Amsco 600 Series Steam Sterilizers - : Application
Amsco 600 Series Steam Sterilizers - : Application
Amsco 600 Series Steam Sterilizers - : Application
APPLICATION
AMSCO 600 Series Steam Sterilizers are configured for
prevacuum and gravity sterilization of heat and moisture-stable
medical devices and their accessories used in healthcare
facilities.
DESCRIPTION
AMSCO 600 Series Steam Sterilizers are equipped with the
latest features in both state-of-the-art technology and ease of
use. Sterilizer is designed to be completely serviced from the
front of the sterilizer, minimizing the footprint.
2
temperatures within cleared 270-275°F (132-135°C) range and • Temperature display and printout can be set to Fahrenheit
dry times from 0-120 minutes to meet various instrument (°F) or Celsius (°C) and displayed/printed to nearest 1°
manufacturer IFUs‡. AMSCO 600 Medium Steam Sterilizers increment.
come standard with a 2-stage electric vacuum pump that • Pressure/vacuum display and printout units can be set
provides a fast, deep vacuum for more efficient drying. to PSIA, PSIG, mbar, or kPa.
† This consists of six factory-set cycles, three test cycles and
26 additional cycles that a Customer can add to meet specific HEALTHCARE CYCLES
processing requirements.
Description
‡ Cycle parameters cannot be adjusted to less than preset, AMSCO 600 Series Steam Sterilizers are provided with
validated settings. prevacuum and gravity cycles designed for steam sterilization
of various types of products such as instruments (in trays) and
NOTE: An under-temperature or over-temperature condition rigid containers, textiles, and products in pouches.
causes the controller to abort the cycle. Pre-Set Cycles
Operator interface control panel, consisting of a touch screen AMSCO 600 is pre-configured with the following cycles:
display and thermal printer, is located on the operating end of
the sterilizer. Pre-set Cycles
• Touch-sensitive screen is a 8.4" (213.4 mm) 8-bit color
graphics display located at an ergonomic height, and easily Sterilize Dy
Cycle Type and Use Temperature
Time Time
accessible to the operator on the right side of the sterilizer.
The control touch screen, from which all sterilizer functions Prevacuum Minutes
are controlled, features a wide viewing angle and high-
visibility backlighting. Instrument Trays 270°F (132°C) 4 30
The display indicates any abnormal conditions that may Single Pack Fabric 270°F (132°C) 4 5
exist, either in or out of cycle. Displayed messages are
complete phrases with no codes to be cross-referenced. Full Load Fabric 270°F (132°C) 4 20
• Thermal printer – High Resolution (8 dots/mm) thermal
Immediate Use 270°F (132°C) 4 1
printer is fast and quiet, printing 25 lines/minute on industrial
grade thermal paper. Printout has a guaranteed shelf-life of Instrument Trays 275°F (135°C) 3 30
25 years and is resistant to exposure from steam, alcohol,
UV and visible light, oil, heat and water. On double-door Gravity
sterilizers, the printer is located on the sterile (unload) side. Instrument Trays 250°F (121°C) 30 30
• Cycle Storage / Cycle Reprint – Sterilizer stores up to the
Test
last 300 cycles on internal memory and alerts the operator
when 270 cycles are saved. These internally stored cycles Warm-up 270°F (132°C) 3 1
can be retrieved, viewed, printed or copied to a USB stick.
The sterilizer includes a USB port below the E-Stop on both Bowie-Dick/Air Removal 270°F (132°C) 3.5 1
the OE and NOE panels. When a port is occupied with a Leak Test 15 Minute Leak Hold Period
USB stick, cycle log data can be saved to the stick.
• Sterile (Unload Side) control panel, located on double- Configurable Cycles
door sterilizers only, includes a full-size touch-sensitive
Configurable operating parameters enable the Customer to
screen identical to the operating end screen.
develop and optimize cycles for a variety of applications. All
Preprogrammed cycles can be started from the sterile side
cycles consist of three separate phases as follows:
control panel. The display concurrently shows the same
information as the sterilizer loading end screen display. • Conditioning phase: this phase removes air from the
chamber prior to the exposure phase. Prevacuum and
Cycle configuration is performed by accessing the change Gravity-displacement Air Removal Cycles are available.
values menu through the operating end touch screen. The After air removal, the chamber heats up to reach the defined
change values menu can be used to adjust cycle values and exposure temperature.
the following operating parameters:
• Exposure phase: The standard process operates with time
• Time display and printout units – AM/PM or 24-hour.
based exposure. The timer runs only when the chamber
• Access code – a four-digit access code is required to reaches and maintains the pre-selected exposure
change cycle values. temperature.
• Sound options – touch screen control includes buttons for • Drying phase: The drying phase uses vacuum to evaporate
selecting a volume level of Low, Medium, or High for: moisture from the chamber and load.
> Touch Tone
> End of Cycle Sound Test Cycles
> Alarm Sound AMSCO 600 sterilizers are pre-programmed with the following
three test cycles.
Each of the three signals can be independently adjusted,
for example, the touch tone can be set to Low, End of Cycle • Warm-Up Cycle is run to heat the sterilizer chamber and
to Medium, and Alarm Sound to High. jacket to set temperature. The operator typically runs a warm
up cycle prior to running a Bowie-Dick/Air Removal Test
A Cycle End Duration button allows the user to set a value Cycle or Leak Test.
within the range of 0 to 90 seconds, determining how long
the End of Cycle Sound will continue to signal.
3
• Bowie-Dick Test Cycle is used to test prevacuum cycles entire perioperative loop achieves higher levels of
for verification of effective air removal in chamber and load, performance.
in other words, the test cycle determines whether the
vacuum system is operating properly so the sterilizer can CS-iQ Sterile Processing Workflow Management Software
be used for pre-vacuum cycles. The cycle is programmed Features:
with pre-set parameters and should be run daily. • Real-time information available hospital-wide to make the
Test packs should be placed in an empty chamber directly right decision at the right time.
over the drain on the bottom rack or shelf. • Paper reduction by managing documentation and using
• Leak Test is a standard test provided for verification of electronic record keeping, including capture of biological
chamber and piping integrity and should be run weekly, with and chemical indicator results.
results recorded.The test uses a 15 minute leak hold period. • Viewing and printing of equipment status reports including
equipment monitoring results and daily maintenance
STEAM SOURCE OPTIONS notifications
• Tracking of functionality and uptime of CSSD equipment.
Integral Electric Steam Generator (45 kW)
• Proactive management through real-time visual dashboards
The optional 45 kW Integral Electrical Steam Generator is
designed to produce clean steam. To prevent generator with reminders for preventative maintenance and daily
malfunction, feed water must be either deionized or RO, per service tasks for CSSD equipment.
ASME Pressure Vessel Code 5.3.
The system is located beneath the sterilizer chamber and ACCESSORIES
includes feed water system, steam generation system and
controls. All generator components can be completely serviced Manual Loading Accessories
from the front of the sterilizer. Manual loading accessories for AMSCO 600 series sterilizers
include the following:
The Integral Electrical Steam Generator is provided with the
following features: • Chamber rails
• The pressure vessel meets the ASME pressure vessel code. • Racks and shelves (39" Chamber Length Units only)
• The AISI 316L stainless-steel vessel is rated for 45 psig • Loading cars
(3.1 bar g) per code requirements and stamped. • Loading car shelves
• The heating element contains stainless-steel AISI 316L • Loading car extraction hook
tubular elements. • Transfer carriages (fixed and variable height)
• The unit is equipped with two float switch style water level • PouchCare (Rack for peel pouches)
sensors.
Refer to the tech data, Manual Loading Accessories for
Feed water booster pump – included on all AMSCO 600 AMSCO 600 Series Medium Steam Sterilizers (p/n 10304914),
sterilizers that are equipped with the integral steam generator for details pertaining to the above items. For details pertaining
option. The booster pump maintains the flow of feed water at specifically to PouchCare, refer to marketing document
a rate sufficient to ensure efficient operation of the sterilizer. M3557EN.
4
ENGINEERING DATA / UTILITY REQUIREMENTS
Drain 1-1/2" ODT drain terminal (Gravity drain must be able to handle peak water capacity of 15 gpm (57 LPM))
Electric 208-240 Volt, 3-phase, 60 Hz, 10 Amps, max. protection 20 Amps. #12 [3.31] AWG
440-480 Volt, 3-phase, 60 Hz, 7 Amps, max. protection 20 Amps. #12 [3.31] AWG
26.5 x 26.5 x 39
35 (16) 310 (141) 2515 (1141) 939V-1 448.8 15.8 27388
(673 x 673 x 991)
26.5 x 26.5 x 51
45 (20) 310 (141) 2850 (1293) 1251V-1 586.9 20.7 35815
(673 x 673 x 1295)
26.5 x 26.5 x 63
55 (25) 310 (141) 3185 (1445) 1563V-1 725.0 25.6 44242
(673 x 673 x 1600)
Compressed Air 1/2" NPT 80-100 psig (5.5 to 6.9 bar psig), dry, oil free, 3 scfm.
1/2" NPT; hot water temperature 70-140°F 21-(60°C). Generator Drain: 1/2" NPT
RO / DI Water
Model 939V-1 4.2 gal/cycle (16 L/cycle)
General
information and Model 1251V-1 5.4 gal/cycle (20 L/cycle)
consumption
Model 1563V-1 6.6 gal/cycle (25 L/cycle)
Optional Stainless
Steel Integral 208-220 VAC, 50/60 Hz, 125 Amps, max. protection 170 Amps, 3-phase minimum, #1 [42.4] AWG
Steam Generator 230-240 VAC, 50/60 Hz, 113 Amps, max. protection 145 Amps, 3-phase minimum, #2 [33.6] AWG
(45 kW) Electric
380-415 VAC, 50/60 Hz, 59 Amps, max. protection 95 Amps, 3-phase minimum, #6 [13.3] AWG
440-480 VAC, 50/60 Hz, 57 Amps, max. protection 75 Amps, 3-phase minimum, #6 [13.3] AWG
5
ENGINEERING DATA / UTILITY REQUIREMENTS
45 PSIG max
CHAMBER 293°F (145°C) 1" NPT
3.1 bar(g) max
36 PSIG
JACKET 282°F (139°C) 1" NPT
2.5 bar(g) max
OPTIONAL INTEGRAL
45 PSIG
ELECTRIC STEAM 293°F (145°C) 1" NPT
3.1 bar(g) max
GENERATOR (ESG)
CUSTOMER IS RESPONSIBLE FOR COMPLIANCE WITH APPLICABLE LOCAL AND NATIONAL CODES
AND REGULATIONS.
6
NOT FOR INSTALLATION*
Total height: Upper Panel
89-1/16" (2263 mm) (cut-away added)
Utility Connections
12"
(304 mm)
Chamber Pressure
Human Machine Gauge
Interface (HMI)
Touchscreen Jacket Pressure
Gauge
Emergency Stop
77"
(1950 mm)
Thermal Printer
Chamber Door‡
D†
39"
1000 mm
7
Associated Documents
Document Title Document Number
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10307698
SLIDING, SD, RECESSED WALL, STEAM HEAT (p/n 11001835)
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10336741
SLIDING, SD, CABINET, STEAM (p/n 11001836)
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10336576
SLIDING, DD, RECESSED 1 WALL, CABINET, STEAM HEAT (p/n 11001837)
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG. Dwg. No. 10308020
SLIDING, DD, RECESSED 2 WALLS, STEAM HEAT (p/n 11001829)
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10802437
SLIDING, DD, RECESSED 1 WALL, CABINET, STEAM GENERATOR (p/n 11008536)
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10786670
SLIDING, DD, RECESSED 2 WALLS, STEAM GENERATOR (p/n 11008534)
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10778442
SLIDING, SD, CABINET, STEAM GENERATOR (p/n 11008531)
EQUIPMENT DRAWING, AMSCO 600, 39", 51" & 63" LG Dwg. No. 10749736
SLIDING, SD, RECESSED WALL, STEAM GENERATOR (p/n 11008529)
This document is intended for the exclusive use of STERIS Customers, including
architects or designers. Reproduction in whole or in part by any party other than
10304921 Rev B ©2019, STERIS Corporation. All rights reserved. a Customer is prohibited.