Digital Strain-Gage Indicators: Web Design by

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

Copyright 1999 - 2002. All rights reserved.

Web design by Sullivan Creative

Laboratory Accuracy to
0.01%
Full 5-digit (99999 Counts)
Resolution
14 Plug-in Options
Sensitivity to 1
Microvolt/Count
Simple, Front Panel
Programming
Superior Temperature
Measurement
Process Voltage and Current Loop
Strain-Gage Measurement
LED Displayed Engineering Units
Made in USA

Digital Strain-gage Indicators

The Series 600 of Digital Strain-gage Indicators is equally compatible with


lower output bonded foil type transducers as with high output
semiconductor transducers (up to 100mV). Each instrument allows
selection from one of four full-scale voltage ranges to optimize sensitivity.

The measuring technique used is the unique voltage-to-frequency


conversion, dual slope method with a true, four-wire ratio measurement.
This method assures accurate and stable readings even in noisy
environments that can occur with fluctuating excitation power. Also, it
allows for use of indicators with floating or grounded transducers.

Three alphanumeric LEDs are supplied either for display of engineering


unit tables (e.g., PSI, KG, LBS) or as "0" for active display of dead zeros.

Digital scaling is standard. Drift-free scaling is set to precise engineering


units and decimal location via the instrument's simple, front panel keyboard
entry.

Dual engineering units (e.g., LBS and KG) are standard. The primary and
secondary engineering unit displays are easily toggled via the front panel.
For example, the primary could be pounds (lbs) while the secondary, a
mathematically proportional kilograms (kg). Because of their math
equivalency, all options function correctly, regardless of display selection.

System calibration can be accomplished using a 3-point "live" load, such as


actual weight on a scale, or via electronic instrument calibration using a DC
voltage source.

To reduce settling indications on the display or option outputs, the display


may be rounded to count by 1, 2, 5, 10, or 100 counts via a simple program
menu selection.

Auto zero (tare) may be accomplished by the auto zero menu command or
while the process is running. A simple, simultaneous push of the arrow
down - (Ø) and ENT keys rezeros the indicator.

Three Base Models

There are three base models of strain-gage indicators from which you can
select. The models vary by their excitation power supply and the availability
of option slots.

The Model 620 is for a user-supplied, external excitation source. It has two
available option slots.

The Model 621 includes a built-in excitation supply capable of driving one
or two 350 bridges. Supply voltage is switch selectable for 5, 10, or
15VDC. One option slot is available.

The Model 622 includes a higher output built-in excitation supply and,
because it comes in a larger double-width chassis, the ability to add up to
four options. This excitation supply provides 5, 10, 15, or 20VDC selection
with the capability to excite up to eight 350 bridges.

Order Models 620G--8 for use with external supply and up to two options,
621G -8 for built-in supply and one option slot, 622g--8 for built-in supply
and four options slots.

NOTE: " " specifies power: 1 = 115VAC, 2 = 230VAC.

Strain-gage Range Table

Range Order Code Accuracy(±)* Resolution


-15mV to +100mV -8 0.01% 1µV per
count
-30mV to +200mV -8 0.01% 1µV per
count
-75mV to +500mV -8 0.01% 1µV per
count
-1500mV to +1V -8 0.01% 1µV per
count
* Accuracy percentage is of reading, ±1digit.

Series 600 Options

Tailor the basic instruments and your applications with options to extend
your measurement capabilities. A complete explanation, any usage
restrictions, and option ordering codes are included in the following
descriptions. Any option can be ordered now or added in the future so your
Series 600 Indicator will never be obsolete.

Available options slot summary: 610 Models: 2, 612 Models: 5, 620


Model: 2, 621 Model: 1, 622 Model: 4.

Multiple inputs. Monitor as many as 24 inputs. The first multiple input card
provides an additional five channels of thermocouple or voltage, or two
channels of RTD, or thermistor inputs. Additional cards add six more
thermocouple or voltage, or three more RTD or thermistor inputs. One card
per Model 610 or a maximum of 4 cards per Model 612 instruments. (This
feature is not for use with current loop or strain-gage versions.) Order
Option -21 for 6 input of thermocouple or voltage, -21 for 3 inputs of RTD or
thermistor.

Differential measurement. This card provides an additional input channel


and the capability to display the difference (sT) between the two channels.

The -22 option card can be specified with thermocouple and voltage input
models. The -23 option is specifically for use with RTD or thermistor
configurations.

This function can be combined with other options: maximum, minimum,


and average deltas when used with math capability, scaled analog output
of differentials when used with analog output, and alarm points based on
differential limits (Cannot be used in same indicator with -20 and -21
options. Not for use with current loop or strain-gage versions.) Order Option
-22 for DELTA and 2 thermocouple or voltage inputs, -23 for DELTA and 2
RTD or thermistor inputs.

Scaling and offset. Specifically designed for use with voltage and current
loop versions of the 610 and 612, this option provides wide range user
scaling of the display to process units.

With this option you can rescale and offset inputs for display in direct
engineering units to ±99999 counts. In addition, you can assign
alphanumeric labels or dead zeros to the displayed units.

For example, 4-20mA could be scaled to display "0.00-255.00 PSI", "LBS",


"KG", "GPM", etc. And it's easy.

Unlike with other instruments, you need not perform any computations or
enter any formulas. Simply enter upper and lower input and output values.
The Series 600 does the rest, adjusting the optimum slope and offset for
you automatically. (This option is not for use with Models 620, 621, or 622
as digital scaling is included as standard.) Order Option -1

Math expressions. Provides display of maximum, minimum, rate of


change, or the timed average value of measurements. Front panel keys
allow you to view the math value or switch back to view the actual
measured value. Order Option -04

Alarms. Each option card provides two alarms with separate limits and
relays. Up to two cards can be specified. Any combination of high, low, or ±
values can be used to trigger the alarm from either actual or math
processed input. Trigger delay and a settable dead-band are included as
standard. Order Option -03 (Up to 2 per instrument.)

DC power. For automotive, marine, aviation, or plant DC power loops, the


Series 600 can be configured for 12 or 24VDC power operation. (Cannot be
used in Models 621 or 622.) Order Option -10 for 9-16VDC, -11 for
18-32VDC (One type per instrument.)

Analog output. For output to recorders or other analog input instruments,


you can choose from two fully isolated linearized and analog outputs. Both
output types, 4-20mA or 0-10VDC, can be fully scaled. Output can be
assigned to direct measurements or math values. Order Option: -06 for
4-20mA, -05 for 0-10VDC

Scanning Option. The scanning option gives the capability to scan


multiple inputs by working with the multi-input cards in your indicator. This
new option requires no programming; thus, no special menu prompts are
given with this option installed.

The display readout shows the value of multi-input card channels at a


default rate of 3.4 seconds a channel. The display rate can be made slower
(6.8 sec) - refer to the configuration instructions to change the rate of
display.

It is recommended that the scanning option not be used with a


communications option such as serial output, etc., since measurement data
cannot be identified as discrete channels. Order Option: -31

Digital Outputs
Three digital output modes are available for output to digital equipment.

Serial output. RS232C or 20mA interface sends display information in


serial ASCII format. Includes user-selectable transmission rates from 300 to
9600 baud, line feed, and carriage return. Order Option -07

IEEE-488 output. For integrating IEEE-488 Bus based measurement or


analytical systems. Main indicator and other options can be programmed
via the IEE-488 Bus. Order Option -08

Series 600 accessories. The Rack Adapter Plate permits mounting of the
Series 600 instruments in a standard 19" (48.2cm) instrument rack.
Different versions have cutouts for 1, 2, or 3 indicators.

Order Accessory -RKM1 for 1 cutout for Models 610, 620, or 621; -RKM2 for
2 cutouts for Models 610, 620, or 621; -RKM3 for 3 cutouts for Models 610,
620, or 621; -RKD1 for 1 cutout for Models 612 or 622.

Model 600 Series General Specifications

Repeatability
±1 digit

Stability with Ambient Temperature


Zero: 0.5µV/°C
Span: 0.005% rdg/°C
Reference Junction automatic built-in,
0.018°/° for 0°-50°C J, K, T, E, N, C,
G, D, CGI,FeCon, and CuCon thermocouples
For R, S, and B thermocouples: 0.03°/°

Noise Rejection
NMRR: 60dB at 50Hz or 60Hz ±0.1Hz
CMRR: 120dB at 50Hz or 60Hz ±0.1Hz

Overload Protection
Power lead to ground: 1500VDC or VAC RMS
Input to ground: 270VDC or VAC RMS
Across inputs: 270VDC or VAC RMS continuous

4-20mA range: 80mA


10-50mA range: 200mA

Input Impedance
Thermocouples: 22M (with 20nA of break
detect current)
RTDs: I1 to VinLo (10M ),
12 to VinLo (12.3K ),
depending on range VinHi to VinLo (500M )
Thermistors: I1 to VinHi (3.2K ),
I1-12, (9.45K )
Voltage: 10M to 500M depending on range
Current: <15
Strain-gage: 10M to 500M depending on range

Environmental Ranges
Operating: 0° to 50°C
Storage: -40° to +65°C
Humidity: 90%RH noncondensing

Display
8 digits, 14-segment alphanumeric red LEDs
0.54" (13.7mm) H.
Also one negative ("-") LED at left of array

Input Connections
Sensors: screw terminal blocks
Multi-input: quick connect
AC power: plug-in
DC power: screw terminal blocks

Alarms Option
2 alarm settings per alarm option -
up to 2 alarm options per instrument,
Form C relay output (1A at 120VAC) reset
auto, manual, remote, and override selectable
alarm delay and deadband

Multi-input Options -20 and -21 Accuracy


Thermocouple: ±10µV to ±70µV depending on
T/C type and measurement point
RTD:
Cu10: ±10m
All others: ±50m
Current loop and thermistor: no effect

Point Update Rate


2 readings per second,
1 reading/second for R, S, B, C, G, D,
and CGI thermocouples

Linearization, Temperature Inputs


100% digital, using variable length,
second-order, polynomial segments

Program Storage
EEROM

Case Construction
Metal, black anodized, extruded aluminum
Reliability
40,000 hours MTBF

Installation
Panel mounting from front, secured at sides
by rail clamps supplied with each indicator

Size (H x W x D)
Single-wide Models 610, 620, 621
Case: 2.63" x 5.34" x 9.87"
(6.7cm x 13.6cm x 25cm)
Bezel: 2.84" x 5.67" (7.2cm x 14.4cm)
Panel cutout: 2.68" x 5.44" (6.8cm x 13.8cm)

Double wide Models 612, 622


Case: 2.63" x 10.69" x 9.87"
(6.7cm x 27.2cm x 25cm)
Bezel: 2.84" x 11.01" (7.2cm x 28.cm)
Panel cutout: 2.68" x 10.79" (6.8cm x 27.4cm)

Weight
Single-wide: 4 lbs (1800 g)
Double-wide: 6 lbs (2800 g) without options

Power
AC power, 90V to 132V: 48Hz to 400Hz,
190V to 262 V: 48Hz to 400Hz,
8.0W typical (without options)
detachable 6-foot (1.8m) power cord is supplied,
two detachable 6-foot power cords are supplied
with Model 622

Excitation Supply
Model 621: 5, 10, and 15 volts, switch selectable
isolated, will drive two parallel 350
strain-gages at 15V (90mA)

Model 622: 5, 10, 15, and 20 volts, switch selectable,


isolated, will drive eight parallel 350
strain-gages at 15V (350mA)

Warranty
1 Year

600 Series Ordering Guide

Base Power Code Range Codes Options Code


Model
610 1= 90-132VAC -1 (Thermocouple) See 600 Series
612 2=190-262VAC -2 (Voltage) Options List for
-4 (Current & Loop) individual details
-5 (RTD) and codes.
-6 (Thermistor)
620
621 -8 (Strain-gage)
622
Examples: 610-1-1-03
612-2-5-21-21-06-04
621-18
622-1-8-16-03-03-07