Amsco 600 Series Steam Sterilizers - : Application

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

AMSCO® 600 SERIES STEAM STERILIZERS

(HEALTHCARE) — NORTH AMERICA USE ONLY

APPLICATION
AMSCO 600 Series Steam Sterilizers are configured for
prevacuum and gravity sterilization of heat and moisture-stable
medical devices and their accessories used in healthcare
facilities.

DESCRIPTION
AMSCO 600 Series Steam Sterilizers are equipped with the
latest features in both state-of-the-art technology and ease of
use. Sterilizer is designed to be completely serviced from the
front of the sterilizer, minimizing the footprint.

Chamber sizes – AMSCO 600 medium steam sterilizers are


available in three sizes, with interior chamber dimensions
(W x H x L) and capacities as follows:

• 26.5 x 26.5 x 39" (673 x 673 x 990 mm); 450L


AMSCO 600 Series Steam Sterilizer
• 26.5 x 26.5 x 51" (673 x 673 x 1295 mm); 588L
• 26.5 x 26.5 x 63" (673 x 673 x 1600 mm); 724L (Typical - details may vary)

U-1 is furnished. The shell and door are constructed to


Clean steam-ready – All AMSCO 600 series sterilizers include withstand a working pressure of 45 psig (3.1 bar).
a fully-jacketed chamber with stainless piping for steam
delivery. FEATURES

Vertical-sliding door(s) – Pneumatically-driven doors are Chamber and Jacket


operated from a button on the touch screen; and travel down The AMSCO 600 series sterilizer chamber has a 26.5 x 26.5"
vertically to open. (673 x 673 mm) cross section and is sized to allow for efficient
high-volume processing of sterilization containers, trays and
Control – includes a STERIS micro-processor based control packs.
system featuring a touch-sensitive screen. See pages 2 and 3 Fully jacketed chamber construction reduces internal
for details. condensation. The chamber is manufactured from AISI 316L
stainless steel, as is the chamber door; and the full structural
STANDARDS jacket is manufactured from AISI 304 stainless steel.
Each AMSCO 600 sterilizer meets applicable requirements of Pressure Vessel
the following listings and standards, and carries the appropriate All pressure vessels are designed and manufactured in
symbols: accordance to the required pressure vessel standards. An
identification nameplate is permanently mounted on each
• Underwriters Laboratory (UL) Standard 61010-1 pressure vessel. Features include:
as certified by Intertek (ETL Registered). • Chamber steam baffle provides uniform steam distribution
• ASME Code, Section VIII, Division 1 for unfired pressure to the chamber.
vessels. The pressure vessel is so stamped; ASME Form • Chamber and jacket are provided with safety relief devices.

Selections Checked Below Apply To This Equipment


AMSCO 600 SERIES MODELS STEAM SOURCE CONNECTIVITY
❑ 939V-1; 39" (990 mm) ❑ Facility Steam ❑ ConnectAssure™
❑ 1251V-1; 51" (1295 mm) ❑ Stainless-Steel Integral Electric Steam Generator ❑ CS-iQ®
❑ 1563V-1; 63" (1600 mm) STERILIZER ELECTRICAL SERVICE LOADING ACCESSORIES
SINGLE DOOR MOUNTING ❑ 208 - 240 V, 60 Hz Sterilizer Manual Loading - refer to Tech Data 10304914
❑ Recessed ❑ 440 - 480 V, 60 Hz Sterilizer ❑ Loading car and transfer carriage
❑ Cabinet enclosed/free-standing ❑ Include Service for Integral Electrical Steam Generator ❑ Rack and shelf (39" Chamber Length Unit Only)
(includes Top Panel) 208 - 240 V, 50/60 Hz ❑ Chamber rails
❑ Rear Cabinet Panel (Option) ❑ Include Service for Integral Electrical Steam Generator ❑ Loading car extraction hook
440 - 480 V, 50/60 Hz (Telescopic: adjustable length: 47-71" (1041-1803 mm)
DOUBLE DOOR MOUNTING (Recessed)
Item ________________________
❑ Through two walls
❑ Through one wall ❑ AIR COMPRESSOR Location(s)___________________
____________________________
10304921 Rev B
• Chamber bottom is sloped to promote drainage of finish in accordance with American Society for Testing and
condensate. Materials (ASTM) (EN 10088-2: 2G).
• Pressure vessel and doors are insulated with 30 mm of • Fascia panels are easy to remove to save time and effort
mineral wool (R-Value = 3.3 hr.ft2.F/BTU). Jacket is covered during maintenance.
with rigid aluminum cladding. Door is covered with stainless- Piping
steel cladding.
AMSCO 600 series sterilizers include four piping subsystems;
• Chamber drain port is provided with removable debris one for each utility supply and one for drainage as follows:
screen for easy cleaning.
• Each pressure vessel is hydrostatic pressure tested. • Steam • Compressed Air • Water • Chamber Drain
• Chamber is designed to operate from full vacuum to a
Manual-shutoff ball valves located on water, steam and air lines
maximum allowable working pressure of 45 psig
isolate the respective utilities when required.
(3.1 bar g).
Steam supply piping and associated valves are made of
• Jacket is designed for a maximum working pressure of
stainless steel for clean steam.
36 psig (2.5 bar g) internal pressure.
Two-stage Vacuum Pump provides quiet and efficient
• Pressure vessel penetrations include compression and operation and deep vacuum for better drying; and results in
Tri Clamp fittings, where applicable.
lower water consumption than is seen with water ejector
• Sterilizer models 1251V-1 and 1563V-1 (the 51 and 63" deep systems.
chamber units) use removable 316L stainless-steel Sensors for Temperature and Pressure
chamber rails. The rails support the loaded car within the
chamber. AMSCO 600 series sterilizers include sensors for independent
control and recording of temperature per AAMI ST8.
• Sterilizer model 939V-1 (the 39" deep chamber unit) can be Additionally, one sensor is used for controlling and recording
equipped with rails for use with loading car and transfer
pressure.
carriage, or with rack and shelves for scenarios where floor
space is limited. Effluent Cooling
Drain effluent discharge is cooled with a mixing tank. Water
Door Design
usage is minimized through thermal detection.
AMSCO 600 Series Steam Sterilizers are available with one or Sterile Air Filter
two vertical sliding doors. Each door is sealed to chamber by
a steam-activated gasket which requires no lubrication. Air entering the chamber is filtered through a 0.2-micron
hydrophobic sterilizer air filter.
• Chamber and jacket pressure gauges are provided on the
front face (non-sterile side) of the sterilizer. A chamber Steam Pressure Regulating Valve (PRV)
pressure gauge is provided on the sterile side of double PRV regulates incoming steam pressure to within the
door units. 42 ±1 psig (2.9 bar g) limit.
• Touch-sensitive safety edge stops doors from closing in Pneumatic Valves
case of obstruction. System pneumatic valves are designed and constructed for
• Control lock-out switch prevents cycle from starting if door long life and low maintenance.
is not fully closed.
MOUNTING ARRANGEMENT
• Door interlocks, on double door units, ensure only one door
can be opened at a time. The sterilizer is designed for either freestanding or recessed
Electrical Design installation. Each sterilizer is height adjustable.
Electrical design includes the following features: AISI 201/AISI 304 stainless-steel side panels and vented top
panel [and optional back panel] are designed to enclose both
• Standard configuration requires one 3-phase connection.
sides, top [and back] of a single door sterilizer to conceal
• Design ensures fail-safe operation in event of an operational internal piping and mechanical and electrical components.
abnormality. Recessed-mounted sterilizers are provided with a trim kit to
• Emergency stop (E-STOP) present by each door for critical isolate the service area and assist with air management.
situations. When pushed, the E-STOP turns off all outputs
from the control and triggers an emergency stop pressed CONTROL SYSTEM
alarm. When released, an emergency stop released alarm
occurs. If in-cycle, an abort screen displays, prompting Design Features
operator to abort the cycle; otherwise the sterilizer returns The proprietary AMSCO 600 control system consists of a real
to an idle condition. time operating system, microcomputer control boards, and
peripheral function circuit boards. An Internal flash memory
Panel and Frame Assembly
stores all cycle memory.
The panel and frame assembly supports pressure vessel,
piping, electrical components and fascia panels. The cycles available on the AMSCO 600 Steam Sterilizer
facilitate efficient processing. The unit offers 35 user-
• Frame is a fully welded structure fabricated from carbon programmable† cycle options, including prevacuum and
steel and painted with corrosion and heat resistant epoxy
gravity cycles designed for steam sterilization of various types
paint. Frame includes adjustable stainless-steel legs for
of products such as instruments (in trays) and rigid containers,
leveling the unit.
textiles, and pouched items. The AMSCO 600 Steam Sterilizer
• Sterilizer is designed and manufactured to allow access for also offers standard test cycles (Warm-up, Bowie-Dick/air
maintenance and service from loading end. removal and leak test). All processing cycles offer user-
• Loading and unloading ends of sterilizer are covered with programmable options, allowing operators to customize
AISI 304 stainless steel fascia panels polished to a No. 3 sterilize times above AAMI recommended minimums,

2
temperatures within cleared 270-275°F (132-135°C) range and • Temperature display and printout can be set to Fahrenheit
dry times from 0-120 minutes to meet various instrument (°F) or Celsius (°C) and displayed/printed to nearest 1°
manufacturer IFUs‡. AMSCO 600 Medium Steam Sterilizers increment.
come standard with a 2-stage electric vacuum pump that • Pressure/vacuum display and printout units can be set
provides a fast, deep vacuum for more efficient drying. to PSIA, PSIG, mbar, or kPa.
† This consists of six factory-set cycles, three test cycles and
26 additional cycles that a Customer can add to meet specific HEALTHCARE CYCLES
processing requirements.
Description
‡ Cycle parameters cannot be adjusted to less than preset, AMSCO 600 Series Steam Sterilizers are provided with
validated settings. prevacuum and gravity cycles designed for steam sterilization
of various types of products such as instruments (in trays) and
NOTE: An under-temperature or over-temperature condition rigid containers, textiles, and products in pouches.
causes the controller to abort the cycle. Pre-Set Cycles
Operator interface control panel, consisting of a touch screen AMSCO 600 is pre-configured with the following cycles:
display and thermal printer, is located on the operating end of
the sterilizer. Pre-set Cycles
• Touch-sensitive screen is a 8.4" (213.4 mm) 8-bit color
graphics display located at an ergonomic height, and easily Sterilize Dy
Cycle Type and Use Temperature
Time Time
accessible to the operator on the right side of the sterilizer.
The control touch screen, from which all sterilizer functions Prevacuum Minutes
are controlled, features a wide viewing angle and high-
visibility backlighting. Instrument Trays 270°F (132°C) 4 30
The display indicates any abnormal conditions that may Single Pack Fabric 270°F (132°C) 4 5
exist, either in or out of cycle. Displayed messages are
complete phrases with no codes to be cross-referenced. Full Load Fabric 270°F (132°C) 4 20
• Thermal printer – High Resolution (8 dots/mm) thermal
Immediate Use 270°F (132°C) 4 1
printer is fast and quiet, printing 25 lines/minute on industrial
grade thermal paper. Printout has a guaranteed shelf-life of Instrument Trays 275°F (135°C) 3 30
25 years and is resistant to exposure from steam, alcohol,
UV and visible light, oil, heat and water. On double-door Gravity
sterilizers, the printer is located on the sterile (unload) side. Instrument Trays 250°F (121°C) 30 30
• Cycle Storage / Cycle Reprint – Sterilizer stores up to the
Test
last 300 cycles on internal memory and alerts the operator
when 270 cycles are saved. These internally stored cycles Warm-up 270°F (132°C) 3 1
can be retrieved, viewed, printed or copied to a USB stick.
The sterilizer includes a USB port below the E-Stop on both Bowie-Dick/Air Removal 270°F (132°C) 3.5 1
the OE and NOE panels. When a port is occupied with a Leak Test 15 Minute Leak Hold Period
USB stick, cycle log data can be saved to the stick.
• Sterile (Unload Side) control panel, located on double- Configurable Cycles
door sterilizers only, includes a full-size touch-sensitive
Configurable operating parameters enable the Customer to
screen identical to the operating end screen.
develop and optimize cycles for a variety of applications. All
Preprogrammed cycles can be started from the sterile side
cycles consist of three separate phases as follows:
control panel. The display concurrently shows the same
information as the sterilizer loading end screen display. • Conditioning phase: this phase removes air from the
chamber prior to the exposure phase. Prevacuum and
Cycle configuration is performed by accessing the change Gravity-displacement Air Removal Cycles are available.
values menu through the operating end touch screen. The After air removal, the chamber heats up to reach the defined
change values menu can be used to adjust cycle values and exposure temperature.
the following operating parameters:
• Exposure phase: The standard process operates with time
• Time display and printout units – AM/PM or 24-hour.
based exposure. The timer runs only when the chamber
• Access code – a four-digit access code is required to reaches and maintains the pre-selected exposure
change cycle values. temperature.
• Sound options – touch screen control includes buttons for • Drying phase: The drying phase uses vacuum to evaporate
selecting a volume level of Low, Medium, or High for: moisture from the chamber and load.
> Touch Tone
> End of Cycle Sound Test Cycles
> Alarm Sound AMSCO 600 sterilizers are pre-programmed with the following
three test cycles.
Each of the three signals can be independently adjusted,
for example, the touch tone can be set to Low, End of Cycle • Warm-Up Cycle is run to heat the sterilizer chamber and
to Medium, and Alarm Sound to High. jacket to set temperature. The operator typically runs a warm
up cycle prior to running a Bowie-Dick/Air Removal Test
A Cycle End Duration button allows the user to set a value Cycle or Leak Test.
within the range of 0 to 90 seconds, determining how long
the End of Cycle Sound will continue to signal.

3
• Bowie-Dick Test Cycle is used to test prevacuum cycles entire perioperative loop achieves higher levels of
for verification of effective air removal in chamber and load, performance.
in other words, the test cycle determines whether the
vacuum system is operating properly so the sterilizer can CS-iQ Sterile Processing Workflow Management Software
be used for pre-vacuum cycles. The cycle is programmed Features:
with pre-set parameters and should be run daily. • Real-time information available hospital-wide to make the
Test packs should be placed in an empty chamber directly right decision at the right time.
over the drain on the bottom rack or shelf. • Paper reduction by managing documentation and using
• Leak Test is a standard test provided for verification of electronic record keeping, including capture of biological
chamber and piping integrity and should be run weekly, with and chemical indicator results.
results recorded.The test uses a 15 minute leak hold period. • Viewing and printing of equipment status reports including
equipment monitoring results and daily maintenance
STEAM SOURCE OPTIONS notifications
• Tracking of functionality and uptime of CSSD equipment.
Integral Electric Steam Generator (45 kW)
• Proactive management through real-time visual dashboards
The optional 45 kW Integral Electrical Steam Generator is
designed to produce clean steam. To prevent generator with reminders for preventative maintenance and daily
malfunction, feed water must be either deionized or RO, per service tasks for CSSD equipment.
ASME Pressure Vessel Code 5.3.
The system is located beneath the sterilizer chamber and ACCESSORIES
includes feed water system, steam generation system and
controls. All generator components can be completely serviced Manual Loading Accessories
from the front of the sterilizer. Manual loading accessories for AMSCO 600 series sterilizers
include the following:
The Integral Electrical Steam Generator is provided with the
following features: • Chamber rails
• The pressure vessel meets the ASME pressure vessel code. • Racks and shelves (39" Chamber Length Units only)
• The AISI 316L stainless-steel vessel is rated for 45 psig • Loading cars
(3.1 bar g) per code requirements and stamped. • Loading car shelves
• The heating element contains stainless-steel AISI 316L • Loading car extraction hook
tubular elements. • Transfer carriages (fixed and variable height)
• The unit is equipped with two float switch style water level • PouchCare (Rack for peel pouches)
sensors.
Refer to the tech data, Manual Loading Accessories for
Feed water booster pump – included on all AMSCO 600 AMSCO 600 Series Medium Steam Sterilizers (p/n 10304914),
sterilizers that are equipped with the integral steam generator for details pertaining to the above items. For details pertaining
option. The booster pump maintains the flow of feed water at specifically to PouchCare, refer to marketing document
a rate sufficient to ensure efficient operation of the sterilizer. M3557EN.

Steam Generator Bypass Valve


A ball valve permits the quick exchange of steam source from PREVENTIVE MAINTENANCE
the integral generator to a stand-alone generator or facility
steam source. Preventive Maintenance programs designed to maximize
equipment uptime can be tailored to meet Customer
CONNECTIVITY OPTIONS requirements. A typical contract provides preventive
maintenance site visits twice per year. This maintenance is
performed by factory trained STERIS Technicians following
AMSCO 600 Sterilizers are equipped for the following
connectivity options. documented preventive maintenance procedures to inspect
and adjust the equipment. The exact Customer requirements
ConnectAssure can be determined by a visit with the STERIS District Service
ConnectAssure Technology™ is an electronic record Manager before the equipment arrives.
management system that collects cycle information from
STERIS equipment to be displayed in an SPD (Sterile A maintenance-training program can be designed to provide
Processing Department) tracking system or in the training on preventive maintenance and first response
ConnectAssure software interface. troubleshooting. Training can be scheduled by contacting
STERIS.
Note: ConnectAssure Data Export is required for the electronic
transmission of cycle data over the network to a SPD tracking
system. For additional information refer to ConnectAssure UTILITY REQUIREMENTS
Technology™ tech data 10515801.
Refer to the following tables for basic information pertaining to
required utilities, connection types, and consumption.
CS-iQ®
Sterile Processing Workflow Management Software
The three modules of CS-iQ software promote productivity, NOTE: For detailed information always refer to associated
traceability and compliance across a department. Functionality equipment drawings.
is added with each module, benefits increase, and the facility
moves from reactive management to proactive planning. The

4
ENGINEERING DATA / UTILITY REQUIREMENTS

Drain 1-1/2" ODT drain terminal (Gravity drain must be able to handle peak water capacity of 15 gpm (57 LPM))

Control and Vacuum Pump

Electric 208-240 Volt, 3-phase, 60 Hz, 10 Amps, max. protection 20 Amps. #12 [3.31] AWG

440-480 Volt, 3-phase, 60 Hz, 7 Amps, max. protection 20 Amps. #12 [3.31] AWG

Size: 3/4" NPT stainless steel fitting.


Facility Steam
Pressure: 50 to 80 psig (3.45 to 5.5 bar), dynamic, condensate free, and 97 to 100% vapor quality.

Steam Consumption Average Peak Sterilizer


Chamber Volume
270°F (132°C) Cycle Consumption Consumption Operating Wt.
Model
Inches (mm) lb (kg) / Cycle lb (kg) / hr lb (kg) Liter Cubic Feet Cubic Inches

26.5 x 26.5 x 39
35 (16) 310 (141) 2515 (1141) 939V-1 448.8 15.8 27388
(673 x 673 x 991)

26.5 x 26.5 x 51
45 (20) 310 (141) 2850 (1293) 1251V-1 586.9 20.7 35815
(673 x 673 x 1295)

26.5 x 26.5 x 63
55 (25) 310 (141) 3185 (1445) 1563V-1 725.0 25.6 44242
(673 x 673 x 1600)

Size: 3/4" NPT. Pressure: 20 to 50 psig (1.4 to 3.5 bar), dynamic.


Feed Water
Temperature: 40-70°F (4-21°C). Peak Flow (all units): 12 gpm (45 Lpm)

Average Standard Water


Consumption Consumption (Average)

Inches (mm) gal (L) / cycle

26.5 x 26.5 x 39 97 (367)


(673 x 673 x 991)

26.5 x 26.5 x 51 112 (424)


(673 x 673 x 1295)

26.5 x 26.5 x 63 128 (484)


(673 x 673 x 1600)

Compressed Air 1/2" NPT 80-100 psig (5.5 to 6.9 bar psig), dry, oil free, 3 scfm.

1/2" NPT; hot water temperature 70-140°F 21-(60°C). Generator Drain: 1/2" NPT
RO / DI Water
Model 939V-1 4.2 gal/cycle (16 L/cycle)
General
information and Model 1251V-1 5.4 gal/cycle (20 L/cycle)
consumption
Model 1563V-1 6.6 gal/cycle (25 L/cycle)
Optional Stainless
Steel Integral 208-220 VAC, 50/60 Hz, 125 Amps, max. protection 170 Amps, 3-phase minimum, #1 [42.4] AWG
Steam Generator 230-240 VAC, 50/60 Hz, 113 Amps, max. protection 145 Amps, 3-phase minimum, #2 [33.6] AWG
(45 kW) Electric
380-415 VAC, 50/60 Hz, 59 Amps, max. protection 95 Amps, 3-phase minimum, #6 [13.3] AWG

440-480 VAC, 50/60 Hz, 57 Amps, max. protection 75 Amps, 3-phase minimum, #6 [13.3] AWG

Boiler Weight (Empty): 237 lb (108 kg)


Boiler Weight
Boiler Weight (Operating): 285 lb (129 kg)

5
ENGINEERING DATA / UTILITY REQUIREMENTS

Additional Noise Level 64 dBa


Specification

SAFETY RELIEF VALVE PARAMETERS


Safety Relief Valve PRESSURE TEMPERATURE CONNECTION

45 PSIG max
CHAMBER 293°F (145°C) 1" NPT
3.1 bar(g) max

36 PSIG
JACKET 282°F (139°C) 1" NPT
2.5 bar(g) max

OPTIONAL INTEGRAL
45 PSIG
ELECTRIC STEAM 293°F (145°C) 1" NPT
3.1 bar(g) max
GENERATOR (ESG)

CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH APPLICABLE LOCAL AND NATIONAL CODES
AND REGULATIONS.

6
NOT FOR INSTALLATION*
Total height: Upper Panel
89-1/16" (2263 mm) (cut-away added)

Utility Connections
12"
(304 mm)

Chamber Pressure
Human Machine Gauge
Interface (HMI)
Touchscreen Jacket Pressure
Gauge

Emergency Stop
77"
(1950 mm)

Thermal Printer
Chamber Door‡

Integral Steam Generator


Fascia Panel (Option)
(1 of several)

D†
39"
1000 mm

† Sterilizer Dimension D varies according as follows:


Chamber Length Dimension D
39" 53" (1342 mm)
51" 65" (1642 mm) Note: Power Box and Control Box not shown, reside behind upper panel.
Enclosure side, top and rear panels not shown.
63" 77" (1942 mm)
Depths shown are for double door sterilizers. ‡Vertical Sliding Chamber Door (on both sides for double-door units)

AMSCO 600 Series Sterilizer

*Refer to applicable equipment drawings for installation details.

7
Associated Documents
Document Title Document Number

AMSCO 600 MAINTENANCE MANUAL Dwg. No. 10330608

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10307698
SLIDING, SD, RECESSED WALL, STEAM HEAT (p/n 11001835)

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10336741
SLIDING, SD, CABINET, STEAM (p/n 11001836)

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10336576
SLIDING, DD, RECESSED 1 WALL, CABINET, STEAM HEAT (p/n 11001837)

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10308020
SLIDING, DD, RECESSED 2 WALLS, STEAM HEAT (p/n 11001829)

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10802437
SLIDING, DD, RECESSED 1 WALL, CABINET, STEAM GENERATOR (p/n 11008536)

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10786670
SLIDING, DD, RECESSED 2 WALLS, STEAM GENERATOR (p/n 11008534)

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10778442
SLIDING, SD, CABINET, STEAM GENERATOR (p/n 11008531)

EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10749736
SLIDING, SD, RECESSED WALL, STEAM GENERATOR (p/n 11008529)

For further information, contact:

The base language of this document is ENGLISH.


STERIS Corporation Any translations must be made from the base
5960 Heisley Road language document.
Mentor, OH 44060-1834 • USA
440-354-2600 • 800-548-4873
www.steris.com

This document is intended for the exclusive use of STERIS Customers, including
architects or designers. Reproduction in whole or in part by any party other than
10304921 Rev B ©2019, STERIS Corporation. All rights reserved. a Customer is prohibited.