Ka01023pen 1614

Download as pdf or txt
Download as pdf or txt
You are on page 1of 32

Brief Operating Instructions

Deltapilot S
FMB70

Hydrostatic level measurement

These Instructions are Brief Operating Instructions; they are not a substitute for the
Operating Instructions pertaining to the device.
Detailed information about the device can be found in the Operating Instructions
and the other documentation:
Available for all device versions via:
– Internet: www.endress.com/deviceviewer
– Smart phone/tablet: Endress+Hauser Operations App

KA01023P/00/EN/16.14
71270391
Deltapilot S PROFIBUS PA

Order code 00X00-XXXX0XX0XXX


Ser. No.: X000X000000
TAG No.: XXX000

Serial number

www.endress.com/deviceviewer Endress+Hauser Operations App

A0023555

2 Endress+Hauser
Deltapilot S PROFIBUS PA Table of contents

Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Installation, commissioning and operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.3 Operational safety and process safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5 Safety icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

2 Product identification. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.1 General installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3.2 Measuring arrangement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.3 Assembling and mounting the "separate housing" version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

4 Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1 Connecting the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Connecting the measuring unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.1 On-site display (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
5.2 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
5.3 On-site operation via on-site display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.4 Configuring the device address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.5 Locking/unlocking operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

6 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
6.1 Position adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
6.2 Level measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
6.3 Pressure measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28

Endress+Hauser 3
Safety instructions Deltapilot S PROFIBUS PA

1 Safety instructions
1.1 Designated use
The Deltapilot S is a hydrostatic pressure transmitter for measuring level and pressure.
The manufacturer accepts no liability for damages resulting from incorrect use or use other than
that designated.

1.2 Installation, commissioning and operation


• The device must only be installed, connected, commissioned and maintained by qualified and
authorized specialists (e.g. electrical technicians) in full compliance with the instructions in
this manual, the applicable norms, legal regulations and certificates (depending on the
application).
• The specialist must have read and understood this manual and must follow the instructions it
contains. If you are unclear on anything in these Brief Operating Instructions, you must read
the Operating Instructions. The Operating Instructions provide detailed information on the
device/measuring system.
• The device may only be modified or repaired if such work is expressly permitted in the
Operating Instructions.
• If faults cannot be rectified, the device must be taken out of service and secured against
unintentional commissioning.
• Do not operate damaged devices. Mark them as defective.

1.3 Operational safety and process safety


• Alternative monitoring measures must be taken to ensure operational safety and process safety
during configuration, testing and maintenance work on the device.
• The device is safely built and tested according to state-of-the-art technology and has left the
factory in perfect condition as regards technical safety. The applicable regulations and
European standards have been taken into account.
• Pay particular attention to the technical data on the nameplate.
• Devices for use in hazardous areas are fitted with an additional nameplate. If the device is to
be installed in an explosion hazardous area, then the specifications in the certificate as well as
all national and local regulations must be observed. The device is accompanied by separate
"Ex documentation", which is an integral part of this Operating Instructions. The installation
regulations, connection values and Safety Instructions listed in this Ex document must be
observed. The documentation number of the related Safety Instructions is also indicated on
the additional nameplate.

4 Endress+Hauser
Deltapilot S PROFIBUS PA Product identification

1.4 Return
Follow the instructions on returning the device as outlined in the Operating Instructions.

1.5 Safety icons

Symbol Meaning

#
Warning!
A warning highlights actions or procedures which, if not performed correctly, will lead to personal injury,
a safety hazard or destruction of the instrument.

"
Caution!
Caution highlights actions or procedures which, if not performed correctly, may lead to personal injury or
incorrect functioning of the instrument.

!
Note!
A note highlights actions or procedures which, if not performed correctly, may indirectly affect operation
or may lead to an instrument response which is not planned.

2 Product identification
The following options are available for identification of the measuring device:
• Nameplate specifications
• Order code with breakdown of the device features on the delivery note
• Enter serial numbers from nameplates in W@M Device Viewer
(www.endress.com/deviceviewer): All information about the measuring device is displayed.
For an overview of the technical documentation provided, enter the serial number from the
nameplates in the W@M Device Viewer (www.endress.com/deviceviewer).

3 Installation
3.1 General installation instructions
# Warning!
The seal is not allowed to press on the process isolating diaphragm as this could affect the
measurement result.

Endress+Hauser 5
Installation Deltapilot S PROFIBUS PA

! Note!
• If a heated Deltapilot S is cooled during the cleaning (e.g. by cold water), a vacuum develops
for a short time, whereby water can penetrate the sensor through the pressure compensation
(1). If this is the case, mount the sensor with the pressure compensation (1) pointing
downwards.
1
FIELD TERMINALS

SLANIMRET DLEIF

1
• Keep the pressure compensation and GORE-TEX® filter (1) free from contaminations.
• Do not use sharp or hard objects to handle or clean the process isolating diaphragm.
• Due to the orientation of the Deltapilot S, there may be a shift in the measured value, i.e.
when the container is empty, the measured value does not display zero. You can correct this
zero point shift either via the "zero" key on the electronic insert, or on the outside of the
device or via the on-site display.See Page 14, Section 4.2.1 "Position of operating
elements", Page 15, Section 4.2.2 "Function of the operating elements" and Page 24, Section
5.1 "Position adjustment".
• To ensure optimal readability of the on-site display, it is possible to rotate the housing up to
380°.
• The on-site display can be rotated in 90° stages.
• Endress+Hauser offers a mounting bracket for installing on pipes or walls.

6 Endress+Hauser
Deltapilot S PROFIBUS PA Installation

3.2 Measuring arrangement


3.2.1 Level measurement

P01-PMP75xxx-11-xx-xx-xx-000
Fig. 1: Measuring arrangement for level

• Always install the device below the lowest measuring point.


• Do not install the device at the following positions:
– in the filling curtain
– in the tank outflow
– in the suction area of a pump
– or at a point in the tank that can be affected by pressure pulses from the agitator
• The calibration and functional test can be carried out more easily if you mount the device
downstream of a shutoff device.
• Deltapilot S must be included in the insulation for media that can harden when cold.

3.2.2 Pressure measurement in gases


• Mount Deltapilot S with shutoff device above the tapping point so that any condensate can
flow into the process.

3.2.3 Pressure measurement in steams


• Mount Deltapilot S with siphon above the tapping point.
• Fill the siphon with liquid before commissioning.
The siphon reduces the temperature to almost the ambient temperature.

3.2.4 Pressure measurement in liquids


• Mount Deltapilot S with the shutoff device below or at the same level as the tapping point.

Endress+Hauser 7
Installation Deltapilot S PROFIBUS PA

3.3 Assembling and mounting the "separate housing" version

6
8
5
3
2

1 r ³ 120 mm

P01-FMB70xxx-11-xx-xx-xx-003
Fig. 2: "Separate housing" version
1 In the "separate housing" version, the sensor is supplied with process connection and cable fitted.
2 Cable with connection jack
3 Pressure compensation
5 Plug
6 Locking screw
7 Housing fitted with housing adapter, included
8 Mounting bracket suitable for wall and pipe mounting, included

Assembly and mounting


1. Connect plug (item 5) into the corresponding connection jack of the cable (item 2).
2. Plug the cable into the housing adapter (item 7).
3. Tighten the locking screw (item 6).
4. Mount the housing on a wall or pipe using the mounting bracket (item 8). When
mounting on a pipe, tighten the nuts on the bracket uniformly with a torque of at least
5 Nm.
Mount the cable with a bending radius (r)  120 mm.

8 Endress+Hauser
Deltapilot S PROFIBUS PA Wiring

4 Wiring
4.1 Connecting the device
# Warning!
Risk of electric shock!
If the operating voltage is > 35 VDC: Dangerous contact voltage at terminals.
In a wet environment, do not open the cover if voltage is present.

# Warning!
Limitation of electrical safety due to incorrect connection!
• Risk of electric shock and/or explosion in hazardous areas! In a wet environment, do not open
the cover if voltage is present.
• When using the measuring device in hazardous areas, installation must comply with the
corresponding national standards and regulations and the Safety Instructions or Installation or
Control Drawings.
• Devices with integrated overvoltage protection must be earthed.
• Protective circuits against reverse polarity, HF influences and overvoltage peaks are installed.
• The supply voltage must match the supply voltage on the nameplate.
• Switch off the supply voltage before connecting the device.
• Remove housing cover of the terminal compartment.
• Guide cable through the gland. Preferably use twisted, screened two-wire cable.
• Connect device in accordance with the following diagram.
• Screw down housing cover.
• Switch on supply voltage.

Endress+Hauser 9
Wiring Deltapilot S PROFIBUS PA


PA PA

➁ ➄
PA PA


P01-xMx7xxxx-04-xx-xx-xx-008
Fig. 3: Electrical connection PROFIBUS PA Observe also the following section.
For devices with plug see Operating Instructions.
1 Housing
2 Internal earth terminal
3 External earth terminal
4 Supply voltage, for version in non-hazardous area = 9...32 V DC
5 Devices with integrated overvoltage protection are labelled OVP (overvoltage protection) here.

4.2 Connecting the measuring unit


4.2.1 Supply voltage
• Version for non-hazardous area: 9...32 V DC

4.2.2 Current consumption


Up to HW Version 1.10:
11 mA ±1 mA, switch-on current corresponds to IEC 61158-2, Clause 21.

As of HW Version 02.00:
13 mA ±1 mA, switch-on current corresponds to IEC 61158-2, Clause 21.

As of Hardware Version 1.10, you will find a label in the device on the electronic insert.

4.2.3 Cable specification


• Use a twisted, screened two-wire cable, preferably cable type A.
• Terminals for wire cross-sections: 0.5...2.5 mm2
• Outer cable diameter: 5...9 mm

10 Endress+Hauser
Deltapilot S PROFIBUS PA Operation

! Note!
For further information on the cable specifications, see Operating Instructions BA00034S
"Guidelines for planning and commissioning PROFIBUS DP/PA", PNO Guideline 2.092
"PROFIBUS PA User and Installation Guideline" and IEC 61158-2 (MBP).

4.2.4 Earthing and screening


Deltapilot S must be earthed, for example by means of the external earth terminal.
Different earthing and screening installation methods are available for PROFIBUS PA networks
such as:
• Isolated installation (see also IEC 61158-2)
• Installation with multiple earthing
• Capacitive installation

5 Operation
5.1 On-site display (optional)
A 4-line liquid crystal display (LCD) is used for display and operation. The on-site display shows
measured values, dialog texts, fault messages and notice messages.
The display of the device can be turned in 90° steps.
Depending on the installation position of the device, this makes it easy to operate the device and
read the measured values.

Endress+Hauser 11
Operation Deltapilot S PROFIBUS PA

Measured value display Parameter


Identification
Function name Value number

Header line
Main line
Unit
Information
line
– + E Symbol
Bargraph
Editing modes

Operating keys
Selection
options

Value that
can be edited

Current measured value


P01-xxxxxxxx-07-xx-xx-xx-011

12 Endress+Hauser
Deltapilot S PROFIBUS PA Operation

The following table illustrates the symbols that can appear on the on-site display. Four symbols
can occur at one time.

Symbol Meaning

Alarm symbol
– Symbol flashing: warning, device continues measuring.
– Symbol permanently lit: error, device does not continue measuring.
Note: The alarm symbol may overlie the tendency symbol.

Lock symbol
The operation of the device is locked. Unlock device, see Page 22, Section 4.5.

Communication symbol
Data transfer via communication

Tendency symbol (increasing)


The primary value of the Transducer Block is increasing.

Tendency symbol (decreasing)


The primary value of the Transducer Block is decreasing.

Tendency symbol (constant)


The primary value of the Transducer Block has remained constant over the past few minutes.

Endress+Hauser 13
Operation Deltapilot S PROFIBUS PA

5.2 Operating elements


5.2.1 Position of operating elements
With regard to aluminium housings and stainless steel housing (T14/T15), the operating key is
located either outside the device under the protection cap or inside on the electronic insert. In
hygenic stainless housings (T17), the operating key is always located inside on the electronic
insert. Additionally, three operating keys are located on the optional on-site display.

➅➆
t
➀ ➁ ➂
on

➀ ➃
off 0%

Sensor
Zero
Display ➄
0%
Zero on on ON CK S D A 0 8

1 2 off off 1 2 3 4 5 6 7 8

PC
1 2 3 4 5 6 7 8
Address

Histo ROM
SW
HW

P01-PMx7xxxx-19-xx-xx-xx-075 P01-xxxxxxxx-19-xx-xx-xx-105
Fig. 4: Operating key, external under the protective flap Fig. 5: Operating key and operating elements, internal
1 Operating key or position adjustment 1 Green LED to indicate value is accepted
(zero point-correction) or total reset 2 Operating key for position adjustment
(zero point-correction) or total reset
3 DIP switch for hardware address
4 Slot for optional display
5 Slot for optional HistoROM®/M-DAT
6 DIP-switch for locking/unlocking
measured-value-relevant parameters
7 DIP-switch for damping on/off

14 Endress+Hauser
Deltapilot S PROFIBUS PA Operation

5.2.2 Function of the operating elements – on-site display not connected

Operating key(s) Meaning

– Position adjustment (zero point correction): Press key for at least 3 seconds. If the
LED on the electronic insert lights up briefly, the pressure applied has been
0% accepted for position adjustment.
See also Page 27 (Level measuring mode) or Page 29 (Pressure measuring
Zero mode).
– Total reset: Press key for at least 12 seconds. If the LED on the electronic insert
P02-xxxxxxxx-19-xx-xx-xx-107
lights up briefly, the reset is being carried out.

Set address in the bus. See also Page 21, Section 4.4 "Configuring the device
on ON CK S D A 0 8

address".
off 01 1 2 3 4 5 6 7 8

1 2 3 4 5 67 8
Address
SW
HW

P01-xxxxxxxx-19-xx-xx-xx-109

– DIP-switch 1: for locking/unlocking measured-value-relevant parameters


τ on Factory setting: off (unlocked)
– DIP switch 2: damping on/off
Factory setting: on (damping on)
1 2 off
P01-xxxxxxxx-19-xx-xx-xx-108

Endress+Hauser 15
Operation Deltapilot S PROFIBUS PA

5.2.3 Function of the operating elements – on-site display connected

Operating key(s) Meaning

O – Navigate upwards in the picklist


– Edit the numerical values and characters within a function

S – Navigate downwards in the picklist


– Edit the numerical values and characters within a function

F – Confirm entry
– Jump to the next item

O F
Contrast setting of on-site display: darker
and

S F
Contrast setting of on-site display: brighter
and

ESC functions:
– Exit edit mode without saving the changed value.
– You are in a menu within a function group. The first time you press the keys
simultaneously, you go back a parameter within the function group. Each time you
O S and press the keys simultaneously after that, you go up a level in the menu.
– You are in a menu at a selection level. Each time you press the keys simultaneously,
you go up a level in the menu.
Note: The terms function group, level and selection level are explained in Section
4.3.1, Page 17.

Set address in the bus. See also Page 21, Section 4.4 "Configuring the device
on ON CK S D A 0 8

address".
off 01 1 2 3 4 5 6 7 8

1 2 3 4 5 67 8
Address
SW
HW

P01-xxxxxxxx-19-xx-xx-xx-109

16 Endress+Hauser
Deltapilot S PROFIBUS PA Operation

5.3 On-site operation via on-site display


5.3.1 Structure of the operating menu
The menu is split into four levels. The three upper levels are used to navigate while you use the
bottom level to enter numerical values, select options and save settings.
For the entire menu see CD-ROM, Operating Instructions BA00356P.
The structure of the OPERATING MENU depends on the measuring mode selected, e.g. if the
"Pressure" measuring mode is selected, only the functions necessary for this mode are displayed.

Measured value

GROUP SELECTION

LANGUAGE MEASURING MODE QUICK SETUP OPERATING MENU



SETTINGS DISPLAY

POSITION ADJUSTMENT BASIC SETUP EXTENDED SETUP

POS. ZERO ADJUST


➃ POS. INPUT VALUE

CALIB. OFFSET

P01-xxxxxxxx-19-xx-xx-xx-145
Fig. 6: Structure of the operating menu
1 1. Selection level
2 2. Selection level
3 Function groups
4 Parameter

Endress+Hauser 17
Operation Deltapilot S PROFIBUS PA

5.3.2 Selecting an option


Example: select "English" as the language of the menu.

On-site display Operation

German is selected as the language. A ✓in front of the


menu text indicates the active option.

P01-xxxxxxxx-19-xx-xx-xx-017

Select English with "+" or "–".

P01-xxxxxxxx-19-xx-xx-xx-033

1. Confirm your choice with "E". A ✓ in front of the


menu text indicates the active option. (English is now
selected as the menu language.)
2. Jump to the next item with "E".

P01-xxxxxxxx-19-xx-xx-xx-034

18 Endress+Hauser
Deltapilot S PROFIBUS PA Operation

5.3.3 Editing a value


Example: adjusting DAMPING VALUE function from 2.0 s to 30.0 s. See also Page 16,
Section 4.2.3 "Function of the operating elements".

On-site display Operation

The on-site display shows the parameter to be changed.


The value highlighted in black can be changed. The "s"
unit is fixed and cannot be changed.

P01-xxxxxxxx-19-xx-xx-xx-023

1. Press "+" or "–" to get to the editing mode.


2. The first digit is highlighted in black.

P01-xxxxxxxx-19-xx-xx-xx-027

1. Use "+" to change "2" to "3".


2. Confirm "3" with "E". The cursor jumps to the next
position (highlighted in black).

P01-xxxxxxxx-19-xx-xx-xx-028

The decimal point is highlighted in black, i.e. you can now


edit it.

P01-xxxxxxxx-19-xx-xx-xx-029

1. Keep pressing "+" or "–" until "0" is displayed.


2. Confirm "0" with "E".
The cursor jumps to the next position.  is displayed
and is highlighted in black. See next graphic.

P01-xxxxxxxx-19-xx-xx-xx-030

Use "E" to save the new value and exit the editing mode.
 See next graphic.

P01-xxxxxxxx-19-xx-xx-xx-031

Endress+Hauser 19
Operation Deltapilot S PROFIBUS PA

On-site display Operation

The new value for the damping is now 30.0 s.


– Jump to the next parameter with "E".
– You can get back to the editing mode with "+" or
"–".

P01-xxxxxxxx-19-xx-xx-xx-032

5.3.4 Taking pressure applied at device as value


Example: performing position adjustment.

On-site display Operation

The bottom line on the on-site display displays the pressure


present, here 3.9 mbar.

P01-xxxxxxxx-19-xx-xx-xx-158

Use "+" or "–" to switch to the "Confirm" option. The active


selection is highlighted in black.

P01-xxxxxxxx-19-xx-xx-xx-159

Use "E" to assign the value (3.9 mbar) to the POS. ZERO
ADJUST parameter. The device confirms the calibration
and jumps back to the parameter, here POS. ZERO
ADJUST (see next graphic).

P01-xxxxxxxx-19-xx-xx-xx-037

Switch to the next parameter with "E".

P01-xxxxxxxx-19-xx-xx-xx-160

20 Endress+Hauser
Deltapilot S PROFIBUS PA Operation

5.4 Configuring the device address


Note the following points:
• An address must be assigned to every PROFIBUS PA device. Only when the address is
configured correctly will the device be recognised by the control system/master.
• Each address may only be assigned once in each PROFIBUS PA network.
• Valid device addresses are in the range from 0 to 125.
• The address 126 set at the factory can be used to check the function of the device and to
connect to a PROFIBUS PA network already in operation. This address then has to be changed
in order to integrate additional devices.
• All devices have the address 126 and software addressing on leaving the factory.
• The FieldCare operating program is delivered with the address 0 (default setting).
There are two ways of assigning the device address to Deltapilot S:
• Using a DP Class 2 master operating program, such as FieldCare or
• On site using the DIP switches

➀ 0%

Zero
Sensor Display

on on ON CK S D A 0 8
2 + 8 = 10
1 2 off off 1 2 3 4 5 6 7 8
PC

1 2 3 4 5 6 7 8
Address
Histo ROM

SW
HW ON CK S D A 08
on
250002272-–

HW-Version: 1
SW-Version: 2

– + E
off 01 1 2 3 4 5 6 7 8

1 2 3 4 5 6 7 8
Address
SW
HW
P01-xxxxxxxx-19-xx-xx-xx-112
Fig. 7: Configuring the device address using the DIP switches
1 If necessary, remove on-site display (optional)
2 Set the hardware address via the DIP switches

5.4.1 Hardware addressing


Hardware addressing is configured as follows:
1. Set DIP switch 8 (SW/HW) to "Off".
2. Configure the address with DIP switches 1 to 7.
3. You have to wait 10 seconds for a change in address to take effect. The device is restarted.

DIP switch 1 2 3 4 5 6 7

Weighting in Position "On" 1 2 4 8 16 32 64

Weighting in Position "Off" 0 0 0 0 0 0 0

Endress+Hauser 21
Operation Deltapilot S PROFIBUS PA

5.5 Locking/unlocking operation


Once you have entered all the parameters, you can lock your entries against unauthorised and
undesired access.
You have the following possibilities for locking/unlocking the operation:
• Via a DIP-switch on the electronic insert, locally on the display (see Page 14, Fig. 4).
• Via the on-site display (optional)
• Via digital communication.
The -symbol on the on-site display indicates that operation is locked. Parameters which refer
to how the display appears, e.g. LANGUAGE and DISPLAY CONTRAST can still be altered.

! Note!
• If operation is locked by means of the DIP-switch, you can only unlock operation again by
means of the DIP-switch. If operation is locked by means of remote operation e.g. FieldCare,
you can only unlock operation again by means of remote operation.
The table provides an overview of the locking functions:

Locking via View/ Modify/write via1) Unlocking via


read
parameter On-site Remote DIP-Switch On-site Remote
display operation display operation

DIP-Switch yes no no yes no no

On-site display yes no no no yes yes

Remote operation yes no no no yes yes

1) Parameters which refer to how the display appears, e.g. LANGUAGE and DISPLAY CONTRAST can still be
altered.

Locking/Unlocking operation via on-site display or remote operation

Locking operation 1. Select INSERT PIN NO. parameter,


Menu path: GROUP SELECTION  OPERATING MENU OPERATION 
INSERT PIN NO.
2. To lock operation, enter "0" for this parameter.

Unlocking operation 1. Select INSERT PIN NO. parameter.


2. To unlock operation, enter "2457" for the parameter.

22 Endress+Hauser
Deltapilot S PROFIBUS PA Commissioning

6 Commissioning
# Warning!
• If a pressure smaller than the minimum permitted pressure is present at the device, the
messages "E120 Sensor low pressure" and "E727 Sensor pressure error - overrange" are output
in succession.
• If a pressure greater than the maximum permitted pressure is present at the device, the
messages "E115 Sensor overpressure" and "E727 Sensor pressure error - overrange" are
output in succession.
• Messages E727, E115 and E120 are "Error"-type messages and can be configured as a
"Warning" or an "Alarm". These messages are configured as "Warning" messages at the
factory. These messages are configured as "Warning" messages at the factory. In applications
where the user is consciously aware of the fact that the sensor range can be exceeded (e.g.
cascade measurement), this setting prevents the transference of status BAD.
• We recommend setting messages E727, E115 and E120 to "Alarm" in the following instances
( see Operating Instructions BA00356P):
– The sensor range does not have to be exceeded for the measuring application.
– Position adjustment has to be carried out that has to correct a large measured error as a
result of the orientation of the device.

! Note!
By default, the device is set up for the Level measuring mode, "Level Easy Pressure" level
selection and "%" as the unit.

Endress+Hauser 23
Commissioning Deltapilot S PROFIBUS PA

6.1 Position adjustment


Due to the orientation of the device, there may be a shift in the measured value, i.e. when the
container is empty, the measured value parameter does not display zero. There are three options
to choose from when performing position adjustment.
(Menu path: GROUP SELECTION  OPERATING MENU SETTINGS POSITION
ADJUSTMENT)

Parameter name Description

POS. ZERO ADJUST (685) Position adjustment – the pressure difference between zero (set point) and the measured
Entry pressure need not be known. (A reference pressure is present at the device.)
Example:
– MEASURED VALUE = 2.2 mbar
– Correct the MEASURED VALUE via the POS. ZERO ADJUST parameter with the
"Confirm" option. This means that you are assigning the value 0.0 to the pressure
present.
– MEASURED VALUE (after pos. zero adjust) = 0.0 mbar
The CALIB. OFFSET parameter displays the resulting pressure difference (offset) by which
the MEASURED VALUE was corrected.
Factory setting:
0

POS. INPUT VALUE Position adjustment – the pressure difference between zero (set point) and the measured
(563) pressure need not be known. (A reference pressure is present at the device.)
Entry
Example:
– MEASURED VALUE = 0.5 mbar
– For the POS. INPUT VALUE parameter, specify the desired set point for the
MEASURED VALUE, e.g. 2 mbar.
(MEASURED VALUEnew = POS. INPUT VALUE)
– MEASURED VALUE (after entry for POS. INPUT VALUE) = 2.0 mbar
– The CALIB. OFFSET parameter displays the resulting pressure difference (offset) by
which the MEASURED VALUE was corrected.
CALIB. OFFSET = MEASURED VALUEold – POS. INPUT VALUE,
here: CALIB. OFFSET = 0.5 mbar – 2.0 mbar = – 1.5 mbar)
Factory setting:
0

CALIB. OFFSET (319) Position adjustment – the pressure difference between zero (set point) and the measured
Entry pressure is known.
Example:
– MEASURED VALUE = 2.2 mbar
– Via the CALIB. OFFSET parameter, enter the value by which the MEASURED VALUE
should be corrected. To correct the MEASURED VALUE to 0.0 mbar, you must enter
the value 2.2 here.
(MEASURED VALUE new = MEASURED VALUEold – CALIB. OFFSET)
– MEASURED VALUE (after entry for calib. offset) = 0.0 mbar
Factory setting:
0

24 Endress+Hauser
Deltapilot S PROFIBUS PA Commissioning

6.2 Level measurement


6.2.1 Quick Setup menu for Level measuring mode – on-site display

! Note!
• Some parameters are only displayed if other parameters are appropriately configured (see the
following table).
• The following parameters are set to the following values in the factory:
– LEVEL SELETION: Level Easy Pressure
– CALIBRATION MODE: Wet
– OUTPUT UNIT or LIN. MEASURAND: %
– EMPTY CALIB.: 0.0
– FULL CALIB.: 100.0
•  For parameter description see CD-ROM, Operating Instructions BA00296P.
• The quick setup is suitable for simple and quick commissioning. If you wish to make more
complex settings, e.g. change the unit from "%" to "m", you will have to calibrate using the
BASIC SETTINGS group.
• See also Page 16, Section 4.2.3 "Function of the operating elements" and Page 17,
4.3 "On-site operation via on-site display".

MEASURED VALUE

GROUP SELECTION

LANGUAGE MEASURING MODE QUICK SETUP OPERATING MENU

LEVEL SELECTION

Pressure Level

POS. ZERO ADJUST


1)
EMPTY CALIB.

1)
1) – LEVEL SELECTION "Level Easy Pressure"
and CALIBRATION MODE "Wet" FULL CALIB.
– LEVEL SELECTION = "Level Standard",
LEVEL MODE = "Linear" and
DAMPING VALUE
CALIBRATION MODE = "Wet"

P01-xMx7xxxx-19-xx-xx-xx-049
Fig. 8: Quick Setup menu for the Level measuring mode

Endress+Hauser 25
Commissioning Deltapilot S PROFIBUS PA

On-site operation

Measured value display


On-site display: Switch from the measured value display to GROUP SELECTION with F.

GROUP SELECTION
Select MEASURING MODE.

MEASURING MODE
Select "Level" option.

LEVEL SELECTION
Select level mode.

GROUP SELECTION
Select QUICK SETUP menu.

POS. ZERO ADJUST


Due to orientation of the device, there may be a shift in the measured value. You correct the MEASURED VALUE via
the POS. ZERO ADJUST parameter with the "Confirm" option, i. e. you assign the value 0.0 to the pressure present.

EMPTY CALIB. 1)
Enter level for the lower calibration point.
For this parameter, enter a level value which is assigned to the pressure present at the device.

FULL CALIB. 1)
Enter level for the upper calibration point.
For this parameter, enter a level value which is assigned to the pressure present at the device.

DAMPING TIME
Enter damping time (time constant ). The damping affects the speed at which all subsequent elements, such as the
on-site display, measured value and OUT value of the Analog Input Block react to a change in the pressure.

1) – LEVEL SELECTION "Level Easy Pressure" and CALIBRATION MODE "Wet"


– LEVEL SELECTION "Level Standard", LEVEL MODE "Linear" and CALIBRATION MODE "Wet"
(Menu path for CALIBRATION MODE: GROUP SELECTION OPERATING MENU SETTINGS BASIC
SETTINGS)

26 Endress+Hauser
Deltapilot S PROFIBUS PA Commissioning

6.2.2 On-site operation – on-site display not connected


If no on-site display is connected, the following functions are possible by means of the key on
the electronic insert or on the exterior of the device:
• Position adjustment (zero point correction)
• Device reset, see also Page 15, Section 4.2.2 "Function of the operating elements", Table.

! Note!
• The operation must be unlocked. See page 22, Section 4.5 "Locking/unlocking operation".
• The pressure applied must be within the nominal pressure limits of the sensor. See
information on the nameplate.
Carry out position adjustment:
1. Pressure is present at device.
2. Press key for at least 3 seconds. See Page 14, Section 4.2.1 "Position of operating
elements".
3. If the LED on the electronic insert lights up briefly, the pressure applied has been accepted
for position adjustment.
If the LED does not light up, the pressure applied was not accepted. Observe the input
limits.

Endress+Hauser 27
Commissioning Deltapilot S PROFIBUS PA

6.3 Pressure measurement


6.3.1 Quick Setup menu for Pressure measuring mode – on-site display

! Note!
See also Page 16, Section 4.2.3 "Function of the operating elements" and Page 17, 4.3 "On-site
operation via on-site display".

MEASURED VALUE

GROUP SELECTION

LANGUAGE MEASURING MODE QUICK SETUP OPERATING MENU

Pressure Level

POS. ZERO ADJUST

DAMPING VALUE

P01-xMx7xxxx-19-xx-xx-de-048
Fig. 9: Quick Setup menu for Pressure measuring mode

On-site operation

Measured value display


On-site display: Switch from the measured value display to GROUP SELECTION with F.

GROUP SELECTION
Select MEASURING MODE.

MEASURING MODE
Select "Pressure" option.

GROUP SELECTION
Select QUICK SETUP menu.

POS. ZERO ADJUST


Due to orientation of the device, there may be a shift in the measured value. You correct the MEASURED VALUE via
the POS. ZERO ADJUST parameter with the "Confirm" option, i. e. you assign the value 0.0 to the pressure present.

DAMPING TIME
Enter damping time (time constant ). The damping affects the speed at which all subsequent elements, such as the
on-site display, measured value and OUT value of the Analog Input Block react to a change in the pressure.

28 Endress+Hauser
Deltapilot S PROFIBUS PA Commissioning

6.3.2 On-site operation – on-site display not connected


If no on-site display is connected, the following functions are possible by means of the key on
the electronic insert or on the exterior of the device:
• Position adjustment (zero point correction)
• Device reset, see also Page 15, Section 4.2.2 "Function of the operating elements", Table.

! Note!
• The operation must be unlocked. See page 22, Section 4.5 "Locking/unlocking operation".
• The pressure applied must be within the nominal pressure limits of the sensor. See
information on the nameplate.
Carry out position adjustment:
1. Pressure is present at device.
2. Press key for at least 3 seconds. See Page 14, Section 4.2.1 "Position of operating
elements".
3. If the LED on the electronic insert lights up briefly, the pressure applied has been accepted
for position adjustment.
If the LED does not light up, the pressure applied was not accepted. Observe the input
limits.

Endress+Hauser 29
Commissioning Deltapilot S PROFIBUS PA

30 Endress+Hauser
Deltapilot S PROFIBUS PA Commissioning

Endress+Hauser 31
KA01023P/00/EN/16.14
71270391
CCS/FM+SGML 9.0
71270391

You might also like