IPA1 Series Infusion Pump User Manual
IPA1 Series Infusion Pump User Manual
IPA1 Series Infusion Pump User Manual
0197
Version:0
September 2017
Contents
1. EXPLANATION OF SYMBOLS AND WARNINGS...............................- 1 -
1.2 WARNINGS......................................................................................... - 1 -
2. TERMS AND DEFINITIONS.................................................................- 4 -
3.1 INTRODUCTION...................................................................................- 4 -
3.2 APPLICATION......................................................................................- 5 -
4. IMPORTANT PROPERTIES.................................................................- 5 -
6.1 STRUCTURE......................................................................................- 12 -
7. CLINICAL OPERATION......................................................................- 15 -
I
8. ALARM SIGNAL FUNCTION..............................................................- 25 -
9. IV SET DEBUGGING..........................................................................- 33 -
II
11.TECHNICAL PARAMETERS..............................................................- 35 -
12.INTERNAL BATTERY.........................................................................- 41 -
13.5 M AINTENANCE................................................................................- 43 -
15.WASTE TREATMENT.........................................................................- 44 -
15.1 BATTERY........................................................................................ - 44 -
16.ELECTROMAGNETIC COMPATIBILITY (EMC)................................- 45 -
17.EMC STATEMENT..............................................................................- 52 -
III
IV
1. Explanation of Symbols and Warnings
Symbol Description
Attention!
Disconnection (used in part of the device
only)
Connection(used in part of the device only)
AC
DC
CF application part
Access to files
Production date
Discard prohibited
Lot number
1.2 Warnings
1
necessary to place any other device over / near the pump, it should be verified that
the pump can run under normal operating conditions.
Avoid using the pump in presence of flammable gases and flammable
anesthetic mixture with air, oxygen or nitrous oxide.
The device can be operated under the environment with no strong
shock and strong electromagnetic fields.
Not suitable for portable infusion.
Prohibited to be used, together with other infusion systems, in the
same infusion tube to the patient.
Be sure that the pump is kept at a distance of no more than 1m from the
allowable installation height of IV set (i.e. transfusion set) above or below the
patient's heart.
All parts and accessories of the device must be the product of specified
model.
The device cannot be used to automatically identify the type of IVset.
The operator/ user must ensure that the IVset is the recommended type and
select the corresponding gears.
The device should be maintained calibrated by a trained professional.
The IVset without recommendation may cause the large deviation of infusion
accuracy or excess of specified flow rate accuracy.
The device belongs to IPX4 Infusion Pump. Do not immerse the device in
water. Contact your authorized dealer for repair, in case of any inflow of
liquid into the device. Continued use of the device is allowed only after the
check by a trained professional.
If the device falls from a height or suffers from strong shock, it is allowed to
be used only after being checked by a trained professional.
Prohibited to press the button with sharp object.
The device has no function of liquid leak detection. The user should
confirm there is no liquid leak in the IV set before use, and stop using
2
it and contact your authorized dealer for repair in the case of
any leakage.
Infusion parameters must be set as determined by the doctor.
Improper dosage of infusion could harm patients.
If IVset is in continuous use, the user should change the extruded
position of IV set every 6 to 8 hours to maintain an accurate infusion rate.
The user must turn off liquid blocking roller before the change of
mounting position of IV set, and turn it on after completion of mounting.
Press “off” button if there is a non-trouble alarm from the device,
press the “on” button to start the operation after the trouble shooting.
Same or similar devices used in any separate area, if used with
alarm presets, may pose potential hazard, such as ICU or operating
room.
Before opening the door for repairs, maintenance personnel must
remove power supply cord.
Except the transducer and cable which are used by the
manufacturer of device or system as the spare parts for internal
components, transducer and cable may result in device or system
emission increase or immunity decrease.
Untrained personnel are prohibited to open the door of pump,
otherwise they would be disqualified from the warranty.
Contact your authorized dealer for maintenance of the pump if its
device and accessories reach the end of their service life, which may
cause a malfunction of the pump.
User should conduct periodical inspection of the pump. If any irregularity and
failure are detected during the normal operation, stop operation of the pump
immediately and contact your local authorized dealer to replace with other
infusion device or mode.
3
2. Terms and Definitions
3.1 Introduction
3.2 Application
4
Intended use: applicable to clinical intravenous infusion in medical
institutions.
Contraindications: Blood transfusion, insulin, analgesia,
chemotherapy and epidural anesthetics.
IPA 1 1 2
P roduct s e rie s
S ub mode l
4. Important Properties
6
5. Basic Parameters, Main Features and
Functions
7
Flow rate -time mode Flow rate -time mode
Flow rate – volume Flow rate – volume mode
mode Time – volume mode
Time – volume mode Dosage mode
1-1200ml/h
b) Infusion speed set
Flow rate<100ml/h,Step value 0.1
range
Flow rate≥100ml/h,Step value 1
c) Mean flow rate
±5%
accuracy
d) Setup delivery 1-9999ml,<100 Step value 0.1,≥100 Step
volume range value 1
e) Setup delivery
±5%
volume accuracy
f) Setup delivery time
00:01-99:59 (hour : minute)
range
8
occlusion Medium 60 KPa±20KPa
o) Maximum trigger-
time of warning
occlusion, at minimum
operating speed and 55min
minimum
warning occlusion
threshold (pressure)
p) Maximum trigger-
time of warning
occlusion, at medium
operating speed and 1min30s
minimum
warning occlusion
threshold (pressure)
q) Maximum trigger-
time of warning
occlusion, at minimum
operating speed and 1h35min
minimum
warning occlusion
threshold (pressure)
r) Maximum trigger-time
of warning occlusion, at
medium
operating speed and 2min55s
maximum
warning occlusion
threshold (pressure)
9
s) Bolus dose,
at medium
operating speed and
1.0ml
minimum
warning occlusion
threshold (pressure)
t) Bolus dose,
at medium
operating speed and
1.6ml
maximum
warning occlusion
threshold (pressure)
u) Maximum capacity
under single fault 1.0ml
condition
1.0-5.0ml/h adjustable,Default1ml/h,Step value
v)KVO flow rate 0.1
Mean flow rate accuracy±5%
Total infused volume / Accumulated infusion
w)Real-time display of
volume
infusion information
Flow rate
10
ab)Enclosing class IPX4
ad) Alarm function Battery exhausted, Flow error, Power off, Power connection,
Recharge the battery, Recharge the battery fully, low ambient
temperature
7
6
5 8
4
3 9
10
F/
11
12
2
1 13
14
12
Table 6-1 Front view
15
16
17
18
13
Internal battery status Occlusion alarm pressure gear
KVO status
Occlusion pressure
selection/night mode switch
OK button
Fast discharge/Bolus
Numerical button
Start/Stop
Zero/Silence(Mute)
7. Clinical Operation
14
IV set initially used in the device must be debugged. For debugging
method, see "9 IV set debugging".
Basic operation steps: Fix the unit/device→Turn on →Setup
parameters→IV set fit→Tube emptying→Start infusion→Infusion
completion→Turn off
Note: Before the infusion, make sure that IV set used can match the
gear selected.
7.1Device Fixation
As shown in Table 7-1, during fixing the device, first loosen the knob,
put the pole of infusion support into fixed stand, adjust the position, and
tighten the knob; Ensure fixed stand is placed horizontally, and pole of
infusion support and fixed stand are placed vertically. The
vertical tilt angle should not exceed 10 degrees if there is a tilt.
User/operator must ensure that the device is placed in a stable and
secure manner.
7.2Power on
15
After power cord is connected, LCD on the front view lights on, press
7.3Setup Parameters
zero the input number by pressing key. Press key for the next
17
parameter setup. Press key for cyclic setup of “mode serial number
> dosage>body weight of patient> drug mass> solution volume> mode serial
No.” parameters. Cursor cyclically points to each parameter logo.
The parameter pointed indicates the parameter selected enters the
value by numeric keypad.
The mode serial No. corresponds to the unit serial number specified
in the list of input mode unit according to the drug dosage unit used. The
parameter “body weight of patient” is not included in the dosage unit, and
there is no need to enter the parameter value of body weight.
dosage mode parameter set, and infusion flow rate parameter value is
generated automatically, then set up time or preset delivery volume.
Flow rate parameter value generated in dosage mode can’t be
modified, e.g. quit dosage mode by modifying device, and clear the
parameters in dosage mode zero. Flow rate parameter value generated
in dosage mode ≤1200.
d) Setting parameter of drop volume is equivalent to flow rate. These
two parameters are automatically converted according to the following
formula. Setting either drop volume or flow rate will be ok.
Formula:
d rip rate(dotp ⁄ min)=Flow rate (ml /h)∗drop volume per ml( drop/ml)/60
e) Set up time by taking “●” as delimiter for hour and minute. Input “●”
after setting “hour “parameter value, then input “minute “parameter value.
18
7.4.1 Fitting of IV set
a) Pour drug solution into drip cup of IV set, reaching 1/2 level.
Open the door (of pump) by pulling the handle of lock; Fix, from top to
bottom, a tube into bubble detector slot and tube slot at the 10cm
below the drip cup.
b) Open tubing clamp, and thread IV set through bayonet of tubing
clamp, keeping IV set vertical.
c) Close the door, and open the drop pause pulley of IV set.
7.4.2 IV set /Solution Container Replacement and Re-installation
a) If replacing IV set or fitting position, close the drop pause pulley
of IV set to avoid the solution gravitational flow.
b) Open the door and tubing clamp, take out IV set, re-install IV
set, then open the drop pause pulley of IV set after fitting.
c) Before replacing solution container, stop the operation, draw out
puncture outfit in IV set, and insert it into solution container; after fixing
the solution container, start the operation of the device after checking.
7.6Infusion Start
19
a) Confirming that IV set is removed and connected to the patient,
press key to start infusion.
b) RUN LED, at1Hzfrequency, flashes. Operation indicator lights
cyclically. IV set starts to accumulate infused volume.
7.7Infusion Completion
When the total infused volume reaches delivery volume, the device
will sound the alarm of setup delivery volume completion and display
“completion” logo in alarm instruction zone, with its flow rate turning into K.V.O
rate automatically and displaying “K.V.O” logo. In case of alarm, turn the
7.8Turning Off
a) After completing the infusion, stop the operation of the pump by pressing
key.
3 consecutive seconds.
Attention: Prohibited to turn off the power when the device is in normal
operation, otherwise, it may cause infusion interruption.
key, and clear zero “total infused volume”, “remaining time” and
22
In night mode, the user may lower the luminance and sound in
decibels of the device, which can reduce the impact on the patient in
the night time.
a) Under the normal infusion state, the user may turn on night
sequence.
d) U000-U021 stands for function (option). Select “setup” option, and
23
e) Select “U014” option. Quit by pressing key, and return to
standby display.
Table 3 Description for user function setup parameters
Function Default
Range Description
items value
U000 — IV set debugging interface —
alarm volume setting
U001 0,1 1
0=low,1=high
Retain memory
U002 0,1 1
0=off,1=on
Drop parameter display switch
U003 0,1 1
0=off,1=on
Run default parameter selection
U005 0,1,2 0=flow rate,1=time,2=preset 0
delivery volume
Default parameter selection
U006 0,1 0=accumulated infusion volume, 0
1= remaining time
IV set lock
U008 0,1 0
0=off,1=on
Key tone switch
U010 0,1 0
0=off,1=on
1-
U011 1200m Bolus default flow rate setting 800
l/h
1.0-
U012 Single bolus maximum 3
10ml
U013 1.0- KVO flow rate 1
24
5.0ml/
h
Return to standby display by
U014 — —
pressing OK key
U016 15-70 IV set parameter/ml 20
Alarm thresholds for
20-
U018 occlusion pressure of IV set at —
1999
various gears
Return to ex-factory parameters
U019 0,1 setting 0
0=off,1=on
U020 0-10 Nearly completion alarm time 0
Occlusion default alarm gears
U021 — 1
1= medium
26
b) When alarm goes off with “occlusion”, press ] button to hold
alarm sound for 2 minutes. Alarm sound goes off again 2 minutes later.
Alarm resets by pressing “stop” button, and " Occl." character is blank,
and the device stops operation, entering“standby”mode.
c) Check if there is any bending and extrusion in the tube, and any
occlusion in filter needles; Restart the infusion after removing the
problem.
d) Check accuracy of alarm: Flow rate by initial setup is 25 ml/h;
Preset delivery volume is 5 ml; Operation is started; In case of artificial
interruption of drug infusion, the device gives “occlusion” alarm, and
“Occl.” is displayed on the LCD, at the same time, the device stops
operation, LED flashes and alarm sounds continuously, which shows
the accurate “occlusion” alarm.
27
“KVO” characters are displayed on the LCD, and the flow rate turns
into K.V.O. rate automatically.
b) When alarm goes off with “completion”, press button to hold
alarm sound for 2 minutes. Alarm sound goes off again 2 minutes later.
Alarm resets by pressing “Stop”, "Completion" and “KVO” characters
are blank, and the device stops operation, entering“standby”mode.
c) Check accuracy of alarm: Flow rate by initial setup is high level;
Preset delivery volume is 5 ml; Operation is started; When
accumulated window value reached 5, the device gives “door open”
alarm; At the same time, the device runs at K.V.O. flow rate, “KVO”
characters are displayed on the LCD, and LED flashes, which shows
the accurate “completion” alarm.
a) When total infused volume reaches the delivery volume, the device
gives near completion alarm, "near completion” character is displayed
on the LCD, and the device continues operating.
30
b) When alarm goes off with “Near completion”, press button to stop
alarm sound. The alarm lasts until the infusion completion.
c) Near completion alarm can prompt user to make preparation
before the infusion. The alarm offending time can be customized, setup
time range : 0-10 minutes, setup method referred to in the article 7.14.
8.10 Power off Alarm
a) When external power is off, the device gives power off alarm,
external power indicator lights out, the device continues operating,
turning into battery power supply automatically, and battery token
flashes.
b) When external power is off, check external power line status,
connect power plug into the device again, alarm resets, and battery token
stops flashing. Contact the manufacturer to replace with new one if it is
judged that device failure results in power disconnection.
31
f) Alarm system of the device is technical alarm.
g) Air-in-line alarm during fast discharge is inhibited.
h) If power outage is less than 30s, alarm setting before the
outage
is automatically restored.
32
When ambient temperature is 15 ℃ below, the device displays
command prompt of low ambient temperature. Snowflake command
prompt lighting up and flashing, the device continues operation. When
ambient temperature is above 15 ℃, the low ambient temperature
command prompt resets automatically, and the prompt is blank
automatically.
9. IV Set Debugging
cyclically.
key.
steps (for 1-2 times) until the consistency between display milliliters
35
Dotted portion in the above graph indicates the setup flow rate ( 1
ml / h herein). Solid line is continuous connection line of mean flow
rate value during the sampling period.
f) Ascent curves of Jerry IV set at medium flow rate in the first two
hours
Dotted portion in the above graph indicates the setup flow rate
( 25 ml / h herein). Solid line is continuous connection line of mean
flow rate value during the sampling period.
g) Trumpet curves of Jerry IV set at lowest flow rate in the 2nd hour
37
k) Ascent curve of Kindly IV set at lowest flow rate in the first two
hours
Dotted portion in the above graph indicates the setup flow rate ( 1
ml / h herein). Solid line is continuous connection line of mean flow
rate value during the sampling period.
l) Ascent curves of Kindly IV set at medium flow rate in the first
two hours
Dotted portion in the above graph indicates the setup flow rate
38
( 25 ml / h herein). Solid line is continuous connection line of mean
flow rate value during the sampling period.
m) Trumpet curves of Kindly IV set at lowest flow rate in the 2nd
hour
39
deviation curve within the 2nd hour;
o) Trumpet curves of Kindly IV set at lowest flow rate in the last
one hour
13.2Cleaning / Disinfection
The series ofbattery requires routine surface cleaning and
disinfection before being used between the patients.
41
Do not use diluents or other organic solvents for cleaning. Do not
use high pressure steam for sterilization. Do not use desiccant or
similar products for drying. Do not immerse the product in water.
It is recommended to use the following cleaning / disinfection
methods. Disinfectants should be used in accordance with the
instructions
Ultraviolet disinfection
Ethylene oxide gas (EOG) disinfection : Placed in the ventilated
place for more than 8 hours after the disinfection
Ozone disinfection
Chlorine dioxide disinfectant surface wiping
Cleaning or sterile towel is used to wipe the surface of device.
Disconnect AC power outlet when cleaning or disinfecting to prevent
the detergent from soaking through the device. Dry the splashed liquid
in time.
Attention: Turn off infusion pump before cleaning, and disconnect
AC power; Prevent liquid from leaking out or soaking through the
device;
Avoid the accumulation of liquid on the pump; Avoid using strong
cleaning agent to avoid damage to outer surface.
Attention: Do not clean the pump in case of any visible cracks,
breaks, or damage to the cover of the pump. Stop the operation
immediately and arrange qualified maintenance personnel to inspect
the pump.
Avoid scratching the bubble detector during the use and cleaning.
Clean it immediately if the solution is found on occlusion sensor and
42
bubble detector to avoid the false alarm due to reduced performance.
13.5 Maintenance
43
14.Transport and Storage
Temperature range:-20℃~+55℃
Relative humidity:≤93%
15.Waste Treatment
15.1 Battery
16.Electromagnetic Compatibility (EMC)
44
electromagnetic environment specified below. the customer or the user of
the IPA1 series infusion pump should assure that it is used in such an
environment.
Emissions test Compliance Electro-magnetic environment-
guidance
The IPA1 series infusion pump
uses RF energy only for its internal
RF emissions function. Therefore its RF emissions
Group 1
CISPR11 are very low and are not likely to
cause any interference in nearby
electronic equipment
RF emissions The IPA1 series infusion pump
Class A
CISPR11 is suitable for use in all
Harmonic emissions Not establishments other than domestic
IEC61000-3-2 applicable and those directly connected to the
Voltage public low-voltage power supply
fluctuation/flicker Not network that supplies buildings used
emissions applicable for domestic purposes.
IEC61000-3-3
Guidance and manufacturer’s declaration – electromagnetic immunity
–for all EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The IPA1 series infusion pump is intended for use in the
electromagnetic environment specified below. The customer or the user of
the IPA1 series infusion pump should assure that it is used in such an
environment.:
IEC 60601 Compliance electro-magnetic
Immunity test
test level level environments: guidance
Electrostatic ±6KV ±8KVContact Floors should be wood,
discharge Contact discharge concrete or ceramic tile. If
IEC61000-4-2 discharge ±15KV air floors are covered with
45
synthetic material, the
±8KV air
discharge relative humidity should
discharge
be at least 30 %.
±2KV ±2 KV Mains power quality
connected to connected to should be that of atypical
Electrical fast power lines power lines commercial or hospital
transient/burst ±1 KV environment.
IEC61000-4-4 connected to not
in/output applicable.
lines
±1 KV Mains power quality
differential- ±1 KV should be that of atypical
mode differential- commercial or hospital
Surge voltage mode voltage environment.
IEC61000-4-5 ±2 KV
common- Not
mode applicable
voltage
Voltage <5%UT, <5%UT, Mains power quality
sags, short lasting half a lasting half a should be that of atypical
interruptions period period commercial or hospital
and voltage (>UT, >95% (>UT, >95% environment. If the user
Variations on sags) sags) of the IPA1 series
power supply 40%UT, 40%UT, infusion pump requires
input lines lasting five lasting five continued operation
IEC61000-4 periods periods during power
-11 (> (>UT, 60% mains interruptions, it is
UT,60% sags) recommended
sags) 70%UT, that the IPA1 series
70%UT, lasting 25 infusion pump be
lasting 25 periods powered from an
46
periods (>UT, 30%
(>UT, 30% sags)
sags) <5%UT,
<5%UT, lasting 5 uninterruptible power
lasting 5 seconds supply or a battery.
seconds (>UT, >95%
(>UT, >95% sags)
sags)
Power Power frequency
frequency magnetic fields should be
magnetic at levels characteristic of
400A/m 400 A/m
field(50/60 atypical location in a
Hz) typical commercial or
IEC61000-4-8 hospital environment.
Note:UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
– for Life-supporting EQUIPMENT and SYSTEMS
Guidance and manufacturer’s declaration—electromagnetic immunity
The IPA1 series infusion pump is intended for use in the
electromagnetic environment specified below. The customer or the user of
the IPA1 series infusion pump should assure that it is used in such an
environment.
47
Than the recommended
separation distance
3 Vrms 3V
calculated from the
( Effective Value)
Radiation RF equation applicable to the
GB/T 150 kHz to frequency of the
IEC61000-4-3 80MHz transmitter.
Except
Recommended
ISM bands
separation distance
10 V150 d=1.2 P
kHz to
10V
80MHzin
ISM bands
P
d=1.2
10V/m
80MHz~ 10V/m
2.5GHz
d=1.2 P 80 kHz to
800MHz
d=2.3 P 800MHz to
2.5GHz
NOTE1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
a The ISM (industrial, scientific and medical) bands between 150 kHz
and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz;
26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz
b The compliance levels in the ISM frequency bands between 150 kHz
and 80 MHz and in the frequency range 80 MHz to 2,5 GHz are
intended to decrease the possibility that mobile/portable
communications equipment could cause interference if it is
inadvertently brought into patient areas. For this reason, an additional
49
factor of 10/3 is used in calculating the recommended separation
distance for transmitters in these frequency ranges.
50
150kHz to 150kHz to 80MHz~ 800MHz~
80MHz 80MHz in 800MHz 2.5GHz
outside ISM d = 1.2 d = 2.3
output power
ISM bands P P
of
bands d = 1.2
transmitter/W
d = 1.2 P
P
51
propagation is affected by absorption and reflection from structures,
objects and people.
17.EMC Statement
IPA1 infusion pumps have been tested, and conformed with medical
equipment standard YY0505-2012.These restrictions are intended to
provide reasonable protection for typical medical devices, to prevent
harmful interference.
a) The device can transmit radio frequency. It may cause harmful
interference to other equipment nearby if the installation and use of the device
are not in accordance with the instructions. However, it can’t be guaranteed
that such interference will not occur in a particular device.
The device can be used by the trained professional medical personnel
only.
If the device causes harmful interference, it can be determined by
turning on / off the device. The user is encouraged to resolve it by
adopting one or more of the following measures:
Readjust the position of receiver
Increase the distance between devices,
Connect this device and etc. to power outlet of different circuits
Consult manufacturer or on-site service technician
b) Take and install specific EMC precautions for medical electrical
devices, and provide service according to the EMC information provided.
Portable and mobile radio-frequency communication equipment may
affect the operation of medical electrical devices, thus medical electrical
devices should be used carefully.
Except the transducers and cables sold as part of internal
52
components of IPA1 infusion pump, use of accessories, transducers and
cables (other than those specified) may result in increased emission or
reduced radiated immunity for equipment or system.
It is not allowed to place any other device over IPA1 infusion pump.
If it is necessary to place any other device over / near the pump, it should be
verified that the pump can run under normal operating conditions.
c) Anti- electrostatic precautions
Do not contact with pins of connectors marked with electrostatic
discharge warning symbols. Do not connect to these connectors unless
electrostatic discharge precautions are adopted.
Warning: Basic contents of training in electrostatic discharge
precautions are prescribed as below:
1) We draw the attention of all the members to the followings:
Do not touch the connectors marked with electrostatic discharge
warning symbols with hands or tools, unless the following electrostatic
discharge precautions are adopted, including:
Methods of preventing accumulation of static charge (such as air
conditioning, humidification, floor conductive coating, non-synthetic
clothing etc.
Human body electrostatic discharge into the frame of device or
system, or into ground or metal object.
Connect human body to the device or system by a wrist belt.
2) All personnel who may touch the connectors marked with
electrostatic discharge warning symbols, including clinical / biomedical
engineering and medical personnel should receive the instructions herein
and training.
Electrostatic discharge training shall include an introduction of physics
related to static charge, and damage to electronic components caused by
53
voltage amplitude that may be generated in normal practice, and an
operator associated with static electricity. More specifically, methods of
preventing accumulation of static charge, and why and how to prompt
human body electrostatic discharge into the frame of device or system, or
into ground or metal object, or methods of connecting themselves to the
device or system or earth by a wrist belt before starting work, should be
specified.
54
55
MEDEVO MEDICAL Co., Ltd.
Registration Address: Room 201, F/L 2, Unit 5, Building 10, Courtyard
20,
KeChuang14th Street,Beijing Economic and Technological Development A
rea (BDA), Beijing
Manufacturing Address: F/L 2, Unit 5, Building 10, Courtyard 20,
KeChuang14th Street,Beijing Economic and Technological Development A
rea (BDA), Beijing
Tel:86-10-87227118
Web: www.medevomed.com
MedNet GmbH