HERAcell 150i 240i With Decontamination Routine PDF
HERAcell 150i 240i With Decontamination Routine PDF
HERAcell 150i 240i With Decontamination Routine PDF
CO2 Incubator
HERAcell® 150i / 240i
with Decontamination Routine
50115191 B / 11.2010
Copyright 2009 Thermo Fisher Scientific. All rights reserved.
These operating instructions are protected by copyright. Rights resulting
thereof, particularly reprint, photomechanical or digital postprocessing or
reproduction, even in part, are only allowed with the written consent of the
Thermo Electron LED GmbH.
This regulation does not apply to reproductions for in-plant use.
The contents of the operating instructions are subject to change without further
notice. Concerning translations into foreign languages, the German version of
these operating instructions is binding.
Trademarks
HERAcell®, iCan™ and iONGUARD™are registered trademarks of Thermo Scientific.
Thermo Scientific is a brand of Thermo Fisher Scientific Inc.
All trademarks mentioned in the operating instructions are the exclusive property
of the respective manufacturers.
Contents
1 General notes 9
1.1 Identification of the device and of the documentation....................................... 10
1.2 Instruction of the operating personnel................................................................ 10
1.3 Applicability of the instructions ........................................................................... 11
1.4 Warranty ............................................................................................................... 11
1.5 Explanation of safety information and symbols ................................................. 12
1.5.1 Safety information and symbols used in the operating instructions........... 12
1.5.2 Additional symbols for safety informations: ................................................ 12
1.5.3 Symbols on the device .................................................................................. 13
1.6 Intended purpose of the device........................................................................... 14
1.6.1 Correct use ..................................................................................................... 14
1.6.2 Incorrect use................................................................................................... 14
1.7 Standards and directives ..................................................................................... 15
1.8 Safety notes on gases.......................................................................................... 16
1.8.1 Safety notes on carbon dioxide (CO2) .......................................................... 16
1.8.2 Safety notes on oxygen (O2) ......................................................................... 17
1.8.3 Safety notes on nitrogen (N2) ........................................................................ 17
50115191 B / 11.2010 3
®
Operating Instructions | HERAcell 150i / 240i
Contents
5 Start-up 49
5.1 Let the device adapt to the ambient conditions ................................................. 49
5.2 Preparing the work space .................................................................................... 49
5.3 Installing the shelf system ................................................................................... 50
5.3.1 Installation/removal of the support rail ......................................................... 50
5.3.2 Installing the shelf supports: ......................................................................... 51
5.3.3 Levelling the device ....................................................................................... 51
5.4 Installing the gas humidification (optional, only for O2/N2-control)................... 51
5.5 Installing the optional bottle turning devices HERAcell® 240i ........................... 52
5.6 Installing the optional center strut HERAcell® 240i for split inserts................... 54
5.7 Gas connection..................................................................................................... 55
5.7.1 Installing gas pressure hoses ........................................................................ 55
5.7.2 Installing devices without the optional gas monitoring system .................. 56
5.7.3 CO2 connection with gas monitoring system (optional) .............................. 57
5.7.4 Combined CO2/O2/N2 connection with gas monitoring system (optional) . 57
5.8 Power supply connection .................................................................................... 59
5.9 Connecting the RS 232 interface ......................................................................... 60
5.10 Connecting the USB interface (optional)........................................................... 60
5.11 Connecting the alarm contact............................................................................ 61
6 Operation 63
6.1 Preparing the device ............................................................................................ 63
6.2 Starting operation ................................................................................................ 64
50115191 B / 11.2010 4
®
Operating Instructions | HERAcell 150i / 240i
Contents
8 Shut-down 100
8.1 Shutting the device down.................................................................................. 100
10 Maintenance 111
10.1 Inspections and checks.................................................................................... 111
10.2 Service intervals ............................................................................................... 111
10.3 Preparing the temperature calibration ............................................................ 112
10.4 Temperature calibration procedure ................................................................ 113
10.5 Preparing the CO2 calibration .......................................................................... 114
10.6 CO2 calibration procedure ............................................................................... 116
10.7 Replacing the gas inlet filter............................................................................. 117
10.8 Replacing the device fuses .............................................................................. 117
10.9 Replacing the door seal ................................................................................... 118
11 Disposal 119
50115191 B / 11.2010 5
®
Operating Instructions | HERAcell 150i / 240i
Contents
50115191 B / 11.2010 6
®
Betriebsanleitung | HERAcell 150i / 240i
Figures
50115191 B / 11.2010 7
®
Betriebsanleitung | HERAcell 150i / 240i
Figures
50115191 B / 11.2010 8
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
50115191 B / 11.2010 9
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
50115191 B / 11.2010 10
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
Original document for translation • The contents of the operating instructions are subject to change
without further notice.
• Concerning translations into foreign languages, the German version of
these operating instructions is binding.
• Keep these operating instructions close to the device so that safety
instructions and important information are always accessible.
Should you encounter problems that are not detailed adequately in these
operating instructions, please contact Thermo Scientific immediately for
your own safety.
1.4 Warranty
Requirements for warranty Thermo Scientific warrants the operational safety and functions of the CO2
incubator only under the condition that:
• the device is operated and serviced exclusively in accordance with its
intended purpose and as described in these operating instructions,
• the device is not modified,
• only original spare parts and accessories that have been approved by
Thermo Fisher Scientific are used,
• inspections and maintenance are performed at the specified intervals.
The warranty is valid from the date of delivery of the device to the operator.
50115191 B / 11.2010 11
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
Harmful liquids!
Electric shock!
Hot surfaces!
Fire hazard!
Explosion hazard!
Suffocation hazard!
50115191 B / 11.2010 12
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
50115191 B / 11.2010 13
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
Fields of application HERAcell® 150i / 240i has been designed for installation and operation in
the following fields of application:
• Laboratories for cytobiological and biotechnological experiments of
safety levels L1, L2, and L3.
• Medical-microbiological laboratories in accordance with DIN 58 956.
• Laboratories in the central area of clinics and hospitals.
Gas supply system The gases required for the incubator (CO2 and / or O2/N2) are supplied to
the device from a separate gas supply system, either from gas cylinders or
from a central pressurized gas container.
The layout of the gas supply system must ensure that the operating
pressure of the gas supply lines can be set to a range between 0.8 bar
(min.) to 1 bar (max.) and that the pressure cannot be changed.
If the device is equipped with the optional gas monitoring system, up to
four devices can be interconnected in line, independent of the of the gas
supply system capacity.
The CO2 incubator is suitable for continuous operation.
50115191 B / 11.2010 14
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
50115191 B / 11.2010 15
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
Installation work:
Any work to supply lines and pressurized gas containers,
cylinders or containers used for storing CO2 or O2/N2 must only
be carried out by expert personnel using the appropriate tools.
Suffocation hazard!
Large amounts of CO2 released into the room atmosphere may
cause suffocation.
If CO2 is released, initiate safety measures immediately!
• Leave the room immediately and do not allow others to
enter the room!
• Inform security service or fire department!
50115191 B / 11.2010 16
®
Operating Instructions | HERAcell 150i / 240i
1 General notes
Oxygen explosion!
O2 may explode in combination with oils, greases, and
lubricants. If highly compressed oxygen comes in contact with
grease- or oil-containing substances, the mixture may explode!
• For cleaning these device components, use only oil- and
grease-free lubricants.
• Keep all connections and components of the oxygen system
free from substances that contain oil, grease, or lubricant!
Fire hazard!
Released oxygen promotes combustion. Do not use open
flames in the vicinity of oxygen-operated systems!
• Do not smoke in the vicinity of oxygen systems.
• Do not expose the components of an oxygen system to
excessive heat.
Suffocation hazard!
Large amounts of N2 released into the room atmosphere may
cause suffocation.
If N2 is released, initiate safety measures immediately!
• Leave the room immediately and do not allow others to
enter the room!
• Inform security service or fire department!
50115191 B / 11.2010 17
®
Operating Instructions | HERAcell 150i / 240i
2.1 Packaging
After the device has been delivered, check the delivery immediately for:
• completeness,
• possible damage.
Damage report If components are missing or damage is found on the device or the
packaging, esp. damages caused by humidity and water, please contact
the linehauler as well as the Technical Support of Thermo Scientific
immediately.
50115191 B / 11.2010 18
®
Operating Instructions | HERAcell 150i / 240i
HERAcell® 150i
Quantity of the delivered components CO2 incubator or CO2/O2 incubator
(in piece) Solid glass door and 3-door gas tight screen and
continuous shelves continuous shelves
Shelf 3 3
Support rail for shelf 4 4
Shelf supports for shelf 6 6
Insert for pressure compensation opening 1 1
Plug for pipe channel 1 1
Power supply cable 1 1
Connector, potential-free contact 1 1
Spare caps, set 1 1
CO2 connecting hose set 1 1
Open-end wrench, 24 mm 1 1
Allen wrench 2 mm for blower wheel 1 1
Allen wrench 3 mm for blower wheel cover 1 1
Operating instructions (CD) 1 1
Water pump 1 1
HERAcell® 150i
Quantity of the delivered components CO2 incubator or CO2/O2 incubator
(in piece) Solid glass door and 3-door gas tight screen and
continuous shelves continuous shelves
O2 connecting hose set 1 1
O2 sensor with set for gas humidification *1) 1 1
CO2 connecting hose set, gas monitoring (optional) 1 1
O2 / N2 connecting hose set, gas monitoring (optional) 1 1
*1) O2 sensor with gas humidification set is packed in a separate carton and placed into
the work space during transport.
50115191 B / 11.2010 19
®
Operating Instructions | HERAcell 150i / 240i
HERAcell® 240i
CO2 incubator or CO2/O2 incubator
Quantity of the delivered
Solid glass door and Solid glass door 6-door gas tight 6-door gas tight
components
continuous shelves and split shelves screen and screen and split
(in piece)
(standard version) (optional) continuous shelves shelves
(standard version) (optional)
Shelf 3 6 3 6
Support rail for shelf 4 6 4 6
Shelf supports for shelf 6 12 6 12
Insert for pressure
1 1 1 1
compensation opening
Plug for pipe channel 1 1 1 1
Power supply cable 1 1 1 1
Connector, potential-free
1 1 1 1
contact
Spare caps, set 1 1 1 1
CO2 connecting hose set 1 1 1 1
Open-end wrench, 24 mm 1 1 1 1
Allen wrench 2 mm for
1 1 1 1
blower wheel
Allen wrench 3 mm for
1 1 1 1
blower wheel cover
Operating instructions (CD) 1 1 1 1
Water pump 1 1 1 1
50115191 B / 11.2010 20
®
Operating Instructions | HERAcell 150i / 240i
HERAcell® 240i
CO2 incubator or CO2/O2 incubator
Quantity of the delivered
components Solid glass door and Solid glass door 6-door gas tight 6-door gas tight
(in piece) continuous shelves and split shelves screen and screen and split
(standard version) (optional) continuous shelves shelves
(standard version) (optional)
O2 connecting
1 1 1 1
hose set
O2 gas humidification
system connecting hose 1 1 1 1
set *1)
Bottle turning device
(motor with four 1 — 1 —
rollers, optional) *2), *3)
CO2 connecting hose set,
1 1 1 1
gas monitoring (optional)
O2/ N2 connecting hose set,
1 1 1 1
gas monitoring (optional)
*1) O2 sensor with gas humidification set is packed in a separate box that is placed into
the work space during transport.
*2) For bottle turning devices, the version with split shelves is not available.
*3) Each level of a bottle turning device is packed in a separate box. The boxes (1 to 4)
are delivered with the device.
50115191 B / 11.2010 21
®
Operating Instructions | HERAcell 150i / 240i
The device must only be operated at locations that meet the particular
ambient conditions listed below:
Requirements:
Location • Draft-free and dry location.
• The minimal distance to adjacent surfaces must be observed on all
sides (see Section 3.3.).
• The operating room must be equipped with appropriate room
ventilation.
• Solid, level, fire-proof surface.
• Vibration-proof substructure (floor stand, lab table) capable of bearing
the weight of the device and of accessories (particularly if several
devices are stacked).
• The electrical system of the device has been designed for an operating
height of up to 2000 m above sea level.
• To ensure a constant incubation temperature of 37 °C, the ambient
temperature must be within a range of +18 °C to +33 °C.
• Relative humidity up to 80% (max.).
• No direct exposure to sunlight.
• Devices that produce excessive heat must not be placed near the
location of the HERAcell® 150i / 240i.
50115191 B / 11.2010 22
®
Operating Instructions | HERAcell 150i / 240i
EMERGENCY STOP!
The mains socket must always be
accessible for the case of an EMERGENCY
STOP.
50115191 B / 11.2010 23
®
Operating Instructions | HERAcell 150i / 240i
3.4 Transport
Lift points:
Lift the device only using the lift points.
3.5 Stacking
HERAcell® 240i
• Place the adapter plate with the receptacle
recesses [8] at the bottom onto the stacking
elements [9] on top of the device. Make sure that
the adapter plate is seated safely on the recesses.
• Place the device to be stacked with the device
stands [6] onto the stacking elements [7] at the top
of the adapter plate.
The devices are secured to one another by their own
weight.
Fig. 3: Stacking
50115191 B / 11.2010 24
®
Operating Instructions | HERAcell 150i / 240i
3.6 Retrofitting/Modifications
Retrofitting options The following components can be retrofitted to the standard version:
HERAcell® 150i
• 3-segment gas-tight screen (the gas-tight screen is installed in place of
the glass door),
• door hinge replacement on outer door and glass door (RH / LH door
opening),
• lockable outer door,
• split inserts,
• infrared (IR) sensor,
• PCB for USB interface.
HERAcell® 240i
• 6-segment gas-tight screen (the gas-tight screen is installed in place of
the glass door),
• door hinge replacement on outer door and glass door,
• lockable outer door,
• split inserts,
• infrared (IR) sensor,
• PCB for USB interface.
Modifications:
Retrofittings and modifications must only be performed by the
Technical Service of Thermo Electron LED.
50115191 B / 11.2010 25
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 26
®
Operating Instructions | HERAcell 150i / 240i
®
Fig. 4: Front view HERAcell 150i
50115191 B / 11.2010 27
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 28
®
Operating Instructions | HERAcell 150i / 240i
®
Fig. 5: Rear view HERAcell 150i
50115191 B / 11.2010 29
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 30
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 31
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 32
®
Operating Instructions | HERAcell 150i / 240i
®
Fig. 7: Front view HERAcell 240i
50115191 B / 11.2010 33
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 34
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 35
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 36
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 37
®
Operating Instructions | HERAcell 150i / 240i
Temperature:
Temperature conditions To ensure undisturbed operation, the temperature in the operating room
must be at least 18 °C and the incubation temperature must exceed this
temperature by at least 3 °C.
The heating system controls the incubation temperature from this
temperature threshold up to 55 °C. The principle of air jacket heating and
the additional, separate heating of the outer door and glass door/gas-tight
screen minimize the risk of condensate formation at the side walls, at the
ceiling of the work space, and at the glass door/gas-tight screen.
Relative humidity:
Water quality The heating of the work space promotes the condensation of the water,
thereby ensuring a constant humidity within the work space. For the
running operation, always keep a sufficient quantity of processed water of
the following quality available:
• HERAcell® 240i: quantity 4.5 l,
• HERAcell® 150i: quantity 3.0 l,
50115191 B / 11.2010 38
®
Operating Instructions | HERAcell 150i / 240i
Termination of warranty!
Using chlorinated tap water or additives that contain chlorine
will void the manufacture warranty. Similarly, the use of
ultrapure water whose conductivity is out of the range of 1 to
20 μS (and whose resistivity is out of the range of 50 kΩ to
1 MΩ) will void the manufacture warranty.
If you should have any questions, please contact Technical
Support of Thermo Fisher Scientific.
CO2 supply:
CO2-Work space atmosphere control To ensure the growth conditions for the cell and tissue cultures, the work
space is supplied with CO2.
The pH of the bicarbonate-buffered culture media largely depends on the
CO2 content of the work space atmosphere.
The CO2 content of the work space atmosphere can be controlled within a
range of 0 to 20%.
The supplied CO2 must have one of the following quality characteristics:
• Purity 99.5% min.,
• medical gas quality.
O2 supply:
O2-Work space atmosphere control If the CO2 incubator is to be operated with more than 21% oxygen, the
work space is supplied with oxygen. The O2 content of the work space
atmosphere can be controlled within a range of 0 to 21%.
N2 supply:
N2-Work space atmosphere control If the oxygen content during operation is to be lowered to less than 21%
(air oxygen content), the work space is supplied with nitrogen. The O2
content of the work space atmosphere can be controlled depending on the
configuration of the sensor within a range of 1% to 21% and 5% to 90%.
50115191 B / 11.2010 39
®
Operating Instructions | HERAcell 150i / 240i
The outer door can only be closed after the glass door
has been latched properly.
50115191 B / 11.2010 40
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 41
®
Operating Instructions | HERAcell 150i / 240i
Gas connection:
The gas supply lines between the device and the gas
supply system are connected using the supplied
connecting hoses. O2/N2 is supplied to the device
through connecting sleeve [2]. CO2 is connected to
sleeve [3].
All process gases must be supplied to the device at a
fixed pressure that has been preset within a range of
0.8 to 1.0 bar and must remain unchanged.
Before the gases are fed into the work space, they flow
through a gas inlet filter with a separation rate of 99.97
% for a particle size of 0.3 μm (HEPA filter quality).
The illustration shows the optional combined gas
connection without the optional gas monitoring
system.
Label:
The label [4] contains information about gas supply,
an alarm contact terminal legend, and notes about the
electrical fusing of the device.
RS 232 interface:
Via the RS 232 interface [5], the incubator can be
connected to the serial interface of a PC. This
connection allows the computer-aided acquisition and
documentation of the major operating parameters
(temperature, CO2/O2/N2 concentration, error codes,
etc.).
50115191 B / 11.2010 42
®
Operating Instructions | HERAcell 150i / 240i
Alarm contact:
External alarm system Fig. 12: The device can be connected to an on-site, external alarm system
(e.g. telephone system, building monitoring system, visual or audible
alarm system).
For this purpose, a potential-free alarm contact [7] is preinstalled in the
device.
Alarm contact:
The alarm contact responds for all errors reported by the
control loops (see chap. 7.13‚ pg. 95).
50115191 B / 11.2010 43
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 44
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 45
®
Operating Instructions | HERAcell 150i / 240i
HERAcell® 150i:
– access port, Ø 42 mm [1]
– pressure compensation opening [2]
HERAcell® 240i:
– access port, Ø 42 mm [3]
– pressure compensation openingt [4]
Operating conditions:
When accessories are to be operated in the
work space of the CO2 incubator, the
ambient condition requirements must be
observed (see table below). The energy
introduced into the work space affects the
beginning of the temperature control
range. When additional heating sources are
introduced into the work space,
condensation (e.g. at the glass door) may
occur.
50115191 B / 11.2010 46
®
Operating Instructions | HERAcell 150i / 240i
Connection assignment:
If the drive roller is not properly connected
to the control unit, the parameters cannot Fig. 16: Bottle turning device
be set correctly on the operating panel of
the device.
50115191 B / 11.2010 47
®
Operating Instructions | HERAcell 150i / 240i
Receptacles:
To prevent corrosion due to moisture in the
work space, unassigned receptacles of the
bottle turning device must always be
plugged with the protective cap.
50115191 B / 11.2010 48
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
7. Upon delivery, the CO2 incubator is not in a sterile state. Before the
initial start-up, the device must be decontaminated.
Before the decontamination is performed, the following work space
components must be cleaned:
• Support rails,
• support hooks,
• shelves,
• bottle turning assembly (optional),
• gas humidification (optional),
• work space surfaces,
• work space seals and gaskets,
• glass door/gas-tight screen.
Decontamination:
For details about the cleaning and disinfection of the device,
(see chap. 9‚ pg. 101).
50115191 B / 11.2010 49
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
50115191 B / 11.2010 50
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
50115191 B / 11.2010 51
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
Use of diagram:
8. The diagram shows the common bottle diameters
as different graphs (ascending straight).
– Set the desired speed at the Y axis. Read the
corresponding set value (in %) from the X axis
at the point of intersection of the Y value with
the specific graph.
50115191 B / 11.2010 52
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
Example:
– Bottle diameter: 56 mm
– Desired speed: 3.5 rpm
– Determined set value: 48%
In the dialog box SPEED BOTTLE TURN. DEVICE, the value 48% is set (see
chap. 7.11.3‚ pg. 85).
50115191 B / 11.2010 53
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
50115191 B / 11.2010 54
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
Gas quality:
The gases must have one of the following
qualities:
• Purity 99.5 % min,
• medical gas quality.
Overpressure!
The operating pressure of the gas applied
to the device must not exceed 1 bar. If the
gas is supplied at a higher pressure, the
valves integral to the device may not close
correctly and the gas supply control may
be impaired.
Set the gas supply to a range between 0.8
bar min. and 1.0 bar max. and make sure
that this pressure setting cannot be
changed!
50115191 B / 11.2010 55
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
CO2 connection:
• For a device with CO2 connection, connect the gas
supply to the sterile filter [2] in accordance with the
connection diagram [1].
50115191 B / 11.2010 56
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
CO2 connection:
For a device that is operated with process gas CO2
and equipped with the optional gas monitoring
system, connect the gas supply lines as follows:
• Distributor connection [5] for the CO2 gas supply
for up to three additional devices. For looping the
gas supply through, interconnect the devices with
a gas pressure hose.
• Connect the gas container B to the upper
connection [2] of the gas monitoring system,
• connect the gas container A to the lower
connection [3] of the gas monitoring system. Fig. 26: Gas connection with gas monitoring system (optional)
• The outlet of the gas monitoring system [4] is
connected at the factory to the gas inlet filter [5] by
a short gas pressure hose [6].
O2/N2 supply:
• Distributor connection [1] for the O2/N2 gas supply
for up to three additional devices. For looping the
gas supply through, interconnect the devices with
a gas pressure hose.
• Connect the gas container B [2] to the upper
50115191 B / 11.2010 57
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
CO2 supply:
• Distributor connection [5] for the CO2 gas supply
for up to three additional devices. For looping the
gas supply through, interconnect the devices with
a gas pressure hose.
• Connect the gas container B to the upper
connection [6] of the gas monitoring system,
• connect the gas container A to the lower
connection [7] of the gas monitoring system.
• The outlet of the gas monitoring system [8] is
connected at the factory to the gas inlet filter [10]
by a short gas pressure hose [12].
50115191 B / 11.2010 58
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
Electric shock!
Contact with current-carrying components
may cause a lethal electric shock.
Before connecting the device to the power
supply, check plug and power supply cable
for damage.
Do not use damaged components when
connecting the device to the power supply!
50115191 B / 11.2010 59
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
Data communication via RS 232 The RS 232 data communication interface has been designed for a cable
connection with 9-pin connector and a contact assignment of 1:1.
Data exchange is accomplished by a predefined structure of command
sequences (see annex at the end of this section).
Data communication via USB As an alternative to data communication via RS 232 interface, the devices
can be equipped with the optional USB interface. The USB interface
complies with Standard USB 1.1 and is compatible with Standard USB 2.0
(full speed).
The interface stepping rate can be changed within the defined baud rates
(9,600, 19,200, 38,400, 57,600 baud).
Data exchange is accomplished via a defined command sequence
structure (frames). The command sequences correspond with the RS 232
interface installation diagram (see annex at the end of this section).
Using the USB interface for data communication between a PC and the
device a driver must be installed first (see annex data communication).
50115191 B / 11.2010 60
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
Skilled work:
Thermo Scientific warrants the operational safety and the
operativeness of the device only if installation and repairs are
performed properly.
The connection of the device to an external alarm system must
only be carried out by adequately trained and authorized
expert electrical/telecommunication personnel!
Function:
When system errors and failures occur in the temperature or gas control
circuits, an alarm message is issued to the connected alarm/monitoring
system. The potential-free contact (1 changeover contact) has been
designed for the following circuits:
Alarm relay:
Contact Contact
Operating states
4-1 4-3
Operating state power failure "off" X O
Operating state power failure "on" O X
Failure: power failure "off" O X
Failure: power failure "on" X O
X: Contact closed / O: Contact open
Switching structure:
The alarm relay switches for all failures reported by the control
loops (see chap. 7.13‚ pg. 95).
50115191 B / 11.2010 61
®
Operating Instructions | HERAcell 150i / 240i
5 Start-up
Connection example:
The connector [5] for the connecting cable is included
in the standard equipment. The values for the
operating voltage of the external circuits and of the
fusing of the alarm system are given in the table on the
previous page.
1. Connect the individual conductors [1] to [4] of the
connecting cable as shown in the wiring diagram.
2. Connect the connector [5] of the alarm system
connecting cable to the interface [6] at the control
box at the rear panel of the device.
50115191 B / 11.2010 62
®
Operating Instructions | HERAcell 150i / 240i
6 Operation
The device must only be released for operation after all major measures for
the start-up have been taken (see Section 5, page 45).
Device check:
Device check Prior to starting operation, the following device components must be
checked for their correct function
• Gas hoses must be seated tight on the connecting filter and must be
secured using a hose clamp.
• The access port must be capped.
• The pressure compensation opening must be permeable, its insert
must be installed in the work space.
• The glass door/gas-tight screen seal must not be damaged.
• The glass door measurement opening must be capped.
• The shelf system components must be installed safely.
• The gas humidification (optional) must be connected to the device-
integral gas supply and aligned parallel to the rear panel.
• The drive rollers of the bottle turning device (optional) must be properly
connected to the corresponding control unit.
• The idle rollers of the bottle turning device (optional) must be
positioned in accordance with the cylinder diameters.
Hygiene regulations:
Prior to any operation, the user must clean and disinfect the
work space in accordance with the hygiene regulations set
forth by the operator to protect the cultures.
The “Principles of good microbiological proceedings” at the
annex of these instructions can be used as safety information
for personnel working with the device.
Water supply:
A sufficient quantity of treated water of the following quality
must always be available for the running operation:
• distilled, fully demineralized, and
• autoclaved, sterile.
When the filling level drops below a minimum value, it must be
possible to top up water during the work process.
Capacities:
• HERAcell® 150i: 3.0 l
• HERAcell® 240i: 4.5 l
50115191 B / 11.2010 63
®
Operating Instructions | HERAcell 150i / 240i
6 Operation
1. Fill the water tray with a sufficient quantity of processed water. Do not
exceed the upper level mark.
2. Make sure that the CO2/O2/N2 supply system valves are open.
3. Turn the device on using the power switch.
4. Set the set values for temperature and CO2 / O2 content on the iCan™
touchscreen.
5. Start contra-con.
Loading:
To ensure sufficient air circulation and even heating of the
samples, the loading surface within the work space should be
used up to 70% max. Voluminous objects in the work space
that dissipate heat may impair heat distribution.
50115191 B / 11.2010 64
®
Operating Instructions | HERAcell 150i / 240i
Fig. 31: The control panel works as a touchscreen Fig. 30: Power switch
(iCan™ touchscreen) and can be controlled
performing light pressure on the pressure-sensitive
areas of the screen using a finger or a blunt pin:
• Temperature display [1],
• CO2 display [3],
• O2 display [5] (optional).
• contra-con key [2]
• menu key [4]
• auto-start key [6],
50115191 B / 11.2010 65
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 66
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 67
®
Operating Instructions | HERAcell 150i / 240i
CO2/O2 control:
Since the CO2 concentration of the air is
nearly 0%, the CO2 control and the control
Fig. 35: Heat-up phase display
loop error monitoring system are disabled
at a set value of 0%.
Since the O2 concentration of the air is
nearly 21%, the O2 control and the control
loop error monitoring system are disabled
at a set value of 21%.
50115191 B / 11.2010 68
®
Operating Instructions | HERAcell 150i / 240i
Key response:
Pressing a key can increase or reduce a value gradually:
• When the – key [4] or the + key [6] is kept depressed, the system
switches to quick run,
• after the key has been kept depressed for more than 3 seconds, the
quick run speed increases.
50115191 B / 11.2010 69
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 70
®
Operating Instructions | HERAcell 150i / 240i
Gas monitoring:
The switching state of the CO2 control loop
does not affect the function of the optional
gas monitoring system. The gas monitoring
system remains active even when the CO2
control is switched off.
For set values of 21% O2, the O2 control is
switched off and error monitoring is
disabled as well.
The control loop error monitoring system
(optional) is enabled.
50115191 B / 11.2010 71
®
Operating Instructions | HERAcell 150i / 240i
Factory settings:
Depending on the type of the O2 sensor,
one of the two O2 control ranges has been
preset at the factory:
Control range I: 1% to 21%
Control range II: 5% to 90%
50115191 B / 11.2010 72
®
Operating Instructions | HERAcell 150i / 240i
The auto-start function is an automated routine for the start and the
subsequent adjustment of the CO2 measuring system. After the start, the
device control adjusts the temperature to the set value while humidity is
generated. When temperature and relative humidity have reached constant
values, the CO2 measuring system is automatically adjusted to these
values, and the work space is supplied with the preset quantity of CO2.
50115191 B / 11.2010 73
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 74
®
Operating Instructions | HERAcell 150i / 240i
Cancellation:
The auto-start routine can be cancelled at
any time!
Press the Stop key [6].
Automatic restart:
The auto-start routine is restarted
automatically if the routine is cancelled due
to one of the following events:
• The glass door is opened, Fig. 39: Activating the auto-start routine
50115191 B / 11.2010 75
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 76
®
Operating Instructions | HERAcell 150i / 240i
• Keypad lock,
• Software versions.
To make a user-specific setting in a dialog box,
navigate through the submenus listed in the
illustrations and open the dialog box.
7.11.1 Settings
The input dialogs of the SETTINGS category comprise
all settings for the individual configuration of the user
display screen and the user interface:
• Keypad lock coding,
• Date/time setting,
• Contrast setting,
• Key tone setting,
• Interface configuration,
• User display screen language setting,
• Reminder interval setting.
50115191 B / 11.2010 77
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 78
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 79
®
Operating Instructions | HERAcell 150i / 240i
Use of HERAline:
If the devices are combined with the
HERAline accessory, the baud rate must be
set to 9,600 baud.
2. To reduce the value: Fig. 45: RS 232 interface baud rate setting
> Press the - key [2].
50115191 B / 11.2010 80
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 81
®
Operating Instructions | HERAcell 150i / 240i
Factory settings:
contra-con decontamination routine 90 days
auto-start routine Off
Service interval 365 days
Fig. 48: Reminder interval setting
50115191 B / 11.2010 82
®
Operating Instructions | HERAcell 150i / 240i
Event display:
The event display [1] uses short, single-line entries
specifying date and time to report the events that were
logged during the operation of the device. The entries
are listed in chronological order with the most recent
event at the top position. The list can be displayed but
not edited. If the event display consists of several
pages, the user can browse through the list. The status
indicator [2] indicates which page of the total number
of pages is currently being displayed.
10 s 22.5 hours
30 s 2.8 days
60 s 5.6 days
120 s 11.2 days
180 s 16.8 days
pg. 94).
50115191 B / 11.2010 83
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 84
®
Operating Instructions | HERAcell 150i / 240i
Troubleshooting:
For a detailed overview of causes for errors
and their correction, please refer to the end
of this chapter!
7.11.3 Options
The input dialogs of the OPTIONS category comprise all
settings of the functional device options:
• Alarm relay,
• Low humidity,
• Gas-tight screen,
• Water level sensor,
• Audible alarm,
• Bottle turning device (optional),
• O2 (optional).
50115191 B / 11.2010 85
®
Operating Instructions | HERAcell 150i / 240i
Low humidity:
The enabling/disabling of the Low
humidity function is entered into the event
list.
If the Low humidity function is enabled, the
corresponding icon [5] is displayed in the
main menu.
50115191 B / 11.2010 86
®
Operating Instructions | HERAcell 150i / 240i
Malfunction:
The conversion to gas-tight screen causes a
change of the control parameters.
Failure to set the gas-tight screen function
in accordance with the actually installed
door version may impair the incubation Fig. 54: Gas tight screen setting
performance.
50115191 B / 11.2010 87
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 88
®
Operating Instructions | HERAcell 150i / 240i
3. To accept and save the setting: Fig. 58: Switching the O2 control on and off
O2 value display:
If the O2 control has been switched off, the
O2 display does not show an actual value (-
- -).
This procedure provides protection for the
O2 sensor.
If the set value is set to 21%, the O2 control
loop is not monitored. This applies to both
O2 control ranges:
• Control range I: 1% to 21%
• Control range II: 5% to 90%
In this case, the O2 display shows the
actual value.
50115191 B / 11.2010 89
®
Operating Instructions | HERAcell 150i / 240i
Gas monitoring:
The switching state of the O2 control loop
does not affect the function of the optional
gas monitoring system. The gas monitoring
system remains active even when the O2
control is switched off or the O2 sensor is
disabled.
Low humidity:
This function indicator shows that the relative
humidity in the work space has been lowered from
approx. 93% to approx. 90%. Fig. 59: Icon description overview
Keypad lock:
This function indicator shows that the keypad lock has
been activated so that currently the settings cannot be
changed.
Overtemperature:
This error indicator shows that the device control has
activated the overtemperature protection and has
switched to backup control.
50115191 B / 11.2010 90
®
Operating Instructions | HERAcell 150i / 240i
Request service:
This indicator shows that the routine service is due.
The display of the icon is controlled by the time entry
in the REMINDER INTERVAL dialog box and it appears
after the reminder message has been confirmed.
50115191 B / 11.2010 91
®
Operating Instructions | HERAcell 150i / 240i
Cylinder changeover:
A manual or automatic changeover of a gas
cylinder is entered into the event display.
50115191 B / 11.2010 92
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 93
®
Operating Instructions | HERAcell 150i / 240i
Logging cycle:
The time interval of the data logging cycle
can be redefined in the INTERVAL TIME EVENT
LOGGER dialog box (see chap. 7.11.2‚
pg. 83).
50115191 B / 11.2010 94
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 95
®
Operating Instructions | HERAcell 150i / 240i
7.13.3 Troubleshooting
The error tables indicate source of error, cause of
error, and possible corrections.
For any communication with the Technical Support of
Thermo Fisher Scientific, please have the device data Fig. 65: Event error overtemperature
ready.
50115191 B / 11.2010 96
®
Operating Instructions | HERAcell 150i / 240i
*1) Error message will be displayed, but not entered in error table.
*2) Error message will be displayed, but not entered in error table.
*3) If failure is detected, a special sample protection controlling will be enabled. To indicate the start of this routine,
the icon overtemperature is displayed. The routine will will be disabled after a power reset.
*4) After changing the set value, error time will be set on 159 min., after door opening on 45 min. ( O2 on 159 min.).
50115191 B / 11.2010 97
®
Operating Instructions | HERAcell 150i / 240i
*4) After changing the set value, error time will be set on 159 min., after door opening on 45 min. ( O2 on 159 min.).
50115191 B / 11.2010 98
®
Operating Instructions | HERAcell 150i / 240i
50115191 B / 11.2010 99
®
Operating Instructions | HERAcell 150i / 240i
8 Shut-down
Contamination hazard!
If the work space surfaces are contaminated, germs can be
transferred to the environment of the device.
In case of a shut-down, the device must be decontaminated!
1. Remove the culture containers and all accessories from the work
space.
2. Pump the water out off the water reservoir (see chap. 9).
3. Fill in 350 ml water and start the contra- con decontamination routine.
4. Turn the device off using the power switch when the contra-con
decontamination routine has been completed.
5. Unplug the power connector and protect it against accidental
reconnection.
6. Close the CO2/O2/N2 supply system shut-off valves.
7. Disconnect the gas pressure hoses from the sleeve at the rear of the
device.
8. Clean the work space and wipe the device dry.
9. Until the device is shut down, the work space must be continuously
ventilated. Leave the glass door and the outer door open and secure
them in this state.
9.1 Cleaning
Incompatible cleaners!
Some device components are made of plastic. Solvents can
dissolve plastics. Strong acids or caustic solutions can cause to
become brittle of the plastic.
For cleaning plastic components and surfaces, do not use
solvents that contain hydrocarbons, solvents with an alcohol
content of more than 10% or strong acids or caustic solutions.
Moisture-sensitive components!
Do not spray cleaning agent onto the touchscreen and the
control box at the rear of the device. When wiping the device
clean, always make sure that moisture does not enter into
these components.
Cleaning display:
Moisture-sensitive display!
Do not spray or wipe the display with cleaner.
Disinfection procedure The following disinfection procedures are compatible with the device:
Wipe/spray disinfection:
The wipe/spray disinfection is used as the standardized manual disinfection
procedure for the device and for all accessories.
Alcoholic disinfectants!
Disinfectants with an alcohol content of more than 10% may
form, in combination with air, easily combustible and explosive
gas mixtures.
When using such disinfectants, avoid open flames or exposure
to excessive heat during the entire disinfection process!
• Use such disinfectants only in adequately ventilated rooms.
• After the disinfectant has been allowed to react, wipe the
cleaned device components thoroughly dry.
Observe safety regulations to avoid fire and/or explosion
hazard caused by alcohol-containing disinfectants (ZH 1/598).
Chloride-containing disinfectants!
Chloride-containing disinfectants can corrode stainless steel.
• Use only disinfectants that do not affect stainless steel!
Electric shock!
Contact with current-carrying components may cause a lethal
electric shock.
Prior to cleaning and disinfection work, disconnect the device
from the power supply!
• Turn the device off using the power switch.
• Unplug the power connector and protect it against
accidental reconnection.
• Check to see if the device is deenergized.
Health hazard!
The surfaces of the work space may be contaminated. Contact
with contaminated cleaning liquids may cause infections.
Disinfectants may contain harmful substances.
When cleaning and disinfecting, always observe the safety
instructions and hygiene regulations!
• Wear safety gloves.
• Wear safety goggles.
• Wear mouth and respiratory system protection gear to
protect the mucous membranes.
• Observe the safety instructions of the manufacturer of the
disinfectant and of the hygiene experts.
Predisinfection:
Predisinfection 1. Remove all samples from the work space and store them in a safe
place.
2. Pump the water off and remove water residues using a cloth.
3. Spray disinfectant onto the surfaces of the work space and of the
accessories or wipe the surfaces clean using disinfectant.
Moisture-sensitive components!
Do not spray disinfectant onto the CO2 sensor in the measuring
cell baseplate or onto the O2/N2 sensor.
of the measuring cell. The wheel and the cover can be autoclaved.
Electric shock!
Contact with current-carrying components may cause a lethal
electric shock.
Prior to removing the blower wheel, switch the device off and
disconnect it from the power supply system.
1. Remove the two retaining screws of the cover using the supplied Allen
wrench (3 mm) and remove the cover.
2. The blower wheel is secured to the axle by a set screw. Remove the
set screw using the Allen wrench (2 mm) and pull the blower wheel
off.
Final disinfection:
Final disinfection 1. Spray disinfectant again onto the work space, the shelf system, and
the removed components and wipe them clean.
2. Allow the disinfectant to react as specified by the manufacturer.
3. Reinstall the shelf system and the removed components.
Functional check:
After the installation, check to see if the blower wheel is
securely attached to the axle and if it can rotate freely, then
secure cover using the screws.
Overtemperature protection:
If the overtemperature protection was enabled on the device,
the contra-con decontamination routine cannot be started
before the fault hes been corrected or reset.
Hot surfaces!
The handle and the screen of the glass door, the interior panel
of the outer door as well as the surfaces of the shelf system and
of the work space become extremely hot during the contra-con
decontamination routine.
During the routine run or immediately after completion of the
run, always wear safety gloves when touching these surfaces!
3. To activate contra-con:
> Press the START key [3].
> The contra-con-instruction menu [2] is
displayed.
1. To interrupt contra-con:
> Press the STOP key [6].
2. To cancel contra-con:
> Press the END key [2].
> The error message is displayed.
> If the error message is confirmed, the system
returns to the main menu.
3. To resume contra-con:
> Press the BACK key [1].
> The system returns to the status display, and
the decontamination routine is resumed.
2. To cancel contra-con:
> Press the END key [2].
> The error message is displayed.
> If the error message is confirmed, the system
returns to the main menu.
10 Maintenance
To ensure the operativeness and the operational safety of the device, the
functions and device components listed below must be checked at
different intervals.
Daily check:
• Gas supply of the CO2 supply system.
• Gas supply of the O2/N2 supply system.
Annual inspection:
• Tightness of the glass door seal.
• Permeability of the pressure compensation opening with insert.
• Functional check of the operating panel and of the device control.
• Electrical safety check in accordance with the relevant national
regulations (e.g. VBG 4).
Functional check:
If safety devices were removed or disabled for inspections, the
device must not be operated before the safety devices have
been reinstalled and checked for their correct function.
3-month service:
• Run auto-start routine and contra-con decontamination routine,
• perform temperature and CO2/O2 comparison measurement.
Annual service:
• Replace gas inlet filter.
• Perform Technical Service service inspection.
Service contract:
Thermo Scientifc offer a device-specific service contact that
comprises all test and service works required.
10 Maintenance
Isothermal container:
Do not use a container filled with water as
isothermal container as the evaporation of
water will result in a lower temperature Fig. 70: Preparing the temperature calibration
reading.
10 Maintenance
Measurement example:
• Temperature set value: 37 °C
Reference temperature: 36.4 °C
10 Maintenance
Value reset:
If the value isn’t changed within the next 30
seconds, the system automatically exits
the menu, and the most recently confirmed
value is preserved.
Suitable instrument:
• Portable IR readout instrument.
10 Maintenance
IR measuring cell:
For devices with infrared (IR) measuring
cells, the CO2 calibration can only be
performed when the CO2 concentration has
been set to 4.0% or more.
Calibration should be performed with the
CO2 set value designated for the work
process (prospective work process value).
10 Maintenance
Measurement example:
• CO2 set value: 5 %
Reference temperature: 5,6 %
Value reset:
If the value isn’t changed within the next 30
seconds, the system automatically exits
the menu, and the most recently confirmed
value is preserved.
Fig. 74: CO2 calibration procedure
10 Maintenance
Fig. 76: The two identical device fuses [4] are installed
in the fuse compartment [1] next to the power plug
receptacle of the device:
1. The fuse holder is secured to the fuse Fig. 76: Device fuse replacement
compartment [1] using two locking tabs [2].
2. To remove the fuse holder, squeeze the two
locking tabs and pull the holder [3] out of the fuse
compartment.
3. Remove the faulty fuse from the fuse holder and
install the new fuse.
4. Slide the fuse holder into the fuse compartment
and press the holder on until the locking tabs are
fully engaged.
10 Maintenance
1. Pull the magnetic seal [3] out of the guide slot [1].
2. Position the new seal at a corner [2] and press the
seal retaining rail [4] into the slot.
3. Make sure that the retaining rail taper is
positioned correctly in the slot [1] and that the
seal is flush with the door frame.
11 Disposal
Contamination hazard!
The device may have been used for treating and processing
infectious substances. Therefore, the device and device
components may have been contaminated.
Prior to disposal, all device components must be
decontaminated.
Decontamination measures • Clean the device components thoroughly and either
disinfect or decontaminate them (depending on application).
• Attach a declaration of non-objection with details of
performed decontamination measures to the items that are
to be disposed of.
Controlled disposal All device components can be disposed of after they have been
decontaminated properly.
Recycling Service
Recycling Service
Thermo Fisher Scientific offer a recycling service for discarded
components at the owner’s expense.
11 Disposal
12 Technical data
HERAcell® 150i
12 Technical data
HERAcell® 150i
12 Technical data
HERAcell® 240i
12 Technical data
HERAcell® 240i
13.1 Interfaces
Windows dialog box 1. Select the INSTALL FROM A LIST OR SPECIFIC LOCATION (ADVANCED) option.
All characters sent and received in the data exchange between a PC and the
incubator HERAcell® 150 i / 240 i are ASCII characters that can be displayed
on a conventional terminal.
This ensures that communication can be set up, controlled, and
programmed easily.
Reading parameters:
Query: ?:aaaa:bb::cc<CR>
or: ?:aaaa:bb:XXXX:cc<CR>
Response: !:aaaa:bb:XXXXX:cc<CR>
where: aaaa = parameter address
bb = amount of payload in this telegram (00 – ff)
cc = checksum: Inverted XOR of all bytes without
checksum and <CR>
XXXX = bb payload bytes
Writing parameters:
Command: !:aaaa:bb:XXXXX:cc<CR>
Response: !:aaaa:bb::cc<CR>
where: aaaa = parameter address
bb = amount of payload in this telegram (00 – ff)
cc = checksum: Inverted XOR of all bytes without
checksum and <CR>
XXXX = bb payload bytes
General parameters General parameters are system values such as date, time, and
motherboard version number.
Error memory data set structure These data sets consist of 11 byte and are encrypted in 21 ASCII characters
prior to data transmission. Example: The byte 0x23 is converted into the
ASCII characters 0x32 (‚2’) and 0x33 (‚3’).
– Byte 1 consists of 1 characters,
– Bytes 2 - 11 consist of 2 characters.
Therefore, a response consists of 1+(10 x 2) = 21 data bytes plus
separator.
A data set always delivers the date, the time, the faulty control loop, the
device status, and the error message.
Example of a response:
00 0b 01 06 0f 37 22 8000 0002
The data logger can store up to 10,000 events. Depending on the setting for
the logging cycle (in sections of seconds), e.g. for a value of 60 s (default
value), the events of about 5 days can be stored.
The data logger stores the following information:
• important user actions, system events, and error messages,
• measured data of the three control loops during incubation operation.
Data logger data set structure Each query command is responded with up to 7 data sets that follow one
another
without a separator. These data sets consist of
16 bytes and are encrypted into 32 ASCII characters prior to transmission.
For example, the byte 0x23 is converted into the ASCII characters:
0x32 (‚2’) and 0x33 (‚3’).
Therefore, a response consists of up to 7 X 16 = 112 bytes, i.e. 224 ASCII
characters.
A data set always delivers the date and the time (without seconds), the
device status and the type of data logger entry (byte 0-7 or ASCII character
0-15).
Also, depending on the entry, the current actual values or set values of the
control loops or other parameters can be entered (byte 8-15 or ASCII
character 16-31).
Example of a response:
Example of a code:
For an example of a code, please refer to the end of this
chapter.
Data logger entry An entry in the data logger is 16 bytes large and has the following structure:
1. byte: indicates the event (e.g. door open 0x31, measured value entry
0x01)
2. byte: day of entry
3. byte: month
4. byte: year
5. byte: hour
6. byte: minute
7. and 8. byte: device status
9. to 16. byte: various data on the event
switch (zahlenstring[i*SIZE_DATA2]){
case EVENT_STATUS:
// check cyclic entries for device errors
status = zahlenstring[6+i*SIZE_DATA2]*0x100+
zahlenstring[7+i*SIZE_DATA2];
if (status & INFO_ERROR){
str_cpy (&excelstring[len], "Error active;", 13);
len += 13;
}
else{
// query all device errors (see chap. 13.6.2‚ pg. 139)
if (status & DOOR_LONG){
str_cpy (&excelstring[len], "Door open too long;",
19);
len += 19;
else {
if (status & DOOR_OPEN){
str_cpy (&excelstring[len], "Door open;", 10);
len += 10;
}
}
// query remaining device errors now
// ·
// ·
// ·
// ·
// ·
// and finally query cyclic nominal values entries without device
// errors
else{
str_cpy (&excelstring[len], "ok;", 3);
len += 3;
}
}
// copy nominal values from numerical string to Excel string
len += num_2_str ((zahlenstring[8+i*SIZE_DATA2]*0x100+
zahlenstring[9+i*SIZE_DATA2]), &excelstring[len]);
excelstring[len ++] = ';';
len += num_2_str ((zahlenstring[10+i*SIZE_DATA2]*0x100+
zahlenstring[11+i*SIZE_DATA2]), &excelstring[len]);
excelstring[len ++] = ';';
len += num_2_str ((zahlenstring[12+i*SIZE_DATA2]*0x100+
zahlenstring[13+i*SIZE_DATA2]), &excelstring[len]);
excelstring[len ++] = ';';
len += num_2_str ((zahlenstring[14+i*SIZE_DATA2]*0x100+
zahlenstring[15+i*SIZE_DATA2]), &excelstring[len]);
excelstring[len ++] = ';';
// enter set values from here
len += num_2_str (SollTemp, &excelstring[len]);
excelstring[len ++] = ';';
len += num_2_str (SollCO2, &excelstring[len]);
excelstring[len ++] = ';';
len += num_2_str (SollO2, &excelstring[len]);
excelstring[len ++] = ';';
len += num_2_str (SollrH, &excelstring[len]);
excelstring[len ++] = ';';
excelstring[len] = '\n';
len += 1;
WriteFile (FileHandle, excelstring, len);
break;
// query the remaining events from here
case EVENT_FORMAT_DATALOG:
WriteFile (FileHandle, excelstring, len);
WriteFile (FileHandle, "Data logger erased;\n",20);
break;
case EVENT_POWER_ON:
// update of set values
SollTemp = zahlenstring[8+i*SIZE_DATA2]*0x100+
zahlenstring[9+i*SIZE_DATA2];
SollCO2 = zahlenstring[10+i*SIZE_DATA2]*0x100+
zahlenstring[11+i*SIZE_DATA2];
SollO2 = zahlenstring[12+i*SIZE_DATA2]*0x100+
zahlenstring[13+i*SIZE_DATA2];
SollrH = zahlenstring[14+i*SIZE_DATA2]*0x100+
zahlenstring[15+i*SIZE_DATA2];
WriteFile (FileHandle, excelstring, len);
WriteFile (FileHandle, "Power on;\n", 10);
break;
case..
// query all events here (see chap. 13.6.2‚ pg. 139)
// cancel 0xFF indicates the end of the data logger
case 0xFF:
WriteFile (FileHandle, "End;\n",5);
}
}
return 0;
}
User surface for data communication The program provides a user surface (only with English menu designation)
for handling the data communication between a device and a connected
PC.
Transmission speeds:
The transmission speed settings in the user menu PRESETTING
and of the device must be identical!
TEST COM
Test Com The submenu TEST COM is used for testing the communication
connection with the settungs defined in the submenu PRESETTING.
1. Data in the two text boxes must be entered in the format DD.MM.YY
(day, month, year).
2. Accept the input data:
> Press the SET button.
ERROR LOGGER
Error Logger The submenu ERROR LOGGER is used for reading the error messages into
the text box of the user surface.
DATA LOGGER
Data Logger The submenu DATA LOGGER is used for reading the event entries into the
text box of the user surface.
SERVICEFILE
Servicefile The submenu SERVICEFILE is used for reading error informations and for
creating a service file from it, saved with the proprietary extension *srf. The
service file is transmitted to the Technical Service of Thermo Fisher Scien-
tific for fault analysis.
PASSWORD
Password The submenu PASSWORD can only be accessed by the service personnel
of Thermo Fisher Scientific.
14 Device log