Aun Infusomat P - Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 78
At a glance
Powered by AI
The document provides information on servicing various models of the Infusomat P pump, including technical details, parts lists, and safety guidelines.

The document is a service manual that provides information for servicing different models of the Infusomat P infusion pump, including technical details, parts lists, procedures, and safety guidelines.

The system overview section on pages 1-1 to 1-9 discusses the design, operation flow chart, functions, voltage supply, signal table, and mains and battery operation of the Infusomat P pump as well as its computer interface and integration with the Braun fluid manager system.

0-8

Infusomat P
Service-Manual

0-

Version 2.0 English


Software: IFPC02001
IFPC02002
IFPe02003
IFPE13002
IFPE13003

This Service-Manual is valid for

Voltage 230 V:

Ord. No.

Infusomat P, German . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2174


Infusomat P, French . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2182
Infusomat P, Dutch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2190
Infusomat P, Italian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2204
Infusomat P, Danish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2298
Infusomat P, Norwegian . . . . . . . . . . . . . . . . . . . . . . . . . 871 2301
Infusomat P, Swedish. . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2310
Infusomat P, Finnish. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2336
Infusomat P, Czech. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2344
Infusomat P, Polish. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2352
Voltages 200 V / 230 V / 240 V, switchable:
Infusomat P, English. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2212
Infusomat P, English. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2379
Infusomat P, Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2263
Infusomat P, Portuguese . . . . . . . . . . . . . . . . . . . . . . . . . 871 2271
Infusomat P, Turkish. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2280
Voltages 100 V / 110 V / 120 V, switchable:
Infusomat P, English. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2387
Infusomat P, Dutch. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2395
Infusomat P, Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . 871 2409
Infusomat P, Portuguese . . . . . . . . . . . . . . . . . . . . . . . . . 871 2417
The complete Service-Manual contains
the following pages:

Page 0-1 to Page 0-8


Page 1-1 to Page 1-10
Page 2-1 to Page 2-6
Page 3-1 to Page 3-12
Page 4-1 to Page 4-20
Page 5-1 to Page 5-2
Page 6-1 to Page 6-2
Page 7-1 to Page 7-2
Page 8-1 to Page 8-6
Page 9-1 to Page 9-2
Page 10-1 to Page 10-4
Page A-1 to Page A-4

0-2

Infusomat P 2.0gb

0-

Important Information on this Manual

Table of Contents

Service Work
Technical Safety Inspections
Current Versions
Revision Service
Non-Liability
ISO 9000 ff/EN 46 000 ff
Repair and Inspection only
by Trained Technicians
Spare Parts and Test Equipment
Setting Off

Page
Page
Page
Page
Page
Page

0-5
0-5
0-5
0-5
0-5
0-6

Page
Page
Page

0-6
0-6
0-6

General Information

Technical Training
Entry for the Product Training
Ordering of Spare Parts and Test Equipment
Return of Spare Parts
Service Hotline
Safety Officer
Translation

Page
Page
Page
Page
Page
Page
Page

0-7
0-7
0-7
0-7
0-7
0-7
0-7

System Overview

Design
Operation Flow Chart
Function
Voltage Supply
Signal Table
Mains Operation
Battery Operation
Alarm Circuit
Pump Unit
Computer Interface
Braun fluid manager system (fm system)
Internal Assignment
Accessories

Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page
Page

1-1
1-2
1-3
1-4
1-5
1-6
1-7
1-7
1-7
1-8
1-9
1-9
1 - 10

Software

Software Update
Approved Software Versions
Error Messages and Alarms
Software Default Values

Page
Page
Page
Page

2-1
2-2
2-3
2-5

Service Program

Structure of the Service Program


Additional Functions with
Inserted Service Plug
Start / Quit the Service Program
Unit Data
History Data
Test
Unit Modifications
Calibration

Page

3-1

Page
Page
Page
Page
Page
Page
Page

3-2
3-3
3-3
3-5
3-6
3-7
3 - 10

Mains Fuses
Battery
Door Lock
Pump Cover
Pump Housing
Housing and Handle

Page
Page
Page
Page
Page
Page

4-1
4-1
4-2
4-3
4-4
4-5

Unit Elements

Infusomat P 2.0gb

0-3

Table of Contents

Controller Board
Rear Panel
Front Panel
Pump Unit
Occlusion Sensor
Air Sensor
Operating Unit
Barcode Label
Frame with Seal

Page
Page
Page
Page
Page
Page
Page
Page
Page

4-6
4-8
4 - 11
4 - 13
4 - 15
4 - 18
4 - 19
4 - 20
4 - 20

Checks after Repair

Page

5-1

Maintenance

Page

6-1

Technical Safety Inspection TSI

Page

7-1

Procedural Instructions for Inspection

1. Visual Inspection
2. Electrical Safety Inspection as
per IEC/EN60 601-1
3. Functional Inspection
4. Pump Unit Inspection

Page

8-1

Page
Page
Page

8-1
8-2
8-3

Test Equipment and Special Tools

Test Equipment and Special Tools

Page

9-1

Spare Parts List

Unit Elements
Colours
Miscellaneous
Accessories
Software Update

Page
Page
Page
Page
Page

10 - 1
10 - 2
10 - 2
10 - 3
10 - 3

Appendix

ESD Recommendations
Workstation
Revision Service-Documentation
Current Information
Order Form

Page
Page
Page
Page
Page

A-1
A-1
A-2
A-2
A-2

0-4

Infusomat P 2.0gb

0-

Important Information on this Manual

The following notes should be observed:


Service Work

Technical Safety Inspections

The present manual is for your information only. The possession of


this manual does not authorize the performance of service work.
Service tasks may only be executed by persons, who
-

have received appropriate training on the system from B.


Braun

are included in the revision service

possess the necessary test equipment and mechanical aids,


and

fulfil the personal requirements (training and knowledge).

B. Braun also recommends training on the technical safety inspections, or to perform at least the steps indicated in the current
version of the manual, as:
-

the TSI requires that the instructions in the manuals are observed

the manuals are a reference for measurements

depending on the unit type, the service program must be


called up which may lead to a dangerous unit condition in the
case of an inappropriate operation. Furthermore, a special
service connector may be necessary.

Current Versions

This manual version corresponds to the state when the manual


was written. Technical changes, especially software modifications
must always be expected. The state of the revision is indicated by
the index number on the title page.

Revision Service

The possession of this manual does not automatically mean inclusion in the revision service. You will be included in the revision
service after:

Non-Liability

Infusomat P 2.0gb

technical training by B. Braun Melsungen or

a written order placed with the sales department of B. Braun


(incurred costs).

B. Braun Melsungen AG does not assume any responsibility for injuries to persons, property damages or other damages caused by:
-

the use of a wrong manual, or a manual which does not correspond to the revision state for maintenance, repair, and
service tasks

the non-inclusion of the service technician in the revision


service

technicians who have not participated in a technical training


course for the specific B. Braun unit.

0- 5

Important Information on this Manual

ISO 9000 ff/EN 46 000 ff

B. Braun is certified in accordance with DIN ISO 9001 and DIN EN


46 001. This certification also includes maintenance and service.
The Infusomat P complies with the IEC/EN60601-1 and
IEC/EN60601-2-24 standards and is certified with the CE label according to the EC directive 93/42/EC.

Repair and Inspection only


by Trained Technicians

Training may only be performed by B. Braun. The possession of the


manual does not authorize the performance of repairs. The instructions on electrostatic sensitive components (ESD standards)
must be observed.
An electrical check must be performed following each repair (see
Checks after Repair pg. 5 - 1).

Spare Parts and Test Equipment

Only use original spare parts from the manufacturer. Do not


tamper with assembly groups which can only be exchanged completely. The spare parts required are listed in the repair instructions.
The inspection personnel is responsible for the calibration of the
test equipment. Original test equipment can be calibrated at the
works of B. Braun. Further information is available upon request.

Setting Off

Additional notes and warnings are set off as follows:


Note
Gives additional or special notes on information and working
steps.
CAUTION
Is used for working steps which may result in damage to the unit,
system or to a connected device.
WARNING
IS USED FOR WORKING STEPS WHICH MAY RESULT IN PERSONAL
INJURY.

0- 6

Infusomat P 2.0gb

0-8

0-

Technical Training

General Information

B. Braun Melsungen AG
Stadtwald Haus 6
D-34 212 Melsungen
Germany
Responsibility
Dieter Gundlach

Tel.:
Fax:

Entry for the Product Training

Only via the responsible representative

Ordering of Spare Parts and Test Equipment

B. Braun Melsungen AG
P.O. Box 11 20
D-34209 Melsungen
Germany

+49 5661 71 - 37 74
+49 5661 71 - 28 81

Hospitals and Suppliers


of Hospitals

Tel.:
Fax:

+49 800 2 27 28 24
+49 5661 71 37 98

Dealer and
Medical Houses

Tel.:
Fax:

+49 5661 71 36 28/29


+49 5661 71 35 50

Return of Spare Parts

B. Braun Melsungen AG
Schwarzenberger Weg 73-79
Wareneingang Werk C
D-34 212 Melsungen
Germany

Service Hotline

Mr. Tippel, Mr. Clobes


Tel.: +49 56 61 71 - 35 25
Fax: +49 56 61 71 - 35 26
E-Mail: [email protected]
E-Mail: [email protected]

Safety Officer

Dr. Wilhelm Wucherpfennig


Fax: +49 56 61 71 - 46 36
E-Mail: [email protected]

Translation

Infusomat P 2.0gb

Brckner GmbH, Germany

0- 7

General Information

For your notes:

0- 8

Infusomat P 2.0gb

1 - 10

1-

System Overview

Design

The Infusomat P is a compact volumetric peristaltic infusion


pump.
Standard delivery rate range 0.1 to 999.9 ml/h
The unit is operated via a membrane keypad. It is equipped with
an LCD (liquid crystal display) for the display of the delivery rate
and the operating support of the user. Two control LEDs display
alarms, and the running of the infusion pump.

Guide for Short Pole Clamp

Drop Sensor

Handle

Peristaltic Pump

Flow Inhibitor
MFC-Socket

Door Lock

LCD Display
Softkeys

Optical
Interface
fm Recessed Plug
Potential Equalization
Bolt
Drop Sensor Socket

Keypad

Barcode Label
Control LEDs

Air Sensor
IP 003

Mains Connection

Fig.: 1 - 1 Unit Design Infusomat P


A barcode label is attached to the left front side of new Infusomat
P unit versions which can be retrofitted on previous devices. This
barcode label is used to read the serial and DIANET type number
via a scanner when the Infusomat P is operated in an fm-system.

Infusomat P 2.0gb

1- 1

System Overview

Operation Flow Chart

Switch On

Tube Selection 5

Standard Operation

Set Delivery Rate

Preselected Volume

Service Plug,
Inserted ... 1

Special Functions ... 3

Preselected Time

Rate Calculation

Start Infusion

Stop Infusion

Dose Calculation 5

Bolus Function

Standby Function

Drug Change

CC Mode 4

Occlusion Pressure

Drop Control

Piggyback 5

Battery Capacity

Data Lock

Alarm Tone 6

Contrast Setting

Clock 5

Change Delivery Rate

Recall Info

Call up Bolus

See service program diagram

Only when preselected time or preselected volume has expired

Please pay attention to the activation in service program

4 Only

available with IFPC

From software IFPe, IFPE

From software IFPE

KOR Mode 2
Fig.: 1 - 2
See Instructions for Use for detailed information.
1- 2

Infusomat P 2.0gb

System Overview

Function

Two independent software-controlled microprocessor systems


control and monitor the hardware. On the basis of their functions,
they are defined respectively as a control and a function processor. Both systems work with independent clock frequencies and
have access to different program and data memories. All safetyrelevant functions are handled by both processors and the results
are counter checked (CF- and FC-latch).
The input via the keypad is fed to both processors. Additionally the
acknowledgement signal of the ON/OFF key is fed to the mains
power supply logic (voltage E/A-TAS). The function processor has
also access to this logic via E/A-INT.
Description of the voltage signals (see Signal Table
pg. 1 - 5).

Fig.: 1 - 3

Infusomat P 2.0gb

1- 3

System Overview

Voltage Supply

The voltage supply is generated either directly from mains, via the
FM plug (14 V connection to the fluid manager system), or via the
MFC connector (11 to 16 V) and as an internal supply via the internal 7.2 V NiCD battery. The mains module is available in three
versions: 230 V, 220 / 230 / 240 V and 100 / 110 / 120 V. The rated
voltage has a tolerance of + 10% to 15 %.
A voltage of 11 V to 18 V is available after transformation and rectification. This voltage is fed to the battery charge circuit and the
unit supply. This is also valid for an external 12 V supply from the
MFC or FM plug. The FET V10 switches between the external and
the internal voltage supply. The transistor V10b works as an ON/
OFF switch for the operating voltages UBS, UMOT and +5V. The
+5V supplies the complete electronics including the double channel microprocessor system. A window comparator constantly
monitors the +5V for undervoltage or overvoltage. The function is
checked during switch-on. The operating voltage UBS supplies the
stepper motor and the UMOT, the stepper motor drive.
The transistor V47 switches the operating voltage UMOT. In case
of an alarm the motor is switched off by V47. Additionally the
switching function of the transistor is checked during the switchon test.
The circuit has two separate assembly groups with separate supply voltages UBA and UBB. The ON/OFF circuit has a retriggerable
delay switch-off. A follow-up charging circuit drives the transistor V10b.
The alarm logic (operating voltage UBB) is an RS latch. This is set
when the unit is running and activates the alarm circuit. The
alarm buzzer and drive are also driven by UBB. The ON/OFF circuit
is activated and the voltage supply is switched on by pressing the
ON/OFF key. The alarm latch is reset simultaneously. A function
test of the voltage monitoring, motor circuit and alarm activation
is performed. The voltage supply is maintained by cyclic self-holding pulses fed to the logic. The alarm latch is also activated.

1- 4

Infusomat P 2.0gb

System Overview

Fig.: 1 - 4

Signal Table
Signal

Meaning

Signal

Meaning

+5V

Voltage Supply Electronic

PKS2

Pump Head Sensor 2

5V-HT

Overvoltage Test

PKSS

Pump Head Sensor Control

5V-LT

Undervoltage Test

PRS

Staff Call Relay Control

AK-I

Battery Charge and Discharge Current

PRS-F

Staff Call Relay Function Channel

AK-LAD

Battery Capacity ON/OFF

PKSS

Pump Head Sensor Control

AK-Test

Battery Test

PRS.RUF

Staff Call Relay Control

ALA-UB

Operating Voltage Alarm

RDE

Rate Display Enable

CS

Chip Select

RES

Power on Reset

DI

Data Input

RES-F

Reset Function Channel

DO

Data Output

RES-K

Reset Control Channel

E/A-INT

ON/OFF by Microprocessor

RTS

Return to Send (DIANET)

E/A-ST

ON/OFF Status

Rx

Receive Data

Table 1 - 1 Signal Table (Part 1 of 2)

Infusomat P 2.0gb

1- 5

System Overview

Signal

Meaning

Signal

Meaning

E/A-TAS

ON/OFF Key

SCK

Serial Data Lock

EDB

Electronic Occlusion Pressure

TD-A0

Text Display Address 0

FMC-F

FM Connection Function Channel

TD-A1

Text Display Address 1

FMC-K

FM Connection Control Channel

TD-A2

Text Display Address 2

UEXT-N

External 12V Supply (-)

TD-A3

Text Display Address 3

UEXT-P

External 12V Supply (+)

TD-E

Text Display Enable

UFM-P

External 12V Supply (FM)

TD-R/W

Text Display Read/Write

LFCL

Air Sensor Clock

TD4

Text Display Data 4

LFDA

Air Sensor Data

TD5

Text Display Data 5

LFS

Air Sensor Signal

TD6

Text Display Data 6

LFSEL

Air Sensor Selection

TD7

Text Display Data 7

LFT

Air Sensor Reset

TSA

Drop Sensor Output

MISO

Serial Data Output Interface

TSCL

Drop Sensor Clock

MOSI

Serial Data Input Interface

TSE

Drop Sensor Receiver

MOTEIN

Motor ON

TSR

Drop Sensor Regulation

MS

Motor Control

TSS

Drop Sensor Control

P-ENA

Port Enable

Tx

Transmit Data

PH0

Phase 0

UBA, UBB

Supply Voltage for Alarm, On/Off Logic, RTC

PH1

Phase 1

UBS

Switched Operating Voltage UB

PH2

Phase 2

UBS-M

UBS Measurement Line

PH3

Phase 3

UMOT

Supply Voltage of Motor Drive

PKS

Pump Cover Sensor

UMOT-M

UMOT Measurement Line

PKS1

Pump Head Sensor 1

URTC

Supply Voltage Clock Module

Table 1 - 1 Signal Table (Part 2 of 2)

Mains Operation

When the unit is connected to mains the unit supply voltage is


switched on for the duration of the switch-off delay time. If the
microprocessor recognizes a sufficient mains voltage for the
charging, the voltage supply is maintained. In this case only a battery balance is carried out, because a key was not pressed. The actual battery capacity and the battery operting hours are displayed
in the LCD. The unit is switched off when the ON/OFF key is
pressed for at least 2 seconds. Thereby the self-holding is triggered and the alarm latch is reset with a delay. After another 20
seconds the unit is switched off, because the pulses are missing.
If the Infusomat P is switched off in mains operation with the ON/
OFF key, the internal mains voltage is still present.
In mains operation the battery function is checked during the
switch-on test. Therefor the charge and discharge current is

1- 6

Infusomat P 2.0gb

System Overview

measured and the charging of the battery is interrupted for the


duration of the measurement.

Battery Operation

The battery function is monitored by the following data: charge


current, discharge current and time, and self-discharge time. In
battery operation the battery function is checked during the
switch-on test. The theoretical load condition is read from the
clock module of the battery. Then the battery is connected to UBS
and the voltage is measured. If the minimum requirements are not
reached a battery alarm is activated.

Alarm Circuit

The alarm generation consists of:


-

Standstill of pump due to switch-off of MS (motor circuit)


and UMOT (motor operating voltage)

Audible alarm due to the drive of the buzzer or the loudspeaker via ALA-AK (control channel) or via ALA-UB

Optical alarm. Is displayed in the LCD and a separate LED display. Additionally the set rate flashes with AAA.A.

Staff call via the MFC staff call cable.

The user must check the optical and audible alarm during the
switch-on test. An alarm must be activated to test the staff call
alarm of the Infusomat P, e.g. open pump cover during operation.

Pump Unit

The pump head is driven by a stepper motor. Each full step of the
motor is realized by 5 microsteps. The motor is driven by an FET
output stage. The function processor controls the motor with the
MS signal. A slot disc which is mounted on the pump head axle is
scanned by two light barriers (PKS1 and PKS2 signal). Thereby the
control microprocessor monitors the direction of rotation and
speed of the pump head.
The pump head position is also determined with the PKS2 signal.
The motor can therefore be accelerated during the withdrawal
phase. Thus a nearly pluse-free flow is realized in the lower delivery range (<100 ml/h). The total pump head cycles and running
time are available in the service program under history data.

Infusomat P 2.0gb

1- 7

System Overview

Mechanical Occlusion Pressure:


The Infusomat P has a linear peristaltic pump. This pump has 12
slides which are driven by a camshaft.
When the pump cover is closed, the pump tube is squeezed (occlusion) by at least one of the slides, independent of the pump
head position. The complete pump unit is mounted behind the
front panel in the cabinet frame. The hinges and the locking bow
for the pump cover are led through the front panel. The pump cover is automatically closed when the operating unit door is closed.
The slides are pressed against the pump cover by a spring system
in the pump unit. Thereby a delivery pressure is realized and mechanically limited by the springs.
If the pressure limit is exceeded there is no volume delivery. The
drop sensor activates an alarm. If one of the springs fails, the
spring system will ensure that an unsafe condition cannot occur
(free flow). The two remaining springs ensure an appropriately
high occlusion pressure.
Electronic Occlusion Pressure:
The electronic occlusion sensor is mounted on the output side of
the pump. A spring pressure loaded slide is seated on the infusion
line. An increase of pressure in the infusion line leads to a deflection of the coil core via the pressure slide. The depth of immersion
is measured inductively. When a preset pressure threshold is
reached the pump drive is switched off, and an alarm is activated.
The electronic occlusion pressure is a single channel circuit. In
case of a failure, the mechanically limited maximum pressure can
be reached.
Motor Switch-Off by Both Processors:
Function processor: MS signal to switch-off the motor drive. Control processor: MOTEIN signal to switch-off the drive of the
motor operating voltage.

Computer Interface

The Infusomat P is equipped with a computer interface. It can be


connected to the optical interface or via the MFC service plug. To
activate the computer operation please ask for a detailed description from B. Braun.
Until software version IFPC: DIANET
From software version IFPe, IFPE DianetStar

1- 8

Infusomat P 2.0gb

System Overview

Braun fluid manager system (fm system)

The Infusomat P can be operated as a stand-alone unit or integrated in an intensive care unit, e.g. the B. Braun fluid manager
system (fm system). It is integrated by simply snapping the unit
into the system.
Mains supply and data communication are automatically connected. Thereby data acquisition and transmission to higher computer system levels are possible.

Internal Assignment

Fig.: 1 - 5

Infusomat P 2.0gb

1- 9

System Overview

Accessories

General
Designation

Ord. No.

Mounting clip for drop chamber TK 2000 . . . . . . . . 3477 3223


Mounting clip for drop chamber Intrafix P . . . . . . . 3477 3215
Universal adapter for drop sensor . . . . . . . . . . . . . . . . 0871 1747
Drop sensor, complete . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 578A
Short infusion pole. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0870 1644
Drop chamber holder . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3088
Mains lead (200-240V~) . . . . . . . . . . . . . . . . . . . . . . . . 3450 2718
Mains lead (100-120V~) . . . . . . . . . . . . . . . . . . . . . . . . 3450 5423
Mains lead USA / CAN . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5393
Universal Clamp
Universal clamp, complete. . . . . . . . . . . . . . . . . . . . . . . 3450 5857
Pole Clamp (rotating) . . . . . . . . . . . . . . . . . . . . . . . . . . 0871 8482

1 - 10

Infusomat P 2.0gb

2-6

2-

Software Update

Designation

Software

2
Ord. No.

Update-Kit IFPC02002 . . . . . . . . . . . . . . . . . . . . . . . . . 3450 6462


Update-Kit IFPe02003 . . . . . . . . . . . . . . . . . . . . . . . . . . On request
Update-Kit IFPE13003 . . . . . . . . . . . . . . . . . . . . . . . . . On request
(only units with controller
board with loudspeaker)
MFC interface line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0871 1661
The higher digit always replaces the lower digit for the revision
level, e.g. IFPE13003 replaces IFPE13002.
Position

1 2 3 4 5 6 7 8 9

Digit

F P E 1 3 0 0 3

Revision Level
Hardware Identification
Software Group

When the software group changes (IFPE13002) the unit functions


are changed, too. Therefore unit users must be informed (e.g. instruct the user and exchange the instructions for use software
coding, e.g. IFPE is on the cover page of the instructions for use.)
Note
Mark the unit after having updated the software! The new software version must be clearly recognizable.

Infusomat P
Fig.: 2 - 1 Coding

Only update from old to new software versions, never in reverse


order (e.g. never update from IFPC02002 to IFPC02001!).
All units used in one ward should have the same software status
and basic setup to avoid operator mistakes.
Note
Software updates must be reported to B. Braun for registration.
Observe the notes of the update program and the supplements!

Infusomat P 2.0gb

2- 1

Software

Approved Software Versions

IFPC02001
-

Basic software
(Must not be used any more. Please contact the Technical
Service of B. Braun).

IFPC02002
-

No optical staff call after power up (fluid manager system)

Optimized air sensor evaluation

Manual bolus volume limited to 99.9 ml

Optimized special bolus function

IFPe02003
-

Preselected volume and time counted down to 0

Long-time compensation of delivery rate

Interface changed to DIANET Star


(not compatible with Dianet)

New special function Piggyback

New special function clock

Storage of alarms in case of malfunctions, which can be recalled in the service program menu 230

Tube selection menu

Calibration occlusion sensor

Tube update

IFPE13002
Only controller board with loudspeaker (see Controller Board
pg. 4 - 6).
-

Alarm volume can be set

History function

IFPE13003
-

2- 2

Optimized dose calculation

Infusomat P 2.0gb

Software

Error Messages and Alarms


7 Segment Display

Alarms of the function processor 80c535 are displayed in the text


field of the LCD. Alarms of the control processor 68HC11 are displayed in the 7 segment display. The alarms help to troubleshoot
unit malfunctions. As not all malfunctions can be considered, unit
malfunctions with different messages, which are not listed, can
be displayed, or there may be no message.
Detected unit alarms are displayed in the text field as Unit
Alarms in the selected language. Additionally the error number is
displayed in the text field.

Text Field
Fig.: 2 - 2
Text Field

Description

100

defective RAM memory

101

UMOT cannot be switched on

102

UMOT still switched on despite overvoltage

103

UMOT still switched on despite MOTEIN=0

104

UMOT still switched on despite undervoltage

105

ON/OFF key pressed longer than 14 sec

106

defective air sensor (calibration value?)

107

defective program memory

108

defective program flow

109

different number of pump head cycles

110

different keypad gaps between 80c838 and 68hc11

111

different program versions between 80c535 and 68hc11

112

defective program flow

113

testbit!=0 out of switch-on test

116

defective program memory - text

117

defective program memory - text does not match with program

118

reset during active operation

119

defective program memory transit time ROM-test

120

defective program memory tube parameters

Table 2 - 1

Infusomat P 2.0gb

2- 3

Software

Control Microprocessor 68hc11


FFxx is displayed in the 7 segment display with flashing dots. FFxx
is the error code.

7 Segment Display Description


FF01

dummy for test

FF02

battery not present / missing battery current

FF03

defective RAM memory

FF04

defective program memory - ROM test error 1

FF05

defective program memory - ROM test error 2

FF06

calibration data error from EEPROM

FF07

pump head cycle not plausible

FF08

failure / inaccuracy of system clock

FF09

failure 100msec system clock

FF10

reset during active operation

FF12

no dynamic pressure sensor signal (EDB)

FF14

defective temperature sensor

FF16

defective membrane keypad (from IFPe02003)

FF17

defective program memory tube parameters

Table 2 - 2

2- 4

Infusomat P 2.0gb

Software

Software Default Values

Unit No.: _______________________

Menu Point
Standard Function

Special Functions

Customer Setting

Depending on Art. No.

_______________________

Alarm Type

Single stage

_______________________

Staff Call

Static without OFF Alarm, without _______________________


switch-on pulse

Ward Identification

Ward Identification

_______________________

Drug 1 ... 9

Drug 1 ... 9

_______________________

Operating Alarms

_______________________

Minimum Delivery Rate

0.1 ml/h

_______________________

Maximum Delivery Rate

999.9 ml/h

_______________________

Maximum Air Rate

1.5 ml/h

_______________________

Maximum Air Bubble

0.30 ml

_______________________

Dose Calculation

Deactivated

_______________________

Bolus

Deactivated

_______________________

Standby

Activated

_______________________

Deactivated

_______________________

Deactivated

_______________________

Occlusion Pressure

Activated

_______________________

Drop Control

Deactivated

_______________________

Piggyback

Deactivated

_______________________

Battery Capacity

Deactivated

_______________________

Data Lock

Deactivated

_______________________

Deactivated

_______________________

Deactivated

_______________________

Activated

_______________________

Interval Bolus Dose

Off

_______________________

Online Rate Setting

On

_______________________

Double Rate Entry

Off

_______________________

Off

_______________________

Occlusion Pressure

High

_______________________

Contrast

Optimum contrast

_______________________

CC Address

_______________________

Drug

Blank

_______________________

Data Lock

Off

_______________________

Standby Time

24h 00min

_______________________

CC Mode

2)

1)

Alarm Tone

1) 3)

Contrast
Clock

1)

Tube Selection
User Data

Default

User Language

Drug Selection

SM Menu

2)

1)

1)

From software IFPE


No longer available in software IFPe, IFPE
3) Not software IFPe
2)

Infusomat P 2.0gb

2- 5

Software

Menu Point
Drop Control

_______________________

Stage 7

_______________________

60sec

_______________________

Bolus Key

On

_______________________

Bolus Rate

999.9 ml/h

_______________________

Air Sensor Calibration Value


Tube group I
(dependent on tube)

130 mV

Must not be changed

Air Sensor Calibration Value


Intrafix Air P
(dependent on tube)

130 mV

Must not be changed

Scale Factor
(dependent on tube)

50 (Intrafix Air)

_______________________

Occlusion Level, Low

_______________________

Occlusion Level, High

12

_______________________

History Function

Activated

_______________________

DIANET Type No.

Depending on unit

_______________________

Unit No.

Depending on unit

_______________________

Operating Hours

Depending on unit

_______________________

Battery Hours

Depending on unit

_______________________

Number of Pump Head Cycles

Depending on unit

_______________________

1) 3)

Dianet Mode Display

Unit Specific Data

Customer Setting

On

Alarm Tone

Calibration Data

Default

1)

1)

From software IFPE


No longer available in software IFPe, IFPE
3) Not software IFPe
2)

2- 6

Infusomat P 2.0gb

3 - 12

3-

Service Program

Structure of the Service Program

Service Plug,
Inserted

Switch On

1)

Short Display:
software version,
user language
Standard Operation
(see operating
flow diagram)

Tube Type Selection2)

Special Functions
(see operating
flow diagram)
Service Program
Activated
Group:
Unit Data

Group:
History Data

Group:
Test

Group:
Unit Modification

Group:
Calibration

Software Version
100.0

Operating Hours
200.0

Air Sensor
300.0

Service Language
400.0

Occlusion Sensor
500.0

Drug
110.0

Battery Op. Hours


210.0

Occlusion Sensor
310.0

User Language
410.0

Scale Factor
510.0

Ward Identification
120.0

Pump Head
Cycles
220.0

Temperature Sensor
320.0

Alarm Tone
420.0

Air Sensor
520.0

Staff Call
430.0

Comp. Head Volume1)


530.0

Serial Number
140.0

Special Function
440.0

Occlusion Level
540.0

DIANET Type No.


150.0

Menu
450.0

Pump Data
550.0

Delivery Rate
460.0

History 3)
560.0

Tube Type 1)
130.0

Alarms
230.0

Air Alarm
470.0

1)

Only software IFPC

2)

From software IFPe / IFPE

3)

From software IFPE

Dianet Mode
Display 2)
480

Fig.: 3 - 1

Infusomat P 2.0gb

3- 1

Service Program

Additional Functions with


Inserted Service Plug

Software Version and User Languages


1.

Insert service plug into the MFC socket at the rear of the unit.

2.

Switch on unit and keep the ON/OFF button pressed


(for max. 15 s).

3.

The software version and user languages are displayed in the


LCD.

4.

The unit is switched on when the ON/OFF button is released.

5.

** appears in the LCD if the service plug is inserted.

Two LEDs are integrated in the service plug:


-

green = Power supply active

red = Alarm

The following conditions are active:


-

The operating alarms are muted.

All special functions are accessible (including the disabled).

The special functions are slightly modified. (Example: SM


battery capacity has key for 0 min/32 min presetting).

The battery capacity display switches between the nominal


and the actual capacity. If the maximum battery capacity is
not reached ? will be displayed before the new nominal capacity.

The tube type will be displayed in the tube selection menu.

Disabling the Occlusion Sensor


The electronic pressure monitoring can be deactivated to check
the mechanical occlusion pressure.
1.

Select Occlusion Pressure with the SM key.

2.

Then select mechanical.

3.

Return to the main menu with END.

A too low pump speed is indicated in the display by Pressure


Alarm with underlined stars.
The display "Pressure Alarm" with underlined questions marks indicates that no tube has been inserted. 1)

1) From software IFPe / IFPE


3- 2

Infusomat P 2.0gb

Service Program

Start / Quit the Service Program

Activate the Service Program


1.

Insert service plug into the MFC socket at the rear of the unit.
- ** appears in the display.

2.

Select Service Program with the SM key. When the service


program is activated the red alarm LED flashes. The LED displays the code number of the selected group and function.

Fig.: 3 - 2
FUNCTION
END
GR+
FU+
OK
Fig.: 3 - 3

Jumps to the initial function


Selects group
Selects function in the
activated group
Activates the selected function or if
necessary skips to the sub-functions
with NEXT

Quit the Service Program


1.

Press END in the main menu. - A data storage query is activated: Save Changes? Yes / No.
Y / N terminates the service program.
Press END to jump to the last function.

2.

Switch off the Infusomat P and remove service plug.

Note
Disconnect the Infusomat P from mains for at least 30 seconds after termination of the service program (memory is deleted). Then
the unit can be switched on again.

Unit Data

Infusomat P 2.0gb

Software Version

Function 100.0

1.

Select sub-functions with NEXT.

2.

The current software version is displayed in the LCD:


- User program version with date
- Language with text version. Further languages with (+).
- Service program version
- Service language with text version
- Tube type version

3.

Return to the initial function with END.

3- 3

Service Program

Drug Name

Function 110.0

Memory for maximum 9 drugs and 20 characters per name.


1.

Display the stored drug names with the NEXT key.

2.

Delete displayed entry with CLR.

3.

Press YES to modify a drug name:


Move cursor to character with NEXT.
Select new character from line 3 with << or >>.

4.

Repeat the procedure for each character.

5.

Return to the initial function with END.

Fig.: 3 - 4

Ward Identification

Function 120.0

Enter and display of a ward specific unit identification. Permanent


display if the unit is connected to mains and switched off.
1.

Delete displayed entry with CLR. Press YES to enter modifications:


Move cursor to character with NEXT.
Select new character from line 3 with << or >>.

2.

Repeat the procedure for each character.

3.

Return to the initial function with END.

Tube Type 1)

Function 130.0

There is only one tube type implemented in version IFPC02002.


This function is for future extensions.
Serial Number

Function 140.0

The displayed serial number must correspond with the number on


the unit type plate, as this number is used in the interface mode.
1.

YES activates the entry mode. Enter via the numeric keypad.

2.

YES stores the changed or new number.

3.

Return to the initial function with END.

DIANET Type Number

Function 150.0

The displayed serial number must correspond with the number on


the unit type plate, as this number is used in the interface mode.
1.

YES activates the entry mode. Enter via the numeric keypad.

2.

YES stores the changed or new number.

3.

Return to the initial function with END.

1)

3- 4

Only software IFPC


Infusomat P 2.0gb

Service Program

History Data

Operating Hour Counter


1.

OK activates the display.

2.

Return to the initial function with END.

Battery Operating Hours


1.

OK activates the display.

2.

Return to the initial function with END.

Pump Head Cycles

Function 200.0

Function 210.0

Function 220.0

Display of the pump head cycles (delivered volume).


1.

OK activates the display.

2.

Return to the initial function with END.

Operating Alarms

Function 230.0

Bit

15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0

The last 20 operating alarms can be recalled.

Code

1 0 0 0 0 0 0 0 0 0 0 0 0 0 1 0

They are displayed as 16 bit binary codes and each bit position can
be set from 0 to 1.

Example for Alarm Pump Cover Open


0

Battery empty (battery alarm)

Pump cover open

Drop alarm

Air alarm

Occlusion alarm

Expired standby time

Interface alarm

KOR end alarm

8 to 14

free

15

Operating alarm

1.

OK activates the alarm display.

2.

Display operating alarms -01 to -20 with the (+) and (-) key.

3.

Delete the operating alarms with CLR.

4.

Return to the initial function with END.

In addition the last unit alarm (GA) will be displayed from software version IFPe, IFPE.

Fig.: 3 - 5

Infusomat P 2.0gb

3- 5

Test

Service Program

Air Sensor

Function 300.0

After exchange check the function of the air sensor.


See TSI-List for permissible check values (see Technical Safety
Inspection TSI pg. 7 - 1).
1.

Press OK. The received signal amplitude is displayed as a


measurement value.
(The test value with NEXT is not important).

2.

Insert an infusion line filled with air and check the maximum
permissible air value.

3.

Insert an infusion line filled with fluid and check the minimum permissible water value.

4.

Return to the initial function with END.

Pressure Sensor

Function 310.0

Test Equipment
2.2 mm template
(see Test Equipment and Special Tools pg. 9 - 1)
1.

Push in the bottom slide of the finger pump.

2.

Press OK twice.

3.

Open unit door.

4.

Note the measurement value (actual value).

5.

Insert the 2.2 mm template and close the unit door.

6.

The new measurement value will be displayed. It must be 15


to 25 digits higher than the first value.

7.

Return to the initial function with END.

If the 15 to 25 digits are not reached, the pressure sensor unit


must be mechanically aligned (see Occlusion Sensor
pg. 4 - 15).
Temperature Sensor

Function 320.0

Display of the measured temperature of both processors. Deviations are possible, the display is only for internal use.

3- 6

1.

OK activates the display.

2.

Return to the initial function with END.

Infusomat P 2.0gb

Service Program

Unit Modifications

Service Language

Function 400.0

English or German can be selected.


1.

OK activates the function.

2.

Select language with NEXT.

3.

Acknowledge with YES.

4.

Return to the initial function with END.

User Language

Function 410.0

Four user languages per language group are available (depending


on software).
1.

OK activates the function.

2.

Select language with NEXT.


The language no. and text version are displayed.

3.

Acknowledge with YES.

4.

Return to the initial function with END.

Alarm Tone

Function 420.0

Different alarm modes can be selected:


-

Single stage

A 10 minutes off alarm can be selected.


In this mode the audible alarm is activated with a delay of 10
minutes.

The activation of the 10 minutes off alarm is only permissible, if:

Infusomat P 2.0gb

the staff call is connected and

the Infusomat P has an attention label (label drawing no.


M007100000F04).

1.

OK activates the function.

2.

Select alarm tone with NEXT.

3.

Acknowledge with YES.

4.

Return to the initial function with END.

3- 7

Service Program

Staff Call

Function 430.0

Select first, whether the switch-on pulse shall be activated (YES/


NO). The switch-on pulse is used to test the staff call line. Different staff call modes can be selected with the NEXT key:
-

Dynamic with OFF Alarm

Dynamic without OFF Alarm

Static without OFF Alarm

For further details see staff call line in the instructions for use.
An additional switch-on pulse (YES/NO) can be activated for each
mode to test the staff call unit.

Fig.: 3 - 6

1.

OK activates the function.

2.

Select staff call type with NEXT.

3.

Acknowledge with YES.

4.

Return to the initial function with END.

Special Functions

Function 440.0

Special functions can be activated in the service program, which


are then available on the user interface. Deactivated special functions will not be displayed. The SM softkey will not be displayed
in standard operation, if all special functions are deactivated. Special functions to be selected, see Fig.: 1 - 1.

3- 8

1.

OK activates the function.

2.

Select special functions with NEXT.

3.

Activate / deactivate the respective function with YES/NO.

4.

Return to the main menu with END.

Infusomat P 2.0gb

Service Program

Menu

Function 450.0

The availability of menus on the user interface can be set.


-

Double rate entry

Online rate entry

Interval bolus

Tube selection

1.

OK activates the function.

2.

Activate / deactivate the decimal function with NEXT.

3.

Acknowledge with YES.

4.

Return to the main menu with END.

Delivery Rate

Function 460.0

The maximum and minimum delivery rates can be set. Range of


adjustment of the delivery rate: 0.1 to 999.9 ml/h

1.

OK activates the function.

2.

Select the min./max. delivery rate with NEXT.

3.

Acknowledge with YES.

4.

Enter value with a numeric key.

5.

Acknowledge with OK.

6.

Return to the initial function with END.

Air Alarm

Function 470.0

The air sensor sensitivity of the air rate alarm in ml/h (total air
alarm) and of the maximum air bubble in ml (single bubble) can
be adjusted.
Setting range air rate: 0.5 to 3.5 ml/h
Setting range air bubble: 0.01 to 0.3 ml/h

Infusomat P 2.0gb

1.

OK activates the function.

2.

Select air rate (ml/h) or air bubble (ml) with NEXT.

3.

Acknowledge with YES.

4.

Enter value with a numeric key.

5.

Acknowledge with OK.

6.

Return to the initial function with END.

3- 9

Service Program

Dianet Mode Display

Function 480.0

When operated with DianetStar the respective DianetStar-mode


(CA, CC, CD) with address 01, e.g. mode CA and address 01 is displayed by: ###CA01###.
The duration of the display after the last data transmission can be
set between 0 and 255 seconds.

Calibration

All safety-relevant parameters are set by the manufacturer. If


these parameters are changed, a new calibration must be performed with calibrated test equipment.
Occlusion Sensor

Function 500.0 *

Measurement and setting of the zero value of the occlusion sensor, Ord. No.: 3450 7353.
1.

Press the OK key twice.

2.

Press the Align key.

3.

Open the unit door and remove tube.

4.

Check the occlusion sensor for cleanliness and function


(briefly operate the sensor and check whether the displayed
measurement value changes).

5.

Wait until the displayed measurement value does not change


any more (+/- one digit). The measurement value must be between 150 and 220 digits.

6.

Acknowledge the measurement value with YES.

7.

Return to the initial function with END.

It is not allowed to enter the zero point value via the keypad.
Occlusion Sensor

Function 500.0 **

Measurement and setting of the zero value "3" of the occlusion


sensor, Ord. No.: 3452 0589. Calibration (see Occlusion Sensor
pg. 4 - 15).
1.

Press the OK key twice.

2.

Press the Align key.

3.

Open the unit door and remove tube.

4.

Check the occlusion sensor for cleanliness and function


(briefly operate the sensor and check whether the displayed
measurement value changes).

Pay attention to the indications in Table 4 - 1 Compatibility List pg. 4 - 6).


* Until software IFPC
** From software IFPe / IFPE
3 - 10

Infusomat P 2.0gb

Service Program

5.

Wait until the displayed measurement value does not change


any more (+/- one digit). The measurement value must be between 250 and 400 digits.

6.

Acknowledge the entry with YES.

7.

Return to the initial function with END.

It is not allowed to enter the zero point value via the keypad.
Scale Factor

Function 510.0

The scale factor can be set in the limits between 40 and 99 digits
for each tube type. Every digit step is equivalent to a 0.5 % modification of the delivery rate. An increase of the scale factor reduces the pump speed, and a decrease increases the pump speed.
1.

Press the OK key twice to activate the function.

2.

Select the tube type with the NEXT key.

3.

"Yes" activates the function for the selected tube type.

4.

The value can be changed with the entry keypad.

5.

Acknowledge the entry with OK.

6.

Return to the initial function with END.

7.

Quit the service program and save changes with YES.

8.

Switch on unit and check the delivery rate (see Delivery Accuracy pg. 8 - 4).
If necessary repeat the delivery rate measurement.

Air Sensor

Function 520.0

Alignment or check of the air sensor value (alarm threshold) (see


Air Sensor pg. 4 - 18).
Note
The value is dependent on the tube type.
1.

Press OK twice and then YES to activate the function.

2.

Press OK again to activate the air sensor value.

3.

The value can be changed with the entry keypad.

4.

Acknowledge the entry with YES.

5.

AIR SENSOR IS SET acknowledges the entry.

6.

Return to the initial function with END.

7.

Quit the service program and save changes with YES.

Compensation Head Volume

Function 530.0 *

Only for test purposes at the manufacturers. The compensation of


the head volume must always be activated.

Infusomat P 2.0gb

3 - 11

Service Program

Occlusion Level

Function 540.0

Setting of the occlusion threshold values for the occlusion levels


"low" and "high".
1.

Press the OK key twice to activate the function.

2.

Select the occlusion level with NEXT.

3.

YES activates the entry mode.

4.

Enter the pressure value via the numeric keys (pay attention
to the displayed limit values!).

5.

Press the YES key to take over the settings.

6.

Return to the initial function with END.

Pump Data

Function 550.0

Note
Compatibility between controller board and pump must be
checked before acknowledging. Only press YES if pump corresponds to controller board.
History

Function 560.0 *

The history function can be activated or deactivated.


1.

Press the OK key twice to activate the function.

2.

Select the history function with NEXT.

3.

ON activates the function. The history protocol memory is initialized (the actual software version and the serial number
are registered).
Press the OFF key to deactivate the function. The history protocol memory is deleted.

4.

Return to the initial function with END.

* Only software IFPC


* From software IFPE
3 - 12

Infusomat P 2.0gb

4 - 20

4-

4.1

Mains Fuses

Unit Elements

Designation

4
Ord. No.

Fuse T 0.16 A for 200 / 230 / 240 V


(10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2847
Fuse T 0.315 A for 100 / 110 / 120 V
(10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 0534
Fuse holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 0979
Note
Only use recommended fuses.
Exchange
1.

Press the expansion clamps at the fuse holder on the recessed


mains plug with a screw driver in direction of the arrows and
pull out fuse holder.

2.

Replace blown fuse and press in fuse holder. Only use recommended fuses.

Check
Safety check, functional check.

4.2

Battery

Designation

Ord. No.

Battery incl. connector 1.2 AH / 7.2 V and holder. . . . 3450 2556


Battery

M4x8
Exchange
1.

Switch off unit and disconnect from mains.

2.

Loosen screw, open the cover of the battery compartment and


remove battery.

3.

Pull off the battery connector.

4.

Assembly is done in reverse order.

5.

After having exchanged the battery the Infusomat P must be


connected to mains, before switching on the unit. Thereby
the charge and discharge currents are aligned.

6.

Charge battery (16 h).

Fig.: 4 - 1
Note
Defective batteries must be orderly disposed of, e.g. send back to
B. Braun (see Return of Spare Parts pg. 0 - 7).
Check
Perform switch-on test in battery operation and check the battery
running time, if necessary.

Infusomat P 2.0gb

4- 1

4.3

Unit Elements

Door Lock

Designation

Ord. No.

Door lock complete with push button . . . . . . . . . . . . . 3450 5601


Spring holder for door lock . . . . . . . . . . . . . . . . . . . . . . 3450 5440
Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790

S M 3x6

Exchange
1.

Open door and unlatch the spring holder.

2.

Remove coutersunk screw and press out the mounting by


pressing the holder for the door lock.

3.

Remove door lock in an upward direction and exchange.

4.

Assembly is done in reverse order.

Check
Pump unit check (only mechanical occlusion pressure).

Spring Holder

IP 013

Fig.: 4 - 2

4- 2

Infusomat P 2.0gb

Unit Elements

4.4

Pump Cover

Designation

Ord. No.

Pump cover with lock . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1916


Blind plug 7.1 mm (10 pcs.) . . . . . . . . . . . . . . . . . . . . . 3477 3207

Hinge Pin for


Pump Cover

Torsion spring in lever/pump cover (5 pcs.) . . . . . . . . 3477 3363


Torsion spring for pump cover (5 pcs.) . . . . . . . . . . . . 3477 3355
Lever (pump cover). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4092
Hinge pin for pump cover . . . . . . . . . . . . . . . . . . . . . . . 3477 3967
Hinge pin for pump cover lever. . . . . . . . . . . . . . . . . . . 3450 5725
Exchange
Tools: Pin punch 1.8 mm, pin punch 6 mm, 2.2 mm template

Pump Cover
Hinge Pin for
Pump Cover Lever

1.

Open door and remove hinge pin with pin punch (1.8 mm)
from below. Do not lose the torsion spring.

2.

Disassemble pump cover.

3.

Insert the torsion spring in new pump cover. Insert the hinge
pin from the top with a pin punch (6 mm).

4.

Check the occlusion sensor with 2.2 mm template in the service program, and align, if necessary.

Check
Torsion Spring
IP 014

Safety check, pump unit check.

Fig.: 4 - 3

Infusomat P 2.0gb

4- 3

4.5

Unit Elements

Pump Housing

Designation

Ord. No.

Pump housing, (cpl.). . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3390


Exchange
1.

Remove 4 tamper-proof caps by piercing a screwdriver


through the caps.

2.

Loosen the screws.

3.

Remove the pump housing.

Note
Do not use the pump housing again, but replace it by a new one.

Pump Housing
Tamper-Proof Caps
IP 038

Fig.: 4 - 4

4- 4

Infusomat P 2.0gb

Unit Elements

4.6

Housing and Handle

Designation
Housing Labelling
German . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
French . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Italian. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Danish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Norwegian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Swedish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Finnish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Portuguese. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Czech . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Polish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Turkish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Ord. No.
3450 1851
3450 1851
3450 1851
3450 1851
3450 3382
3450 3382
3450 3404
3450 3404
3450 3404
3450 3404
3450 3382
3450 3358
3450 3358
3450 3382

Foot stand complete with rubber feet . . . . . . . . . . . . . 3450 5415


Rubber feet (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3096
Unit handle with O-rings and PT screws . . . . . . . . . . . 3450 3450
Exchange

O-Ring
PT5.0x16

1.

Remove battery (see Battery pg. 4 - 1).

2.

Loosen 4 screws from the foot stands and remove foot stands.

3.

Remove the safety seal from the rear panel, break tamperproof cap and remove screw.

4.

Slidly widen the sides at the bottom of the housing and pull
off to the top.

5.

Lay the unit on the front side and assemble in reverse order.

6.

Safety seal the rear panel screw after functional check.

Check
Safety check.

4 x M4x8
Fig.: 4 - 5

Infusomat P 2.0gb

4- 5

4.7

Unit Elements

Controller Board

Designation

Ord. No.

Distance sleeve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3366


Cable Output Motor

Air Sensor Connector


Motor Connector
Flex Cable LCD Module
Occlusion Sensor
Connector
Rear Panel
Connector

Loudspeaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8848


Controller board with buzzer, raw material No. 3810 7660,
software IFPC, can be updated to IFPe, if desired.
Language Group*

New Part

Exchange

Ord. No.:3450 2041 Ord. No.:3488 0895

Ord. No.:3450 2050 Ord. No.:3488 0909

Ord. No.:3450 2068 Ord. No.:3488 0917

Ord. No.:3450 2076 Ord. No.:3488 0925

Ord. No.:3450 2084 Ord. No.:3488 0933

Controller board with loudspeaker, volume control and history


function, raw material no. 3810 8003,
software IFPE:

Reed Sensor
Connector

Fig.: 4 - 6

Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3447

Battery Connector
Loudspeaker / Buzzer
Press snap-in pin together

Language Group*

New Part

Exchange

Ord. No.:3450 8767 Ord. No.:3488 0941

Ord. No.:3450 8775 Ord. No.:3488 0950

Ord. No.:3450 8783 Ord. No.:3488 0968

Ord. No.:3450 8791 Ord. No.:3488 0976

Ord. No.:3450 8805 Ord. No.:3488 0984

CAUTION
The controller boards (raw material no. 3810 7660 and 3810 8003)
are not compatible.

Controller Board
Raw Material No.

Software

Buzzer

Occlusion Sensor Calibration Device Finger Pump, cpl.


Order No.
Order No.

3810 7660

IFPC

Buzzer

3450 7353

no

3450 1746

3810 7660, with update

IFPe

Buzzer

3452 0589

yes

3452 0597

3810 8003

IFPE

Loudspeaker

3452 0589

yes

3452 0597

Table 4 - 1 Compatibility List

* Language Group:
A German, French, Dutch, Italian
B English, Dutch, Spanish, English
C Danish, Norwegian, Swedish, Finnish
D Spanish, Portuguese, English, Turkish
E Czech, Polish, Hunguarian, English
4- 6

Infusomat P 2.0gb

Unit Elements

If the controller board (raw material no. 3810 7660) is to be replaced by the new controller board (raw material no. 3810 8003)
the occlusion sensor (3451 0589) has also to be installed (see
Occlusion Sensor pg. 4 - 15). An exchange in reverse order
is not permitted.
Exchange
1.

Remove battery (see Battery pg. 4 - 1).

2.

Dismount housing (see Housing and Handle pg. 4 - 5).

3.

Press snap-in pin together at the distance sleeves and remove


the board carefully.

4.

Pull off connector (see figure).

5.

Exchange the board or buzzer.

6.

Assembly is done in reverse order. Be careful with the optical


components. Insert the board into the lower guide parallel to
the base plate. (Otherwise problems with the optical interface can occur.)

Note
If Calibration Defective is displayed after having exchanged the
controller board, check that the correct board (risk of mix-up) was
assembled.
Check
Safety check, functional check.

Infusomat P 2.0gb

4- 7

4.8

Unit Elements

Rear Panel

Designation

Ord. No.

Rear panel with screws (M3) and seal . . . . . . . . . . . . . 3450 1860


Cover for optical interface. . . . . . . . . . . . . . . . . . . . . . . 3477 3164
Strip seal for rear panel . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3142
MFC connector board . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3374
Potential equalization bolt . . . . . . . . . . . . . . . . . . . . . . 3477 0550
fm recessed plug (3 pin) . . . . . . . . . . . . . . . . . . . . . . . . 3477 3177
Screw 30x8 for fm recessed plug (20 pcs.). . . . . . . . . . 3477 3185
U Washer 3.2 (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3193
Mains module 200/230/240 V . . . . . . . . . . . . . . . . . . . . 3450 1886
Mains module 100/110/120 V . . . . . . . . . . . . . . . . . . . . 3450 1894
Mains module 230 V . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 190A
Drop sensor socket incl. cable and plug . . . . . . . . . . . . 3450 1878

Fig.: 4 - 7

Pin
No.

Signal Name

Pin 1 Uext-

Function
Input of external supply voltage,
connection of shield

Pin 2 not assigned

Fig.: 4 - 8

Pin 3 GND

Reference level 0V

Pin 4 Staff call

Output open collector or standard


74HC level each with 220 -series
resistance

Pin 5 Ub

Output supply voltage

Pin 6 MFC-KAD

Analogue input K

Pin 7 Emergency Off

Input interface

Pin 8 MFC-FAD

Analogue input F

Pin 9 Tx

Transmit line interface

Pin 10 STB

Output with 470 -series resistance, activation of MFC circuit

Table 4 - 2 MFC Pin Assignment


4- 8

Infusomat P 2.0gb

Unit Elements

Pin
No.

Signal Name

Function

Pin 11 Rx

Receive line interface

Pin 12 Uext+

Input of external supply voltage

Table 4 - 2 MFC Pin Assignment


Exchange Rear Panel
1.

Remove battery (see Battery pg. 4 - 1).

2.

Dismount housing (see Housing and Handle pg. 4 - 5).

3.

Pull off the rear panel connectors and loosen both screws on
the unit bottom.

Rear Panel

M4x8
Fig.: 4 - 9

Infusomat P 2.0gb

4- 9

Unit Elements

Exchange MFC Connector Board


Tools: Special socket spanner M18
1.

Remove nuts at the fm recessed plug (red/blue).

2.

Pull off the connector to the mains power supply and drop
sensor, see Fig.: 4 - 7.

3.

Loosen the MFC socket with special socket spanner M18.

4.

Assembly is done in reverse order.

M18

Fig.: 4 - 10
Exchange fm Recessed Plug
Cable, Red

1.

Cable, Blue

Loosen screws and nuts.

PT 3x8
Note
The new recessed plug must be slightly moveable after assembly.
Exchange Potential Equalization Bolt
1.

Remove nuts with ring spanner.

2.

Exchange the bolt.

2x M4 each
2x M6
Fig.: 4 - 11

4 - 10

Infusomat P 2.0gb

Unit Elements

Exchange Mains Module


Power Supply Unit

Tools: Screw driver Torx T10


1.

Loosen 3 screws and remove the power supply unit.

Note
The voltage selection for switchable power supplies is at the fuse
element.
Exchange Drop Sensor Socket

3 X Torx M3x10

Fig.: 4 - 12

1.

Loosen nut and exchange the drop sensor socket.

2.

Safety lock with Sicomet 50.

3.

Make ground connection.

Check
Safety check, functional check.

4.9

Front Panel

Designation

Ord. No.

Front panel without clamp lever and torsion spring . . 3450 2092
Snap-In Pin
(Rear Side)

Circular seal 45 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3126


Tamper-proof caps 10 mm (50 pcs.) . . . . . . . . . . . . . . 3477 3134
Cover 6.4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3412
Clamp lever with torsion spring and pin 4x20 . . . . . . 3450 3420
Torsion spring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3439
Reed sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1754
Exchange
Tools: Pin punch

Tamper-Proof Cap

Reed Sensor
(Rear Side)
Snap-In Pin
(Rear Side)
IP 023

Fig.: 4 - 13

Infusomat P 2.0gb

1.

Remove battery (see Battery pg. 4 - 1).

2.

Dismount housing (see Housing and Handle pg. 4 - 5).

3.

Disassemble pump cover (see Pump Cover pg. 4 - 3).

4.

Disassemble pump housing (see Pump Housing


pg. 4 - 4).

5.

Remove countersunk screws by piercing a screwdriver


through the caps to loosen the countersunk screws.

6.

Press reed sensor out of the holder on the rear side, see Fig.:
4 - 14.

7.

Press both snap-in pins inwards (rear side top and bottom).
Then remove the front panel to the front.

8.

Exchange the flow inhibitor, i.e. pull straight pin with M3


screw. Prior to assembly slightly grease the plastic surfaces
with Molikote Platislip PG 602 grease.
4 - 11

Unit Elements

9.

Insert circular seal. Assembly is done in reverse order. Snapin pins manually. Insert reed sensor and seat the cables firmly
with cable ties.

Note
Always use a new pump housing.
Check
Safety check, pump unit check.

4 - 12

Infusomat P 2.0gb

Unit Elements

4.10 Pump Unit

Designation

Ord. No.

Note
When replacing the finger pump adhere to Table 4 - 1 Compatibility List pg. 4 - 6).
Finger pump (without motor) including pump, . . . . . 3450 1746
pump cover, seal membrane and occlusion sensor
board, software IFPC
(installed up to unit serial no. 5505)
Finger pump (without motor) including pump, . . . . . 3452 0481
pump cover, seal membrane and occlusion sensor
board, software IFPe, IFPE
(installed from unit serial no. 5506)
Finger pump (without motor and board) . . . . . . . . . . 3452 0597
incl. pump, pump cover,
and seal membrane
Motor with pinion for finger pump . . . . . . . . . . . . . . . 3450 1924
Exchange
M3x6
Motor

1.

Remove battery (see Battery pg. 4 - 1).

2.

Dismount housing (see Housing and Handle pg. 4 - 5).

3.

Disassemble front panel (see Front Panel pg. 4 - 11).

4.

Loosen motor screws, pull off the motor connector and remove motor from pump unit.

Snap-In Pin
Note
Do not loosen the assembly screw on the lower occlusion sensor
board! Otherwise the pump has to be calibrated with the 2.2 mm
template.
5.

Pull off the connector from the controller board.

6.

Press down snap-in pins (from the inside) and tilt the pump
forward and unhinge.

7.

Assembly is done in reverse order.

Pump Unit
CAUTION
Due to the improved occlusion sensor the finger pump to be installed is dependent on the controller board and the software.

Reed Sensor
Cable Tie

Note
Fig.: 4 - 14

Infusomat P 2.0gb

Be sure that the cable of the reed sensor and air sensor do not
hinder the function of the occlusion sensor and the monitoring of
the motor speed (slot disc).
4 - 13

Unit Elements

The complete pump unit (pump without motor, pump cover, membrane, boards) was calibrated by B.Braun. After a complete exchange the unit data and user data must be entered in the service
program again as the data memory is on the pump board:
8.

Enter unit and user data in the EEPROM.

Serial number according to the type plate

DIANET type number according to the type plate


If necessary enter:

Drug name

Ward identification

Alarm tone

Delivery rate min./max.

Air alarm: air bubbles in ml and air rate in ml/h respectively

User language

Special functions (ON/OFF)

Menu

Staff call type


Note

If data is not entered, Calibration Defective may be displayed after the unit is switched on again.
9.

After ending the service program save the data.

The counters for operating hours, battery operation and pump


head cycles are reset to zero when the pump unit is exchanged. If
Calibration Defective is displayed, select the service program
and quit with Save? Yes.
Check
Safety check, pump unit check.

4 - 14

Infusomat P 2.0gb

Unit Elements

4.11

Occlusion Sensor

Test Equipment

Ord. No.

Template 2.2 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0770 5034


Calibration device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0770 501A
Designation

Ord. No.

Note
When replacing the occlusion sensor pay attention to Table 4 - 1
Compatibility List pg. 4 - 6).
Occlusion sensor (cpl.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7353
Software IFPC
(installed up to unit serial no. 5505)
Occlusion sensor (cpl.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3452 0589
Software IFPe, IFPE
(installed from unit serial no. 5506)
(only delivered when a pressure
calibration device is available)
Exchange
Light Barrier
Board

CAUTION
Screw (1)

Occlusion Sensor
cpl.

Plain
Washer

The occlusion sensor unit is mechanically very sensitive. It must


not be dismounted, which means that the occlusion sensor board
and the spring element must not be detached from the occlusion
sensor support (plastic part). Make sure that the movement of the
coil core inside the coil is not hindered.
The occlusion sensor consists of the light barrier board with EEPROM and the occlusion sensor board with holder. Coil core and
bending element are screwed on this holder. Both boards are connected via a flat cable.
1.

Dismount the rear panel (see Rear Panel pg. 4 - 8) or


the pump unit (see Pump Unit pg. 4 - 13).

2.

Pull off the connecting cable to the microprocessor board.

3.

Loosen and remove screw (1) of the light barrier board.

4.

Loosen screw (2) of the occlusion sensor and remove it together with plain washer and serrated lock washer. Then remove the occlusion sensor with occlusion sensor board and
light barrier board.

5.

Mount new occlusion sensor.

Serrated Lock Washer


Screw (2)

Occlusion Sensor Board


Do not loosen screw!

Fig.: 4 - 15

CAUTION
The occlusion sensor to be installed is dependent on the controller
board and the software.

Infusomat P 2.0gb

4 - 15

Unit Elements

Note
Make sure that the occlusion sensor slide can smootly run in the
guides and that the coil core can move inside the coil.
6.

Calibrate occlusion sensor:


a) Call the pressure display in the service program (see
Pressure Sensor Function 310.0 pg. 3 - 6).
b) Loosen hexagon socket screw (M 3x6) on the occlusion
sensor unit (with a 2.5 mm Allen key).
c) Push the occlusion sensor unit with board slightly forward or backward.
d) Tighten the Allen screw.
e) The new measurement value will be displayed (repeat if
necessary until the value is increased by 15 to 25 digits).
f)

After calibration return to the initial function with END.


- Select SAVE NO. (Do not operate the YES key.)

g) Switch off the unit.


7.

Assembly is done in reverse order.

After a complete exchange the unit data and user data must be
entered in the service program again as the data memory is on the
lightbarrier board.
8.

Enter unit and user data in the EEPROM.

Serial number according to the type plate

DIANET type number according to the type plate


If necessary enter:

Drug name

Ward identification

Alarm tone

Delivery rate min./max.

Air alarm: air bubbles in ml and air rate in ml/h respectively

User language

Special functions (ON/OFF)

Menu

Staff call type


Note

If data is not entered, Calibration Defective may be displayed after the unit is switched on again.
9.

4 - 16

Calibrate occlusion sensor, Ord. No. 3450 7353: (see Occlusion Sensor Function 500.0 * pg. 3 - 10).
Infusomat P 2.0gb

Unit Elements

10. Calibrate occlusion sensor, Ord. No. 3452 0589:


Note
To calibrate the occlusion sensor, Ord. No. 3452 0589 calibration
weights are required. An alignment with infusion lines is not permitted.
a) Clean the pump front side.
b) Place the Infusomat P without mains lead and drop sensor and the service plug inserted horizontally (front facing upwards) in the cellular packing of the calibration
device.
c)

Call function 500 in the service program (see Pressure


Sensor Function 310.0 pg. 3 - 6).

d) Activate the calibration function with "Align".


e) Select calibration value low "1" by pressing ">>".
f)

Open the pump cover.

g) Fasten the holders for the calibration weights under the


pump cover and let snap in at the locking bow.
h) Position weight 1 (100 g) for calibration value low "1"
(1.04 N) carefully on the occlusion sensor. The weight
must be placed without any friction in the holder.
i)

Acknowledge calibration value low "1" with ">>" and set


calibration value high "2" with weight 3 (166g).

j)

Acknowledge calibration value high "2" by pressing ">>"


and remove weight.

k)

Acknowledge calibration value "3" "Zero point" with


"Yes".

l)

Remove the holder.

Note
If the calibration values are known the values can also be entered
manually.
m) Call function 500 in the service program (see Occlusion
Sensor Function 500.0 ** pg. 3 - 10).
n) Open the data entry window with "yes".
o) Select Calibration value low "1" and Calibration value
high "2" by pressing <<" and >>". Enter the calibration
values via the keypad and acknowledge with yes.
p) Quit function in the service program and save data.
q) Register the changed pressure values in the unit book.

Infusomat P 2.0gb

4 - 17

Unit Elements

r)

Check the electronic occlusion pressure (see Electronic


Occlusion Pressure (Occlusion Sensor) pg. 8 - 3).

11. Quit function in the service program and save data.


12. Check the electronic occlusion pressure (see Electronic Occlusion Pressure (Occlusion Sensor) pg. 8 - 3).
Check
Safety check, functional check.

4.12 Air Sensor

Designation

Ord. No.

Air sensor incl. connector . . . . . . . . . . . . . . . . . . . . . . . 3450 193A


Exchange
1.

Remove battery (see Battery pg. 4 - 1).

2.

Dismount housing (see Housing and Handle pg. 4 - 5).

3.

Dismount controller board (see Controller Board


pg. 4 - 6).

4.

Disassemble front panel (see Front Panel pg. 4 - 11).

5.

Press the complete sensor from behind and out of the frame.

6.

Insert new sensor and seat the cables firmly with cable ties.

Note
Be sure that the cables of the reed sensor and air sensor do not
hinder the function of the occlusion sensor and the monitoring of
the motor speed (slot disc).
Check
After exchange of the air sensor:

Air Sensor

Fig.: 4 - 16

Check the air value, please refer to the TSI-list (see Technical Safety Inspection TSI pg. 7 - 1).

Check the water value; please refer to the TSI-list (see Technical Safety Inspection TSI pg. 7 - 1).

Check the calibration value (alarm threshold) and adjust if


necessary; refer to the TSI-list (see Technical Safety Inspection TSI pg. 7 - 1).

Safety check, functional check.

4 - 18

Infusomat P 2.0gb

Unit Elements

4.13 Operating Unit

Designation

Ord. No.

Membrane keypad with support plate and seal . . . . . 3450 1800


Support Plate

Zero Force Insertion Connector

LCD module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1819


Flex cable, preformed . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8830
Frame incl. pressure spring and magnet . . . . . . . . . . . 3450 1835
Flexible cable 42 mm (5 pcs.) . . . . . . . . . . . . . . . . . . . . 3477 3347
Hinge unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5571
Hinge pin (3 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5580
Magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5849
Exchange
Tools: Screw driver Torx T6

Flex Cable, Preformed


LCD Module

1.

Remove battery (see Battery pg. 4 - 1).

2.

Disassemble the door lock.

3.

Loosen countersunk screw and bridge.

4.

Remove tamper-proof caps (6 pieces) on the door frame by


piercing a screwdriver through the caps to loosen the countersunk screws.

5.

Unlatch the zero force insertion connector and loosen the


flex cable.

Fig.: 4 - 17

Frame Operating Unit


Note
The position of the flex cable must not be changed, i.e. the preformed section must be in the hinge area (pivot). Mark the cable,
if necessary.
6.

Disassemble either LCD module or support plate with membrane keypad or door hinge pins respectively and exchange
the door frame.

7.

Assembly is done in reverse order. Pay attention to the correct


direction of the door hinge pin during assembly.

Note

Hinge Pin
6 x PT2.5x14
Fig.: 4 - 18

Infusomat P 2.0gb

Do not kink either of the flex cables. Push the contacts to the stop
of the zero force insertion connector and lock in same position.
Check
Electrical safety, functional check, pump unit check.

4 - 19

Unit Elements

4.14 Barcode Label

Designation

Ord. No.

Barcode label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 9070


(see Order Form pg. A - 2)
Exchange
1.

Remove old barcode label, if existing.

2.

Clean the adhesion surface with an alcoholic cleaning agent


and let dry.

3.

Loosen barcode label from the base material and stick it on.

Note
Destroy the type plate delivered.
Check
Check that serial number and pump symbol in the plain text field
of the barcode label correspond with the type plate on the pump
of the Infusomat P.

Barcode Label
Fig.: 4 - 19

4.15 Frame with Seal

Designation

Ord. No.

Frame with seal plate . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1770


Exchange
1.

Disassemble all parts as described before and exchange the


frame with seal plate.

Check
Electrical safety, functional check, pump unit check.

4 - 20

Infusomat P 2.0gb

5-2

5-

Checks after Repair

Depending on the work carried out, perform the relevant check


blocks (1., 2., 3 and / or 4.).

1. Visual Inspection

OK after visual inspection

2. Safety Inspection
As per IEC/EN
60 601-1
Mains voltage ____ V AC
Protective conductor resistance
incl. mains
cable < 0.2
____
Insulation resistance
>> 2 M
____
Earth leakage current
30 A
____ A

3. Functional Inspection

Switch on Unit:
Self-test
Control lamps
Compare with Display:
Set delivery rate
Battery Test:
Switch mains/battery/
mains
Switch on in battery mode
and check self-test
Air Sensor:
0.4 ml air bubbles alarm
Air value
Water value
Calibration value (alarm
threshold), adjust if necessary
Values see TSI

4. Pump Unit Inspection

Room Temperature 20-28 C


Electronic Occlusion Pressure:
Check alarm with switchoff pressure
low / high
Zero value setting
Service program function
500.0
Mechanical Occlusion Pressure
Flow inhibitor
Pressure check 0.8bar
no free flow
Delivery accuracy

Drop Sensor:
Simulate occlusion alarm
(alarm with closed roller
clamp)
Simulate free flow (alarm)
Staff call
Alarm suppression
Observe the procedure information (see Procedural Instructions
for Inspection pg. 8 - 1)!

Infusomat P 2.0gb

5- 1

Checks after Repair

For your notes:

5- 2

Infusomat P 2.0gb

6-2

6-

Maintenance

It is recommended every 2 years. In addition to the technical safety inspection, perform the following inspection points:
1.

Check the rubber feet and if necessary exchange.

2.

Check easy running of the pump cover, lock mechanism and


door.

3.

Check easy running of the flow inhibitor, clean and if necessary exchange pressure springs.

4.

Check seal membrane and if necessary exchange.

5.

Check the drop sensor optic and spring mechanics and clean,
if necessary.

6.

Open unit. Internal visual inspection. Clean the seal surfaces


and if necessary exchange seal strip.

7.

Check mechanical occlusion pressure and if necessary calibrate.

8.

Check electronic occlusion pressure and if necessary calibrate


(see Occlusion Sensor pg. 4 - 15).

9. Assemble and seal unit ready for operation.

Infusomat P 2.0gb

6- 1

Maintenance

For your notes:

6- 2

Infusomat P 2.0gb

7-2

7-

Technical Safety Inspection TSI


Index c
(Kopiervorlage - Gertedokumentation beifgen)

Checklist for Technical Safety Inspection Every 24 Months


User

Unit: Infusomat P
Manufacturer: B. Braun Melsungen AG
Observe the service manual and the instructions for use. All measured values are to be documented.
Accessories used should be included in testing. Make exclusive use of calibrated measuring instruments.
Article No.

1. Visual Inspection
Cleanliness, completeness,
damage
Pump sealing diaphragm
Softkeys, rubber feet
Control unit, lock mechanism, pump cover
Flow inhibitor
Mains cable and mains plug
connector
MFC connector incl. MFC
cable
Drop sensor line and
plug connector
Check voltage values
100/110/120 V = T 0.315 A
200/230/240 V = T 0.16 A

Unit No.

Year of Procurement

2. Safety Inspection
as per IEC/EN60601-1

4. Functional Inspection

Check mains
voltage
____ V AC
Protective conductor resistance
incl. mains
cable < 0.2
____
Insulation resistance
>> 2 M
____
Earth leakage current
30 A
____ A

Switch on Unit:
Self-test
All symbols in LCD
Control lamps
Compare with Display:
Set delivery rate
Set volume
Set time
Press every key once
Battery Test:
Switch mains/battery/mains
Switch on in battery mode
3. Accessories Used
and check self-test
MFC staff call lead
Air Sensor:
_______________________
(Check with Intrafix Air P Ord.
_______________________
No. 0406 2957, in temperature
range 20C...25C)
0.4 ml air bubbles alarm
Air value max. 65mV
Water value min. 455mV
Alarm threshold = 130 mV
check and if necessary enter

Drop Sensor:
Simulate occlusion alarm
with closed roller clamp
Simulate free flow (alarm)
Electronic Occlusion Pressure:
Check alarm with switchoff pressure
low and high
Mechanical Occlusion Pressure:
max. 2.9 bar
____ bar
> 1.8 bar
____ bar

Staff call
Pump cover alarm
Alarm suppression
Flow inhibitor
pressure check up to 0.8bar
no free flow
Check delivery accuracy according to service manual
CAUTION: Charge battery!

Applied infusion line


Type: ___________________

Inspection performed by:


Manufacturer: ___________________

Test Result: Defects found which could endanger patients, users or third parties:
Measures to be taken:
Repair
___________________
Special Features / Documentation:

Yes

No

Unit handed over to/on:

Date / Signature:
Next deadline:

M664 00 00 92 F04 C 38912201


Infusomat P 2.0gb

7- 1

Technical Safety Inspection TSI

For your notes:

7- 2

Infusomat P 2.0gb

8-6

8-

Procedural Instructions for Inspection

1. Visual Inspection

Operating unit, lock mechanism, pump cover, seal membrane,


flow inhibitor. Door lock: easy opening and closing, correct top
and bottom locking.
Pump cover must automatically open when the unit door is
opened.

2. Electrical Safety Inspection as


per IEC/EN60 601-1

Protective Conductor Resistance


Protective conductor resistance < 0.2 . incl. mains lead.
Measurement points:
-

Potential equalization bolt

Bolt for door lock

Unit housing:
a) If the unit is not sealed countersunk screw at the rear of
the unit.
b) If the unit is sealed, remove lacquer from one of the holes
in the foot stands.
Note

Do not use the foot stand assembly screws as alternative measurement points.
Document the largest value.
Insulation Resistance
Insulation resistance >> 2 M
Measurement with 500 V between shorted mains connectors and
potential equalization bolt.
Earth Leakage Current
Earth leakage current 30 A incl. mains cable.
Measurement under standard conditions at the protective conductor of the mains cable. Two measurements (one with changed
poles).
Document the largest value.

Infusomat P 2.0gb

8- 1

Procedural Instructions for Inspection

3. Functional Inspection

Switch-on Test
Switch-on test keypad and display: check correct procedure.
Alarm tone, display: rate 000.0, VOL, TIME etc., display of all pixel,
contrast, display light, alarm and operating LED.
Battery Test
Switch mains/battery/mains: Interrupt mains supply twice in intervals of 1 second. Pay attention to the switch-over in the display. The unit must not switch to malfunction.
Note
Running time minimum 30 minutes after charging of 16 hours.
Air Sensor

Fig.: 8 - 1

Set rate to 400 ml/h.


Then inject 0.4 ml air bubble (inject piece). An alarm must be activated.
Drop Sensor
Set rate to 400 ml/h.
-

Simulate occlusion: Clamp infusion line in front of the pump


to prevent any drops. Alarm in less than 5 sec.

Simulate free flow: Press the bottom part of the drop chamber together, to generate a jet. Immediate alarm.

Staff Call
-

Connect MFC test plug

Generate an alarm (e.g. open pump cover during operation),


red LED is on in the test plug

If dynamic is set, 1 sec.

If static is set, until the alarm is acknowledged

Alarm Suppression
Press the alarm key. The current alarm is suppressed for 2 minutes.

8- 2

Infusomat P 2.0gb

Procedural Instructions for Inspection

4. Pump Unit Inspection

General
-

Room temperature 20 - 28 C

Use infusion line Intrafix AIR P (PVC) only once.

Connect manometer to the outlet side of the pump and position the manometer approx. to medium height of the Infusomat P.

Vent manometer (filled with water) 100 mm or 160 mm.

Measurement range: 4 bar

The results differ according to different measurement procedures.


With electronic measurement devices the values may be approx.
0.1 bar higher than indicated.
Electronic Occlusion Pressure (Occlusion Sensor)
Switch off drop control.
Set 600 ml/h and deliver in an open system for one minute.
Close the system and build-up pressure against a manometer with
100 ml/h.
Occlusion sensor threshold low. . . . . . . . . . . . . 400 to 1000 mbar
Occlusion sensor threshold high . . . . . . . . . . . 1000 to 1600 mbar
Note
The pressure threshold can be changed in the service program (not
recommended).
Default setting:
- low pressure: 0.6 bar
- high pressure: 1.2 bar
Tolerance range:
- set value: 0.3 bar.
Mechanical Occlusion Pressure

Fig.: 8 - 2

Set occlusion sensor threshold to mechanical (test plug).

Switch off drop control.

Build up pressure with 600 ml/h. Then measurement with 100


ml/h.

Read 90 seconds after start of delivery:

Measurement upper pressure value


Measurement lower pressure value
Measured values (see Technical Safety Inspection TSI
pg. 7 - 1).
After the measurement inspection the mechanical setting must
be switched off! The electronic occlusion pressure is not automatically activated again when the service plug has been disconnected.
Infusomat P 2.0gb

8- 3

Procedural Instructions for Inspection

Mechanical Pressure Setting


Check pump pressure. If the deviation is max. 0.3 bar from set
range, the pump can be calibrated.
Perform pressure measurement. Calibrate the pressure range at
the set screw with an Allen key 2.5 mm. to the upper pressure value of 2.4 bar.
After the measurement inspection the mechanical setting must
be switched off! The electronic occlusion pressure is not automatically activated again when the service plug has been disconnected.
Flow Inhibitor
Switch to stop at high pressures. Then open unit door. The pressure must stay above 0.8 bar.
Delivery Accuracy
Temperature 22 C
Rate > 1 ml/h
Typically 5% at a measurement of more than 8 hours.

8- 4

Infusomat P 2.0gb

Procedural Instructions for Inspection

Determination of Delivery Rate


Alternative procedure (12 minutes short measurement)
Temperature 20 - 28 C
Reference Height
80 2 cm

Measurement Equipment:
-

500 ml glass bottle, vented

Intrafix AIR P (PVC), drop chamber filled 2/3

Graduated cylinder 25 ml, accuracy 0.4 ml

Infusion solution NaCl or distilled water .

Procedure:

80 0.4 ml

Use a new infusion line for every measurement.

Check the system for narrow sections or kinks.

Delivery rate 125 ml/h.

The system run-in time is 1 - 2 minutes, but not longer than


2 minutes.

Insert the outlet cannula in the graduated cylinder.

Measurement rate:

125 ml

Measured volume:

25 ml

Simultaneously start stop watch and Infusomat P.

Stop when the 25 ml mark is reached and read time.

As shown in the graph the delivery behaviour slightly changes


over the infusion time. Consequently the tolerances must be
in the upper range when the delivery accuracy ( 5% over 8
hours) is checked with a short measurement over 12 minutes.

Expected rate deviation due to the short time method:


Software IFPC:+4.5%
tolerance range -0.5% bis +9.5% of the set rate.
Software IFPe / IFPE:+2.5%
tolerance range -2.5% bis +7.5% of the set rate.

Fig.: 8 - 3

Fig.: 8 - 4 Graph

Infusomat P 2.0gb

8- 5

Procedural Instructions for Inspection

Measurement Time

Deviation
%

Rate
ml/h

12 min

37.9 sec

-5

118,75

12 min

30.0 sec

-4

120,00

12 min

22.3 sec

-3

121,25

12 min

14.7 sec

-2

122,50

12 min

7.3 sec

-1

123,75

12 min

0.0 sec

125,00

11 min

52.9 sec

+1

126,25

11 min

45.9 sec

+2

127,50

11 min

39.0 sec

+3

128,25

11 min

32.3 sec

+4

130,00

11 min

25.7 sec

+5

131,25

11 min

19.7 sec

+6

132,50

11 min

12.9 sec

+7

133,75

11 min

6.7 sec

+8

135,00

11 min

0.6 sec

+9

136,25

10 min

54.5 sec

+10

137,50

10 min

48.5 sec

+11

138,75

10 min

42.9 sec

+12

140,00

10 min

26,1

+15

143,75

Table 8 - 1 Measurement Examples


The inspection and calibration - if required - must be performed
with the infusion line Intrafix AIR P (PVC)!
Alternative Measurement Procedure
Inspection of the delivery rate with a weight measurement. Avoid
errors due to evaporation!
Measurement Equipment:
-

Scales

Accuracy 0.1 g . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 min


Delivery Rate Determination:

8- 6

Set the delivery rate to 125 ml/h.

The run-in time is 1 - 2 minutes.

Insert the outlet cannula in container and simultaneously


start stop watch and Infusomat P.

After the time has expired stop Infusomat P and stop watch.

Immediately determine the delivery rate.


Infusomat P 2.0gb

9-2

9-

Test Equipment and Special Tools

Test Equipment and Special Tools

For Repair / for Technical Safety Inspection (TSI)


Order No.
Test equipment case Infusomat fm (complete) . . . . . 0770 1527
with:
Template 4 mm (for adjustment after exchange
of the pump cover) (for Infusomat fmS) . . . . . . . 0770 1489
Pin punch 1.8 mm x 160 mm (for hinge pin/
disassembly of the pump cover) . . . . . . . . . . . . . . 0770 1446
Pin punch 6 mm x 125 mm (for hinge pin/
assembly of the pump cover). . . . . . . . . . . . . . . . . 0770 1454
Flat tool 100 x 20 mm (for assembly/
disassembly of the tube adapter
of the Infusomat fmS) . . . . . . . . . . . . . . . . . . . . . . 0770 1462
Special socket spanner M18 (for disassembly of
the recessed plug). . . . . . . . . . . . . . . . . . . . . . . . . . 0770 1497
Manometer 0 - 4 bar . . . . . . . . . . . . . . . . . . . . . . . 0770 1357
Service plug (red) . . . . . . . . . . . . . . . . . . . . . . . . . . .0770 0709
MFC service plug. . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1215
OIL test infusion line (for Infusomat fmS) . . . . . . 0770 1500
Template 2.2 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0770 5034
Calibration device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0770 501A
Screw driver Torx T6
Screw driver Torx T10

Infusomat P 2.0gb

9- 1

Test Equipment and Special Tools

For your notes:

9- 2

Infusomat P 2.0gb

10 - 4

10 -

Spare Parts List

10

Unit Elements
Designation

Ord. No.

Mains Fuses
Fuse T 0.16 A for 200 / 230 / 240 V
(10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2847

Designation
Finnish. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Portuguese. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Czech . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Polish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Turkish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Ord. No.
3450 3404
3450 3382
3450 3358
3450 3358
3450 3382

Fuse T 0.315 A for 100 / 110 / 120 V


(10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 0534

Foot stand complete with rubber feet . . . . . . . . . . . . . 3450 5415

Fuse holder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 0979

Rubber feet (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3096


Unit handle with O-rings and PT screws . . . . . . . . . . . 3450 3450

Battery
Battery incl. connector plug
1.2 Ah / 7.2 V and holder . . . . . . . . . . . . . . . . . . . . 3450 2556

Controller Board

Door Lock

Buzzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3447

Door lock complete with push button . . . . . . . . . . . . . .3450 5601

Loudspeaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8848

Spring holder for door lock . . . . . . . . . . . . . . . . . . . . . . 3450 5440


Mounting for door lock . . . . . . . . . . . . . . . . . . . . . . . . . 3477 2790

Distance sleeve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3366

Rear Panel
Rear panel with screws (M3) and seal . . . . . . . . . . . . . 3450 1860

Pump Cover

Cover for optical interface . . . . . . . . . . . . . . . . . . . . . . 3477 3164

Pump cover with lock . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1916

Strip seal for rear panel . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3142

Blind plug 7.1 mm (10 pcs.) . . . . . . . . . . . . . . . . . . . . . .3477 3207

MFC connector board . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3374

Torsion spring in lever/pump cover (5 pcs.) . . . . . . . . 3477 3363

Potential equalization bolt . . . . . . . . . . . . . . . . . . . . . . 3477 0550

Torsion spring for pump cover (5 pcs.) . . . . . . . . . . . . 3477 3355

fm recessed plug (3 pin) . . . . . . . . . . . . . . . . . . . . . . . . 3477 3177

Lever (pump cover) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4092

Screw 30x8 for fm recessed plug (20 pcs.) . . . . . . . . . 3477 3185

Hinge pin for pump cover . . . . . . . . . . . . . . . . . . . . . . . 3477 3967

U Washer 3.2 (20 pcs.). . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3193

Hinge pin for pump cover lever . . . . . . . . . . . . . . . . . . 3450 5725

Mains module 200/230/240 V. . . . . . . . . . . . . . . . . . . . 3450 1886


Mains module 100/110/120 V . . . . . . . . . . . . . . . . . . . . 3450 1894

Pump Housing

Mains module 230 V . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 190A

Pump housing, (cpl.) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3390


Housing
Housing Labelling
German . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
French . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Dutch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Italian . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
English . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Spanish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Danish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Norwegian. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Swedish . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Infusomat P 2.0gb

Drop sensor socket incl. cable and plug . . . . . . . . . . . . 3450 1878


Front Panel

3450 1851
3450 1851
3450 1851
3450 1851
3450 3382
3450 3382
3450 3404
3450 3404
3450 3404

Front panel without clamp lever and torsion spring . . 3450 2092
Circular seal 571 mm / 45 mm . . . . . . . . . . . . . . . . . . . 3477 3126
Tamper-proof caps 10 mm (50 pcs.) . . . . . . . . . . . . . . 3477 3134
Pump housing, cpl. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3390
Cover 6.4 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3412
Clamp lever with torsion spring and pin 4x20 . . . . . . 3450 3420
Torsion spring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 3439
Reed sensor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1754

10 - 1

10

Designation

Spare Parts List

Ord. No.

Pump Unit
Note
When replacing the finger pump adhere to Table 4 - 1 Compatibility List pg. 4 - 6).

Designation

Ord. No.

Operating Unit
Membrane keypad with support plate and seal. . . . . . 3450 1800
LCD module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1819
Flex cable, preformed . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8830
Frame incl. pressure spring and magnet . . . . . . . . . . . 3450 1835

Finger pump (without motor) including pump, . . . . . 3450 1746


pump cover, seal membrane and occlusion sensor
board, software IFPC
(installed up to unit serial no. 5505)
Finger pump (without motor) including pump, . . . . . 3452 0481
pump cover, seal membrane and occlusion sensor
board, software IFPe, IFPE
(installed from unit serial no. 5506)
Finger pump (without motor and board) . . . . . . . . . . 3452 0597
incl. pump, pump cover,
and seal membrane

Flexible cable 42 mm (5 pcs.) . . . . . . . . . . . . . . . . . . . . 3477 3347


Hinge unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5571
Hinge pin (3 mm) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5580
Magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5849
Barcode Label
Barcode label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 9070
(see Order Form pg. A - 2)
Frame with Seal
Frame with seal plate . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 1770

Motor with pinion for finger pump . . . . . . . . . . . . . . . 3450 1924


Colours
Occlusion Sensor
Occlusion sensor (cpl.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 7353
Software IFPC
(installed up to unit serial no. 5505)
Occlusion sensor (cpl.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3452 0589
Software IFPe, IFPE
(installed from unit serial no. 5506)
(only delivered when a pressure
adjusting device is available)

Touch-up pen RAL 9001 (white) . . . . . . . . . . . . . . . . . . 3450 6977


Touch-up pen RAL 7032 (grey) . . . . . . . . . . . . . . . . . . . 3450 6985
Miscellaneous
Assembly screw for display board
PT 2.5x14 (10 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3100
Screw PT 3x10 Torx (self-forming) . . . . . . . . . . . . . . . . 3450 0960
Rubber feet grey (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . 3477 3096
Attention label fuse T 0.16 A (5 pcs.) . . . . . . . . . . . . . 3450 0898

Air Sensor

Screw M 6x8 for fm recessed plug (20 pcs.) . . . . . . . . 3477 3185

Air sensor incl. connector . . . . . . . . . . . . . . . . . . . . . . . 3450 193A

U Washer 3.2 (20 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3193

10 - 2

Infusomat P 2.0gb

10

Spare Parts List

Accessories
Universal Clamp

Designation

Ord. No.

Universal clamp, complete . . . . . . . . . . . . . . . . . . . . . . 3450 5857


Universal clamp base body . . . . . . . . . . . . . . . . . . . . . . 3450 8325
Threaded rod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8333
Star handle body for universal clamp. . . . . . . . . . . . . . 3450 8384
Safety bow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 8341
Safety hook. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 5865
Plate (2 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3450 2610
Connection cap D12 mm (5 pcs.) . . . . . . . . . . . . . . . . . 3477 4149
Bellows (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3477 3274
Pressure spring (5 pcs.) . . . . . . . . . . . . . . . . . . . . . . . . . 3477 4165
Note

Fig.: 10 - 1 Universal Clamp

For installation of the plate an appropriate press-in tool is required.

Software Update
Update-Kit IFPC02002 . . . . . . . . . . . . . . . . . . . . . . . . . 3450 6462
Update-Kit IFPe02003 . . . . . . . . . . . . . . . . . . . . . . . . . . On request
Update-Kit IFPE13003 . . . . . . . . . . . . . . . . . . . . . . . . . On request
(only units with controller
board with loudspeaker)
MFC interface line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .0871 1661

Infusomat P 2.0gb

10 - 3

10

Spare Parts List

For your notes:

10 - 4

Infusomat P 2.0gb

A-4

A-

Appendix

ESD Recommendations

A
Semiconductors can be destroyed by electrostatic discharge. Especially MOS components can be damaged by interference from
electrostatic fields, even without discharge via contact. This type
of damage is not immediately recognizable. Unit malfunctions
can even occur after a longer period of operation.
Every workstation must be equipped according to the recommendations with the necessary static protective measures, if ESD
components or boards are handled.

Workstation

Each workstation must be equipped with a conductive table surface. The conductive surface, the soldering iron or the soldering
stations must be grounded via protective resistors.
Chairs must be of antistatic design. The floor or floor mats should
be of electrically conductive material.
Personnel must wear conductive wristbands which are connected
to a central ground potential via protective resistors, e.g. the
ground contact of a wall outlet. Furthermore it is recommended
that personnel wear cotton clothing and electrically conductive
shoes to prevent electrostatic charge.

Fig.: A - 1

Infusomat P 2.0gb

A- 1

Appendix

Revision Service-Documentation

Version 2.0
This manual has been completely revised. The most important
changes are listed below:
-

Changed manual structure

New TSI list with new air alarm value

New paragraph occlusion sensor (exchange / calibration)

New test equipment for occlusion sensor

New software

New spare parts

Total list of spare parts

Current Information

Order Form

A- 2

The following form can be used as master to order type plates or


barcode labels.

Infusomat P 2.0gb

Appendix

FAX

(Master)

B. Braun Melsungen AG

Fax-No: (05661) 75 - 38 09

attn. Mr. Lohr


MT-PR-DE08C

Hospital: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

P.O. Box 1120

Ward: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

34 209 Melsungen

Street: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
zip code / town: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

email: [email protected]

Person responsible: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Tel.: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Order Barcode Label Order No. 3450 9070


for Infusomat P (DIANET-Type-Number 00203)
We herewith order the barcode label / type plate with the following serial numbers (the serial number is indicated
on the type plate of the pump).
Serial Number

Serial Number

Serial Number

Serial Number

We guarantee that we will fix the barcode label / type plates to the corresponding pumps in accordance with the
Service Manual and its enclosures and will carry out the necessary test steps. Barcode labels or type plates that
are not required will be destroyed.
Date: ____________________

Infusomat P 2.0gb

Signature:

A- 3

Appendix

For your notes:

A- 4

Infusomat P 2.0gb

You might also like