Cze Series: High Voltage Power Supply

Download as pdf or txt
Download as pdf or txt
You are on page 1of 31

R

CZE SERIES

High Voltage Power Supply


MODEL :
SERIAL# :
DATE :

SPELLMAN
HIGH VOLTAGE ELECTRONICS
CORPORATION
475 Wireless Blvd.
Hauppauge, New York, 11788

+1(631) 630-3000*FAX: +1(631) 435-1620*


E-mail: [email protected]
Website: www.spellmanhv.com

CZE MANUAL 118031-001 Rev D


CZE2000 AUTO-REVERSING
POWER SUPPLY
S P E L L M A N H I G H V O LTA G E E L E C T R O N I C S C O R P O R AT I O N
PAGE 1 OF 2

• Ideal for Electrospinning


• 0-30kV Remotely Programmable
• 0-300µA Remotely Programmable
• Polarity Reversible Upon Command
in <1 Sec at No Load
• Low Stored Energy, Current Limited Output
• Cost Effective Modular Design
www.spellmanhv.com/manuals/CZE2000

Spellman’s CZE2000 modular high voltage power supply Line Regulation:


is ideal for OEM usage. It is specifically designed to meet 0.01% for a 10% input voltage change
the needs of applications requiring a hot switched Load Regulation:
reversible output voltage. The output polarity of the unit 0.01% for a full load change
can be quickly and safely reversed via the Polarity Ripple:
Control Signal provided on the interface connecter. 0.1% Vp-p
Both the output voltage and current are fully adjustable
Stability:
via ground referenced remote programming signals such 0.02% per 8 hours (after 1/2 hr warmup)
that 0 to 10Vdc corresponds to 0 to 100% rated output
voltage and current. NL Time Constant:
100ms
Remote motioning functionality is provided by voltage
and current test points such that 0 to 10Vdc corresponds Stored Energy:
to 0 to 100% rated voltage and current. Additionally 0.2 Joules at 30kV
remote polarity and mode indicators provide a compre- Temperature Coefficient:
hensive overview of power supply operation. 100ppm/°C
Excellent load and line regulation specifications along Operating Temperature:
with outstanding stability and low ripple assure a stable 0°C to 40°C
high voltage output for consistent process results. Storage Temperature:
-40°C to 85°C
TYPICAL APPLICATIONS
Humidity:
Electrospinning
10% to 85% RH, non condensing
Mass Spectrometry
Capillary Electrophoresis Cooling:
Electrostatic Research Convection cooled
Dimensions:
3.5”H x 5”W x 10”D (8.9cm x 12.7cm x 25.4cm).
SPECIFICATIONS Weight:
Input Voltage: 6.2lbs. (2.8kg)
24Vdc, ±10% Interface Connector:
Input Current: 25 pin male D connector
Less than 1 amp
Efficiency: HV Output Connector:
75% typical Detachable 36˝ (0.91m) cable provided

Output Voltage: Regulatory Approvals:


See selection table Compliant to 2004/108/EC, The EMC Directive and
2006/95/EC, The Low Voltage Directive, UL/CUL
Output Current: recognized file E148969
See selection table
Polarity:
Auto reversible via command
Power:
10 watts, maximum

Corporate Headquarters
Hauppauge, New York USA www.spellmanhv.com 128076-001 REV. D

+1-631-630-3000 FAX: +1-631-435-1620 Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
e-mail: [email protected] Copyright © 2002 Spellman High Voltage Electronics Corp.
CZE2000 AUTO-REVERSING
POWER SUPPLY
S P E L L M A N H I G H V O LTA G E E L E C T R O N I C S C O R P O R AT I O N

PAGE 2 OF 2

CZE2000 SELECTION TABLE DIMENSIONS: in.[mm]


MAXIMUM RATING MODEL NUMBER
FRONT VIEW
kV mA
5 2.0 CZE 5PN2000 JST CONNECTOR JBY-25P-1A1A
SHV P/N 105708-652
10 1.0 CZE 10PN2000
15 0.67 CZE 15PN2000
20 0.50 CZE 20PN2000 .79 ± .02
30 0.30 CZE 30PN2000 [20.1] J1

2.50 ± .02 3.50 ± .02


[63.5] J2
[88.9]
CZE2000 25 PIN MALE D CONNECTOR
PIN SIGNAL PARAMETERS 1.00
± .02
HV OUTPUT [25.4]
1 +24Vdc Return Power Return 2.50 ± .02
2 +24Vdc Return Power Return [63.5]
5.00 ± .02
3 +24Vdc Return Power Return [127.0]
4 HV Enable/Inhibit Open or <1Vdc = HV OFF, >3.4Vdc
(up to 15Vdc) = HV ON
5 Voltage Test Point 0 to 10Vdc = 0 to 100% rated output,
Zout =10kΩ, 1% TOP VIEW
6 Current Test Point 0 to 10Vdc = 0 to 100% rated output, 2.50[63.50] #8-32 SELF-CLINCHING NUT
Zout =10kΩ, 1% FLUSH-HEAD STUD
6.000 ± .010 4 PL.
7 Chassis Ground Ground
8 Remote Voltage Control 0 to 10Vdc = 0 to 100% Rated Output, Zin =10MΩ .944 ± .02
9 Remote Current Control 0 to 10Vdc = 0 to 100% Rated Output, Zin =10MΩ
10 +10Vdc Reference Output +10Vdc, 4mA maximum
11 Signal Return Signal Return 3.000 ± .010
12 Polarity Control Open or >3.4Vdc (up to 15Vdc) = Positive
Polarity. Grounded or <1Vdc = Negative Polarity
13 Positive Polarity Indicator +24Vdc sourced through a 100Ω series limiting
resistor. +24Vdc = active signal
14 +24Vdc Input Power Input
15 +24Vdc Input Power Input
16 Chassis Ground Ground
17 Negative Polarity Indicator +24Vdc sourced through a 100Ω series limiting
resistor. +24Vdc = active signal BOTTOM VIEW
18 I Mode Indicator Open collector pulled up internally to +15Vdc
through 2.7kΩ resistor with a 470Ω limiting 2.50 ± .03 6.000 ± .010 1.42 ± .02
resistor in series. Transistor OFF = signal active [63.5] [152.4] [36.1]
19 V Mode Indicator Open collector pulled up internally to +15Vdc
through 2.7kΩ resistor with a 470Ω limiting .940 ± .02
resistor in series. Transistor OFF = signal active [23.9]
HIGH VOLTAGE ELECTRONICS CORP.
MADE IN USA

20 Return Current Test Point 0 to 10Vdc = 0 to 100% rated output current, as


measured returned from load. Zout =10kΩ, 1% 3.000 ± .010
SPELLMAN
Plainview, NY 11803

21 Load Return High Voltage Return Point. Required for GFI [76.2]
SERIAL NO.

4-40 FIXED
PART NO:

circuit functionality
OUTPUT:
MODEL:

INPUT:

FEMALE
22 Ground Fault Indicator Open collector pulled up internally to +15Vdc SCREWLOCKS
through 4.7kΩ resistor with a 470Ω limiting
resistor in series. Transistor OFF = signal active
.24 ± .00 8–32 X .50 LG.
23 Spare No Connection INTERNAL THREADS 8-32 X .28 .03 GND STUD
24 Spare No Connection DEEP TYP. 4 PL. [6.09]
25 Spare No Connection

Corporate Headquarters
Hauppauge, New York USA www.spellmanhv.com 128076-001 REV. D

+1-631-630-3000 FAX: +1-631-435-1620 Spellman High Voltage is an ISO 9001:2008 and ISO 14001:2004 registered company
e-mail: [email protected] Copyright © 2002 Spellman High Voltage Electronics Corp.
IMPORTANT SAFETY PRECAUTIONS

SAFETY
THIS POWER SUPPLY GENERATES VOLTAGES THAT ARE DANGEROUS AND MAY BE FATAL.
OBSERVE EXTREME CAUTION WHEN WORKING WITH THIS EQUIPMENT.

High voltage power supplies must always be grounded.

Do not touch connections unless the equipment is off and the


Capacitance of both the load and power supply is discharged.

Allow five minutes for discharge of internal capacitance of the power supply.

Do not ground yourself or work under wet or damp conditions.

SERVICING SAFETY
.

Maintenance may require removing the instrument cover with the power on.

Servicing should be done by qualified personnel aware of the electrical hazards.

WARNING note in the text call attention to hazards in operation of these units
that could lead to possible injury or death.

CAUTION notes in the text indicate procedures to be followed to avoid possible


damage to equipment.

Copyright © 2000, Spellman High Voltage Electronics Corporation. All Rights Reserved.
This information contained in this publication is derived in part from proprietary and patent data. This information has
been prepared for the express purpose of assisting operating and maintenance personnel in the efficient use of the
model described herein, and publication of this information does not convey any right to reproduce it or to use it for
any purpose other than in connection with installation, operation, and maintenance of the equipment described.

118004-001 REV. B
WICHTIGE SICHERHEITSHINWEISE

SICHERHEIT
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT LEBENSGEFÄHRLICHE HOCHSPANNUNG.
SEIN SIE SEHR VORSICHTIG BEI DER ARBEIT MIT DIESEM GERÄT.

Das Hochspannungsnetzteil muß immer geerdet sein.

Berühren Sie die Stecker des Netzteiles nur, wenn das Gerät ausgeschaltet ist und die elektrischen
Kapazitäten des Netzteiles und der angeschlossenen Last entladen sind.

Die internen Kapazitäten des Hochspannungsnetzteiles benötigen ca. 5 Minuten, um sich zu entladen.

Erden Sie sich nicht, und arbeiten Sie nicht in feuchter oder nasser Umgebung.

SERVICESICHERHEIT
Notwendige Reparaturen können es erforderlich machen, den Gehäusedeckel während des Betriebes zu
entfernen.

Reparaturen dürfen nur von qualifiziertem, eingewiesenem Personal ausgeführt werden.

“WARNING” im folgenden Text weist auf gefährliche Operationen hin, die zu Verletzungen oder zum Tod
führen können.

“CAUTION” im folgenden Text weist auf Prozeduren hin, die genauestens befolgt werden müssen, um
eventuelle Beschädigungen des Gerätes zu vermeiden.

118004-001 REV. B
PRECAUTIONS IMPORTANTES POUR VOTRE SECURITE

CONSIGNES DE SÉCURITÉ
CETTE ALIMENTATION GÉNÈRE DES TENSIONS QUI SONT DANGEUREUSES ET PEUVENT ÊTRE FATALES.
SOYEZ EXTRÊMENT VIGILANTS LORSQUE VOUS UTILISEZ CET ÉQUIPEMENT.

Les alimentations haute tension doivent toujours être mises à la masse.

Ne touchez pas les connectiques sans que l’équipement soit éteint et que la capacité à la fois de la charge et de
l’alimentation soient déchargées.

Prévoyez 5 minutes pour la décharge de la capacité interne de l’alimentation.

Ne vous mettez pas à la masse, ou ne travaillez pas sous conditions mouillées ou humides.

CONSIGNES DE SÉCURITÉ EN CAS DE REPARATION


La maintenance peut nécessiter l’enlèvement du couvercle lorsque l’alimentation est encore allumée.

Les réparations doivent être effectuées par une personne qualifiée et connaissant les risques électriques.

Dans le manuel, les notes marquées « WARNING » attire l’attention sur les risques lors de la manipulation de ces
équipements, qui peuvent entrainer de possibles blessures voire la mort.

Dans le manuel, les notes marquées « CAUTION » indiquent les procédures qui doivent être suivies afin d’éviter
d’éventuels dommages sur l’équipement.

118004-001 REV. B
IMPORTANTI PRECAUZIONI DI SICUREZZA
SICUREZZA
QUESTO ALIMENTATORE GENERA TENSIONI CHE SONO PERICOLOSE E
POTREBBERO ESSERE MORTALI.
PONI ESTREMA CAUTELA QUANDO OPERI CON QUESO APPARECCHIO.

Gli alimentatori ad alta tensione devono sempre essere collegati ad un impianto di terra.

Non toccare le connessioni a meno che l’apparecchio sia stato spento e la capacità interna
del carico e dell’alimentatore stesso siano scariche.

Attendere cinque minuti per permettere la scarica della capacità interna dell’alimentatore
ad alta tensione.

Non mettere a terra il proprio corpo oppure operare in ambienti bagnati o saturi d’umidità.

SICUREZZA NELLA MANUTENZIONE.

Manutenzione potrebbe essere richiesta, rimuovendo la copertura con apparecchio


acceso.
La manutenzione deve essere svolta da personale qualificato, coscio dei rischi elettrici.

Attenzione alle AVVERTENZE contenute nel manuale, che richiamano all’attenzione ai


rischi quando si opera con tali unità e che potrebbero causare possibili ferite o morte.

Le note di CAUTELA contenute nel manuale, indicano le procedure da seguire per evitare
possibili danni all’apparecchio.

118004-001 REV. B
Table of Contents
PAGE
1. INTRODUCTION
1.1 Description Of The CZE Series ...........................................................................1
1.2 CZE Specifications ..............................................................................................1
1.3 Standard Features.................................................................................................2
1.4 Options.................................................................................................................2
1.5 Interpreting The Model Number ..........................................................................3
2. INSPECTION & INSTALLATION
2.1 Initial Inspection ..................................................................................................4
2.2 Mechanical Installation ........................................................................................4
3. OPERATING INSTRUCTIONS
3.1 Operation..............................................................................................................7
3.2 Standard Features.................................................................................................7
4. PRINCIPLES OF OPERATION
4.1 Chassis .................................................................................................................15
4.2 High Frequency Inverter ......................................................................................15
4.3 High Voltage Circuits ..........................................................................................16
4.4 Control Circuits....................................................................................................16
4.5 Polarity Reversal ..................................................................................................16
4.4 Options.................................................................................................................16
5. OPTIONS
5.1 Floating Ground ...................................................................................................17
5.2 Input Voltage........................................................................................................17
5.3 No Slow Start.......................................................................................................17
5.4 Non-Standard Slow Start .....................................................................................17
5.5 Extra Length Output Cable ..................................................................................17
5.6 Auxiliary Outputs.................................................................................................17
5.7 Custom Designed Models ...................................................................................17
6. MAINTENANCE
6.1 Periodic Servicing................................................................................................19
6.2 Performance Test .................................................................................................19
6.3 High Voltage Dividers .........................................................................................19

CZE MANUAL ii 118030-001 REV D


7. REPLACEMENT PARTS
7.1 Replacement Parts................................................................................................20
7.2 Correspondence and Ordering Spare Parts ..........................................................20
8. FACTORY SERVICE
8.1 Warranty Repairs .................................................................................................21
8.2 Factory Service Procedures..................................................................................21
8.3 Ordering Options and Modifications ...................................................................21
8.4 Shipping Instructions ...........................................................................................21
APPENDIX
A. Specification Controls (Custom Models Only)
B. Schematics

LIST OF FIGURES
2.1 CZE1000 Dimensions & Connector Interface .....................................................5
2.2 CZE2000 Dimensions & Connector Interface .....................................................6
3.1 Typical Operating Set Up CZE1000 ....................................................................10
3.2 Typical Operating Set Up CZE2000 ....................................................................10
3.3A Remote Programming Via Voltage Source..........................................................11
3.3B Remote Programming Via External Resistance...................................................11
3.4 Remote Monitor ...................................................................................................12
3.5 HV ON/OFF Control ...........................................................................................12
3.6 Polarity Control....................................................................................................13
3.7 Polarity Indicator..................................................................................................13
3.8 Mode Indicator.....................................................................................................14
3.9 GFI & Load Return ..............................................................................................14
3.7A Floating Ground Interface Circuit........................................................................18
LIST OF TABLES

1.1 Options.................................................................................................................2
2.1 Output Voltage and Current ................................................................................6

CZE MANUAL iii 118030-001 REV D


Chapter 1

INTRODUCTION
1.1 Description Of The CZE Series · Current Regulation:
Load Regulation: 0.05% ±25uA from 0 voltage to

T he CZE series of high voltage power supplies


provides very well regulated, low ripple high voltage
in a highly efficient, compact design. The
full voltage.
Line Regulation: ±0.01% of full current over the
specified input voltage range.
dramatically reduced size of these modules, compared to
traditional high voltage modules, is obtained by a state of · Ripple:
the art off-line resonant converter. The resonant converter 0.1% p-p of output voltage.
utilizes a unique control scheme, which allows constant
frequency operation while maintaining high efficiency. · Polarity:
High operating frequency allows for very low ripple and Reversible via TTL Signal in Polarity Control Signal
excellent dynamic response capabilities. Pin.
The DC output voltage and current are controllable over · Stability:
the full range of operation. Monitoring and control
signals are provided for simple, yet flexible control of the 0.01% per hour after 1/2 hour warm up.
power supply. The CZE series operates from 24VAC · Temperature Coefficient:
±10%. (User should specify the input voltage at the time
of ordering). The CZE series operates at 9W continuous. 100 ppm per °C.
The series does not require forced air cooling. The
· Temperature:
ambient temperature must be kept below the maximum
rating as specified in 1.2. Consult Spellman Sales Operating: 0°C to +50°C
Department for higher power capabilities of the CZE Storage: -40°C to +80°
series.
The standard warranty applies to the CZE series modules. · Monitoring:
Consult factory about the warranty for custom modules. 0 to 10VDC corresponding to 0 to 100% of output
voltage
1.2 CZE Specifications 0 to 10VDC corresponding to 0 to 100% of output
· Output Control: Voltage and current are externally current.
programmable over the entire range from zero to 0 to 10VDC corresponding to 0 to 100% of load
maximum rating via 0-10VDC reference or return current input.
potentiometer. (Other scale factors for control and Ground Fault Indication (GFI): +15V = Fault;
monitoring can be provided i.e. 1 volt per 10KV, 1 Output Impedance: = 4.7kOhma.
volt per 1mA etc.). (Other scale factors available).

· Input
· Control:
Voltage: 24Vdc ±10%. Consult the factory for
other input voltages. Control signal for high voltage enable/disable.

· Voltage Regulation: IMPORTANT


Load Regulation: 0.01% of full voltage for a no
load to full load change. This control signal in not a safety
Line Regulation: ±0.01% of full voltage over the interlock and should not be used for
specified input voltage range.
protection from high voltage generation
for safety purposes.

CZE MANUAL 1 118030-001 REV D


1.3 Standard Features Remote Monitor: Allows remote monitoring of the
output voltage and current.
The CZE series incorporates several standard features
designed to optimize user operation. High Voltage Enable/Disable: Allows remote ON/OFF
control of the high voltage.
Slow Start: Provides a gradual increase in high voltage
output until the maximum set point is reached. Various +10VDC Reference: A +10VDC reference is provided
slow start times can be accommodated. Consult the for remote programming via a potentiometer or resistive
Spellman’s Sales Department for information on slow divider.
start options. (The slow start time will be a function of set
point, input voltage and load conditions).
1.4 Options
The options available are listed in Table 1.1. See section
Polarity Control: A TTL compatible input controls the
5 for more information on the options along with
polarity of the high voltage output. A logic level HIGH
operating and set-up instructions. With few exceptions,
will give a Positive, a LOW a Negative. If left open, the
these options can be retrofitted to your power supply at
unit will default to Positive.
the factory in a short time. For price and retrofit
G.F.I.: The Ground Fault Interrupt will give an arrangements, contact the Spellman Sales Department.
indication if there is more than approximately a 10%
CODE DISCRIPTION
difference between the return current test point and the
delivered current test point. • FG Floating Ground
Polarity Indicator: Two outputs indicate whether • Alternate Test Point Scaling
positive or negative polarity has been selected.
• Test Point Polarity Reversal
Mode Indicators: The two mode outputs indicate which
control loop is in control, the voltage or the current • NSS No Slow Start
regulation circuits. • SS(X) Non-Standard Slow Start
• LL(X) Extra Length Output Cable
1.3.1 Remote Operating Features
• Non-Standard Output Connector
Remote Programming: Allows remote adjustment of the
output voltage and current via an external voltage source. Table 1.1 Options

CZE MANUAL 2 118030-001 REV D


1.5 Interpreting the Model CZE 30 P/N 1000/2000 / FG / X#
Number:
The model number of the power supply describes its Series
Custom
"X" Number
capabilities. After the series name is: Name Option
Maximum Series of
CZE
Voltage
(1) The maximum voltage in kilovolts. Polarity
Reversible
(2) P/N indicated reversibility.
(3) The Series of CZE 1000 or 2000
(4) The option code for all options that are
included.
(5) Custom “X” number representing details Maximum Ratings Model
listed in a separate specification control kV mA
Number
drawing.
5 2 CZE5PN1000/2000
10 1 CZE10PN1000/2000
15 0.67 CZE15PN1000/2000
20 0.50 CZE20PN1000/2000
30 0.3 CZE30PN1000/2000

CZE MANUAL 3 118030-001 REV D


Chapter 2

Inspection and Installation

I nitial inspection and preliminary check-out


procedures are recommended. For safe
operation, please follow the step-by-step
2.2 Mechanical Installation
The CZE series module power supplies are
procedures described in Chapter 3, Operating designed for installation into existing or newly
Instructions. developed OEM equipment. The power supply
can also easily fit into bench top applications or
2.1 Initial Inspection test set requirements. Standard unit dimensions
are shown in Figure 2.1
Inspect the package exterior for evidence of
damage due to handling in transit. Notify the For custom mounting requirements or specific
carrier and Spellman immediately if damage is package size requirements consult Spellman’s
evident. Do not destroy or remove any of the Sales Department. Spellman has many package
packing material used in a damaged shipment. designs available, or can design a specific
After unpacking, inspect the panel and chassis enclosure for your requirements.
for visible damage.
The CZE series utilizes solid encapsulations for
Fill out and mail the Warranty Registration card corona free operation. No periodic maintenance
accompanying the unit. Standard CZE high is required. Due to conservative voltage spacing
voltage power supplies and components are design, periodic maintenance is not required
covered by warranty. Custom and special order
models (with an X suffix in the model number)
are also covered by warranty.

CZE MANUAL 4 118030-001 REV D


PIN FUNCTION Remote Programming:
1 Chassis Ground & 24Vdc Return Permits remote adjustment of the output voltage
and current via an external voltage source. By
2 +24Vdc Input adjusting the voltage source from 0V to 10V the
3 High Voltage Enable/Inhibit desired output may be selected.

4 Voltage Test Point: 10V±2% = 0 to Rated Remote Monitor:


Output Test points, pins 4 and 5 of the input connector,
5 Current Test Point: 10V±2% = 0 to Rated monitor voltage and current outputs. The output
Output polarity is positive from 0V to 10V equal to 0 to
100% of the output ±0.5%. Output impedance is
6 Remote Voltage Control (0 to 10V) 10k Ohms.
7 Remote Current Control (0 to 10V) High Voltage Enable/Inhibit:
8 +10.24Vdc Reference The external TTL signal allows the user to inhibit
the power supply output without switching the
9 Test Point & Remote Programming Return HIGH VOLTAGE OFF.
10 Polarity Control Signal
11 Positive Polarity Indicator
12 Negative Polarity Indicator
13 Current Mode Indicator
14 Voltage Mode Indicator
15 Return Current Test Point
16 Load Return
17 Ground Fault Indication
18 Spare (Shield Ground Option)
Note: If Enable/Inhibit is not used, connect Pin 3 to Pin 8.

Figure 2.1 CZE1000 Dimensions & Connector Interface

CZE MANUAL 5 118030-001 REV D


PIN FUNCTION Remote Programming:
1,2,3 Chassis Ground & 24Vdc Return Permits remote adjustment of the output voltage and
current via an external voltage source. By adjusting the
4 High Voltage Enable/Inhibit voltage source from 0V to 10V the desired output may be
5 Voltage Test Point: 10V±2% = 0 to Rated selected.
Output Remote Monitor:
6 Output Current Test Point: 10V±2% = 0 to Test points, pins 5 and 6 of the input connector, monitor
Rated Output voltage and current outputs. The output polarity is
7 Chassis Ground positive from 0V to 10V equal to 0 to 100% of the output
±0.5%. Output impedance is 10k Ohms.
8 Remote Voltage Control
High Voltage Enable/Inhibit:
9 Remote Current Control
The external TTL signal allows the user to inhibit the
10 +10.24Vdc Reference power supply output without switching the HIGH
VOLTAGE OFF.
11 Test Point and Remote Programming Return
12 Polarity Control Signal
13 Positive Polarity Indicator
14, 15 +24Vdc Input
16 Chassis Ground
17 Negative Polarity Indicator
18 I – Mode Indicator
19 V- Mode Indicator
20 Current Return Test Point
21 Load Return
22 Ground Fault Indication
23,24,25 Spares
Note: If Enable/Inhibit is not used, connect Pin 4 to Pin
10.

Figure 2.2 CZE2000 Dimensions & Connector Interface

CZE MANUAL 6 118030-001 REV D


Chapter 3

Operating Instructions
3.1 Operation A) Insure that the high voltage cable is properly
terminated to the load. Insure that all circuits connected
WARNING to the high voltage output are safely interlocked against
accidental contact. Insure external load is discharged.
THIS EQUIPMENT GENERATES B) Check the input voltage rating on the serial
DANGEROUS VOLTAGES THAT MAY BE nameplate of the supply and make certain that this is the
FATAL. PROPER GROUNDING OF ALL HIGH rating of the available power source. CZE series modules
VOLTAGE EQUIPMENT IS ESSENTIAL. operate on 28Vdc unless ordered with a different input
voltage option. External fusing is recommended for the
DC input power. For the CZE series a 2 AMPHERE
WARNUNG SLO-BLO type fuse must be used.
DIESES GERÄT ERZEUGT C) PROPER GROUNDING TECHNIQUE: The
LEBENSGEFÄHRLICHE HOCHSPANNUNG. chassis of high voltage power supplies must be grounded,
ALLE HOCHSPANNUNGSGERÄTE MÜSSEN preferably to a water system ground using copper pipe or
other earth ground. See Figure 3.1 and 3.2 for a typical
ÜBER EINE GEEIGNETE ERDUNG
operating setup. The return line from the load should be
VERFÜGEN. connected to the power supply chassis or to Load Return
pin.
IMPORTANT D) Options Note: See section 5 for hook up and
Before connecting the power supply to the operating instructions for the options on your unit.
AC line, follow this step-by-step procedure. Custom models may also require set up changes.
Do not connect the power supply to the AC E) Hook-up: Connect Input control and monitoring
line until Step G is reached. connections as described in this manual.
Failure to follow these procedures may void F) For initial turn-on, program the voltage and
the warranty. current for zero output. Connect the enable/disable signal
to disable by grounding the Enable line. Set the desired
ACHTUNG polarity.
BEVOR SIE DAS
G) The DC input power may now be connected.
HOCHSPANNUNGSNETZEIL AN DIE
STROMVERSORGUNG ANSCHLIESSEN, H) Enable the power supply via the enable/disable
MÜSSEN FOLGENDE PUNKTE GEPRÜFT logic signal.
WERDEN. I) Slowly program the output voltage and current to
SCHLIESSEN SIE DAS desired level. Monitor the output voltage and current via
the monitoring test points. Note equipment operation is
HOCHSPANNUNGSNETZEIL NICHT AN DIE
normal, i.e. Load is behaving as predicted.
SPANNUNGSVERSORGUNG BEVOR PUNKT
G ERREICHT IST. EVENTUELLE J) To turn high voltage off, use the enable/disable
AUFTRETENDE BESCHÄDIGUNG DES signal. If equipment is to be kept off for extended
periods, disconnect power supply from the DC INPUT
GERÄTES DURCH NICHT BEFOLGEN
source.
DIESER ANWEISEN KANN ZUM VERLUST
DES GARANTIEANSPRUCHES FÜHREN.

CZE MANUAL 7 118030-001 REV D


berühren oder zu reparieren, da
WARNING
lebensgefährliche Hochspannungen in Innern
AFTER TURNOFF, DO NOT HANDLE THE erzeugt werden.
LOAD UNTIL THE CAPACITANCE HAS
BEEN DISCHARGED! 3.2 Standard Features
LOAD CAPACITANCE MAY BE DISCHARGED BY
A note on remote interface circuitry and remote signal
SHORTING TO GROUND.
grounding: whenever possible, electrical isolation should
be provided when interfacing with any high voltage power
WARNUNG supply. For enable/disable signal connections, an isolated
relay or optocoupler should be used. If possible, analog
Nach drm Ausschalten des Gerätes die Last erst programming and monitoring signals should be isolated
berühren wenn diese vollständig entladen ist. via analog isolation amplifiers. Spellman application
Die elektrische Kapazität der Last kann durch engineers are available to assist in interface circuitry
einen Kurzschluß zur Erde entladen werden. design. All interface cables should be properly shielded.
All power supply signals should be referenced to the
WARNING power supplies signal ground or power supply chassis
ground
THE VOLTAGE MONITOR ON THE POWER
REMOTE PROGRAMMING: Allows remote
SUPPLY FRONT PANEL DOES NOT READ THE
adjustment of the output voltage and current via an
OUTPUT VOLTAGE WHEN THE POWER IS
external voltage source. A 0 to 10VDC signal is supplied
TURNED OFF, EVEN IF A CHARGE STILL EXISTS
to Pin 6/8 of the control input connector for voltage
ON THE LOAD.
programming. A 0 to 10 VDC signal is supplied to Pin
7/9 of the control input connector for current
WARNUNG programming. Programming signals should be referenced
to Pin 9/11 of the control input connector signal ground.
Der Spannungsmonitor arbeitet nicht bei By adjusting the voltage source from 0 volts (zero output)
abgeschalteter Versorgungspannung, auch nicht to 10 volts (full rated output) the desired output can be
wenn die Last noch aufgeladen ist. selected. See Figure 3.3A for wiring diagram and
specifications.
CAUTION An alternate method of controlling the output remotely is
by using external resistance such as a potentiometer or a
ALWAYS OPERATE THE UNIT WITH THE COVER
resistor network. See Figure 3.3B for wiring diagram.
ON. DO NOT ATTEMPT TO ACCESS OR REPAIR
ANY INTERNAL CIRCUITS. DANGEROUS AND Remote Monitoring: Test points are made available for
LETHAL VOLTAGES ARE GENERATED INSIDE monitoring the voltage and current output. The test points
THE MODULE. are always positive regardless of the output polarity,
where zero (0) to 10 (10) volts equals 0-100% of the
output. See Figure 3.4 for test point wiring.
CAUTION
.
Betreiben Sie das Hochspannungsnetzteil
ausschließlich mit geschlossenem Gehäuse.
Versuchen Sie nicht die internen Schaltkreise zu

CZE MANUAL 8 118030-001 REV D


current drawn by the incandescent bulb (maximum current
WARNING is 50mA). If it is desired to drive an LED instead of an
It is extremely dangerous to use this incandescent bulb, simply add a 3.3k 1/4W resistor in
series with the output and the LED.
circuit to inhibit high voltage generation
for the purpose of servicing or Voltage and Current Mode Signals: +15V is provided
through a 3.3K resistor to drive an LED. +15V on pin 19
approaching any area of load indicates that the voltage regulator mode is selected.
considered unsafe during normal use. +15V on pin 18 indicated that the current regulation mode
is selected. The internal control circuitry employs
automatic crossover between voltage and current modes.
WARNUNG In the voltage mode, the current level is set and the
voltage programming is used to control the output. If the
Es ist sehr gefährlich diese Schaltung load current in increased to the point where it exceeds the
preset level, the unit will cross over automatically into
zum abschalten der Spannung fur current mode. In current mode operation, the voltage
Reparatur oder Service Zwecke zu programming is preset and current control controls the
benutzen. output. If the load current is reduced to the point that the
voltage exceeds the preset level, then the output will
automatically cross over to voltage mode. When the unit
Reverse Polarity Control Signal: This input signal is crosses over from either current to voltage mode or
intended to be driven from TTL logic and will cause the voltage to current mode, the mode signal will switch with
output to switch polarity. The unit will automatically it.
sequence all internal functions to switch output polarity Control Current Test Point: +10V on pin 8/10 = Full
with minimal stress. Greater than 3.4V (or open) on pin Scale Current ±2%. The return current circuitry allows a
10 yields Positive High Voltage. Less than 1V on pin 10 separate path to measure actual load current. The load is
yields Negative High Voltage. returned to the load return, pin 16/21 of the connector and
Operation is as follows: upon reverse command, the a buffer test point is provided on pin 15/20. The standard
inverter in instantaneously disabled and the output begins current test point provided on pin 5/6 measures the current
to decay from its set point to zero. At no load, this time delivered to the load while the return current test point
constant is 0.1 seconds. At approximately 10% output, measures the current returned to the power supply by the
the logic circuitry will switch the primary and secondary load. By comparing these two signals, any differences
relays and enable the inverter. The output then slows up between them must be going into leakage. Therefore, if
to its set value time constant. Obviously, the time to this leakage exceeds some level, the unit could be shut
complete the switch operation is a function of output down.
voltage setting and load current. The maximum switch Ground Fault Indication: +15 V on pin 17/22 = Fault.
time for set points above 3kV occurs at no load and 30kV. Output impedance = 4.7K. This feature compares the
Under this condition, that time is a constant 100 return load current to the delivered load current and
milliseconds, regardless of output voltage or current provides a fault signal if the difference between these
setting. signals (leakage current) exceeds 20uA. This fault signal
Polarity Indicator: +24V on pin 11/13 indicates positive can be used to shut down the whole system or through
polarity selected, +24V on pin 12/17 indicates negative appropriate circuitry to shut down the power supply either
polarity selected. Output impedance is 100 ohms. This is by using the remote enable or by removing the +24V input
intended to drive an incandescent lamp. Internally, a relay power. The output circuit for this fault indication is an
contact switches the 100ohm resistor to the +24V. open collector with a 4.7K pull up resistor
Therefore, the indicator bulb should be connected
between the output and ground. Please note that the
power supply input current will increase by the amount of

CZE MANUAL 9 118030-001 REV D


Figure 3.1 CZE1000 Typical Operating Set Up

Figure 3.2 CZE2000 Typical Operating Set up

CZE MANUAL 10 118030-001 REV D


Figure 3.3A CZE Remote Programming Via Voltage Source
____________________________________________________________________________________

Figure 3.3B CZE Remote Programming Via External Resistance

CZE MANUAL 11 118030-001 REV D


Figure 3.4 CZE Remote Monitoring
____________________________________________________________________________________

Figure 3.5 CZE HV ON/OFF Control

CZE MANUAL 12 118030-001 REV D


Figure 3.6 CZE Polarity Control
____________________________________________________________________________________

Figure 3.7 CZE Polarity Indicators

CZE MANUAL 13 118030-001 REV D


Figure 3.8 CZE Mode Indicators
____________________________________________________________________________________

Figure 3.9 GFI and Load Return

CZE MANUAL 14 118030-001 REV D


Chapter 4

PRINCIPLES OF OPERATION
T he CZE Series of high voltage power supplies utilizes
sophisticated power conversion technology.
Advanced analog and power conversion techniques
Kenntnis über mögliche Gefahren beim Ungang mit der
Hochspannung haben. Versuchen Sie nicht interne
Schaltkreise oder elektrische Bauelemente, die an das
are used in the CZE series. The intention of the Principles Hochspannungsnetzteil angeschlossen sind, zu berühren.
of Operation is to introduce the basic function blocks of Vergewissern Sie sich, daß alle Restladungen, die vor
the CZE series power supply. For details on a specific oder nacj dem Betreib des Hochspannungsnetzteils
circuit, consult Spellman’s Engineering Department. vrhanden sind, entladen werden. Weitere Informationen
finden Sie der IEEE-Vorschrift Nr. 510-1983.
The CZE series power supply is basically an DC to DC
power converter. Within the power supply, conversions
of DC to high frequency AC, then to high voltage DC take
place. WARNING
Typical CZE series power supplies comprise a few basic To reduce the risk of fire, replace fuse with same type and
building blocks. These are: 1) High frequency quasi- rating.
resonant inverter, 2) High voltage transformer and
rectifier circuits, and 3) Control and monitoring circuits.
The following is a brief description of each building
block. WARNING
Um die Brandgefahr zu verringern, muss die Sicherung
WARNING durch eine neue gleichen Typs ersetzt werden.
The energy levels used and generated by the
power supply can be lethal! Do not attempt to
operate the power supply unless the user has a 4.1 Chassis
sufficient knowledge of the dangers and hazards
of working with high voltage. Do not attempt to The CZE is a compact, high efficiency, high voltage
approach or touch any internal or external power supply. The power supply can supply up to 50
circuits or components that are connected or watts of DC power. (Output power capability may be
have been connected to the power supply. Be higher or lower depending upon model ordered). Output
voltages of up to 65kV can be generated.
certain to discharge any stored energy that may
be present before and after the power supply is 4.2 High Frequency Inverter
used. Consult IEEE recommended practices for
safety in high voltage testing #510-1983. The CZE series uses a resonant converter operating in a
zero current switching, resonant flyback and forward
converter topology. MOSFET transistors switch DC
WARNUNG voltage to the resonant circuit. Typical operating
frequency is in the range of 20-35KHz depending on
Die verwendete und erzeugte Engergie des model. Control of the resonant circuit output is done by
Hochspannungsnetzteiles kann tötlich sein! Betreiben Sie the low voltage control circuits. The resonant circuitry
das Hochspannungsnetzeil nur wenn Sie ausreichendes consists of a resonant inductor and step-up transformer.

CZE MANUAL 15 118030-001 REV D


4.3 High Voltage Circuits 4.4 Control Circuits
The high voltage transformer is a step-up type. The Control circuits are used for regulation, monitoring, pulse-
secondary of the high voltage transformer is connected to width, control, slow-start and inhibit control. Feedback
the high voltage rectifier circuit. The rectifier circuit will signals are calibrated and buffered via general purpose
vary depending upon the rated output voltage. For lower OP-AMPS. Pulse width control is accomplished by a
output voltage, a full bridge or doubler circuit is used. typical PWM type control I.C. Logic enable/disable is
For higher voltages, a half wave Cockroft-Walton provided by a logic gate I.C. Regulators generate +/- 15V
multiplier is used. A feedback signal is generated by the and 10VDC.
high voltage resistor divider. This feedback signal is sent
to control circuits to provide voltage regulation and 4.5 Polarity Reversal Circuits
monitoring. A current sense resistor is connected at the
The CZE1000 and CZE2000 polarity reversal circuits
low voltage end of the rectifier circuit. The circuit sense
operate as follows: A polarity reversal command is given;
signal is sent to the control circuits to provide current
at this point the output begins to decay. At a safe level a
regulation and monitoring.
high voltage relay switches to the opposite polarity output
In the SMS series, an auxiliary voltage sense signal is section and ramps to its preset level.
generated by the overvoltage protection feedback resistor.
This is used to provide overvoltage protection. 4.6 Options
The high voltage rectifier output is connected to an R-C Due to the variations of models and options provided in
type filter to reduce high frequency ripple components. the CZE series, details of actual circuits used may differ
The filter is then connected to the output limiting resistors. slightly from above descriptions. Consult Spellman’s
These resistors limit the peak surge current in the event an Engineering Department for questions regarding the
arc or discharge occurring. The limiting resistor output is principles of operations for the CZE series.
connected to the output cable or connector provided.

CZE MANUAL 16 118030-001 REV D


Chapter 5

OPTIONS
T he options available for this power supply
described in this section. Interface diagrams
shown where required. Options are specified
are
are
by
The no slow start option causes the output voltage of the
power supply to rise (within approximately 500 msec) to
the rated voltage upon Power Up.
including the option code in the model number as
described in Section 1.4. 5.4 Non-Standard Slow Start
5.1 Floating Ground FG SS(x)
The non-standard slow start option allows the gradual rise
The floating ground option allows isolation of the power time of the output voltage to be different from the
supply common from chassis ground. A typical standard. To order the option, place the time desired in
application for FG is for connection of an external meter seconds after the suffix letter; i.e. SS(10) denotes a 10
or circuit into the FG signal point. This circuit can second rise time. (THESE TIMES ARE
monitor actual current flow at a low voltage level. See APPROXIMATE).
Figure 5.1 for typical connection to the floating ground
terminal.
5.5 Extra Length Output Cable
IMPORTANT LL(ft)
Standard output cable is 18” long. Other lengths may be
The power supply common must be specified.
connected to chassis ground through a
low impedance circuit. The power 5.6 Auxiliary Outputs AO
supply common is clamped to chassis
The specific AUXILIARY output requirements should be
ground using a power zener diode. This discussed with Spellman’s Sales Department..
diode is not intended to clamp under
continuous operation. On higher 5.7 Custom Designed Models X(#)
current units the power dissipation
Units built to customer specifications are assigned an X
within the zener diode can be excessive
number be the factory. If this unit is an X model,
if allowed to clamp. specification control sheet is added at the end of this
instruction manual.
5.2 Input Voltage IV
CZE power supplies with different input voltage options Spellman welcomes the opportunity to tailor its units to fit
will operate from the input voltage specified plus or minus your requirements or to develop new products for your
a tolerance, typically ±10%. applications. Contact the Spellman Sales Department
with your needs.
5.3 No Slow Start NSS

CZE MANUAL 17 118030-001 REV D


Figure 5.1 Floating Ground Interface Circuit

CZE MANUAL 18 118030-001 REV D


Chapter 6

MAINTENANCE
T his section describes periodic
performance testing procedures.
servicing and
WARNING

WARNING HOCHSPANNUNG IST GEFÄHRLICH!

THIS POWER SUPPLY GENERATES VOLTAGES NUR QUALIFIZIERTES PERSONAL DARF DIESE
THAT ARE DANGEROUS AND MAY BE FATAL. PRÜFUNGEN DURCHFÜHREN.

OBSERVE EXTREME CAUTION WHEN WORKING High voltage test procedures are described in Bulletin
WITH HIGH VOLTAGE. STP-783, Standard Test Procedures for High Voltage
Power Supplies. Copies can be obtained from the
Spellman Customer Service Department. Test equipment,
WARNUNG including an oscilloscope, a high impedance voltmeter,
DIESES HOCHSPANNUNGSNETZTEIL ERZEUGT and a high voltage divider such as the Spellman HVD-100
LEBENSGEFÄHRLICHE HOCHSPANNUNG! or HVD-200, is needed for performance tests. All test
components must be rated for operating voltage.
SEIEN SIE SEHR VORSICHTIG BEIM ARBEITEN
MIT HOCHSPANNUNG!
6.3 High Voltage Dividers
High voltage dividers for precise measurements of output
6.1 Periodic Servicing voltage with an accuracy up to 0.1% are available from
Spellman. The HVD-100 is used for voltages up to
The CZE series does not require periodic maintenance or 100KV. The HVD-200 measures up to 200KV. The
servicing. Spellman divider is designed for use with differential
voltmeters or high impedance digital voltmeters. The high
6.2 Performance Test input impedance is ideal for measuring high voltage low
current sources, which would be overloaded by traditional
WARNING lower impedance dividers

HIGH VOLTAGE IS DANGEROUS.


ONLY QUALIFIED PERSONNEL SHOULD
PERFORM THESE TESTS.

CZE MANUAL 19 118030-001 REV D


Chapter 7

REPLACEMENT PARTS
7.1 Replacement Parts 7.2 Correspondence And
Ordering Spare Parts
S Pellman provides parts and assemblies for its high
voltage power supplies but recommends that only
qualified personnel perform the repair. High voltage
Each power supply has an identification label on the rear
of the chassis that bears its model and serial number.
is dangerous; even minor mistakes in repairs can have When requesting engineering or applications information,
serious consequences. please state the model and serial number of the power
supply. If specific components or circuit sections are
When requesting parts please give the model number and involved in the inquiry, it is helpful to indicate the
serial number of the power supply. component symbol number(s) shown on the applicable
schematic diagram.

When ordering spare parts, please specify the part’s


description, the part’s reference designation or part
number, and the model and serial number of the unit.

CZE MANUAL 20 118030-001 REV D


Chapter 8

FACTORY SERVICE
8.1 Warranty Repairs A preliminary estimate for repairs will be given by phone
During the Warranty period, Spellman will repair all units by Customer Service. A purchase order for this amount is
free of charge. The Warranty is void if the unit is worked requested upon issuance of the RMA Number. A more
on by other than Spellman personnel. See the Warranty in detailed estimate will be made when the power supply is
the rear of this manual for more information. Follow the received at the Spellman Repair Center. In the event that
return procedures described in Section 8.2. The customer repair work is extensive, Spellman will call to seek
shall pay for shipping to and from Spellman. additional authorization from your company before
completing the repairs.
8.2 Factory Service Procedures
8.3 Ordering Options And
Spellman has a well-equipped factory repair department.
If a unit is returned to the factory for calibration or repair, Modifications
a detailed description of the specific problem should be Many of the options listed in Chapter 5 can be retrofitted
attached. into Spellman power supplies by our factory. For prices
For all units returned for repair, please obtain an and arrangements, contact the Spellman Sales
authorization to ship from the Customer Service Department.
Department, either by phone or mail prior to shipping.
When you call, please state the model and serial numbers, 8.4 Shipping Instructions
which are on the plate on the rear of the power supply, All power supplies returned to Spellman must be sent
and the purchase order number for the repair. A Return shipping prepaid. Pack the units carefully and securely in
Material Authorization Code Number (RMA Number) is a suitable container, preferably in the original container, if
needed for all returns. This RMA Number should be available. The power supply should be surrounded by at
marked clearly on the outside of the shipping container. least four inches of shock absorbing material. Please
Packages received without an RMA Number will be return all associated materials, i.e. high voltage output
returned to the customer. The Customer shall pay for cables, interconnection cables, etc., so that we can
shipping to and from Spellman. examine and test the entire system.
All correspondence and phone calls should be directed to:
Spellman High Voltage Electronics Corp.
475 Wireless Boulevard
Hauppauge, New York 11788
TEL: (631) 630-3000 FAX: (631) 435-1620
E-Mail: [email protected]
http://www.spellmanhv.com

CZE MANUAL 21 118030-001 REV D


SPELLMAN HIGH VOLTAGE ELECTRONICS
WARRANTY
Spellman High Voltage Electronics (“Spellman”) warrants that all power supplies it manufactures will be
free from defects in materials and factory workmanship, and agrees to repair or replace, without charge, any
power supply that under normal use, operating conditions and maintenance reveals during the warranty
period a defect in materials or factory workmanship. The warranty period is twelve (12) months from the
date of shipment of the power supply. With respect to standard SL power supplies (not customized) the
warranty period is thirty-six (36) months from the date of shipment of the power supply.

This warranty does not apply to any power supply that has been:
• Disassembled, altered, tampered, repaired or worked on by persons unauthorized by Spellman;
• subjected to misuse, negligent handling, or accident not caused by the power supply;
• installed, connected, adjusted, or used other than in accordance with the original intended application and/or
instructions furnished by Spellman.

THE FOREGOING WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING
THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.

The buyer’s sole remedy for a claimed breach of this warranty, and Spellman’s sole liability is limited, at
Spellman’s discretion, to a refund of the purchase price or the repair or replacement of the power supply at
Spellman’s cost. The buyer will be responsible for shipping charges to and from Spellman’s plant. The
buyer will not be entitled to make claim for, or recover, any anticipatory profits, or incidental, special or
consequential damages resulting from, or in any way relating to, an alleged breach of this warranty.

No modification, amendment, supplement, addition, or other variation of this warranty will be binding unless
it is set forth in a written instrument signed by an authorized officer of Spellman.

Factory Service Procedures


For an authorization to ship contact Spellman’s Customer Service Department. Please state the model and
serial numbers, which are on the plate on the rear panel of the power supply and the reason for return. A
Return Material Authorization Code Number (RMA number) is needed from Spellman for all returns. The
RMA number should be marked clearly on the outside of the shipping container. Packages received without
an RMA Number may delay return of the product. The buyer shall pay shipping costs to and from Spellman.
Customer Service will provide the Standard Cost for out-of-warranty repairs. A purchase order for this
amount is requested upon issuance of the RMA Number (in-warranty returns must also be accompanied by
a “zero-value” purchase order). A more detailed estimate may be made when the power supply is received
at Spellman. In the event that the cost of the actual repair exceeds the estimate, Spellman will contact the
customer to authorize the repair.

Factory Service Warranty

Spellman will warrant for three (3) months or balance of product warranty, whichever is longer, the repaired
assembly/part/unit. If the same problem shall occur within this warranty period Spellman shall undertake all
the work to rectify the problem with no charge and/or cost to the buyer. Should the cause of the problem be
proven to have a source different from the one that has caused the previous problem and/or negligence of
the buyer, Spellman will be entitled to be paid for the repair.

Spellman Worldwide Service Centers

For a complete listing of Spellman’s Global Service facilities please go to:


http://www.spellmanhv.com/customerservice/service.asp

101520-007 REV D

You might also like