DSO1060 Manual
DSO1060 Manual
DSO1060 Manual
USER’S MANUAL
1060/1200/1600/1600H
www.hantek.com.cn
DSO1000 SERIES
USER’S MANUAL
1060 / 1200 / 1600 / 1600H
1060/1200/1600/1600H
Content
DSO1000 SERIES
WARNING: Warning statements identify conditions or practices that could result in injury
or loss of life.
■ Use Proper Power adapter. Use only the power cord specified for this product
and certified for the country of use.
1060/1200/1600/1600H
■ Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
■ Use Proper Probe. To avoid shock hazard, use a properly rated probe for your
measurement.
■ Do Not Operate With Suspected Failures. If suspected damage occurs with the
device, have it inspected by qualified service personnel before further operations.
Product features:
■ Dual Channel, Bandwidth:
60MHz (DSO1060)
200MHz (DSO1200)
600MHz (DSO1600)
600MHz (DSO1600H)
DSO1000 SERIES
1060/1200/1600/1600H
DSO1000 SERIES
General Check
When you have got a new DSO1000 series scope meter, it is suggested that you should
perform a general inspection on the instrument according to the following steps:
User’s Interface
The first thing to do with a new scope meter is to know its front panel. This chapter
helps to be familiar with the layout of the keys and how to use them. Read this chapter
carefully before further operations.
1060/1200/1600/1600H
DSO1000 SERIES
Description
1. LCD Display
2. F1~F5: Sets or switch options for the menu
3. Arrow Keys
4. HORI: Shows Horizontal menu
5. TRIG: Shows Trigger menu
6. LEVEL: Adjust the trigger level
7. RUN/STOP: key for running or stopping the operation
8. AUTO: Be used for auto setting under the oscilloscope operation mode
1060/1200/1600/1600H
Display screen
1 15 3 4 5 6 7 8
9
21
10
20 11
12
19 13
2
18 14
17
16
Description
1. Shows Brand
10
DSO1000 SERIES
11
1060/1200/1600/1600H
Input Connections
See the following Figure 1-4:
Description:
1. The power adapter is supplied for AC power supply and battery recharging.
2. Multi-meter input jacks, including four circular banana jacks .The four circular jacks
are used for voltage,resistance,mA range current and A range current inputs.
3. Multi-meter test lead.
4. Scope Meter probes.
5. Scope Meter input channels.
12
DSO1000 SERIES
Function Check
Perform this quick functional check to verify that your scope meter is operating correctly.
Set the probe attenuation of the scope meter to 10X.To do this, push CH1→Probe→10X
.
3. Push the AUTO button. Within a few seconds, a square wave will display
(approximately 1 kHz 2 V peak- to- peak).
13
1060/1200/1600/1600H
To compensate probes
Perform this adjustment to match the characteristics of the probe and the channel input.
This should be performed whenever attaching a probe to any input channel the first time.
1. From CH1 menu, set the probe attenuation to 10X (press CH1→Probe→10X).
Set the switch to X10 on the probe and connect it to CH1 of the scope meter.
When using the probe hook-tip, insert the tip onto the probe firmly to ensure proper
connection.
Attach the probe tip to the probe compensator connector and the reference lead to the
ground pin, Select CH1, and then press AUTO.
2. Check the shape of the displayed waveform.
Correctly Compensated
Over compensated
14
DSO1000 SERIES
Under Compensated
3. If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for
the flattest square wave being displayed on the scope meter.
4. Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the perfection of
the insulated cable, and do not touch the metallic portions of the probe head while it is
connected with a voltage source.
Press the AUTO button, the scope meter turns on and scales all channels that have
signals applied, and it selects a time base range based on the trigger source. The trigger
source selected is the lowest numbered input that has a signal applied. The scope meter
has two-channels input . Connect a signal to the CH1 input.
The scope meter may change the current setting to display the signal. It will automatically
adjust the vertical and horizontal scaling, the trigger coupling, type, position, slope, level,
and mode settings.
15
1060/1200/1600/1600H
Menu Operation
The following example shows how to use the tools menus to select a function, as shown
in the following figure.
16
DSO1000 SERIES
1. Press the MENU ON/OFF key to display the Function Menu on the bottom of the
screen and the corresponding optional settings on the bottom. Press MENU ON/
OFF again to hide the Function Menu.
2. Choose one key from F1 to F5 and press it to change function setting.
17
1060/1200/1600/1600H
■ The or button changes the time base in a 1-2-5 step sequence, and displays
thevalue in the status bar.
2. Move signal horizontally.
■ The or button moves displayed signal horizontally on waveform window. It sets
the trigger point position.
18
DSO1000 SERIES
19
1060/1200/1600/1600H
1. Press the CH1 or CH2 button and then the function menu appears at the bottom of the
screen.
2. Select and press key from F1 to F5 keys to make different settings.
Now, you can find the menu that looks like the following figure 2-1.
The CH1 menu (Page 1/2)
20
DSO1000 SERIES
2/2
Back to the previous menu page
3/3
Go to the next menu (not include DSO1060)
21
1060/1200/1600/1600H
1. Change Volt/DIV
It is the default setting of Volts/Div in a 1-2-5-step sequence from 1mV/div,2mV/div,5mV/
div or 10mV/div, 20mV/div, 50mV/div,..., to 1V/div,2 V/div,5 V/div.
The Volt/DIV will be displayed in the status bar on the bottom of the screen.
22
DSO1000 SERIES
23
1060/1200/1600/1600H
24
DSO1000 SERIES
Figure 2-6 Probe Setting
25
1060/1200/1600/1600H
4. Invert a waveform
The displayed waveform reverses 180 degrees relatively to the ground potential.
Press CH1 or CH2 → F5 → F3, to turn on/off the Invert.
26
DSO1000 SERIES
Press CH1→F5→F4→OFF, to set up bandwidth limit to “OFF” status. The scope meter
is set to full bandwidth and passing the high frequency component in the signal.
The waveform is displayed as Figure 2-9.
27
1060/1200/1600/1600H
Press CH1→F5→F4→ON, to set up bandwidth limit to “ON” status. It will reject the
frequency component higher than 20MHz.
Press CH1→F5→F5→F2→ON, to set up bandwidth limit to “ON” status. It will reject the
frequency component higher than 100MHz.This function is not used in DSO1060.
The waveform is displayed as Figure 2-10:
28
DSO1000 SERIES
6. Math Setting
The mathematic functions include “add”, “subtract”, “multiply”, “division”, and “FFT” for
CH1 and CH2. The mathematic result can be measured by the grid and the cursor.
The mathematic window is displayed as figure 2-11.
29
1060/1200/1600/1600H
2/2
Back to the previous menu page
30
DSO1000 SERIES
Figure 2-13 The multiplication
7. FFT
The FFT (Fast Fourier Transform) process converts a time-domain signal into its
frequency components mathematically.
31
1060/1200/1600/1600H
2/2
Back to the previous menu page
Note:
1. Signals that have a DC component or offset can cause incorrect FFT waveform
component magnitude values. To minimize the DC component, choose AC Coupling
on the source signal.
2. To reduce random noise and aliases components in repetitive or single-shot events,
set the oscilloscope acquisition mode to average.
3. To display FFT waveforms with a large dynamic range, use the “dBVrms” scale. The
“dBVrms” scale displays component magnitudes using a log scale.
FFT Window
The scope meter provides four FFT windows. Each window is a trade-off between
frequency resolution and amplitude accuracy. It depends on the desired
32
DSO1000 SERIES
measurement and the source signals characteristics to determine the window to use.
Use the following guidelines to select the best window.
8. REF
Reference Waveforms are saved waveforms to be selected for display. The reference
function will be available after saving the selected waveform to non-volatile memory.
Press M/R button to display the Reference waveform menu.
Select menu Enable->ON to enter reference function window. The reference function
window is displayed in Figure 2-15.
33
1060/1200/1600/1600H
Internal
Press M/R → REF → InternaI and go to the following menu.
REF menu when using internal memory
34
DSO1000 SERIES
External
Press M/R→REF→External and go to the following menu.
The External REF menu (Page 1/2)
Figure 2-18 The External menu
35
1060/1200/1600/1600H
Note:
The Reference function is not available in X-Y mode.
36
DSO1000 SERIES
center.
Horizontal position changes the displayed waveform position, relative to the trigger point.
Press HORI button to show the horizontal system menu. The settings of this menu are
listed in the following table.
1. TIME/DIV
Use this button to select the horizontal time/div (scale factor) for the main.
1 2 3
Marks Indicator
1. The current running status.
2. The trigger position in the memory.
3. The current waveform window’s position in the memory.
4. The trigger position in the current waveform window.
5. The horizontal time base (main time base).
ALT Mag
The ALT Mag is a magnified portion of the waveform window. Use ALT Mag to locate
and horizontally expand part of the main waveform window for a more detailed (higher
horizontal resolution) analysis of signal. The ALT Mag time base setting cannot be set .
38
DSO1000 SERIES
Description:
1. Wave to be horizontally expanded.
2.The magnified Wave.
3.The time base of ALT Mag.
4.The main time base.
X-Y Format
This format is useful for studying phase relationships between two signals. CH1 in the
39
1060/1200/1600/1600H
40
DSO1000 SERIES
Trigger Modes
The scope meter provides four trigger modes: Edge, Pulse , Alternative and Video.
Edge: An edge trigger occurs when the trigger input passes through a specified
voltage level in the specified slope direction.
Pulse: Use this trigger type to catch pulses with certain pulse width.
Alternative: Trigger on non-synchronized signals.
Video: Trigger on video-synchronized signals.
Press TRIG → F1 → Edge button to show the edge trigger menu as the following
table.
The Edge trigger menu (Page 1/2)
41
1060/1200/1600/1600H
Press TRIG → F1 → Pulse button to show the pulse trigger menu as the following table.
42
DSO1000 SERIES
Note: The Pulse width adjust range is10ns ~ 10s. When the condition is met, it will
trigger and acquire the waveform.
Press TRIG → F1 → ALT button to show the ALT trigger menu as the following table.
The ALT Trigger menu (Type Edge)(Page 1/2)
43
1060/1200/1600/1600H
Edge
Type Set Edge/Pulse Trigger as the trigger type
Pulse
Rising Trigger on rising edge
Slope
Falling Trigger on falling edge
1/2
Go to the next menu page
1 2 3 4 5
11
10
12
13 14
ON
HF Reject Reject high frequency signals or not
OFF
50% Set the trigger level to the center of the signal
2/2
Back to the previous menu page
45
1060/1200/1600/1600H
The Video trigger mode is designed to captrue the video signal type to display stable
NTSC or PAL/SECAM standard composite video waveforms.
Press TRIG → F1 → Video button to show the Video trigger menu as the following
table.
46
DSO1000 SERIES
2/2
Back to the previous menu page
47
1060/1200/1600/1600H
Term interpretation
■ Auto:
This sweep mode allows the scope meter to acquire waveforms even when it does
not detect a trigger condition. If no trigger condition occurs while the scope meter is
waiting for a specific period (as determined by the time-base setting), it will force itself
to trigger.
When forcing invalid triggers, the scope meter cannot synchronize the waveform,
and the waveform seems to roll across the display. If valid triggers occur, the display
becomes stable on the screen. Any factor results in the un-stability of waveforms can
be detected by Auto Trigger, such as the output of Power supply.
NOTE: When horizontal control is set more than 50 ms/div, Auto mode allows the
oscilloscope to capture without trigger signal.
■ Normal:
The Normal mode allows the scope meter to acquire a waveform only when it is
triggered. If no trigger occurs, the scope meter keeps waiting, and the previous
waveform, if any, will remain on the display.
■ Single:
In Single mode, after pressing the RUN/STOP key, the scope meter waits for trigger.
While the trigger occurs, the scope meter acquires one waveform then stop.
48
DSO1000 SERIES
Wave
Setup
Bitmap
49
1060/1200/1600/1600H
CSV
Factory
50
DSO1000 SERIES
Internal Memory
External Memory
51
1060/1200/1600/1600H
Factory
The scope meter has default settings and can be recalled at any time.
Memory location
Specify the memory location to save/recall the waveforms and setups.
Load
Recall saved waveforms, setups and default settings.
Save
Save waveforms and setups.
Note:
1. Select save stores not only the waveforms, but also the current scope meter.
2. The oscilloscope can store 15 settings permanently and can restore at any time.
Utility Function
52
DSO1000 SERIES
Press the Utility button to show the menu of the settings in the Utility system.
53
1060/1200/1600/1600H
54
DSO1000 SERIES
1/5
Go to the next menu page
The Utility menu table (Page 2/4 and Page 2/5)
Menu Setting Description
2/4
Back to the previous menu page
2/5
Back to the previous menu page
2/5
Go to the next menu page
3/5
Back to the previous menu page
5Min
10Min
Shut
20Min Set the shut down time
Down
30Min
Infinite
ON
Sound Turn beeper sound on/off
OFF
Calibrate Calibrate the scope meter
3/4
Go to the next menu page
55
1060/1200/1600/1600H
3/5
Go to the next menu page
4/5
Back to the previous menu page
4/5
Go to the next menu page
Calibrate
The calibration adjusts the internal circuitry to get the best accuracy. Use these functions
to calibrate the vertical and horizontal systems.
For maximum accuracy at any time, run this calibration if the ambient temperature
changes by 5°C or more.
56
DSO1000 SERIES
Note:
The scope meter must have been working or warm-up at least 30-minutes before
running calibration to get best accuracy.
The scope meter will calibrate parameter of vertical system (CH1, CH2).
Pass/Fail
The Pass/Fail function monitors changes of signals and output pass or fail signals by
comparing the input signal is within the pre-defined mask.
57
1060/1200/1600/1600H
Mask setting
58
DSO1000 SERIES
The Mask setting menu table (Page 2/2) When the save as internal memory
Menu Setting Description
2/2
Back to the previous menu page
The Mask setting menu (Page 2/2) When the save as external memory
59
1060/1200/1600/1600H
Waveform Recorder
Waveform recorder records input waveforms from CH1 and CH2, with a maximum record
length of 1000 frames.
60
DSO1000 SERIES
1/2
Go to the next menu page
61
1060/1200/1600/1600H
1/2
Go to the next menu page
62
DSO1000 SERIES
The Record menu when the save to Internal memory (Storage Mode) (Page 2/2)
Menu Setting Description
Save Save recorded waveform to internal memory location
Load Recall recorded waveform from internal memory location
Back to the Utility menu page
The Record menu when the save as external memory (Storage Mode) (Page 2/2)
Menu Setting Description
Create File Create a new file
Delete File Delete a selected file
Load Recall recorded waveform from external memory location
Language:
DSO1000 series scope meters have multi-language user menu, choose as your desire.
Signal Measurement
Press the MEAS button to display the menu of the automatic measurements settings.
63
1060/1200/1600/1600H
64
DSO1000 SERIES
65
1060/1200/1600/1600H
Note: The results of the automatic measurements will be displayed on the bottom of
the screen. Maximum 3 results could be displayed at the same time. When there is no
room, the next new measurement result will make the previous results moving left out of
screen.
Cursor Measurement
The screen displays two parallel cursors. Move the cursors to make custom voltage
or time measurements of the signal. The values are displayed on the boxes below the
menu. Before using cursors, make sure to set the Signal Source as the channel for
measuring.
Press CURSOR to show the cursor menu as the following:
66
DSO1000 SERIES
The scope meter measures the Y or X coordinate values of the cursors, and the
increments between the two cursors.
The values will be automatically displayed on the right upper corner of screen when the
cursor function menu is hidden or displaying other menus.
67
1060/1200/1600/1600H
68
DSO1000 SERIES
69
1060/1200/1600/1600H
IO Set
Press UTILITY→ F5 → F5→ F5 → F5→ F2 button to show the menu of IO Set
The IO Set menu (LAN)
70
DSO1000 SERIES
Press the button F2 in the IO Set menu(LAN) ,then a small window pops up.
You can select the Arrow keys to move the cursor position in which you want to modify
parameter, then press the Enter key to pop up a soft key board by which you can set
what you want .
71
1060/1200/1600/1600H
Press the button F2 in the IO Set menu(RS232),you can modify Baud Rate value.
72
DSO1000 SERIES
The Scope Meter set the vertical, horizontal, and triggers controls at the best status
automatically. Also, you can adjust the controls to meet your measurements to optimize
the waveform display.
To measure the frequency and Vpp, you can do these steps as followings:
1. Press the key MEAS , Select Source -> CH1 and select Type ->Voltage, then press
the key F5 and then press the key F2 to make a Pk to Pk test. The Vpp value will
display on the bottom of the waveform interface.
2. Press the key MEAS, select Source ->CH1and select type->Time, then Press the
key F5 and then press the key F2 to make a Frequency test. The Frequency value will
display on the bottom of the waveform interface.
3. Click the item Clear in the Measure menu, To clear the measurements on the
waveform interface.
See the measurements in the figure 3-1.
73
1060/1200/1600/1600H
74
DSO1000 SERIES
75
1060/1200/1600/1600H
Sinθ = A/B or C/D, where θ = phase shift (in degrees) between the two signals.
From the formula above:
θ must be in the range of (0~π/2) or (3π/2~2π) if the main axis of the ellipse is between
I and III quadrant, . If the main axis is at II and IV quadrant, θ must be in the range of
(π/2~π) or (π~3π/2).
Figure 3-4 FFT Measurement (Cursor Type X)
77
1060/1200/1600/1600H
Figure 3-5 The Pass/Fail Window
79
1060/1200/1600/1600H
To capture a single event, it needs to gather some pre-test knowledge of the signal in
order to set up the trigger level and slope correctly. For example, if the event is derived
from TTL logic, a trigger level of 2 Volts should work on a rising edge.
80
DSO1000 SERIES
CHAPTER 4: Multimeter
About this chapter
This chapter provides an introduction to the multimeter functions of DSO1000 Series.
The introduction gives guides to show how to use the menus and perform basic
measurements.
1
2
7 8 9
Figure 4-1 Multimeter operation window
Description
1) Battery electric quantity indictor.
2) Measurement mode indictors:
81
1060/1200/1600/1600H
Then, the screen will look like the following figure 4-2.
82
DSO1000 SERIES
83
1060/1200/1600/1600H
Then, the screen will look like the following figure 4-4.
84
DSO1000 SERIES
Figure 4-4 On-off Measurement
Then, the screen will look like the following figure 4-5.
85
1060/1200/1600/1600H
Then ,the screen will look like the following figure 4-6.
86
DSO1000 SERIES
Then, the screen will look like the following figure 4-7.
87
1060/1200/1600/1600H
Then, the screen will look like the following figure 4-8.
88
DSO1000 SERIES
Figure 4-8 DC current Measurement for 600 mA
89
1060/1200/1600/1600H
Then, the screen will look like the following figure 4-10.
90
DSO1000 SERIES
Then, the screen will look like the following figure 4-11.
91
1060/1200/1600/1600H
Then, the screen will look like the following figure 4-12.
92
DSO1000 SERIES
The default range mode of the instrument is automatic range. Suppose you are
using the DC voltage mode, to switch to the manual range, perform the following
steps:
1) Press F3 key to enter the manual range mode and then Manual is displayed on the
top of the screen.
2) Under the manual range mode, the measuring range is increased by a stage when
pressing F4 key each time, and when reaching the highest stage, it jumps to the
lowest stage by pressing F4 key once again.
3) Press F3 key to switch back to the automatic range mode and then Auto is displayed
on the top of the screen.
93
1060/1200/1600/1600H
94
DSO1000 SERIES
Chapter 5: Troubleshooting
1.The scope meter cannot start.
1) Check the power cable connection.
2) Ensure the power switch is turned on.
3) After the above inspection, restart the scope meter.
4) If the problem still remains, please contact us for help.
2. The measurement result is 10 times higher or lower than the value expected.
Check if the probe attenuation is the same as the channel attenuation.
3. The waveform is displayed on the screen but is not stable under the scope
meter mode.
1) Check the Trigger Source and notice if it is set to the channel in use.
2) Check the Trigger Type.
4. When select the average sampling in the sampling mode or select a long
display time in the display mode under the scope meter, the display speed is
slow.
It is normal in the above case.
95
1060/1200/1600/1600H
Chapter 6: Specifications
Vertical
Channels 2
DSO1060: 60MHz
DSO1200: 200MHz
Bandwidth DSO1600: 600MHz
DSO1600H: 600MHz
DSO1060: 5.8ns
DSO1200: 1.7ns
Rise Time DSO1600: 0.58ns
DSO1600H: 0.58ns
Input Impedance Resistance: 1M;Capacitance: 15pF
DSO1060: 10mv/div to 5v/div
DSO1200: 10mv/div to 5v/div
Input Sensitivity DSO1600: 1mv/div to 5v/div
DSO1600H: 1mv/div to 5v/div
Input Coupling AC,DC with Ground Level Indicator
Vertical Resolution 8bits
Memory Depth 32k at single channel; 16k at dual channels
Maximum Input 300V(DC+AC Peak)
Horizontal
DSO1060: 150MSa/s
DSO1200: 500MSa/s
Sampling DSO1600: 1GSa/s
DSO1600H: 2GSa/s
Equivalent Sampling 50GSa/s
DSO1060: 5ns/div~1000s/div
DSO1200: 2ns/div~1000s/div
Time Base
DSO1600: 1ns/div~1000s/div
DSO1600H: 1ns/div~1000s/div
Time Base Precision ±50ppm
Trigger
96
DSO1000 SERIES
Source CH1,CH2
X-Y Mode
X-Axis Input CH1
Y-Axis Input CH2
Phrase Shift Max.3 degree
Meter Mode
Maximum Resolution 6000 Counts
Meter Specification
97
1060/1200/1600/1600H
98
DSO1000 SERIES
Display
TFT LCD Type 5.7 Inch width LED Backlight Display
Display Resolution 240 (Vertical) X 320 (Horizontal) pixels
Interface
USB USB 2.0
Optional RS232,LAN
Power Source
AC 100V ~ 240V, 50Hz ~ 60Hz;
Line Voltage Range
DC Input:8.5VDC,1500mA
Battery Power(Installed) 6 hours(Li-ion Battery)
Mechanics
Dimension 245 x 163 x 52 (mm)
Weight 1.2 Kg
Other
GND Reference Scope meter and Multimeter Independence
Caution
Technical specifications may be jubject to variations without notice
99
1060/1200/1600/1600H
Chapter 7: Appendix
Appendix A: Accessories
Appendix B: Warranty
We warrant our products’ mainframe and accessories in materials within the warranty
period. During the period concerned,
We guarantee the free replacement or repair of products which are approved defective.
To get repair service or obtain a copy of the whole warranty statement, please contact
with your nearest sales and service office.
We do not provide any other warranty items except the one being provided by this sum-
mary and the warranty statement. The warranty items include but not being subjected to
the hinted guarantee items related to tradable characteristic and any particular purpose.
We will not take any responsibility in cases regarding to indirect, particular and ensuing
damage.
Appendix C: Maintenance
General Care
Do not store or leave the instrument in where the LCD display will be exposed to direct
sunlight for long periods of time.
Caution: To avoid damage to the instrument or probes, do not expose them to sprays,
liquids, or solvents.
Cleaning
If this instrument requires cleaning, disconnect it from all power sources and clean it with
a mild detergent and water. Make sure the instrument is completely dry before recon-
necting it to a power source.
To clean the exterior surface, perform the following steps:
1) Remove loose dust on the outside of the instrument and probes with a lint- free cloth.
Take care to avoid scratching the clear plastic display filter.
2) Use a soft cloth dampened with water to clean the instrument.
Notice: To avoid damage to the surface of the instrument or probes, do not use any
abrasive or chemical cleaning agents.
Storage of Scope meter
If the test tool is to be stored for a long time, it is required to charge the lithium battery
100
DSO1000 SERIES
before storage.
Charging the scope meter
The lithium battery is possibly not charged when delivery. To make the battery with
enough electric quantity, it must be charged for 4 hours (the test tool must be turned off
during charging). The battery can supply power for 6 hours after being charged com-
pletely.
To avoid superheat of battery during charging, the environment temperature is not al-
lowed to exceed the permissible value given in technical specification.
Note:
No hazard will occur even connecting the charger for a long time, e.g. during a whole
weekend. The instrument can automatically switch to slowly charging status.
Replacing the Lithium Battery Unit
It is usually not required to replace the battery unit. But when it is required to replace it,
only qualified personnel can carry out this operation, and only use the same specification
lithium battery.
101
1060/1200/1600/1600H
102