Picospritzer Manual
Picospritzer Manual
Picospritzer Manual
SAFETY
Please read this entire set of instructions before attempting to use this manual for intracellular
or extracellular ejections. This instrument should only be used as specified by the
manufacturer. Use other than as specified may present a safety hazard. It is intended for
research purposes only. Not for use on humans or for diagnostic purposes.
Please do not remove the cover as hazardous voltages may be exposed. Use only the electric
cord supplied with the instrument or one of equal capacity.
For continuing safe operation of the unit periodic inspections of the connections in the pressure
system should be made to ensure they are all still tight, particularly after a period of inactivity or
if the system is moved.
This instrument is intended for use in an indoor environment between 10 and 40 degrees
Celsius (50 to 104 degrees Fahrenheit) with no more than 90% humidity (non-condensing).
EQUIPMENT SPECIFICATIONS
Power Requirements: 100 to 240 Volts AC
50 to 60 HZ
0.8 Amps @ 115 Volts AC
0.6 Amps @ 230 Volts AC
The instrument is equipped with a universal power supply which will operate on the full
range of voltages between 100 and 240 volts AC.
TABLE OF CONTENTS
1.0 INTRODUCTION
5.0 BACKGROUND
6.0 REFERENCES
APPENDIX 2 – DRAWINGS
Please read this entire set of instructions before attempting to use this instrument for
intracellular or extracellular ejections. These instructions cover the currently produced
system (P/N 052-0500-900).
Intracellular applications range from femtoliter ejections of RNA into small cells to
nanoliter ejections into oocytes. Extracellular applications range from picoliter
applications of dilute neuroactive substances onto neurons during intracellular recording
to the presentation of chemical stimuli to whole animals.
The system comes complete with high speed valves and the necessary tubing
assemblies (less pipette and holder). It is designed to fit a standard 19" relay rack. The
Picospritzer III comes fully adjusted and ready to use.
1. Position the control unit in the desired location. Position the remote valve box within
two feet of the experimental site (VELCRO® is supplied for mounting) and within 6 feet
of the control unit.
2. Connect the black ¼ inch pressure tubing to the side port on the front panel regulator
and to a source of clean dry compressed gas (air, nitrogen, CO2, or argon) at a
maximum of 150 PSI (10bar). DO NOT USE OXYGEN OR COMBUSTIBLE GASES.
The fitting on the regulator is a quick- connect type. The end of the tubing should be cut
cleanly and inserted fully into the opening in the fitting. An o-ring seals around the
outside of the tube. To remove the tubing from the fitting you must first vent the pressure
from the line then press the ring on the end of the fitting (around the tubing) toward the
body of the fitting while pulling on the tubing.
3. Use the 6-foot long 1/8-inch tubing assembly to connect the front panel regulator to
the remote valve box. The connection at the regulator is also a quick-connect type while
the remote valve box is threaded (1/4-28 male). There should be a ferrule pressed in to
this end of the tubing.
4. The 3-foot long 1/16-inch tubing assembly is used to connect the remote valve box
to a pipette holder (not supplied). Both ends of this assembly have ¼-28 nuts and
ferrules pre-assembled on it.
PLEASE NOTE: An important detail in arranging the location of the solenoid in relation
to the pipette holder is to maintain a loose coil in the interconnecting small bore (1/16
Pulse Durations:
3 to 999 milliseconds, 1 ms intervals
0.1 to 99.9 seconds, 0.1 sec. intervals
0.1 to 99.9 minutes, 0.1 min. intervals
External Inputs:
5-volt TTL input BNC jack per channel.
The unit is activated when the power switch is in the ON position and the red pilot light is
on. The unit must be connected to a source of clean, dry compressed gas at a pressure
at least as high as that desired for ejections. Clockwise rotation of the knurled knob on
the right of the rack mount panel will increase the pressure (indicated on adjacent
gauge) available for injection purposes. This unit contains a self-bleeding regulator so
that the pressure may be reduced by simply rotating the pressure control knob counter-
clockwise.
The system is rated to operate up to 100 PSI (6.9 bar). One can be assured that all the
connections are gas-tight by rotating the knob on the front panel so that the meter
indicates approximately 80 PSI of pressure and then shutting off the main source of
pressure to the panel. If a gradual reduction of pressure is observed, it indicates that
there is a leak somewhere between the remote valve housing and the main pressure
source. The site of this leak may be found by carefully listening for a hissing sound or by
application of a dilute soap solution (or “snoop”) and watching for the formation of
bubbles at various connections. It is essential to have leak proof connections throughout
the system to assure reproducible results and to avoid unnecessary loss of gases.
Volume ejected is a linear function of both pulse duration and pressure (see references
1, 3). The pulse duration has a greater dynamic range with more accurate and
reproducible settings. Thus, pulse duration will be changed frequently during the course
of the experiment while pressure settings will be changed relatively little on a daily basis.
4.3 INDICATORS
The Picospritzer III contains four LED indicators. “ON” is the power indicator. The
“PULSE” indicator shows that the timer is active; the channel 1 and channel 2 indicators
show when each channel is active.
4.8 MARKER
The time mark provides a convenient indication of the duration of the pulse with respect
to a biological signal much like that of the “artifact” associated with iontophoresis. It is a
5-volt TTL signal controlled by channel one. The time mark not only provides a useful
indication of the duration of the pulse, but it may also serve as a sync-out signal for
triggering the sweep of an oscilloscope, etc.
Since the duration setting is easier to use and more repeatable it is usually the one that
is varied. An operating pressure is selected and kept that way for a given
experiment. You may need to make a few test droplets at different pressures and
durations with a given pipette to determine what operating pressure gives you the
desired range by varying just the duration. When setting the pressure, it is important to
do so by increasing the value. If you over shoot your desired set point always back
down to less than that point and then raise it back up.
The volume ejected during a single pulse depends on the applied pressure, the pulse
duration, AND the pipette tip diameter. Every pipette must be calibrated by optically
measuring the droplet sizes that result from varying the pressure or duration.
A typical pipette will only hold 1 to 4 microliters of media. Doing one ejection in the
microliter range may be possible but multiple ejections are not. Doing multiple ejections
to total 1 to 2 microliters is possible but the user still must calibrate the system, know the
number of ejections that will be required to reach the desired volume, and keep track of
the number of ejections completed
5.0 BACKGROUND
In 1977, Dr. R.E. McCaman and associates provided a complete description (1) of a
pressure ejection system that utilized a high speed valve. This valve continues to be the
heart of the pressure system offering very precise control of ejection volumes (in the
picoliter range) and ejection times (in the millisecond range).
These systems have been used for intracellular as well as extracellular ejections. In
listing advantages of the pressure system, these investigators emphasize that the linear
As you find additional uses for your Picospritzer, please send us a reprint for addition to
our reference section so that others may benefit from your experience.
6.0 REFERENCES
References describing the use and unique advantage of pressure systems in several
types of experimentation in the field of neurobiology, cell biology, and biophysics are:
1. McCaman, R.E., Mc Kenna, D.G. and Ono, J.K. “A pressure system for intracellular
and extracellular ejections of picoliter volumes.” Brian Research 136:141 (1977).
2. Sakaki, M., Sakai, H. and Woody, C.D. “Intracellular staining of cortical neurons by
pressure micro-injection of horseradish peroxidase and recovery by core biopsy.” Exp.
Neurol. 58:138 (1978)
3. Sakai, M., Swartz, B.E. and Woody, C.D. “Controlled micro release of
pharmacological agents: measurements of volume ejected in vitro through fine tipped
glass microelectrodes by pressure.” Neuropharmacol. 18:209 (1979)
4. Dufy, B., Vincent J-D. Fluery, H., Pasquier, P., Gourdji, D., and Tixler- Vidal, A.
“Membrane effects of thyrotropin-releasing hormone and estrogen shown by intracellular
recording from pituitary cells.” Science. 204.509 (1979)
5. Tauc, L., Hoffman, A., Tsuji, S., Hinzen, D. and Faille, L. “Transmission abolished in
a cholinergic synapse after injection of acetylcholinesterase into the presynaptic neuron.”
Nature, 250, 496 (1974)
6. Chang, J.J., Gelperin, A., and Johnson, F.H. “Intracellulary injected aequorin detects
transmembrane calcium flux during action potentials in an identified neuron from the
terrestrial slug.” Brain Research 77:431 (1974)
10. Eisenstadt, M., Goldman, J.E., Kandel, E.R., Koike, H., Koester, J. and Schwartz, J.
“Intrasomatic injection of radioactive precursors for studying transmitter synthesis in
identified neurons of Aplysia.” Proc. Nat. Acad. Sci. 70:3371 (1973)
11. Schwartz, J.H. “Synthesis, axonal transport and release of acetylcholine by identified
neurons of Aplysia.” Proc. Nat. Acad. Sci. 70:3371 (1973)
12. Thompson, E.B., Schwartz, J.H. and Kandel, E.R. “A radio autographic analysis in
the light and electron microscope of identified Aplysia neurons and their processes after
intrasomatic injection of L-H3-Fucose.” Brain Research.112:251 (1976)
13. Baux, G., Simonneau, M. Tauc, L. “Transmitter release, Ruthenium red used to
demonstrate a possible role of sialic acid containing substrates.” J. Physiol. 291, 161-
178 (1979)
14. Sakai, M., Sakai, H., and Woody, C.D. “Sampling distribution of morphologically
identified neurons of coronal- pericruciate cortex of awake cats following intracellular
injection of HRP.” Brain Research. 152: 329-333 (1978)
15. Amaral, D.G. and Price, J.L. “An air pressure system for the injection of tracer
substances into the brain.” Jrnl. Neuroscience Methods. 9:35-34 (1983)
16. Waitzman, D.M., Sliakov, V.L., DePalma-Bowles, S. and Ayers, A.S. “Effects of
reversible inactivation of the primate mesencephalic reticular formation. I. Hypermetric
goal-directed saccades.” Journal of Neurophysiology (2000)
https://journals.physiology.org/doi/full/10.1152/jn.2000.83.4.2260
17. Pham, L.N., Dionne, M.S., Shiarsu-Hiza, M. and Schneider, D.S. “A specific primed
immune response in drosophila is dependent on phagocytes.” PLOS
https://journals.plos.org/plospathogens/article?id=10.1371/journal.ppat.0030026 (2007)
18. Kim, L.K., Choi, U.Y., Cho, H.S., Lee, J.S., Lee, W., Kim, J., Jenog, K., Shim, J.,
Kim, J. and Kim, Y. “Down-regulation of NF-κB target genes by the AP-1 and STAT
complex during the innate immune response in drosophila.” PLOS
https://journals.plos.org/plosbiology/article?id=10.1371/journal.pbio.0050238 (2007)
19. Kuhlman, S.J. and Huang, Z.J. “High-resolution labeling and functional manipulation
of specific neuron types in mouse brain by cre-activated viral gene expression.” PLOS
https://journals.plos.org/plosone/article?id=10.1371/journal.pone.0002005 (2008)
20. Gibbs, G.M., Orta, G., Reddy, T., Koppers, A.J., Martinez-Lopez, P., Vega-Beltran,
The diameter of the droplet ejected in air was measured with an ocular micrometer in a
dissecting microscope. The volume was calculated for droplets formed by varying pressure or
pulse duration parameters. The following graphs demonstrate that each pipette can be
calibrated by varying these two major determinants of the volume ejected.
The X-intercept (15 MSEC) represents the mechanical lag time of the particular solenoid
valve.
The X-intercept (7.5PSI) represents the minimum pressure necessary for ejection and is a
characteristic of a particular pipette.
The points of this graph are highly correlated (R=.97) with the independently determined
specific activity (6.5 x 106 CPM/mL) of the radioactive solution.
FIGURE 3A:
FIGURE 3B:
Problems of desensitization can be circumvented with a pressure pipette since the pipette is
usually filled with agonists in the concentrations of 10-6 to 10-3 M in contrast to iontophoretic
pipettes. Braking current is not necessary for the pressure pipette, thus avoiding inconsistent
ejection of compounds.
FIGURE 3D:
Comparison of response to iontophoretic (I) and pressure (P) applied ACh in a drug study. Both
responses are equally antagonized by hexamenthonium, even though the ACh in the pressure
pipette is ejected in a droplet of normal artificial seawater. Similar results are obtained during
substitution experiments. The arrows in the figure point to an input resistance test pulse.
2. I used to purchase a 0-10, 0-30 or 0-60psi regulator, are those options still available?
The 0-10, 0-30, or 0-60psi regulator options are no longer available, only the 0-100psi
configuration is available. Please see Appendix 1 for alternative regulator information.
3. Can I trigger the two channels independently using two separate footswitches?
The Picospritzer III can only handle one footswitch (the footswitch controls the internal timer).
You can trigger them independently using an external trigger. Please see Section 4.7 Second
Channel for more details.
The Picospritzer III will only provide the same pressure to both valve boxes.
Check the capillary tip size. Typically, we recommend a tip size between 1 and 5 microns. Once
the resistance of the tip is higher than the capillary effect of the pipette, material will no longer
be drawn into the pipette. Keep in mind that other factors such as pressure, timing, and the
nature of the material being ejected, may come into play.
6. What is the shortest pulse duration the Picospritzer III can support?
A minimum of 3 milliseconds should be used to ensure the valve in the valve box can fully
actuate for injection.
7. What are the approximate shipping weights and dimensions of the Picospritzer III
The Picospritzer III typically ships in a box with dimensions of 24” x 14” x 9” and weighs
approximately 10 lbs. Actual shipping weights and dimensions may vary.
ALTERNATIVE REGULATORS:
PIPETTE HOLDERS:
PIPETTES:
FOOTSWITCH: