Grids and Scatter Reduction
Grids and Scatter Reduction
Grids and Scatter Reduction
LECTURE NOTES
BY
Ugwoke Eze, B.Sc., (Hons) Rad., M.Sc., Ph.D., (Nig) (Medical Imaging).
• Reducing the field size also reduces the amount of scatter radiation
produced.
Grid Construction
• In 1913, Dr. Gustave Bucky, an American radiologist, designed a grid
in an attempt to remove scatter and improve contrast.
• The first grid was a rather crude design with strips running in two
directions which left a checkerboard pattern superimposed over the
patient’s anatomy.
• At the time of the exposure, an electromagnet pulls the grid to one side
and then releases it.
• The grid will oscillate in a circular motion within the grid frame for
approximately 20 to 30 seconds before ceasing motion.
Moving Grid Disadvantages
and Advantages
• The early use of grids demonstrated grid lines or the checkerboard
pattern largely due to the large, thick strips and interspace material.
• Although cleaning up scatter and improving the contrast of the image,
the grid lines were not acceptable as they distracted from the image.
• The moving grid was designed to remove the grid lines;
• however, there were some disadvantages with the design and
placement of moving grids:
Disadvantages
• Figure11. Grid conversion. (A) The AP image of the abdomen shows low contrast due to
no grid being used. This is an acceptable image for the soft tissue but does not demonstrate
the bone very well. (B) The same image, except an 8:1 grid, has been used. Notice the
shorter scale of gray which demonstrates the bone. (C) A 12:1 grid has been used; the scale
Required Change in mAs Following a
Change of Grids contd
• The radiographer must remember the following factors when selecting
the correct grid for the exam:
• This grid error can be avoided if the radiographer correctly places the
x-ray tube in the center of the grid;
• 2. Optical Density: Correct density in the center and on one side, low
density on one side.
• Possible Causes: a. Grid center not aligned with central axis;
b. Grid not perpendicular to central axis.
Alternative Method to Reduce Scatter
• Air Gap Technique
•The air gap technique is an alternate scatter reduction method and
can be used instead of a grid.
•The air gap technique uses an increased OID to reduce scatter
reaching the image receptor.
• This technique is used in lateral C-Spine and chest radiographs.
•Remember that the patient is the source of scatter and the increased
OID causes much of the scattered radiation to miss the image
receptor.
•This eliminates the need for using a grid to reduce scatter and to
improve the contrast of the image.
Air Gap Technique contd
• A major disadvantage of the air gap technique is the loss of sharpness, which
results from the increased OID.
• The air does NOT filter out the scattered x-rays.
• An OID of at least 6 in is required for effective scatter reduction with the air
gap technique.
• When the OID is increased, there is an increase in magnification and a
reduction in detail.
• To compensate for this, a longer SID is utilized with a small focal spot size.
• Because of the longer SID, the mAs must be increased to maintain
radiographic density.
• The patient dose does not increase because the intensity of the x-ray beam
will decrease with the longer SID.
Chapter Summary
• Radiation leaving the patient is a combination of transmitted and
scatter radiation.
• Scatter decreases contrast and depends on field size, patient thickness,
beam energy, or kVp.
• The primary methods of scatter reduction are reduction in field size
and the use of a grid.
• In special applications, such as C-spine imaging, an air gap can be
used to reduce scatter.
• Grids are constructed of alternating strips of a radiopaque material
such as lead and a radiolucent material such as aluminum or plastic.
Chapter Summary contd
• Using a grid with a higher grid ratio increases the patient dose, increases
the Bucky factor, increases the film contrast, and requires an increase in
mAs.
• The Bucky factor is the ratio of mAs with the grid to mAs without the
grid.
• High grid ratios are not used in portable imaging because the alignment
is very critical.
• The types of grid errors include off-center, off-level, off-focus, and
placing the grid upside down.
• These errors can be avoided when the radiographer properly centers the
tube with the Bucky and places the cassette properly with the tube side
of the cassette toward the patient.
Case Study
• Todd performed a portable abdominal radiograph using a 40 inch SID,
5:1 grid, 85 kVp, and 20 mAs. Upon reviewing the image, Todd
noticed that the overall appearance of the anatomy was not what he
expected.
• The image lacked sufficient contrast and had a long scale of gray
which obscured some anatomy.
• There also appeared to be less density over the whole image than what
he expected and the spine was not in the middle of the image.
• Todd determined that the image would need to be repeated, but first he
had to decide the factors that would need to be changed to produce a
diagnostic image.
Critical Thinking Questions
• 1. Did Todd use the correct grid ratio for the technical factors he used?
• 2. With a higher kVp exam, should Todd have used a high-ratio grid?
• 3. If a change in grid ratio is determined, what will the new mAs be?
• 4. How will Todd address the placement of the spine in the image and
how does this relate to the overall decrease in density on the image?
Answers
• Todd determined that using 85 kVp required him to use a higher ratio grid, as a 5:1
grid will not effectively clean up scatter when the kVp is above 80 kVp.
• He will need to use a 12:1 grid for the repeat.
• This will correct the lack of contrast by cleaning up more scatter and will produce a
shorter scale of gray.
• When converting from a 5:1 grid to a 12:1 grid, Todd will need to make an increase
in mAs.
• To determine exactly how much he will need to use the Bucky factor or grid
conversion formula,
• mAs = mAs (GCF )/(GCF ).
2 1 2 1
• The placement of the spine indicated that the grid was not placed
completely underneath the patient, and the lack of exposure density
indicated that the tube was not centered with the grid, causing the grid
to be off-centered toward one side.
• This will place the spine in the center of the image and will prevent the
off-centering error which decreased the amount of exposure to the
grid.
• Todd feels confident that with these changes the next abdomen
radiograph will be a diagnostic image.
Conclusion
• Ponder over books and read voraciously if you want greatness and
note that there is no end to reading. …Ugwoke Eze.
• The more you read the more educated you will be and the more you
realize how little you know. …..Ugwoke Eze.