DET3TA, DET3TC, DET3TD and DET4TD - EN
DET3TA, DET3TC, DET3TD and DET4TD - EN
DET3TA, DET3TC, DET3TD and DET4TD - EN
INTRODUCTION 3
G SAFETY WARNINGS 4
G BATTERY INSTALLATION 5
GENERAL DESCRIPTION 6
INSTRUMENT ILLUSTRATIONS 11
BATTERY 18
Battery replacement 20
FUSE 21
DESCRIPTION OF TESTS 22
CAT IV 100 V Over voltage category IV is for equipment installed at or near the origin of the electrical
supply to a building.
100 V refer to the rms phase-to-earth voltage that this instrument can withstand to
the overvoltage category IV rating
Note: The safety warnings provided in this document are indicative of safe practice and shall not be considered
exhaustive. Additionally, they are not intended to replace local safety procedures where the instrument is being
used.
Note: This user guide uses the terms ‘earth’ and ‘ground’ interchangeably.
2 DET Series www.megger.com
Introduction
Thank you for purchasing the Megger Earth Ground Tester.
For your own safety and to get the maximum benefit from your instrument, please ensure that you read and understand the
following safety warnings and instructions before attempting to use the instruments.
This user manual describes the operation and functions of the DET series of Earth Ground Testers:
■ DET3TA
■ DET3TC
■ DET3TD
■ DET4TD2
■ DET4TR2
■ DET4TC2
■ DET4TCR2
■ MCC CLAMP
■ MVC CLAMP
■ CALIBRATION ADAPTOR
Megger Limited
Archcliffe Road
Dover
Kent
CT17 9EN
England
Megger Limited reserves the right to change the specification of these instruments at any time without prior notice.
■ The CAT IV 100 V rating is dependant on the use of a leadset of at least this rating, for example the optional two
wire leadset part number 1001-858.
■ Do not leave the instrument connected to the system under test when not in use.
■ Do not touch circuit connections and exposed metalwork of an installation or equipment under test.
■ Do not touch the earth stakes, test leads and their terminations (including connections to the earthing system under
test) if an installation earth fault can arise unless adequate precautions are taken.
■ Do not touch the earth stakes, test leads and their terminations (including connections to the earthing system under
test) while the instrument is switched on.
■ Do not operate the instrument or connect it to any external system if it shows any visible signs of damage or if it has
been stored for prolonged periods in unfavourable conditions.
■ Do not operate the instrument or connect it to any external system if the battery compartment or casing is open or
any parts of the case (including keypad, selector switch, display window, etc.) are missing.
■ Special precautions are necessary when operating in situations where “live” earths may be encountered: isolation
switches and fuses (not supplied with this instrument) must be used.
■ Special precautions are necessary when working near high tension systems (MV and HV): rubber gloves and shoes
(not supplied with this instrument) should be worn.
■ Special precautions are necessary when working in wet conditions or in agricultural areas: observe the local safety
standards and take all necessary special precautions applicable to the particular location and do not touch the test
leads with bare hands.
■ Always disconnect the instrument from the earthing system under test while batteries are being changed or the fuse
replaced.
■ Always replace batteries and fuses with parts of the correct type and rating.
■ Do not replace the rechargeable cells in the DET4TR2 and DET4TCR2 with non-rechargeable “dry” cells.
■ Do not use any other charging equipment other than that supplied by Megger for use with the DET4TR2 and
DET4TCR2.
■ Do not operate the charging equipment supplied with the DET4TR2 and DET4TCR2 in damp or wet environments or
outside.
CAT IV - Measurement category IV: Equipment connected between the origin of the low-voltage mains supply and the
distribution panel.
CAT III - Measurement category III: Equipment connected between the distribution panel and the electrical outlets.
CAT II - Measurement category II: Equipment connected between the electrical outlets and the user’s equipment.
NOTE: THE INSTRUMENT MUST ONLY BE USED BY SUITABLY TRAINED AND COMPETENT PERSONS!
Users of this equipment and/or their employers are reminded that National Health and Safety Legislation requires them
to carry out valid risk assessments of all electrical work so as to identify potential sources of electrical danger and risk of
electrical injury such as inadvertent short circuits. Where the assessments show that the risk is significant then the use of
fused test leads may be appropriate.
■ All persons involved must be trained and competent in isolation and safety procedures for the system to be worked
on. They must be clearly instructed not to touch the earth electrode, test stakes, test leads, or their terminations if
any ‘Live’ earths may be encountered. It is recommended that they wear appropriate rubber gloves, rubber soled
shoes, and stand on a rubber mat.
■ The earth electrode under test should be isolated from the circuit it is protecting before testing commences. If this is
not possible, ART may be used to measure electrode resistance.
■ The instrument terminals should be connected to the system under test through isolation switches that are rated to
handle the likely maximum fault voltages and currents that could be encountered at the installation. The isolation
switch must be open whilst any personal contact is made with the remote test stakes, or the connecting leads, e.g.
when changing their position.
■ The instrument terminals should be connected to the system under test through fuses that are rated to handle the
likely maximum fault voltages and currents that could be encountered at the installation.
G BATTERY INSTALLATION
WARNING : Whenever battery cells are being fitted or replaced, there should be no connections to the
instrument terminals and the instrument should be switched off.
WARNING : To avoid damage by leaking electrolyte or deep discharge, do not leave cells fitted in an instrument
which will remain unused for extended periods of time.
WARNING : Incorrect battery cell polarity can cause electrolyte leakage, resulting in damage to the instrument.
If the battery condition indicator does not show a full charge when battery cells are new, a cell may be
reversed.
WARNING : DET4TR2 and DET4TCR2 ONLY Warning: Only use the cells recommended by Megger, part number
1000-530.
WARNING : Do not attempt to recharge non-rechargeable (dry) cells in the DET4TR2 and DET4TCR2. Doing so
will result in instrument damage and may cause personal injury.
WARNING : Only use the charging equipment provided by Megger for use with this instrument.
■ The DET3TA provides 2 and 3-wire measurements and offers an analogue display.
■ The DET3TC and DET3TD provide 2 and 3-wire measurements and offer a digital display.
■ The DET4TD2, DET4TR2, DET4TC2 and DET4TCR2 provide 2, 3 and 4-wire measurement and offer a digital display.
■ The DET4TC2 and DET4TCR2 offer variable frequency testing (94 Hz, 105 Hz, 111 Hz and 128 Hz) and
measurement to 200 kΩ for 2P and 3P measurements.
The DET3TC, DET4TC2 and DET4TCR2 can use an optional current clamp (MCC CLAMP) to measure conductor current and
electrode (rod) resistance without disconnection, leaving the installation earthing system intact (Attached Rod Technique,
ART).
Additionally, the DET4TC2 and DET4TCR2 can drive an optional voltage-inducing clamp (MVC CLAMP) which, in
conjunction with the MCC CLAMP, can be used to make stakeless measurements of the earthing system.
DET4TCR2
DET4TD2
DET4TR2
DET4TC2
DET3TD
DET3TA
DET3TC
Feature
Automatic C stake check n n n n n n
Automatic P stake check n n n n n n
Manual P stake check n
Automatic noise check n n n n n n
Manual noise check n
Noise rejection (40 V pk-pk) n n n n n n n
Variable frequency test n n
2-wire test n n n n n n n
3-wire test n n n n n n n
4-wire test n n n n
2 kΩ range n n n n n n n
20 kΩ range n n n n
200 kΩ range n n
No disconnect testing (ART) n n n
Stakeless measurement n n
Voltmeter (ground noise voltage
n n n n n n n
measurement)
Current meter n n n
LCD display n n n n n n
Backlit display n n
Moving coil meter n
IP54 rated n n n n n n n
EN61010-1 100 V CAT IV n n n n n n n
Built-in battery charger n n
■ DET instrument
■ Test leads (for 3-wire instruments, lead lengths are: 3 m, 10 m and 15 m; for 4-wire instruments:
3 m, 10 m, 10 m and 15 m)
■ Test stakes (for 3-wire instruments: 2 stakes are supplied; for 4-wire instruments, 4 stakes are supplied)
■ Warranty card
Batteries
The Megger DET series instruments are supplied with batteries fitted. When batteries become exhausted, refer to the
section on battery replacement or recharging.
Warning: Do not switch the instrument on with the battery cover removed.
Inspection
Before each use of the instrument, visually inspect the instrument case, test leads, stakes and connectors to confirm that
their condition is good, with no damaged or broken insulation.
DET3TA:
1. Press and hold the TEST button and switch instrument ON to the V setting using the selector switch.
2. The needle will sweep across the scaleplate and return to the rest position. [On older DET3TC and DET3TD
instruments, the firmware version will be displayed followed by ‘tst’].
3. Release TEST button. The maximum output test voltage will be displayed, either ‘50 V’ or ‘25 V’.
4. To toggle between the maximum output test voltages, press the TEST button.
5. Switch off instrument when the desired maximum test voltage is displayed
NOTE : For ART mode, the output voltage is automatically set to 25 V and cannot be changed. Some older
versions of the DET3TC and DET3TD do not have the the 25 V/50 V button. The procedure for changing the
output voltage is as per the DET3TA.
ARTP Situation is suitable for making ART measurements [DET3TC, DET4TC2 and DET4TCR2 only].
ARTO Situation is not suitable for making ART measurements [DET3TC, DET4TC2 and DET4TCR2 only].
>100V Indicates that the ground noise voltage exceeds the instrument measurement capability (test is inhibited).
RpP Potential stake (P stake) is within range for accurate measurement.
RpO Potential stake (P stake) resistance exceeds range for accurate measurement.
RcP Current stake (C stake) is within range for accurate Measurement.
RcO Current stake (C Stake) resistance exceeds range for accurate measurement.
Ground noise voltage is within range for accurate measurement of resistance.
V P
Ground noise voltage exceeds range for accurate measurement of resistance.
V O
MCC CLAMP is connected; MVC CLAMP is connected [DET3TC, DET4TC2 and DET4TCR2 only].
P
MCC CLAMP is not connected; MVC CLAMP is not connected [DET3TC, DET4TC2 and DET4TCR2 only].
O
Sufficient MCC CLAMP current [DET3TC, DET4TC and DET4TCR only].
I P
Insufficient MCC CLAMP current [DET3TC, DET4TC2 and DET4TCR2 only].
I O
Rc
Current stake (C stake) is within range for accurate measurement.
Rc
Instrument is performing pre-measurement checks.
Rc
DET3TA
Multi-function analogue display Test lead
connections (at
rear
Rc OK
Resistance measurement
settings (2P and 3P)
Selector switch
Spike Test Battery check setting
button button voltage setting
Resistance scale
Voltage scale
Selector switch
Voltage
and current
Test button 25 V/50 V output measurement
voltage select button settings
Output Warning –
voltage Ruptured fuse refer to user Battery charge
indicator indicator manual indicator
Measuring range
Selector switch
Voltage
measurement
Test button 25 V/50 V output output
voltage select button settings
Output Warning –
voltage Ruptured fuse refer to user Battery charge
indicator indicator manual indicator
Measuring range
Selector switch
Output Warning –
voltage Ruptured fuse refer to user Battery charge indicator
indicator indicator manual
Measuring range
Selector switch
Voltage measurement
output settings
Test button 25 V/50 V output
voltage select button
Output Warning –
Ruptured fuse refer to user Battery charge
voltage
indicator manual indicator
indicator
Measuring range
Selector switch
Voltage measurement
settings
Frequency Backlight
Test button 25 V/50 V
select button control button
output voltage
select button
Output Warning –
voltage Ruptured fuse refer to user
indicator indicator manual Battery charge indicator
IART measurement
indicator
(Possible/Not possible)
Measuring range
Selector switch
Frequency Backlight
Test button 25 V/50 V select button control button
output voltage
select button
Output Warning –
voltage Ruptured fuse refer to user
indicator indicator manual Battery charge indicator
ART measurement
indicator
(Possible/Not possible)
Measuring range
Some charge remaining, but instrument may auto power down at any time.
Some change remaining, but instrument may auto power down at any time.
Note: If the DET4TR2 and DET4TCR2’s rechargeable batteries are replaced with alkaline batteries the instrument
may automatically disable its charging circuit to prevent inadvertent charging of non-rechargeable cells. Follow
the instructions ‘Re- enabling the DET4TR2 and DET4TCR2 charger circuit’ to switch the charger back on when
the NiMH cells are re-inserted.
Battery status indication (DET3TA)
1. Set the rotary selector switch to the position.
WARNING : Do not attempt to operate the instrument whilst the external AC/DC adaptor is connected.
WARNING : Do not attempt to recharge non-rechargeable (dry) cells in the DET4TR2 and DET4TCR2. Doing so
will result in instrument damage and may cause personal injury. Only use the cells recommended by Megger.
1. To avoid the possibility of electric shock, switch instrument OFF and disconnect the instrument from any electrical
circuits.
2. Move the slide on the terminal panel until the external AC/DC adaptor socket is exposed.
5. The charging cycle will commence and last approximately 17 hours. Follow the progress using the charge status LED
as shown in Figure 2..
NOTE: The instrument ambient temperature should be between +10 ºC (50 ºF) and +40 ºC (105 ºF) during the
charging cycle.
Fig 2: charge status LED progression
Battery charging complete.
Battery type
DET3TA, DET3TC, DET3TD, DET4TD2, DET4TC2:
8 x AA (LR6) 1.5 V Alkaline.
Megger part number: 25511-841.
DET4TR2, DET4TCR2:
8 x AA (LR6) 1.2 V NiMH 2000 mAhr.
Megger part number: 1000-530.
Battery replacement
WARNING : Do not operate instrument with the battery cover removed.
WARNING : Incorrect battery cell polarity can cause electrolyte leakage, resulting in damage to the instrument.
If the battery condition indicator does not show a full charge when battery cells are new, a cell may be
reversed.
WARNING : To avoid damage by leaking electrolyte or deep discharge, do not leave cells fitted in an
instrument, which will remain unused for extended periods of time.
1. To avoid the possibility of electric shock, switch instrument OFF and disconnect the instrument from any electrical
circuits.
2. The rear cover must not be opened if the test leads are connected.
3. To avoid the possibility of electric shock, do not press the TEST button or touch the fuse when changing batteries.
4. To remove the rear cover, release the screw at the bottom of the cover and lift the cover upwards.
6. Refit new batteries of the correct type observing the correct polarity as marked on the battery compartment.
7. Replace the instrument back cover and secure by tightening the retaining screw.
1. Switch the instrument on in to the 4P position while holding down the TEST button.
The display will carry out a self test then briefly display the software version number.
2. The screen will show the letters ‘tst’ – release the TEST button.
3. The state of the charger circuit is shown by either a cross or a tick being shown below the letters ‘CHg’. A cross
indicates that the charger circuit is disabled.
4. If the cross is shown, re-enable the charger by pressing the TEST button once (the cross should change to a tick).
Fuse status indication (DET3TC, DET3TD, DET4TD2, DET4TR2, DET4TC2 and DET4TCR2)
The fuse blown status indicator is displayed whilst the instrument is switched on and performing a test as shown:
f Fuse blown
Rc Fuse blown.
Fuse type
All instruments: 500 mA (F), HBC (50 kA, 600 V), 32mm x 6mm.
Megger part number: 25950-056.
Fuse replacement
WARNING : Do not operate instrument with the battery cover removed.
1. To avoid the possibility of electric shock, switch instrument OFF and disconnect the instrument from any electrical
circuits.
2. The rear cover must not be opened if the test leads are connected.
3. To remove the rear cover, release the screw at the bottom of the cover and lift the cover upwards.
4. To avoid the possibility of electric shock, do not press the TEST button when changing the fuse.
7. Replace the instrument back cover and secure by tightening the retaining screw
Suitable applications
Where there is doubt about a particular application, reference should be made to the advice and guidance contained in the
publication, “Getting Down to Earth” which is available from Megger, part number 21500-072.
The DET range of instruments can be used in agricultural locations (as per IEC 61557-5). In these circumstances, the output
voltage should be switched to 25 V in order to comply with the standard.
The DET range of instruments makes resistance measurements using a switched DC signal at a frequency of 128 Hz. The
DET4TC2 and DET4TCR2 provide the user with the option to select 94 Hz,
Note about measurements: when connecting the instrument to electrodes, ensure that any leads and cables are fully
unwound and laid out without loops.
The DET injects an a.c. current of known magnitude into the system under test and measures the voltage developed across
it as shown in Figure 3. The system resistance is a simple ratio as per Ohm’s Law, i.e. R=V/I.
Fig 3: schematic for two-terminal resistance measurement
The DET instruments automatically link the C1-P1 and C2-P2 terminals when a two-terminal test is selected
The DET injects an a.c. current of known magnitude into the system under test and measures the voltage developed across
it as shown in Figure 4. The system resistance is a simple ratio as per Ohm’s Law. In this case, the potential stake is moved by
fixed increments in a straight line between the electrode under test and the current stake. At each location, the resistance is
calculated as R=V/I. A graph of resistance versus potential stake position is plotted and the resistance of the electrode under
test is taken to be the point at which the curve is flattest.
Empirical testing has shown that with suitably positioned stakes, this method can be shortened by placing the potential
stake at a distance of approximately 62% between the electrode under test and the current stake, i.e. at A = 0.62 x B.
In this diagram, the C1 and P1 terminals are connected together at the electrode under test. This is the “three-terminal with
lead null” configuration which is only applicable in four-terminal testers. This configuration allows the resistance of the P1
lead to the electrode under test to be “nulled” out. For three-terminal testers or when lead null is not required, only the P1
terminal (or X terminal on a three-terminal instrument) connects to the electrode under test. This is shown in Figure 5.
Fig 5: schematic for three-terminal resistance measurement without lead null
By using a current transducer (the Megger MCC CLAMP) to measure the current flowing through the electrode under test
as a fraction of the total test current injected, the instrument can determine the individual resistance. This arrangement is
shown in Figure 6.
Fig 6: schematic for three-terminal resistance measurement using ART without lead null
In this configuration, the injected test current I splits along two paths into I1 (flowing into the connected earthing system)
and I2 (flowing into the electrode under test, i.e. I=I1+I2. The resistance of the electrode under test is calculated as R=V/I2
or R=V/ (I-I1). The current transducer (MCC CLAMP) measures I2 and feeds this value back to the instrument.
A defined test voltage is injected into the system using the MVC CLAMP, inducing a current, I, to flow and be measured
by the MCC CLAMP. The model shown in Figure 7 can be simplified to the resistance of the electrode under test, Rx and
the resistance of the other electrodes in parallel, i.e. R1 || R2 || … || Rn. Therefore, the current induced by the test voltage
is I=V/[Rx+(R1 || R2 || … || Rn)]. It follows that as the resistance of the other electrodes in parallel approaches zero, then the
resistance measured, approaches the value of the electrode under test.
Fig 7: schematic for two-clamp stakeless resistance measurement
For this test, the relative spacing and depth of the stakes is important. When configured as shown in Figure 8, the soil
resistivity can be calculated from the resistance value, R, displayed on the instrument as p = 2 x π x A x R.
4. The ground noise voltage reading will be displayed as shown in Figure 10.
Fig 10: example ground noise voltage reading (DET4 display shown)
Note:
■ The instrument will display the warning triangle and an excessive noise voltage indicator above
40 V pk-pk (14 Vrms).
■ The instrument will display the warning triangle and an over-range condition above 100 V.
4. The ground noise voltage reading will be indicated on the scaleplate as shown in Figure 12.
Fig 12: example ground noise voltage reading
5. DET4TC2 and DET4TCR2 only: Select the desired test frequency using the Hz button.
6. Press and release the TEST button (by holding the TEST button, the resistance measurement will be continually
updated).
7. The instrument will perform pre-measurement check, the status of which will be indicated on the display.
Note:
■ The test voltage used to make the two-terminal resistance reading is a.c. and may not be suitable for continuity
testing according to some local regulations.
■ The instrument will display the warning triangle and an excessive noise voltage indicator if the ground noise voltage
is above 40 V pk-pk (14 Vrms).
■ The instrument will display the warning triangle and a voltage over-range condition if the ground noise voltage is
above 100 V – no resistance reading is possible under these conditions.
6. If the ground noise voltage is less than 40 V pk-pk (14 V rms), press and hold the TEST button. [Testing will be
inhibited if the ground noise voltage exceeds 40 V pk-pk.]
7. The Rc LED will flash to indicate that the instrument is carrying out pre-measurement checks.
8. If the Rc LED does not illuminate following the pre-measurement checks, this indicates that the instrument fuse has
ruptured.
9. If the Rc LED illuminates solidly following the pre-measurement checks, this indicates that a two-terminal
measurement can be made.
10. The two-terminal resistance reading will be indicated on the scaleplate as shown in Figure 17.
Fig 17: example two-terminal resistance reading
Note:
■ The test voltage used to make the two-terminal resistance reading is a.c. and may not be suitable for continuity
testing according to some local regulations.
5. DET4TC2 and DET4TCR2 only: Select the desired test frequency using the Hz button.
6. Press and release the TEST button [by holding the TEST button, the resistance measurement will be continually
updated].
7. The instrument will perform pre-measurement check, the status of which will be indicated on the display.
Note:
■ The instrument will display the warning triangle and an excessive noise voltage indicator if the ground noise voltage
is above 40 V pk-pk (14 Vrms).
■ The instrument will display the warning triangle and a voltage over-range condition if the ground noise voltage is
above 100 V – no resistance reading is possible under this condition.
2. Select the desired test voltage using the procedure in the section on General Operating Instructions.
5. The ground noise voltage will be indicated on the scaleplate as shown in Figure 21.
Fig 21: ground noise voltage indication
6. If the ground noise voltage is less than 40 Vpk-pk (14 Vrms), press and hold the Rp button. [Testing will be inhibited
if the ground noise voltage exceeds 40 Vpk-pk.]
7. The P stake resistance will be indicated on the scaleplate: the needle will move into the green region (Figure 22) if
the P stake resistance is within limits for an accurate measurement and into the red region (Figure 23) if the P stake
resistance exceeds the limits for an accurate measurement.
Fig 22: P stake resistance OK (green region)
8. If the P stake resistance is acceptable (needle in the green region of the scaleplate), release the Rp button.
10. The Rc LED will flash to indicate that the instrument is carrying out pre-measurement checks.
11. If the Rc LED does not illuminate following the pre-measurement checks, this indicates that the C stake resistance
exceeds the limits for an accurate measurement or that the instrument fuse has ruptured.
12. If the Rc LED illuminates solidly following the pre-measurement checks, this indicates that a three-terminal
measurement can be made.
13. The three-terminal resistance reading will be indicated on the scaleplate as shown in Figure 24.
Note:
■ No resistance reading is possible if the ground noise voltage exceeds 100 V.
2. Select the desired test voltage using the procedure in the section on General Operating Instructions.
The pictures above show the DET3TD connected in the usual slope measurement, and the equivalent circuit. Re is the
resistance of the Earth electrode under test, and Rp and Rc are the Potential and Current spike resistances.
The instrument measures Re by passing a current between the X and C terminals via Rc (the current spike resistance) and
measures the voltage between the X and P terminals. As this is high impedance, there is no (or negligible) volt drop across
Rp and therefore this is the volt drop across Re. Knowing the current and voltage the resistance can be calculated.
Significant leakage or noise currents can be passing through Re so a 128Hz square wave current is used, and the phase
sensitive detector is used to measure the voltage giving excellent 50 and 60Hz ground noise voltage rejection.
When the potential or current spike resistances is greater than the value in the table below the reading will be shown with
0.1 Ω, 1 Ω or 0.1 k Ω resolution depending on the test selected.
A more precise result can be obtained when this happens by reducing the spike resistance.
5. DET4TC2 and DET4TCR2 only: Select the desired test frequency using the Hz button.
6. Press and release the TEST button [by holding the TEST button, the resistance measurement will be continually
updated].
7. The instrument will perform pre-measurement check, the status of which will be indicated on the display.
8. The three-terminal resistance with lead-null reading will be displayed as shown in Figure 26.
Fig 26: example three-terminal resistance with lead-null reading (DET4 display shown)
Note:
■ The instrument will display the warning triangle and an excessive noise voltage indicator if the ground noise voltage
is above 40 V pk-pk (14 Vrms).
■ The instrument will display the warning triangle and a voltage over-range condition if the ground noise voltage is
above 100 V – no resistance reading is possible under this condition.
5. DET4TC2 and DET4TCR2 only: Select the desired test frequency using the Hz button.
6. Press and release the TEST button [by holding the TEST button, the resistance measurement will be continually
updated].
7. The instrument will perform pre-measurement check, the status of which will be indicated on the display.
9. A soil resistivity value can be derived from the resistance reading and the stake geometry.
Note:
■ The instrument will display the warning triangle and an excessive noise voltage indicator if the ground noise voltage
is above 40 V pk-pk (14 Vrms).
■ The instrument will display the warning triangle and a voltage over-range condition if the ground noise voltage is
above 100 V – no resistance reading is possible under this condition.
5. The ground current flowing in the conductor will be displayed as shown in Figure 30.
Fig 30: example ground current reading (DET4 display shown)
Note:
■ The instrument will display the warning triangle above 2 A – no ART testing is possible under this condition.
■ The instrument will display the warning triangle and an over-range condition above 20 A – no ART testing is
possible under this condition.
5. DET4TC2 and DET4TCR2 only: Select the desired test frequency using the Hz button
6. Press and release the TEST button [by holding the TEST button, the resistance measurement will be continually
updated].
7. The instrument will perform pre-measurement checks, the status of which will be indicated on the display.
8. The three-terminal resistance reading using ART will be displayed as shown in Figure 32.
Fig 32: example three-terminal resistance reading using ART (DET4 display shown)
9. Under certain circumstances, the instrument may display a noise warning as shown:
This means that interference has been detected which may impair the accuracy of the measurement. In particular, the
reading could be lower than the actual resistance. The resistance of the electrode or system must be verified by an
alternative method.
Note:
■ The instrument will display the warning triangle and an excessive noise voltage indicator if the ground noise voltage
is above 40 V pk-pk (14 Vrms).
■ The instrument will display the warning triangle and a voltage over-range condition if the ground noise voltage is
above 100 V – no resistance reading is possible under this condition.
■ The instrument will display the warning triangle above 2 A – no ART testing is possible under this condition.
36 DET Series www.megger.com
■ The instrument will display the warning triangle and an over-range condition above 20 A – no ART testing is
possible under this condition.
■ Ensure that the MCC CLAMP jaw mating surfaces are free of dust and contamination and that they contact
completely when the MCC CLAMP is closed.
■ Currents carried by conductors in close proximity to the MCC CLAMP may affect calibration and reduce the
accuracy of measurements made.
Three-terminal resistance measurement with lead-null using ART (DET4TC2 and DET4TCR2)
In certain circumstances, the resistance of the test leads connecting to the electrode under test may itself form a significant
part of the MCC
measurement electrode resistance. This effect can be eliminated using the lead-null technique described in this
section.
Before making a measurement using ART, please follow the procedure contained in the section on MCC CLAMP calibration.
9. Under certain circumstances, the instrument may display a noise warning as shown:
This means that interference has been detected which may impair the accuracy of the measurement. In particular, the reading
could be lower than the actual resistance. The resistance of the electrode or system must be verified by an alternative method.
■ The instrument will display the warning triangle and an excessive noise voltage indicator if the ground noise voltage
is above 40 V pk-pk (14 Vrms).
■ The instrument will display the warning triangle and a voltage over-range condition if the ground noise voltage is
above 100 V – no resistance reading is possible under this condition.
■ The instrument will display the warning triangle above 2 A – no ART testing is possible under this condition.
■ The instrument will display the warning triangle and an over-range condition above 20 A – no ART testing is
possible under this condition.
■ Ensure that the MCC CLAMP jaw mating surfaces are free of dust and contamination and that they contact
completely when the MCC CLAMP is closed.
■ Currents carried by conductors in close proximity to the MCC CLAMP may affect calibration and reduce the
accuracy of measurements made.
Two-clamp stakeless measurement (DET4TC2 and DET4TCR2)
Before making a stakeless measurement, please follow the procedures contained in the sections on MCC CLAMP
calibration.
3. Close the MCC CLAMP around the conductor under test. Ensure the arrow on the side of the jaw is pointing in the
same direction as the arrow on the MVC CLAMP.
4. Close the MVC CLAMP around the conductor under test. Ensure the arrow on the side of the jaw is pointing in the
same direction as the arrow on the MCC CLAMP.
5. Ensure a minimum separation of 100 mm between the MCC CLAMP and MVC CLAMP.
7. DET4TC2 and DET4TCR2 only: Select the desired test frequency using the Hz button.
8. Press and release the TEST button [by holding the TEST button, the resistance measurement will be continually
updated].
9. The instrument will perform pre-measurement checks, the status of which will be indicated on the display.
11. Under certain circumstances, the instrument may display a noise warning as shown:
This means that interference has been detected which may impair the accuracy of the measurement. In particular, the
reading could be lower than the actual resistance. The resistance of the electrode or system must be verified by an
alternative method.
Note:
■ The instrument will display the warning triangle above 2 A – no “Stakeless” testing is possible under this condition.
■ The instrument will display the warning triangle and an over-range condition above 20 A – no “Stakeless” testing is
possible under this condition.
■ Ensure that the MVC CLAMP and MCC CLAMP jaw mating surfaces are free of dust and contamination and that
they contact completely when the MVC CLAMP and MCC CLAMP are closed.
■ Currents carried by conductors in close proximity to the MVC CLAMP and MCC CLAMP may affect calibration and
reduce the accuracy of measurements made.
■ If the MVC CLAMP opens at any time after the TEST button is pressed, the test will be aborted.
4. Press and hold the TEST button and switch instrument ON to the A setting using the selector switch.
6. Press and release the TEST button until a “0” reading is obtained on the display.
7. Close the MCC CLAMP around the calibration lead, which connects the C1 and C2 terminals as shown in Figure 38.
[X and C terminals for DET3TC]
Fig 38: MCC CLAMP 100% calibration
8. Press and release the TEST button until a “100” reading is obtained on the display.
10. The instrument is now calibrated to the MCC CLAMP and the calibration stored in non-volatile memory.
3. Close the MCC CLAMP around one loop of the two-clamp calibration adaptor.
4. Close the MVC CLAMP around the other loop of the two-clamp calibration adaptor.
5. Ensure a minimum separation of 100mm between the MCC CLAMP and MVC CLAMP.
7. Press and release the TEST button [by holding the TEST button, the resistance measurement will be continually
updated].
8. The instrument will perform pre-measurement checks, the status of which will be indicated on the display.
9. The two-clamp resistance reading will be displayed and should match the value written on the two-clamp calibration
adaptor.
Note:
1. Ensure that the MCC CLAMP and MVC CLAMP jaw mating surfaces are free of dust and contamination and that
they contact completely when the units are closed.
2. Currents carried by conductors in close proximity to the MCC CLAMP may affect calibration and reduce the
accuracy of measurements made.
4. Press and release the TEST button [by holding the TEST button, the resistance measurement will be continually
updated].
5. The instrument will perform pre-measurement checks, the status of which will be indicated on the display.
6. The resistance reading will be displayed and should match the value written on the calibration adaptor.
41 DET Series www.megger.com
PREVENTIVE INSTRUMENT MAINTENANCE
1. The DET series instruments require very little maintenance.
Voltage:
±25 V or ±50 V at 94 Hz, 105 Hz, 111 Hz and 128 Hz
(DET4TC2 and DET4TCR2)
Current:
4.5 mA or 0.45 mA
(DET3TA, DET3TC, DET3TD, DET4TD2 and DET4TR2)
Current:
4.5 mA or 0.45 mA and 0.045 mA (DET4TC2 and DET4TCR2)
ART:
0.01 Ω to 2000 Ω (DET3TC)
0.01 Ω to 200 kΩ (DET4TC2, DET4TCR2)
Stakeless:
0.01 Ω to 200 Ω (DET4TC2, DET4TCR2)
DET4TC2 DET4TCR2
Rp limit: 200 kΩ (50 V output voltage)
Rc limit: 200 kΩ (50 V output voltage)
Limits reduced to 100 kΩ for 25 V output voltage
Limits reduced to 5 kΩ for 0.01 resolution
DET4TCR2, DET4TR2
– 8 off AA (LR6) NiMH rechargeable cells
Operational uncertainties
Refer to www.megger.com
Weight: 1 kg / 2.2 lb
Safety IEC61010-2-032
EMC IEC61326-1
EMC IEC61326-1
ELECTRICAL SPECIFICATION
Value: 25 Ω ± 0.1%
MECHANICAL SPECIFICATION
Operating temperature -10 °C to +50 °C / 14 °F to 122 °F
0% to 85% RH at +35 °C / 95 °F
ELECTRICAL SPECIFICATION
Value: 25 Ω ± 1%
MECHANICAL SPECIFICATION
Operating temperature -10 °C to +50 °C (14 °F to 122 °F)
0% to 85% RH at +35 °C / 95 °F
Note: Any unauthorized prior repair or adjustment will automatically invalidate the Warranty.
Megger operate fully traceable calibration and repair facilities, ensuring your instrument continues to provide the high
standard of performance and workmanship you expect. These facilities are complemented by a worldwide network of
approved repair and calibration companies, which offer excellent in-service care for your Megger products.
■ Serial number, to be found on the underside of the case or on the calibration certificate.
2. Make a note of the RA number. A returns label can be emailed or faxed to you if you wish.
4. Ensure the returns label is attached, or that the RA number is clearly marked on the outside of the package and on
any correspondence, before sending the instrument, freight paid, to Megger. Copies of the original purchase invoice
and packing note should be sent simultaneously by airmail to expedite clearance through customs. In the case of
instruments requiring repair outside the warranty period, an immediate quotation can be provided when obtaining
the RA number.
WEEE
The crossed out wheeled bin placed on the Megger products is a reminder not to dispose of the product at the end of it’s
product life with general waste.
Batteries
The crossed out wheeled bin placed on the batteries is a reminder not to dispose of them with general waste at the end of
their life.
This product contains either six 1.5v alkaline AA batteries (DET3TD, DET3TC, DET4TD2, and DET4TC2) or six 1.2v NiMH
batteries (DET4TR2, DET4TCR2) located in the battery compartment on the rear of the instrument
Spent alkaline or NiMH AA batteries are classified as Portable Batteries and should be disposed of in the UK in accordance
with Local Authority requirements.
For disposal of batteries in other parts of the EU contact your local distributor. Megger is registered in the UK as a producer
of batteries. The registration number is BPRN00142
Declaration of Conformity
Hereby, Megger Instruments Limited declares that radio equipment manufactured by Megger Instruments Limited described
in this user guide is in compliance with Directive 2014/53/EU. Other equipment manufactured by Megger Instruments
Limited described in this user guide is in compliance with Directives 2014/30/EU and 2014/35/EU where they apply.
The full text of Megger Instruments EU declarations of conformity are available at the following internet address:
megger.com/eu-dofc.
Manufacturing sites
Megger Limited Megger GmbH Megger Valley Forge
Archcliffe Road Weststraße 59 400 Opportunity Way
Dover 52074 Aachen Phoenixville,
Kent GERMANY PA 19460
CT17 9EN T. +49 (0) 241 91380 500 USA
ENGLAND E. [email protected] T. +1 610 676 8500
T. +44 (0)1 304 502101 F. +1 610 676 8610
F. +44 (0)1 304 207342
The company reserves the right to change the specification or design without prior notice.
The Bluetooth® word mark and logos are registered trademarks owned by Bluetooth SIG Inc., and is
used under licence.
DET3TA--DET3TC--DET3TD--DET4TD2--DET4TR2--DET4TC2--DET4TCR2_UG_en_V19 08 2023