SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

TECHNOLOGY & METROLOGY TEGMETRO S.A.


Juan Domingo N65-85 y Manuel Guizado
Quito, Ecuador 170301
Henry Arizaga Phone: +593-02-6007779

CALIBRATION

Valid To: December 31, 2022 Certificate Number: 4150.01

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 5:

I. Acoustics & Vibration

Parameter/Range Frequency CMC2, 6, 8 () Comments

Sound Level – Measuring


Equipment

(74 to 84) dB (125 to 1000) Hz 0.51 dB Comparison using


(84 to 94) dB 0.51 dB portable reference
(94 to 104) dB 0.51 dB calibrator sound
(74 to 114) dB 0.51 dB level
(104 to 114) dB 0.24 dB

Sound Level3 – Measuring


Equipment

(74 to 84) dB (1 to 4) kHz 0.59 dB Comparison using


(84 to 94) dB 0.59 dB portable reference
(94 to 104) dB 0.59 dB calibrator sound
(74 to 114) dB 0.59 dB level
(104 to 114) dB 0.35 dB

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 1 of 21


II. Chemical

Parameter/Equipment Range CMC2 () Comments

Conductivity Meters3 1 µS/cm 0.3 µS/cm Conductivity


10 µS/cm 0.6 µS/cm standard solution

Conductivity Meters3 100 µS/cm 2 µS/cm Conductivity


1000 µS/cm 4.6 µS/cm standard solution
1413 µS/cm 4.6 µS/cm

pH Meters3 4 pH Unit 0.011 pH Unit pH buffer solution


7 pH Unit 0.011 pH Unit
10 pH Unit 0.011 pH Unit

III. Dimensional Testing

Parameter/Equipment Range CMC2 () Comments

One Dimensional8 – Up to 25 mm 0.000 82 mm Vision system with


Length Up to 1 in 32 µin microscope

Up to 25 mm 0.000 45 mm
Up to 1 in 180 µin Precision micrometer

Up to 25 mm 0.0015 mm Indicator
Up to 1 in 59 µin

Up to 150 mm 0.0019 mm Micrometer


Up to 6 in 75 µin

Up to 200 mm 0.011 mm Caliper


Up to 8 in 0.000 43 in

Up to 450 mm 0.011 mm Height gage


Up to 17.7 in 0.000 43 in

Up to 5000 mm 0.61 mm Tape master


Up to 197 in 0.024 in

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 2 of 21


IV. Dimensional

Parameter/Equipment Range CMC2, 6 () Comments

Calipers – Outside, Inside,


& Depth

Digital Up to 600 mm 8.2 µm Gage blocks


(Up to 24) in 0.000 32 in
Analog
Up to 600 mm 16 µm
(Up to 24) in 0.000 63 in

Outside, Inside, & Depth Up to 25 mm 0.35 µm Gage blocks


Micrometers – Digital & (25 to 600) mm (0.012L - 0.052) µm
Analog
Up to 1 in 14 µin
(1 to 24) in (12L - 2.0) µin

Three Contact Inside (12.7 to 114.3) mm 0.9 µm Gage rings


Micrometers & Bore Gages (0.5 to 4.5) in 35 µin
– Digital & Analog
127 mm 2.2 µm
5 in 86 µin

(139.7 to 152.4) mm 1.0 µm


(5.5 to 6) in 38 µin

Mechanical Thickness Up to 50 mm 0.8 µm Gage blocks


Gages Up to 2 in 32µin

Indicators Up to 25 mm 1.2 µm Gage blocks


Up to 1 in 47 µin

Analog & Digital Weld Up to 50 mm 20 µm Gage blocks


Gages
Up to 2.7 in 0.000 79 in Caliper

Oil Gauging Tape Up to 30 m 200 µm Master tape measure


Measures

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 3 of 21


Parameter/Equipment Range CMC2, 6 () Comments

Tape Measures Up to 30 m 210 µm Master tape measure


(30 to 40) m 300 µm
(40 to 50) m 420 µm
(50 to 60) m 520 µm
(60 to 70) m 600 µm
(70 to 80) m 670 µm
(80 to 90) m 730 µm
(90 to 100) m 790 µm

Pi Tape (10 to 150) mm 0.019 mm


(0.39 to 5.9) in 0.000 75 in Ring gages

(150 to 3600) mm 0.12 mm Tape master


(5.9 to 141.7) in 0.0047 in

Multi-Step Thickness Up to 25 mm 1.1 µm ASTM E797/E797M:


Gage Blocks Up to 1 in 40 µin gage blocks &
micrometer as
comparator

Ultrasonic Thickness (2 to 25) mm 0.018 mm Multi-step thickness


Gages3 (0.0079 to 1) in 0.000 71 in gage blocks

Ultrasonic Flaw Detectors3 (25.4 to 100) mm 0.013 mm IIW Type 1 reference


(1 to 4) in 0.000 51 in block

(100 to 250) mm 0.015 mm


(4 to 9.8) in 0.000 59 in

Coating Thickness Gage3 25.1 µm 1 mils 0.38 µm Coating thickness


53.4 µm 2.1 mils 15 µin standard
128.2 µm 5 mils
260 µm 10.2 mils
261.6 µm 10.3 mils
538.5 µm 21.2 mils
770.2 µm 30.3 mils
1580.3 μm 62.2 mils

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 4 of 21


Parameter/Equipment Range CMC2, 6 () Comments

Rulers Up to 1 m 0.17 mm Master tape measure

Reference Blocks for Up to 450 mm 15 µm ASTM E164:


Calibration of Ultrasonic Up to 17.7 in 610 µin calipers
Testing Instruments

Feeler/Thickness Gages3 Up to 2 mm 0.76 µm Precision micrometer


Up to 0.08 in 30 µin

Force Testing Machine (0.5 to 25) mm 2 μm Digital indicator


Crosshead Travel3
(25 to 150) mm 8.4 μm Digital caliper

V. Electrical – DC/Low Frequency

Parameter/Equipment Range CMC2, 7 () Comments

DC Voltage3 – Generate (0 to 100) mV 33 µV/V + 10 µV Transmille 1000A


(0.1 to 1) V 62 µV/V + 26 µV
(1 to 10) V 0.062 mV/V + 0.24 mV
(10 to 100) V 0.062 mV/V + 2.4 mV
(100 to 1000) V 0.062 mV/V + 24 mV

DC Voltage3 – Measure Up to 100 mV 10 µV/V + 0.5 µV HP 3458A


(0.1 to 1) V 14 µV/V + 0.26 µV
(1 to 10) V 14 µV/V + 1.4 µV
(10 to 100) V 14 µV/V + 43 µV
(100 to 1000) V 14 µV/V + 470 µV

DC Current3 – Generate (0 to 100) μA 0.15 nA/μA + 27 nA Transmille 1000A


(0.1 to 1) mA 0.25 μA/mA + 0.063 μA
(1 to 10) mA 0.25 μA/mA + 0.63 μA
(10 to 100) mA 0.25 μA/mA + 6.3 μA
(0.1 to 1) A 0.22 mA/A + 0.11 mA
(1 to 10) A 0.44 mA/A + 1.4 mA

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 5 of 21


Parameter/Equipment Range CMC2, 7 () Comments

DC Current3 – Measure (10 to 100) μA 19 μA/A + 0.0009 μA HP 3458A


100 μA to 1 mA 21 μA/A + 0.0052 μA
(1 to 10) mA 21 μA/A + 0.05 μA
(10 to 100) mA 32 μA/A + 1.1 μA
100 mA to 1 A 110 μA/A + 2.5 μA

(1 to 3) A 0.94 mA/A + 0.24 mA Fluke 8845A


(3 to 10) A 1.4 mA/A + 1.7 mA

DC Current3 – Clamp-On Up to 50 A 0.8 A Transmille 1000A &


Meters (>50 to 100) A 1.1 A EA002 turn coil
(>100 to 200) A 1.8 A adapter
(>200 to 300) A 2.4 A
(>300 to 400) A 3.1 A
(>400 to 500) A 3.7 A

Resistance3 – Generate (0 to 10) Ω 0.11 mΩ/Ω + 1.4 mΩ Transmille 1000A


(10 to 100) Ω 0.11 mΩ/Ω + 14 mΩ
(0.1 to 1) kΩ 0.11 Ω/kΩ + 0.14 Ω
(1 to 10) kΩ 0.11 Ω/kΩ + 1.4 Ω
(10 to 100) kΩ 0.11 Ω/kΩ + 14 Ω
(0.1 to 1) MΩ 0.11 kΩ/MΩ + 0.14 kΩ
(1 to 10) MΩ 0.42 kΩ/MΩ + 1.0 kΩ

Resistance3 – Measure (0 to 10) Ω 0.014 mΩ/Ω + 0.09 mΩ HP 3458A


(10 to 100) Ω 0.014 mΩ/Ω + 0.8 mΩ
(0.1 to 1) kΩ 0.013 mΩ/Ω + 2.6 mΩ
(1 to 10) kΩ 0.013 mΩ/Ω + 20 mΩ
(10 to 100) kΩ 14 mΩ/kΩ + 150 mΩ
(0.1 to 1) MΩ 18 Ω/MΩ + 2.9 Ω
(1 to 10) MΩ 0.07 kΩ/MΩ + 0.1 kΩ

(1 to 100) MΩ 7.8 kΩ/MΩ + 24 kΩ Fluke 8845A

Parameter/Range Frequency CMC2, 7 () Comments

AC Voltage3 – Generate

60 Hz & 1 kHz (0.1 to 1) V 0.64 mV/V + 0.21 mV Transmille 1000A


(1 to 10) V 0.60 mV/V + 2.6 mV
(10 to 100) V 0.69 mV/V + 21 mV
(0.1 to 1) kV 0.60 V/kV + 0.42 V

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 6 of 21


Parameter/Range Frequency CMC2, 7 () Comments

AC Voltage3 – Measure

1 kHz (0.1 to 1) V 58 μV/V + 18 μV HP 3458A

20 Hz to 1 kHz (1 to 10) V 0.03 mV/V + 1 mV


(1 to 20) kHz 0.1 mV/V + 0.6 mV
(20 to 100) kHz 0.8 mV/V + 0.2 mV
100 kHz to 1 MHz 9.4 mV/V + 6.4 mV

20 Hz to 1 kHz (10 to 100) V 0.18 mV/V + 3 mV


(100 to 700) V 0.21 mV/V + 250 mV

AC Current3 – Generate

60 Hz & 1 kHz (0 to 0.1) mA 0.17 μA/mA + 0.40 µA Transmille 1000A


(0.1 to 1) mA 0.081 μA/mA+ 0.79 µA
(1 to 10) mA 0.040 μA/mA+ 8.0 µA
(10 to 100) mA 0.57 μA/mA+ 73 µA
(0.1 to 1) A 0.57 mA/A+ 0.73 mA
(1 to 10) A 0.39 mA/A+ 15 mA

AC Current3 – Measure

1 kHz Up to 10 mA 1.1 μA/mA + 5.6 μA Fluke 8845A


(10 to 100) mA 0.7 μA/mA + 47 μA
(100 to 400) mA 0.45 μA/mA + 530 μA
400 mA to 1 A 720 μA/A + 480 μA
(1 to 3) A 0.91 mA/A + 3.6 mA
(3 to 10) A 0.88 mA/A + 11 mA

AC Current3 – Clamp-On
Meters

56 Hz Up to 50 A 0.8 A Transmille 1000A &


(>50 to 100) A 1.1 A EA002 turn coil
(>100 to 200) A 1.8 A adapter
(>200 to 300) A 2.4 A
(>300 to 400) A 3.1 A
(>400 to 500) A 3.7 A

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 7 of 21


Parameter/Frequency Range CMC2, 7 () Comments

Insulation Resistance3 –

Fixed Points 100 kΩ 0.12 kΩ Decade resistance


200 kΩ 0.35 kΩ
300 kΩ 0.35 kΩ
400 kΩ 0.58 kΩ
500 kΩ 0.81 kΩ
600 kΩ 0.81 kΩ
700 kΩ 1 kΩ
800 kΩ 1 kΩ
900 kΩ 1 kΩ
1 MΩ 0.0024 MΩ
5 MΩ 0.0058 MΩ
10 MΩ 0.13 MΩ
20 MΩ 0.04 MΩ
30 MΩ 0.04 MΩ
40 MΩ 0.072 MΩ
50 MΩ 0.072 MΩ
60 MΩ 0.092 MΩ
70 MΩ 0.092 MΩ
80 MΩ 0.13 MΩ
90 MΩ 0.13 MΩ
100 MΩ 0.25 MΩ
200 MΩ 0.48 MΩ
300 MΩ 0.72 MΩ
400 MΩ 0.96 MΩ
500 MΩ 1.2 MΩ
600 MΩ 1.7 MΩ
700 MΩ 1.7 MΩ
800 MΩ 2.1 MΩ
900 MΩ 2.1 MΩ
1 GΩ 0.0091 GΩ
5 GΩ 0.062 GΩ
10 GΩ 0.12 GΩ
20 GΩ 0.26 GΩ
30 GΩ 0.38 GΩ
40 GΩ 0.63 GΩ
50 GΩ 0.63 GΩ
60 GΩ 0.88 GΩ
70 GΩ 0.88 GΩ
80 GΩ 1.3 GΩ
90 GΩ 1.3 GΩ
100 GΩ 1.3 GΩ

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 8 of 21


VI. Fluid Quantities

Parameter/Equipment Range CMC2 () Comments

Hydrometry – Cuckow method

Hydrometers (600 a 1000) kg/m³ 0.12 kg/m³


(1000 a 1500) kg/m³
(1500 a 2000) kg/m³

API Hydrometers and (-1 a 101) API 0.011 API


Thermo-hydrometers
(API Scale for
Petroleum Products)

Volume – Laboratory 0.5 mL 0.0017 mL Gravimetric method


Glassware Single 1 mL 0.0027 mL using analytical balance
Volume Pipettes 2 mL 0.0033 mL per TM-CR-03 based ISO
5 mL 0.005 mL 8655-6, ISO 4787, NIST
10 mL 0.0066 mL SOP 14
20 mL 0.01 mL
25 mL 0.01 mL
50 mL 0.017 mL
100 mL 0.027 mL

One-Mark Volumetric 1 mL 0.0082 mL Gravimetric method


Flasks 2 mL 0.0082 mL using analytical balance
5 mL 0.0082 mL per TM-CR-03 based ISO
10 mL 0.0082 mL 8655-6, ISO 4787, NIST
20 mL 0.013 mL SOP 14
25 mL 0.013 mL
50 mL 0.02 mL
100 mL 0.033 mL
200 mL 0.05 mL
250 mL 0.05 mL
500 mL 0.082 mL
1000 mL 0.13 mL
5000 mL 0.4 mL

Graduated Pipettes 0.1 mL 0.002 mL Gravimetric method


0.2 mL 0.002 mL using analytical balance
0.5 mL 0.002 mL per TM-CR-03 based ISO
1 mL 0.0023 mL 8655-6, ISO 4787, NIST
2 mL 0.0033 mL SOP 14
5 mL 0.0058 mL
10 mL 0.0091 mL
20 mL 0.033 mL
25 mL 0.033 mL

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 9 of 21


Parameter/Equipment Range CMC2 () Comments

Graduated Measuring 5 mL 0.011 mL Gravimetric method


Cylinders 10 mL 0.013 mL using analytical balance
25 mL 0.082 mL per TM-CR-03 based ISO
50 mL 0.14 mL 8655-6, ISO 4787, NIST
100 mL 0.2 mL SOP 14
250 mL 0.36 mL
500 mL 0.57 mL
1000 mL 0.96 mL
2000 mL 1.5 mL

Plastic Graduated 10 mL 0.033 mL Gravimetric method


Measuring Cylinders 25 mL 0.11 mL using analytical balance
50 mL 0.19 mL per TM-CR-03 based ISO
100 mL 0.26 mL 8655-6, ISO 4787, NIST
250 mL 0.47 mL SOP 14
500 mL 0.7 mL
1000 mL 1.1 mL
2000 mL 1.9 mL
4000 mL 7.3 mL

Burettes 1 mL 0.0020 mL Gravimetric method


2 mL 0.0033 mL using analytical balance
5 mL 0.0033 mL per TM-CR-03 based ISO
10 mL 0.0066 mL 8655-6, ISO 4787, NIST
25 mL 0.010 mL SOP 14
50 mL 0.017 mL
100 mL 0.033 mL

Pyknometers 1 mL 0.0023 mL Gravimetric method


2 mL 0.0023 mL using analytical balance
5 mL 0.0023 mL per TM-CR-03 based ISO
10 mL 0.0023 mL 8655-6, ISO 4787, NIST
25 mL 0.005 mL SOP 14
50 mL 0.010 mL
100 mL 0.010 mL

Centrifuge Tube 0.1 mL 0.0066 mL Gravimetric method


0.3 mL 0.010 mL using analytical balance
0.5 mL 0.017 mL per TM-CR-03 based ISO
1 mL 0.017 mL 8655-6, ISO 4787, NIST
2 mL 0.033 mL SOP 14
3 mL 0.033 mL
5 mL 0.066 mL

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 10 of 21


Parameter/Equipment Range CMC2 () Comments

Centrifuge Tube (cont) 10 mL 0.17 mL Gravimetric method


25 mL 0.34 mL using analytical balance
100 mL 0.57 mL per TM-CR-03 based ISO
8655-6, ISO 4787, NIST
SOP 14

Imhoff Cone (0 to 2) mL 0.033 mL Gravimetric method


(>2 to 10) mL 0.17 mL using analytical balance
(>10 to 40) mL 0.34 mL per TM-CR-03 based ISO
(> 40 to 100) mL 0.66 mL 8655-6, ISO 4787, NIST
(>100 to 1000) mL 3.3 mL SOP 14

Water Trap 2 mL 0.0082 mL Gravimetric method


5 mL 0.017 mL using analytical balance
10 mL 0.033 mL per TM-CR-03 based ISO
25 mL 0.033 mL 8655-6, ISO 4787, NIST
SOP 14

Piston Operated
Volumetric Apparatus –

Piston Pipettes (>10 to 20) µL 0.031 µL Gravimetric method


50 µL 0.042 µL using analytical balance
100 µL 0.094 µL per TM-CR-03 based ISO
200 µL 0.34 µL 8655-6, ISO 4787, NIST
500 µL 1.3 µL SOP 14
1000 µL 2.5 µL
2000 µL 5.7 µL
5000 µL 14 µL
10 000 µL 21 µL

Piston Operated
Volumetric Apparatus –

Piston Burettes 1 mL 0.0010 mL Gravimetric method


2 mL 0.0020 mL using analytical balance
5 mL 0.0050 mL per TM-CR-03 based ISO
10 mL 0.0070 mL 8655-6, ISO 4787, NIST
20 mL 0.014 mL SOP 14
25 mL 0.018 mL
50 mL 0.025 mL
100 mL 0.030 mL

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 11 of 21


Parameter/Equipment Range CMC2 () Comments

Piston Operated
Volumetric Apparatus –
(cont)

Dispensers (>0.01 to 0.02) mL 0.000 034 mL Gravimetric method


0.02 mL 0.000 034 mL using analytical balance
0.05 mL 0.000 042 mL per TM-CR-03 based ISO
0.1 mL 0.000 095 mL 8655-6, ISO 4787, NIST
0.2 mL 0.000 60 mL SOP 14
0.5 mL 0.0012 mL
1 mL 0.0029 mL
2 mL 0.0058 mL
5 mL 0.012 mL
10 mL 0.029 mL
25 mL 0.058 mL
50 mL 0.12 mL
100 mL 0.29 mL
200 mL 0.29 mL

VII. Magnetic Quantities

Parameter/Equipment Range CMC2 () Comments

Magnetometer³ (1 to 80) Gauss 0.85 Gauss Gaussmeter Master


Gaussmeter 0.89 Gauss FW BELL 5170
Positive and Negative 0.96 Gauss
Polarity 1.1 Gauss
1.7 Gauss
2.3 Gauss

VIII. Mechanical

Parameter/Equipment Range CMC2 () Comments

Vibration3 – Measuring Back-to-back comparison


Equipment using portable reference
calibrator (shaker and
Peak Acceleration 0.4 g pk (3.92 m/s2) 0.030 g pk (0.29 m/s²) transducer)
@ 7 Hz

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 12 of 21


Parameter/Equipment Range CMC2 () Comments

Vibration3 – Measuring Back-to-back comparison


Equipment (cont) using portable reference
calibrator (shaker and
Peak Acceleration 0.8 g pk (7.84 m/s2) 0.038 g pk (0.37 m/s²) transducer)
@ 10 Hz

1 g pk (9.81 m/s2) 0.039 g pk (0.38 m/s²)


@ 30 Hz to 2 kHz

1 g pk (9.81 m/s2) 0.054 g pk (0.53 m/s²)


(>2 to 10) kHz

Pressure Gages,
Calibrators3 –

Pneumatic (-9 to 0) psig 0.0035 psig Precision pressure


(-62.1 to 0) kPa 0.024 kPa indicator mensor CPG
2500 & vacuum
transducer mensor CPT
6020

( -5 to 5) mbar 0.0059 mbar Digital Pressure Gauge


0.59 Pa Additel ADT681-05-DP2
BAR

(0 to 15) psig 0.0018 psig Precision pressure


(0 to 0.103) kPa 0.013 kPa indicator mensor CPG
2500 & pressure
(15 to 87) psig 0.011 psig transducer mensor CPR
(0.103 to 0.60) kPa 0.077 kPa 2550

Hydraulic (87 to 1015.26) psi 0.007 % of reading Oil operated deadweight


(0.6 to 7.0) MPa tester, DH-budenberg
CPB 5800 (low pressure
piston)
(1015.26 to 10 152.0) 0.0063 % of reading
psig Oil operated dead-weight
(7 to 70) MPa tester, DH-budenberg
CPB 5800 (high pressure
piston)

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 13 of 21


Parameter/Equipment Range CMC2 () Comments

Pressure Transmiters3 –

Pneumatic (-9 to 0) psig 0.0053 psig Precision pressure


(-62.1 to 0) kPa 0.036 kPa indicator mensor CPG
2500; vacuum transducer
mensor CPT 6020 &
amperemeter 6 ½ digits
Fluke 8845A

( -5 to 5) mbar 0.0059 mbar Digital Pressure Gauge


0.59 Pa Additel ADT681-05-DP2
BAR amperemeter 6 ½
digits Fluke 8845A

(0 to 15) psig 0.0062 psig Precision pressure


(0 to 103.4) kPa 0.043 kPa indicator mensor CPG
2500; pressure transducer
mensor CPR 2550 &
amperemeter 6½ digits
Fluke 8845A

(15 to 87) psig 0.037 psig Oil operated dead-weight


(103.4 to 599.8) kPa 0.26 kPa tester DH-budenberg CPB
5800 (low pressure piston)
& amperemeter 6½ digits
fluke 8845A

Hydraulic (87 to 1015.26) psig 0.041 % of reading Oil operated dead-weight


(599.8 to 7000) kPa tester DH-budenberg CPB
5800 (low pressure piston)
(1015.26 to 10 152) psig 0.041 % of reading & amperemeter 6½ digits
(7 to 70) MPa fluke 8845A

Pressure Safety Valve (-9 to -1) psi 0.4 % of reading Vacuum & pressure
(PSV´s) & Pressure (-62.1 to -1) kPa indicator
Switch3
(1 to 10 000) psi 0.1 % of reading
(1 to 68.95) MPa

Scales & Balances3 (10 to 100) mg 0.013 mg OIML Class E2 weights


100 mg to 100 g 0.02 mg
100 mg to 200 g 0.082 mg

100 mg to 2100 g 0.91 mg OIML Class F1 weights


(1 to 10 000) g 9.1 mg

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 14 of 21


Parameter/Equipment Range CMC2 () Comments

Scales & Balances3 (10 to 32 000) g 91 mg OIML Class M1 weights


(cont) (50 to 61 000) g 0.9 g
(0.5 to 160) kg 9.1 g
(2 to 200) kg 18 g
(2 to 500) kg 45 g

(2 to 1000) kg 83 g

(2 to 2000) kg 0.17 kg

Mass –

OIML Classes F2, 1 mg 0.0091 mg Mass comparator & ASTM


M1, M2 & M3 2 mg 0.0091 mg Class F1 weights
5 mg 0.0091 mg
10 mg 0.0091 mg
20 mg 0.0091 mg
50 mg 0.0091 mg

OIML Classes F1, 0.1 g 0.0093 mg Mass comparator & ASTM


F2, M1, M2 & M3 0.2 g 0.0093 mg Class E2 weights
0.5 g 0.0094 mg
1g 0.011 mg
2g 0.012 mg
5g 0.013 mg
10 g 0.018 mg
20 g 0.028 mg
50 g 0.063 mg
100 g 0.15 mg
200 g 0.25 mg
500 g 0.6 mg
1000 g 1.2 mg

OIML Classes F2, 2 kg 9.6 mg Mass comparator & ASTM


M1, M2 & M3 Class F1 weights

OIML Classes M1, 5 kg 91 mg


M2 & M3 10 kg 110 mg
20 kg 310 mg

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 15 of 21


Parameter/Equipment Range CMC2 () Comments

Torque Tools & (1 to 10) Nꞏm (0.0005ꞏreading + 0.0038) Nꞏm AWS TT-QC-100I-
Torque Wrenches 1/4 torque transducer
and display

(10 to 100) Nꞏm (0.0028ꞏreading + 0.12) Nꞏm AWS ITI-1000


torque transducer &
display

(100 to 1000) Nꞏm (0.0039ꞏreading + 0.37) Nꞏm AWS ITS-750 torque


transducer & display

Torque Transducers³ Up to 1,13 Nꞏm (8.92 × 10-5X + 4.97 × 10-5) Nꞏm Comparison with
Torque Indicator³ torque arm and
(1.13 to 11.3) Nꞏm (5.27 × 10-4X – 1.65 × 10-4) Nꞏm OIML class F1, F2 &
M1 weights
(11.3 to 113) Nꞏm (6.62 × 10-4X – 1.21× 10-2) Nꞏm
X = reading value
(113 to 1016) Nꞏm (1.76 × 10-3X – 2.31 × 10-1) Nꞏm

Force transducers (0.044 to 1) kN (6.0 × 10-4X + 5.4 × 10-6) kN Comparison with


with indicator³ – reference Precision
load cells
Tension & (1 to 50) kN (1.0 × 10-3X + 3.3 × 10-3) kN
Compression X = reading value

(50 to 500) kN (9.0 × 10-4X + 1.8 × 10-3) kN

Force indicators³ – (44 to 1000) N 0.13 % of reading Comparison with


Tension & reference Precision
Compression (>1 to 5) kN 0.19 % of reading load cells

(>5 to 50) kN 0.19 % of reading

(>50 to 500) kN 0.12 % of reading

Force – Tension & (44 to 1000) N 0.12 % of reading Precision load cells
Compression Testing
Machines3 (>1 to 5) kN 0.14 % of reading

(>5 to 50) kN 0.13 % of reading

(>50 to 500) kN 0.12 % of reading

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 16 of 21


Parameter/Equipment Range CMC2 () Comments

Force transducers³ – (0.044 to 1) kN (5.61 × 10-4X + 5.41 × 10-6) kN Comparison with


Tension & reference precision
Compression (1 to 50) kN (9.8 × 10-4X + 3.0 × 10-3) kN load cells

(50 to 500) kN (8.02 × 10-4X + 1.94 × 10-3) kN X = reading value

Force indicators³ – (44 to 1000) N 0.12 % of reading Comparison with


Tension & reference precision
Compression (>1 to 5) kN 0.14 % of reading load cells

(>5 to 50) kN 0.13 % of reading

(>50 to 500) kN 0.12 % of reading

Force Testing (10 to 840) mm/min 0.12 % of reading Stopwatches,


Machine Crosshead indicator & caliper
Speed3

IX. Optical Quantities

Parameter/Equipment Range CMC2 () Comments

Light Meters³ –

Light Visible (20 to 3000) lux (8.6 × 10-5X + 1.0) % of Reference standard
(Luxmeter) reading lamp

X = reading value

Light UVA (2162 to 4148.33) µW/cm² 2.9 % of reading Radiometer master


(Radiometer)

Spectrophotometers3 –

Absorbance:
λ: (235 to 350) nm (0.09 to 0.28) A 0.0037 A Potassium
(0.19 to 0.57) A 0.0045 A Dichromate
(0.29 to 0.87) A 0.0049 A quartz cells
(0.39 to 1.15) A 0.0057 A
(0.48 to 1.45) A 0.0067 A

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 17 of 21


Parameter/Equipment Range CMC2 () Comments

Spectrophotometers3 –
(cont)

Absorbance:
Neutral
λ: (440 to 635) nm (0.52 to 0.56) A 0.0027 A density glass
(0.66 to 0.77) A 0.0027 A filters
(0.95 to 1.11) A 0.0027 A

Transmittance:

λ: (440 to 635) nm (7.71 to 11.22) % 0.047 % of transmittance


(16.81 to 21.71) % 0.10 % of transmittance
(27.44 to 30.34) % 0.17 % of transmittance

Wavelength:
Holmium glass
SBW: (0.1, 0.25, 0.5, (241 to 638) nm 0.12 nm filter
1, 1.5, 2, 3, 4, 5) nm
Didymium glass
SBW: (0.1, 0.25, 0.5, (440 to 880) nm 0.12 nm filter
1, 1.5, 2, 3, 4, 5) nm

X. Thermodynamics

Parameter/Equipment Range CMC2, 4 () Comments

Digital Thermometers3 (-30 to -20) °C 0.038 °C Digital indicator whit


(-20 to 0) °C 0.035 °C thermocouple type “R/S”
(0 to 140) °C 0.042 °C and temperature bath
(140 to 231) °C 0.059 °C
(231 to 420) °C 0.037 °C
(420 to 450) °C 1.3 °C
(450 to 700) °C 1.4 °C
(700 to 950) °C 1.4 °C
(950 to 1200) °C 1.8 °C

Mechanical Thermometers3 (-30 to 420) °C 0.15 °C Digital thermometer &


temperature bath

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 18 of 21


Parameter/Equipment Range CMC2, 4 () Comments

Liquid-In-Glass (-30 to 100) °C 0.07 °C Digital thermometer &


Thermometers3 (100 to 150) °C 0.30 °C temperature bath
(150 to 200) °C 0.40 °C
(200 to 300) °C 0.80 °C
(300 to 420) °C 1.0 °C

Temperature Transmiters3 (-30 to -20) °C 0.05 °C Digital thermometer &


(-20 to 0) °C 0.048 °C amperemeter, &
(0 to 50) °C 0.056 °C temperature bath
(50 to 100) °C 0.067 °C
(100 to 150) °C 0.08 °C
(150 to 200) °C 0.10 °C
(200 to 300) °C 0.14 °C
(300 to 420) °C 0.18 °C

IR Thermometers3 (-15 to 0) °C 0.49 °C Fluke 4180 black body

(>0 to 35) °C 0.43 °C Fluke 4181 black body


(>35 to 75) °C 0.44 °C
(>75 to 100) °C 0.47 °C
(>100 to 150) °C 0.56 °C
(>150 to 200) °C 0.65 °C
(>200 to 300) °C 0.90 °C
(>300 to 400) °C 1.2 °C
(>400 to 500) °C 1.5 °C

Temperature3 – Measure

Ovens, Liquid Baths, Dry (-80 to 420) ºC 0.039 °C Digital thermometer


Wells, Weather Cameras, (420 to 700) ºC 1.9 °C
Freezers, Stoves, (700 to 1000) °C 2.9 °C
Incubators, Muffles,
Refrigerators, Autoclaves

Sterile Rooms, Premises


with Controlled
Environments, Stability
Rooms, Warehouses, Cold
Rooms, Trucks

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 19 of 21


Parameter/Equipment Range CMC2, 4 () Comments

Relative Humidity – (20 to 90) % RH 2.1 % RH Thermo-hygrometer


Thermo Hygrometer Vaisala HM40 &
humidity chamber

Humidity3 – Measure, (20 to 90) % RH 2.3 % RH Thermo-hygrometer


Humidity Chambers Vaisala HM40

XI. Time & Frequency

Parameter/Equipment Range CMC2, 4 () Comments

Non-Contact Tachometers (100 to 3000) RPM 0.14 RPM Transmille 1000ª,


(4000 to 60 000) RPM 1.8 RPM multicabrator &
Transmille EA003
optical tachometer
adapter

Frequency – Measuring (0 to 100) Hz 0.01 Hz Transmille 1000A


Equipment (100 to 1) kHz 0.1 Hz multicalibrador
(1 to 10) kHz 0.6 Hz
(10 to 100) kHz 3 Hz

Stopwatches, Hours meters Up to 32 768 s (9.6 × 10-8X + 2.2×10-3) Comparison with


& Timers³ s reference stopwatches

X = reading value
_______________________________________________

1
This laboratory offers commercial calibration service and field calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 20 of 21


3
Field calibration service is available for this calibration. Please note the actual measurement uncertainties
achievable on a customer's site can normally be expected to be larger than the Calibration and
Measurement Capability Uncertainty (CMC) found on the A2LA Scope. Allowance must be made for
aspects such as the environment at the place of calibration and for other possible adverse effects such as
those caused by transportation of the calibration equipment. The usual allowance for the uncertainty
introduced by the item being calibrated, (e.g., resolution) must also be considered and this, on its own,
could result in the actual measurement uncertainty achievable on a customer’s site being larger than the
CMC.
4
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that generate the values in the ranges indicated for the
listed measurement parameter.
5
This scope meets A2LA’s P112 Flexible Scope Policy.
6
In the statement of CMC, L is the numerical value of the nominal length in the units indicated.
7
The stated measured values are determined using the indicated instrument (see Comments). This
capability is suitable for the calibration of the devices intended to measure or generate the measured value
in the ranges indicated. CMC’s are expressed as either a specific value that covers the full range or as a
percent or fraction of the reading plus a fixed floor specification.

(A2LA Cert. No. 4150.01) Revised 08/26/2022 Page 21 of 21


Accredited Laboratory
A2LA has accredited

TECHNOLOGY & METROLOGY TEGMETRO S.A.


Quito, ECUADOR
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017
General requirements for the competence of testing and calibration laboratories. This laboratory also meets the
R205 – Specific Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates
technical competence for a defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 9th day of April 2021.

_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 4150.01
Valid to December 31, 2022

For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.