Manual ASpect CS EN 2011-09
Manual ASpect CS EN 2011-09
Manual ASpect CS EN 2011-09
User's manual
Service: Analytik Jena AG
Customer Services
Konrad-Zuse-Str. 1
07745 Jena
Germany
This documentation describes the state of this product at the time of publishing. It need not
necessarily agree with future versions of the product.
Subject to change!
Contents
1 ASpect CS Software .................................. 5 4.1.7.1 Entry of concentrations for manually
prepared standards ..................................... 45
1.1 User manual conventions .............................. 5
4.1.7.2 Entry of concentrations for standards
1.2 Starting and exiting ASpect CS ..................... 6
prepared with autosampler .......................... 46
1.3 General information on operation .................. 7
4.1.8 Statistics tab – Defining statistical analysis . 48
1.3.1 The workspace .............................................. 7
4.1.9 QCS tab – Quality Control Samples ............ 50
1.3.2 The Help function .......................................... 7
4.1.10 QCC tab – Quality control during the
1.3.3 The menu bar ................................................ 7 analytical process ....................................... 54
1.3.4 Frequently used control elements ................. 8 4.1.11 Output tab – screen view, printout and
saving options ............................................. 56
1.3.5 Function keys ................................................ 9
4.2 Saving and opening methods ...................... 57
1.3.6 Choosing a printer ....................................... 10
1.4 General measurement procedure ................ 10 5 Sequences ................................................ 58
5.1 Creating a sequence ................................... 58
2 Preparatory settings - Choosing the
analytical task .......................................... 11 5.1.1 Specific dialog functions in Sequence
window ........................................................ 58
3 Customizing ASpect CS .......................... 14
5.1.2 Specifying sequences of samples and
3.1 General settings .......................................... 14 actions......................................................... 60
3.1.1 View options ................................................ 14 5.2 Saving and opening sequences .................. 62
3.1.2 Save paths .................................................. 15
6 Sample information files ......................... 63
3.1.3 Export options ............................................. 16
6.1 Create information data for samples and
3.1.4 Options for continuous ASCII export ........... 16 QC samples ................................................ 63
3.1.5 Options for analysis sequence .................... 17 6.1.1 Elements in Sample ID window ................... 63
3.2 Specifying units, QC and stock standards ... 19 6.1.2 Editing sample ID ........................................ 65
3.2.1 Specifying units of measurements ............... 19 6.1.3 Edit QC sample IDs..................................... 65
3.2.2 Specifying stock standards and 6.2 Saving, opening and importing sample
QC samples ................................................. 20 information .................................................. 65
4 Methods .................................................... 21 7 Starting an analysis / Calculating
4.1 Creating method parameters ....................... 21 results ....................................................... 67
4.1.1 Lines tab – Choosing element lines ............. 22 7.1 Starting the analysis process ...................... 67
4.1.2 Flame tab – Flame parameters ................... 24 7.2 Interrupting, stopping, continuing the
analysis process ......................................... 69
4.1.3 Furnace tab – Parameters for graphite
tube furnace ................................................ 26 7.3 Repeating an individual measurement ........ 70
4.1.4 Hydride tab – Specifying the hydride 7.4 Specifics for solid sample analysis with
system ......................................................... 27 SSA 600 ...................................................... 70
4.1.5 Set auto sampler - Sample transport tab ..... 32 7.4.1 Preparing samples for automated analysis . 73
4.1.5.1 Auto sampler for flame and Hydride/ 7.4.2 Re-analyze samples for solid analysis ........ 74
HydrEA technique........................................ 32 7.5 Rinsing the system ...................................... 75
4.1.5.2 Autosampler for graphite tube technique 7.6 Reprocessing analysis results ..................... 75
(solution analytics) ....................................... 35
7.7 Evaluating measurements parallel to
4.1.5.3 Autosampler for graphite tube technique running analyses ......................................... 76
(solid analysis) ............................................. 37
7.8 Displaying results and analysis progress .... 77
4.1.6 Evaluation tab – Spectral range and
7.8.1 Sequence/Results tab ................................. 77
background corrections ............................... 40
7.8.2 Sequence tab .............................................. 77
4.1.7 Calib. tab – Calibration parameters ............. 42
7.8.3 Results tab .................................................. 78 8.4.1 Control of hydride system functions ........... 125
7.8.4 Overview tab ............................................... 80 8.4.2 Testing the hydride system for errors ........ 127
7.8.5 Solid tab ...................................................... 81 8.5 Autosampler .............................................. 129
7.9 Opening, closing and printing 8.5.1 Autosampler AS 51/52s ............................. 129
analysis results............................................ 81
8.5.1.1 Specify autosampler AS 51/52s ................. 129
7.10 Sample details and spectra ......................... 82
8.5.1.2 Technical parameters of the autosampler
7.10.1 Displaying details of single values AS 51/52s .................................................. 131
of samples ................................................... 83
8.5.1.3 Setting insertion depth and dosing speed
7.10.2 Displaying spectra ....................................... 85 for AS 51/52s ............................................. 132
7.10.2.1 Showing single spectra and varying 8.5.1.4 Functional test of the autosampler
background correction ................................. 85 AS 51/52s .................................................. 133
7.10.2.2 3D presentation of the spectra of single 8.5.1.5 Adjust autosampler AS 51/52s .................. 135
values .......................................................... 89
8.5.1.6 Position overview in the autosampler
7.10.2.3 Line identification ........................................ 90 AS 51/52s .................................................. 136
7.11 Calibration ................................................... 91 8.5.1.7 Supply of reagents for sample ................... 136
7.11.1 Graphic presentation of calibration curve .... 92 8.5.2 Micro pipetter unit MPE 60 ........................ 137
7.11.2 Displaying calibration results ....................... 92 8.5.2.1 Specify autosampler MPE 60 .................... 138
7.11.3 Modifying the calibration curve .................... 94 8.5.2.2 Technical parameters of the autosampler
MPE 60 ...................................................... 139
7.12 Quality control ............................................. 95
8.5.2.3 Set insertion depth and dosing speed of
7.12.1 Parameters of QC charts ............................ 95
the MPE 60 ................................................ 141
7.12.2 Displaying QC charts .................................. 97
8.5.2.4 Automatic depth correction for MPE 60 ..... 141
8 Controlling and monitoring 8.5.2.5 Aligning MPE 60 to graphite tube furnace . 143
spectrometer and accessories ............... 99
8.5.2.6 Function test of autosampler ..................... 144
8.1 Spectrometer............................................... 99
8.5.2.7 Adjusting the MPE 60 ................................ 146
8.1.1 Device function test on contrAA .................. 99
8.5.2.8 Position overview of the MPE 60 ............... 147
8.1.2 Measuring a spectral peak at a selected
wavelength ................................................ 101 8.5.3 Solid autosampler SSA 600 ....................... 148
8.1.3 Starting continuous measurements ........... 101 8.5.3.1 Function test of solid sampler .................... 149
8.1.4 Correcting peak offsets ............................. 103 8.5.3.2 Alignment of solid sampler ......................... 150
1 ASpect CS Software
ASpect CS is the control and data analysis software for the High Resolution Con-
tinuum Source Atomic Absorption Spectrometer
• contrAA 300 for flame technique and hydride/Hg-cold vapour technique
• contrAA 600 for graphite tube technique and hydride/Hg cold vapour tech-
nique and HydrEA technique
• contrAA 700 – a multi-function system for flame technique, graphite tube
technique, HydrEA technique and hydride/Hg cold vapour technique
The following accessories from Analytik Jena are supported by this software:
• AAS Autosamplers AS 51s and AS 52s
• SSA 600 solid autosampler
• MPE 60 micro pipetting unit for graphite tube technique
• Hydride system HS60 / HS60A, HS55 / HS55A and hydride injector HS50
(hydride/Hg cold vapor technique)
• SFS 6 Injection Switch (flame technique)
The software supports the unique features of HR-CS AAS, such as simultaneous
background correction by acquiring spectral information in the vicinity of the analy-
sis line, fast element/line change and the fast sequential multielement measure-
ment at continuous sample flow in flame mode. The method parameters for the
measurement procedures can be optimized to the specific demands of the sample
to be analyzed.
The obtained data can be recalculated, exported to various file formats and printed
out.
Intended use
ASpect CS software exclusively serves to control the above mentioned high resolu-
tion continuum source atomic absorption spectrometer and to analyze the data
obtained with this device.
The manufacturer does not assume any liability for problems or damage caused by
the non-intended use of ASpect CS.
ASpect CS and the devices to be controlled by it may only be operated by appro-
priately qualified and instructed personnel. The use must be familiar with the infor-
mation given herein and in the user’s manual of the device.
Exiting ASpect CS
To exit the application, activate menu command File / Exit.
Alternatively, you may close the program in the Main Settings window by a
click on the [Exit program] button.
If, at this time, method, sequence or sample information data files are open
that have not been saved yet, you will be informed accordingly. If you want to
save these files, click on [Yes].
A request for a software controlled system cleaning is displayed when working
with a hydride system.
If the xenon lamp is still on, a request for the switch-off of the lamp will appear.
After you have switched off the lamp, keep a wait time of 30 s before switching
off the contrAA. The countdown of the wait time is graphically displayed on the
screen. Afterwards, you can switch off the contrAA by hardware.
Menu Description
File Creating, opening and saving method, sequence and sample
information files.
Opening results files.
Setting up a printer and printout.
Starting offline or online program instance.
Activating the Main Settings window.
Exiting the application.
Direct recall of method and sequence files opened recently.
Edit Copying and pasting the contents of textboxes and input fields.
Copying selected rows of the result list to the clipboard.
Deleting the contents of the result list.
Buttons
The function of icon buttons is indicated by means of tool tips displayed when the
mouse pointer rest on the corresponding button.
Lists
In some of the windows, values are to be entered directly in a list. Dependent on
the type of entry, the list cell behaves like an input field, a selection list, or a spin
box.
To select a row of a table, click on the corresponding row in the first table
column highlighted by a gray background. Afterwards, you can move the line
cursor with the [↑] and [↓] buttons.
To change the width of a column move the mouse pointer to the correspond-
ing border line in the column head until it turns into a double-headed arrow.
Keeping the left mouse button depressed, you can then drag the border line to
adjust the desired width.
In input fields, the following functions are additionally available:
• [F2] activates the edit mode. In this mode, the [←] and [→] keys are used
for editing character by character. Renewed pressing of [F2] reactivates
the standard mode where the cursor keys are used to navigate between
the cells.
Edit / Copy • Text can be copied to the Windows clipboard via menu command Edit /
Copy or key combination [Ctrl+C] and inserted via menu command Edit /
Edit / Insert
Insert or key combination [Ctrl+V].
Graphs
In graphs, you can open a context menu by clicking the right mouse button. This
menu provides options for copying either the graph or the entire window to the
Windows clipboard.
In several graphic windows, additional icon buttons are accessible:
Activates the zoom mode. With the left mouse button pressed you can
select an area of the graph to be zoomed in.
Deactivates the zoom mode and resets the graph to the original scale.
Activates the text mode. Keeping the left mouse button pressed you
can draw a frame and enter text that shall be added to the graph.
A double click on the existing text opens the window for editing or de-
leting the text. With Ctrl+right mouse button you can move the text.
Activates the select mode in graphs of signal curves or spectra. Click-
ing the left mouse button adds labels to the measuring points; pressing
Shift+right mouse button deletes all labels.
Technique
In this field, define the technique to be used for the specific task. The technique
selected here, determines various options accessible for the further program run.
Task
For measurement tasks and after a change of the technique, device and accesso-
ries must be reinitialized (see the description of the [Initialize] button further be-
low). Reinitialization of the device is not necessary if you select Data management
or Options.
Available accessory
Before you can actually start a measurement, device and PC must be connected to
each other. Verify that the spectrometer and the accessory to be used are con-
nected and ready for operation. Clicking the [Initialize] button releases the detec-
tion and the configuration of the spectrometer and the accessory unit(s) dependent
on the technique selected.
On exiting the Main Settings window with the [OK] button, the system checks the
state of initialization and informs the operator of the necessity of initialization, if
necessary.
Detected accessory units are listed by name. If an accessory is identified as Not
initialized, its initialization was skipped because of the technique selected (e.g. the
hydride system if the flame technique has been selected); if --- is displayed, the
respective accessory was not detected.
In the Date and Time field, the current system time of the PC is displayed.
In the Laboratory field, you can type in a name of maximally 30 characters. The
name entered last is saved and printed out as information on result reports.
If you use the optionally installable user management, the Operator input field
shows the registered user. If you do not use user privilege management, you may
enter the operator’s name manually (30 characters).
For trainings and demonstration purposes, you may operate ASpect CS without an
AAS device being connected. To this end, activate the Simulation checkbox.
Then, all device functions (including data acquisition and analysis) will be run in
simulation mode.
3 Customizing ASpect CS
3.1 General settings
In the Options window, you can choose the following settings, which apply to the
complete operation of ASpect CS:
• Display options
• Save paths of files
• Parameters for data export
• General measurement settings
The options selected in this window remain set after exiting and restarting ASpect
CS.
Extras / Options... Activate the Options window with menu command Extras / Options... or by a
click on the
button.
Alternatively, in the Main settings window, you may activate the Options op-
tion button in the Task field and click on [OK] (→ "Preparatory settings -
Choosing the analytical task" p. 11).
Display
Table colors
A click on the [...] button opens the color selection dialog. There, you may choose
predefined or newly defined colors as list background.
Use PrtScr key By default, the printout of the hardcopy of the screen is start-
for hardcopy ed with [F5]. In this case, the [PrtScr] key of the keyboard is
function used for the Windows clipboard function.
If this checkbox is activated, the [PrtScr] button starts the
printout of the screen hardcopy.
This function becomes active only after restarting ASpect CS.
Scientific Mode Activates spectrum presentation.
If this option has been deactivated, the functions for the
presentation and editing of spectrum data are not accessible.
With the [Default settings] button, all currently set options are reset to default set-
tings.
Signals
Button [...] opens a color selection dialog. Pre-defined or custom colors can be
selected for displaying the net signal or the background signal (graphite furnace
technique).
With the [Add…] button, you can create new subfolders to the Results folder for
saving results. Besides, it is possible to delete and rename empty folders here.
Options
In the Result fields for export field, you may define which columns of the result
table shall be exported to the ASCII file. All exports all columns of the selected and
non-selected tabs of the result list, only selected fields opens a list in which you
can choose the columns to be exported.
A click on the [Default settings] button resets all options and stored window loca-
tions to default settings.
File name
[Method The file name corresponds to the name of the method. The file
name].csv name extension used is ".csv".
The file is saved to the default Export/Import path (Options -
Paths).
[Results file The file name corresponds to the name of the results file. The
name].csv file name extension is ".csv".
The file is saved to the default Export/Import path (Options -
Paths).
other You may freely define file name and save path. A click on the
[...] button opens the file selection dialog for the selection of an
existing file.
Continuous ex- The spectra are exported as CSV files to the specified path.
port of spectra The filename is created automatically using "Row number-
(CSV) Sample name-Line-Replicate", e.g. 0007-Sample-Al309-
02.csv.
Export path A click on the [...] button opens the path selection dialog for the
selection of an export path.
Offset of optical The analysis sequence is stopped if the optical system is not
system correctly adjusted.
Please check the correct function of the spectrometer (→
"Device function test on contrAA" p. 99).
Invalid calibra- If activated, the analysis sequence will be stopped if the cali-
tion function bration function could not be calculated.
Monotony of If activated, the calibration points will not be tested for monoto-
calibration ny.
points The monotony test serves to determine if higher standard con-
centrations also lead to higher measured values.
Autosampler If activated, the signals from the fill level sensors of the au-
level sensors tosampler and the waste bottle will be checked.
Display windows
Other
Take-up compo- While the graphite tube cools down, the MPE takes up next
nents during samples. This option may be used to accelerate a measuring
cooling phase sequence.
Beep after end of A beep will sound as soon as the graphite tube has completely
cooling phase cooled down.
Stop after trans- Stops a running measurement program if active and if the
former overheat- transformer for graphite tube technique was found to be in an
ing overheated state. The program will not continue in this case.
Interrupts a running program on detecting a transformer over-
heat situation if disabled. Resumes the interrupted program
after the transformer has cooled down.
Readjust wave- If activated the wavelength is readjusted for each replcate
length before measurement. This improves precision (default is active).
each measure-
ment
Formation re- If activated a message is displayed when starting a measure-
quired after ment without tube formation after opening the furnace or
opening furnace switching on the instrument.
The [Append] and [Insert] buttons serve to append user-defined units to the end
of the list or insert them above the currently selected row. With the [Delete but-
ton], you can delete user-defined units, the preferred default units, however, can-
not be deleted. To permanently save any changes on this tab, click on Save.
4 Methods
4.1 Creating method parameters
To create a new method, activate menu command File / New Method....
This will bring up a selection dialog with the following options:
File / New Meth- Based on de- Opens the Method window for the entry of new method
od... fault parameters parameters and with editable default settings for calibration
and statistics.
Based on cur- Opens the Method window with the currently set method
rent parameters parameters.
Based on saved Activates the Load method dialog.
method… After the selection of the method, its parameters are dis-
played in the Method window.
Choose the desired option and confirm your choice with [OK] to open the
Method window either directly or via the Load method dialog.
Lines tab
Order Order in which the analysis lines within a group are measured.
With the [Opt.] button, you can have their order optimized auto-
matically, i.e. the lines will be measured by increasing fuel gas
flows.
With graphite tube technique, Order defines the order in which
lines will be measured.
This selection dialog contains the periodic system with the selectable elements
being displayed as blue buttons. Additionally, it shows the line table with the follow-
ing columns:
By activating the Element or the Line option button, the line table is sorted
increasingly by either the chemical symbol or the wavelength.
If you click on an element symbol in the periodic system (blue buttons are
selectable elements), the lines of the selected element will be displayed in the
line table.
Alternatively, you may enter the element symbol in the Select Element text
box.
To have the complete element list displayed again, delete the entry in the Se-
lect Element text box.
The selected elements/lines will appear on the table in the Method - Lines win-
dow.
Note
You may also choose several lines of different sensitivities of a single element.
Flame tab
Fig.6 Method - Flame window with burner and gas flow settings
Flame
Note
When working with auxiliary oxidant, optimize the flame parameters manually (→
"Manual flame optimization", p. 108).
Burner/nebulizer
Furnace tab
To serve as default settings for atomization of the various elements using EA tech-
nique, the respective furnace program data from the cookbook are preset. Regard-
less of that, a furnace program can be edited for any analytical line in the Furnace
window and a currently selected method can be transferred (→ "Editing a furnace
program" p. 112).
The related list includes the following furnace program parameter items:
Control buttons
[Edit furnace Opens the Furnace – Furnace program window that provides
program] a full display of the furnace program. Furnace parameter set-
tings can be adapted for any element line subject to analysis
(→ "Editing a furnace program" p. 112).
[Accept furnace Validates the parameter settings of a marked analytical line for
program] all subsequent lines in the list.
[Modif.+Extras] Opens the Furnace – Modif.+Extras window with setting op-
tions for additional modifiers, enrichment and thermal pre-
treatment of element lines (→ "Matrix modifiers, selections for
enrichment and pretreatment" p. 115).
Note
The commands for additional washing or loading of the hydride system are re-
leased from the Hydride System window (→ "Hydride system", p. 125).
Mode
You can choose among different modes depending on the equipment of the hy-
dride system.
Pump speed Four speed levels (1 - 4) are available for the transport of the
level sample (in continuous mode) and the components.
In continuous mode, the supplied sample volume is determined
on the basis of the pump speed level selected and the reaction
time.
System cleaning
System cleaning may be selected optionally after every sample measurement.
Operation times
The operation times are to be adjusted dependent on the selected operating mode.
All operation times are entered in seconds.
Load time Time in which the sample pump loads the sample tube up to
the two-valve assembly with sample. This time is needed only
for the first measurement of a new sample.
Prewash time Time for purging the beaker with argon before the reaction
(for the analysis of hydride formers).
The prewash time is used to purge off air in order to prevent
an oxy-hydrogen reaction in the following reaction.
Reaction time Time in which the sample pump pumps sample into the reac-
tor. This is the crucial parameter for the supplied sample vol-
ume and the measuring sensitivity.
Pump time Time in which reductant is pumped into the beaker in order to
start a reaction.
AZ wait time Time directly preceding the baseline (auto zero) adjustment.
Wash times 1 … 3 Times used to convey the reaction gas by means of the argon
flow. The transport paths are different in the individual phases
for the various operating modes.
The transport paths can be presented graphically.
Heat. Time collec- Time in which the collector heating is on in order to release
tor the enriched Hg.
Cool. Time collec- Time in which the collector is ventilated in order to cool it
tor down for a new enrichment cycle.
Gas flows Defines the argon flow flowing in the phases displayed left of
it. The adjusted gas flow applies until a new gas flow can be
entered. The gas flow can be switched over differently often
for the different operating modes. The gas paths for the indi-
vidual phases are illustrated in the graphic of the analysis
process on the hydride system (→ "Hydride system – Graph
of analysis process", p. 30). The gas flows are adjustable in
three steps from 5 to 15 Liters/hour.
Batch Parameters
Sample volumes Here, enter the volumes of the sample being in the beaker.
Enrichment cycles To be defined for the batch mode with Hg enrichment on the
collector.
Define the number of beakers the contents of which is en-
riched.
[Plot] button
By a click on the [Plot] button, you can open the graphic presentation of the gas
paths for the individual phases of the analysis process.
Hydride tab Click on the [Plot] button in the Method - Hydride window to open the Plot win-
dow. This window shows a graphic presentation of the programmed analysis pro-
cess.
Fig. 9 Plot window: Graphic representation of analysis process with hydride system
The individual phases of the analysis process are shown in colors in the process
chart. By clicking on a particular phase shown by a colored field, the corresponding
parameters appear in the Parameters field, and the adjusted gas flow is illustrated
in the Gas paths and flows field. The process is determined by the operating
mode selected on the Hydride tab.
Graphic for If sample statistics has been activated, the different processes
can be displayed for the first, the next and the last measure-
ment.
Gas paths and This flow diagram shows the gas paths of the hydride system.
flows The modules of the system are shown with their connection
tubes (for argon and reaction gas). The modules include React.
(Reactor), GLS (Gas-Liquid-Separator), Coll. (gold collector),
Batch (Batch module) and Cell or furnace.
In the process graph, click on the phase the gas flow of which
you want to get displayed. The gas path is marked in red and
the argon flow displayed numerically in L/h.
Parameters This field serves to display and edit the operation times, meas-
uring times and the corresponding gas flows of a selected
phase of the process.
Click on the phase of interest. The name and the numerical
value of the operation or measuring time /gas flow will be dis-
played and can be changed. The process graph will be updated
accordingly when you change a parameter.
Line Select displayed line data.
Fig. 10 Method – Sample transport window with activated autosampler for flame
techniques
The Method – Sample Transport window allows you to specify parameters for the
autosamplers:
• turning a sampler on
• Wash modes
• automatic dilution steps as part of an analysis
• using Flame technique:
Involvement of SFS 6 injection switch or HS 50 hydride injector
Accessories
Injection switch The SFS 6 Injection switch can be used in combination with an
SFS autosampler or in manual mode. On the one hand, it allows for
permanent intake of wash or carrier solution and for the burner
to be held at a constant temperature by the aerosol, on the oth-
er hand it allows for reproducible measurements of small sam-
ple volumes.
The injection module can also be used for processing time con-
stant signals (mean integration).
Injection time: The time the valve of the SFS 6 opens the
sample path to nebulize the sample and convey the
aerosol to the burner. The time depends on the
highest concentration to be expected. Typical val-
ues: 0,5 ... 2.0 s.
Load time: The time needed to fill the sample aspiration
path between sample cup and injection module with
new sample.
Hydride injector The hydride injector HS 50 is a purely pneumatic batch system
HS50 for manual operation. It consists of batch installation and cell
holder with quartz cell. The reductionagent solution is trans-
ported pneumatically from the supply bottle into the reaction
tank. The quartz cell is heated by the flame. The operation
times of HS50 are controlled by the AAS software in a simple
way.
Peak area mode as well as peak height mode is possible.
The measurement procedure is divided in following parts: Pre-
wash – Autozero – Reaction/Integration.
Reaction time Set reaction time.
During the reaction phase reaction agent is trans-
ferred to the reaction beaker. The measurement
signal acquisition starts simultaneously. The integra-
tion time has to be set in a way to acquire the total
signal.
Prewash time Set prewash time.
During the prewash time the reaction beaker is
flushed free from air. The prewash phase is omitted
for the determination of Hg because the argon flow
is necessary in order to transport the Hg out of the
sample.
Sample volume Define sample volume.
Delay time Time which is needed to transport the sample to the atomiza-
tion unit (e.g flame or reaction chamber in the hydride system).
The time is essentially determined by the length of the sample
tubes. This time is needed to convey the sample to the flame.
With AS 52s, dilution is performed in the mixing cup. Required volumes are math-
ematically determined as part of a program sequence, depending on the extinction
value for undiluted solution state. Once calculated, the analyte volume is intro-
duced into the mixing cup and the mixing cup is then filled with diluent solution to
the predefined fill level. The amount of diluent solution is drawn from the storage
bottle for diluent solution.
Wash mode off Wash mode switched off. No rinsing performed au-
tomatically.
after sample Washes after each sample, but not within a sta-
tistical series.
Wash time Time in which the rinsing agent is aspirated in the wash cup.
Includes washing of tube path and burner-nebulizer system.
Mixing cup cy- Number of wash cycles for mixing cup.
cles Fills mixing cup with wash liquid/diluent solution and drains it
again in one cycle.
Controlled cleaning
Where analyzed samples are of a kind that result in exceeding the calibration
graph working range by more than 10 %, the graphite tube, the burner-nebulizer
system (flame technique) or the hydride system may be washed as may be appro-
priate for the currently selected technique, in order to remove contamination from a
preceding measurement. During the wash, the absorbance/emission is measured
in order to check the cleaning results.
Automatic controlled cleaning should be performed following measurement of high-
ly concentrated samples, notably, with Dilution on conc. exceeding mode active.
Note
Controlled cleaning can also be defined as part of a sequence, independently of a
concentration exceeded situation.
(cannula, sample tube, sample injector SFS6 and the burner-nebulizer system).
Surplus amounts of washing liquid will flow off into the waste bottle.
The mixing cup of AS 52s is cleaned by filling with wash liquid/diluent solution and
draining it again in one single cycle.
Sample transport
tab
Fig.11 Method – Sample transport window with settings for the MPE 60
The Method – Sample Transport window allows you to specify parameters for
these autosamplers:
• turning a sampler on
• Wash modes
• automatic dilution steps as part of an analysis
Delay time Moving the calculated time of the measurement start by the
delay time This time is needed to convey the sample to the
flame.
Use autosampler Use the connected and initialized autosampler.
This option must always be activated in the graphite tube tech-
nique.
Wash mode off Wash mode switched off. No rinsing performed auto-
matically.
between samples
Washes after each sample, but not within a statistical
series.
between measurements
Cleaning after each statistic run
after each component
The autosampler is washed after transfer of each
component into the graphite tube (modifier, standard,
sample, etc.).
Wash cycles Number of wash cycles per wash, 1 to 5
Mixing cup cy- Number of wash cycles for mixing cup.
cles Fills mixing cup with wash liquid/diluent solution and drains it
again in one cycle.
Controlled cleaning
If samples are analyzed which result in the calibration graph working range being
exceeded by more than 10%, the graphite tube may cleaned by baking out in order
to remove contamination from the previous measurement. During the cleaning, the
absorbance/emission is measured in order to check the cleaning results.
Automatic controlled cleaning should be performed following measurement of high-
ly concentrated samples, notably, with Dilution on conc. exceeding mode active.
Note
Controlled cleaning can also be defined as part of a sequence, independently of a
concentration exceeded situation.
Autosampler
Note
In the display of the line spectra, the pixels can be selected based on the graphic
presentation for the background correction and transferred to the method.
Fig.14 Method– Calib. window: Selection of calibration method and line-specific pa-
rameters
Standard preparation
Blank correction
This function is available only for the methods of standard addition and addition
calibration.
Note
With the methods of standard addition and addition calibration, always the linear
calibration curve will be used.
Intercept
Set zero The calibration curve exactly intercepts the measured zero
point.
calculate The zero value is included in the calculation as any other cali-
bration point.
none All calibration points are taken into account with the same
weighting.
1/conc Calibration points obtained with lower concentrations are to be
weighted more strongly.
1/SD Points with lower deviations within several repeat measure-
ments of a standard are taken into account with a stronger
weighting (requires: activated Sigma statistics option).
1/(SD*conc) Combination of the calculation methods 1/conc and 1/SD.
none Uses all measured and non-deleted calibration points for the
calculation of the curves. Calibration points are neither labeled
nor eliminated.
Elim. outliers If calibration points are outside the calculated prognosis range,
outliers will be eliminated by means of an F-test (a test to ascer-
tain whether the exclusion of a point results in a significant im-
provement of the residual scattering):
• An F-test is performed for the calibration point being far-
thest outside the prognosis range.
If the exclusion of this point does not lead to a significant
improvement of the residual scattering, the point will be
included and the calibration point not further optimized.
• If the exclusion of this point results in a significant im-
provement, the calibration point will be defined as outlier
(marked in the table by "!", in the graph marked by red
color) and the calibration recalculated without this point.
• Then, another F-Test is performed for the calibration
point now being farthest outside the prognosis range.
This procedure is repeated until all outliers have been
eliminated.
Concentration unit
Enter the concentration units for the element to be analyzed in the Unit column.
Note
With the contrAA it is possible to choose several analysis lines of different sensitivi-
ties for one element and thus extend the working range. For every analysis line, a
separate calibration curve is determined. In the calibration table, the calibrations for
different lines of an element are distinguished by an index following the element
symbol. The index number corresponds to the sequence in the element table, e.g.
Cu1, Cu2.
Calibration tabel for flame technique Caibration table für graphit furnace technique
Fig.15 Calibration table for standard methods with manually prepared standards
Use [Concentration] on the Calib tab to open the Calibration table window.
Define the number of standards in the various list fields above the table, as
required for your selected calibration method.
Maximum number of standards: 65.
Calbration table for flame technique Calibration tabel for graphite furnace technique
Fig. 16 Calibration table for standard procedures with automatically prepared stand-
ards
Statistics type
Deactivated Includes all values of the statistics series for the determination
of the mean value.
Activated Eliminates outliers by excluding them from the calculation of
statistical quantities. The thus found mean values in the result
table are marked by "!".
From the Reaction list box, choose the further procedure to be run if the error
limit is exceeded:
With QC Sample, QC Std. and QC Stock selected as QC sample type:
If you selected QC-Blank as QC sample type, you may only choose among the
flag, next method and Stop reaction options described above.
For the QC Trend and QC Matrix QC sample types, no reaction is provided.
In the list of QC samples, define the line-specific parameters for every element
line dependent on the QC sample type selected:
next method Accessible only for Kalib. check and Recal. check options.
Stops the measurement of the currently running method and
starts the next method, if the error limits were exceeded.
Stop Accessible only for Calib. check and Recal. check options.
Stops the measurement of the currently running method, if the
error limits were exceeded..
Tube control
Only for graphite tube technique.
Specifies settings for graphite tube monitoring.
By activating the Method and/or Spectra option, the corresponding data will
be saved additionally together with the analysis results.
In the table below, define the number of decimal places for the display and
printout of absorbance and concentration values, and the order in which the
analysis lines shall appear on the printout.
For the development of analysis methods, it is advisable to save the spectra as
well. You will get important information from them, e.g. on potential matrix or inter-
ference problems. Recalculation with modified method parameters without saved
spectra will be possible only based on the calculated absorbance values. So, for
instance, it is not possible anymore to change the data points for background cor-
rection, because the spectral information on the individual pixels is missing.
For a proven routine method, however, saving of spectra will not be necessary.
Note
If in the Options – Analysis sequence window, you activated the Always save
spectra option, the spectra will always be saved independent of the settings in the
method parameters (→ "Options for analysis sequence", p. 17).
Saving methods
The command for saving the current method parameters can be given in different
ways.
In the Method window, click on the [Save] button.
Alternatively, you may activate menu command File / Save / Method....
File / Save /
Method This will bring up the Save method window.
In the Name textbox, type the desired method name.
In the Cat (category) textbox, you may optionally enter an additional identifier
of maximally three characters to facilitate later method search in the database.
In the Description textbox, you can optionally enter information on the meth-
od.
Activate the option Save calibration data, if you want to save the calibration
data with the method.
Save the method with [OK].
On doing so, the method will be saved to the database. If you choose an existing
method name, the existing method will not be overwritten, but a new version
created in the database. To remove methods from the database, you must explic-
itly delete them!
Note
The method may also be saved to the results file of the measurement.
To save the method along with the results file, you must have activated the Save
with results - Method option in the Method - Output window. After having opened
the results file, you may also reload the method.
Opening methods
File / Activate menu command File / Open method... or, on the toolbar, click on the
Open method folder symbol right of the Method field.
Alternatively, you may click on the [Open] button in the Method window.
This will bring up the Load method window.
Choose the desired method from the list.
By the entry in the Cat textbox, you can define that only methods of the speci-
fied category are displayed.
If you want to see methods of all categories, delete the entry in the Cat field.
Activate the Current version only check box, if with methods of the same
name you want to see the method of the highest version number only.
Open the selected method with [OK].
5 Sequences
A sequence is based on a loaded method, which contains the information on the
type of calibration, statistical analyses, quality control, etc.
The sequence contains the information on the samples and actions in their order of
processing. Some sample describing data such as sample name and position on
the sample tray may also be entered directly. For permanent storage, however, the
sample describing data must be saved as sample information file.
Buttons
[New] Adds a new row to the end of the list and opens the Edit Se-
quence window.
[Insert] Inserts a new row at the selected place of the list.
[Delete] Deletes the selected row.
[Delete Table] Deletes the complete sequence table.
[Samples] Opens the window Sample ID.
[Sequ.<-QC- Transfers QC sample information (e.g. names and sampler
samples] positions) to window Sample ID (→ "Create information data for
samples and QC samples" p. 63).
The information from the QC sample ID table are entered in
the sequence table. The first row with the new sample ID is to
be defined in the From row textbox.
[Sequence<- Transfers information on sample name and place on the tray of
Sample] the autosampler from the Sample ID window (→ "Create infor-
mation data for samples and QC samples" p. 63).
The information from the Sample ID table is entered in the se-
quence table. The first row with the new sample ID is to be de-
fined in the From row textbox.
Sample/Action Description
Samples Measures the number of samples specified under Number.
QC samples Measures a QC sample and evaluates it as specified in the
Method.
From the appearing list box, choose one of the QC sample
types specified in the Method – QCS window. The parame-
ters of the QC sample are displayed in the opposite field.
Reference Measures a reference.
Reag. Blank Measures a blank.
QC Blank DL Measures a blank for the determination of the detection and
determination limits according to the blank method.
Calibration Measures the calibration samples and runs the calibration
according to the options defined in the method.
Recalibration Measures the calibration sample provided for recalibration
and performs the recalibration.
Special action Actions which do not directly involve the measurement of
samples. These actions can occur for each line, only for
the first line or only for the last line in the method as pre-
ferred. For list of possible special actions see below.
Load method Loads a saved method, e.g. to start another element analysis
within the sequence.
Click on [...] to open the Open method window (→ "Saving
and opening methods", p. 57). Choose one of the stored
methods.
The following special actions can also be inserted in the measurement pro-
cess:
After you have chosen all elements of the sequence, click on [OK] to return to
the Sequence window.
In the list, enter the Elements to be analyzed.
When using the autosampler:
Define the position of the samples on the autosampler. The positions of cali-
bration and QC samples are automatically transferred from the method. How-
ever, you may change the positions here.
Note
It is advisable to enter the data of the samples to be analyzed in the Sample ID
window and transfer them afterwards to the sequence list (→ "Create information
data for samples and QC samples" p. 63).
Saving sequences
In the Sequence window, click on [Save].
Alternatively, you may activate menu command File / Save / Sequence.
File / Save / Se-
quence This will bring up the Save Sequence window.
In the Name textbox, type the desired sequence name.
As default, "Sequence" and the current date are offered, e.g. "Sequence-08-
15-05".
In the Cat (category) textbox, you may optionally enter an additional identifier
of maximally three characters to facilitate later sequence search in the data-
base.
In the Description textbox, you can optionally enter information on the se-
quence.
Save the sequence with [OK].
On doing so, the sequence will be saved to the database. If you choose an exist-
ing sequence name, the existing method will not be overwritten, but a new
version created in the database. To remove sequences from the database, you
must explicitly delete them!
Opening sequences
File / Open Se- In the Sequence window, click on [Open].
quence ....
Alternatively, you may activate menu command File / Open Sequence .... or,
on the toolbar, click on the folder symbol right of the Sequence field.
This will bring up the Open Sequence window.
Choose the desired sequence from the list.
By the entry in the Cat textbox, you can define that only sequences of the
specified category are displayed.
If you want to see sequences of all categories, delete the entry in the Cat field.
Activate the Current version only check box, if with sequences of the same
name you want to see the sequence of the highest version number only.
Open the selected sequence with [OK].
Method Develop- .
ment / Sample Alternatively, you may open the Sample ID window using the menu commands
ID... Method Development / Sample ID... or File / New Sample Information File.
File / New Sample The meaning of buttons and symbols contained in the Sample ID window, which
Information File are also used in other windows of this application, are described in Section
"Frequently used control elements" p. 8.
The program pops up brief information (tool tips) on the table while you move the
mouse pointer across the button.
Sample information
Pos. Position of sample on autosampler
Name Sample name
The entry is optional.
Maximum number of characters: 20
Pre-DF Pre-dilution factor. Factor by which the original sample was
diluted before being placed on the autosampler or supplied to
the spectrometer, if no autosampler is used. The entry of the
factor is required for the calculation of the concentration of the
original sample (Conc.2 in the result table).
Note!
For solid analytics, known weighed sample portions must be specified in the Re-
sult – Solid window (→ "Specifics for solid sample analysis with SSA 600" p. 70).
QC sample information
This tab contains the QC sample information.
The columns correspond to the columns of the tab Sample information. Addition-
ally information about the QC type is contained in column Type. The columns Unit
is omitted because QC sample units are defined in the method. Blank correction is
not available for QC samples.
Buttons
[New] Adds the number of rows specified in the # textbox to the end
of the list.
[Insert] Inserts the number of rows specified in the # textbox at the
selected place of the list.
[Delete] Deletes the number of rows specified in the # textbox from the
selected row of the list on.
# Input field for the entry of the number of rows to be added,
inserted, or deleted.
Name Here, enter the desired name for the result database.
New file/list With this option activated, you must enter a
new file name. The program will check if the file
name already exists. Existing files cannot be over-
written.
Append to file/list New results will be appended to an
existing results file. Click on [...] to open a selection
window where you can choose an existing results
file from the displayed list.
Folder Choose the save path for the results file. The default folder for
saving of results is shown in the Options – Folder window
(click on ).
Description Here, you may enter an additional comment that is saved along
with the analysis results.
The entry is optional.
Switch off lamp When the measurement is stopped because of an error the Xe
after error stop lamp is automatically switched off..
[OK] Starts the measurement.
The file contains the results of measurement and evaluation and the sample ID
information. If selected accordingly in the method settings, method and spectrum
data are saved simultaneously with the results. In this case, method parameters
are saved to the result database, while spectra are saved to a separate file of the
same file name, but a different extension in order to save storage capacity.
The result database is saved with the extension ".tps". Spectrum files get the ex-
tension ".SPK".
Starting measurements
Start the measurement routine using menu command Routine / Start Se-
quence or click on .
In the Start Sequence window, choose a file name for the results file.
You may optionally save the results to a new file or append it to an existing file.
Routine / Start Overwriting of an existing file is not possible.
Sequence
After you selected the file name, the measurement routine will start according to
the settings selected in the method and the sequence.
If you use an autosampler, the measurement runs automatically.
In the case of manual sample feed without autosampler, follow the prompts for
the provision of samples displayed on the screen.
Additionally, in the symbol bar at the side up to three large buttons are displayed
during measurements:
[Activate Scrap- The scraper cleans the burner head between two measure-
er] ments within a statistics series of a sample.
[Method (read Turns the method window on. A method can only be read, it
only)] cannot be changed.
[Sequence Turns the sequence window on. A sequence may be expanded
Samples] while measurement is going on. The sequence window includes
a [Sample ID] control button that allows you to edit further
sample ID data.
[Stop Sequence] Instantly stops the analysis process.
[Display win- Opens (if closed) or closes (if open) the report and graphic win-
dows] dows during the measurement.
Routine / Stop
With menu command Routine / Stop or button (red button), you can stop
… / Break the analysis process instantly (recommended for flame analyses or as emer-
… / Continue gency stop in hydride analyses).
With menu command Routine / Break or button (yellow button), you can
request for an interruption of the analysis process. After this request was de-
tected, the button is grayed out. The currently running hydride measurement
will be finished first. Then, a stop of the analysis process is released (recom-
mende for graphite furnace technique and hydride technique).
With menu command Routine /... or the button, you can continue a
stopped/interrupted routine.
This will bring up the Continue Sequence dialog, in which you are informed of
the status of actions before the interruption.
If you want to modify the method before continuing, activate the Continue with
modified method check box. On doing so, a new method entry is made to the
results file.
The measurement may be continued in one of the following ways:
If known, the following settings can be directly entered at the sample table:
• Weighed portion if sample preparation occurred on external scales
• Deadweight (tare)
• Marking for completely dosed sample platform
Preparing samples
Re-analyze a sample
Select tab Results.
Double-click on the sample. The window Single values (→ "Sample details
and spectra" p. 82) is opened.
Select the sample in the table.
Select [Mark for re-measurement].
Close window Single values.
Repeat the procedure for all samples which shall be re-analyzed.
Go back to tab Solid.
Select [Prepare re-measurement].
For all marked samples additional entries for re-analysis are added to the
sample table.
Prepare the samples for analysis according to the platform assignment.
Start the measurement with [Re-measure single val.].
The recalculated results for the re-analyzed samples are added to the results
table in window Solid.
Flame technique
The sampler tube is immersed in the wash cup for rinsing. If the Injection Switch
has been selected as accessory unit, the switch will open. This is to ensure that the
sample path is rinsed, too. The rinse pump is continuously delivering fresh wash
solution.
Confirm the selection of the results file with [OK] and start the reprocessing
action.
Note
Reprocessed values may be optionally saved to a new database or appended to
an existing results file. The originally obtained results are preserved unless the
original file is deleted. Thus, it is impossible to manipulate the original data.
.
You may further process the results, e.g. by loading sample details or the calibra-
tion function (→ "Displaying details of single values of samples" p. 83, "Calibration"
p. 90).
Note
In reprocessing, the recalculated results are saved to a new database. For access-
ing the original results file these must be re-opened.
The measurement results and the sequence are extensively displayed in the result
window in the background of the workplace.
The presentation on different tabs in the Results window provides a good overview
of measurement results and statistical analyses.
The following tabs are selectable:
• Sequence/Results (contents of Sequence and Results tab on a single
tab)
• Sequence (displays the current sequence)
• Results (presentation of measurement results)
• Overview (summary of measurement results)
The status bar of the result window shows the file name of the current results file.
Note
After the measurement, you can remeasure a selected sample. To this end, you
must have marked the sample row in the sequence. Then, click on on the
toolbar.
Abs./Time table
The table contains the absorbance values and the statistical analyses according to
the selected method options (Method - QCC window).
Conc.1 table
The Conc.1 table shows the analyzed concentrations of the samples as supplied
to the AAS.
Conc.2 table
The Conc.2 table shows the concentrations of the original samples. In calculating
Conc.2, the sample information data (see Sample ID table below) are considered:
• Pre-dilution
• Weighed portion of solids
• Conversion factors for other units
QC Res. table
The QC Res. table shows the results of QC samples: Rated and actual values of
concentration, recovery rates (not for blank) and reactions to possible deviations
(all types except blank)
Error table
If errors occur during the measurements, the corresponding measurements are
marked in red in all tables. In the Error table, the respective error is indicated in
writing.
Sample ID table
The Sample ID table contains the sample information data (→ "Create information
data for samples and QC samples" p. 63).
The print icon opens the window Print Overview from which the printout pro-
cess can be started. In addition to the printout, the data can be optionally saved as
TXT-, HTML- or PDF files (→ "Creating and printing result reports" p.152).
name and its serial number, also the used analysis technique, the software
version and the optional description are displayed.
If, in later steps of operation, you need the sample information data, activate
the Extract sample information check box.
Conform your choice with [OK].
The results file is being loaded and displayed in the result window.
Note
The name of the currently loaded results file appears in the status bar of ASpect
CS.
Note
Sequences transferred to the process and result window remain displayed there.
Note
Activate the spectra-saving option in the Method - Output window (→ "Output tab
– screen view, printout and saving options", p. 56).
To open the Single values window, double-click with the left mouse button on
the desired sample row of the result table.
Alternatively, right-click on the row of interest of the result table. In the appear-
ing context menu, click on Detail results....
View / Detail re- Another alternative to open Single values window is to mark the respective
sults... row of the table and then activate menu command View / Detail results....
Sample data
Fe303 (Example) Analyzed element line
No. Number of measurement
Name Sample name
Abs. /Ems. Absorbance or emission value (averaged over all single values)
SD Standard deviation (Sigma statistics). This parameter is dis-
played independently of the statistical method chosen for the
measurement (Sigma/Median).
RSD Relative Standard Deviation (Sigma statistics). This parameter
is displayed independently of the statistical method chosen for
the measurement (Sigma/Median).
Date / Time Date and time of the measurement selected in the table
In the graph right of the table, the signal curve of the marked single value is shown
as a function of time. The number of measuring points corresponds to the number
of measured spectra. Spectrum number and Time (read time) were specifically
defined for each element line in the Method - Lines window (→ "Lines tab –
Choosing element lines", p. 22).
If desired, you may manually exclude a single value from the calculation of the
sample average.
To this end, mark the single value to be excluded in the table.
Click on [Exclude] to exclude the value from the calculation of the sample av-
erage for result reprocessing.
To reinclude the previously excluded single value in average calculation, click
on [Reinclude].
Note
By activating the Grubbs outlier test option, outliers among single values can be
detected and eliminated automatically during the analysis.
Buttons
You can select the next line or the next sample of the results list for display by us-
ing the buttons:
Previous sample
Previous line
Next line
Next sample
Graphic presentation
In the wavelength plot, the abscissa is graduated in Pixels. Three red wavelength
values indicate the upper and the lower limit of the measured spectral range and
the position of the measurement pixel (peak of element line). The data points for
background correction are marked by gray lines. The measurement pixel is set off
by a continuous red line. The range of evaluation pixels is marked in light red.
Activates the text mode. Holding the left mouse button de-
pressed draw a frame and type in text, which can be added to
the graph.
A double click on existing text opens the window for editing or
deleting the text. Holding the Ctrl-key and the right mouse but-
ton depressed, you can shift existing text across the graph.
Select button. Activates the selection mode in signal or spectral
plots. On activation, measuring points are marked in the graph.
With the left mouse button, you can add a vertical line cursor to
the graph at the desired position.
BG parameters
Eval. Pixels Number of pixels used for the evaluation of the measured sig-
nal. The integral that represents the measurement result is
calculated over the measured values of these pixels.
dynam. BGC fit Automatic searching of correction pixels.
Spectral corrections
Abs. (Mean) Displays the mean value of the sample absorption (Abs
(corr.)).Using graphite tube technique displays additional the
background absorption (Abs (BG)).
[Line parame- Loads lines parameters for background correction and spectral
ters] evaluation from the method or sends them to the method.
[3D plot] Open an additional window for spectra display.
Select the next line or the next sample of the results list for dis-
play (as in window Single values).
The left side of the Plots tab and its options are identical with those of the Pro-
cessing tab. On the right side, the plot of the family of spectra is shown three-
dimensionally at variable tilts.
The spectrum/pixels selected on the left side is/are marked by a light green on the
plot.
The upper list displays the entries found in the line database for the current spec-
tral region. Selecting a row sets the plot cursor on the corresponding wavelength.
Clicking on a spectral region or dragging with the mouse searches element line at
the selected position in the line database. The element symbol is displayed next to
the cursor and the corresponding list entry is selected.
The lower list shows the entries found in the molecule spectra database for the
current spectral region. Selecting a line displays the corresponding molecule spec-
trum. Factor multiplies the spectrum with the entered value while Offset shifts the
spectrum. The check box Show sum spectrum activates an additional list column
for displaying the sum of selected spectra. These functions may be used for model-
ling complex background spectra.
7.11 Calibration
Calibration is carried out during sample measurement according to the options
selected in the sequence. The calibration curves and functions can be viewed and
edited after the measurement.
Note
After the opening of stored results, the calibration can be displayed only if the
method has been saved along with the results.
Line Select the element line the calibration of which shall be dis-
played.
Calibration func- Display of the used calibration function (set in the Method -
tion Calib. window).
The calibration function is specific for every element line.
You may select a new calibration function from the list box. The
calibration will then be recalculated accordingly.
Curve colors:
Note
The statistical analysis is to be defined in the method.
To exclude individual calibration standards from the calculation, select the standard
in the table by a mouse click and then click on [Delete Std.].
The measurement is only marked as deleted and can be reactivated at any time.
Residuals tab
The graph on the Residuals tab shows the deviations of the calibration points from
the calculated calibration curve and the limits of the prognosis range.
tion and determination can be displayed and printed in the Calibration window.
They are not saved in the result protocol.
Blank method
The standard deviation of the blank is determined within the sample measurement.
For this purpose, the measurement of the blank (QC Blank DL) is included in the
sequence (→"Specifying sequences of samples and actions", p. 60).
Calculation instructions for the blank method:
The blank is to be measured 11 x. From the obtained values, the absolute standard
deviation SD of the blank is determined. The following formulas apply to the limits
of detection and determination:
Limit of Detection (LOD) LOD = 3 * SD / (slope of calibration curve)
Limit of determination (LOQ) LOQ = 10.0 * SD / (slope of calibration curve)
The limits of detection and determination according to the blank method are dis-
played in percent.
QC charts pa-
rameters tab
The content of the QC charts is defined in the QC window on the QC charts pa-
rameters tab and can be customized to the requirements of your laboratory re-
garding the frequency of the entries.
Entries
Number prep. The preparation period is a number of QC chart entries that are
period used for the calculation of control limits (C) and error limits (E).
The preparation period always contains the older chart entries.
If set to 0 (no prep. period), all entered QC data will be includ-
ed in the calculation of control and error limits.
Max. number of Total number of entries on a QC chart.
entries A full chart must be renewed (see below).
Enter scheme All values Enters every performed QC action.
1 value / day Enters the last QC action of the day only.
2 values / day Enters the first and the last QC action of the
day only.
Definition of "day": A "day" corresponds to one day according to the PC clock, i.e.
in the course of a day, any previous entry on the QC chart will be overwritten by a
new QC value; however, when a new day begins, a new entry will be generated.
Chart display
Specification of the type of QC evaluation to be displayed on the QC chart:
Accept prep. Accepts the preparation period of the old chart for application
period, delete to the new chart and deletes remaining values.
remain
Last values > The values of the old chart measured last represent the prepa-
new prep. period ration period of the new chart; all other values will be deleted
from the chart. New measured values will be evaluated based
on the newly created preparation period.
Delete all, new All values will be deleted. New measured values will first fill the
prep. period preparation period.
[Process] Deletes the charts according to the option selected above.
Point size The individual points are displayed as circles. Choose a higher
point size for larger circles.
Connect points Connects the points on the graph with each other by a line.
If you click on a measured value on the graph, a window is opened with the follow-
ing information on this measured value.
Parameters tab
In the Spectrometer – Parameters window, you may check basic device functions
and start a test measurement at a selected wavelength.
Basically, all control lamps must be active and lighting green, when the contrAA
has been switched on and initialized.
Device functions are tested by a click on the corresponding buttons. If the test run
was successful, the lamp lights green.
Instrument data
Display of the connected AAS device and the installed software version.
Lamp
Wavelength
Corrections
Measurement
Read time [s] Entry of the total time for spectral measurement.
[Measurement] Start the measurement based on the adjusted parameters.
Emission meas- If deactivated, the absorbance spectrum will be measured.
urement If activated, the emission spectrum will be measured.
Always open If activated, the measurement in window Spectrometer are
shutter always performed with opened shutter. When checking the
system with closed shutter, e.g. for dark current measurement,
the control has to be deactivated.
Note: The setting has no effect for measurements performed
outside of the window. In all other cases the shutter is automat-
ically controlled.
Always request Performs a reference measurement before every sample
reference measurement.
Spectrum tab
Fig.34 Spectra – Adjust wavelength window: wavelength adjustment for peak offsets
Peak pick
[Delete adjust- Deletes the entry for the current analysis wavelength from the
ment value] line/wavelength file.
8.2 Flame
Selected technique: Flame
In the Flame window, you can test individual functions of the burner/nebulizer sys-
tem and individually adjust the parameters for the elements.
Method Develop- Activate the Flame window using menu command Method Development /
ment / Flame... Flame... or icon button
Status
Siphon The fill level of the siphon (drain trap) below the mixing cham-
ber is monitored. The drain trap serves to drain non-nebulized
liquid.
OK The drain trap is filled with liquid up to the overflow
port.
Error The fill level of the drain trap is insufficient.
The drain trap must always be filled sufficiently to prevent
backfiring of the flame particularly that of the nitrous oxide
flame. In case of error prompt remove the burner head. Fill
distilled water in the mixing chamber till the error prompt disap-
pears.
Waste bottle Fill level of waste bottle. (If you do not use a waste bottle with
fill level sensor, plug the supplied jumper plug into the corre-
sponding connector).
OK The waste bottle is not full (correct state).
Error The waste bottle is full and needs to be emptied.
Measurements are possible; on starting a meas-
urement, however, a warning message will be dis-
played.
Pressures
Function tests
The test functions are accessible only if the flame is extinguished. The availability
of the test functions depends on the context.
Flame / Scraper
Settings
In the Settings field, you may edit the gas flow settings:
Note
The attached burner is automatically detected by the burner sensor. Switching be-
tween C2H2/air and C2H2/N2O is possible only if the flame is burning, and if the
50-mm burner has installed.
Manual optimiza-
tion tab
Line Choose the element line for which the flame shall be optimized.
You may choose among the wavelengths defined in the meth-
od.
[Set] Moves the monochromator to the selected wavelength.
Fuel [L/h] Adjustment of fuel flow.
Burner height Adjustment of burner height relative to the optical axis of the
[mm] light path of the xenon lamp.
Oxidant (total) Adjustment of total oxidant flow, sum of the flows of oxidant
[L/h] and auxiliary oxidant.
Oxidans [L/h] Display of the oxidant flow in L/h.
Fuel/Oxidant Display of the fuel/oxidant flow ratio.
[Start] Starts the measurement. The signal is continuously being rec-
orded.
The button label turns into [Stop].
[Stop] Stops the measurement.
[Accept values] Accepts the found flame parameters for the selected element
line and transfers them to the method.
Graph Display of signal curve.
Ordinate expansion is possible via the input fields at the Y-axis.
Absorbance Display of current absorbance value.
Maximum Display of the maximum absorbance value of the running
measurement.
Procedure
1. From the Line list box, choose the desired element line.
2. Click on [Set] to drive the monochromator to the wavelength of this line.
3. Immerse the aspiration hose of the nebulizer in the sample solution.
4. Start the measurement with [Start].
5. Vary the Fuel setting by means of the arrow buttons and watch the signal
change in the graph and in the Absorbance field. Adjust the fuel rate until ab-
sorbance is maximal.
6. In the same way, adjust the Burner height, until you have found the maxi-
mum absorbance.
7. When using the flame with auxiliary oxidant, vary the Oxidant (total) parame-
ter until again you obtain the maximum signal level.
8. Repeat steps 5 to 7, until you can no longer notice a relevant increase of the
signal level.
9. Finish the measurement with [Stop].
10. Click on [Accept] to transfer the found parameters for the selected element
line to the method.
In the same way, find the appropriate parameters for all element lines defined in
the method.
Note
There is a small delay between the modifation of the parameters and the signal
response.
The used optimization algorithm changes the settings for fuel flow and burner
height with the aim to obtain an increase in the signal level. This algorithm takes
into account the interactions between the two parameters (fuel flow and burner
height). The optimization procedure will be stopped if in three successive steps no
further increase in signal level is achieved. This means that either the maximum
signal level was reached or the change of the parameters is of no effect on the
signal level.
Line Choose the element line for which the flame shall be optimized.
You may choose among the wavelengths defined in the meth-
od.
Position Position of the test sample on the autosampler.
[Start] Starts the measurement. The signal is continuously being rec-
orded.
The button label turns into [Stop].
All lines Perfoms the optimization for all lines of the method with the
sample solution. The sample must contains all analytes.
[Stop] Stops the measurement.
[Delete] Deletes the found values.
[Load] Loads optimized flame parameters.
[Save] Saves optimized flame parameters.
[Accept values] Accepts the found flame parameters for the selected element
line and transfers them to the method.
5. For single line optimization: After successful optimization the parameters can
be transferred to the method by using [Accept values] if the respective set-
ting has not been activated in the start dialog.
8.3 Furnace
Selected technique: graphite tube technique
Working in the Furnace window, you are able to check individual graphite tube
furnace functions and make settings for analysis of individual elements.
Method Develop-
Use a Method Development / Furnace... command or click the
ment / Furnace...
Furnace Program
tab
Tabular display
A tabular display will list all working steps which are part of a currently selected
furnace program. The parameters on display are:
Program steps:
Temperature settings:
Gas supply
Inj. If marked with a "*", the particular sample (gas in HydrEA tech-
nique) will not be introduced into the graphite tube before com-
pletion of this step (pipetting into preheated tube).
Selections for modifiers for matrix delimitation and thermal pretreatment must be
made for each specific line. Up to five modifiers can be specified for analysis of a
given element line.
Matrix modifiers
During each measurement, a maximum of five different matrix modifiers can be
added. These can be activated by clicking into the relevant modifier checkbox.
In order to prevent errors due to carryover effects, analytical components are rec-
orded in the following standard order:
• Blank (in case of dilution)
• Modifier 1
• ... further modifiers (if specified)
• Sample solution.
Outputs to the graphite tube occur in reverse order, i.e. the sample will be the first
to be injected. As the other components are being supplied, residual sample matter
will be washed from the dosing tube and injected into the graphite tube. This
standard order of sample (first) and modifiers (then) may be modified if necessary.
Required parameter entries for modifiers are:
Enrichment
For enrichment, a given furnace program is repeatedly performed until the speci-
fied step is reached (column E/P). As part of each cycle, the sample quantity which
is specified per sample table will be injected and pretreated, the tube is then cooled
to room temperature and the next sample volume is injected. This procedure al-
lows greater sample volumes to be placed into the furnace. Modifier volume is in-
jected only once.
Specificable enrichment modes are:
Thermal pretreatment
For thermal pretreatment of analytical solution and/or modifiers, this option will be
executed up to the specified working step of a furnace program. At the end of this
step, the remaining components will be injected into the preheated tube.
For SSA 600 with liquid dosing, a modifier or liquid samples are automatically
pipetted. Up to three modifiers may be specified in this case.
Results display
The results of optimization can be displayed in a results window.
Graphic representation is facilitated by Furnace – Furnace program window tools.
Available for displaying are the particular autozero value and the measured extinc-
tion value:
Note
The graphic are confirm with presetting in the Option – View window.
Performing optimization
An autosampler is required to perform this series of measurements.
Optimization goal
Use [<] / [>] control buttons to move the vertical cursor over the graphic area
or use arrow keays as necessary to adjust optimal temperature. Then click
[Accept] to transfer this value into the furnace program.
Repeat this procedure for all other analytical lines included in the current
method.
The plot shows the various steps of a furnace program. The red line indicates the
actual furnace temperature during program execution.
During program execution, the following values are output for digital display:
Once the Graphite tube coating checkbox is marked, the following input parame-
ters will be enabled for selections to run this procedure.
Note
Before removing a graphite tube which is still in a state fit for operation, you are
recommended to record its operating data from the Graphite tube subarea in or-
der to be able to restore these values for automatic subsequent updating after the
tube has been installed again.
Clean furnace
Clean furnace is a single-step cleaning program. It works with a gas rate setting
that is fixed to maximal flow. Current parameter settings which can be modified
here will be saved with the method.
Test furnace
Checks the furnace for sensor errors. A measurement will be aborted on occur-
rence of an error situation listed here (with a corresponding error message output
to the screen).
To trigger an error test, click [Test]. On successful completion of testing, the result
will be reported by a green control lamp, on unsuccessful completion by a red con-
trol lamp.
Pump control
Note
If on switch on of one of the two pumps none of the two valves 3 or 4 is active,
valve 3 will be switched on automatically to prevent the liquid from flowing back.
If the sample pump is being active, the components pump will be activated, too, to
avoid liquid impoundage in the gas/liquid separator.
Control of cell
Heat value Parameter for the bake-out temperature of the gold collector.
The value is factory-adjusted. It should not be varied unless the
thermal behavior of the gold-collector heating has changed. A
higher value corresponds to a higher temperature.
[Set] Saves the displayed heat value as new value in the hydride
system.
Error test
The result of the test is indicated by green (for a successful test) and red (test
failed) pseudo LEDs:
To start the error test, click on the [Test] button.
A negative test result may have the following causes:
8.5 Autosampler
8.5.1 Autosampler AS 51/52s
The autosampler is an optional accessory. You may use device-bound au-
tosamplers depending on the operating mode of the spectrometer and your specific
Method Develop- analytical needs. The type of involved autosampler is identified by ASpect LS as
ment / part of an initialization routine.
Autosampler... In the Autosampler window,
• the connected autosampler model is displayed,
• the autosampler can be adjusted,
• the autosampler can be additionally rinsed / initialized again,
• various functions can be tested.
The parameters directly relevant to the analysis (charging of sample tray, dilution,
mixing and wash steps) are to be specified in the method, the sequence and the
sample identification data.
Open the Autosampler window using menu command Method Development
/ Autosampler... or the icon button
The parameters for washing and controlled cleaning are taken from the currently
loaded method (→ "Auto sampler for flame and Hydride/HydrEA technique" pg.
32). Conversely, however, changes to the parameters made on the Autosampler
– Parameters tab, are of no influence on the entries in the method.
Autosampler type
Wash
Wash mode off Wash mode switched off. No rinsing performed au-
tomatically.
after sample Washes after each sample, but not within a sta-
tistical series.
Wash time Wash Time in which the rinsing agent is aspirated in the wash cup.
cup
Mixing cup cy- Number of rinse cycles for the mixing cup.
cles In a rinse cycle the mixing cup is filled with wash liquid / diluent
and then emptied again.
Controlled cleaning
Special functions
Wash mix cup Wash mix cup separately outside a sequence measurement.
Volume Enter volume for cleaning.
Start Wash mixing cup.
Techn. Parame-
• depth of submergence of cannula / dosing tube into the various cups
ters tab • working speed of dosing unit
• alignment of autosampler arm in relation to sample cups
For the individual cup types the following actions are taken into account:
Cup Action
Sample cup Take up of sample by dosing unit or aspiration by nebulizer
(flame technique) or peristaltic pump (hydride technique).
Special cup Take up or aspiration of special samples.
Mix cup Dosing volume of analyte and diluent, as well as take up of sam-
ples after dilution.
Wash cup Rinsing of cannula and aspiration path (only with flame and hy-
dride technique).
Table subarea
Using the controls in the Table field, you can change the parameters of the select-
ed table row.
Tracker/Rotator
Driving the sampler arm to the various positions:
Cup No. Drives to the cup position selected from the corresponding list
box.
Wash position Drives to the wash cup.
Mixing position Only for AS 52s
Drives to the mixing cup
Pipetter
Only AS 52s
Dipping arm
Lowering the sampler arm at the position selected under Tracker/Rotator.
Depth Adjust the depth by which the sampler arm can be lowered.
Pumps
Wash pump Start the pump for supply from wash cup.
Mixing cup Starts the pump for pumping out the mixing cup.
pump
Test programs
Tests are provided to test the autosampler with pre-configured, dry running test
programs. The cups the sampler arm is driven to during the test must be empty!
When the test programs are finished, you are informed of the test success.
Start the selected test program with the button [Start]
Error test
These tests serve to test the autosampler for sensor errors. If one of the error
states listed here appears, every measurement will be aborted (on the screen a
corresponding error message will be displayed).
The error test is started with the [Test] button. If the test was successful, the indi-
cator will light green; if the test fails, it will light red.
If a test fails, this may have the following causes:
Wash bottle lev- Reports excessively low fill level of storage bottle (bottle emp-
el ty).
Diluent bottle Reports excessively low fill level of storage bottle (bottle emp-
level ty).
Tracker/Rotator Swivel drive of sampler arm and rotary drive of tray are defec-
tive.
Tray identifica- Sample tray not detected.
tion
Pipettor (drive) Dosing pump error.
Pipettor (vol- The volume taken up by the pipettor was too large.
ume)
[Adjust sampler] Opens the Adjust sampler dialog for readjustment of the au-
tosampler.
(→ section "Adjust autosampler AS 51/52s" pg. 135)
Depth Dipping depth in mm. With this field, the dipping arm can be
driven into or out of the particular cup. The immersion depth for
Mixing position and Tray position 1 is no optimizable parame-
ter. The possibility of changing the depth only serves to facilitate
assessing the cup center.
Dipping arm Buttons for changing the sampler arm position. Alternatively,
[]/[] you may press arrow keys "Arrow right" and "Arrow left" on the
PC keyboard.
Sampler tray Buttons for rotating the sampler tray. Alternatively, you may use
[]/[] the arrow keys "Arrow up" and "Arrow down" on the PC key-
board.
[Save] Saves the new parameters for the selected position. The set-
tings remain valid until you save new adjustment states.
Wash pump
Only samplers with newer firmware versions
Autosampler type
Wash modes
Wash mode off Wash mode switched off. No rinsing performed automat-
ically.
after sample Washes after each sample, but not within a sta-
tistical series.
between runs Washes after each measurement, including
within a statistical series.
Controlled cleaning
Special functions
Wash mix cup Rinse mixing cup separately outside the measurement.
Volume Enter volume for cleaning.
Start Wash mixing cup.
Techn. Parame-
• depth of submergence of cannula / dosing tube into the various cups
ters tab • working speed of dosing unit
• alignment of autosampler arm in relation to sample cups
• the automatic depth adjustment for volume decrease during the analysis
• the alignment of the MPE 60 to the graphite tube furnace.
For the individual cup types the following actions are taken into account:
Cup Action
Sample cup Take up samples through doser.
Special cup Take up special samples.
Mix cup Dosing volume of analyte and diluent, as well as take up of
samples after dilution.
Wash cup Rinsing of cannula.
Graphite tube Inject sample or special sample into the graphite tube.
Table subarea
Using the controls in the Table field, you can change the parameters of the select-
ed table row.
Automat. depth Automatically adapts the depth, to which the dosing tube will
correction submerge into sample or special cups, to a new volume.
MPE alignment
With regard to individual special cups or sample cups, the following parameters
can be specified:
Position Set cup position on the tray, for which the three displayed val-
ues: volume, depth and diameter apply. Settings must be made
individually for each cup that is intended to be modified.
Volume Displays the amount of sample volume already retrieved from a
cup or allows input of lacking sample volume balance into a cup
not completely filled.
This reading is software-updated after each sample take-in se-
quence.
Depth Displays the amount of additional depth due to overall sample
withdrawal up to this moment.
This value is recalculated after each sample take-in sequence.
Total immersion depth is the sum of a specified immersion
depth (Depth column on Techn. Parameters tab) and the addi-
tional depth shown here.
Diameter Indicates the cup diameter.
This value is used as input for calculation of the depth, based
on the amount of withdrawn volume.
Diameter For cups with dimensions at variance against those of standard
checkbox cups.
Allows you to specify a diameter for each cup individually if
marked.
[Delete vol- Resets volume values for all special cups or sample cups to 0.
umes]
[Reset] Resets all volumes and the depth values of all cups to 0, resets
diameters to their latest via [OK] saved values.
Sample cups Applies settings for maximal immersion depth to sample cups if
marked.
Special cups Applies settings for maximal immersion depth to special cups if
marked.
Depth [mm] Defines the maximal allowed immersion depth of the au-
tosampler into a sample cup or special cup.
Caution:Please note that the autosampler will at once dip
down to the new depth level if the Set checkbox is marked!
Make sure that the autosampler motion path is not blocked.
A modified setting for immersion (dipping) depth should be
saved with [Save].
Position The sample tray position in which immersion depth settings are
checked.
Set The dosing tube will dip into a cup that was specified via Posi-
tion and to a depth that was set via Depth [mm]. The immer-
sion depth can be visually checked. Caution:When the check
box Adjust is marked, the autosampler will be dipped immedi-
ately to the given depth! Make sure that the autosampler mo-
tion path is not blocked.
Tracker/Rotator
Driving the sampler arm to the various positions:
Cup No. Drives to the cup position selected from the corresponding list
box.
Wash position Drives to the wash cup.
Mixing position Move the mixing cup (for MPE 60).
Tube position Move graphite tube.
Pipetter
Speed Adjust the pipetter speed (→ "Set insertion depth and dosing
speed of the MPE 60" pg. 141).
Volume [µL] Adjust the pipetting volume to be taken up.
Valve to bottle Switches the valve to the bottle.
The valve switches the flow between the diluent bottle and the
sample. In switching, you must hear the valve click.
[Take up] Take-up the selected volume with the currently selected pipetter
speed.
[Sispense] Dispense the selected volume with the currently selected pipet-
ter speed.
[Reset] The volume taken up is dispensed again.
Dipping arm
Lowering the sampler arm at the position selected under Tracker/Rotator.
Depth Adjust the depth by which the sampler arm can be lowered.
Pumps
Wash pump Start the pump for supply from wash cup.
Mixing cup pump Starts the pump for pumping out the mixing cup.
Test programs
Tests are provided to test the autosampler with pre-configured, dry running test
programs. The cups the sampler arm is driven to during the test must be empty!
When the test programs are finished, you are informed of the test success.
Start the selected test program with the button [Start]
Error test
These tests serve to test the autosampler for sensor errors. If one of the error
states listed here appears, every measurement will be aborted (on the screen a
corresponding error message will be displayed).
The error test is started with the [Test] button. If the test was successful, the indi-
cator will light green; if the test fails, it will light red.
If a test fails, this may have the following causes:
Wash bottle level Reports excessively low fill level of storage bottle (bottle emp-
ty).
Waste bottle lev- Fill level of waste bottle is too high (bottle is full).
el
Tracker/Rotator Swivel drive of sampler arm and rotary drive of tray are defec-
tive.
Tray ident. Sample tray not detected.
Pipetter (drive) Dosing pump error.
Pipetter (volume) The volume taken up by the pipettor was too large.
Techn. Parame-
An adjustment must be carried out in exceptional cases (e.g. following inappropri-
ters tab
ate transport), if the sampler arm no longer immerses centrally in the cups. Read-
justment is carried out computer-controlled in the Adjust sampler dialog (accessi-
ble via the [Adjust sampler] button on the Autosampler – Functional Test tab).
Depth [mm] Dipping depth in mm. With this field, the dipping arm can be driv-
en into or out of the particular cup. The immersion depth for Mix-
ing position and Tray position 1 is no optimizable parameter.
The possibility of changing the depth only serves to facilitate
assessing the cup center.
Dipping arm Buttons for changing the sampler arm position. Alternatively, you
[]/[] may press arrow keys "Arrow right" and "Arrow left" on the PC
keyboard.
Sampler tray Buttons for rotating the sampler tray. Alternatively, you may use
[]/[] the arrow keys "Arrow up" and "Arrow down" on the PC key-
board.
[Save] Saves the new parameters for the selected position. The settings
remain valid until you save new adjustment states.
Sample preparation items which are different from those in solution analytics (de-
termination of sample platform deadweight, dosing and weighing of sample) can be
specified on the Solid tab of the result window (→ "Solid tab" pg. 81).
Status / Buttons Displays SSA 600 status and button that was actuated since the
latest query in appropriate color (green, orange, red).
Using [Update], the status can be queried again and the button
display be updated.
Move to posi- Performs motion to the position selected via the Pos. list field.
tion Thus no platform is taken up or put down.
Rotate tray Rotates the sample tray into a selected position.
Transport Means that SSA 600 transports a platform from a given position
to the next starting and target position.
If Take-up platform is activated, the gripper will pick up a plat-
form.
If Put-down platform is activated, the gripper will place a plat-
form down in its target location.
Gripper Opens or closes the gripper.
Lower cannula Moves the cannula downwards.
Balance Determines the weight of a platform in a specified position on
the tray (Pos.)
Weighing with tare
Before a platform can actually be weighed, the scale
must be tared. #1 indicates the tared weight. #2 con-
tains the weight of the platform (also with dosed
sample as may be the case).
[Internal calibration]
During a calibration cycle of this type, the internal
calibration graph of the scales is redetermined.
To achieve this, the scales are initially reset, their
zero-point is determined and an internal weight is
weighed. The values obtained for zero-point and in-
ternal weight will provide the input for determination
Adjustment tab
• Control and adjustment of motion to individual autosampler positions
• Adjustment of the SSA 600 to the position of the graphite tube furnace
• SSA 600 with dosing unit: Automatic depth correction for taking in matrix
modifiers and special liquid samples.
• Test of liquid dosing
.
On the window following information are shown:
• name of the current result file
• information about the selected result file
• all method versions that belong to the results file (only if the method is
currently loaded or it was saved along with the results file)
For printing of a stored result with open the standard window open and
choose the result file.
Print window using [Print].
Select output target and report options (→ "Selecting output target and report
options" p. 153).
Start the printout using [Print].
Note
The PDF converter available in ASpect CS does not convert graphics (spectra or
calibration curves). If you should need these properties, install your own PDF con-
verter. Select Printer as target medium and choose the PDF converter as the
printer to be used.
Results tab
On the report, only the selected options will be included.
Types
Sample type Adds the measurement results for the sample type to the re-
port.
Calibration Adds the calibration results to the report.
QC and other Adds the measurement results for QC samples and further
samples sample types to the report.
Actions Adds performed actions to the report. This settings documents
the Special actions defined in the sequence (pause, display-
ing calibration curves).
Results
Signals Absorbance signal vs. time curve (only for the 1st measurement
of a statistical series).
Page break
Signals
In the Y-scale (Abs.) field, you can set the scaling for the graphic presentation of
the measured signal.
Other tab
In the Userdefined header text field, you may enter a text of up to 125 characters.
If you activate the Page break after method parameters option, a page break is
made after the presentation of the method parameters.
If you activate the Open saved document in application option, result reports
saved as TXT, HTML, or PDF file will be opened in the application associated with
the corresponding file extension.
With the setting selected in the Scaling [%] field you can vary the size of the report
printout. In this way it is possible to adapt it to other paper sizes or your specific
analysis reports.
Print preview
In the Print window, click on [Print preview...] to open the Print preview win-
dow.
The Print preview window contains options for instant printout and page zooming.
In this window, you can manage methods, sequences and results. Choose the de-
sired type of data from the Type list box.
Note
To select several data records of the database window simultaneously, hold the
Ctrl or Shift key depressed.
Open the Data – Data management window using menu command Extras /
Data... or icon button
Extras / Data...
.
From the Type list box, choose the type of data record to be managed: Meth-
od or Sequence.
Click on [Delete].
The database window is being updated displaying only the remaining data rec-
ords.
With data records of the same name, the version number is kept. The versions
will not be renumbered.
.
From the Type list box, choose the Lines/wavelength file option.
Click on [Copy…] to open the Select directory window.
Choose the target folder and confirm your choice with [OK].
The lines/wavelength files are being copied to the selected target folder.
Edit
Copy visible Copies the visible sample results of the current table.
columns only
(Strg+C)
Copy all col- Copies the sample results of all tables.
umns
Column titles If activated (check mark), the copy action includes the column
headers.
Select the samples from the desired table of the results list.
Holding the Ctrl or Shift key depressed, choose the samples by mouse clicks
on the respective row.
To select all rows, use menu command Edit / Select All.
If necessary, activate menu command Edit / Column Titles to have the col-
umn headers included in the copy action.
Activate the desired menu command to copy the results to the Windows clip-
board.
10 User Management
An optional software package for ASpect CS user management is available. It pro-
vides extended functionality for operation of ASpect CS in accordance with the
FDA requirements on Electronic Records & Electronic Signatures (21 CFR Part
11).
The user management package allows for one administrator level and four user
levels. A user with ASpect CS administrator rights may perform any of the following
action with the help of this user management package:
• flexible system configuration (code word and login gidelines, audit trailing,
signatures, data directories)
• creation of a user level for each user with a stepped pattern of user rights
• assignment of code words for access to ASpect CS software
• assignment of personalized working directories for methods, sequences
and results to users
• inspection and exporting of created audit trail (events log sheet)
Once the user management package is installed and configured, the System
menu item in ASpect CS will be active. You can use this menu item to access one
or more desired functions of user identity management.
Any change in user data will be permanently saved in an encoded data base on
exiting a pertaining window.
Note
In order to meet safety requirements, a Microsoft Windows 2000/XP operating sys-
tem with adequate configuration resources must be available. This applies to file
access rights and other setting actions of a kind that should be performed by an
authorized system administrator.
Level 1
Level 1 users have unlimited access to all Aspect CS functions, but are denied
access to user identity management.
Level 2
Same as level 1 users, except:
• Deletion of methods (M1 ID code)
Level 3
Same as level 2 users, except:
• Saving of methods (creating methods in a method data base) (M2 ID code)
• Saving of sequences (creating sequences in a sequence data base) (P2 ID
code)
• Changes in records regarding wavelengths (W1 ID code)
Level 4
Same as level 3 users, except:
• Changes in method parameters (E1 ID code)
(users of this category can only load previously created methods and sequences
and perform measurement).
• ID code will be referred to in operating advice.
System / User Trigger a System / User Management... menu command or launch user man-
Managment agement via an entry button of this name in the starting menu.
On completion of login procedures, the User Management window will open.
The window contains a list of registered user names with assigned passwords. It
shows the user profile details for a selected user in its right-hand subarea.
Control buttons
[New ...] Creates new user.
Opens the Add user data window.
[Set default set- Sets the default settings (factory settings) for a given selection.
tings] The settings for other areas will be unaffected.
Login/password
Number of login Shows the number of invalid login attempts (max. 10).
attempts On excession of the maximum allowed number, the ASpect CS
session will be terminated after a cetrain waiting time and must
be restarted for another login. An entry (warning) is added to
the audit trail file.
Minimum user Defines the minimum prescribed number of characters for a
name length newly created user name
Maximum number of characters: 10
The ASpect-CS working directory and the audit trail directory provide specification
options.
Paths
Audit Trail
Inactive (no en- No entries will be added to the audit trail file (event log sheet).
tries)
Active Entries will be added to the audit trail file (event log sheet).
Allow measuring When active, this checkbox allows a measurement to be trig-
only with saved gered in ASpect CS only if a method was loaded and no
methods change was made in this method since it was saved for the last
time.
Signatures
User-specific If active, this box will create the user’s own working directory in
data path this format: \ASpect-Working directory\User name.
As part of an initial login to ASpect CS the appropriate directory
structure will be created.
Use e-signature Allows the user to electronically sign for measured results if
active.
Disable user ID Prohibits the use of a user name if active.
User names can be temporarily disabled. A disabled (as op-
posed to a removed) user name cannot be reassigned to an-
other newly created user.
User must Will prompt the user to change his/her password as the next
change pass- ASpect CS session is triggered.
word at next lo-
gon
Note
The size of an audit trail file has no influence on system speed. However, for a file
size of 20 MB and beyond, a note will be displayed, prompting you to archive this
file and delete your entries.
[Yes] Deletes the audit trail file. The audit trail list is empty and
begins to be filled with new entries.
[Cancel] The audit trail file will not be deleted. New entries will be
added to the existing list.
System / Change Select the System / Change password... menu command in Aspect CS.
password
The Change password dialog screen opens.
Enter your previous password at the input field, then enter the new password
twice and click [OK].
On successful entry, a “Password changed“ message will appear.
Issued by States the full user name and ID user of the individual who
signed this file.
Signed on: Date/time of signature granting
Status The signature state may take on one of the following meanings:
Valid Signature and result data are complete and correct.
Calculated checksums of a file reveal no variance
against the checksums contained in the signature file
at the moment of signing.
Invalid (signature file missing or invalid)
Failure to find signature file that belongs to data set
or signature file contains faults.
Invalid (TPS data)
Results file was modified following signature. Com-
parison between newly calculated checksums and
previously saved checksums reveals variances.
Invalid (SPK data)
A file with raw spectrum data was modified after be-
ing signed. Comparison between newly calculated
checksums and previously saved checksums reveals
variances.
Meaning The meaning of signatures
Remarks For comment text (optional)
11 Supplement
11.1 Overview of markings used in the display of values
Remark Meaning Values Output
> Cal The mean value is larger than the working Mean val- Sequence and
range of the calibration curve. ues results window
< Cal The mean value is smaller than the working Mean val- Sequence and
range of the calibration curve. ues results window
< LOD The value is smaller than the limit of detec- Mean val- Sequence and
tion. ues results window
< LOQ The value is smaller than the limit of deter- Mean val- Sequence and
mination and larger than the limit of detec- ues results window
tion.
RSD! The mean value of sample or standard is Mean val- Sequence and
outside the range of the specified relative ues results window
standard deviation.
RR! The mean value of sample or standard is Mean val- Sequence and
outside the range of the specified relative ues results window
range.
Factor! Limit of recalibration factor for the calibra- Calibration Sequence and
tion curve was exceeded curve results window
R2(adj.) The quality of regression R2(adj.) of the Calibration Sequence and
calibration curve falls below the specified curve results window
value. Calibration
curve window
#MAN. The single value of sample or standard was Single val- Single values
manually excluded from the calculation of ues window
the mean value.
#COR. The single value of sample or standard was Single val- Single values
automatically excluded from the calculation ues window
of mean values by the Grubbs outlier test
Abscorr=lg(I0/Ipeak)
I0 – Mean value of the pixels within the slit, excluding the measurement pixel +/-
10 pixels.
Ipeak – Measurement pixel
Limits
Flame and other molecule structures that would be covered by a separate refer-
ence spectrum are not compensated. Due to the inhomogeneous CCD illumination,
the absorbance spectrum may have a falling or a rising baseline.
This correction corresponds to the broadband D2 correction of line AAS.
Static:
The BG correction points are set manually or from a list in the range of center pixel
+/-0.5 x measuring range.
Dynamic:
The BG correction points are found by an algorithm.
• Aim: Identification of those pixels that do not significantly differ from base-
line noise. To this end, the gradients of the observed pixel to the +/-3
neighboring pixels are calculated and tested for gradient change.
The BG correction points of the single spectra are summarized. If a certain
BG correction pixel is found with a defined frequency, this pixel will be
used as BG correction point for the mean-value spectrum.
• Side conditions: If all BG correction points lie on one side of the measure-
ment pixel, an error message will be output. In the case of one-sided cor-
rection, the extrapolation of the polynomial may lead to large errors. The
correction point algorithm will detect a gradient only if the absorbance dif-
ference between four adjacent pixels exceeds a defined amount (>2E-
4/pixel).
• The range of the measurement pixel +/- 10 pixels is excluded from the
search.
Limits
The algorithm used has fundamental problems with overlapping structures extend-
ing over the entire range of the spectral slit, as in this case no region is dominated
from baseline noise. Here, a dynamic fit of frequency is carried out, i.e. the deci-
sion criterion is gradually relaxed. If this does not result in a sufficient number of
BG correction points, static pixels will be set, which are saved in the line table as
presetting.
Assumptions
• The sum spectrum is the weighted sum of the spectra of the pure sub-
stances and the noise.
• Other wavelength regions (e.g. additional analyte bands that do not exist in
the correction spectrum) do not exist, or they are excluded.
• No or only very little wavelength shifts between sample and correction
spectrum as these lead to artifacts.
The flame structure spectrum is calculated from the reference spectrum according
to Abscorr=lg(I0/I0-Offset); for further molecule corrections, the spectra of the pure sub-
stances are required.
The basis for the calculation is the multivariate classical calibration:
y=X*b+e m – Number of wavelengths/pixel
n – Number of pure substance spectra
y – Sum spectrum (mx1)
X – Matrix of pure substance spectra (mxn)
b – Coefficient vector
+ + -1
b(estimated)=X y with X =(X’*X) X’*y (pseudo inverse)
The product of pure substance spectrum and coefficient can then be subtracted
from the sample spectrum:
Note
Refer to Welz et al.: ”High-Resolution Continuum Source AAS“ for potential inter-
fering elements or molecular bands.
Example
The zinc line at 213.857 nm may be compromised by NO molecular bands or an
iron line (see diagram view). NO molecular bands may be caused, among other
reasons, by high concentration of HNO3 in a sample solution.
Note
The concentration values of matrix components need not necessarily match those
in the respective samples. But concentration values must be high enough for
mean-value spectra to clearly show extinction values. As a compulsory precondi-
tion for correct spectral correction, only one component may be measured as pure
substance at a time.
Example
For a zinc line that is overlapped by HNO3 and iron, an HNO3 solution of higher
concentration (e.g. by 5%) and an iron solution (some hundred milligrams per liter)
will be measured as sample.
Assign a name to the spectrum in question and complete this process by se-
lecting [Save]. For testing the model the check box “Test model (no save
prompt)” can be selected.
Repeat these working steps for each matrix component.
Use the [Add] button to open a selection screen of previously saved correction
spectra. Use a mouse-click to select a correction spectrum and complete this pro-
cess by clicking [Load]. Repeat these working steps for each individual correction
spectrum.
Activate the Highlight corrected spectrum option in order to check the result-
ing sample spectrum for absence of overlap effects.
For testing the model select the check box Test model (no save prompt) and
return with [Close] tot he window Spectra.
Note! The model is not saved!
Select again [New/modify model] to open the Spectral corrections window.
Click [Save] to open the window for saving a correction model. Assign a name
to the model you want to save and complete this process by selecting [Save].
Note
By clicking the [Mask.] button and keeping the left mouse-key depressed at the
same time, you can mark certain ranges of a spectrum you want excluded from
calculation of a correction model. By default, the analyte line range is masked (pix-
els 92 to 110). More ranges may have to be masked in cases where no pure sub-
stances were available for determination of correction spectra and where impurities
may occur in varying portions.
Open the current Method screen (via icon), then open the Evaluation
tab.
Select Spectral corrections in this screen to open the related window.
Select a desired Model in the analyte line dialog field and exit this window via
[OK].
Once a method has been saved, all subsequent measurements using this method
will integrate and follow this generated correction model. Previous measurements
can also be re-calculated with the help of this method so no new measurement is
required.
12 Index F
Flame
A Adjustment 105
Automatic optimization 110
Absorbance measurement 42 Manual optimization 108
Adjust autosampler AS 51/52s 135 Flame window 105
Ashing 114 Furnace program
Atomize 114 Graphic 120
Autosampler Trial run 120
graphite tube technique 35 Furnace Program
Solid analytics 37 Edit 112
Autosampler AS 51/52s Furnace window 112
Dosing speed 132
Insertion depth 132 G
Autosampler MPE 60
Adjustment 146 Gas adaption 114
Depth adjustment 141 Graphit tube
Dosing speed 141 Coating 121
Insertion depth 141 Graphite tube
Autosampler SSA 600 Clean 124
Depth adjustment 141 Graphite tube
Autosampler window 129, 137 Sensor error 124
Autosamplers Graphite tube furnace
Flame technique 32 Format 123
Hydride and HydrEA techniques. 32 Graphite tube lifetime 123
Graphs 9
B
H
Background correction 87
Burner Hydride system
Function test 105 Cleaning 125
Buttons Heating the cell 126
generally used 8 Initialize 125
Loading 125
C Window 125
Calbratione curve I
LOD/LOQ tab 93
Parameters tab 93 Injection step 114
Residual tab 93 Integration mode 22
Calibration Intensity 102
Delete Standard 93
Calibration table L
Automatic preparation 46
Limits of detection 93
Manual preparation 45
Limits of determination 93
Cleaning
Line identification 90
controlled 34, 36
Line/Wavelength file 103
Lines
D User-defined 24
Depth correction Lines/wavelength file
MPE 60 140 View 161
Dosing speed Lists 8
MPE 60 140
M
E Menu command ?
Energy 102 Contents and Index 7
Enrichment step 115 Search ... 7
Menu command Edit