Electromiografo Marca Xltec Modelo Neuromax 1004
Electromiografo Marca Xltec Modelo Neuromax 1004
Electromiografo Marca Xltec Modelo Neuromax 1004
TECHNICAL MANUAL
Page 2
XLTEK NeuroMax
Publisher’s Notice
105352 Rev E
XLTEK NeuroMax Technical Manual
The information in this publication is provided for reference only. All information contained in this publication is believed to
be correct and complete. Excel-Tech Ltd. shall not be liable for errors contained herein nor for incidental or consequential
damages in connection with the furnishing, performance, or use of this material. All product specifications, as well as the
information contained in this publication, are subject to change without notice.
This publication may contain or reference information and products protected by copyrights or patents and does not
convey any license under the patent rights of Excel-Tech Ltd., nor the rights of others. Excel-Tech Ltd. does not assume
any liability arising out of any infringements of patents or other rights of third parties.
All rights reserved. This document contains confidential or proprietary information of Excel-Tech Ltd. No part of this
document may be reproduced or transmitted in any form or by any means without the written permission of Excel-Tech
Ltd.
Excel-Tech Ltd. makes no warranty of any kind with regard to this material, including but not limited to the implied
warranties of merchantability and fitness of a particular purpose.
Page 1
XLTEK NeuroMax
TABLE OF CONTENTS
1.1. INTRODUCTION 13
1.1.1. INTENDED USE 13
1.2. CUSTOMER SERVICE 13
1.3. USING THE MANUAL 14
Page 2
XLTEK NeuroMax
4. EMG TESTS 45
5. OTHER TESTS 50
Page 3
XLTEK NeuroMax
6. AV STIM 1000 73
8. ADMINISTRATIVE FUNCTIONS 83
10.1. SOFTWARE 95
Page 4
XLTEK NeuroMax
10.3. ELECTROMYOGRAPHY 95
INDEX 118
Page 5
XLTEK NeuroMax
The NeuroMax and its accessories have been designed to comply with the following national
and international standards.
Page 6
XLTEK NeuroMax
IEC 61000-4-5:1995 / EN 61000-4- Electromagnetic compatibility (EMC) - Part 4-5: Testing and
5:2006 measurement techniques - Surge immunity test
IEC 61000-4-6:1996 / EN 61000-4- Electromagnetic compatibility (EMC) - Part 4-6: Testing and
6:2007 measurement techniques - Immunity to conducted
disturbances, induced by radio-frequency fields
IEC 61000-4-8 / EN 61000-4-8 Electromagnetic compatibility (EMC) - Part 4-8: Testing and
measurement techniques - Power frequency magnetic field
immunity test
IEC 61000-4-11:2004 / EN 61000-4- Electromagnetic compatibility (EMC) - Part 4-11: Testing and
11:2004 measurement techniques - Voltage dips, short interruptions
and voltage variations immunity tests
IEC 61000-3-2:2005 / EN 61000-3- Electromagnetic compatibility (EMC) - Part 3-2: Limits -
2:2006 Limits for harmonic current emissions
Page 7
XLTEK NeuroMax
Page 8
XLTEK NeuroMax
Page 9
XLTEK NeuroMax
Table 3 - Electromagnetic Immunity – for EQUIPMENT and SYSTEMS that are not LIFE-
SUPPORTING
The NeuroMax 1004/1002 is intended for use in the electromagnetic environment specified below. The customer or the user of
the NeuroMax 1004/1002 should assure that it is used in such an environment
Conducted RF 3 Vrms
150 kHz to 80 MHz
150kHz to 80MHz
IEC 61000-4-6
Complies
80MHz to 800MHz
Radiated RF 3 V/m
80 MHz to 2.5 GHz
IEC 61000-4-3 800MHz to 2.5GHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the NeuroMax 1004/1002 is used
exceeds the applicable RF compliance level above, the NeuroMax 1004/1002 should be observed to verify normal
operation. If abnormal operation is observed, additional measures may be necessary, such as re-orienting or
relocating the NeuroMax 1004/1002.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Page 10
XLTEK NeuroMax
Note: This equipment has been tested and found to comply with the limits for a Class A
digital device, pursuant to part 15 of the FCC Rules. These limits are designed to
provide reasonable protection against harmful interference when the equipment is
operated in a commercial environment. This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instruction manual, may cause harmful interference to radio communications. Operation
of this equipment in a residential area is likely to cause harmful interference in which
case the user will be required to correct the interference at his own expense.
Page 11
XLTEK NeuroMax
1. WELCOME TO THE
NEUROMAX
Congratulations, you have purchased the NeuroMax
from XLTEK, one of the world's top manufacturers of
neurodiagnostic equipment and software.
Page 12
XLTEK NeuroMax
1.1. INTRODUCTION
XLTEK is commitment to continual product
improvements and quality design to meet the needs
of our clients. We thus encourage all feedback and
any suggestions you have regarding any aspect of
the EMG system, the manual, our line of
accessories, and our support services.
Page 13
XLTEK NeuroMax
Page 14
XLTEK NeuroMax
Page 15
XLTEK NeuroMax
2.2. SPECIFICATIONS
Specification Value
General Description
Dimensions 4” (H) x 13” (W) x 13” (D)
Weight 11.4lbs (5.2kg)
Colour White
Materials Injection moulded PC ABS
chassis
External Connectors Headbox, Headphones,
Printer, Footswitch
(pneumatic)
Channels 4 (NeuroMaxCE 1004)
2 (NeuroMaxCE 1002)
Sampling Rate 60 kHz
Display
Size and Type 12.1” active matrix colour LCD
screen
Resolution 1024 x 768
Electrical Specifications
Maximum Rated 92 VA
Input Power
Heat Loss 92 Watts
Maximum Supply ± 10% for either 50 Hz or 60
Current Tolerance Hz transformer*
Electrical Supply ± 10% for either 50 Hz or 60
Frequency Tolerance Hz transformer*
Insulation Class and Class 1, Type BF
Type
Power Input Voltage, 120 VAC, 60 Hz, 1A
Frequency, and 230 VAC, 50 Hz, 0.4 A
Rated Current
Mains Connection Protectively grounded
detachable power supply cord
CMRR ~105 dB
Page 16
XLTEK NeuroMax
Stimulator Specifications
Maximum Simulator ~450 V
Voltage
Stimulator Current ± 1%
Accuracy
Stimulator Duration ± 1%
Accuracy
Voltage Accuracy on ± 1%
Screen
Page 17
XLTEK NeuroMax
2.3.1. WARNINGS
Warnings MUST be followed when using the
equipment. Warnings apply to conditions which can
injure the patient and/or the operator.
Page 18
XLTEK NeuroMax
Page 19
XLTEK NeuroMax
2.3.2. CAUTIONS
Cautions must be noted when using the equipment.
Cautions apply to conditions which may damage the
NeuroMax.
Page 20
XLTEK NeuroMax
Page 21
XLTEK NeuroMax
878-02-02 Type BF
Equipment
417-5007 Power On
Medical Device
Directive CE Mark
0086 93/42/EEC
Canadian
Standards
Not Applicable Association
(indicates safety
approval by)
Page 22
XLTEK NeuroMax
Page 23
XLTEK NeuroMax
Page 24
XLTEK NeuroMax
Page 25
XLTEK NeuroMax
Page 26
XLTEK NeuroMax
AV Stimulator parameters
KEY FUNCTION
Space Bar if in full screen mode, averaging and
single acquisition mode, deletes the
latest acquisition from the average
X post processing artifact extraction
KEY FUNCTION
TRIGGER go to Triggered from FreeRun
1,2
F go to FreeRun from Triggered
Page 27
XLTEK NeuroMax
KEY FUNCTION
A change the Acceptance level of the
signal (entered as a percentage of
the A/D converter full scale value)
D Display only certain traces
F Flip to show the next or previous set
of data in the data box
L set-up the left ear for Audio
threshold test (Left = 15 dB nHL . Audio tests
Stimulus, Right = 0 dB Noise) are only valid when
set-up the right ear for Audio the AVStimulator is
R attached and in Audio
threshold test (Left = 0 dB Noise, Stimulation mode.
Right = 15 dB nHL Stimulus)
Space Bar show live trace
Page 28
XLTEK NeuroMax
Page 29
XLTEK NeuroMax
Page 30
XLTEK NeuroMax
Page 31
XLTEK NeuroMax
key .
2. From the Administrative Functions Menu, select
Patient Directory and then press the Select
key.
3. Select the function and press the Select key.
Page 32
XLTEK NeuroMax
Page 33
XLTEK NeuroMax
Page 34
XLTEK NeuroMax
Data Box
Assigned Trace Area
Page 35
XLTEK NeuroMax
key.
Page 36
XLTEK NeuroMax
Page 37
XLTEK NeuroMax
Page 38
XLTEK NeuroMax
Page 39
XLTEK NeuroMax
Page 40
XLTEK NeuroMax
Page 41
XLTEK NeuroMax
Page 42
XLTEK NeuroMax
Page 43
XLTEK NeuroMax
Page 44
XLTEK NeuroMax
4. EMG TESTS
This chapter will deal with the operation of the
NeuroMax in its EMG (Electromyography) Studies
mode of operation. The NeuroMax has many views
of the data on screen to view single traces, multiple
traces in a raster, single traces with multiple
triggered sweeps captured, compressed EMG data
in the real-time window. The NeuroMax
incorporates two triggers that may be adjusted
independently. The first trigger is amplitude/slope.
The second trigger acts as a discriminator that
allows you to look at smaller potentials
Page 45
XLTEK NeuroMax
Page 46
XLTEK NeuroMax
Page 47
XLTEK NeuroMax
Page 48
XLTEK NeuroMax
Page 49
XLTEK NeuroMax
5. OTHER TESTS
The NeuroMax conducts several other tests,
including Evoked Potentials, Blink Reflex,
Incremental Stim, Heart Rate variability, and P 300.
Page 50
XLTEK NeuroMax
Page 51
XLTEK NeuroMax
Page 52
XLTEK NeuroMax
Page 53
XLTEK NeuroMax
Page 54
XLTEK NeuroMax
Channe
C3‟/4' (-) VS Fz (+)
l1
Channe
Cv5 (-) VS Fz (+)
l2
Channe
C3‟/C4‟ (-) VS Fz (+)
l1
Channe
Cv7 (-) VS Fz (+)
l2
Channe
Erb's (-) VS Fz (+)
l3
Channe
Elbow (-) VS Fz (+)
l4
Channe
Cz (-) VS Fz (+)
l1
Channe Pop fos ipsi Pop fos contra
VS
l2 (-) (+)
Channe
Cz (-) VS Fz (+)
l1
Channe
Cv5 (-) VS L3 (+)
l2
Page 55
XLTEK NeuroMax
Page 56
XLTEK NeuroMax
Channe
Cz (-) VS Fz (+)
l1
continue.
3. In Section 2, choose Blink using the Select
arrows and then press the Select key to
continue.
4. In Section 3, the Bilateral side is defaulted.
Press the Select key to open the Blink Test
screen.
Page 57
XLTEK NeuroMax
Page 58
XLTEK NeuroMax
Page 59
XLTEK NeuroMax
Page 60
XLTEK NeuroMax
Page 61
XLTEK NeuroMax
Page 62
XLTEK NeuroMax
Page 63
XLTEK NeuroMax
Page 64
XLTEK NeuroMax
Example:
Median 2CH NCS
Active Ch 1 2
Channel 1 On On
CV Segments Segmental Total
LFF(Hz) 5.0 30
HFF(Hz) 2.0k 2.0k
Notch Filter Off On
Gain (uv/div) 5k 20
Lat1 take off take off
Lat2 neg phase peak
Amp bsln to peak peak to peak
Rec Site APB Index
Stim Site1 Wrist Wrist
Stim Site2 Elbow Elbow
Stim Site3 Axilla Axilla
Etc.
Page 65
XLTEK NeuroMax
Page 66
XLTEK NeuroMax
Page 67
XLTEK NeuroMax
Page 68
XLTEK NeuroMax
Page 69
XLTEK NeuroMax
Page 70
XLTEK NeuroMax
Page 71
XLTEK NeuroMax
Page 72
XLTEK NeuroMax
6. AV STIM 1000
The AV Stim is used with Evoked Potentials and
operates with both the NeuroMax 1002 and the
NeuroMax 1004, and with subsequent operating
systems.
The red
stimulus indicator
The green power
blinks on when the
indicator lights up
NeuroMax powers up
when AVSTIM
and flashes at the set
1000 is powered on.
stimulus rate.
NeuroMax
Audio Out
Connect to NeuroMax
LEDGoggles
VGA
Update Port
Page 73
XLTEK NeuroMax
Page 74
XLTEK NeuroMax
6.4.1. WARNINGS
Warnings MUST be followed when using the
equipment. Warnings apply to conditions, which can
injure the patient and/or the operator.
6.4.2. CAUTIONS
Page 75
XLTEK NeuroMax
Page 76
XLTEK NeuroMax
Page 77
XLTEK NeuroMax
In a unilateral study, 3. Have the test subject cover the right eye and
4 sets of 2 channels fixate towards the center of the checkerboard
can be collected in
this manner. field with the left eye.
4. Press the Start/Stop Control key and begin the
acquisition.
5. Press the Averager key to begin averaging.
6. Once a set is completed, press the Averager
key again to collect the next set. If you are
doing a Bilateral test
7. To test the Right side, press the Test Menu key, (collecting waveforms
on the same menu
select the Right side in the Choose Side section screen), switch eyes
of the Test Screen, and then press the Select and repeat steps 3-6.
key to continue.
8. Cover left eye and repeat steps 2-6.
Page 78
XLTEK NeuroMax
Channe
Oz (-) VS Fpz (+)
l1
Channe 10cm lateral to
Use these tables to VS Fpz (+)
set your Montage l2 O1 (-)
levels.
Channe
Oz (-) VS Fpz (+)
l1
Channe 10cm lateral to
VS Fpz (+)
l2 O2 (-)
Page 79
XLTEK NeuroMax
Page 80
XLTEK NeuroMax
7.1.1. CREATING AN
ELECTROMYOGRAPHY SUITE
You can set the default parameters to create a
personalized EMG suite that fits your specifications.
Suite.
6. Use the Left/Right arrows to place the yellow
highlight over the next muscle you wish to add to
the suite and then press the Select key.
Continue until all muscles have been added.
7. Press the Test Menu key twice to save and exit.
Page 81
XLTEK NeuroMax
Page 82
XLTEK NeuroMax
8. ADMINISTRATIVE FUNCTIONS
The NeuroMax Administrative Functions are
designed to help you customize the NeuroMax and
to manage the information acquired by the tests.
REPORT FUNCTIONS
Patient Recall, print, and edit stored
Directory patient files
Memory Delete files and erase
Management memory
Batch Print Print all non-printed reports
System Set global options including
Options date, time, units of measure,
printer type, artifact rejection,
external stimulator, and more
Edit Report Edit macros, select fields
Format included in the report, and
report layout
Edit Site Name Edit the available default
List stimulation sites
Edit EMG Edit available notes for
Notepad scoring EMG
Edit Patient Edit the field definitions for
Info. Fields the Patient Info screen
Page 83
XLTEK NeuroMax
Page 84
XLTEK NeuroMax
Future
Expansion
Print All Test Print all performed test
Screens waveforms in condensed
format.
Print Letter Print the history and
interpretation in a letter
format with the referring
doctor‟s name.
Save File to Moves highlighted files to
USB USB device. See section
8.1.2. for details.
Exit Directory Exit to Administration Menu.
Page 85
XLTEK NeuroMax
Page 86
XLTEK NeuroMax
Page 87
XLTEK NeuroMax
Page 88
Xltek NeuroMax
Page 89
Xltek NeuroMax
Name/Institution:
Tel: Fax: Email:
Additional Contacts: Name Position
Basic Functions
Connections for NeuroMax: Keyboard Layout:
USB Printer Function Hard Keys
Headbox Menu Hard Keys
Stim Probe Select Switch
Accessories Arrow Keys
USB Mass Storage Trigger Arrows
Foot Switch Stimulus Intensity
AV STIM Functions: Opening a Patient File
External Trig Electrode Impedance Testing
Hot Keys
Power Supply of NeuroMax:
AVStim:
Electromyography
Basic Functions: Advanced Functions: Creating Editing:
Recording (Start/Stop) Review Defaults
Settings: (Gain/Timebase) Turns and Amplitudes Triggered Defaults
Audio Adjustments Motor Unit Analysis (1,2) EMG Suites
Notepad Free Run to Triggered Muscles
Suites (key function)
Programs:
Free Run
Triggered
Page 90
Xltek NeuroMax
Administrative Functions
Patient Directory: Memory Management:
Search Utility
Patient Callback Patient Directory
Test Directory Delete All Files
Print Report Delete All Printed files
View Report
Delete File System Options:
Edit Patient Info 1) Report Format
Transfer to USB -Report Header -Report Fields
Print Test Screens -Interpretation Macros -Report Layout
2) Site Name List
Batch Print: 3) EMG Notepad
4) Patient Information Fields
Report Menu
While in an active patient file, press REPORT. The following functions should now be available:
View Report Print Report
Edit Report (host directory) Print Report and Waveforms
Interpretation Save Current Patient to USB
Accessories
Main Contact (for ordering Consumable items): Please indicate style of Needle currently used:
Name: Disposable Monopolar
Phone: Disposable Concentric
Disposable Injectible Monopolar
Please indicate the length and gauge of the needles used:
Please indicate the number of EMG's performed monthly:
Please indicate if you are using: Please indicate your current Accessories Supplier:
Disposable Electrodes
or Reusable Electrodes
for Nerve Conduction Studies
Are you currently receiving Preferential Pricing for a Needle Contract? Yes No
?? - Would you be interested in hearing how Accessories are able to enhance the quality of
your studies, as well as save you time and money? Yes No
Page 91
Xltek NeuroMax
XLTEK University
?? - Would you be interested in hearing how XLTEK University is able to give you one-on one, on-site
training, that can help you in the areas of, but not limited to:
equipment use, basic and advanced clinical training, and insight on clinical correlation?
Yes No
COMMENTS
Please Confirm
Date Completed:
Page 92
Xltek NeuroMax
10. APPENDIX 2:
TROUBLESHOOTING
Symptom Cause Solution
No visual display or audio alarms at Power switch not turned on. Plug in power cord.
power-up. Turn power switch on.
Electrode impedance is too high. Electrode dislodged from patient, but not Insert dislodged electrode; tape down in place.
> 10 KΩ for subdermal electrodes. completely out.
> 25 KΩ for Prass paired electrodes. High resistance in electrode. Remove and replace with new electrode.
> 05 KΩ for EMG tube. Electrode pin not firmly inserted into patient Check connection at Patient Interface box.
> 40 KΩ for hookwire electrode. interface.
Electrode impedance ≤ 0.1 KΩ Positive and negative electrodes touching below Remove and relocate electrodes.
surface of skin.
Extremely low impedance, particularly in EMG Use “tap test” near electrodes to evoke EMG or artifact
tubes. (increase threshold, decrease volume for test). If activity
is noted on channel in question, proceed.
Electrode reading is: Electrode laying on skin surface. Re-insert electrode in question.
(+ or -) Off Electrode placement insecure.
or Dirty electrode tip. Remove and replace electrode in question.
Δ==== Electrode cable is broken.
Electrode pin disconnected from patient interface. Check connection to Patient Interface box.
Electrode difference is greater than Dirty electrode. Remove and replace electrode for appropriate channel
2KΩ (Subdermal electrodes) or with highest impedance reading first.
10KΩ (Prass Paired electrodes). Mismatched pair. Remove and replace electrode in question.
Unequal placement.
Interference on anesthesia equipment Measuring current on NIM® EMG Electrodes. Try an alternate EKG Lead set.
(EKG Monitor).
Electrode Check (Electrode Screen selected). Deselect Electrode Screen.
Muting function active. See Excessive Muting (Symptom Column).
With Stimulator active. Turn the NIM® Stimulator to 0.0 mA when not needed.
Incrementing Probe will not adjust Loose connector. Check connector is properly aligned and fully seated (See
stimulation. System Set Up / Patient Interface Set Up).
- -
Stimulus keeps changing (run away). Bad Incrementing Probe. Replace Incrementing Probe or disconnect STIM
CONTROL connector and manually adjust stimulus at
touch screen.
Electrosurgical interference. Muting Detector Probe not connected. Check Muting Detector Probe connections.
Muting Detector Probe input insufficient. Loop the unit cable through muting detector.
Electrosurgical grounding inadequate. Check electrosurgical grounding pad on patient.
Source of interference unidentified. Identify source of interference; then eliminate or separate
NIM 3.0 system or Patient Interface cable too Maintain separation between electrosurgical cable
close to electrosurgical unit or its cables.
For less coupling, coil up the Muting Detector Probe next
Excessive Muting. Unit receiving excessive signal into the Muting Disconnect the muting detector completely.
Detector Probe or electrode leads.
Rhythmic Artifact. Pacemakers – Pace Pulse. Relocate electrode ground and stimulus return to patients
shoulder (Acromion).
Inadequate muting. Signal from electrosurgical unit inadequate to Loop the electrosurgical unit cable and clip the muting
cause muting. detector over the doubled cable.
No response to direct stimulation. Inadequate stimulus intensity. Increase stimulus intensity.
Paralyzing anesthetic in use. Eliminate paralyzing anesthetic.
White Stimulation (+) electrode has fallen out or Check that Stimulus Measure is approximately the same
is not connected. value as the Stimulus setting. Re insert electrode in
-
question.
Probe (electrode) not connected. Check stimulator anode (+) and cathode (-) connections.
Patient safety fuse blown. Check fuse in STIM 1 (EMG) Patient Interface Replace if
STIM 1 (EMG) Patient Interface fuse REF necessary.
8253075.
Not holding probe on nerve long enough. Hold probe tip to nerve for at least 1 second.
Nerve not contacted. Check stimulator tip for obstruction. Replace if
necessary. Check location of stimulation.
Volume control too low. Check and correct all settings volume, event threshold,
Event threshold set too high. stimulus intensity.
Excessive current shunting in surgical field. Remove fluids from surgical stimulating area.
No electrodes in innervated muscle tissue. Nerve Place channel electrodes in muscle to be monitored.
not stumbled. Check EMG tube placement if applicable.
Page 93
Xltek NeuroMax
Unexpected responses when not Unexplained continuous “train” EMG response. Identify and eliminate possible source of “train”
directly stimulating nerve. stimulation:
Cold irrigation.
Laser heat.
Retraction on nerve or muscles being recorded.
Patient waking from anesthesia.
Nerve drying.
Ultrasonic aspirator.
Nerve or monitoring area being stimulated or Identify and eliminate source of inadvertent
manipulated by thermal or mechanical means. manipulation.
Metal-to-metal discharge artifact. Determine response type from waveform pattern on
50 ms screen.
Intertwined recording electrode and stimulator Disentangle recording electrode and stimulator cables.
wires.
Inadvertent manipulation of electrode wires, Check area near recording electrodes for excessive
Patient Interface cable, or recording area on stretching from tape, drapes, etc.
patient.
Electrical interference from other equipment. Check for intermittent stimulation from anesthesiologist
(i.e., hand-held electrical stimulator).
Move NIM 3.0 system away from source of interference.
Make sure Patient Interface cable and electrode wires do
not cross other electrical equipment or cables.
Page 94
Xltek NeuroMax
10.1. SOFTWARE
The following sections show you how to verify the software version on your
machine, how to update your software, and how to install NeuroMax Loader.
NOTE: ANY CUSTOMIZATION OF TEST DEFAULTS AND REPORT FORMAT WILL NOT
BE AFFECTED.
10.3. ELECTROMYOGRAPHY
This section gives you the basic steps for creating an electromyography suite and
how to use it.
Page 95
Xltek NeuroMax
1. From Main Menu screen, select Administrative Functions and then select
Patient Directory.
2. Using the Select arrows, choose Patient Directory, highlight patient name,
and then select Test Directory in the Functions column.
3. Choose the test to edit and press the Select key, choose Recall and Edit
File.
1. From the Main Menu screen, select Administrative Functions and then
select Edit Report Format.
2. From the Edit Report Format Menu, select Edit Interpretation Macros to
view the 30 titles that can be edited. Each title represents one interpretation
macro.
3. To edit a title, choose it using the Select arrows. A red cursor appears in the
lower box. Next use the keyboard to enter your prepared macro into the
highlighted position and then press the Select key to enter the macro.
4. When finished, press the Select key once to move onto the next macro.
5. To exit from the Edit Interpretation Macros screen, highlight EXIT from
Interpretation Macros and press the Select key.
Page 96
Xltek NeuroMax
1. From the Main Menu screen, select Administrative Functions and then
select Patient Directory.
2. Highlight the patient name and press the Select key.
3. Choose Test Directory from the Functions column and press the Select
key.
4. From the Test Directory, highlight Interpretation and press the Select key
to open the Interpretation window.
5. Type in your entry or enter a macro.
6. To enter a macro, press the Notes key and then the two-digit number that
represents the title of the new macro to enter.
7. Press the Select key to exit from the Interpretation screen.
The electrical stimulus pulse and surface electrodes give rise to an artifact
consisting of an initial spike and a longer lasting tail which often interferes with
the recorded signal. It has 4 sources:
Page 97
Xltek NeuroMax
The NeuroMax takes care of the first two variables. You can check the
following:
Electrode impedance should be kept below 10k (green range) for the best
results in recording low amplitude signals (<50uV peak to peak). Higher
impedance allows the incursion of power line interference (60Hz or 50Hz) and
shock artifacts. Both of these phenomena can obscure your short latency
sensory response. At these impedance levels, it becomes necessary to turn on
your notch filter. Impedance greater than 100K red level with sensory responses
below 100 uV peak to peak are invariably difficult to interpret.
Page 98
Xltek NeuroMax
The skin is a living tissue. As such it is made of various layers. The top three
layers are of crucial importance as they are highly nonconductive. The three
layers are the actual dermis, the dead covering of dry scaly dead skin, and a
layer of oils, sebum, and dirt.
Normal SNAPs require no skin preparation. Clients who have callused, greasy,
dirty, or dry skin require an extra effort on the part of the Physician/Technologist
to obtain good results. Typically washing the skin area of interest (both
recording and stimulus sites) with warm soapy water is adequate. For more
difficult sites or where washing is difficult or inappropriate, rubbing with of an
alcohol swab may be more efficacious. Remember, you are removing these
three layers so use of gentle force may be required. We recommend the use of
a mild abrasive such as Omni-Prep (pumice suspended in a mild detergent), the
hard end of a Q-tip, or rough tissue paper for more difficult situations.
Please be consistent. Always use the same type of electrodes for active and
reference, otherwise your results may not be consistent.
The most common type of electrodes are either metal discs (with conductive
paste) or disposable self-adhesive electrodes. Both are usually about 0.5 to 3
cm2 in area (0.1 to 0.5 inch2).
Page 99
Xltek NeuroMax
The Active electrode (Black - Cathode) is to be placed over the nerve towards
the cathode of the stimulator. The Reference electrode (Red + Anode) is placed
2 to 3cm (distance is set by your procedure) distal to the active electrode
(relative to the cathode of the stimulator), again over the nerve. Make sure that
the skin surface between all 3 electrodes is dry. In addition, stainless steel clip
type electrodes with conductive paste may be used.
We do not recommend the use of steel coil electrodes for recordings if you are
unfamiliar with their use. Slipping the coils onto the digit will disperse the gel
between the two electrodes, usually causing an electrode bridge and
subsequent shorting between the electrodes. Also, using spring coils or spring
electrodes can increase the surface area and also disperse the signal since part
of the surface area is over a silent area of the digit. Stimulus artifact may be
reduced in these situations by covering the segment of the clip that is not over
the nerve with a piece of gauze or tissue paper.
10.5.6. STIMULATOR
Though the NeuroMax can stimulate well by simply pressing the stimulator
prongs against the skin, if the message Stimulator impedance high appears
on the screen you will be required to add a small drop of conducting gel to the
stimulator prongs. Do not spread the gel otherwise the current will spread over
a larger skin area, effectively diminishing the current under the prongs forcing
you to increase the applied current and causing stimulus artifact problems.
Keep the stimulus current to "just the right amount." Excessive current causes
the patient to feel pain and the NeuroMax to pick up excessive stimulus artifact.
Furthermore, tense patients contract their muscles and inject EMG activity into
the sensory recording. This will make your NCV look very rough making correct
interpretation impossible.
Page 100
Xltek NeuroMax
With individual unshielded recording lead wires, the best results are achieved
when the active and reference lead wires are of the same length and are
loosely twisted together up to the point where they have to branch off to reach
the electrode placements. They should be kept away from other cables and
from the unit's enclosure to minimize interference from power lines and other
external sources.
Similarly, when stimulating with two individual cables from the NeuroMax Stim
Probe, those wires should also be twisted together and kept away from
recording lead wires. This will ensure no increase in stimulus artifact due to
coupling between cables.
In environments with high interference levels, better SNAPs are recorded if you
use the NeuroMax shielded acquisition cable. This cable takes advantage of
NeuroMax‟s unique active differential shield drivers producing efficient shielding
without compromising on common mode rejection and signal bandwidth. This
cable plugs into the 5 pin DIN connector on the headbox and provides alligator
clips for active, reference and ground electrodes.
XLTEK's design differs from other manufacturer's shielded recording cables that
do not support differential shield driving technology and instead rely on simply
grounding their shields. Cables of this type add common mode capacitance to
the input lines causing degradation of high frequency common mode rejection
which in turn causes an increase in stimulus artifact and an increased
susceptibility to high frequency interference. Depending upon cable length,
signal bandwidth may also be compromised. Such cables may reduce power
line interference at the expense of an increase in stimulus shock artifact,
increased high frequency interference, and reduced bandwidth.
Sometimes the obvious solution is the right one. When all else fails, check the
following:
Page 101
Xltek NeuroMax
2. Try running a surface EMG with the volume on. The EMG acts as an
oscilloscope, if there is noise in the system, you may be able to spot it here.
3. Unplug any other devices on the same circuit such as printers, mechanical
beds, vacuum cleaners, or other potential sources of leakage current.
4. A medical grade ground should be installed in your clinic. Some clinics do
not have properly grounded electrical systems.
5. If the patient is tense, ask her to relax.
6. Frequent inspections of your cables will help prevent recording problems.
7. Always have a backup acquisition cable and a backup ground.
8. Are any of the electrodes touching? If so, they are causing a short circuit
and you will develop an artifact.
9. Try pressing X on the keyboard to mathematically remove the shock artifact
from the waveform in the acquisition screen.
10. Rotate the anode around the cathode. This might shunt the shock artifact
away from the recording electrodes.
11. Increase or decrease the pulse duration of the stimulus.
12. Average the sensory response.
1. Use an alcohol prep pad to swab the skin to remove any excess sweat or
lotion.
2. Use a mild abrasive paste or pumice tape and gently rub the skin over the
stimulation site to remove dead skin and then swab with the alcohol pad.
3. Instruct patients ahead of time to avoid applying lotion to the skin on the
morning prior to their test.
4. Make sure that the sticky electrodes are adhering properly to the skin. If they
have become loose, replace them with new ones.
Page 102
Xltek NeuroMax
5. Make sure that the Stim probe has some gel on the tips and that there is not
gel all over the patient's skin causing a bridge.
6. Make sure that the patient is properly grounded--good sticky electrode, wet
Velcro ground strap, and good connections in the headbox.
7. When in doubt, try another channel.
10.6. AV STIM
1. Attach the short cable between the NeuroMax and the AV Stim and then
turn on the power to the NeuroMax. When the NeuroMax recognizes the AV
Stim it will provide power to it.
2. When the power is turned on, the green power light on the AV Stim lights up
and the red stimulus light flashes once confirming the initialization of the AV
Stim with the NeuroMax.
3. The printer port of the AV Stim is not active with this version of the
NeuroMax.
10.7. PRINTING
If you are having printing problems, try the following:
10.7.1. NEUROMAX
1. Shut down NeuroMax and printer and then disconnect all cables.
2. Reconnect all cables, turn on the printer, and then turn on NeuroMax.
3. From the Main Menu screen, select Administrative Studies and then
select Patient Directory.
4. Select the patient file and then select Print Report.
10.7.2. PRINTER
Follow the instructions from the printer manufacturer to print a test page.
Does the printer need a new toner cartridge?
Download files to USB mass-storage device and print through Microsoft
Word or XLTEK Archiver on a PC.
Page 103
Xltek NeuroMax
CAUTION:
DISCONNECT THE POWER CORD FROM THE SYSTEM AND THE WALL BEFORE
CLEANING. USE A LINT-FREE CLOTH. DO NOT USE ABRASIVE CLEANERS ON
ANY SYSTEM COMPONENT(SEE NEUROMAX ENCLOSURE BELOW).
Be careful not to allow any excess fluid to seep into the internal electronic
components of the system; be especially careful around the grills located both
on the front panel near the top (i.e. the speaker grill) and on the left side of the
NeuroMax near the back, as well as around the back panel connections and the
headphone jack on the right side near the front.
Page 104
Xltek NeuroMax
Page 105
Xltek NeuroMax
10.8.2. KEYBOARD
The keyboard is made up of discrete keys and cannot be easily cleaned. If any
material were to slip between the keys the NeuroMax can be held upside down
to allow those pieces to drop. Also, individual keys can be cleaned carefully
with Q-tips. Avoiding any sustained exposure to extreme temperatures (above
50o Celsius or below -20o Celsius) will help prolong the integrity of the keyboard.
10.8.3. SCREEN
The screen itself is protected with a front cover made from a mylar-type
material, however the screen uses glass internally, so care should be exercised
not to drop or bump the screen, as this may cause the glass to crack or break.
The screen can be cleaned by wiping with a damp cloth with no water dripping
off or a mild detergent; though a soft lint-free cloth is recommended for cleaning
the screen.
CAUTION:
DO NOT LEAVE ANY CABLES CONNECTED TO THE BACK PANEL WHEN
TRANSPORTING THE NEUROMAX; DOING SO MAY CAUSE THE CONNECTORS
TO COME LOOSE, AND/OR MALFUNCTION. THE BACK PANEL AND
CONNECTORS SHOULD ONLY BE CLEANED WITH A DRY, SOFT, LINT-FREE
CLOTH.
Page 106
Xltek NeuroMax
The stimulus probe body and cable can be cleaned with a soft, lint-free cloth or
a damp cloth soaked in water, mild detergent, or a cold sterilizing agent.
Determine cable integrity regularly. The stimulus probe headpieces can be
cleaned in the same fashion as the body, but extra care should be taken to
remove all electrode gels or pastes that have been used on the headpieces;
avoid any long term build-up of gels or pastes on the stimulus probe
headpieces, as this may interfere with the optimal performance of the stimulus
probe. If the various headpieces do not fit snugly into the probe body, replace
them immediately to ensure optimal performance.
Page 107
Xltek NeuroMax
1. Check the lower right corner of the screen, in the stimulator status area, to
see that the "STIMULATOR ON" message is displayed. If the display reads
"OFF", press the START/STOP switch once to activate the stimulator.
2. Check the stimulator status area to determine that the intensity level, in mA,
is appropriately set for the current test.
3. Check the connection between the stimulator probe cable and the headbox to
ensure that the fit is snug.
4. Check the connection between the headbox and the NeuroMax (at the back
on the lower right side of the NeuroMax to ensure that the headbox is properly
connected - be sure to tighten the thumbscrews so that the headbox cable does
not come loose.
5. Check that the skin between the two stimulating electrodes (i.e. anode and
cathode) is clean and free from any layer of conducting gel or paste that might
create a bridge between the anode and cathode.
6. Press MAIN MENU to exit to the main menu, and then re-enter the current
test to reset all software and hardware settings for the stimulus.
7. Check that the "Stimulator Impedance High" error message does not appear
when you attempt to stimulate. If it does, clean the stimulus site on the skin and
prepare the site with some type of abrasion (e.g. Omni-Prep) to decrease the
impedance level.
1. Make sure that the patient electrodes are connected to the correct amplifier
channel in the headbox. In all test modes, the screen will display the currently
active channel beside the test name in the data table area; check that the active
channel corresponds to where the patient electrodes are connected in the
headbox.
Page 108
Xltek NeuroMax
2. Check that the patient electrodes fit properly into the headbox (not loosely),
and also check that there are no apparent breaks in the patient electrode
cables.
3. Check the patient electrode impedance levels - if the impedance levels are
too high, you may need to replace the electrode, electrode cable, or simply re-
prep the recording site on the skin. Make sure that the positive and negative
recording electrode sites have similar impedance values.
4. Check the connection between the headbox and the NeuroMax to make sure
that it is tight.
5. Check the screen gain and timebase settings to ensure that they are
appropriate for the current test. You may also want to check the LFF, HFF, and
Notch Filter settings, as well as any sweep or trigger delays being used.
6. Press MAIN MENU to exit to the main menu, and then re-enter the current
test to reset all software and hardware settings for the amplifiers
B. In Dual Channel Mode, check for all of the above situations, and then check
the following:
1. If the current test calls for averaging, check that the "AVERAGER ON"
message is displayed on the screen. If it is not shown, press the AVERAGER
hardkey.
2. If the AVERAGER is "ON", check the raw data display to ensure that
appropriate data is actually being acquired through the amplifiers (Press space
bar to display raw data). If the raw data display is flat, then no data is being
acquired by the amplifier(s).
1. Check the patient electrode impedance levels - if the impedance levels are
too high, you may need to replace the electrode, electrode cable, or simply re-
prep the recording site on the skin. Make sure that the positive and negative
recording electrode sites have similar impedance values.
2. Check the integrity of the ground electrode cable. If it appears broken or bent
at all, try replacing it.
3. Position the ground electrode between the stimulating electrode (or probe)
and the active recording electrode.
Page 109
Xltek NeuroMax
4. If you are using a non-disposable, metal ground electrode plate, check that
there is no corrosion or build up of gel on either the surface of the plate or the
contact between the plate and the cable.
5. Check that the amplifier gain setting is appropriate for the current test. It is
possible that you may be over-amplifying the signal being recorded, and thus
over amplifying the stimulus artifact.
6. If you are using single shot stimulus mode, or even if you are using train
mode with a low number of pulses per train, try using AVERAGED acquisition
(i.e. press the AVERAGER hardkey to turn on the Averager) to reduce the
influence of the variable "noise" of the Stimulus artifact.
1. Make sure that the patient electrodes are connected to the correct amplifier
channel in the headbox. In all test modes, the screen will display the currently
active channel beside the test name in the data table area; check that the active
channel corresponds to where the patient electrodes are connected in the
headbox.
2. Check that the patient electrodes fit properly into the headbox (not loosely),
and also check that there are no apparent breaks in the patient electrode
cables.
3. Check the patient electrode impedance levels - if the impedance levels are
too high, you may need to replace the electrode, electrode cable, or simply re-
prep the recording site on the skin. Make sure that the positive and negative
recording electrode sites have similar impedance values.
4. Check the connection between the headbox and the NeuroMax to make sure
that it is tight.
Page 110
Xltek NeuroMax
1. Check both ends of the AC power cord to ensure they are connected to the
unit and the wall properly (use only an approved hospital grade, 3 pronged
outlet for AC connection).
2. Check that the green LED for "power On" is lit up (this can be found below
the screen on the left side of the front screen panel). If the power on LED
appears to be lit and the screen is not on, call XLTEK service.
3. Check the fuse which is located beside the ON/OFF switch on the back
panel. If the fuse needs replacing, replace only with two T3.15A/250 V fuse.
Page 111
Xltek NeuroMax
There are also a few NeuroMax error messages which may appear from time to
time when you are using the NeuroMax EMG system. Most of these messages
have simply been designed to warn you that less-than-perfect testing conditions
exist, or that you must do something to continue in a test mode. If any of these
messages are unclear, please call your XLTEK representative. The following
messages are presented in alphabetical order.
-OR-
Page 112
Xltek NeuroMax
2. Press TEST MENU, and re-select the current test to acquire a new set of
data while automatically storing the old set of data.
Once again this error message appears simply to let you know that you must
proceed in a particular fashion, so that no test data is ever inadvertently erased.
"Printer Error"
This message only appears when you attempt to either print a screen copy or a
final report. The printer is either not connected to the NeuroMax, or it is not
powered on. When the NeuroMax is initially powered on, AC power is
immediately supplied to the printer; however, if for some reason the printer is
turned off during testing, you must press the printer's POWER ON button to re-
initialize the printer.
"Out of Range"
Page 113
Xltek NeuroMax
You have either entered an invalid number for the number of pulses in the train
or You have entered an invalid number for the stimulus frequency. The
maximum number of pulses allowed in any single train is 60 pulses; the
minimum number is 2 pulses. Therefore the number of pulses that can be
entered is between 2 and 60 pulses. The stimulus frequency range is nominally
0.0 Hz to 50.0 Hz; the number entered must be lower than 50.0. You may
choose any value below 50.0, accurate to one decimal place (i.e. 27.8 Hz is
acceptable).
For patient safety and comfort however, the maximum stimulus frequency is
also dependent upon the pulse duration chosen, such that at higher stimulus
pulse durations, the maximum frequency available is decreased (see Table 9.1
below).
Table 9.1 - Maximum Stimulus Frequencies at all Pulse Durations
Page 114
Xltek NeuroMax
appear. Press the left or right arrows of the SELECTOR SWITCH and see that
as you highlight different tests for the selected nerve, the error message will
appear and disappear, depending on which test is highlighted. As an example, if
you choose RADIAL nerve and then highlight "H Reflex", the error message
appears, because no H reflex test has been defined for the radial nerve. If you
now press the left arrow three times to move to SNC, this test has been defined,
so the error message will disappear.
Page 115
0
Encl osure/Basic insulation
11.
EMG
Type BF
Speaker
Applied Part
Digital I/ O EMG Input
Xltek NeuroMax
Line Out
Audio Am plifie r / Stim 1
Headpho ne Accessorie s Type BF
Applied Part
Hand sw itch Digital I/ O St im 1Ou t
Mute Prob e Stim 2
Type BF
Applied Part
Meta l Shield Parts with Protective Earth
Digital I/ O St im 2Ou t
Video Card
Video Output
(Neuro only)
Ou tput to Miniscreen
PC/104 Plus Bus Ty pe F
Et hernet
Applied Part
Rear Panel DSP
USB1/U SB2 I/O
PC/104 Plus Bus
USB3
BLOCK DIAGRAM
Auxiliar y
Power Disp lay
Outlet
Fl
exible cord Backlight
Po wer: Po wer: Power
wi thMains
Primary Secondar y Distributio n
Plug
Page 116
Switch Fuse s
Reinforced insulation Protect ively Earthe d
between the primary Intermediate Ci rcuits
and seco ndary of the
po wer suppl y
Protec tive
Earth
Metal Shield Parts wi thProtec tive Eart h
Potent ia l
Equalization
Terminal Re v A 8-25-2009
Xltek NeuroMax
3.3V/5V
VCCA HR1x3(3.3V:Close;5V:Open)
CON1
3
2
1
L1 1
V_CAL R27 R25 R24 CAL_SIG RST
POWER
2 3V3
22µH/±10%/1.7ohm/80mA(LB2012T220K/8030984) 3
10k/1% 1M/1% 1M/1% 4
5V
C27
C26
C25
CAL voltage d ivider 1:20000 CAL_SIG should be PWM output GND
1 5
Square wa ve Ca li bration Signal GND
2 R26 When 5V p ower suppl y i s used: Z1 PWR 6 VIN
100R/1% Frequency is set to be approx: (10-14)Hz
GYX-SD-TC0805SURK(RED)
100nF/X7R/10%
5Vp-p +/-10% => 250µVp-p +/-10%
10µF/6.3V/0805
10nF/X7R/10%
HR1x2(Open)
BZV55C6V2 NA(PN1X6)
When 3.3V power sup pl y is use d:
AIN_SEL CON2
GND_CAL R28 3.3Vp -p +/-10% => 1 65µVp-p +/-10 % R29 2 1 1 A0
2k/1% 4 3 2
ANALOG
10k/1% A1
6 5 3 A2
8 7 4 A3
V_REF 10 9 5 A4
12 11 6
A5
HR2x6(1-2:Close) NA(PN1X6)
CON3
1
D0
2 D1
D4/D9 3 D2
DIGITAL
HR1x3(D4:Open;D9:Close) 4 D3
VCCA D4 5
D4
6 D5
D9 7 D6
8 D7
3
2
1
VCCA VCCA R19 NA(PN1X8)
330R CAL_SIG CON4
C10 C22 1 D8
3V_REF_ADC 2
D9
3 D10
DIGITAL
100nF/X7R/10% 100nF/X7R/10%
R22 R20 4
SR1
C21
C20
C19
C18
REF_E D11
8
8
IC2A 3 10k/1% 2k/1% C17 5
HR1x2(Open) D12
C11 1.5V_REF_BUF 1 C 6
D13
2 1.5V_REF 7
1uF/X5R/10%
10nF/X7R/10%
100nF/X7R/10%
100nF/X7R/10%
10µF/6.3V/0805
TLC277IDR GND
100nF/X7R/10% 3V_REF_ADC
1
2
R 8
C23 AREF
4
4
A
C12 100nF/X7R/10%
R23 R21 NA(PN1X8)
AP431SA
10k/1% 10k/1%
100nF/X7R/10% C24
+
V_REF V_REF
47µF/10%/6.3V/TANT(C)/ESR/0.25ohm(TR3C476K6R3C0250/1754044)
Instrumental Amplifier High-Pass filter OAmp with regulated gain! High-Pass filter 3rd order "Besselworth" filter, fc = 40Hz.
1 pole G=1+R12/(TR1+R11) 1 pole
G=5*(1+R8/R7)=10 G= (R17/R16)+1= 3.56
High Voltage protection HF rejection Fc=0.16Hz G=5,76...101 Fc=0.16Hz
R5 NA(3.3M) V_REF
R7 10k/1% R8 10k/1%
R1 R3 2k/1% C28 NA C13 10nF/X7R/10%
CH1_IN-
2k/1% IC1A
U1 R30 R31 IC1B
C2
PGS1
VCCA C7
100pF/COG/5%
1 1uF/X5R/10% 2 7 R18
C1
A3
BAV199(1156415)
3
VIN- RG 8 1uF/X5R/10% R16 6
VIN+ SHUTDOWN 4 C29 NA 3.9k/1%
C3
BAV199(1156415)
2 2k/1% 10nF/X7R/10% 7 R12 100k/1%
100pF/COG/5%
C14
CH1_IN+
1-L R11
2
V_REF D2 1M/1% 3 20k 1 C8 1nF/X7R/10% 1M/1%
SCJ325P00XG0B02G V_REF INA321EA
10nF/X7R/10% C15 1nF/X7R/10% C16
R2 R4 2k/1% C4 100nF/X7R/10% 3 1 1k/1%
1uF/X5R/10%
2k/1% R6 NA(3.3M) V_REF C5 1uF/X5R/10%
IC2B
6 R9
TLC277IDR
7
5
100k/1%
Page 117
Xltek NeuroMax
INDEX
EOG Protocol
configuring ........................................................... 72
A EOG Test
Administrative Functions ......................................... 83 about .................................................................... 72
AEP EP Test
impedance check................................................ 77 setting up ............................................................. 51
protocols .............................................................. 77 ERG Protocol
AV Stim configuring ........................................................... 71
cautions ................................................................ 75 ERG Test
AV Stim 1000 ........................................................... 73 about .................................................................... 71
calibration & maintenance ................................. 76
connecting ........................................................... 74 F
front panel ............................................................ 73
rear panel............................................................. 73 Files
warnings............................................................... 75 RTF ...................................................................... 88
warnings & cautions overview .......................... 75 Function Keys .......................................................... 25
F-Wave Test
about .................................................................... 40
B hints & features................................................... 43
Batch Print ................................................................ 86 performing ........................................................... 41
Blink Reflex Test
about .................................................................... 57 H
conducting ........................................................... 57
setting up ............................................................. 57 Hot Keys ................................................................... 27
H-Reflex Response
acquiring .............................................................. 44
C H-Reflex Test
setting up ............................................................. 44
Calibrating
NeuroMax .......................................................... 115 HRV Test
about .................................................................... 61
stimulator ........................................................... 115
conducting ........................................................... 61
Cautions .................................................................... 20
setting up ............................................................. 61
Control Keys ............................................................. 24
D I
Incremental Stim Test
Defaults
about .................................................................... 58
editing test ........................................................... 81
conducting ........................................................... 60
setting test ........................................................... 80
setting up ............................................................. 58
values................................................................... 60
E In-Service Checklist ................................................ 90
Introduction............................................................... 13
EMG
acquiring .............................................................. 46
activating triggers ............................................... 49 K
analyzing motor units ......................................... 50
free run ................................................................. 48 Key Pad .................................................................... 24
saving a test ........................................................ 50
triggered features ............................................... 49 M
turns & amplitude................................................ 48
EMG Notepad Macro
edit ........................................................................ 87 creating ................................................................ 96
EMG Suite using pre-defined ............................................... 97
creating .......................................................... 81, 95 Main Menu ................................................................ 23
using ..................................................................... 95 Manual
EMG Tests using..................................................................... 14
about .................................................................... 45 Memory Management ............................................. 85
setting up ............................................................. 45 Multi-Channel EMG/IOM ........................................ 64
Page118
Xltek NeuroMax
S W
Select Keys ............................................................... 25 Warnings................................................................... 18
SEP Test ................................................................... 53 Warnings & Cautions
protocols .............................................................. 54 about .................................................................... 17
Site Name List Waveform
edit ........................................................................ 87 checking ............................................................ 116
Software Welcome ................................................................... 12
checking version ................................................. 93
updating ............................................................... 93
Page119
A Total Service Solution