Anton-Paar-Dma-48-Instruction-Handbook (Densimetro)
Anton-Paar-Dma-48-Instruction-Handbook (Densimetro)
Anton-Paar-Dma-48-Instruction-Handbook (Densimetro)
DMA 48
Density / Specific Gravity /
Concentration Meter
Instruction Handbook
DMA 48
INSTRUCTION HANDBOOK
Neither is any liability assumed for damages resulting from the use of the information contained
herein.
Anton Paar GmbH does not make a commitment to update the information contained herein.
/&> 2. SO
CONTENTS
1. INTRODUCTION 7
2. SAFETY INSTRUCTIONS 8
8. ADJUSTMENTS 28
A25IB01
10. MEASUREMENTS 45
A25IB01
1. INTRODUCTION
A25IB01
2. SAFETY INSTRUCTIONS
• This handbook does not purport to address all of the safety problems
associated with the use of the instrument and samples. It is the
responsibility of the user of the instrument to establish safety and health
practices and determine the applicability of reguiatory limitations prior to
use.
• Anton Paar GmbH warrants the proper functioning of the DMA 48, if
no unauthorized adjustments have been made in mechanical parts,
electronic parts and software, and the following points are adhered to.
• Do not use the DMA 48 for any purpose other than described in the
handbook.
A25IB01
• Due to the nature of the measurement the measuring results not only
depend upon the correct use and functioning of the DMA 48, but may
also be influenced by other factors. Therefore it is advised, that the
analysis results are plausibility tested before consequential actions are
taken.
A25IB01
3. SYMBOLS IN THE HANDBOOK
10 A25IB01
4. UNPACKING AND CHECKING THE
SUPPLIED ITEMS
HINTS:
• The DMA 48 has been tested and packed carefully before shipment.
However, damage may occur during transport.
1 Keyboard 62414
1 Power Cord Europe 65146
or USA 52656
or UK 61865
1 Instruction Handbook English 67796
or German 64532
1 Accessory Box 64947
4m Hose 3 x 5 mm Silicone 50814
2 Fuses 1.6 A 55500
3.15 A 51015
8 Syringes 2 ml, Luer tip 51974
2 Adapter Record UNF1/4" 63275
1 Adapter Record Luer Female 63276
1 Adapter Record Luer Male 64780
2 Adapter Luer 1/4" UNF 64792
2 Adapter UNF 1/4" Luer Male 64793
2 Male Luer Plug 63865
2 Adapter Luer Male 63863
A25IB01 11
Hose 3x5 mm
Silicons Fuses Syringe
12 A25IB01
5. INSTALLING AND PREPARING THE
DMA 48 FOR OPERATION
HINTS:
• It is not necessary to level the DMA 48, nor to ensure that the
installation location is completely free of vibrations.
NOTE:
A strong built-in cooling fan dissipates heat through the bottom of the
DMA 48. Care must therefore be taken that the airflow at the bottom is not
blocked.
A25IB01 13
5.2 Filling Equipment
Fig. 5.2-1
Attaching the 1. Take two Teflon adapters out of the accessory box.
adapters
HINT:
Except when using the sample changers SP1 or SP3 (optional), one
adapter Record Luer female and one adapter Record Luer male are
used.
3. Screw in the Teflon adapters (see fig. 5.2-1) carefully, using your fingers
only.
NOTE:
Make sure that the tip of each Teflon adapter is directed into the center
of the appropriate bore of the measuring cell.
14 A25IB01
4. Tighten the adapters until some resistance against revolving can be felt.
NOTE:
5. Continue to turn the adapters to the next possible position for locking
the port lock (90° turn at the most).
NOTE:
During transportationof the DMA 48 theadapters should be removed,.
Otherwise the glass measuring eel! may be damaged.
7. Check the connection of the adapters to the measuring cell for tightness
by closing one adapter tightly with a finger.
8. Fill air under moderate pressure through the other adapter using a 2 ml
plastic syringe from the accessory box.
• If the connections are tight, the plunger of the syringe will be slowly
pushed back by the pressure in the measuring cell.
• If the connections are leaking, no pressure was built up in the
measuring cell and the plunger will not move. Tighten the adapters
again.
A25IB01 15
5.3 The Air Pump
HINT:
The air pump incorporated in the DMA 48 is used for drying the measuring
cell after cleaning it
Connecting the 1. Cut a piece of approx. 750 mm length off the silicone hose contained in
air pump the accessory box of the DM A 48.
2. Attach the silicone hose to the nozzle on the rear of the DMA 48 marked
-AIR OUT1 (see fig. 5.3-1).
Fig. 5.3-1
3. Attach an adapter Luer male (found in the accessory box) to the other
end of the silicone hose (see fig. 5.3-2).
Fig. 5.3-2
16 A25IB01
NOTE:
Never connect the siiicone hose to the "AIR IN" nozzle of the air pump
in order to prevent damage of the built-in air pump (e.g. due to sample
contamination).
HINT:
a desiccator has to be connected to the "AIR IN" nozzle of the air pump.
If the measuring temperature is 15°C> the relative humidity at ambient
temperature must be lower accordingly.
Otherwise humidity condensations in the measuring cell may cause
adjustment and measurement errors.
Suitable desiccators are available at Anton Paar.
1. Connect the keyboard to the DMA 48. The keyboard connector is Connecting the
located at the left side of the DMA 48. cables
2. If a PC is used, plug the interface cable into the upper 9-pin connector
at the rear of the DMA 48 (interface A, see chapter 6.2).
3. If a printer is used, plug the interface cable into the lower 9-pin
connector at the rear of the DMA 48 (interface B, see chapter 6.2).
4. Connect the power inlet of the DMA 48 (see chapter 6.2) to the mains
supply using the power cord.
NOTE:
Before plugging in the DMA 48 and switching it on, make sure that both
the correct line voltage and frequency are given and appropriate fuses
(technical data see appendix F) are used. If large voltage fluctuations
are to be expected, the use of a constant voltage source (UPS) is
recommended.
A25IB01 17
6. FUNCTIONAL ELEMENTS OF THE DMA 48
B CD
cJereity meier
period
g/cm3
(t devB
B renwle
C
act. l e m *t •D
PAAR
HADE n AUSTRIA
G H
Fig. 6.1-1
18 A25IB01
A Light bars to designate measured value shown on the display:
period (oscillation period of the measuring cell)
g/cm3 (measured density)
A (custom function A)
B (custom function B)
C (custom function C)
D Light bars:
X (flashes, if the measuring value is not stable)
dev. B (deviation B, see chapter 8.1 and chapter 8.2)
remote (indicates that an external measuring cell is activated)
E Set temperature
F Actual temperature
H Measuring cell light switch (turns cell light OFF and ON)
A25IB01 19
6.2 The Rear of the DMA 48
Fig. 6.2-1
20 A25IB01
A . "AIR OUT" nozzle of the air pump
C Fuse holder
D Power inlet
F Type plate
A25IB01 21
6.3 Keyboard Functions
Fig. 6.3-1
CLR • To clear incorrect entries while " J " has not been
pressed
• To clear an incorrect adjustment (if the displayed
deviation B is out of range)
• To clear error messages
• To switch a frozen (flashing) display back to normal
• • Decimal point
- Negative sign
22 A251B01
7. PUTTING THE DMA 48 INTO
OPERATION
NOTE:
• Before switching the DMA 48 on, make sure that both the correct line
voltage and frequency are given and appropriate fuses (technical data
see appendix F) are used, if large voltage fluctuations are to be
expected, the use of a constant voltage source (UPS) is recommended.
• The nonfused earth conductor of the power cord (or power inlet).has to
be connected to earth.
HINTS:
• For clear view of the measuring cell and the displays, remove the
colorless protective cover of the front foil by lifting it up in one corner
and pulling it off.
4 Turn on the DMA 48 using the "power" switch on the front of the
DMA 48.
• If the light and/or the pump are turned on, turn them off by using the
corresponding switches.
• After turning on the power all displays of the DMA 48 will light up.
• A "C" will be shown on the last digit of the large display, if the
adjustment is older than 7 days and the adjustment warning has not
been turned off. This "C" will disappear after a new adjustment.
• "CAL" will be shown on the iast three digits of the large display 7 days
before the expiry of the adjustment, and if the adjustment warning has
not been turned off. This "CAL" will disappear after a new adjustment
• If only "period" values can be displayed and "-CAL-" is shown for all
other selections, then no valid adjustment is available for the set
temperature. "-CAL-" has to be acknowledged by "J", and then a new
adjustment has to be carried out.
A25IB01 23
7.2 Master Reset
HINTS:
• A master reset has to be carried out if initial parameters, e.g. time and
date, have to be changed.
• Carrying out a master reset will clear all adjustments, parameters and
printer settings!
Performing a 1. TurnofftheDMA48.
master reset
2. Press the master reset key at the rear of the DMA 48 (see chapter 6.2)
using e.g. a small screwdriver, and turn on the "power" switch on the
front of the DMA 48 at the same time.
4. Enter the initial parameters according to the order of the following table.
24 A25IB01
Key in Display
F39 P39
J 0
1 1
J
F51 P51
J (time)
142035 14.20.35
J
F52 P52
J (date)
951119 95.11.19
J
4. F53: User specified number of days xxxx (e.g. 360 days) for
adjustment validity:
F53 = 0 Default value, which stands for 100 days
F53 = xxxx Adjustment validity for 20 - 9999 days
F53 P53
j 0
360 360
J
A25I801 25
5. F54 = 0: If the sample changer SP3 is connected each sample will
be filled once (default value)
F54 = 1: If the sample changer SP3 is connected each sample will
be filled twice
F54 P54
J 0
1 1
J
F55 P55
J 0
1 1
J (lamp test)
(period)
Putting the To put the DMA 48 into operation after a master reset proceed in the
DMA 48 into following order:
operation after a
26 A25IB01
7.3 Setting the Measuring Temperature
HINT:
The measuring temperature can be selected within -10°Cto +70°C (upon
request+80 °C).
3. Key in: 20 J
HINTS:
• The display marked "act. temp." shows the actual temperature of the
solid state thermostat, the display marked "set temp." shows the
selected measuring temperature.
• "X" at the front of the DMA 48 flashes until the sample tempera ture
agrees with the set temperature and the oscillation period of the
measuring cell is stable.
A25IB01 27
8. ADJUSTMENTS
HINTS:
- After the expiry of the validity ot the adjustment: after a set number
of days the adjustment will expire automatically (see master reset
chapter 7.2). Default value is 100 days. Only the oscillation period of
the measuring cell will be displayed then. Attempts to switch the
display to other settings will cause the display to show "-CAL-". This
reading has to be acknowledged by " J " and then anadjustment has
to be performed.
28 A251B01
8.1 Adjustment with Air and Water
HINTS:
• The adjustment with air and water has to be completed within one day.
1. Use a syringe or a proper sampling system (e.g. a sample changer) to Adjusting with
clean the measuring cell carefully by rinsing it with water or a proper air and water
solvent to remove sample residues.
3. Attach the air hose (see chapter 5.3) to the measuring cell.
HINT:
Check whether the air humidity does not exceed the limits given in
chapter 5.3 for working without a desiccator.
4. Turn on the air pump using the "pump" switch on the front of the
DMA 48.
5. Let dry air blow through the measuring cell for at least 10 minutes.
8. Turn on the measuring cell light and check the measuring cell. Make
sure there is no obvious moisture or dirt in the measuring cell.
A25IB01 29
9. Make sure that the DMA 48 is set to the correct measuring temperature
(see chapter 7.3) and that the actual temperature corresponds to the
selected measuring temperature.
HINT:
The DMA 48 should be turned on and set to the correct measuring
temperature for at least two hours prior to adjustment
10. Wait until the DMA 48 is stable ("Xn does not flash).
11. Press "F500" and " J" to select the oscillation period as display reading.
12. Press "F100" and "J". This starts the air adjustment routine.
13. The density of air at an air pressure of 1013.25 mbar <=hPa) at the
selected measuring temperature will be displayed.
• If all measurements are done within 0.75 and 1.25 g/cm3 and the
DMA 48 is located at an altitude above sea level of less than
1000 m, then no correction of the displayed air density is necessary.
Otherwise select the correct air density according to appendix G in
respect to the local air pressure and the set measuring temperature.
• In case of a correction overwrite the value and acknowledge the
correct value with "J".
• In case of agreement acknowledge the displayed value with
"j".
14. A red "X" and a green "g /cm3" will be flashing on the front of the
DMA 48. This indicates that the air adjustment procedure is in progress.
15. Wait until the air adjustment is finished (about 2 minutes). This is given
as soon as "g/cm3" and the red "X" on the front.of the DMA 48 do not
flash anymore.
16. Note down the oscillation period of the measuring cell (F500) for air in
the adjustment report (see appendix I).
17. Take a clean syringe with a Luer tip and fill it with bi-distilled water.
HINTS:
30 A25IB01
18. Slowly inject the water into the measuring cell through the adapter
Record Luer female until drops are coming out of the adapter Record
Luer male. Leave the syringe in the adapter.
HINT:
A minimum of 2 ml of bi-distilled water should be filled.
21. Press "F101" and "J". This starts the water adjustment routine.
22. The density of water at your measuring temperature will appear on the
display. Check the given value {see density table of water in
appendix G), and acknowledge the value with " J " .
23. A red "X" and a green "g /cm 3 " will be flashing on the front of the
DMA 48. This indicates that the water adjustment is in progress.
24. Wait until the water adjustment is finished {about 3 minutes). This is
given as soon as " g / c m 3 " a n c l * n e " x " o n t n e f r o n t o f t n e DMA 48 stop
flashing. At the same time "dev. B" and a number wiil be displayed.
25. • If this is the very first adjustment at.the selected temperature, or the
first adjustment at the selected temperature after a master reset, the
display should now show a number between 1.3 and 1.6.
If the number on the display is within this range, press " J " . The
adjustment is now complete. Go to step 26.
A25IB01 31
• If the number on your display is outside these ranges, some part of
the adjustment failed and the entire adjustment procedure needs to
be re-done. Press "CLR" to abort the adjustment procedure. Check
again the measuring cell for cleanliness, and perform a complete
. . cleaning procedure if necessary. Then start from the beginning. If
then the number of dev. B is identical to the previously obtained
dev. B {± 0.0002), then press "J".
26. Before removing the adjustment water from the measuring cell:
• Note down the oscillation period of the measuring cell (F 500) for
water in the adjustment report (see appendix I).
• Also note down the apparatus constants A and B with 5 places after
the decimal point. These constants can be displayed and printed out
by using function F 110 (see chapter 8.4).
• Also plot the value of the apparatus constant Bin-the appropriate
diagram (copy form see appendix I). After several adjustments
check the quality of the adjustment procedure by examining the
value of the apparatus constant B (see appendix I).
27. Rinse the measuring cell with ethanol and dry it using the built-in air
pump (see steps 2. to 7.).
32 A25IB01
8.2 Adjustment with Two Standards of Known
Density
NOTE:
Before filling any sample into the DMA 48, make sure that all wetted parts,
being made of Teflon and borosilicate glass, are resistant to the sample.
HINTS:
• Density values of the two standards must differ by at least 0.01 g/cm3,
the density must be known better than ±0.00005 g/cm3.
• The density values of the two standards should bracket the total .density
range required.
1. Use a syringe or a proper sampling system (e.g. a sample changer) to Adjusting with
clean the measuring cell carefully by rinsing it with water or proper air and water
solvent to remove sample residues.
3. Attach the air hose (see chapter 5.3) to the measuring cell.
HINT:
Check whether the air humidity does not exceed the limits given in
chapter 5.3 for working without a desiccator.
4. Turn on the air pump using the "pump" switch on the front of the
DMA 48.
5. Let air blow through the measuring cell for at least 10 minutes.
A25IB01 33
8. Turn on the measuring cell light and check the measuring cell. Make
sure there is no obvious moisture or dirt in the measuring cell.
9. Make sure that the DMA 48 is set to the correct measuring temperature
(see chapter 7.3) and that the actual temperature corresponds to the
selected measuring temperature.
HINT:
The DMA 48 should be turned on and set to the correct measuring
temperature for at least two hours prior to adjustment.
10; Wait until the DMA 48 is stable ("Xu does not flash).
11. Press "F500" and "J u to select the oscillation period as display reading.
13. Slowly inject the density standard 1 into the measuring cell through the
adapter Record Luer female until drops are emerging from the adapter
Record Luer male. Leave the syringe in the adapter.
HINT:
A minimum of 2 mi of density standard 1 should be filled.
14. Before starting the adjustment with density standard 1, check the
following:
16. Press "F102" and "J". This starts the adjustment routine for density
standard 1.
34 A25IB01
17. Key in the density value for density standard 1 at the selected
measuring temperature and acknowledge the correct value with " J " .
18. A red "X" and a green "g /cm 3 " will be flashing on the front of the
DMA 48. This indicates that the adjustment procedure is in progress.
19. Wait until the adjustment is finished (about 3 minutes). This is given as
soon as qg/cm3" and the red "X" on the front of the DMA 48 do not flash
anymore.
20. Note down the oscillation period of the measuring cell (F500) for density
standard 1 in the adjustment report (see appendix I).
24. Attach the air hose (see chapter 5.3) to the measuring cell.
HINT:
Check whether the air humidity does not exceed the limits given in
chapter 5.3 for working without a desiccator.
25. Turn on the air pump using the "pump" switch on the front of the
DMA 48.
26. Let air blow through the measuring cell for at least 10 minutes.
29. Turn on the measuring cell light and check the measuring cell. Make
sure there is no obvious moisture or dirt in the measuring cell.
A25IB01 35
30. lTake a clean syringe and fill it with density standard 2.
31. Slowly inject the density standard 2 into the measuring cell through the
adapter Record Luer female until drops are emerging from the adapter
Record Luer male. Leave the syringe in the adapter.
HINT:
A minimum of 2 mi of density standard 2 should be filled.
32. Before starting the adjustment with density standard 2, check the
following:
34. Press "F103" and "J". This starts the adjustment routine for density
standard 2.
35. Key in the density value for density standard 2 at the selected
measuring temperature and acknowledge the correct value with" J".
36. A red "X" and a green "g /cm3" will be flashing on the front of the
DMA 48. This indicates that the adjustment procedure is in progress.
37. Wait until the adjustment for density standard 2 is-finished (about 3
-'"• minutes). This is given assoon as ''g/crn?" and.the "X".on.the..front of the
DMA 48 stop flashing. At the same time "dev. B" and a number will be
displayed.
38. • If this is the very first adjustment at the selected temperature, or the
first adjustment after a master reset, the display should now show a
number between 1.3 and 1.6.
If the number on the display is within this range, press "J". The
adjustment is now complete. Go to step 39.
36 A25IB01
• If highly viscous standards and/or standards with very small
differences in density have been used, then the limits of 1.3 to 1.6
may be exceeded. Press "J" and measure again both standards.
The correct densities (F505) have to be displayed. If not, restart at
step 1.
• If this is not the first adjustment with the same standards at the
selected temperature, the number on the display should be between
-0.0002 and +0.0002, if the previous adjustment was valid.
If so, press "J". The adjustment is now complete. Go to step 39.
39. Before removing the density standard 2 from the measuring cell:
• Note down the oscillation period of the measuring cell (F 500) for
density standard 2 in the adjustment report (see appendix I).
• Also note down the apparatus constants A and B with 5 places after
the decimal point. These constants can be displayed and printed out
by using function F110 (see chapter 8.4).
• Also plot the value of the apparatus constant B in the appropriate
diagram (copy form see appendix I). After several adjustments
check the quality of the adjustment procedure by examining the
value of the apparatus constant B (see appendix I). The results may
only be significant if always the same standards are applied for
adjustment.
40. Rinse the measuring cell with an appropriate cleaning solvent and dry it
with the built-in air pump (see steps 22. to 28.).
A25IB01 37
8.3 Storage of Several Adjustments
HINTS:
• When storing the 21st adjustment the oldest one will be deleted.
3B A25IB01
8.4 Displaying and Printing the Adjustment Data
HINT:
Adjustment data of the currently used adjustment at the selected measuring
temperature can be displayed and printed using function F110. If the printer
does not print, try keying in F604 (see chapter 9.2).
F110
CAL PARAMETER
CP 0 = 2 0 . 0 Adjustment temperature
CP 1 = 19950706 Date of adjustment
CP •a 1.4550767 E-7 Apparatus constant A
CP 4 =* 1 . 4 1 6 7 3 7 4 E 0 Apparatus constant B
CP 5 = 1 Status CP 5 = 1 internal cell
CP5 = 2 external cell with temperature
transmission
CP5 = 3 external cell without temperature
transmission
A25IB01 39
8.5 Possible Errors During the Adjustment
Procedure
4. Measuring cell not perfectly dry before air adjustment. There are drops
of liquid or humidity condensations in the measuring cell.
5. Air bubbles in the measuring cell at adjustment with water or other liquid
standard.
Make sure that liquid standards are free of gas bubbles.
AQ A25IB01
9. DISPLAY AND PRINTER SETTINGS
2)
F 520 J custom function 2 B
2)
F530 J custom function 3 C
A25IB01 41
9.2 Printer Settings
Connecting the 1. Plug the printer cable into the lower 9-pin connector at the rear of the
printer and DMA 48 (interface B).
entering the
printer settings 2. Connect the printer to the mains.
Sample No.:4
DMA48
Date: 1995 02 27
Time: 10 39 36
Temperature: 20.0
Cell: Internal
Period: 3690.37
Density 0.9962
42 A25IB01
Keyboard Meaning Printout Units
Function
time Time HH MM SS
1)
F600 cancels F605, F607 and F608.
2
> F604 has to be entered if the input buffer of the printer has a capacity of less than 256 bytes
(technical data see printer handbook).
3
> If certain entries from F610 to F623 are not made, the corresponding data will not be printed
out.
4
) Printouts corresponding to F620 to F623 will occur only if there is a valid adjustment.
5
> Printouts corresponding to F621 to F623 will occur only if custom functions are programmed
and the measuring values are within the limits for the corresponding custom function.
A25IB01 43
Printout of current measuring results:
1. Key in: #1
HINT:
A sample number with a maximum of 3 digits can be entered.
1. Key in: #4
HINT:
A sample number with a maximum of 3 places can be entered.
HINT:
If an optional FRS is contained and turned on, this will also start the
FRS.
3. After stability has been reached and the printout has occured, the
display will freeze the value and start to flash.
44 A25IB01
10. MEASUREMENTS
HINTS:
• If the adjustment has been made with other standards but water, a
larger deviation may occur for water, in this case check for the
repeatability of this value only, or use the adjustment standards for
checking.
1. Fill the dry and clean measuring cell with degassed, distilled water. Checking the
adjustment with
2. Wait until temperature equilibrium is reached. This is given as soon as water
the red "X" at the front of the DMA 48 stops flashing.
• If the value of the true density deviates more than ±0.0001 g/cm3
from the corresponding value of the table, then clean and dry the
measuring cell (see chapter 11) and fill it again with degassed,
distilled water.
• If the value of true density still deviates more than ±0.0001 g/cm3
from the corresponding value of the table carry out a new
adjustment (see chapter 8).
A25IB01 45
10.2 Sample Preparation
NOTE:
- make sure that all wetted parts, being made of Teflon and
borosilicate glass, are resistant to the sample.
work out a proper cleaning procedure for the measuring cell (see
chapter 11).
HINTS:
46 A25IB01
10.3 Filling of the Sample
HINT:
There are several choices to fill the samples into the measuring cell:
• by gra vity, using a funnel at the inlet and a valve (e.g. a clamp to block a
flexible hose) at the outlet
• automatically, using the Anton Paar Filling and Rinsing System FRS
(see FRS handbook)
• automatically, using the Anton Paar Sample Changers SP1 or SP3 (see
SP1 orSP3 instruction handbook).
Waste
Fig. 10.3.1-1
A251B01 47
NOTE:
• Make sure that there is a waste bottle at the outlet of the DMA 48.
• Never fill the samples without the filling equipment (see chapter 5.2) in
order to avoid glass breakage of the measuring celt.
• Attach the syringe to the adapter Record Luer female of the filling
equipment.
• Proper filling of the measuring cell can be observed through the window
with the cell tight being switched on. After the fiiiing procedure turn off
the cell light.
• Take care that the entire visible part of the measuring cell is filled with
sample. For this, a sample amount of approximately 1 ml is necessary.
• The sample has to be filled into the measunng ceil by pushing the
plunger of the syringe slowly and continuously, thereby avoiding the
formation of tiny (invisible) gas bubbles which influence the measuring
result.
• Leave the syringe in the filling position, in order that the filled-in sample
does not leak out.
48 A25IB01
10.3.2 Filling with a Peristaltic Pump
Sample
Waste
Fig. 10.3.2-1
NOTE:
• Use two adapters Record Luer male for the hose connections.
• Make sure that there is a waste bottle at the outlet of the peristaltic
pump.
• Make sure that the pump hose is compatible with all samples and
cleaning liquids.
• Turn off the pump after filling a proper amount of sample and wait for
the instrument to reach stability.
A25IBO1 49
10.4 Actual Measurement, Measuring Results
NOTE:
make sure that the cell light is turned off during measurement.
wait until the red "X" on the front of the DMA 48 stops flashing. Then
the measuring result is valid.
• If a printer is connected and activated (see chapter 9.2) and F605 has
been entered as printer setting, a printout will occur every time the "X"
stops flashing.
Date 1996 2 4
Time 11 15 56
Temperature 20.0
Cell: Internal
Period: 3901.78
Density 1.0434
• Clean and dry the measuring cell after each measurement (see
chapter 11), if very different samples are to be measured. For similar
samples it is usually sufficient to replace the old sample by larger
amount (e.g. 10 ml) of the new sample.
50 A25IB01
10.5 Possible Measurement Errors
The display shows Air bubbles in the Take care that the sample
fluctuating values measuring cell. Sample is completely degassed.
was introduced into the Introduce the sample
"Xn on the front of the measuring cell too fast. slowly into the measuring
DMA 48 flashes This may result in the cell.
formation of small Slowly fill more sample
Error indication "E-00-02" (microscopic) gas until all air bubbles are
after 10 minutes bubbles. pushed out.
Take care that the sample
Error indication "-OSO" temperature is similar to
the measuring
temperature.
If the filling is performed
by peristaltic pump, loose
hoses or connections can
cause the introduction of
gas bubbles. Check the
tightness of all hoses and
connections.
A25IB01 51
The display reading The measuring cell is not Clean the measuring cell
slowly drifts. clean. (see chapter 11).
52 A25IB01
10.6 The Influence of Viscosity on Density
Measurement
• For samples with a viscosity T\ between 100 and 500 mPa.s the density
has to be corrected according to following equation:
= x
Pcouecled Pmeasured
A25IB01 53
11. CLEANING AND DRYING OF THE
MEASURING CELL
Cleaning and 1. Fill the measuring cell with cleaning liquid 1 by using a syringe with Luer
Drying cone, and move the plunger of the syringe in and out several times. This
creates air bubbles which improve the cleaning action. Instead of a
syringe, any other suitable device like a peristaltic pump can be used.
NOTE:
3. Fill the measuring cell with cleaning liquid 2 by using a syringe with Luer
tip, and move the plunger of the syringe several times in and out.
NOTE:
5. Attach the air hose (see chapter 5.3) to the adapter Record Luer female.
HINT:
Check whether the air humidity does not exceed the limits given in
chapter 5.3 for working without a desiccator.
54 A25IB01
6. Tum on the air pump using the "pump" switch on the front of the
DMA 48.
7. Let air blow through the measuring cell for several minutes.
A25IB01 55
12. CUSTOM FUNCTIONS
HINTS:
If the density related value is out of range of the custom function, " "
is shown on the display of the DMA 48.
Examples:
F-RHO = A1
where p is the density of the sample and A1, A2, A3, A4 and A5 are the
polynomial coefficients, which can be entered in the DMA 48 using the
keyboard.
56 A25IB01
APPENDIX A: TROUBLE SHOOTING
• "C"
"CAL"
"-CAL-" see chapter 7.1
E-xx-yy
xx... error location
yy... error code
A25IB01 57
Display Cause Correction
E-00-02 No stable measuring values 1. Make sure that there are. no.
within 10 minutes. bubbles in the sample.
2. Fill the measuring cell slowly and
completely.
3. Make sure that the cell light Is off.
E-00-50 Error at the processor. Turn off the DMA 48. Turn it on
E-00-90 again. If the problem reappears,
E-00-95 contact your local Anton Paar
E-00-99 representative.
E-02-yy Error at the temperature Turn off the DMA 48. Turn it on after
controller system. 15 minutes again. If the problem
reappears, contact your local Anton
Paar representative-
E-03-yy Error at the frequency controller Turn off the DMA48..Turn it.on
system. again. If the problem reappears,
contact your local Anton Paar
representative.
E-06-yy Error at the optional SP1 or SP3 See SP1 or SP3 handbook.
sample changer.
58 A25IB01
APPENDIX B: DATA TRANSFER TO AN
EXTERNAL COMPUTER
HINTS:
• The DMA 48 offers a data output function, which allows the measuring
values to be transmitted to an external computer.
HINT:
In the following, transmitted characters always appear as defined above.
A25IBO1 59
External Computer DMA 48 Comment
60 A251B01
1
> D(213) shows the current condition of the measurement in the following
format:
D(213) = 1xxx1yyyE+00
xxx = 000 The measuring value is stable and valid, if the value
foryyy<128.
yyy: The measuring values are invalid, if the value foryyy > 127.
If custom functions are programmed, valid data will occur as soon as the
measurement is stable (see D(213J) and remain unchanged unless the
measurement becomes unstable or a printout is made.
If problems occur during data transfer, the external computer has to transmit
the following:
<001>
<127>
<127>
<127>
After this the whole data transfer procedure has to be repeated from the
beginning.
A25IB01 61
APPENDIX C: EXTERNAL MEASURING
CELLS
HINT:
The DMA 48 can also be used as evaluation unit for Anton Paar external
measuring cells DMA 401, DMA 512 and DMA 602.
DMA512P.DMA602:
DMA 48
DMA512P
DMA 602
DMA 48
brown
loooooooo
I Er. 0-
DMA 401 YH while OND 8
HINT:
Connect the external measuring cell before activation.
62 A25IB01
APPENDIX D: COOLING WATER FOR
MEASUREMENTS BELOW
10 °C
HINT:
At low measuring temperatures (10 °C or lower) additional water cooling is
recommended. A slow flow of cooling water assists the built-in solid state
thermostat.
1. Connect one cooling water nozzle on the rear of the DMA 48 (see Connecting the
chapter 6.2) to the water pipe using a hose. water pipe
2. Connect a second hose to the other cooling water nozzle and put the
other end of the hose Into the sink.
3. Turn on the cold water (less than 20 °C) at a flow rate of 0.3 -
0.5 I/minute.
NOTE:
Do not block the water outlet. A maximum pressure of 0.3 bar (5psi) must
not be exceeded.
A25IB01 63
APPENDIX E: SETTING THE STABILITY
CRITERION
HINTS:
• The "X" at the front of the DMA 48 will stop flashing, if the oscillation
period of the measuring cell is stable fora certain time. This time can
be selected in steps of 10 from 5 to 85 seconds by setting F350,
2. Key In: 2J
64 A25IB01
APPENDIX F: TECHNICAL DATA
SPECIFICATIONS
Power; 80 VA
Mains voltage according to technical data shield ±10 %
Frequency according to technical data shield ±2 Hz
Fuses: glass tube fuses 5 x 20 mm; DIN 41662; ratings according to technical
data shield
GND TXD
All interfaces are designed to comply with SELV (Separated Extra-Low Voltage) requirements
according to EN 60950 and must not be connected to any interfaces other than those complying
with SELV.
A25IB01 65
Measuring range: 0 to 3 g/cm3
Uncertainty:
Density (after correct on-site
adjustment and calibration): ±1*10" 4 g/cm3
Temperature: ±0.1 °C
Repeatability:
Density: ±3 * 1 0 5 g/cm3
Temperature: ±0.01 °C
Weight: approx. 25 kg
Temperature limits for emergency shut off: The DMA 48 turns itself off,
• as soon as a temperature of
81 °C in the measuring cell
is exceeded.
• as soon as a temperature
between 56 °C and 65 °C is
exceeded at the ground
plate of the housing.
66 A25IB01
APPENDIX G: DENSITY TABLES
Density of Dry Air
At the temperature t in [°C] and the pressure p in [mbar] or [hPa] the density
p of air in [g/cm3] is calculated using
0.0012930 p
1 +0.00367 x t 1013.25
The numbers are valfd for a CO2 content in air of 0.03 % by volume; the
numbers change by ±1/19000 for every change in CO2 volume content of
±0.0001.
Literature:
Kohlrausch: Praklische Physik, Bd. 3, Tafeln, 22. Auflage (1968), pg. 39, B.GTeubner SluHgart
A25IB01 67
Density of Dry Air (-10 °C to +80 °C)
Composition of dry air in [v/v]: 78.110 % N 2 ; 20.938 % O 2 ; 0.916 % Ar; 0.033 % CO 2 ; 0.002 % Ne
Literature:
DIN 51 757 (04.1994): Testing of mineral oils and related materials; determination of density
68 A25IB01
Density of Water (0 °C to 80 °C)
t°c 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
0 .99984 .99985 .999B5 .99986 .99987 .99987 .99988 .99988 .99989 .99989
1 .99990 .99990 .99991 .99991 .99992 .99992 .99993 .99993 .99993 .99954
2 .99994 .99994 .99995 .99995 .99995 .99995 .99996 .99996 .99996 ^99996
3 .99996 .99997 .99997 .99997 .99997 .99997 .99997 .99997 .99997 .99997
4 .99997 .99997 .99997 .99997 .99997 .99997 .99997 .99997 .99997 .99997
5 .99996 .99996 .99996 .99996 .99996 .99995 .99995 .99995 .99995 .99994
6 .99994 .99994 .99993 .99993 .99993 .99992 .99992 .99991 .99991 .99991
7 .99990 .99990 .99989 .99989 .99988 .99988 .99987 .99987 .99986 .999B5
8 .99985 .99984 .99984 .99983 .99962 .99982 •999B1 .99980 .99980 .99979
9 .99978 .99977 .99977 .99976 .99975 .99974 .99973 .99973 -99972 .99971
10 .99970 .99969 .99968 .99967 .99966 .99965 .99964 .99963 .99962 .99961
11 .99960 .99959 .99958 .99957 .99956 .99955 .99954 .99953 .99952 .99951
12 .99950 .99949 .99947 .99946 .99945 .99944 .99943 .99941 .99940 .99939
13 .9993B .99936 -99935 .99934 .99933 .99931 .99930 .99929 .99927 .99926
14 .99924 .99923 .99922 .99920 .99919 .99917 .99916 .99914 .99913 .99911
15 .99910 .99908 .99907 .99905 .99904 .99902 .99901 .99899 .99897 .99896
16 .99894 .99893 .99891 .99869 .998B8 .99386 .99884 .99883 .99881 .99879
17 .99877 .99876 .99874 .99872 .99870 .99869 .99867 .99865 .99863 .99861
18 .99859 .99858 .99856 .99854 .99852 .99850 .99848 .99846 .99844 .99842
19 .99840 .99838 .99836 .99835 .99833 .99831 .99828 .99626 .99824 .99822
20 .99820 .99B1B .99B16 .99814 .99812 .99810 .99808 .99806 .99803 .99801
21 .99799 .99797 .99795 .99793 .99790 .99788 .99786 .99784 .99781 .99779
22 .99777 .99775 .99772 .99770 .99768 .99765 .99763 .99761 .99758 .99756
23 .99754 .99751 .99749 .99747 .99744 .99742 .99739 .99737 .99734 .99732
24 .99730 .99727 .99725 .99722 .99720 .99717 .99715 .99712 .99709 .99707
25 .99704 .99702 .99699 .99697 .99694 .99691 .99689 .99686 .99683 .99681
26 .99678 .99676 .99673 .99670 .99667 .99665 .99662 .99659 .99657 .99654
27 .99651 .99648 .99646 .99643 .99640 .99637 .99634 .99632 .99629 .99626
28 .99623 .99620 .99617 .99615 .99612 .99609 .99606 .99603 .99600 .99597
29 .99594 .99591 .99588 .99585 .99582 .99579 .99577 .99574 .99571 .99568
30 .99564 .99561 .9955B .99555 .99552 .99549 .99546 .99543 .99540 .99537
31 .99534 .99531 .99528 .99524 .99521 .99518 .99515 .99512 .99509 .99506
32 .99502 .99499 .99496 .99493 .99490 .99486 .99483 .99480 .99477 .99473
33 .99470 .99467 .99463 .99460 .99457 .99454 .99450 .99447 .99444 .99440
34 .99437 .99433 .99430 .99427 .99423 .99420 .99417 .99413 .99410 .99406
35 .99403 .99399 .99396 .99393 .99389 .99386 .99382 .99379 .99375 .99372
36 .99368 .99365 .99361 .99358 .99354 .99350 .99347 .99343 .99340 .99336
37 .99333 .99329 .99325 .99322 .99318 .99314 .99311 .99307 .99304 .99300
38 .99296 .99292 .99289 .99285 .99281 .99278 .99274 .99270 .99267 .99263
39 .99259 .99255 .99252 .99248 .99244 .99240 .99236 .99233 .99229 .99225
40 .99221 .99217 .99214 .99210 .99206 .99202 .99198 .99194 .99190 .99186
A25IB01 69
t°c 0.0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9
41 .99183 .99179 .99175 .99171 .99167 .99163 .99159 .99155 .99151 .99147
42 .99143 .99139 .99135 .99131 .99127 .99123 .99119 .99115 .99111 .99107
43 .99103 .99099 .99095 .99091 .99087 .99083 .99079 .99075 .99071 .99066
44 .99062 .99058 .99054 .99050 .99046 .99042 .99038 .99033 .99029 .99025
45 .99021 .99017 .99012 .99008 .99004 .99000 .98996 .98991 .98987 .98983
46 .98979 .98974 .98970 .98966 .93962 .98957 .9B953 .98949 .98944 .98940
47 .98936 .98931 .98927 .98923 .98918 .98914 .98910 .98905 .98901 .98897
48 .98892 .96888 .98883 .98879 .98875 .98870 .93866 .96861 .98B57 .98852
49 .98B4B .9S843 .96839 .98835 .98830 .98826 .93821 .98817 •98B12 .98808
50 .98803 .98799 .98794 .98789 .98785 .98780 .93776 .98771 .98767 .98762
51 .98757 .98753 .98748 .98744 .98739 .98734 .98730 .98725 .98721 .9B716
52 .98711 .98707 .98702 .98697 .9B693 .9868B .9B6B3 .98679 .98674 .98669
53 .98664 .9866 .98655 .98650 .98646 .98641 .98636 .98631 .98626 .98622
54 .98617 .98612 .98607 .98603 .98598 .98593 .98588 -985B3 .98579 .98574
55 .98569 .98564 .98559 .98554 .98549 .98545 .98540 .9B535 .98530 .9B525
56 .9B52O .98515 .98510 .98505 .98500 .98495 .98491 .9B4B6 .98481 .9B476
57 .98471 .98466 .98461 .98456 .98451 .98446 .98441 .9B436 .98431 .98426
58 .98421 .98416 .98411 .98406 .98401 .98396 .98391 .98385 .98380 .98375
59 .9B370 .98365 .98360 .98355 .98350 .98345 .98340 .98334 .98329 .98324
60 .98319 .98314 .98309 .98304 .98298 .98293 .98288 .98283 .98278 .98273
61 .98267 .98262 .98257 .98252 .98246 .9B241 .98236 .98231 .98226 .98220
62 .98215 .98210 .98204 .98199 .98194 .98189 .98183 .98178 .98173 .98167
63 .98162 .98157 .98151 .98146 .98141 .98135 .98130 .98125 .9B119 .98114
64 .98109 .98103 .98098 .98092 .98087 .98082 .98076 .98071 .98065 .98060
65 .98055 .98049 .98044 .98038 .98033 .98027 .96022 .98016 .98011 .98005
66 .98000 .97994 .97989 .97983 .97978 .97972 .97967 .97961 .97956 .97950
67 .97945 .97939 .97934 .97928 .97923 .97917 .97911 .97906 .97900 .97895
68 .97889 .97883 .97878 .97872 .97867 .97861 .97855 .97850 .97844 .97B38
69 .97833 .97827 .97821 .97816 .97810 .97804 .97799 .97793 .97787 .97782
70 .97776 .97770 .97764 .97759 .97753 .97747 .97742 .97736 .97730 .97724
71 .97719 .97713 .97707 .97701 .97695 .97590 .97684 .97678 .97672 .97666
72 .97661 .97655 .97649 .97643 .97637 .97631 .97626 .97620 .97614 .97608
73 .97602 .97596 .97590 .97585 .97579 .97573 . .97567 .97561 .97555 .97549
74 .97543 .97537 .97531 .97525 .97519 .97514 .97508 .97502 .97496 .97490
75 .97484 .97478 .97472 .97466 .97460 .97454 .97448 .97442 .97436 .97430
76 .97424 .97418 .97412 .97406 .97400 .97393 .97387 .97381 .97375 .97369
77 .97363 .97357 .97351 .97345 .97339 .97333 .97327 .97320 .97314 .97308
78 .97302 .97296 .97290 .97284 .97278 .97271 .97265 .97259 .97253 .97247
79 .97241 .97234 .97228 .97222 .97216 .97210 .97203 .97197 .97191 .97185
80 .97178 .97172 .97166 .97160 .97153 .97147 .97141 .97135 .97128 .97122
Literature:
Bettin, H.; Spieweck, F.: Die Dichte das Wassere als Funktion der Temperatur nach Einfuhrung der Intemationalen
Temperaturskala von 1990. PTB-Mitt. 100 (1990), pg.195-196
70 A25IB01
APPENDIX H: SUMMARY OF KEYBOARD
COMMANDS
Adjustment Commands
Parameter Commands
A25IB01 71
Display Commands
Printer Commands
Printout Unit
72 A25IB01
APPENDIX I: ADJUSTMENT REPORT (COPY FORM)
A25IB01 73
Copy Form: Checking the Adjustment Procedure Using Apparatus Constant B
Measuring temperature: °C
B
actual
B
start I
I
B
start + 0.0003
B
start + 0.0002
B
start + 0.0001
B Number of
start
5 o- 5
i
>o i
5
i
0 ;!5 Adjustments
B
start - 0.0001
I
B
start - 0.0002
B
start - 0.0003
B
start -0.0004
Checking the Adjustment Procedure Using
Apparatus Constant B
2. Note down the value of Bs,art at the origin of coordinates in the diagram.
4. After several adjustments have been made, check the quality of the
applied adjustment procedures by the aid of the given example.
A25IB01 75
Adjustment Report Serial Number of the DMA 48:
Air Water
(Density Standard 1) (Density Standard 2) Apparatus Constants (F110)
Date Temperature
Period Period
F500 F500 A B
Example: Checking the Adjustment Procedure Using Apparatus Constant B
Measuring temperature: 20 °C
B
actual
B
start + 0.0004 -i
B
start +0-0003 -j
B s t a r t +0.0002 -
B
start +0-0001 -
B s t a r t = 1.4570 Number of
Adjustments
B
acceptable
start - ° - 0 0 0 1 ~
improper adjustment procedures
(e.g. water for adjustment improperly degassed or
measuring cell not totally dried for air adjustment)
B
start
B
start - 0 - 0 0 0 3 ~
B
start - 0 - 0 0 0 4 "
APPENDIX J: DOCUMENT NUMBERS
78 A25IB01