Sitarc - C - 0135
Sitarc - C - 0135
Sitarc - C - 0135
2. DIGITAL CALIPERS
L.C. 0.01 mm 0 to 200 mm 7.5 µm Using Caliper Checker as per
0 to 300 mm 8.5 µm IS 3651 – Part 2
0 to 600 mm 12.5 µm
3. VERNIER DEPTH
GAUGES
L.C. 0.02 mm 0 to 150 mm 12.5 µm Using Grade 0 and K Slip
0 to 200 mm 14.0 µm Gauges as per IS 4213
0 to 300 mm 15.0 µm
0 to 600 mm 17.0 µm
5. VERNIER HEIGHT
GAUGES
L.C. 0.02 mm 0 to 300 mm 14.5 µm Using Height Master /
0 to 600 mm 17.0 µm Caliper Checker /
0 to 1000 mm 22.1 µm Slip Gauges as per IS 2921
6. DEGITAL HEIGHT
GAUGES
L.C. 0.01 mm 0 to 300 mm 10.1 µm Using Height Master /
0 to 600 mm 15.0 µm Caliper Checker / Slip
0 to 1000 mm 20.0 µm Gauges as per IS 2921
7. PISTOL CALIPERS
L.C. 0.01 mm 0 to 100 mm 60.0 µm Using Grade 0 Slip Gauges
as per IS 3651 – Part 1
8. EXTERNAL
MICROMETERS
L.C. 0.01 mm 0 to 25 mm 5.8 µm Using Grade 0 Slip Gauges /
25 mm to 100 mm 6.5 µm Long Slip Gauges /
100 mm to 300 mm 7.5 µm ULM as per IS 2967
300 mm to 600 mm 11.0 µm
600 mm to 1000 mm 15.0 µm
9. DIGITAL MICROMETERS
L.C. 0.001 mm 0 to 25 mm 1.0 µm Using Grade 0 Slip Gauges /
25 mm to 50 mm 1.47 µm Long slip Gauges /
50 mm to 100 mm 2.67 µm ULM as per IS 2967
100 mm to 150 mm 2.88 µm
150 mm to 200 mm 3.25 µm
Laboratory Scientific and Industrial Testing and Research Centre, 83 & 84,
Avarampalayam Road, K.R. Puram, Coimbatore
Accreditation Standard ISO/IEC 17025:2005
11. INTERNAL /
STICK MICROMETERS
L.C. 0.01 mm 0 to 25 mm 6.0 µm Using Grade 0 Slip Gauges /
25 to 50 mm 7.0 µm Long Slip Gauges as per
50 mm to 100 mm 7.2 µm IS 2966
100 mm to 300 mm 7.5 µm
300mm to 500 mm 9.0 µm
30. TACHOMETER 100 rpm to 3000 rpm 3 rpm Using Digital Tachometer
( CONTACT TYPE) Calibration Unit
31. TACHOMETER 100 rpm to 10000 rpm 2.5 rpm Using Digital Tachometer
( NON CONTACT TYPE) Calibration Unit
32. PRESSURE GAUGES & 0.5 bar to 60 bar 1.3% FS Using Digital Pressure
PRESSURE 60 bar to 1000 bar 0.38 % FS Calibrator
TRANSMITTER
6.2 kg/cm2 to 35 kg/cm2 0.42% FS Using Hydraulic Dead
35 kg/cm2 to 700 kg/cm2 0.1% FS Weight Tester
33. VACUUM GAUGES & - 710 mmHg to -30 mmHg 0.37% FS Using Vacuum Dead Weight
VACUUM TRANSMITTER Tester
* Measurement Capability is expressed as an uncertainty () at a confidence probability