Spectralis Service Manual: This Is An Additional Procedure of The
Spectralis Service Manual: This Is An Additional Procedure of The
Spectralis Service Manual: This Is An Additional Procedure of The
SPECTRALIS
Service Manual
Version 6
Addition 2
Spectralis power supply box: parts, versions, and replacement
Table of Contents
1 Power supply box ............................................................................................................................. 5
1.1 Spectrometer unit ..................................................................................................................... 5
1.2 Sapphire laser unit ................................................................................................................... 6
1.3 Power supply base ................................................................................................................... 7
2 Versions ........................................................................................................................................... 8
2.1 Spectralis OCT with Sapphire laser........................................................................................ 8
2.2 Spectralis OCT or Spectralis HRA OCT with blue laser diode .......................................... 10
2.3 Spectralis HRA with blue laser diode ................................................................................... 13
2.4 Spectralis HRA with Sapphire laser ..................................................................................... 14
2.5 Heidelberg Retina Angiograph 2 (HRA 2) ............................................................................ 15
3 Replacements ................................................................................................................................ 16
3.1 Power supply adapter board ................................................................................................. 16
3.2 Power supply board ............................................................................................................... 18
3.3 I-module A and power supply controller.............................................................................. 19
3.4 I-gate ........................................................................................................................................ 23
3.5 Carrier and sensor board ...................................................................................................... 24
3.5.1 Sensor board .................................................................................................................. 24
3.5.2 Carrier board .................................................................................................................. 25
3.6 I-module D ............................................................................................................................... 27
Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 4 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
CAUTION!
ESD (electrostatic discharge)
Only work in Electrostatic Protected Area
Use in the field an ESD service kit
Electrostatic discharge (ESD) is the sudden and momentary electric current that flows between two
objects at different electrical potentials. Prevention of ESD bases on Electrostatic Protective Area
(EPA). EPA can be a small working station or a large manufacturing area. The main principle of an
EPA is that there are no highly charging materials in the vicinity of ESD sensitive electronics, all
conductive materials are grounded, workers are grounded, and charge build-up on ESD sensitive
electronics is prevented. For service jobs in non-electrostatic protected areas we highly recommend
the use of an ESD service kit. With this kit the worker is grounded, and the electronic boards and
components should be placed on the grounded ESD mat.
_______________ Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 5 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
1. 2. 3. 4. 5.
Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 6 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
1. Sapphire laser
2. Sapphire laser controller
3. AOM modulator
4. Transmission filter
5. Four axis manipulator
6. Sapphire controller adapter board
7. AOM driver
1. 2. 3. 4. 5. 6. 7.
_______________ Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 7 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
To get access to the base, you need to remove the spectrometer unit and the Sapphire laser unit.
Depending of the configuration of the unit you may have no spectrometer unit or no Sapphire laser
unit. This is described as well.
2 Versions
2.1 Spectralis OCT with Sapphire laser
4.
5. 4.
3.
Place the cover as illustrated below. Protect the top of the power supply housing.
Now carefully turn the spectrometer unit upside down. Take extremely care of the
cables.
CAUTION!
Do not pitch or crush those cables here!
_______________ Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 9 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
1.
1.
3.
1. 1.
2.
CAUTION!
2.2 Spectralis OCT or Spectralis HRA OCT with blue laser diode
6.
5.
4.
3.
zoomed detail
Hold the Spectrometer block with the left hand. Take care of the cables, and continue with the
next steps:
1. 2.
1. cut cable ties
2. remove upper plug
3. remove lower plug
zoomed detail
3.
CAUTION!
Carefully remove
the Spectrometer block
completely.
_______________ Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 13 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
CAUTION!
1. 1.
3.
1. 1.
2.
CAUTION!
This unit was delivered from 2003 till mid of 2008, the serial number is HRA2-00675 or lower.
The power supply box of the HRA2 is almost the same, but the power supply adapter board
was not installed. The access is the same as for a Spectralis HRA with Sapphire laser.
Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 16 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
3 Replacements
3.1 Power supply adapter board
Depending of the type of unit, you may have to cut cable ties before you disconnect the cables.
Spectralis HRA
_______________ Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 17 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
All connectors are accessible now. Remove all of them. Depending of the Spectralis type, you have to
disconnect 2, 3, or 5 cables from the power supply adapter board.
3. 3.
2.
4. 4.
Replace the power supply board. Continue in reversed order to install the new board.
_______________ Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 19 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
2.
1.
Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 20 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
Now you can start to unscrew the nuts from the plugs on the backside of the power supply box.
Finally take the power supply controller with the piggy backed I-module A board out of the power
supply box. Place all nuts and washers in clearly arranged order to verify the correct installation.
Now you can start to dismount of the I-module A from the power supply controller.
Unscrew the 3
slotted screws.
Remove the
screws and
washers.
CAUTION!
Take care when you reinstall the two boards. Place the I-Module A correct to pins of the power supply
controller.
bottom side
upper side
_______________ Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 23 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
3.4 I-gate
1. Disconnect this cable on the I-gate power Typically the I-gate is fixed only with 3 nuts
supply board. Due to a difficult access there is no nut
Remove one or both I-fibers. used underneath the power entry module
2. Use a 5.5mm nut driver to unscrew the nuts. Remove the I-gate.
2.
1.
When you install the new I-gate, take care of the ground cable and the blue/red cable
with the black insulating. Do not put the I-gate metal socket on to the cables.
Additional procedure of Spectralis Service Manual
Art. No.: 96 084-001 of 26 August 2011 Page 24 / 27 96049-006_SPECTRALIS_Service
Manual_V006_Addition2_EN.doc
Before you fix the three nuts, align the metal socket in a way
that you have the FireWire connectors straight and centered in
the housing cutout.
Finally connect all removed cables.
The sensor of the new sensor board is covered with a glass plate!
The carrier board is piggy backed to the I-module A board. Remove it carefully!
Take care when you install the new carrier board. There are three rows where the carrier
board is connected to the I-module A. Verify all pins fit correctly into their contacts.
Press the two boards and verify they are pressed completely together
3.6 I-module D
To get access to the I-module D, you have to remove the carrier board, see chapter 3.5.2
Disconnect the orange I-fiber. To remove the plug from the connection jack, you have to press
the clip on the lower side carefully with a screw driver, see below