Troubleshooting QMatrix QM - QC
Troubleshooting QMatrix QM - QC
Troubleshooting QMatrix QM - QC
Software Exceptions
When a software exception occurs in the QMatrix program a message window pops up
and shows the available information about the exception.
All software exceptions are also recorded in a logfile located in the installation directory
of QMatrix: %InstallDir%\QMatrix\Status\RQuantron.log.
Software exceptions can help you to analyse problems and can assist you during
service activities. How you can handle with the most common software exceptions is
described below. The possible solutions are listed in a “most common happened”
ordered.
Errorgroup:
TCP/IP Socket-Error
Description:
The QMatrix program could not establish a connection to the instrument. The QMatrix
program turns into FAILURE (offline) mode as shown in the status bar below.
The QMatrix program can now be used for configuration purposes only.
Running measurements are not possible in FAILURE mode. The QMatrix program
must be restarted in order to get back into READY mode again.
Solutions:
Check if instrument’s Main Switch (red) is turned on.
-or-
Check if network cable coming out of the instrument is plugged into correct port on PC
site. Usually an additional PCI network adapter is installed in the PC for communication
with the instrument.
-or-
Check the two LED’s beside the RJ45 plug on PC’s additional network adapter. If the
network cable is connected and the instrument in on, one LED should be continuously
on, indicating the established network link. The other LED should flash (glow),
indicating network traffic.
-or-
Check the configuration of the additional network adapter by using the ipconfig
command on the commandline (Start->Run->cmd: ipconfig /all). One of the network
adapters should be configured like this:
Ethernet adapter Local Area Connection:
Errorgroup:
TCP/IP Socket-Error
Description:
A command that was requested to be executed by the Application-Controller has not
been acknowledged in time. There seems to be a problem in the Application Controller
while processing a command.
Solutions:
Send the Logfile to [email protected] and contact the QUANTRON Service.
The Logfile is usually located at:
%InstallDir%\QMatrix\Status\RQuantron.log
Errorgroup:
TCP/IP Socket-Error
Description:
The given name (Host or Servicename) could not be translated to an IP-Address.
Solutions:
In file C:\WINDOWS\SYSTEM32\DRIVERS\ETC\HOSTS:
In C:\WINDOWS\SYSTEM32\DRIVERS\ETC\SERVICES:
Use the standard Windows text editor (Notepad.exe) for adding these lines to the end
of the files.
Hint: Sometimes these files are replaced during a network setup by the customer.
Attention: These filenames have no extension, do not assign any extension when
saving with notepad!
Errorgroup:
Data-Acquisition-Error
Description:
An expected trigger did not occur during data acquisition.
Solutions:
Check if the DAQ (68 pin SCSI) cable coming out of the instrument is plugged and
threaded properly on the installed DAQ board on PC site.
-or-
Check the DAQ (68 pin SCSI) cable for damages, maybe one or more pins in the plug
on the cable site are out of shape or broken.
-or-
Check the fibre-optic connection between Readout-Master and Source-Master board.
-or-
Check the fibre-optic connections between Readout-Master and DAQ-Interface board.
Errorgroup:
Data-Acquisition-Error
Description:
The QMatrix program could not process the acquired data fast enough.
Solutions:
Check if another Windows program is running on the Instrument PC which consumes a
lot of CPU time. This can be evaluated by using the Task Manager (press Ctrl+Alt+Del
and choose Task Manager). Sometimes printing routines may consume much process
time with a higher priority than the data acquisition process is running on.
Errorgroup:
Data-Acquisition-Error
Description:
This error occurs when data transfer from the hardware to the driver cannot keep up
with the input clock rate.
Solutions:
When the PC wakes up after the Windows “Standby-Mode” this error occurs during the
next measurement. The computer must be restarted before the next measurement.
Please do not enable any Standby-Mode in Windows.
-or-
Check the fibre-optic connections between Readout-Master and DAQ-Interface board.
If the two fibre-optics (Trigger & Clock) are exchanged, this exception message will
occur as well.
Errorgroup:
Data-Acquisition-Error
Description:
An internal error occurred while calling a DAQ board’s driver function.
Solutions:
Reinstall the DAQ board’s driver software from CD (OmniCD) that was shipped with the
Instrument-PC. The name of the board is DT321.
-or-
Replace the DAQ board.
Errorgroup:
QDL-Error
Description:
This error is generated by the QDL-Server running on the Application-Controller. An
unknown command was received by the QDL-Server or a parameter for a received
command was not specified or has an invalid value.
Solutions:
There might be a version conflict between the QMatrix program and the QDL-Server.
Update the QDL-Server as described below.
1. Exit QMatrix
2. Make sure the instrument is switched on
3. Start the Windows-Explorer and navigate to
%InstallDir%\QMatrix\Tools\Q-APPCTRL\FTP_Update\
4. Doubleclick with the left mousebutton on the link:
(NOT FTP_Update.bat)
5. After this link has been executed, switch off the instrument and wait 5 seconds
6. Switch on the instrument
7. Start QMatrix
-or-
Send the Logfile to [email protected] and contact the QUANTRON Service.
The Logfile is usually located at:
%InstallDir%\QMatrix\Status\RQuantron.log
Errorgroup:
I2C-Error
Description:
This error is generated by the QDL-Server running on the Application-Controller. An
error occurred while accessing the I2C bus which connects the I/O-Board to the
Source-Master.
Solutions:
Check, if voltage on I/O-Board’s power supply plug is ok (->see electrical drawings)
-or-
Check the blue cable between the I/O-Board and the Source-Master.
-or-
Replace the I/O-Board.
Errorgroup:
DMB-Error
Description:
This error is generated by the QDL-Server running on the Application-Controller. An
error occurred while accessing the DMB (DIN MEASUREMENT BUS) which connects
one or more Drive-Controllers to the Application Controller.
Solutions:
Check, if voltage on Drive-Controller’s power supply plug is ok (->see electrical
drawings and ISEL Product Information)
-or-
Check the cable between the Drive-Controller and the Application-Controller (15 pin
DSub).
-or-
Replace the Drive-Controller.
Errorgroup:
Function Parameter Error
Description:
This error is generated by the IO-System when a requested steppermotor position is
out of the specified range.
Solutions:
If this exception occurs during the start of QMatrix and after a QMatrix software update,
check the file %InstallDir%\QMatrix\Settings\IOSignals.ini for this entry and modify it
like:
[IO/SIGNALS/OUT_STEPPER_MASK]
UseDefaultValue = 0
Description:
There were no intensities measured during measurement.
Solutions:
If changes to the method were made:
• Check if all sequences that are related to the measurement channels are
configured with a measure-time greater than 0 seconds (Edit->Measuretimes).
• Check if all channels are related to a measured sequence
• Check if all channels are set to valid integrator numbers
• Check if there are any warnings displayed in the channel list
Chiptool.exe
This program is provided by BECK, the manufacturer of the one-chip computer that is
used on the Application-Controller. The program is located at:
%InstallDir%\QMatrix\Tools\Q-APPCTRL\BECK\Chiptool_V4018\Chiptool.exe
After starting the program it scans the network for BECK controllers and lists the found
ones. This program should only be used for diagnostic purposes.
Start->Programs->QUANTRON->QMatrix->Tools->QUANTRON Helpdesk
Basics
ISL LIGHT enables you to see your customer's desktop over the internet when
help is needed. Using ISL LIGHT there will be no more doubts what your customer is
talking about over the phone. You can see your client's desktop on the screen and you
are able to remotely control his keyboard and mouse. ISL LIGHT can be used as an
online support for your software products and services, or for distance collaboration
when preparing presentations, business plans or other documents. You can also
exchange files as simply as using drag and drop.
Easy to use
ISL LIGHT can be used in a few seconds. No installation or configuration is needed.
The only thing your customer needs is an internet connection (ISDN or faster) to
download the ISLClientLight.exe from your website. You lead your customer over the
phone through the process of establishing connection.
Security
Connection over ISL LIGHT cannot be established until unique session number is
entered. Secure SSL connection and file transfer. When the connection is terminated,
the session number becomes obsolete. Please read ISL LIGHT Security Statement for
details.
Requirements
- MS Windows or Linux operating system
- internet access and any web browser
- Port 7615 must be open for outgoing connections on client site when using a firewall
Benefits
- access any PC worldwide
- efficient online customer support
See detailed documentation for using the QUANTRON Helpdesk system in file:
QUANTRONHelpdeskUserGuide.pdf