The Multiwave PRO microwave reaction system delivers the sample preparation Convenient tool-free handling
solutions you need to obtain excellent trace analysis results. Multiwave PRO masters Opening and sealing Multiwave PRO’s rotors, vessels and sensors is quick and simple - just
high temperatures and provides comprehensive reaction control features. Its wide use your hands. This tool-free handling is unique to Anton Paar and simplifies this frequently
repeated work step.
range of accessories allows digestion, leaching, oxygen combustion, solvent
extraction, drying, evaporation, and UV digestion with one single system.
Fit for pharma
Anton Paar supports you with an instrument-specific qualification package for Multiwave PRO.
Superior trace elemental analysis begins with outstanding sample preparation.
This package includes DQ, IQ, OQ and PQ, Risk Analysis, 21 CFR Part 11 checklist, Deviation
Even when using the best analytical equipment, you still need to optimize sample
List, Traceability Matrix and userSOP. The documentation fulfills the requirements of GMP,
preparation conditions to obtain reliable measurement values. With over 40 years
21 CFR Part 11, GAMP 5, USP<1058> and covers all steps of a complete instrument
of experience, Anton Paar is your trusted resource.
qualification.
SMART VENT
Lip-type seal vessel technology vessel
Pressure jacket
High-end
Rotor 64MG5 41HVT56 24HVT50 24 HVT80 16MF100 16HF100 8NXF100 8NXQ80
Use Rotor 8N to obtain the highest digestion quality. Rotor 8N masters
temperatures up to 300 °C at pressures up to 80 bar for longer than 2 hours.
Continuous pressure monitoring of all eight sample vessels ensures safe
and precise reaction control, even with critical samples such as ceramics,
semiconductors, ores, slags, ashes, plastics, oil, grease, coal and samples No. of vessels 64 41 24 24 16 16 8 8
from refractories.
T in 16 T in all
Workhorse Reaction control
positions positions
T in all positions T and p in all positions
Rotor 24HVT50 and Rotor 24HVT80 with vessels using SMART VENT technology offers
you reliable pressure control and the possibility to safely digest samples with different Liner material
Glass
PTFE-TFM PTFE-TFM PTFE-TFM PTFE-TFM PTFE-TFM PTFE-TFM n.a.
(PTFE seal)
reaction behavior in one run. The internal temperature of each vessel can be controlled
without the need of any optional sensor. These rotors are ideal for the digestion of
various kinds of samples, including food and biological samples, waste water and
sludge, soil and sediments, EPA procedures, environmental samples, agricultural
samples, cosmetic and pharmaceutical samples. Pressure jacket n.a. n.a. n.a. n.a. PEEK-GF Ceramic Ceramic Quartz
The 41-position rotor meets the demand for high sample throughput and eliminates Volume 5 mL 56 mL 50 mL 80 mL 100 mL 100 mL 100 mL 80 mL
your bottlenecks in sample preparation. The three-part vessel is designed for tool-free
handling, simple cleaning, a long lifetime and low cost of consumables. Rotor 41HVT56
can be used for the digestion of various kinds of samples, including food and biological
samples, environmental samples, agricultural samples, cosmetic and pharmaceutical HF resistant No Yes Yes Yes Yes Yes Yes No
samples.
Wireless temperature measurement and control Wireless pressure measurement and control
An infrared sensor measures the temperature at the base of each Based on the SMART VENT technology used in the HVT vessels
the overpressure is safely and reliably released. Thus the maximum
reaction vessel. If the temperature rises too high, Multiwave PRO
Sample temperature is not limited by the applied sample amounts. This
Reaction reduces microwave power so that the temperature does not concept permits the digestion of large sample quantities and the
vessel
exceed a preset limit. digestion of samples with different reaction behavior in one run.
Infrared Overpressure
sensor
By controlling the internal temperature of each vessel By measuring the pressure in a reference vessel
IR data
transmission
Precise control of digestion processes in each vessel. The result: quick and precise control, even of fast and
spontaneous reactions.
Rotor 24HVT50, Rotor 24HVT80 and Rotor 41HVT56 vessels provide
the internal temperature in each vessel without the need for an optional Sensor A hydraulic pressure sensor with or without temperature sensor
electronics
sensor. The digestion runs can be controlled based on different is integrated into one of the reaction vessels. It measures the
Sample
Reaction temperature control strategies. The hottest sample, the coldest sample pressure and the pressure increase rate every 20 milliseconds.
vessel Sensors If one of the preset limits is exceeded, the microwave power
or the average temperature of all samples can be used as references.
Infrared Hydraulic is immediately reduced and, if required, the cooling effort is
sensor transducer increased.
By measuring the temperature in a reference vessel By measuring the pressure in all vessels every
IR data Immersion IR data transmission Sensor electronics
transmission thermometer 20 milliseconds
Quick and precise control, even of fast and spontaneous
reactions. The result: An extremely quick and simultaneous pressure
measurement allows you to apply the highest
Sensor A temperature sensor is integrated into one of the reaction vessels. possible temperatures for acid digestion.
electronics
With Multiwave PRO you can define temperature ramps, making
it a valuable tool for digestion of unknown samples and method Critical situations emerging from spontaneous reactions, for
Sensors development. example. are intercepted due to the high frequency of pressure
measurements. Multiwave PRO immediately reduces microwave
power and the unwanted reaction is stopped before it leads to
Hydraulic transducer pressure release.
Special Solutions
Some samples require special treatment. Multiwave PRO provides
unique methods which are efficient alternatives to digestion.
Cone-type seal
Oxygen
You want to digest combustible You need to digest liquid You want to extract organic
Reflux rinses
inner surface solids quickly and cleanly? samples containing organic compounds?
compounds?
Microwave-induced oxygen Microwave-assisted extraction
combustion (MIC) XQ80
UV digestion
Efficient alternative to slow and tedious
Unique, clean and quick method. Unique microwave-powered UV lamps solvent extraction methods.
Halogens or metals are trapped in a low- produce radicals which digest the Improves performance and throughput of
concentration absorption solution which Solvent HPLC-based or GC-based analysis.
sample.
can be measured without dilution. The resulting low-concentration solution Compliant with US-EPA and ASTM
Suitable for all combustible solids such can be measured without dilution. methods.
Sample pellet
as wood, paper, coal, food or polymers. Sample Reduces extraction times from hours to
UV lamp For ultra-trace analysis of e.g. seawater,
Parallel combustion of up to 8 samples in effluents, sewage, body fluids or minutes.
Absorption highly resistant quartz vessels. beverages. Suitable for extractions of PCBs, PAHs and
solution It replaces steel combustion bombs. Stir bar hydrocarbons from environmental and food
Liquid sample Low analytical blanks.
samples, derivatization reactions prior to
analysis and polymer extractions.
Filter disk
Filter disk
You want to accelerate your
evaporation procedures? You want to dry samples before
digestion?
Holder for Microwave-assisted evaporation
4 vials You want to run protein
To external
hydrolysis in a fraction of the scrubber Microwave drying
Rotor 8EVAP simplifies and
time?
accelerates the microwave-
Rotor 1DRY efficiently dries samples.
assisted evaporation of acids and
Microwave-assisted protein Efficient drying: four times faster than
concentration of aqueous sample
hydrolysis with conventional methods.
solutions.
Provides samples without carbonization
Protein hydrolysis in less than 1 hour. The external scrubber neutralizes the
or contamination.
For milligram to gram amounts. acid vapors, washing out up to 95 %.
Humidity and unwanted odors are
Precise temperature control during Prior to digestion, sample solutions
HCl removed via the exhaust system.
reaction. can be concentrated to
Inert gas atmosphere applicable. increase analyte levels and improve
digestion efficiency.
Protein sample After evaporation, just transfer the
liner into the digestion rotor.
Ceramic
plate
Exhaust
system