) on the bed frame (H) with the two covers (C)
4. Connect the hilow column cable to the extension cable inside the central
rail.
5. Install the two plastic bearings (N) onto the new hilow column (D).
7. Operate the hilow function until the foot hilow column is the same height
as the bed.
SHOCK HAZARD:
The equipotential link and the grounding of the metal parts of the bed
depend on the correct installation of the earth conductor and grounding
spring blades. Personal injury could result if the installation is not
correct and/or if the earth continuity is compromised.
9. Check that the earth continuity of the accessible metal parts is correct (see
“Leakage current” on page 6-5 in the Preventive Maintenance Schedule).
10. To make sure the CareAssist™ Bed operates correctly, do the “Function
Checks” on page 2-4.
Removal
1. Set the brakes.
3. Raise the right, foot end siderail to the up and locked position.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
4. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
5. Remove the foot frame cover (A) (see figure 4-19 on page 4-39).
6. Remove the cover (B) from the bed right side of the foot hilow column (C).
7. Disconnect the cable (D) from the brake detection switch (F).
8. Open the cable track located under the foot crossbar on the frame, and
remove the brake detection switch cable.
9. Remove the screw (E) securing the brake detection switch (F) to the frame.
Replacement
1. Perform the removal procedure in reverse order.
Ratchet
Extension
13 mm wrench
10 mm hex key
8 mm pin punch
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
WARNING:
Use correct lifting techniques. Failure to do so can cause personal
injury or equipment damage.
5. Using a second person, lift the foot end of the bed and place a piece of
wood under the foot hilow column.
6. Using a second person, lift the head end of the bed and place a piece of
wood under head the hilow column.
7. Remove the foot frame cover (A) (see figure 4-20 on page 4-41).
9. From the left side of the bed, drive the brake bar hex rod (C) out of the left
foot caster.
NOTE:
Driving the brake bar hex rod out of the caster will break the white plastic
bushing on the right side brake/steer tube.
11. From the left side of the bed, drive the brake bar hex rod out of the left head
caster.
NOTE:
Driving the brake bar hex rod from the caster will break the white plastic
bushing on the right side brake/steer tube.
13. Remove the brake/steer link bar (E) from the frame.
14. Remove the plastic bushing (F) from the head and foot end brake bars.
Replacement
1. Insert the new bushings (F) into the brake/steer link bar (E).
NOTE:
The plastic bushings will be between the brake/steer link bar and the frame.
3. Align the head end brake bar with the bushing in the brake/steer link bar
(E).
4. Drive the head end brake bar into the brake/steer link bar (E), seating it
inside the bushing.
5. Align the foot end brake bar (D) with the bushing in the brake/steer link bar
(E).
6. Drive the foot end brake bar (D) into the brake/steer link bar (E), seating it
inside the bushing.
WARNING:
Use proper lifting techniques. Failure to do so can cause personal injury
or equipment damage.
9. Using a second person, remove the pieces of wood from under the hilow
columns.
10. To make sure the CareAssist™ Bed operates correctly, do the “Function
Checks” on page 2-4.
4.13 Caster
Tools required: 13 mm socket
Ratchet
Extension
10 mm hex key
Medium screwdriver
Scissor jack
Removal
NOTE:
The removal and replacement procedure is the same for both the head and foot
end casters.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
5. Using the scissor jack, raise the bed high enough to remove the caster .
NOTE:
When the caster disengages from the brake bar hex rod, it will fall from the
socket.
4
Replacement
NOTE:
Brake/steer casters have a black washer on top. Brake casters have a silver
plated washer on top.
1. Hold the new caster (F) so the window (I) with the cam (H) showing is
towards the foot end of the bed.
2. Insert the new caster (F) into the socket (G) so the window with the cam
showing is towards the foot end of the bed.
WARNING:
Ensure the caster is correctly positioned with respect to the movements
of the hexagonal bar. Failure to do so may cause the brake and steer
system to malfunction. Personal injury or equipment damage could
occur.
3. Install the new caster (F) and socket (G) into the bed (B) so the window
with the cam showing is towards the foot end of the bed.
5. Install the bolt (E) and nut (D) through the socket (G).
b. Rotate the cam inside the caster clockwise until the caster is braked.
e. Repeat as needed.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
NOTE:
The SideCom® Communication System P.C. board is mounted on the
interface box cover.
4. Remove the nine screws (A) securing the interface box cover (B) to the
interface box (C) (see figure 4-22 on page 4-48). 4
5. Lower the box cover (B).
7. Remove the six screws (F) securing the interface P.C. board (E) to the box
cover (B).
Replacement
1. Perform the removal procedure in reverse order.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
NOTE:
The interface P.C. board is mounted on the interface box cover.
4. Remove the nine screws (A) securing the interface box cover (B) to the
interface box (C) (see figure 4-23 on page 4-50).
7. Remove the six screws (E) securing the interface P.C. board (D) to the box
cover (B).
Replacement
1. Perform the removal procedure in reverse order.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) and washers (B) from the power supply cover
(C) (see figure 4-24 on page 4-52).
Replacement
1. Perform the removal procedure in reverse order.
NOTE:
Removing the switch panel from the siderail requires the installation of a new
switch panel.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
4. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
4
battery to time out.
CAUTION:
Use extreme care when removing the switch panel. Failure to do so
could cause damage to the seating area of the switch panel on the
siderail.
5. Starting on the bottom edge, remove the switch panel (A) from the siderail
(B) (see figure 4-25 on page 4-54).
6. Remove the six screws (C) securing the stiffener (D) to the siderail (B).
7. Disconnect the switch panel cable (E) from the siderail control P.C. board
(F).
8. Disconnect the siderail cables (G) from the siderail control P.C. board (F).
9. Remove the siderail P.C. board (F) from the stiffener (D).
WARNING:
Isopropyl alcohol is flammable and toxic to skin, eyes, and the
respiratory tract. Do not use near an open flame. Do not use in confined
areas. Personal injury may occur.
10. Using rags and isopropyl alcohol, clean the mounting area where the
switch panel (A) was located.
Replacement
1. Connect the new switch panel cable (E) to the siderail control P.C. board
(F).
2. Connect the siderail cables (G) to the siderail control P.C. board (F).
3. Install the siderail control P.C. board (F) onto the stiffener (D).
5. Install the six screws (C) to secure the stiffener (D) to the siderail (B).
6. Remove the wax paper backing from the new switch panel (A).
7. Install the new switch panel (A) onto the siderail (B).
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
Replacement
1. Make sure the new pendant has the same functions as the pendant being
replaced.
NOTE:
This procedure is for moving the pendant mount from one side of the bed to the
other side of the bed.
Removal
1. Set the brakes.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
2. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Cut and remove the cable ties securing the pendant mount to the bed.
Replacement 4
1. Move the pendant mount to the opposite side of the bed.
Removal
NOTE:
The tape switch assembly consists of two tape switches connected to a single
cable, if one tape switch is inoperable, both tape switches must be replaced.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
4. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
6. Using the punch, punch out the center of the four rivets (A) securing the
switch (B) to the seat section (C) (see figure 4-26 on page 4-59).
8. Using the punch, punch out the center of the four rivets (A) securing the
switch (B) to the thigh section (D).
10. Remove the seat section (C) and knee section (D).
11. Remove the two cable covers (E) from the central rail.
a. Remove the nine screws (F) securing the interface box cover (G) to the
c. Disconnect the switch cable from the interface P.C. board (H).
a. Remove the two screws (J) from the retainers (L) (see figure 4-27 on
page 4-60).
c. Remove the three screws (M) that attach the power supply cover (N) to
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
e. Disconnect the battery cable from the power supply P.C. board.
f. Remove the power supply cover (N) from the bed.
g. Disconnect the switch cable from the interface P.C. board (P).
Replacement
NOTE:
The tape switch assembly consists of two tape switches connected to a single
cable that connects to the interface P.C. board.
1. Route the switch cable through the central rail toward the head end of the
bed to the interface box (A model beds) or power supply (B model beds).
4. Lay one switch (B) on the seat section (C) (see figure 4-26 on page 4-59).
5. Install four plastic rivets (A) to secure the switch (B) to the seat section (C).
7. Install four plastic rivets (A) to secure the switch (B) to the knee section
(D).
Removal
1. Set the brakes.
WARNING:
Do not work under an unsupported load. Install appropriate supports.
Failure to do so could cause personal injury or equipment damage.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
6. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
7. Remove the two screws (A) from the retainers (C) (see figure 4-28 on page
4-64).
9. Remove the three screws (D) that attach the power supply cover (E) to the
power supply (F).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
11. Disconnect the battery cable from the power supply P.C. board.
12. Remove the power supply cover (E) from the bed.
13. Disconnect the head hilow column cable from the power supply P.C.
board.
15. Remove the screw (H) that attaches the hilow column ground wire (I) to
the bed (J) (see figure 4-29 on page 4-65).
16. Remove the four screws (K) and washers (L) that attach the column (M) to
the bed (J).
17. Cut and remove the cable ties that attach the power cable, load cell cables,
and communication cable to the wire guides (N).
18. Remove the two screws (O) and washers (P) from the column cover (Q).
20. Remove the screw (R) from the load cell bracket (S).
21. Move the column (M) towards the head end of the bed until the load cell
bracket (S) is clear of the weigh frame (T).
22. Remove the load cell bracket (S) and rollers (U) from the column (M).
23. Put the load cell bracket (S) and rollers (U) on the weigh frame (T).
Replacement
1. Put the new hilow column (M) in position on the bed (J). Make sure the
power cable towards the left side of the bed (J).
2. Connect the cable to the power supply P.C. board (connector 6).
3. Disconnect the foot hilow column from the power supply P.C. board.
5. Use the Bed Hilow control to raise the hilow column to the same height as
the bed.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
6. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
7. Connect the foot hilow column to the power supply P.C. board.
8. Put the mark on the power cord and communication cable so the edges of
the marks are even with the edge of the wire guide.
9. Put the load cell cables along the side of the power cord and
communication cable.
12. Operate the Trendelenburg function to make sure there is no tension in the
cables on the column.
Removal
1. Set the brakes.
WARNING:
Do not work under an unsupported load. Install appropriate supports.
Failure to do so could cause personal injury or equipment damage.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
8. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
9. Remove the two screws (A) from the retainers (C) (see figure 4-30 on page
4-69).
11. Remove the three screws (D) that attach the power supply cover (E) to the
4
power supply (F).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
13. Disconnect the battery cable from the power supply P.C. board.
14. Remove the power supply cover (E) from the bed.
1
15. Install the Vice Grips® into the head hilow channel.
16. Disconnect the column cable from the power supply P.C. board.
17. Tie the string around the end of the column cable.
18. Remove the cover (F) from the bottom of the foot hilow column (G) (see
figure 4-31 on page 4-70).
20. Remove the screw (I) that attaches the hilow column ground wire (J) to the
bed (K).
21. Remove the four screws (L) and washers (M) that attach the foot hilow
23. Remove the four screws (O) that attach the foot hilow column (G) to the
24. Remove the two screws (Q) and washers (R) that attach the column cover
26. Remove the screw (T) from the load cell bracket (U).
27. Mark the locations of where the cable ties, load cell cables, and the column
28. Cut and remove the cable ties that attach the load cell cables to the weigh
frame (V).
4
29. Move the column (G) towards the foot end of the bed (K) until the load cell
30. Remove the load cell bracket (U) from the column (G).
31. Put the load cell bracket on top of the weigh frame (U).
Replacement
1. Put the new foot hilow column (G) in position on the bed (K). Make sure
the two covers (F) are toward the foot end of the bed.
2. Remove the cover (F) on the base of the new column (G).
5. Pull the string so the column cable goes to the power supply.
8. Disconnect the head hilow column from the power supply P.C. board.
10. Use the Bed Hilow control, to raise the hilow column to the same height as
the bed.
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
11. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
12. Connect the head hilow column to the power supply P.C. board.
15. Operate the Trendelenburg function to make sure there is no tension in the
NOTE:
The load cell bracket assembly must be changed as an assembly. The
individual load cells are not replaceable.
Removal
1. Set the brakes.
WARNING:
Do not work under an unsupported load. Install appropriate supports.
Failure to do so could cause personal injury or equipment damage.
5. Place the jack stands under the articulating frame near the load cells being
replaced.
6. Remove the two screws (A) and washers (B) that attach the column cover
(C) to the column (D) (see figure 4-32 on page 4-74).
8. Remove the screw (E) that attaches the load cell bracket (F) to the column
(D).
9. Lower the bed until the load cell bracket (F) is clear of the column (D).
SHOCK HAZARD:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
10. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
11. Remove the two screws (H) from the retainers (J).
13. Remove the three screws (K) that attach the power supply cover (L) to the
power supply (M).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
4
15. Disconnect the battery cable from the power supply P.C. board.
16. Remove the power supply cover (L) from the bed.
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
18. Disconnect the load cells from the scale P.C. board (N).
NOTE:
P1 and P4 for head end load cells, and P2 and P3 for foot end load cells.
19. Tie the string around the ends of the load cell cables.
20. Cut and remove the cable ties that attach the load cell cables to the bed.
21. Remove the two screws (O) that attach the two wire guides (P) to the load
cell bracket (F).
23. Remove the load cell cables from the column (D).
24. Remove the load cell bracket (F) from the weigh frame (G).
Replacement
1. Do the Removal procedure in reverse order.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
5. Slide the pod (B) towards the head end of the siderail (C).
8. Remove the four screws (D) from the back of the pod (B).
11. Disconnect the cable (F) from the P.C. board (G).
Replacement
1. Do as follows to the new pod:
a. Remove the four screws (D) from the back of the pod (B).
2. Put the new pod (B) in position next to the cable (F) that comes out of the
siderail (C).
6. Install the four screws (D) to attach the pod back (E) to the pod (B).
7. Install the weldment (H), removed earlier, into the spring (I).
9. Slide the pod (B) towards the foot end of the siderail (C) so the weldment
4
10. Install the spacer (A).
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
5. Slide the pod (B) towards the head end of the siderail (C).
7. Slide the pod (B) towards the foot end of the siderail (C), and remove it
from the siderail (C).
8. Remove the four screws (D) from the back of the pod (B).
11. Disconnect the cable (F) from the P.C. board (G).
12. Disconnect the P.C. board (G) from the switch panel cable (H).
Replacement
1. Connect the new P.C. board (G) to the switch panel cable (H).
5. Install the four screws (D) to attach the pod back (E) to the pod (B).
7. Slide the pod (B) toward the foot end of the siderail (C) so the weldment
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the four screws (A) that attach the battery cover (B) to the power
supply (C) (see figure 4-35 on page 4-84).
Replacement
1. Do the removal procedure in reverse order.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) from the retainers (C) (see figure 4-36 on page
4-86).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
8. Disconnect the battery cable from the power supply P.C. board.
9. Remove the power supply cover (E) from the power supply (F).
10. Note the positions of the cables connected to the power supply P.C. board
(G).
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
12. Disconnect the cables from the power supply P.C. board (G).
14. Pull the P.C. board (G) off the standoff (H).
15. Remove the P.C. board (H) from the power supply (F).
Replacement
1. Do the removal procedure in reverse order.
2. Do the “Function Checks” on page 2-4.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) from the retainers (C) (see figure 4-37 on page
4-88).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
8. Disconnect the battery cable from the power supply P.C. board.
9. Remove the power supply cover (E) from the power supply (F).
10. Note the positions of the cables connected to the communication P.C.
board (G).
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
12. Disconnect the cables from the communication P.C. board (G).
15. Remove the P.C. board (G) from the power supply (F).
Replacement
1. Do the removal procedure in reverse order.
2. Do the “Function Checks” on page 2-4.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) from the retainers (C) (see figure 4-38 on page
4-90). 4
5. Remove the retainers (C).
6. Remove the three screws (D) that attach the power supply cover (E) to the
power supply (F).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
8. Disconnect the battery cable from the power supply P.C. board.
9. Remove the power supply cover (E) from the power supply (F).
10. Note the positions of the cables connected to the line filter (G).
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
13. Remove the two screws (H) that attach the line filter (G) the bracket (H).
Replacement
1. Do the removal procedure in reverse order.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) from the retainers (C) (see figure 4-39 on page
4-92).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
8. Disconnect the battery cable from the power supply P.C. board.
10. Note the positions of the cables connected to the scale P.C. board (G).
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
12. Disconnect the cables from the scale P.C. board (G).
14. Pull the P.C. board (G) off the standoffs (H).
15. Remove the P.C. board (G) from the power supply (F).
Replacement
1. Do the removal procedure in reverse order.
2. Calibrate the scale, refer to “Scale Calibration” on page 6-11.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) from the retainers (C) (see figure 4-40 on page
4-94).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
8. Disconnect the battery cable from the power supply P.C. board.
9. Remove the power supply cover (E) from the power supply (F).
10. Note the positions of the cables connected to the interface P.C. board (G).
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
12. Disconnect the cables from the interface P.C. board (G).
14. Pull the P.C. board (G) off the standoffs (H).
15. Remove the P.C. board (G) from the power supply (F).
Replacement
1. Do the removal procedure in reverse order.
2. Do the “Function Checks” on page 2-4.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) from the retainers (C) (see figure 4-41 on page
4-96).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
8. Disconnect the battery cable from the power supply P.C. board.
10. Note the positions of the cables connected to the transformer (G).
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
13. Remove the screw (H) that secures the transformer (G) the power supply
(F).
14. Remove the washer (I) and spacers (J) from the transformer (G).
Replacement
1. Do the removal procedure in reverse order.
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
4. Remove the two screws (A) from the retainers (C) (see figure 4-42 on page
4-98).
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
8. Disconnect the battery cable from the power supply P.C. board.
10. Cut and remove the wire tie (G) that attaches the power cord (H) to the
bracket (I).
11. Cut and remove the cable ties that attach the power cord (H) to the wire
guide (J).
12. Remove the two screws (K) that attach the power cord (H) to the bed (L).
13. Remove the power cord (H) from the bed (L).
Replacement
1. Put the mark (M) on the power cord (H) so the edge of the mark (M) is
even with the edge of the wire guide (J).
2. Put the other mark (M) on the power cord (H) so the edge of the mark (M)
is even with the enclosure entrance on the power supply (F).
Removal
1. Set the brakes.
WARNING:
Unplug the bed from its power source. Failure to do so could cause
personal injury or equipment damage.
3. Unplug the bed from its power source. Let the bed sit for 60 seconds for the
battery to time out.
NOTE:
The power supply cover is connected to the power supply P.C. board by the
battery cable.
e. Disconnect the battery cable from the power supply P.C. board.
f. Remove the power supply cover (E).
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could cause component damage.
g. Put on the antistatic strap.
h. Remove the both fuses (G) from the interface P.C. board (H) (see
4
figure 4-44 on page 4-102).
NOTE:
Replace both fuses at the same time.
NOTE:
The large fuses are for the 100 V through 127 V beds. The smaller fuses are for
the 220 V through 240 V beds.
NOTE:
The fuse holder is a twist type fuse holder.
d. Remove the fuse (L).
Replacement
1. Do the removal procedure in reverse order.
NOTES:
Parts List
(812) 934-8472
Terms:
• Net 30 days
• F.O.B. Batesville, IN
• Prepaid shipping charges added to invoice
• All orders shipped ground transportation unless specified
NOTE:
To eliminate possible delays or incorrect billings, do not return any items
without a Return Material Authorization (RMA) number. When a return is
requested, an RMA packet is included with each order. This packet includes an
RMA number, instructions, and a shipping label. If an RMA number is not
available, obtain one by phoning Hill-Rom Technical Support at
(800) 445-3720.
Exchange Policy
The following are policies for in-warranty and out-of-warranty exchanges from
Hill-Rom.
In-Warranty Exchanges
In some cases, Hill-Rom will request that parts/products be returned for
inspection. When this occurs, you are expected to return parts/products within
30 days of receipt of the exchange part. If you fail to return the inoperative
parts/products within the 30 day period, Hill-Rom will invoice your facility for
NOTE:
requests to be returned.
In some cases, the invoice accompanying the parts will show the full selling
price (only for internal use at Hill-Rom). Do not confuse this price with your
price.
Do not return any parts without an RMA number. When parts/products have
been requested to be returned, Hill-Rom will include an RMA packet with the
Out-of-Warranty Exchanges
You are expected to return the inoperative parts/products within 30 days of
receipt of the exchange part. Hill-Rom will include an RMA packet with the
parts/products shipment. If an RMA number is not available, obtain one by
phoning Hill-Rom Technical Support at (800) 445-3720. Hill-Rom will
5
invoice your facility for the full selling price of the parts/products. Upon return
of the inoperative parts/products, Hill-Rom will issue a credit to your facility
for the difference between the exchange price and the full selling price of
the parts/products.
Warranty
HILL-ROM COMPANY, INC.
LIMITED WARRANTY
Hill-Rom Company, Inc. (Hill-Rom) has a long tradition of providing superior products and service to our customers. Our
goal is “Total Customer Satisfaction”. In that spirit, Hill-Rom is proud to offer the following warranty.
GENERAL WARRANTY (APPLICABLE UNLESS A SPECIFIC WARRANTY IS LISTED)
Hill-Rom warrants to the original purchaser that its products and replacement parts shall be free from defects in material and
workmanship for a period of one (1) year from date of delivery. Hill-Rom’s obligation under this warranty is expressly
limited to supplying replacement parts and/or service for, or replacing, at its option, any product which is, in the sole
discretion of Hill-Rom, found to be defective. In addition to the foregoing one year warranty, Hill-Rom warrants to the
original purchaser that the frame and welds on its products will be free from structural defects for the life of the product. Any
product upgrade or modification initiated by Hill-Rom does not affect the original product warranty.
SPECIFIC WARRANTIES
MATTRESS WARRANTIES
Hill-Rom warrants to the original purchaser that its mattress product shall be free from defects in material and workmanship
for a period of two (2) years from date of delivery. However, electro mechanical mattress components (compressors, valves,
printed circuit boards, hoses, and couplers) are covered by the general one (1) year warranty.
EXPENDABLES WARRANTIES
A sixty (60) day limited warranty from date of delivery applies to expendable parts such as cushions, coverlets, software
diskettes, locator badge batteries, dome light incandescent bulbs, overhead fluorescent tubes, heating elements, temperature
probes, filter sheets, and microspheres. This warranty is limited to replacement of the parts covered.
TO OBTAIN PARTS AND SERVICE
In the United States, call Hill-Rom Technical Support Department at (800) 445-3720, Monday through Friday. In Canada,
call Hill-Rom Technical Support Department at (800) 267-2337, Monday through Friday. Outside the United States and
Canada, call your authorized Hill-Rom Distributor. In order to expedite service, we request you furnish the following
information: customer identification number, product model number, serial number, and description of problem. A qualified
specialist will provide, via telephone (United States and Canada), or FAX (Outside the United States and Canada),
troubleshooting assistance for facility personnel and provide necessary parts to make repairs. If troubleshooting determines
the need for on-site technical service, a qualified service representative will be dispatched. Replacement of non-technical
items will be the responsibility of the customer. If requested by Hill-Rom, products or parts for which a warranty claim is
made shall be returned prepaid to Hill-Rom’s factory.
OUT OF WARRANTY EXCHANGE POLICY
After the expiration of the original warranty, upon request, Hill-Rom will ship as a replacement, components such as selected:
motors and printed circuit boards, for like units returned to Hill-Rom by the original purchaser at a substantial savings. Please
call Hill-Rom Technical Support Department for current pricing.
PARTS AVAILABILITY POLICY
Hill-Rom will offer parts for new and remanufactured products for ten (10) years from date of sale; for communications
Note: Some original component parts and assemblies may not be available; functional equivalents may be substituted.
THE FOREGOING WARRANTIES ARE EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES
AND IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF
MERCHANTABILITY AND FITNESS OF PURPOSE. HILL-ROM’S OBLIGATION UNDER THESE
WARRANTIES SHALL NOT INCLUDE ANY LIABILITY FOR LOSS OF PROFITS, DIRECT, INDIRECT OR
CONSEQUENTIAL DAMAGES OR DELAYS. Some states, provinces, or countries do not allow the exclusion or
limitation of incidental or consequential damages, so the above exclusion or limitation may not apply. Any improper or
negligent use, any alterations or repairs not in accordance with Hill-Rom’s manuals or performed by others in such manner as
in Hill-Rom’s sole judgment affects the product materially and adversely, shall void these warranties. These warranties do not
cover failures due to misuse, abuse, neglect, or lack of routine maintenance. No employee or representative of Hill-Rom is
authorized to change these warranties in any way or grant any other warranty unless in writing and signed by a Hill-Rom
officer. These warranties provide specific legal rights; but, there may be other available rights, which vary from state to state,
province to province, or country to country.
Revised July 6, 2001
Hill-Rom Company, Inc., 1069 State Route 46 E, Batesville, IN 47006-9167
NOTES:
NOTES:
NOTES:
17
(1170)
CA0001 (1170) 1 Weldment, intermediate frame
5
18 40421 (1170) 6 Mount, cable tie
19 72854 (1170) 1 Backrest weldment
20 S30555K (1170) 1 Cover, short cable tray
21 S30548K (1170) 1 Cover, long cable tray
22 71494 (1170) 2 Weldment, drainage hook
23 VI0020 (1170) 1 Screw
24 VI0361 (1170) 1 Nut
25 S11002A (1170) 1 Pivot pin, CPR
26 CA0020 (1170) 1 Bracket
NOTES:
NOTES:
20
71061 (1170)
71656 (1170)
1
1
Siderail assembly, foot end, lh
Label
5
21 QD1653A 1 Level ball
(1170)
22 72792 (1170) 1 Spring, siderail, foot
NOTES:
NOTES:
15
(1170)
VI0361 (1170) 4 Nut
5
16 QD1545A 4 Bumper
(1170)
17 132188 (1170) 1 Weldment, intermediate frame
18 72274 (1170) 4 Mount, cable tie
19 72854 (1170) 1 Backrest weldment
20 S30555K (1170) 1 Cover, short cable tray
21 S30548K (1170) 1 Cover, long cable tray
22 71494 (1170) 2 Weldment, drainage hook
23 VI0609A (1170) 2 Screw
NOTES:
NOTES:
20
71061 (1170)
71656 (1170)
1
1
Siderail assembly, foot end, lh
Label
5
21 QD1653A 1 Level ball
(1170)
22 72792 (1170) 1 Spring, siderail, foot
NOTES:
NOTES:
WARNING:
Follow the product manufacturer’s instructions. Failure to do so could
result in personal injury or equipment damage.
SHOCK HAZARD:
Unplug the unit from its power source. Failure to do so could result in
personal injury or equipment damage.
SHOCK HAZARD:
Do not expose the unit to excessive moisture. Personal injury or
equipment damage could occur.
CAUTION:
Do not use harsh cleaners, solvents, or detergents. Equipment damage
could occur.
General Cleaning
Clean the unit with a lightly dampened cloth and ordinary disinfectants. Do not
use excessive liquid.
Steam Cleaning
Do not use any steam cleaning device on the CareAssist™ Bed. Excessive
moisture can damage mechanisms in this unit.
6
Disinfecting
When there is visible soilage and between patients, we recommend that you
disinfect the unit with a tuberculocidal disinfectant. (For customers in the US,
the disinfectant should be registered with the Environmental Protection
Agency.)
Do not spray the disinfectant onto power cord terminals or bed control
switches.
Component Handling
CAUTION:
To help prevent component damage, make sure your hands are clean,
and only handle the P.C. board by its edges.
CAUTION:
When handling electronic components, wear an antistatic strap. Failure
to do so could result in component damage.
CAUTION:
For shipping and storage, place the removed P.C. board in an antistatic
protective bag. Equipment damage can occur.
P.C. Board
When servicing the P.C. board, follow good handling practices. Mishandling a
P.C. board can cause the following:
• P.C. board damage
• Shortened P.C. board life
• Unit malfunctions
Lubrication Requirements
There are no lubrication requirements for the CareAssist™ Bed.
Preventive Maintenance
WARNING:
Only the facility-authorized personnel should service the CareAssist™
Bed. Servicing performed by unauthorized personnel could result in
personal injury or equipment damage.
Every 5 years, the bed exit tape switches should be replaced by a facility-
authorized maintenance personnel.
Function Procedure
Overall condition, Check the overall condition of the bed.
frame and welded Check that the structure and welded assemblies are in good working
assemblies condition and that there are no impacts or corrosion.
Do any necessary repairs or paint retouches, replace parts if
necessary.
Observe the symmetry of the bed and check that the bed frame and
base are not twisted. Do any necessary repairs or paint retouches.
Check that all labels are installed and can be read, replace them if
necessary.
Headboard and Check the appearance, attachment and safety of the headboard and
footboard footboard. Replace if necessary.
AC power cable Check that there are no cuts, scrapes, or breaks in the AC power
cable or connector. Replace the AC power cable if necessary.
Leakage current Unplug the bed from its power source.
Connect the bed to the safety tester and connect this device to the
power outlet.
Check that the bed is correctly supplied by the test device: control
unit LEDs lit (unlock the functions if necessary).
Measure the leakage current measures without operating any
function.
The value must be less than 100 µA.
Check the AC power cable and power supply unit if the value is
outside of the specifications (0.2 Ohms).
Replace the AC power cable or power supply unit if necessary.
Patient pendant Disconnect the patient pendant and check the condition of the con
nector. Then reconnect or replace the pendant.
Test each of the buttons to check that they activate the correct func
tion and that they do not work intermittently by pressing each button
6
for several seconds. Each movement must be continuous.
Replace the pendant if necessary.
Function Procedure
Hilow columns Inspect the column assembly for the presence and tightness of the
attachment screws and the snap ring at the bottom of the column.
Repair as necessary.
Fully raise and lower the bed frame one time.
Make sure there is no friction or abnormal noises and that no audible
overload indication can be heard during the movement.
Make sure the Bed Not Down position indicator illuminates on the
control pendant and goes out when the bed is in the low position.
Replace the defective column(s) in the event of a malfunction.
Troubleshoot in the event of a doubt.
Head section motor Inspect the actuator assembly. Make sure the pins and retaining clips
are intact and not missing.
Fully raise and lower the head section. Make sure there is no friction
or abnormal noises and that no audible overload indication can be
heard during the movement.
Repair as necessary.
Plastic sleep deck Fully raise and lower the head section.
Check for binding during the head section movement.
Make sure the hard surface and its drive system are in good
condition.
Make sure there is no friction or abnormal noises.
Replace any damaged parts.
Remove each hard surface (except the head section) and check its
condition.
Replace if necessary.
Function Procedure
CPR release Inspect the handles, cables and CPR mechanism on the head motor.
Make sure the screws are present and fully tightened.
Make sure the two CPR release handles, cable and CPR mechanism
on the head motor are in good condition.
Raise the head section to the high position, then activate one of the
CPR releases. Make sure t the head section lowers. Adjust the CPR
cable as necessary. Do the same tests on the other side of the bed.
Release the CPR handles and make sure the mechanism locks
correctly by operating the Head Up control for a few seconds. The
elevation movement must be observed.
Replace the head section motor as necessary.
Head section gas Inspect the assembly of the gas springs. Check that the screws are
springs present and fully tightened.
Make sure there is no oil on its shaft.
Fully raise the head section. Activate the CPR function. Make sure
the head section lowers rapidly to the intermediate position then
gradually to the low position.
Replace as necessary.
Knee section motor Inspect the knee section motor. Make sure the pins and retaining
clips are not missing.
Fully raise and lower knee section. make sure there is no friction or
abnormal noises and that no audible overload indication can be heard
during the movement.
Replace as necessary.
Automatic Contour Make sure the knee section rises up to mid-height when the head
section is raised from the low position, and that it lowers
automatically when the head section lowers and the head section
reaches the low position.
Replace the actuator or defective unit if necessary.
Preventive Maintenance
Chapter 6: General Procedures
Function Procedure
Foot section motor Raise the knee section to the mid-height position.
Inspect the foot actuator. Make sure the pins and retaining rings are
present.
Fully raise and the foot section.
Make sure there is no friction or abnormal noises and that no audible
overload indication can be heard during the movement.
Replace as necessary.
Dining Chair™ Make sure the Chair, head section, thigh section and foot section
Position functions are not locked out.
Activate the chair function and make sure the three sections to the
required position.
Return the sleep deck to the flat position.
Troubleshoot as required.
Trendelenburg/ Activate the Trendelenburg function.
Reverse Make sure there is no friction or abnormal noises and that no audible
Trendelenburg overload indication can be heard during the movement.
Activate the Reverse Trendelenburg function.
Make sure there is no friction or abnormal noises and that no audible
overload indication can be heard during the movement.
Replace as necessary.
Battery Make sure that the bed is unplugged from its power source.
Operate all the functions using the caregiver controls.
Charge the batteries or replace them if necessary.
Check the date indicated on the batteries, replace the battery if the
date is over three years.
Scale System Calibrate the scale. Refer to“Scale Calibration” on page 6-11.
Bed Exit System Ensure the Bed Exit System works properly.
Replace worn or defective parts as needed.
Function Procedure
Siderails Make sure the head and foot siderails are not bent or twisted.
Make sure the latch mechanism operates correctly. An audible click
must be heard.
Remove the siderail cover, and make sure the mounting screws are
tight.
Inspect the cable routing for pinching, binding, and damage.
Make sure all functions on the caregiver control operate correctly.
Repair or replace the siderail as necessary.
Foot extension Operate the foot extension to make sure the movement is correct and
that it locks properly into the different positions when the lever is
released.
Repair or replace the foot extension as required.
Pivot points Make sure the articulations of the bed operate without making any
noise (high low column bearings, head and thigh sections).
Lubricate as necessary.
Casters Check for cuts, wear and quality of the tread, etc.
Replace if necessary.
Braking and steering Check the brakes to see whether the bed moves when the brake bar is
pressed. Repair as necessary.
Make sure the steering mechanism. Replace or adjust the steering
control elements of the steering caster if necessary. Replace the
caster if necessary.
Plug the bed into an appropriate power source and set the brake bar
to neutral. Make sure the alarm is heard.
Set the brakes on the bed. make sure the alarm stops sounding.
Accessories Check the proper operation of any accessories installed on the bed.
Replace any missing or damaged parts.
SideCom® Inspect and test the communication junction box.
Communication
System
Make sure the SideCom® Communication System features operate
correctly. Inspect the communication cable, including the male and
6
female pins in the plug.
Replace as necessary.
Preventive Maintenance
Chapter 6: General Procedures
Date Functions
Hill-Rom
Overall condition
Head and footboards
Manufacturer
AC power cord
Leakage current
Patient pendant
Hilow columns
Head section motor
Plastic sleep deck
CPR release
Head section gas spring
Model Number
Foot extension
Pivot points
Casters
Braking and steering
Accessories
SideCom®
Communication System
this Page
Total Cost for
Scale Calibration
Bed Exit System
Labor Time:
Repair Cost:
Inspected by:
Remarks:
Applicable
N=Not
O=Okay
Replace
R=Repair or
A=Adjust
C=Clean
L=Lube
Legend
The scale calibration weight must be between 100 and 250 lb (45 kg and
113 kg).
1. Make sure the bed is at a comfortable height to add and remove the
weights.
6. Remove the four screws (A) that attach the battery cover (B) to the power
supply (C) (see figure 6-1 on page 6-12).
SHOCK HAZARD:
The potential for electrical shock exists with electrical equipment.
Failure to follow facility protocols may cause death or serious personal
injury.
9. Press and release the calibration switch (D) on the scale P.C. board (E).
The display changes to 100.0.
10. Use the +/- switches (next to the calibration switch) to adjust the weight up
or down to match the calibration weight.
11. When the correct weight is shown, press and release the calibration switch.
12. The display should change to CAL0. Do not touch the bed as it zeroes. 6
13. When the bed is zeroed, the display changes to CAL1 and one tone sounds.
14. Add the calibration weight to the left head of the bed over the load cell. Do
not touch the bed.
NOTE:
For best results, put the weight over the load cell.
15. After approximately 30 seconds the display changes to CAL2, and two
tones will sound.
NOTE:
The time may change because of the environmental conditions.
17. After approximately 30 seconds, the display changes to CAL3 and three
tones sound.
18. Add the calibration weight to the left foot of the bed, over the load cell. Do
not touch the bed.
19. After approximately 30 seconds, the display changes to CAL4 and four
tones sound.
21. After approximately 30 seconds, the display changes to CAL5 and five
tones sound.
22. Add the calibration weight to the right foot of the bed, over the load cell.
Do not touch the bed.
23. After approximately 30 seconds, the display changes to CAL6 and six
tones sound.
25. After approximately 30 seconds, the display changes to CAL7 and seven
tones sound. 6
26. Add the calibration weight to the right head of the bed over the load cell.
Do not touch the bed.
27. The bed will beep 1 time then display the current weight if enabled.
29. Make sure the weight shown on the scale display matches the calibration
weight.
32. Install the battery cover (B) on the power supply (C).
33. Install the four screws (A) to attach the battery cover (B) to the power
supply (C).
Accessories
For CareAssist™ Bed accessories, see table 7-1 on page 7-1.
7
CareAssist™ Bed Service Manual (MAN330 REV 2) Page 7 - 1
7.1 Infusion Support System—P158
Chapter 7: Accessories
Each Infusion Support System can support one infusion pump plus two liters
of intravenous solution.
Installation
WARNING:
Do not exceed ISS pole weight capacity. Personal injury or equipment
damage could occur.
WARNING:
When lowering the upper section of an ISS pole, always grasp and hold
the upper section of the pole before twisting the release knob. Failure to
do so could result in personal injury or equipment damage.
1. Install the P163 adapter into any of the four equipment sockets.
CAUTION:
Do not mount infusion pumps on the lower section of an IV rod.
Interference with head section articulation could result.
Removal
Do the installation procedure in reverse order.
The oxygen tank holder attaches to the bed in a vertical position. The Oxygen
Tank Holder accommodates one E-size oxygen tank with a regulator. The
mounting points are located at each corner of the bed and permit the affixed
oxygen tank holders to pivot.
WARNING:
If the oxygen tank holder is placed at the foot end of the bed, make sure
the Knee Up/Down controls are locked out. Failure to do so can result in
caregiver, patient, or visitor injury if the foot section fully lowers and the
holder becomes dislodged from the bed.
Installation
1. Install the mounting bar vertically into an equipment socket at any corner
of the bed.
NOTE:
The thumbscrew keeps the oxygen tank from rotating in the holder.
Removal
1. Loosen the thumbscrew that holds the tank securely in the holder.
3. Lift up on the tank holder, and remove it from the equipment socket.
7
CareAssist™ Bed Service Manual (MAN330 REV 2) Page 7 - 3
7.3 Standard IV Pole—P2217
Chapter 7: Accessories
The IV pole is a removable, telescopic pole that installs at any of the four
corners of the system in the holes provided.
WARNING:
Do not exceed IV rod weight capacity, personal injury or equipment
damage could result.
WARNING:
When lowering the upper section of an IV rod, always grasp and hold
the upper section of the pole before pulling the release knob. Failure to
do so could result in personal injury or equipment damage.
Installation
Install the IV rod in an equipment socket at any corner of the bed.
Removal
Reverse the installation procedure to remove the IV Rod.
Installation
1. Extend the foot section (A) (see figure 7-1 on page 7-5).
m330_121
Removal
Reverse the installation procedure to remove mattress pad extender.
7
CareAssist™ Bed Service Manual (MAN330 REV 2) Page 7 - 5
7.5 Mattress—P731EA3
Chapter 7: Accessories
7.5 Mattress—P731EA3
Tools required: None
Installation
1. Lower all four siderails.
Removal
Reverse the installation procedure to remove the mattress.
Caregiver Control