General Information 0
An on-site water supply (e.g., chiller, central cooling water supply) is required for operating
the MR system. The primary chilled water supply must apply with the specifications listed
in this document, to guarantee a failure-free operation of the water cooling circuit, and the
water cooled components listed in the table (Water cooled components / p. 173).
Three solutions of on-site water supply are feasible:
NOTE In Germany, DIN 1988 "Technical guideline for tap water installa-
tions" should be part of the planning phase.
Local/country specific regulations/codes will apply in all other
countries.
NOTE The pipe/hose dimension between the MR chiller and IFP or cen-
tral water supply and SEP must be 2 inches. If other diameters are
used, the function of the water cooling circuit will be affected.
Check the use of the correct pipe/hose size during the plan-
ning/installation phase.
NOTE Only the KKT chiller type KCC 215 with the IFP panel is used as a
dedicated MAGNETOM Avanto chiller.
For all other chiller types or brands the SEP cabinet must be
installed.
NOTE The delivered water hoses for connecting the SEP primary circuit
are equipped with shut-off valves for the forward and return pipes.
This is to be able to change the SEP cabinet and maintain the pri-
mary strainer, if required.
The 2” on-site pipes for the primary water connection must have a
2” female thread.
The following components are connected to the secondary water cooling circuit.
Tab. 17 Water cooled components
Volume flow
Component Refrigerating capacity
(SEP / KKT-Chiler with IFP)
Gradient coil / Tas-C 25 l/min / 30 l/min < 30.0 kW
GPA 55 l/min / 60 l/min < 20.0 kW
ACS 25 l/min < 5.0 kW
RFPA 10 l/min < 2.0 kW
Refrigerator 10 l/min < 8.0 kW
NOTE If the system is switched “off” (i.e., water pump is not running), a
static overpressure of 1.0 to 1.5 bar is present in the secondary
water circuit.
NOTE Pay attention to the use/not use of the materials listed in the table
for the pipework (Piping material specification / p. 174).
NOTE The water circuit between KKT chiller and IFP or central water
supply and SEP has to be flushed before connecting.
The cooling water used has to meet our specifications.
CAUTION The pipe/hose dimension between the KKT chiller and IFP or cen-
tral water supply and SEP must be minimum 2” in size. If other
diameters are used, the function of the water cooling circuit will
be affected.
Check the use of the correct pipe/hose size during the
planning/installation phase.
NOTE The 2” water pipes are required for new system installations with
SEP configuration.
If the existing water pipes comply with our water requirement
specifications, they do no have to be modified.
The on-site primary water supply pipes (forward and return) must
be equipedp with a 2” female pipe thread.
NOTE All pipes for the primary water supply must be identified with
labels. The flow direction of the cooling media has to be identified
as well. Local requirements have to be fulfilled.
NOTE The 2” adapter ((Fig.1: Pos. 2) DIN-Inch) for preparing the on-site
primary water connection prior system delivery can be ordered
with
PN.: 10018861
NOTE The following specifications are valid for all chilled water circuits.
NOTE System with dedicated MAGNETOM Avanto KKT KCC 215 Chiller:
The water circuit must be filled with minimum 35% to maximum
38% ethylene glycol.
Never use propylene glycol or automobile anti-freeze.
NOTE If the KKT chiller will not be part of the system configuration a
SEP cabinet must be used.
Only two possibilities can be chosen for the system cooling con-
figuration:
1: SEP cabinet or
2: KKT KCC 215 chiller plus IFP
The IFP (Interface and filter panel) is part of the MR system if the local building environ-
ment does not supply central cooling water. A dedicated KKT MR chiller has to be used in
this case. This KKT MR chiller (KKT type KCC 215) is mandatory for this type of chilled
water supply.
1. The value of 63kW includes the heat dissipation of the built in pump.
2. Flow rate between KKT chiller and IFP/ACC; can not be affected.
3. Flow rate between on-site chilled water supply and SEP.
4. The supply temperature from the dedicated KKT chiller; can not be affected.
5. The difference in temperaure from the primary water inlet temperature (into the SEP) to the primary
water outlet (out of the SEP) is app. 10K (e.g. primary water inlet is 10°C the primary outlet will be app.
22°C). This is valid at maximum heat dissipation to water.
6. This specification is valid for all load change conditions with a floating integration time of 30 s.; can not
be affected.
7. This specification is valid for all load change conditions with a floating integration time of 30 s.
8. Typical 0.8 bar.
NOTE A tab water connection within 20.0m distance to the SEP, IFP, ACC
cabinet has to be available.
This is for filling up the secondary water circuit.