Sick GM31
Sick GM31
Sick GM31
GM 31 Product File
SO2
NH3
NO
Temp
NO2
GM
GM31
31Models
Models Meas.
Meas.Components
Components
GM
GM31
31-1-1 SO2
SO2
GM
GM31
31-2-2 SO2
SO2and
andNO
NO
GM
GM31
31-3-3 SO2,
SO2,NONOand
andNO2
NO2
GM
GM31
31-4-4 NO
NO
GM
GM31
31-5-5 SO2,
SO2,NONOand
andNH3
NH3
GM
GM31
31-7-7 NO
NOand
andNO2
NO2
Cement Industry
- Continuous emission monitoring
- SO2 and NOx generating processes
- NH3 slip determination
Other Branches
Petrochemical - Glas Industry
Industry - Paper Industry
- Continuous emission - Sugar Industry
monitoring - ...
- Process control
GE02 GM 31 Gas Analyzer 11/99
GM31 - Certificates
US. EPA
Small dimensions
Simple commissioning
High reliability
Contents
GE02 GM 31 Gas Analyzer 11/99
GM31 System with Standard Components
Temperature
sensor
up to 8 anlalog inputs
(0-20mA /50 Ohm)
up to 8 analog outputs
(0,2,4 - 20mA,
max.700 Ohm)
up to 12 relay outputs
(48V/1A)
Simulation
Front window
Condenser
UV
transmitter
lamp
Deflection mirror
Diode row
Optical grid
During the control cycle the zero-point reflector remains in the light path.
The cycle is active for at least 4.5 min and is flagged with a status signal.
NO Cuvette
Check of the spectral characteristics of diode array
Conversion in general:
Small dimensions
Probe length and active path min / max: see flange and probe type
Protection class: IP 65
Condition
Contents of under pressure at mounting location: max. - 60 hPa
GE02 GM 31 Gas Analyzer 11/99
GM31 - Probe Types
Temperature sensor
Temperature sensor
Filter element
Temperature sensor
Ceramic or
Teflon + Ceramic
Ceramic filter
Temperature sensor
Temperature sensor
GPP with
Directed GPP with EPA Probe
Teflon +
Purge Air Ceramic with Audit
Ceramic
Probe ( HT ) Filter Cell
Filter
Flue gas temperature [°C] <550 <260 <250 <250
Suitability of calibration with test
x x x
gases acc. EPA
Suitability for extreme disturbances
x x
of flow profile in the duct
Purge air requirement x x
Velocity of flue gas [m/s] >1 any any >7
Wet flue gas application x x x
Overpressure [hPa] <60 <120 <120 <60
SO2, NO, SO2, NO, SO2, NO, SO2, NO,
Measuring components
NO2, NH3 NO2, NH3 NO2, NH3 NO2*, NH3*
Duct diameter [mm] >300 >300 >300 >800
Gas concentration sum [g/m3B] <14 <14 <14 <8
Dust concentration [g/m3B] <15 <30 <30 <10
Available lengths of probes [m] 0.9/ 1.5/ 2.0 0.9/ 1.5/ 2.0 0.9/ 1.5/ 2.0 2.0
500
Available active measuring 1250/800/
300/ 500 300/ 500 (AuditCell
distances [mm] 500/ 300
300)
* upon request
GE02 GM 31 Gas Analyzer 11/99
GMP31 Probe Selection
n
Temp >550°C
• Calibration with test gases
acc EPA part 75 required ? n n
• High dust concentration ?
• Purge air not allowed ?
y • Gas velocity <1m/s ?
y
y
y
• Temp >260°C ?
y n
• Temp >260°C ?
n
GPP ?
y n
n
- Dry flue gas ?
y - Wet !
- Temp <250°C !
Contents
GE02 GM 31 Gas Analyzer 11/99
GM31 with GPP31 (Gas Permeable Probe)
PT1000
Filter element
Flue gas
PT1000
Filter element
Flue gas
WATLOW
1
160 2
RDY
SET
SET <<
<< >>
>>
PT1000
Gas consumption for 1 test point during gas test from 15 l on depending on the
application. Time necessary for 1 test point during gas test max. 65 s
Overpressure inside of filter element during gas test approx.. 20 hPa
acc. EPA
Equipment
• Test gas cylinders (N2 and measured gases);
concentrations corresponding to measuring range
• pressure gauge / pressure reducer
• leading 1/8‘‘ with connector
• pressure monitor, 1000 hPa, resolution 0.5 hPa
• Laptop with MEPA31 and RS232 cable
• 3.5‘‘ diskette for data recording
Procedure
• Connect the N2 gas.
• Start the MEPA31 programm, and record the current measuring data.
• Adjust pressure reducer to 0 bar, open the valve completely.
• Adjust pressure to 3 bar and wait approx. 30s.
• If reading has not dropped to zero increase pressure in steps until zero is reached.
• Increase pressure by approx. 0.5bar.
• Notice the pressure of the gauge and pressure monitor as default values.
The differential pressure will be between 2mbar and 30mbar.
• Close the valve on the pressure reducer.
• Connect the test gas and adjust the pressure default values.
• Wait for stabilizing of readings and record the measuring values.
• Repeat with other test gas concentrations if necessary.
Contents
GE02 GM 31 Gas Analyzer 11/99
GM31 - Gas Filter Box - Cal. Check
The gas filter box is used for checking the systems with standard probe or
probe with directed purge air.
The test gas needed can be obtained by division of the filter box measuring range
thought the max. length of audit cell (150mm) and, if necessary, conversion to ppm:
It should be avoided to use gases with higher concentrations in order to esure that no
testing out of the calibration range can take place.
Contents
GE02 GM 31 Gas Analyzer 11/99
GM31 - Additional Information
continues
IO - Calculator
Measuring
Reset Status Malfunction
RS232 / Modem
Service
1 2
Service board 1 = Service
2 = Measuring
Contents
GE02 GM 31 Gas Analyzer 11/99
GPP31 - Maintenance - 1
Heating control
and
2 gas connectors
Contents
GE02 GM 31 Gas Analyzer 11/99
GM31 Control Unit TCU - Examples
Example GM 31-2 formulas in TCU:
for NO conversion:
RV02 = MV02 * (273+MV03)/273 * 1013/p [hPa] * 100/(100-H2O[%]) / 1.34
with
For p[hPa] and H2O[%] use either substitute values (fix values stated
by the customer) or values read in via analog inputs Al.
Reckoning values RV01 and RV02 can be proceedeed to the Analog Outputs
e.g. AO 1 and AO 2 of TCU in order to get analog outputs in ppm.