04 Calibration Manual For Instruments of BalClor BWMS-REV1.0
04 Calibration Manual For Instruments of BalClor BWMS-REV1.0
04 Calibration Manual For Instruments of BalClor BWMS-REV1.0
版
1.0 2016/11/299 张晓
晓磊 李樱 徐凤麒
Initial Publish
版次 日期 起草
草 审核 批准 修改-说明
明
Rev. Date Drafteed By Revviewed By Approved Byy Moddifications-Ob
bservations
船
船舶压载
载水管
管理系
系统技
技术文件
Technical Document of Ballast Watter Managemeent System
DOC.
D
N
NO X C B W M S J Z S C W J
TITLE
T
标
标题
合 tract No.:
合同号/Cont
船 s:
船级社/Class
发布:青岛双瑞
青 瑞海洋环境
境工程股
股份有限
限公司
Issued: Sunrui Maarine Enviroonment Eng
gineering Coo.,Ltd
内部 部编码:
Interrnal No.:
本文件产权
本 权属于青岛双 双瑞海洋环 环境工程股份 份有限公司 司。未获本公司书面许 许可,任何 何人不得擅自 自
使
使用、复制 制、传递或泄 泄露该文件 件。This doccument is the t propertyy of Sunrui, It must no
ot be used、
r
reproduced、
、transmitteed or disclosed withoutt permission
n of Sunrui.
XCBWMSJZSCWJ REV1.0 Page 1 / Total21
Recommended
Possible Calibration
No. Name of Instrument Calibration
Location Method
Cycle
Refer to
2 Pressure Transmitter/Switch EDU/AFU 1 year
Page 3
Refer to
4 Temperature Sensor EDU 1 year
Page 5
5 Conductivity Meter /
Refer to
5.1 CHEMINS DDM-200 EDU 6 months
Page 8
Refer to
5.2 HACH 3433
Page 11
No need for
6 Flow Meter Ballast Pipe/EDU /
calibration
No need for
7 Magnetic Liquid Level Meter EDU/ANU /
calibration
Refer to
8 Hydrogen Sensor Near EDU 1 year
Page 15
Refer to
9 Chlorine Sensor EDU 1 year
Page 17
10 TRO Analyzer /
10.2 HF CLX-Ex/Ex2 /
Note:
a. All the instruments have been calibrated and tested before equipped in the vessel.
XCBWMSJZSCWJ REV1.0 Page 2 / Total21
2. Visual Inspection
The surface of instrument case and outer components (terminal knob, panel, switch etc.)
should be smooth and clean, nameplates and signs should be clean and bright.
3. Calibration Procedure
a. Connect the instrument with the external circuit which could reveal its on/off state.
b. Connect the pressure gauge or vacuum pressure gauge with the pressure check meter
tightly, increase the pressure slowly and smoothly, until to the specified calibration
point and hold for 3 minutes, then keep on increasing the pressure to measurement
upper limit, after that lower the pressure to Zero point smoothly and steadily.
Observe the display value of pressure transmitter at calibration point, the difference
between display value and pressure calibration point value should satisfy the
accuracy requirement.
d. Label the instrument and fill out the instrument calibration record sheet.
4. Judgment on the calibration results
The error value should be no more than the allowable error value specified in the
nameplate. If the nameplate specifies no allowable error value or accuracy level, the error
value should be no more than 1.5% of the absolute value of setting point.
5. Return to SUNRUI
If you hope to carry out the adjustment by SUNRUI or need assistance in repair, please
[email protected] and provide information about the instrument quantity and condition.
1. Instruments Preparation
2. Visual Inspection
The surface of instrument case and outer components (terminal knob, panel, switch etc.)
should be smooth and clean, nameplates and signs should be clean and bright.
3. Calibration Procedure
a. Connect the instrument with the external circuit which could reveal its on/off state.
b. Connect the instrument with the pressure check meter tightly, increase the pressure
slowly and smoothly, until to the specified calibration point and hold for 3 minutes,
then keep on increasing the pressure to measurement upper limit, after that lower the
pressure to Zero point smoothly and steadily. Observe the display value of pressure
transmitter at calibration point, the difference between display value and pressure
d. Label the instrument and fill out the instrument calibration record sheet.
4. Judgment on the calibration results
The error value should be no more than the allowable error value specified in the
nameplate. If the nameplate specifies no allowable error value or accuracy level, the error
value should be no more than 1.5% of the absolute value of setting point.
5. Return to SUNRUI
If you hope to carry out the adjustment by SUNRUI or need assistance in repair, please
[email protected] and provide information about the instrument quantity and condition.
1. Instruments Preparation
2. Visual Inspection
The surface of instrument case and outer components (terminal knob, panel, switch etc.)
should be smooth and clean, nameplates and signs should be clean and bright.
3. Calibration Procedure
a. Connect the instrument with the external circuit which could reveal its on/off state.
b. Connect the high pressure port of the pressure differential switch with the pressure
check meter tightly, and connect the low pressure port of pressure differential switch
to air. Increase the pressure slowly and smoothly (prevent from overshooting or
reverse stroke), when the pressure reach to the point which is higher than set point
and keep steady, observe the action (recovery) value of the set point, and then
decrease the pressure slowly and smoothly, when the pressure reach to the point
which is lower than set point and keep steady, observe the action (recovery) value of
d. Label the instrument and fill out the instrument calibration record sheet.
4. Judgment on the calibration results
The error value should be no more than the allowable error value specified in the
nameplate. If the nameplate specifies no allowable error value or accuracy level, the error
value should be no more than 1.5% of the absolute value of setting point.
5. Return to SUNRUI
If you hope to carry out the adjustment by SUNRUI or need assistance in repair, please
[email protected] and provide information about the instrument quantity and condition.
1 Monkey Wrench 2
2. Calibration Method
In order to calibrate the instrument, we could compare the instrument with thermometer.
3. Visual Inspection
The surface of instrument case and outer components (terminal knob, panel, switch etc.
should be smooth and clean, nameplates and signs should be clean and bright.
4. Calibration Procedure
a. Connect the instrument with the external circuit which could reveal its on/off state.
b. Fill clean fresh water into a clean container; room temperature water could be accepted.
d. Place the sensor inside the solution. Hold the sensor from touching the container.
Make sure the sensor region totally dipping inside the solution.
f. Wait until the reading of sensor is steady, place the thermometer into the solution, and wait
g. Compare the reading between temperature transmitter and thermometer, the difference
h. After the calibration, the instruments should be settled to original positions, then fill out the
The 4 to 20 mA endpoint values are independent of one another and may be adjusted in the
field. For example, to reduce the 20 mA endpoint value from the factory setting of 100°C, but to
XCBWMSJZSCWJ REV1.0 Page 6 / Total21
allow the4 mA endpoint to remain at 0°C, perform only the steps listed in 5.2 below.
NOTE: The RED wire, which is not connected during normal 4to 20 mA operations, assumes
100°C/85°C submersible). (Be sure to allow sufficient time for the sensor to equilibrate to this
temperature.)
b. With power applied as described in Picture 1 and Picture 2, connect the RED wire to the
WHITE wire for 15 seconds. (After about 10 seconds the output will drop to 3.6 mA and remain
c. Disconnect the RED wire from the WHITE wire; The 4 mA endpoint has been adjusted.
NOTE: The output will act as a switch if the 4 and 20 mA endpoints are set very near to the
same value.
100°C/85°C submersible). (Be sure to allow sufficient time for the sensor to equilibrate to this
temperature.)
b. With power applied as described in Picture 1 and Picture 2, connect the RED wire to the
BLACK wire for 15 seconds. (After about 10 seconds the output will rise to 22 mA and remain
c. Disconnect the RED wire from the BLACK wire; the 20 mA endpoint has been adjusted.
NOTE: The output will act as a switch if the 4 and 20 mA endpoints are set very near to the
c. Apply power as described in Picture 1 and Picture 2, but with the RED wire connected to
(After about 10 seconds the output will drop to 3.6 mA and remain there until the RED wire
is disconnected.)
XCBWMSJZSCWJ REV1.0 Page 7 / Total21
d. Disconnect the RED wire from the WHITE wire; factory settings have been restored.
Picture 1
Picture 2
6. Return to SUNRUI
If you hope to carry out the adjustment by SUNRUI or need assistance in repair, please
contact SUNRUI before returning the instrument. Please send E-mail to [email protected]
and provide information about the instrument quantity and condition. Please clean and carefully
1 Monkey Wrench 1
2. Visual Inspection
The surface of instrument case and outer components (terminal knob, panel, switch etc.
should be smooth and clean, nameplates and signs should be clean and bright.
a. Before testing, power on 1~2 minutes to warm-up, then testing in 35PSU standard seawater.
If there is big value difference, you can fine-tune the Slope. Ensure the value within permissible
error range.
b. Take out the sensor from the cup, cleaning in distilled water and drying with a clean cloth,
put it in 5PSUstandard seawater to test. If the value is large difference, you can fine-tune the Zero.
c. Take out the sensor from the cup, cleaning in distilled water and drying with a clean cloth;
put it in 20PSUstandard seawater to test the accuracy of the value. If the error exceeds 2.0PSU,
Notes: If the standard seawater was contained in a cup, the electrode cannot touch the wall of
the cup (including the cup’s bottom and sidewall); the distance is at least 20mm to prevent
4. Calibration Procedure
0~100PSU salinity transmitter calibration, take two-wire 4-20mA output as example, the
(Note: If thee probe expossed to the air, the showing value is invaalid.)
b. Full-scalle calibration (Transmittter has been calibrated att the factory,, don’t adjusst the
5. Return to SU
UNRUI
contact SUNRUI before returning the instrument. Please send E-mail to [email protected]
and provide information about the instrument quantity and condition. Please clean and carefully
1 Monkey Wrench 1
2. Visual Inspection
The surface of instrument case and outer components (terminal knob, panel, switch etc.
should be smooth and clean, nameplates and signs should be clean and bright.
3. Zero Calibration
Define the unique ZERO of conductivity sensor with Zero calibration standard procedure.
The Zero should be defined in advance, and then reference solution or process sample could be
1) Pull the sensor out from the process solution. Wipe the sensor with clean towel.
4) If the password in Safe Menu of controller has been set, then input the password.
OPTION DESCRIPTION
Active Send the output value of the active measurement during the calibration
procedure.
Hold Remain the output value of sensor in active measurement during the
calibration procedure.
Transfer Send the pre-set output value during the calibration procedure. Please
9) Fail — value overrun. Make sure the sensor dry, and repeat the Zero Calibration
XCBWMSJZSCWJ REV1.0 Page 12 / Total21
Procedure.
11) If the operator ID in the Calibration Options has been set to YES, input the
OPTION DESCRIPTION
Yes The sensor has not been calibrated by the controller. The running days
Calibrate and adjust the sensor reading, to compare with a reference solution. Use the
Note: If it is the first time to calibrate the censor, Zero calibration should be finished in
advance.
2) Place the sensor inside the reference solution. Hold the sensor from touching the
container. Make sure the sensor region totally dipping inside the solution (Drawing 3). Stir the
3) Wait until the temperature of sensor equal to the solution. If the temperature
difference between the process solution and reference solution are huge, this procedure may
4) Press MENU, and then select “Sensor Setup”, [“Select Sensor”], “Calibrate”.
6) If the password in Safe Menu of controller has been set, then input the password.
OPTION DESCRIPTION
Send the output value of the active measurement during the calibration
Active
procedure.
Send the pre-set output value during the calibration procedure. Please
Transfer
reference the controller User Manual to change the pre-set value.
11) Success — Sensor has been calibrated and ready to measure the sample. The
12) Fail— slope or deviation value overrun. Repeat the Calibration Procedure with
new solution. For more details, please refer to No.83 of Maintenance and No.83 of Trouble
Shooting.
14) If the operator ID in the Calibration Options has been set to YES, input the
OPTION DESCRIPTION
Yes The sensor has not been calibrated by the controller. The running days
16) Place the sensor back inside the process solution, and press ENTER. The output
XCBWMSJZSCWJ REV1.0 Page 14 / Total21
signal return to active state, and then the sample value would be show on the measurement
screen.
Note: If the output mode has been set on HOLD or CONVERT, select the delay time of the
5. Return to SUNRUI
If you hope to carry out the adjustment by SUNRUI or need assistance in repair, please
contact SUNRUI before returning the instrument. Please send E-mail to [email protected]
and provide information about the instrument quantity and condition. Please clean and carefully
Calibraation Proocedure of
o Hydroogen sen
nsor
BR
RAND: RAE
R MODEL:
M S
SP-1102
1. Instrumentts Preparatioon
No. Item
m Qty. Notee.
1 Monnkey Wrenchh 2
2. Visual Insp
pection
should be
b smooth annd clean, nam
meplates and signs should be
b clean and bbright.
b. The com
mponents shoould be clean and intact, an
nd there is noo erosion or distortion.
The detectorr already calibrated in facttory, the adju w should be done for the
ustment below
off the
t water-prooof cover for sensor,
s the dissplay value on
o controller should
s be “0000” in pure aiir,
4. Gas Calibration
After zeroing, keep the flow rate of the standard sample of the inflammable gas of known
concentration within the range of about 500ml/min with a flow meter; then, direct it to the
attached calibration gas cover through a pipe and then cover the detector sensor with it with it
ventilated for about one minute. Adjust gain potentiometer “WS” till the displayed value is the
same as the concentration value of the standard gas. Shut off the standard gas supply and take off
5. Return to SUNRUI
If you hope to carry out the adjustment by SUNRUI or need assistance in repair, please
contact SUNRUI before returning the instrument. Please send E-mail to [email protected]
and provide information about the instrument quantity and condition. Please clean and carefully
a. The surface of instrument case and outer components (terminal knob, panel, switch etc.
should be smooth and clean, nameplates and signs should be clean and bright.
b. The components should be clean and intact, and there is no erosion or distortion.
3. Zero Calibration
3.1 Preparations for zero calibration
● Confirm the instrument operating environment is an environment with clean air. Otherwise,
a clean air shall be fed into the instrument using standard air bottle through calibration hood.
recommended to carry out zero calibration after 30-minute warming up upon energization).
● Click the MODE key. Input the password (the default is 1111) through the "+" or"-" key
and clicking the MODE key, press and hold the MODE key for confirmation. If the foregoing
● Let the cursor select through the "+" or "-" key and click the MODE key to enter the
"Zero Calibration" status. The figure at the lower part of the picture means the time value to be
counted down (this value is determined based on the type of the gas sensor and is 60 in general).
XCBWMSJZSCWJ REV1.0 Page 18 / Total21
Zero Calibration
● The instrument returns to the main menu after automatic zero adjustment. At this moment,
the instrument and controller display ZERO and the instrument outputs 4mA.
4. Gas Calibration
The measurement accuracy of the detector can be ensured only after completing two-point
calibration, so it is required to carry out gas calibration after zero calibration during detector
maintenance. The gas calibration shall be carried out using standard gas with some concentration,
so it is necessary to set the concentration value of the standard gas to be used before gas
calibration
After entering the main menu (operation is the same as above-mentioned), let the cursor
select through the "+" or "-" key and click the MODE key with 4mA outputted from the
instrument to enter the "Gas Calibration" status as shown in the lower figure. The figure at the
lower part of the picture means the gas concentration value to be set. Set the value at the frame
position to the known concentration value of the calibration gas through the "+" or "-" key and
Gas Calibration
Don't carry out gas calibration operation for the detector before setting the calibration value.
Please note: two-point calibration shall be carried out as per the sequence of zero calibration →
setting of calibration value → gas calibration, especially the zero calibration must be carried out
XCBWMSJZSCWJ REV1.0 Page 19 / Total21
● A pressure relief valve is mounted on the standard gas bottle to be used and shall be a
pressure relief valve within the flow adjustment range specified in the inspection specification or a
● The standard gas hose is used to connect the pressure relief valve to the standard gas hood
b) Calibration operation
●Click the MODE key to enter "Gas Calibration “status. The figure at the lower part of the
picture means the time value to be counted down (this value is determined based on the type of the
Gas Calibration
● After automatic gain adjustment for the instrument is completed, "Calibration is successful"
will be displayed, as shown in the lower figure. The figure at the lower part of the picture means
the time value to be counted down. At this moment, the instrument display and controller shall
XCBWMSJZSCWJ REV1.0 Page 20 / Total21
indicate the standard gas concentration value within the error range and the instrument outputs the
corresponding current. If "Calibration fails" is displayed, the instrument outputs 22mA. Please
Calibration done
● Close the pressure relief valve and remove the calibration hood, meaning the calibration is
completed.
When the detector becomes a part of the system that also includes other controllers, if the
display terminal of the detector isn't consistent with controller display value due to the deviation
from detector output reference or from full-scale current or for other reason, the deviation can be
corrected through adjusting 4mA or 20mA output current. However, it is required to determine the
reason for such deviation before such adjustment. Therefore, this operation shall be completed
with the help of confirmed accurate current measuring tools; otherwise, improper adjustment will
Setting operation:
● After entering the main menu (operation is the same as above-mentioned), let the cursor
select through the "+" or "-" key and click the MODE key to enter "4mA Calibration"
status as shown in the lower figure. The current meter displays the output current value of the
detector circuit.
● Adjust the output current value to 4.00mA through the "+" or "-" key.
XCBWMSJZSCWJ REV1.0 Page 21 / Total21
4mA Calibration
● Click the MODE key for confirmation. The detector returns to the main menu and the
Setting operation:
● After entering the main menu (operation is the same as above-mentioned), let the cursor
select through the "+" or "-" key and click the MODE key to enter "20mA Calibration"
status as shown in the lower figure.. The current meter displays the output current value of the
detector circuit.
● Adjust the output current value to 20.00mA through the "+" or "-" key.
20mA
C lib i
● Click the MODE key for confirmation. The detector returns to the main menu and the
6. Return to SUNRUI
If you hope to carry out the adjustment by SUNRUI or need assistance in repair, please
contact SUNRUI before returning the instrument. Please send E-mail to [email protected]
and provide information about the instrument quantity and condition. Please clean and carefully