Grids and Scatter Reduction

Download as pptx, pdf, or txt
Download as pptx, pdf, or txt
You are on page 1of 79

Grids and Scatter Reduction

LECTURE NOTES

BY
Ugwoke Eze, B.Sc., (Hons) Rad., M.Sc., Ph.D., (Nig) (Medical Imaging).

(Professor of Radiography and Medical Imaging)


 University of Nigeria, Nsukka  
Objectives
• Upon completion of this chapter, the student will be able to:
• 1. Describe the effect of scatter on radiographic contrast.
• 2. Identify the factors that affect the amount of scatter.
• 3. Identify methods of scatter reduction.
• 4. Describe the construction of an anti-scatter grid.
• 5. Identify the types of grids.
• 6. Explain the types of grid errors.
Key Terms
• ● air gap technique
• ● Bucky factor
• ● Bucky grids
• ● crossed grids
• ● exit radiation
• ● focused grids
• ● grid
• ● grid cutoff
• ● grid frequency
• ● grid ratio
• ● oscillating grid
• ● parallel grids
Introduction
• Most of the x-ray photons entering a patient undergo Compton scattering
before they exit the patient.
• Only those photons that do not have an interaction in the patient are
useful in forming the diagnostic image.
• Scattered photons contain no useful information and create fog on the
image, which decreases the contrast of the image.
• Contrast is one of the most important characteristics of image production.
• Contrast provides the many shades of gray which make a manifest image
possible.
• Grids are the most widely used anti-scatter devices in radiology because
they eliminate scatter before it can reach the film.
Purpose of the Grid
• A grid is a device used to absorb scatter radiation before it can reach
the image receptor and fog the film.
• Grids are used with radiographic exams which require a high kVp
setting;
• this high kVp causes a high intensity beam and results in more
Compton interactions.
• Controlling the amount of scatter radiation that reaches the image
receptor controls the contrast in the image.
• Therefore, grids clean up scatter, which improves image contrast and
makes the manifest image visible.
Exit Radiation
• Exit radiation is a combination of transmitted and scattered radiation that
passes through the patient.
• Transmitted radiation undergoes no interaction in the patient.
• The transmitted radiation passes through the patient with no change in
direction or loss of energy.
• Scattered radiation has undergone at least one Compton scattering interaction.
• The Compton interaction causes the scattered radiation to change directions
and lose energy before leaving the patient.
• Scattered radiation has lower energy and is emitted in all directions from the
patient.
• The scattered radiation is the source of exposure to personnel in the room
during fluoroscopy and portable examinations.
Effect of Scatter on Contrast
• Scattered radiation reduces radiographic contrast by adding a general
background density over the entire image.
• Because the photons have changed direction they no longer relay
anatomic information to the image receptor.
• For this reason, scattered radiation provides no diagnostic information.
• The presence of scatter lowers image contrast which creates a long
scale of gray that makes objects more difficult to see in the
radiographic image.
Factors That Affect the Amount of Scatter
• Scatter depends on patient thickness, x-ray photon energy, and field
size.
• Scatter does not depend on SID or focal spot size.
• The radiographer can change the x-ray photon energy by changing the
kVp and the field size by adjusting the collimator to reduce scatter.
• Careful selection of the field size and kVp can significantly reduce the
amount of scatter and improve image quality.
Patient Thickness and Atomic Number
• The patient is the source of scatter which is degrading the image.
• An increase in tissue thickness increases the amount of scatter
because there are more atoms in the thicker tissue available for
interactions.
• The greater number of these interactions produces more scatter.
• It is sometimes possible to reduce the patient or part thickness and
reduce the amount of scatter with some examinations.
• Compression in mammography for example, reduces tissue thickness
and scatter.
Patient Thickness and Atomic Number 2
• The grid is designed to absorb the unwanted scatter radiation that occurs
with larger, thicker body parts and with procedures that use higher kVp
techniques.
• Radiographers must keep in mind the following to determine if a grid is
needed for a given procedure:
• 1. Body part thickness >10 cm
• 2. kVp above 60
• When either of these factors applies to the patient, a grid must be used to
clean up scatter radiation.
• Another source of scatter radiation is the type of tissue being irradiated.
Patient Thickness and Atomic Number 3
• Tissue with a higher atomic number will absorb more of the x-ray
beam than tissue with a lower atomic number.
• Bone is an example of high atomic number tissue.
• Bone will absorb more of the beam and produce less scatter because
bone absorbs more photons photoelectrically.
• Soft tissue has a lower atomic number and less ability to absorb the
photons, which then creates more Compton scatter radiation.
X-ray Beam Energy
• kVp is the factor that controls x-ray energy or the penetrability
• of the beam.
• Increasing the kVp results in more forward scatter exiting the patient
and striking the image receptor.
• Decreasing the kVp decreases the x-ray beam energy and the amount
of scatter.
• However, lower energy x-rays have decreased penetration and result in
higher patient doses because more x-rays are absorbed in the patient.
• The kVp selected must be tailored to the body part under examination.
Field Size
• Images of smaller fields have less scatter because there are fewer
interactions taking place.
• Larger field sizes produce more scatter because the larger area results
in more tissue being irradiated.
• Decreasing the field size decreases the area of tissue available for x-
ray interactions.
• Smaller field sizes result in less scattered radiation and higher image
contrast.
Collimation
• The purpose of collimation is to define the size and shape of the
primary x-ray beam striking the patient and
• to provide a visible light field which outlines the x-ray field.
• A light-localizing collimator consists of two pairs of lead shutters that
are adjusted to intercept x-ray photons outside the desired x-ray field.
• The top pair of shutters absorbs off-focus radiation as it leaves the
anode.
• The bottom set of shutters has two sets of shutters which allow the
radiographer to independently adjust the longitudinal and transverse
edges of the field.
Collimation contd
• This allows for infinite possibilities when matching the field size to the
patient’s anatomy.
• Some collimators have an iris-like shutter that can approximate a circular
field.
• A light source is located off the x-ray beam axis and a mirror directs the
light through the shutters.
• The x-ray beam passes through the mirror with very little attenuation.
• The distance from the mirror to the light source is equal to the distance
from the mirror to the tube focal spot, so the light and radiation fields are
the same distance from the patient surface.
• The collimator adjustment controls have indicators to show the field size in
centimeters or inches at different SIDs.
Collimation contd
• The size of the light field should never exceed the size of the image receptor
since this would cause primary radiation to directly strike the table and create
more scatter.
• Positive beam limitation (PBL) collimators automatically adjust the x-ray beam
to the size of the image receptor.
• Sensors in the cassette holder (sometimes called the Bucky tray) detect the size
of the image receptor and adjust the collimator shutters to match the cassette
size.
• A PBL collimator prevents selecting a field larger than the image receptor.
• The field size can be reduced to limit the field to the area of interest and
decrease scatter.
• Limiting the field size also improves image quality and minimizes patient dose.
Scatter Reduction Techniques
• The two most important methods of reducing scatter are beam
collimation and grids.

• As previously stated, reducing scatter radiation improves radiographic


contrast.

• Reducing the field size also reduces the amount of scatter radiation
produced.
Grid Construction
• In 1913, Dr. Gustave Bucky, an American radiologist, designed a grid
in an attempt to remove scatter and improve contrast.

• The first grid was a rather crude design with strips running in two
directions which left a checkerboard pattern superimposed over the
patient’s anatomy.

• Despite the checkerboard artifact, the grid removed scatter and


improved the contrast of the image.
Materials
• A grid is a thin, flat, rectangular device that consists of alternating
strips of radiopaque and radiolucent materials.
• The radiopaque material is usually lead foil due to its high atomic
number and increased mass density.
• Lead foil is easy to shape and is relatively inexpensive;
• these properties along with the high atomic number make it the best
material to clean up scatter.
• A grid absorbs scatter from the exit radiation before it reaches the
image receptor.
• The grid is located between the patient and the image receptor.
Materials contd
• Scattered x-ray photons scatter in various angles as they leave the
patient and are preferentially attenuated by the lead foil strips, because
they are not parallel to the grid interspaces.
• The interspace material, which is the radiolucent space between the
lead strips, is made of plastic fiber or aluminum.
• Aluminum has a higher atomic number and produces less visible grid
lines on the image.
• The use of aluminum increases the absorption of primary x-ray
photons in the interspace, especially at low kVp.
• Higher mAs factors are required to maintain image density but
increase patientbdose by approximately 20%.
Materials contd
• Plastic fiber interspace materials are preferred to aluminum interspace grids
because there is no need to increase patient dose with these types of grids.
• There are positive and negative attributes to each type of interspace
material, therefore both are manufactured.
• The radiolucent interspace materials allow transmitted x-rays to reach the
image receptor while intercepting or absorbing scattered x-ray photons.
• Only x-ray photons that are parallel or almost parallel with the interspaces
pass through the grid.
• When to use a grid is a matter of professional judgment.
• As a general rule, grids are employed when the body part is >10-cm thick.
Grid Ratio
• The grid ratio has a major influence on the grid’s ability to clean up scatter
and improve contrast.
• The amount of scattered radiation removed by the grid depends on the height
of, and the distance between, the lead strips.
• The amount of scatter eliminated, often called the scatter cleanup, depends
on the grid ratio.
• The grid ratio is the ratio of the height of the lead strips to the distance
• between the lead strips or the thickness of the interspace.
• The formula to determine the grid ratio is:
• GR = h/D
• When the height of the grid is constant, the grid ratio can be increased by
decreasing the space between the lead strips.
Grid Ratio contd
• To decrease the grid ratio of a grid with a constant height, the lead
strips would be placed farther apart.
• High grid ratios are made by reducing the width of the interspace
material, increasing the height of the lead strip, or using both methods.
• With high grid ratios higher exposure factors are required to permit a
sufficient number of x-ray photons to reach the image receptor.
• The higher ratio grids result in higher patient dose.
• Grid ratios typically range from 5:1 to 16:1.
• Grids with higher grid ratios remove more scattered radiation but are
much more difficult to align properly.
Grid Ratio contd
• More attention must be paid to make sure that the grid and x-ray tube
are properly aligned to avoid errors in positioning.
• For this reason, portable examinations are usually taken using grid
ratios of <12:1.
• General radiography examinations utilize an 8:1 or 10:1 grid ratio.
• The high-ratio grid absorbs scatter that is at a small angle to the lead
strip, meaning that less scatter reaches the image receptor.
CRITICAL THINKING
• A grid is made of lead with 40 mm thick placed between aluminum
interspace material 350 mm thick.
• The height of the grid is 3.5 mm.
• Determine the grid ratio. (Hint: first change the grid height to
micrometers.)
• Answer
• GR = h/D
• GR = 3,500 mm/350
• GR = 10:1
Grid Frequency
• The grid frequency is the number of lead strips per centimeter or inch.
• Grids with thinner strips have higher grid frequencies because the lead strips
are closer together.
• The lead strips of a high frequency grid are less visible on the radiographic
image.
• The higher the grid frequency the thinner the strips of interspace material
must be and the higher the grid ratio.
• Typical grid frequencies range from 60 to 200 lines per inch.
• As the grid frequency increases, greater technical factors are required to
produce an image with sufficient density; this results in greater patient dose.
• This occurs because as the grid frequency and grid ratio increase, there are
more lead strips absorbing the photons, which means fewer photons are
Types of Grids
• There are two types of grids, parallel and focused.
• Parallel grids have parallel lead and interspace strips running
• parallel to each other.
• Parallel grids are sometimes called linear grids because the lead and
interspace strips run in one direction.
• Focused grids have the leadstrips that are parallel at the center of the
grid, and as the strips move away from the center of the grid, they
become progressively more angled.
• If imaginary lines were extended from the lead strips toward a fixed focal
distance, the lines would meet at a point, and this is called the
convergence line.
Types of Grids contd
• The distance from the front surface of the grid to the convergence line
is called the grid radius.
• The focused grid is designed to match the divergence of the x-ray
beam.
• This distance is known as the focal distance of the grid.
• The divergent rays transmitted from the x-ray source pass through the
focused grid interspaces while the scattered x-rays are intercepted.
• When using a focused grid, the x-ray tube must be located along the
length of the strips.
Types of Grids contd
• If the tube were placed so that the divergence of the beam ran
perpendicular to the strips, grid cutoff would occur.
• Grid cutoff can affect a portion of the image or the whole image and
results in reduced density or total absence of film exposure.
• The term defines what occurs to the primary x-ray photons, they are
cut off from the image receptor.
• Grid cutoff can also occur with parallel grids if the tube and grid are
misaligned.
Parallel Grids
• Parallel grid cutoff arises because parallel grids are constructed with
the lead strips parallel to the central axis of the x-ray beam, but x-ray
photons diverge from the focal spot.
• Grid cutoff occurs because x-rays near the edge of the field are not
parallel to the lead strips and are attenuated.
• Parallel grid cutoff is greatest when the grid is used with short SID or
with a large image receptor because the x-ray beam has a wide
divergence at a shorter SID.
• The pronounced angle of divergence will cause more of the primary
beam to be attenuated in the lead strips and grid cutoff will be seen
along the outer edges of the image.
Parallel Grids contd
• Parallel grids should be used with smaller field sizes and longer SIDs
to reduce grid cutoff.
• The divergence of the beam is less with longer a SID, in other words,
the beam will be straighter.
• This works well as the beam is more parallel to the grid strips and less
attenuation will take place.
• Parallel grid cutoff produces a film which has the correct intensity in
the center but is lighter at both edges.
Focused Grids
• Focused grids eliminate grid cutoff at the edges of the field because the
lead strips are angled toward the center and converge at the focal
distance.
• When the x-ray source is in line with the center of the grid and located at
the grid focal distance, there is no grid cutoff because the transmitted
radiation passes through the radiolucent interspaces.
• Focused grids are recommended for examinations that must use large
fields or short SIDs.
• An example would be abdominal imaging that requires both the large
field size and short SID to produce an acceptable image.
• Typical focused grid distances are 100 and 180 cm (40 inch and 72 inch).
Focused Grids contd
• Each focused grid must be used with the appropriate focal range for which it
was designed.
• The focal range includes short, medium, and long focal ranges where each is
designed to be used with a specific SID.
• Mammography uses the short focal range grid, whereas chest radiography
requires the long focal range grid.
• Focused grids can be used at distances within about 13 cm of the focal
distance with no noticeable grid cutoff.
• Low grid ratio focused grids allow more latitude or leeway in the alignment of
the tube to the grid before grid cutoff occurs.
• The higher the grid ratio, the less latitude the radiographer has with grid and
tube alignment.
Focused Grids contd
• Focused grids that are used at other than the proper SIDs will show grid
cutoff.
• Focused grids used outside the focal range show a decreased intensity toward
both edges.
• Focused grids located off center or not perpendicular to the central ray show
reduced intensity on only one side of the image.
• The images in Figure 9 were taken with a 8:1 focused grid at 40 inch SID;
image A is centered appropriately, while image B is off center 6 cm.
• Although the grid cutoff is subtle, there is a noticeable blurring of bony
markings.
• The images demonstrate the increased degree of grid cutoff when the tube is
off-centered from the grid.
Focused Grid cutoff

• Figure 9. Grid cutoff. (A) 40 inch SID on center. (B) 40 inch


SID, 6 cm off laterally.
Crossed Grids
• Crossed grids overcome the limitation of linear grids where the grids clean up scatter
in only one direction.
• These grids are made by placing two linear grids on top of each other with the grid
lines perpendicular to each other.
• Crossed grids are more efficient than linear grids at cleaning up
scatter.
• The crossed grids will clean up at least twice the amount of scatter when compared to
a linear or parallel grid.
• An 8:1 crossed grid will clean up more scatter than a 16:1 linear grid.
• The crossed grid is constructed by placing two 8:1 linear grids together.
• Although these grids are capable of cleaning up more scatter, the radiographer must
use precise positioning of the center of the grid to the center of the x-ray beam.
• Grid cutoff will occur if the crossed grid and x-ray beam are not perfectly lined up.
Stationary Grids
• Stationary grids are used in radiography departments for mobile
examinations, upright imaging, or horizontal beam views.

• Most radiography departments have a supply of stationary grids, some


are directly mounted to the front of a cassette, while others are
specially designed cassettes with the grid built-in called a grid
cassette.

• When using stationary grids, the radiographer must be aware of the


grid ratio and type of grid, whether linear or focused.
Grid Movement
• Stationary grids with low grid frequencies produce noticeable grid lines on the final
image.
• One way to eliminate these grid lines is to move the grid during the exposure.
• This motion of the grid blurs out the grid lines, so they are not
• noticeable.
• Moving grids are termed Bucky grids, named after one of the inventors, Dr.
Gustave Bucky.
• Bucky’s grid design was improved upon by Dr. Hollis Potter, a Chicago radiologist,
in 1920 when he placed the strips in one direction, made the strips thinner and
designed a device that allowed the grid to move during the exposure.
• The device is called the Potter-Bucky diaphragm, Bucky diaphragm, and Bucky
grid.
• The design is still in use today.
Reciprocating and Oscillating
• Bucky grids are located directly under the table yet above the image
receptor.
• Moving grids are typically focused grids which move when the
exposure is being made.
• There are two types of moving grids used today:
• ● Reciprocating grid: the reciprocating grid is driven by a motor.
• During the exposure, it moves back and forth multiple times.
• The grid moves no more than 2 to 3 cm at a time.
• A selector at the control panel activates the grid motor.
• If the motor is not activated, the grid is stationary during exposure and
Reciprocating and Oscillating movements
contd
• ● Oscillating grid: the oscillating grid is suspended in the center of a
frame by four spring-like devices.

• At the time of the exposure, an electromagnet pulls the grid to one side
and then releases it.

• The grid will oscillate in a circular motion within the grid frame for
approximately 20 to 30 seconds before ceasing motion.
Moving Grid Disadvantages
and Advantages
• The early use of grids demonstrated grid lines or the checkerboard
pattern largely due to the large, thick strips and interspace material.
• Although cleaning up scatter and improving the contrast of the image,
the grid lines were not acceptable as they distracted from the image.
• The moving grid was designed to remove the grid lines;
• however, there were some disadvantages with the design and
placement of moving grids:
Disadvantages

• Grid mechanism: mechanical mechanism that was subject to failure.


• ● Increased OID: the distance between the patient and film was increased
due to the size of the grid mechanism.
• The increased OID creates magnification and blurring of the image.
• ● Motion: the nature of the moving grid increased the motion of the film
holder, if not operating perfectly the result would be additional image blur.
• ● Longer exposure time: the exposure time must be lengthened to allow the
motion of the moving grid to have the intended effect of removing the grid
lines.
• For some patients, this may be detrimental due to their inability to hold
still or to hold their breath and the entire exposure.
Advantages
• The advantages of the moving grid have greatly improved overall image
quality and far outweigh the disadvantages.
• Advantages of the moving grid include the following:
• ● Motion blur: grid mechanisms which are operating properly will
completely blur out grid lines, making the overall image more diagnostic.
• ● Use in radiography: the moving grid is consistently used in radiography for
body parts which measure >10 cm.
• This industry standard allows consistent imaging from one facility to the next.
• Grids are very useful in radiographic exams where the tissue thickness is
above 10 cm, as the tissue thickness increases so does the amount of scatter.
• A skilled radiographer will be able to determine the proper grid for the
examination that will be performed.
Grid Selection
• Selecting the correct grid for a specific procedure requires consideration for
the type of examination and the amount of kVp to be used.
• Examinations that require kVp settings over 95 kVp require a high-ratio
grid for maximum cleanup of scatter; higher kVp results in increased
amounts of scatter.
• High-ratio grids are more efficient at absorbing scatter, which results in less
exposure to the image receptor.
• For some examinations, this could have a negative effect on the image, so
with high-ratio grids, the exposure factors will need to be increased.
• mAs is typically increased to maintain image density and gray scale,
• but the payoff is increased patient dose.
• In other words, the more efficient the grid is at cleaning up scatter, the
Bucky Factor
• The Bucky factor measures how much scatter is removed by the grid and how
the technique factors must be adjusted to produce the same optical density.
• Scattered radiation accounts for a portion of the density on the final
radiograph.
• If a grid removes some scatter, the exposure factors must be increased to
compensate for the decrease in x-ray photons reaching the image receptor.
• The Bucky factor or grid conversion factor (GCF) is the ratio of the mAs
required with the grid to the mAs without the grid to produce the same optical
density.
• The Bucky factor is always >1.
• The Bucky factor depends on the grid ratio and the grid frequency but is
usually in the range 3 to 5.
Bucky Factor contd
• This means that adding a grid requires an increase in mAs by a factor of
3 to 5 to obtain the same optical density compared to a non-grid
technique.
• The use of a grid increases the patient dose by the Bucky factor.
• The Bucky factor is used to calculate the necessary change in mAs when
a grid is added or when changing to a grid with a different grid ratio.
• The GCF will increase with higher grid ratios and higher kVp settings.
• The Bucky factor is mathematically represented as:
• GCF = mAs with the grid / mAs without the grid.
TABLE 12.1 GRID RATIOS AND
ASSOCIATED BUCKY FACTORS OR GCF
GRID RATIO Bucky Factor or GCF
None 1
5:1 2
6:1 3
8:1 4
10:1 or 12:1 5
16:1 6
CRITICAL THINKING
• A satisfactory AP knee radiograph was produced using 7 mAs at 75
kVp without a grid. A second image is requested using an 8:1 grid.
Using Table 12.1, what mAs is needed for the second image?
• Answer
• GCF = mAs with the grid / mAs without the grid.
• 4 = x mAs/7mAs
• xmAs = 4 x 7 = 28 mAs
Required Change in mAs
Following a Change of Grids
• If a grid with a different grid ratio is used in a follow-up examination,
the change in mAs is given by the ratios of the Bucky factors:
• mAs2 = mAs1[GCF2/GCF1]

• where mAs1 is the original mAs, mAs2 is the new mAs,

• GCF1 is the original GCF, and GCF2 is the new GCF.


CRITICAL THINKING 1
• An examination taken in the department used 30 mAs with a 12:1 grid.
What mAs should be used for the follow-up portable examination
taken with an 8:1 grid?
• Answer
• mAs2 = mAs1[GCF2/GCF1]
• mAs2 = 30[4/5]
• mAs2 = 24 mAs
CRITICAL THINKING 2
• A satisfactory abdominal radiograph was produced using an 8:1 grid, 25 mAs,
and 90 kVp. Due to the long scale of contrast in the image, a second image
was requested with a 12:1 grid. Using Table 12.1, what will be the new mAs
for the second image?
• Answer
• mAs2 = mAs1[GCF2/GCF1]
• mAs2 = 25[5/4]
• mAs2 = 31.3 mAs
• The second image will utilize a higher mAs setting and there will be more
cleanup of scatter resulting in an image with more contrast and less gray scale.
Required Change in mAs Following a
Change of Grids contd
• As seen in the above examples, a change in mAs is required when the
grid ratio changes.
• The major disadvantage with using a high-ratio grid is the increased
patient dose(Fig.11).
• When using low-ratio grids with low kVp, there is still a concern for
patient dose.
• The proper selection of the appropriate grid will increase image
contrast and the diagnostic quality of the image.
Grid conversion: Illustrative Radiographs

• Figure11. Grid conversion. (A) The AP image of the abdomen shows low contrast due to
no grid being used. This is an acceptable image for the soft tissue but does not demonstrate
the bone very well. (B) The same image, except an 8:1 grid, has been used. Notice the
shorter scale of gray which demonstrates the bone. (C) A 12:1 grid has been used; the scale
Required Change in mAs Following a
Change of Grids contd
• The radiographer must remember the following factors when selecting
the correct grid for the exam:

• 1. High grid ratios increase patient dose.


• 2. High kVp examinations typically use high-ratio grids.
• 3. Low kVp examinations result in increased patient dose because
more radiation is attenuated.
• 4. High kVp examinations result in decreased patient dose.
Required Change in mAs Following a
Change of Grids contd
• The utilization of grids for many examinations has required an
increase in technical factors when compared to non-grid technical
factor settings.
• The exposure time, mA, or kVp must be increased to provide adequate
optical density on the image.
• Of these factors, kVp has been standardized in many departments
• and the mAs is changed to compensate for the patient’s body habitus.
Grid Errors
• Modern x-ray rooms use the moving grid in the table and
• upright Buckys.
• Improper usage of the moving Bucky will result in a poor radiographic
image.
• Many errors occur because the design is a focus grid where the
likelihood of an error is more common.
• Grid errors can be avoided if the radiographer properly centers the x-
ray tube with image receptor at the correct SID
• and if the moving mechanism is function normally.
Grid Errors contd
• The Potter-Bucky diaphragm or Bucky is mounted underneath the table
top directly above the image receptor.
• The grid in the Bucky must move side to side or in a circular pattern
perpendicular to the lead strips to effectively blur out the grid lines.
• If the Bucky were to move in the same direction as the lead strips, the
grid lines would be visible on the image.
• Improper positioning of the grid will always produce grid cutoff.
• The grid must be placed perpendicular to the central ray to eliminate grid
cutoff.
• The central ray can be angled to the grid providing it is angled along the
long axis of the grid strips but not across or perpendicular to the lead
Grid Errors contd
• Grid cutoff rarely occurs in the radiology department where fixed
cassette holders or Bucky trays are routinely used.
• Grid alignment is more critical with high grid ratios and can be a
serious problem with portable radiographs.
• Careful positioning of grids during portable examinations is especially
important because even slight misalignments will produce noticeable
grid cutoff.
• Grid cutoff during portable examinations is a major cause of retakes.
Grid Errors contd
• Grid errors occur most frequently because of improper positioning of
the x-ray tube and grid.
• The grid will function correctly when the x-ray tube and grid are
precisely lined up with each other.
• When the radiographer is not careful and misaligns the tube and grid,
the following errors will occur:
• ● Off-level error
• ● Off-center error
• ● Off-focus error
• ● Upside down error
Off-level Error
• An off-level error will occur when the x-ray tube is angled across or perpendicular to the
grid strips.
• This can result from improper tube or grid positioning.
• Improper tube positioning occurs when the central ray is directed across the long axis of
the table.
• Improper grid positioning most commonly occurs with stationary grids which are used
for mobile procedures or decubitus imaging, for example, when a patient is lying on a
grid for a mobile pelvis examination and the patient’s weight is not evenly distributed on
the grid causing the grid to angle underneath the patient.
• When the vertical x-ray beam is aligned to the angled grid, off-level grid error will
occur.
• The image will demonstrate a decrease in density over the whole image.
• Figure12 illustrates the angle of the grid compared to the vertical beam and the resultant
radiographic image.
Off-level Error Illustrated

• A radiographic image demonstrating off-level grid error.


Off-center Error
• The x-ray tube must be centered along with the center of a focused
grid to prevent off-center error, also called off-axis and lateral
decentering grid error.
• The center lead strips in a focused grid are perpendicular and the lead
strips become more angled away from center as the strips get closer to
the edges of the grid.
• The focused grid is designed to match the divergence of the x-ray
beam.
• When the x-ray tube is off-centered laterally, the perpendicular portion
of the x-ray beam will intersect the angled grid strips causing a
decrease in exposure across the image.
Off-center Error contd
• As demonstrated in Figure13, the divergence of the beam will not line
up with the angle of the lead strips.

• This grid error can be avoided if the radiographer correctly places the
x-ray tube in the center of the grid;

• in some equipment, the tube will lock in place when correctly


positioned to the detent in the middle of the table.
Off-center grid error illustrated

• A radiographic image demonstrating off-center grid error.


Off-focus Error
• The off-focus error results when the focused grid is used with a SID
that is out of the focal range and not specified for the grid.
• Figure14 illustrates what happens when a focused grid is not used in
the proper focal range.
• Unlike the other grid errors, off-focus grid errors are not uniform
across the entire image, rather there is severe grid cutoff at the
periphery of the image.
• Positioning the grid at the proper focal distance is more crucial with
high-ratio grids because these grids have less positioning latitude than
low-ratio grids.
Off-focus grid error illustrated

• A radiographic image demonstrating off-focus grid error.


Upside-down Error
• This type of error is readily seen and identified immediately.
• When the grid is placed upside down, the lead strips are not angled
toward the center of the grid.
• The result will be severe grid cutoff on either side of the center of the
image.
• The x-ray beam will pass through the central axis of the grid and will be
attenuated by the lead strips that are angled in the opposite direction of
the beam divergence (Fig.15).
• Focused grids are clearly marked on the tube side of the grid.
• The radiographer has to merely look at the grid to know which surface
needs to face the tube.
Upside-down grid error illustrated

• A radiographic image demonstrating upside-down grid error.


GRID ARTIFACT APPEARANCES
AND THEIR POSSIBLE CHANGES
• 1. Optical Density: Correct density in the center, lower density on
both sides of image.
• Possible Causes: a. Parallel grid at too short SID; b. Upside-down
focused grid; c. Focused grid outside focal distance range.

• 2. Optical Density: Correct density in the center and on one side, low
density on one side.
• Possible Causes: a. Grid center not aligned with central axis;
b. Grid not perpendicular to central axis.
Alternative Method to Reduce Scatter
• Air Gap Technique
•The air gap technique is an alternate scatter reduction method and
can be used instead of a grid.
•The air gap technique uses an increased OID to reduce scatter
reaching the image receptor.
• This technique is used in lateral C-Spine and chest radiographs.
•Remember that the patient is the source of scatter and the increased
OID causes much of the scattered radiation to miss the image
receptor.
•This eliminates the need for using a grid to reduce scatter and to
improve the contrast of the image.
Air Gap Technique contd

• A major disadvantage of the air gap technique is the loss of sharpness, which
results from the increased OID.
• The air does NOT filter out the scattered x-rays.
• An OID of at least 6 in is required for effective scatter reduction with the air
gap technique.
• When the OID is increased, there is an increase in magnification and a
reduction in detail.
• To compensate for this, a longer SID is utilized with a small focal spot size.
• Because of the longer SID, the mAs must be increased to maintain
radiographic density.
• The patient dose does not increase because the intensity of the x-ray beam
will decrease with the longer SID.
Chapter Summary
• Radiation leaving the patient is a combination of transmitted and
scatter radiation.
• Scatter decreases contrast and depends on field size, patient thickness,
beam energy, or kVp.
• The primary methods of scatter reduction are reduction in field size
and the use of a grid.
• In special applications, such as C-spine imaging, an air gap can be
used to reduce scatter.
• Grids are constructed of alternating strips of a radiopaque material
such as lead and a radiolucent material such as aluminum or plastic.
Chapter Summary contd
• Using a grid with a higher grid ratio increases the patient dose, increases
the Bucky factor, increases the film contrast, and requires an increase in
mAs.
• The Bucky factor is the ratio of mAs with the grid to mAs without the
grid.
• High grid ratios are not used in portable imaging because the alignment
is very critical.
• The types of grid errors include off-center, off-level, off-focus, and
placing the grid upside down.
• These errors can be avoided when the radiographer properly centers the
tube with the Bucky and places the cassette properly with the tube side
of the cassette toward the patient.
Case Study
• Todd performed a portable abdominal radiograph using a 40 inch SID,
5:1 grid, 85 kVp, and 20 mAs. Upon reviewing the image, Todd
noticed that the overall appearance of the anatomy was not what he
expected.
• The image lacked sufficient contrast and had a long scale of gray
which obscured some anatomy.
• There also appeared to be less density over the whole image than what
he expected and the spine was not in the middle of the image.
• Todd determined that the image would need to be repeated, but first he
had to decide the factors that would need to be changed to produce a
diagnostic image.
Critical Thinking Questions
• 1. Did Todd use the correct grid ratio for the technical factors he used?
• 2. With a higher kVp exam, should Todd have used a high-ratio grid?
• 3. If a change in grid ratio is determined, what will the new mAs be?
• 4. How will Todd address the placement of the spine in the image and
how does this relate to the overall decrease in density on the image?
Answers
• Todd determined that using 85 kVp required him to use a higher ratio grid, as a 5:1
grid will not effectively clean up scatter when the kVp is above 80 kVp.
• He will need to use a 12:1 grid for the repeat.
• This will correct the lack of contrast by cleaning up more scatter and will produce a
shorter scale of gray.
• When converting from a 5:1 grid to a 12:1 grid, Todd will need to make an increase
in mAs.
• To determine exactly how much he will need to use the Bucky factor or grid
conversion formula,
• mAs = mAs (GCF )/(GCF ).
2 1 2 1

• 20 mAs (5/2) = 50 mAs will be the new mAs.


Answers contd
• The new factors of 85 kVp and 50 mAs will provide the necessary
amount of photons to have a lower contrast image.

• The placement of the spine indicated that the grid was not placed
completely underneath the patient, and the lack of exposure density
indicated that the tube was not centered with the grid, causing the grid
to be off-centered toward one side.

• This error caused a decrease in density over the whole image.


Answers contd
• On the repeat, Todd will need to make certain that the grid is placed
equally underneath the patient and that the tube is placed in the center
of the patient.

• This will place the spine in the center of the image and will prevent the
off-centering error which decreased the amount of exposure to the
grid.

• Todd feels confident that with these changes the next abdomen
radiograph will be a diagnostic image.
Conclusion
• Ponder over books and read voraciously if you want greatness and
note that there is no end to reading. …Ugwoke Eze.

• The more you read the more educated you will be and the more you
realize how little you know. …..Ugwoke Eze.

• Quality and comprehensive education in addition to rich native


wisdom can make one great in life. ……Ugwoke Eze

• Many thanks for Listening.

You might also like