Ventilator: CCV-700B User'S Manual
Ventilator: CCV-700B User'S Manual
Ventilator: CCV-700B User'S Manual
USER’S MANUAL
Safety Guide
1. Before operating the ventilator, please read this manual thoroughly, and make sure
that all the cautions and instructions are strictly followed.
2. To avoid any unexpected ill effects on patients in case the power is cut off suddenly,
make sure the ventilator is equipped with a built-in backup battery before operation.
3. Main power supply to the ventilator should be equipped with a protective ground. If
this cannot be ensured, apply the backup battery.
4. DO NOT operate the ventilator under circumstances that contain flammable, explosive
or narcotic air, since the unit is not explosion proof. Do not let liquid enter the
mechanism since the unit is not waterproof. If there is any liquid splashing on the unit,
please wipe it immediately.
5. Before putting the ventilator into use, make sure it has been cleaned thoroughly and
disinfected, and each function is in good order. NEVER continue to operate the
ventilator which has been detected to be malfunctioning.
6. Before being connected to the windpipe of a patient, the ventilator should be
stationed, adjusted and tested on a simulated lung firstly.
7. The electrical wire and breathing tubes should be well laid in order not to interfere with
personnel’s movements in the sickroom. The airway cannot be under any pressure in
case it is out of shape and blocked. Do not move the ventilator casually when it is in
use so as to avoid such accidents as the airway or wire falling off and the oxygen
cylinder toppling over.
8. There must be qualified medical personnel guarding on the spot during the operation
of the ventilator. Pay attention to the working status of ventilator and humidifier,
meanwhile, pay attention to the patient’s Life Indicator and vigor analyzing data, and
adjust the ventilator to the most appropriate status for the patient.
9. Although the ventilator is considered technically mature and highly reliable in design,
any instruments have the possibility to malfunction unexpectedly. For the sake of the
patient’s safety, please do prepare a standby ventilator of good and reliable
conditions.
Figure & Type Matter
3. Technical ················································································3
6. Troubleshooting······································································· 9
11. Others···················································································· 12
1.Overview
1.1Scope of Application and Main Features
such functions as time, volume cycling, pressure limit, etc. It is mainly intended for
providing ventilation support to a critically ill patient during the life threatening phase and
ensuring the going-through of the dangerous period by the patient and smooth treatment
respiratory function of the patient, and also provides ventilation assistance for the patient
during the recovery from a disease or operation. Its main features are as follows:
A.Gas drive and electrical control, time-pressure switching and pressure limit control.
B. A high-brightness LED digital display is used to present the control frequency, tidal
C.A highly sensitive and responsive pressure sensor and a flow sensor are used to
measure, control and display the airway pressure and gas flow rate and the ventilator is
D.In case of an abnormality to the ventilator or misoperation, the ventilator can raise a
Requirements and Test Methods for Medical Electrical Equipment to operate in Climatic
Environment Group II and Mechanical Environment Group II. Its normal operating
— — Gas source requirement: medical oxygen source with a pressure ranging from
The CCV-700B ventilator is gas driven by medical compressed oxygen and compressed
air. In the inspiratory phase, two streams of compressed gas (compressed oxygen and
compressed air) flow into a high performance air-oxygen mixer to form a mix of oxygen
and air with a certain pressure. Such mix of oxygen and air flows into a high performance
electrically controlled inspiratory proportional valve and is delivered through the
inspiratory circuit of the ventilator into the airway in the patient for mechanical ventilation.
In the expiratory phase, the gas exhaled by the patient reaches the expiration control
valve through a filter and an expiratory circuit to be discharged into the atmosphere.
During such process, a high performance proportional valve, a highly sensitive flow
sensor, a pressure sensor and a single-chip microcomputer control system are used and
control in the time fixed, volume controlled and constant pressure modes is realized by
adjusting the airway pressure and airway flow applied to the patient in a closed loop
mode.
3.Technical Features
——End-inspiratory plateau;
——SIPPV
——IPPV
——IMV
——SIMV
——PEEP
——SPONT
——SIGH
——MANUAL
3.2 Technical Data
—Tidal volume range: no less than 50 to 1200ml, permissible deviation: ±20 %.
——Maximum minute ventilation: ≥ 18 L/min, permissible deviation: ±20 %.
deviation: ±15 %.
——Oxygen consumption: the variation in the gas pressure in the cylinder should be
less than or equal to 1.5MPa/h when the ventilator operates on a 12250KPa / 40L medical
permissible deviation: +1 s, -2 s.
——Sigh (deep breath): the inspiration time should be no less than 1.5 times of the
original setting.
Alarm
Alarm Item Alarm Condition Alarm Form
Level
The gas source
Low gas pressure drops to a
source level below
pressure 0.2MPa.
Range: 1.0 ~
Upper
6.0KP a, Immediate alarm; “high level alarm”
airway
High permissible indicator flickers raising an audible
pressure
deviation: ±20% alarm.
limit
(90% of set value)
Continuous The airway The maximum delay is no longer than
positive pressure exceeds 17s. The “high level alarm” indicator
pressure 15hpa. flickers raising an audible alarm.
The alarm will be raised after one
Low tidal breathing cycle. The “medium level
Lower than 50ml
volume alarm” indicator flickers raising an
audible alarm.
The airway
Lower pressure drops to The alarm will be raised after a delay
airway 0.5KPa. The of 4 to 15s. The “medium level alarm”
pressure Medium permissible indicator flickers raising an audible
limit deviation is ±0.2 alarm.
KPa.
Upper The throughput is Immediate alarm. The “medium level
throughput higher than alarm” indicator flickers raising an
limit 18L/min. audible alarm.
Battery The battery voltage
voltage is lower than 10.5V.
3.4 Battery
This ventilator is equipped with an internal backup power supply of the voltage 12V ± 10%,
rated capacity 7Ah and maximum current 2A. The fully recharged battery can support the
operation of this ventilator for a duration of no less than 30min. In case of a utility main
failure, this ventilator can automatically switch to operate on the internal power supply.
——Type I device.
——Type B application
——Common (IPX0)
——Continuous operation.
E)This product is not equipped with an application part for protection against defibrillation
F)This product is not equipped with signal output and input parts.
damage during transportation and that contents of the packing box are consistent with the
packing list. Keep properly the damping cushions in the packing box for use in another
transportation.
Learn how to use the front and rear control panels of this ventilator . Check the position of
the pointer of the pressure gauge. If the pointer is not at zero point, then adjust the zero
Check the medical compressed oxygen source and make sure that its pressure ranges
from 280 to 600KPa and flow rate is 50L/min. If you supply oxygen with a cylinder, then
you also need to check and make sure that the cylinder is sufficient of oxygen, and that
the pressure reducer on the cylinder functions well and is correctly installed.
Check and make sure that the single-phase AC power supply used for this ventilator is of
the AC 220V±10% voltage and securely grounded for protection and make sure that the
Before the first use, you need to check whether the corresponding components are
washed and disinfected following the method as specified in Section 8 of this user
manual.
4)Connect two gas guiding screw tubes to the inspiration outlet and expiration inlet of the
ventilator respectively, connect the other ends of these tubes to a tee tube and connect
5)Connect the pressure signal interface on the tee tube to the “pressure signal input
interface” on the front panel of the ventilator circuit casing by using a tube.
6)Install the flow sensor between the “expiration screw tube adapter” and screw tube and
connect the output signal wire of the sensor to the “flow signal input interface” on the front
panel of the ventilator circuit casing.
7)Turn the “tidal volume adjustment” knob to the middle position.
8)Set the IPPV frequency to 20times/min.
9)Set the “inspiration triggering pressure” to -0.2 kPa.
10)Set the airway pressure limit to 4.0kPa.
11)Connect the medical compressed oxygen source with a pressure ranging from 280 to
600kPa.
12)When the above adjustment is completed, you can connect the power supply of the
ventilator.
4.3 Test Operation of Ventilator
After connecting the gas source and power supply of the ventilator and turning on the
power to start up the ventilator, you should observe:
1)That the ventilator operates in the ventilation mode in which it is shut down previously.
2)Such parameters as “spontaneous breathing frequency”, “overall respiratory rate”,
“throughput”, etc. have to be displayed one minute after the startup as they are refreshed
once every minute.
3)The indication of the airway pressure does not exceed 4.0kPa.
4)The expiration indicator and inspiration indicator flicker alternatively and you can hear
the close-open sound of the solenoid valve in the ventilator. The frequency at which the
solenoid valve closes and opens and the indicators flicker is the indicative value of the
“control frequency”. Also, you can see that the height of the water column in the test lung
increases and decreases alternatively at the control frequency.
5) The indicative value of the tidal volume on the test lung is substantially the same as that
of the “tidal volume” on the ventilator and relative error between both does not exceed
±20 %.
Self-test Interface
The system will automatically test four parameters: 1. Whether the pressure of the air
source ranges from 280 to 600kPa. 2. Whether the oxygen source pressure ranges rom
280 to 600kPa. 3. Whether the oxygen concentration sensor fails. 4. Whether the
If self-test of the above mentioned four parameters is passed, a “√” mark will be displayed
following each parameter, the screen stays for 3s and then the system automatically
enters the weight setting interface. If any of these parameters fails the self test, such
parameter will be shown in red background with a “×” mark, and the ventilator will halt on
the self-test screen and wait for medical staffs to solve and confirm the problems of the
parameter. The medical staffs may feed back the problems to our After-sales Service for
solution. With the “×” problem unsolved, you can press and hold the Mode key to
Caution: if self-test is not passed under the condition that the air
source is not connected and only the oxygen source is
connected, you can mandatorily have the ventilator operate.
Operating Interface
After entering the operating interface, the ventilator starts to operate at the preset
parameters.
The operating interface consists of the testing area, waveform area, status
The alarm message is the prompt information sent by the ventilator after it identifies an
abnormality. An alarm message in red indicates a high priority alarm and an alarm
The testing area includes parameters needing to be set in various modes and
real-time parameters tested. Such parameters are defined as follows:
Frequency: this refers to the overall respiration rate and is namely the mechanical
control frequency.
Peak pressure value: maximum pressure in the airway during each breath cycle.
Airway resistance: pressure difference generated per unit of flow rate in the airway.
Lung compliance: this refers to the change in the lung volume which is caused when
unit pressure changes.
Operation method for parameter setting interface: move the jog cursor to the option
needing to be set, press the jog key and the corresponding option is reversely displayed.
Select the corresponding value by moving cursor left or right and press the jog
Upon completion of setting of all parameters, turn the cursor to the OK key, press the
jog key and the ventilator operates at the newly set parameters. If you select Return, then
the ventilator continues to operate at the original parameters and switch to the operating
interface.
For the setting method for the alarm setting interface, you can refer to the instructions
for the setting of the modes and parameters. The contents and sequence of the setting on
the alarm setting interface are as follows:
Volume setting: 1-10, increase of the number means the volume increases
Alarm history query: as shown in the figure above, the specific time when
different alarms occur will be shown if you press the “Alarm History” key, a
maximum of 100 alarms can be stored, the alarm is in the “time->alarm contents”
format and also high and medium level alarms will be shown in red and yellow
respectively.
Upon completion of setting of all alarm parameters, turn the cursor to the OK key, press
the jog key and the ventilator operates at the new alarm parameters. If you select Return,
then the ventilator continues to operate at the original parameters and switches to the
operating interface.
If you press and hold the System key for 3s on the operating interface, the system setup
interface pops up
Upon completion of setting of all system parameters, turn the cursor to the OK or
Return key, press the jog key and the ventilator switches to the operating interface.
Instructions for Film Panel Switches
6 functional keys are located on the front panel of the ventilator and are as defined
below:
Silence key: in case of an alarm, pressing this key can keep the alarm silenced for
120s. If the time is up and the alarm is not disarmed, then the ventilator continues to
System key: if you press and hold this key for more than 2s, then the system setting
Mode key: if you press this key on the operating interface, then the parameter setting
Alarm key: if you press this key on the operating interface, then the alarm setting
Standby key: If you press and hold this key for more than 2s on the operating
interface, the ventilator stops operating. If you press and hold this key for more than
2s again, the ventilator immediately operates at the parameters as set on the system.
The SIPPV mode is the default mode in which the ventilator operates when it is
started up.
Such mode is mainly intended for patients with no or weak and intermittent spontaneous
breathing. If the patient shows no spontaneous breathing, the respirator provides
intermittent positive pressure ventilation of the patient following the set parameters, and
this ventilation mode is namely called controlled ventilation mode. When the spontaneous
synchronized with the spontaneous breathing in the patient, and this ventilation mode is
namely called assisted ventilation mode. The control ventilation and assisted ventilation
Operating parameters should be preset on the test lung following the procedures as
follows:
1)Connect the gas source and power and confirm that the ventilator operates in the
2)Adjust the “Adjust IPPV Frequency” knob and the “Control Frequency” digital display
4)Adjust the “Adjust Tidal Volume” knob, observe the “Tidal Volume” indicator and set the
tidal volume to the value as needed. For adult patients, make the initial setting based on
the value of 10mL per kg of body weight and then fine tune the value according to the
5)The airway pressure indicator shows the variation in the airway pressure in real time.
Carefully adjust the “Airway Pressure Limit” according to the airway pressure peak to set
the airway pressure limit to a level slightly higher than the peak pressure.
6 ) Set the “Inspiration Triggering Pressure”. When the spontaneous breathing in the
patient is recovered to a certain extent, the inspiration triggering pressure will provide a
ventilation synchronization signal to the respirator. At the same time, each time when the
patient takes a breath spontaneously the inspiration triggering pressure indicator flickers
once. Generally, the inspiration triggering pressure can be set to be a level 0.1kPa lower
than the minimum airway pressure when the patient has no spontaneous breathing.
7 ) Adjust the “PEEP” knob, observe the minimum airway pressure displayed when
expiration ends and judge whether the setting of the PEEP is appropriate or not. The
the ventilator provides one ventilation at a big tidal volume (no less than 1.5 times of the
Only upon completion of the above setting can you remove the test lung and connect the
After connecting the ventilator with the airway in the patient, you should carefully
observe the symptoms and lung inflation of the patient and further fine adjust the
operation status of the respirator according to the monitor instrument and arterial blood
Press the “Select Ventilation Mode” key to enable the “Control” indicator to be on and the
ventilator enters this operating mode. The “Spontaneous Breathing Frequency” digital
display provides no presentation due to the condition that no spontaneous breath is taken
and other displays still show the corresponding contents. The setting of the operating
In this mode, you can still select (or not select) the PEEP and Sigh functions.
In the event that the AC power supply to the ventilator is down, the ventilator can operate
the emergency battery. The output voltage of the battery gradually drops during operation.
If such output voltage drops to a level which is insufficient to drive the ventilator to operate,
you should replace the battery with one with sufficient power in time. If you cannot replace
with a new battery or in an urgent case when you cannot find a new battery, you can apply
Such mode needs to be operated by a physician with rich clinical experiences. That
physician should press the “Manually Controlled Ventilation” button at a certain rhythm to
simply maintain the respiration of the patient. Each time the physician presses the button
the ventilator ventilates the patient once and such parameters as the ventilation time, tidal
volume, circuit pressure, etc. are completely manually controlled by the physician.
When pressing the button, the physician must pay close attention to the lung inflation of
the patient and indication on the airway pressure gauge. The ventilator uses an pneumatic
pressure gauge and such gauge will not be affected in case of a sudden power supply
failure.
The ventilator should be immediately washed and disinfected after it is used and then
should be necessarily serviced and maintained.
6.Troubleshooting
20 and upper pressure limit to 4kPa. Block the patient end of the tee tube. Under this
condition, the upper airway pressure limit alarm should occur each time when the
If the problem cannot be solved after these measures are taken, please notify MINA
Remove the power plug of the ventilator from the AC power outlet, open the ventilator and
expose the safety valve on the ventilation circuit, and seal the patient end of the tee tube
and inlet and outlet on the ventilator back panel. Press and hold the Manually Controlled
Ventilation button and observe the airway pressure gauge. When the pressure gauge
indication should range from 5.5 to 6kPa, the safety valve begins to release gas. If the
pressure gauge indication is too high or low when the safety valve operates, please
contact MINA Medical or its authorized service agency for adjustment and repair.
7.3 Humidifier
During the use of the humidifier, you should pay attention to observing the temperature of
the output gas and water volume in the humidifier. If the patient inhales gas with too high
temperature, adverse reactions will occur and even respiratory tract burn will be caused. If
the water volume is insufficient, then the humidifier may be burnt out.
7.5 Fuse
The fuse of the ventilator (fuse tube) is installed on the rear panel.
If the “Operate on Battery” indicator goes on (indicating that the ventilator is powered by
the battery) and the power grid is not down, then we can judge that the fuse is burnt out.
You must replace the fuse following the procedures as below.
1)Do not turn the power switch of the ventilator in operation. You can simply extract the
plug of the ventilator from the power outlet and the ventilator automatically switches to
operate on the battery and then replace the fuse.
2) Screw off the cover of the fuse seat by using a screwdriver and replace the fuse.
3)Fuse tube specification: 1A glass casing fuse tube of Φ5×20 mm dimension.
the respiratory cannula of the patient with a new or disinfected tube and wash and
After the patient does not use the ventilator, you should ultimately disinfect it: thoroughly
wash and disinfect the ventilator and then install it for reuse. The ventilator which is not
used for a long time must be washed and disinfected before it is reused.
Work records for the washing and disinfection should be kept for archiving and query.
10.1Transportation Conditions
——The original packing box and damping cushion of the ventilator should be used
as specified.
— — Standardize the transportation based on the indications and symbols on the
11.Others
The pressure reducer used on the oxygen cylinder and humidifier for warming and
humidifying the output gas of the ventilator are not standard configuration of the ventilator
and need to be otherwise purchased.
You are welcomed to make consultation by call and letter, and when necessary we can
provide further technical data.