Turbidity & TSS Probe BA00461CEN - 1513

Download as pdf or txt
Download as pdf or txt
You are on page 1of 40

BA00461C/07/EN/15.

13 Products Solutions Services


71219749

Operating Instructions
Turbimax CUS51D
Sensor for turbidity and solids content
About this document

Notes on safety icons


The structure, signal words and safety colors of the signs comply with the specifications of
ANSI Z535.6 ("Product safety information in product manuals, instructions and other
collateral materials").

Safety message structure Meaning


! DANGER This symbol alerts you to a dangerous situation.
Cause (/consequences) Failure to avoid the situation will result in a fatal or serious
Consequences if safety injury.
message is not heeded
‣ Corrective action
! WARNING This symbol alerts you to a dangerous situation.
Cause (/consequences) Failure to avoid the situation can result in a fatal or serious
Consequences if safety injury.
message is not heeded
‣ Corrective action
! CAUTION This symbol alerts you to a dangerous situation.
Cause (/consequences) Failure to avoid this situation can result in minor or
Consequences if safety medium injury.
message is not heeded
‣ Corrective action
NOTICE This symbol alerts you to situations that can result in
Cause/situation damage to property and equipment.
Consequences if safety
message is not heeded
‣ Action/note

Symbols

Additional information, tips


Permitted or recommended
Forbidden or not recommended

Endress+Hauser
Turbimax CUS51D

Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . . . 4 10 Technical Data . . . . . . . . . . . . . . . . . . . . 35
1.1 Requirements for the personnel . . . . . . . . . . . . . . . 4 10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
1.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 10.2 Performance characteristics . . . . . . . . . . . . . . . . . 36
1.3 Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 10.3 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
1.4 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 10.4 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
1.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 10.5 Mechanical construction . . . . . . . . . . . . . . . . . . . . 38

2 Incoming acceptance and product Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39


identification . . . . . . . . . . . . . . . . . . . . . . 6
2.1 Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Product identification . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.4 Certificates and approvals . . . . . . . . . . . . . . . . . . . . 6

3 Installation . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Installation instructions . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Installation examples . . . . . . . . . . . . . . . . . . . . . . 10
3.4 Post-installation check . . . . . . . . . . . . . . . . . . . . . 14

4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Connecting to the transmitter . . . . . . . . . . . . . . . 15
4.2 Post-connection check . . . . . . . . . . . . . . . . . . . . . 15

5 Device description . . . . . . . . . . . . . . . . . 16
5.1 Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5.2 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . 16
5.3 Measuring methods . . . . . . . . . . . . . . . . . . . . . . . 19
5.4 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.6 Stability criterion . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.7 Cyclic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

6 Diagnostics and troubleshooting. . . . . 30

7 Maintenance . . . . . . . . . . . . . . . . . . . . . . 31
7.1 Cleaning the sensor . . . . . . . . . . . . . . . . . . . . . . . 31

8 Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.1 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

9 Accessories . . . . . . . . . . . . . . . . . . . . . . . 33
9.1 Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
9.2 Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
9.3 Compressed air cleaning . . . . . . . . . . . . . . . . . . . 34
9.4 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

Endress+Hauser 3
Safety instructions Turbimax CUS51D

1 Safety instructions

1.1 Requirements for the personnel


‣ Installation, commissioning, operation and maintenance of the measuring system must
only be carried out by trained technical personnel.
‣ The technical personnel must be authorized by the plant operator to carry out the
specified activities.
‣ The electrical connection may only be performed by an electrical technician.
‣ The technical personnel must have read and understood these Operating Instructions
and must follow the instructions they contain.
‣ Measuring point faults may only be rectified by authorized and specially trained
personnel.
Repairs not described in the enclosed Operating Instructions may only be carried out
directly at the manufacturer's or by the service organization.

1.2 Designated use


CUS51D is a sensor designed to measure the turbidity and solids content of water and
wastewater.
The sensor is particularly suited for use in the following applications:
• Turbidity measurement in the outlet
• Suspended solids in the activated sludge basin and in the recirculation
• Suspended solids in the sludge treatment
• Filterable solids in the outlet
Any other use than the one described here compromises the safety of persons and the entire
measuring system and is not permitted.
The manufacturer is not liable for damage caused by improper or non-designated use.

1.3 Workplace safety


As the user, you are responsible for complying with the following safety conditions:
• Regulations for explosion protection
• Installation instructions
• Local standards and regulations
Electromagnetic compatibility
With regard to electromagnetic compatibility, this device has been tested in accordance with
the applicable European standards for industrial applications.
The electromagnetic compatibility indicated only applies to a device that has been connected
in accordance with the instructions in these Operating Instructions.

1.4 Operational safety


‣ Before commissioning the entire measuring point, make sure all the connections are
correct. Ensure that electrical cables and hose connections are not damaged.
‣ Do not operate damaged products, and safeguard them to ensure that they are not
operated inadvertently. Mark the damaged product as defective.
‣ If faults cannot be rectified, the products must be taken out of service and secured against
unintentional commissioning.

4 Endress+Hauser
Turbimax CUS51D Safety instructions

1.5 Product safety


The product is designed to meet state-of-the-art safety requirements, has been tested and
left the factory in a condition in which it is safe to operate. Relevant regulations and
European standards have been observed.

Endress+Hauser 5
Incoming acceptance and product identification Turbimax CUS51D

2 Incoming acceptance and product


identification

2.1 Incoming acceptance


‣ Make sure the packaging is undamaged!
‣ Inform the supplier about any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
‣ Make sure the contents are undamaged!
‣ Inform the supplier about damage to the contents. Keep the damaged products until the
matter has been settled.
‣ Check that the order is complete and agrees with your shipping documents.
‣ The packaging material used to store or to transport the product must provide shock
protection and humidity protection. The original packaging offers the best protection.
Also, keep to the approved ambient conditions (see "Technical data").
‣ If you have any questions, please contact your supplier or your local sales center.

2.2 Product identification

2.2.1 Nameplate
The nameplate contains the following information:
• Manufacturer data
• Order code
• Extended order code
• Serial number
• Operating conditions
• Safety icons
Compare the order code on the nameplate with your order.

2.2.2 Identifying the product


The order code and serial number of your device can be found in the following locations:
• On the nameplate
• On the front page of these Operating Instructions
• In the delivery papers
To find out the version of your device, enter the order code indicated on the nameplate
in the search screen at the following address:
www.products.endress.com/order-ident

2.3 Scope of delivery


The scope of delivery comprises:
• 1 sensor Turbimax CUS51D in the ordered version
• 1 Operating Instructions BA00461C/07/EN
If you have any questions, please contact your supplier or your local sales center.

2.4 Certificates and approvals


Declaration of conformity
The product meets the requirements of the harmonized European standards. It thus
complies with the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the 4 symbol.

6 Endress+Hauser
Turbimax CUS51D Installation

3 Installation

3.1 Dimensions

a0013179

Fig. 1: Dimensions

See the "Accessories" section for the dimensions of the cleaning unit

Endress+Hauser 7
Installation Turbimax CUS51D

3.2 Installation instructions

3.2.1 Measuring system


A complete measuring system comprises:
• Turbidity sensor Turbimax CUS51D
• Transmitter Liquiline
• Assembly:
– Assembly Flexdip CYA112 and holder system Flexdip CYH112 or
– Retractable assembly , e.g. Cleanfit CUA451

3
2 1

4
1

a0012965

Fig. 2: Measuring system with immersion assembly (example)

1 Holder system Flexdip CYH112 4 Assembly Flexdip CYA112


2 Transmitter Liquiline 5 Turbidity sensor Turbimax CUS51D
3 Weather protection roof

8 Endress+Hauser
Turbimax CUS51D Installation

1
4

a0013334

Fig. 3: Measuring system with immersion assembly (example)

1 Transmitter Liquiline 4 Assembly Flexdip CYA112


2 Weather protection roof 5 Turbidity sensor Turbimax CUS51D
3 Holder system Flexdip CYH112

1 2

4 3
a0012964

Fig. 4: Measuring system with retractable assembly


(example)
1 Turbidity sensor Turbimax CUS51D
2 Transmitter Liquiline
3 Retractable assembly Cleanfit CUA451
4 Flow direction

Endress+Hauser 9
Installation Turbimax CUS51D

3.3 Installation examples

3.3.1 Immersion operation


Fixed installation with wastewater assembly

3
3 2 1

2 4
1

1
5

a0013383 a0012965

Fig. 5: Installation secured on railing Fig. 6: Installation with upright post


1 Turbidity sensor Turbimax CUS51D 1 Flexdip CYH112 holder
2 Flexdip CYA112 wastewater assembly 2 Multichannel transmitter Liquiline CM44x
3 Flexdip CYH112 holder 3 Weather protection cover
4 Railing 4 Flexdip CYA112 wastewater assembly
5 Turbidity sensor Turbimax CUS51D

This type of installation is particularly suitable for strong or turbulent flow


(>0.5 m/s (1.6 ft/ s)) in the basin or channels.

10 Endress+Hauser
Turbimax CUS51D Installation

Installation with chain retainer

3 3
2 1

a0013384 a0013386

Fig. 7: Chain retainer on railing Fig. 8: Chain retainer on upright post


1 Flexdip CYA112 wastewater assembly with Turbimax 1 Flexdip CYH112 holder
CUS51D turbidity sensor 2 Multichannel transmitter Liquiline CM44x
2 Chain 3 Weather protection cover
3 Flexdip CYH112 holder 4 Flexdip CYA112 wastewater assembly
4 Railing 5 Turbidity sensor Turbimax CUS51D

The chain retainer is particularly suitable for applications that require a sufficient distance
between the mounting location and the edge of the aeration basin. As the assembly is freely
suspended, any vibration of the upright post is practically ruled out.
The swinging movement of the chain retainer enhances the self-cleaning effect of the
optical window.

Endress+Hauser 11
Installation Turbimax CUS51D

Cleaning unit

a0013259

Fig. 9: Turbimax CUS51D sensor with cleaning unit

The cleaning unit is particularly suitable for clear water and media containing fats/oils that
tend to cause heavy buildup.
Mount the cleaning unit as follows:
1. Fit the cleaning unit onto the sensor as far as it will go.
2. Locate the two LEDs (they are installed at an angle and have a bright enclosure).
3. Position the cleaning unit in such a way that the nozzle is located at the side of the two
LEDs (see  å 10).
4. Fix the cleaning unit in place with the securing screw (max. torque: 0.5 Nm
(0.37 lbf ft)).
5. Insert the compressed air hose of the compressor into the hose connection.

1
1 1
2 2

a0013410 a0013411

Fig. 10: Aligning the cleaning unit Fig. 11: Fixing the cleaning unit
1 LEDs 1 Hose connection
2 Nozzle 2 Securing screw

12 Endress+Hauser
Turbimax CUS51D Installation

3.3.2 Pipe installation

Arrow 1 shows the flow direction.


The installation angle  must not exceed 90°.
The recommended installation angle is 75°.
The optical windows of the sensor have to be aligned
parallel to the flow direction ( = 90°) or face the
flow direction ( < 90°).
For manual insertion/retraction of the assembly the
medium pressure may not exceed 2 bar (29 psi).

a0013009

Fig. 12: Installation with retractable assembly

The installation angle is 90°.


Turbidity measurement < 200 FNU will result in
erroneous measuring results due to backscattering of
pipe wall.

a0015169

Fig. 13: Installation with flow assembly CYA251

Endress+Hauser 13
Installation Turbimax CUS51D

The following figure illustrates various installation positions in pipes and indicates whether
they are permitted or not.

3
5
1 4

a0013011

Fig. 14: Orientation and installation positions (with retractable assembly CUA451)

• The pipeline diameter must be at least 100 mm (4") if reflective materials (e.g. stainless
steel) are used. An on-site calibration is recommended.
• Install the sensor in places with uniform flow conditions.
• The best installation location is in the ascending pipe (it. 1). Installation is also possible in
the horizontal pipe (it. 5).
• Do not install the sensor in places where air may collect or foam bubbles form (it. 3) or
where suspended particles may settle (it. 2).
• Avoid installation in the down pipe (it. 4).
• Turbidity measurement < 200 FNU will result in erroneous measuring results due to
backscattering of pipe wall. Therefore a multipoint calibration is recommended.
• Avoid installations behind pressure reduction steps which can outgas.

3.4 Post-installation check


‣ Sensor and cable undamaged?
‣ Cap undamaged?
‣ Compliance with permissible sensor installation position?
‣ Is the sensor installed in an assembly and is not suspended from the cable?
‣ Avoid moisture by rain by putting the protective cap on the assembly?

14 Endress+Hauser
Turbimax CUS51D Wiring

4 Wiring
! WARNING
Device is energized
Improper connection can cause injury or death.
‣ The electrical connection must only be carried out by a certified electrician.
‣ Technical personnel must have read and understood the instructions in this manual and
must adhere to them.
‣ Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.

4.1 Connecting to the transmitter


The sensor will be connected to the transmitter as follows:
• With the M12 plug (version: fixed cable, M12 plug) or
• With the fixed cable connected to the terminal strips (version: fixed cable, end sleeves):

a0012460

Fig. 15: Sensor connection

The maximum cable length is 100 m (328 ft).

4.2 Post-connection check

Instrument status and specifications Remarks

Are the sensor, assembly, junction box or cable damaged? Visual inspection

Electrical connection Remarks

Does the supply voltage of the transmitter match the specifications on


the nameplate?

Are the installed cables strain-relieved and not twisted ?

Is the cable type route completely isolated ? Power cable/weak current cable

Are the power supply and signal cable correctly connected to the Use the connection diagram of the
transmitter ? transmitter.
Long enough length of cable core stripped and correct in terminal? Check seating (pull slightly)

Are all the screws terminals properly tightened ? Tighten

Are all the cable entries installed, tightened and sealed ? For cable entries lateral: cable loops
downwards for water to be able to drip
Are all the cable entries installed downwards or lateral ?
off.

Endress+Hauser 15
Device description Turbimax CUS51D

5 Device description

5.1 Sensor design


The sensor is designed for the continuous in-situ measurement of turbidity and solids
content.
The sensor is designed as a 40 mm sensor that can be operated directly and completely in
the process without the need for further sampling (in situ).
All the necessary modules are contained in the sensor:
• Power supply
• Light sources
• Detectors detect the measuring signals, digitize them and process them to form a
measured value.
• The sensor microcontroller is responsible for controlling the internal processes and
transmitting the data.
All the data - including the calibration data - are stored in the sensor. The sensor can thus
be precalibrated and used at a measuring point, calibrated externally, or used for several
measuring points with different calibrations.

5.2 Measuring principle


For turbidity measurement a light beam is sent through the medium and is diverted from its
original direction by optically denser particles, e.g. solid matter particles. This process is also
called scattering.

1 Light source
2 Light beam
3 3 particle
4 Scattered light
2

4
1

4
a0013159

Fig. 16: Deflection of the light

The impinging light will be scattered in different angles. Two angles are of interest in this
matter:
• The scattered light in the 90° direction is less influenced by the size of the particles.
• The scattered light in the 135° direction gives enough information also at a high number
of particles.

16 Endress+Hauser
Turbimax CUS51D Device description

135° 90°

1 2 3
a0012986

Fig. 17: Principle operating mode of the turbidity sensor


1 Light source
2 135° light receiver
3 90° light receiver

If only a small number of particles is in the medium, most of the light will be scattered to the
90° channel and less light will be scattered to the 135° channel. When the number of
particles increases the relationship will change (more light scattered to the 135° channel,
less light scattered to the 90°channel).

Ir
1

135°

90°

FNU
a0013158

Fig. 18: Signal distribution depending on the number of particles


Ir Relative intensity

Endress+Hauser 17
Device description Turbimax CUS51D

The turbidity sensor CUS51D is equipped with two independent sensor units that are
arranged in parallel. The application-specific analysis of both signals results in stable
measured values.

4 6

3 5

1 2

a0012966

Fig. 19: Arrangement of the light souces and the light receivers
1,2 Light sources 1 and 2
3,5 135° light receivers
4,6 90° light receivers

This allows the optimal turbidity and suspended solids measurement:


• For low turbidity values preferably the 90° channel is used.
• For average and high turbidity values and for suspended solids measurement the 135°
channel is used.
• The dual sensor technology allows operation with a large range of soiling compensation,
e.g. suspended solids measurement in the activated sludge basin (basis: four-beam pulsed
light).
Based on the chosen application the appropriate model is used automatically inside the
sensor.

The available sensor types differ in their measuring ranges and therefore in the
selection of the available applications.

18 Endress+Hauser
Turbimax CUS51D Device description

5.3 Measuring methods


Four-beam pulsed light method
The method is based on two light sources and four light receivers. Long-life LEDs are used
as monochromatic light sources. To eliminate interference from extraneous light sources,
these LEDs are pulsed.
Two measuring signals are detected at the four light receivers. The eight measuring signals
are processed in the sensor and are converted into turbidity units and solids concentrations.
The four-beam pulsed light method compensates the sensor soiling as well as the wearing
of the optical components.
The number of the used signals depends on the application.

S1 E1-135 E1-90

S2 E2-135 E2-90
a0012987

Fig. 20: Four-beam pulsed light method

S1 S2Light sources
E9090° channel light-receiver
E135135° channel light-receiver

90° scattered light method


The measurement uses a wavelength of 860 nm like described in ISO 7027 / EN 27027.
The transmitted light beam is scattered by the solid matter particles in the medium. The
scattered beams are detected by scattered light receivers which are arranged at an angle of
90 ° to the light sources. The turbidity of the medium is determined by the amount of the
scattered light.

S = Light source
E E = Light receiver
P = Particle

P a

S
a0013168

Fig. 21: Principle of the 90° scattered light method

Endress+Hauser 19
Device description Turbimax CUS51D

135° backscattered light method


The transmitted light beam is scattered by the solid matter particles in the medium. The
backscattered beams are detected by scattered light receivers, which are arranged next to
the light sources. The turbidity of the medium is determined by the amount of backscattered
light.
This method is used to measure high turbidity values.

I0 = Intensity of transmitted light


Is = I0 · C · A · f(a) IS = Intensity of backscattered light
A = Geometric factor
C = Concentration
P = Particle
IS f() = Angle dependence
E1 a P

I0

S1

a0013304

Fig. 22: Principle of backscattered light method

20 Endress+Hauser
Turbimax CUS51D Device description

5.4 Application
The sensor enables measurement in a variety of ways that are adapted to suit the specific
measurement problems. The method is set by the choice of application and the reference
model.

Application type "Clear water"

Application Method Measuring range

Formazine 135° turbidity, single channel 0 to 4000 FNU


Display range up to 9999 FNU

Kaolin 135° turbidity, single channel 0 to 5 g/l

TiO2 (titanium oxide) 4-beam pulsed light, 135° 0 to 150 g/l

SiO2 (silicon dioxide) 4-beam pulsed light, 135° 5 to 100 g/l

a0020168-en

Application type "Solids"

Application Method Measuring range

Thin sludge 135° turbidity, single channel 0 to 5 g/l

Sludge activation 4-beam pulsed light, 90° 2 to 15 g/l

Return sludge 4-beam pulsed light, 135° 3 to 50 g/l

Digested sludge / ooze 135° turbidity, single channel 5 to 100 g/l / 300 g/l

a0020169-en

Endress+Hauser 21
Device description Turbimax CUS51D

Application areas

Model name Application Unit Compensation


*

Formazine Process water, sewage treatment plant outlet FTU / FNU

Kaolin Filterable solids, process water, sewage treatment mg/l ; g/l;


plant outlet, activated sludge in low concentration ppm; %

SiO2 SiO2, mineralic solids (sands) g/l; ppm; % X

TiO2 TiO2, (white medium) g/l; ppm; % X

Thin sludge From activated sludge down to clear water g/l; ppm; %

Activated sludge Activated sludge basin and comparable medium g/l; ppm; % X

Excess sludge Excess sludge, primary sludge, thickened sludge g/l; ppm; % X

Digested sludge Digested sludge, black - homogeneous g/l; ppm; %

* compensation of contamination with four-beam pulsed light

Detailed information about application areas are described in the following section
"Calibration".

22 Endress+Hauser
Turbimax CUS51D Device description

5.5 Calibration
The sensor is precalibrated on leaving the factory. As such, it can be used in a wide range of
applications (e.g. clear water measurement) without the need for additional calibration. The
factory calibrations are based on "three-point calibration". The kaolin and formazine
applications are already fully calibrated and can be used without further calibration.
All other applications are pre-calibrated with reference samples and need to be calibrated to
the appropriate application.
In addition to the factory calibration data, which cannot be modified, the sensor has five
other data records to be used for storing process calibrations.

5.5.1 Selecting the application


On the CM44x, select the appropriate application for your field of application during
initial commissioning. This application can be adopted during calibration or can be
reselected by you.

Wastewater application

Waste activated sludge (WAS)


(formazine, FTU)

Sludge activation

Digested sludge
Diluted sludge

(g/l, % of TS)

(g/l, % of TS)
(mg/l, g/l)

(mg/l, g/l)

(mg/l, g/l)
Turbidity

Kaolin
Wastewater treatment plant inlet (X) (X) X

Primary sludge extraction, primary clarification X (X)

Aeration basin, range 0 - 5 g/l, e.g. SBR X

Aeration basin, range 2 - 15 g/l X (X)

Recirculation line X

Waste activated sludge extraction X X

Sludge thickener (primary sludge) X (X)

Digester inlet X (X)

Digester outlet (sludge) (X) X

Wastewater treatment plant outlet X X (X)

Sand filter control X X

Depending on the arrangement of the facility, individual measuring points can be


declared as Ex zones.

Endress+Hauser 23
Device description Turbimax CUS51D

Process water application

Waste activated sludge (WAS)


(formazine, FTU)

Digested sludge
(g/l, % of TS)

(g/l, % of TS)
Dilute sludge
(mg/l, g/l)

(mg/l, g/l)
(ppm, g/l)

(ppm, g/l)
Turbidity

Kaolin

TiO2
SiO2
Inlet X X

Process control X X

Filter rinsing X X (X)

Sedimentation tank X X X

Process water X X X X

Process sludges X X X X

5.5.2 Choosing the type of calibration


1 to 5 points can be calibrated for all applications.
The following recommendation describes the common types of calibration.

Model Single-point calibration Two-point or multipoint


(in the medium) (outside the medium)

Formazine X

Kaolin X

SiO2 X

TiO2 X

Dilute sludge X

Activated sludge X

Waste activated sludge X


(WAS)

Digested sludge X

With the "dilute sludge" model, measurements can be carried out in the aeration basin
up to approx. 5 g/l. This model has the advantage of being able to be calibrated at a
single point in the process during operation.

24 Endress+Hauser
Turbimax CUS51D Device description

Effects of the various types of calibration


Factory calibration Factory calibration
New calibration curve New calibration curve
Calibration point Calibration point

Signal

Signal
0 0

0 2 4 6 8 0 2 4 6 8
Suspended solids (g/l) Suspended solids (g/l)
a0020210-en a0020211-en

Fig. 23: Single-point calibration Fig. 24: Two-point calibration

Single-point calibration causes a change in Two-point calibration causes a change in


the slope. the slope and the zero point. This type of
calibration is used if the measured value
changes across a wide range.
Factory calibration Factory calibration
New calibration curve New calibration curve
Calibration point Calibration point

Δ5

Δ4
Signal

Signal
Δ2 Δ3

Δ1

0 0

0 2 4 6 8 0 2 4 6 8
Suspended solids (g/l) Suspended solids (g/l)
a0020213-en a0020214-en

Fig. 25: Three-point calibration Fig. 26: Five-point calibration

With three-point calibration, a new With four- or five-point calibration, the


calibration curve is drawn through all three calibration curve between the calibration
calibration points. This leads to higher points is determined. This leads to better
accuracy in the calibrated range. alignment with the selected calibration
points; however, it does not improve
accuracy.

Single-point and two-point calibration are based on the data record stored internally in the
device.
Calibration at three or more points always causes the measuring curve to be recalculated.

With multipoint calibration, the calibration points must always cover the complete
measuring range of the application.
Furthermore, no points outside the specified measuring range of the application may
be selected.
In this way, a calibration with zero water (0 g/l) usually results in unusable calibrations
for the following applications:
•Activated sludge
•Waste activated sludge (WAS)
•Digested sludge
•SiO2

Endress+Hauser 25
Device description Turbimax CUS51D

5.5.3 Single-point calibration


With single-point calibration, the sensor can remain immersed in the process medium.
Proceed as follows:
1. Remove a sample of the medium in close proximity to the sensor for the laboratory
measurement.
2. Send the sample to the laboratory to measure the turbidity or solids content.
3. Select a data record in the CM44x transmitter.
4. Start the calibration as simultaneously as possible to the sampling and enter the
laboratory value of the sample as the set point.
If there is no laboratory value available during calibration, enter an approximate value
as the set point. As soon as the laboratory value is available, you can edit the set point
on the transmitter retrospectively.

5.5.4 Multipoint calibration


! CAUTION
Risk of injury due to acid or medium
‣ Switch off the cleaning unit before removing the sensor from the medium.
Sample preparation:

1. Remove a sample from the process (e.g.


10 liters - bucket).

a0020482

2. Wait until the sludge parts have settled.

a0020483

3. Remove the residual water (if possible).


You now have a sample with an increased
concentration. Stir the sample to increase
the homogeneity.

a0020484

4. Remove part of the sample for analysis in


the laboratory.

a0020485

5. Add a defined part of the sample (e.g.


2 liters) to the calibration vessel (black
bucket)
Stir the sample again to maintain the
homogeneity.
a0020486

26 Endress+Hauser
Turbimax CUS51D Device description

Preparing the CUS51D sensor for calibration:


1. Clean the optical components (windows) of the sensor with water and a brush.
2. Put the sensor into the calibration vessel. When doing this, observe the following
instructions.
• Do not immerse the sensor vertically into the sample. Immerse the sensor at an angle. By
doing this, you will prevent air bubbles building up around the windows.

a0020487

Fig. 27: Immersing the sensor

• The LEDs on the sensor should be directed towards the middle of the calibration vessel.
• The minimum clearance between the sensor and the wall of the vessel is 10 mm
(0.4 inches).
• The distance to the floor of the vessel should be as large as possible, but at the same time,
the sensor must be immersed at least 10 mm (0.4 inches).

a0020448

Fig. 28: Positioning the sensor


1 Beam direction of the LEDs
2 LEDs

• Fix the sensor in this position (ideally with a laboratory stand).


During calibration, make sure the medium is as homogeneous as possible. A magnetic
stirrer would be the best solution.

Endress+Hauser 27
Device description Turbimax CUS51D

During calibration, note the following points:


• The calibration points should cover the complete measuring range.
• Be extremely careful when determining the laboratory measured value (the quality of the
laboratory measurement has a direct impact on the accuracy of the sensor).
• Use a graduated cylinder to measure the volumes for the sample and the dilution water
with the highest possible accuracy.
• Air bubbles on the optical components distort the calibration result significantly and must
therefore be removed before each calibration.
• Ensure that the medium is always well mixed (homogeneity).
• Avoid changes in temperature during calibration, and make sure the dilution water and
the medium are the same temperature, as far as possible.
• The position of the sensor should not be changed during calibration.
• The calibration set points can also be edited retrospectively in the CM44x (e.g. if the
reference value of the laboratory measurement is not known at the time of calibration).
Carrying out a calibration (example of a three-point calibration):
1. Select a data record in the CM44x transmitter.
2. Wait for at least one minute (for stabilization).
3. Start the calibration for measuring point 1 (example: 2 l sample with a concentration of
6 g/l).
Enter the value of the sample determined in the laboratory as the set point (example:
6 g/l) or adopt the suggested value and adapt it later to the laboratory value.

4. Dilute the sample to a ratio of 1:2 (add 2 l water to produce 3 g/l in the example)

H 2O

5. Avoid the build-up of air bubbles under the sensor.


6. Calibrate measuring point 2.
Enter half of the laboratory value as the set point or edit the value later.
7. Dilute the sample to a ratio of 1:3 (add 2 l water to produce 2 g/l in the example)

H 2O

8. Avoid the build-up of air bubbles under the sensor.


9. Calibrate measuring point 3.
Enter a third of the laboratory value as the set point or edit the value later.
You can also carry out the calibration in increasing concentrations (less advisable).

28 Endress+Hauser
Turbimax CUS51D Device description

5.6 Stability criterion


During the calibration process, the measured values returned by the sensor are checked to
ensure they remain constant. The stability criterion defines the maximum amount the
measured values may deviate during calibration and still be accepted.
The data comprise:
• The maximum permissible deviation in temperature measurement
• The maximum permissible deviation in the measured value in %
• The minimum timeframe in which these values must be maintained
As soon as the stability criteria for the signal values and the temperature are reached, the
calibration is continued. If these criteria are not fulfilled within the maximum time frame of
5 minutes, no calibration takes place, and a warning is displayed.
The stability criteria monitor the quality of the individual calibration points in the course of
the calibration. The aim is to achieve the best possible calibration quality in a short
timeframe while taking external conditions into account.

For calibrations in the field under tough weather and environmental conditions, the
measured value windows selected can be large and the timeframe short.

5.7 Cyclic cleaning


Compressed air is most suitable for cyclic cleaning. The optional cleaning unit is either
ready-supplied or can be retrofitted, and is fitted on the sensor head. It operates at a rate of
50 l/min (13.2 US gal/min)
The following settings are recommended for the cleaning unit:

Type of fouling Cleaning interval Cleaning duration

Severe fouling with rapid buildup 5 min 10 s

Low degree of fouling 10 min 10 s

Endress+Hauser 29
Diagnostics and troubleshooting Turbimax CUS51D

6 Diagnostics and troubleshooting


Troubleshooting must take account of the whole measuring system:
• Transmitter
• Electrical leads and connectors
• Assembly
• Sensor
The possible causes of failure listed in the following table primarily refer to the sensor.

Problem Check Remedy

No display, no sensor Mains voltage at transmitter? Connect mains voltage.


reaction Sensor connected correctly? Set up correct connection.
Film formation on optical windows? Clean the sensor.

Reading too high or Film formation on optical windows? Clean the sensor.
too low Sensor calibrated? Calibrate.

Reading greatly Check mounting location. Select other mounting location.


fluctuating

Please pay special attention to the instructions on handling errors in the Operating
Instructions of the transmitter. Check the transmitter, if necessary.

30 Endress+Hauser
Turbimax CUS51D Maintenance

7 Maintenance
You have to perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the system, the installation conditions and the
medium in which measurement takes place.
! CAUTION
Risk of injury caused by acid or medium
‣ Switch off the cleaning unit before removing the sensor from the medium.

7.1 Cleaning the sensor


Sensor fouling can affect the measurement results and even cause a malfunction.
The sensor must be cleaned at regular intervals to ensure reliable measurement results. The
frequency and intensity of the cleaning process depends on the medium.
Clean the sensor:
• As specified in the maintenance schedule
• Before every calibration
• Before returning the sensor for repair

Type of fouling Cleaning measure

Lime deposits Immerse the sensor in 1-5 % hydrochloric acid (for a


few minutes).

Dirt particles on the optical windows Use a cloth to clean the optical windows.

You must rinse the sensor thoroughly with water after cleaning.

Endress+Hauser 31
Repair Turbimax CUS51D

8 Repair

8.1 Return
The device must be returned if repairs or a factory calibration are required, or if the wrong
device has been ordered or delivered. According to legal regulations, Endress+Hauser, as an
ISO-certified company, is required to follow certain procedures when handling returned
products that are in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions on the internet site:
www.services.endress.com/return-material

8.2 Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Please observe local regulations.

32 Endress+Hauser
Turbimax CUS51D Accessories

9 Accessories

9.1 Assemblies
Wastewater assembly Flexdip CYA112
• Modular assembly system for sensors in open basins, channels and tanks
• Versions in stainless steel or PVC
• Ordering per product structure (--> Online configurator: www.products.endress.com/
cya112)
• Technical Information TI00432C/07/EN
Retractable assembly Cleanfit CUA451
• Retractable assembly with ball valve; for turbidity sensors; material: stainless steel
• Ordering per product structure (--> Online configurator, www.products.endress.com/
cua451)
• Technical Information TI00369C/07/EN
Flow assembly Flowfit CYA251
• Connection: see product structure
• Material: PVC-U
• Order as per product structure

9.2 Holder
Holder system Flexdip CYH112 for water and wastewater assembly CYA112
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• The holder system CYH112 works for nearly any type of fixing - fixing on the floor, wall or
directly on a rail.
• Material: stainless steel
• Ordering acc. to product structure (--> Online configurator: www.products.endress.com/
cyh112)
• Technical Information TI00430C/07/EN

Endress+Hauser 33
Accessories Turbimax CUS51D

9.3 Compressed air cleaning


Cleaning system with pressurized air
• Connection: 6 or 8 mm (metric) or 6.35 mm (¼")
• Materials: POM/V4A
• 6/8 mm order number: 71110782
• 6.35 mm (¼") order number: 71110783

A B 72 (2.83)

(0.70)
17.8
(0.70)
17.8
65.8 (2.59)

77.3 (3.04)

(0.45)
11.5

48 (1.89)
Ø 50 (1.97)

Ø 40.1
(1.58)
mm (inch)

a0013261

Fig. 29: Cleaning system


A Version 6 mm
B Version 6.35 mm (¼")

Compressor
• For cleaning system
• 230 V AC order number: 71072583
• 115 V AC order number: 71194623

9.4 Transmitter
Liquiline CM442/CM444/CM448
• Multiple-channel transmitter for the connection of digital sensors with Memosens
technology
• Power supply: 100 to 230 V AC, 24 V AC/DC
• Universally upgradeable
• SD card slot
• Alarm relay
• IP 66, IP67, NEMA 4X
• Ordering per product structure (--> Online configurator on product page)

34 Endress+Hauser
Turbimax CUS51D Technical Data

10 Technical Data

10.1 Input
Measured variable Turbidity
Solids content
Temperature

Measuring range CUS51D-**C1 Application


Turbidity 0 to 4000 FNU Formazine
display range up to 9999 FNU
Solids content 0 to 4 g/l Kaolin, filterable solids
Temperature -20 to +80 °C (-4 to +176 °F)

CUS51D-**D1
Turbidity 0 to 4000 FNU Formazine
display range up to 9999 FNU
Solids content 0 to 300 g/l Solids content according to chosen
0 to 30 % application (see list)
Temperature -20 to +80 °C (-4 to +176 °F)

Measuring range of solids content:


For solids the achievable measuring ranges are depending of the actual medium and
can deviate from the recommended operating ranges. Nonhomogeneous media
produce fluctuations in measured values and restrict the measuring range.

Endress+Hauser 35
Technical Data Turbimax CUS51D

10.2 Performance characteristics


Maximum measured error Turbidity < 2% of the measured value or 0.1 FNU (the respectively larger value
is valid)
Solids < 5 % of the measured value or 1 % of full scale ( the respectively
larger value is valid); valid for sensors in the calibrated measuring
range

Wavelength 860 ± 30 nm

Applications The sensor is factory calibrated in the application "formazine" and hereof derived for "kaolin
filterable solids)". Further precalibrated applications are optimized for the corresponding
medium.
The calibration can be performed up to 5 points.

Application water Recommended working CUS51D-


ranges
C1 D1

Factory calibration formazine 0 to 4000 FNU X X

Factory calibration kaolin 0 to 4 g/l X X

Application SiO2 5 to 100 g/l X

Application titanium dioxide 0.2 to 150 g/l X

a0020168-en

Application wastewater Recommended working CUS51D-


ranges
C1 D1

Application thin sludge 0 to 5 g/l X

Application activated sludge 0.5 to 15 g/l X

Application excess sludge 3 to 50 g/l X

Application digested sludge / ooze 5 to 100 g/l / 300 g/l X

36 Endress+Hauser
Turbimax CUS51D Technical Data

a0020169-en

For solids the achievable measuring ranges are depending of the actual medium and
can deviate from the recommended operating ranges.

Factory calibration FNU, FTU and solids concentration according to the application table
Standard: 3 points

Drift Thanks to electronic control the sensor works drift compensated in a wide range.

Limit of detection Application Measuring range Limit of detection

0 to 50 FNU 0.006 FNU


Formazine
0 to 9999 FNU 0.4 FNU

Kaolin 0 to 4000 mg/l 0.85 mg/l

10.3 Environment
Ambient temperature -20 to 60 °C (-4 to 140 °F)
range

Storage temperature -20 to 70 °C (-4 to 158 °F)

Ingress protection IP 68 (test conditions: 1 m (3.3 ft) water column during 60 days, 1 mol/l KCl)

10.4 Process
Process temperature -5 to 50 °C (23 to 120 °F)
max. 80 °C (175 °F) short term(1 h)

Process pressure 0.5 to 10 bar (7 to 145 psi) absolute

Minimum flow No minimum flow required.


Make sure that there is a sufficient turbulence for solids with a tendency to sedimentation.

Endress+Hauser 37
Technical Data Turbimax CUS51D

10.5 Mechanical construction


Dimensions See "Installation conditions"

Weight approx. 0.7 kg (1.5 lbs) without cable

Materials Sensor Stainless steel 1.4404 (AISI 316 L) or


Stainless steel 1.4571 (AISI 316 L)
Optical windows Sapphire
O-rings EPDM

Process connections G1 and NPT ¾''

38 Endress+Hauser
Turbimax CUS51D

Index
A N
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . 37
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 O
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Operational safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
P
C Pipe installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Multipoint. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Process calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Single point. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Checking Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Process temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Product identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Compressed air cleaning. . . . . . . . . . . . . . . . . . . . . . . 29, 34 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Connection
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 R
Requirements for the personnel . . . . . . . . . . . . . . . . . . . . . 4
D Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Declaration of conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 S
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Sensor design. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Single-point calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Stability criterion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
E Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . 4
T
F Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Factory calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Transmitter connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
H
Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
I
Identifying the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 W
Immersion operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Wavelength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Ingress protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Input Workplace safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–8
Installation examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . 36
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Minimum flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Multipoint calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

Endress+Hauser 39
71219749

www.addresses.endress.com

You might also like