Four-Channel Sound & Vibration Analyser


The SVAN 958A is an unique four-channel instrument offering All measurement results are stored in non-volatile 32 MB internal
20kHz-band sound & vibration analysis. It is a perfect choice for all memory or USB memory stick and can be easily downloaded to a PC
applications that require simultaneous class 1 noise measurements & with SvanPC++ software.
tri-axial vibration assessment. SVAN 958A with RS 232 interface (SV 55) can be offered with
Each of four input channels can be independently configured for GPRS modem or LAN & WLAN connection module. Together
sound or vibration detection with different filters and RMS detector with SvanPC++_RC remote communication software, these
time constants giving users an enormous measurement flexibility. interfaces provide easy remote access to instrument settings
The real advantage of SVAN 958A is the capability to perform & data over Internet and local area network.
advanced analysis simultaneously to the level meter mode. Instrument is powered from four AA standard or rechargeable
In practise this allows to obtain broad-band results such as Leq, RMS, batteries as well as from the external DC power source or USB
LMax, LMin, LPeak together with four-channel analysis like FFT or interface.
octave band analysis. List of available analyser functions includes FFT, Robust case and light weight design accomplish the exceptional
1/1 or 1/3 octave, cross spectra, sound intensity, RT 60 and more. The features of this new generation instrument.
4-channel time-domain signal recording to .wav format is available
as well as an exceptional feature!

Features
 Four-channel, 20 kHz real-time, simultaneous sound and vibration
measurements
 Analyser options:
 FFT analysis up to 1600 lines in 20 kHz band

 1/1 and 1/3 octave real-time analysis

 Time-domain signal recording

 Sound intensity

 FFT cross spectra

 RT 60 reverberation time

 Acoustic dose meter

 Human Vibration measurements meeting ISO 8041:2005


including VDV and MTVV
 Sound level meter class 1, IEC 61672
 Advanced data logger including spectral analysis
 USB memory stick providing almost unlimited logging capability
 Advanced trigger and alarm functions
 USB 1.1 Client and USB Host interfaces
 RS 232 for modems support (GPRS, Ethernet, WLAN)
 Powered by four AA standard or rechargeable batteries
 Easy in use, hand held, light weight and robust

continuous innovation
SVAN 958A
Technical Specification
Vibration Level Meter & Analyser
Standards ISO 8041:2005, ISO 10816-1
Meter Mode RMS, VDV, MTVV or Max, Peak, Peak-Peak
Simultaneous measurement in up to four channels with independent set of filters and detector constants

Analyser (option) 1/1 octave real-time analysis, 15 filters with centre frequencies from 1 Hz to 16 kHz (class 1, IEC 61260)

1/3 octave real-time analysis, 45 filters with centre frequencies from 0.8 Hz to 20 kHz (class 1, IEC 61260)

FFT analysis up to 1600 lines with Hanning, Kaiser-Bessel or Flat Top window

FFT cross spectra measurements

RPM rotation speed measurements parallel to the vibration measurement (1 ÷ 99999)
and more…
Filters Wd, Wk, Wc, Wj, Wm, Wb, Wg (ISO 2631), Wh (ISO 5349), HP1, HP3, HP10, Vel1, Vel3, Vel10,
VelMF, Dil1, Dil3, Dil10, KB (DIN 4150)
RMS & RMQ Detectors Digital true RMS & RMQ detectors with Peak detection, resolution 0.1 dB
Time constants: from 100 ms to 10 s
Accelerometer (option) SV 84 triaxial high sensitivity accelerometer for ground or building vibration measurements (1 V/g)
SV 38 low-cost triaxial accelerometers for whole-body measurements (1 V/g MEMS type)
Measurement Range Accelerometer dependent (with SV 84: 0.0005 ms-2 RMS ÷ 50 ms-2 PEAK)
Frequency Range 0.5 Hz ÷ 20 kHz; accelerometer dependent

Sound Level Meter & Analyser


Standards Class 1: IEC 61672-1:2002
Meter Mode SPL, Leq, SEL, Lden, Ltm3, Ltm5, Statistics - Ln (L1-L99), LMax, LMin, LPeak
Simultaneous measurement in three profiles per channel (up to four channels)
with independent set of filters and detector time-constants

Analyser (option) 1/1 octave real-time analysis, 15 filters with centre frequencies from 1 Hz to 16 kHz (Type 1, IEC 61260)

1/3 octave real-time analysis, 45 filters with centre frequencies from 0.8 Hz to 20 kHz (Type 1, IEC 61260)
RT 60 reverberation time analysis in 1/3 octave bands

FFT analysis up to 1600 lines with Hanning, Kaiser-Bessel or Flat Top window

FFT cross spectra measurements
Sound Intensity measurements
and more…
Weighting Filters A, C, Z and G
RMS Detector Digital true RMS detector with Peak detection, resolution 0.1 dB
Time constants: Slow, Fast, Impulse
Microphone (option) MK 250, Class 1, 50 mV/Pa, prepolarised 1/2" condenser microphone with SV 12L preamplifier
SV 25, Class 2, dose meter, ceramic 1/2" microphone with integrated preamplifier
Measurement Range Total Dynamic Range: 16 dBA RMS ÷ 140 dBA Peak
Linearity Range (IEC 61672): 26 dBA RMS ÷ 140 dBA Peak
Frequency Range 0.5 Hz ÷ 20 kHz; microphone dependent, with MK 250 microphone: 3.5 Hz ÷ 20 kHz

Basic Data
Input IEPE type (channels 1, 2, 3 - LEMO 4-pin & channel 4 - TNC connector)
Dynamic Range 100 dB, 4 x 20 bits A/D converters
Frequency Range 0.5 Hz ÷ 20 kHz, sampling rate 48 kHz

Data Logger Time-history logging to internal memory or USB memory stick
Time-domain signal recording on USB memory stick (option)
Display Super contrast (10000:1) OLED 2.4” colour display (320 x 240 pixels)
Memory 32 MB non-volatile flash type, external USB memory stick (not included)
Interfaces USB 1.1 Client, USB 1.1 Host, RS 232 (option: SV 55 required)
Extended I/O - AC output (1 V Peak) or Digital Input/Output (Trigger / Pulse)

Power Supply Four AA batteries (alkaline) operation time > 10 h (6.0 V / 1.6 Ah)

Four AA rechargeable batteries (not included) operation time > 14 h (4.8 V / 2.6 Ah)

SA 17A external battery pack (option) operation time > 24 h
External power supply 6 V DC ÷ 24 V DC (1.5 W)
USB interface 500 mA HUB
Environmental Conditions Temperature from -10 oC to 50 oC
Humidity up to 90 % RH, non-condensed
Dimensions 140 x 82 x 42 mm
Weight 510 grams with batteries
 
each function parallel to the meter mode with USB 1.1 Host function not active and backlight off
Continuous product development and innovation are the policy of our company. Therefore, we reserve the right to change the specifications without prior notice.

DISTRIBUTOR: