Nellcor N-560 - Service Manual
Nellcor N-560 - Service Manual
Nellcor N-560 - Service Manual
ai 1/24/06 9:25:48 AM
CM
MY
CY
CMY
K
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor, OxiCliq,
OxiBand, Dura-Y, MAX-FAST, SatSeconds, PediCheck, Oxismart and OXIMAX are trademarks of Nellcor
Puritan Bennett Inc.
To obtain information about a warranty, if any, contact Nellcor’s Technical Services Department, or your
local representative.
Purchase of this instrument confers no express or implied license under any Nellcor Puritan Bennett patent to
use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or more of the following US patents and foreign equivalents: 4,802,486; 4,869,254; 4,928,692;
4,934,372; 4,960,126; 5,078,136; 5,485,847; 5,743,263; 5,865,736; 6,035,223; 6,298,252; 6,463,310; 6,591,123;
6,675,031; 6,708,049; 6,801,797; Re 35,122.
Contents
Contents
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vii
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Warnings ....................................................................................................................................................1
Cautions .....................................................................................................................................................1
Notes ..........................................................................................................................................................1
Manual Overview ........................................................................................................................................2
Description of N-560 Pulse Oximeter .........................................................................................................2
Related Documents ....................................................................................................................................4
Routine Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Cleaning .....................................................................................................................................................5
Periodic Safety Checks ..............................................................................................................................5
Functional Checks ......................................................................................................................................5
Battery ........................................................................................................................................................6
Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introduction .................................................................................................................................................7
Equipment Needed .....................................................................................................................................7
Performance Tests .....................................................................................................................................8
Battery Charge ......................................................................................................................................8
Power-Up Performance .........................................................................................................................8
Power-On Self-Test (POST) ............................................................................................................9
Power-On Defaults and Alarm Limit Ranges .................................................................................11
General Operation ...............................................................................................................................16
LED Excitation Test ........................................................................................................................16
Operation with a Live Subject ........................................................................................................17
Alarm Volume Control ....................................................................................................................18
Pulse Tone Volume Control ...........................................................................................................19
Pulse Oximetry Functional Tests .........................................................................................................19
Introduction ....................................................................................................................................20
Initial Setup ....................................................................................................................................20
Test #1: BPM .................................................................................................................................21
Test #2: SpO2 ................................................................................................................................22
Test #3: Modulation Level ..............................................................................................................23
Test #4: Light .................................................................................................................................25
Safety Tests ..............................................................................................................................................26
Audible Alarm Settings and Service Functions . . . . . . . . . . . . . . . . . . . . 29
Introduction ...............................................................................................................................................29
Audible Alarm Settings .............................................................................................................................29
Alarm Silence State .............................................................................................................................30
Alarm Silence Duration ........................................................................................................................30
Alarm Volume ......................................................................................................................................30
N-560 Menu ..............................................................................................................................................30
Figures
Figure 1: N-560 Front Panel ...................................................................................................................3
Figure 2: N-560 Rear Panel ...................................................................................................................3
Figure 3: N-560 Controls ........................................................................................................................9
Figure 4: Adjusting High %SpO2 Alarm Limit .......................................................................................12
Figure 5: Adjusting Low %SpO2 Alarm Limit .......................................................................................13
Figure 6: Adjusting High Pulse Rate Alarm Limit .................................................................................13
Figure 7: Adjusting Low Pulse Rate Alarm Limit ..................................................................................14
Figure 8: Adjusting SatSeconds Alarm Limit ........................................................................................15
Figure 9: N-560 Controls ......................................................................................................................29
Figure 10: Year Month Display ...............................................................................................................36
Figure 11: Day Hour Display ..................................................................................................................36
Figure 12: Minute Second Display .........................................................................................................37
Figure 13: N-560 Rear Panel Screws .....................................................................................................55
Figure 14: N-560 Case Screws ..............................................................................................................55
Figure 15: Top Case Removal ...............................................................................................................56
Figure 16: Separated Cases ..................................................................................................................56
Figure 17: Front case Wiring ..................................................................................................................57
Figure 18: Front Case Mounting Screws ................................................................................................58
Figure 19: Front Case Bracket ...............................................................................................................58
Figure 20: Speaker Removal ..................................................................................................................59
Figure 21: Front PCB Removed .............................................................................................................59
Figure 22: NELL-3 Cable Connections ...................................................................................................60
Figure 23: NELL-3 Removal ...................................................................................................................60
Figure 24: Front case Wiring ..................................................................................................................61
Figure 25: Front Case Mounting Screws ................................................................................................61
Figure 26: Main PCB Removal ...............................................................................................................62
Figure 27: Main PCB Removed ..............................................................................................................62
Figure 28: Power Supply Wiring .............................................................................................................63
Figure 29: Power Supply Screws ...........................................................................................................64
Figure 30: Power Supply Removed ........................................................................................................64
Figure 31: Fan Wires ..............................................................................................................................65
Figure 32: Fan Bracket Removal ............................................................................................................65
Figure 33: Fan and Bracket Removal .....................................................................................................66
Figure 34: Fan Screws ...........................................................................................................................66
Figure 35: Battery Screws ......................................................................................................................67
Figure 36: Battery Cover ........................................................................................................................67
Figure 37: Battery Connector .................................................................................................................67
Figure 38: Battery Removal ....................................................................................................................68
Figure 39: Ground Wire ..........................................................................................................................69
Figure 40: Equipotential Ground Terminal .............................................................................................69
Figure 41: Equipotential Ground Terminal Removed .............................................................................70
Figure 42: Exploded View ......................................................................................................................73
Figure 43: Packing .................................................................................................................................76
Figure 44: Data Port Pin Layout .............................................................................................................98
Figure 45: Real-Time Printout ..............................................................................................................100
Ta b l e s
Table 1: Equipment Needed .................................................................................................................7
Table 2: N-560 Menu Selections .........................................................................................................30
Table 3: Institutional Default Values ....................................................................................................37
Table 4: Factory Default Settings ........................................................................................................38
Table 5: Problem Categories ..............................................................................................................44
Table 6: Power Problems ....................................................................................................................45
Table 7: Button Problems ....................................................................................................................46
Table 8: Display/Alarms Problems ......................................................................................................47
Table 9: Operational Performance Problems ......................................................................................48
Table 10: Data Port Problems ...............................................................................................................49
Table 11: Error Codes ...........................................................................................................................50
Table 12: Spare Parts and Accessories ................................................................................................71
Table 13: Tone Definition ......................................................................................................................80
Table 14: Cable Compliance .................................................................................................................87
Table 15: Electromagnetic Emissions ...................................................................................................88
Table 16: Electromagnetic Immunity .....................................................................................................88
Table 17: Electromagnetic Immunity .....................................................................................................90
Table 18: Recommended Separation Distances Between Portable and Mobile RF Communication
Equipment and the N-560 ............................................................................................91
Table 19: Data Port Pinouts ..................................................................................................................97
Table 20: Status Codes .......................................................................................................................103
Table 21: NELL-3 J1 Pinouts ..............................................................................................................110
Table 22: NELL-3 J2 Pinouts ..............................................................................................................111
Warnings
Warnings are identified by the WARNING symbol shown above and a line
above and below the warning text. The word WARNING and all warning text
are boldfaced.
Warnings alert the user to potential serious outcomes (death, injury, or
adverse events) to the patient or user.
Cautions
Cautions are identified by the CAUTION symbol shown above. The word
CAUTION and the caution text are boldfaced.
Cautions alert the user to exercise care necessary for the safe and effective use
of the N-560™.
Notes
Note: Sensor LED light emissions fall within Class 1 level, according to IEC
60825-1:2001. No special safety precautions are required.
Manual Overview
This manual contains information for servicing the Nellcor® model N-560
pulse oximeter. Only qualified service personnel should service this product.
Before servicing the N-560, read the operator's manual carefully for a
thorough understanding of operation.
2
Introduction
Related Documents
To perform test and troubleshooting procedures and to understand the
principles of operation and circuit analysis sections of this manual, you must
know how to operate the N-560. Refer to the N-560 operator's manual. To
understand the various Nellcor sensors that work with the N-560, refer to the
individual sensor's directions for use.
The latest versions of the operator’s manual and the service manual are posted
on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
Spare parts and accessories are posted on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/Apartweb/main/PartAcceMenu.html
4
Routine Maintenance
Cleaning
CAUTION: Do not spray, pour, or spill any liquid on the N-560, its
accessories, connectors, switches, or openings in the chassis.
For surface-cleaning and disinfecting follow your institution's procedures or:
• The N-560 may be surface-cleaned by using a soft cloth dampened
with either a commercial, nonabrasive cleaner or a solution of 70%
alcohol in water, and lightly wiping the surfaces of the N-560.
• The N-560 may be disinfected using a soft cloth saturated with a 10%
solution of chlorine bleach in tap water.
Before attempting to clean an SpO2 sensor, read the directions for use
enclosed with the sensor. Each sensor model has cleaning instructions specific
to that sensor.
2. Inspect safety labels for legibility. If the labels are damaged, contact
Nellcor’s Technical Services Department, 1.800.635.5267, or your local
Nellcor representative.
Functional Checks
1. If the N-560 has been visibly damaged or subjected to mechanical shock
(for example, if dropped), perform the performance tests. See
Performance Tests on page 8.
2. Perform the electrical safety tests detailed in Safety Tests on page 26. If
the N-560 fails these electrical safety tests, refer to Troubleshooting on
page 43.
3. Inspect the fuses for proper value and rating (F1 and F2 = 2 amp,
250 volts).
Battery
Nellcor recommends replacing the N-560's battery every two years. When the
N-560 is going to be stored for two months or more, remove the battery prior
to storage. To replace or remove the battery, refer to Disassembly Guide on
page 53.
If the N-560 has been stored for more than 30 days, charge the battery as
described in Battery Charge on page 8. A fully discharged battery requires
6 hours to receive a full charge. The battery is being charged whenever the
N-560 is plugged into an AC power source.
6
Per for mance Verification
Introduction
This section discusses the tests used to verify performance following repairs
or during routine maintenance. All tests can be performed without removing
the N-560 cover. All tests except the battery charge and battery performance
tests must be performed as the last operation before the N-560 is returned to
the user.
If the N-560 fails to perform as specified in any test, repairs must be made to
correct the problem before the N-560 is returned to the user.
Equipment Needed
Performance Tests
Note: This section is written using Nellcor factory-set defaults. If your
institution has preconfigured custom defaults, those values will be displayed.
Factory defaults can be restored. See Menu Item 10 (Restore Factory Default
Settings as Power-On Default) on page 38.
Battery Charge
Perform the following procedure to fully charge the battery.
1. Connect the N-560 to an AC power source.
2. Verify that the N-560 is off and that the AC Power indicator is lit.
Power-Up Performance
The power-up performance tests verify the following N-560 functions:
• Power-On Self-Test (POST) on page 9
8
Performance Verification
1. Turn on the N-560 by pressing the Power On/Off button for greater than
0.5 seconds.
Display Sound
5 (in pulse rate left window) low priority alarm
6 (in pulse rate center window) medium priority alarm
0 (in pulse rate right window) high priority alarm
n (in SpO2 left window)
n (in SpO2 center window)
5. Once the display test portion of POST is complete, the N-560 version is
displayed for approximately two seconds.
10
Performance Verification
Note: The version above is only a sample. Check your N-560 for the current
version installed.
The version number is needed when calling Nellcor's Technical
Services Department, 1.800.635.5267, or your local Nellcor representative for
technical assistance. Write down the number and have it available prior to
requesting technical assistance.
6. If the N-560 detects an internal problem during the POST, an alarm tone
sounds and the N-560 displays an Error Code and corresponding number.
See Troubleshooting on page 43.
WARNING: If you do not hear the POST pass tone, do not use the N-560.
Note: In addition to serving as the POST pass verification, the POST pass
tone also functions as an audible confirmation that the speaker is performing
properly. If the speaker does not function, the alarm sounds cannot be heard.
4. Verify that the N-560 emits a single beep and the %SpO2 display
indicates an upper alarm limit of “100” for about three seconds.
Note: “HI” in the display window indicates an upper alarm limit that is
being adjusted, and a “LO” in the window indicates that a low alarm limit is
being adjusted.
6. Press the SpO2 Alarm Limit button. Within 3 seconds, press and hold
the Adjust Down button. Verify that the %SpO2 display reduces to a
minimum of one point above the %SpO2 low alarm limits setting.
Note: A decimal point to the right of the value in either window display
indicates that the alarm limits are not power-on default values.
7. Press the SpO2 Alarm Limit button two times rapidly (twice within three
seconds). Verify that the N-560 emits two beeps and that the %SpO2
display indicates “85” for three seconds. Verify that “LO” is displayed in
the pulse rate window.
12
Performance Verification
8. Press the SpO2 Alarm Limit button two times rapidly. Within three
seconds, press and hold the Adjust Down button and verify that the
%SpO2 display reduces to a minimum of “20.” Press and hold the Adjust
Up button and verify that the %SpO2 display cannot be raised past the
upper alarm limit setting minus one point.
11. Wait for POST to be completed. The %SpO2 and pulse rate alarm limits
will be reset to the default values.
12. Press the Pulse Rate Alarm Limit button. Verify that the N-560 emits a
single beep, the Pulse Rate display indicates an alarm limit of “170”
and that the %SpO2 display window shows “HI” for about 3 seconds.
13. Press the Pulse Rate Alarm Limit Button. Within 3 seconds press and
hold the Adjust Down button. Verify that the Pulse Rate display
reduces to a minimum of one point above the pulse rate low limit alarm
setting.
14. Press the Pulse Rate Alarm Limit button two times (twice within 3
seconds). Verify that the N-560 emits two beeps and that the Pulse Rate
display indicates an alarm limit of “40” for 3 seconds. Verify that “LO” is
displayed in the %SpO2 display.
15. Press the Pulse Rate Alarm Limit button two times rapidly. Within 3
seconds press and hold the Adjust Down button. Verify that the Pulse
Rate display cannot be raised past the upper alarm limit setting minus
one point.
16. Turn the N-560 off by pressing the Power On/Off button.
18. Wait for POST to be completed. The %SpO2 and pulse rate alarm limits
will be reset to the default values.
19. Press the Pulse Rate Alarm Limit button two times rapidly. Within
three seconds press and hold the Adjust Up button. Verify that the Pulse
Rate display cannot be adjusted above the upper alarm limit setting
minus one point.
20. Press the SatSeconds Alarm Limit button. Verify that the N-560 emits a
single beep and that the Pulse Rate display indicates an alarm limit of
“OFF” for three seconds. Verify that “SAS” is displayed in the %SpO2
display.
14
Performance Verification
21. Press the SatSeconds Alarm Limit button. Within three seconds press the
Adjust Up button or the Adjust Down button. Verify that the Pulse
Rate display changes between 10, 25, 50, 100, and OFF.
22. Press the Power On/Off button to turn the N-560 off.
23. Press the Power On/Off button to turn the N-560 on.
24. Press and release the %SpO2 Alarm Limit button. Verify that the
%SpO2 display indicates an alarm limit of “100.”
25. Press the %SpO2 Alarm Limit button two times rapidly. Verify that the
%SpO2 display indicates an alarm limit of “85.”
26. Press and release the Pulse Rate Alarm limit button. Verify that the
Pulse Rate display indicates an alarm limit of “170.”
27. Press the Pulse Rate Alarm Limit button two times rapidly. Verify that
the Pulse Rate display indicates an alarm limit of “40.”
28. Press the SatSeconds Alarm Limit button. Verify that the Pulse Rate
display indicates an alarm limit of “OFF.”
29. Press the Power On/Off button to turn the N-560 off.
General Operation
examine LED intensity control. The red LED is used to verify intensity
modulation caused by the LED intensity control circuit.
1. Sensor Port
2. Connect a DOC-10 pulse oximetry cable to the N-560 sensor port (1).
5. Leave the sensor open with the LEDs and photo detector visible.
6. After the N-560 completes its normal power-up sequence, verify that the
sensor LED is brightly lit.
16
Performance Verification
7. Slowly move the sensor LED in proximity to the photo detector element
of the sensor (close the sensor slowly). Verify, as the LED approaches the
optical sensor, that the LED intensity decreases.
8. Open the sensor and notice that the LED intensity increases.
9. Repeat step 7 and the intensity will again decrease. This variation is an
indication that the microprocessor is in proper control of LED intensity.
10. Press the Power On/Off button to turn the N-560 off.
1. Sensor Port
2. Connect a DOC-10 pulse oximetry cable to the N-560 sensor port (1).
5. Press the Power On/Off button to turn the N-560 on and verify that the
N-560 is operating.
2. Press the Adjust Down button, while holding the Alarm Silence
button, until an alarm volume setting of “1” is displayed. Verify that the
alarm volume has decreased but is still audible.
3. Press the Adjust Up button, while holding the Alarm Silence button, to
increase the alarm volume setting to a maximum value of “10.” Verify
that the volume increases.
4. Press the Adjust Down button, while holding the Alarm Silence
button, until a comfortable audio level is obtained.
18
Performance Verification
2. Press the Adjust Down button and verify that the beeping pulse tone
decreases until it is no longer audible.
Introduction
Initial Setup
1. Sensor Port
1. With the N-560 turned off, connect the DOC-10 pulse oximetry cable to
the sensor port.
20
Performance Verification
2. Connect the SRC-MAX tester to the other end of the DOC-10 cable.
• be in SpO2 alarm
• display an %SpO2 of 75 (pass criteria is 73 to 77 %SpO2)
• display a pulse rate of 60 (pass criteria is 57 to 63 bpm)
• Pulse Amplitude indicator - display low level modulation (low
amplitude Pulse Amplitude indicator)
2. The N-560 bpm will increase to 200 and stabilize at 200 bpm. The test
pass criteria is 197 to 203 bpm.
• 75 %SpO2
• 200 bpm (pass criteria is 197 to 203 bpm)
• alarm
• Pulse Amplitude indicator - low level modulation
4. Press the SRC-MAX Pulse Rate select button. The SRC-MAX Pulse Rate
60 LED will light.
5. The N-560 pulse rate will decrease to 60 and stabilize at 60 bpm. The test
pass criteria is 57 to 63 bpm.
• 75 %SpO2
• 60 bpm (test pass criteria is 57 to 63 bpm)
• alarm
• Pulse Amplitude indicator - low level modulation
2. The N-560 will display three dashes until the SRC-MAX stabilizes at 90
%SpO . The test pass criteria is 88 to 92 %SpO2.
2
22
Performance Verification
5. The N-560 will display three dashes until the SRC-MAX stabilizes at 75
%SpO . The test pass criteria is 73 to 77 %SpO2.
2
2. The N-560 Pulse Amplitude indicator will fill (10 bars) with each pulse
beat.
4. Perform Test #1: BPM on page 21. The Pulse Amplitude indicator
should indicate high level modulation.
5. Perform Test #2: SpO2 on page 22. The Pulse Amplitude indicator
should indicate high level modulation.
• 75 %SpO2
• 60 bpm
• alarm
• Pulse Amplitude indicator - low level modulation
9. Perform Test #1: BPM on page 21. The Pulse Amplitude indicator
should indicate low level modulation.
24
Performance Verification
10. Perform Test #2: SpO2 on page 22. The Pulse Amplitude indicator
should indicate low level modulation.
1. Press the SRC-MAX Light Level selection button. The SRC-MAX Light
Level LED will light.
2. The N-560 Pulse Amplitude indicator will fill (10 bars) and stabilize
and illuminate 4 bars with each pulse beat.
4. Perform Test #1: BPM on page 21. The Pulse Amplitude indicator
should indicate low level modulation.
5. Perform Test #2: SpO2 on page 22. The Pulse Amplitude indicator
should indicate low level modulation.
6. Press the SRC-MAX Light Level selection button. The SRC-MAX Light
Level LED will light.
• 75 %SpO2
• 60 bpm
• alarm
• low level modulation
9. Perform Test #1: BPM on page 21. The Pulse Amplitude indicator
should indicate low level modulation.
10. Perform Test #2: SpO2 on page 22. The Pulse Amplitude indicator
should indicate low level modulation.
Safety Tests
The N-560 safety tests meet the standards of, and are performed in accordance
with, IEC 60601-1 (EN 60601-1, Amendment 1, Amendment 2) and UL
2601-1, for instruments classified as Class 1 and TYPE BF and ANSI/AAMI
Standard ES1.
Applicable tests for these standards are listed below. Technicians must be
familiar with the standards applicable to their institution and country. Test
equipment and its application must comply with the applicable standard.
• Ground Integrity, see page 92 for test values.
• Earth Leakage Current, see page 92 for test values.
• Enclosure Leakage Current, see page 92 for test values.
• Patient Applied Risk Current, see page 93 for test values.
• Patient Isolation Risk Current, see page 93 for test values.
26
Performance Verification
Note : Patient Applied Risk Current and Patient Isolation Risk Current. The
leakage test lead from the test equipment must be connected to the N-560
SpO2 Sensor Port using a male 9-pin “D” type connector that has all pins
shorted together.
28
Audible Alarm Settings and Service Functions
Introduction
Alarm Volume
1. Press and hold the Alarm Silence button until “VOL” is displayed in the
%SpO2 window.
N-560 Menu
Table 2 lists the menu selections available on the N-560. Items in BOLD are
the default settings.
30
Audible Alarm Settings and Service Functions
The menu items listed below are options that are available to the operator.
These options can be used to print or clear trend data, and to change the
configuration of data from the data port. Changes can be made in menu
options 1 through 5 while monitoring a patient. Menu items beyond 5 cannot
be accessed when a sensor is connected to the N-560. Unless changes are
saved as power-on default values, they will be lost when the N-560 is turned
off. When the N-560 is turned on again, it will begin to use the power-on
32
Audible Alarm Settings and Service Functions
default values that have been stored. Factory-set power-on default values are
listed in Table 4 on page 38.
2. Press the Adjust Up or Adjust Down button to select the menu item
number. Menu items 4, 5, 6, 7, 8, 11, 12, 13, and 14 have sub-options
within them that can be selected by first pressing the SatSeconds Alarm
Limit button, and then pressing the Adjust Up or Adjust Down button.
The sub-option being selected is displayed in the %SpO2 display window.
3. Once adjustments have been made within a menu item, the SatSeconds
Alarm Limit button can be used to initiate the current selection. To save
the current settings as power-on default values, refer to the procedure
outlined in Menu Item 9 (Save Current Values as Power-On Defaults) on
page 37.
displayed. For more information about trend printouts, refer to the N-560
Operator’s manual.
Use the Adjust Up or Adjust Down button to select the desired baud rate.
The baud rates will be displayed in the %SpO2 window as 24 (2400), 96
(9,600), or 192 (19,200). The protocol setting (menu item 5; EPP mode)
determines the baud rate that can be selected:
• Option 1 ASCII = 19,200
• Option 2 External equipment communications = 2,400, 9,600, 19,200
When the desired option is indicated in the %SpO2 display, press the
SatSeconds Alarm Limit button to initiate the current selection.
When the desired option is indicated in the %SpO2 display, press the
SatSeconds Alarm Limit button to initiate the current selection.
Note: Menu items greater than 5 cannot be accessed when a valid sensor is
connected to the N-560.
Service menu options can be accessed only when the sensor is disconnected
from the N-560. Only qualified service personnel should access these options.
Refer to Accessing Menu Items on page 33 for instructions on how to access
the menu options and make selections within them.
Pressing the SatSeconds Alarm Limit button again will cause the Day Hour
display to appear. See Figure 11.
36
Audible Alarm Settings and Service Functions
Press the SatSeconds Alarm Limit button to select Hour and the next press of
the SatSeconds Alarm Limit button will select the Minute Second display. See
Figure 12.
Press the SatSeconds Alarm Limit button to select Second and the next press
of the SatSeconds Alarm Limit button enters the selected date and time into
the N-560. The N-560 sounds three beeps to indicate that the date time is set.
Parameter Range
SpO2 high alarm limit low limit plus 1 to 100%
SpO2 low alarm limit 80% to high limit minus 1
SatSeconds Off, 10, 25, 50, 100 seconds
Pulse rate high alarm limit low limit plus 1 to 250 bpm
Pulse rate low alarm limit 30 bpm to high limit minus 1
Alarm volume 1 to 10
Alarm silence duration 30, 60, 90, 120 seconds
Parameter Range
Alarm silence restrictions None, sound reminder, do not allow alarms off
Blip volume 0 to 10
Language English
Serial port baud rate 2400, 9600, 19200
Serial port mode ASCII, external equipment communications
Event in-sensor trend format SpO2, SpO2 and pulse rate
RS-232 nurse call polarity Normally high, normally low
If an invalid tone is heard instead of the triple beep, the current settings were
not changed. An invalid tone is a single low-pitched tone.
Note: Current values will not be stored in memory as institutional defaults if
power is interrupted before exiting this menu option.
Note: When the operator changes an alarm limit to a value other than a
institutional default value, a decimal point is displayed to the right of the
parameter whose alarm limit was changed.
Factory
Parameter Range Default
Setting
%SpO2 Upper Alarm Lower Alarm Limit plus 1 to 100%
Limit 100%
%SpO2 Lower 20% to Upper Alarm Limit 85%
Alarm Limit minus 1
Pulse Rate Upper Lower Alarm Limit plus 1 to 170 bpm
Alarm Limit 250 bpm
38
Audible Alarm Settings and Service Functions
Factory
Parameter Range Default
Setting
Pulse Rate Lower 30 bpm to Upper Alarm Limit 40 bpm
Alarm Limit minus 1
Alarm Silence On or Off On
Reminder
Alarm Silence Off, 30, 60, 90, 120 seconds 60 seconds
Duration
Alarm Silence • None None
Restrictions
• Sound reminder
• Do not allow alarms off
Alarm Sound 1, 2, 3 2 (IAW
Selector IEC60601-1-8)
Alarm Volume 1 to 10 4
If Option “1” Off is selected, all visual display is turned off. Audio sound
cannot be set to off when Option “1” is selected. When any button is pressed,
all visual displays turn on.
40
Audible Alarm Settings and Service Functions
5. Simultaneously press SpO2 Alarm Limit button and Pulse Rate Alarm
Limit button for over three seconds. Menu Item 1 screen appears.
7. Press the SatSeconds Alarm Limit button to save all settings as the
institutional default settings.
8. The N-560 sounds three beeps, indicating that defaults are reset.
42
Troubleshooting
Troubleshooting
Introduction
This section explains how to troubleshoot the N-560 if problems arise. Tables
are supplied that list possible N-560 difficulties, along with probable causes,
and recommended actions to correct the difficulty.
Troubleshooting Guide
Problems with the N-560 are categorized in Table 5 on page 44. Refer to the
paragraph indicated for further troubleshooting instructions.
Note: Taking the recommended actions discussed in this section will correct
the majority of problems you may encounter. However, problems not covered
here can be resolved by calling Nellcor’s Technical Services or your local
Nellcor representative.
• No power-up on AC and/or DC
• Fails power-on self-test
• Powers down without apparent cause
2. Buttons Buttons on page 46
Power
Power problems are related to AC and/or DC. Table 6 lists recommended
actions to power problems.
44
Troubleshooting
Buttons
Table 7 lists symptoms of problems relating to nonresponsive buttons and
recommended actions. If the action requires replacement of a PCB, see
Disassembly Guide on page 53.
46
Troubleshooting
Display/Alarms
Table 8 lists symptoms of problems relating to non-functioning displays and
audible tones or alarms and recommended actions. If the action requires
replacement of a PCB or module, refer to Disassembly Guide on page 53.
Operational Performance
Table 9 lists symptoms of problems relating to operational performance (no
error codes displayed) and recommended actions. If the action requires
replacement of a PCB or module, refer to Disassembly Guide on page 53.
48
Troubleshooting
Data Port
Table 10 lists a symptom for data port problems and recommended actions. If
the action requires replacement of the User Interface PCB, refer to
Disassembly Guide on page 53.
Error Codes
When the N -560 detects a defective sensor connected to the N-560, the N-560
displays an error code of “SEN Err.” The sensor should be replaced and the
N-560 power should be cycled.
50
Troubleshooting
Error
Meaning
Code
Error
Meaning
Code
52
Disassembly Guide
Disassembly Guide
Introduction
Prior to Disassembly
1. Turn the N-560 off by pressing the Power On/Off button.
54
Disassembly Guide
3. Place the N-560 up-side down and remove the five screws holding the
cases together. See Figure 14.
5. Carefully separate the N-560’s top case from its bottom case, being
careful not to stress the wire harnesses between the cases. See Figure 15.
6. Carefully disconnect the two cables from the main PCB. See Figure 16
items 1 and 2.
56
Disassembly Guide
2. Perform the procedure Separating Top and Bottom Cases on page 54.
3. Disconnect the SpO2 cable connector from the NELL-3 PCB. See
Figure 17 item 1.
4. Disconnect the speaker wire connector, item 2, from the main PCB.
5. Disconnect the ribbon cable connector, item 3, from the front case PCB.
6. Remove the bracket screw connected to the main PCB. See Figure 18
item 1.
8. Remove the front panel assembly from the top case assembly.
9. Remove the bracket and screw connected to the front PCB. See Figure 19
item 1.
10. Remove the two screws holding the speaker to the front case. See
Figure 20 items 1 and 2.
58
Disassembly Guide
11. Carefully lift the front PCB out of the front case. See Figure 21.
NELL-3 Removal
1. Perform the procedure Prior to Disassembly on page 54.
2. Perform the procedure Separating Top and Bottom Cases on page 54.
3. Disconnect the front case SpO2 cable connector from the NELL-3 PCB
connector. See Figure 22 item 1.
4. Disconnect the main PCB cable connector from the NELL-3 PCB
connector, item 2.
5. Carefully lift the NELL-3 PCB out of the top case. See Figure 23.
2. Perform the procedure Separating Top and Bottom Cases on page 54.
60
Disassembly Guide
3. Disconnect the SpO2 cable connector from the NELL-3 PCB. See
Figure 17 item 1.
4. Disconnect the speaker wire connector, item 2, from the main PCB.
5. Disconnect the ribbon cable connector, item 3, from the front case PCB.
6. Remove the bracket screw connected to the main PCB. See Figure 18
item 1.
8. Remove the front panel assembly from the top case assembly.
9. Disconnect the NELL-3 to main PCB cable from the main PCB
connector. See Figure 26 item 1.
10. Disconnect the main PCB to front PCB cable from the main PCB
connector item 2.
11. Remove the three remaining screws holding the main PCB to the top case.
12. Carefully lift the main PCB out of the top case. See Figure 27.
62
Disassembly Guide
2. Perform the procedure Separating Top and Bottom Cases on page 54.
3. Unplug the wiring harness connector from the power supply, item 1.
5. Unscrew the four screws from the power supply. See Figure 29.
6. Lift the power supply out of the bottom case. See Figure 30.
Fan Removal
1. Perform the procedure Prior to Disassembly on page 54.
2. Perform the procedure Separating Top and Bottom Cases on page 54.
64
Disassembly Guide
3. Unwrap the two fan wires, Figure 31 item 1, from the wiring harness.
5. Remove the two screws holding the fan bracket to the bottom case. See
Figure 32.
6. Lift the fan and bracket out of the bottom case. See Figure 33.
7. Remove the 4 screws attaching the fan to the fan bracket. See Figure 34.
Battery Removal
1. Perform the procedure Prior to Disassembly on page 54.
2. Perform the procedure Separating Top and Bottom Cases on page 54.
3. Remove the four battery screws from the bottom of the bottom case. See
Figure 35.
66
Disassembly Guide
5. Lift the battery cover out of the bottom case. See Figure 38 item 1.
6. Lift the battery out of the case. See Figure 38, item 2.
2. Perform the procedure Separating Top and Bottom Cases on page 54.
68
Disassembly Guide
4. Remove the remaining nut and washer from the equipotential ground
terminal. See Figure 40.
5. Remove the equipotential ground terminal from the bottom case. See
Figure 41.
70
Spare Parts
Spare Parts
Introduction
The current version of this manual is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
Spare parts are shown in Table 12 and Figure 42.
72
Spare Parts
74
Packing for Shipment
Introduction
To ship the N-560, for any reason, follow the instructions in this section.
General Instructions
Pack the N-560 carefully. Failure to follow the instructions in this section may
result in loss or damage not covered by any applicable Nellcor warranty. If the
original shipping carton is not available, use another suitable carton.
Prior to shipping the N-560, contact your supplier or local Nellcor office
(Technical Services Department) for a RGA number. Mark the shipping
carton and any shipping documents with the RGA number. Return the N-560
by any method that provides proof of delivery.
2. Place the N-560 into the shipping carton and seal the carton with packing
tape.
3. Label the carton with the shipping address, return address, and RGA
number.
76
Packing for Shipment
3. Fill the bottom of the carton with at least two inches of packing material.
4. Place the bagged N-560 on the layer of packing material and fill the box
completely with packing material.
6. Label the carton with the shipping address, return address, and RGA
number.
78
Specifications
Specifications
Performance
P
Measurement Range
SpO2 1% to 100%
Pulse Rate 0 and 20 beats per minute (bpm) to 250 bpm
Perfusion Range 0.03% to 20%
Accuracy and Interference Tolerance
Saturation
Adults1 70 to 100% ±2 digits
Neonate 70 to 100% ±3 digits
Low Perfusion2 70 to 100% ±2 digits
Pulse Rate
- -
vary by sensor type. Refer to the Sensor Accuracy Grid. The Sensory
Accuracy Grid is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
2
Specification applies to N-560 performance. Reading accuracy in
the presence of low perfusion (detected IR pulse modulation
amplitude <1.5%) was validated using signals supplied by a
patient simulator. SpO2 and pusle rate values were varied across
the monitoring range including weak signal conditions and
compared to the known true saturation and pulse rate of the input
signals.
80
Specifications
Electrical
Instrument
Power Requirements 100 to 240 volts AC, 25 to 39 volt/amps to be
compliant with IEC 60601-1 sub-clause 10.2.2
Fuses qty 2, 2 A, 250 volts, slow-blow, IEC (5 x 20 mm)
Battery
The battery provides at least eight hours of battery life when new and fully charged
with no alarms, no serial data, while using a pulse simulator set for 60 bpm, high
light and low modulation.
Type Nickel metal hydride
Voltage 9.6 Volts DC, 3.8 AH
Recharge 6 hours
Shelf Life 2 months, new, fully charged battery
After 2 months storage the N-560 will run for 50% of stated
battery life
82
Specifications
Battery
Complies 91/157/EEC
With
Sensors
Electrical/Optical Nellcor Pulse oximetry sensors contain light emitting diodes
Specifications (LEDs) that emit red (~660 nm) and infrared (~900 nm) light,
with a total optical output power of less than 15 mW. This
information of sensor wavelength range can be especially
useful to clinicians, for example, those performing
photodynamic therapy.
Environmental Conditions
Operation
Temperature 50 ºF to 113 ºF (10 ºC to 45 ºC)
Altitude/Barometric Pressure -390 m to 3,012 m
(-1,280 ft. to 9,882 ft.)
70 kPa to 106 kPa
(20.6 in. Hg to 31.3 in. Hg)
Relative Humidity 15% to 95% non-condensing to be compliant
with IEC 60601-1, sub-clause 44.5
84
Specifications
Physical Characteristics
Compliance
Manufacturer’s Declaration
This product complies with EN/IEC60601-1-2:2001.
WARNING: The use of accessories, sensors, and cables other than those
specified may result in increased emission and/or create invalid readings
of the N-560.
86
Specifications
88
Specifications
seconds
Power frequency 3 A/m Complies Power frequency
(50/60 Hz) magnetic fields should
magnetic field be at levels
characteristic of a
IEC 61000-4-8 typical location in a
typical commercial or
hospital environment.
Note: U
T is the AC mains voltage prior to application of the test level.
90
Specifications
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: At 80 MHz to 800 MHz, the separation distance for the higher
frequency range applies.
Safety Tests
Ground Integrity
100 milliohms or less
92
Specifications
94
Data Port Interface P rotocol
Introduction
The data port, located at the rear of the N-560, provides interfacing
capabilities for:
• printing N-560 data
• displaying N-560 data on a computer
Protocol
The data port supports two communication protocols:
• Option 1 = real-time ASCII for printouts
• Option 2 = enables communication with the external equipment.
Refer to the external equipment documentation for the interface
procedures.
Menu item 5 allows the user to choose between the communication protocols.
To access menu item 5:
1. Disconnect the sensor cable.
2. Press both the SpO2 Alarm Limit and Pulse Rate Alarm Limit buttons
simultaneously for three seconds.
Baud Rate
2. Press both the SpO2 Alarm Limit and Pulse Rate Alarm Limit buttons
simultaneously for 3 seconds.
96
Data Port Interface P rotocol
Data is transmitted in the RS-232 and RS-422 formats. RS-232 data can be
transmitted a maximum of 25 feet (7.62 meters). The RS-422 data can be
transmitted a maximum of 4,000 feet (1,219.2 meters). The pin outs for the
data port are listed in Table 19.
TXD+ and TXD- are the differential transmit data pair. RXD+ and RXD- are
the differential receive pair.
The pin layouts are illustrated in Figure 44. The conductive shell is used as
earth ground. An AMP connector is used to connect to the data port. Use
AMP connector (AMP P/N 747538-1), ferrule (AMP P/N 1-747579-2) and
compatible pins (AMP P/N 66570-2). The serial cable must have a braided
shield providing 100% coverage, such as Beldon cable (Beldon P/N 9616) or
equivalent. Connectors at both ends of the serial cable must have the shield
terminated to the full 360 degrees of the connector's metal shell.
CAUTION: Do not create sharp bends in the cable; this may tear or
break the shield.
98
Data Port Interface P rotocol
Real-Time Printout
When a real-time printout is being transmitted, a new line of data is printed
every two seconds. Every 25th line will be a Column Heading line. A Column
Heading line will also be printed any time a value in the Column Heading line
is changed. A real-time printout is shown in Figure 45.
Note: Printouts are available only if the N-560 is running on AC power.
Column Heading
100
Data Port Interface P rotocol
Printout Source
Data in the highlighted box above represents the source of the printout, in this
case, the N-560.
The next data field tells the user the software level (Version 1.0.0.0) and a
software verification number (CRC XXXX). Neither of these numbers should
change during normal operation. The numbers will change if the N-560 is
serviced and receives a software upgrade.
Alarm Limits
The last data field in the top line indicates the upper and the lower alarm
limits for %SpO2, the pulse rate (PR), and the SatSeconds alarm limit. In the
example above, the low alarm limit for SpO2 is 85% and the upper alarm limit
is 100%. Pulse rate alarm limits are 40 bpm (lower) and 170 bpm (upper). The
SatSeconds alarm limit is set to Off (0 sec.).
Column Headings
Actual column headings are in the third row of the Column Heading line.
Patient data, from left to right, are the time that the chart was printed, the
current %SpO2 value being measured, the current pulse rate in beats per
minute (bpm), the current Pulse Amplitude (PA), and the operating status of
the N-560.
Time Tag
Time Tag represents a real-time clock in: Day, Month, Year, and 24-hour
clock. The clock is maintained by either AC or battery power.
Patient Data
Patient data and the operating status of the N-560 are highlighted in the
patient data display. Parameter values are displayed directly beneath the
heading for each parameter. In this example, the %SpO2 is 100 and the pulse
rate (PR) is 190 beats per minute. The asterisk (*) next to the 190 indicates
that 190 beats per minute is outside of the alarm limits, indicated at the
far-right end in the top row, for pulse rate. If no data for a parameter is
available, three dashes (- - -) will be displayed in the printout.
The number under PA is an indication of pulse amplitude. The number can
range from 0 to 254 and will typically range around 45. There are no alarm
parameters for this value. It can be used for trending information and indicates
a change in pulse volume, pulse strength, or circulation.
Operating Status
102
Data Port Interface P rotocol
The Status column indicates alarm conditions and operating status of the
N-560. The PH in this example indicates a Pulse Rate Upper alarm. The
Status column can have as many as four codes displayed in one line of data.
The status codes are listed in Table 20.
Code Meaning
AO Alarm Off
AS Alarm Silence
BU Battery in Use
LB Low Battery
LM Loss of Pulse with Interference
LP Loss of Pulse
MO Interference
PH Pulse Rate Upper Limit Alarm
PL Pulse Rate Lower Limit Alarm
PS Pulse Search
SD Sensor Disconnect
SH Saturation Upper Limit Alarm
SL Saturation Lower Limit Alarm
--- No Data Available
* Alarm Parameter Being Violated
104
Technical Discussion
Technical Discussion
Oximetry Overview
The N-560 Pulse Oximeter measures functional oxygen saturation by
measuring the light absorption of tissue, bone, and blood during the pulsatile
cycle. Red and infrared sensor LED’s are used as light sources. A photodiode,
acting as a photodetector, senses each received wavelength of light; namely,
incident light less the amount absorbed by tissue, for both the red and infrared
channels. It also feeds corresponding electrical signals from the sensor to the
N-560. These signals are then processed by the N-560 software engine to
produce SpO2 and pulse rate data. Bone, tissue, pigmentation, and venous
vessels normally absorb a constant amount of light over time. The arteriolar
bed normally pulsates and absorbs variable amounts of light during the
pulsations. The ratio of light absorbed is translated into a measurement of
functional oxygen saturation (SpO2).
Because a measurement of SpO2 is dependent upon light from the sensor,
excessive ambient light can interfere with this measurement.
Specific information about ambient conditions, sensor application, and patient
conditions is contained throughout this manual.
Pulse oximetry is based on two principles: that oxyhemoglobin and
deoxyhemoglobin differ in their absorption of red and infrared light (i.e.,
spectrophotometry), and that the volume of arterial blood in tissue (and hence,
light absorption by that blood) changes during the pulse (i.e.,
plethysmography). A pulse oximeter determines SpO2 by passing red and
infrared light into an arteriolar bed and measuring changes in light absorption
during the pulsatile cycle. Red and infrared low-voltage light-emitting diodes
(LED) in the oximetry sensor serve as light sources; a photodiode serves as
the photo detector.
Because oxyhemoglobin and deoxyhemoglobin differ in light absorption, the
amount of red and infrared light absorbed by blood is related to hemoglobin
oxygen saturation. To identify the oxygen saturation of arterial hemoglobin,
the N-560 uses the pulsatile nature of arterial flow. During systole, a new
pulse of arterial blood enters the vascular bed, and blood volume and light
absorption increase. During diastole, blood volume and light absorption reach
their lowest point. The N-560 bases its SpO2 measurements on the difference
between maximum and minimum absorption (i.e., measurements at systole
and diastole). By doing so, it focuses on light absorption by pulsatile arterial
blood, eliminating the effects of nonpulsatile absorbers such as tissue, bone,
and venous blood.
Automatic Calibration
Because light absorption by hemoglobin is wavelength dependent and
because the mean wavelength of LEDs varies, an oximeter must know the
mean wavelength of the sensor's red LED to accurately measure SpO2. During
manufacturing, the mean wavelength of the red LED is encoded in the sensor.
During monitoring, the N-560's software uses the coefficients that are
appropriate for the wavelength of that individual sensor's red LED; these
coefficients are then used to determine SpO2. The N-560’s software obtains
coefficients when the N-560 is turned on, periodically thereafter, and each
time a new sensor is connected.
Additionally, to compensate for differences in tissue thickness, the light
intensity of the sensor's LEDs is adjusted automatically.
106
Technical Discussion
Signal Processing
The N-560 takes advantage of increased microprocessing power with
advanced mathematical algorithms. OXIMAX advanced signal processing
allows the N-560 to provide accurate saturation and pulse rate values, even
during difficult monitoring conditions of low perfusion and signal
interference.
Interference is a term that was formally introduced in ISO9919.2005,
“Medical electrical equipment — Particular requirements for the basic safety
and essential performance of pulse oximeter equipment for medical use”.
Interference may arise from a number of sources, alone or in combination,
and the term refers to the effects on the incoming red and IR signals caused by
physiological conditions, medical procedures, or external agents. The effect
108
Technical Discussion
OXIMAX Technology
NELL-3 Interface
The NELL-3 interfaces with the N-560 through two connections. Figure 47
illustrates the interconnection, and Table 21 and Table 22 identify the signals.
110
Technical Discussion
Circuit Analysis
The following paragraphs discuss the operation of each of the printed circuit
boards within the N-560 pulse oximeter. (Refer to the appropriate schematic
diagram at the end of this section, as necessary).
AC Input
The N-560 allows the user to connect the N-560 to AC power ranging from
100 volts AC to 240 volts AC.
AC power enters directly the N-560’s Power Supply. A 2-amp fuse protects
both the “Line” and “Neutral” lines. These user-non accessible fuses are
located in a fuse drawer, which is part of the Power Supply located in the
N-560.
112
Technical Discussion
Battery
A nickel metal hydride battery is used in the N-560. It is rated at 9.6 volts DC,
3.8 amp hours. When new and fully charged, the battery will operate the
N-560 for eight hours. A new battery will last 15 minutes from the time the
low battery alarm is declared until the N-560 is shut down due to battery
depletion.
The battery can withstand 400 charge/discharge cycles. Recharging the
battery to full capacity takes 6 hours.
Changing from AC to battery power does not interrupt the normal monitoring
operation of the N-560. When the N-560 is running on battery power, the data
port will be turned off.
Main PCB
The Main PCB is the heart of the N-560. All functions except the unregulated
DC power supply, display, and keypad reside on the main PCB. The
following text covers the key circuits of the main PCB.
Controlling Hardware
There is one microprocessor on the Main PCB, a Microchip Technology
PIC17C756A.
CPU
The PIC17C756A is the main controller of the N-560. The PIC17C756A
controls the front panel display, data storage, N-560 status, sound generation,
monitor, and controls the N-560's power. The PIC17C756A also controls data
port communication and communicates with the NELL-3 SpO2 Module.
Battery voltage is checked periodically by the processor. If the processor
determines that the battery voltage is below 8.9 ± 0.2 volts DC, a low battery
alarm is declared by the CPU. If battery voltage on the Main PCB is measured
below 8.2 ± 0.2 volts DC, the N-560 will automatically turn off. The user will
be unable to begin monitoring a patient if the battery voltage remains below
this point. If either event occurs, plug the N-560 into an AC source for 6 hours
to allow the battery to fully charge.
The PIC17C756A also controls a set of dry contacts provided by a relay on
the Main PCB. The relay will function normally on AC power or on internal
battery power.
When the CPU sends a tone request, three items are used to determine the
tone that is sent to the speaker. First, pulse tones change with the %SpO2
value being measured. The pulse beep tone will rise and fall with the
measured %SpO2 value. Second, three levels of alarms, each with its own
tone, can occur: high, medium, and low priority. Third, the volume of the
alarm is user adjustable. Alarm volume can be adjusted from level 1 to level
10, with level 10 being the highest volume.
The PIC17C756 controls the display driver and the buttons. By pressing any
of seven keys on the keypad the operator can access various functions of the
N-560. The PIC17C756A will recognize the keystroke and change the display
as required. Any changes made by the operator (such as: alarm limits, pulse
114
Technical Discussion
beep volume) are used by the N-560 until it is turned off. Default values will
be restored when the N-560 is turned back on.
Patient Data is stored by the N-560 and can be downloaded to a printer
through the data port provided on the back of the N-560. An in-depth
discussion of the data port is covered in the on
Data Port Interface Protocol
page 95.
Charging Circuits
The power supply will charge the battery any time the N-560 is connected to
AC power. The voltage applied to the battery is 14 ±0.15 volts DC.
Visual patient data and N-560 status are provided by the Front Panel Display.
At power up, all indicators are illuminated to allow verification of their proper
operation.
There are two sets of three, 7-segment displays. One set displays %SpO2 and
the other displays pulse rate. A decimal point immediately to the right of
either display indicates that an alarm limit for that parameter is no longer set
at the power-on default value.
Between the two sets of 7-segment displays is a 10-segment Pulse
Amplitude indicator. The Pulse Amplitude indicator illuminates with each
pulse beat. The number of segments illuminated indicates the relative signal
strength of the pulse beat. A tone will accompany each pulse beat. The sound
of the tone changes pitch with the %SpO2 level being measured.
Below the 10-segment Pulse Amplitude indicator is a 8-segment display.
The 8-segment display illuminates with SatSecond values. The first segment
of the 8-segment display indicates the SatSecond alarm on/off. The number of
segments illuminated indicates the relative values of the SatSecond.
Seven LEDs and icons are also located on the Front Panel Display part. An
LED illuminated next to an icon indicates a function that is active. Functions
indicated by the LEDs are AC/Battery Charging, Low Battery, Alarm Silence,
Interference, Pulse Search, Sensor Off, and Sensor Message.
Button Interface
The Button Interface is mounted as part of Front Panel PCB. Seven keys
allow the operator to access different functions of the N-560.
These keys allow the user to select and adjust the alarm limits, cycle power to
the N-560, and silence the alarm. Alarm volume and alarm silence duration
can also be adjusted via the keypad. A number of other functions can be
accessed by pressing the SatSeconds Alarm Limit button and Pulse Rate
Alarm Limit button simultaneously and then selecting the desired option with
the Adjust Up or Adjust Down button. These functions are discussed in
greater detail in Audible Alarm Settings and Service Functions on page 29.
116
Technical Discussion
TP28 TP31
T POINT M1 T POINT M1
04-05-18 CR11
2K
VIN BAT+
1
CD1 CR16 0_3216 CC2
CC1 MBRS340 22uF_35VEL CD3
0.1uF 2 1 S1B
BAT-
CD2
S1B TP59 T POINT M1 CL1
+5VC Alternative
100uH
CU2 1 2
8 1 CR15 CR15-1
1
VIN VOUT CQ1 0.33_2W_F 0.33_2W_F_SMD
7 2 CR10 B772 CD4
GND GND 100 CC10
Alternative
3
6 3 1 MBRS340 330uF/16V
GND GND TP58 Alternative
CR12
T POINT M1
5 4 100K_F
NC NC
LM78L05A/SOIC
CC3 1 TP29
3
T POINT M1 CR13
22uF_35VEL 2 CQ2 100K_F
CC4 CC5 2N3904/TO-92
47uF_16VEL 0.1uF
1
LED1 LED2
TP30 CR9
1 200 Battery Absent LOW LOW
T POINT M1
+5VC +5VC Fast charge pending
LOW FLASH
CC11 Discharge-before-charge in
Open progress
CU1 +5VC
CR1 CR16 1 16
30K Open 2 _DCMD _INH 15 Fast charging LOW HIGH
3 DSEL DIS 14
CR17 0 4 VSEL MOD 13 CR14
5 TM1 VCC 12 CC6 66.5k_F Fast charge complete HIGH LOW
6 TM2 VSS 11 0.1uF
7 TCO LED2 10
THERM TS LED1
8 9
CR4 BAT SNS
510k bq2004 CR8 LD2
CC7 CC8
2K LED2
0.1uF 0.1uF
CR2 1 2
2.4K
CR6 1 2
+5VF2
+5VF2 TP57
TP56 T POINT M2
+5V NELL-3 1
+5VF2
T POINT M2 R12 +12VF
1
C8 3.9K J5 C15 C58
120pF 1uF 120pF
2 1
R9 TP22
TP20 4 3
U7 T POINT M2
T POINT M2 1 8
6 5
N.C Vcc C14 8 7
R8 2 7 390
ANODE ENABLE 1uF C16 10 9
3 6
TX2 4 CATHODE Vout 5 TX2F 47pF 12 11
1
1
N.C GND 14 13
390
6N137 CON14A
+5VF2
+5V RX2F
TX2F
C9
120pF
C49 C50
NELL-3 Connector
R10 TP23
TP21 U8 10pF 10pF
T POINT M2 T POINT M2
8 1
390 7 Vcc N.C 2 R11
6 ENABLE ANODE 3
RX2 5 Vout CATHODE 4 RX2F
1
1
GND N.C 390
6N137
+12VF +5VF2
U9
U21 8 1
1 22 VIN VOUT
2 GND NA 21 7 2
3 NA NA 20 GND GND
+5V VIN NA
4 19 6 3
5 NA NA 18 C1 GND GND C2
6 NA NA 17 5 4 C62
7 NA NA 16 47pF NC NC 47uF_16VEL 47pF
NA NA C63
8 15
9 NA NA 14 LM78L05A/SOIC
10 0V +V 13
11 NA NA 12
-V NA
47uF_16VEL
C64 NTV0512
1uF_3216
Isolation DC/DC
LP1
800V Arrestor
+5V +5VF1
R15 U14
C17 U11 4 1
120pF C+ C+ JDP1
1 8 C21 C22
N.C Vcc 390
2 7 1uF 5 3 1uF
R14 3 ANODE ENABLE 6 C- C- 1
TX1 4 CATHODE Vout 5 TX1F 12 13 U13 1 9
1
1
390 N.C GND RX1F O1 R1 1 8 2 9
TP24 TP26 9 8 2 L1I L1O 7 2 10
6N137 O2 R2 L2I L2O 10
T POINT M3 T POINT M3 3 6 3
1
11 14 4 L3I L3O 5 3 11
TX1F I2 T1 L4I L4O 4 11
+5VF1 TP15 4
+5V 10 7 ZJYS51R5-M4PA(T) 12
TX1F_N I1 T2 T POINT M3 5 12
6 2 TP54 TP55 5 13
C118 V- V+ T POINT M3 6 13
C67 U27 R28 T POINT M3 6
120pF 10pF MAX232S 14
1 8 7 14
390 C119
N.O.
1
2 N.C Vcc 7 C23 C24 7 15
R29 10pF
1
3 ANODE ENABLE 6 8 15
Nurse Call 2 CATHODE Vout TX1F_N 1uF 1uF N.C. 8
1
4 5 C25
1
1
10pF C68
Serial
0.1uF
TP53
T POINT M3
Communication
+5V +5VF1
R30 R31 Connector
4.7_2012 4.7_2012
C70
C18 1 TP14
R16 U12 T POINT M3 C80
120pF
390 8 1 2.2nF
7 Vcc N.C 2 +5VF1
6 ENABLE ANODE 3 R17 0.1uF
RX1 5 Vout CATHODE 4 RX1F U30 MAX489E/SO
1
1
GND N.C 390
1 14 TP52
TP25 TP27 NC VCC T POINT M3
6N137
T POINT M3 T POINT M3 2 13
RX1F RSOUT NC U32
3 12 1 8
1
_RXENAB RXD+ 2 L1I L1O 7
4 11 3 L2I L2O 6
ITX422EN TXENAB RXD- 4 L3I L3O 5
+5V +5VF1 5 10 L4I L4O
TX1F TXIN TXD-
6 9 ZJYS51R5-M4PA(T)
C121 GND1 TXD+
C81
R37
1
120pF U31 7 8
1 8 10pF GND2 NC
390 C122
2 N.C Vcc 7 TP51
R38 10pF
3 ANODE ENABLE 6 T POINT M3
TX422EN CATHODE Vout ITX422EN
1
4 5
1
C20
U19
1 22
2 GND NA 21 0.1uF
NA NA R39
3 20 +5V R1 R2
+5V VIN NA
4 19 470
5 NA NA 18 4.7_2012 4.7_2012
6 NA NA 17
7 NA NA 16 C38
8 NA NA 15
9 NA NA 14
10 0V +V 13 2.2nF
11 NA NA 12 +5VF1
-V NA
U24
NTV0505 1 8
2 Anode N.C 7 N.O.
3 Cathod N.C 6
4 Anode N.O 5 N.C.
Nurse Call 1 Cathod N.O COMMON
1
AQW614
TP48 Relay
T POINT M3 Figure 51: Main PCB Isolation
for Interface Schematic Diagram
N-560 121
Technical Discussion
PRR3
5.6K_F
+5VB +5V
TP9
4
T POINT M4 PRU1 LM2675-ADJ PRR2 TP12 TP13
1.5K_F 1 T POINT M4 1 T POINT M4
FB
1 PL2
7 CB PRC4 PRL1 SB3040
INPUT_V VIN PRD7
ON/_OFF
10nF BDS8040D-330M
1
8 1 2
GND
VSW
S1B PRC2
PRR5 PRC1 PRC3 PRC6 47uF_16VEL
6
2.7k_3216 47uF_35VEL 0.47uF_3216 PRD1 0.1uF
(1/4w) MBRS340 PRC5
100uF_16V_
TA_lowESR
PRD2
Power IN
1 2
VIN
TP10 PL3 SB3040
S1B T POINT M4 J1
PRD3 S1B
PQ1 2 1 1 SMPS Input
2
TIP42C/TO-220 C E
1
3
4
B CON4
J8
1
2 Battery Input
F1 3
4
50T 250V
PRR4 PRD5 C 5
2
T2AL
1 2 B PQ2 PRD6 PRD4 S1B CON5
POWER ON
2N3904/TO-92 2 1
220_2012 S1B
S1B E THERM
BAT-
1
BAT+
PRR1
680_2012
TP17 +5V
C32 T POINT M5
0.47uF 47uF_16VEL 47uF_16VEL
RG0 C33 C34 C35 C36 C37
1
R20 0.1uF
5.1K
+5V 47uF_16VEL 47uF_16VEL
U16
TP18
1 VCC GND1
6 C82 120pF
R18 C30 D5 T POINT M5
4
10K 0.1uF B140 R21 J7
1 2 3 U17A 1K 2 5 S1
1
+ 1 4 VI2 VO1 8 1
2 VI4 VO2 2
S2
1
C29 R19 C31 - LM324/SO CON2
RB2 R25 820pF 100K_F 22nF TP16 7 NC GND2
3 TP19
11
5K_POT T POINT M5 T POINT M5
TDA7052A
Connector
J3
SEG a 1 A1 B1
2 SEG b
SEG c 3 A2 B2
4 SEG d
SEG e 5 A3 B3
6 SEG f
SEG g 7 A4 B4
8 DP
C83 C85 C87 C89 C91 DIGIT 1 9 A5 B5
10 DIGIT 2 C92 C93 C94 C95 C96 C97 C98 C99 C100 C101
10pF 10pF
C86
10pF 10pF 10pF
DIGIT 3 11 12 DIGIT 4 10pF 10pF 10pF 10pF 10pF 10pF 10pF 10pF 10pF 10pF
C84 C88 C90 13 A6 B6 14
10pF 10pF 10pF 10pF DIGIT 5 15 A7 B7 16 DIGIT 6
DIGIT 7 17 A8 B8 18 DIGIT 8
RF0 19 A9 B9 20 RF1
10pF 10pF 10pF 10pF RF2 21 A10 B10 22 RF3
C103 C105 C107 C109 RE0 23 A11 B11 24 RE1
10pF 10pF 10pF 10pF 10pF RE2 25 A12 B12 26 RC3 10pF 10pF 10pF 10pF 10pF 10pF
C102 C104 C106 C108 C110 RC4 27 A13 B13 28 RC5 C111 C112 C113 C114 C115 C116
RC6 29 A14 B14 30 RC7
RB0 31 A15 B15 32 RB1
RB4 33 A16 B16 34 RB5
RB3 35 A17 B17 36
37 A18 B18 38
RC2 39 A19 B19 40
41 A20 B20 42
43 A21 B21 44 RE3
A22 B22
DIN 44_AB-H
C117 +5V
C28 anteld
10pF
J6 MR1
47uF_35VEL 4.7K
Alarm SW
Alarm SW
Power OUT
MC1
+5VB 1uF
PHONEJACK STEREO SW
Power IN
Foot SW/Antenna Phone Jack
VIN Front BD Connection Figure 53: MainPCB Interface and
Speaker Control Schematic Diagram
N-560 125
Technical Discussion
+5VA +5VA
+5V
1
C71 C72 C73 C74 C75 C76 C77 C78
TP4 D4
TP5 B140 2.2nF 2.2nF 2.2nF 2.2nF 10nF 10nF 10nF 10nF
T POINT M6
T POINT M6 +5VA
U3
8 1
2
Power OUT VIN VOUT
1
7 2 D3
49
37
20
GND GND
2
B140 U1
C4 6 3 C5 51
GND GND OSC2
VDD
VDD
VDD
VDD
1uF_3216 1uF
5 4 45
NC NC RA2/SS 46 SCL TP11 +5V
RA1 RA3/SDI/SDA 43 SDA T POINT M6
LM78L05A/SOIC RA4/RX1/DT1 42 RX1
RA5/TX1/CK1 TX1 Q18
59 2N2222A
RB0
1
RB0/CAP1 58 R24
TP6 RB1/CAP2 54
RB1 100 +5VB
+5V T POINT M6 RB2/PWM1 57 RB2
RB3/PWM2 RB3
2
1
56 C61
RB4
1
RB4/TCLK12 55 J4 47uF_16VEL
U4 L26 RB5/TCLK3 47
RB5 CON2
1 8 R6 R7 C79 0.56uH_2012 RB6/SCK 48 C26
A0 VCC 0.1uF RB7/SDO Nurse Call 1 0.1uF
2 7 2K 2K 3
A1 WP RC0/AD0 67 RC0 U15
3 6 RC1/AD1 66 RC1 12 19
A2 SCL RC2/AD2 65 RC2 RD0 ID0 VDD
4 5 C7 RC3/AD3 64 RC3 +5V 11 27
VSS SDA 47pF RC4/AD4 63 RC4 RD1 ID1 SEG a SEG a
24LC256/SOIC +5V XTAL1 RC5/AD5 62 RC5 13 3
32MHz 16 RC6/AD6 61 RC6 RD2 ID2 SEG b SEG b
17 MCLR/Vpp RC7/AD7 RC7 R3 14 1
50 TEST 11 RD3 ID3 SEG c SEG c
OSC1 RD0/AD8 10 RD0 10K 10 25
C6 60 RD1/AD9 9 RD1 1 RD4 ID4 (!SHUTDOWN) SEG d SEG d
U5 47pF 44 RA0/INT RD2/AD10 8 RD2 TP8 T POINT M6 6 2
RA1 RA1/TOCKI RD3/AD11 RD3 RD5 ID5 (!DECODE) SEG e SEG e
1
1 8 R40 R41 7
A0 VCC 29 RD4/AD12 6 RD4 RT1 5 24
2 7 2K 2K +5VA 30 AVss RD5/AD13 5 RD5 10k_THER RD6 ID6 (HEXA/!CODE B) SEG f SEG f
A1 WP AVDD RD6/AD14 4 RD6 7 26
RD7/AD15 RD7 RD7 ID7 (DATA COMING) SEG g SEG g
t
3 6
A2 SCL 12 4
4 5 RE0/ALE 13 RE0 DP DP
RE1
2
VSS SDA RE1/OE 14
24LC256/SOIC +5V RE2/WR 15 RE2
RE3/CAP4 RE3 8 15
28
INPUT_V RC0 !WRITE DIGIT 1 DIGIT 1
RF0/AN4 27 RF0 R22 9 16
RF1/AN5 26 RF1 22.1K_F RC1 MODE DIGIT 2 DIGIT 2
RF2/AN6 25 RF2 23
U6 RF3/AN7 24 RF3 1 TP7 DIGIT 3 DIGIT 3
1 8 R42 R43 RF4/AN8 23 T POINT M6 20
A0 VCC RF5/AN9 22 DIGIT 4 DIGIT 4
2 7 2K 2K RF6/AN10 21 TX422EN R23 17
A1 WP RF7/AN11 Alarm SW 6.81K_F DIGIT 5 DIGIT 5
3 6 34 22
A2 SCL RG0/AN0 33 RG0 DIGIT 6 DIGIT 6
4 5 RG1/AN1 32 Nurse Call 2 21
VSS SDA RG2/AN2/VREF- 31 DIGIT 7 DIGIT 7
24LC256/SOIC +5V RG3/AN3/VREF+ 38 18
RG4/CAP3 39 anteld DIGIT 8 DIGIT 8
RG5/PWM3 41 POWER ON
RG6/RX2/DT2 40 RX2
Vss
Vss
Vss
Vss
28
RG7/TX2/CK2 TX2 VSS
PIC17C756/PLCC ICM7228AIPI
U18
68
53
36
19
1 8 R44 R45 TP1
A0 VCC 2K
2 7
2K T POINT M6 Single ICM7228 Driver
A1 WP
3 6 Main Control Unit
A2 SCL SCL
1
4 5
VSS SDA SDA
1
24LC256/SOIC
TP2
T POINT M6
Serial EEPROM
TP3
T POINT M6 +5V
1
XTAL2 U22
32.768kHz 1 8 R4 R5
X1 Vcc 2k 2k
2 7
X2 SQW/OUT
3 6
VBAT SCL
4 5
GND SDA
1
BT1 DS1307
3V LITHIUM
2
N-560 +
C36 C33 44
2
J7
J3 2
MAIN_BD_XB 1
2004. 05. 21
TP19
C82 TP18 C61 1 C89 C87 C85 C83 C110 C109 C102 C117 43 TP12
U16 C26 C28
C103
TP10
C37 C90 C88 C86 C84 TP3
CUTTING AREA
C91
C116
C108
C107
C106
C105
C104
C112
C113
XTAL2
C111
TP17 C35 R25 C97 C96 C95 C94 C93 C92 C114 C115
EA
C101
U15
C98
C100
C99
TP16 C34 C74
AR
U21
R18
U1
BC7
BC1
G
R21
IN
C64
TT
BC8
C30 C29
U17
CU
R20
D5
C32
C31 U5
14 13 U22
U18 U4
R19
C63
CUTTING AREA
C16 BC2 TP2
C79
TP6
C71 BC4 R5 R4
C7
C58 BC5
D4
BC3
BC6
TP13 U6 BC10
TP1
U9 C5
C15 TP57 BC9
R43
R42
R45
R44
U3 R7 R6
R41
R40
J5 C50
TP25
TP49
C1 C67 TP24 R14
LP1 RT1 TP44 R29 C17 R16 R38 C81
CC10
C62
U12
TP56 D3 L26
R11
C2 R9 TP31
TP5 C73
TP23 U31
U8
U7
TP26 U11
R17
R37
CUTTING C77 R48 TP50
2 1 CUTTING AREA AREA
TP45 TP14
C68
R30
R15 C121
C25
C22
C18
C123
R24
C120
TP27
CL1 TP7 TP11
Q18 C122
C119 C118 C23 C21 C20
TP21 PRD1 PQ1 C75 R23 R22 R1
C19
PRR4 U19
R10 CD4 PRD5 F1
C24
C8 R8 TP20 PRR2
CD3 R2 C80
TP59 PRC4 R31 TP46
CR10
CR11
C38
CQ1
U14 TP15 C70
CR12
PRR1 U30
CR3
TP28
PRU1 PRC3
PRD4
E
TP9
TP53
PRD6
PRL1
CUTTING AREA
TP51 TP52
C
CR14 CR2 TP58
PRR5
PQ2
CR16
CR13
CC5 PRC6
CC8 CR17
PRD2
CU1 TP48
BT1
CC7
TP55
BC11
PRD7
TP30
CD2 15 9
E
1
FRF0
FRF1
1
FDIGIT 1 FDIGIT 2 FDIGIT 3
TP41
T POINT F1
1
470
2 2 2 2 2 2
2N2907A 2N2907A 2N2907A 2N2907A 2N2907A 2N2907A
3
14
13
12
10
11
FU2
9
DIG1 AN(G)
DIG2 AN(G)
DIG3 AN(G)
8 4
DIG1 AN(R)
DIG2 AN(R)
DIG3 AN(R)
FSEG a A DP FDP
7 6
FSEG b B G FSEG g
3 5
FSEG c C F FSEG f
2 1
FSEG d D E FSEG e
BI-FND
TP42
BPM Display Digits T POINT F1
FRF2
1
FRF3
1
FDIGIT 4 FDIGIT 5 FDIGIT 6
TP43
T POINT F1
1
2 2 2 2 2 2
2N2907A 2N2907A 2N2907A 2N2907A 2N2907A 2N2907A
3
3
14
13
12
10
11
FU3
9
DIG1 AN(G)
DIG2 AN(G)
DIG3 AN(G)
8 4
DIG1 AN(R)
DIG2 AN(R)
DIG3 AN(R)
FSEG a A DP FDP
7 6
FSEG b B G FSEG g
3 5
FSEG c C F FSEG f
2 1
FSEG d D E FSEG e
Figure 56: Front Panel PCB
BI-FND 7-Segment Display
Schematic Diagram
N-560 131
Technical Discussion
FSD1
FSEG a FR45 FD7
TP33 FR36 1 2
GLED SpO2 FRC7
T POINT F2 4.7K 270
FSW1 1N4148 AC in (Green) Indicator
FSD2 FD1
FRB0 FR31
FSEG b ACLED
1
FC4 2.7K_3216(1/4w)
GLED GLED
1uF
FSD3 Low Battery (Orange) Indicator
FSEG c FD2 FR32
GLED +5V FRC6
820
OLED
FSD4
FSEG d Motion (Orange) Indicator
TP35 FR38 FD3 FR33
GLED T POINT F2 4.7K BPM
FDIGIT 8 FRC5
FSW3 820
FSD5 OLED
FRB1
FSEG e Sensor Off (Orange) Indicator
1
FC6
GLED FD4 FR34
1uF FRC4
FSD6
FSEG f OLED 820
GLED +5V
Pulse Search (Orange) Indicator
FD5 FR35
FSD7
FRC3
FSEG g 820
TP37 FR40 OLED
GLED T POINT F2 4.7K Adjust UP
Sensor Adjust (Orange) Indicator
FSW5
FSD8 FD8
FRB3 FR46
FDP FRE3
1
FC8 820
GLED OLED
1uF
Alarm Silence (Orange) Indicator
FD6 FR44
Blip Display BarGraph Array FRE2
+5V OLED 820
FU1
IN-Sensor trend (Green) Indicator
1 20 FR29 470 TP38 FR41 FD9 FR47
FRE0 a+ a- T POINT F2 4.7K Adjust DOWN
2 19 FR30 470
FRC2
FRE1 FSW6 270
b+ b- GLED
3 18 FRB4
FSEG a
1
c+ c- FC9
4 17
d+ d- FSEG b
1uF
5 16
e+ e- FSEG c
6 15
f+ f- FSEG d
7 14 +5V
g+ g- FSEG e
8 13
h+ h- FSEG f
9 12 TP39 FR42
i+ i- FSEG g T POINT F2 Sat Second
4.7K
10 11 FSW7
j+ j- FDP
FRB5
1
Bar Graph Array FC10
1uF
FDIGIT 7
N-560 133
Technical Discussion
FR9
+5V
10K
2
FQ1
FR1 FJ1
1 3
FSEG a
2 1
15 2N2907A B1 A1
4 3
B2 A2
6 5
B3 A3
FR10 8 7
B4 A4
10 9
+5V B5 A5
12 11
10K B6 A6
2
14 13
FQ2 B7 A7
16 15
FR2 B8 A8
1 3 18 17
FSEG b FRF1 B9 A9 FRF0
2N2907A 20 19
15 FRF3 B10 A10 FRF2
22 21
FRE1 B11 A11 FRE0
24 23
FRC3 B12 A12 FRE2
26 25
FR11 FRC5 B13 A13 FRC4
28 27
+5V FRC7 B14 A14 FRC6
30 29
10K FRB1 B15 A15 FRB0
32 31
FRB5 B16 A16 FRB4
2
34 33
FQ3 B17 A17 FRB3
FR3 36 35
B18 A18
1 3 38 37
FSEG c B19 A19 FRC2 +5V +5V +5V
2
+5V
2
2N2907A 40 39
15 B20 A20 FQ15 FQ13 FQ11 FQ9
42 41
B21 A21 3 1
44 43 3 1 3 1 3 1
B22 A22
FR12 2N2222A 2N2222A 2N2222A 2N2222A
+5V
10K
2
DIN 44_AB-H
FQ4 ACLED
FR4
1 3
FSEG d
FRE3
15 2N2907A FC1
1uF
FR13
+5V
10K
2
FQ5 1 TP32
FR5
1 3 T POINT F3 FDIGIT 7 FDIGIT 5 FDIGIT 3 FDIGIT 1
FSEG e
15 2N2907A
+5V
FR14
+5V
10K
2
2
+5V
2
FC2
2
FR6 FQ10 FQ12 FQ14 FQ16
FSEG f
1 3 1uF Power ON/OFF
FSW9 3 1 3 1 3 1 3 1
15 2N2907A
2N2222A 2N2222A 2N2222A 2N2222A
FQ7 FSW8
FR7
1 3
FSEG g
2N2907A
Alarm Silence
15
FR16
+5V
10K
2
FU1
FU4 FU5 FU6 FU7
FSW9
FD
8
6
W
FS
FD1
FSW5
FD2
FSD1
D8
FD3
FS
FS
D2
FD4
N-560 FSW1 FSW3 FSW6
FSD7
FSD3
P1008-0
FD5
2001. 12. 02 FSW7
FS
FS
D4
D6 FSD5
FD8
T1
MDT15-20
F1 TH1 1M/0.25W D21
N.C. BD1 L21
L AC1 + R2
+14V/1.0A
250V/2A 10D-9 C8 B3R5-15.5T
C2 120K S3L20U
R1 C1 1W R21
INPUT + 400V + + C23 + + C24
47uF SXE100uF 2K 1J104
85 VAC - 264 VAC C9 SHL C21 C22 35V 1W
F2 LF1 LB1 LXV470uF LXV470uF
LF2 AC2 - 35V 35V
N DWF-141C 275V D1
250V/2A 224LF-2023
S1VB60 UF4007 GND
D
IC1 C4
C TOP223Y 2KV
101 R22 R23
R3 S 200 8.2K
6.8
PC1
+ C5 TLP621
47uF R25 (H11A817)
35V 1K C25
4.7uF/50V
IC21
TL431
C6 R24
1J104 1.2K
D2
VR21
C7 1K
1SS133
PC1
TLP621
(H11A817) 102/400VAC/Y1
PB - MD15 - 14
Fine - Suntronix
Made in Korea
R22 R24 IC21
C24
CN1
R23 L N
VR21
C25 INLET
R21 JP3 F1 F2
JP4
250V/T2A
D2
R25
LF2
C6
L21
C23 PC1 HS1
250V/T2A
C22 C7
C21 R3 C8
JP2
IC1 R1 C9
C5
D1 R2
C1
C4 C3
TH1
D21
T1 JP1
C2 BD1 LF1
142
Index
Symbols D
--- 103 data port interface protocol 95
* 103 data port problems 49
description of N-560 pulse oximeter 2
A disassembly guide 53
display/alarms problems 47
AC input theory 112
accessing menu items 33 E
accessories 71
alarm limit ranges 11 earth leakage current 26
alarm off 103 enabling the data port 95
alarm parameter being violated 103 enclosure leakage current 26
alarm silence 103 equipment needed 7
alarm silence duration 30 equipotential terminal removal 68
alarm silence state 30 error codes 49
alarm volume 30 exploded view 73
alarm volume control 18
AO 103 F
AS 103
audible alarm settings 29 fan removal 64
automatic calibration 106 front panel 3
front panel display theory 116
B front panel PCB and controls theory 116
functional checks 5
battery 6 functional versus fractional saturation 106
battery charge 8
battery in use 103 G
battery removal 66
battery theory 113 ground integrity 26
baud rate 96
block diagram 112 H
block diagram theory 112
BU 103 hospital type environments 2
button interface theory 116
button problems 46 I
interference, terminology 107
C
cautions 1 L
charging circuits theory 115 LB 103
circuit analysis 111 LED excitation test 16
cleaning 5 LM 103
connecting to the data port 97 loss of pulse 103
controlling hardware theory 114 loss of pulse with interference 103
CPU theory 114 low battery 103
LP 103
144
Index
T
technical discussion 105
test #1
SpO2 22
test #2
BPM 21
test #3
modulation level 23
test #4
light 25
transport
altitude 83
barometric pressure 83
relative humidity 83
temperature 83
troubleshooting 43
troubleshooting guide 43
W
warnings 1
Authorized Representative
Tyco Healthcare UK LTD
154 Fareham Road
Gosport PO13 0AS, U.K.
Rx ONLY
10006636A-0206