P 900 Pump
P 900 Pump
P 900 Pump
ÄKTAdesign
Pump P-900
User Manual
ÄKTA™
Important user information
All users must read this entire manual to fully understand
the safe use of the component.
WARNING!
The WARNING! sign highlights instructions that
must be followed to avoid personal injury. It is
important not to proceed until all stated conditions
are met and clearly understood.
CAUTION!
The Caution! sign highlights instructions that must
be followed to avoid damage to the product or
other equipment. It is important not to proceed
until all stated conditions are met and clearly
understood.
Note
The Note sign is used to indicate information
important for trouble-free and optimal use of the
product.
CE Certifying
This product meets the requirements of applicable CE-
directives. A copy of the corresponding Declaration of
Conformity is available on request.
The CE symbol and corresponding declaration of
conformity, is valid for the instrument when it is:
– used as a stand-alone unit, or
– connected to other CE-marked GE Healthcare
instruments, or
– connected to other products recommended or
described in this manual, and
– used in the same state as it was delivered from
GE Healthcare except for alterations described in this
manual.
Recycling
This symobl indicates that the waste of electrical
and electronic equipment must not be disposed
as unsorted municipal waste and must be
collected separately. Please contact an authorized
representative of the manufacturer for information
concerning the decommissioning of equipment.
Contents
Contents
1 Introduction
1.1 General .....................................................................................................7
1.1.1 Accessories ...........................................................................................................8
1.2 Safety .....................................................................................................................9
2 Installation
2.1 Unpacking ............................................................................................ 11
2.2 General precautions........................................................................ 11
2.3 Connecting electrical signal cables .........................................11
2.3.1 Connecting to chart recorder (if used) .................................................. 12
2.4 Connecting to communication link ..........................................12
2.5 Connecting to supply voltage ....................................................12
2.6 Installing the piston seal rinsing system ...............................13
2.7 Connecting the inlet tubing .........................................................14
2.8 Purging the pump ............................................................................ 14
2.9 Running-in the new pump ...........................................................15
2.9.1 General ................................................................................................................ 15
2.9.2 Running-in ......................................................................................................... 15
2.10 Connecting the pump outlet tubing ........................................16
2.11 Installation of accessories ..........................................................16
3 Operation
3.1 On/off ..................................................................................................... 19
3.2 Menu selection and settings .......................................................19
3.2.1 Menu selection ................................................................................................ 19
3.2.2 Return to main menu ................................................................................... 20
3.2.3 Select value ....................................................................................................... 20
3.3 Main menu overview ......................................................................20
3.4 Starting and stopping the pump ...............................................21
3.4.1 Main operating menu ................................................................................... 21
3.4.2 Mode changing menu .................................................................................. 21
3.4.3 Preparation before starting ....................................................................... 21
3.4.4 Emergency stop .............................................................................................. 22
3.5 Setting the flow rate and starting the pump ......................22
3.6 Setting concentration B.................................................................22
3.7 Running a simple gradient ...........................................................23
3.8 Ending the run and storage ........................................................23
3.9 Changing eluent ................................................................................ 24
3.10 Restart after power failure ...........................................................24
4 Maintenance
4.1 Periodic maintenance ....................................................................25
4.2 Cleaning-in-place ............................................................................. 25
4.3 General care ....................................................................................... 25
4.3.1 General recommendations for all eluents ........................................... 26
4.3.2 Additional recommendations for aqueous eluents ......................... 26
4.4 Replacing the piston seal .............................................................26
4.5 Running-in a new piston seal .....................................................31
4.6 Replacing a damaged piston ......................................................32
4.7 Cleaning the inlet and outlet check valves ..........................32
4.7.1 Cleaning after removal ................................................................................ 32
5 Trouble shooting
5.1 General .................................................................................................. 35
5.2 Faults and actions ............................................................................35
5.3 Error messages .................................................................................. 36
5.4 Checking the pump pressure......................................................37
5.5 Removing trapped air bubbles from the pump .................38
A Description
A.1 Instrument ........................................................................................... 39
A.2 Pump head .......................................................................................... 41
A.3 Pump principle ................................................................................... 43
A.4 Outlet manifold .................................................................................. 43
A.5 Pressure monitoring ........................................................................44
A.6 Using an external chart recorder .............................................44
B Menus
B.1 Check menu ........................................................................................ 45
B.1.1 Checking piston stroke ................................................................................. 45
B.1.2 Check pump operation time ...................................................................... 45
B.1.3 Check number of piston strokes .............................................................. 45
B.1.4 Check Service Mode ...................................................................................... 45
B.1.5 Calibrate Pressure Offset ............................................................................ 46
B.2 Setup menu ......................................................................................... 46
B.2.1 Set pressure limit ............................................................................................ 46
B.2.2 Set pressure full scale on recorder ......................................................... 46
B.2.3 Set pressure unit ............................................................................................. 46
B.2.4 Set gradient base ........................................................................................... 46
B.2.5 Set gradient mode ......................................................................................... 47
B.2.6 Set pump type ................................................................................................. 47
B.2.7 Set compression compensation .............................................................. 47
B.2.8 Set mixer operation ....................................................................................... 48
B.2.9 Set valve operation ........................................................................................ 48
B.2.10 Setup language ............................................................................................... 48
B.2.11 Setup unit number ......................................................................................... 49
B.2.12 Setup display angle....................................................................................... 49
B.3 Alarm timer .......................................................................................... 49
B4 Service displays................................................................................. 50
B.5 Menu overview .................................................................................. 51
C Technical specifications
C.1 Operating data .................................................................................. 53
C2 Physical data ...................................................................................... 54
6 Short instructions
12 59
1 Introduction
1.1 General
Pump P-900 is the collective name for a pump family. The family members
are:
P-9
00
Out to system
P-901, P-903
All P-900 models are high performance laboratory pumps for use in liquid
chromatography and other applications where accurately controlled liquid
flow is required. Double mode can be selected to double the flow rate range.
• P-901
• P-903
• The pumps are equipped with 2 pump modules, A and B, including two
pump heads each. This allows for binary gradients with high pressure
mixing.
• Low pulsation.
All versions are run with the same software. Normal or double mode, as well
as the size of the pump heads are stated in a Setup meny, see Reference
information B.2.
1.1.1 Accessories
Accessories can be connected to the pumps, and can be controlled from the
pumps or from a PC running UNICORN™ version. 2.10 or higher.
• Mixer M-925.
1.2 Safety
IMPORTANT! Pump P-900 is intended for laboratory use only, not for
clinical or in vitro use, or for diagnostic purposes.
2 Installation
2.1 Unpacking
Unpack the instrument and check the items against the supplied packing
list. Inspect the items for obvious damage which may have occurred during
transportation.
It is recommended that all packing materials should be retained if onward
transport of the instrument is expected.
Valve 1
Valve A
Valve B
Pressure,
analog
out 0 - 1V UniNet 2 Remote UniNet 1 Mains
WARNING! Do not block the rear panel of the system. The mains power
switch must always be easy to access.
Pin Signal
1 Signal
2 Signal ground
3–6 Not used (Reserved for factory testing)
Note: The signal cable is delivered with protective covers on each wire.
Do not remove the protective covers from unused connections as a
short circuit may disturb the measurements.
2 Set the recorder to 0–1 V input, full scale.
3 Refer to section B.2.2 for a description of setting the full scale pressure.
The rinsing system for the piston seals must be installed before use.
20% ethanol should be passed around the piston and piston seal on the
non–pressure side of the piston seal. This protects the piston seals and
pump heads by preventing a build-up of deposits from the solvents.
1 Connect the tubing for the piston seal rinsing system. The connections
are simple plug-in fittings.
2 The three lengths of wide bore tubing connect the pump heads in series,
connecting the top of one pump head to the underside of the next. The
wide bore tubing with the check valve should be fitted between the right
pump head of pump module A and the left pump head of pump module
B. The arrow on the check valve must point towards pump A.
A B 3 Two lengths of narrow bore tubing are also supplied with the pump.
Connect the tubing with a Union Luer connector to the underside of
P-901, P-903
the left pump head on pump A. Connect the other to the top of the
right pump head of pump module B and immerse the tubing in a flask
containing rinsing solution (20% ethanol).
5 Loosen the syringe and immerse the tubing in the rinsing solution.
Purge tubing
1 Fill a reservoir flask with distilled water. Immerse the inlet tubings of both
pump modules, with filters, in the water.
Note: Never place the reservoir flask below the level of the pump inlet.
3 Connect the male Luer connector at the other end of the purge tubing to
the left purge valve at pump module module A.
4 Turn the purge valve counterclockwise half a turn to open it and slowly
draw eluent to the syringe.
5 When fluid starts to enter the syringe continue to draw a few millilitres
before closing the purge valve. Check that there is no air left in the inlet
tubing.
The pump heads are filled with 20% ethanol for protection during
transportation. This is removed by running-in the pump with a liquid such
as distilled water which is miscible with 20% ethanol. After running-in, the
pump is primed for the chromatographic run.
Note: If, for any reason, the pumpheads are dry, actions according
to section 4.5 Running in a new piston seal must be carried out
instead!
2.9.2 Running-in
1 Purge the pump, see section 2.8 Purging the pump.
3 Check that the outlet tubing are not blocked, and press OK in the pump
synchronization display.
4 Set the flow rate in menu Set Flow Rate. Run both pumps at the same
time by setting the concentration to 50% in menu SetConcentrationB
under menu SetConc./Gradient. Set the pump in Run mode. Run the
pump at the flow rates and times specified below.
1 Remove the blind plug from the pump head outlet manifold. Connect the
outlet tubing using 1/16” Fingertight connectors. Place the end of the
tubing in a drain or waste collection vessel.
Outlet tubings
3 Set the flow rate to 10 ml/min for P-901, and to 1 ml/min for P-903.
4 Set the pump in Run mode. Allow the pump to run for 1–2 minutes to
prime the new outlet tubing.
0, 6
M-9
25
SV-903 valve for buffer selection is connected to Valve 1. The valve is set
on/off via the pump sub menu Set Valve 1 under the Setup menu. See
Reference information B.2.8.
Valve 1
Valve A
Valve B
Pressure,
analog
out 0 - 1V UniNet 2 Remote UniNet 1 Mains
Out to system
0, 6
M-9
25
3 Operation
3.1 On/off
Switch on the instrument at the mains switch on the rear panel. At switch on
the instrument performs a selftest and then asks for synchronization.
Open the purge valves (see section 2.8) to avoid high pressure during
synchronisation which is done at the speed of 0.1 ml/min, and press OK.
The display will show Synchronizing pumps and after approx. 1 minute the
display shows the main operating menu with the pump in End mode. Close
the purge valves. All parameters are factory set to default values.
If a menu has sub levels, the sub menu is displayed by pressing the OK-
button. Pressing the ESC-button moves back one menu level.
Main menu 1
Main menu 2 OK OK
Sub menu 2.1 Sub menu 2 .1.1
ESC ESC
Main menu 3
Sub menu 2.2
Main menu 4 OK
Sub menu 2.3 Sub menu 2.3.1
ESC
Main menu 5
Sub menu 2.3.2
Main menu 1
ESC
Main menu 2 ESC ESC ESC ESC
When setting numerical values the cursor moves up to the next digit if the
dial is turned quickly in one direction, to simplify entering large values. The
cursor moves back one place to the right every two seconds if the dial is not
turned. The text or numerical value displayed is accepted by pressing the K-
button. To cancel, press the ESC-button.
Main operating menu: The menu is accessed from all positions by pressing
the ESC-button repeatedly.
Setup language, pressure limits, etc. See Reference information section B.2.
Set different timer options. The pump can be started or stopped at set times.
See Reference information section B.3.
CAUTION! Before the start of each run, always ensure that there is an
adequate supply of eluent in the reservoirs. Never allow the pump to run
dry, since this will affect the lifetime of the piston seals.
1 Check that there is sufficient eluent present for the run, and that the
solvent filter is fully immersed. If the eluent is to be changed,
see section 3.9 Changing eluent
Note: The pump may not work if the buffer vessels are sealed. Do not
close the vessels off completely.
ÄKTAdesign Pump P-900 User Manuall 18-1120-04 Edition AE 23
3 Operation
2 If there is air in the tubing or the pump, prime the pump. Refer to
section 2.8 Purging the pump.
3 Check that there is sufficient liquid (20% ethanol) in the rinsing system
reservoir.
3 Select the mode menu, select Run, press OK. The pump speed gradually
increases to deliver the set flow rate.
4 To change the flow rate while the pump is running, perform steps 1 and
2 again.
5 To stop the pump, select the mode menu, select End, press OK.
Current
concentration
0%
Gradient
started
2 The display shows Set Concentration B. Turn the dial to select menu Set
Gradient Length, press OK.
4 Select menu Set GradientTarget. Set the gradient target value in %B,
press OK. If the pump is in Run mode the gradient will start immediately.
5 At the end of the set gradient length time, the pump continues to run at
the target concentration.
If no further runs are planned, the pump should be flushed immediately with
pure eluent. If aqueous buffers have been in use, flushing with pure distilled
water is particularly important to prevent salt precipitation.
Weekend and Long time storage: Flush the pump with water and then fill it
with 20% ethanol.
2 Transfer the inlet tubing into the new eluent or into the intermediate
liquid.
3 Run the pump at a flow rate and time as specified in the table in
section 2.9.
4 Stop the pump. If an intermediate liquid is being used, transfer the inlet
tubing into the final eluent and repeat step 3 with the new eluent.
4 Maintenance
Note: The piston seals have a limited life time depending on the flow rate,
pressure and eluents used. The seals are regarded as consumables
and are available as a service kit. The wear of the piston seals is
not covered by the warranty of the instrument. A typical lifetime is
around 1000 run hours
WARNING! Remove liquid or dirt from the system surface using a cloth
and, if necessary, a mild cleaning agent.
4.2 Cleaning-in-place
Pump a cleaning or sanitizing agent through the pump. The standard
recommendation is to pump 1 M NaOH for 30 minutes and then wash out
with buffer or distilled water.
WARNING! When using hazardous chemicals, make sure that the entire
system has been flushed thoroughly with bacteriostatic solution, e.g.
NaOH, and distilled water before service and maintenance.
• Wipe the surface regularly with a damp cloth. Do not allow spilt liquid to
dry on the instrument.
• Remove dirt from the surface using a cloth and a mild cleaning agent.
• Let the system dry completely before using it.
If, at any time, air is observed in the inlet tubing, purge the pump by opening
the appropriate purge valve and running the pump at max. flow rate for
about 1 minute. If there is a lot of air in the inlet tubing, stop the pump and
perform the procedure in section 2.8 Purging the pump.
Note: The power must be switched OFF when removing and refitting the
pump heads.
Note: Always replace the piston seals on both pump heads at the same
time. An even better practice is to replace all four piston seals.
Note: After a new seal has been installed, the pump should be run in, see
section 4.5 Running-in a new piston seal.
Note: Before disassembling the pump heads move all input buffers bottles
below the level of the pump heads to prevent siphoning.
1 Switch off the pump at the mains power switch on the back panel.
Remove the piston seal rinsing system. The connectors are simple plug-
in fittings.
3 If the pump is equipped with Switch Valves SV-903, first remove the switch
valves and disconnect the tubing. Unscrew the two white knurled screws
Outlet valve under the pump to release the support bracket locking the inlet manifold
into the inlet valve. Gently lower and remove the complete manifold.
4 Using the Allen key, unscrew and completely remove one of the two
Allen screws locking the pump head in position.
5 When unscrewing the second locking screw, push firmly on the front
face of the pump head to compensate for the pressure of the piston
return spring. Hold the pump head firmly to prevent it from twisting.
Remove the second screw and, without allowing the pump head to twist
sideways, carefully pull it out.
6 Place the pump head face down on the bench. Pull out the piston
together with the return spring.
7 Inspect the piston and return spring for sign of damage. If damaged,
they should be replaced.
8 Wipe the piston with a clean cloth. If salt solutions have been used
the piston may be slightly corroded. This corrosion can be removed
with a rubber eraser. If it cannot be wiped or rubbed clean, scrape
off any deposits with a scalpel or razor blade. Inspect the piston with
a magnifying glass for scratches. Replace with a new piston if any
scratches or cracks are found.
9 Remove the two screws securing the drain plate and the rinse chamber.
Remove and discard the rinsing membrane. Remove the rinse chamber.
For P-903, remove also the support washer.
Rinsing membrane
Drain plate
Rinse chamber
11 The pump head, rinse chamber and drain plate should be carefully
rinsed or cleaned in an ultrasonic bath, if available. If dirt can be seen on
any surfaces, the inlet and outlet check valves should be removed and
cleaned separately (see section 4.7)
12 Wet the new seal slightly and place it in the hole on the pump head and
press it down into position with a hard flat object For P-903, refit the
support washer on top of the new seal.
13 With the pump head still facing downwards on the bench, place the
rinse chamber onto the head with the rinse ports in line with the inlet
and outlet check-valves. The conical depression in the rinse chamber
should be facing upwards, ready to accept the new rinsing membrane.
Fit the rinsing membrane with the conical face downwards.
UP UP
Rinse membrane Rinse membrane
Support washer
Inlet Inlet
check valve Jointing ring check valve
Piston seal
Piston seal
Outlet Outlet
check valve check valve
P-901 P-903
14 Place the drain plate on top of the assembly. Use the two screws to lock
the complete assembly together.
Note: Align the drainage hole in the drainage plate with the inlet check
valve (the opposite side of the pump head marked OUT/UP).
Inlet check
valve
15 Wipe clean the piston and remove all finger prints. Wet the piston
UP
and then insert it into the return spring. With the pump head facing
downwards on the bench, insert the piston into the pump head by
pushing it gently but firmly vertically downwards into the seal.
CAUTION! Do not push the piston at an angle to the head and DO NOT
twist the piston.
16 Turn the head so that the inlet valve and drainage hole are facing
downwards and the text UP/OUT on the pump head is facing upwards.
Mount the complete pump head over the locating pins on the front
panel. Locate the metal end of the piston and the spring towards the
drive cam.
Hold the pump head firmly against the side panel of the housing with
one hand. Do not allow the assembly to twist under pressure from the
return spring. Using the Allen key, fit and tighten one of the Allen screws.
Fit and tighten the remaining Allen screw.
17 Reconnect the outlet tubing to the outlet check valve and the manifold
block as described above.
20 The pump should now be purged and the new piston seal carefully run-
in following the instructions in section 4.5.
1 Ensure that the reservoir is filled with sufficient eluent. Immerse the
inlet tubing in the eluent. The reservoir should be placed at least 30 cm
above the pump inlet.
3 Connect the male Luer connector at the other end of the purge tubing
to the left purge valve at pump module module A.
4 Turn the purge valve counterclockwise half a turn to open it and slowly
draw eluent to the syringe.
5 When fluid starts to enter the syringe continue to draw a few millilitres
before closing the purge valve. Check that there is no air left in the inlet
tubing.
If a damaged piston has been in operation, the piston seal will be destroyed
and should also be replaced. To replace the piston and the seal follow the
instructions in section 4.4 Replacing the piston seal.
In addition to the spare parts listed in section 4.4, the following are required
(see Reference information D for code no.):
Record the pressure (see section 5.4) and identify the faulty check valve by
observing which pump head is delivering the flow (see section B.1.1).
First try to clean the check valves in–place on the pump head by pumping
100% methanol for approximately 10 minutes. If this does not correct the
problem, follow the instructions for removing and then cleaning the valves.
Note: Change solvent to distilled water and flush out all salt before
removingthe check valves.
2 Use the 13 mm wrench to remove the valve from the pump head.
CAUTION! Handle the check valves with care when they have been
removed from the pump heads to prevent loss of any internal
components.
4 Refit the check valves. The inlet check valve (with a lip for the manifold
and a larger diameter opening) is fitted to the side marked IN of the
pump head. Tighten the valves until fully finger-tight and then use the
13 mm wrench to tighten a further 1/3rd (110o) of a turn. Do not over-
tighten the valves since damage to the internal components can occur.
6 Purge the pump carefully and check that the pumping action has been
corrected. See section 2.8.
5 Trouble shooting
5.1 General
When contacting GE Healthcare for support, state the program version of
the instrument, shown for 2 seconds during switch-on.
Fault Action
No text on the front display 1 Check that the mains cable is onnected and
the power switch is in ON-position I.
2 Check the mains power supply.
Erratic flow, noisy baseline
signal, irregular pressure trace
Air bubbles passing 1 Check that there is sufficient eluent present in
the through or trapped in the pump reservoirs.
2 Check all connections for leaks.
3 Follow the instructions in section 5.5 Removing
trapped air bubbles from the pump.
Inlet or outlet check valves 1 Follow the instructions in section 4.7 Cleaning
not functioning correctly the inlet and outlet check valves.
2 Clean the valves in-place. If improvement is not
seen, remove the valves and clean again.
Piston seal leaking 1 Replace the piston seal according to the
instructions in section 4.4 Replacing the piston
seal.
Blockage or part blockage 1 Flush through to clear blockage.
2 If necessary, replace tubing.
3 Check inlet tubing filter. It can become clogged
if unfiltered buffers or samples are applied.
See instructions for flushing through at the end
of the run in section 3.8 Ending the run and
storage.
Liquid leaking between the Piston seal or rinsing membrane incorrectly fitted
or worn pump head and the 1 Replace or re-install the seal or membrane.
side panel 2 Run-in carefully, see section 4.5 Running-in a
new piston seal.
Fault Action
Low eluent flow and noise 1 Disassemble pump head and examine the
as the pistons move piston spring according to section
4.4 Replacing the piston seal. Replace if
necessary.
2 If spring is corroded, check piston seal and
rinse membrane. Ensure that piston rinsing
system is always used when working with
aqueous buffers with high salt concentration.
3 Check the piston for damage. If damaged,
replace the piston according to section
4.6 Replacing a damaged piston.
4 Remember to replace the piston seal and rinse
membrane with new items.
Leaking connection and/or 1 Unscrew the connector and check if it is worn
crystalized material around or incorrectly fitted. If so replace the connector.
a connector 2 Gently tighten the connector with your fingers.
Error in external chart recorder 1 Check the chart recorder in accordance with its
manual.
Message Action
ERROR No mixer 1 Switch off the instrument.
Check connection! 2 Check the connections to the mixer.
ERROR in mixer. 3 Switch on the instrument.
Check mixer!
ERROR in valve x 1 Switch off the instrument.
Check connections! 2 Check the connections to the valve(s).
3 Switch on the instrument.
ERROR Pump too warm 1 Switch off the instrument to let it cool.
Check fans! 2 Clean or clear the front air inlet.
3 Switch on the instrument and check that the
fans are running.
ERROR Sync failure 1 Switch off the instrument.
Please restart! 2 Switch on the instrument. Remember to
open the purge valves before synchronization.
ERROR key (OK) 1 Switch off the instrument.
ERROR key (Esc) 2 Check all connections.
ERROR key (OK+Esc) 3 Switch on the instrument.
ERROR 100
ERROR 109-113
ERROR 120-121
ERROR 106-108 1 Switch off the instrument.
ERROR 118 2 Check all UniNet1 and UniNet 2 connections.
3 Switch on the instrument.
• leaking connections
• piston damaged
If the problem remains, try to remove the air bubbles in accordance with the
method described below. Use 100% methanol.
A Description
A.1 Instrument
Pump P-900 is a high performance laboratory pump for use in liquid
chromatography and other applications where accurately controlled liquid
flow is required.
An eluent in an external vessel is drawn into the inlet manifold by the action
of the pump. Twin reciprocating pump heads work in unison to deliver a
smooth, low-pulsation flow from the pump outlet manifold.
• P-901
• P-903
• The pumps are equipped with 2 pump modules, A and B, including two
pump heads each. This allows for binary gradients with high pressure
mixing.
• Low pulsation.
Valve 1
Valve A
Valve B
Pressure,
analog
out 0 - 1V UniNet 2 Remote UniNet 1 Mains
Connector/switch Function
Pressure analogue out 0-1V Pressure signal to chart recorder
UniNet 2 Connection to mixer M-925 and UniNet 2 network
Remote Input for remote on/off and output for pressure alarm
UniNet 1 Connection UniNet-1 network
Mains Supply voltage, grounded
0/1 Instrument on/off switch
Valve A, B Connection to switch valves. A and B used for
BufferPrep. When pump used for one pump
gradients as stand-alone, valve A should be used.
Valve 1 Connection for an additional switch valve for buffer
selection when used as stand–alone
The inlet to each pump head is fed from a common manifold. Solvent is
drawn up into the pump head through a non-return check valve by the
action of the piston being withdrawn from the pump chamber.
Outlet check
valves Outlet manifold Outlet, to mixer
Outlet check valve
Rinse chamber
inlet
U
Inlet manifold
Inlet Inlet check
valve
On the delivery stroke of the piston, the inlet valve is sealed by the pressure
developed and eluent is forced out through a similar check valve at the
outlet.
UP UP
Rinse membrane Rinse membrane
Support washer
Inlet Inlet
check valve Jointing ring check valve
Piston seal
Piston seal
Outlet Outlet
check valve check valve
P-901 P-903
The pistons are actuated by cams (eccentrics) driven by the motors. Force
for the retraction of the pistons is provided by coil springs. The length of
stroke of the pistons is fixed and changes in the flow rate are made by
varying the speed of the drive motor.
Leakage between the pump chamber and the drive mechanism is prevented
by a piston seal. The pistons and seals are continuously lubricated by the
presence of eluent. To prevent any deposition of salts from aqueous eluents
on the piston and to prolong the life of the seals, the low pressure chamber
behind the pump head can be flushed continuously with a low flow of
distilled water. The flexible rinsing membrane sweeps the piston and seals
the rinsing system.
Outlet tubing
Purge tubing
Purge valve
The output pressure is shown in the display together with the flow rate.
A continuous recording of the pressure may be made using the analog
output. For the protection of the system, a maximum and minimum
pressure limit can be set.
The external chart recorder output for pressure is 0–1 V. The full scale
range is set under sub menu Set Pr Full Scale under Setup, see Reference
information B.2.2.
B Menus
B.1 Check menu
B.1.1 Checking piston stroke
To enable trouble shooting it is possible to check which pump module head
that delivers flow.
Pump P-900
V1.00
B Reference information
Date of Maintenance
?
Buzzer Test
Setup Pr. Full Scale 1 Select sub menu Setup Pr. Full Scale, press OK.
Setup Pr. Full Scale
Setup Pr. Full Scale 2 Set the value, press OK.
Setup Pr. Full Scale
B.2.3 Set pressure unit
Sets the pressure unit used in the display.
Setup Pressure Unit
Setup Pressure Unit 1 Select sub menu Setup Pressure Unit, press OK.
Setup Pressure Unit
Note: Selecting one pump gradient mode disables the B pump module.
Gradient mixing is not possible in double mode.
Setup Pump-B Type 3 Select sub menu Setup Pump-B Type, press OK.
Setup Compr. comp 2 Select sub menu Setup Compr. comp, press OK.
(off) off on
3 Select On to activate, or Off to deactivate compensation. Press OK.
Setup Comp A
(4.0E-10) 4.0 4 Enter a compensation value between 4.0 and 15.0 for pump module A.
Setup Comp B 5 Enter a compensation value between 4.0 and 15.0 for pump module B.
(4.0E-10) 4.0
IN IN
NC NO NC NO
GB = British English
D = German
F = French
E = Spanish
I = Italian
3 Select viewing angle (->\ Up, ->| Mid or ->/ Down), press OK.
2 Set the action to take place. Press OK to select action. Buzzer will
Alarm/Timer action generate an audible alarm for 15 s and a message. Run will start the
(Bzz) Buzzer Run End
pump at the set flow rate, End will stop the pump, each generating one
beep and a message.
Set Alarm 12:32:21
(0) 00.00.00 3 Use the sub menu Set Alarm, if you want to set an alarm at a fixed time.
Press OK to enter the time value in the form HH.MM.SS, pressing the OK
button after entering each time unit.
Set Timer
(0) 00.00.00 4 If you want to set a count-down timer, turn the dial to select sub menu
Set Timer. Press OK to enter the countdown value in the form HH.MM.SS,
pressing the OK button after entering each time unit.
Alarm/Timer 12:30:52
(Bzz12:33:00)
5 Press ESC button to return to the Alarm/Timer menu which now shows
the set alarm time or count-down time as BzzHH:MM:SS.
Bzz12:41:29 12:41:49 6 When the alarm time is due or the count-down timer reaches 00:00:00,
!! Alarm time !!
an alert display is shown and the instrument beeps, until the OK button is
pressed.
Set Clock
(00:36:53) 00:36:53
Bzz12:41:29 12:41:49
!! Alarm time !!
Set Clock
The alarm timer is based on the internal instrument clock which can be set
(00:36:53) 00:36:53 in the Set Clock menu placed after the Alarm/Timer menu. The clock will be
reset when power is turned OFF.
Alarm/Timer off? An already set alarm/timer function can be reset by pressing OK in the
(Buzz 05:15:00)
menu Alarm/Timer off?
B. 4 Service displays
The instrument has service displays for use by authorised service personnel.
Enter Access Code!
If the service display Enter Access Code: is accidentally selected, press the
ESC-button to exit to the normal operation display.
Pause 1.00ml/min
C Technical specifications
C.1 Operating data
The specifications are valid only at pressures higher than 0.2 MPa.
12 Short instructions
The following short instructions are intended as a guide to users who are
fully familiar with safety precautions and operating instructions described
in this manual. The instructions assumes that the instrument is installed
according to the installation instructions.
P-9
00
Selftest
Pump synchronization
End 0.00ml/min
0.00MPa 0.0%B
1 Switch on the instrument by means of the mains switch on the
rear panel.
Set Flow Rate
2 Press OK to start synchronisation.
Run 2.50ml/min
All goods and services are sold subject to the terms and conditions of sale
GE Healthcare Bio-sciences AB of the company within GE Healthcare which supplies them. GE Healthcare
Björkgatan 30 reserves the right, subject to any regulatory and contractual approval, if
required, to make changes in specifications and features shown herein, or
751 84 Uppsala discontinue the product described at any time without notice or obligation.
Contact your local GE Healthcare representative for the most current
Sweeden information.
GE Healthcare Bio-Sciences AB
Björkgatan 30, 751 84 Uppsala, Sweden
GE Healthcare UK Ltd
Amersham Place, Little Chalfont, Buckinghamhire, HP7 9NA, UK
GE Healthcare Bio-Sciences KK
Sanken Bldg. 3-25-1, Huakunincho, Shinjuku-ku, Tokyo 169-0073, Japan
Asia Paci.c Tel: +852 2811 8693 Fax: +852 2811 5251 • Australasia Tel: + 61 2 9899 0999 Fax: +61 2 9899 7511 • Austria Tel: 01/57606-1619 Fax: 01/57606-1627 • Belgium Tel: 0800 73 888
Fax: 03 272 1637 • Canada Tel: 800 463 5800 Fax: 800 567 1008 • Central, East, & South East Europe Tel: +43 1 982 3826 Fax: +43 1 985 8327 • Denmark Tel: 45 16 2400 Fax: 45 16 2424 • Finland &
Baltics Tel: +358-(0)9-512 39 40 Fax: +358 (0)9 512 39 439 • France Tel: 01 69 35 67 00 Fax: 01 69 41 96 77 • Germany Tel: 0761/4903-490 Fax: 0761/4903-405 • Italy Tel: 02 27322 1 Fax: 02 27302
212 Japan Tel: +81 3 5331 9336 Fax: +81 3 5331 9370 • Latin America Tel: +55 11 3933 7300 Fax: +55 11 3933 7304 • Middle East & Africa Tel: +30 210 9600 687 Fax: +30 210 9600 693 • Netherlands
Tel: 0165 580 410 Fax: 0165 580 401 • Norway Tel: 815 65 555 Fax: 815 65 666 • Portugal Tel: 21 417 7035 Fax: 21 417 3184 • Russia & other C.I.S. & N.I.S Tel: +7 (095) 232 0250, 956 1137
Fax: +7 (095) 230 6377 • South East Asia Tel: 60 3 8024 2080 Fax: 60 3 8024 2090 • Spain Tel: 93 594 49 50 Fax: 93 594 49 55 • Sweden Tel: 018 612 1900 Fax: 018 612 1910 • Switzerland
Tel: 0848 8028 12 Fax: 0848 8028 13 • UK Tel: 0800 616928 Fax: 0800 616927 • USA Tel: 800 526 3593 Fax: 877 295 8102
imagination at work
User Manual 18-1120-04 AE