TM VolumatMCAgiliaCE Int LF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 140

MEDICAL DEVICES

Volumat MC Agilia
Technical Manual
Revisions table
TM Volumat MC Agilia_int : ref TM 3897
valid for pumps with serial numbers from 20135026 till…

Date Revision Chapter Description


07/07/08 0 All index "c" only MT Volumat MC AgiliaCreation

15/12/09 1 Index "d" Charte 2009


spare parts catalogue removed from this
document, a document is available for spare
parts.
1.5.5 Update symbol of protection defibrillation

Spelling corrections
09/07/13 2 Index "e" Precautions to be taken for flow rate test were
added.
Table of contents

1 Introduction ...................................................................................................... 7
1.1 General introduction .............................................................................. 7
1.2 Operation diagram ................................................................................. 8
1.3 Precautions for use................................................................................ 9
1.4 Operation safety ..................................................................................... 9
1.5 Technical characteristics .................................................................... 10
1.5.1 Electrical details ....................................................................................................... 10
1.5.2 Electronic details ...................................................................................................... 10
1.5.3 Mechanical details.................................................................................................... 10
1.5.4 Material characteristics ............................................................................................ 10
1.5.5 Conformity to standards ........................................................................................... 11

2 Description and operation ............................................................................. 13


2.1 Physical description ........................................................................... 13
2.1.1 The display board..................................................................................................... 14
2.1.2 CPU board ............................................................................................................... 16
2.1.3 The power supply board........................................................................................... 20
2.1.4 The mains board ...................................................................................................... 23
2.1.5 The air detection board ............................................................................................ 24
2.2 Operational description ....................................................................... 25
2.2.1 Administration set check and maintenance sub-assembly ...................................... 25
2.2.2 Pumping sub-assembly............................................................................................ 25
2.2.3 External connection sub-assembly .......................................................................... 25

3 Description of the menus .............................................................................. 27


3.1 Options menu ....................................................................................... 27
3.2 Calibration menu .................................................................................. 29
3.3 After-Sales Service test menu............................................................. 29
3.4 Maintenance options menu ................................................................. 29

4 Preventive maintenance ................................................................................ 31


4.1 Recommendations ............................................................................... 31
4.2 Maintenance schedule ......................................................................... 31
4.2.1 Preventive maintenance........................................................................................... 31
4.2.2 Quality control .......................................................................................................... 31

TM_VolumatMCAgiliaCE_intTDM.fm 3
4.3 Checks ................................................................................................. 33
4.3.1 Accessing the tests .................................................................................................. 33
4.3.2 Visual check ............................................................................................................. 34
4.3.3 Test 1, identification ................................................................................................. 35
4.3.4 Test 2, Maintenance................................................................................................. 36
4.3.5 Test 3, Maintenance messages ............................................................................... 37
4.3.6 Test 4, Events .......................................................................................................... 37
4.3.7 Test 5, Upstream pressure....................................................................................... 38
4.3.8 Test 6, Downstream pressure .................................................................................. 38
4.3.9 Test 7, Door ............................................................................................................. 39
4.3.10 Test 8, battery information ....................................................................................... 40
4.3.11 Test 9, Signalisation ................................................................................................. 41
4.3.12 Test 10, Keyboard.................................................................................................... 41
4.3.13 Test 11, Ageing test ................................................................................................. 42
4.3.14 Test 12, Battery life .................................................................................................. 43
4.3.15 Test 13, Battery test ................................................................................................. 43
4.3.16 Test 14, Air sensor ................................................................................................... 44
4.3.17 Test 15, Clamp motor............................................................................................... 45
4.3.18 Test 16, Flow rate .................................................................................................... 46
4.3.19 Test 17, Flow rate correction.................................................................................... 47
4.3.20 Test 18, Occlusivity .................................................................................................. 47
4.3.21 Test 19, Clamp ......................................................................................................... 47
4.3.22 Checking the presence limit alarm ........................................................................... 48
4.3.23 Mains/battery operation check ................................................................................. 48
4.3.24 Battery life check ...................................................................................................... 49
4.3.25 Electrical test ............................................................................................................ 49
4.4 Maintenance options ........................................................................... 50
4.4.1 Access to test........................................................................................................... 50
4.4.2 SAV 1, Maintenance date ........................................................................................ 52
4.4.3 SAV 2, Initial set up .................................................................................................. 52
4.4.4 SAV 3, Data log event.............................................................................................. 53
4.4.5 Quality Control Certificate ........................................................................................ 55
4.5 Flow rate check .................................................................................... 57
4.5.1 Measuring by computer............................................................................................ 57
4.5.2 Measuring with a scale............................................................................................. 60
4.5.3 Measuring with a test tube ....................................................................................... 62
4.6 Cleaning and disinfecting ................................................................... 65
4.7 Storage.................................................................................................. 65
4.8 Use of the internal battery.................................................................. 66

5 Troubleshooting..............................................................................................67
5.1 Breakdown guide ................................................................................. 67
5.2 Errors messages.................................................................................. 73

44 TM_VolumatMCAgiliaCE_intTDM.fm
6 Intervention procedures ................................................................................ 77
Procedure N°1: Battery and battery door ................................................. 79
Procedure N°2: Upper case........................................................................ 81
Procedure N°3: Display board ................................................................... 85
Procedure N°4: CPU board ....................................................................... 89
Procedure N°5: Power supply board......................................................... 93
Procedure N°6: Base kit ............................................................................. 97
Procedure n°7 : Air detector board ......................................................... 103
Procedure N°8: Pumping system ............................................................ 107
Procedure N°9: Motor ............................................................................... 113
Procedure N°10: Pressure sensor ........................................................... 119
Procedure N°11: OCS clamp motor......................................................... 125

7 Calibrations................................................................................................... 129
7.1 Calibrations procedure ...................................................................... 129
7.1.1 Access to the calibrations ...................................................................................... 129
7.1.2 Etal 1, Pressures.................................................................................................... 131
7.1.3 Etal 2, Door ............................................................................................................ 133
7.1.4 Etal 3, Air sensor.................................................................................................... 134
7.1.5 Elal 4, Flow rate ..................................................................................................... 135

8 Spare parts catalogue .................................................................................. 137

TM_VolumatMCAgiliaCE_intTDM.fm 5
66 TM_VolumatMCAgiliaCE_intTDM.fm
1 Introduction
1.1 General introduction
Volumat MC Agilia is the advanced volumetric infusion pump in the Agilia range, the family
of intuitive infusion devices developed by Fresenius Kabi.
Volumat MC Agilia offers a wide range of flow rates (from 1 to 1500ml/h in macro mode and
from 0.1 to 100ml/h in micro mode) as well as programming modes that suit all infusion
protocols, from the simplest to the most advanced.
As for all devices of the Agilia range, special attention has been paid to the ergonomics of
the pump in order to make it easy and pleasant to use.
Volumat Agilia can be programmed in three different modes:
„ No drug name,
„ Drug labelling,
„ Vigilant Drug'Lib.
The drop detector is optional. The pump is equipped with a patented auto-test called "OCS
test" (Occlusivity Check System) that checks the correct working order of the pump in
association with its administration set, thus preventing any risk of free flow.
The pump works with the Volumat Lines: these dedicated sets are equipped with a silicone
pumping segment and a SafeClip that automatically clamps the line to avoid any risk of free-
flow when the door is opened.
The Volumat MC Agilia pump has its own control keyboard and a LCD screen. It has an
internal battery which ensures a minimum battery life of 8 hours for a flow rate of 125ml/h.
A large choice of configuration parameters enables optimal use of the functions according to
the needs of each department.
Volumat MC Agilia is fitted with an infrared cell at the back. Fresenius Kabi also offers a
Link+ Agilia system enabling several Agilia pumps to communicate to a communications
system via the infrared cell and Link+ Agilia.
Volumat MC Agilia can be:
„ used as a stand alone device thanks to the mounted "swinglock" clamp that allows its
fixation on a rail or a pole.
„ stacked together with other Agilia devices thanks to its integrated locking system allowing
three Volumat MC Agilia pumps to be stacked on top of one another to facilitate
transportation. Easy to carry, Volumat MC Agilia is ideal for transporting patients.
Introduction

va01_002e_int.fm 7
1.2 Operation diagram
Introduction

8 va01_002e_int.fm
1.3 Precautions for use

The ! symbol visible on the device, indicates that the relevant operator guide should be
read thoroughly.
Volumat MC Agilia has been tested in accordance with the standards applicable to
electromagnetic compatibility of the medical devices. Its immunity makes helps to ensure
correct operation. The limitation of the radiation emitted avoids undesirable interference with
other equipment such as EEG, ECG, etc. If a Volumat MC Agilia pump is placed near
devices such as surgical equipment HF, X-rays, mobile phones or Wifi points, then minimal
distances between equipment must be respected.
The device must not be used in the presence of inflammable anaesthetic agents due to a risk
of explosion. It should always be used away from all risk areas.
The device could be disturbed by pressure or pressure variations, mechanical shocks, heat
ignition sources, etc. Please contact our After-Sales Department, if you wish to use devices in
specific conditions. The pump must be used in a horizontal and stable position to work
correctly.
The physiological effects of medicine may be influenced by the characteristics of the device
and the associated tubing (the constitution material is normally listed on the tubing
packaging). Check that the prescriptions are compatible with the characteristics of the
trumpet curves and with the occlusion alarm setting times in relation to the programmed flow
rate.
The device uses a Lithium Ion Rechargeable battery. Incorrect handling of a Lithium Ion
battery by a non qualified personnel may cause leakage, heat, smoke, explosion or fire, which
could result in deterioration of performance or failure. This may also damage the protection
device installed in the battery pack, resulting in damage to the equipment or injury to the user.

Fresenius Kabi cannot, in any circumstances, be held responsible for any medical or other
problem due to incorrect use of the device.
Kindly consult the Instructions for use manual for more details.
Use conditions:
„ Temperature : 5°C to 40°C / 41°F to 104°F.
„ Humidity : 20% to 90%, no condensation.
„ Atmospheric pressure : 700 hPa to 1060 hPa.

1.4 Operation safety


As soon as it is in operation, the device ensures continuous surveillance of its functions. Any
internal fault or any procedural anomaly is immediately detected. Nevertheless, abnormal
functioning of the device, without a defined cause, must always be brought to the attention of
the qualified staff in your establishment or our Technical Service.
Volumat MC Agilia pump is equipped with an internal battery that ensures normal
functioning during a mains power cut. In addition, 2 safety fuses protect the electrical mains
concerned.
Introduction

va01_002e_int.fm 9
1.5 Technical characteristics

1.5.1 Electrical details


„ Power supply : 100V - 240V~/ 50-60Hz with functional earth
„ Max. consumption : 180mA
„ Max. power : 15VA
„ Fuse : T1AH 250V - 2xAT internal to power supply unit
„ Battery : 7.2V - 2.2Ah. (LI-ion)
„ External power supply : 9V DC. Power > 15W.
„ Communication port and 11-30 V DC power.

1.5.2 Electronic details


Volumat MC Agilia pump contains 5 electronic boards:
„ Power supply board
„ Mains board
„ CPU board
„ Display board
„ Air detection board

1.5.3 Mechanical details


„ Dimensions: H x W x D: 135 x 190 x 170mm
„ Weight: approximately 2kg

1.5.4 Material characteristics

Components Materials
1 Screw cap (2) Polyamide
2 Nut bolt button Polyamide
3 M8 moulded screws Polyamide
4 Angle bracket ABS
5 Battery door ABS
6 Right and left flange Polycarbonate
7 Pumping block plate Polyamide
8 Moulded cover over light guide ABS
9 Light guide Polycarbonate
10 Window Polycarbonate
11 Moulded base ABS
12 Base right tube guide Polyamide 6
13 Base left tube guide Polyamide 6
14 Cover central hook Polyamide 6
15 Cover Right and left hook Polyamide 6
16 Infrared window Polycarbonate
17 Excentric bolt POM
18 Bolt front case POM
Introduction

19 Bolt rear case POM


20 Bolt button POM
21 Bolt button pen POM

10 va01_002e_int.fm
Components Materials
22 Air detector support ABS
23 Door ABS
24 Door plate Polyarylamide
25 Moulded lever + hooks Polyarylamide
26 Right and left door hinge Polyamide 6
27 OCS fuse Polyacetal
28 OCS unclamping finger Polyamide
29 Mechanical spring cap Polyamide
30 OCS detector cradle Polyarylamide
31 Pumping block Polyamide
32 Star motor ring Polyacetal
33 Motor pulley Polyamide
34 Cam shaft pulley Polyamide
35 Belt idler pulley Polyamide
36 Belt idler lever Polyacetal
37 Rotation control disk Polyamide
38 Pumping finger PA6-12
39 Pumping block Polyarylamide
40 Moulded cam shaft PA66
41 Bearing motor support Polyarylamide
42 Magnet support ABS
43 Membrane TF3 STE
44 Membrane frame Polypropylene
45 Back foot LSR
46 Front foot LSR
47 RJ11 stopper TPE EPTR
48 Connector stopper TPE EPTR
49 Star motor shock absorber Elastomere

1.5.5 Conformity to standards

Conform to the 93/42/CE Medical Directive. IP22 Protection against


0459 splashing liquid.
Safety of Electro Conform to EN/IEC 60601-1 and EN/IEC 60601-2-24 Protection against
Medical leakage current:
Equipement Defibrillation-proof type
CF applied part.
EMC Conform to EN/IEC 60601-1-2 and EN/IEC 60601-2-24 Protection against
(ElectroMagnetic electric shocks: class II.
Compatibility) Functional earth.
The functional earth is directly connected to the mains plug. It reduces the residual current
which could cause interference to ECG or EEG devices.

Detailed information concerning electromagnetic compatibility is available in the


Introduction

"Guidance and manufacturer’s declaration on EMC" chapter of the Instructions For Use
manual.

va01_002e_int.fm 11
Introduction

12 va01_002e_int.fm
2 Description and operation
2.1 Physical description

Angle bracket
Air bubbles detector Cover

Base Door

Control panel

Volumat MC Agilia pump is made up of 3 main parts: an angle bracket, a cover and a base.
„ The cover contains:
† a display board.
„ The base supports:
† a CPU board,
† an air bubble detector board,
† an OCS system,
† a pump motor mechanism,
† an air bubble detection,
Description and operation

† a clamp system.
„ The angle bracket supports:
† a power supply board,
† a mains board,
† a battery.

va02_002e_int.fm 13
2.1.1 The display board
The display board is situated underneath the front of the command panel and contains all the
essential parts for man-machine interaction:
„ Keyboard
„ Synoptics and control indicators
„ LCD screen
„ Buzzer

J2
J1
J3

Display board
This board is connected to the other parts by means of connectors.

J1 connector to keyboard

Pin Signal Description


1 A_LED Ground
2 COL0 Keyboard column 0
3 COL1 Keyboard column 1
4 COL2 Keyboard column 2
5 LIG0 Keyboard line 0
6 LIG1 Keyboard line 1
7 LIG2 Keyboard line 2
8 ON/OFF Keyboard ON/OFF key
9 CTR Keyboard control or line 3
10 GND Ground
Description and operation

J2 connector to CPU board

Pin Signal Description


A1 GND Ground
B1 VBAT Battery power supply or "+10V mains"
A2 +3V3- Diagnostic signal (correct boards connection)
PERM_PROT
B2 - Not connected

14 va02_002e_int.fm
J2 connector to CPU board

Pin Signal Description


A3 GND Ground
B3 +3V3 +3.3V power supply
A4 TON/OFF Keyboard ON/OFF key
B4 MISO SPI bus MISO signal
A5 GND Ground
B5 MOSI SPI bus MOSI signal
A6 GND Ground
B6 CLK SPI bus clock signal
A7 GND Ground
B7 RESET/ Control signal
A8 RSLCD Control signal (intended for LCD display)
B8 CSLCD/ Control signal
A9 CSLDCD/ Control signal
B9 CD_BuZ_AFF Buzzer command
A10 GND Ground
B10 S_ANA_DOOR Analogue signal sent by "DOOR" sensor
A11 +3V3 +3.3V power supply
B11 GND Ground
A12 - Not connected
B12 PRES_CARTES Diagnostic signal (correct board connection)
A13 VBAT Battery power supply or "+10Vmains"
B13 GND Ground

J3 to display connector

Pin Description
1 A_LED Backlight LED anode
2 C_LED Backlight LED cathode
3 - Not connected
4 - Not connected
5 GND Ground, indicating serial access to LCD
6 MOSI LCD serial data line
7 CLK LCD serial line clock
8 GND Not used by the LCD in serial access
9 GND Not used by the LCD in serial access
10 GND Not used by the LCD in serial access
11 GND Not used by the LCD in serial access
12 GND Not used by the LCD in serial access
13 GND Not used by the LCD in serial access
14 GND Not used by the LCD in serial access
Description and operation

15 GND Not used by the LCD in serial access


16 RSLCD Command selection /LCD data
17 RESET/ LCD reset
18 CSLCD/ LCD Chip select
19 +3V3 +3.3V power supply
20 GND Ground

va02_002e_int.fm 15
2.1.2 CPU board
The CPU board is made up of a main micro controller, FLASH and RAM memories and
various analogue or power functions, namely:
„ Two single step bipolar motor commands
„ A power supply cut out - elevator type (BOOST)
„ Adjustment of and acquisition from different sensors
„ Different lines of communication (synchronous or non-synchronous serial link, optional
Bluetooth connection).
The CPU board is connected to the display board by means of the J2 connector.
A ribbon cable connects it to the power supply board by means of the J14 connector.

J11

J7 J3 J10

J4 on the inside
J13 J12

J9

J14
J2 J1

CPU board

J1 connector to pumping motor

Pin Description
1 +BOBINEA + motor A coil
2 -BOBINEA - motor A coil
3 +BOBINEB + motor B coil
4 -BOBINEB - motor B coil

J2 to OCS motor connector

Pin Description
1 +BOBINEA + motor A coil
Description and operation

2 -BOBINEA - motor A coil


3 +BOBINEB + motor B coil
4 -BOBINEB - motor B coil

16 va02_002e_int.fm
J3 connector to downstream pressure sensor

Pin Description
1 GND Ground
2 +3V3_PS Power supply +3.3V filtered for pressure sensor
3 V+ + Pressure sensor resistance bridge
4 GND Ground
5 V- - Pressure sensor resistance bridge
6 GND Ground

J4 connector to upstream pressure sensor

Pin Description
1 GND Ground
2 +3V3_PS Power supply +3.3V filtered for pressure sensor
3 V+ + Pressure sensor resistance bridge
4 GND Ground
5 V- - Pressure sensor resistance bridge
6 GND Ground

J7 to CLAMP connector board

Pin Description
1 GND Ground
2 MC_OPTO2/ Opto 2 status
3 MC_CD_OPTO Opto 2 power supply
4 GND Ground
5 MC_OPTO1/ Opto 1 status
6 MC_CD_OPTO Opto 1 power supply
7 GND Ground
8 /S_CLAMP CLAMP opto status
9 /S_CD_CLAMP CLAMP opto power supply
10 GND Ground

J9 to JTAG connector (micro programming)

Pin Description
1 +3V3 +3.3V power supply
2 +3V3 +3.3V power supply
3 TCK "Input" JTAG port signal
4 TDO "Output" JTAG port signal
5 TRST/ "Input" JTAG port signal
Description and operation

6 TDI "Input" JTAG port signal


7 MR/ "Input" JTAG port signal
8 GND Ground
9 TMS "Output" JTAG port signal
10 GND Ground

va02_002e_int.fm 17
J10 connector to opto motor

Pin Description
1 GND Ground
2 MP_OPTO/ Motor opto rotation state
3 MP_CD_OPTO Opto power supply
4 GND Opto power supply return

J11 connector to display board

Pin Description
A1 GND Ground
B1 VBAT Battery power supply or + "10V mains"
A2 +3V3- Diagnostic signal (correct boards connection)
PERM_PROT
B2 - Not connected
A3 GND Ground
B3 +3V3 Power supply +3V3
A4 TON/OFF Keyboard ON/OFF key
B4 MISO SPI bus MISO signal
A5 GND Ground
B5 MOSI SPI bus MOSI signal
A6 GND Ground
B6 CLK SPI bus clock signal
A7 GND Ground
B7 RESET/ Control signal
A8 RSLCD Control signal (intended for LCD display)
B8 CSLCD/ Control signal
A9 CSLDCD/ Control signal
B9 CD_BUZ_AFF Buzzer command
A10 GND Ground
B10 S_ANA_DOOR Analogue signal sent by the "DOOR" sensor
A11 +3V3 Power supply +3.3V
B11 GND Ground
A12 - Not connected
B12 PRES_CARTES Diagnostic signal (correct boards connection)
A13 VBAT Battery power supply or "+10V mains"
B13 GND Ground

J12 connector to drop detector

Pin Description
Description and operation

1 GND Ground
2 PULSE_goutte Sensor return signal
3 GND Ground
4 GND Ground
5 CDLED_goutte Sensor power supply
6 GND Ground

18 va02_002e_int.fm
J13 connector to air detector

Pin Description
1 GND Ground
2 DA_ANA_AIR Analogue signal sent by the air detector
3 GND Ground
4 DA_DISCH Air detector reset command
5 +3V3 +3.3V power supply
6 DA_GAIN Air detector gain adjustment command
7 +3V3 +3.3V power supply
8 DA_SCL I²C bus clock signal between main micro controller
and air detector
9 DA_SDA I²C bus data signal between main micro controller
and air detector oscillator
10 GND Ground

J14 connector to power supply board

Pin Description
A1 GND Ground
B1 VBAT Battery power supply or "+10V mains"
A2 +3V3-PERM Power supply +3.3V backup for date stamper
B2 - Not connected
A3 GND Ground
B3 +3V3 +3.3V power supply
A4 TON/OFF Keyboard ON/OFF key
B4 GND Ground
A5 SCL_ALIM I²C bus clock signal between main micro controller
(CPU) and secondary (Power supply)
B5 SDA_ALIM I²C data signal between main micro controller
(CPU) and (Power supply)
A6 GND Ground
B6 RXIR Infrared cell reception signal
A7 TXIR Infrared cell emission signal
B7 GND Ground
A8 SCK Serial bus clock signal
B8 GND Ground
A9 RXD External plug data reception signal
B9 TXD External plug data emission signal
A10 GND Ground
B10 CD_BUZZ_ALIM Buzzer command
A11 +3V3 +3.3V power supply
Description and operation

B11 GND Ground


A12 - Not connected
B12 PRES_CARTES Diagnostic signal (good board connection)
A13 VBAT Battery power supply or "+10V mains"
B13 GND Ground

va02_002e_int.fm 19
2.1.3 The power supply board
The power supply board is made up of a micro controller (ATmega48, as a secondary in
relation to the one on the CPU board and in relation to its attributed functions) and the
different analogue or power functions, namely:
„ Management of the power supply for all the boards
„ A battery charger
„ An Infrared interface
„ Different lines of communication (internal and external to the boards)
„ A buzzer
The power supply board is situated on the angle bracket. It is used to supply power to the
electronic unit from the 100V - 240V CA network or, externally, from 9V CC. It also charges a
2.2Ah battery.

J1

J2 on the inside

circular J4 on the inside


connector

J6
J3 on the inside

Power supply board


This board is connected to the other parts by means of connectors.

J1 to CPU board connector

Pin Description
A1 GND Ground
B1 VBAT Battery power supply or "+9V mains"
A2 +3V3-PERM Power supply +3.3V backed up for the time stamper
B2 - Not connected
A3 GND Ground
Description and operation

B3 +3V3 Power supply +3.3V


A4 TON/OFF Keyboard ON/OFF key
B4 GND Ground
A5 SCL_ALIM Signal Clock of the bus I²C clock signal between main
(CPU) and secondary micro controller (Power supply)
B5 SDA_ALIM Signal Data of the I²C bus data signal between main
(CPU) and secondary (Power supply) micro controller
A6 GND Ground

20 va02_002e_int.fm
J1 to CPU board connector

Pin Description
B6 RXIR Infrared cell reception signal
A7 TXIR Infrared cell emission signal
B7 GND Ground
A8 SCK Serial bus clock signal
B8 GND Ground
A9 RXD External plug data reception signal
B9 TXD External plug data emission signal
A10 GND Ground
B10 CD_BUZ_ALIM Buzzer command
A11 +3V3 +3.3V power supply
B11 GND Ground
A12 - Not connected
B12 PRES_CARTES Diagnostic signal (good board connection)
A13 VBAT Battery power supply or "+9V mains"
B13 GND Ground

J2 connector to mains connector

Pin Signal Description


1 Neutral Mains power supply
2 Phase Mains power supply
3 GND Ground

J3 connector to secondary micro controller

Pin Signal Description


1 MISO “Output” SPI programming signal
2 +3V3 +3.3V power supply
3 SCK “Input” SPI programming signal
4 MOSI “Input” SPI programming signal
5 /RST “Input” SPI programming signal
6 GND Ground

J4 connector to internal battery

Pin Signal Description


1 +BATTERIE +Battery
2 ANA_CTN +CTN
3 GND Ground battery and CTN reference
Description and operation

va02_002e_int.fm 21
J5 connector to external plug

Pin Signal Description


1 SCK Serial connection clock
2 TXD Serial connection data emission
3 RXD Serial connection data reception
4 +5V Accessories power supply (supplied by the board)
5 GND Ground
6 RELAIS Relay command (open collector)
7 GND Ground
8 +10V0 External power supply (supplied by the board)
9 GND Ground
10 GND Ground
Description and operation

22 va02_002e_int.fm
2.1.4 The mains board
The mains board contains the mains plug, the ECG filter as well as F1 and F2 fuses.
It is the interface between the mains and the power supply board. The mains board is situated
on the angle bracket. It enables the power supply board to be changed without the need for
soldering.

J2

J1 on the inside
F2 F1

Mains board
This board is connected to the other parts by means of connectors.

J1 to power supply board connector

Pin Signal Description


1 Neutral Mains power supply
2 Phase Mains power supply
3 GND Ground

J2 connector to mains

Pin Signal Description


1 Neutral Mains power supply
2 Phase Mains power supply
3 GND Functional earth

Description and operation

va02_002e_int.fm 23
2.1.5 The air detection board
The air detection board is made up of an input amplifier and an output amplifier which are
linked to the ceramic elements for reception and emission.
The air detection board is fixed to the base. It transmits the air bubbles volume in the
administration set in the form of analogue data to the CPU board.

J1
J3 J2

J2

Air detection board


This board is connected to the other parts by means of connectors.

J1 to CPU board connector

Pin Description
1 GND Ground
2 DA_ANA_AIR Analogue signal sent by the air detector
3 GND Ground
4 DA_DISCH Air detector command reset
5 +3V3 +3.3V power supply
6 DA_GAIN Air detector adjustment gain command
7 +3V3 +3.3V power supply
8 DA_SCL I²C bus clock signal between main micro controller
and air detector oscillator
9 DA_SDA I²C bus data signal between main micro controller
and the air detector oscillator
10 GND Ground

J2 to emission connector
J3 to receiver connector
Description and operation

24 va02_002e_int.fm
2.2 Operational description
From an operational point of view, the Volumat MC Agilia pump is made up of 3 sub-
assemblies:
„ a sub-assembly to maintain and to check the correct positioning of the administration set.
„ a motorisation sub-assembly.
„ an external connection sub-assembly.

2.2.1 Administration set check and maintenance sub-assembly


The administration set is positioned along the front part panel held in place by the door.
The front part is equipped with four detection systems:
„ A detector to check the position of the door (closed /open).
„ A photoelectric cell to detect the presence of the administration set clamp.
„ An ultra-sound detector to detect the presence of air bubbles in the administration set.
„ Detection of the pressure is ensured by two piezo-resistant sensors.

2.2.2 Pumping sub-assembly


The pumping sub-assembly contains the peristaltic pumping mechanism.
The mechanism is made up of a camshaft that creates a wave-like movement in a bank of
12 "fingers". The displacement of these ’fingers’ is controlled by the CPU board, while turning
it displaces the liquid at the programmed flow rate.
The camshaft is driven by a stepper-motor and a reduction gearbox assembly via a toothed
belt.
A photoelectric cell and a rotation palette are situated on the camshaft end. To motor/pumping
system control.

2.2.3 External connection sub-assembly


Volumat MC Agilia has 2 plugs situated at the back of the angle bracket:
„ a mains supply plug
„ an 8 pins Binder plug which, together with various specific Fresenius Kabi accessories,
enables the following functionalities:
† Continuous external power supply
† Nurse call
† Serial link.
An infrared cell is located at the back of the device. It permits information to be exchanged
with the Link+ Agilia: the information can then be transmitted by dedicated communication
cables.
Description and operation

va02_002e_int.fm 25
Description and operation

26
va02_002e_int.fm
3 Description of the menus
3.1 Options menu
The options menu enables the Volumat MC Agilia to be adapted to the specific needs of
each department. It accesses the menus that customise the infusion parameters.
Fresenius Kabi recommends the presence of its qualified staff, or a member of your
Technical Department, who will assist you in implementing the configuration procedures you
wish to select.

The options mode is activated


by holding down the "Menu" key when the device is switched on.

The following options menu is displayed:


„ User,
„ Service,
„ Maintenance.

Each menu is made up of sub-menus.


The sub-menus can be accessed by using the "ENTER" key.

To access some of the sub-menus, you need to type in an access code. This code can be
defined and/or de-activated in the Service mode. As long as the device is kept switched on,
the code does not need to be keyed in again.

Main Menu Key in code Sub-menu Sub-menu


User
User options menu
Service Code 0200
Service options menu
Maintenance Code XXXX
Test
Maintenance options test
menu
Maintenance options
Maintenance options menu
Calibration
Calibration menu
Description of the menus

Only the maintenance menu is described in this document.


For more information concerning the other menus, kindly consult the
Instructions For Use manual.

va03.1_002e_int.fm 27
Description of the menus

28
va03.1_002e_int.fm
Description of the menus

3.2 Calibration menu

The calibration menu is reserved for qualified staff only. Its access is protected by a
code. To find out about the operating mode for the various calibrations, refer to the
"Calibrations" chapter.
This menu enables the calibration of:
„ Etal 1, Pressures,
„ Etal 2, Door,
„ Etal 3, Air sensor,
„ Etal 4, Flow rate.

3.3 After-Sales Service test menu


The After-Sales Service test menu is reserved for qualified personnel. It enables a series of
checks that can be carried out on the Volumat MC Agilia to confirm that it is operating
correctly. (See chapter "Checks".)

The After-Sales Service tests can also be carried out more quickly and simply from
a PC using our maintenance software. (Contact our Technical Service department
for more details.)
The After-Sales Service test menu contains a series of 19 tests or checks:
„ Test 1, Identification,
„ Test 2, Maintenance,
„ Test 3, Maintenance messages,
„ Test 4, Events,
„ Test 5, Upstream pressure,
„ Test 6, Downstream pressure,
„ Test 7, Door,
„ Test 8, Battery information,
„ Test 9, Indication,
„ Test 10, Keyboard,
„ Test 11, Ageing test,
„ Test 12, Battery life,
„ Test 13, Battery test,
„ Test 14, Air sensor,
„ Test 15, Clamp motor,
„ Test 16, Flow rate,
„ Test 17, Flow rate correction,
„ Test 18, Occlusivity,
„ Test 19, Clamp.
Description of the menus

3.4 Maintenance options menu


Maintenance options menu allows to enter parameters for the 3 following options:
„ SAV 1, Maintenance date,
„ SAV 2, Initial set up,
„ SAV 3, Data log event.

va03.3&4_002e_int.fm 29
Description of the menus

30
va03.3&4_002e_int.fm
4 Preventive maintenance
4.1 Recommendations
The device can only be checked, serviced or repaired by Fresenius Kabi or by a certified and
approved maintenance service. Any abnormal functioning of the pump must be brought to the
attention of your in-house qualified technical personnel or our Technical Service.
Should you need to return the pump to our Technical Service, it should be cleaned,
disinfected and very carefully packaged, preferably in its original packaging, before being
sent.
For all information concerning the repair and use of the pump, kindly contact our Technical
Service or our Sales department.
Fresenius Kabi is not liable for loss or damage to the pump during its transport to our
Technical Service.

4.2 Maintenance schedule

4.2.1 Preventive maintenance


In order to maintain the pump’s performance, a Preventive Maintenance inspection must be
carried out every 3 years. This procedure, which includes changing the battery, replacing the
membrane, should be carried out by a qualified technician.
Any abnormal functioning or failure must be reported to the qualified technical staff in your
organisation or to our Technical Service. In these instances, the pump should not be used.

Important:
If these maintenance procedures are not observed, the pump’s correct operation will
be impaired.

4.2.2 Quality control


At the request of the health organisation, a quality check will be carried out every 12 months.
A quality check (not included in the guarantee) consists of different inspection procedures as
defined in the pump’s Technical Manual. Only a qualified technician may perform the quality
check which must be performed using Fresenius Kabi software.
For more information, kindly contact our Technical Service.

Recycling of obsolete batteries and devices: Before disposal, remove


battery from the device. Batteries and devices with this label must not be
disposed of with the general waste. They must be collected separately
and disposed of according to local regulations. For further information
Preventive maintenance

pertaining to waste processing regulation, contact your local Fresenius


Kabi.

va04.1&2_001e_int.fm 31
Preventive maintenance

32 va04.1&2_001e_int.fm
Maintenance préventive

4.3 Checks
A quality control certificate is available at the end of this section.

In order to ensure the smooth running of the checking procedure, recharge the
battery for 16 hours beforehand.

4.3.1 Accessing the tests

Keyboard description.
Key Function
ON switches on the pump.
OFF switches it off when pressed for over three seconds.
SILENCE ALARM

STOP stops infusion.

Menu/History accesses the test mode.

The selection keys scroll through the numbers and letters on the
tenths, units and tens values and allow one to move to the next
menu.
OK validates your test selection or moves the cursor.
enter starts up infusion.
exit exits the screen and returns to the "previous" menu.
Bolus or prime

C/Back, returns to the "previous" menu.

Maintenance préventive

va04.3_002e_int.fm 33
Moving to the After-Sales Service test mode.
„ Press "Menu/History" key.
„ Keep it pressed down and press "ON/OFF".
„ When the "Options" main menu displays,
release "Menu/History" and "ON/OFF" keys.

„ Choose your "Maintenance" menu using the central


selection keys.
„ Press "enter" to confirm.

„ The "Maintenance" menu displays.


„ Key in the code.

„ Press "OK" to confirm.

„ Choose the "Test" menu using the central selection keys.

„ Press "enter" to confirm.


† Navigate through this sub-menu using the central
selection keys.

You may exit the configuration mode at any time by pressing


"Menu/History"; this function cancels the modifications.

4.3.2 Visual check


Check the general appearance of the case and labels and check for shock impact damage.
Maintenance préventive

34 va04.3_002e_int.fm
4.3.3 Test 1, identification
This test displays:
the device information screen, serial numbers screen, language information screen, zone
information screen and boot information screen.

„ Test 1, press "enter".


† The main menu "Identification" displays.

„ Scroll through the information using the central selection


keys.
† device information:
Œ device name
Œ version/revision and software production date
Œ device code and product serial number
† serial numbers:
Œ CPU board serial number
Œ power supply board serial number
† language information:
Œ language name, version and revision
Œ date the language file was created
† zone information:
Œ zone parameter number, version and revision
Œ date the parameter file was created
Œ date of the last manual modification
† boot information:
Œ boot version/revision
Œ boot creation date
„ Press "exit" to return to the "Tests" menu.

Maintenance préventive

va04.3_002e_int.fm 35
4.3.4 Test 2, Maintenance
This test displays:
the running time since the last maintenance service in hours, days or months; the date of the
last maintenance service; the total running time in hours, days or months and the number of
cycles executed by the pumping unit.

„ Test 2, press "enter".


† The main "Maintenance" menu displays.
† Scroll through the information using the selection keys.
† Choose between the proposed options "Clear"
or "Exit".

„ "Clear", using the left selection key.


The following "Maintenance" menu displays.

For this menu:


† Running time:
Œ the number of operating hours is reset to "0".
† Date:
Œ the date of the last maintenance is reset to the day’s date.
† Total running time:
Œ the total running time since the beginning remains
incremented.
† Total distance:
Œ the number of cycles executed since the start remains
incremented.

„ "Exit", leaves the screen and returns the user to the last "Test" menu.
Maintenance préventive

36 va04.3_002e_int.fm
4.3.5 Test 3, Maintenance messages
This test displays information concerning the Technical Service interventions.

Test 3, press "enter".


† The "Messages" main menu displays.

„ Scroll through the information, from screen to screen, using


the selection keys.

„ Press "exit" to return to the "Test" menu.

It is necessary to use "Product maintenance software" to write messages in Test 3.

4.3.6 Test 4, Events


This test displays the list of the alarms stamped with date and time.

„ Test 4, press "enter".


† The "Events" main menu displays.

„ Scroll through the information, from screen to screen, using


the selection keys.

„ Press "enter" to display the alarm type.

„ Press "Menu/History" to return to the "Test" menu.


Maintenance préventive

va04.3_002e_int.fm 37
4.3.7 Test 5, Upstream pressure
This test displays the upstream pressure value in mV, the calibration values of the different
upstream pressures in mV, the date and calibration counter.

„ Test 5, select "enter".


† The "Upstream pressure" main menu displays.

„ Scroll through the information, from screen to screen, using


the selection keys.
† Conversion:
Œ Sensor value in mV
Œ Calibration door open in mV
Œ Calibration 0 mmHg in mV
Œ Sensor zero adjustment offset in LSB
† Date:
Number:
Œ Date and number of calibration.

„ Select "exit" to return to the "Test" menu.

Refer to chapter 7, "Calibration / Calibration procedure / Etal 1 pressure", if one of the


values is out of range.

4.3.8 Test 6, Downstream pressure


This test displays:
the downstream pressure value in mV, the calibration values for the different downstream
pressures in mV, the date and calibration counter.

„ Test 6, press "enter".


† The main menu "Downstream pressure"
displays.

„ Scroll through the information, from screen to screen, using


the selection keys.
† Conversion:
State:
Maintenance préventive

Œ Sensor value mV,


Œ Door open calibration in mV,
Œ Calibration 0 mmHg in mV,
Œ Calibration 760 mmHg in mV,
Œ Sensor zero adjustment offset in LSB,
† Date:
Number:
Œ Date and number of calibrations.

38 va04.3_002e_int.fm
„ Select "exit" to return to the "Test" menu.

Refer to chapter 7, "Calibration / Calibration procedure / Etal 1 pressure", if one of the


values is out of range.

4.3.9 Test 7, Door


This test displays:
the door value in mV and its state, the calibration values in mV, the date and calibration
counter.

„ Test 7, press "enter".


† The main menu "Door" displays.

„ Scroll through the information, from screen to screen, using


the selection keys.
† Conversion
State
Œ Door open calibration value in mV,
Œ Intermediary door calibration value in mV,
Œ Door closed calibration value in mV.
† Date,
Number:
Œ Calibration date,
Œ Number of calibrations,

„ Press "exit" to return to the "Test" menu.

Refer to chapter 7, "Calibration/Calibration procedure / Etal 2 door, if one of the values is


out of range.
Maintenance préventive

va04.3_002e_int.fm 39
4.3.10 Test 8, battery information
This test displays:
the battery running time displayed in hours, days and months;
the number of discharge cycles equivalent to the number of hours of battery use / 7h;
the start date for the battery running time counter; the battery voltage in mV and
in 1/100 volts; the battery temperature in mV and in 1/10°C.
The current charge/discharge in LSB and mA.

„ Test 8, select "enter".


† The "Battery info" main menu displays.
† Choose between the "Clear"
and "exit" options.

Select "Clear" to reset to 0 after all battery replacements.

„ "Clear", using the left selection key

† The "Battery Info" menu displays, as shown.

† Duration:
Œ the number of battery operating hours is reset to "0",
resetting the number of cycles to"0".
† Cycles:
Œ the number of cycles is reset to "0".
† Start date:
Œ Only the total battery running time since the beginning
remains incremented.

„ Scroll through the information, from screen to screen, using


the selection keys.
† Conversion:
Voltage:
Œ Voltage value at the battery terminals and conversion
in mV.
† Conversion:
Temperature:
Œ battery temperature value and conversion in mV.
† Conversion:
Maintenance préventive

Current:
Œ the value of the current consumed or on charge and
its conversion in LSB.

„ Press "exit" to return to the "Test" menu.

40 va04.3_002e_int.fm
4.3.11 Test 9, Signalisation
This test checks that the indicator lights and buzzer are functioning correctly.

„ Test 9, press "enter".


† The main screen displays.
† All LED and the mains LED are lit up.

„ Press "OK" to start the test.


† The "Signalisation" main menu displays.
Œ Check that the LED, LCD and the two buzzer tones
are functioning correctly.

„ "exit", exits the screen and returns to the "Test" menu for the
entire duration of the test.

4.3.12 Test 10, Keyboard


This test checks that the keyboard is functioning correctly.

„ Test 10, press "enter".


† The "Keyboard" main menu displays.

„ Press all the keys one by one, for example:


† Press "Bolus/Purge"
Œ the corresponding circle on the screen becomes dark.

„ Press the "Menu" key for a long time to exit the screen and
return to the "Test" menu for the entire duration of the test.
Maintenance préventive

va04.3_002e_int.fm 41
4.3.13 Test 11, Ageing test
This text checks the motor block/pumping system is functioning correctly as well as the pump
occlusivity.

„ Test 11, press "enter".


† The main "Ageing test" menu displays.

When the pump is fitted with an administration


set, open the administration set clamp by pressing
the "Bolus" key.

„ Select your submenu using the "OK" key.


† Flow rate: between 0.1 and 1200ml/h
Modify the value using the following selection keys:
Œ for the hundreds and tens, use the double arrows to
increase or decrease
Œ for the units, increase or decrease using the single
arrows.
† Press "OK" to confirm and to move to the next step.

† Mode: Go forward or synchro or go forwards/backwards


or backwards
Modify your choice using the single arrow selection keys.
† Press "OK" to confirm and move to the next step.

† Steps numbers or turns number:


Content or value lying between 1 and 9999
Modify the value using the single arrow selection keys
† Press "OK" to confirm and move to the next step.

† Current: high, low, auto or error.


Modify the value using the single arrow selection keys,
† Press "OK" to confirm and move to the next step.

„ Press "OK" to confirm and start up the test.

† The following, "Ageing test", menu displays:


Œ The flow rate and current category for the
motor command are displayed at the top
of the screen.
Œ The number of steps decreases according to
the actual change for the number of steps.

„ Press "exit" to return to the "Test" menu.


Maintenance préventive

During the test, the rotation control is activated and the device sets in error if a problem is
detected.
In a test with a defined number of steps/cycles or using a synchro mode, the test ends
when the requested position is reached.

42 va04.3_002e_int.fm
4.3.14 Test 12, Battery life
This test displays:
the amount of battery charge; the date of the last automatic extinction; the pump’s battery
running time before the pre-alarm and battery alarm were triggered; the alarm’s running time
till the battery is off.

„ Test 12, press "enter".


„ The "Battery life" main menu displays.

„ Scroll through the information, from screen to screen, using


the selection keys.
† Charge
Auto extinction
† Pre-alarm
Alarm
Cut-out
„ Press "exit" to return to the"Test" menu.

"Test 12" always displays information concerning the previous discharge. If the battery is
replaced, all the parameters on test 8 must be reset to zero.

4.3.15 Test 13, Battery test


This test discharges the battery completely and measures the time taken for the alarms and
pre-alarms to be triggered, as well as the duration of the power cut.

„ Test 13, press "enter".


„ If the pump is connected to the mains,
the "Battery test/Disconnect mains" menu displays.

Carrying out the battery test does not require the presence
of an operator to note the battery life.
At the end of the battery life test, the device switches off.
To read the battery life, connect the device
to the mains power and select "test 12" :
the battery life should be > 6h30.
Maintenance préventive

Once the battery life has been obtained, test 13 accelerates the battery discharge. If the
battery is changed, all the parameters for test 8 must be reset to zero.

va04.3_002e_int.fm 43
4.3.16 Test 14, Air sensor
This test displays:
the air detector value in mV, the calibration values in mV, the date and calibration counter.

„ Test 14, press "enter".


† The main "Air sensor" menu displays.

„ Scroll through the information, from screen to screen, using


the selection keys.
† Conversion:
State: water/air
Œ Air sensor value in mV
Œ Calibration without administration set in mV
Œ Calibration with administration set and air in mV
Œ Calibration with administration set and water in mV
Œ Frequency during calibration
Œ Frequency shift variance.
† Date:
Number:
Œ Date and number of calibrations.

„ Press "exit" to return to the "Test"menu.

Refer to chapter 7, "Calibration / Calibration procedure / Etal 3 calibration of the air


detector", if the values are out of range.
Maintenance préventive

44 va04.3_002e_int.fm
4.3.17 Test 15, Clamp motor
This test checks the clamp motor is functioning correctly when carrying out displacement.

„ Test 15, press "enter".


† The main "Clamp motor" menu displays.

„ Select your sub-menu using the "OK" key.

† Mode: declamping, continuous, go back, go forward or


initialise.
Modify the choice using the central selection keys.

† Press "OK" to validate and move to the next step.

† Steps numbers: Auto or a value from 1 to 1000


Modify the value using the single arrow selection keys,

† Press "OK" to validate and move to the next step.

† Frequency: Auto or a value lying between 10 and 500Hz


Modify the value using the single arrow selection keys.

„ Press "OK" to validate and start up the test.

† The "Clamp motor" menu displays, as shown.


Œ The displacement mode, and the motor
position for the motor command appear at the
top of the screen.

Œ The number of steps decreases according to


the actual change for the number of steps.

During the test, the rotation control is activated and the device set in error if a problem is
detected.
In a test with a defined number of steps/cycles or using a "declamping" displacement
mode, the test ends when the requested position is reached.

„ Press "exit" to return to the "Test" menu.


Maintenance préventive

va04.3_002e_int.fm 45
4.3.18 Test 16, Flow rate
This test checks the flow rate coefficient performed in the flow rate calibration procedure.
(Etal 4)
„ Carry out a flow rate test that conforms to the EN-60-601-2-24 standard
(see chapter 4.5" Flow rate check").
„ Check that the flow rate error lies between -5% and +5%.

Before any flow rate test, carry out an infusion of 25ml to warm up the tube.

„ Test 16, press "enter".


† The "Flow rate" main menu displays.
The "default values" are:
Œ Flow rate 100ml/h
Œ Volume 25ml
Œ Duration 0h15

The IS flow rate software dedicated to this procedure is available from Fresenius Kabi.

Any flow rate test using the "default values" menu ends with a minimum/maximum volume
indication that eases the checking of the result obtained when this test is carried out with a
scale or a test tube.

Refer to chapter 7, "Calibration/Calibration procedure / Etal 4 flow rate calibration", if the


value is out of range.
Maintenance préventive

46 va04.3_002e_int.fm
4.3.19 Test 17, Flow rate correction
This test displays the flow rate corrective coefficient on the device, the date and
calibration counter.

„ Test 17, press"enter".


† The main "Flow rate correction" menu displays.

„ Scroll through the information, from screen to screen, using


the selection keys.
† Correction:
Œ The device flow rate corrector coefficient is displayed as a
percentage.
† Date:
Number:
Œ Calibration date and number.

„ Press "exit" to return to the "Test" menu.

4.3.20 Test 18, Occlusivity


This test is dedicated exclusively to product development.

4.3.21 Test 19, Clamp


This test displays the presence or absence of the administration set clamp, and its position.

„ Test 19, press "enter".


† The main "Clamp" menu displays.

Clamp checking, test 19.


„ Door open without administration set, check
† Clamp: absent
† Position: open.
„ Insert the administration set with the clamp in open position,check
† Clamp: present
† Position: open.
„ Close the door with administration set, check
† Clamp: present
Maintenance préventive

† Position: closed.
„ Open the door, check
† Clamp: present
† Position: closed.

Check that the administration set clamp is closed.


„ Press "exit" to return to the "Test" menu.

va04.3_002e_int.fm 47
4.3.22 Checking the presence limit alarm
To carry out this operation, exit the test mode pressing "ON/OFF" key.
„ Carry out the following assembly:

Bag

1m

50 cm
Manometer
3-way stopcock

T
Pump

Container

„ Press "ON/OFF" key.


† Place the administration set on the device and close the
door.
† Place the manometer at the tube outlet.
† Select a flow rate of 125ml/h.
† Adjust the pressure threshold to 750mmHg.
Œ Open the 3-way stopcock to a container.
Œ Carry out the infusion and leave it running
for 30 seconds.
Œ Position the 3-way stopcock on the manometer side.
Œ Check that the alarm is triggered for a value
of 1bar ± 0.15bar.
† Stop the infusion cycle by pressing "STOP".

When the value of the measured pressure, lies outside of the limit selected, recalibrate
the pressure sensor (see "Etal 1 Upstream and downstream pressure sensors
calibration").

4.3.23 Mains/battery operation check


„ To carry out this procedure, exit the test mode pressing "ON/OFF" key.
Maintenance préventive

„ Connect the device to the mains.


† Check the functioning of mains presence LED (plug-
shaped indicator).

„ Disconnect the device from the mains.


† Check that the mains indicator goes out.

48 va04.3_002e_int.fm
4.3.24 Battery life check
„ Carry out a complete battery recharge before proceeding with the test.
The duration of the charging procedure is:
† <6h device switched off
† <20h device functioning
(See test 12, Battery life)

„ When the test has been carried ou automatically, as described in test 13, the battery life
must be greater than 6.30 hours.
„ When the test is carried out in normal functioning mode, the battery life must be an
average of 8h for a flow rate of 125ml/h.

4.3.25 Electrical test


Carry out the electrical test according to EN/IEC 60601-1 standard.

Maintenance préventive

va04.3_002e_int.fm 49
4.4 Maintenance options

4.4.1 Access to test

Keyboard description.
Key Function
ON, switches on the device.
OFF stops the device by pressing for more than three seconds.
Menu/History accesses the test mode.

The selection keys are used to scroll through the numbers and
letters on the tenths, units and tens and move to the next menu.

OK, validates your choice or moves the curser.


enter, starts up infusion.
exit, exits the screen and returns to the preceding menu
C/Back, returns to the "previous" menu.
Maintenance préventive

50 va04.3_002e_int.fm
Moving to the Maintenance configuration
„ Press "Menu/History".
„ Keep it pressed down while pressing "ON/OFF" key.
„ When the "Options" main menu displays,
release "Menu/History" and "ON" keys.

„ Select "Maintenance" menu using the selection keys.


„ Press "enter" to confirm.

„ The "Maintenance" menu displays.


„ Key in the code.

„ Press "OK" to confirm.

„ Choose the "Maintenance options" menu using the


selection keys.

„ Press "enter" to confirm.


† You can browse through this sub-menu using the central
selection keys.

You can exit the configuration mode at any time by pressing "Menu".
This function cancels the modifications.

Maintenance préventive

va04.3_002e_int.fm 51
4.4.2 SAV 1, Maintenance date
This screen selects the date for the next maintenance.

„ The "Maintenance date" main menu displays.

„ Modify the day using the selection keys:


† for the tens, increase and decrease using the double
arrows.
† for the units, increase and decrease using the single
arrows.

† Press "OK" to confirm and move to the next step.

„ Modify the month using the selection keys:


† increase and decrease using the single arrows,
Œ press the "downward double arrow" for a clear month.

† Press "OK" to confirm and move to the next step.

„ Proceed as above to modify the year.


„ Press "OK" to confirm.

4.4.3 SAV 2, Initial set up


This screen activates the initial configuration when the device is next switched on.

„ The "Initial set up" main menu displays.

„ Modify your selection using "yes" or "no":


† "YES", using a single left arrow.

† "NO", using a single right arrow.

„ Press "OK" to confirm.


Maintenance préventive

52 va04.3_002e_int.fm
4.4.4 SAV 3, Data log event
This test displays the list of the alarms stamped with date and time (like Test 4).
„ The "Data log event" main menu displays.

„ Scroll through the information, from screen to screen, using


the selection keys.

„ Press "enter" to display the alarm type.


„ Press "Exit" to return to the "Test" menu.

Maintenance préventive

va04.3_002e_int.fm 53
Maintenance préventive

54
va04.3_002e_int.fm
Maintenance préventive

4.4.5 Quality Control Certificate


Use this table to summarise the results obtained from the various tests.
These tests can be carried out using a software dedicated to the Agilia devices.

Device type: Code: Device serial n°:

Conformity
N° Procedure Value Obtained
Yes No
1 „ Check the general appearance of the case, label containing serial number and
product code.
2 „ OCS check:
† Switch the device on.
† Install an administration set filled with water.
† Check OCS.
3 „ Check the upstream pressure (without drip detector): *********** ..........................
† Start a 100ml/h infusion without a drop detector.
† Clamp the administration set between the drop chamber and the pump.
† Check that the Upstream occlusion alarm is triggered before 30 seconds.
† Unclamp the tube.
† Press "silence alarme".
† "Start" displays.
Refer to chapter 7 "Calibration / Calibration procedure / EtAL1 pressure
calibration", if one of the values is out of range.
4 „ Check the downstream pressure.
† Install the water manometer at the administration set outlet.
† Set the pressure threshold at 750mmHg.
† Start infusion at 100ml/h.
† Check that the Downstream occlusion alarm is triggered at 750mmHg +125
(1bar + 0.15).
Refer to chapter 7 "Calibration / Calibration procedure / EtAL1 pressure
calibration", if one of the values is out of range.
5 „ Check the drop detector "if present".
6 „ Check the air detector.
† Start the test with at a flow rate of 500ml/h,
† Create an air bubble in the administration set with a minimum length of 5 cm.
† Check the presence of the air bubble alarm (sound and visual) when the air
bubble passes through the air bubble detector.
Refer to chapter 7, "Calibration / Calibration procedure / EtAL3 calibration of the
air detector", if the values are out of range.
7 „ Display the running time and the last maintenance date, test 2 (in hours, days or
months). ************************************************************* ..........................
Maintenance préventive

8 „ Check the LCD screen, all light indicators and buzzer, test 9. ** ..........................
9 „ Check the keyboard, test 10. ************************************** ..........................

vafiche_001e_int.fm 55
Conformity
N° Procedure Value Obtained
Yes No
10 „ Check the occlusivity, test 11.
† Connect the water manometer at the administration set outlet.
† Press the Bolus key to unclamp the administration set.
† Start the test with at a flow rate of 500ml/h and confirm all other steps.
† Check that no motor error is triggered and that the maximum pressure lies
between 1.9bar and 2.5bar in maximum 45 seconds.
11 „ Check the clamp motor, test 15.
† Validate the clamping and unclamping.
† Check for absence of motor alarm.
12 „ Check the clamp, test 19.
† Close the door without the administration set.
† Check the message:
Œ clamp "absent",
Œ position "opened".
† Insert the tube and close the door.
† Check the message:
Œ clamp "present",
Œ position "closed".
13 „ Check the flow rate.
Before any flow rate test, carry out an infusion of 25 ml to ’warm up’ the
administration set.
† Carry out a flow rate test in conformity with EN-60-601-2-24 standard.
† Check that the flow rate error lies between - 5% and + 5%.
Refer to chapter 7, "Calibration / Calibration procedure / Etal 4 flow rate
calibration", if the value is out of range.
14 „ Check the battery life, test 13:
† Recharge the battery for 6 hours (device off).
† Carry out the test according to test 13.
† Note the battery running time information, test 12: *********** ..........................
it should be > 6h30
15 „ Carry out the electrical test according to EN / IEC 60601-1 standard

Name: Date: Visa:

Observations:
Maintenance préventive

56 vafiche_001e_int.fm
Preventive maintenance

4.5 Flow rate check

4.5.1 Measuring by computer


Measuring the flow rate using a computer requires Fresenius Kabi IS Flow rate software.
For further information, contact our Technical Service.
The operating mode defined in this software follows the infusion pump norm protocol which
is adapted to the user’s software.

Use a new administration set [otherwise, make a hole in the filter of the administration set
(if equipped with) and use this set for maximum 10 calibrations].
Do not recycle medical water.
Keep the container clean changing it as often as necessary.
Make sure the environmental temperature is between 18°C and 30°C.

To acquire the IS flow rate software, contact the Fresenius Kabi Customer service
department.

Equipment needed
„ Scale connected to a micro-computer:
Flow rate value Scale sensitivity
x < 5ml/h 1/10000th
5ml/h < x < 30ml/h 1/1000th
x > 30ml/h 1/100th
„ Multi-scale acquisition programme
„ Test tube or beaker with 1ml graduations
„ Liquid: distilled water, oil.
„ Catheter extension with Luer Lock end (length 100cm, internal diameter 2.5mm)
„ Needle:
Flow rate value Type of needle
x < 30ml/h G26
x > 30ml/h G18 or G21 Preventive maintenance

va04.4_001e_int.fm 57
Installation
„ Following the diagram below:

Catheter extension bracket support

50 cm

1 cm

RS 232
Anti-vibration measuring table
cable

Observe the equipment horizontal installation plan.


„ Fill the container with 600ml or more of distilled water.
„ Connect an administration set with the container.
„ Connect the canula with the distal end
of the administration set.
„ Fit the administration set on to the pump.
„ Fill the administration set with water. Take care that
no air is in the administration set before the measurement
is started.
„ Fill the test tube or the beaker with some water ensuring
the canula is dipped in the water (> 1cm).
„ Add several drops of oil to create a greasy film on the
surface of the water. This way the user will avoid any
measurement error due to evaporation of the water.
„ Place the test tube or the beaker in the centre of the scale
platform ensuring is not in touch with the scale.
„ Place the canula inside the test tube or the beaker ensuring
that it is dipped in the water.

The infusion line (needle/catheter extension) must not rest on the scale/test tube
unit.
„ Press "ON" (device on mains power mode).
† Prime the infusion line using the "BOLUS" key.
† Check for the absence of air bubbles.
Preventive maintenance

58 va04.4_001e_int.fm
Operating mode

The software operates according to the operating mode described in the


EN 60601-2-24 standard for infusion pumps.

„ Start the acquisition programme for the scale.


„ Enter the necessary programme data without confirming the
flow rate.
„ Adjust the device to the specific flow rate.
„ Confirm the flow rate on the micro-computer to enable self-
adjustment of the scale.
„ Start infusion by pressing "VALIDATION", when 00.00
appears on the scale display.
„ At the end of the specified time, note the error percentage
displayed on the screen.

Preventive maintenance

va04.4_001e_int.fm 59
4.5.2 Measuring with a scale

Use a new administration set [otherwise, make a hole in the filter of the administration set
(if equipped with) and use this set for maximum 10 calibrations].
Do not recycle medical water.
Keep the container clean changing it as often as necessary.
Make sure the environmental temperature is between 18°C and 30°C.

Equipment needed
„ Chronometer
„ Scale:
Flow rate value Scale sensitivity
x < 5ml/h 1/10000th
5ml/h < x < 30ml/h 1/1000th
x > 30ml/h 1/100th
„ Test tube or beaker with 1ml graduations.
„ Liquid: distilled water, oil
„ Catheter extension with Luer Lock end (length 100cm, internal diameter 2.5mm)
„ Needle:
Flow rate value Type of needle
x < 30ml/h G26
x > 30ml/h G18 or G21

Installation
„ Following the diagram below:

Catheter extension bracket support

50 cm

1 cm

Anti-vibration measuring table

Observe the equipment horizontal installation plan.


Preventive maintenance

60 va04.4_001e_int.fm
„ Fill the container with 600ml or more of distilled water.
„ Connect an administration set with the container.
„ Connect the canula with the distal end of the administration
set.
„ Fit the administration set on to the pump.
„ Fill the administration set with water. Take care that no air is
in the administration set before the measurement is started.
„ Fill the test administration set or the beaker with some water
ensuring the canula is dipped in the water (> 1cm).
„ Add several drops of oil to create a greasy film on the
surface of the water. This way the user will avoid any
measurement error due to evaporation of the water.
„ Place the test tube or the beaker in the centre of the scale
platform ensuring is not in touch with the scale.
„ Place the canula inside the test tube on the beaker ensuring
that it is dipped in the water.

The infusion line (needle/catheter extension) must not rest


on the scale/test tube unit.
„ Press "ON" (device on mains power mode).
† Prime the infusion line using the "BOLUS" key.
† Check for the absence of air bubbles.

Operating mode
„ Carry out a flow rate selection.

For low flow rates (< 5ml/h), confirm or wait 1 hour for infusion to stabilise. For high flow
rates, wait 10 to 30 minutes after the end of the infusion.
„ Carry out the scale taring: 00.00 g.
„ Start the infusion by pressing "VALIDATION", and start the
chronometer (if necessary, note the chronometer value).
„ After one hour, press "STOP" to stop the test.
„ Note the value (in grams) of the "infused" liquid.
„ Calculate the difference between the theoretical value and
the real value.

1gram = 1ml.
„ The error percentage is calculated according to the following
formula:
( Measured value – Theoretical value )
× 100 = Error percentage
Preventive maintenance

---------------------------------------------------------------------------------------------------
Theoretical value

va04.4_001e_int.fm 61
4.5.3 Measuring with a test tube

Use a new administration set [otherwise, make a hole in the filter of the administration set
(if equipped with) and use this set for maximum 10 calibrations].
Do not recycle medical water.
Keep the container clean changing it as often as necessary.
Make sure the environmental temperature is between 18°C and 30°C.

Material needed
„ Chronometer
„ Test tube or beaker with 1ml graduations
„ Liquid: distilled water, oil
„ Catheter extension with Luer Lock end (length 100cm, internal diameter 2.5cm)
„ Needle:
Flow rate value Type of needle
x < 30ml/h G26
x > 30ml/h G18 or G21

Installation
„ Following the diagram below:

Catheter extension support bracket

50 cm

1 cm

Anti-vibration measuring table

Observe the material horizontal installation plan.

„ Fill the container with 600ml or more of distilled water.


„ Connect a giving set with the container.
„ Connect the canula with the distal end of the giving set.
„ Fit the giving set on to the pump.
Preventive maintenance

„ Fill the giving set with water. Take care that no air is in the
tube before the measurement is started.
„ Fill the test tube or the beaker with some water ensuring the
canula is dipped in the water (> 1cm).
„ Add several drops of oil to create a greasy film on the
surface of the water. This way the user will avoid any
measurement error due to evaporation of the water.
„ Place the canula inside the test tube or the beaker ensuring
that it is dipped in the water.

62 va04.4_001e_int.fm
Operating mode
„ Carry out a flow rate selection.

For low flow rates (< 5ml/h), confirm and wait 1 hour for the infusion to stabilise. For
higher flow rates, wait 10 to 30 minutes after the end of the infusion.
„ Start the infusion by pressing "VALIDATION" and start the
chronometer (if necessary, note the chronometer value).
„ When the volume has been totally infused in the test tube,
calculate the difference between the theoretical value and
the real value:
Real flow rate = 50ml / time in hours
„ The error percentage is calculated according to the following formula:
( Measured value – Theoretical value )
--------------------------------------------------------------------------------------------------- × 100 = Error percentage
Theoretical value

Preventive maintenance

va04.4_001e_int.fm 63
Preventive maintenance

64
va04.4_001e_int.fm
Preventive maintenance

4.6 Cleaning and disinfecting


The device is part of the patient's immediate environment. It is advisable to clean and
disinfect the device’s external surfaces on a daily basis in order to protect patient and staff.
„ Disconnect the device from its main supply before cleaning.
„ Do not place in an AUTOCLAVE nor IMMERSE the device, Do not let liquids enter the
device’s casing.
„ If the device is placed in a high contamination risk unit, it is advisable to leave it in the room
during disinfecting, after having disinfected it with amoist cloth.
„ Use a cloth soaked in DETERGENT-DISINFECTANT, previously diluted with water if
required, to destroy micro organisms.
† Avoid abrasive scrubbing which could scratch the casing.
† Do not rinse or wipe the surfaces.

Do not use :
†TRICHLOROETHYLENE-DICHLOROETHYLENE.
†AMMONIA.
† AMMONIUM CHLORIDE.
† CHLORINE AND AROMATIC HYDROCARBON.
† ETHYLENE DICHLORIDE-METHYLENE.
† CHLORIDE-CETONE.
These aggressive agents could damage the plastic parts and cause device malfunction.

Take care with ALCOHOL based SPRAYS (20% - 40% alcohol). They lead to
tarnishing and small cracks in the plastic, and do not provide the necessary
cleaning prior to disinfection.
Disinfecting SPRAYS may be used, in accordance wih the manufacturer recommendation,
from a distance of 30 cm of the device, avoid the accumulation of the product in liquid form.
Please contact the appropriate service, responsible for cleaning and disinfecting products, in
your establishment for further details.

4.7 Storage
The device should be stored in a dry, cool place. In case of prolonged storage, the battery
should be disconnected via the battery access flap situated underneath the device.This
should be done by a qualified technician.
„ Storage conditions and carrying:
† Temperature: -10°C to +60°C / 14°F to 140°F.
† Atmospheric pressure: 500hPa to 1060hPa.
† Humidity: 10% to 90%, no condensation.

Fully recharge the battery before using the device to avoid any risks caused by micro power
Preventive maintenance

cuts in the mains supply and to ensure maximum autonomy.

va04.5&6_001e_int.fm 65
4.8 Use of the internal battery
This device is provided with Li-Ion battery. When the device is disconnected from the mains, it
automatically switches to battery mode.
Before starting for the first time, charge the battery for approx. 5 hours by connecting the
power supply cord without using the device.
If possible, if not using device for an extended period (higher than 2 months), remove it from
the device and store it in a place according to recommended use conditions. If not possible to
remove battery or for a short period (lower than 2 months), it is recommended to charge the
battery at least once a month, by leaving the device connected on mains power supply 8
hours at least (device OFF).
You can also recharge a Lithium-ion battery whenever convenient, without the full charge /
discharge cycle necessary to get full capacity when fully charged.

In order to maximize the life time and the performances of the battery:
„ Use and store in a cool place.
„ In use, leave the device connected to mains power supply to maintain the charge of the
battery and the maximum capacity when possible.

Lithium Ion rechrageable battery, handle with care !


„ Do not incinerate or place near an open flame.
„ Do not drop, crush, puncture, modify or disassemble battery.
„ Do not use the battery that is severely scarred or deformed.
„ Do not short terminals.
„ Do not expose to high temperatures.
„ Do not replace by a battery other than that specified by manufacturer.
„ Do not charge or discharge otherwise than in the device.
Preventive maintenance

66 va04.5&6_001e_int.fm
5 Troubleshooting
5.1 Breakdown guide

Displayed
Meaning Cause Action
message
Door opened „ The Hall effect „ Check that the magnet is
alarm sensor detects that present on the door
the door is open „ Calibrate the door sensor
„ Replace the door sensor
Administration „ The pump does „ Check the position of the
set position not detect the administration set and the clamp.
alarm presence of the Close the door.
safety clamp „ Check the clamp detection
opto and replace if necessary
Administration „ Incorrect „ Check the administration set
set insertion administration set position
alarm position „ Check that the magnet is
present on the door
„ Calibrate the door sensor
„ Replace the door sensor

„ At start-up: „ Install the administration set


administration set and close the door
not installed or
door opened
OCS test failure „ The pumping „ Check the administration set
alarm mechanism has position
not managed to „ Check the pressure sensors
raise the pressure calibration
in the line during „ Check the integrity of the
the OCS test. membrane and replace it if
necessary
„ Check the installation of the
administration set and its integrity
„ Check the integrity of the door
„ If the problem persists,
contact the Technical Service
Air bubble alarm „ The volume of „ Remove the air bubbles
air detected in the present by rebooting the
line exceeds the administration set
limit „ Check the air bubble size
setting
„ Calibrate the air detector
„ Replace the air detector

„ Air bubble „ Remove the air bubble by


Troubleshooting

detected at start- priming the administration set


up or in stop mode

va05.1_001e_int.fm 67
Displayed
Meaning Cause Action
message
Flow rate „ Flow rate has „ Check the flow rate and
flashing alarm been modified but confirm with the OK key
has not been
confirmed
End of infusion „ The remaining „ Check if the volume remaining
pre-alarm VTBI is less than in the container corresponds to
5% of the initially the remaining VTBI. If necessary,
programmed VTBI prepare a container for a new
or 5 minutes or infusion sequence. If a drop
5 ml remain before detector is present, the end of
the initially infusion pre-alarm is disactivated
programmed VTBI and can be defined as an option.
is reached
End of infusion „ VTBI delivered. „ Press Stop to define the new
alarm KVO activated infusion parameters (if
conforming with necessary).
the configuration
Drop detector „ The drop „ Check the drop detector
alarm (if used detector is not position
with this correctly „ Check the integrity of the drop
accessory) positioned detector
„ Check the drop detector
connection
„ Try the device using another
drop detector
„ If the trouble keeps up, check
the connection on the device
„ Replace the component if
necessary

„ The drop „ Connect the drop detector to


detector has been the pump and to the drip
connected then chamber
disconnected
„ The drop
detector is
disconnected
when the pump is
configured to
function with it
Under flow rate „ Flow rate „ Check the container
alarm detected by the „ Check the roller clamp
drop detector is „ Check that liquid forms
lower than the set properly approx. 20 drops / ml
flow rate
Troubleshooting

68 va05.1_001e_int.fm
Displayed
Meaning Cause Action
message
Over flow rate „ Flow rate „ Check the administration set
alarm detected by the position
drop detector is „ Check the drop detector
greater than the position
set flow rate „ Check the liquid volume in the
drip chamber
„ Check the fluid temperature
„ Check that liquid forms
properly approx. 20 drops / ml
Uncontrolled „ At the start or „ Check the drop detector and
flow rate alarm when stopped, free the administration set
flow is detected by installations
the drop detector
High flow rate „ Upper soft limit „ Check the liquid volume in the
warning exceeded, drip chamber
conforming with „ Check the fluid temperature
the drug „ Check that liquid forms
parameters set in properly approx. 20 drops / ml
the drug library
Weak flow rate „ Lower soft limit „ Check the liquid volume in the
warning exceeded, drip chamber
conforming with „ Check the fluid temperature
the drug „ Check that liquid forms
parameters properly approx. 20 drops / ml
defined in the drug
library
Pressure „ In-line pressure „ Check whether the infusion
increase rises. This warning line is occluded (clamp or 3-way
warning can be selected as stopcock closed, catheter
an option blocked, line creased, etc.)
Pressure drop „ Fall in pressure „ Check the downstream Luer
warning in the infusion line. lock connection and the integrity
This warning can of the whole line
be selected as an
option
Occlusion pre- „ In-line pressure „ Check the infusion line
alarm has reached „ Define the pressure threshold
50 mmHg below
the programmed
threshold
Troubleshooting

va05.1_001e_int.fm 69
Displayed
Meaning Cause Action
message
Upstream „ The pressure „ Check the container and the
occlusion alarm level in the line
pumping system „ Check the roller clamp
upstream line is „ Check that the administration
too low set is not pinched
„ The container is „ Check the height of the
empty container
„ The roller „ Check the air inlet stopper
clamp is closed (when used with a bottle)
„ The air inlet „ Calibrate the pressure
stopper is closed sensors
(when used with a
„ Replace the upstream
bottle)
pressure sensor
„ Only if the drop
detector is absent.
The upstream line
pressure is too
weak
Downstream „ The „ Check whether the infusion
occlusion alarm downstream line is occluded
pressure for the „ Check that the pressure
pumping system threshold is set at an appropriate
has exceeded the level
programmed „ Calibrate the pressure
pressure threshold sensors
„ Replace the downstream
pressure system
Battery pre- „ Weak battery „ Connect the device to the
alarm charge level mains

Battery alarm „ Battery „ Immediately connect the


discharged device to the mains
(automatic switch „ Replace the battery if there is
off of the device in not enough life left
5 minutes)
Battery „ Uncharged „ Connect to mains power
uncharged battery supply and wait till the battery is
alarm charged
Mains „ Mains „ Press on silence to stop the
disconnected disconnection warning
warning „ Check that there is enough
battery charge left for the
expected duration of the infusion
„ If the disconnection location
cannot be found:
Troubleshooting

† check the mains power


supply connection
Preventive „ Maintenance „ Carry out a complete check
maintenance date reached for the device and update the
requested maintenance date (test 2)

70 va05.1_001e_int.fm
Problem Cause Recommended action
Flow rate variance „ A mechanical part of the „ Carry out a flow rate test,
too great pumping mechanism or the calibrate the infusion corrector
CPU board has been coefficient
changed without the „ Redo the test to ensure that the
calibration of the infusion correction is right
corrector coefficient being
carried out
Display default: „ LED and/or connectors „ Check the general state of the
segments, LED defective display board
„ Check the contrast
„ Check the connectors
„ Replace the display board
The device „ Battery is completely „ Check the battery connection
switches off when discharged „ Charge the battery
it is disconnected
„ Replace the battery
from the mains
The mains „ Mains fuses are out of „ Replace the mains fuses
indicator does not order „ Check that the mains connector
light up has been correctly connected
„ Replace the power supply board
Battery alarm „ Battery is completely „ Check the charging voltage
while it has been discharged „ Replace the battery
correctly charged
The device „ Defective keyboard „ Check and replace the keyboard
switches on or off if necessary
on its own „ Defective CPU board „ Replace the CPU board and
reconfigure completely
The keyboard „ Defective keyboard „ Check and replace the keyboard
keys do not work if necessary
After a fall „ Mechanical elements „ Check the integrity of the
damaged mechanical unit and the cases
„ Check the state of the battery
Troubleshooting

va05.1_001e_int.fm 71
Troubleshooting

72 va05.1_001e_int.fm
Diagnostic

5.2 Errors messages

Error
Description Recommended action
Code
Software and electronic command faults
01 „ Rotation check „ Check correct functioning of the motor.
„ Check the flexible Cl Opto and CPU board
connection.
„ Check the correct functioning of the motor
optical switch rotation.
„ Replace the CPU board.
„ Carry out a complete configuration using
the maintenance software.

10 „ RAM „ Replace the CPU board.


„ Carry out a complete configuration using
the maintenance software.

11 „ Flash „ Flash memory deleting or programming


problem.
„ Replacing the CPU board.
„ Carry out a complete configuration using
the maintenance software.

12 „ Resources „ Crc/16 problem.


„ Carry out a complete configuration for the
device using the maintenance software. If the
problem persists, replace the CPU board and
carry out a complete configuration.

13 „ Version /secondary Crc16 „ Secondary Crc/16 µC version is incorrect.


µC check „ Carry out a complete configuration using
the maintenance software.

14 „ Secondary µC starting up „ There is a problem of the secondary µC


again. watch dog.
„ Replace the power supply board.

15 „ Secondary µC „ Check the state of the power supply/CPU


communication board connection ribbon cable.
„ Replace the power supply board.

16 „ Disconnected battery „ Check the battery connection.


„ Check the battery charging voltage.
„ Replace the battery.

17 „ Battery charge „ Check the status of the battery.


„ Check the battery charge voltage.
„ Check the battery.
„ Replace the battery.
„ Replace the power supply board.

18 „ Mains presence „ Check the status of the mains fuses.


„ Replace the power supply board.
Diagnostic

va05.2_001e_int.fm 73
Error
Description Recommended action
Code
19 „ Battery temperature „ Check the battery status.
„ Replace the battery.

20 „ 3.3 V missing of power „ Check the status of the ribbon cable


supply connection between power supply board and
CPU board.

21 „ Secondary µC / coulometer „ Contact the Technical Service.


power supply board
communication

22 „ Clamp opto switch „ Check the opto functioning.


„ Check the status of the ribbon cable
connection.
„ Replace the opto switch.

23 „ Clamp motor optical switch „ Check the functionality of the clamp opto
motor switch.
„ Check the flexible CI.
„ Replace faulty part(s).

24 „ Unclamping „ Select test 15.


„ Check the status of the clamp motor.
„ Check the connector.
„ Replace the motor.

25 „ Door Hall effect sensor „ Select test 7.


„ Check the "open or closed" door detection.
„ Check the Hall effect; change the Hall
effect if necessary.
„ Perform a calibration of the door.

26 „ Board missing „ Check the connection between the display


board and the CPU board.
„ Check the connection between the power
supply board and the CPU board.
„ Replace the faulty part(s).

27 „ LCD black-light missing „ Check the connection between the display


board and the Back light.
„ Replace the LCD display.
„ If the problem persists, replace the display
board.

28 „ Keyboard „ Check the keyboard test 10.


„ Replace the keyboard and check its
functionality.
„ If the problem persists, replace the display
board.

29 „ Analogue/digital converter „ Replace the CPU board.


„ Carry out a complete configuration check
Diagnostic

using the maintenance software.

74 va05.2_001e_int.fm
Error
Description Recommended action
Code
30 „ Date Time stamp „ Check the frequency of the X1 quartz
which drives the date time stamp. Its
frequency must equal 32.768 kHz.
„ Check the frequency of the X2 quartz. Its
frequency must equal 32.768 kHz.
„ Replace one of the cristals if necessary.
„ Replace the CPU board, the Date Time
stamp is integrated in the main
Microcontroller.
„ Carry out a complete configuration using
the maintenance software.

31 „ Administration sets „ Carry out a complete configuration using


parameters the maintenance software.

32 „ Default "language" loading „ Carry out a complete configuration using


configuration the maintenance software.

33 „ Air detector „ Select test 14.


„ Check the status of the air detector using
an administration set filled with water and air.
„ Calibrate the detector.

34 „ Infusion flow rate check „ Carry out a complete configuration using


the maintenance software.

35 „ Motor period check „ Carry out a complete configuration using


the maintenance software.

36 „ Motor rotation direction „ Carry out a complete configuration using


check the maintenance software.

37 „ OFF key „ Carry out a complete configuration using


the maintenance software.
„ Replace the power supply board to check
the functionality of the secondary µC.
„ Replace the CPU board to check the
functionality of the main µC.
„ Replace faulty parts.

38 „ OFF key locked „ Check the functionality of the keyboard


OFF key.
„ Replace the keyboard if necessary.
„ Replace the power supply board and
check the secondary µC functionality.

39 „ Impossible to switch off „ Replace the power supply board.

40 „ Drop detector „ Detector detection unstable.


„ Check the CPU board/detector connection.
„ Replace the CPU board
„ Carry out a complete configuration using
the maintenance software.
Diagnostic

va05.2_001e_int.fm 75
Error
Description Recommended action
Code
41 „ Upstream pressure sensor „ Select test 5.
out of range „ Check the change in the state of the
sensor by applying pressure to the sensor.
„ Check the connections.
„ Calibrate the upstream pressure sensor.

42 „ Downstream pressure „ Select test 6.


sensor out of range „ Check the change in state of the sensor by
applying a pressure on the sensor.
„ Check the connections
„ Calibrate the downstream pressure
sensor.

99 „ Activation of the watch dog „ Replace the power supply board to check
the functionality of the secondary µC.
„ Replace the CPU board to check the
functionality of the main µP.
„ Carry out a complete configuration, when
replacing the CPU board, using the
maintenance software.
Diagnostic

76 va05.2_001e_int.fm
6 Intervention procedures
This chapter details all the dismantling and re-assembling procedures.

Intervention procedures

va06.0_001a_int.fm 77
Intervention procedures

78
va06.0_001a_int.fm
Intervention procedure

Procedure N°1: Battery and battery door

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Dismantling

„ Unscrew the 2 fixing screws (marker 1) holding the battery door in place.
„ Remove the battery door (marker 2).

2
1

1
Intervention procedure

va06.0_0110a_int.fm 79
„ Disconnect the J4 connector and remove the battery (marker 3).
3

J4

Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).
Intervention procedure

80 va06.0_0110a_int.fm
Intervention procedures

Procedure N°2: Upper case

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

Intervention procedures

Do not touch the electronic boards with your hands.

va06.0_0012e_int.fm 81
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Pull out the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1
4
4 5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
„ Disconnect the flat cable connected to the CPU board from the display board at J2.

J2
Intervention procedures

82 va06.0_0012e_int.fm
Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).

Intervention procedures

va06.0_0012e_int.fm 83
Intervention procedures

84
va06.0_0012e_int.fm
Fiches d’intervention

Procedure N°3: Display board

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

Intervention procedures

Do not touch the electronic boards with your hands.

va06.0_0022c_int.fm 85
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Pull out the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1
4 4
5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
„ Disconnect the flat cable connected to the CPU board from the display board at J2.

6
J1
6
J2
Intervention procedures

86 va06.0_0022c_int.fm
„ Disconnect the keyboard/display board connection at (J1).
„ Unscrew the 4 fixing screws (marker 6) holding the board to the upper cover.
„ Remove the display board/display unit.
„ Disconnect the display board /display connection at (J3).
„ Unclip and separate the display board from the display.

J3

Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).

Intervention procedures

va06.0_0022c_int.fm 87
Intervention procedures

88
va06.0_0022c_int.fm
Fiches d’intervention

Procedure N°4: CPU board

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

Intervention procedures

Do not touch the electronic boards with your hands.

va06.0_0032e_int.fm 89
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Pull out the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1
4
4
5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the display board at J2.
Intervention procedures

90 va06.0_0032e_int.fm
„ Disconnect the flat cables connected to the CPU board at:
† J2 and J13
† J7, J3 and J10
† J4 and J12
† J1
„ Unscrew the fixing screws (marker 6) holding the CPU board to the flange.
J7 J13
J3 J10
J2

J4

J12 J1

„ Remove the CPU board unit being careful not to pull out the flat cable on the inside.
„ Disconnect the flat cables (J14 and J11) if necessary.

J11

Intervention procedures

J14

va06.0_0032e_int.fm 91
Re-assembling

When replacing the CPU board, carry out a complete configuration of the device. Each
Volumat MC Agilia has its own CPU board model. Be careful not to invert the different
Volumat MC Agilia references. Order the one that corresponds to your device.

„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).
Intervention procedures

92 va06.0_0032e_int.fm
Fiches d’intervention

Procedure N°5: Power supply board

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 14" wrench
„ 1 anti-static bracelet

Maintenance level:
Level 3, specialist technician (see biomedical technical training documents).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

Fiches d’intervention

Do not touch the electronic boards with your hands.

va06.0_0042e_int.fm 93
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Lift the stopper and loosen the nut from the plug (marker 3).
2
J1

„ Unscrew the Torx screw (marker 9) from the holding board.


„ Remove the power supply board.
9
Fiches d’intervention

94 va06.0_0042e_int.fm
Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).

Don’t forget to set insulator between the power supply and the mains boards.

Fiches d’intervention

va06.0_0042e_int.fm 95
Fiches d’intervention

96 va06.0_0042e_int.fm
intervention procedure

Procedure N°6: Base kit

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet.

Maintenance level :
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

intervention procedure

Do not touch the electronic boards with your hands.

va06.0_0052e_int.fm 97
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Remove the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1
4 4
5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the display board at J2.
intervention procedure

98 va06.0_0052e_int.fm
„ Disconnect the flat cables connected to the CPU board at:
† J2 and J13
† J7 and J3 and J10
† J4 and J12
† J1
„ Unscrew the holding screws (marker 6) holding the CPU board to the flange.
J7 J13
J3 J10
J2

J4

J12 J1

„ Remove the CPU board.

intervention procedure

va06.0_0052e_int.fm 99
„ Unscrew the button screw (marker 7) and pull it out.
† Push the button and turn it to the left and remove it.
† Remove the spring.
„ Unscrew the 2 screws (marker 8) on either side of the right and left flanges being careful
not to damage the air bubble board.
8

„ Pivot the device on its front side.


„ Unscrew the screw (marker 9) positioned under the base.
„ Remove the pumping plate (marker 10).
„ Unscrew the screws (marker 11) positioned under the base.

10

11
9

11
intervention procedure

100 va06.0_0052e_int.fm
„ Remove the left flange (marker 12).
„ Take out the right flange (marker 13).

12

13

„ Separate the base kit (marker 14) from the motor kit (marker 15).

14 15 intervention procedure

va06.0_0052e_int.fm 101
Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).
intervention procedure

102 va06.0_0052e_int.fm
Intervention procedure

Procedure n°7 : Air detector board

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet

Maintenance level:
Level 3, specialist technician, (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

Intervention procedure

Do not touch the electronic boards with your hands.

va06.0_0062c_int.fm 103
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Remove the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1
4
4 5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the display board at J2.

J2
Intervention procedure

104 va06.0_0062c_int.fm
„ Unscrew the 2 screws (marker 6) of the air detector board (marker 7).

7 6

„ Remove the board being careful not to pull out the flat cable.
„ Disconnect the flat cable connected to the board at J1.

J1

Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
Intervention procedure

„ Calibrate the air bubble detector.


„ Carry out the regular servicing tests (see Quality Control Certificate).

va06.0_0062c_int.fm 105
Intervention procedure

106 va06.0_0062c_int.fm
Fiches d’intervention

Procedure N°8: Pumping system

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet.

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation)

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them (3 under the angle bracket).
1

1 1

Fiches d’intervention

Do not touch the electronic boards with your hands.

va06.0_0072c_int.fm 107
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Remove the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base
2
J1

4
4
5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
„ Disconnect the flat cable leading from the CPU board to the display board at J2.
Fiches d’intervention

108 va06.0_0072c_int.fm
„ Disconnect the flat cable connected to the CPU board at:
† J2 and J13
† J7, J3 and J10
† J4 and J12
† J1
„ Unscrew the holding screws (marker 6) holding the CPU to the flange.
J7 J13
J3 J10
J2

J4

J12 J1

„ Remove the whole CPU board being careful not to pull out the flat cable on the inside.

Fiches d’intervention

va06.0_0072c_int.fm 109
„ Unscrew the button screw (marker 7) and pull it out.
† Push the button and turn it to the left and remove it.
† Pull out the spring.
„ Unscrew the screw (marker 8) holding the left flange.

„ Pivot the device on its front


„ Unscrew the Torx screw (marker 9) positioned under the base.
„ Remove the pumping plate (marker 10).
„ Unscrew the screw (marker 11) positioned under the base

10

11
Fiches d’intervention

110 va06.0_0072c_int.fm
„ Remove the left flange (marker 12).

12

„ Remove the pumping system.

Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).
Fiches d’intervention

va06.0_0072c_int.fm 111
Fiches d’intervention

112 va06.0_0072c_int.fm
Intervention procedure

Procedure N°9: Motor

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet
„ 1 4" wrench
„ 1 small cruciform screwdriver

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1
Intervention procedure

Do not touch the electronic boards with your hands.

va06.0_0082c_int.fm 113
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Remove the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1

4
4 5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
„ Disconnect the flat cable leading from the CPU board to the display board at J2.
Intervention procedure

114 va06.0_0082c_int.fm
„ Disconnect the flat cables leading to the CPU board at:
† J2 and J13
† J7 and J3 and J10
† J4 and J12
† J1
„ Unscrew the holding screws (marker 6) holding the CPU board to the flange.
J7 J13
J3 J10
J2

J4

J12 J1

„ Remove the whole CPU board being careful not to pull out the flat cable on the inside.

Intervention procedure

va06.0_0082c_int.fm 115
„ Pivot the device on its front.
„ Unscrew the button screw (marker 7) and pull it out.
† Push the button and turn it to the left and remove it.
† Pull out the spring.
„ Unscrew the screw (marker 8) holding the left flange.

„ Unscrew the Torx screw (marker 9) positioned under the base.


„ Remove the pumping plate (marker 10).
„ Unscrew the screw (marker 11) positioned under the base.

10

11
Intervention procedure

116 va06.0_0082c_int.fm
„ Remove the left flange (marker 12).

12

„ Pull out the pumping system.

Intervention procedure

va06.0_0082c_int.fm 117
„ Unscrew the 4 motor fixing nuts (marker 12).

12
12

12
12

„ Pull out the 4 screws, and remove the motor with its silenblocks.

Re-assembling
„ Put the motor in position, motor wires on the left side of the mechanical bloc using a weak
loctite ref. 248 to fix the motor with its nuts.
„ Moderately tighten the motor fixing nuts in a diametrically opposite sequence (marker 12).

12
12

12
12
Intervention procedure

„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).

118 va06.0_0082c_int.fm
Intervention procedure

Procedure N°10: Pressure sensor

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

Intervention procedure

Do not touch the electronic boards with your hands.

va06.0_0092c_int.fm 119
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Remove the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1
4
4 5

„ Remove the upper case (marker 5), being careful not to pull out the display board flat
cable.
Intervention procedure

120 va06.0_0092c_int.fm
„ Disconnect the flat cable leading from the CPU board to the display board at J2.
„ Disconnect the flat cables leading to the CPU board at:
† J2 and J13
† J7, J3 and J10
† J4 and J12
† J1
„ Unscrew the holding screws (marker 6) holding the CPU board to the flange.
J7 J13
J3
J10
J2

J4

J12 J1

„ Remove the whole CPU board being careful not to pull out the flat cable on the inside.

Intervention procedure

va06.0_0092c_int.fm 121
„ Unscrew the button screw (marker 7) and pull it out.
† Push the button and turn it to the left and remove it.
† Pull out the spring.
„ Unscrew the screw (marker 8) holding the left flange.

„ Pivot the device on its front side.


„ Unscrew the Torx screw (marker 9) positioned under the base.
„ Remove the pumping plate (marker 10).
„ Unscrew the screw (marker 11)positioned under the base.

10

11
Intervention procedure

122 va06.0_0092c_int.fm
„ Pull out the left flange (marker 12).
„ Pull out the pumping system.

12

„ Unscrew the 2 upstream sensor and downstream sensor screws (markers 13 and 14).

14
13

Intervention procedure

va06.0_0092c_int.fm 123
„ Pull out the sensor.

Re-assembling
„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Calibrate the sensors.
„ Carry out the regular servicing tests (see Quality Control Certificate).
Intervention procedure

124 va06.0_0092c_int.fm
Intervention procedure

Procedure N°11: OCS clamp motor

Safety:
For safety reasons, the technician should not intervene when the device is connected to the
mains voltage.
Disconnect the power supply cord from the mains.
Switch off the device.

Disconnect the battery before opening the device (see procedure n°1).

Material needed:
„ 1 Torx T-10 screwdriver
„ 1 anti-static bracelet.

Maintenance level:
Level 3, specialist technician (see biomedical technical training documentation).

Procedure:

Access
„ Turn the device upside down.
„ Unscrew the 3 Torx screws (marker 1) and remove them.
1

1 1

Intervention procedure

Do not touch the electronic boards with your hands.

va06.0_0102e_int.fm 125
Dismantling

When working with electronic components, we recommend wearing an anti-static bracelet


connected to earth and working on a foam anti-static mat.

„ Remove the angle bracket (marker 2) being careful not to pull out the CPU board flat
cable.
„ Disconnect the flat cable connecting the CPU board to the power supply at J1.
„ Remove the RJ45 stopper (marker 3).
„ Unscrew the 2 screws (marker 4) holding the upper case (marker 5) to the base.
2
J1
4
4 5

„ Remove the upper case (marker 5) being careful not to pull out the display board flat cable.
Intervention procedure

126 va06.0_0102e_int.fm
„ Unscrew the 2 fixing screws (marker 6) holding the motor to the flange.

This procedure should be handled with care in order to avoid any damage of the air
bubble detector board/flat cable when using a scew driver.

„ Using a small flat screwdriver, apply a light pressure in order to disengage the motor
positioning spigot (marker 7).

Intervention procedure

va06.0_0102e_int.fm 127
„ Remove the motor (marker 7).

Re-assembling

Check the right position of the OCS motor fins when assembling the motor kit. The fins
should be left side OCS optical sensor board.
„ Position the OCS motor into the flask, wires position as indicated by marker 8 (lef side and
at the bottom).
„ Screw the 2 fixing screws (marker 6) holding the motor to the flask.
„ Check that the motor is clipped correctly (marker 9).

8 6
9
Intervention procedure

„ Carry out the dismantling procedure in reverse order to re-assemble the unit.
„ Carry out the regular servicing tests (see Quality Control Certificate).

128 va06.0_0102e_int.fm
7 Calibrations
7.1 Calibrations procedure

The calibrations menu is reserved for qualified staff only. Its access is protected by a
secret code.

7.1.1 Access to the calibrations

Keyboard description
Key Function
ON switches on the device.
OFF switches off the device when held down for more than three
seconds.
Menu/History accesses the test mode.

The selection keys scroll through the numbers and letters on the
LCD screen using the tenths, units and tens values.
Activate or de-activate the infusion menu.
OK confirms the test selection or moves the cursor.
start starts the infusion
exit exits the screen and returns to the "Test" menu.
C/Back, returns to the "previous" menu.

Calibrations

va07.1_002e_int.fm 129
How to activate the "Calibration" configuration mode
„ Press "Menu/History".
„ Keep it held down while pressing "ON".
„ When the "Options" main menu displays,
release "Menu" and "ON" keys.

„ Select your "Maintenance" menu using the selection keys.

„ Press "enter" to confirm.

„ The "Maintenance" menu displays.


„ Key in your code.

„ Press "OK" to confirm.

„ Select the menu using the central selection keys.


† Calibration
„ Press "enter" to confirm.
† Navigate through this sub-menu using the central
selection keys.

You may exit the configuration mode at any time by pressing "Menu";
this function cancels the modifications.
Calibrations

130 va07.1_002e_int.fm
7.1.2 Etal 1, Pressures
Calibrating the pressure consists of calibrating the upstream and downstream pressure in
several phases.

A new administration set must be used for this procedure.

„ Etal 1, press "enter".


The main "Pressures" menu displays.
† Open the door and remove the administration set.

† Press "OK" to reset the search to zero.

„ The calibration values for upstream and downstream


pressure appear in mV starting at 0.

† In "open door" phase.

† Press "OK" to validate and move to the next step.

„ In "Pressures - door closed: unclamp" phase displays.


† Position the administration set and close the door.

† Press "OK" to move to the next step.

Œ Unclamping in progress.

Œ The motor starts positioning

„ The "Calibrating pressures" menu, 0 mmHg with a tube


appears in mV,
the manometer should indicate "0" mmHg (0 bar)
Calibrations

† Press "OK" to validate and move to the next step.

va07.1_002e_int.fm 131
„ The "pressures calibration" menu, 760mmHg with tube
appears in mV.
† Apply a pressure of 760mmHg (of around 1bar) at the
tube outlet.
† Press "OK" to validate and move to the next step.

„ The "Pressures calibration" menu, 0mmHg with tube


appears in mV.
† Decrease the pressure to 0mmHg (0bar).

† Press "OK" to validate and move to the next step

† Press a second time to confirm and save the calibration


values.

† If the values are out of the ranges, a message specify the


non-conformity appears.

† Restart the calibration procedure.


Calibrations

132 va07.1_002e_int.fm
7.1.3 Etal 2, Door
Calibrating the door consists of calibrating the door’s Hall Effect sensor in several phases.

„ Etal 2, press "enter".


† The main "Door" menu displays.

† Open: open the door.


Œ Press "OK" to validate and move to the next step.
The value is read and recorded. It should lie between
1400 < X < 1800mV.

† Intermediate: place the calibration tool referenced


067059 with the door in the intermediate position.

Œ Press "OK" to validate and move to the next step.


Œ The value is read and recorded.

† Closed: shut the door.

Œ Press "OK" to validate and move to the next step.


Œ The value is read and recorded. This value should lie
between 1500 < X < 2700mV.
† Press again to confirm and save the calibration values.

† If the values are out of range, a message specifies the


non-conformity.

† Restart the calibration procedure.


Calibrations

va07.1_002e_int.fm 133
7.1.4 Etal 3, Air sensor
The air sensor calibration consists of calibrating the air detector board in several phases

A new tube must be used for this procedure.

„ Etal 3, press "enter".


† The main menu "Air sensor" displays.
† Place an administration set filled with water in position
and shut the door.

† Press "OK" to launch the resonance search.


† Wait the stop of resonance phase

„ Resonance frequency calibration phase with a tube and


water:
† Water tube: the air detector value in mV.

† Frequency: current frequency value in KHz.


Œ Water tube: the air detector value in mV. The value
must lie between 600 < X < 2400mV.
Œ Frequency: the resonance frequency for this phase.

† Press "OK" to validate and move to the next phase.

„ Resonance frequency calibration phase, without a tube.


† Remove the tube and shut the door.
Œ Without the tube: the air detector value in mV. The
value must be X < 300mV.
Œ Frequency: the resonance frequency for this phase.
† Press "OK" to validate and move to the next phase.

„ Calibration resonance frequency phase, with tube and air


† Put a tube with air in place and close the door.
Œ Air detector value in mV. The value must lie
between X < 300mV.
Œ Frequency: resonance frequency for this phase.
† Press "OK" to confirm and move to the next phase.

† Press "OK" again to confirm and save the calibration


values.

† If the values are out of range, a message specifies the


Calibrations

non coformity.

† Restart the calibration procedure.

134 va07.1_002e_int.fm
7.1.5 Elal 4, Flow rate
The flow rate calibration consists in defining the device coefficient corrector to calibrate the
pump.

A new tube must be used for this procedure.

„ Etal 4, press "enter".


† The main menu "Flow rate" displays.
Œ Flow rate: the infusion flow rate, 100ml/h, enables the
speed of the motor rotation to be defined.
Œ Turns: the number of motor cycles to execute 100.

† Perform a flow rate measurement according to instructions


described in paragraph "4.5 Flow rate check".
† Press "OK" to validate and move to the next phase.

† "Unclamping" and "Motor clamp start-up" phase.

„ Coefficient corrector calibration.


Œ Flow rate: infusion flow rate, 100ml/h, enables the
speed of the motor rotation to be defined.
Œ Cycles: number of motor cycles to execute100.

† Calibration in progress, wait till the end of the cycle.

† The volume measured by the pump is displayed.


† If necessary, correct the value, using the selection keys,
after comparing it with the measured volume obtained
during the same test.

† Press "OK" to validate and move to the next phase.

† The correction is displayed in %, it is calculated from the


corrective coefficient programmed in the device .
The value of the corrective coefficient must lie between
80 < X <110.

† Press "OK" to save the flow rate correction.


Calibrations

va07.1_002e_int.fm 135
Calibrations

136 va07.1_002e_int.fm
8 Spare parts catalogue
Refer to the spare parts catalogue.

Spare parts catalogue

va08.0_001d_int.fm 137
Spare parts catalogue

138 va08.0_001d_int.fm
Useful addresses
CUSTOMER SERVICE
Fresenius Vial
Le Grand Chemin, 38590 Brézins
Tel.: 33 (0) 4 76 67 11 11
Fax: 33 (0) 4 76 67 11 34

TRAINING DEPARTMENT /
TECHNICAL DEPARTEMENT
Fresenius Vial Tél. : 33 (0) 4 76 67 10 76
Le Grand Chemin, 38590 Brézins or 33 (0) 4 76 67 11 40
or 33 (0) 4 76 67 60 73
Fax : 33 (0) 4 76 67 11 22

This document may contain inaccuracies or typographical errors.


Modifications may thus be made and will be included in later editions.
Because of the evolution of the standards, lawful texts and material, the characteristics
indicated by the text and the images of this document are applicable only for the device that it
accompanies.
This document may not be reproduced in whole or in part without the written consent
of Fresenius Kabi.
Volumat®, Vigilant® and Agilia® are registered trademarks in the name of Fresenius Kabi
in selected countries.

Fresenius Vial S.A.S. - siège social : Le Grand Chemin - 38590 Brézins (FRANCE)
Société par Actions Simplifiées au capital de 13 744 520 Euros
SIREN Grenoble B 408 720 282.
Design and construction : SEDOC

Technical Manual Volumat MC Agilia_int


Fresenius Vial S.A.S
Le Grand Chemin
38590 Brézins
France
Tel: +33 (0)4 76 67 10 10
Fax: +33 (0)4 76 67 11 34
www.fresenius-kabi.com A Company of Fresenius Kabi

You might also like