Turbidity & TSS Probe BA00461CEN - 1513
Turbidity & TSS Probe BA00461CEN - 1513
Turbidity & TSS Probe BA00461CEN - 1513
Operating Instructions
Turbimax CUS51D
Sensor for turbidity and solids content
About this document
Symbols
Endress+Hauser
Turbimax CUS51D
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . . . 4 10 Technical Data . . . . . . . . . . . . . . . . . . . . 35
1.1 Requirements for the personnel . . . . . . . . . . . . . . . 4 10.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
1.2 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 10.2 Performance characteristics . . . . . . . . . . . . . . . . . 36
1.3 Workplace safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 10.3 Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
1.4 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 10.4 Process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
1.5 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 10.5 Mechanical construction . . . . . . . . . . . . . . . . . . . . 38
3 Installation . . . . . . . . . . . . . . . . . . . . . . . . 7
3.1 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Installation instructions . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Installation examples . . . . . . . . . . . . . . . . . . . . . . 10
3.4 Post-installation check . . . . . . . . . . . . . . . . . . . . . 14
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
4.1 Connecting to the transmitter . . . . . . . . . . . . . . . 15
4.2 Post-connection check . . . . . . . . . . . . . . . . . . . . . 15
5 Device description . . . . . . . . . . . . . . . . . 16
5.1 Sensor design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
5.2 Measuring principle . . . . . . . . . . . . . . . . . . . . . . . 16
5.3 Measuring methods . . . . . . . . . . . . . . . . . . . . . . . 19
5.4 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.5 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
5.6 Stability criterion . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.7 Cyclic cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . 31
7.1 Cleaning the sensor . . . . . . . . . . . . . . . . . . . . . . . 31
8 Repair. . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.1 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
8.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
9 Accessories . . . . . . . . . . . . . . . . . . . . . . . 33
9.1 Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
9.2 Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
9.3 Compressed air cleaning . . . . . . . . . . . . . . . . . . . 34
9.4 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Endress+Hauser 3
Safety instructions Turbimax CUS51D
1 Safety instructions
4 Endress+Hauser
Turbimax CUS51D Safety instructions
Endress+Hauser 5
Incoming acceptance and product identification Turbimax CUS51D
2.2.1 Nameplate
The nameplate contains the following information:
• Manufacturer data
• Order code
• Extended order code
• Serial number
• Operating conditions
• Safety icons
Compare the order code on the nameplate with your order.
6 Endress+Hauser
Turbimax CUS51D Installation
3 Installation
3.1 Dimensions
a0013179
Fig. 1: Dimensions
See the "Accessories" section for the dimensions of the cleaning unit
Endress+Hauser 7
Installation Turbimax CUS51D
3
2 1
4
1
a0012965
8 Endress+Hauser
Turbimax CUS51D Installation
1
4
a0013334
1 2
4 3
a0012964
Endress+Hauser 9
Installation Turbimax CUS51D
3
3 2 1
2 4
1
1
5
a0013383 a0012965
10 Endress+Hauser
Turbimax CUS51D Installation
3 3
2 1
a0013384 a0013386
The chain retainer is particularly suitable for applications that require a sufficient distance
between the mounting location and the edge of the aeration basin. As the assembly is freely
suspended, any vibration of the upright post is practically ruled out.
The swinging movement of the chain retainer enhances the self-cleaning effect of the
optical window.
Endress+Hauser 11
Installation Turbimax CUS51D
Cleaning unit
a0013259
The cleaning unit is particularly suitable for clear water and media containing fats/oils that
tend to cause heavy buildup.
Mount the cleaning unit as follows:
1. Fit the cleaning unit onto the sensor as far as it will go.
2. Locate the two LEDs (they are installed at an angle and have a bright enclosure).
3. Position the cleaning unit in such a way that the nozzle is located at the side of the two
LEDs (see å 10).
4. Fix the cleaning unit in place with the securing screw (max. torque: 0.5 Nm
(0.37 lbf ft)).
5. Insert the compressed air hose of the compressor into the hose connection.
1
1 1
2 2
a0013410 a0013411
Fig. 10: Aligning the cleaning unit Fig. 11: Fixing the cleaning unit
1 LEDs 1 Hose connection
2 Nozzle 2 Securing screw
12 Endress+Hauser
Turbimax CUS51D Installation
a0013009
a0015169
Endress+Hauser 13
Installation Turbimax CUS51D
The following figure illustrates various installation positions in pipes and indicates whether
they are permitted or not.
3
5
1 4
a0013011
Fig. 14: Orientation and installation positions (with retractable assembly CUA451)
• The pipeline diameter must be at least 100 mm (4") if reflective materials (e.g. stainless
steel) are used. An on-site calibration is recommended.
• Install the sensor in places with uniform flow conditions.
• The best installation location is in the ascending pipe (it. 1). Installation is also possible in
the horizontal pipe (it. 5).
• Do not install the sensor in places where air may collect or foam bubbles form (it. 3) or
where suspended particles may settle (it. 2).
• Avoid installation in the down pipe (it. 4).
• Turbidity measurement < 200 FNU will result in erroneous measuring results due to
backscattering of pipe wall. Therefore a multipoint calibration is recommended.
• Avoid installations behind pressure reduction steps which can outgas.
14 Endress+Hauser
Turbimax CUS51D Wiring
4 Wiring
! WARNING
Device is energized
Improper connection can cause injury or death.
‣ The electrical connection must only be carried out by a certified electrician.
‣ Technical personnel must have read and understood the instructions in this manual and
must adhere to them.
‣ Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
a0012460
Are the sensor, assembly, junction box or cable damaged? Visual inspection
Is the cable type route completely isolated ? Power cable/weak current cable
Are the power supply and signal cable correctly connected to the Use the connection diagram of the
transmitter ? transmitter.
Long enough length of cable core stripped and correct in terminal? Check seating (pull slightly)
Are all the cable entries installed, tightened and sealed ? For cable entries lateral: cable loops
downwards for water to be able to drip
Are all the cable entries installed downwards or lateral ?
off.
Endress+Hauser 15
Device description Turbimax CUS51D
5 Device description
1 Light source
2 Light beam
3 3 particle
4 Scattered light
2
4
1
4
a0013159
The impinging light will be scattered in different angles. Two angles are of interest in this
matter:
• The scattered light in the 90° direction is less influenced by the size of the particles.
• The scattered light in the 135° direction gives enough information also at a high number
of particles.
16 Endress+Hauser
Turbimax CUS51D Device description
135° 90°
1 2 3
a0012986
If only a small number of particles is in the medium, most of the light will be scattered to the
90° channel and less light will be scattered to the 135° channel. When the number of
particles increases the relationship will change (more light scattered to the 135° channel,
less light scattered to the 90°channel).
Ir
1
135°
90°
FNU
a0013158
Endress+Hauser 17
Device description Turbimax CUS51D
The turbidity sensor CUS51D is equipped with two independent sensor units that are
arranged in parallel. The application-specific analysis of both signals results in stable
measured values.
4 6
3 5
1 2
a0012966
Fig. 19: Arrangement of the light souces and the light receivers
1,2 Light sources 1 and 2
3,5 135° light receivers
4,6 90° light receivers
The available sensor types differ in their measuring ranges and therefore in the
selection of the available applications.
18 Endress+Hauser
Turbimax CUS51D Device description
S1 E1-135 E1-90
S2 E2-135 E2-90
a0012987
S1 S2Light sources
E9090° channel light-receiver
E135135° channel light-receiver
S = Light source
E E = Light receiver
P = Particle
P a
S
a0013168
Endress+Hauser 19
Device description Turbimax CUS51D
I0
S1
a0013304
20 Endress+Hauser
Turbimax CUS51D Device description
5.4 Application
The sensor enables measurement in a variety of ways that are adapted to suit the specific
measurement problems. The method is set by the choice of application and the reference
model.
a0020168-en
Digested sludge / ooze 135° turbidity, single channel 5 to 100 g/l / 300 g/l
a0020169-en
Endress+Hauser 21
Device description Turbimax CUS51D
Application areas
Thin sludge From activated sludge down to clear water g/l; ppm; %
Activated sludge Activated sludge basin and comparable medium g/l; ppm; % X
Excess sludge Excess sludge, primary sludge, thickened sludge g/l; ppm; % X
Detailed information about application areas are described in the following section
"Calibration".
22 Endress+Hauser
Turbimax CUS51D Device description
5.5 Calibration
The sensor is precalibrated on leaving the factory. As such, it can be used in a wide range of
applications (e.g. clear water measurement) without the need for additional calibration. The
factory calibrations are based on "three-point calibration". The kaolin and formazine
applications are already fully calibrated and can be used without further calibration.
All other applications are pre-calibrated with reference samples and need to be calibrated to
the appropriate application.
In addition to the factory calibration data, which cannot be modified, the sensor has five
other data records to be used for storing process calibrations.
Wastewater application
Sludge activation
Digested sludge
Diluted sludge
(g/l, % of TS)
(g/l, % of TS)
(mg/l, g/l)
(mg/l, g/l)
(mg/l, g/l)
Turbidity
Kaolin
Wastewater treatment plant inlet (X) (X) X
Recirculation line X
Endress+Hauser 23
Device description Turbimax CUS51D
Digested sludge
(g/l, % of TS)
(g/l, % of TS)
Dilute sludge
(mg/l, g/l)
(mg/l, g/l)
(ppm, g/l)
(ppm, g/l)
Turbidity
Kaolin
TiO2
SiO2
Inlet X X
Process control X X
Sedimentation tank X X X
Process water X X X X
Process sludges X X X X
Formazine X
Kaolin X
SiO2 X
TiO2 X
Dilute sludge X
Activated sludge X
Digested sludge X
With the "dilute sludge" model, measurements can be carried out in the aeration basin
up to approx. 5 g/l. This model has the advantage of being able to be calibrated at a
single point in the process during operation.
24 Endress+Hauser
Turbimax CUS51D Device description
Signal
Signal
0 0
0 2 4 6 8 0 2 4 6 8
Suspended solids (g/l) Suspended solids (g/l)
a0020210-en a0020211-en
Δ5
Δ4
Signal
Signal
Δ2 Δ3
Δ1
0 0
0 2 4 6 8 0 2 4 6 8
Suspended solids (g/l) Suspended solids (g/l)
a0020213-en a0020214-en
Single-point and two-point calibration are based on the data record stored internally in the
device.
Calibration at three or more points always causes the measuring curve to be recalculated.
With multipoint calibration, the calibration points must always cover the complete
measuring range of the application.
Furthermore, no points outside the specified measuring range of the application may
be selected.
In this way, a calibration with zero water (0 g/l) usually results in unusable calibrations
for the following applications:
•Activated sludge
•Waste activated sludge (WAS)
•Digested sludge
•SiO2
Endress+Hauser 25
Device description Turbimax CUS51D
a0020482
a0020483
a0020484
a0020485
26 Endress+Hauser
Turbimax CUS51D Device description
a0020487
• The LEDs on the sensor should be directed towards the middle of the calibration vessel.
• The minimum clearance between the sensor and the wall of the vessel is 10 mm
(0.4 inches).
• The distance to the floor of the vessel should be as large as possible, but at the same time,
the sensor must be immersed at least 10 mm (0.4 inches).
a0020448
Endress+Hauser 27
Device description Turbimax CUS51D
4. Dilute the sample to a ratio of 1:2 (add 2 l water to produce 3 g/l in the example)
H 2O
H 2O
28 Endress+Hauser
Turbimax CUS51D Device description
For calibrations in the field under tough weather and environmental conditions, the
measured value windows selected can be large and the timeframe short.
Endress+Hauser 29
Diagnostics and troubleshooting Turbimax CUS51D
Reading too high or Film formation on optical windows? Clean the sensor.
too low Sensor calibrated? Calibrate.
Please pay special attention to the instructions on handling errors in the Operating
Instructions of the transmitter. Check the transmitter, if necessary.
30 Endress+Hauser
Turbimax CUS51D Maintenance
7 Maintenance
You have to perform maintenance tasks at regular intervals.
We recommend setting the maintenance times in advance in an operations journal or log.
The maintenance cycle primarily depends on the system, the installation conditions and the
medium in which measurement takes place.
! CAUTION
Risk of injury caused by acid or medium
‣ Switch off the cleaning unit before removing the sensor from the medium.
Dirt particles on the optical windows Use a cloth to clean the optical windows.
You must rinse the sensor thoroughly with water after cleaning.
Endress+Hauser 31
Repair Turbimax CUS51D
8 Repair
8.1 Return
The device must be returned if repairs or a factory calibration are required, or if the wrong
device has been ordered or delivered. According to legal regulations, Endress+Hauser, as an
ISO-certified company, is required to follow certain procedures when handling returned
products that are in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures and
conditions on the internet site:
www.services.endress.com/return-material
8.2 Disposal
The device contains electronic components and must therefore be disposed of in accordance
with regulations on the disposal of electronic waste.
Please observe local regulations.
32 Endress+Hauser
Turbimax CUS51D Accessories
9 Accessories
9.1 Assemblies
Wastewater assembly Flexdip CYA112
• Modular assembly system for sensors in open basins, channels and tanks
• Versions in stainless steel or PVC
• Ordering per product structure (--> Online configurator: www.products.endress.com/
cya112)
• Technical Information TI00432C/07/EN
Retractable assembly Cleanfit CUA451
• Retractable assembly with ball valve; for turbidity sensors; material: stainless steel
• Ordering per product structure (--> Online configurator, www.products.endress.com/
cua451)
• Technical Information TI00369C/07/EN
Flow assembly Flowfit CYA251
• Connection: see product structure
• Material: PVC-U
• Order as per product structure
9.2 Holder
Holder system Flexdip CYH112 for water and wastewater assembly CYA112
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• The holder system CYH112 works for nearly any type of fixing - fixing on the floor, wall or
directly on a rail.
• Material: stainless steel
• Ordering acc. to product structure (--> Online configurator: www.products.endress.com/
cyh112)
• Technical Information TI00430C/07/EN
Endress+Hauser 33
Accessories Turbimax CUS51D
A B 72 (2.83)
(0.70)
17.8
(0.70)
17.8
65.8 (2.59)
77.3 (3.04)
(0.45)
11.5
48 (1.89)
Ø 50 (1.97)
Ø 40.1
(1.58)
mm (inch)
a0013261
Compressor
• For cleaning system
• 230 V AC order number: 71072583
• 115 V AC order number: 71194623
9.4 Transmitter
Liquiline CM442/CM444/CM448
• Multiple-channel transmitter for the connection of digital sensors with Memosens
technology
• Power supply: 100 to 230 V AC, 24 V AC/DC
• Universally upgradeable
• SD card slot
• Alarm relay
• IP 66, IP67, NEMA 4X
• Ordering per product structure (--> Online configurator on product page)
34 Endress+Hauser
Turbimax CUS51D Technical Data
10 Technical Data
10.1 Input
Measured variable Turbidity
Solids content
Temperature
CUS51D-**D1
Turbidity 0 to 4000 FNU Formazine
display range up to 9999 FNU
Solids content 0 to 300 g/l Solids content according to chosen
0 to 30 % application (see list)
Temperature -20 to +80 °C (-4 to +176 °F)
Endress+Hauser 35
Technical Data Turbimax CUS51D
Wavelength 860 ± 30 nm
Applications The sensor is factory calibrated in the application "formazine" and hereof derived for "kaolin
filterable solids)". Further precalibrated applications are optimized for the corresponding
medium.
The calibration can be performed up to 5 points.
a0020168-en
36 Endress+Hauser
Turbimax CUS51D Technical Data
a0020169-en
For solids the achievable measuring ranges are depending of the actual medium and
can deviate from the recommended operating ranges.
Factory calibration FNU, FTU and solids concentration according to the application table
Standard: 3 points
Drift Thanks to electronic control the sensor works drift compensated in a wide range.
10.3 Environment
Ambient temperature -20 to 60 °C (-4 to 140 °F)
range
Ingress protection IP 68 (test conditions: 1 m (3.3 ft) water column during 60 days, 1 mol/l KCl)
10.4 Process
Process temperature -5 to 50 °C (23 to 120 °F)
max. 80 °C (175 °F) short term(1 h)
Endress+Hauser 37
Technical Data Turbimax CUS51D
38 Endress+Hauser
Turbimax CUS51D
Index
A N
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . 37
Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 O
Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Operational safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
P
C Pipe installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Multipoint. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Process calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Single point. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Checking Process pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Process temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Product identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Compressed air cleaning. . . . . . . . . . . . . . . . . . . . . . . 29, 34 Product safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Connection
Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 R
Requirements for the personnel . . . . . . . . . . . . . . . . . . . . . 4
D Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Declaration of conformity. . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 S
Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Sensor design. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Drift . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Single-point calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Stability criterion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
E Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Electromagnetic compatibility . . . . . . . . . . . . . . . . . . . . . . . 4
T
F Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Factory calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Transmitter connection . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
H
Holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
I
Identifying the product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 W
Immersion operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Wavelength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Ingress protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Input Workplace safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–8
Installation examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Materials. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . 36
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring methods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Minimum flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Multipoint calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Endress+Hauser 39
71219749
www.addresses.endress.com