Non-contact Infrared Forehead Thermometer Dear customer, We thank you for having purchased Microlife FR1DZ1, the thermometer that takes a fast, precise body temperature reading from a distance. Microlife FR1DZ1 is a high quality product incorporating the latest technology and tested in accordance with international standards. With its unique technology, the Microlife FR1DZ1 can provide a stable, heat-interference-free reading with each measurement. The instrument performs a self-test every time it is switched on to always guarantee the specified accuracy of measurement. The Microlife FR1DZ1 allows a simple and quick measurement of body temperature. Please keep in mind that, like any skin temperature, the temperature at the forehead and temporal area might differ from the core body temperature (e.g. obtained by rectal measurement). For example a basic physiologic effect called vasoconstriction can occur in the early stages of fever, resulting in a cool skin effect, so the recorded temperature using a forehead thermometer may be suspiciously low. If the measurement result is not consistent with the patients finding or suspiciously low, repeat the measurement every 15 minutes or double check the result by another core body temperature measurement, e.g. use an infrared eardrum thermometer or a rectal pen-type thermometer. Skin temperature of babies reacts very quickly to the ambient temperature. Therefore, do not take a measurement while or after nursing, as the skin temperature may be lower than the core temperature.

Table of Contents 1. Features 2. Important Safety Instructions 3. Product Overview 4. Icons and display sequence 5. Directions of Use 6. Reference site 7. Switching between C and F 8. Recalling Readings in Memory Mode 9. Error Messages 10. Cleaning and Storage 11. Battery Replacement 12. Guarantee 13. Technical Specifications 14. Website

1. Features Non-contact Measurement within seconds The innovative infrared sensor measures temperature without even touching the subject for a safe and sanitary measurement within seconds. Human and Object Measurement Conveniently measures the temperature of a human being or an object with its sliding switch. Customized and Localized Option of Measurement Reference site selection function enables you to take measurement for readings comparable to readings taken at conventional measuring sites (rectal, oral, axillar) with a common pen-type thermometer. Tracking light Blue tracking light easily keeps you concentrated to the measured area while scanning the forehead above the eye brow with the thermometer at a distance.
1

Fever Alarm Red LCD glow and ten short beeps alert the patient that he/she may have a fever. Multiple Reading Recalls Recall the last 30 readings when you enter the Memory Mode. Auto-Display Memory The last reading is automatically displayed for 2 seconds when the unit is switched on. 2. Important Safety Instructions This instrument may be used only for the purpose described in this booklet. The manufacturer cannot be held liable for damage caused by incorrect application. Never immerse the thermometer into water or other liquids (not waterproof). For cleaning, please follow the instructions in the Cleaning and Storage section. Keep the instrument away from direct exposure to the sun and keep it in a dust-free, dry area with a temperature between -20 to 50 C (-4.0 122 F). Do not use the thermometer if there are signs of damage on the probe area or on the unit itself. If damaged, do not attempt to repair the instrument. Please contact the nearest Microlife customer service center. This thermometer consists of high-quality precision parts. Do not drop the unit! Protect it from severe impact. Do not twist the unit or the measuring probe. Warning: The use of this thermometer is not a substitute for consultation with your physician.

3. Product Overview Side View Back View

No. 1

Item Display area

Mode Power Off

Start button Power on

Description Displays measurements and other thermometer symbols Press and release immediately to enter Memory Mode. Press and hold for 3 seconds to switch between C and F . Press to measure the temperature. Press to switch the thermometer on or off. Covers the battery for protection. Focus on the subject to measure temperature. Infrared sensor to measure the temperature. Sliding the switch down and up again activates the reference site selection. See Selecting the Reference Site. Slide the switch up to activate the body mode. Slide the switch down to activate the object mode. There is a screw on the battery cover to protect the cover.
3

3 4 5 6 7

O/I button Battery cover Tracking light Probe

Slide switch

Power on Power on/off

screw

4. Icons and Control Display The icons on the display indicate different settings of the thermometer: Setting Description Icon The thermometer will take or has taken a Body mode body temperature. The thermometer will take or has taken an Object mode object temperature. The thermometer will take or has taken a Celsius display temperature in degree Centigrade. The thermometer will take or has taken a Fahrenheit display temperature in degree Fahrenheit. When this icon flashes, the body temperaOral reference site ture measured corresponds to the oral temperature. Axillary reference site When this icon flashes, the body temperature measured corresponds to the axillary temperature. Rectal reference site When this icon flashes, the body temperature measured corresponds to the rectal temperature. The thermometer displays a stored temperMemory recall ature from the memory.
In different stages of the operation sequence, the display shows different types of information. Display area Sequence Description Press the O/I button to turn on the unit. All segPower On ments are displayed for 2 seconds. Last Memory The last reading is displayed for 2 seconds with M icon, measuring Mode icon (body or object), and reference site icon(rectal, oral or axillar) The C or F icon is continuously flashing. The unit Ready is ready for measurement with measuring Mode icon (body or object), and reference site icon (rectal, oral or axillar) in Body mode. Measurement The reading is shown on the display area with measuring Mode icon (body or object), and referCompleted ence site icon(rectal, oral or axillar) in Body mode. After 2 seconds, the C or F icon flashes again. Further Measurement The unit is now ready for the next measurement.
4

5. Directions of Use Measuring Body Temperature 1.Make sure the mode switch is in the upper position to enter the Body Mode. 2.Press the O/I button. The LCD is activated and performs an all segment check. 3.The last readings are displayed on the LCD for 2 seconds.

4.The thermometer is ready for measurement when A short beep sound is heard The C/F and the reference site icon on the LCD are flashing. 5.For starting the measurement, aim the probe at the center of the forehead with any distance of no more than 5cm. If the forehead area is covered with hair, sweat or dirt, please remove them beforehand to improve the reading accuracy. 6. Then keep pressing the Start button. The blue tracking light is activated as a reference of area under detection.You can take measurement at any distance within 5 cm seeing the blue light. Steadily move the thermometer towards the temple to detect the body temperature. As the sensor scans the area, quick beep sounds will be emitted as indicator of temperature detecting process. The beep sounds will slow down to indicate that the body temperature has already been detected. 7.As soon as beeps slow down, please release the Start button to get the result. Stop aiming the probe. The display area shows the measured temperature. 8.Body temperature intelligent analysis. If the detected temperature is lower than fever level, a long beep is emitted and the result is displayed with a green backlight. If the detected temperature is equal to or higher than fever level, ten short beeps
5

will be emitted while the result is displayed with a red backlight to alert the patient of possible fever. Site option Rectal Oral Underarm Fever level 37.8C/100.04F 37.5C/99.5F 37.2C/98.96F

(Based on physician recommendation, temperatures taken at different body sites belong to varied ranges. Fever levels also comply with this logic.) Measuring Object Temperature 1.Make sure that the mode switch is in the lower position to enter the Object Mode. 2.Press the O/I button. The LCD is activated and performs an all segment check. 3.The last reading is displayed on the LCD for 2 seconds.

4.The thermometer is ready for measurement when A short beep sound is heard C/F icon on the LCD is flashing. 5.Aim the probe at the center of the object with any distance of no more than 5 cm. If there is water, dust or dirt, please wipe them beforehand to improve the reading accuracy. 6.Then keep pressing the Start button. The blue tracking light is activated and the sensor measures the temperature. 7.Release the Start button after a few seconds to get the result and a long beep will be heard. Stop aiming the probe. The display area shows the measured temperature. Note: Patients and the thermometer should stay in steady-state room conditions for at least
6

200 180 160 140 120 100 80 60 40

30 minutes before performing a measurement. Wait at least 2 minutes between consecutive measurements and the thermometer must be turned off and on again. Dont take a measurement while or immediately after nursing a baby. Dont use the thermometer in high humidity environments. Patients should not drink, eat, or exercise before/while taking the measurement. Before aiming the probe, clean the scan area; remove dirt, hair or sweat. Before aiming the probe at an object, remove water, dust or dirt. Dont move the thermometer away from the measuring area before hearing the termination beep. If the probe gets dirty during or after the measurement, use an alcohol swab to carefully clean the probe before putting it away for storage. Always take the temperature in the same location, since the temperature readings may vary from different locations. In the following situations, it is recommended to perform a reference measurement with a common pen-type thermometer(make sure the measuring site, rectal, oral, or axillar, is the same habitual site also selected for FR1DZ1) : 1.If the measurement result is surprisingly low/high 2.Newborn infants in the first 100 days. 3.Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical. 6. Reference Site Microlife FR1DZ1 specially enables you to take measurements and judge the readings according to your local habits. This is based on the considerations that people of different regions have their customs about where to take the measurements with an ordinary pen-type thermometer. (Rectal, oral, or axillar) Microlife FR1DZ1 can measure for three types of readings comparable to readings measured at such three conventional measuring sites as rectal, oral, and axillar with an ordinary pen-type thermometer. We call this function as reference site. To set the reference site: 1.Turn on the unit in Body Mode and wait until the C or F symbol is flashing to indicate that the thermometer is ready for measurement. 2.Slide the mode switch down and up again (if the slide switch is already down, you need to slide up, down and up again). The present reference site icon flashes while the other two icons are steady.
7

3.Press the Start button repeatedly to scroll between the three measuring sites.

A short beep sound is heard when changing the reference site. 4.The symbol for the selected mode flashes and after 5 seconds, the flashing icon will be set as the reference site and the unit will automatically enter the ready for measurement mode. 7. Switching between C and F This thermometer can display the measured temperature in either Celsius or Fahrenheit. To switch between C and F, turn the unit off. Press and hold the Start button for 3 seconds. The flashing C or F icon is shown on the LCD. To switch the measurement scale, press the Start button again. Settings are finalized after 5 seconds and the thermometer will automatically enter the ready for measuring mode. 8. Recalling Readings in Memory Mode The thermometer can recall the last 30 readings. To recall readings, do the following: 1.Press and release the Start button in the off mode to enter the Memory Mode. The icon flashes. 2.Press and release the Start button again and 1 is shown briefly on the LCD, then the latest reading is displayed.

3.Press and release the Start button again and 2 is shown briefly on the LCD, then the second latest reading is displayed.

4.Press and release the Start button repeatedly to recall the readings in succession. Up to 30 readings can be recalled.

Pressing the Start button after the last 30 readings will resume the sequence from reading 1. 9. Error Messages The thermometer is equipped with predefined error messages to notify the user of measurement discrepancies. Display area Display Meaning Measured temperature is too high. Measured temperature is too low. Ambient temperature is too high. Ambient temperature is too low. Error function display Possible cause and remedy Measured temperature is higher than: 42.2 C or 108.0 F (Body Mode) 100 C or 212 F (Object Mode) Measured temperature is lower than: 34.0 C or 93.2 F (Body Mode) 0 C or 32.0 F (Object Mode) Ambient temperature is higher than 40.0 C or 104.0 F (Body and Object Mode) Ambient temperature is lower than 16.0 C or 60.8 F (Body Mode) 5.0 C or 41.0 F (Object Mode) When the system has a malfunction.

Blank display

Check if the batteries are loaded correctly. The battery level is low. Please replace the batteries immediately.

Low battery indication

10. Cleaning and Storage Use an alcohol swab or cotton tissue moistened with 70% Isopropyl alcohol to clean the thermometer casing and the measuring probe. Ensure that no liquid enters the inte9

rior of the thermometer. Never use abrasive cleaning agents, thinners or benzene for cleaning. Never immerse the thermometer into water or other cleaning liquid. Be careful not to scratch the surface of the probe lens and the display. Remove the batteries from the instrument if it is not required for extended periods of time to avoid potential damage to the thermometer resulting from leaking batteries. 11. Battery Replacement The thermometer requires 2 DC3V (AAA) batteries. Replace the batteries when the flashing battery icon appears on the display. Remove the battery cover with the screwdriver. To remove the batteries, insert the tip of a screwdriver into the slot as indicated to move the batteries out of the battery compartment.

12. Guarantee This high-quality measuring device is covered by a 2-year-guarantee from the date of purchase. The guarantee is valid only on presentation of the guarantee card completed by the dealer (see back) confirming date of purchase or the purchasing receipt. Warranty claims must be filed within the guarantee period.The guarantee covers the instrument; batteries and packaging are not included. Opening or altering the instrument invalidates the guarantee.The guarantee does not cover damages caused by improper handling, discharged batteries, accidents or non-compliance with the operating instructions. This product was manufactured with the utmost care according to international quality standards. Should you have a reason for complaints despite this, please send the device, accompanied by the completed Guarantee Card with the dealers stamp as well as original proof of purchase directly or through your medical supplier to your nearest Microlife Distributor. Name and address of responsible dealer:
10

13. Technical Specifications Type: Measuring Range And Accuracy: Resolution: Display: Acoustic: Non-contact Forehead Thermometer FR 1DZ1 Refer to the following table 1 0.1C/0.1F LCD, 4-digits plus special icons a. The unit is turned ON and ready for measurement: 1 short beep sound. b. Completion of Measurement: Body Mode: 1 long beep sound if the reading is less than fever level (refer to diagram #1); 10 short beep sounds if the reading is equal to or higher than fever level (refer to diagram #1). Object Mode: 1 long beep sound c. System error or malfunction: 3 short beeps d. Fast and slow beeps during measurement. a. Auto-Display the last measured temperature when the unit is turned on. b. 30 readings recall in the Memory Mode. a. The display is lighted GREEN for 4 seconds, when the unit is turned ON. b. The display is lighted GREEN for 5 seconds, when a measurement has been completed with a reading of less than fever level. c. The display is lighted RED for 5 seconds, when a measurement has been completed with a reading equal to or higher than fever level.

Memory:

Backlight:

Storage/transport Temperature: -20 C to +50 C (-4 F to 122 F) Automatic Switch Off: Approx. 3 minutes after the last measurement has been taken. Battery: Two DC 3V (AAA) batteries Dimensions: 146.93mm (L) x 38.14mm (W) x 21.34 mm (H) Weight: 48g (with battery), 45g (without battery) Reference to standards Device standard: Device corresponds to the requirements of the standard for infrared thermometers ASTM E 1965
11

IEC 60601-1 IEC 60601-1-2 Electromagnetic Compatibility: Device fulfills the stipulations of the standard IEC 60601-1-2

The stipulations of EU-Directive 93/42/EEC for Medical Devices Class IIa have been fulfilled. Table 1 TECHNICAL CHARACTERISTICS Body mode Forehead measuring range Ambient temperature working range From 36C to 39C (96.8F to 102.2F) 39.1 C to 42.2C(102.4F to 108.0F) Object mode Measuring range Ambient temperature working range From 36C to 39C (96.8F to 102.2F) From 0C to 35.9C (32.0F to 96.6F) and from 39.1C to 100C (102.4F to 212.0F) 0 C to 100C /32F to 212? 5C to 40C /41F to 102F +/-0.2C /0.4F +/-1C /2F C/F scale 34C to 42.2C(93.2F to 108.0F) 16C to 40C(60.8F to 102F) +/-0.2C /0.4F

From 34.0C to 35.9C(93.2C to 96.6F) and from +/-0.3C /0.6F

12

14. www.microlife.com Detailed user information about our thermometers and blood pressure monitors as well as services can be found at www.microlife.com.

Batteries and electronic instruments must be disposed of in accordance with the locally applicable regulations, not with domestic waste.

Read the instructions carefully before using this device.


Type BF applied part Reference number Lot number Manufacturer