Fresenius Kabi Optima PT Vs ST Technical Manual PDF
Fresenius Kabi Optima PT Vs ST Technical Manual PDF
Fresenius Kabi Optima PT Vs ST Technical Manual PDF
Optima pt/vs/st
Technical Manual
Revision table
TM Optima PT-VS-ST_Int : ref NT1741 / from serial n° 18180100 till...
Optima_PT-VS-ST_enTDM.fm 3
4.3.8 Checking the total operating period ......................................................................... 35
4.3.9 Checking the analogue sensor input........................................................................ 35
4.3.10 Checking the software version ................................................................................. 36
4.3.11 Checking the motor clamp (only Optima VS) .......................................................... 36
4.3.12 Checking the calibration values ............................................................................... 37
4.3.13 Checking the infusion motor..................................................................................... 37
4.3.14 Checking the last 10 events before a locking alarm or error .................................... 38
4.3.15 Checking the pressure alarm limit ............................................................................ 38
4.3.16 Flow rate check ........................................................................................................ 39
4.3.17 Air detector check .................................................................................................... 39
4.3.18 Battery autonomy check........................................................................................... 39
4.3.19 Continuity test .......................................................................................................... 40
4.3.20 Contrast check ......................................................................................................... 40
4.3.21 Timestamp check ..................................................................................................... 40
4.3.22 Quality Control Certificate ........................................................................................ 41
4.4 Cleaning and disinfecting ................................................................... 43
4.5 Storage.................................................................................................. 44
5 Troubleshooting.............................................................................................. 45
5.1 Breakdown guide ................................................................................. 45
5.2 Error messages ................................................................................... 49
6 Intervention procedures................................................................................. 53
N°1, Procedure: CPU board ....................................................................... 55
N°2, Procedure: Display board .................................................................. 59
N°3, Procedure: Pumping system ............................................................. 63
N°4, Procedure: LCD display .................................................................... 67
N°5, Procedure: Air detector board........................................................... 71
N°6, Procedure: Pressure sensor.............................................................. 75
N°7, Procedure: OCS clamp motor (if OCS is present) ........................... 79
N°8, Procedure: Power supply board ....................................................... 83
N°9, Procedure: Battery ............................................................................. 85
N°10, Procedure: Front cover, base and angle bracket .......................... 87
7 Calibration ....................................................................................................... 91
7.1 Calibration procedure.......................................................................... 91
7.1.1 Access to the calibrations ........................................................................................ 91
7.1.2 EtAL.1 Door calibration....................................................................................... 92
7.1.3 EtAL.2 Calibration of the flow rate corrective coefficient .................................... 93
7.1.4 EtAL.9 Calibrating the upstream and downstream pressure sensors ................ 94
44 Optima_PT-VS-ST_enTDM.fm
1 Presentation
1.1 General presentation
The Optima pump is a new pump based on the same pumping mechanism as the MVP
pump. It is used for infusion of intravenous agents.
There are three types of the Optima pump:
PT: for standard PVC tubes with a drop detector.
VS: for standard PVC tubes. The drop detector is optional. The pump is equipped with an
automatic occlusion checking system (OCS).
ST: for silicon tubes equipped with a drop detector.
The Optima pump has its own control keyboard and distinct displays. It has an internal
battery which ensures a minimum autonomy of 4 hours for a flow rate of 125 ml/h.
A choice of easily accessible configurations enables optimal use of the functions according to
the needs of each department.
Presentation
optima01_002b_en.fm 5
1.2 Operation diagram
Presentation
6 optima01_002b_en.fm
1.3 Precautions for use
The symbol ! which appears on the device’s condensed user instructions recommends a
complete reading of the user instructions, in accordance with norm EN 60 601-1.
Fresenius Vial cannot, in any circumstances, be held responsible for any medical or other
problem due to incorrect use of the device.
Kindly consult the user manual for more details.
optima01_002b_en.fm 7
1.5.4 Conformity and norms
8 optima01_002b_en.fm
2 Description and operation
2.1 Physical description
The Optima pump is made out of a front cover, a base and an angle bracket.
! The front cover contains:
" a display board linked to the control panel,
" a CPU board,
" an air detector board,
" the OCS system if OCS is present.
optima02_002b_en.fm 9
2.1.1 Display board and front panel
The display board is situated under the control panel and contains all the essential parts for
the man-machine interaction:
! keyboard,
! synoptic and control indicators,
! 7-segment displays,
! LCD screen.
Pin n°1
J1
Pin n°1
J2
J3
Pin n°1
11 GND
12 CSAFCL Display selected
13 GND
14 A0AFF LCD address
15 GND
16 RESET/
17 GND
10 optima02_002b_en.fm
J1 connector to CPU board
J2 connector to LCD
J3 connector to keyboard
optima02_002b_en.fm 11
2.1.2 The CPU board
The CPU board contains a 80C32 micro-processor. It is connected to the display board by
means of the J1 connector.
J1 J2 J4
J3
Pin n°1
J7
J6
J9 J10
J8
J11 J12
CPU board
Info: Pin 1 of connector J1 is opposite to that indicated by the CPU board screen printing.
J1 display board connector
16 RESET/
17 GND
18 GNDTON/OFF Signal ON/OFF
19 VBAT Battery voltage
20 5V Power supply
12 optima02_002b_en.fm
J12 drip detector connector
Pin Signal Description
1 Pulse-drip Low signal yield
2 GND
3 GND
4 CDLED-drip Control LED 5V/ 40 mA
J4 motor connector
Pin Description
1 +Reel A
2 - Reel A
3 +Reel B
4 - Reel B
J8 battery connector
Pin Description
1 + DC motor
2 - DC motor
optima02_002b_en.fm 13
J2 sensors connector
14 optima02_002b_en.fm
2.1.3 The power supply board and battery
The power supply board is positioned on the angle bracket. It supplies power to the electronic
unit from the 230 VAC network. It also ensures a battery charge of 2.7/3 Ah.
J6
J4 J5
J3
U1
U2
C3
Bars code
F2 L1
J2
J1
J2 Mains connector
Pin Description
1 PH Phase (100 - 240 Vac)
2 N Neutral (100 - 240 Vac)
Pin Description
1 9V Sect 9V output convertor AC/DC
2 GND
3 9V Ext 9V external power supply not used
Description and operation
Pin Description
1 +Vbat
2 +Thermal sensor
3 GND
4 VBAT NP
optima02_002b_en.fm 15
J6 battery input connector
Pin Description
1 + VBAT
2 VBAT NP
3 + Thermal sensor
4 - Thermal sensor
5 GND
Description and operation
16 optima02_002b_en.fm
2.2 Operational description
From an operational point of view, the Optima pump is made up of 3 sub-assemblies:
! a sub-assembly to maintain to check the correct positioning of the tube,
! a motorisation sub-assembly,
! an external connection sub-assembly.
optima02_002b_en.fm 17
Description and operation
18
optima02_002b_en.fm
3 Description of the menus
3.1 Parameter configuration menu
The configuration menu adapts the Optima to the particular needs of each department. It
accesses the menus that customise the infusion parameters.
Fresenius Vial recommends the presence of its qualified staff, or a member of your Technical
Department, who will assist you in implementing the configuration procedures you wish to
select.
To access some of the sub-menus, you need to type in an access code. This code can be
defined and/or de-activated in the Service configuration mode. As long as the device is kept
switched on, the code does not need to be keyed in again.
optima03.1_001b_en.fm 19
Test TEst.1 operating period and maintenance date
SAV.3 Beep
20 optima03.1_001b_en.fm
3.1.1 Access to the menu
Key Function
ON switches on the device.
OFF switches off the device and exits the configuration mode when
pressed for more than three seconds.
SILENCE ALARM
The selection keys scroll through the numbers and letters, the tenths,
units and tens values.
Activate or de-activate choices.
ENTER moves the cursor.
optima03.1_001b_en.fm 21
3.1.2 SAV1, preventive maintenance configuration
" By validating once again, the new running time and date
are memorised and you may return to the After-sales
Service menu.
By pressing "STOP" you can cancel the modification and return to the After-
sales Service configuration menu.
You can exit the configuration mode at any time by pressing the "OFF" key.
The tubes which have already been calibrated for pressure are marked with a "v".
" When the device is switched on, the selected tube will
appear on the display unit.
By pressing "STOP", the modification is cancelled and you can return to the
After-sales Service configuration menu.
Description of the menus
You can exit the configuration mode at any time by pressing the "OFF" key.
22 optima03.1_001b_en.fm
3.1.4 SAV3, "Beep"
This option is available on Optima pump equipped with software version 1.8e minimum.
! + 10 s. (not activated)
! + 0 s. (activated)
" Select your choice using the key.
# 10 s: the audible alarm "Beep" is triggered every 10
seconds.
# 0 s: the audible alarm "Beep" is triggered every 0
second.
" Press "START" to record your selection.
By pressing "STOP", the modification is cancelled and you can return to the
After-sales Service configuration menu.
You can exit the configuration mode at any time by pressing the "OFF" key.
optima03.1_001b_en.fm 23
Description of the menus
24
optima03.1_001b_en.fm
Description of the menus
The calibration menu is reserved for qualified staff only. Its access is protected by a
secret code. To find out about the operating mode for the various calibrations, refer
to the "Calibrations" chapter.
optima03.3_001b_en.fm 25
Description of the menus
26
optima03.3_001b_en.fm
Description of the menus
The After-Sales Service tests can also be carried out more quickly and simply from a PC
using our maintenance software. (Contact our After-Sales Service department.)
A log of more than 760 time-stamped events can be consulted using our maintenance
software "VP Control". (Contact our After-Sales Service department.)
optima03.4_002b_en.fm 27
Description of the menus
28
optima03.4_002b_en.fm
4 Preventive maintenance
4.1 Recommendations
The Optima pump can only be checked, serviced or repaired by Fresenius Vial or by a
certified and approved maintenance service. Any abnormal functioning of the pump must be
brought to the attention of your in-house qualified technical personnel or our After-Sales
Service.
Should you need to return the pump to our After-Sales Service, it should be cleaned,
disinfected and very carefully packaged, preferably in its original packaging, before being
sent.
For all information concerning the repair and use of the pump, kindly contact our After-Sales
Service or our Sales department.
Fresenius Vial is not liable for loss or damage to the pump during its transport to our After-
Sales Service.
Important: If these maintenance procedures are not observed, the pump’s correct
operation will be impaired.
optima04.1&2_001c_en.fm 29
Preventive maintenance
30 optima04.1&2_001c_en.fm
Preventive maintenance
4.3 Checks
In order to ensure the preventive maintenance follow-up, a regular servicing inspection is
recommended every 12 months (see "Quality Control Certificate").
Recharge the battery during 16h before starting the checking procedure !
Keyboard description.
Key Function
ON switches on the device.
OFF switches off the device when pressed for over three seconds.
SILENCE ALARM
The selection keys scroll through the numbers and letters on the tenths,
units and tens values, and activate or de-activate the infusion menus.
optima04.3_002c_en.fm 31
4.3.2 Visual check
Check the general aspect of the case and labels and check for shock impact damage.
The test is OK if all the indicator lights have lit up and the buzzer has gone off.
Preventive maintenance
You may interrupt the test at any time by pressing "STOP" and returning
to the test selection.
If the display is faulty, replace the display board (see "Display Board procedure).
32 optima04.3_002c_en.fm
4.3.5 Keyboard check
This test checks ! tESt.3, press "START".
that the keyboard
keys are
functioning " The following screen appears:
correctly. " Press each key one by one,
# The keys that have been pressed appear as a
reverse video image on the LCD screen.
If two keys or more are held down simultaneously, "beep! beep! beep!" will be heard,
indicating an error.
optima04.3_002c_en.fm 33
Alarm or error codes description:
Alarm Description
A10 Battery alarm
A12 End of infusion alarm
A16 Downstream occlusion alarm
A17 Air detection alarm
A18 Air volume alarm
A19 Upstream occlusion alarm
A20 Tube absence alarm
A21 Mobile clamp absence alarm.
A22 Door alarm
A23 Drip detector absence alarm
A24 Drip detector alarm: under flow
A25 Drip detector alarm: over flow
A26 Free flow
A27 Occlusion problem during auto-set
OFF Normal stop
FAIL Abnormal stop
Error Description
Er10 Internal RAM
Er20 External RAM
Er30 Check-sum Flash
Er40 Access (E2prom) Eeprom
Er50 Access Adc
Er60 Tube parameters
Er70 Motor frequency
Er80 Keyboard matrix (key press non functioning)
Er90 Check-sum Flash or software version µC secondary
Er01 Rotation (absence top turn)
Er11 Rotation (white or black too great)
Er21 Rotation (incorrect number of turns)
Er31 Automatic clamping motor
Er41 Rotation (presence top turn)
Er14 Period calculation parameters verification
Er24 Rotation direction verification
Er44 Incoherence quartz µC principal and µC secondary
Preventive maintenance
34 optima04.3_002c_en.fm
Error Description
Er65 Absence LCD display back light
Er85 Restart µC secondary
Er16 Date and time in µC secondary verification
Er56 Association condition output/step number
Er17 Keyboard matrix (key press and columns inactive)
Er27 Keyboard check default (path not fine tuned)
Er37 ON/OFF key locked
Er47 Switching off impossible
Er28 Language file incompatible
If the value is out of limits, recalibrate the force sensor. (See "EtAL9 upstream and
downstream pressure sensors calibration").
Preventive maintenance
optima04.3_002c_en.fm 35
4.3.10 Checking the software version
This test displays ! tESt.A, press "START".
the software
version and
" The LCD screen displays:
revision numbers.
# the name of the Optima pump.
# the software version.
# the date.
" Press the tenths keys to continue the test.
You may interrupt the test at any moment by pressing "STOP" and
returning to the test selection.
Preventive maintenance
36 optima04.3_002c_en.fm
4.3.12 Checking the calibration values
This test displays ! tEStE, press "START".
the calibration
values recorded in
the EEPROM. " Move to another value using the keys.
Display Description
Por1 Hall effect sensor, door closed without tube.
The value between brackets after the title is the number of calibrations carried out. The
display of the upstream and downstream pressure values is carried out according to the
number of active tubes.
" Press "START" to confirm the flow rate and start up the
motor.
The test may be interrupted at any time by pressing the "STOP" key and
Preventive maintenance
optima04.3_002c_en.fm 37
4.3.14 Checking the last 10 events before a locking alarm or error
This test enables ! teSt.J, press "START".
the visualisation of " The number of times the machine has been switched on
the codes for the since the last locking error is displayed on the LCD
last 10 events that screen.
have occurred
# "A" for an alarm. (See table teSt.5)
during operation of
the device before # "E" for an error. (See table teSt.5)
the last locking
error. " By validating once again, a new test may be seleted.
1 m 50cm
PU M P
M a n o m e te r
T a p 3 w a ys
C o n ta in e r
! Press "ON".
" Position a tube in the device and lock the door inserting
the clamp.
" Position the manometer at the outlet of the tube.
" Select a flow rate of 125 ml/h.
" Adjust the pressure threshold to 750 mmHg.
# Let the tap run in a container.
# Start up the infusion and leave it running for 30
seconds.
# Position the tap to the manometer.
# Check that the alarm is triggered for a value of 1 bar ±
0.15 bar.
" Stop the infusion by pressing "STOP".
Preventive maintenance
If the measured pressure value, depending to the limit selected is out of range, recalibrate
the pressure sensors. (See"ETAL9 Calibration of the upstream and downstream
pressure sensors".)
38 optima04.3_002c_en.fm
4.3.16 Flow rate check
! Connect the Optima pump to the mains.
Before any flow rate test, run out an infusion of 20 ml to stabilise the flow rate.
To carry out this test, we recommend to use the IS Flow rate software available with
Fresenius Vial.
optima04.3_002c_en.fm 39
4.3.19 Continuity test
Use a multimeter.
To carry out this procedure, exit the test mode. Press "OFF.
! Using a calibrated multimeter in ohms, check the presence of an electrical resistance
greater than 10 MΩ between:
" between the phase and equipotential pin.
" between the neutral and equipotential pin.
" By pressing the confirm key, you can exit the test.
" By pressing the confirm key you can exit the test.
Preventive maintenance
40 optima04.3_002c_en.fm
Preventive maintenance
Conformity
N° Procedure Obtained value
Yes No
1 ! Check the general appearance of the case, labels, serial number and product
code.
2 ! Check the contrast.
If a fault is observed, refer to chapter ’Contrast Check’.
3 ! Check the clock. If an adjustment is needed, refer to chapter on datestamp
checking.
4 ! Display the operating period, tESt.1and the maintenance date.
5 ! Check the indicator lights, tESt.2.
6 ! Check the keyboard, tESt.3.
7 ! Display the total running time, tESt.6.
8 ! Occlusion check, tESt.F
" Connect the water manometer to the tube outlet.
" Start up test F.
" Check that there is no motor error and that the maximum pressure is above
2 bar (1500 mmHg) for the version PT and ST,
above 1.6 bar for the version ST.
9 ! Check clamping and unclamping, tESt.C. (if OCS is present) .
" Confirm the two functions.
" Check for any absence of motor alarm.
10 ! Check the drop detector (if the detector is present).
" Start up an infusion.
" Check the (green) drop indicator lights up each time a drop appears in the drop
chamber.
" Disconnect the detector.
" Check if the drop alarm is activated.
11 ! Check the air detector.
" Create an air bubble in the tube of at least 5 cm long.
" Check the presence of the air bubble alarm (sound and visual) while the air
bubble is in the air bubble detector.
12 ! Check the downstream pressure.
" Install a calibrated water manometer at the tube outlet.
Preventive maintenance
optfiche_002b_en.fm 41
Conformity
N° Procedure Obtained value
Yes No
13 ! Check the upstream pressure. (Only Optima VS)
" Start an infusion at 100 ml/h without drop detector.
" Clamp the tube between the drop chamber and the pump.
" Check that the upstream occlusion alarm is activated within 10 seconds.
" Unclamp.
" Check that the confirmation indicator lights up.
14 ! Check the clamp.
" Place a water pipe in the device.
" Close and open the door.
" Check that the clamp lock the tube.
# There should be no dripping at the tube outlet.
15 ! Check the battery operation.
" Disconnect the mains lead.
" Check the presence of the battery indicator.
" Reconnect the mains lead.
" Check if the battery indicator is switched off and that the mains indicator lights
up.
16 ! Check the flow rate.
Before any flow rate test, run an infusion of 20 ml to stabilise the flow rate.
" Carry out a flow rate check in accordance with the norm EN 60601-2-24.
" Check that the flow rate error is between - 5% and + 5%.
If there is too much discrepancy, refer to chapter 7 "Calibration / Calibration
procedure / EtAL.2 calibrating the flow rate corrector coefficient".
17 ! Check the battery autonomy.
" Start an infusion at 125 ml/h for 4 hours.
18 ! Carry out an electrical safety test according to norm EN 60601-1-2
Observations:
Preventive maintenance
42 optfiche_002b_en.fm
Preventive maintenance
Do not use :
" ¨TRICHLOROETHYLENE-DICHLOROETHYLENE.
" AMMONIA.
" AMMONIUM CHLORIDE.
" CHLORINE AND AROMATIC HYDROCARBON.
" ETHYLENE DICHLORIDE-METHYLENE
" CHLORIDE-CETONE
These aggressive agents could damage the plastic parts and cause device malfunction.
Take car also with ALCOHOL based SPRAYS (20% -40% alcohol) .They lead to
tarnishing and small cracks in the plastic, and do not provide the neceeeary
cleaning prior to disinfection.
optima04.5&6_001c_en.fm 43
4.5 Storage
The device should be stored in a dry, cool place. In case of prolonged storage, the battery
should be disconnected via the battery access flap situated underneath the device.This
should be done by a qualified technician
! Storage conditions and carrying:
" Temperature: -10°C to 60°C / 14°F to 140°F.
" Atmospheric pressure: 500 hPa to 1060 hPa.
" Humidity: 10% to 90%, no condensation.
! Use conditions:
" Temperature: 5°C to 40°C / 41°F to 104°F.
" Atmospheric pressure: 700 hPa to 1060 hPa.
" Humidity: 20% to 90%, no condensation.
Fully recharge the battery before using the equipment to avoid any risks caused by micro
power cuts in the mains supply and to ensure maximum autonomy.
Preventive maintenance
44 optima04.5&6_001c_en.fm
5 Troubleshooting
5.1 Breakdown guide
Message on
Meaning Causes Action
LCD display
Door open ! Hall affect sensor ! Check set installation
alarm detects that the door ! Check the presence of the
is open door magnet
! Calibration of the door
sensor
! Replace the door sensor
Clamp ! The Optima does ! Check set installation
position not detect the ! Check the securing clamp
alarm presence of the set micro switch
securing clamp ! Replace it if necessary
Set ! Wrong position of ! Check set installation
insertion the set ! Check the presence of the
alarm door magnet
! Calibration of the door
sensor
! Replace the door sensor
OCS test ! Occlusivity test ! Check set installation
Alarm not successful. The ! Check the door and carry on
pumping mechanism a calibration if necessary
does not succeed to ! Verify the good condition of
increase the the membrane. Replace it if
pressure in the IV necessary; refer to the
set during the OCS maintenance protocol : PM 0094
test
Air bubble ! Air volume ! Prime the line to remove the
alarm detected is higher air
Than the limit ! Check set installation on the
programmed air detector Verify the air
! Set is not adjustment of the air size
properly inserted in programmed
the air detector ! Replace the air detector or
air bubble board
Troubleshooting
optima05.1_001a_en.fm 45
Message on
Meaning Causes Action
LCD display
Drop alarm ! The container is ! Verify the volume remaining
(If drop empty on the container
detector is ! The rolling clamp ! Verify if the roller clamp is
present) is closed open
! Over-flow , under ! Verify if the air inlet cap if
flow, free flow open when the pump is used
! The position of with a bottle
the drip sensor ! Verify the positioning of the
! Drop chamber is drop detector
full ! Verify the volume of the liquid
! Drop detector is in the drop chamber
disconnected while ! Verify the IV set position
the pump is ! Verify the drip sensor is well
configured to be connected
used with ! Verify the connector if it's
! The drip sensor damaged
is connected and ! Replace the drip sensor
disconnected
Free flow ! Over-flow ! Check set installation
alarm ! Set not properly ! Check drip sensor
(in case of positioned positioning
use with
drip
sensor)
46 optima05.1_001a_en.fm
Message on
Meaning Causes Action
LCD display
Air alarm ! The IV set is ! Remove the IV set and re-
present on the air install it
bubble detector but ! Carry on a calibration of the
not on the pumping door
mechanism
! The pump runs
over 24 hours, no air
bubble have
detected by the
sensor
Battery ! Low battery ! Connect the pump to the
alarm voltage mains power
! Check the charging voltage,
pump connected to the mains.
! Replace the battery
Preventive maintenance Preventive ! Preventive ! Proceed to the complete
24/10/2003 passed
maintenan maintenance date or check up of the pump and
ce time is passed update the maintenance date or
requested time ( test 1)
Mains LED's does ! Mains fuses out of ! Replace Check the mains fuses
not switch ON order ! Replace the power supply board
! Power supply out of ! If the problem remains, replace the
order CPU board
optima05.1_001a_en.fm 47
Problems Causes Action
Battery alarm while ! Battery is totally ! Check the charging voltage
it has been discharged ! Replace the battery
completely charged
Battery alarm ! Power supply or CPU ! Check the charging voltage
whereas the boards are faulty ! Check the power supply mains
Optima is fuses
connected to the
! Replace the power supply if
mains
necessary.
! If the problem remains, replace the
CPU board and carry on a complete
configuration
The module MVP ! Keyboard is defective ! Check the keyboard ( test 3)
turns ON or turns ! Replace the keyboard if necessary
OFF alone ! CPU board is ! Check that there was no liquid
defective infiltration on the CPU board
! Replace the CPU board and carry
on a complete configuration of the
device
Some keys from ! Keyboard is defective ! Check the keyboard (test 3)
keyboard out of ! Replace the keyboard if necessary
order
After a fall ! Mechanical elements ! Check that the mechanical
damaged assembly and the cases are intact
Pump noisy ! Power supply and ! Fix the wires as it shown on the
CPU wires are closed to technical manual
the motor (earth wire
also)
Troubleshooting
48 optima05.1_001a_en.fm
Troubleshooting
Er31 OCS DC motor failure ! Check the motor clamp for OCS and
rubber O ring
! Put some grease on the clamp finger
threaded screw. Replace the rubber
O ring if necessary (p/n: 174256)
Troubleshooting
optima05.2_002c_en.fm 49
Error codes Meaning Action
Er14 Motor period calculation ! Reconfigure the device with VP
anomaly Control software
! Replace the CPU board and carry
on a complete configuration of the
device
Er16 ! Date / time failure ! Check the date and the time clock
! Replace the CPU board and Carry
on a complete configuration of the
Troubleshooting
50 optima05.2_002c_en.fm
Error codes Meaning Action
Er17 ! Keyboard matrices default ! Replace the keyboard
! Check that there is no liquid
presence inside the device which could
damage keyboard and electronics
boards
Er47 ! OFF key out of order ! Replace the CPU board and carry
(no way to switch OFF the on a complete configuration of the
device) device via VP Control software
Troubleshooting
optima05.2_002c_en.fm 51
Troubleshooting
52 optima05.2_002c_en.fm
6 Intervention procedures
This chapter lists all the dismantling and re-assembling procedures.
Intervention procedures
optima06.0_001a_en.fm 53
Intervention procedures
54 optima06.0_001a_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima angle bracket to the
base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump to its original position.
1
2 2
optima06.0_0010b_en.fm 55
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
J1
3
J9 J6
J10
Intervention procedures
J8
! Remove the board from its housing, being careful with the display board J3 flat cable.
! Remove the CPU board.
56 optima06.0_0010b_en.fm
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation.
Correct
Incorrect
! Positioning of the wires connecting the power supply board and the CPU board.
Intervention procedures
Right Wrong
Wrap an adhesive silicon cord around the CPU board J12 and J11 connectors.
optima06.0_0010b_en.fm 57
! Perform a complete configuration for a new board using VP Control Maintenance
Software.
Each Optima pump has its own type of "display and CPU board"; it is important that you
do not mix up the different Optima pump references and that you order the one that
corresponds to your device.
58 optima06.0_0010b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician (see biomedical technical training documentation).
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1) connecting the Optima pump angle bracket to
the base.
! Remove the angle bracket, being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump back to its original position.
1
1
2
2
! Using an Allen key, gently slide out the two screws located under the door.
! Remove the door.
! Unscrew the 4 cruciform screws (marker 2 on Fig. 1 and 2) connecting the front cover to
the base.
! Remove the front cover, being careful not to pull out the CPU board flat cable.
optima06.0_0020b_en.fm 59
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
! Disconnect the J1 connector connecting the display board to the CPU board.
! Disconnect the display board J3 connector.
! Unscrew the cruciform screw (marker 3) situated on the display board, connecting it to the
front cover.
! Remove the display board.
J3
Intervention procedures
J1
J2
3
60 optima06.0_0020b_en.fm
Re-assembling
Each Optima pump has its own type of "display and CPU board"; it is important that you
do not mix up the different Optima pump references and that you order the one that
corresponds to your device.
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
When fixing the display board, ensure that the plastic inserts are not tightened too much
or they will be damaged.
When positioning the front case, ensure that the sealing is perfectly positioned in its
housing after re-assembling.
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. Also check that the earth
wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo on procedure sheet n°1.)
Intervention procedures
optima06.0_0020b_en.fm 61
Intervention procedures
62 optima06.0_0020b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump back to its original position.
1
1
2 2
! Using an Allen key gently slide out the two screws located under the door from their
Intervention procedures
groove.
! Remove the door.
! Unscrew the 4 cruciform screws (marker 2 on Fig. 1 and 2) connecting the front cover to
the base.
! Remove the front cover, being careful not to pull out the CPU board flat cable.
optima06.0_0030b_en.fm 63
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig.2
Dismantling
3
3 3
3
3
Intervention procedures
4 4
64 optima06.0_0030b_en.fm
! Unscrew the cruciform screw (marker 5), paying attention to the washer.
! Remove the pumping system.
5
Re-assembling
When replacing the mechanical bloc, it’s recommended to read the following instruction.
! If a good metal washer and a metal pulley are already mounted, let them in place.
old aluminium pulley with washer New aluminium pulley without washer
Check that there is no axial free play on the mechanical block cam shaft.
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
LWhen closing the pump ensure that the 2 cables connecting the power supply
Intervention procedures
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo in Procedure n°1.)
optima06.0_0030b_en.fm 65
Intervention procedures
66 optima06.0_0030b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump to its original position.
1
2 2
optima06.0_0040b_en.fm 67
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
J2
Intervention procedures
68 optima06.0_0040b_en.fm
Re-assembling
! Recover the self-adhesive block from the old LCD.
! Position the self-adhesive block on the LCD display, above the LCD display label.
! Position the LCD display on the display board.
! Press down the display on the display board so that the block adheres to the display
board.
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo on Procedure n°1)
.
! Carry out regular servicing checks. (See Quality Control Certificate.)
Intervention procedures
optima06.0_0040b_en.fm 69
Intervention procedures
70 optima06.0_0040b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump to its original position.
1
2 2
optima06.0_0050b_en.fm 71
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
Dismantling
! Unclip the non-oxidizing aluminium plate (marker 3) holding the air detector board in
place.
! Unclip the board.
3
Intervention procedures
72 optima06.0_0050b_en.fm
! Disconnect the J9 connector (marker 4) connecting the CPU board to the air detection
board.
! Disconnect the connector (marker 5) connecting the board to the air detector.
5
4
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the new unit.
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support. Incorrect positioning of one of these cables
could result in a noise being heard when the pump is in operation. (See photo in
Procedure n°1.) Carry out regular servicing checks.
Intervention procedures
optima06.0_0050b_en.fm 73
Intervention procedures
74 optima06.0_0050b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump to its original position.
1
1
2
2
optima06.0_0070b_en.fm 75
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
76 optima06.0_0070b_en.fm
Dismantling
2
3
2
4 4
5
Intervention procedures
optima06.0_0070b_en.fm 77
! Separate the pressure sensors from their support (see arrows)
! Unsolder the pressure sensor wires (marker 6).
! Remove the pressure sensors.
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
After re-assembling the pressure sensors, calibrate them. (See ETAL9, Calibrating
the upstream and downstream pressure sensors.)
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo in Procedure n°1.)
78 optima06.0_0070b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure
Access
! Pivot the Optima pump over the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima angle bracket to the
base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump to its original position.
2 1
2
optima06.0_0081b_en.fm 79
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig. 2
80 optima06.0_0081b_en.fm
Dismantling
4
3
4
4
4
6
5
5
5 Intervention procedures
optima06.0_0081b_en.fm 81
Re-assembling
Carry out the dismantling procedure in reverse order to reassemble the unit.
When re-assembling the motor clamp, apply a drop of silicon grease around the screw
path (marker 6).
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo in Procedure n°1.)
82 optima06.0_0081b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima pump angle bracket to
the base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
1
Intervention procedures
optima06.0_0090b_en.fm 83
! Unscrew the 2 cruciform screws (marker 4) holding the power supply shelf in place.
! Disconnect the earth wire (marker 5) and the mains connector (marker 6).
! Remove the power supply board.
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo in Procedure n°1.)
.
Intervention procedures
84 optima06.0_0090b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed:
! 1 Posidriv Z1 screwdriver
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.)
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1) connecting the Optima pump angle bracket to
the base.
! Remove the angle bracket, being careful not to pull out the connectors.
1
Intervention procedures
optima06.0_0100b_en.fm 85
Dismantling
! Unscrew the 2 cruciform screws (marker 2) holding the battery in place.
! Disconnect the battery.
! Remove the battery.
Re-assembling
Carry out the dismantling procedure in reverse order to re-assemble the unit.
We recommend carrying out a complete charging and discharging cycle to guarantee that
the battery will function correctly.
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo in Procedure n°1.)
86 optima06.0_0100b_en.fm
Intervention procedures
Safety:
For safety reasons, the technician should not intervene while the device is in operation.
Material needed
! 1 Posidriv Z1 screwdriver
! 1 anti-static bracelet
! 1 soldering iron
! Soldering wire "RADIEL Sn60Pb RI 1" (no cleaning necessary) or equivalent
Maintenance level:
Level 2, specialist technician. (See biomedical technical training documentation.).
Procedure:
Access
! Pivot the Optima pump on the top of the front cover.
! Unscrew the 2 cruciform screws (marker 1), connecting the Optima angle bracket to the
base.
! Remove the angle bracket being careful not to pull out the connectors.
! Disconnect the J4 and J3 connectors, as well as the functional Earth.
! Return the Optima pump to its original position.
1
1
2
2
Intervention procedures
optima06.0_0110b_en.fm 87
! Disconnect the CPU board flat cable: J2 connector and J4 connector.
2
2
Fig.2
88 optima06.0_0110b_en.fm
Dismantling the angle bracket
! Dismantle the battery. (See corresponding Procedure.)
! Dismantle the power supply board. (See corresponding Procedure.)
When closing the pump, ensure that the 2 cables connecting the power supply
board and the CPU board are positioned above the motor. In addition, check that the
earth wire is next to the metal support.
Incorrect positioning of one of these cables could result in a noise being heard when
the pump is in operation. (See photo in Procedure n°1.)
Intervention procedures
optima06.0_0110b_en.fm 89
Intervention procedures
90 optima06.0_0110b_en.fm
7 Calibration
7.1Calibration procedure
The calibration menu is reserved for qualified staff only. Its access is protected with
a secret code.
Keyboard description.
Key Function
ON switches on the device.
OFF switches off the device when pressed for more than three
seconds.
SILENCE ALARM
The selection keys scroll through the numbers and letters on the LCD
screen using the tenths, units and tens values.
Activate or de-activate the infusion menu.
ENTER confirms the selected calibration, or moves the cursor.
Calibration
optima07.1_001c_en.fm 91
How to activate the calibration mode
! Keep the "MODE" button pressed, during the device switch on.
" The various menus are displayed.
! Using the selection keys, select "Maintenance" on the
display.
! Press "START".
" Por.3: position a tube filled with water and close the door.
# Press "START".
The value is read and stored in the EEPROM.
Press "STOP" key to exits the menu without memorising the calibration
values.
Calibration
92 optima07.1_001c_en.fm
7.1.3 EtAL.2 Calibration of the flow rate corrective coefficient
Before any flow rate test, carry out an infusion of 20 ml to stabilise the flow rate. The
adjustment range allows a correction up to -10% to +10%.
This menu adjusts ! Place a tube filled with water on the device, taking into
the device flow account the positioning of the clamp, and close the door.
rate according to ! Select a flow rate of 100 ml/h.
the used tube. ! Start the infusion.
! Perform a flow rate measurement according to the norm
601-1-2-24.
! Note the programmed flow rate and the obtained flow rate.
! EtAL.2, press "START".
" Using the selection keys, enter the programmed flow rate.
" Press the "ENTER" key to move from one field to another.
" Enter the measured flow rate.
" Press "START" to start the calculation of the new
corrective coefficient.
# if the coefficient lies between 0.9 and 1.1, you may
return to the configuration menu.
# if the coefficient does not lies between 0.9 and 1.1
the "Error" message appears.
Pressing "STOP" exits the menu without calculating the new coefficient.
Calibration
optima07.1_001c_en.fm 93
7.1.4 EtAL.9 Calibrating the upstream and downstream pressure
sensors
This menu records The Optima can operate with various tube types. The choice of tube is made through:
the pressure value
! the "MODE" key, set selection when the pump is on normal mode.
in the EEPROM
according to the ! maintenance menu and SAV 2
selected tube. The pressure calibration must be done for each type of tube active on the pump.
If a pressure sensor, the door sensor or CPU board has been changed, the calibration
values must be reset to zero for each tube. Where the CPU board has been changed, the
calibration must be carried out using maintenance software during the configuration
procedure.
Installation:
! According to the diagram below:
Pocket
1 m 50 cm
Pum p
M a n o m e te r
T a p 3 w a ys
C o n ta in e r
The manometer must be placed at the same level as the Optima pump.
Calibration
94 optima07.1_001c_en.fm
! EtAL.9, press "START".
If the calibration of the pressure sensors should be performed only for the tube selected,
it’s recommennded to choose "NO".
When choosing "YES" a calibration for all the tubes programmed should be carried out
using VP control maintenance sofware.
When changing the force sensors, the calibration should be done for all the tube available
on the pump.
The calibration described below is concerns only the active tube chosen through mode
key / set selection or maintenance / SAV 2.
Calibrating the downstream and upstream pressure sensor with the door open.
" PaV.1,PaM.1: open the door of the Optima
pump, no tube installed.
optima07.1_001c_en.fm 95
Calibrating the downstream pressure sensor at 0 mm Hg with the door closed, filled
tube installed.
" Position the tube according to the preceding diagram and
open the 3-channel tap in the direction of the Pump-
Manometer.
" Press "START", to warm up the tube.
" Wait until the pressure reaches around 750 mm Hg
(1 bar +/- 0.05 bar), and open the 3-channel tap in the
pump-container direction to lower the pressure
(0 mm Hg).
" Press "STOP" to stop infusion.
Calibrating the downstream pressure sensor at 750 mm Hg with the door closed filled
tube installed.
" Position the 3 channel tap in the direction pump-manometer.
" Press "START" to start up the pump.
" Wait until the pressure reaches 750 mm Hg ±30 mm Hg
(1 bar +/- 0.05 bar).
" Press "STOP" to stop infusion.
Calibrating the upstream pressure sensor at 0 mm Hg with the door closed filled
tube installed.
" Keep the door closed.
" Open the 3-channel tap in the direction pump-container.
" Press "ENTER",
to record the pressure, when the pressure level
is back to 0 mm Hg.
Press "STOP" key to exit the menu without memorising the calibration
values.
Calibration
96 optima07.1_001c_en.fm
8 Spare parts catalogue
8.1 Upper cover
107
100
104
102
109
110
106
101
105
111
112 103
108
optima08.1_002c_en.fm 97
Spare parts catalogue
98 optima08.1_002c_en.fm
Spare parts catalogue
8.2 Base
207
205
203
... 200
201
202
204
...
....
optima08.2_002c_en.fm 99
....
....
....
....
....
.... ....
....
....
212 211
217 214
213
215
216
218
100 optima08.2_002c_en.fm
Spare parts catalogue
305
306
302 307
300
301
303
308
optima08.3_001b_en.fm 101
Spare parts catalogue
102 optima08.3_001b_en.fm
Useful addresses
TRAINING / SERVICE DEPARTMENT
Fresenius Kabi Tél. : 00 33 (0) 4 76 67 10 76
Le Grand Chemin, Tél. : 00 33 (0) 4 76 67 11 40
38590 Brézins Fax : 00 33 (0) 4 76 67 11 22
France
This document may not be reproduced in part or in whole without the written permission of
Fresenius Kabi